Stimpod NMS450X IFU XM400 21A04 11
Stimpod NMS450X IFU XM400 21A04 11
2
• Large current densities associated with failing electrodes may cause superficial • The Stimpod enclosure should not be opened under any circumstances. Opening
burns. the unit will void the warranty.
• For acceleromyography, the Stimpod is designed to be compatible with standard STIMPOD (NMS 410/450X) conforms to the following standards:
ECG electrodes, however, for high currents the use of a dedicated NMT electrode • IEC 60601-1, IEC 60601-2-10, IEC 60601-2-40
such as the Xavant XT45008 is recommended. • IEC 60601-1-2: CISPR 11 Group1 class A; IEC 61000-4-2; IEC 61000-4-3
• Electrodes that have current densities exceeding 2mA/cm2 may require special • ISO 13485, Directive 93-42-EEC
attention of the operator.
• This product must be stored at 0 – 50°C.
• This product must be transported in the carry case provided.
• This product and all the accessories have been certified latex free.
• Inspect all parts for any damage or manipulation. Never use any damaged or
manipulated part!
• If an electrically conductive surface of the Stimpod device or its cables are exposed,
such electrically conductive surface may shock a person handling it. Do not use such
a device or accessory, please contact the manufacturer for repair.
• The refractory period delay is set at a default value to prevent the user from
repeating stimulation while the nerve synapse is recovering from effects of the
previous stimulation. A refractory period of less than 12 seconds in TOF mode is not
advisable as measurements might not represent the effect of blocking agents on the
neuromuscular junction.
Application Specification:
• The patient population includes patients of all ages, weight and nationality. Patient
health and state is described in contraindications, warnings and cautions.
• The user must be a medical professional with knowledge of anatomy.
• The use environment requirements of the device such as the medical practice and
operating room is described in the guidance and manufacturers declaration.
• The device can be used on any part of the body except for limitation described in
warnings and cautions or identified in section 3 for the NMT mode.
Warranty:
• The Stimpod (device only) carries a 24 month warranty against manufacturing defects,
provided that the device was used in accordance with the operating instructions.
• The cables included in the Stimpod kit carry a 6 month warranty against
manufacturing defects, provided that the cables were used in accordance with the
operating instructions.
3
Contents
1. Getting to know the STIMPOD (NMS 410/450X) 5 3.6) Adjusting the Current 22
1.1) Device Description 5 3.7) Adjusting Stimulation Mode 22
1.2) Device Layout 6 3.8) Adjusting the Twitch/ Tetanus Frequency 22
1.3) Screen Layout 7 3.9) Train of Four Mode (TOF) 23
1.4) Accessories 8 3.10) Double Burst Mode (DB) 23
1.5) Warning Screens 10 3.11) Post Tetanic Count Mode (PTC) 24
1.6) Open Circuit Detection 12 3.12) Supra Maximal Current Mode (SMC) 24
1.7) Auto Shutdown 12 3.13) Auto Mode 25
1.8) Symbols 12 3.14) Twitch Mode 26
3.15) Tetanus Mode 26
2. Nerve Locating/Mapping Mode (NMS 410/450X) 13
2.1a) Adjusting the Current in LOC mode 14 4. Setting up Device Defaults 27
2.1b) Adjusting the Current in MAP mode 15 4.1) Setup Menu 28
2.2) Adjusting the Pulse Width 16 4.2) User Settings 28
2.3) Proximity Indicator 16 4.3) NMT Settings 29
2.4) Adjusting the Twitch Frequency 16 4.4) Locate Settings 30
4
Getting to Know the STIMPOD
(NMS410/450X)
1
1.1) Device Description
The STIMPOD NMS450X is a quantitative Neuromuscular Transmission (NMT) monitor utilizing either tri-axial accelerometry or electromyography to provide real-time quantitative feedback.
The Stimpod NMS 410 as well as Stimpod NMS 450X is also a precision nerve locating tool used for localizing specific neural pathways. Localization of nerves by electrical stimulation
involves connecting the nerve stimulator to a conducting needle through which local anaesthetics can be injected. The distance of the needle (cathode) from the nerve can be estimated
by establishing the minimum threshold current required, to facilitate a neuromuscular response.
CAUTION: This device should only be used by a qualified physician with appropriate knowledge in anaesthesia.
The sale or purchase of the device is restricted to licensed medical practitioners, as governed by the law of the country/state in which he/she practices, or where the device is
to be used.
