0% found this document useful (0 votes)
270 views22 pages

Hetek Flow Sampler User Manual

The document is a user manual for the Hetek Flow Sampler, a portable instrument for measuring gas leak rates. It quantifies leaks by sampling at high flow rates and measuring natural gas concentration to calculate leak rates. The instrument is intrinsically safe for use in hazardous locations and is operated inside a padded backpack for portability. The manual provides specifications, operating instructions, maintenance procedures, and safety information for proper use of the instrument.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
270 views22 pages

Hetek Flow Sampler User Manual

The document is a user manual for the Hetek Flow Sampler, a portable instrument for measuring gas leak rates. It quantifies leaks by sampling at high flow rates and measuring natural gas concentration to calculate leak rates. The instrument is intrinsically safe for use in hazardous locations and is operated inside a padded backpack for portability. The manual provides specifications, operating instructions, maintenance procedures, and safety information for proper use of the instrument.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

HETEK FLOW SAMPLER

User Manual

Version 1.7 – July 2023


Document #6.11

Our Solutions. Your Safety.


Hetek Flow Sampler
User Manual

Revision History
Rev # Date Revised By Detail
1.1 May 6th, 2022 Ashwin Mohan New SPP
1.2 Aug 4th, 2022 Ashwin Mohan Update of Specification - Blower Flow Rate
1.3 Dec 23rd, 2022 Ashwin Mohan Update of Sections 1, 3, 4, 10, 12, 13 to include
updated imagery, warning messages, and
specifications in accordance with certification
1.4 Jan 19th, 2023 Ashwin Mohan Addition of grounding for electrostatic hazards
1.5 Mar 20th, 2023 Ashwin Mohan Certification nomenclature added
1.6 April 24th, 2023 Ashwin Mohan Best Practices for Battery and Screen Messages
1.7 July 25th, 2023 Ashwin Mohan ATEX Certification nomenclature added

Contents
WARRANTY ............................................................................................................................................. 2
1. Introduction .................................................................................................................................... 3
2. Standard Materials .......................................................................................................................... 5
2.1 Optional Items ......................................................................................................................... 5
3. Attachments .................................................................................................................................... 6
4. Specifications................................................................................................................................... 8
Table 1. Technical Datasheet ............................................................................................................... 8
5. Grounding ....................................................................................................................................... 9
6. Operation ...................................................................................................................................... 10
7. Measurement Process ................................................................................................................... 14
8. Purge Mode ................................................................................................................................... 15
9. Access Records .............................................................................................................................. 16
10. Upload Data – Instrument to Computer ..................................................................................... 16
11. Verification Test ......................................................................................................................... 17
12. Calibration ................................................................................................................................. 17
13. Battery Use ................................................................................................................................ 19
14. Errors and Troubleshooting........................................................................................................ 20
Table 2 – Error Codes......................................................................................................................... 20

15. Manufacturing and Green Policy ......................................................................................... 21

16. Authorized Service Centres ................................................................................................. 21

1
Hetek Flow Sampler
User Manual

WARRANTY

Hetek Solutions Inc. (“Hetek”) warrants to the buyer that at the time of delivery, this product will be free
from defects in material and manufacture and will conform substantially to Hetek’s applicable
specifications. Hetek’s liability and buyer’s remedy under this warranty are limited to repair or
replacement, at Hetek’s option, of this Product or parts thereof returned to the seller at the factory of
manufacture and shown to Hetek’s reasonable satisfaction to have been defective, provided that
written notice of the defect shall have been given by buyer to Hetek Solutions Inc. within one (1) year
after the date of delivery of this product.

Hetek warrants to buyer that it will convey good title to this product. Hetek’s liability and Buyer’s
remedy under this warranty of title are limited to the removal of any title defects or, at the election of
Hetek, to the replacement of this product or parts thereof that are defective in title. Hetek or its
authorized service centre will be responsible for all the warranty work and service of the instrument.

THE FOREGOING WARRANTIES ARE EXCLUSIVE AND ARE GIVEN AND ACCEPTED IN LIEU OF (I) ANY AND
ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE: AND (II) ANY
OBLIGATION, LIABILITY, RIGHT, CLAIM OR REMEDY IN CONTRACT OR TORT, WHETHER OR NOT ARISING
FROM HETEK’S NEGLIGENCE, ACTUAL OR IMPLIED. The remedies of the Buyer shall be limited to those
provided herein to the exclusion of any and all other remedies including, without limitation incidental or
consequential damages. No agreement varying or extending the foregoing warranties, remedies or this
limitation will be binding upon Hetek. unless in writing, signed by a duly authorized officer of Hetek.

