0% found this document useful (0 votes)
50 views140 pages

GWR FMP54

Uploaded by

ahmed elkholany
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
50 views140 pages

GWR FMP54

Uploaded by

ahmed elkholany
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 140

TI01001F/00/EN/23.

17 Products Solutions Services


71384596

Technical Information
Levelflex FMP51, FMP52,
FMP54
Guided wave radar

Level and interface measurement in liquids

Application
• Rod, rope or coax probe
• Process connection: Starting 3/4" thread, flange or process connections for hygiene
requirements (Tri-Clamp, 11851)
• Process temperature: –196 to +450 °C (–320 to +842 °F)
• Process pressure: –1 to +400 bar (–14.5 to +5 800 psi)
• Maximum measuring range: Rod 10 m (33 ft); rope 45 m (148 ft); coax 6 m (20 ft)
• Accuracy: ±2 mm (±0.08 in)
• International explosion protection certificates; WHG; marine approval; steam
boiler approval; EN10204-3.1
• Linearity protocol (3-point, 5-point)

Your benefits
• Reliable measurement even for changing product and process conditions
• HistoROM data management for easy commissioning, maintenance and
diagnostics
• Highest reliability due to Multi-Echo Tracking
• Hardware and software developed according to IEC 61508 (up to SIL3)
• Seamless integration into control or asset management systems
• Intuitive user interface in national languages
• Easy proof test for SIL and WHG
• Heartbeat Technology™
Levelflex FMP51, FMP52, FMP54

Table of contents

Important document information . . . . . . . . . . . . . . . 4 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . 76


Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Function and system design . . . . . . . . . . . . . . . . . . . 6 Process temperature range . . . . . . . . . . . . . . . . . . . . . . 77
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Process pressure range . . . . . . . . . . . . . . . . . . . . . . . . . 77
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Dielectric constant (DC) . . . . . . . . . . . . . . . . . . . . . . . . 78
Expansion of the rope probes through temperature . . . . . . 78
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Mechanical construction . . . . . . . . . . . . . . . . . . . . 79
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Blocking distance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Tolerance of probe length . . . . . . . . . . . . . . . . . . . . . . . 87
Measuring frequency spectrum . . . . . . . . . . . . . . . . . . . 14 Surface roughness of AlloyC-coated flanges . . . . . . . . . . . 87
Shortening probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Materials: GT18 housing (stainless steel, corrosion-
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
resistant) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Materials: GT19 housing (plastic) . . . . . . . . . . . . . . . . . . 90
Linearization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Materials: GT20 housing (die-cast aluminum, powder-
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
coated) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Materials: Process connection . . . . . . . . . . . . . . . . . . . . 93
Materials: Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Materials: Mounting bracket . . . . . . . . . . . . . . . . . . . . . 96
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Materials: Adapter and cable for remote sensor . . . . . . . . 97
Device plug connectors . . . . . . . . . . . . . . . . . . . . . . . . . 30 Materials: Weather protection cover . . . . . . . . . . . . . . . . 98
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . 35
Operation with remote display and operating module
Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
FHX50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Integration in tank gauging system . . . . . . . . . . . . . . . . 104
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . 36
SupplyCare inventory management software . . . . . . . . . 105

Performance characteristics . . . . . . . . . . . . . . . . . . 37 Certificates and approvals . . . . . . . . . . . . . . . . . . 108


Reference operating conditions . . . . . . . . . . . . . . . . . . . 37
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Reference accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
RoHS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
RCM-Tick marking . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Reaction time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Influence of ambient temperature . . . . . . . . . . . . . . . . . 40
Dual seal according to ANSI/ISA 12.27.01 . . . . . . . . . . . 108
Influence of gas layer . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Functional Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Gas phase compensation with external pressure sensor
Overfill prevention . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
(PROFIBUS PA / FOUNDATION Fieldbus) . . . . . . . . . . . . 41
Sanitary compatibility . . . . . . . . . . . . . . . . . . . . . . . . . 109
Gas phase compensation with reference signal (option for
AD2000 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
FMP54) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
NACE MR 0175 / ISO 15156 . . . . . . . . . . . . . . . . . . . . 109
NACE MR 0103 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 ASME B31.1 and B31.3 . . . . . . . . . . . . . . . . . . . . . . . 109
Mounting requirements . . . . . . . . . . . . . . . . . . . . . . . . 44 Pressure equipment with allowable pressure
≤ 200 bar (2 900 psi) . . . . . . . . . . . . . . . . . . . . . . . . . 109
Pressure equipment with allowable pressure >
Operating conditions: Environment . . . . . . . . . . . . 69
200 bar (2 900 psi) . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . 69
Steam boiler approval . . . . . . . . . . . . . . . . . . . . . . . . . 110
Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . 69
Marine certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Telecommunications . . . . . . . . . . . . . . . . . . . . . . . . . 110
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
CRN approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Altitude according to IEC61010-1 Ed.3 . . . . . . . . . . . . . . 76
Track record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Test, Certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Product documentation on paper . . . . . . . . . . . . . . . . . 113
Cleaning the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

2 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Other standards and guidelines . . . . . . . . . . . . . . . . . . 114

Ordering information . . . . . . . . . . . . . . . . . . . . . . 115


Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . 115
3-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . 116
5-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . 117
Customized parametrization . . . . . . . . . . . . . . . . . . . . 119
Tagging (TAG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119

Application Packages . . . . . . . . . . . . . . . . . . . . . . 120


Heartbeat Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . 120
Heartbeat Verification . . . . . . . . . . . . . . . . . . . . . . . . 121
Heartbeat Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . 122

Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . 123
Communication-specific accessories . . . . . . . . . . . . . . . 131
Service-specific accessories . . . . . . . . . . . . . . . . . . . . . 132
System components . . . . . . . . . . . . . . . . . . . . . . . . . . 132

Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . 133
Standard documentation . . . . . . . . . . . . . . . . . . . . . . . 133
Supplementary documentation . . . . . . . . . . . . . . . . . . . 133
Safety documentation . . . . . . . . . . . . . . . . . . . . . . . . . 134

Registered trademarks . . . . . . . . . . . . . . . . . . . . . 138

Patents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139

Endress+Hauser 3
Levelflex FMP51, FMP52, FMP54

Important document information


Symbols Safety symbols

Symbol Meaning

DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will
result in serious or fatal injury.

WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in serious or fatal injury.

CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in minor or medium injury.

NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.

Electrical symbols

Symbol Meaning

Direct current

Alternating current

Direct current and alternating current

Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.

Protective Earth (PE)


A terminal which must be connected to ground prior to establishing any other
connections.
The ground terminals are situated inside and outside the device:
• Inner ground terminal: Connects the protectiv earth to the mains supply.
• Outer ground terminal: Connects the device to the plant grounding system.

Symbols for certain types of information

Symbol Meaning

Permitted
Procedures, processes or actions that are permitted.

Preferred
Procedures, processes or actions that are preferred.

Forbidden
Procedures, processes or actions that are forbidden.

Tip
Indicates additional information.

Reference to documentation.

A Reference to page.

Reference to graphic.

Visual inspection.

4 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Symbols in graphics

Symbol Meaning

1, 2, 3 ... Item numbers

1. , 2. , 3. … Series of steps

A, B, C, ... Views

A-A, B-B, C-C, ... Sections

Hazardous area
- Indicates a hazardous area.

Safe area (non-hazardous area)


. Indicates the non-hazardous area.

Symbols at the device

Symbol Meaning

Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.

Temperature resistance of the connection cables


Specifies the minimum value of the temperature resistance of the connection cables.

Endress+Hauser 5
Levelflex FMP51, FMP52, FMP54

Function and system design


Measuring principle Basic principles
The Levelflex is a "downward-looking" measuring system that functions according to the ToF method
(ToF = Time of Flight). The distance from the reference point to the product surface is measured.
High-frequency pulses are injected to a probe and led along the probe. The pulses are reflected by
the product surface, received by the electronic evaluation unit and converted into level information.
This method is also known as TDR (Time Domain Reflectometry).

20 mA
100%
D
E
LN F

4 mA
0%

A0011360

1 Parameters for level measurement with the guided radar


LN Probe length
D Distance
L Level
R Reference point of measurement
E Empty calibration (= zero)
F Full calibration (= span)

If, for rope probes, the DC value is less than 7, then measurement is not possible in the area of
the straining weight (0 to 250 mm (0 to 9.84 in) from end of probe; lower blocking distance).

6 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Dielectric constant
The dielectric constant (DC) of the medium has a direct impact on the degree of reflection of the
highfrequency pulses. In the case of large DC values, such as for water or ammonia, there is strong
pulse reflection while, with low DC values, such as for hydrocarbons, weak pulse reflection is
experienced.

Input
The reflected pulses are transmitted from the probe to the electronics. There, a microprocessor
analyzes the signals and identifies the level echo which was generated by the reflection of the high-
frequency pulses at the product surface. This clear signal detection system benefits from over 30
years' experience with pulse time-of-flight procedures that have been integrated into the
development of the PulseMaster® software.
The distance D to the product surface is proportional to the time of flight t of the impulse:
D = c · t/2,
where c is the speed of light.
Based on the known empty distance E, the level L is calculated:
L=E–D
The reference point R of the measurement is located at the process connection. For details see the
dimensional drawing:
• FMP51: →  81
• FMP52: →  84
• FMP54: →  85
The Levelflex possesses functions for interference echo suppression that can be activated by the user.
They guarantee that interference echoes from e.g. internals and struts are not interpreted as level
echoes.

Output
The Levelflex is preset at the factory to the probe length ordered so that in most cases only the
application parameters that automatically adapt the device to the measuring conditions need to be
entered. For models with a current output, the factory adjustment for zero point E and span F is 4
mA and 20 mA, for digital outputs and the display module 0 % and 100 %. A linearization function
with max. 32 points, which is based on a table entered manually or semi-automatically, can be
activated on site or via remote operation. This function allows the level to be converted into units of
volume or mass, for example.

Endress+Hauser 7
Levelflex FMP51, FMP52, FMP54

Interface measurement
When the high-frequency pulses hit the surface of the medium, only a percentage of the
transmission pulse is reflected. In the case of media with a low DC1, in particular, the other part
penetrates the medium. The pulse is reflected once more at the interface point to a second medium
with a higher DC2. The distance to the interface layer now can also be determined taking into
account the delayed time-of-flight of the pulse through the upper medium.

c = c0 = 300 000 km/s

LL

C0
DK1 (DC1) C1 =
DK1

LI

DK2 (DC2)

A0011178

2 Interface measurement with the guided radar


LL Level complete
LI Level interface
R Reference point of measurement

In addition, the following general conditions must be observed for interface measurement:
• The DC of the upper medium must be known and constant 1). The DC can be determined with the
aid of the DC manual CP00019F or the "DC Values App". In addition, if the interface thickness is
existing and known, the DC can be calculated automatically via FieldCare.
• The DC of the upper medium may not be greater than 10.
• The DC difference between the upper medium and lower medium must be >10.
• The upper medium must have a minimum thickness of 60 mm (2.4 in).
• Emulsion layers in the area of the interface can strongly attenuate the signal. However, emulsion
layers up to 50 mm (2 in) are admissible.
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
• the Endress+Hauser DC manual (CP01076F)
• the Endress+Hauser "DC Values App" (available for Android and iOS)

1) For FMP55: Under certain conditions measurement is possible even with a changing DC. For details please contact your Endress+Hauser
representative.

8 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Life cycle of the product


Engineering
• Universal measuring principle
• Measurement unaffected by medium properties
• Hardware and software developed according to SIL IEC 61508
• Genuine, direct interface measurement
Procurement
• Endress+Hauser being the world market leader in level measurement guarantees asset protection
• Worldwide support and service
Installation
• Special tools are not required
• Reverse polarity protection
• Modern, detachable terminals
• Main electronics protected by a separate connection compartment
Commissioning
• Fast, menu-guided commissioning in only 6 steps
• Plain text display in national languages reduces the risk of error or confusion
• Direct local access of all parameters
• Short instruction manual at the device
Operation
• Multi-echo tracking: Reliable measurement through self-learning echo-search algorithms taking
into account the short-term and long-term history in order to check the found echoes for
plausibility and to suppress interference echoes.
• Diagnostics in accordance with NAMUR NE107
Maintenance
• HistoROM: Data backup for instrument settings and measured values
• Exact instrument and process diagnosis to assist fast decisions with clear details concerning
remedies
• Intuitive, menu-guided operating concept in national languages saves costs for training,
maintenance and operation
• Cover of the electronics compartment can be opened in hazardous areas
Retirement
• Order code translation for subsequent models
• RoHS-conforming (Restriction of certain Hazardous Substances), unleaded soldering of electronic
components
• Environmentally sound recycling concept

Endress+Hauser 9
Levelflex FMP51, FMP52, FMP54

Measuring system General notes on probe selection


• Normally use rod or coax probes for liquids. Rope probes are used in liquids for measuring ranges >
10 m (33 ft) ( > 4 m (13 ft) for FMP52) and with restricted ceiling clearance which does not allow
the installation of rigid probes.
• For interface measurement, ideally coax probes or rod probes in a bypass/stilling well are used.
• Coax probes are suited to liquids with viscosities of up to approx. 500 cst. Coax probes can measure
most liquefied gases, as of a dielectric constant of 1.4. Moreover, installation conditions, such as
nozzles, tank internal fittings etc., have no effect on the measurement when a coax probe is used.
A coax probe offers maximum EMC safety when used in plastic tanks.

Probe selection
The various types of probe in combination with the process connections are suitable for the following
applications 2):

Levelflex FMP51

Type of probe Rod probe Rope probe Coax probe 1)

A0011387 A0011359

A0011388

Feature 060 - Probe: Option: Option: Option:

AA 8 mm (316L) LA 4 mm (316) UA ... mm (316L)

AB 1/3" (316L) LB 1/6" (316) UB ... inch (316L)

AC 12 mm (316L) MB 4 mm (316) with UC ... mm (AlloyC)


center rod

AD 1/2" (316L) MD 1/6" (316) with UD ... inch (AlloyC)


center rod

AL 12 mm (AlloyC)

AM 1/2" (AlloyC)

BA 16 mm (316L)
BC divisible

BB 0.63 in (316L)
BD divisible

Max. probe length 10 m (33 ft) 2) 45 m (148 ft) 6 m (20 ft)

For application level and interface level and interface level and interface
measurement in liquids measurement in liquids measurement in liquids

1) Punched for process connections G1-1/2" or flange; multiple holes for 316L, one hole for AlloyC
2) Maximum probe length for indivisible rod probes: 4 m (13 ft)

2) If required, rod and rope probes can be replaced. They are secured with Nord-Lock washers or a thread coating. For further information on
service and spare parts please contact the Endress+Hauser service.

10 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Levelflex FMP52

Type of probe Rod probe Rope probe

A0011357

A0011358

Feature 060 - Probe: Option: Option:

CA 16 mm (PFA>316L) OA 4 mm (PFA>316), max. 150 mm nozzle height,


center rod

CB 0.63 in (PFA>316L) OB 4 mm (PFA>316), max. 300 mm nozzle height,


center rod

OC 1/6" (PFA>316), max. 6 in nozzle height, center


rod

OD 1/6" (PFA>316), max. 12 in nozzle height,


center rod

Max. probe length 4 m (13 ft) 45 m (148 ft)

For application level and interface level and interface measurement in corrosive liquids
measurement in corrosive
liquids

Levelflex FMP54

Type of probe Rod probe Rope probe Coax probe

A0011357 A0011359

A0011358

Feature 060 - Probe: Option: Option: Option:

AE 16 mm (316L) LA 4 mm (316) UA ... mm (316L)


BA
BC

AF 0.63 in (318L) LB 1/6" (316) UB ... inch (316L)


BB
BD

Max. probe length 10 m (33 ft) 1) 45 m (148 ft) 6 m (20 ft)

For application level and interface level and interface level and interface
measurement in liquids measurement in liquids measurement in liquids

1) Maximum probe length for indivisible rod probes: 4 m (13 ft)

Endress+Hauser 11
Levelflex FMP51, FMP52, FMP54

Input
Measured variable The measured variable is the distance between the reference point and the product surface.
Subject to the empty distance entered "E" the level is calculated.
Alternatively, the level can be converted into other variables (volume, mass) by means of
linearization (32 points).

Measuring range The following table describes the media groups and the possible measuring range as a function of
the media group.

Levelflex FMP51, FMP54


1)
Measuring range
Media group DC (εr) Typical liquids bare metallic bare metallic coax probes
rod probes rope probes

1 1.4...1.6 condensed gases, e.g. N2, CO2 on request 6 m (20 ft)

2 1.6...1.9 • liquefied gas, e.g. propane • one-piece: 4 m (13 ft) 15 to 22 m (49 to 72 ft) 6 m (20 ft)
• solvent • divisible: 10 m (33 ft)
• Freon
• palm oil

3 1.9...2.5 mineral oils, fuels • one-piece: 4 m (13 ft) 22 to 32 m (72 to 105 ft) 6 m (20 ft)
• divisible: 10 m (33 ft)

4 2.5...4 • benzene, styrene, toluene • one-piece: 4 m (13 ft) 32 to 42 m 6 m (20 ft)


• furan • divisible: 10 m (33 ft) (105 to 138 ft)
• naphthalene

5 4...7 • chlorobenzene, chloroform • one-piece: 4 m (13 ft) 42 to 45 m 6 m (20 ft)


• cellulose spray • divisible: 10 m (33 ft) (138 to 148 ft)
• isocyanate, aniline

6 >7 • aqueous solutions • one-piece: 4 m (13 ft) 45 m (148 ft) 6 m (20 ft)
• alcohols • divisible: 10 m (33 ft)
• ammonia

1) The measuring range for interface measurement is limited to 10 m (33 ft).

12 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Levelflex FMP52
1)
Measuring range
Media group DC (εr) Typical liquids PFA-coated PFA-coated
rod probes rope probes

1 1.4...1.6 condensed gases, e.g. N2, CO2 — —

2 1.6...1.9 • liquefied gas, e.g. propane 4 m (13 ft) 9 to 14 m (30 to 46 ft)


• solvent
• Freon
• palm oil

3 1.9...2.5 mineral oils, fuels 4 m (13 ft) 14 to 21 m (46 to 69 ft)

4 2.5...4 • benzene, styrene, toluene 4 m (13 ft) 21 to 28 m (69 to 92 ft)


• furan
• naphthalene

5 4...7 • chlorobenzene, chloroform 4 m (13 ft) 28 to 32 m (92 to 105 ft)


• cellulose spray
• isocyanate, aniline

6 >7 • aqueous solutions 4 m (13 ft) 32 to 45 m


• alcohols (105 to 148 ft)
• acids, alkalis

1) The measuring range for interface measurement is limited to 10 m (33 ft).

• Reduction of the max. possible measuring range through buildup, above all of moist products.
• Due to the high diffusion rate of ammonia it is recommended with gas-tight bushing 3) for
measurements in this medium.

3) always contained in FMP54 ; optionally available for FMP51/52

Endress+Hauser 13
Levelflex FMP51, FMP52, FMP54

Blocking distance The upper blocking distance (= UB) is the minimum distance from the reference point of the
measurement (mounting flange) to the maximum level.

R
SD UB
100%
E

F
LN

0%

A0011279

3 Definition of blocking distance and safety distance


R Reference point of measurement
LN Probe length
UB Upper blocking distance
E Empty calibration (= zero)
F Full calibration (= span)
SD Safety distance

Blocking distance (factory setting):


• with coax probes: 0 mm (0 in)
• with rod and rope probes up to 8 m (26 ft): 200 mm (8 in)
• with rod and rope probes exceeding a length of 8 m (26 ft): 0.025 * (length of probe)
The specified blocking distances are preset on delivery. Depending on the application these
settings can be changed.
For rod and rope probes and for media with DC > 7 (or generally for stilling well/bypass
applications) the blocking distance may be reduced to 100 mm (4").
Within the blocking distance, a reliable measurement can not be guaranteed.
A safety distance SD can be defined in addition to the blocking distance. A warning is generated
if the level rises into this safety distance.

Measuring frequency 100 MHz to 1.5 GHz


spectrum

14 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Output
Output signal HART

Signal coding FSK ±0.5 mA over currency signal

Data transmission rate 1200 Baud

Galvanic isolation Yes

PROFIBUS PA

Signal coding Manchester Bus Powered (MBP)

Data transmission rate 31,25 KBit/s, voltage mode

Galvanic isolation Yes

FOUNDATION Fieldbus

Signal coding Manchester Bus Powered (MBP)

Data transmission rate 31,25 KBit/s, voltage mode

Galvanic isolation Yes

Switch output
For HART devices, the switch output is available as an option. See product structure, feature 20:
"Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output"
Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output.

Endress+Hauser 15
Levelflex FMP51, FMP52, FMP54

Switch output

Function Open collector switching output

Switching behavior Binary (conductive or non-conductive), switches when the programmable switch
point is reached

Failure mode non-conductive

Electrical connection U = 16 to 35 VDC, I = 0 to 40 mA


values

Internal resistance RI < 880 Ω


The voltage drop at this internal resistance has to be taken into account on
planning the configuration. For example, the resulting voltage at a connected relay
must be sufficient to switch the relay.

Insulation voltage floating, Insulation voltage 1 350 VDC to power supply aund 500 VAC to ground

Switch point freely programmable, separately for switch-on and switch-off point

Switching delay freely programmable from 0 to 100 s, separately for switch-on and switch-off
point

Number of switching corresponds to the measuring cycle


cycles

Signal source • Level linearized


device variables • Distance
• Terminal voltage
• Electronic temperature
• Relative echo amplitude
• Interface linearized 1)
• Interface distance 1)
• Upper interface thickness 1)
• Relative interface amplitude 1)
• Diagnostic values, Advanced diagnostics

Number of switching unlimited


cycles

1) only if an interface measurement is active

Signal on alarm Depending on the interface, failure information is displayed as follows:


• Current output (for HART devices)
– Failsafe mode selectable (in accordance with NAMUR Recommendation NE 43):
Minimum alarm: 3.6 mA
Maximum alarm (= factory setting): 22 mA
– Failsafe mode with user-selectable value: 3.59 to 22.5 mA
• Local display
– Status signal (in accordance with NAMUR Recommendation NE 107)
– Plain text display
• Operating tool via digital communication or service interface (CDI)
– Status signal (in accordance with NAMUR Recommendation NE 107)
– Plain text display

Linearization The linearization function of the device allows the conversion of the measured value into any unit of
length or volume. Linearization tables for calculating the volume in cylindrical tanks are pre-
programmed. Other tables of up to 32 value pairs can be entered manually or semi-automatically.

Galvanic isolation All circuits for the outputs are galvanically isolated from each other.