5
1.2) Device Layout
1 Cable Connector
Insert the combined Nerve Mapping / Locating
Multi Functional Clip
Cable or the NMT cable to activate the relevant
mode. 1
en
men
3
ter
Flashing Green: Stimulus delivered.
Flashing Red: Open Circuit.
4
5 Pause Button
5
NMS 410/450X (LOC/MAP Mode)
Press to Stop / Start Stimulation. 6
NMS 450X (NMT Mode)
NMS 450X
Press and release to elicit a single stimulation.
Press and hold to activate a repeated stimulation.
7
6 The Wheel
Adjust current in the main operating mode.
Navigate the Setup Menus.
6
1.3) Screen Layout
en
men
ter
Diagnostic Screen NMS 410/450X (LOC / MAP Mode) Diagnostic Screen NMS 450X (NMT Mode)
• Indication of • Real-time quantitative feedback
current setting • Proximity indicator arrow • Countdown timer for repeat mode
7
1.4) Accessories
WARNING: Use of cables or other accessories other than those supplied with the STIMPOD may result in serious injury.
NOTE: ECG Electrodes and Nerve Locating needles are not included in this package.
CAUTION: A sterile wipe should be applied to the Nerve Mapping Probe prior to use.
8
EMG Accessories (STIMPOD NMS450X)
9
1.5) Warning Screens
Cable Not
Replace Batteries
Recognised
Open Circuit
Detected
Press Enter
Open Circuit Detected: To Confirm
This warning informs the user that the contact points of the connected accessory are not
forming a closed circuit. This warning will be accompanied by a red flashing LED, every Current Adjustment in NMT Mode:
time the unit attempts to stimulate. When attempting to adjust the current while in NMT mode the device will show this
message, requesting the user to confirm the adjustment of the current.
Refractory
Delay Active Stimulation In
Progress
Refractory Delay Active:
Once a TOF, DB, or PTC stimulation is performed, the refractory period timer is initiated. Stimulation in Progress:
During the countdown time it will not be possible to perform another stimulation. If Warning shown when the play button is pressed during stimulation.
another stimulation is attempted, this warning screen will appear.
10
EMI Warning
Incorrect
Electrode Placement
No NMT
Data received
Stimulation aborted
Stimulation aborted:
This warning is shown when:
• SMC: Measured current is more than 10% less or more than the set current.
• NMT: The stimulation was interrupted.
11
1.6) Open Circuit Detection 1.8) Symbols
The Stimpod performs impedance measurements at regular intervals to detect whether
the connection between the STIMPOD and the patient comprises a closed circuit.
Open Circuit
Detected
10%
10% 0°C0°C
1.7) Auto Shutdown Type BF Non-sterile Humidity limitation Temperature limit
Applied Part
STIMPOD will shut down after 10 minutes of no user or patient interaction.
12
Nerve Locating/Mapping
Mode (NMS 410/450X)
2
Locating mode (LOC) This mode offers the user the means to do nerve mapping and locating without having
to switch or unplug cables.
Localisation of nerves by electrical stimulation involves connecting the nerve stimulator
to a conducting, locating needle (not supplied) through which local anaesthetics can When inserting the Nerve Mapping / Locating Cable, the STIMPOD will default to the Nerve
be administered. This procedures involves subcutaneous stimulation of the motor Mapping current range (0-5mA). Current will be directed to the Nerve Mapping probe and
component of the relevant peripheral nerve, to ‘locate’ the nerve. the STIMPOD will attempt to stimulate. The moment that the needle penetrates the skin
the cable will sense it and inform the STIMPOD. The stimpod will switch to Nerve locating
• Select this mode by inserting the Nerve Locating / Mapping cable. mode. If contact between the needle and patient is broken, and the nerve mapping probe
• The STIMPOD will automatically default to the Nerve Locating current range (0.00 - touches the patient, STIMPOD will switch back to the nerve mapping mode and start
5.00 mA) and display the ‘ LOC ’ indicator. monitoring the needle - patient connection again. Whenever the nerve mapping probe
and the nerve locating needle simultaneously make contact with the patient the needle
will have first priority.
Mapping Mode (MAP)
• This mode is selected when the Nerve Mapping / Locating cable is inserted.
Percutaneous nerve mapping enables the anaesthesiologist to map out a particular
superficial nerve prior to nerve location with the needle. This is accomplished by
stimulating the motor component of the relevant peripheral nerve percutaneously with
the nerve mapping probe. This technique ensures a higher success rate for directing the
needle to the correct nerve.