NOTICE
Product improvements and enhancements are continuous; therefore, the specifications and information
contained in this document may change without notice. Hetek Solutions Inc. shall not be liable for errors
contained herein or for incidental or consequential damages in connection with the furnishing,
performance, or use of this material. No part of this document may be photocopied, reproduced, or
translated to another language without the prior written consent of Hetek Solutions Inc.

www.hetek.com

2
Hetek Flow Sampler
User Manual

1. Introduction
The Hetek Flow Sampler is a field based, portable, battery-powered instrument designed to determine
the volumetric flow rate of gas leaks or fugitive emissions around various above grade natural gas
infrastructure such as pipe fittings, valve packings, and compressor seals. Settings include facilities such
as natural gas storage, compressor stations, pressure regulation stations, and industrial facilities.
Produced by Hetek Solutions Inc. (“Hetek”), the Hetek Flow Sampler is certified for intrinsic safety for
use in hazardous locations according to applicable regulatory standards across North America (see
Section 4, Table 1 for more details).

1.1 Specified Use


The Hetek Flow Sampler (also referred in this document as the “instrument” or “HFS”) is a gas leak
quantification device. Typically used in facilities containing natural gas infrastructure, it quantifies the
leak by sampling at high flow rate(s) to capture all the gas leaking from the component along with
surrounding air. By measuring the flow rate of the sampling stream and the natural gas concentration
within that sample stream, the gas leak rate is calculated using Equation 1 and outputted as parameter –
QLEAK on the instrument display.

The Hetek Flow Sampler is meant to be used after a gas leak detection survey has been conducted using
approved methods for the purposes of pinpointing the source of the leak such as using a portable gas
detector or optical gas imaging (OGI) camera. Once the leaking components have been identified, the
user can measure the rates of those leaks using the Hetek Flow Sampler with knowledge of the exact
leaking component, thus being able to accurately capture the leak using appropriate accessories.

The instrument is packaged and operated while it is inside a padded backpack, enabling the operator to
carry it especially when climbing ladders or entering a confined space. The handheld display allows the
user to perform the instrument’s main functions which is connected via a 6-foot coiled cable.

� WARNING: The Hetek Flow Sampler is not to be used in any application that is beyond its intended
purpose or beyond the scope of its specifications. Failure to adhere to this warning can result in
personal injury, damage to the equipment, or reduced instrument performance.

� WARNING: The Hetek Flow Sampler is not a life safety device meant for the personal protection of
the user or others. A multi-gas personal safety monitor is to be used in conjuncture with the Hetek Flow
Sampler to warn the user of hazardous gases that could be present in the area.

� WARNING: Use only the following battery pack.


• Hetek battery pack – Part Number 1840-1003
• Hetek approved battery packs are supplied with an identifying label
• Replacement or additional battery packs supplied by Hetek

� WARNING: Use of unauthorized parts will void certification and warranty of the device.

3
Hetek Flow Sampler
User Manual

𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄𝑄 𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿 = 𝐹𝐹𝐹𝐹𝐹𝐹𝐹𝐹 𝑅𝑅𝑅𝑅𝑅𝑅𝑅𝑅 × �𝐺𝐺𝐺𝐺𝐺𝐺𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠𝑠 − 𝐺𝐺𝐺𝐺𝐺𝐺 𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵𝐵 � × 10−2 (𝐸𝐸𝐸𝐸. 1)

where:
Quantified Leak = rate of gas leak from source (cfm or lpm)
Flow Rate = blower flow rate (cfm or lpm)
Gassample = concentration of gas from leak source (%)
Gasbackground = background gas concentration (%)

𝑄𝑄2𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿 − 𝑄𝑄1𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿
1 − 2 (%) = � � × 100 (𝐸𝐸𝐸𝐸. 2)
𝑄𝑄2𝐿𝐿𝐿𝐿𝐿𝐿𝐿𝐿

Equation 2 is used to calculate the difference in the QLEAK values after a 2-stage test. It served as an
indicator of the quality of sample that was collected with differing blower flow rates. The Q1 – Q2 (%)
value will be displayed on the results screen at the end of each 2-staged test and outputted in the data
log file.

1.2 Actual and Standard Quantities


The final quantification output of the leak will be an actual quantity which is affected by field conditions.
A standard quantity considers temperature and pressure conditions. If a standard quantity is desired,
then Equation 3 will be required to convert the final QLEAK value in relation to standard temperature and
pressure. This calculation to be tabulated separately by the user.
𝑃𝑃𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴 528˚𝑅𝑅
𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 = 𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴 ∗ � �∗� � (𝐸𝐸𝐸𝐸. 3)
14.7𝑝𝑝𝑝𝑝𝑝𝑝 𝑇𝑇𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴𝐴
where:

SCFM = standard cubic feet per minute


ACFM = actual cubic feet per minute – QLEAK
PACTUAL = barometric pressure (psi)
TACTUAL = temperature in Rankine (°R)

TACTUAL and ACFM are both outputs from the Hetek Flow Sampler, on the display after a measurement is
complete and on the DATA LOG file on a computer when logging is enabled.
PACTUAL can be obtained from local weather station (station pressure).

