NISTIR 6969 - Good Laboratory Practice For The Quality Assurance of The Measurement Process (March 2003)
NISTIR 6969 - Good Laboratory Practice For The Quality Assurance of The Measurement Process (March 2003)
GLP 1
The validity of tests and calibrations should be monitored with quality control procedures.
Statistical techniques are used to record and monitor charted measurement results to permit the
detection of trends. The metrologists and laboratory management should plan and review the
results from quality assurance monitoring.
Other steps taken to ensure the quality of the measurement process may include, but are not
limited to:
• the regular use of Standard Reference Materials (SRMs) and /or internal quality
control using secondary reference materials;
• participation in interlaboratory comparisons (round robins);
• test replications with same and/or different methods;
• recalibration of retained items;
• correlation of different characteristics of an item; and
• proper calibration intervals.
Each measurement parameter in the laboratory’s scope of activities should be reviewed and
analyzed to determine the validity of the measurement process.
The standards and the measurement process for each parameter must be in a state of statistical
control. Statistical control means that the variability of the measurement process is known and
stable; when a process is in statistical control, we can assume that the reported measurement
uncertainties are valid. The National Institute of Standards and Technology provides technical
guidance and support to the State weights and measures laboratories to develop measurement
control programs that provide measurement assurance. The objective of these programs is to
evaluate the entire measurement process including:
• procedures;
• standards;
• equipment;
• personnel; and
• environment.
GLP 1 Page 1 of 2
March 2003
While other quality assurance programs could meet these objectives, the control programs
developed for measurement assurance greatly increase the comprehensiveness of the program.
The State laboratories typically provide measurement services in the disciplines of mass,
volume, and length. Some laboratories provide services in other measurement areas.
Approximately 89 % of their workload is in mass standards calibration.1 Mass calibration
demands the greatest precision and is the first discipline for which a measurement assurance
program was developed and implemented. Nevertheless, all measurement disciplines should
have a measurement assurance system in place.
The most recent improvement in assuring the quality of each measurement parameter in the State
Laboratories is the incorporation of a Process Measurement Assurance Program (PMAP).2
Measurement results that are collected over several years may be statistically evaluated with
current results being compared to results from previous years. Any observed problems or
changes in the measurement results are investigated and if necessary, corrective action can be
taken. Ongoing monitoring establishes a continuous, comprehensive, internal measurement
assurance program in a State laboratory.
Data from internal measurement assurance programs is compared to the results of inter-
laboratory comparisons (round robins) in which the laboratory participates as part of the
Regional Measurement Assurance Program (RMAP) groups. (These groups consist of the
metrologists of the State laboratories and their counterparts in private industry in a geographic
region of the country. Six regional measurement assurance groups operate in the United States
and its territories. )
The strength of the measurement assurance approach lies in its applicability to a wide variety of
measurements with sufficient flexibility to permit each measurement control program to be
tailored to the particular needs of a given measurement area. The sophistication of the control
program depends on the criticality of the measurement.
1
State Laboratory Program Workload Survey, 1999, NIST/OWM & NCWM Metrology Subcommittee
2
See NISTIR 6176 and SOP 30.
GLP 1 Page 2 of 2
December 2003
GLP 4
A number of States have provisions in their weights and measures laws that require the periodic
submission of their State standards to NIST for calibration. Those provisions are based on an
early version of the Model Law (1911), which was considered appropriate for the circumstances
that prevailed prior to the establishment of the New State Standards Program by NIST. Periodic
calibration is necessary on a regular, periodic basis, and also when measurement control results
from internal control charts or external round robins indicate questionable data.
Standards of mass, volume, and length, fabricated from modern materials, kept in the controlled
environment of a State metrology laboratory under the custody of trained metrologists, are
generally stable and not subject to change. Moreover, the cooperative NIST-State audit
programs often identify changes in ample time for corrective action in the unlikely event that
such a change should occur. These same programs provide the necessary evidence of the ability
to provide traceable measurement results at a level of confidence sufficient for the need.
Moreover, the process of packing, shipping, and unpacking exposes the standards to unnecessary
hazards that could result in damage, compromising their integrity. The return and re-calibration
could take several months causing an unavailability of State services that would be disruptive to
the performance of the mission of the State laboratories.
In order to develop a policy for the guidance of and implementation by all 50 States regarding
this subject, the following actions are recommended:
1. All States should recognize the fact that periodic return of their State
standards to NIST for re-calibration is essential to comply with ISO/IEC
17025, GMP 11, and when: 1) data regarding traceability is unavailable;
2) charted measurement results indicate that the standards are out of
control; 3) measurement results on round robins or performance tests are
suspect; or 4) NIST advises the State of the need.
"He (the director) shall maintain the State standards in good order and
shall submit them, at least once in ten years, to the National Institute of
Standards and Technology for calibration,"
GLP 4 Page 1 of 2
December 2003
The approach described above is recommended by NIST because each State that participates in
the Weights and Measures Division Laboratory Recognition Program has the capability of
evaluating its own State standards with the necessary documentation referencing the national
standards. The Laboratory Recognition Program provides interaction between the State standards
laboratories and NIST, ensuring satisfactory laboratory conditions suitable for primary standards
in addition to the proper use of NIST procedures in standards calibration. Thus, each State
program is evaluated and, if found in compliance, is Recognized as being capable of performing
the measurements listed on the Certificate of Measurement Traceability.
GLP 4 Page 2 of 2
March 2003
GLP 9
A calibration is not complete until the expanded uncertainty associated with the calibration is
determined and reported. Each Standard Operating Procedure (SOP) includes information
regarding the calculation of uncertainties. The expanded uncertainty is generally reported with
approximately a 95 % confidence interval. The confidence interval is determined by multiplying
a coverage factor (usually 2) times the root sum square of properly combined Type A and Type
B evaluated components according to the ISO/IEC Guide to the Expression of Uncertainty in
Measurement (GUM).
Zeros, which follow a decimal point, when there are only zeros ahead of the decimal point, are
not considered significant figures.
Rounding Rules
Use the following rules to round measurement data, consistent with its significance:
1. When the digit next beyond the one to be retained is less than five, keep the
retained figure unchanged. For example: 2.541 becomes 2.5 to two significant
figures.
2. When the digit next beyond the one to be retained is greater than five, increase the
retained figure by one. For example: 2.453 becomes 2.5 to two significant
figures.
3. When the digit next beyond the one to be retained is exactly five, and the retained
digit is even, leave it unchanged; conversely if the digit is odd, increase the
retained figure by one (even/odd rounding). Thus, 3.450 becomes 3.4 but 3.550
becomes 3.6 to two significant figures.
GLP 9 Page 1 of 3
March 2003
4. When two or more figures are to the right of the last figure to be retained,
consider them as a group in rounding decisions. Thus, in 2.4(501), the group
(501) is considered to be greater than 5 while for 2.5(499), (499) is considered to
be less than 5.
Several examples to illustrate the proper method of reporting corrections and uncertainties
follow.
Example 1
Suppose the correction for a weight is computed to be 1.3578 mg and the uncertainty is 0.5775
mg. First, round the uncertainty to two significant figures, that is, 0.58 mg. Then state the
correction as 1.36 mg. Notice that the uncertainty and the correction express the same number of
decimal places. Report the correction as 1.36 mg ± 0.58 mg.
Example 2
The volume of a given flask is computed to be 2000.714431 mL and the uncertainty is 0.084024
mL. First, round the uncertainty to two significant figures, that is, 0.084 mL. (Do not count the
first zero after the decimal point.) Round the calculated volume to the same number of decimal
places as the uncertainty statement, that is, 2000.714 mL. Report the volume as 2000.714 mL ±
0.084 mL.
Example 3
The correction for a weight is computed to be 4.3415 mg and the uncertainty is 2.0478 mg.
First, round the uncertainty to two significant figures, that is, 2.0 mg. (Notice that two
significant figures are shown. The zero is a significant figure since it follows a non-zero
number.) Then, round the correction to the same number of decimal places as the uncertainty
statement, that is, 4.3 mg. Report the correction as 4.3 mg ± 2.0 mg.
Example 4
The correction for a weight is computed to be 285.41 mg and the uncertainty is 102.98 mg.
Because this uncertainty is a large number, we first convert both values to the next larger
commonly reported unit (i.e., 0.28541 g and 0.10298 g respectively). First, round the
uncertainty to 0.10 g. (The first nonzero digit (1) is the first significant figure and the remaining
digits are rounded to the nearest number following the first nonzero digit.) Then, round the
correction to the point where the rounding occurred in the uncertainty statement. Round the
correction to 0.29 g. Report the correction as 0.29 g ± 0.10 g.