16 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Protocol-specific data HART

Manufacturer ID 17 (0x11)

Device type ID 0x1122

HART specification 7.0

Device description files (DTM, Information and files at:


DD) • www.endress.com
• www.fieldcommgroup.org

HART load Min. 250 Ω

HART device variables The measured values can be freely assigned to the device variables.
Measured values for PV (primary variable)
• Level linearized
• Distance
• Interface 1)
• Interface distance 1)
• Upper interface thickness 1)
• Electronic temperature
• Relative echo amplitude
• Relative interface amplitude 1)
Measured values for SV, TV, FV (second, third and fourth variable)
• Level linearized
• Distance
• Interface linearized 1)
• Interface distance 1)
• Upper interface thickness 1)
• Terminal voltage
• Electronic temperature
• Absolute echo amplitude
• Relative echo amplitude
• Absolute interface amplitude 1)
• Relative interface amplitude 1)
• Calculated DC

Supported functions • Burst mode


• Additional transmitter status

1) only if an interface measurement is active

Wireless HART data

Minimum start-up voltage for device version "2-wire; 4-20mA HART" 1): 17.5 V

Minimum start-up voltage for any other device version: 16.0 V

Start-up current 3.6 mA

Start-up time 45 s

Minimum operating 11.4 V


voltage

Multidrop current 3.6 mA

Set-up time 1s

1) Ordering feature 020: "Power supply; output", option A

Endress+Hauser 17
Levelflex FMP51, FMP52, FMP54

PROFIBUS PA

Manufacturer ID 17 (0x11)

Ident number 0x1558

Profile version 3.02

GSD file Information and files at:


• www.endress.com
GSD file version • www.profibus.org

Output values Analog Input:


• Level linearized
• Distance
• Interface 1)
• Interface distance 1)
• Upper interface thickness 1)
• Terminal voltage
• Electronic temperature
• Absolute echo amplitude
• Relative echo amplitude
• Absolute interface amplitude 1)
• Relative interface amplitude 1)
• Calculated DC
Digital Input:
• Extended diagnostic blocks
• Status output PFS Block

Input values Analog Output:


• Analog value from PLC (for sensor block external pressure and
temperature)
• Analog value from PLC to be indicated on the display
Digital Output:
• Extended diagnostic block
• Level limiter
• Sensor block measurement on
• Sensor block save history on
• Status output

Supported functions • Identification & Maintenance


Simple device identification via control system and nameplate
• Automatic Ident Number Adoption
GSD compatibility mode with respect to the previous device Levelflex M
FMP4x
• Physical Layer Diagnostics
Installation check of the PROFIBUS segment and the Levfelflex FMP4x via
terminal voltage and telegram monitoring
• PROFIBUS Up-/Download
Up to 10 times faster reading and writing of parameters via PROFIBUS Up-/
Download
• Condensed Status
Simple and self-explanatory diagnostic information due to categorization of
diagnostic messages

1) only if an interface measurement is active

18 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

FOUNDATION Fieldbus

Manufacturer ID 0x452B48

Device type 0x1022

Device Revision 0x01

DD Revision Information and files at:


• www.endress.com
CFF Revision • www.fieldcommgroup.org

Device Tester Version (ITK 6.01


Version)

ITK Test Campaign Number IT080500

Link Master (LAS) capable yes

Link Master / Basic Device yes; default: Basic Device


selectable

Node address Default: 247 (0xF7)

Features supported Following methods are supported:


• Restart
• ENP Restart
• Setup
• Linearization
• Self Check

Virtual Communication Relationships (VCRs)

Number of VCRs 44

Number of Link Objects in VFD 50

Permanent entries 1

Client VCRs 0

Server VCRs 10

Source VCRs 43

Sink VCRs 0

Subscriber VCRs 43

Publisher VCRs 43

Device Link Capabilities

Slot time 4

Min. inter PDU delay 8

Max. response delay 5

Endress+Hauser 19
Levelflex FMP51, FMP52, FMP54

Transducer Blocks
Block Content Output values

Setup Transducer Block Contains all parameters for a standard commissioning • Level or volume 1)
procedure (Channel 1)
• Distance (Channel 2)

Advanced Setup Contains all parameters for a more detailed no output values
Transducer Block configuration of the device

Display Transducer Block Contains all parameters for the configuration of the no output values
display module

Diagnostic Transducer Contains diagnostic information no output values


Block

Expert Configuration Contains parameters which require detailed no output values


Transducer Block knowledge of the functionalities of the device

Expert Information Contains information about the state of the device no output values
Transducer Block

Service Sensor Transducer Contains parameters which can only be operated by no output values
Block Endress+Hauser service personnel

Service Information Contains information on the state of device which is no output values
Transducer Block relevant for service operations

Data Transfer Transducer Contains parameters which allow to backup the device no output values
Block configuration in the display module and to restore it
into the device.

1) depending on the configuration of the block

20 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Function Blocks
Block Content Number of Number of Execution Functionality
permanent instantiable time
blocks blocks

Resource Block The Resource Block contains all the 1 0 - enhanced


data that uniquely identify the field
device. It is an electronic version of
a nameplate of the device.

Analog Input The AI block takes the 2 3 25 ms enhanced


Block manufacturer's input data, selected
by channel number, and makes it
available to other function blocks at
its output.

Discrete Input The DI block takes a discrete input 1 2 20 ms standard


Block value (e.g. indication of an level
limit), and makes it available to
other function blocks at its output.

PID Block The PID block serves as 1 1 25 ms standard


proportional-integral-derivative
controller and is used almost
universally to do closed-loop-
control in the field including
cascade and feedforward.

Arithmetic This block is designed to permit 1 1 25 ms standard


Block simple use of popular measurement
math functions. The user does not
have to know how to write
equations. The math algorithm is
selected by name, chosen by the
user for the function to be done.

Signal The signal characterizer block has 1 1 25 ms standard


Characterizer two sections, each with an output
Block that is a non-linear function of the
respective input. The non-linear
function is determined by a single
look-up table with 21 arbitrary x-y
pairs.

Input Selector The input selector block provides 1 1 25 ms standard


Block selection of up to four inputs and
generates an output based on the
configured action. This block
normally receives its inputs from AI
blocks. The block performs
maximum, minimum, middle,
average and ‘first good’ signal
selection.

Integrator The Integrator Function Block 1 1 25 ms standard


Block integrates a variable as a function
of the time or accumulates the
counts from a Pulse Input block.
The block may be used as a
totalizer that counts up until reset
or as a batch totalizer that has a
setpoint, where the integrated or
accumulated value is compared to
pre-trip and trip settings,
generating discrete signals when
these settings are reached.

Analog Alarm 1 1 25 ms standard


Block

Up to 20 blocks can be instantiated in the device altogether, including the blocks already
instantiated on delivery.

Endress+Hauser 21
Levelflex FMP51, FMP52, FMP54

Power supply
Terminal assignment 2-wire: 4-20mA HART

! 2- w
4-2 ire le
0 v el
HA mA
[21 RT
4-2
0m
A ]
ope
A

n Sp
a
711 re pa

HA 20mA
08x rt
xx

RT
4...
+
1 -
1 2 4 5 6

mm
2

10
+ +


# 4...20 mA
3 B

7 +
1-c 1 -
1 2 4 5 han
nel 4-2 2
ove 0m
[16
] rvo A
ltage
pro
tec
tion
+ +


# 4...20 mA
3
A0011294

4 Terminal assignment 2-wire; 4-20mA HART


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Active barrier with power supply (e.g. RN221N): Observe terminal voltage
2 HART communication resistor (≥ 250 Ω): Observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device: Observe maximum load
5 Cable screen; observe cable specification
6 4-20mA HART (passive): Terminals 1 and 2
7 Overvoltage protection module
8 Terminal for potential equalization line
9 Cable entry

22 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

2-wire: 4-20mA HART, switch output

10
9
8

% 2- w
4-2 ire
0
HA mA
[02 RT
PF
S
A /03
] ope
n Sp
a
711 re pa

HA 20 mA
08x rt
3+ ++ xx

31 --

RT
4...
42 +
7
1 -

mm
2 6
4-

10
+ +

-
≥250 Ω 4...20 mA
$
1 2 3 4 5 B
3+ 11 +
3 -
4-2 4
0m +
2-c
A/ 1 -
han 4-2 2
4- nel
ove F IEL 0m
DB A/
[17 rvo
] ltag
ep US
rote
ctio
n
+ +

-
≥250 Ω 4...20 mA
$
1 2 3 4 5
A0013759

5 Terminal assignment 2-wire; 4-20mA HART, switch output


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Active barrier with power supply (e.g. RN221N): Observe terminal voltage
2 HART communication resistor (≥ 250 Ω): Observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device: Observe maximum load
5 Cable screen; observe cable specification
6 4-20mA HART (passive): Terminals 1 and 2
7 Switch output (open collector): Terminals 3 and 4
8 Terminal for potential equalization line
9 Cable entry for 4-20mA HART line
10 Cable entry for switch output line
11 Overvoltage protection module

Endress+Hauser 23
Levelflex FMP51, FMP52, FMP54

2-wire: 4-20mA HART, 4-20mA

14
13
12

# 2- w
4-2 ire le
0 v el
HA mA
[04 RT
4-2
0m
A /05
]
A
ope

A
n Sp

20m
a
711 re pa

HA 20mA
08x rt
xx
+

4 ...
3 -

RT
11

4 ...
4 +
1 -

mm
2
+ +

10
1 " –
-
4...20 mA
9 8 4
+ +
2 " –
-
4...20 mA
3 5 6 7 4
B
+ +
1 " –
- +
3 -
4...20 mA 4-2 4
10 0m +
9 8 4 2-c
A/ 1 -
han
nel 4- 2
[17
ove FIE 20mA
] rvo LD /
ltag
ep
B U
ro S
tec
tion
+ +
2 " –
-
4...20 mA
3 5 6 7 4
A0013923

6 Terminal assignment 2-wire, 4-20 mA HART, 4...20mA


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Connection current output 2
2 Connection current output 1
3 Supply voltage for current output 1 (e.g. RN221N); Observe terminal voltage
4 Cable screen; observe cable specification
5 HART communication resistor (≥ 250 Ω): Observe maximum load
6 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
7 Analog display device ; observe maximum load
8 Analog display device ; observe maximum load
9 Supply voltage for current output 2 (e.g. RN221N); Obeserve terminal voltage
10 Overvoltage protection module
11 Current output 2: Terminals 3 and 4
12 Terminal for the potential equalization line
13 Cable entry for current output 1
14 Cable entry for current output 2

This version is also suited for single-channel operation. In this case, current output 1 (terminals
1 and 2) must be used.

24 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

4-wire: 4-20mA HART (10.4 to 48 VDC)

13
12
11

! 2- w
4-2 ire
0
HA mA
[08 RT
A ]

= V
o pe

48
n Sp

4...
a
711 re pa

HA 20 mA
08x rt

10.
L+ xx

1 L-

RT
4...
2 +
10 9
3 -

mm
4 8

10
7

+ !
-
≥250 Ω 4...20 mA
6
1 2 3 4 5
A0011340

7 Terminal assignment 4-wire; 4-20mA HART (10.4 to 48 VDC)

1 Evaluation unit, e.g. PLC


2 HART communication resistor (≥ 250 Ω): Observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device: Observe maximum load
5 Signal cable including screening (if required), observe cable specification
6 Protective connection; do not disconnect!
7 Protective earth, observe cable specification
8 4...20mA HART (active): Terminals 3 and 4
9 Supply voltage: Terminals 1 and 2
10 Supply voltage: Observe terminal voltage, observe cable specification
11 Terminal for potential equalization
12 Cable entry for signal line
13 Cable entry for power supply

Endress+Hauser 25
Levelflex FMP51, FMP52, FMP54

4-wire: 4-20mA HART (90 to 253 VAC)

13
12
11

! 2- w
4-2 ire
0
HA mA
[09 RT
A ]

~
3V
ope
n

..25
Sp
a
711 re pa

HA 20 mA
9 0.
08x rt
xx
L
1 N

RT
4...
2 +
10 9
3 -

mm
4 8

10
7

+ !
-
≥250 Ω 4...20 mA
6
1 2 3 4 5
A0018965

8 Terminal assignment 4-wire; 4-20mA HART (90 to 253 VAC)

1 Evaluation unit, e.g. PLC


2 HART communication resistor (≥ 250 Ω): Observe maximum load
3 Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
4 Analog display device: Observe maximum load
5 Signal cable including screening (if required), observe cable specification
6 Protective connection; do not disconnect!
7 Protective earth, observe cable specification
8 4...20mA HART (active): Terminals 3 and 4
9 Supply voltage: Terminals 1 and 2
10 Supply voltage: Observe terminal voltage, observe cable specification
11 Terminal for potential equalization
12 Cable entry for signal line
13 Cable entry for power supply

LCAUTION
To ensure electrical safety:
‣ Do not disconnect the protective connection (6).
‣ Disconnect the supply voltage before disconnecting the protective earth (7).
Connect protective earth to the internal ground terminal (7) before connecting the supply
voltage. If necessary, connect the potential matching line to the external ground terminal (11).
In order to ensure electromagnetic compatibility (EMC): Do not only ground the device via the
protective earth conductor of the supply cable. Instead, the functional grounding must also be
connected to the process connection (flange or threaded connection) or to the external ground
terminal.
An easily accessible power switch must be installed in the proximity of the device. The power
switch must be marked as a disconnector for the device (IEC/EN61010).

26 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

PROFIBUS PA / FOUNDATION Fieldbus

! 2-
F4IEwire
-20L lev
mD
ABU el
FA
P IE/F S
06/ LD
[2 F B PFS
A 07]
2 U S
op en Sp
a
7110 re pa

US
0283x4 rt
3+ ++ x5x7

B
31 --

/LFDF
PIAE
42 +

F
2
1 2-

mm
4-

10
3
1

B
3+ 6 +
3 -
4-2 4 +
0m
2-c
A/ 1 -
han
nel 4-2 2
4- [17
ove F IEL 0m
A
] rvo
ltag DB /
1 ep US
rote
ctio
n

A0011341

9 Terminal assignment PROFIBUS PA / FOUNDATION Fieldbus


A Without integrated overvoltage protection
B With integrated overvoltage protection
1 Cable screen: Observe cable specifications
2 Switch output (open collector): Terminals 3 and 4
3 PROFIBUS PA / FOUNDATION Fieldbus: Terminals 1 and 2
4 Terminal for potential equalization line
5 Cable entries
6 Overvoltage protection module

Endress+Hauser 27
Levelflex FMP51, FMP52, FMP54

Connection examples for the switch output


For HART devices, the switch output is available as an option. See product structure, feature 20:
"Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output"
Devices with PROFIBUS PA and FOUNDATION Fieldbus always have a switch output.

+
3+ 1
2
Ri + 3+
-
Ri
4-
A0015909
4-
 10 Connection of a relay A0015910

 11 Connection of a digital input


Suitable relays (examples):
• Solid-state relay: Phoenix Contact OV-24DC/480AC/5 with mounting 1 Pull-up resistor
rail connector UMK-1 OM-R/AMS 2 Digital input
• Electromechanical relay: Phoenix Contact PLC-RSC-12DC/21

For optimum interference immunity we recommend to connect an external resistor (internal


resistance of the relay or Pull-up resistor) of < 1 000 Ω.

28 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

HART loop converter HMX50


The dynamic variables of the HART protocol can be converted into individual 4 to 20 mA sections
using the HART loop converter HMX50. The variables are assigned to the current output and the
measuring ranges of the individual parameters are defined in the HMX50.

HART

+1 10
+2 11 HMX50
12
-3
16
4
17
5 18
6
7
8
9

13
14

15

19
20

21
FXA195

22 23 24 HART

24V DC
A0023287

 12 Connection diagram for HART loop converter HMX50 (example: passive 2-wire device and current outputs
connected as power source)

The HART loop converter HMX50 can be acquired using the order number 71063562.
Additional documentation: TI00429F and BA00371F.

Endress+Hauser 29
Levelflex FMP51, FMP52, FMP54

Device plug connectors For the versions with fieldbus plug connector (M12 or 7/8"), the signal line can be connected
without opening the housing.

Pin assignment of the M12 plug connector


Pin Meaning
4 3 1 Signal +

2 not connected
1 2 3 Signal -

A0011175
4 Ground

Pin assignment of the 7/8" plug connector


Pin Meaning
1 3 1 Signal -

2 Signal +
2 4 3 Not connected

A0011176
4 Screen

30 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Power supply An external power supply is required.


Various supply units can be ordered from Endress+Hauser.

2-wire, 4-20mA HART, passive

2-wire; 4-20mA HART 1)


"Approval" 2) Terminal voltage U at the device Maximum load R, depending on the supply voltage U0 at the
supply unit

• Non-Ex 11.5 to 35 V 3) R [W]


• Ex nA
• Ex ic 500
• CSA GP

Ex ia / IS 11.5 to 30 V

0
10 20 30 35 U0 [V]
11.5 22.5
A0035511

• Ex d / XP 13.5 to 30 V 4) R [Ω]
• Ex ic[ia]
• Ex tD / DIP
500

0
U0 [V]
10 20 30
13.5 24.5
A0034969

1) Feature 020 of the product structure: option A


2) Feature 010 of the product structure
3) For ambient temperatures Ta≤ -30 °C (-22 °F) a minimum voltage of 14 V is required for the sartup of the device at the MIN error current (3,6
mA). For ambient temperatures Ta ≥ 60 °C (140 °F) a minimum voltage of 12V is required for the sartup of the device at the MIN error current
(3,6 mA). The startup current can be parametrized. If the device is operated with a fixed current I ≥ 4,5 mA (HART multidrop mode), a voltage of
U ≥ 11,5 V is sufficient throughout the entire range of ambient temperatures.
4) For ambient temperatures Ta ≤ -20 °C (-4 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).

Endress+Hauser 31
Levelflex FMP51, FMP52, FMP54

2-wire; 4-20 mA HART, switch output 1)


"Approval" 2) Terminal voltage U at the device Maximum load R, depending on the supply voltage U0 at the
supply unit

• Non-Ex 13.5 to 35 V 3) R [Ω]


• Ex nA
• Ex nA[ia]
500
• Ex ic
• Ex ic[ia]
• Ex d[ia] / XP
• Ex ta / DIP
• CSA GP

• Ex ia / IS 13.5 to 30 V 3)
• Ex ia + Ex d[ia] / IS + XP 0
10 20 30 U0 [V]
13.5 24.5 35
A0034971

1) Feature 020 of the product structure: option B


2) Feature 010 of the product structure
3) For ambient temperatures Ta≤ -30 °C (-22 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).

2-wire; 4-20mA HART, 4-20mA 1)


"Approval" 2) Terminal voltage U at the device Maximum load R, depending on the supply voltage U0 at the
supply unit

any Channel 1:

13.5 to 30 V 3) 4) R [Ω]

500

0
U0 [V]
10 20 30
13.5 24.5
A0034969

Channel 2:

12 to 30 V R [W]

500

0
10 20 30 U0 [V]
12 23
A0022583

1) Feature 020 of the product structure: option C


2) Feature 010 of the product structure
3) For ambient temperatures Ta ≤ -30 °C (-22 °F) a minimum voltage of 16 V is required for the startup of the device at the MIN error current (3.6
mA).
4) For ambient temperatures Ta ≤ -40 °C (-40 °F), the maximum terminal voltage must be restricted to U ≤ 28 V.

32 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Polarity reversal Yes


protection

Admissible residual ripple USS < 1 V


at f = 0 to 100 Hz

Admissible residual ripple USS < 10 mV


at f = 100 to 10000 Hz

Endress+Hauser 33
Levelflex FMP51, FMP52, FMP54

4-wire, 4-20mA HART, active

"Power supply; Output" 1) Terminal voltage Maximum load Rmax

K: 4-wire 90-253VAC; 4-20mA HART 90 to 253 VAC (50 to 60 Hz), overvoltage 500 Ω
category II

L: 4-wire 10,4-48VDC; 4-20mA HART 10.4 to 48 VDC

1) Feature 020 of the product structure

PROFIBUS PA, FOUNDATION Fieldbus

"Power supply; Output" 1) "Approval" 2) Terminal voltage


3)
E: 2-wire; FOUNDATION Fieldbus, switch output • Non-Ex 9 to 32 V
G: 2-wire; PROFIBUS PA, switch output • Ex nA
• Ex nA[ia]
• Ex ic
• Ex ic[ia]
• Ex d[ia] / XP
• Ex ta / DIP
• CSA GP
3)
• Ex ia / IS 9 to 30 V
• Ex ia + Ex d[ia] / IS + XP

1) Feature 020 of the product structure


2) Feature 010 of the product structure
3) Input voltages up to 35 V will not spoil the device.

Polarity sensitive No

FISCO/FNICO compliant Yes


according to IEC 60079-27

Power consumption "Power supply; Output" 1) Power consumption

A: 2-wire; 4-20mA HART < 0.9 W

B: 2-wire; 4-20mA HART, switch output < 0.9 W

C: 2-wire; 4-20mA HART, 4-20mA < 2 x 0.7 W

K: 4-wire 90-253VAC; 4-20mA HART 6 VA

L: 4-wire 10,4-48VDC; 4-20mA HART 1.3 W

1) Feature 020 of the product structure

Current consumption HART

Nominal current 3.6 to 22 mA, the start-up current for multidrop mode can be parametrized (is
set to 3.6 mA on delivery)

Breakdown signal adjustable: 3.59 to 22.5 mA


(NAMUR NE43)

PROFIBUS PA

Nominal current 14 mA

Error current FDE (Fault 0 mA


Disconnection Electronic)

34 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

FOUNDATION Fieldbus

Device basic current 15 mA

Error current FDE (Fault 0 mA


Disconnection Electronic)

FISCO

Ui 17.5 V

Ii 550 mA

Pi 5.5 W

Ci 5 nF

Li 10 μH

Power supply failure • Configuration is retained in the HistoROM (EEPROM).


• Error messages (incl. value of operated hours counter) are stored.

Potential equalization No special measures for potential equalization are required.


If the device is designed for hazardous areas, observe the information in the documentation
"Safety Instructions" (XA).

Terminals • Without integrated overvoltage protection


Plug-in spring terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• With integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)

Cable entries Connection of power supply and signal line


To be selected in feature 050 "Electrical connection"
• Gland M20; Material dependent on the approval:
– For Non-Ex, ATEX, IECEx, NEPSI Ex ia/ic:
Plastics M20x1.5 for cable ⌀5 to 10 mm (0.2 to 0.39 in)
– For Dust-Ex, FM IS, CSA IS, CSA GP, Ex nA:
– For Ex d:
No gland available
• Thread
– ½" NPT
– G ½"
– M20 × 1.5
• Plug M12 / Plug 7/8"
Only available for Non-Ex, Ex ic, Ex ia

Connection of remote display FHX50


Dependent on feature 030: "Display, Operation":
• "Prepared for display FHX50 + M12 connection":
M12 socket
• "Prepared for display FHX50 + custom connection":
M16 cable gland

Cable specification • Devices without integrated overvoltage protection


Pluggable spring-force terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Devices with integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)
• For ambient temperature TU≥60 °C (140 °F): use cable for temperature TU +20 K.

Endress+Hauser 35
Levelflex FMP51, FMP52, FMP54

HART
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the
plant.
• For 4-wire devices: Standard device cable is sufficient for the power line.

PROFIBUS
Use a twisted, screened two-wire cable, preferably cable type A.
For further information on the cable specifications, see Operating Instructions BA00034S
"PROFIBUS DP/PA: Guidelines for planning and commissioning", PNO Guideline 2.092
"PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP).

FOUNDATION Fieldbus
Endress+Hauser recommends using twisted, shielded two-wire cables.
For further information on the cable specifications, see Operating Instructions BA00013S
"FOUNDATION Fieldbus Overview", FOUNDATION Fieldbus Guideline and IEC 61158-2 (MBP).

Overvoltage protection If the measuring device is used for level measurement in flammable liquids which requires the use of
overvoltage protection according to DIN EN 60079-14, standard for
test procedures 60060-1 (10 kA, pulse 8/20 μs), an overvoltage protection module has to be
installed.

Integrated overvoltage protection module


An integrated overvoltage protection module is available for 2-wire HART as well as PROFIBUS PA
and FOUNDATION Fieldbus devices.
Product structure: Feature 610 "Accessory mounted", option NA "Overvoltage protection".

Technical data

Resistance per channel 2 × 0.5 Ω max.

Threshold DC voltage 400 to 700 V

Threshold impulse voltage < 800 V

Capacitance at 1 MHz < 1.5 pF

Nominal arrest impulse voltage (8/20 μs) 10 kA

External overvoltage protection module


HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection.
For detailed information please refer to the following documents:
• HAW562: TI01012K
• HAW569: TI01013K

36 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Performance characteristics
Reference operating • Temperature = +24 °C (+75 °F) ±5 °C (±9 °F)
conditions • Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi)
• Humidity = 60 % ±15 %
• Reflection factor ≥ 0,8 (water surface for coax probe, metal plate for rod and rope probe with min.
1 m (40 in) diameter)
• Flange for rod or rope probe ≥ 300 mm (12 in) diameter
• Distance to obstacles ≥ 1 m (40 in)
• For interface measurement:
– Coax probe
– DC of the lower medium = 80 (water)
– DC of the upper medium = 2 (oil)

Reference accuracy Typical data under reference operating conditions: DIN EN IEC 61298-2 / DIN EN IEC 60770-1,
percentage values in relation to the span.