13
When using the Nerve Locating / Mapping Cable
6 Proximity Indicator
Pulse Width Setting 2 7 Stimulating Frequency Setting
Adjust using Menu / Pulse width button Adjust using Enter / Hz width button
en
men
ter
2.1a) Adjusting the Current in LOC mode 1 0.6 - 2.0mA Default 0.2mA
2.0 - 5.0mA Default 0.5mA
Current Mode Options: Linear Mode, Non-Linear Mode
Default: Linear Increments can be adjusted in the Setup Menus
The Linear Mode is called linear because one ‘click’ on the Wheel will correspond with one Non-Linear Mode:
increment as set in the specific current range. The Linear Mode allows the user to selectNMS 450X
individual incrementing options for the three different current ranges. The Non-Linear Mode facilitates the non-linear nature of the current intensity versus the
distance from the nerve. This mode allows the user to define 20 adjustment positions in
Default Current range: terms of Current (mA) and Pulse Width (ms). If correctly implemented each adjustment
0.00 - 5.00mA adjustable in the following default increments: position should afford the user with a relatively linear progression in terms of the distance
0.0 - 0.6mA Default 0.1mA from needle tip to the nerve.
14
Default Current and Pulse Width Range: NOTE: Because the 20 definable positions include both current and pulse width settings,
As shown in table 1 in section 4.4 circle the Wheel to select the pulse width cannot be adjusted independently in this mode. This is indicated on the
predetermined current and pulse width positions sequentially. screen by the fact that pulse width is not highlighted.
6 Proximity Indicator
Pulse Width Setting 2 7 Stimulating Frequency Setting
Adjust using Menu / Pulse width button Adjust using Enter / Hz width button
en
men
ter
2.1b) Adjusting the Current in MAP mode 1 NOTE: The STIMPOD will automatically default to the Nerve Mapping current range
(0-20mA) and display the ‘ MAP ’ indicator.
Current Range: 0 - 20mA adjustable in 1mA increments.
Circle the Wheel to adjust current.
NMS 450X 15
2.2) Adjusting the Pulse Width 2 2.4) Adjusting the Twitch Frequency 7
Options: 0.05ms, 0.1ms, 0.3ms, 0.5ms, 1ms Options: 1Hz, 2Hz, 5Hz
Default: 0.05ms Default: 2Hz
Press Menu/Pulse Width button to toggle between different Pulse Widths. Press Enter/Hz button to toggle between different stimulating frequencies.
The proximity indicator notifies the user that the target charge range has been reached.
This function allows the user to set up an upper and lower limit of charge. When
contraction is elicited at the set charge, this indicator should indicate to the user that the
needle has reached the desired proximity to the nerve. This proximity is indicated both
visually and audibly.
Visual Indication:
• Visually indicated in the diagnostics screen by two arrowheads.
• Arrowhead indicating the lower threshold points upwards.
• Arrowhead indicating the upper threshold points downwards.
• The dotted line representing the selected current will be positioned between the two
arrowheads if the target charge range is entered.
Audible Indication:
• A successful stimulus above the proximity range will make a single beep.
• A successful stimulus within the proximity range will make a double beep.
• A successful stimulus below the proximity range will make a triple beep.
16
Neuromuscular Transmission
Monitoring (NMT) Mode (NMS 450X)
3
3.1) Introduction to NMT Monitoring 3.2) Cables and Sensors for NMT Monitoring
Monitoring Neuromuscular Blocking Agent involves stimulating a neural pathway The Stimpod NMS450X makes provision for two different sensor technologies for NMT
which facilitates the contraction of an appendage. Based on the relative strength of the Monitoring namely Acceloromyography (AMG) and Electromyography (EMG).
contraction which is the result of a stimulus of specific intensity or waveform, it is possible
to draw conclusions about the efficacy of an injected Neuromuscular Blocking Agent.
NMT Monitoring Cable (AMG)
The Stimulation Modes used in NMT are as follows: Train-of-Four, Double Burst, Post
Tetanic Count, Supra Maximal Current and Auto modes. In this case the NMT Monitoring Cable is fitted with a tri-axial accelerometer that is
attached to the contracting appendage of the patient to measure the strength of the
• These modes can be selected when a NMT Monitoring cable is inserted into the contraction resulting from the applied electrical stimulus.
Stimpod.
NMT Monitoring Cable (EMG)
In this case the NMT Monitoring Cable is fitted with an EMG electrode that is attached to
the contracting appendage of the patient to measure the action potential of the muscle
due to the applied electrical stimulus.