Example 5
The correction for a weight is computed to be 285.41 mg and the uncertainty is 33.4875 mg.
First, round the uncertainty to two significant figures, that is 33 mg. Then, round the correction
GLP 9 Page 2 of 3
March 2003
to the same number of decimal places as the uncertainty statement, that is, 285 mg. Report the
correction as 285 mg ± 33 mg.
Example 6
Rather than stating the uncertainty value with each correction, it is also proper to place the
correction values in a column headed by the words "Correction" or "Error," etc., and place the
uncertainties (without plus or minus signs) in a column headed "Expanded Uncertainty".
GLP 9 Page 3 of 3
March 2003
GLP 11
Large cast iron or steel weights should be painted both for their protection and to preserve their
mass integrity. Unprotected weights are subject to corrosion. Furthermore, when corrosion is
present, the extent and any changes resulting from it may be difficult to estimate.
Thin even coats of aluminum paint are recommended for this purpose. Spray applications are
best if large weights or a number of small weights are to be painted. In preparation for painting,
a weight should be cleaned and loose scale should be removed using a wire brush (DO NOT
remove old paint and corrosion by means of sand blasting or pressure washing. Severe damage
may result.). The painting should be done before the weights are calibrated or tolerance tested,
unless arrangements are made to obtain "as found" values. If “as found” values are desired, two
tests may be required.
Painting should be done in an area reserved for this purpose, or at least in a place which is
removed from laboratory measurement operations. Painting operations must be in compliance
with applicable safety standards. The weights should be protected from dust or dirt
contamination while the coating is drying.
GLP 11 Page 1 of 1
March 2003
GMP 1
The damping of the beam oscillations of an equal-arm balance is generally very slow.
Consequently, it is practical to read the turning points (the highest and lowest value of the beam
oscillation) on the graduated scale and use the sum of the turning points as the observation value
rather than wait for the beam to come to rest. This GMP requires that the graduated scale be
numbered such that adding weights to the left arm increases the readings.
Suppose the graduated scale has 20 graduations that are numbered from 0 to 20. The loads on
the balance arms should be adjusted so that the sum of the turning points is approximately twice
the midscale reading. In this example, the sum of the turning points should be within one
division of twenty. Turning points should be estimated to at least 0.1 division in this example,
which is typical of the Russell balance. This means that the final rest point is approximately 10,
the midscale reading. Motion should be induced to the beam so that the turning points can be
read easily. Care should be taken so that the beam does not hit limiting stops during its normal
oscillation while turning points are being read. The amount of the beam oscillation is not critical
although a span of from three to ten divisions is adequate.
Once motion has been induced for the beam oscillation, wait for at least one complete oscillation
cycle for the beam motion to stabilize. After this time, the turning points can be read. The
readings may begin with either the high or low turning point. The turning points for at least two
consecutive oscillation cycles should be recorded. The turning points should reveal a consistent
pattern of slow damping; that is, the turning points should gradually converge to the eventual rest
point. For example, if the last high turning point was greater than the previous high turning
point (assuming that the readings normally drop on successive readings), this would indicate that
something has interfered with beam oscillation, hence the last reading was not valid with respect
to the previous readings. Under these circumstances, turning points should continue to be read
until a consistent damping pattern has been obtained. In some cases, the balance oscillation will
dampen so slowly that the same readings may be obtained for several oscillations before a
decrease is observed. These readings are valid and may be used to compute the sum of the
turning points.
When at least four satisfactory turning points have been obtained (two high and two low turning
points), all but the last three readings should be discarded. This will leave two high and one low
turning point or vice versa. The two readings for the high or low turning points, as the case may
be, should be averaged and added to the single turning point to obtain the sum of the turning
points. The sum should be carried out to two decimal places if the second decimal place digit is
nonzero. As an example, the following readings were obtained: 15.5, 4.3, 15.4, 4.4. The sum of
4.3 + 4.4
the turning points is computed as + 15.4 = 19.75 .
2
GMP 1 Page 1 of 1
March 2003
GMP 10
Good laboratory techniques and practices, as well as good standards and balances, are required
to obtain good mass measurements. Errors can arise from a variety of sources, such as balance
drift, air currents, lack of thermal equilibrium, magnetic or electrostatic fields, and to
manipulative skills using the balance. The magnitude and significance of these errors depends
upon the laboratory environment, the balance, and the accuracy required for the measurement.
Different methods of weighing are used to eliminate or reduce the effects of sources of error.
Particular sources of error and ways to eliminate the errors are discussed with the weighing
procedures. The accuracy required for a measurement and the criticality of the measurement
often influence the choice of the weighing method used in the measurement. Regardless of the
procedure used, several practices should be followed to promote good measurements.
First, the balance should be installed in a laboratory having good temperature and humidity
control. The requirements for temperature stability increase as more precision is needed in the
measurement. For high precision measurement, the temperature should not vary by more than
± 1 °C in a 24 h period throughout the year and should vary by less than 0.5 °C during the course
of any one measurement series (typically less than 1 h). General laboratory measurements can
probably withstand temperature variations of ± 2 °C per day. Varying temperatures result in
balance drift and in unequal temperatures for the air, balance, and objects to be weighed. The
relative humidity should be maintained between 40 % and 60 %. If the relative humidity falls
significantly below 40 %, electrostatic discharge may be generated both in and near the balance.
This can result in erratic readings and make precision weighing impossible. If precision mass
standards are cleaned, they should stabilize in the laboratory environment for seven to ten days.
Thermal and environmental equilibrium are critical for mass calibrations. This is why each mass
SOP will provide instruction that the environment must be stable and the weights set in or near
the balance for 24 hours prior to a calibration. The test objects, balances, reference standards,
and air should all be in thermal equilibrium. When possible, the objects to be weighed and the
reference standards should be placed either in the balance chamber or adjacent to the balance so
all can come to the same thermal equilibrium. If there is a lack of thermal equilibrium,
convection currents will be generated when an object is on the balance pan and an erroneous
reading may result. These types of errors are likely to go unnoticed when check standards are
handled the same way, so care must be taken to reduce the likelihood of their occurrence. Tests
have shown that these types of errors depend upon the temperature differences among the objects
and the balance, and on the geometry of the objects being weighed. On 20 g standards of greatly
differing geometry, studies have shown that errors as large as 0.2 mg can occur for a 4 °C
temperature difference among the standards and the balance. Errors as large as 3 mg have also
been observed at 1 kg loads when standards were stored in a cabinet, and unknown test items
GMP 10 Page 1 of 6
March 2003
near the balance where a temperature gradient of 2 °C was present (when the uncertainty was
less than 0.1 mg).
The balance must be installed in an area free from air currents. Balances should be installed
away from heating/cooling vents. It is not adequate to merely close vents when weighing
because disrupting the temperature control system may result in temperature changes near the
balances.
Handling Weights
The masses of standard weights or objects weighed can be affected significantly by the manner
in which they are handled. Human contact can leave grease or oily films that affect the mass at
the time of measurement and can even cause permanent damage due to corrosion.
Small weights should never be touched by hand, but handled using forceps, clean gloves, or
swatches of cloth. In the latter case, the cloth must be lint free. Instructions for cleaning weights
and for removing adhering foreign material are described in GMP 5.
Large weights of lower tolerance classes (NIST Class F) may be handled by bare hands. Large
weights are a source of special problems. Fork lifts, portable cranes, hoists, or even overhead
cranes may have to be used to move the weights within the laboratory and on or off the balances.
Laboratory personnel must become expert in using such equipment, as necessary, to avoid
damage to the laboratory facilities, to the balances used, and even to the weights themselves.
Special hoist/crane hydraulics or multi-speed systems are available to gently set large weights in
place on large comparators to avoid damage. The problem of temperature equilibrium for large
weights is qualitatively the same as for small weights and needs consideration on an individual
basis.
Large weights must be clean at the time of use, but cleaning may be a problem. Minimally, they
should be inspected to ensure that foreign material is not present. Cleaning is addressed further
in GMP 5.
Operation
Analytical balances are high precision instruments and should be operated carefully. Excessive
shock can damage a balance. Avoid shock loading the balance. Gently place all weights in the
center of the weighing pan/platform. The dials on mechanical balances should be turned slowly
and gently. Careful balance operation will improve the repeatability of measurements.
Mechanical analytical balances are provided with partial and full release positions. The partial
release position is used when determining if an unknown load will provide an on-scale reading.