Output: digital analog 1)

Accuracy (Sum of non- Level measurement: ±0.02 %


linearity, nonrepeatability • Measuring distance up to 15 m (49 ft): ±2 mm (±0.08 in) 3)
and hysteresis) 2) • Measuring distance > 15 m (49 ft): ±10 mm (±0.39 in)

Interface measurement:
• Measuring distance up to 500 mm (19.7 in):
±20 mm (±0.79 in)
• Measuring distance > 500 mm (19.7 in): ±10 mm (±0.39 in)
• If the thickness of the upper medium is < 100 mm (3.94 in):
±40 mm (±1.57 in)

Non-repeatability 4) ≤ 1 mm (0.04 in)

1) Add error of the analogous value to the digital value.


2) If the reference conditions are not met, the offset/zero point arising from the mounting conditions may be
up to ±16 mm (±0.63 in). This additional offset/zero point can be compensated for by entering a
correction (parameter "level correction") during commissioning.
3) For probes with centering stars, the accuracy may deviate close to the centering stars.
4) The non-repeatability is already considered in the accuracy.

Endress+Hauser 37
Levelflex FMP51, FMP52, FMP54

Differing from this, the following measuring error is present in the vicinity of the lower probe
end:

D
80 (3.15)
60 (2.36) DC = 2
40 (1.57)
DC = 80
20 (0.79)
0
-20 (-0.79)
-40 (-1.57)
-60 (-2.36)
-80 (-3.15)

50 (1.97)

100 (3.94)

150 (5.91)

250 (9.84)

300 (11.8)
A

20 (0.79)

200 (7.87)
40 (1.57)

80 (3.15)

A0021480

 13 Measuring error at the end-of-probe for rod and coax probes


A Distance from probe end [mm(in)]
D Measuring error: Sum of non-linearity, non-repeatability and hysteresis

D
80 (3.15)
60 (2.36) DC = 80
40 (1.57)
20 (0.79)
0
-20 (-0.79)
-40 (-1.57)
-60 (-2.36)
-80 (-3.15)
50 (1.97)

100 (3.94)

150 (5.91)

200 (7.87)

300 (11.8)

A
0
20 (0.79)

250 (9.84)

A0021482

 14 Measuring error at the end-of-probe for rope probes


A Distance from probe end
D Measuring error: Sum of non-linearity, non-repeatability and hysteresis

38 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

D
120 (4.72)
100 (3.94)
80 (3.15) DC = 2
60 (2.36) DC = 80
40 (1.57)
20 (0.79)
0
-20 (-0.79)
-40 (-1.57)
-60 (-2.36)
-80 (-3.15)
-100 (-3.94)
-120 (-4.72)

50 (1.97)

150 (5.91)

300 (11.8)
A

0
30 (0.79)

100 (3.94)

200 (7.87)

250 (9.84)
A0021483

 15 Measuring error at the end-of-probe for probes with metallic centering disk (product structure: feature
610 "Accessory mounted", option OA, OB or OC)
A Distance from probe end [mm(in)]
D Measuring error: Sum of non-linearity, non-repeatability and hysteresis

If for rope probes the DC value is less than 7, then measurement is not possible in the area of
the straining weight (0 to 250 mm from end of probe; lower blocking distance).

Endress+Hauser 39
Levelflex FMP51, FMP52, FMP54

In the area of the upper probe end, the measuring error is as follows (rod/rope only):

R
D

40 (1.57)

30 (1.18) DC = 2
DC = 80

10 (0.39)

2 (0.08)
0 (0)
- 2 (- 0.08)

- 10 (-0.39)

- 30 (- 1.18)

- 40 (- 1.57)

100 (3.94) 200 (7.87)


A0015091

 16 Measuring error at the upper end of the probe; dimensions: mm (in)


D Sum of non-linearity, non-repeatability and hysteresis
R Reference point of measurement
DC Dielectric constant

Resolution • digital: 1 mm
• analog: 1 μA

Reaction time The reaction time can be parametrized. The following step response times (as per DIN EN IEC
61298-2 / DIN EN IEC 60770-1) 4) are valid if the damping is switched off:

Level measurement

Probe length Sampling rate Step response time

< 10 m (33 ft) 3.6 measurements/second < 0.8 s

< 40 m (131 ft) ≥ 2.7 measurements/second <1s

Interface measurement

Probe length Sampling rate Step response time

< 10 m (33 ft) ≥ 1.1 measurements/second < 2.2 s

Influence of ambient The measurements are carried out in accordance with DIN EN IEC 61298-3 / DIN EN IEC
temperature 60770-1
• digital (HART, PROFIBUS PA, FOUNDATION Fieldbus): average TK = 0.6 mm/10 K
For devices with remote sensor 5) there is an additional offset of ±0.3 mm/10K (±0.01 in/10K) per
1 m (3.3 ft) of the remote cable.
• analog (current output):
– zero point (4 mA): average TK = 0.02 %/10 K
– span (20 mA): average TK = 0.05 %/10 K

4) According to DIN EN IEC 61298-2 / DIN EN IEC 60770-1 the response time is the time which passes after a sudden change of the input signal
until the output signal for the first time assumes 90% of the steady-state value.
5) Product structure: Feature 600, options MB, MC or MD)

40 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Influence of gas layer High pressures reduce the propagation velocity of the measuring signals in the gas/vapor above the
fluid. This effect depends on the kind of gas/vapor and of its temperature. This results in a
systematic measuring error that gets bigger as the distance increases between the reference point of
the measurement (flange) and the product surface. The following table illustrates this measured
error for a few typical gases/vapors (with regard to distance; a positive value means that too large a
distance is being measured):

Gas layer Temperature Pressure

°C °F 1 bar (14.5 psi) 10 bar (145 psi) 50 bar (725 psi) 100 bar (1450 psi) 200 bar (2900 psi) 400 bar (5800 psi)

Air 20 68 0.00 % 0.22 % 1.2 % 2.4 % 4.9 % 9.5 %

200 392 -0.01 % 0.13 % 0.74 % 1.5 % 3.0 % 6.0 %

400 752 -0.02 % 0.08 % 0.52 % 1.1 % 2.1 % 4.2 %

Hydrogen 20 68 -0.01 % 0.10 % 0.61 % 1.2 % 2.5 % 4.9 %

200 392 -0.02 % 0.05 % 0.37 % 0.76 % 1.6 % 3.1 %

400 752 -0.02 % 0.03 % 0.25 % 0.53 % 1.1 % 2.2 %

Gas layer Temperature Pressure

°C °F 1 bar 2 bar 5 bar 10 bar 20 bar 50 bar 100 bar 200 bar
(14.5 psi) (29 psi) (72.5 psi) (145 psi) (290 psi) (725 psi) (1450 psi) (2900 psi)

Water 100 212 0.26 % - - - - - - -


(saturated
steam) 120 248 0.23 % 0.50 % - - - - - -

152 306 0.20 % 0.42 % 1.14 % - - - - -

180 356 0.17 % 0.37 % 0.99 % 2.10 % - - - -

212 414 0.15 % 0.32 % 0.86 % 1.79 % 3.9 % - - -

264 507 0.12 % 0.26 % 0.69 % 1.44 % 3.0 % 9.2 % - -

311 592 0.09 % 0.22 % 0.58 % 1.21 % 2.5 % 7.1 % 19.3 % -

366 691 0.07 % 0.18 % 0.49 % 1.01 % 2.1 % 5.7 % 13.2 % 76 %

Gas phase compensation PROFIBUS and FOUNDATION Fieldbus devices can receive the signal of an external pressure sensor
with external pressure through the bus and use it to perform a pressure dependent time-of-flight correction. In the case of
sensor (PROFIBUS PA / saturated steam in the temperature range from 100 to 350 °C (212 to 662 °f), for example, the
FOUNDATION Fieldbus) measuring error of the distance measurement can be reduced by this method from up to 29 %
(without compensation) to less than 3 % (with compensation).

Endress+Hauser 41
Levelflex FMP51, FMP52, FMP54

Gas phase compensation Under high pressure the propagation speed of microwave signals is reduced in the steam (polar
with reference signal (option media) above the liquid to be measured. As a result, the Levelflex indicates the level too low
for FMP54) →  41.
As an option Levelflex is available in a version with automatic gas phase correction, which corrects
this error (feature 540 "Application Package", option EF: "Gas Phase Compensation Lref= 300 mm" or
EG: "Gas Phase Compensation Lref= 550 mm"). This version of the Levelflex generates a reference
reflection in the distance Lref from the flange by a diameter step of the probe rod. The reference
reflection must be at least 150 mm (6") above the highest level. By means of the shift of the
reference reflection the actual propagation speed is measured and the level value will be
automatically corrected.

(11.8/21.7)
(11.8/21.7)

300/550
300/550

Lref
Lref

Dref
iD

LN
LN

max max
150 (5.91)

150 (5.91)
min min

A0014534

 17 FMP54 with reference signal for gas phase compensation; dimensions: mm (in)

Coax probes with reference reflection can be installed in any tank (free in the tank or into a
bypass). Coax probes are completely mounted and adjusted ex works. After mounting they are
ready for use, additional settings are not necessary.
Rod probes are only recommended if the installation of a coax probe is not possible (e.g. if the
bypass diameter is too small).
Rod probes with reference reflection are only suited for mounting in stilling wells and side
gauges (bypasses). The diameter Dref of the probe rod in the range of the reference distance Lref
must be chosen depending on the pipe inner diameter iD, see table below. In the range of the
reference distance Lref the pipe has to be cylindrical; changes of the cross section, for example at
flanged connections, are only allowed up to 5% of the inside diamter iD.
Additionally, the settings must be checked by an expert person after mounting and corrected if
necessary.

Inner diameter iD of the stilling well/bypass Diameter Dref of the rod probe within the reference distance
Lref

40 mm (1.57") ≤ iD < 45 mm (1.77") 22 mm (0.87")

45 mm (1.77") ≤ iD < 70 mm (2.76") 25 mm (0.98")

70 mm (2.76") ≤ iD < 100 mm (3.94") 30 mm (1.18")

42 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Limitations for coax/rod probes

Maximum probe length LN • For rod probes:


LN ≤ 4 000 mm (157 in)
• For coax probes:
LN ≤ 6 000 mm (236 in)

Minimum probe length LN LN > Lref + 200 mm (Lref + 7.7")

Reference distance Lref 300 mm (11.8") or 550 mm (21.7"), see feature 540 of the product
structure

Maximum level relative to sealing Lref + 150 mm (Lref + 6")


surface of flange

Minimum dielectric constant of the DC > 7


medium

Area of application
Level measurements with high pressure for measuring ranges up to a few meters in polar media with
a dielectric constant DC > 7 (e.g. water or ammonia), which would cause a high measuring error
without the compensation.
The accuracy of measurement at reference conditions is the higher the larger the reference length
Lref and the samller the measuring range:

45

40

35
Lref = 300 mm
30
Delta [mm]

25

20

Lref = 550 mm
15

10

5
0 500 1000 1500 2000 2500 3000 3500 4000
D [mm]

A0014535

D   Distance from liquid surface to lower edge of the flange


Delta  Measuring error

If there are fast changes in pressure, there may be an additional error, since the mesured reference
distance is filtered with the time constant of the level measurement. In addition, conditions of
imbalance - for example due to heating - may cause density gradients within the medium and
condensation of steam at the probe. As a result, the level readings at different locations inside the
tank may vary slightly. Caused by this application influences the measuring error may be increased
by a factor up to 2 to 3.

Endress+Hauser 43
Levelflex FMP51, FMP52, FMP54

Mounting
Mounting requirements Suitable mounting position

1 2 3

A 4

A0012606

 18 Mounting requirements for Levelflex

Mounting distances
• Distance (A) between wall and rod or rope probe:
– for smooth metallic walls: > 50 mm (2 in)
– for plastic walls: > 300 mm (12 in) to metallic parts outside the vessel
– for concrete walls: > 500 mm (20 in), otherwise the available measuring range may be reduced.
• Distance (B) between rod or rope probe and internal fittings in the vessel: > 300 mm (12 in)
• When using more than one Levelflex:
Minimum distance between the sensor axes: 100 mm (3.94 in)
• Distance (C) from end of probe to bottom of the vessel:
– Rope probe: > 150 mm (6 in)
– Rod probe: > 10 mm (0.4 in)
– Coax probe: > 10 mm (0.4 in)
For coax probes the distance to the wall and to internal fittings is arbitrary.

44 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Additional conditions
• When mounting in the open, a weather protection cover (1) may be installed to protect the device
against extreme weather conditions.
• In metallic vessels: Preferably do not mount the probe in the center of the vessel (2), as this would
lead to increased interference echoes.
If a central mounting position can not be avoided, it is crucial to perform an interference echo
suppresion(mapping) after the commissioning of the device.
• Do not mount the probe in the filling curtain (3).
• Avoid buckling the rope probe during installation or operation (e.g. through product movement
against silo wall) by selecting a suitable mounting location.
With suspended rope probes (probe end not fixed at the bottom) the distance between the
probe rope and internal fittings in the tank must not fall below 300 mm (12") during the entire
process. A sporadic contact between the probe weight and the cone of the vessel, however, does
not influence the measurement as long as the dielectric constant of the medium is at least DC =
1.8.
When mounting the electronics housing into a recess (e.g. in a concrete ceiling), observe a
minimum distance of 100 mm (4 inch) between the cover of the terminal compartment /
electronics compartment and the wall. Otherwise the connection compartment / electronics
compartment is not accessible after installation.

Endress+Hauser 45
Levelflex FMP51, FMP52, FMP54

Applications with restricted mounting space

Mounting with remote sensor


The device version with a remote sensor is suited for applications with restricted mounting space. In
this case the electronics housing is mounted at a separate position from which it is easier accessible.

A B

6 Nm 6 Nm
(4.42 lbf ft) (4.42 lbf ft)

rmin = 100 (4) C


C
rmin = 100 (4)

6 Nm 6 Nm
(4.42 lbf ft) (4.42 lbf ft)

A0014794

A Angled plug at the probe


B Angled plug at the electronics housing
C Length of the remote cable as ordered

• Product structure, feature 600 "Probe Design":


– Option MB "Sensor remote, 3m/9ft cable"
– Option MC "Sensor remote, 6m/18ft cable"
– Option MB "Sensor remote, 9m/27ft cable"
• The remote cable is supplied with these device versions
Minimum bending radius: 100 mm (4 inch)
• A mounting bracket for the electronics housing is supplied with these device versions. Mounting
options:
– Wall mounting
– Pipe mounting; diameter: 42 to 60 mm (1-1/4 to 2 inch)
• The connection cable has got one straight and one angled plug (90°). Depending on the local
conditions the angled plug can be connected at the probe or at the electronics housing.
Probe, electronics and connection cable are adjusted to match each other. They are marked by a
common serial number. Only components with the same serial number shall be connected to
each other.

46 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Divisible probes

14 mm

14 mm

A0021647

If there is little mounting space (distance to the ceiling), it is advisable to use divisible rod probes (
16 mm).
• max. probe length 10 m (394 in)
• max. sideways capacity 30 Nm
• probes are separable several times with the following lengths of the individual parts:
– 500 mm (20 in)
– 1 000 mm (40 in)
• torque: 15 Nm

Endress+Hauser 47
Levelflex FMP51, FMP52, FMP54

Notes on the mechanical load of the probe

Tensile load limit of rope probes

Sensor Feature 060 Probe Tensile load limit [kN]

FMP51 LA, LB Rope 4mm (1/6") 316 5


MB, MD

FMP52 OA, OB, OC, OD Rope 4mm (1/6") PFA>316 2

FMP54 LA, LB Rope 4mm (1/6") 316 10

Bending strength of rod probes

Sensor Feature 060 Probe Bending strength [Nm]

FMP51 AA, AB Rod 8mm (1/3") 316L 10

AC, AD Rod 12mm (1/2") 316L 30

AL, AM Rod 12mm (1/2") AlloyC 30

BA, BB, BC, BD Rod 16mm (0.63") 316L divisible 30

FMP52 CA, CB Rod 16mm (0.63") PFA>316L 30

FMP54 AE, AF Rod 16mm (0.63") 316L 30

BA, BB, BC, BD Rod 16mm (0.63") 316L divisible 30

Bending load (torque) through fluid flow


The formula for calculating the bending torque M impacting on the probe:
M = cw ⋅ ρ/2 ⋅ v2 ⋅ d ⋅ L ⋅ (LN - 0.5 ⋅ L)

with:
cw: Friction factor
ρ [kg/m3]: Density of the medium
v [m/s]: Velocity of the medium perpendicular to the probe rod
d [m]: Diameter of the probe rod
L [m]: Level
LN [m]: Probe length

Calculation example

Friction factor cw 0,9 (on the assumption of a turbulent current - high


Reynolds number)

Density ρ [kg/m3] 1000 (e.g. water)

Probe diameter d [m] 0,008

L = LN (worst case) LN
v L

d
A0014175

48 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Bending torque [M] on rod probes, diameter 8mm (1/3”)


20.0
18.0
16.0

Bending torque [Nm]

max. bending torque


14.0
12.0
10.0
8.0
6.0
4.0
2.0
0.0 v=0.5m/s
0.4 0.8 1.2 1.6 2 2.4 2.8 3.2 3.6 4 v=0.7m/s
Probe length [LN] in meters v=1.0m/s

A0014182-EN

Bending strength of coax probes

Sensor Feature 060 Process connection Probe Bending strength


[Nm]

FMP51 UA, UB Thread G¾ oder NPT¾ Coax 316L, Ø 21,3 60


mm

• Thread G1½ or Coax 316L, Ø 42,4 300


NPT1½ mm
• Flange

UC, UD Flange Coax AlloyC, Ø 300


42,4 mm

FMP54 UA, UB • Thread G1½ or Coax 316L, Ø 42,4 300


NPT1½ mm
• Flange

Endress+Hauser 49
Levelflex FMP51, FMP52, FMP54

Notes on the process connection


Probes are mounted to the process connection with threaded connections or flanges. If during this
installation there is the danger that the probe end moves so much that it touches the tank floor or
cone at times, the probe must, if necessary, be shortened and fixed down →  56.

Threaded connection

A0015121

 19 Mounting with threaded connection; flush with the container ceiling

Seal
The thread as well as the type of seal comply to DIN 3852 Part 1, screwed plug form A.
They can be sealed with the following types of sealing rings:
• Thread G3/4": According to DIN 7603 with the dimensions 27 x 32 mm
• Thread G1-1/2": According to DIN 7603 with the dimensions 48 x 55 mm
Please use a sealing ring according to this standard in the form A, C or D and of a material that is
resistant to the application.
For the length of the screwed plug refer to the dimensional drawing:
• FMP51: →  81
• FMP54: →  85

50 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Nozzle mounting

£150 (6)
H
ø £150 (6)
A0015122

H Length of the center rod or the rigid part of the rope probe

• Permissible nozzle diameter: ≤ 150 mm (6 in).


For larger diameters the near range measuring capability may be reduced.
For nozzles ≥ DN300: →  54.
• Permissible nozzle height 6): ≤ 150 mm (6 in).
For a larger height the near range measuring capability may be reduced.
Larger nozzle heights may be possible in special cases (see sections "Center rod for FMP51 and
FMP52" and "Rod extension/centering HMP40 for FMP54").
• The end of the nozzle should be flush with the tank ceiling in order to avoid ringing effects.
With thermally insulated vessels the nozzle should also be insulated in order to prevent
condensate formation.

6) Larger nozzle heights on request

Endress+Hauser 51
Levelflex FMP51, FMP52, FMP54

Center rod for FMP51 and FMP52


For rope probes it may be necessary to use a version with center rod in order to prevent the probe
rod from coming into contact with the nozzle wall. Probes with center rod are available for FMP51
and FMP52.

Probe Max. nozzle height (= length of the center rod) Option to be selected in feature 060 ("Probe")

FMP51 150 mm LA

6 inch LB

300 mm MB

12 inch MD

FMP52 150 mm OA

6 inch OC

300 mm OB

12 inch OD

52 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Rod extension/centering HMP40 for FMP54


For FMP54 with rope probes the rod extension/centering HMP 40 is available as an accessory
→  125. It has to be used if otherwise the probe rope comes into contact with the lower edge of
the nozzle.
For FMP54 with rope probes the rod extension/centering HMP 40 is available as an accessory. It has
to be used if otherwise the probe rope comes into contact with the lower edge of the nozzle.
This accessory consists of the extension rod corresponding to the nozzle height, on which a
centering disk is also mounted if the nozzles are narrow or when working in bulk solids. This
component is delivered separately from the device. Please order the probe length
correspondingly shorter.
Centering disks with small diameters (DN40 and DN50) may only be used if there is no
significant build-up in the nozzle above the disk. The nozzle must not become clogged by the
product.

Endress+Hauser 53
Levelflex FMP51, FMP52, FMP54

Installation in nozzles ≥ DN300


If installation in ≥ 300mm/12" nozzles is unavoidable, installation must be carried out in accordance
with the following sketch.

1
3
2

A0014199

1 Lower edge of the nozzle


2 Approx. flush with the lower edge of the nozzle (± 50 mm/2")
3 Plate
4 Pipe  150 to 180 mm (6 to 7 inch)

Nozzle diameter Plate diameter

300 mm (12") 280 mm (11")

≥ 400 mm (16") ≥ 350 mm (14")

54 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Mounting cladded flanges


For cladded flanges of FMP52, observe the following:
• Use flange screws according to the number of flange holes.
• Tighten the screws with the required torque (see table).
• Retighten the screws after 24 hours or after the first temperature cycle.
• Depending on process pressure and process temperature check and retighten the screws at
regular intervals.
Usually, the PTFE flange cladding also serves as a seal between the nozzle and the device
flange.

Flange size Number of screws Recommended torque [Nm]

minimum maximum

EN

DN40/PN40 4 35 55

DN50/PN16 4 45 65

DN50/PN40 4 45 65

DN80/PN16 8 40 55

DN80/PN40 8 40 55

DN100/PN16 8 40 60

DN100/PN40 8 55 80

DN150/PN16 8 75 115

DN150/PN40 8 95 145

ASME

1½"/150lbs 4 20 30

1½"/300lbs 4 30 40

2"/150lbs 4 40 55

2"/300lbs 8 20 30

3"/150lbs 4 65 95

3"/300lbs 8 40 55

4"/150lbs 8 45 70

4"/300lbs 8 55 80

6"/150lbs 8 85 125

6"/300lbs 12 60 90

JIS

10K 40A 4 30 45

10K 50A 4 40 60

10K 80A 8 25 35

10K 100A 8 35 55

10K 100A 8 75 115

Endress+Hauser 55
Levelflex FMP51, FMP52, FMP54

Securing the probe

Securing rope probes

B
2
A

A0012609

A Sag of the rope: ≥ 1 cm per 1m of the probe length (0.12 inch per 1 ft of the probe length)
B Reliably grounded end of probe
C Reliably isolated end of probe
1: Mounting and contact with a bolt
2 Mounting kit isolated

• The end of the probe needs to be secured under the following conditions:
if otherwise the probe sporadically comes into contact with the wall of the vessel, the outlet cone,
internal fittings or other parts of the installation.
• The end of probe can be secured at its internal thread
rope 4 mm (1/6"), 316: M 14
• The fixing must be either reliably grounded or reliably insulated. If it is not possible to mount the
probe weight with a reliably insulated connection, it can be secured using an isolated eyelet, which
is available as an accessory.
• In order to prevent an extremely high tensile load (e.g. due to thermal expansion) and the risk of
rope crack, the rope has to be slack. Make the rope longer than the required measuring range such
that there is a sag in the middle of the rope that is ≥ 1cm/(1 m rope length) [0.12 inch/(1 ft rope
length)].
Tensile load limit of rope probes: →  48

56 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Securing rod probes


• For WHG approvals: For probe lengths ≥ 3 m (10 ft) a support is required.
• In general, rod probes must be supported if there is a horizontal flow (e.g. from an agitator) or in
the case of strong vibrations.
• Rod probes may only be supported at the end of the probe.