17
3.3) Stimulation Sites for NMT Monitoring
Electrode placement relies on the cathode (black electrode clip) to be as close to the
targeted nerve as possible in order to effectively depolarize the nerve. The anode (red
electrode clip) should be away from the targeted nerve. 1
18
3.4) AMG Patient Setup
1 Check Batteries & Switch Device On 2 Connect NMT Monitoring Cable (AMG) 3 Prepare Skin for Monitoring
Power on device
Patient Setup
4 Attach Electrodes
A
M
G
19
3.5) EMG Patient Setup
1 Check Batteries & Switch Device On 2 Connect NMT Monitoring Cable (EMG) 3 Prepare Skin for Monitoring
Power on device
Patient Setup
E
M
G
20
Refractory Period Delay
The three modes: TOF, DB and PTC are subject to refractory period delays, providing a
safety period which prevents the user from repeating stimulation while the nerve synapse
is recovering from the effects of the previous stimulation.
Refractory timer
Immediately after stimulation in one of these modes, the countdown timer is activated and Only displayed after a single
shown on the screen. If repeat mode is activated, only the repeat timer will be displayed stimulation is completed and is
on the screen because the repeat period for the repeat timer will always be larger than the removed after it reaches 00:00.
refractory period timer. If an attempt is made to stimulate while the refractory timer is active
a warning screen will be displayed reminding the user that the refractory period is active.
Default refractory period delays are as follows for the three modes:
TOF: 15 seconds
DB: 1 minute
PTC: 2 minutes
21
When using the NMT Mode (NMS 450X)
3 Warning: Notifies the user of a discrepancy
between current setting and average current
of actual stimulus.
4 If sound is enabled: STIMPOD will beep
every time that a stimulus was delivered. The
pitch of the sound will follow the current
Current Setting 1 intensity level.
Adjust using wheel 5 Average current of actual stimulus
A E
Stimulation Mode 2 6 Cable Type Indentifier: M or M
G G
en
men
ter
3.6) Adjusting the Current 1 3.8) Adjusting Twitch / Tetanus Frequency
Default Current range: 0 - 80 mA adjustable in 5mA increments.
Circle the Wheel to adjust current. N/A
When the Stimpod starts up in NMT mode, it will revert to the NMT current that was used
previously. In order to adjust the current, circle the adjustment wheel, the current setting
will flash and start to adjust accordingly. At the same time a warning screen will appear
prompting you to press ‘enter’ to confirm. Press ‘enter’ to confirm the current change Twitch Mode: Options (Stimulating Frequency): 1Hz, 2Hz, 5Hz
within 2 seconds. NMS 450X
Default: 2 Hz
Press Enter/Hz button to toggle between different stimulation frequencies.
22
3.9) Train of Four Mode (TOF) 3.10) Double Burst Mode (DB)
u u
en
en
men
men
ter
ter
The TOF stimulation comprises four, square waves with a pulse width of 200 microseconds, The DB stimulation comprises a burst of three square waves of 200microseconds pulse
500 milliseconds apart. width, 20 milliseconds apart, followed by another burst of three square waves, 750
milliseconds later.
Selecting TOF Mode:
• Ensure that an NMT Monitoring Cable (AMG/EMG) is inserted in the Stimpod. Selecting DB Mode:
• Ensure that TOF is one of the selected Active Modes in the NMT Settings menu (4.3). • Ensure that an NMT Monitoring Cable (AMG/EMG) is inserted in the Stimpod.
• Press the ‘Mode’ button until ‘TOF’ is shown
NMS 450Xon the display. • Ensure that DB is one of the selectedNMSActive
450X Modes in the NMT Settings menu (4.3).
• Press the ‘Mode’ button until ‘DB’ is shown on the display.
Real-time Patient Data Displayed:
• The relative contraction strength caused by each stimulus is indicated graphically in Real-time Patient Data Displayed:
the diagnostic screen as shown in the picture. • The relative contraction strength caused by each stimulus is indicated graphically in
• In the case that all four contractions could be measured, the percentage of measured the diagnostic screen as shown in the picture.
contraction strength of the fourth stimulus compared to the first stimulus will be • The percentage of the measured contraction strength of the second contraction
displayed in the diagnostic screen. compared to the first contraction will be displayed in the diagnostic screen.
• If less than four contractions were measurable, the number of contractions that could
be identified will be displayed, i.e. 2/4.