The balance beam has limited movement in this position. The partial release position provides
some protection to the balance when the dialed-in weights are not close to the actual mass placed
on the pan. It is good practice to arrest the pan each time a dial is being changed to protect the
balance from shock loading. It is acceptable to change the dial representing the smallest built-in
GMP 10 Page 2 of 6
March 2003
weights when in the partial release position because the small weight changes should not result
in shock loading of the balance.
When releasing the pan to either the full or partial release position, the action should be done
slowly and carefully. The objective is to minimize disturbances to the balance as much as
possible.
Similarly, all loads should be placed on the balance pan carefully and centered on the pan.
When a mechanical balance is released, the beam goes through a series of oscillations. The
frequency of these oscillations diminishes as time passes until they are almost imperceptible to
the naked eye. At this point, optimal stabilization is achieved. This stabilization of the balance
typically lasts for a very short period of time, after which the balance reading will usually drift.
A similar situation occurs when a mass is placed on the pan of an electronic balance. Therefore
readings should be taken at the precise moment of achieving balance stability. This interval
between the releasing of a pan on a mechanical balance, or the placing of a mass on an electronic
balance, and the reading of the observation, varies from balance to balance. Stabilization time
differs for all balances, even among those of the same type and model. Knowledge of the
instrument is critical to correctly establish this time interval. Although manufacturers will
usually state this value, it is necessary for the metrologist to verify its reliability. Many
electronic balances have a stability signal incorporated into the design, but this also must be
verified. All measurements in a calibration should be performed at the same time interval, and
within the shortest time possible.
All balances should be exercised before taking readings. A load equal to the load to be measured
should be placed on the balance, a reading taken, and the pan arrested, if appropriate, or the
weight removed from electronic balances. This operation should be repeated several times
before readings are taken for measurement purposes. Once the balance has been "warmed-up",
better repeatability will be achieved. Balances can be very accurate even when used without
being exercised first, but improved results can be obtained by going through a "warm-up"
procedure. If the larger variation present in initial weighings on a balance that has not been
exercised is not significant to the measurement, the warm-up procedure may be minimized.
To determine the repeatability of measurements when a balance has not been exercised versus its
performance after being exercised, and to determine how much warm-up time is required, begin
measurements starting with the unexercised balance and record the readings. Repeat a series of
measurements until you have obtained several measurements after the balance has been
exercised. This test can be repeated over several days using the same objects and the results
compared. The readings obtained when using an unexercised balance are likely to show a
slightly larger variation than those obtained after the balance has been exercised. Balance drift is
likely to be larger initially and then reach a steady state when the balance has been "warmed-up".
GMP 10 Page 3 of 6
March 2003
For mass calibrations, the unknown object must be compared to a known mass standard.
Comparison weighing eliminates the errors of the built-in weights, reduces disturbances during
the measurement because dial settings are not changed during the measurement, and can cancel
the effect of drift by selecting the appropriate weighing design. Comparing the unknown, X, to a
standard, S, eliminates the built-in weights from the measurement: thus, the built-in weights act
only as counterweights; they do not affect the difference measured between X and S.
Consequently, the dial settings must not be changed during a comparison measurement;
otherwise the built-in weights would be part of the measurement.
When comparison measurements are made on a single pan mechanical balance, all readings are
taken from the optical scale. The unknown and the standard must have nearly the same mass so
that the difference between them can be measured on the optical scale. If the masses of the
unknown and the standard are significantly different, small mass standards are used as tare
weights with either the unknown or the reference standard or both to obtain an observed mass
difference that is significantly less than one-fourth the range of the optical scale.
The first readings for the standard and the unknown in a comparison on a single pan balance
should fall in the first quarter of the optical scale, but well ahead of zero, so the balance drift will
not result in negative values for any readings. Although negative numbers may be used in
calculations, they are avoided to simplify calculations and reduce calculation errors. Because the
sensitivity weight may have a mass as large as one-half the range of the optical scale and the
measured difference between the standard and the unknown may be as large as one-fourth the
range of the optical scale, it is necessary to obtain the first two readings in the first quarter of the
optical scale so all readings will remain on-scale for the measurement. In this way it is not
necessary to change the dial settings to measure the difference between the standard and the
unknown.
GMP 10 Page 4 of 6
March 2003
Measurements made on a full electronic balance are simplified because there are no built-in
weights to consider. Although many electronic balances are equipped with a built-in calibration
weight, the weight is not involved in the comparison weighing.
The principles for comparison weighing on a full electronic balance are the same as when using
a single pan mechanical balance. The balance indications are used to measure the mass
difference between the standard and the unknown, and a sensitivity weight is used to establish
the mass value for a digital division on the balance. Since there are no built-in weights in the full
electronic balance, the entire range of the digital indications can be considered for "optical scale"
of the balance.
For comparison weighing the standard and the unknown should be "nearly the same mass."
Since a full electronic balance has a much larger range for indicating mass values, the masses do
not have to be as close together as when a mechanical balance is being used. When using an
electronic balance, the difference in mass between the standard and unknown should be less than
0.05 % of the balance capacity. Tare weights that are known standards should be used if the
masses are not sufficiently close together. The sensitivity weight used to determine the mass per
digital division should have a mass that is at least 10 to 20 times the mass difference between the
standard and the unknown but not exceeding 1 % of the balance capacity. For high precision
weighing, air buoyancy corrections must be made for all objects used in the intercomparison.
Care must be taken when weighing magnets or objects having electrostatic charges. A magnetic
field will likely affect results of a mass measurement. A magnet is attracted to ferromagnetic
materials in the balance and balance chamber. The magnetic field may also affect the magnetic
field generated in an electronic balance that utilizes the principle of magnetic force restoration as
its method of measurement.
Weights made of ferromagnetic material can become magnetized during manufacture or during
use if they are placed in a strong magnetic field. Magnetized weights can result in measurement
errors that may go undetected. If a measurement problem is suspected, the weights should be
checked for magnetism and may have to be rejected if excessively magnetized.
Electrostatic fields can also cause errors in measurements. If there is a static charge in a
mechanical balance, the balance may give erratic readings and lack repeatability. If the object
GMP 10 Page 5 of 6
March 2003
being weighed has a significant electrostatic charge, it may result in measurement errors and
may leave a static charge on the balance. Electrostatic charges are of particular concern when
plastic containers are placed on the balance.
Care should be taken to remove electrostatic charges from objects being weighed by grounding
the objects, if necessary, before placing them on the balance. To prevent the build-up of static
electricity in a balance, the relative humidity in the laboratory should be maintained between
40 % and 60 %. The water vapor in the air will serve to drain off electrostatic charges from the
balance.
Balances utilizing the magnetic force restoration principle for weighing should be checked to
verify that the magnetic field generated by the magnetic cell in the balance does not exist around
the balance pan. If the shielding of the magnetic cell is inadequate, measurement errors may
occur when weighing ferromagnetic objects or when the balance is placed on a surface
comprised of ferromagnetic material.
GMP 10 Page 6 of 6
March 2003
GMP 11
1. Introduction
1.1. Purpose
Measurement processes are dynamic systems and often deteriorate with time or use.
The design of a calibration system would be incomplete without some established
means of determining how often to calibrate instruments and standards. A
calibration performed only once establishes a one-time reference of uncertainty.
Recalibration detects uncertainty growth and serves to reset values while keeping a
bound on the limits of errors. A properly selected interval assures that an item will
receive recalibration at the proper time. Proper calibration intervals allow
specified confidence intervals to be selected and they support measurement
traceability. The following practice establishes calibration intervals for standards
and instrumentation used in measurement processes.
1.2. Prerequisites
1.3. Safety
2. Methodology
2.1. Summary
2.2. Apparatus
None.
GMP 11 Page 1 of 6
March 2003
2.3. Procedure
Critical Parameters
Components that contribute more than 25 % of a measurement’s uncertainty are
identified as critical parameters. To ensure an accurate evaluation of
performance, calibration intervals are determined to meet a 99 % reliability
target. Critical parameters are checked and defined below:
Balance Performance
Mass Standards
Mass standards are dynamic with use. Wear, contamination and other factors
can cause drift from accepted values. Thus, the following intervals have been
set:
GMP 11 Page 2 of 6
March 2003
Length Critical Parameters
Length standards are dynamic with use. Wear, contamination and other factors
can cause drift from accepted values. The following intervals have been set due
to these factors:
Volume standards are dynamic with use. Wear, contamination and other factors
can cause drift from accepted values. Calibration intervals are as follows:
GMP 11 Page 3 of 6
March 2003
Table 3. Calibration intervals for volume standards
Item Initial Cal Interval (months) Source
P100 gal standard 60 NIST
P25 gal standard * 60 Lab
P5 gal standard * 12 Lab
Glassware-
Autopipetes 5 L to 100 mL 120 Lab
*Gravimetric calibration for volumes 5 gallon or smaller, and all “slicker plate”
standards. Laboratory must be qualified for performing gravimetric calibrations.
