Admixture

The document summarizes the measurement of uncertainty for testing the pH and relative density of an admixture. It identifies sources of uncertainty including repeatability, calibration buffers/hydrometer, and thermometer. It calculates the standard uncertainty from each source, combines them, and determines the expanded uncertainty of 1.0887 or ±0.14 at a 95% confidence level for pH testing and 0.0086 or ±0.0086 for relative density testing.

Uploaded by

Mishra dtu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
36 views5 pages

Admixture

The document summarizes the measurement of uncertainty for testing the pH and relative density of an admixture. It identifies sources of uncertainty including repeatability, calibration buffers/hydrometer, and thermometer. It calculates the standard uncertainty from each source, combines them, and determines the expanded uncertainty of 1.0887 or ±0.14 at a 95% confidence level for pH testing and 0.0086 or ±0.0086 for relative density testing.

Uploaded by

Mishra dtu
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 5

Global Test House Pvt. Ltd.

Plot No.87/2&9 , Opp. Metro Pillar No.617, Industrial Area


Rohtak Road Mundka, New Delhi- 110041

Measurement of Uncertainty
(As per NABL 141 Guideline)

Item: Admixture Date of Test: 16-26/10/2022


Test Parameter: pH Temperature: 31 ± 2°C

Source of Uncertainty:
U1 Repeatability Type A Normal Distribution Normal Distribution
U2 pH Buffer 4 Type B Uncertainty from Calibration ………...Do…………….
Certificate at 95% CL
2
U3 pH Buffer 7 Type B ………...Do……………. ………...Do…………….
U4 pH Buffer 9 Type B ………...Do……………. ………...Do…………….
U5 Thermometer Type B ………...Do……………. ………...Do…………….

Standard Deviation
S.No. I (Test Observation/Readings) (I-A) Deviation (I-A)²
√∑ (I-A)²/N-1
1 6.50 -0.098 0.009604
√0.093280/4
2 6.52 -0.078 0.006084
3 6.80 0.202 0.040804
4 6.72 0.122 0.014884
0.1527
5 6.45 -0.148 0.021904
Avg (A) 6.60 0.093280

Std. Uncertainty= Standard Deviation/√n 0.0683

Relative Uncertainty %

Std. Uncertainty at Average test value


Relative Uncertainty %
S.No. Source of Uncertainty Std.Uncertainty/Max. value of machine/equipment X Avg.
Std. Uncertainty/AverageX100
test value
1 Repeatabilty 0.0683 0.0683/6.60X100 1.0351
2 pH Buffer 4 0.005X4/4 0.0050 0.005/4X100 0.1250
3 pH Buffer 7 0.005X7/7 0.0050 0.005/7X100 0.0714
4 pH Buffer 9 0.005X9/9 0.0050 0.005/9X100 0.0556
5 Thermometer 0.1500X25/50 0.0750 0.075/25X100 0.3000

Uncertainty Budget

Probable Relative Unc.


S.No. Sources of Uncertainty Type Value (X) Divisor Std. Unc. Degree of Freedom
Distribution %
U1 Repeatabilty A Normal 0.0683 1 0.0683 1.0351 4
U2 pH Buffer 4 B Normal 0.0050 2 0.0100 0.1250 α
U3 pH Buffer 7 B Normal 0.0050 2 0.0100 0.0714 α
U4 pH Buffer 9 B Normal 0.0050 2 0.0100 0.0556 α
U5 Thermometer B Normal 0.1500 2 0.3000 0.3000 α

Combined Uncertainty = √(UI)² + (U2)² + (U3)² + (U4)² + (U5)²


Combined Uncertainty = √(1.0351)² + (0.1250)² + (0.0714)² + (0.0556)² + (0.3000)²
= 1.0887
Expanded Uncertainty = 2X1.0887 = 2.177
(At 95% Confidence level) ±2.177
K=2