øa

øb

»50 (1.97)
3

ø<25 (1.0)

»3 (0.12)

»50 (1.97) 6

mm (in)

A0012607

1 Probe rod, uncoated


2 Sleeve bored tight to ensure electrical contact between the rod and sleeve!
3 Short metal pipe, e.g. welded in place
4 Probe rod, coated
5 Plastic sleeve, e.g. PTFE, PEEK or PPS
6 Short metal pipe, e.g. welded in place

 probe  a [mm (inch)]  b [mm (inch)]

8 mm (1/3") < 14 (0.55) 8.5 (0.34)

12 mm (1/2") < 20 (0.78) 12.5 (0.52)

16 mm (0.63in) < 26 (1.02) 16.5 (0.65)

NOTICE
Poor grounding of the end of probe may cause measuring errors.
‣ Apply a narrow sleeve which has good electrical contact to the probe.
NOTICE
Welding may damage the main electronics module.
‣ Before welding: Ground the probe and dismount electronics.

Endress+Hauser 57
Levelflex FMP51, FMP52, FMP54

Securing coax probes


For WHG approvals: For probe lengths ≥ 3 m (10 ft) a support is required.

A0012608

Coax probes can be supported at any point of the outer tube.

58 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Special mounting conditions

Bypasses and stilling wells


In bypass and stilling well applications it is recommended to use a centering disks or stars.

A B

C
1

ød
2

(3.94)
³100
3 ød

(3.94)
³100
øD
2
4

A0012615

 20 Dimensions: mm (in)
A Mounting in a stilling well
B Mounting in a bypass
C Center washer or centering star
1 Metallic center washer (316L) for level measurement
2 Fixing screw; torque: 25 Nm ± 5 Nm
3 Non-metallic centering star (PEEK, PFA) for interface measurement
4 Minimum distance between end of probe and lower edge of the bypass; see table below

Allocation of probe type and center washer or centering star to pipe diameter
Feature 610 - Accessory mounted

Center washer Pipe


Application Option Type of probe Centering star

 d [mm (in)] Material  D [mm (in)]

Level measurement OA Rod probe 75 (2,95) 316L DN80/3" to DN100/4"

OB Rod probe 45 (1,77) 316L DN50/2" to DN65/2½"

OC Rope probe 75 (2,95) 316L DN80/3" to DN100/4"


1)
Level or interface OD Rod probe 48...95 (1,89...3,74) PEEK ≥ 50 mm (2")
measurement 2)
OE Rod probe 37 (1,46) PFA ≥ 40 mm (1.57")

1) Operation temperature: –60 to +250 °C (–76 to 482 °F)


2) Operation temperature: –200 to +250 °C (–328 to +482 °F)

Minimum distance between end of probe and lower edge of the bypass
Type of probe Minimum distance

Rope 10 mm (0.4 in)

Rod 10 mm (0.4 in)

Coax 10 mm (0.4 in)

Endress+Hauser 59
Levelflex FMP51, FMP52, FMP54

• Pipe diameter: > 40 mm (1.6") for rod probes


• Rod probe installation can take place up to a diameter size of 150 mm (6 in). In the event of larger
diameters, a coax probe is recommended.
• Side disposals, holes or slits and welded joints that protrude up to approx. 5 mm (0.2") inwards do
not influence the measurement.
• The pipe may not exhibit any steps in diameter.
• The probe must be 100 mm longer than the lower disposal.
• Within the measuring range, the probe must not get into contact with the pipe wall. If necessary,
secure the probe by retaining or tensioning. All rope probes are prepared for tensioning in
containers (tensioning weight with anchor hole).
• If a metallic center washer is mounted at the end of the probe, it enables a reliable recognition of
the end-of-probe signal (see feature 610 of the product structure).
Note: For interface measurements only use the nonmetallic centering star made of PEEK or PFA
(feature 610, options OD or OE).
• Coax probes can always be applied if there is enough mounting space.
For bypasses with condensate formation (water) and a medium with low dielectric constant
(e.g. hydrocarbons):
In the course of time the bypass is filled with condensate up to the lower disposal and for low
levels the the level echo is superimposed by the condensate echo. Thus in this range the
condensate level is measured instead of the correct level. Only higher levels are measured
correctly. To prevent this, position the lower disposal 100 mm (4 in) below the lowest level to
be measured and apply a metallic centering disk at the height of the lower edge of the lower
disposal.
With heat insulated tanks the bypass should also be insulated in order to prevent condensate
formation.
For information on bypass solutions from Endress+Hauser please contact your Endress+Hauser
sales representative.

60 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Installation in horizontal and upright cylindrical tanks

A0014141

• Any distance from wall, as long as occasional contact is prevented.


• When installing in tanks with a lot of internals or internals situated close to the probe: use a coax
probe.

Endress+Hauser 61
Levelflex FMP51, FMP52, FMP54

Underground tanks

A0014142

Use a coax probe for nozzles with large diameters in order to avoid reflections at the nozzle wall.

62 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Installation at an angle

LN
a

A0014145

• For mechanical reasons, the probe should be installed as vertically as possible.


• With inclined installations the probe length has to be adjusted in dependence to the installation
angle.
– Up to LN = 1 m (3.3 ft): α = 30°
– Up to LN = 2 m (6.6 ft): α = 10°
– Up to LN = 4 m (13.1 ft): α = 5°

Endress+Hauser 63
Levelflex FMP51, FMP52, FMP54

Non-metallic vessels

A0012527

1 Non-metallic vessel
2 Metal sheet or metal flange

To ensure reliable measurements in non-metallic vessels:


• Select an instrument version with metal flange (minimum size DN50/2").
• Or: mount a metal sheet with a diameter of at least 200 mm (8 in) to the probe at the process
connection. Its orientation must be perpendicular to the probe.
A metallic surface is not required for coax probes.

64 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Plastic or glass tanks: Mounting the probe externally at the wall

1
3

A0014150

1 Plastic or glass tank


2 Metall sheet with threaded sleeve
3 No free space between tank wall and probe!

Requirements
• The dielectric constant of the medium must be at least DC > 7.
• The tank wall must be non-conductvie.
• Maximum wall thickness (a):
– Plastic: < 15 mm (0.6")
– Glass: < 10 mm (0.4")
• There may be no metallic reinforcements fixed to the tank.
Mounting conditions:
• The probe must be mounted directly to the tank wall (no open space)
• A plastic half pipe with a diameter of approx. 200 mm (8"), or some other protective unit, must be
affixed externally to the probe to prevent any influences on the measurement.
• If the tank diameter is less than 300 mm (12"):
A metallic grounding sheet must be installed at the opposite side of the tank. The sheet must be
conductively connected to the process connection and cover about the half of the vessel's
circumference.
• If the tank diameter exceeds 300 mm (12"):
A metal sheet with a diameter of at least 200 mm (8") must be mounted to the probe at the
process connection. Its orientation must be perpendicular to the probe (see above).

Endress+Hauser 65
Levelflex FMP51, FMP52, FMP54

Vessels with heat insulation


If process temperatures are high, the device must be included in normal tank insulation to
prevent the electronics heating up as a result of heat radiation or convection. The insulation
may not exceed beyond the points labeled "MAX" in the drawings.

2 3

1
MAX MAX

A0014653

 21 Process connection with thread - FMP51


1 Tank insulation
2 Compact device
3 Sensor remote (feature 600)

2 3
40 (1.57)

40 (1.57)
1
MAX MAX

mm (in)

A0014654

 22 Process connection with flange - FMP51, FMP52


1 Tank insulation
2 Compact device
3 Sensor remote (feature 600)

XT 280 °C (536 °F)* HT 450 °C (842 °F)


2 3
2 3
MAX MAX MAX MAX
1 1

A0014657

 23 Process connection with thread - FMP54, sensor version XT and HT


1 Tank insulation
2 Compact device
3 Sensor remote (feature 600)
* The XT version is not recommended for saturated steam above 200 °C (392 °F). Use the HT version instead.

66 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

XT 280 °C (536 °F)* HT 450 °C (842 °F)

2 3
2 3
MAX MAX MAX MAX
1 1

A0014658

 24 Process connection with flange - FMP54, sensor version XT and HT


1 Tank insulation
2 Compact device
3 Sensor remote (feature 600)
* The XT version is not recommended for saturated steam above 200 °C (392 °F). Use the HT version instead.

Endress+Hauser 67
Levelflex FMP51, FMP52, FMP54

Replacing a displacer system in an existing displacer chamber


FMP51 and FMP54 are a perfect replacement for a conventional displacer system in an existing
displacer chamber. Endress+Hauser offers flanges that suit Fisher and Masoneilan displacer
chamber for this purpose (special product for FMP51; feature 100, options LNJ, LPJ, LQJ for FMP54).
Thanks to menu-guided local operation, commissioning the Levelflex only takes a few minutes.
Replacement is also possible when partially filled, and wet calibration is not required.
Your benefits:
• No moving parts, thus zero-maintenance operation.
• Not sensitive to process influences such as temperature, density, turbulence and vibrations.
• The rod probes can be shortened or replaced easily. In this way, the probe can be easily adjusted on
site.

A0014153

1 Flange of the displacer chamber

Planning instructions:
• In normal cases, use a rod probe. When installing into a metallic displacer chamber up to 150 mm,
you have all the advantages of a coax probe.
• It must be ensured that the probe does not come into contact with the side wall. Where necessary,
use a center washer or centering star at the lower end of the probe (feature 610 of the product
structure).
• The center washer or centering star must be adapted as accurately as possible to the internal
diameter of the displacer chamber to also ensure perfect operation in the area of the probe end.
Additional information on interface measurement
• In the case of oil and water the centering disk or centering star should be positioned at the lower
edge of the lower disposal (water level).
• The pipe may not exhibit any steps in diameter. Use the coax probe where necessary.
• In the case of rod probes, it must be ensured that the probe does not come into contact with the
wall. If necessary, use a center washer or centering star at the end of the probe.
• A plastic centering star has to be used for interface measurement (feature 610, options OD and
OE).

68 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Operating conditions: Environment


Ambient temperature range Measuring device –40 to +80 °C (–40 to +176 °F)

Measuring device (option –50 to +80 °C (–58 to +176 °F)


for FMP51 and FMP54)

Local display –20 to +70 °C (–4 to +158 °F), the readability of the display may be impaired at
temperatures outside the temperature range.

Connection cable (for max. 100 °C (212 °F)


"Probe Design" = "Sensor
remote")

Remote display FHX50 –40 to 80 °C (–40 to 176 °F)

Remote display FHX50 –50 to 80 °C (–58 to 176 °F) 1)


(option)

1) This range is valid if option JN "Ambient temperature transmitter –50 °C (–58 °F)" has been selected in
ordering feature 580 "Test, Certificate". If the temperature is permanently below –40 °C (–40 °F), failure
rates may be increased.

When operating the device in the open with strong sunlight:


• Mount the device in a shady position.
• Avoid direct sunlight, especially in warmer regions.
• Use a weather protection cover (see accessories).

Ambient temperature limits The following diagrams take into account only function requirements. There may be further
restrictions for certified device versions. Please refere to the separate Safety Instructions →  134.

Endress+Hauser 69
Levelflex FMP51, FMP52, FMP54

With a temperature (Tp) at the process connection the admissible ambient temperature (Ta) is
reduced according to the following diagram (temperature derating):

Temperature derating for FMP51 with threaded connection G¾ or NPT¾

[°C] ([°F])Ta
Ta
+80 (+176)
GT20: +58(+136)
Tp
GT18: +48(+118)
A: 4...20 mA HART GT19: +29(+84)

-40/-50 (-40/-58)³ Tp
-40 +82 +200 [°C]
(-40) (+180) (+392) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +54(+129)
C: 4...20 mA HART
4...20 mA GT18: +48(+118)

K: 90...253 VAC GT19: +29(+84)

L: 10.4...48 VDC
-40/-50 (-40/-58)³ Tp
-40 +74 +200 [°C]
(-40) (+165) (+392) ([°F])
[°C] ([°F])Ta
+81 (+178)
GT20: +57(+134)
GT18: +48(+118)

G 1: PROFIBUS PA
FOUNDATION Fieldbus
-40 (-40) Tp
-40 +81 +200 [°C]
(-40) (+178) (+392) ([°F])
[°C] ([°F])Ta
+79 (+174)
GT20: +54(+129)
GT18: +48(+118)
2
G : PROFIBUS PA
FOUNDATION Fieldbus
Switch output
-40 (-40) Tp
-40 +79 +200 [°C]
(-40) (+174) (+392) ([°F])
A0013687

GT18 = stainless steel housing A = 1 current output Ta = ambient temperature 3)


GT19 = plastic housing C = 2 current outputs Tp = temperature at the process connection
GT20 = aluminum housing G1, G2 = PROFIBUS PA 1) 2)
K, L = 4-wire

1) G1: switch output not used


2) G2: Switch output used
3) Ta may be as low as –50 °C (–58 °F) for ordering feature 580 "Test,Certificate" = JN "Ambient temperature
transmitter –50 °C (–58 °F)"; only available for 2-wire HART devices

70 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP51 with threaded connection G1½ or NPT1½

[°C] ([°F])Ta
Ta
+80(+176)
GT20: +57(+135)
Tp
GT18: +49(+120)

GT19: -34(-29) A: 4...20 mA HART GT19: +20(+68)


GT18: -37(-34)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +82 +200 [°C]
(-58) (-40) (+180) (+392) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +54(+129)
C: 4...20 mA HART
4...20 mA GT18: +49(+120)

K: 90...253 VAC +20(+68)

GT19: -34(-29) L: 10.4...48 VDC GT19:


GT18: -37(-34)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +74 +200 [°C]
(-58) (-40) (+165) (+392) ([°F])
[°C] ([°F])Ta
+81 (+178)
GT20: +57(+134)
GT18: +49(+120)
G 1: PROFIBUS PA
FOUNDATION Fieldbus
GT18: -37(-35)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +81 +200 [°C]
(-58) (-40) (+178) (+392) ([°F])
[°C] ([°F])Ta
+79 (+174)
GT20: +54(+129)
2
PROFIBUS PA GT18: +49(+120)
G :
FOUNDATION Fieldbus
Switch output
GT18: -37(-35)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +79 +200 [°C]
(-58) (-40) (+174) (+392) ([°F])
A0014121

GT18 = stainless steel housing A = 1 current output Ta = ambient temperature 3)


GT19 = plastic housing C = 2 current outputs Tp = temperature at the process connection
GT20 = aluminum housing G1, G2 = PROFIBUS PA 1) 2)
K, L = 4-wire

1) G1: switch output not used


2) G2: Switch output used
3) Ta may be as low as –50 °C (–58 °F) for ordering feature 580 "Test,Certificate" = JN "Ambient temperature
transmitter –50 °C (–58 °F)"; only available for 2-wire HART devices

Endress+Hauser 71
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP51 with flange

[°C] ([°F])Ta
Ta
+80(+176)
GT20: +59(+138)
Tp
GT18: +53(+127)
GT19: -34(-29) A: 4...20 mA HART GT19: +33(+91)
GT18: -37(-34)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +82 +200 [°C]
(-58) (-40) (+180) (+392) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165)
GT20: +55(+131)
C: 4...20 mA HART GT18: +51(+124)
4...20 mA
GT19: +33(+91)
GT19: -34(-29) K: 90...253 VAC
GT18: -37(-34)
GT20: -38(-36) L: 10.4...48 VDC
-40(-40)³ Tp
-50 -40 +74 +200 [°C]
(-58) (-40) (+165) (+392) ([°F])
[°C] ([°F])Ta
+81 (+178)
GT20: +58(+136)
GT18: +53(+127)
1
G : PROFIBUS PA
FOUNDATION Fieldbus
GT18: -37(-35)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +81 +200 [°C]
(-58) (-40) (+178) (+392) ([°F])
[°C] ([°F])Ta
+79 (+174)
GT20: +55(+131)
2
PROFIBUS PA GT18: +51(+124)
G :
FOUNDATION Fieldbus
Switch output
GT18: -37(-35)
GT20: -38(-36)
-40(-40)³ Tp
-50 -40 +79 +200 [°C]
(-58) (-40) (+174) (+392) ([°F])
A0013689

GT18 = stainless steel housing A = 1 current output Ta = ambient temperature 3)


GT19 = plastic housing C = 2 current outputs Tp = temperature at the process connection
GT20 = aluminum housing G1, G2 = PROFIBUS PA 1) 2)
K, L = 4-wire

1) G1: switch output not used


2) G2: Switch output used
3) Ta may be as low as –50 °C (–58 °F) for ordering feature 580 "Test,Certificate" = JN "Ambient temperature
transmitter –50 °C (–58 °F)"; only available for 2-wire HART devices

72 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP52

T a [°C] ([°F])Ta
+80(+176)
GT20: +60(+140)
Tp GT18: +53(+127)
GT19: +38(+100)
GT19: -36(-33) A: 4...20 mA HART
GT18: -37(-35)
GT20: -38(-36)
-40(-40) Tp
-50 -40 +82 +200 [°C]
(-58) (-40) (+180) (+392) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +56(+133)
C: 4...20 mA HART GT18: +51(+124)
4...20 mA GT19: +38(+100)
GT19: -36(-33) K: 90...253 VAC
GT18: -37(-35)
GT20: -38(-36) L: 10.4...48 VDC
-40(-40) Tp
-50 -40 +74 +200 [°C]
(-58) (-40) (+165) (+392) ([°F])
[°C] ([°F])Ta

+81(+178)
GT20: +59(+138)
GT18: +53(+127)

G 1: PROFIBUS PA
FOUNDATION Fieldbus
GT18: -37(-35)
GT20: -38(-36)
-40(-40) Tp
-50 -40 +81 +200 [°C]
(-58) (-40) (+178) (+392) ([°F])
[°C] ([°F])Ta

+79(+174)
GT20: +56(+133)
GT18: +51(+124)
2
G : PROFIBUS PA
FOUNDATION Fieldbus
Switch output
GT18: -37(-35)
GT20: -38(-36)
-40(-40) Tp
-50 -40 +79 +200 [°C]
(-58) (-40) (+174) (+392) ([°F])
A0013633

GT18 = stainless steel housing A = 1 current output Ta = ambient temperature


GT19 = plastic housing C = 2 current outputs Tp = temperature at the process
GT20 = aluminum housing G1, G2 = PROFIBUS PA 1) connection 2)
K, L = 4-wire

1) For PROFIBUS PA and FOUNDATION Fieldbus the temperature derating depends on the usage of the
switch output. (G1: switch output not connected; G2: switch output connected).
2) For saturated steam applications the process temperature should not exceed 150 °C (302 °F). For higher
process temperatures use FMP54.

Endress+Hauser 73
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP54 - version XT up to +280 °C (+536 °F )

T a [°C] ([°F])Ta
+80(+176)
Tp GT20: +62(+144)
GT18: +56(+133)
GT19: +42(+108)
GT19: -5(+23) A: 4...20 mA HART
GT18: -16(+3)
GT20: -23(-9)
-40/-50 (-40/-58)³ Tp
-196 -40 +82 +280 [°C]
(-321) (-40) (+180) (+536) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +59(+138)
GT18: +52(+126)
C: 4...20 mA HART
GT19: +42(+108)
4...20 mA
GT19: -5(+23)
K: 90...253 VAC
GT18: -16(+3)
GT20: -23(-9) L: 10.4...48 VDC
-40/-50 (-40/-58)³ Tp
-196 -40 +74 +280 [°C]
(-321) (-40) (+165) (+536) ([°F])
[°C] ([°F])Ta
+81 (+178)
GT20: +61(+142)
GT18: +55(+130)
G 1: PROFIBUS PA
FOUNDATION Fieldbus
GT18: -16(+3)
GT20: -23(-9)
-40 (-40) Tp
-196 -40 +81 +280 [°C]
(-321) (-40) (+178) (+536) ([°F])
[°C] ([°F])Ta
+79 (+174)
GT20: +58(+137)
GT18: +55(+126)
2
G : PROFIBUS PA
FOUNDATION Fieldbus
Switch output
GT18: -16(+3)
GT20: -23(-9)
-40 (-40) Tp
-196 -40 +79 +280 [°C]
(-321) (-40) (+174) (+536) ([°F])
A0013631

GT18 = stainless steel housing A = 1 current output Ta = ambient temperature 3)


GT19 = plastic housing C = 2 current outputs Tp = temperature at the process connection
GT20 = aluminum housing G1, G2 = PROFIBUS PA 1) 2)
K, L = 4-wire

1) G1: switch output not used


2) G2: Switch output used
3) Ta may be as low as –50 °C (–58 °F) for ordering feature 580 "Test,Certificate" = JN "Ambient temperature
transmitter –50 °C (–58 °F)"; only available for 2-wire HART devices

74 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Temperature derating for FMP54 - version HT up to +450 °C (+842 °F )

[°C] ([°F])Ta
Ta
+80(+176)
GT20: +56(+133)
Tp GT18: +54(+129)

A: 4...20 mA HART GT19: +42(+108)


GT19: -22(-8)
GT18: -27(-17)
GT20: -28(-18)
-40/-50 (-40/-58)³ Tp
-196 -40 +82 +450 [°C]
(-321) (-40) (+180) (+842) ([°F])
[°C] ([°F])Ta +79
(+174)
GT18/20: +79 (+174)
GT19: +74 (+165) GT20: +53(+127)
C: 4...20 mA HART GT18: +51(+124)
4...20 mA GT19: +38(+100)
GT19: -22(-8) K: 90...253 VAC
GT18: -27(-17)
GT20: -28(-18) L: 10.4...48 VDC
-40/-50 (-40/-58)³ Tp
-196 -40 +74 +450 [°C]
(-321) (-40) (+165) (+842) ([°F])

[°C] ([°F])Ta

+81(+178)
GT20: +55(+131)
GT18: +53(+127)
G 1: PROFIBUS PA
FOUNDATION Fieldbus
GT18: -27(-16)
GT20: -28(-18)
-40 (-40) Tp
-196 -40 +81 +450 [°C]
(-321) (-40) (+178) (+842) ([°F])
[°C] ([°F])Ta

+79(+174)
GT20: +52(+126)
GT18: +50(+123)
2
G : PROFIBUS PA
FOUNDATION Fieldbus
Switch output
GT18: -27(-16)
GT20: -28(-18)
-40 (-40) Tp
-196 -40 +79 +450 [°C]
(-321) (-40) (+173) (+842) ([°F])
A0013632

GT18 = stainless steel housing A = 1 current output Ta = ambient temperature 3)


GT19 = plastic housing C = 2 current outputs Tp = temperature at the process connection
GT20 = aluminum housing G1, G2 = PROFIBUS PA 1) 2)
K, L = 4-wire

1) G1: switch output not used


2) G2: Switch output used
3) Ta may be as low as –50 °C (–58 °F) for ordering feature 580 "Test,Certificate" = JN "Ambient temperature
transmitter –50 °C (–58 °F)"; only available for 2-wire HART devices

Endress+Hauser 75
Levelflex FMP51, FMP52, FMP54

Storage temperature –40 to +80 °C (–40 to +176 °F)


Option for FMP51 and FMP54: –50 to +80 °C (–58 to +176 °F) 7)

Climate class DIN EN 60068-2-38 (test Z/AD)

Altitude according to • Generally up to 2 000 m (6 600 ft) above MSL.