Volume transfer is acceptable above 5 gallon.
*May be a “slicker plate” type. None are hand-held, “dump” style, test measures.
Initial Intervals
Temperature standards are dynamic with use. Shock, contamination and other
factors can cause drift from accepted values. Recalibration intervals are as
follows:
Secondary Parameters
3. Calculations
If no control charts are available, the laboratory’s Technical Manager will assign
the initial interval based on engineering evidence, manufacturer′s specifications,
NIST recommendations, and experience.
GMP 11 Page 5 of 6
March 2003
3.2. Adjustment of Intervals
3.2.2. The intervals will be adjusted by taking the following factors into consideration
as appropriate:
• calibration history;
• measurement assurance data;
• data for the population of similar equipment;
• NIST recommendations;
• statistical analysis methods; and
• manufacturer′s recommendations.
4. Assignment of Uncertainty
The uncertainty associated with the establishment of calibration intervals is not included as a
part of the uncertainties associated with the respective measurement SOPs. See SOP 29 for
calculating uncertainties for standards and the instruments.
GMP 11 Page 6 of 6
March 2003
GMP 12
Good laboratory practices, use of proper standards and equipment, and selection of standard
operating procedures are essential for providing calibration results with accurate and traceable
values with appropriate and suitable uncertainties. The following matrix recommends SOPs based
on the parameter, type of test items, and level of uncertainty needed.
Mass
Railroad test cars SOP 27, Railroad Test Cars using a Master
Track Scale
Weight carts SOP 33, Calibration of Weight Carts
(References SOP 4 and SOP 7)
HB 105-1, Class F Class F SOP 8, Modified Substitution - may be used if
ASTM, OIML Class 5, 6, 7 expanded uncertainty is less than 1/3 of the
Class M1, M2, M3 tolerance
e.g., 10 kg to 250 kg SOP 7, Single Substitution - to be used, as a
(≥500 lb) cast iron minimum, if conditions given for SOP 8 cannot
1 mg to 5 kg be met
(1 µlb to 10 lb) NOTE: Balances and standards must be
stainless steel selected properly for these conditions to be met.
GMP 12 Page 2 of 4
March 2003
Length
Tapes SOP 11, Tape to Tape OR
SOP 12, Bench Method (lower uncertainties)
Rigid Rules SOP 10, Rigid Rule
Pi Tapes SOP 23, Pi Tape Calibration
Liquid-in-Glass Thermometers
HB 105-6 Field standards for SOP 25
weights and measures
Timing Devices
HB 105-5 Field standards for SOP 24
weights and measures
Traffic Speed Gun Tuning Forks
For highway official SOP 22
use
Measurement Assurance
All Process Use of process measurement assurance
Measurement SOP 30 programs
Assurance
Use of check standards in procedure
Mass SOP 5, 28
TN 952, TN 844, NISTIR 5672
Mass SOP 4, 6, 7, 8 SOP 9
Length SOP 10, 11, 12, 23 Redundancy built into procedures
SOP 13, 14, 15, 16, 18, SOP 17, laboratory check standards OR
Volume
19, 21, 26 SOP 20, range charts
Temperature SOP 25 Use of check standards in procedure
Uncertainty
All parameters All SOPs SOP 29, Calculation of Uncertainty
GMP 12 Page 3 of 4
March 2003
GMP 12 Page 4 of 4
March 2003
GMP 13
1. Introduction
1.1 Purpose
Test numbers issued by NIST should not be used nor required as proof of the
adequacy or traceability of a test or measurement. Having a NIST number does not
provide evidence that the measurement value provided by another organization is
traceable. (See section 1.3.3.)
1.2 Prerequisites
1.2.2 Measurement uncertainty. The measurement uncertainty for each step in the
traceability chain must be calculated according to defined methods and must
be stated so that an overall uncertainty for the whole chain may be calculated
(see SOP 29);
1.2.4 Competence. The laboratories or bodies performing one or more steps in the
chain must supply evidence of technical competence (e.g., by maintaining
appropriate training records, participating in interlaboratory comparisons,
and by demonstrating that they are accredited by a recognized accreditation
body);
1.3 Responsibility
1.3.1 Provider. Providing support for the claim of traceability of the result of a
measurement or value of a standard is the responsibility of the calibration
provider. Calibration reports and certificates must contain a traceability
statement.
1.3.3 Use of, or reference to, official test numbers of a national metrology institute.
Having an authentic test number does not provide assurance or evidence that
the measurement value provided by another organization is traceable. Not
only must there be an unbroken chain of comparisons, but each measurement
must be accompanied by a statement of uncertainty associated with the value.
Test report numbers should not be used nor required as proof of the
adequacy or traceability of a test or measurement. National and international
standards dealing with test and measurement quality requirements, such as
ANSI/NCSL Z 540-1, ISO 10012, ISO/IEC 17025 and the ISO 9000 series
do not require the use or reporting of specific test numbers to establish
traceability.
GMP 13 Page 2 of 11
March 2003
1.4 Safety
2. Methodology
2.1 Summary
2.2 Procedure
The charts in this GMP show examples of the traceability hierarchy for mass, length,
volume, and temperature measurement disciplines. Each laboratory must define the
exact traceability system in their operating quality manual. (A worksheet is included
as an Appendix to help in the definition of the traceability system.)
3. Calculations
4. Assignment of Uncertainty
The uncertainty associated with reported calibration values is included within the uncertainty
analysis for each SOP and in SOP 29, Calculating and Reporting Uncertainties.
GMP 13 Page 3 of 11
March 2003
Mass - Option A
NIST
P 1- kg.
P 1-kg..
P 100 g.
W 1-kg. C 1-kg.
P 100 g..
W 1-kg.. C 1-kg..
W 100 g. C 100 g.
W 100 g.. C 100 g..
Remaining Client
Laboratory Echelon I, II
Standards Calibration
Metric/Avdp Workload
Laboratory Field
Level Working
Standards
Client
Echelon III
Calibration
Workload
GMP 13 Page 4 of 11
March 2003
Mass - Option B
NIST
Laboratory
Standards
Metric/Avdp
Client
Laboratory Field
Echelon I, II
Level Working
Calibration
Standards
Workload
Client
Echelon III
Calibration
Workload
GMP 13 Page 5 of 11
March 2003
Mass - Option C
NIST
Accredited
Laboratory
Laboratory
Primary
Standards
Metric/Avdp
Laboratory Field
Level Working
Standards
Client
Echelon III
Calibration
Workload
GMP 13 Page 6 of 11
March 2003
Volume - Option A
NIST
Volume Group or
Mass Group
Mass Standards
Laboratory
Mass Standards
Metric/Avdp
Volume Standards
5 gal , 25 gal,
100 gal
Client
Gravimetric
Client Volume Volume Calibration
Transfer (5 gal or smaller)
Calibration
Workload
(5 gal or larger)
GMP 13 Page 7 of 11
March 2003
Volume - Option B
NIST
Volume Group
GMP 13 Page 8 of 11
March 2003
Length
NIST
Client
Length Bench
Calibration
Working Standard
Workload
Client
Calibration
Workload
GMP 13 Page 9 of 11
March 2003
Temperature
NIST
Laboratory
Working Standards
Client
Calibration
Workload
GMP 13 Page 10 of 11
March 2003
GMP 13 Page 11 of 11
March 2003
GMP 14
1. Introduction
1.1. Purpose
The following practice will guide you through the process of selecting and
using a correct sensitivity weight for mass determination procedures.
1.2. Prerequisites
1.2.1. Verify that valid calibration certificates are available for the masses to
be used as standards, sensitivity weights, and tare weights.
1.2.3. Verify that the operator is familiar with the design and the operation
of the balances and familiar with weighing procedures.
1.3. Safety
1.3.1. Handling of large or small weights can represent a hazard to either the
weights or personnel if the weights are dropped.
GMP 14 Page 1 of 4
March 2003
2. Methodology
2.1. Summary
2.2. Apparatus
2.3.3. Determine the need for tare weights if the difference between X and S
exceeds the values shown in Table 1.
GMP 14 Page 2 of 4
March 2003
2.3.4. Select tare weights, if necessary, making sure that the difference
between X and S, with the appropriate tare weights, do not exceed the
values shown in Table 1.
*A sensitivity weight is not required if the electronic mass comparator that is used has been
tested (with supporting data available) to determine that the balance has sufficient accuracy,
resolution, repeatability, and stability so that no advantage is gained using a sensitivity
weight. For example, any possible errors must be less than the last digit retained in the
expanded uncertainty. When a mass comparator is used without a sensitivity weight, the
sensitivity must be periodically verified and documented (e.g., prior to each use).