Expanded Uncertainty ±2.177% @ 6.60

± Mµ = 0.14 @ 6.60

Approved by

Global Test House Pvt. Ltd. Document:GTHPLQR-24: Measurement of Uncertainty


Issue No.: 01 Issue Date: 25/08/2021 Rev. No.: 00 Rev. Date: Page No.: 01 of 01
Prepared By: QM Issued By: QM Approved By: TM Controlled Copy
Global Test House Pvt. Ltd.
Plot No.87/2&9 , Opp. Metro Pillar No.617, Industrial Area
Rohtak Road Mundka, New Delhi- 110041

Measurement of Uncertainty
(As per NABL 141 Guideline)

Item: Admixture Date of Test: 16-26/10/2022


Test Parameter: Relative Density Temperature: 31 ± 2°C

Source of Uncertainty:
U1 Repeatability Type A Normal Distribution Normal Distribution
U2 Hydrometer Type B Uncertainty from Calibration ………...Do…………….
Certificate at 95% CL
2

Standard Deviation
S.No. I (Test Observation/Readings) (I-A) Deviation (I-A)²
√∑ (I-A)²/N-1
1 1.10 -0.008 0.000064
√0.001480/4
2 1.13 0.022 0.000484
3 1.08 -0.028 0.000784
4 1.12 0.012 0.000144
0.0192
5 1.11 0.002 0.000004
Avg (A) 1.108 0.001480

Std. Uncertainty= Standard Deviation/√n 0.0086

Relative Uncertainty %

Std. Uncertainty at Average test value


Relative Uncertainty %
S.No. Source of Uncertainty Std.Uncertainty/Max. value of machine/equipment X Avg.
Std. Uncertainty/AverageX100
test value
1 Repeatabilty 0.0086 0.0086/1.108X100 0.7764
B Hydrometer 0.0006X1.166/2 0.00032 0.0032/1.166X100 0.0275

Uncertainty Budget

Probable Relative Unc.


S.No. Sources of Uncertainty Type Value (X) Divisor Std. Unc. Degree of Freedom
Distribution %
U1 Repeatabilty A Normal 0.0086 1 0.0086 0.7764 4
U2 Hydrometer B Normal 0.0006 2 0.0011 0.0275 α

Combined Uncertainty = √(UI)² + (U2)²


Combined Uncertainty = √(0.0086)² + (0.0275)²
= 0.7769
Expanded Uncertainty = 2X0.7769 = 1.554
(At 95% Confidence level) ±1.554
K=2

Expanded Uncertainty ±1.554% @ 1.108

± Mµ = 0.02 @ 1.108

Approved by

Global Test House Pvt. Ltd. Document:GTHPLQR-24: Measurement of Uncertainty


Issue No.: 01 Issue Date: 25/08/2021 Rev. No.: 00 Rev. Date: Page No.: 01 of 01
Prepared By: QM Issued By: QM Approved By: TM Controlled Copy
Global Test House Pvt. Ltd.
Plot No.87/2&9 , Opp. Metro Pillar No.617, Industrial Area
Rohtak Road Mundka, New Delhi- 110041

Measurement of Uncertainty
(As per NABL 141 Guideline)

Item: Admixture Date of Test: 16-26/10/2022


Test Parameter: Dry Material Content, % Temperature: 31 ± 2°C

Source of Uncertainty:
U1 Repeatability Type A Normal Distribution Normal Distribution
U2 Weighing Balance 200 g Type B Uncertainty from Calibration ………...Do…………….
Certificate at 95% CL
2
U3 Oven Type B ………...Do……………. ………...Do…………….