Please Note: Equation 2 uses standard temperature at 528°R (20°C or 68°F). Variations exist.

1.3 Other Conversions


Various fugitive emissions management programs require reporting in other units or parameters. The
following formulas aim to achieve that, and their descriptions are as follows: Equation 3 is the
conversion from imperial to metric units, Equation 4 yields the annual metric volume, and Equation 5 is
the conversion between the temperature units Rankine and Celsius.

4
Hetek Flow Sampler
User Manual

𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 = 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 × (0.028317 𝑠𝑠𝑠𝑠𝑠𝑠/𝑠𝑠𝑠𝑠𝑠𝑠) (𝐸𝐸𝐸𝐸. 4)

𝑆𝑆𝑆𝑆𝑆𝑆 60𝑚𝑚𝑚𝑚𝑚𝑚 8766ℎ𝑟𝑟


= 𝑆𝑆𝑆𝑆𝑆𝑆𝑆𝑆 ∗ � �∗� � (𝐸𝐸𝐸𝐸. 5)
𝑦𝑦𝑦𝑦𝑦𝑦𝑦𝑦 ℎ𝑟𝑟 𝑦𝑦𝑦𝑦
Where:

SCM = standard cubic metre


SCMM = standard cubic metre per minute

𝑇𝑇°𝑅𝑅 = (𝑇𝑇℃ + 273.15) × 1.8 (𝐸𝐸𝐸𝐸. 6)

2. Standard Materials
The following materials are included in every Hetek Flow Sampler unit (P/N 1840-0000):

• Main Unit
• Quick Guide
• Backpack and Accessory Bag (1840-0058)
• Batteries – Qty. 2 (1840-1003)
• Charger (1840-0100)
• Crevice Tool (see Figure 1) (1840-1082)
• Cone Collection Tool (see Figure 2) (1840-0061)
• Sample Collection Bag (see Figure 3) (1840-0059)
• 6-foot hose (1840-2004)
• 12-foot hose (1840-2005)
• Flange Tool (see Figure 4) (1840-0060)
• Communication USB cable (1840-1067)
• Grounding Clamp and Cable (see Figure 5) (1840-0064)
• Firmware Upgrade Kit (1840-1112, 1840-1113, 1840-1114)

2.1 Optional Items


The following optional items are available and can be ordered separately (Hetek part #):

• Cotton Filters for Sensors (Box of 10) (6100-6119)


• Inlet Filter (1840-1001)
• Calibration Kit: Regulators, Case, Gases (Available in North America only) (1840-0013)
• Batteries (1840-1003)
• Charger (1840-0100)

5
Hetek Flow Sampler
User Manual

3. Attachments
The Hetek Flow Sampler comes with four (4) attachments that enables the user to capture leaks from
various fittings, locations, and positions. The appropriate attachment must be selected by the user and
attached to the end of the instrument’s hose assembly (either the 6-foot or 12-foot hose can be used),
of which the hose’s other end is attached to the inlet, located at the top of the instrument.

� WARNING: Attachments can have the effect of concentrating the leak. Ensure that enough air is
present at first when sampling to support catalytic mode of the sensor. Flooding the sensor with high
concentrations of methane in the beginning can result in erroneously low readings and prevent
transition to thermal conductivity mode.

3.1 Crevice Tool (Part #1840-1082)


Used when the leak source is confined to a narrow space.

Figure 1. Crevice Tool

3.2 Cone Collection Tool (Part # 1840-0061)


A tool meant to be used over small fittings and vents.

Figure 2. Cone Collection Tool

6
Hetek Flow Sampler
User Manual

3.3 Sample Collection Bag (#1840-0059)


A durable collection bag that is versatile enough to be used over various fittings, over components that
contain multiple leak sources in a small area or that is hard to pinpoint or access. Once the sample
collection bag is over the leaking component, use the drawstring to close the end of the bag.

Figure 3. Sample Collection Bag

�WARNING: Do not close the off the bag completely, allow air to flow into the bag to replace the
volume of gas and air being removed because of the instrument’s intake during the test.

3.4 Flange Tool (Part #1840-0060)


Wrap the flange tool over a flange and ensure that the hose is connected to the flange tool at the top of
the flange to best capture the leak as natural gas is lighter than air.

Figure 4. Flange Tool

Tip: At high leakage rates or when the ability to capture all fugitive emissions from a leaking source is
unknown, it is recommended to use an optical gas imaging (OGI) camera to view the measurement
being conducted. This ensures that all gas has been captured by an attachment used with the HFS.