23
3.11) Post Tetanic Count (PTC) 3.12) Supra Maximal Current (SMC)
u u
en
en
men
men
ter
ter
The PTC stimulation comprises a tetanus stimulation followed by a delay and a number of SMC mode is used to find the optimal current for the stimulating electrodes placement.
twitches. (Default settings are as shown below) The SMC stimulation comprises of 16 1Hz twitches at increasing current intensities from
5-80mA in 5mA increments. (For Facial mode, this is limited to 8 twitches from 5-40mA).
Defaults: The device will then determine the supra maximal current and change the set current to
Tetanus: 50Hz for 5 seconds that value.
Delay: 3 seconds
Twitch: 20 twitches at 1Hz NMS 450X Important: SMC mode should only be performed
NMS 450X on a non-paralyzed patient.
24
3.13) Auto Mode
u u
en
en
men
men
ter
ter
Auto mode is used to conduct Full Case NMT Monitoring. This is achieved through a These zones are:
series of alternating TOF and PTC stimulation sequences to determine the depth-of-block • Recovered: A TOF ratio of larger than 90%
throughout the duration of the procedure. • Minimal: A TOF ratio of between 40% and 90%.
• Shallow: A TOF ratio less than 40%, but all pulses present.
Selecting Auto Mode • Moderate: A TOF count of 1 to 3.
• Ensure that an NMT Monitoring Cable (AMG/EMG) is inserted in the Stimpod. • Deep: A PTC Count of 1 or more.
• Ensure that “Auto” is one of the selectedNMS
Active
450X modes in the NMT Settings menu (4.3)
• Profound: A PTC count of 0. NMS 450X
• Press the ‘Mode’ button until ‘Auto’ is shown on the display.
Depth-of-Block Zones
The depth-of-block can be classified in terms of six identifiable zones, where each zone
indicates the extent of muscle paralysis the patient is experiencing due to the presence of
a muscle relaxant.
25
3.14) Twitch (TWI) 3.15) Tetanus (TET)
N/A N/A
u u
en
en
men
men
ter
ter
The Twitch stimulation comprises a 200 microseconds square wave pulse. If the ‘Play/ The Tetanus stimulation comprises series of 200 microseconds square wave pulses
Pause’ button is pressed the twitch will repeat at the selected frequency. repeated at a 50Hz or 100Hz repetition rate or frequency.
26
Setting up Device Defaults
4
The setup menu allows the user to customize device parameters. Access the Setup Menu
by pressing and holding the menu button. To exit the menu, the menu button may be
pressed again. A block at the top of each menu indicates which menu is currently active.
The menu is controlled using the menu button, the enter button and the wheel. Two
blocks at the bottom of the screen indicates what action will occur when a button is
pressed, the block bottom left corresponds to the menu button and the block bottom
right with the enter button.
In general, the menu button will cause the menu to go back to the previous state or
cancel the current action, while the enter button will act on the current menu item (select
/ accept / toggle). Clockwise rotation on the wheel will generally mean next (menu item /
option) or increase (selected value), while counter clockwise will generally mean previous
(menu item / option) or decrease (selected value).
27
4.1) Setup Menu Language options: English (default), French, Italian, Dutch, Spanish, Portuguese, German,
Swedish, Danish, Greek, Czech and Polish.
The setup menu contains 3 sub menus, the contents of each will be covered in detail
in the next sections. Use the wheel to select an item and the enter key to enter the Volume options: Off, Low, Medium (default) and High. During operation, an icon on the
corresponding menu. Menu button to exit. main device screen will indicate the currently selected volume.
1) User Settings – Settings regarding the user’s preferences, such as language, buzzer
volume, backlight settings and user info. Backlight options: Off, 5s (on for 5 seconds after last activity), 60s (on for 60 seconds after
2) NMT Settings – Settings regarding NMT modes such as TOF or PTC. These settings last activity) and Always On.
include refractory and repeat timers, active modes and whether the device is
currently being used with a facial electrode or not. Note: The battery life will be drastically shortened if Always On is selected.
3) Locate Settings – Settings regarding Locate mode, including proximity indicator
levels and current mode settings (linear / non-linear). User Info: The User Info menu item will open another menu page which allows the user
to enter two lines of personal information of 20 characters each. This menu is controlled
using two modes: Edit mode and Scroll mode. To change from Edit mode to Scroll mode,
press the menu button and from Scroll mode to Edit mode, press the enter button. In Scroll
mode the menu button will close the User Info menu and return to the User Settings menu.