Standard Deviation
S.No. I (Test Observation/Readings), % (I-A) Deviation (I-A)²
√∑ (I-A)²/N-1
1 34.65 -0.682 0.465124
√9.067480/4
2 36.05 0.718 0.515524
3 37.60 2.268 5.143824
4 34.56 -0.772 0.595984
1.5056
5 33.80 -1.532 2.347024
Avg (A) 35.33 9.067480

Std. Uncertainty= Standard Deviation/√n 0.6733

Relative Uncertainty %

Std. Uncertainty at Average test value


Relative Uncertainty %
S.No. Source of Uncertainty Std.Uncertainty/Max. value of machine/equipment X Avg.
Std. Uncertainty/AverageX100
test value
1 Repeatabilty 0.6733 0.6733/35.33X100 1.9057
2 Weighing Balance 200 g 0.0018X170/200 0.001530 0.001530/170X100 0.0009
3 Oven 0.1500X105/250 0.0630 0.0630/105X100 0.0600

Uncertainty Budget

Probable Relative Unc.


S.No. Sources of Uncertainty Type Value (X) Divisor Std. Unc. Degree of Freedom
Distribution %
U1 Repeatabilty A Normal 0.6733 1 0.6733 1.9057 4
U2 Weighing Balance 200 g B Normal 0.0018 2 0.0036 0.0009 α
U3 Oven B Normal 0.1500 2 0.3000 0.0600 α

Combined Uncertainty = √(UI)² + (U2)² + (U3)²


Combined Uncertainty = √(1.9057)² + (0.0009)² + (0.0600)²
= 1.907
Expanded Uncertainty = 2X1.907 = 3.813
(At 95% Confidence level) ±3.813
K=2

Expanded Uncertainty ± 3.813% @ 35.33 %

± Mµ = 1.35 % @ 35.33 %

Approved by

Global Test House Pvt. Ltd. Document:GTHPLQR-24: Measurement of Uncertainty


Issue No.: 01 Issue Date: 25/08/2021 Rev. No.: 00 Rev. Date: Page No.: 01 of 01
Prepared By: QM Issued By: QM Approved By: TM Controlled Copy
Global Test House Pvt. Ltd.
Plot No.87/2&9 , Opp. Metro Pillar No.617, Industrial Area
Rohtak Road Mundka, New Delhi- 110041

Measurement of Uncertainty
(As per NABL 141 Guideline)

Item: Admixture Date of Test: 16-26/10/2022


Test Parameter: Ash Content, % Temperature: 31 ± 2°C

Source of Uncertainty:
U1 Repeatability Type A Normal Distribution Normal Distribution
U2 Weighing Balance 200 g Type B Uncertainty from Calibration ………...Do…………….
Certificate at 95% CL
2
U3 Oven Type B ………...Do……………. ………...Do…………….
U4 Muffle Furnace Type B ………...Do……………. ………...Do…………….

Standard Deviation
S.No. I (Test Observation/Readings), % (I-A) Deviation (I-A)²
√∑ (I-A)²/N-1
1 2.56 0.068 0.004624
√0.019880/4
2 2.50 0.008 0.000064
3 2.41 -0.082 0.006724
4 2.56 0.068 0.004624
0.0705
5 2.43 -0.062 0.003844
Avg (A) 2.49 0.019880

Std. Uncertainty= Standard Deviation/√n 0.0315

Relative Uncertainty %

Std. Uncertainty at Average test value


Relative Uncertainty %
S.No. Source of Uncertainty Std.Uncertainty/Max. value of machine/equipment X Avg.
Std. Uncertainty/AverageX100
test value
1 Repeatabilty 0.0315 0.3729/10.252X100 1.2652
2 Weighing Balance 200 g 0.0018X50/200 0.000450 0.000450/50X100 0.00090
3 Oven 0.1500X105/250 0.0630 0.0630/105X100 0.0600
4 Muffle Furnace 1.6X600/1200 0.8000 0.8000/600X100 0.1333

Uncertainty Budget

Probable Relative Unc.