7
Hetek Flow Sampler
User Manual

4. Specifications
Table 1. Technical Datasheet
Specification Detail
Product Hetek Flow Sampler
Manufacturer Hetek Solutions Inc.
2085 Piper Lane, London, ON N5V 3S5 Canada
Display Information • Date and Time ● Blower flow rate
• Battery Level ● Difference between 2 stages
• Sample Concentration ● Leak Temperature
• Background Concentration ● Quantified Leak Rate
Display Screen Size 3-inch LCD Display
User Controls 4 Pushbutton keys: ESCAPE; ↑ (up arrow); ↓ (down arrow); ENTER
Data Output Data log file and Calibration log file (.CSV format)
Memory 800 records for Data; 1000 records for Calibration
Communication USB cable from Hetek Flow Sampler to a digital device
Measured Values • Blower Flow Rate ● Sample Gas Concentration
• Battery Level ● Background Concentration
Calculated Values • Quantified Leak (Stages 1 and 2)
• Difference between stage 1 and stage 2 (Automatic and Manual)
Humidity 0 – 95% relative humidity (non-condensing)
Operating Temperature - 20°C to 40°C (-4°F to 104°F)
Storage Temperature - 20°C to 45°C (-4°F to 113°F); - 40°C to 90°C (without battery)
Blower Flow Rate Maximum ≈ up to 9.0 CFM (255 LPM) at full battery charge
Stage 2 is 70 – 80% of the flow rate of Stage 1
Sensor Catalytic Oxidation Mode: 0 to 5% by volume methane
Thermal Conductivity Mode: 5 to 100% by volume methane
Leak Measurement 1.5 – 140 LPM (0.052 – 5.0 CFM) : Automatic 2-Stage & Manual 2-Stage
0.5 – 140 LPM (0.017 – 5.0 CFM) : Manual 1-Stage
Battery Standard Quantity: Two (2) with every Hetek Flow Sampler
4.8 V Nickel-Cadmium Rechargeable Battery
Run Time: 5 hours (per battery); Recharge Time: 12 hours
Electrical Parameters Um = 5.36 V; Pi = 15 W; Li = 50 μH, Ci = 100 μF
Weight Enclosure with Display: 19.90 lbs (9.05 kgs)
Dimensions Enclosure: 12” L x 16” H x 7.5” D (30.48 cm L x 40.64 cm H x 19.05 cm D)
Flow Measurement Differential pressure across orifice
Accuracy Sensor: ± 5%
Flow Rate: ±5%
Calculated Leak Rate: ±10%
Certified against the CSA C22.2 No.60079-11 (2nd edition), CSA C22.2 No. 60079-0 (4th edition)
following standards and UL 60079-0 (7th edition), UL 60079-11 (6th edition), UL 913 (8th edition)
classification ratings for IS Class I, Division 1, Group D, T3
intrinsic safety in Class I, Zone 0 AEx ia IIA T3 Ga
hazardous locations Ex ia IIA T3 Ga
across North America 2900 II 1G Ex ia IIA T3 Ga
and Europe

8
Hetek Flow Sampler
User Manual

5. Grounding
Static discharges can be controlled by proper static bonding and grounding wherever flammable gases
and dusts are present. The goal of static bonding and grounding is to safely remove built-up static
electricity without having a static discharge event. This is achieved by attaching grounding clamps, that
are bonded to a terminal point on the Hetek Flow Sampler via a spiral cable and lug, to a known and
verified ground point in the field such as a grounding rod or cable.

Please Note: In order to prevent a spark while grounding, it is recommended to first ground the
equipment prior to switching the Hetek Flow Sampler ON. When not in use, the grounding clamp should
be stored inside the backpack.

5.1 Grounding Clamp and Cable Assembly (Part #1840-0064)


�WARNING: The enclosure must be protected to avoid ignition hazard due to impact or friction.
• The device must be protected from high electrostatic charge environments and processes.
• In cases where electrostatic conditions are unknown, the device must be protected with a
dissipative guard.
• To reduce the possibility of a static discharge, the Hetek Flow Sampler must be grounded
while conducing a leak measurement.

Before conducting a
measurement, attach the
ground clamp to the
nearest earth ground.

Grounding Rod

Figure 5. Hetek Flow Sampler with Grounding Clamp and Cable Assembly

�WARNING: The ground clamp may experience wear over time such as dulling of its contact points
and weakening of its spring, which can lose its overall functional effectiveness over time. In such an
event, it is recommended to replace it with Hetek Part #1840-0064, as this component is specifically
designed for grounding with its strong spring and sharp contact point to penetrate any coating and/or
rust to reach the underlying metal as well as to attach itself firmly to the grounding material.

9
Hetek Flow Sampler
User Manual

6. Operation
� WARNING: The instrument must be turned ON in a clean atmosphere, or upwind from a hazardous
area. The battery must be connected and disconnected in a non-hazardous area.