28
Timers: The timer items are changed in a two-step process, changing the number of
minutes first and then the number of seconds. Use the wheel to navigate to the item
which should be changed, then press the enter key to start editing.
In scroll mode, the wheel can freely change the cursor position. Any character can be
step 1
selected and then changed in edit mode by pressing the enter button.
In step 1, the wheel will change the number of minutes. The enter button will accept the
value and continue to step 2, while the menu button will revert to the previous timer value
and cancel editing. In step 2, the wheel will change the number of seconds. The enter
button will accept the timer value and conclude editing. The menu button will return the
process to step 1.
step 1
Active modes: This can be used to limit the amount of NMT modes which are available
during operation to limit the amount of button presses necessary to get to the required
mode. Pressing the enter button will toggle the item between active and inactive. The
available modes include TOF, DB, PTC, TWI, TET, SMC and Auto.
step 2
29
Facial Mode: This item is toggled (On or Off ) by pressing the enter key. When on, the Current mode - Linear: The linear mode is called linear because one ‘click’ on the wheel will
resultant current determined by SMC will be limited to a 40mA maximum. correspond with one increment as set in the specific current range. In the linear mode,
Pulse Width is not affected when turning the wheel. Pulse Width is selected when the
Menu/Pulse Width button is pressed. The Linear Mode essentially allows the user to select
4.4) Locate Settings different increment options for each of the three different current ranges.
Settings which are relevant to the locating mode is managed through this menu.
It contains the proximity indicator expandable menu item, current mode item and
depending on the selected current mode either linear current mode range settings or an
item to navigate to the non-linear settings.
The increment for each range can be set by scrolling to the range to be changed, initiating
edit using the enter button and then selecting a different option using the wheel. Each
range has preset options which can be selected.
Current Mode – Non-linear: The Non-Linear Mode facilitates the non-linear nature of the
Proximity Indicator: The upper and lower limits of the proximity indicator as described current intensity versus the distance from the nerve. The required current intensity is
in 2.1 can be set by expanding the proximity indicator menu item, selecting the limit to proportional to the square of the distance from the electrode to the nerve fibre.
be changed using the wheel, then pressing enter to start editing and then changing the
value using the wheel. When the desired value is reached, the enter button will accept the This mode allows the user to define 20 adjustment positions in terms of current (mA) and
change. Alternatively, to discard the change press Cancel (menu button). pulse width (ms). If correctly implemented each adjustment position should afford the
user a relatively linear progression in terms of the distance from needle tip to the nerve.
30
Current Pulse Width Charge These values can be viewed or changed in the Non-linear Settings menu. The charge value
Position for each or the twenty positions are displayed graphically, and the current and pulse width
(mA) (ms) (µC)
value of each position can be viewed by scrolling to that position using the wheel.
1 0.3 0.1 0.03
2 0.43 0.1 0.043
3 0.58 0.1 0.058
4 0.76 0.1 0.076
5 0.97 0.1 0.097
6 1.2 0.1 0.12
7 1.4 0.1 0.14
To change the charge at a specific position, a 2 step process can be followed. Select the
8 1.7 0.1 0.17 position which should be changed using the wheel and press the enter button to start
step 1. The current is changed using the wheel in this step and the menu button will
9 2 0.1 0.2
cancel the change. When satisfied with the current, the enter button can be used to
10 2.3 0.1 0.23 continue to step 2: changing the pulse width. The pulse width for the selected position
can be changed using the wheel. The menu button will return to step 1 and the enter
11 2.7 0.1 0.27
button will accept the change and conclude editing.
12 3 0.1 0.3
13 3.4 0.1 0.34
14 3.8 0.1 0.38
15 4.3 0.1 0.43
16 4.8 0.1 0.48
step 1
17 1.8 0.3 0.54
18 2.1 0.3 0.63
19 2.4 0.3 0.72
20 2.7 0.3 0.81
Table 1 Default non-linear settings
step 2
31
45 Technical Notes
5.1) Performance Test
_._ _
Before operating and using the device a performance test must be carried out at the
site of use. The performance test described below is in compliance with the German § 5
MPBetreibV directive. ! Insert Cable
Insert Cable
32
• The LED should flash RED and no audible feedback should be heard. 5.1.2) Combined Nerve Mapping/Nerve Locating Mode
• Short-circuit the needle connector and the ECG connector.
The following screen should appear on the display. • Insert the Nerve Mapping/Nerve Locating Cable.
The following screen should appear on the display.