The sensitivity weight is used to ensure that the mass differences determined
with the optical scale, or electronic range, have valid accuracy and traceability.
The sensitivity weight calibrates the range of use of the balance used for
making the mass determinations. Using a sensitivity weight provides us with a
sensitivity value in terms of mass units per division.
mass units M sw
sensitivity = =
divisions deflection
GMP 14 Page 3 of 4
March 2003
3. Calculations
No special calculations are associated with this practice. See each mass SOP for
calculation of sensitivity within the procedure.
4. Uncertainty
No uncertainty calculations are associated with this practice. See the appropriate SOP
for the calculations of uncertainty. (The uncertainty of the sensitivity weight does not
need to be included in calculations of uncertainty since the uncertainty value is
distributed across the range of use. However, it does no harm to incorporate it in
uncertainty calculations when spreadsheets are set up to handle all of the data.)
GMP 14 Page 4 of 4
March 2003
SOP 1
1. Introduction
1.1. Test/calibration reports are the visible outputs of the testing laboratory. They should be
prepared with utmost care to ensure that they accurately convey all information
pertaining to the testing so that reports may be used with maximum benefit by all
concerned. Carefully prepared test reports will contain or refer to all information
necessary to justify the test results.
1.2. The test report may consist of filling in the blanks in a form in the case of a routine
measurement. A more detailed report, including narrative information, may be required
for special calibrations or tests.
1.3. Regardless of the final form, the test report must contain the basic information
described in the following sections.
2. Content
2.2. Name and address of the laboratory, or location at which tests were performed.
2.3. Unique identification of the test report or calibration certificate, and on each page an
identification in order to ensure that the page is recognized as part of the test report or
calibration certificate, and a clear identification of the end of the report or certificate.
2.5. Method used – Describe how test was made by reference to SOP(s). In the absence of
SOP’s, brief but informative descriptions of the methodology should be included.
Information describing deviations from previously agreed upon procedure must also be
included.
2.6. Description of, the condition of, and unambiguous identification of the item calibrated.
A laboratory number should be assigned and attached to each test item at the time of its
acceptance for testing. The use of the laboratory number will facilitate the internal
control of test items during the testing process.
2.7. Date of receipt of calibration item where this is critical to the validity and application of
the results, and the date of performance of calibration.
SOP 1 Page 1 of 6
March 2003
2.8. Calibration results with the units of measurement in tabular or other convenient form.
(When an instrument has been repaired or adjusted the calibration results before and
after repair or adjustment, if available, are reported.)
2.10. Conditions (e.g., environmental) under which the calibrations were made that have an
influence on the measurement results.
2.14. Additional information which may be required by specific methods, clients or groups of
clients.
2.15. Name, title, and signature of person authoring the report or certificate. Other signatures
may be required, at the discretion of the laboratory director. Each signer accepts
his/her share of responsibility for the contents of the report.
2.16. Where relevant, a statement to the effect that the results relate only to the items tested
or calibrated.
2.17. Hard copies of test reports should also include the page number and total number of
pages.
2.18. A statement specifying that the test report or calibration certificate shall not be
reproduced except in full, without written approval of the laboratory.
3. Recording
3.1. File all test reports in a systematic manner for ease of retrieval, as necessary.
3.2. Retain copies of all test reports for a minimum period of five years, until superseded by
a subsequent report, or as stated in the laboratory quality manual, or until deemed by
the laboratory director as having no future value.
SOP 1 Page 2 of 6
March 2003
Report of Test
Issued by
Name of Testing Laboratory
Laboratory Report No. ______________
Submitted by:_________________________________________________________________________
Date: _______________________________________________________________________________
________________________________________________________________________________________
Test Results*:
As Found After Adjustment
________________________________________________________________________________________
Reference Information:
Traceability: ________________________________________________________________________
Data
Reference: ________________________________________________________________________
*Report, as appropriate
The results stated on this report relate only to the items specifically identified.
This test report or calibration certificate shall not be reproduced except in full, without written approval of the laboratory.
SOP 1 Page 3 of 6
March 2003
A B C COMPANY
Report of Calibration
Report Number:
Name of Device:
Submitted by:
Calibration (date) .
The ambient conditions were °C, % relative humidity and mm Hg Barometric pressure.
The item tested was/was not in tolerance at time of calibration. Adjustments are noted and any out of tolerance data are
attached.
Data:
The primary standards to which the above data are traceable are identified in this report. The calibration of these
standards is traceable to the National Institute of Standards and Technology. The cycling and certification of all
standards of measurement at this facility meet the requirements of ISO/IEC 17025.
The results stated on this report relate only to the items specifically identified.
This test report or calibration certificate shall not be reproduced except in full, without written approval of the laboratory.
SOP 1 Page 4 of 6
March 2003
BUREAU OF STANDARDS
PO Box 12345
City, State 12345-1234
REPORT OF CALIBRATION
FOR
1 kg to 10 mg weight kit
(Twenty-one metric weights)
SUBMITTED BY
Uncertainty statement:
The combined standard uncertainty includes the standard uncertainty reported for the standard, the standard uncertainty for the measurement
process, the standard uncertainty for any uncorrected errors associated with buoyancy corrections, and a component of uncertainty to account for
any observed deviations from NIST values that are less than surveillance limits. The combined standard
SOP 1 Page 5 of 6
March 2003
Traceability statement:
The Standards of the Compliant Calibration Laboratory are traceable to the National Metrology Institute, and are part of a comprehensive
measurement assurance program for ensuring continued accuracy and measurement traceability within the level of uncertainty reported by this
laboratory. The laboratory test number identified above is the unique report number to be used in referencing measurement traceability for
artifacts identified in this report only.
Supplemental Information
Twenty one metric weights from 1 kg to 10 mg, marked ASTM Class 4. Weights from 1 kg to 1 g: two-piece weights, with assumed density of
8.0 g/cm3. Weights from 500 mg to 50 mg: sheet weights, with assumed density of 16.6 g/cm3. Weights from 30 mg to 10 mg: sheet weights,
with assumed density of 2.7 g/cm3.
Artifacts showed evidence of improper handling. Fingerprints and dents were visible on the surface of the weights.
Artifacts were cleaned with cheesecloth and ethyl alcohol. Thermal equilibrium time/conditions: ten days next to balances in mass lab.
Procedure used:
Temperature: 20.1 °C to 20.2 °C Barometric Pressure: 752.7 mmHg Relative Humidity: 43.35 % to 43.40 %
Date artifacts were received: February 15, 2001 Date of report preparation: March 3, 2002
Date of test: February 25, 2001
Due date per customer’s request: February 25, 2002
Josh Balani II
Test performed by: Josh Balani II
Metrology Expert
Member: ARMAP
NCSLI
NCWM
ASQ
This document does not represent or imply endorsement by the Compliant Calibration Laboratory, NMI, or any agency of the State and/or
national governments. This document may not be reproduced, except in full, without the written permission of the Compliant Calibration
Laboratory.
SOP 1 Page 6 of 6
March 2003
SOP 2
1. Introduction
1.1. Purpose
If uncorrected, the effect of air buoyancy on the objects being compared is frequently
the largest source of error in mass measurement. This SOP provides the equations to
be used to correct for the buoyant effect of air. The significance of the air buoyancy
correction depends upon the accuracy required for the measurement, the magnitude
of the air buoyancy correction relative to other sources of error in the overall
measurement process, and the precision of the mass measurement. An air buoyancy
correction should be made in all high accuracy mass determinations. The
gravimetric volume procedure uses a high accuracy mass determination with the
corresponding buoyancy corrections. The Appendix to this SOP provides a brief
theoretical discussion of this subject.
1.2. Prerequisites
1.2.1. Verify that (true) mass values or corrections are available for the standards
used in the measurement process. The calibration values for the mass
standards must be traceable to NIST.
1.2.2. Verify that the thermometer, barometer, and hygrometer used have been
calibrated and are in good operating condition as verified by periodic tests or
cross-checks with other standards.
2. Methodology
SOP 2 Page 1 of 21
March 2003
2.3.1. Estimate the magnitude of the air buoyancy correct, MABC, using the following
formula:
MABC = ( ρ a − ρ n ) ( Vx − Vs )
1
The barometer, thermometer, and hygrometer are used to determine the air density at the time of the
measurement. The air density is used to make an air buoyancy correction. The accuracies specified are
recommended for high precision calibration. Less accurate equipment can be used with only a small
degradation in the overall accuracy of the measurement.