The instrument is switched on using the ON/OFF toggle switch. This initiates the start up sequence. The
start up sequence involves:

• Hetek Logo
• Product Name and Firmware version
• Calibration Due Reminder
• Purging of sensors

6.1 Display
A 3-inch, LCD screen will be the main user interface for the instrument’s functionality navigation. The
screen comes with a pair of magnets on the back, which is meant to assist the user in performing
measurements hands-free, by placing the display on a ferromagnetic material such as a pipe or ladder.

Figure 6. User Interface

The instrument is operated using the 4 buttons present on the user interface, display module. The
buttons and their functionalities are as follows:

• ESCAPE: exits out of the menu or function


• ↑: Increasing/next selection
• ↓: Decreasing/next selection
• ENTER: selects the option highlighted

Please Note: The ↓↑ keys will advance the menu selection in increments in the same direction. The
cursor will appear on one line of the screen and always highlighted in dark pixels.

6.2 Verification Test


After purging during start-up, the user should perform a verification test once before daily operational
use by selecting the option VERIFY present in the CALIBRATION menu to validate the output of the
sensors against a known sample of certified gas. It is recommended to verify using methane or natural
gas with a balance of air from a certified cylinder of 2.5% volume of gas. Air is required because the
presence of oxygen is required for the sensor to operate in the catalytic mode (0 – 5% Vol. gas in air).

10
Hetek Flow Sampler
User Manual

When the instrument has not been calibrated for over 30 days, it will prompt a warning for “Calibration
Overdue” as seen in Figure 6. The user will be required to perform a calibration to remove this warning.
Additional information regarding calibration will be in Section 8: Calibration.

Figure 7. Calibration Overdue Message

The user can select the ESCAPE key to exit out of the calibration menu onto the main menu.

6.3 Main Menu


The main menu will showcase the Background (%), Sample (%), Leak (⁰C or ⁰F), Blower Rate, Status Line,
and the START option at the bottom of the display that is highlighted with a dark background. Upon
selection, using the ENTER key this for the START option will proceed to begin a measurement test.

Figure 8. Selector on START

When selecting the ↑ or ↓ arrow keys at the main menu, the SETTINGS option is displayed. Upon
selection using the ENTER key this for the SETTINGS option the sub-menu has the following options:

Figure 9. Selector on SETTINGS

1. Flow Units: offers the ability to change units of measurement and output
a. Cubic Feet per minute (CFM)
b. Liters per minute (LPM)

11
Hetek Flow Sampler
User Manual

2. Temp. Units: ability to change units of temperature


a. Celsius °C
b. Rankine °R
c. Fahrenheit °F

3. Calibration: verification testing, calibration, and deletion of records


a. Verify
b. Calibrate Sensor
c. Erase Previous Calibration Log
d. Erase Entire Calibration Log

4. Data Upload: the user can select either data or calibration file to be uploaded at a time
a. Data Log
b. Calibration Log

5. Purge: a manual purge function to clear out all gas from the instrument
a. Start/Stop: displays the background gas and sample gas percentages

6. Erase Data Log: data log can be deleted by user

7. Set Time/Date: adjust date/time in the location of the measurement occurrence

Figure 10. Setting the date/time


8. Access Records: enables the user to access records in the field for a limited number of tests (6).
a. Calibration Records
b. Leak Test Records

6.4 Measurement Test Overview and Modes


Typically, two measurements are performed at different flow rates (automatic 2-stage and manual 2-
stage modes) to ensure that the instrument is capturing all the gas that is escaping from the component.
The first measurement is taken at the highest possible flow rate, followed by a second measurement at
a flow rate that is approximately 70–80% of the first. If the two calculated leak rates are within 10% of
each other, then it can be assumed that all gas has been captured during the test.

12
Hetek Flow Sampler
User Manual

The gas sample is drawn into the instrument through a flexible hose with various attachments that are
to be connected to the end of the hose based on the attachment’s ability to encapsulate the leaking
component.

Please Note: The background gas sample line (clear tubing that is attached to the sampling hose end by
the inlet of the instrument) must be connected to the BACKGROUND port on the instrument when
performing the tests.

The main unit consists of a blower that draws in air from and around the component being tested
through a flexible hose and into the instrument. The sample is first passed through an orifice restrictor
where the measured pressure differential is used to calculate the sample’s actual flow rate. Next, a
portion of the sample is drawn in and directed to a combustible sensor that measures the sample’s
methane concentration. A second identical combustible sensor channel measures the background
methane level within the vicinity of the leaking component. The final element in the sampling system is
a blower that exhausts the gas sample back into the atmosphere away from the sampling area. The
measured flow rate and the measured methane levels (both sample and background levels) are used to
calculate the leak rate of the component being tested, with all measured and calculated values being
displayed on the hand-held display unit.

Figure 11. Start Test screen


6.4.1 Automatic 2-Stage
In the Automatic 2-Stage mode, the unit performs a leak rate measurement upon the user selecting
START, upon which the measurement begins at first at a high flow rate for 1 minute, and then
automatically shifts to a lower flow rate (70 – 80% of max. rate) and performs another measurement for
another 1 minute. This mode is used to measure leaks that are easily identifiable and captured.