S.No. Sources of Uncertainty Type Value (X) Divisor Std. Unc. Degree of Freedom
Distribution %
U1 Repeatabilty A Normal 0.0315 1 0.0315 1.2652 4
U2 Weighing Balance 200 g B Normal 0.0018 2 0.0036 0.00090 α
U3 Oven B Normal 0.1500 2 0.3000 0.0600 α
U4 Muffle Furnace B Normal 1.6000 2 3.2000 0.1333 α

Combined Uncertainty = √(UI)² + (U2)² + (U3)² + (U4)²


Combined Uncertainty = √(1.2652)² + (0.00090)² + (0.0600)² + (0.1333)²
= 1.274
Expanded Uncertainty = 2X1.274 = 2.547
(At 95% Confidence level) ± 2.547
K=2

Expanded Uncertainty ± 2.547% @ 2.49%

± Mµ = 0.063 % @ 2.49 %
Approved by

Global Test House Pvt. Ltd. Document:GTHPLQR-24: Measurement of Uncertainty


Issue No.: 01 Issue Date: 25/08/2021 Rev. No.: 00 Rev. Date: Page No.: 01 of 01
Prepared By: QM Issued By: QM Approved By: TM Controlled Copy
Global Test House Pvt. Ltd.
Plot No.87/2&9 , Opp. Metro Pillar No.617, Industrial Area
Rohtak Road Mundka, New Delhi- 110041

Measurement of Uncertainty
(As per NABL 141 Guideline)

Item: Admixture Date of Test: 16-26/10/2022


Test Parameter: Chloride, % Temperature: 31 ± 2°C

Source of Uncertainty:
U1 Repeatability Type A Normal Distribution Normal Distribution
U2 Weighing Balance 200 g Type B Uncertainty from Calibration ………...Do…………….
Certificate at 95% CL
2
U3 Oven Type B ………...Do……………. ………...Do…………….

Standard Deviation
S.No. I (Test Observation/Readings), % (I-A) Deviation (I-A)²
√∑ (I-A)²/N-1
1 0.017 -0.001 0.000001
√0.000014/4
2 0.017 -0.001 0.000001
3 0.016 -0.002 0.000004
4 0.020 0.002 0.000004
0.0019
5 0.020 0.002 0.000004
Avg (A) 0.018 0.000014

Std. Uncertainty= Standard Deviation/√n 0.0008

Relative Uncertainty %

Std. Uncertainty at Average test value


Relative Uncertainty %
S.No. Source of Uncertainty Std.Uncertainty/Max. value of machine/equipment X Avg.
Std. Uncertainty/AverageX100
test value
1 Repeatabilty 0.0008 0.0008/0.018X100 4.6481
2 Weighing Balance 200 g 0.0018X50/200 0.00045 0.00045/50X100 0.00090
3 Oven 0.1500X150/250 0.0900 0.0900/150X100 0.0600

Uncertainty Budget

Probable Relative Unc.


S.No. Sources of Uncertainty Type Value (X) Divisor Std. Unc. Degree of Freedom
Distribution %
U1 Repeatabilty A Normal 0.0008 1 0.0008 4.6481 4
U2 Weighing Balance 200 g B Normal 0.00180 2 0.0036 0.00090 α
U3 Oven B Normal 0.1500 2 0.3000 0.0600 α

Combined Uncertainty = √(UI)² + (U2)² + (U3)²


Combined Uncertainty = √(4.6481)² + (0.00090)² + (0.0600)²
= 4.648
Expanded Uncertainty = 2X4.648 = 9.297
(At 95% Confidence level) ±9.297
K=2

Expanded Uncertainty ± 9.297% @ 0.0018%

± Mµ = 0.002 % @ 0.018 %

Approved by

Global Test House Pvt. Ltd. Document:GTHPLQR-24: Measurement of Uncertainty


Issue No.: 01 Issue Date: 25/08/2021 Rev. No.: 00 Rev. Date: Page No.: 01 of 01
Prepared By: QM Issued By: QM Approved By: TM Controlled Copy