SOP 2 Page 2 of 21
March 2003
2.4. Procedure
2.4.1. Record the temperature, pressure, and relative humidity at the start and at the
end of the measurement process as near the location of the measurement as
necessary and practical. If these parameters change significantly during the
measurement process, it may be necessary to wait for more stable operating
conditions or to use average values to compute the air density. Use of the
average environmental values may influence the uncertainty of the
measurement result and must be evaluated for significance.
2.4.2. Determine the air density using the equation given in Section 8 of the
Appendix to this SOP.
3. Calculations
3.1. Calculate the mass, Mx, of the unknown weight, X, using the following equation,
where d represents the “difference” obtained with buoyancy corrections applied to
the sensitivity weight.
ρ
M s 1 − a + d
ρs
Mx =
ρ
1 − a
ρx
If tare weights were carried with X and/or S, use the following equation:
ρ ρa
M s 1 − a + d + M t s 1 − − M t 1 − ρa
ρs ρt x ρt x
Mx = s
ρ
1 − a
ρx
SOP 2 Page 3 of 21
March 2003
3.2. If reporting the conventional mass2, CMx, compute it using the following.
0.0012
M x 1 −
ρ
CM x = x
0.0012
1−
8.0
3.3. If reporting the apparent mass, AMx, versus brass, compute it using the following.
0.0012
M x 1 −
ρ
AM x vs brass = x
0.0012
1−
8.3909
3.4. The conventional and apparent mass values are related by the following:
0.0012
AM x vs brass 1 −
8.3909
CM x =
0.0012
1−
8.0
2
Conventional Mass: “The conventional value of the result of weighing a body in air is equal to the mass of a
standard, of conventionally chosen density, at a conventionally chosen temperature, which balances this body at this
reference temperature in air of conventionally chosen density.” The conventions are: reference density 8.0 g/cm3;
reference temperature 20 °C; normal air density 0.0012 g/cm3. Conventional mass was formerly called “Apparent
Mass versus 8.0 g/cm3” in the United States. See OIML IR 33 (1973, 1979), under revision.
SOP 2 Page 4 of 21
March 2003
4. Assignment of Uncertainty
The uncertainty in determining the air buoyancy correction is usually negligible relative to
the precision of the measurement process itself. Consequently, the uncertainty for the
measurement is based upon the uncertainty for the measurement process used. The
uncertainty in the air density equation as given in numerous periodicals is 0.0012 mg/cm3
(or about 0.1 % of normal air density.)
SOP 2 Page 5 of 21
March 2003
Appendix A
by Richard S. Davis3
1. Introduction
In performing measurements of mass, the balance or scale used acts as a force transducer.
The force produced by an object to be weighed in air has two components: one
proportional to the mass of the object, and the other proportional to its volume. The latter
component, or buoyant force, may under some circumstances be large enough to require
correction. The following shows under what circumstances buoyancy corrections are
required as well as how they are made.
2. Scope
The method for applying buoyancy corrections presented below applies to mass
measurements made in air. The density of air is computed from auxiliary measurements of
temperature, pressure and relative humidity after which the buoyancy corrections are
calculated directly from the Principle of Archimedes. The following weighing situations
are considered.
3. Summary of Method
3
Richard S. Davis, formerly of the National Institute of Standards and Technology, Mass Group.
accuracies if the unknown object whose mass is to be determined has a density that differs
widely from that of the standards (weighing of water, for example). Many mass standards
are calibrated in terms of a so-called "apparent mass" [conventional mass] scale (See
Chapter 7.3, Handbook 145). Use of this scale does not indiscriminately eliminate the
need for buoyancy corrections as is sometimes assumed.
5. Terminology
5.2. Mass
The term "mass" is always used in the strict Newtonian sense as a property intrinsic
to matter. Mass is the proportionality constant between a force on a material object
and its resulting acceleration. This property is sometimes referred to as "true mass",
"vacuum mass", or "mass in vacuum" to distinguish it from conventional [apparent]
mass.
The mass of material of a specified density that would exactly balance the unknown
object if the weighing were carried out at a temperature of 20 °C in air of density
0.0012 g/cm3. The mass, MN, of an object, N, is related to its apparent mass MN,A by
the equation:
0.0012
M N 1 −
ρN
M N ,A =
0.0012
1 −
ρ B
4
Pontius, P. E., Mass and Mass Values, NBS Monograph 122, 1974, pp 12-23, 26-33 described the concept of
apparent mass. The term conventional mass is described by OIML R 33, the Conventional Value of Mass in Air.
There are at present two apparent mass scales in wide use. The older is based on
ρB = 8.4000 g/cm3 at 0 °C with a coefficient of volumetric expansion of 0.000054/ °C
and the more recent (Conventional Mass) specifies ρB = 8.0000 g/cm3 at 20 °C. The
quantity MN,A is a function of the particular conventional or apparent mass scale,
which has been used in its calculation. OIML IR 33 only recognizes Conventional
Mass.
5.4. Sensitivity
ρ
M sw 1 − a
M sw − ρ a Vsw ρ sw
sensitivity = =
∆R ∆R
6. Apparatus
In order to ascertain the density of air at the balance, the following measuring instruments
are necessary: thermometer, barometer, and hygrometer. Ideally, these instruments should
be placed in or next to the balance case (as near the measurement location as is practical).
It may only be practical for the thermometer or temperature sensor to actually be placed
inside the balance chamber. A calculator or computer will be extremely useful for this
procedure.
7. Procedure
Weigh the unknown object as directed by the balance manufacturer or in accordance with
accepted procedure. Record the temperature, pressure and relative humidity of the air in
the balance at the time of weighing. Do not correct the barometric pressure to sea level.
8. Calculation
The density of air, in g/cm3, can be approximated for lesser accuracy from the
following formula:5
Small errors (of order 0.01 %) in this equation occur for locations well above sea level or
under conditions in which the concentration of carbon dioxide differs greatly from the
global average. See the references for a more general formulation of the equation.
5
Jones, F.E., "The Air Density Equation and the Transfer of the Mass Unit," Journal of Research, National Institute
of Standards and Technology, Vol. 83, 1978, p. 419.
pM a
ρ = (1 − 0.3780 xv )
ZRT
ρ sv
xv = ( h / 100 ) f
p
[ ] + Tp ( d )
2
p
Z =1− a0 + a1t + a2t 2 + ( b0 + b1t ) xv + ( c0 + c1t ) xv
2
2
+ exv
2
Calculate the density of air at the balance during the weighing. Then determine the
mass of the unknown, Mx, as follows:
6
CIPM References: Giacomo, P. Metrologia 18: 33-40 (1982), Davis, R.S., Metrologia 29: 67-70 (1992).
or
ρ
M s 1 − a + sensitivity ∆M opt
ρs
Mx = (3b)
ρ
1 − a
ρx
or
7
In general, V(t) = V20 [1 + ß(t - 20 °C)] where t is the temperature of the weight, V20 is the volume at 20 °C, and ß
is the coefficient of volumetric expansion.
If the mass and volumes of the standards have been adjusted to a conventional mass
scale, then
ρ
CM s 1 − a + sensitivity ∆M opt
ρB
Mx ≈ (4)
ρ
1 − a
ρx
8.4 If a single-pan balance with built-in weights is used, it is probable that the built-in
weights have been adjusted on an apparent mass or conventional mass scale8.
Determine which apparent mass scale has been used and calculate the mass of the
unknown from the equation
ρ
M D 1 − a + sensitivity M opt
ρB
Mx ≈ (5)
ρ
1 − a
ρx
If the balance has been used only as a comparator, that is, to compare the mass of the
unknown object with that of some external standard, then:
8
Schoonover, R. M. and Jones, F. E., "Air Buoyancy in High-Accuracy Weighing on Analytical Balances," Anal.
Chem., 53, 1981, p. 900.
For some balances, operation requires that the user restore the balance to null by
means of a manually controlled dial. The portion of the mass reading controlled by
this dial should be treated, for purposes of buoyancy corrections, as an optical scale.
8.5 If a single-pan balance with full-range electronic control is used, the following
should be noted. As part of its calibration, the electronic gain has been adjusted by
means of a weight of known mass. For example, if the range of electronic control is
100 g, the electronics have been adjusted so that a 100-g standard weight produces
an indication of precisely 100 g. This procedure effectively builds an apparent mass
calibration into the balance. The reference density of the apparent mass scale is the
density of the standard mass used for the calibration and the reference air density is
the air density at the time of calibration.
ρ
M R 1 − a
ρc
Mx = (6)
ρ
1 − a
ρx
If the balance includes both an electronic control system and built-in weights, the
buoyancy considerations for the built-in weights are as described in section 8.2 and
the considerations for the electronically determined mass are those given directly
above.