IEC61010-1 Ed.3 • Above 2 000 m (6 600 ft) if the following conditions are met:
– Ordering feature 020 "Power supply; Output" = A, B, C, E or G (2-wire versions)
– Supply voltage U < 35 V
– Supply voltage of overvoltage category 1

Degree of protection • With closed housing tested according to:


– IP68, NEMA6P (24 h at 1.83 m under water surface) 8)
– For plastic housing with transparent cover (display module): IP68 (24 h at 1.00 m under water
surface) 9)
– IP66, NEMA4X
• With open housing: IP20, NEMA1
• Display module: IP22, NEMA2
Degree of protection IP68 NEMA6P applies for M12 PROFIBUS PA plugs only when the
PROFIBUS cable is plugged in and is also rated IP68 NEMA6P.

Vibration resistance DIN EN 60068-2-64 / IEC 60068-2-64: 20 to 2 000 Hz, 1 (m/s2)2/Hz

Cleaning the probe Depending on the application, contamination or buildup can accumulate on the probe. A thin, even
layer only influences measurement slightly. Thick layers can dampen the signal and then reduce the
measuring range. Severe, uneven buildup, adhesion e.g. through crystallization, can lead to incorrect
measurement. In this case, we recommend that you use a non-contact measuring principle, or check
the probe regularly for soiling.

Electromagnetic Electromagnetic compatibility to all relevant requirements of the EN 61326- series and NAMUR
compatibility (EMC) recommendation EMC (NE21). For details see declaration of conformity. 10). If only the analogue
signal is used, unshielded interconnection lines are sufficient for the installation. In case of using the
digital signal (HART/ PA/ FF) use shielded interconnection lines.
Use a shielded cable when working with a digital communications signal.
Max. fluctuations during EMC- tests: < 0.5 % of the span.
When installing the probes in metal and concrete tanks and when using a coax probe:
• Interference emission to EN 61326 - x series, electrical equipment Class B.
• Interference immunity to EN 61326 - x series, requirements for industrial areas and NAMUR
Recommendation NE 21 (EMC)
The measured value can be affected by strong electromagnetic fields when installing rod and rope
probes without a shielding/metallic wall, e.g. in plastic and wooden silos.
• Interference emission to EN 61326 - x series, electrical equipment Class A.
• Interference immunity: the measured value can be affected by strong electromagnetic fields.

7) This range is valid if option JN "Ambient temperature transmitter –50 °C (–58 °F)" has been selected in ordering features 580 "Test, Certificate". If
the temperature is permanently below –40 °C (–40 °F), failure rates may be increased.
8) also valid for the "Sensor remote" version
9) This restriction is valid if the following options of the product structure have been selected at the same time: 030 ("Display, Operation") = C
("SD02") or E ("SD03"); 040 ("Housing") = A ("GT19").
10) Can be downloaded from www.endress.com.

76 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Process
Process temperature range The maximum permitted temperature at the process connection is determined by the O-ring version
ordered:

Device O-ring material Process temperature Approval

FMP51 FKM (Viton GLT 37559) –30 to +150 °C (–22 to +302 °F)

–40 to +150 °C (–40 to +302 °F)


only in combination with feature 610 "Accessory
Mounted" option model NC "Gas-tight feed through"

EPDM (70C4 pW FKN or –40 to +120 °C (–40 to +248 °F)


E7515)

FFKM (Kalrez 6375) 1) –20 to +200 °C (–4 to +392 °F) 2)

FVMQ (FVMQ 70C79) –50 to 130 °C (–58 to 260 °F)

FMP52 — –50 to +200 °C (–58 to +392 °F); completely coated FDA, 3A,
EHEDG, USP
Cl. VI 3)

FMP54 Graphite Version XT: –196 to +280 °C (–321 to +536 °F) 4)

Version HT: –196 to +450 °C (–321 to +842 °F)

1) Recommended for steam applications


2) Not recommended for saturated steam above 150 °C (302 °F). Use FMP54 instead.
3) Plastic parts in contact with process have been tested according to USP <88> Class VI-70°C.
4) Not recommended for saturated steam above 200 °C (392 °F). Use the HT version instead.

For FMP52: High process temperatures (> 150 °C (302 °F)) may possibly accelerate diffusion of
the process medium through the probe coating, which may reduce the operating time.
The sensor material 1.4404/316L is resistant against attacks by internal corrosion according to
AD 2000 - instruction sheet W2 for operating temperatures up to 400 °C (752 °F) and
operating times of 100,000 hours (11.4 years). In case of higher temperatures the suitability of
the material must be checked by the operator. Acids, in particular, may cause corrosion attacks.
With uncoated probes, the medium temperature may be higher, under the condition that the
maximum process temperature specified in the table above is not exceeded at the process
connection.
However, when using rope probes, the stability of the probe rope is reduced by structural
changes at temperatures above 350 °C (662 °F).

Process pressure range Device Process pressure

FMP51 –1 to 40 bar (–14.5 to 580 psi)

FMP52 –1 to 40 bar (–14.5 to 580 psi)

FMP54 –1 to 400 bar (–14.5 to 5 800 psi)

Endress+Hauser 77
Levelflex FMP51, FMP52, FMP54

FMP54 - process pressure depending on the process temperature

[bar] ([psi]) p

420 (6090)

400 (5800)

380 (5510)

350 (5075)

320 (4640)

300 (4350) Tp
-196 -100 0 +100 200 300 400 450 [°C]
(-321) (-148) (+32) (+212) (+392) (+572) (+752) (+842) ([°F])

A0014005

p = process pressure Tp = process temperature


This range may be reduced by the selected process connection. The pressure rating (PN)
specified on the flanges refers to a reference temperature of 20 °C, for ASME flanges 100 °F.
Pay attention to pressure-temperature dependencies.
Please refer to the following standards for the pressure values permitted for higher
temperatures:
• EN 1092-1: 2007 Tab. G.4.1-x
With regard to their temperature stability properties, the materials 1.4435 and 1.4404 are
grouped under 13E0 in EN 1092-1:2007 Tab. G.3.1-1. The chemical composition of the two
materials can be identical.
• ASME B 16.5a - 2013 Tab. 2-2.2 F316
• ASME B 16.5a - 2013 Tab. 2.3.8 N10276
• JIS B 2220

Dielectric constant (DC) • Coax probes: DC (εr) ≥ 1.4


• Rod and rope probe: DC (εr) ≥ 1.6 (when installing in pipes DN ≤ 150 mm (6 in): DC (εr) ≥ 1.4)

Expansion of the rope probes Elongation through temperature increase from 30 °C (86 °F) to 150 °C (302 °F): 2 mm / m rope
through temperature length

78 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Mechanical construction
Dimensions Dimensions of the electronics housing

R110 98 (3.86)*
78 (3.07) 90 (3.54)

07)
(4.
3.5

146 (5.75)
ø108.5 (4.27)

142 (5.59)
115.25 (4.54)
ø10
A0011666

 25 Housing GT18 (316L); Dimensions in mm (in)


*for devices with integrated overvoltage protection.

R110
99 (3.9)*
78 (3.07) 90 (3.54)
)
.17
6 (4

144 (5.67)
ø106 (4.17)
ø10

113.5 (4.47)
A0011346

 26 Housing GT19 (Plastics PBT); Dimensions in mm (in)


*for devices with integrated overvoltage protection.

R110 98 (3.86)*
78 (3.07) 90 (3.54)
07)
(4.

ø112 (4.41)

145.5 (5.73)
3.5

141.5 (5.57)
ø108.5 (4.27)
ø10

61 (2.4)

A0020751

 27 Housing GT20 (Alu coated); Dimensions in mm (in)


*for devices with integrated overvoltage protection.

Endress+Hauser 79
Levelflex FMP51, FMP52, FMP54

Dimensions of the mounting bracket

A B

ø42...60
(1.65...2.36)

52 (2)
122 (4.8) 127...140
161 (6.34) (5...5.51)
162...175
(6.38...6.89)

(3.4)
(2.8)

86
70

A0014793

 28 Mounting bracket for the electronics housing: dimensions: mm (in)


A Wall mounting
B Pipe mounting

For the "Sensor remote" device version (see feature 060 of the product structure), the mounting
bracket is part of the delivery. If required, it can also be ordered as an accessory (order code
71102216).

Dimensions of the connection piece for the remote probe

122(4.8)
(2.05)
52

15 (0.6)

rmin = 100 (4)

ø10
(0. .2
4)
(2.4)
61
43 (1.7)

ø82.5
(3.25)
A0023856

 29 Connection piece for the remote probe; dimensions: mm (in); Length of the connecting cable: as ordered

80 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

FMP51: Dimensions of process connection (G¾,NPT¾) and probe

(0.98)
25
(2.72)
SW36

69
AF36
R G¾
NPT¾

E F G H

SW7 SW7 SW7


AF7 AF7 AF7

ø4 (0.16) ø4 (0.16)
ø8 (0.31) ø22 (0.87)
LN

150 (5.91)
20 (0.79)

12 (0.47)
4 (0.16)

2 (0.08)

M14 ø29.5 (1.16) 8 (0.31)


ø22 (0.87) ø75 (2.95)

A0012645

 30 FMP51: Process connection / probe; dimensions: mm (in). Unit of measurement mm (in)


B Thread ISO G3/4 or ANSI MNPT3/4 (Feature (100)
E Rope probe 4mm or 1/6" (Feature 060)
F Rope probe 4 mm or 1/6"; centering disk optional (Features 060 and 610)
G Rod probe 8mm or 1/3" (Feature 060)
H Coax probe (Feature 060); with venting hole Ø approx. 6 mm (0.24 in)
LN Length of probe
R Reference point of the measurement

Endress+Hauser 81
Levelflex FMP51, FMP52, FMP54

FMP51: Dimensions of process connection (G1½,NPT1½,flange) and probe

B C D
78 (3.1)

78 (3.1)

78 (3.1)
(0.98)

(1.02)
25

26
SW55 SW55
AF55 AF55

R
G1½ NPT1½

I J K L M N

(1.54)
39
SW10
AF10 SW14
SW10 AF14
AF10 SW10 ø16 (0.63)
ø4 (0.16) AF10

1000 (39.4)
LN

ø4 (0.16) 500 (19.7)


SW14
AF14

ø12 (0.47)

40 (1.57)
SW14
AF14
2.2 (0.09)
min. 50 (1.97)
max. 551 (21.7)
max. 1051 (41.4)
150 (5.91)
20 (0.79)

12 (0.47)
4 (0.16)

P
9 (0.35)

Q Q
S

M14 ø29.5 (1.16) T 2.77 (0.11) 2 (0.08)


ø22 (0.87) ø75 (2.95) ø12.7 (0.5) ø12 (0.47)
ø42.2 (1.66) ø42.4
(1.67)
A0012756

 31 FMP51: Process connection / probe; dimensions: mm (in). Unit of measurement mm (in)


B Thread ISO228 G1-1/2 (Feature 100)
C Thread ANSI MNPT1-1/2 (Feature 100)
D Flange ANSI B16.5, EN1092-1, JIS B2220 (Feature 100)
I Rope probe 4mm or 1/6" (Feature 060)
J Rope probe 4mm or 1/6"; centering disk optional (Features 060 and 610)
K Rod probe 12mm or 1/2"; centering disk optional, see table below (Features 060 and 610)
L Rod probe 16 mm or 0.63in, 20" or 40" divisible; centering disk optional, see table below (Feature 060 and
610)
M Coax probe; AlloyC (Feature 060); with venting hole Ø approx. 8 mm (0.3 in)
N Coax probe; 316L (Feature 060); with venting holes Ø approx. 10 mm (0.4 in)
LN Length of probe
P Thickness of the centering star; table of values: see below
Q Diameter of the centering star; table of values: see below
R Reference point of the measurement
S Thickness of the center washer or centering star; table of values: see below
T Diameter of the center washer or centering star; table of values: see below

82 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Center washer or centering star


Ordering feature 610 Meaning Thickness Diameter
"Accessory mounted"

OA Rod center washer 316L; pipe diameter DN80/3" + DN100/4" S = 4 mm (0.16 in) T = 75 mm (2.95 in)

OB Rod center washer 316L; pipe diameter DN50/2" + DN65/2-1/2" S = 4 mm (0.16 in) T = 45 mm (1.77 in)

OC Rod center washer 316L; pipe diameter DN80/3" + DN100/4" S = 4 mm (0.16 in) T = 75 mm (2.95 in)

OD Rod centering star PEEK; interface measurement; pipe diameter S = 7 mm (0.28 in) T = 48 to 95 mm (1.9 to 3.7 in)
DN50/2" + DN100/4"

OE Rod centering star PFA; interface measurement; pipe diameter P = 10 mm (0.39 in) Q = 37 mm (1.46 in)
DN40/1-1/2" + DN50/2"

Note on AlloyC flanges


AlloyC flanges always have an additional thread, even if they are not used with a coax probe.
Affected options of ordering feature 100 "Process connection": AEM, AFM, AGM, AQM, ARM, ASM,
ATM, CEM, CFM, CGM, CQM, CRM, CSM, CTM.

M40
8 (0.31)

A0035223

 32 Dimensions of AlloyC flanges; Dimensions: mm (in)

Endress+Hauser 83
Levelflex FMP51, FMP52, FMP54

FMP52: Dimensions of process connection and probe

A B C D E
64 91
2.52 3.58

(3.19)
(3.23)

(3.31)
(3.23)

(3.23)

81
82

84
82

82 R

ø16 ø16 (0.63) ø16 (0.63) ø16 (0.63) ø16 (0.63)


(0.63) ø68.5 (2.7)
ø50.5 (1.99)

F G

ø16 (0.63)
ø16 (0.63)
LN

ø4 (0.16)

ø37 (1.46)
118 (4.65)

ø22 (0.87)

A0012757

 33 FMP52: Process connection / probe; dimensions: mm (in). Unit of measurement mm (in)


A Tri-Clamp 1-1/2" (Feature 100)
B Tri-Clamp 2" (Feature 100)
C Tri-Clamp 3" (Feature 100)
D DIN 11851 (Dairy coupling) DN50 (Feature 100)
E Flange ANSI B16.5, EN1092-1, JIS B2220 (Feature 100)
F Rod probe 16mm or 0.63 in , PFA>316L (Feature 060); optionally with centering star (Feature 610)
G Rope probe 4mm or1/6", PFA>316 (Feature 060)
LN Length of probe
R Reference point of the measurement

84 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

FMP54: Dimensions of process connection and probe

A B C D
132 (5.2)

132 (5.2)

342 (13.5)
315 (12.4)
SW60 SW60
267 (10.5)
SW60 AF60 SW60 AF60
240 (9.45)

AF60 AF60
159 (6.26)

159 (6.26)
ø90 ø90
ø90 ø90 (3.54) (3.54)
(3.54) (3.54)

R
SW14
37 (1.46)

SW14 SW14

37 (1.46)
SW14
8 (0.31)

8 (0.31)

8 (0.31)

8 (0.31)
AF14
AF14 AF14 AF14

M40x1 M40x1 M40x1 M40x1

E F G H I

ø16 (0.63)
SW14 SW14
AF14 AF14
ø16 (0.63) 1000 (39.4)
500 (19.7)

SW14
AF14
LN

ø4 (0.16) ø4 (0.16)
SW14
SW14 AF14
AF14

SW14
2.2 (0.09)
max. 1051 (41.4)
min. 50 (1.97)

M AF14
150 (5.91)
20 (0.79)

12 (0.47)

L
9 (0.35)
4 (0.16)

2 (0.08)
M
N

M14 ø29.5 (1.16) O ø16 (0.63)


ø22 (0.87) ø75 (2.95) ø42.4 (1.67)

A0012778

 34 FMP54: Process connection / probe; dimensions: mm (in). Unit of measurement mm (in)


A Thread ISO228 G1-1/2 or ANSI MNPT1-1/2; XT 280 °C (Features 100 and 090)
B Flange ANSI B16.5, EN1092-1, JIS B2220; XT 280 °C (Features 100 and 090)
C Thread ISO228 G1-1/2 or ANSI MNPT1-1/2; HT 450 °C (Features 100 and 090)
D Flange ANSI B16.5, EN1092-1, JIS B2220; HT 450°C (Features 100 and 090)
E Rope probe 4mm or 1/6" (Feature 060)
F Rope probe 4mm (1/6"), centering disk optional (Features 060 and 610)
G Rod probe 16mm or 0.63in; centering disk optional, see table below (Features 060 and 610)
H Rod probe 16mm or 0.63in ; 20" or 40" divisible; centering disk optional, see table below (Feature 060 and
610)
I Coax probe (Feature 060); with venting holes Ø approx. 10 mm (0.4 in); with centering disc for devices with
the "Gas Phase Compensation" application package (ordering feature 540, options EF or EG)
LN Length of probe
L Thickness of the centering star; table of values: see below
M Diameter of the centering star; table of values: see below

Endress+Hauser 85
Levelflex FMP51, FMP52, FMP54

N Thickness of the center washer or centering star; table of values: see below
O Diameter of the center washer or centering star; table of values: see below
R Reference point of the measurement

Center washer or centering star


Ordering feature 610 Meaning Thickness Diameter
"Accessory mounted"

OA Rod center washer 316L; pipe diameter DN80/3" + DN100/4" N = 4 mm (0.16 in) O = 75 mm (2.95 in)

OB Rod center washer 316L; pipe diameter DN50/2" + DN65/2-1/2" N = 4 mm (0.16 in) O = 45 mm (1.77 in)

OC Rod center washer 316L; pipe diameter DN80/3" + DN100/4" N = 4 mm (0.16 in) O = 75 mm (2.95 in)

OD Rod centering star PEEK; interface measurement; pipe diameter N = 7 mm (0.28 in) O = 48 to 95 mm (1.9 to 3.7 in)
DN50/2" + DN100/4"

OE Rod centering star PFA; interface measurement; pipe diameter L = 10 mm (0.39 in) M = 37 mm (1.46 in)
DN40/1-1/2" + DN50/2"

86 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Tolerance of probe length Rod and coax probes

Over [m (ft)] — 1 (3,3) 3 (9,8) 6 (20)

Up to [m (ft)] 1 (3,3) 3 (9,8) 6 (20) —

Admissible tolerance [mm -5 (-0,2) -10 (-0,39) -20 (-0,79) -30 (-1,18)
(in)]

Rope probes

Over [m (ft)] — 1 (3,3) 3 (9,8) 6 (20)

Up to [m (ft)] 1 (3,3) 3 (9,8) 6 (20) —

Admissible tolerance [mm -10 (-0,39) -20 (-0,79) -30 (-1,18) -40 (-1,57)
(in)]

Surface roughness of AlloyC- Ra = 3.2 μm; lower surface roughness levels are available on request.
coated flanges
This value is valid for flanges with "AlloyC>316/316L"; see product structure, feature 100 "Process
connection". For other flanges the surface roughness matches the corresponding flange standard.

Shortening probes If necessary, probes can be shortened. When doing so, the following must be observed:

Shortening rod probes


Rod probes must be shortened if the distance to the container floor or outlet cone is less than
10 mm (0.4 in). The rods of a rod probe are shortened by sawing at the bottom end.
Rod probes of FMP52 can not be shortened as they are coated.

Shortening rope probes


Rope probes must be shortened if the distance to the container floor or outlet cone is less than
150 mm (6 in).
Rope probes of FMP52 can not be shortened as they are coated.

Shortening coax probes


Coax probes must be shortened if the distance to the container floor or outlet cone is less than
10 mm (0.4 in).
Coax probes can be shortened max. 80 mm (3.2 in) from the end. They have centering units
inside, which fix the rod centrally in the pipe. The centerings are held with borders on the rod.
Shortening is possible up to approx. 10 mm (0.4 in) below the centering unit.

Endress+Hauser 87
Levelflex FMP51, FMP52, FMP54

Weight Housing
Part Weight

Housing GT18 - stainless steel approx. 4.5 kg

Housing GT19 - plastic approx. 1.2 kg

Housing GT20 - aluminium approx. 1.9 kg

FMP51 with threaded connection G¾ or NPT¾


Part Weight Part Weight

Sensor approx. 0.8 kg Rod probe 8 mm approx. 0.4 kg/m probe length

Rope probe 4 mm approx. 0.,1 kg/m probe length Coax probe approx. 1.2 kg/m probe length

FMP51 with threaded connection G1½/NPT1½ or flange


Part Weight Part Weight

Sensor approx. 1.2 kg + weight of flange Rod probe 16 mm approx. 1.1 kg/m probe length

Rope probe 4 mm approx. 0.1 kg/m probe length Coax probe approx. 3.0 kg/m probe length

Rod probe 12 mm approx. 0.9 kg/m probe length

FMP52
Part Weight Part Weight

Sensor approx. 1.2 kg + weight of flange Rope probe 4 mm approx. 0.5 kg/m probe length

Rod probe 16 mm approx. 1.1 kg/m probe length

FMP54
Part Weight Part Weight

Sensor - version approx. 6.7 kg + weight of flange Rope probe 4 mm approx. 0.1 kg/m probe length
XT

Sensor - version approx. 7.7 kg + weight of flange Rod probe 16 mm approx. 1.6 kg/m probe length
HT

Coax probe approx. 3.5 kg/m probe length

88 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Materials: GT18 housing


1
(stainless steel, corrosion-
resistant) 2.1
2.2

5.2
6
5.1
7
4 3 9 8
A0013788

No. Part Material

1 Housing CF3M similar to 316L/1.4404

2.1 Cover of the electronics compartment • Cover: CF3M (similar to 316L/1.4404)


• Window: glass
• Cover seal: NBR
• Seal of the window: NBR
• Thread-coating: Graphite-based lubricant varnish

2.2 Cover of the terminal compartment • Cover: CF3M (similar to 316L/1.4404)


• Cover seal: NBR
• Thread-coating: Graphite-based lubricant varnish

3 Cover lock • Screw: A4


• Clamp: 316L (1.4404)

4 Lock at the housing neck • Screw: A4-70


• Clamp: 316L (1.4404)

5.1 Dummy plug, cable gland, adapter or plug • Dummy plug, depending on the device version:
(depending on the device version) – PE
– PBT-GF
• Cable gland: 316L (1.4404) or nickel-plated brass
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)

5.2 Dummy plug, cable gland or adapter (depending on • Dummy plug: 316L (1.4404)
the device version) • Cable gland: 316L (1.4404) or nickel-plated brass
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM

6 Dummy plug or M12 socket (depending on the • Dummy plug: 316L (1.4404)
device version) • M12 socket: 316L (1.4404)

7 Pressure relief stopper 316L (1.4404)

8 Ground terminal • Screw: A4


• Spring washer: A4
• Clamp: 316L (1.4404)
• Holder: 316L (1.4404)

9 Nameplate • Plate: 316L (1.4404)


• Groove pin: A4 (1.4571)

1) For the version with M12 plug the sealing material is Viton.
2) For the version with 7/8" plug, the sealing material is NBR.

Endress+Hauser 89
Levelflex FMP51, FMP52, FMP54

Materials: GT19 housing


1
(plastic)
2.1
2.2

5.2
6
5.1
7
4 9 8
A0013788

No. Part Material

1 Housing PBT

2.1 Cover of the electronics compartment • Cover glass: PC


• Cover frame: PBT-PC
• Cover seal: EPDM
• Thread-coating: Graphite-based lubricant varnish

2.2 Cover of the terminal compartment • Cover: PBT


• Cover seal: EPDM
• Thread-coating: Graphite-based lubricant varnish

4 Lock at the housing neck • Screw: A4-70


• Clamp: 316L (1.4404)

5.1 Dummy plug, cable gland, adapter or plug • Dummy plug, depending on the device version:
(depending on the device version) – PE
– PBT-GF
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)

5.2 Dummy plug, cable gland or adapter (depending on • Dummy plug, depending on the device version:
the device version) – PE
– PBT-GF
– Nickel-plated steel
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM

6 Dummy plug or M12 socket (depending on the • Dummy plug: Nickel-plated brass (CuZn)
device version) • M12 socket: Nickel-plated GD-Zn

7 Pressure relief stopper Nickel-plated brass (CuZn)

8 Ground terminal • Screw: A2


• Spring washer: A4
• Clamp: 304 (1.4301)
• Holder: 304 (1.4301)

9 Adhesive nameplate Plastic

1) For the version with M12 plug the sealing material is Viton.
2) For the version with 7/8" plug, the sealing material is NBR.