Open Circuit
! Detected
• The LED should flash GREEN and if sound is enabled in the menu a beep should be
heard each time a stimulus is delivered. • The LED should flash RED and no audible feedback should be heard.
• Stimulus should occur at the set frequency. (1,2 or 5 Hz). • Short-circuit the Nerve Mapping probe and the ECG connector.
• Use the adjusting wheel and slowly increase the current to 5.00mA. The following screen should appear on the display.
• Monitor that the stimulating waveform, measured and displayed in the diagnostic
window is square. The top part of the square wave should also touch the dotted line,
which represents the current setting as shown below.
• The LED should flash GREEN and if sound is enabled in the menu a beep should be
heard each time a stimulus is delivered.
• Stimulus should occur at the set frequency. (1,2 or 5 Hz).
• Use the adjusting wheel and slowly increase the current to 20mA.
• Monitor that the stimulating waveform, measured and displayed in the diagnostic
window is square. The top part of the square wave should also touch the dotted line,
which represents the current setting as shown below.
In order to test the Nerve Locating connection and device functionality follow the
instructions in 5.1.1.
33
5.1.3) NMT Monitoring Mode (AMG)
(NMS450X only)
Offener Stromkreis
Open Circuit
! • Separate the red and black electrode connectors to cause an open circuit between
erkannt
Detected
them.
• The LED should flash GREEN in accordance with the four stimulations.
• Each stimulation should be accompanied by an audible ‘beep’.
• In ‘Diagnostic window’ four bars of different heights should indicate that the
accelerometer detected movement.
• Monitor the actual current delivered to ensure that the warning sign does not appear.
34
5.1.4) NMT Monitoring Mode (EMG)
(NMS450X only)
Offener Stromkreis
Open Circuit
! • Separate the red and black electrodes to create an open circuit.
erkannt
Detected
35
5.2) Specifications
Operating Modes: Nerve Locating Mode Nerve Mapping Mode NMT Mode
NMS 410/450X NMS 410/450X NMS 450X
Current Range 0.00 - 5.00 mA ± 5% 0 - 20mA ± 5% 0 - 80mA ± 5%
Pulse Width Options 0.05ms, 0.1ms, 0.3ms, 0.5ms, 1ms ± 5% 0.05ms, 0.1ms, 0.3ms, 0.5ms, 1ms ± 5% 0.2ms ± 5%
Maximum Stimulation Voltage 100V 400V 400V
Stimulus Monophasic square wave Monophasic square wave Monophasic square wave
Stimulating Frequency 1Hz , 2Hz, 5Hz ± 5% 1Hz , 2Hz, 5Hz ± 5% 1Hz, 2Hz, 5Hz, 50Hz, 100Hz ± 5%
Load Impedance 0 kOhm - 20 kOhm 0 kOhm - 20 kOhm 0 kOhm - 5 kOhm
Technical Specifications NMS 410/450X 5.3) Cleaning and Disinfecting STIMPOD NMS 410/450
Device Classification Class IIa, Type BF
Cleaning: Soap and water, applied with a damp cloth is suitable to clean and
Power Supply 4 x AAA alkaline batteries
disinfect the STIMPOD. It is imperative that no moisture penetrates the STIMPOD.
Power Consumption 17mA Disinfecting: Any commercially available methanol - free disinfectant in an ethyl
Waveform Constant Current, Monophasic Square alcohol base can be used for disinfection.
Wave
Weight 130g
Dimensions 145mm x 90mm x 30mm
Operating Temperature 10 - 40 ° Celsius
Storage and Transport Temperature 0 - 50 ° Celsius
Operating Humidity 90% Relative Humidity
Transport and Storage Humidity 90% Relative Humidity
Operating Atmospheric Pressure 50 – 106kPa
Transport and Storage Atmospheric Pressure 50 – 106kPa
36
5.2) Guidance and Manufacturers Declaration
Guidance and manufacturers declaration – electromagnetic emissions– for all equipment and systems
The STIMPOD NMS 410/450X is intended for use in electromagnetic environment specified below. The customer or user of the STIMPOD NMS 410/450X should assure that it is used
in such an environment
RF Emissions Group 2 – Class A The STIMPOD NMS 410/450X must emit electromagnetic energy in order to perform its
CISPR 11 intended function. Nearby electronic equipment may be affected.