8.6 Top-loading balances may be considered a form of single-pan balance and the
appropriate procedure for buoyancy correction followed.
9. Precision
The contribution of the random error of the evaluation of air density to the precision of
mass measurement may be estimated as follows:
δρ a ( V x − Vs )
Table A-10. Variables for above equation
Variable Description
δρa random error of evaluation of ρa
volume of standards, if weighing by substitution
Vs
Vs = MD / DB, if using the built-in weights on a single pan balance.
Vx volume of object weighed
The quantity, δρa will have contributions from the measurements of temperature, pressure
and relative humidity which are required for the calculation of ρa. Equation (1) may be
used to estimate the effects of imprecision in measurements of P, t, and U. It is unrealistic
to expect δρa /ρa ever to be less than 0.05 % even using the best techniques available.
10. Accuracy
If the standards have been calibrated in terms of conventional mass, complete neglect of
buoyancy corrections will produce an error in the measured result Mx of order:
(
CM s 1.2 x 10 −3 − ρ a ) 1
−
1 ρB − ρ x ρa
+
ρ B
(7)
ρs ρB ρx
Use of equation (4), on the other hand, introduces only an error of approximately
1
( )
CM s 1.2 x 10 −3 − ρ a −
1
(8)
ρs ρB
It is a requirement for manufacture that the actual density of standard weights be near
enough to the assumed density of the apparent mass scale to which they are adjusted that
the magnitude of (8) will always be small under normal conditions in laboratories near sea
level.
The fact that there are two apparent mass scales widely used - one based on density
8.0 g/cm3 and an older one based on 8.4 g/cm3 - means that some caution is required on the
part of the user. Conventional mass is generally preferred and reported for all calibrations
where mass standards will be used to calibrate weighing instruments. For the most accurate
work, the apparent mass scale should be abandoned in favor of substitution weighing with
standards of known mass and volume.
The user must decide the accuracy required of the particular mass measurement and choose
a buoyancy correction technique commensurate with that accuracy.
The same considerations, which apply to the accuracy of buoyancy corrections in weighing
by substitution, are easily extended to the other types of weighing indicated above.
There are many factors, which affect the accuracy of a mass measurement. The above has
dealt only with those arising from problems of buoyancy.
Appendix B
Examples
Example 1:
The weight set of Table B-1 is used with an equal-arm balance to find the mass of a piece of
single-crystal silicon. The following weights were used to balance the silicon: 10 g, 3 g. The
balance pointer showed the silicon side to be light by 3.5 divisions. The 10 mg weight of Table
1 was used to find the sensitivity of the balance. When the weight was added, the pointer moved
10.3 divisions. At the time of the weighing, the following measurements were taken:
P = 748.1 mm Hg
t = 22.3 °C
U = 37 % relative humidity
Answer:
−5 315.56
( 22.3 + 273.15 )
es = 1.314 6 x 10 x e
9
= 20.194 mm Hg
The density of silicon at 20 °C is 2.329 1 g/cm3 and its coefficient of linear expansion is
0.000 002 6 /°C.
sensitivity =
( 0.009 999 ) = 0.000 971 g/division = 0.971 mg/division
10.3
Mx =
[ 13.000 172 − ( 1.647 ) ( 1.172 x 10 ) ] + [ ( 0.971 x 10 ) ( − 3.5 ) ]
−3 −3
1.172 x 10 −3
1 −
2.329 1
Mx = 13.001 385 g
Example 2:
Let us again consider the weighing performed in Example 1. This time, all we know about our
weight set is that it has been adjusted to the 8.4 apparent mass scale at the best available
commercial tolerance.
Mx = 13.001 329 g
For routine weighing, it is sometimes satisfactory to assume that the temperature is 20 °C and the
density of air is 1.2 x 10-3 g/cm3. Had this been done, the computed value for the silicon would
be
Mx = 13.001 442 g
Example 3:
Another piece of silicon is measured on a single-pan microbalance. The balance weights were
adjusted by the manufacturer to the conventional mass scale. The sensitivity of the balance has
been determined to be exactly 1.000. This particular laboratory is well above sea level. At the
time of the weighing, the following measurements were recorded:
P = 612.3 mm Hg
t = 23.4 °C
U = 23 % relative humidity
The balance reading was 15.00 g on the built-in weights and 0.000 358 g on the optical screen.
What is the mass of the silicon?
Answer:
es = 21.59 mm Hg
0.956 x 10−3
( 15.00 + 0.000 358 ) 1 −
8.0
Mx =
0.956 x 10 −3
1 −
2.329 1
Mx = 15.004 724 g
Example 4:
The built-in weights in Example 3 are actually stainless steel of density 7.78 g/cm3 at 20 °C.
What is the approximate error caused by using the apparent mass scale?
Answer:
( )
15 g 0.2 x 10 −3 g / cm 3
1
3
−
1
8.0 g / cm 3
= 11 µg
7.78 g / cm
This discrepancy, though larger than the precision of the best analytical balances, is actually well
within the tolerance of Class 1 weights.
SOP No. 4
1.0 Introduction
1.1. Purpose
1.2. Prerequisites
1.2.1. Verify that valid calibration certificates are available for the standards
used in the test.
1.2.2. Verify that the standards to be used have sufficiently small standard
uncertainties for the intended level of calibration. Primary standards
should not be used at this level.
1.2.3. Verify that the balance is in good operating condition with sufficiently
small process standard deviation as verified by a valid control chart or
preliminary experiments to ascertain the performance quality when a new
balance is put into service.
1.2.5. Verify that the laboratory facilities comply with the following minimum
conditions to meet the expected uncertainty possible with this procedure.
SOP 4 Page 1 of 18
December 2003
2. Methodology
2.2. Summary
The double substitution procedure is the same for all of the balances mentioned
above, but the adjustment of the balance to prepare for the intercomparison and
the selection of the sensitivity weight differs slightly depending upon the balance
used. When steps specific to a particular balance are required, they are given in
subsections of the procedure identified by a, b, and c along with the balance type.
2.3.1. Precision balance with sufficient capacity and sensitivity for the
calibrations planned.
SOP 4 Page 2 of 18
December 2003
2.3.3. Calibrated small standard weights with recent calibration certificates and
values traceable to NIST to be used as tare weights.
2.3.5. Forceps to handle the weights, or gloves to be worn if the weights are
moved by hand.
2.3.6. Stop watch or other timing device to observe the time of each
measurement.
2.4. Symbols
1The barometer, thermometer, and hygrometer are used to determine the air density at the time of the measurement.
The air density is used to make an air buoyancy correction. The accuracies specified are recommended for high
precision calibration. Less accurate equipment can be used with only a small degradation in the overall accuracy of
the measurement.
SOP 4 Page 3 of 18
December 2003
Symbol Description
2.5. Procedure
2.5.1.1. Place the test weight and standards in the balance chamber or
near the balance overnight to permit the weights and the balance
to attain thermal equilibrium.
1
a. Single-pan mechanical balance - less than the range
10
of the optical scale.
1
c. Combination balance - less than the range of the
10
digital indications.
1
d. Mass comparator – less than digital range
10
SOP 4 Page 4 of 18
December 2003
SOP 4 Page 5 of 18
December 2003
1 S + ts O1
2 X + tx O2
3 X + tx + sw O3
4 S + ts + sw O4
All observations should be recorded on suitable data sheets, such as those in the
appendix. Record the laboratory ambient temperature, barometric pressure, and
relative humidity.
2.5.2.5. Compare the two differences (O2 – O1) and (O3 – O4); they
should not differ from one another by more than 2 standard
deviations of the balance for this process and load. If this
difference is exceeded, reject the data and redo the
SOP 4 Page 6 of 18
December 2003
1 X + tx O1
2 S + ts O2
3 S + ts + sw O3
4 X + tx + sw O4
3.1. No air buoyancy correction. Calculate the conventional mass correction, Cx, for
the test weight as follows, according to the optional sequence used. In each case,
the conventional mass corrections for the standard weight(s), Cs, the conventional
mass of the tare weights, CM t s and CM t x , and the conventional mass of the
sensitivity weight, CMsw, are included. The symbols Ns and Nx refer to the
nominal values of S and X, respectively. If no tare weights, and equal nominal
values are used, those terms may all be deleted from the equations.
.