90 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Materials: GT20 housing


1
(die-cast aluminum, powder-
coated) 2.1
2.2

5.2
6
5.1
7
4 3 9 8
A0013788

Nr. Part Material

1 Housing, RAL 5012 (blue) • Housing: AlSi10Mg(<0,1% Cu)


• Coating: Polyester

2.1 Cover of the electronics compartment; RAL 7035 • Cover: AlSi10Mg(<0,1% Cu)
(gray) • Window: Glass
• Cover seal: NBR
• Seal of the window: NBR
• Thread-coating: Graphite-based lubricant varnish

2.2 Cover of the terminal compartment; RAL 7035 • Cover: AlSi10Mg(<0,1% Cu)
(gray) • Cover seal: NBR
• Thread-coating: Graphite-based lubricant varnish

3 Cover lock • Screw: A4


• Clamp: 316L (1.4404)

4 Lock at the housing neck • Screw: A4-70


• Clamp: 316L (1.4404)

5.1 Dummy plug, cable gland, adapter or plug • Dummy plug, depending on the device version:
(depending on the device version) – PE
– PBT-GF
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)

5.2 Dummy plug, cable gland or adapter (depending on • Dummy plug, depending on the device version:
the device version) – PE
– PBT-GF
– Nickel-plated steel
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM

6 Dummy plug or M12 socket (depending on the • Dummy plug : Nickel-plated brass (CuZn)
device version) • M12 socket: Nickel-plated GD-Zn

7 Pressure relief stopper Nickel-plated brass (CuZn)

Endress+Hauser 91
Levelflex FMP51, FMP52, FMP54

Nr. Part Material

8 Ground terminal • Screw: A2


• Spring washer: A2
• Clamp: 304 (1.4301)
• Holder: 304 (1.4301)

9 Adhesive nameplate Plastic

1) For the version with M12 plug the sealing material is Viton.
2) For the version with 7/8" plug, the sealing material is NBR.

92 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Materials: Process Endress+Hauser supplies DIN/EN flanges and threaded process connections made of stainless
connection steel according to AISI 316L (DIN/EN material number 1.4404 or 1.4435). With regard to their
temperature stability properties, the materials 1.4404 and 1.4435 are grouped under 13E0 in
EN 1092-1:2007 Tab. G.3.1-1. The chemical composition of the two materials can be identical.

Levelflex FMP51

Threaded connection Flange


No. Material
G¾, NPT¾ G1½ NPT1½ DN40...DN200 DN40...DN100

1.1 316L (1.4404)


1.1 1.2 1.2 Alloy C22 (2.4602)

1.1 1.1 1.1


2 2 2 ASME: 316/316L
EN: 316L (1.4404)
JIS: 316L (1.4435)
3 3 3 3 3 4
A0013850 A0013852 A0013849 A0013854 A0013910
3 Ceramic Al2O3 99.7 %

4 Cladding: Alloy C22 (2.4602)

Levelflex FMP52

Flange Dairy coupling Tri-Clamp


No. Material Approval
EN/ASME/JIS DN50 (DIN 11851) 3" 2" 1½"

1 316L (1.4404)

1 2 ASME: 316/316L
1 1 EN: 316L (1.4404)
3 1
2
3
JIS: 316L (1.4435)
5 3 3 3 316L (1.4404)
4 4 4 4 4 4 Coating 2 mm (0.08 in): PTFE (Dyneon USP Cl. VI 1)
A0013868
A0013865
A0013866 A0013867
A0013869
TFM1600)

5 304L (1.4307)

1) Plastic parts in contact with process have been tested according to USP <88> Class VI-70°C.

Levelflex FMP54

Threaded connection G1½, NPT1½ Flange


No. Material
HT version XT version HT version XT version

1 316L (1.4404)

1 1 2 316L (1.4404)
1 1
3 ASME: 316/316L
EN: 316L (1.4404)
2
7 2 7 2 JIS: 316L (1.4435)
2
7 7 3
4 Alloy C22 (2.4602)
3 4
4
5
4 5 5 316L (1.4404)
4 5 6
5 6 6 Nordlock washer: 1.4547
6 A0013883
6 A0013882

A0013880 A0013881
7 Ceramic Al2O3 99.7 %, pure graphite

Endress+Hauser 93
Levelflex FMP51, FMP52, FMP54

Materials: Probe Levelflex FMP51

Rope probe Rod probe Coax probe

 4 mm
(1/6")  12.7 Thread Thread No. Material
 4 mm  8 mm Thread
with mm (1/2") G1-1/2 G1-1/2
(1/6") (1/3") G3/4
centering AlloyC AlloyC 316L
disk

1.1 316L (1.4404)


1.1
1 1.2 Alloy C22 (2.4602)
3
1.2 2 316 (1.4401)
1.1 1.1
1.1 1.2 3 316L (1.4404)
2
4 Set screw: A4-70
2 5 Screw for tightening: A2-70

6 Socket cap screw: A4-80


4
3 7 Disk: 316L (1.4404)
9.1 9.2 9.1
1.1 4 8 Set screw: A4-70
10 10 10
5 9.1 Rod: 316L (1.4404)
5 6
9.2 Alloy C22 (2.4602)
A0031287 A0013855 A0013838 A0013912 A0013856 A0013911
A0017244
10 Centering star: PFA

Levelflex FMP51

Rod probe

 12 mm (1/2")  16 mm (2/3") No. Material


316L divisible

1 316L (1.4404)

1 3 Connecting bolt: Alloy C22 (2.4602)


1
3
Nordlock washer: 1.4547

11 Hexagon head screw: A4-70

Nordlock washer: 1.4547


1
12 Centering star, PEEK

3 Center washer, 316L (1.4404)


13 13 13 Center washer, PFA

1
12

11
A0013860 A0013861

94 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Levelflex FMP52

Rod probe Rope probe


No. Material
 16 mm (2/3") coated  4 mm (1/6") coated

1 316L (1.4404)
1 1
2 Coating 2 mm (0.08 in): PFA (Daikin PFA AP230)

2 3 Rope: 316 (1.4401)

Coating 0.75mm (0.03 in): PFA (Daikin PFA AP230)

2 4 Core: 316L (1.4435)


2
5 PFA (Daikin PFA AP230), centering star

5 2

2
A0013870 A0013871

Levelflex FMP54

Rope probe Rod probe

 4 mm (1/6")  16 mm (2/3") Coax probe No. Material


 16 mm (2/3")
 4 mm (1/6") with centering with centering
divisible
disk disk

1.1 316L (1.4404) 1)


1 2 316 (1.4401)
1.1
3 Connecting bolt: Alloy C22 (2.4602)
1.1 9
1.1 Nordlock washer: 1.4547
2
4 Set screw: A4-70
3
2 13 5 Screw for tightening: A2-70
12 9
13 6 Socket cap screw: A4-80
4 10
3 11 7 Disk: 316L (1.4404)

1.1 4 A0013885 9 8 Set screw: A4-70

5 9 Rod: 316L (1.4404)


5 6 A0013886
10 Centering star: ceramic Al2O3 99.7 %
A0031287 A0013855 A0013864

11 Hexagon head screw: A4-70

Nordlock washer: 1.4547

12 Centering star, PEEK

Center washer, 316L (1.4404)

13 Centering star, PFA

1) For probes with gas phase compensation this is also the material of the reference rod.

Endress+Hauser 95
Levelflex FMP51, FMP52, FMP54

Materials: Mounting bracket

12

12

11
10

A0015143

Mounting bracket for version "Sensor remote"

Position Part Material

10 Bracket 316L (1.4404)

11 Bracket 316Ti (1.4571)

Screw/nuts A4-70

Distance sleeves 316Ti (1.4571) or 316L (1.4404)

12 Half-shells 316L (1.4404)

96 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Materials: Adapter and cable


for remote sensor

4 3

A0021722

Adapter and cable for version "Sensor remote"

Position Part Material

1 Cable FRNC

2 Sensor adapter 304 (1.4301)

3 Clamp 316L (1.4404)

Screw A4-70

4 Loop 316 (1.4401)

Crimp sleeve Aluminum

Nameplate 304 (1.4301)

Endress+Hauser 97
Levelflex FMP51, FMP52, FMP54

Materials: Weather
protection cover 1

2
3

7 4
5
6
A0015473

No Part: Material

1 Protection cover: 316L (1.4404)

2 Molded rubber part (4x): EPDM

3 Clamping screw: 316L (1.4404) + carbon fibre

4 Bracket: 316L (1.4404)

5 • Cheese head screw: A4-70


• Nut: A4
• Spring washer: A4

6 Ground terminal
• Screw: A4
• Spring washer: A4
• Clamp: 316L (1.4404)
• Holder: 316L (1.4404)

7 • Washer: A4
• Cheese head screw: A4-70

98 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Operability
Operating concept Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Operating languages
• English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык (Russian)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• Bahasa Indonesia
• tiếng Việt (Vietnamese)
• čeština (Czech)
Feature 500 of the product structure determines which of these languages is preset on delivery.

Quick and save commissioning


• Interactive wizard with graphical interface for easy commissioning via FieldCare/DeviceCare
• Menu guidance with brief explanations of the individual parameter functions
• Standardized operation at the device and in the operating tools
Integrated data storage device (HistoROM)
• Enables transfer of configuration when changing electronic modules
• Records up to 100 event messages in the device
• Records up to 1000 measured values in the device
• Saves the signal curve on commissioning which can later be used as a reference.
Efficient diagnostics increase measurement reliability
• Remedy information is integrated in plain text
• Diverse simulation options and line recorder functions

Endress+Hauser 99
Levelflex FMP51, FMP52, FMP54

Local operation Operation Pushbuttons Touch Control


with

Order code for Option C "SD02" Option E "SD03"


"Display;
Operation"

A0032219 A0032221

Display 4-line display 4-line display


elements white background lighting; switches to red in
event of device error

Format for displaying measured variables and status variables can be individually configured

Permitted ambient temperature for the display: –20 to +70 °C (–4 to +158 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.

Operating local operation with 3 push buttons (, , external operation via touch control; 3 optical
elements ) keys: , , 

Operating elements also accessible in various hazardous areas

Additional Data backup function


functionality The device configuration can be saved in the display module.

Data comparison function


The device configuration saved in the display module can be compared to the current device
configuration.

Data transfer function


The transmitter configuration can be transmitted to another device using the display module.

Operation with remote


1 2
display and operating
module FHX50

A0032215

 35 FHX50 operating options


1 Housing of the remote display and operating module FHX50
2 Display and operating module SD02, push buttons; cover must be removed
3 Display and operating module SD03, optical keys; can be operated through the glass of the cover

100 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Remote operation Via HART protocol

2 3 6 8

1 4 5 7
A0028746

 36 Options for remote operation via HART protocol


1 PLC (programmable logic controller)
2 Transmitter power supply unit, e.g. RN221N (with communication resistor)
3 Connection for Commubox FXA191, FXA195 and Field Communicator 375, 475
4 Field Communicator 475
5 Computer with operating tool (e.g. DeviceCare/FieldCare , AMS Device Manager, SIMATIC PDM)
6 Commubox FXA191 (RS232) or FXA195 (USB)
7 Field Xpert SFX350/SFX370
8 VIATOR Bluetooth modem with connecting cable
9 Transmitter

Via PROFIBUS PA protocol

2 3

PROFIBUS DP
ENDRESS + HAUSER

PROFIBUS PA
T

4 4 4
A0015775

1 Segment coupler
2 Computer with Profiboard/Proficard and operating tool (e.g. DeviceCare/FieldCare
3 PLC (Progrommable Logic Controller)
4 Transmitter
5 Additional functions (valves etc.)

Endress+Hauser 101
Levelflex FMP51, FMP52, FMP54

Via FOUNDATION Fieldbus

IN

BT
FF-HSE
1
FF-H1 SB 2
PS BT
LD 4

3
BT ENDRESS + HAUSER

– +
%

PS

FF-H1
BT

1
2

A0017188

 37 FOUNDATION Fieldbus system architecture with associated components


1 FFblue Bluetooth modem
2 Field Xpert SFX350/SFX370
3 DeviceCare/FieldCare
4 NI-FF interface card

IN Industrial network

FF-HSE High Speed Ethernet

FF-H1 FOUNDATION Fieldbus-H1

LD Linking Device FF-HSE/FF-H1

PS Bus Power Supply

SB Safety Barrier

BT Bus Terminator

102 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Via service interface (CDI)

1
A0032466

1 Service interface (CDI) of the measuring device (= Endress+Hauser Common Data Interface)
2 Commubox FXA291
3 Computer with DeviceCare/FieldCare operating tool

Endress+Hauser 103
Levelflex FMP51, FMP52, FMP54

Integration in tank gauging The Endress+Hauser Tank Side Monitor NRF590 provides integrated communications for sites with
system multiple tanks, each with one or more sensors on the tank, such as radar, spot or average
temperature, capacitive probe for water detection and/or pressure sensors. Multiple protocols out of
the Tank Side Monitor guarantee connectivity to nearly any of the existing industry standard tank
gauging protocols. Optional connectivity of analog 4...20 mA sensors, digital I/O and analog output
simplify full tank sensor integration. Use of the proven concept of the intrinsically safe HART bus for
all on-tank sensors yields extremely low wiring costs, while at the same time providing maximum
safety, reliability and data availability.

1 2 3

4 5

8 NXA820

16...253 VAC

A0016590

 38 The complete measuring system consists of:


1 Tankvision workstation
2 Commubox FXA195 (USB) - optional
3 Computer with operating tool (ControlCare) - optional
4 Level measuring device
5 Temperature measuring device
6 Tank Side Monitor NRF590
7 Pressure measuring device
8 Tankvision Tank Scanner NXA820

104 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

SupplyCare inventory SupplyCare is a web-based operating program for coordinating the flow of material and information
management software along the supply chain. SupplyCare provides a comprehensive overview of the levels of
geographically distributed tanks and silos, for instance, providing complete transparency over the
current inventory situation, regardless of time and location.
Based on the measuring and transmission technology installed onsite, the current inventory data are
collected and sent to SupplyCare. Critical levels are clearly indicated and calculated forecasts provide
additional security for material requirements planning.
The main functions of SupplyCare:
Inventory visualization
SupplyCare determines the inventory levels in tanks and silos at regular intervals. It displays current
and historical inventory data and calculated forecasts of future demand. The overview page can be
configured to suit the user's preferences.
Master data management
With SupplyCare you can create and manage the master data for locations, companies, tanks,
products and users, as well as user authorization.
Report Configurator
The Report Configurator can be used to create personalized reports quickly and easily. The reports
can be saved in a variety of formats, such as Excel, PDF, CSV and XML. The reports can be
transmitted in many ways, such as by http, ftp or e-mail.
Event management
Events, such as when levels drop below the safety stock level or plan points, are indicated by the
software. In addition, SupplyCare can also notify pre-defined users by e-mail.
Alarms
If technical problems occur, e.g. connection issues, alarms are triggered and alarm e-mails are sent to
the System Administrator and the Local System Administrator.
Delivery planning
The integrated delivery planning function automatically generates an order proposal if a pre-set
minimum inventory level is undershot. Scheduled deliveries and disposals are monitored
continuously by SupplyCare. SupplyCare notifies the user if scheduled deliveries and disposals are not
going to be met as planned.
Analysis
In the Analysis module, the most important indicators for the inflow and outflow of the individual
tanks are calculated and displayed as data and charts. Key indicators of material management are
automatically calculated and form the basis for optimizing the delivery and storage process.
Geographical visualization
All the tanks and the tank inventories are represented graphically on a map (based on Google Maps).
The tanks and inventory situations can be filtered by tank group, product, supplier or location.
Multi-language support
The multi-language user interface supports 9 languages, thereby enabling global collaboration on a
single platform. The language and settings are recognized automatically using the browser settings.

SupplyCare Enterprise
SupplyCare Enterprise runs by default as a service under Microsoft Windows on an application server
in an Apache Tomcat environment. The operators and administrators operate the application via a
Web browser from their workstations.

Endress+Hauser 105
Levelflex FMP51, FMP52, FMP54

2
3

10
5 7 8
9

.
A0034288

 39 Example of inventory management platform with SupplyCare Enterprise SCE30B


1 SupplyCare Enterprise (via Web browser)
2 SupplyCare Enterprise installation
3 SupplyCare Enterprise on mobile devices (via Web browser)
4 Ethernet/WLAN/UMTS
5 Fieldgate FXA42
6 Power supply 24 V DC
7 Modbus TCP via Ethernet as server/client
8 Converter from Modbus to HART Multidrop
9 HART Multidrop
10 4 x 4 to 20 mA analog input (2-wire/4-wire)

106 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Cloud-based application: SupplyCare Hosting


SupplyCare Hosting is offered as a hosting service (software as a service). Here, the software is
installed within the Endress+Hauser IT infrastructure and made available to the user in the Endress
+Hauser portal.

1
3
WWW

4
10
5 7 8
9

.
A0034289

 40 Example of inventory management platform with SupplyCare Hosting SCH30


1 SupplyCare Hosting installation in Endress+Hauser data center
2 PC workstation with Internet connection
3 Warehouse locations with Internet connection via 2G/3G with FXA42 or FXA30
4 Warehouse locations with Internet connection with FXA42
5 Fieldgate FXA42
6 Power supply 24 V DC
7 Modbus TCP via Ethernet as server/client
8 Converter from Modbus to HART Multidrop
9 HART Multidrop
10 4 x 4 to 20 mA analog input (2-wire/4-wire)

With SupplyCare Hosting, users do not need to make the initial software purchase or install and run
the IT infrastructure needed. Endress+Hauser constantly update SupplyCare Hosting and enhance
the capability of the software in conjunction with the customer. The hosted version of SupplyCare is
thus always up-to-date and can be customized to meet different customer requirements. Other
services are also offered in addition to the IT infrastructure and the software that is installed in a
secure, redundant Endress+Hauser data center. These services include defined availability of the
global Endress+Hauser Service and Support Organization and defined response times in a service
event.

Endress+Hauser 107
Levelflex FMP51, FMP52, FMP54

Certificates and approvals


Currently available certificates and approvals can be called up via the product configurator.

CE mark The measuring system meets the legal requirements of the applicable EC guidelines. These are listed
in the corresponding EC Declaration of Conformity together with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.

RoHS The measuring system complies with the substance restrictions of the Restriction on Hazardous
Substances Directive 2011/65/EU (RoHS 2).

RCM-Tick marking The supplied product or measuring system meets the ACMA (Australian Communications and Media
Authority) requirements for network integrity, interoperability, performance characteristics as well
as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic
compatibility are met. The products are labelled with the RCM- Tick marking on the name plate.

A0029561

Ex approval The devices are certified for use in hazardous areas and the relevant safety instructions are provided
in the separate "Safety Instructions" (XA) document. Reference is made to this document on the
nameplate.
The separate documentation "Safety Instructions" (XA) containing all the relevant explosion
protection data is available from your Endress+Hauser Sales Center. Correlation of
documentations to the device →  134.

Dual seal according to The devices have been designed according to ANSI/ISA 12.27.01 as dual seal devices, allowing the
ANSI/ISA 12.27.01 user to waive the use and save the cost of installing external secondary process seals in the conduit
as required by the process sealing sections of ANSI/NFPA 70 (NEC) and CSA 22.1 (CEC) These
instruments comply with the North-American installation practice and provide a very safe and cost-
saving installation for pressurized applications with hazardous fluids.
Further information can be found in the Safety Instructions (XA) of the relevant devices.

Functional Safety Used for level monitoring (MIN, MAX, range) up to SIL 3 (homogeneous redundancy), independently
assessed by TÜV Rhineland as per IEC 61508. Other information see documentation SD00326F:
"Functional Safety Manual".

Overfill prevention WHG


DIBt Z-65.16-501

108 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Sanitary compatibility The following table shows which device versions meet the requirements of 3A-Sanitary Standard No.
74 and are certified by the EHEDG.

FMP52

Feature Option 3A EHEDG

060: Probe CA  
Rod 16mm PFA>316L

CB  
Rod 0.63 in PFA>316L

100: Process connection TAK  


Tri-Clamp ISO2852 DN40-51 (2"), PTFE>316L

TLK  
Tri-Clamp ISO2852 DN70-76.1 (3"), PTFE>316L

TNK  
Tri-Clamp ISO2852 DN38 (1-1/2"), PTFE>316L

Suitable fittings and seals must be used to ensure hygiene-compliant design according to the
specifications of 3A and EHEDG.
The gap-free connections can be cleaned without residue using the usual cleaning methods (CIP
and SIP).
Wetted non-metallic parts of the FMP52 conform to FDA 21 CFR 177.1550 and USP Class VI.

AD2000 • For FMP51/FMP54:


The wetted material 316L (1.4435/1.4404) corresponds to AD2000 - W2/W10.
• For FMP52/FMP55:
The pressure retaining material 316L (1.4435/1.4404) corresponds to AD2000 - W2/W10.
• Declaration of conformity: see product structure, feature 580, option JF.

NACE MR 0175 / ISO 15156 For FMP51, FMP54:


• The wetted, metallic materials (except for ropes) comply with the requirements of NACE MR
0175 / ISO 15156.
• Declaration of conformity: see product structure, feature 580, option JB.

NACE MR 0103 For FMP51, FMP54:


• The wetted, metallic materials (except for ropes) comply with the requirements of NACE MR
0103 / ISO 17495.
• The certificate of compliance is based on NACE MR 0175.
The hardness and the intercrystalline corrosion have been tested, as well as the heat treatment
(solution annealed) has been carried out. The used materials comply therefore with the
requirements of NACE MR 0103 / ISO 17495.
• Declaration of conformity: see product structure, feature 580, option JE.

ASME B31.1 and B31.3 • The dimensions, materials of construction, pressure / temperature ratings and identification
markings of the device comply with the requirements of AMSE B31.1 and ASME B31.3
• Declaration of conformity: see product structure, feature 580, option KV.

Pressure equipment with Pressure instruments with a flange and threaded boss that do not have a pressurized housing do not
allowable pressure fall within the scope of the Pressure Equipment Directive, irrespective of the maximum allowable
≤ 200 bar (2 900 psi) pressure.

Reasons:
According to Article 2, point 5 of EU Directive 2014/68/EU, pressure accessories are defined as
"devices with an operational function and having pressure-bearing housings".
If a pressure instrument does not have a pressure-bearing housing (no identifiable pressure chamber
of its own), there is no pressure accessory present within the meaning of the Directive.

Endress+Hauser 109
Levelflex FMP51, FMP52, FMP54

Pressure equipment with Pressure equipment designated for application in every process fluid having a pressurized volume V
allowable pressure > of <0.1 l and a max. allowable pressure PS > 200 bar (2 900 psi) shall satisfy the essential safety
200 bar (2 900 psi) requirements set out in Annex I of the Pressure Equipment Directive 2014/68/EU. According to
Article 13 pressure equipment shall be classified by category in accordance with Annex II. The
conformity assessment of the pressure equipment shall be determined by the category I under
consideration of the above-mentioned low pressurized volume. These devices shall be provided with
CE marking.

Reasons:
• Pressure Equipment Directive 2014/68/EU, Article 13 and Annex II
• Pressure equipment directive 2014/68/EU, Commission's Working Group "Pressure", Guideline
A-05

Note:
A partial examination shall be performed for pressure instruments that are part of safety equipment
for the protection of a pipe or vessel from exceeding allowable limits (equipment with safety
function in accordance with Pressure Equipment Directive 2014/68/EU, Article 2, point 4).
The conformity assessment was performed according to module A; and proof of static strength and
fatigue strength according to EN 13445 and AD2000.
The FMP54 is not suitable for use with unstable gases at nominal pressures above 200 bar (2900
psi).