6.4.2 Manual 2 Stage


In the Manual 2-Stage mode, the user starts a measurement at a high flow rate, waits for a duration
(usually longer than a minute) to attain a stable leak rate reading, and then manually switches to the
lower flow rate by pressing ENTER to save. The unit then lowers the flow rate, at which time the user
makes a second leak rate measurement that can be stopped at the discretion of the user by pressing
ENTER to save & stop. This mode is used when additional control of the measurement process is
required.

6.4.3 Manual 1-Stage


In the Manual 1-Stage mode, the user has the ability to adjust the flow rate as required using the UP ↑
and DOWN ↓ arrow keys. This mode is used when additional control of the measurement process is
required such as measuring leaks at lower leak flow rates.

13
Hetek Flow Sampler
User Manual

7. Measurement Process
When exiting the setting menu, the user views the main screen to begin the measurement process. The
user can toggle the options here between START and SETTINGS using the ↑ or ↓ keys. The user
initiates the measurement by pressing the ENTER key when START is displayed.

Figure 12. Data Logging Nomenclature

The user will be prompted with a message to LOG DATA?, upon which the user can toggle the options of
YES or NO using the ↓ or ↑ arrow keys whether the instrument should record data that will eventually
be outputted in an Excel spreadsheet. When selecting NO, the user must manually record the data when
the results of the test are outputted on the display at the end of the measurement. If digital record of
the data is desired, then select the option YES, which will prompt the user to input:
ASSET ID : XXXXXXXX
LOCATION ID : XXXXXXXX
Where ASSET ID requires the user to input the nomenclature of the asset which is being measured on
and the LOCATION ID is the name of the facility in which the series of measurements are taking place.
The ASSET ID will be displayed on the screen as well as on the digital output, the data log spreadsheet.

7.1 Start Test


The start test screen will appear after inputting the ASSET ID and LOCATION ID or when selecting NO at
the LOG DATA? screen. There will be three options displayed for the user to select according to their
preference of measurement modes as mentioned from sections 5.4.1 – 5.4.3.

1. Automatic 2 Stage
2. Manual 2 Stage
3. Manual 1 Stage

After selection one of the above options, the option highlighted will be STOP and this must be changed
using the ↑ or ↓ keys for the option START to appear. Press the ENTER key to begin the test.

1. Automatic 2 Stage
The automatic 2 stage will begin with Stage 1 which is the high blower flow rate measurement
that will last for 1 minute, followed by Stage 2 which is a lower blower flow rate (70-80% of the
first stage) that will last for 1 minute. The results will be displayed at the end.

2. Manual 2 Stage
In cases where the user desires more control of the measurement over 2 stages, the Manual 2
Stage Mode is an available option. In this mode of operation, the user determines the time
spent to gather a measurement of the emission for both stages. Upon satisfaction at the first
stage the user must press the ENTER button to begin Stage 2.

14
Hetek Flow Sampler
User Manual

3. Manual 1 Stage
The Manual 1 Stage is used when a single stage measurement is desired with the ability to
adjust the blower flow rate. The ↑ or ↓ keys can be used to select with a HIGH or LOW option.
Then the user must select and enter START to begin a test as the default is the STOP option.

The blower flow rate can be adjusted using the ↑ or ↓ keys that will change with an increment
of 5% in comparison to full scale.

7.2 Result
Upon conclusion of the test, the higher value between Q1LEAK and Q2LEAK rate should be noted at the leak
quantification flow rate of that measurement. Typically, this is Q2LEAK as the second flow rate takes in
lesser air than the faster flow rate. The user must also ensure that the 1 – 2 (%) is within 10% to ensure
the accuracy of the test. The parameters as seen in Figure 12 that are displayed on the screen are:
Background Gas (%), Sample Gas (%), Q1LEAK, Q2LEAK, Temperature (°), Blower flow rate, and 1 – 2 (%).

Figure 13. Sample Final Screen.

Please Note: The blower flow rate is with respect to the rate and mode at which the test concluded. It
does not represent the average rate of the blower during the test.

8. Purge Mode
The purpose of this mode to manually purge the instrument in case the purge during start up is not
adequate. This step will need to be performed when there is excessive gas in the area. The user can view
the readings on the display until both sensors reach zero. It is recommended to perform this purge in a
clean atmosphere as indicated by one’s personal multi-gas monitor.

Figure 14. Purge Screen

15
Hetek Flow Sampler
User Manual

9. Access Records
Data and calibration records that have been stored can be accessed on the display by selecting “ACCESS
RECORDS” and selecting the appropriate record to view – either leak test or calibration records. The
past six (6) records will be displayed on the screen, with the remaining records viewable on the
spreadsheet once uploaded from the instrument to a computer.