The STIMPOD NMS 410/450X is suitable for use in all establishments, other than
domestic establishments and may be used in domestic establishments and those
directly connected to the public low voltage power supply network that supplies
buildings used for domestic purposes, provided the following warning is heeded:
Guidance and manufacturers declaration – electromagnetic immunity- for all equipment and systems
The STIMPOD NMS 410/450X is intended for use in the electromagnetic environment specified below. The customer or the user of the STIMPOD NMS 410/450X should assure that it
is used in such an environment
Immunity Test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge (ESD) ± 6 kV contact ± 6 kV contact Floors should be wood, concrete or ceramic tile. If the floors are covered with synthetic
IEC 61000-4-2 ± 15 kV air ± 15 kV Air material, the relative humidity should be at least 30%
Power frequency (50/60 Hz) 30 A/m 50 Hz Power frequency magnetic fields should be at levels characteristic of a typical location
magnetic field 30 A/m (Effective) in a typical commercial or hospital environment.
IEC 61000-4-8
37
Guidance and manufacturer’s declaration – electromagnetic immunity
The STIMPOD NMS410/NMS450X is intended for use in the electromagnetic environment specified below. The customer or the user of the STIMPOD NMS 410/450X should assure that
it is used in such an environment.
Immunity IEC 60601 test level Compliance level Electromagnetic environment – guidance
test
Portable and mobile RF communications equipment should be used no closer to any
part of the STIMPOD NMS 410/450X, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the transmitter.
3 V/m (10V/m Home Healthcare) at 3 V/m (10V/m Home Healthcare) at d = 1,2 80 MHz to 800 MHz
Radiated RF 80-2,700MHz, AM Modulation. And 80-2,700MHz, AM Modulation. And
IEC 61000-4-3 9-28V/m at 385-6000MHz, Pulse 9-28V/m at 385-6000MHz, Pulse d = 2,3 800 MHz to 2,5 GHz
Mode and other Modulation (upon Mode and other Modulation (upon
Risk Analysis). Risk Analysis). where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer and d is the recommended separation distance in metres
(m).
38
Recommended separation distances between portable and mobile RF communications equipment and
the STIMPOD NMS410/450X
The STIMPOD NMS 410/450X is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the STIMPOD NMS
410/450X can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
STIMPOD NMS 410/450X as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum output power m
W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
Not applicable d = 1,2 d = 2,3
0,01 - 0,12 0,23
0,1 - 0,38 0,73
1 - 1,2 2,3
10 - 3,8 7,3
100 - 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Guidance and manufacturers declaration – electromagnetic immunity – for equipment and systems that are non- life supporting
The STIMPOD NMS 410/450X is intended for use in the electromagnetic environment specified below. The customer or the user of the STIMPOD NMS 410/450X should assure that it
is used in such an environment
Immunity Test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Radiated immunity 80MHz – 1GHz @ 3V/m & 10V/m 80MHz – 1GHz @ 3V/m & 10V/m Portable and mobile RF communications equipment can affect MEDICAL
80MHz - 2.5GHz 1GHz – 2.5GHz @ 10V/m 1GHz – 2.5GHz @ 10V/m ELECTRICAL EQUIPMENT and should be used no closer to any part of the
equipment, including cables, than the recommended separation distance.
39
Products & Accessories
STIMPOD NMS410 KIT: PRECISION NERVE LOCATOR Product Code XT-41011
(Including Nerve Locating/Mapping Cable, Carry Case, IFU)
STIMPOD NMS450 KIT: QUANTITATIVE NMT MONITOR AMG (1.8M) Product Code XT-45021
(Including Nerve Locating/Mapping cable, NMT Monitoring Cable AMG (1.8m), Carry Case, IFU)
STIMPOD NMS450 KIT: QUANTITATIVE NMT MONITOR EMG (1.8M) Product Code XT-45021B
(Including NMT Monitoring Cable EMG (1.8m), Carry Case, IFU)
NERVE LOCATING/MAPPING CABLE Product Code: XT-41014
NMT MONITORING CABLE AMG (1.8m) Product Code: XT-45025
NMT MONITORING CABLE AMG (3.5m) Product Code: XT-45025A
NMT MONITORING CABLE EMG (1.8m) Product Code: XT-45003
NMT MONITORING CABLE EMG (3.5m) Product Code: XT-45003A
40
41
Unit 102, The Tannery Industrial Park, 309 Derdepoort Rd
Silverton, Pretoria, South Africa, 0184
Tel: +27 (0) 12 743 5959, E-mail: [email protected]
(01)6009880396221(10)xxxxxxxxxx Web: www.xavant.com
42