Steam boiler approval The FMP54 is approved as a limiting device for high water (HW) and low water (LW) for liquids in
containers which are subject to the requirements of EN12952-11 and EN12953-9 (certified by TÜV
Nord).
Product structure: Feature 590 "Additional Approval", option LX "Steam boiler approval".
For details refer to Safety Instructions SD00349F and Planning Recommendations SD01071F.
Device with steam boiler approval always have a SIL approval as well.

Marine certificate Device Marine certificate 1)

GL ABS LR BV DNV

FMP51     

FMP52     

FMP54 - - - - -

1) see oredering feature 590 "Additional approval"

Telecommunications Complies with part 15 of the FCC rules for an unintentional radiator. All probes meet the
requirements for a Class A digital device.
In addition, all probes in metallic tanks as well as the coax probe meet the requirements for a Class B
digital device.

CRN approval Some device versions have a CRN approval. Devices are CRN approved if the folowing two conditions
are met:
• The device has a CSA or FM approval (Product structure: Feature 010 "Approval")
• The device has a CRN approved process connection according to the following table.

Feature 100 of the product structure Approval

AAJ NPS 2" Cl. 600 RF, 316/316L flange ASME B16.5

ABJ NPS 3" Cl. 600 RF, 316/316L flange ASME B16.5

AEJ NPS 1-1/2" Cl. 150 RF, 316/316L flange ASME B16.5

AEK NPS 1-1/2" Cl. 150, PTFE>316/316L flange ASME B16.5

AEM NPS 1-1/2" Cl. 150, AlloyC>316/316L flange ASME B16.5

110 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Feature 100 of the product structure Approval

AFJ NPS 2" Cl. 150 RF, 316/316L flange ASME B16.5

AFK NPS 2" Cl. 150, PTFE>316/316L flange ASME B16.5

AFM NPS 2" Cl. 150, AlloyC>316/316L flange ASME B16.5

AGJ NPS 3" Cl. 150 RF, 316/316L flange ASME B16.5

AGK NPS 3" Cl. 150, PTFE>316/316L flange ASME B16.5

AGM NPS 3" Cl. 150, AlloyC>316/316L flange ASME B16.5

AHJ NPS 4" Cl. 150 RF, 316/316L flange ASME B16.5

AHK NPS 4" Cl. 150, PTFE>316/316L flange ASME B16.5

AJJ NPS 6" Cl. 150 RF, 316/316L flange ASME B16.5

AJK NPS 6" Cl. 150, PTFE>316/316L flange ASME B16.5

AKJ NPS 8" Cl. 150 RF, 316/316L flange ASME B16.5

AOJ NPS 4" Cl. 600 RF, 316/316L flange ASME B16.5

AQJ NPS 1-1/2" Cl. 300 RF, 316/316L flange ASME B16.5

AQK NPS 1-1/2" Cl. 300, PTFE>316/316L flange ASME B16.5

AQM NPS 1-1/2" Cl. 300, AlloyC>316/316L flange ASME B16.5

ARJ NPS 2" Cl. 300 RF, 316/316L flange ASME B16.5

ARK NPS 2" Cl. 300, PTFE>316/316L flange ASME B16.5

ARM NPS 2" Cl. 300, AlloyC>316/316L flange ASME B16.5

ASJ NPS 3" Cl. 300 RF, 316/316L flange ASME B16.5

ASK NPS 3" Cl. 300, PTFE>316/316L flange ASME B16.5

ASM NPS 3" Cl. 300, AlloyC>316/316L flange ASME B16.5

ATJ NPS 4" Cl. 300 RF, 316/316L flange ASME B16.5

ATK NPS 4" Cl. 300, PTFE>316/316L flange ASME B16.5

ATM NPS 4" Cl.300, AlloyC>316/316L flange ASME B16.5

AZJ NPS 4" Cl. 900 RF, 316/316L flange ASME B16.5

A6J NPS 2" Cl. 1500 RF, 316/316L Flansch ASME B16.5

A7J NPS 3" Cl. 1500 RF, 316/316L flange ASME B16.5

A8J NPS 4" Cl. 1500 RF, 316/316L flange ASME B16.5

GGJ Thread ISO228 G1-1/2, 316L

GIJ Thread ISO228 G1-1/2, 200bar, 316L

GJJ Thread ISO228 G1-1/2, 400bar, 316L

RAJ Thread ANSI MNPT1-1/2, 200bar, 316L

RBJ Thread ANSI MNPT1-1/2, 400bar, 316L

RGJ Thread ANSI MNPT1-1/2, 316L

TAK Tri-Clamp ISO2852 DN40-51 (2"), 3A, EHEDG, PTFE>316L

TDK Tri-Clamp ISO2852 DN40-51 (2"), PTFE>316L

TFK Tri-Clamp ISO2852 DN70-76.1 (3"), PTFE>316L

TJK Tri-Clamp ISO2852 DN38 (1-1/2"), PTFE>316L

TLK Tri-Clamp ISO2852 DN70-76.1 (3"), 3A, EHEDG, PTFE>316L

TNK Tri-Clamp ISO2852 DN38 (1-1/2"), 3A, EHEDG, PTFE>316L

• Process connections without CRN approval are not included in this table.

Endress+Hauser 111
Levelflex FMP51, FMP52, FMP54

• Refer to the product structure to see which process connections are available for a specific
device type.
• CRN approved devices are marked with the registration number 0F14480.5C on the
nameplate.

Track record FMP5x is the upgrade model of the corresponding FMP4x series.

112 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Test, Certificate Feature 580 Designation Available for


"Test,
Certificate"

JA 3.1 Material certificate, wetted metallic parts, EN10204-3.1 inspection • FMP51


certificate • FMP54

JB Conformity to NACE MR0175, wetted metallic parts • FMP51


• FMP52
• FMP54

JD 3.1 Material certificate, pressurized parts, EN10204-3.1 inspection FMP52


certificate

JE Conformity to NACE MR0103, wetted metallic parts • FMP51


• FMP52
• FMP54

JF Conformity to AD2000, wetted metallic parts: • FMP51


Material of all wetted/pressurized parts conform to AD2000 (Technical • FMP52
rules W2, W9, W10) • FMP54

JN Ambient temperature transmitter –50 °C (–58 °F) • FMP51


• FMP54
 Devices with this option undergo a routine check test (startup test at
–50 °C (–58 °F)).

KD Helium leak test, internal procedure, inspection certificate • FMP51


• FMP54

KE Pressure test, internal procedure, inspection certificate • FMP51


• FMP52
• FMP54

KG 3.1 Material certificate+PMI test (XRF), internal procedure, wetted metallic • FMP51
parts, EN10204-3.1 inspection certificate • FMP54

KP Liquid penetrant test AD2000-HP5-3(PT), wetted/pressurized metallic • FMP51


parts, inspection certificate • FMP54

KQ Liquid penetrant test ISO23277-1 (PT), wetted/pressurized metallic parts, • FMP51


inspection certificate • FMP54

KR Liquid penetrant test ASME VIII-1 (PT), wetted/pressurized metallic parts, • FMP51
inspection certificate • FMP54

KS Welding documentation, wetted/pressurized seams • FMP51


• FMP54
consisting of
• Welding drawing
• WPQR (Welding Procedure Qualification Report)
• WPS (Welding Process Specification)
• WQR (Manufacturer Declaration)

KV Conformity to ASME B31.3: • FMP51


The dimensions, materials of construction, pressure / temperature ratings • FMP52
and identification markings of the device comply with the requirements of • FMP54
ASME B31.3

Test reports, declarations and material test certificates can be obtained electronically from the
W@M Device Viewer :
Enter the serial number from the nameplate (www.endress.com/deviceviewer)
This applies to the options of the following ordering features:
• 550 "Calibration"
• 580 "Test, Certificate"

Product documentation on Test reports, declarations and material certificates can be ordered as hardcopies via ordering feature
paper 570 "Service", option I7 „Product documentation on paper“. The printouts will be enclosed in the
product delivery.

Endress+Hauser 113
Levelflex FMP51, FMP52, FMP54

Other standards and • EN 60529


guidelines Degrees of protection by housing (IP code)
• EN 61010-1
Protection Measures for Electrical Equipment for Measurement, Control, Regulation and
Laboratory Procedures.
• IEC/EN 61326
"Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC
requirements)
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment.
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with
analog output signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics
• NAMUR NE 107
Status classification as per NE107
• NAMUR NE 131
Requirements for field devices for standard applications
• IEC61508
Functional safety of electrical/electronic/programmable electronic safety-related systems

114 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Ordering information
Ordering information Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate"
-> Select your country -> Click "Products" -> Select the product using the filters and search field ->
Open product page -> The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop

Endress+Hauser 115
Levelflex FMP51, FMP52, FMP54

3-point linearity protocol The following notes must be taken into account if option F3 ("3 point linearity protocol") has
been selected in feature 550 ("Calibration").
Depending on the probe the 3 points of the linearity protocol are defined as follows:

A
1 1

2 2

LN
B
3 3

C
A0021843

A Distance from reference point R to first measuring point


B Measuring range
C Distance from end of probe to third measuring point
LN Length of probe
R Reference point of the measurement
1 First measuring point
2 Second measuring point (centrally between first and third measuring point)
3 Third measuring point

Rod or coax probe 1) Divisible rod probe Rope probe Rope probe
LN ≤ 6 m (20 ft) LN > 6 m (20 ft) LN ≤ 6 m (20 ft) LN > 6 m (20 ft)

Position of 1st measuring • FMP51/FMP52/FMP54 withoput gas phase compensation/ A = 350 mm (13.8 in) A = 350 mm (13.8 in)
point FMP55:
A = 350 mm (13.8 in)
• FMP54 with gas phase compensation, Lref = 300 mm (11 in):
A = 600 mm (23.6 in)
• FMP54 with gas phase compensation, Lref = 550 mm (21 in):
A = 850 mm (33.5 in)

Position of 2nd measuring centrally between 1st and centrally between 1st and 3rd centrally between 1st and centrally between 1st and
point 3rd measuring point measuring point 3rd measuring point 3rd measuring point

Position of 3rd measuring measured from the measured from the top end of the measured from the measured from the top
point bottom end of the probe: probe: bottom end of the probe: end of the probe:
C = 250 mm (9.84 in) A+B = 5 750 mm (226 in) C = 500 mm (19.7 in) A+B = 5 500 mm (217 in)

Minimum measuring B ≥ 400 mm (15.7 in) B ≥ 400 mm (15.7 in) B ≥ 400 mm (15.7 in) B ≥ 400 mm (15.7 in)
range

Minimum length of probe LN ≥ 1 000 mm (39.4 in) LN ≥ 1 000 mm (39.4 in) LN ≥ 1 250 mm (49.2 in) LN ≥ 1 250 mm (49.2 in)

1) also valid for divisible rods

The position of the measuring points may vary by ±1 cm (±0.04 in).

• For rod and rope probes the linearity check is performed with the complete device.
• For divisible rod probes a reference probe is used instead of the original one.
• For coax probes the electronics is mounted to a reference rod probe and the linearity check is
performed to this configuration.
• The linearity is checked under reference conditions.

116 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

5-point linearity protocol The following notes must be taken into account if option F4 ("5 point linearity protocol") has
been selected in feature 550 ("Calibration").
The five points of the linearity protocol are evenly distributetd across the measuring range (0% to
100%). In order to define the measuring range, Empty calibration (E) and Full calibration (F) have
to be specified 11).
The following restrictions have to be taken into account when defining E and F:

A
E
B

F
C
A0014673

Sensor Minimum distance Minimum measuring


between reference point range
(R) and 100% level

FMP51 A ≥ 250 mm (10 in) B ≥ 400 mm (16 in)

FMP51 A ≥ 350 mm (14 in) B ≥ 400 mm (16 in)


Rope probe with center rod, max. 300 mm (12 in)
nozzle height 1)

FMP52 A ≥ 250 mm (10 in) B ≥ 400 mm (16 in)

FMP52 A ≥ 350 mm (14 in) B ≥ 400 mm (16 in)


Rope probe with center rod, max. 300 mm (12 in)
nozzle height 2)

FMP54 w/o gas phase compensation A ≥ 250 mm (10 in) B ≥ 400 mm (16 in)

FMP54 with gas phase compensation, Lref = 300 mm A ≥ 450 mm (18 in) B ≥ 400 mm (16 in)

FMP54 with gas phase compensation, Lref = 550 mm A ≥ 700 mm (28 in) B ≥ 400 mm (16 in)

1) Product structure: Feature 060 "Probe", Option MB or MD


2) Product structure: Feature 060 "Probe", Option OB or OD

11) If E and F are not specified, probe dependent default values will be used instead.

Endress+Hauser 117
Levelflex FMP51, FMP52, FMP54

Type of probe Minimum distance from Maximum value for


end of probe to 0% level "empty calibration"

Rod (not divisible) C ≥ 100 mm (4 in) E ≤ 3.9 m (12.8 ft)

• Coax C ≥ 100 mm (4 in) E ≤ 5.9 m (19.4 ft)


• Rod (divisible)

Rope C ≥ 1 000 mm (40 in) E ≤ 23 m (75 ft)

• For rod and rope probes the linearity check is performed with the complete device.
• For coax probes the electronics is mounted to a reference rod probe and the linearity check is
performed to this configuration.
• The linearity is checked under reference conditions.
The selected values of Empty calibration and Full calibration are only used to record the
linearity protocol and are reset to their probe specific default values thereafter. If values
different from the default are required, they must be ordered as a customized parametrization
→  119.

118 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Customized parametrization If the option IJ "Customized parametrization HART", IK "Customized parametrization PA" or IL
"Customized parametrization FF" has been selected in feature 570 "Service", customer specific
presettings can be selected for the following parameters:

Parameter Communication Selection list / range of


values

Setup → Distance unit • HART • in


• PA • ft
• FF • mm
• m

Setup → Empty calibration • HART 0 to 45 m (0 to 147 ft)


• PA
• FF

Setup → Full calibration • HART 0 to 45 m (0 to 147 ft)


• PA
• FF

Setup → Adv. Setup → Current output 1/2 → Damping HART 0 to 999.9 s

Setup → Adv. Setup → Current output 1/2 → Failure mode HART • Min
• Max
• Last valid value

Setup → Adv. Setup → Current output 1/2 → Burst mode HART • Off
• On

Tagging (TAG) Ordering feature 895: Marking

Option Z1: Tagging (TAG), see additional spec.

Position of the measuring To be selected in the additional specifications:


point marking • Tag plate Stainless Steel
• Self-adhesive paper label
• Supplied label/plate
• RFID TAG
• RFID TAG + Tag plate Stainless Steel
• RFID TAG + Self-adhesive paper label
• RFID TAG + Supplied label/plate

Definition of the measuring To be defined in the additional specifications:


point designation 3 lines containing up to 18 characters each
The measuring point designation appears on the selected label and/or the
RFID TAG.

Designation in the Electronic The first 32 characters of the measuring point designation
Name Plate (ENP)

Designation on the display The first 12 characters of the measuring point designation
module

Endress+Hauser 119
Levelflex FMP51, FMP52, FMP54

Application Packages
Heartbeat Diagnostics Availability
Available in all device versions.

Function
• Continuous self-monitoring of the device.
• Diagnostic messages output to
– the local display.
– an asset management system (e.g. FieldCare/DeviceCare).
– an automation system (e.g. PLC).

Advantages
• Device condition information is available immediately and processed in time.
• The status signals are classified in accordance with VDI/VDE 2650 and NAMUR recommendation
NE 107 and contain information about the cause of the error and remedial action.

Detailed description
See Operating Instructions of the device (→  133); chapter "Diagnostics and trouble shooting"

120 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Heartbeat Verification Availability


Available for the following options of feature 540 "Application Package":
• EH: Heartbeat Verification + Monitoring
• EJ:   Heartbeat Verification

Device functionality checked on demand


• Verification of the correct functioning of the measuring device within specifications.
• The verification result provides information about the condition of the device: Passed or Failed.
• The results are documented in a verification report.
• The automatically generated report supports the obligation to demonstrate compliance with
internal and external regulations, laws and standards.
• Verification is possible without interrupting the process.

Advantages
• No onsite presence is required to use the function.
• The DTM 12) triggers verification in the device and interprets the results. No specific knowledge is
required on the part of the user.
• The verification report can be used to prove quality measures to a third party.
• Heartbeat Verification can replace other maintenance tasks (e.g. periodic check) or extend the
test intervals.

SIL/WHG-locked devices 13)

• The Heartbeat Verification module contains a wizard for the proof test which must be performed
at appropriate intervals for the following applications:
– SIL (IEC61508/IEC61511)
– WHG (German Water Resources Act)
• To perform a proof test, the device must be locked (SIL/WHG locking).
• The wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.
In the case of SIL-locked and WHG-locked devices, it is not possible to perform verification
without additional measures (e.g. by-passing of the output current) because the output current
must be simulated (Increased safety mode) or the level must be approached manually (Expert
mode) during subsequent re-locking (SIL/WHG locking).

Detailed description
SD01872F

12) DTM: Device Type Manager; controls device operation via DeviceCare, FieldCare or a DTM-based process control system.
13) Only relevant for devices with SIL or WHG approval: order code 590 ("Additional approval"), option LA ("SIL") or LC ("WHG").

Endress+Hauser 121
Levelflex FMP51, FMP52, FMP54

Heartbeat Monitoring Availability


Available for the following options of feature 540 "Application Package":
EH: Heartbeat Verification + Monitoring

Function
• In addition to the verification parameters, the corresponding parameter values are also logged.
• Existing measured variables, such as the echo amplitude, are used in the Foam detection and
Build-up detection wizards.
The Foam detection and Build-up detection wizards cannot be used together.

"Foam detection" wizard


• The Heartbeat Monitoring module contains the Foam detection wizard.
• This wizard is used to configure automatic foam detection, which detects foam on the product
surface on the basis of the reduced signal amplitude. Foam detection can be linked to a switch
output in order to control a sprinkler system, for example, which dissolves the foam.
• This wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.

"Build-up detection" wizard


• The Heartbeat Monitoring module contains the Build-up detection wizard.
• The wizard is used to configure automatic buildup detection, which detects the buildup of deposits
on the probe on the basis of the reduced signal amplitude.
• This wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.

Advantages
• Early detection of changes (trends) to ensure plant availability and product quality.
• Use of information for the proactive planning of measures (e.g. cleaning/maintenance).
• Identification of undesirable process conditions as the basis to optimizing the facility and the
processes.
• Automated control of measures to remove foam or buildup.

Detailed description
SD01872F

122 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Accessories
Device-specific accessories Weather protection cover

Accessory Description

Weather protection cover

222 (8.74)
A0015466

300 (11.8) 275 (10.8)


255.1 (10) 164 (6.46) 54
(2
.13
)
55
5)
(1.

4)
38

35°
65°
(6.1
156

A0015472

 41 Weather protection cover; Dimensions: mm (in)

 Enclosed",
The weather protection cover can be ordered together with the device (product structure, feature 620 "Accessory
option PB "Weather Protection Cover").
Alternatively, it can be separately ordered as an accessory; order code 71162242.

Endress+Hauser 123
Levelflex FMP51, FMP52, FMP54

Mounting bracket for the electronics housing

Accessory Description

Mounting bracket for the A B


electronics housing

ø42...60
(1.65...2.36)

52 (2)
122 (4.8) 127...140
161 (6.34) (5...5.51)
162...175
(6.38...6.89)
(3.4)
(2.8)

86
70

A0014793

 42 Mounting bracket for the electronics housing; Dimensions: mm (in)


A Wall mounting
B Pipe mounting

 the
For the "Sensor remote" device version (see feature 060 of the product structure), the mounting bracket is part of
delivery. If required, it can also be ordered as an accessory (order code 71102216).

124 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Extension rod / centering HMP40

Accessory Description

Extension rod / centering


HMP40
• can be used for:
FMP54
• Admissible
temperature
at lower nozzle edge:
– without center 1
washer:
no limitation
– with center washer:
-40 to 150 °C
(-40 to 302 °F)
• Additional information:
SD01002F 2 3
A0013597

1 Nozzle height
2 Extension rod
3 Center washer

010 Approval:

A A: Non-hazardous area

M M: FM DIP Cl.II Div.1 Gr.E-G N.I., zone 21,22

P P: CSA DIP Cl.II Div.1 Gr.G + coal dust N.I.

S S: FM Cl.I, II, III Div.1 Gr.A-G N.I., zone 0,1,2,20,21,22

U U: CSA Cl.I, II, III Div.1 Gr.A-G N.I., zone 0,1,2

1 1: ATEX II 1G

2 2: ATEX II 1D

020 Extension rod; nozzle height:

1 115mm; 150-250mm / 6-10"

2 215mm; 250-350mm / 10-14"

3 315mm; 350-450mm / 14-18"

4 415mm; 450-550mm / 18-22"

9 Special version, TSP-no. to be spec.

030 Center washer:

A Not selected

B DN40 / 1-1/2", inside-d. = 40-45mm, PPS

C DN50 / 2", inside-d. = 50-57mm, PPS

D DN80 / 3", inside-d. = 80-85mm, PPS

E DN80 / 3", inside-d. = 76-78mm, PPS

G DN100 / 4", inside-d. = 100-110mm, PPS

H DN150 / 6", inside-d. = 152-164mm, PPS

J DN200 / 8", inside-d. = 210-215mm, PPS

K DN250 / 10", inside-d. = 253-269mm, PPS

Y Special version, TSP-no. to be spec.

Endress+Hauser 125
Levelflex FMP51, FMP52, FMP54

Mounting kit, isolated

Accessory Description

Mounting kit, isolated


can be used for
• FMP51
• FMP54

A0013586

 43 Scope of supply of the mounting kit:


1 Insulating sleeve
2 Eye-bolt

For reliably insulated fixing of the probe.


Maximum process temperature: 150 °C (300 °F)
For rope probes 4 mm (¹⁄₆ in) or 6 mm (1/4 in) with PA>steel:
• Diameter D = 20 mm (0.8 in)
• Order-No.: 52014249
For rope probes 6 mm (¹⁄₄ in) or 8 mm (1/3 in) with PA>steel:
• Diameter D = 25 mm (1 in)
• Order-No.: 52014250
Due to the risk of electrostatic charge, the insulating sleeve is not suitable for use
in hazardous areas. In these cases the fixing must be reliably grounded.

 The mounting kit can also be ordered directly with the device (see the
Levelflex product structure, feature 620 "Accessory Enclosed", option PG
"Mounting kit, isolated, rope").

126 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Centering star

Accessory Description

Centering star PEEK  48 10 (0.39)


to 95 mm (1.89 to
3.74 inch)

3)
.6
can be used for

(0
6
• FMP51

ø1
7 (0.28)
• FMP54

max. 50 (1.97)
ø48 (1.89)
ø75 (2.95)
ø95 (3.74)
A0014576

The centering star is suitable for probes with a rod diameter of 16 mm (0.6 in) and
can be used in pipes from DN50 (2") up to DN100 (4"). Markings on the 4-leg
centering star ensure a simple tailoring. Hence the centering star can be adapted to
the pipe diameter. See also Operating Instructions BA00377F/00/A2.
• Material of centering star: PEEK (statically dissipative)
• Material of retaining rings: PH15-7Mo (UNS S15700)
• Admissible process temperature: –60 to +200 °C (–76 to +392 °F)
• Order code: 71069064

 Iflower
the centering star is inserted in a bypass, it must be positioned below the
bypass outlet. This has to be accounted for when choosing the probe
length. Generally, the centering star should not be mounted higher than 50
mm (1.97") from the probe end. It is recommended not to insert the PEEK
centering star in the measuring range of the rod probe.