Figure 15. Access Records

10. Upload Data – Instrument to Computer


Connect the USB cable with its USB-B connector into the DATA PORT located on the right side of the
instrument and the USB-A connector into the computer. Turn ON the instrument using the ON/OFF
switch. Follow these steps on the computer:

1. A window should appear on the screen stating “You need to format the disk in drive F: before
you can use it. Do you want to format it? Please select Format Disk.
2. Select “Start” under Format USB Drive (F:)
3. A warning window will appear that will state “Formatting will erase ALL data on this disk. To
format the disk, click OK, To quit, click, CANCEL.
4. A window will appear stating “Format Complete”. Click OK. Click CLOSE.
5. The window should appear with either the data_log file or cali_log file. Only one file can be
uploaded at a time, if a different file is desired then the user must access the menu of the
instrument and select the option desired under DATA UPLOAD. Then unplug the USB and plug it
back in and the desired file will appear in a window.

Figure 16. Data Upload of the Calibration Log

16
Hetek Flow Sampler
User Manual

Figure 17. Sample Data Log File for Manual 1-Stage

11. Verification Test


A verification test can be performed with the Hetek Flow Sampler using the VERIFY function present in
the sub-menu under CALIBRATION. See Section 5.2 for more information.

Figure 18. Calibration Sub-Menu

The user can apply gas using a demand flow regulator to the SAMPLE and BACKGROUND ports after
selecting the VERIFY option. The user may press the ENTER or ESCAPE key upon satisfaction.

Figure 19. Verify Function

12. Calibration
A calibration must be performed every 30 days or after verification testing when it is discovered that the
reading is beyond ±10% of the span or actual gas input. A calibration ensures the optimal performance
of the instrument and its effectiveness in the field. The gases that are applied to the instrument’s leak
and background sensors are 2.5% of methane gas with a balance of air and 100% volume methane.

17
Hetek Flow Sampler
User Manual

Please Note: Do not use a gas cylinder containing 2.5% volume of methane or natural gas with the
balance of nitrogen as the sensor in catalytic mode relies on oxygen to function. Balance of air is desired.
It is recommended to use a demand flow regulator with clear tubing to deliver gas to the instrument.

Gas should be applied to the ports on the top of the instrument: SAMPLE (relates to the leak sensor) and
BACKGROUND (related to background sensor), only one calibration to each port must be performed at a
time. A two-point calibration will be performed for the two sensors:

• Background sensor at 2.5%


• Background sensor at 100%
• Leak sensor at 2.5%
• Leak sensor at 100%

Figure 20. Calibration Gas-Sensor Selection


1. Before switching ON the instrument, ensure that the instrument is not connected to the
calibration gases via either the SAMPLE or BACKGROUND ports, instrument should be switched
on in clean air.
2. Turn ON the instrument and wait until the warmup sequences are completed.
3. Select Calibrate Sensor
4. Select the gas sensor to be calibrated
5. Using the ↑↓ keys, select the sensor to be calibrated along with its gas level. These keys can
also be used to adjust the span or the applied gas concentration if the cylinder is of a different
concentration than the default 2.5% and 100%.
6. Apply the calibration gas by connecting the tubing from the appropriate SAMPLE or
BACKGROUND ports to the demand flow regulator that is connected to the gas cylinder. Begin
the calibration process and the sensor pumps will draw the gas into the instrument.
7. Wait until the reading stabilizes.
8. Select ENTER to calibrate the sensor to the appropriate level. The status line should indicate
“Calibration Pass” upon a successful calibration.
9. Disconnect the clear tubing and wait until the measured gas reading falls to zero.
10. Press ESCAPE to exit out to the sensor calibration menu.
11. Repeat this procedure to calibrate in other gas levels and sensors.
12. If the status indicates “Calibration Fail” then the calibration must be performed again, or the
sensor will need to be sent to an authorized service centre for service.

The calibration record will be saved and digitally outputted as seen in Figure 20 with a PASS or FAIL.

18
Hetek Flow Sampler
User Manual

Figure 21. A typical calibrated record that is uploaded from the instrument.

Please Note: If a calibration has been incorrectly performed, then the calibration record can be
eliminated with the “Erase Previous Calibration Log” option.

Figure 22. Erase of Previous Calibration Log

13. Battery Use


�WARNING: Do not charge or disconnect the battery pack in a hazardous atmosphere. Store in a cool
and dry area. Do not allow water to enter the battery housing and never submerge battery in water.

�WARNING: Only use the Nickel-Cadmium battery (Hetek P/N: 1840-1003) and its associated
standard battery charger (Hetek P/N: 1840-0100) supplied with the Hetek Flow Sampler. Alternates or
misuse may result in bodily injury or death.

Please Note: It is a best practice to charge the battery from a low charge to full (indicated by the green
light on the charger) and disconnect thereafter, for maximum battery life. The charger does not switch
to trickle rate mode; thus, avoid leaving the battery in the charger longer than necessary.

�WARNING: When changing batteries, use the rotating switch to turn the connected battery OFF
first, prior to disconnecting it from the Hetek Flow Sampler. Similarly, the replacement battery must be
in the OFF position when connecting it to the Hetek Flow Sampler.

The battery must be plugged into the 3-prong plug that is suspended from the instrument. Once
plugged, the battery has a rotating switch as additional safety feature to turn it ON or OFF. It is advised
to turn it OFF when the Hetek Flow Sampler is not in use.

• The battery is made of Nickel-Cadmium that can provide up to 500 charge/discharge cycles.
• Infrequently used battery packs should be fully charged, initially, then recharged periodically
once the battery gets depleted to a low charge with time. This helps maintain battery integrity.
• Allowing a battery to self-discharge during extended storage will not harm the battery. Batteries
subjected to prolonged storage (6 months or more) may lose their ability to hold a full charge.

19
Hetek Flow Sampler
User Manual

• Battery capacity can be evaluated by operating the blower for four hours and checking hourly
that the airflow is maintained at 6 CFM or greater. Several charge/discharge cycles may restore
battery capacity.
• Only use the standard battery charger provided with the Hetek Flow Sampler. Connect the
charger into a regulated 120v/60Hz outlet and then insert charging lead into the battery pack.
• The LED light on the charger will turn on, indicating that the battery is connected.

14. Errors and Troubleshooting


The following errors may appear over the course of using the instrument. The error code will be
outputted in the spreadsheet, and the error message will be outputted on the display.

Figure 23. Status message on display with “Sensor Inlet Fault”

The list of their associated causes and solutions are given, however, if these errors persist then contact
Hetek Solutions or its authorized repair centres for technical support.

Table 2 – Errors and Messages


Code Error Message Cause Potential Solution
E Sample Error When the percent difference between Perform an additional test or in
Stage#1 and Stage#2 is greater than 10% manual mode
P Sensor Inlet Fault Blockage along the path to the sensor Turn off the instrument. Ensure
port filters are clean, removal of any
physical obstacles. Turn on the
instrument.
H High Background The background gas level is greater than Repeat measurement after
Gas % the sample gas level purging in a clean atmosphere
F Sample Fault Blockage along the flow path to the inlet Removal of any physical
sample port obstacles
Z Check Sensor Sensor did not properly zero upon start- Purge in a clean area
Zero up
C Calibration Due Duration of calibration is past 30 days Perform a calibration
L Leak Sensor Error Sensor does not read 0 after repetitive Send to an authorized service
manual purging in a clean atmosphere centre
Low Battery Battery is less than 20% and all data must Charge battery and replace with
be saved prior to shutting down a fully charged one
Memory Full No space remaining Download and then delete data
log records on HFS

20
Hetek Flow Sampler
User Manual

Calibration Pass Calibration Successful


Calibration Fail Calibration Unsuccessful Retry calibration or send to an
authorized service centre

Please Note: When encountering incorrect readings such as 0 CFM as the blower flow rate appears to
be operating correctly, please cycle power by switching the HFS OFF, then turning it ON again.

15. Manufacturing and Green Policy


The Hetek Flow Sampler is designed, manufactured, and assembled in North America at Hetek Solutions’
manufacturing facility based in London, Ontario, Canada. The product supports the following:

• Environmental stewardship, combatting climate change, as it will be used to accurately measure


emissions of a major greenhouse gas that will be reported for reduction or elimination by repair.
• Energy efficiency with a longer battery duration and life cycle. Minimal accessory and packing
materials used thus reducing its carbon footprint by reducing its shipping weight and size.
• Low cost of ownership with high quality, durable and lightweight materials used when possible.
• Strengthening local economies by supporting production teams consisting of professional
engineers, technologists, certified technicians, apprenticeships, and training opportunities.
Electronic components designed and produced by OES Inc. based in London, Ontario.
• Promotion of fair wages, labour, and employment practices.

16. Authorized Service Centres


It is recommended to send the Hetek Flow Sampler to an authorized service centre annually for service.
Contact Hetek Solutions Inc. for an authorized service centre located outside of Canada.

CANADA

HETEK SOLUTIONS INC.

HEAD OFFICE WESTERN CANADA ATLANTIC CANADA


2085 Piper Lane 2 Floor – 8725 53rd Avenue
nd
145 McNamara Drive
London, Ontario Edmonton, Alberta Paradise, Newfoundland
N5V 3S5 T6E 5E9 A1L 0A7

Website: www.hetek.com
Toll Free: 1 (855) 298 4473
Phone: (519) 659 1144
Email: customerservice@hetek.com

Fill out an RMA form for instrument service: http://www.hetek.com/repair-form/

Produced in Canada.

21

You might also like