 The PEEK centering star can also be ordered directly with the device (see the
Levelflex product structure, feature 610 "Accessory mounted", option OD). In
this case it is not fastened by the retaining rings. Instead it is fixed to the end
of the probe rod by a hexagonal headed bolt (A4-70) and a Nord-Lock
washer (1.4547).

Endress+Hauser 127
Levelflex FMP51, FMP52, FMP54

Accessory Description

Centering star PFA

10 (0.39)
•  16.4 mm (0.65 in)
•  37 mm (1.46 in)
can be used for

A: ø16.4 (0.65)
B: ø37 (1.46)
• FMP51
• FMP52
• FMP54

A0014577

A For 8 mm (0.3 in) probes


B: For 12 mm (0.47 in) and 16 mm (0.63 in) probes

The centering star is suitable for probes with a rod diameter of 8 mm (0.3 in),
12 mm (0.47 in) and 16 mm (0.63 in) (also coated rod probes) and can be used in
pipes from DN40 (1½") up to DN50 (2"). See also Operating Instructions
BA00378F/00/A2.
• Material: PFA
• Admissible process temperature: –200 to +200 °C (–382 to +392 °F)
• Order code
– Probe 8 mm (0.3 in): 71162453
– Probe 12 mm (0.47 in): 71157270
– Probe 16 mm (0.63 in): 71069065

 The PFA centering star can also be ordered directly with the device (see the
Levelflex product structure, feature 610 "Accessory mounted", option OE).

Accessory Description

Centering star PEEK, ⌀ Ø4


48 to 95 mm
(1.9 to 3.7 in)

5,5
can be used for
• FMP51
• FMP54
5
8
Ø4

Ø75

Ø9
5

A0035182

The centering star is suitable for probes with a rope diameter of 4 mm (¹⁄₆ in) (also
coated rod probes). See also Operating Instructions SD01961F.
• Material: PEEK
• Admissible process temperature: –60 to +250 °C (–76 to +482 °F)
• Order code
– 71373490 (1 pc.)
– 71373492 (5 pcs.)

 The PFA centering star can also be ordered directly with the device (see the
Levelflex product structure, feature 610 "Accessory mounted", option OD).

128 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Remote display FHX50

Accessories Description

Remote display FHX50

2.
1.
4.
3.

A0019128

• Material:
– Plastic PBT
– 316L/1.4404
– Aluminum
• Degree of protection: IP68 / NEMA 6P and IP66 / NEMA 4x
• Suitable for display modules:
– SD02 (push buttons)
– SD03 (touch control)
• Connecting cable:
– Cable supplied with device up to 30 m (98 ft)
– Standard cable supplied by customer up to 60 m (196 ft)
• Ambient temperature range: –40 to 80 °C (–40 to 176 °F)
• Ambient temperature range (option): –50 to 80 °C (–58 to 176 °F) 1)

 • IfL ortheM).remote display should be used, order the device version "Prepared for display FHX50" (feature 030, version
For the FHX50, you must select option A: "Prepared for display FHX50" under feature 050 "Measuring
device version".
• If the device version "Prepared for display FHX50" was not originally ordered and a FHX50 display is to be
retrofitted, you must select version B "Not prepared for display FHX50" under feature 050: "Measuring device
version" when ordering the FHX50. In this case, a retrofit kit for the device is supplied with the FHX50. The kit
can be used to prepare the device so that the FHX50 can be used.

 FHX50
Use of the FHX50 may be restricted for transmitters with an approval. A device can only be retrofitted with the
if the option L or M ("Prepared for FHX50") is listed under Basic specifications, item 4 "Display, operation" in
the Safety Instructions (XA) for the device.
Also pay attention to the Safety Instructions (XA) of the FHX50.

 Retrofitting is not possible on transmitters with:


• An approval for use in areas with flammable dust (dust ignition-proof approval)
• Ex nA type of protection

 For details, see document SD01007F.


1) This range is valid if option JN "Ambient temperature transmitter –50 °C (–58 °F)" has been selected in ordering feature 580 "Test, Certificate". If
the temperature is permanently below –40 °C (–40 °F), failure rates may be increased.

Endress+Hauser 129
Levelflex FMP51, FMP52, FMP54

Overvoltage protection

Accessory Description

Overvoltage protection for


2-wire-devices
OVP10 (1 channel)
OVP20 (2 channel)

A0021734

Technical data
• Resistance per channel: 2 * 0.5 Ωmax
• Threshold DC voltage: 400 to 700 V
• Threshold impulse voltage: < 800 V
• Capacitance at 1 MHz: < 1.5 pF
• Nominal arrest impulse voltage (8/20 μs): 10 kA
• Suited for wire cross-sections: 0.2 to 2.5 mm2 (24 to 14 AWG)

 Ordering with device


The overvoltage protection module is preferably ordered with the device. See product structure, feature 610
"Accessory mounted", option NA "Overvoltage protection". Separate ordering of the module is only necessary if a
device is to retrofitted with the overvoltage protection.

 Order code for retrofitting


• For 1-channel devices (feature 020, option A)
OVP10: 71128617
• For 2-channel devices (feature 020, option B, C, E or G)
OVP20 : 71128619
Hosuing lid for retrofitting
In order to keep the necessary safety distances, the housing lid needs to be replaced if the device is retrofitted with
the overvoltage protection. Depending on the housing type, the order code of the suitable lid is as follows:
• GT18 housing: Lid 71185516
• GT19 housing: Lid 71185518
• GT20 housing: Lid 71185516

 Restrictions for retrofitting


Depending on the approval of the transmitter the usage of the OVP module may be restricted. A device may only be
retrofitted with an OVP module if the option NA (overvoltage protection) is quoted unter Optional Specifications in
the Safety Instructions (XA) pertaining to the device.

 For details refer to SD01090F.

130 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Communication-specific Accessory Description


accessories
Commubox FXA195 For intrinsically safe HART communication with FieldCare via the USB interface.
HART
 For details refer to Technical Information TI00404F

Accessory Description

Commubox FXA291 Connects Endress+Hauser field devices with CDI interface (= Endress+Hauser
Common Data Interface) to the USB interface of a computer.
Order code: 51516983

 For details refer to Technical Information TI00405C

Accessory Description

HART Loop Converter Evaluates the dynamic HART variables and converts them to analog current signals
HMX50 or limit values.
Order code: 71063562

 For details refer to Technical Information TI00429F and Operating


Instructions BA00371F

Accessory Description

WirelessHART Adapter Connects field devices to a WirelessHART network.


SWA70 The WirelessHART adapter can be mounted directly at a HART device and is easly
integrated into an existing HART network. It ensures safe data transmission and
can be operated in parallel with other wireless networks.

 For details refer to Operating Instructions BA00061S

Accessories Description

Connect Sensor Fully integrated, battery-powered gateway for simple applications with SupplyCare Hosting.
FXA30/FXA30B Up to 4 field devices with 4 to 20 mA communication (FXA30/FXA30B), serial Modbus
(FXA30B) or HART (FXA30B) can be connected. With its robust design and ability to run for
years on the battery, it is ideal for remote monitoring in isolated locations. Version with LTE
(USA, Canada and Mexico only) or 3G mobile transmission for worldwide communication.

 BA01710S.
For details, see "Technical Information" TI01356S and Operating Instructions

Accessories Description

Fieldgate Fieldgates enable communication between connected 4 to 20 mA, Modbus RS485 and
FXA42 Modbus TCP devices and SupplyCare Hosting or SupplyCare Enterprise. The signals are
transmitted either via Ethernet TCP/IP, WLAN or mobile communications (UMTS). Advanced
automation capabilities are available, such as an integrated Web-PLC, OpenVPN and other
functions.

 BA01778S.
For details, see "Technical Information" TI01297S and Operating Instructions

Accessories Description

SupplyCare Inventory management software that visualizes levels, volumes, masses, temperatures,
Enterprise pressures, densities or other tank parameters. The parameters are recorded and transmitted
SCE30B by means of gateways of the type Fieldgate FXA42.
This Web-based software is installed on a local server and can also be visualized and
operated with mobile terminals such as a smartphone or tablet.

 BA00055S
For details, see "Technical Information" TI01228S and Operating Instructions

Endress+Hauser 131
Levelflex FMP51, FMP52, FMP54

Accessories Description

SupplyCare Inventory management software that visualizes levels, volumes, masses, temperatures,
Hosting SCH30 pressures, densities or other tank parameters. The parameters are recorded and transmitted
by means of gateways of the type Fieldgate FXA42, FXA30 and FXA30B.
SupplyCare Hosting is offered as a hosting service (Software as a Service, SaaS). In the
Endress+Hauser portal, the user is provided with the data over the Internet.

 BA00050S.
For details, see "Technical Information" TI01229S and Operating Instructions

Accessory Description

Field Xpert SFX350 Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
fieldbus devices in the non-Ex area.

 For details, see Operating Instructions BA01202S

Accessory Description

Field Xpert SFX370 Field Xpert SFX370 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
fieldbus devices in the non-Ex area and the Ex area.

 For details, see Operating Instructions BA01202S


Service-specific accessories Accessory Description

DeviceCare SFE100 Configuration tool for HART, PROFIBUS and FOUNDATION Fieldbus devices

 Technical Information TI01134S


 • DeviceCare is available for download at
www.software-products.endress.com. The download requires a
registration in the Endress+Hauser software portal.
• Alternatively, a DeviceCare DVD can be ordered with the device. Product
structure: Feature 570 "Service", Option IV "Tooling DVD (DeviceCare
Setup)".

FieldCare SFE500 Endress+Hauser's FDT-based Plant Asset Management tool.


Helps to configure and maintain all field devices of your plant. By supplying status
information it also supports the diagnosis of the devices.

 Technical Information TI00028S


System components Accessory Description

Graphic Data Manager The graphic data manager Memograph M provides information on all the relevant
Memograph M process variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on an SD card or USB stick.

 For details refer to Technical Information TI00133R and Operating


Instructions BA00247R

RN221N Active barrier with power supply for safe separation of 4 to 20 mA current circuits.
Provides bi-directional HART transmission.

 For details refer to Technical Information TI00073R and Operating


Instructions BA00202R

RNS221 Transmitter supply for 2-wire sensors or transmitters exclusively for non-Ex areas.
Provides bi-directional communication using the HART communication sockets.

 For details refer to Technical Information TI00081R and Operating


Instructions KA00110R

132 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the
2-D matrix code (QR code) on the nameplate.

Standard documentation Levelflex FMP51, FMP52, FMP54

Correlation of documentations to the device:


Device Power supply, Communication Document type Document code
output

FMP51, A, B, C, K, L HART Technical Information TI01001F


FMP52,
FMP54 Operating Instructions BA01001F

Brief Operating Instructions KA01077F

Description of Device GP01000F


Parameters

G PROFIBUS PA Technical Information TI01001F

Operating Instructions BA01006F

Brief Operating Instructions KA01079F

Description of Device GP01001F


Parameters

E FOUNDATION Fieldbus Technical Information TI01001F

Operating Instructions BA01052F

Brief Operating Instructions KA01107F

Description of Device GP01015F


Parameters

Supplementary Application Package 1) Document type Document code


documentation
• EH: Heartbeat Verification + Monitoring Special documentation SD01872F
• EJ: Heartbeat Verification

1) Feature 540 of the product structure

Device Document type Document code

Fieldgate FXA520 Technical Information TI369F

Tank Side Monitor NRF590 Technical Information TI402F

Operating Instructions BA256F

Description of Device BA257F


Parameters

Description Document type Document code

Continuous level measurement in liquids and bulk solids Competence brochure CP00023F
Selection and engineering guide for the process industry

Planning instructions for steam boiler approval Special documentation SD01071F


For Levelflex M FMP45 and Levelflex FMP54

Endress+Hauser 133
Levelflex FMP51, FMP52, FMP54

Safety documentation Safety Instructions (XA)


Depending on the approval, the following Safety Instructions (XA) are supplied with the device. They
are an integral part of the Operating Instructions.

Feature 010 Approval Available for Feature 020: "Power Supply; Output"
1)
A B 2) C 3) E 4)/G 5) K 6)/L 7)

BA ATEX II 1G Ex ia IIC T6 Ga • FMP51


• FMP52 XA00496F XA01125F XA01126F XA00516F -
• FMP54

BB ATEX II 1/2G Ex ia IIC T6 Ga/Gb • FMP51


• FMP52 XA00496F XA01125F XA01126F XA00516F -
• FMP54

BC ATEX II 1/2G Ex d[ia] IIC T6 Ga/Gb • FMP51 XA00499F XA00499F XA00499F XA00519F XA01133F
• FMP52
• FMP54

BD ATEX II 1/3G Ex ic[ia] IIC T6 Ga/Gc • FMP51 XA00497F XA01127F XA01128F XA00517F -
• FMP52
• FMP54

BE ATEX II 1D Ex t IIIC Da FMP54 XA00501F XA00501F XA00501F XA00521F XA00501F

BF ATEX II 1/2D Ex t IIIC Da/Db FMP54 XA00501F XA00501F XA00501F XA00521F XA00501F

BG ATEX II 3G Ex nA IIC T6 Gc • FMP51


• FMP52 XA00498F XA01130F XA01131F XA00518F XA01132F
• FMP54

BH ATEX II 3G Ex ic IIC T6 Gc • FMP51


• FMP52 XA00498F XA01130F XA01131F XA00518F -
• FMP54

BL ATEX II 1/3G Ex nA[ia] IIC T6 Ga/Gc • FMP51 XA00497F XA01127F XA01128F XA00517F XA01129F
• FMP52
• FMP54

B2 ATEX II 1/2G Ex ia IIC T6 Ga/Gb, 1/2D Ex ia IIIC Da/Db • FMP51 XA00502F XA00502F XA00502F XA00522F -
• FMP52
• FMP54

B3 ATEX II 1/2G Ex d[ia] IIC T6 Ga/Gb, 1/2 D Ex t IIIC • FMP51 XA00503F XA00503F XA00503F XA00523F XA01136F
Da/Db • FMP52
• FMP54

B4 ATEX II 1/2G Ex ia IIC T6 Ga/Gb, Ex d[ia] IIC T6 Ga/Gb • FMP51 XA00500F XA01134F XA01135F XA00520F -
• FMP52
• FMP54

CD CSA C/US DIP Cl.II,III Div.1 Gr.E-G FMP54 XA00529F XA00529F XA00529F XA00570F XA00529F

C2 CSA C/US IS Cl.I,II,III Div.1 Gr.A-G, NI Cl.1 Div.2, Ex ia • FMP51 XA00530F XA00530F XA00530F XA00571F XA00530F
• FMP52
• FMP54

C3 CSA C/US XP Cl.I,II,III Div.1 Gr.A-G, NI Cl.1 Div.2, Ex d • FMP51 XA00529F XA00529F XA00529F XA00570F XA00529F
• FMP52
• FMP54

FB FM IS Cl.I,II,III Div.1 Gr.A-G, AEx ia, NI Cl.1 Div.2 • FMP51 XA00531F XA00531F XA00531F XA00573F XA00531F
• FMP52
• FMP54

FD FM XP Cl.I,II,III Div.1 Gr.A-G, AEx d, NI Cl.1 Div.2 • FMP51 XA00532F XA00532F XA00532F XA00572F XA00532F
• FMP52
• FMP54

FE FM DIP Cl.II,III Div.1 Gr.E-G FMP54 XA00532F XA00532F XA00532F XA00572F XA00532F

GA EAC Ex ia IIC T6 Ga • FMP51 XA01380F XA01380F XA01380F XA01381F XA01380F


• FMP52
• FMP54

134 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Feature 010 Approval Available for Feature 020: "Power Supply; Output"
1)
A B 2) C 3) E 4)/G 5) K 6)/L 7)

GB EAC Ex ia IIC T6 Ga/Gb • FMP51 XA01380F XA01380F XA01380F XA01381F XA01380F


• FMP52
• FMP54

GC EAC Ex d[ia] IIC T6 Ga/Gb • FMP51 XA01382F XA01382F XA01382F XA01383F XA01382F
• FMP52
• FMP54

IA IEC Ex ia IIC T6 Ga • FMP51


• FMP52 XA00496F XA01125F XA01126F XA00516F -
• FMP54

IB IEC Ex ia IIC T6 Ga/Gb • FMP51


• FMP52 XA00496F XA01125F XA01126F XA00516F -
• FMP54

IC IEC Ex d[ia] IIC T6 Ga/Gb • FMP51 XA00499F XA00499F XA00499F XA00519F XA01133F
• FMP52
• FMP54

ID IEC Ex ic[ia] IIC T6 Ga/Gc • FMP51 XA00497F XA01127F XA01128F XA00517F -


• FMP52
• FMP54

IE IEC Ex t IIIC Da FMP54 XA00501F XA00501F XA00501F XA00521F XA00501F

IF IEC Ex t IIIC Da/Db FMP54 XA00501F XA00501F XA00501F XA00521F XA00501F

IG IEC Ex nA IIC T6 Gc • FMP51


• FMP52 XA00498F XA01130F XA01131F XA00518F XA01132F
• FMP54

IH IEC Ex ic IIC T6 Gc • FMP51


• FMP52 XA00498F XA01130F XA01131F XA00518F -
• FMP54

IL IEC Ex nA[ia] IIC T6 Ga/Gc • FMP51 XA00497F XA01127F XA01128F XA00517F XA01129F
• FMP52
• FMP54

I2 IEC Ex ia IIC T6 Ga/Gb, Ex ia IIIC Da/Db • FMP51 XA00502F XA00502F XA00502F XA00522F -
• FMP52
• FMP54

I3 IEC Ex d [ia] IIC T6 Ga/Gb, Ex t IIIC Da/Db • FMP51 XA00503F XA00503F XA00503F XA00523F XA01136F
• FMP52
• FMP54

I4 IEC Ex II 1/2G Ex ia IIC T6 Ga/Gb, Ex d[ia] IIC T6 Ga/Gb • FMP51 XA00500F XA01134F XA01135F XA00520F -
• FMP52
• FMP54

KA KC Ex ia IIC T6 Ga • FMP51 XA01169F - XA01169F - -


• FMP52
• FMP54

KB KC Ex ia IIC T6 Ga/Gb • FMP51 XA01169F - XA01169F - -


• FMP52
• FMP54

KC KC Ex d[ia] IIC T6 • FMP51 - - XA01170F - -


• FMP52
• FMP54

MA INMETRO Ex ia IIC T6 Ga • FMP51 XA01038F XA01038F XA01038F - XA01038F


• FMP52
• FMP54

MC INMETRO Ex d[ia] IIC T6 Ga/Gb • FMP51 XA01041F XA01041F XA01041F - XA01041F


• FMP52
• FMP54

ME INMETRO Ex t IIIC Da FMP54 XA01043F XA01043F XA01043F - XA01043F

Endress+Hauser 135
Levelflex FMP51, FMP52, FMP54

Feature 010 Approval Available for Feature 020: "Power Supply; Output"
1)
A B 2) C 3) E 4)/G 5) K 6)/L 7)

MH INMETRO Ex ic IIC T6 Gc • FMP51 XA01040F XA01040F XA01040F - XA01040F


• FMP52
• FMP54

NA NEPSI Ex ia IIC T6 Ga • FMP51 XA00634F XA00634F XA00634F XA00640F XA00634F


• FMP52
• FMP54

NB NEPSI Ex ia IIC T6 Ga/Gb • FMP51 XA00634F XA00634F XA00634F XA00640F XA00634F


• FMP52
• FMP54

NC NEPSI Ex d[ia] IIC T6 Ga/Gb • FMP51 XA00636F XA00636F XA00636F XA00642F XA00636F
• FMP52
• FMP54

NF NEPSI DIP A20/21 T85…90oC IP66 FMP54 XA00637F XA00637F XA00637F XA00643F XA00637F

NG NEPSI Ex nA II T6 Gc • FMP51 XA00635F XA00635F XA00635F XA00641F XA00635F


• FMP52
• FMP54

NH NEPSI Ex ic IIC T6 Gc • FMP51 XA00635F XA00635F XA00635F XA00641F XA00635F


• FMP52
• FMP54

N2 NEPSI Ex ia IIC T6 Ga/Gb, Ex iaD 20/21 T85…90°C • FMP51 XA00638F XA00638F XA00638F XA00644F XA00638F
• FMP52
• FMP54

N3 NEPSI Ex d[ia] IIC T6 Ga/Gb, DIP A20/21 T85…90°C • FMP51 XA00639F XA00639F XA00639F XA00645F XA00639F
IP66 • FMP52
• FMP54

8A FM/CSA IS+XP Cl.I,II,III Div.1 Gr.A-G • FMP51 XA00531F XA00531F XA00531F XA00572F XA00531F
• FMP52 XA00532F XA00532F XA00532F XA00573F XA00532F
• FMP54

1) A: 2-wire; 4-20mA HART


2) B: 2-wire; 4-20mA HART, switch output
3) C: 2-wire; 4-20mA HART, 4-20mA
4) E: 2-wire; FOUNDATION Fieldbus, switch output
5) G: 2-wire; PROFIBUS PA, switch output
6) K: 4-wire 90-253VAC; 4-20mA HART
7) L: 4-wire 10,4-48VDC; 4-20mA HART

For certified devices the relevant Safety Instructions (XA) are indicated on the nameplate.

136 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Ex-marking in case of connected FHX50 remote display


If the device is prepared for the remote display FHX50 (product structure: feature 030: Display,
Operation", option L or M), the Ex marking of some certificates changes according to the following
table 14):

Feature 010 ("Approval") Feature 030 ("Display, Operation") Ex-marking

BE L or M ATEX II 1D Ex ta [ia] IIIC T500 xx°C Da

BF L or M ATEX II 1/2 D Ex ta [ia Db] IIIC Txx°C Da/Db

BG L or M ATEX II 3G Ex nA [ia Ga] IIC T6 Gc

BH L or M ATEX II 3G Ex ic [ia Ga] IIC T6 Gc

B3 L or M ATEX II 1/2G Ex d [ia] IIC T6 Ga/Gb,


ATEX II 1/2D Ex ta [ia Db] IIIC Txx°C Da/Db

IE L or M IECEx Ex ta [ia] IIIC T500 xx°C Da

IF L or M IECEx ta [ia Db] IIIC Txx°C Da/Db

IG L or M IECEx Ex nA [ia Ga] IIC T6 Gc

IH L or M IECEx Ex ic [ia Ga] IIC T6 Gc

I3 L or M IECEx Ex d [ia] IIC T6 Ga/Gb,


IECEx Ex ta [ia Db] IIIC Txx°C Da/Db

14) The marking of certificates not mentioned in this table are not affected by the FHX50.

Endress+Hauser 137
Levelflex FMP51, FMP52, FMP54

Registered trademarks
HART®
Registered trademark of the FieldComm Group, Austin, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
FOUNDATIONTM Fieldbus
Registered trademark of the FieldComm Group, Austin, Texas, USA
KALREZ®, VITON®
Registered trademark of DuPont Performance Elastomers L.L.C., Wilmington, USA
TEFLON®
Registered trademark of E.I. DuPont de Nemours & Co., Wilmington, USA
TRI CLAMP®
Registered trademark of Alfa Laval Inc., Kenosha, USA
NORD-LOCK®
Registered trademark of Nord-Lock International AB
FISHER®
Registered trademark of Fisher Controls International LLC, Marshalltown, USA
MASONEILAN®
Registered trademark of Dresser, Inc., Addison, USA

138 Endress+Hauser
Levelflex FMP51, FMP52, FMP54

Patents
This product may be protected by at least one of the following patents.
Further patents are pending.

US Patents EP Patents

5.827.985 ---

5.884.231 ---

5.973.637 ---

6.087.978 955 527

6.140.940 ---

6.481.276 ---

6.512.358 1 301 914

6.559.657 1 020 735

6.640.628 ---

6.691.570 ---

6.847.214 ---

7.441.454 ---

7.477.059 ---

--- 1 389 337

7.965.087 ---

Endress+Hauser 139
*71384596*
71384596

www.addresses.endress.com

You might also like