3.1.1. Optional Sequence A (SXXS)
⎡ ( O 2 - O1 ) + ( O 3 - O 4 )⎤⎡ CM sw ⎤
C x = C s + CM t s - CM t x + ⎢ ⎥⎦ ⎢ O - O ⎥ + N s - N x
⎣ 2 ⎣ 3 2 ⎦
SOP 4 Page 7 of 18
December 2003
⎡ ( O1 - O 2 ) + ( O 4 - O 3 )⎤⎡ CM sw ⎤
C x = C s + CM t s - CM t x + ⎢ ⎥⎦ ⎢ O - O ⎥ + N s - N x
⎣ 2 ⎣ 3 2 ⎦
3.2.1. Calculate the air density, ρa, as described in the Appendix to SOP No. 2.
3.2.2. Calculate the mass Mx of the test weight, and its mass correction Cx using
the mass of the standard weight(s), the tare weights and the sensitivity
weights according to the optional sequence used.
⎡ ⎛ ρa ⎞ ⎤
⎢ Msw ⎜⎜ 1 - ⎟⎟ ⎥
⎛ ρ ⎞ ⎛ ρ ⎞ ⎛ ρ ⎞ ⎡
⎟⎟ + Mts ⎜ 1 - a ⎟ - Mtx ⎜ 1 - a ⎟ + ⎢
( O2 - O1 ) + ( O3 - O4 ) ⎤ ⎢ ⎝ ρsw ⎠ ⎥
Ms ⎜⎜ 1 - a ⎥ ⎢
⎝ ρs ⎠ ⎜ ⎟
ρts ⎠ ⎜ ⎟
ρtx ⎠ ⎣ 2 ⎦ O3 - O2 ⎥
⎝ ⎝ ⎢ ⎥
⎢⎣ ⎥⎦
Mx =
⎛ ρ ⎞
⎜⎜ 1 - a ⎟⎟
⎝ ρx ⎠
⎡ ⎛ ρ ⎞ ⎤
⎢ M sw ⎜⎜ 1 - a ⎟⎟ ⎥
⎛ ρ ⎞ ⎛ ρ ⎞ ⎛
⎟ - Mt ⎜ 1 - ρa
⎞
⎟ + ⎡ ( O1 - O2 ) + ( O4 - O3 ) ⎤ ⎢ ⎝ ρ ⎠ ⎥
M s ⎜⎜ 1 - a ⎟⎟ + Mts ⎜ 1 - a ⎢ ⎥ ⎢
sw
⎝ ρs ⎠ ⎜ ρts ⎟ x ⎜ ρtx ⎟ ⎣ 2 ⎦ O3 - O2 ⎥
⎝ ⎠ ⎝ ⎠ ⎢ ⎥
Mx = ⎣⎢ ⎦⎥
⎛ ρ ⎞
⎜⎜ 1 - a ⎟⎟
⎝ ρx ⎠
Cx = Mx - Nx
where Nx is the nominal value for X.
SOP 4 Page 8 of 18
December 2003
3.2.5. If requested, the apparent mass versus the reference density of brass may
be calculated. This value should only be used when calibrating
mechanical balances that have been adjusted to this reference density.
⎛ ρ ⎞
M x ⎜⎜ 1 - n ⎟⎟
⎝ ρx ⎠
AM x vs brass =
⎛ ρn ⎞
⎜1- ⎟
⎝ 8.3909 ⎠
4. Measurement Assurance
4.1. Duplicate the process with a suitable check standard (See GLP 1, SOP 9, SOP 30,
and Sec. 7.4)
4.2. Plot the check standard value and verify that it is within established limits; a t-test
may be incorporated to check observed value against accepted value.
4.3. The mean of the check standard is used to evaluate bias and drift over time.
4.4. Check standard observations are used to calculate the standard deviation of the
measurement process, sp.
5. Assignment of Uncertainty
2 Conventional Mass: “The conventional value of the result of weighing a body in air is equal to the mass of a
standard, of conventionally chosen density, at a conventionally chosen temperature, which balances this body at this
reference temperature in air of conventionally chosen density.” The conventions are: reference density 8.0 g/cm3;
reference temperature 20 °C; normal air density 0.0012 g/cm3. Conventional mass was formerly called “Apparent
Mass versus 8.0 g/cm3” in the United States. See OIML IR 33 (1973, 1979), under revision.
SOP 4 Page 9 of 18
December 2003
(RSS), and the expanded uncertainty, U, reported with a coverage factor of two (k=2), to
give us an approximate 95 percent level of confidence. See SOP 29 for the complete
standard operating procedure for calculating the uncertainty.
5.1. The expanded uncertainty for the standard, U, is obtained from the calibration
report. The combined standard uncertainty, uc, is used and not the expanded
uncertainty, U, therefore the reported uncertainty for the standard will usually
need to be divided by the coverage factor k.
5.2. The value for sp is obtained from the control chart data for check standards using
double substitution measurements. (See SOP No. 9.)
5.3. Other standard uncertainties usually included at this calibration level include
uncertainties associated with calculation of air density and standard uncertainties
associated with the density of the standards used.
5.4. The expanded uncertainty, U, must be ≤ 1/3 of the tolerance applicable as per
ASTM E617-97 and OIML R111 to classify mass standards.
6. Report
SOP 4 Page 10 of 18
December 2003
Appendix
Double Substitution Data Sheet
(Optional Sequence A)
SXXS
Laboratory data and conditions:
Operator
Date Temperature
Balance Pressure
Nominal Load Relative Humidity
Standard deviation of the process, from control chart, sp
*Mass Correction = True Mass if using buoyancy correction. Mass Correction = Conventional Mass if NOT using
buoyancy correction. Density is used only with buoyancy corrections.
Observations:
Observation No. Weights Balance Observations, Units
Time:
1 (O1) S + ts
2 (O2) X + tx
3 (O3) X + tx + sw
4 (O4) S + ts+ sw
Time:
SOP 4 Page 11 of 18
December 2003
Appendix
Double Substitution Data Sheet
(Optional Sequence B)
XSSX
Laboratory data and conditions:
Operator
Date Temperature
Balance Pressure
Load Relative Humidity
Process standard deviation from control chart, sp
*Mass Correction = True Mass if using buoyancy correction. Mass Correction = Conventional Mass if
NOT using buoyancy correction. Density is used only with buoyancy corrections.
Observations:
Observation No. Weights Balance Observations, Units
Time:
1 (O1) X + tx
2 (O2) S + ts
3 (O3) S + ts+ sw
4 (O4) X + tx + sw
Time:
SOP 4 Page 12 of 18
December 2003
*Mass Correction = True Mass if using buoyancy correction. Mass Correction = Conventional Mass if NOT using
buoyancy correction. Density is used only with buoyancy corrections.
Observations:
Observation No. Weights Balance Observations, Units mg
Time: 8:35 AM
1 (O1) S + ts 1 268
2 (O2) X + tx 1 821
3 (O3) X + tx + sw 6 798
4 (O4) S + ts+ sw 6 245
Time: 8:47 AM
SOP 4 Page 13 of 18
December 2003
Use equation 3.2.2.1 for optional sequence A (SXXS) with buoyancy corrections3:
⎡ ⎛ ρa ⎞ ⎤
⎢ Msw ⎜⎜ 1 - ⎟⎟ ⎥
⎛ ρ ⎞ ⎛ ρ ⎞ ⎛ ρ ⎞ ⎡
⎟⎟ + Mts ⎜ 1 - a ⎟ - Mtx ⎜ 1 - a ⎟ + ⎢
( O2 - O1 ) + ( O3 - O4 ) ⎤ ⎢ ⎝ ρ ⎠ ⎥
Ms ⎜⎜ 1 - a ⎥ ⎢
sw
⎝ ρs ⎠ ⎜ ⎟
ρts ⎠ ⎜ ⎟
ρtx ⎠ ⎣ 2 ⎦ O3 - O2 ⎥
⎝ ⎝ ⎢ ⎥
⎢⎣ ⎥⎦
Mx =
⎛ ρ ⎞
⎜⎜ 1 - a ⎟⎟
⎝ ρx ⎠
⎡ ⎛ 0.0011795 ⎞ ⎤
⎢ 0.0049773 ⎜ 1 - ⎟⎥
⎛ 0.0011795 ⎞ ⎡ ( 1.821- 1.268 ) + ( 6.798- 6.245 ) ⎤ ⎢ ⎝ 8.5 ⎠⎥
9.999321 ⎜ 1 - ⎟+0 - 0+⎢ ⎥ ⎢
⎝ 8.0 ⎠ ⎣ 2 ⎦ 6.798- 1.821 ⎥
⎢ ⎥
⎣ ⎦
MX =
⎛ 0.0011795 ⎞
⎜ 1- ⎟
⎝ 7.84 ⎠
Cx = Mx – Nx
3 Keep in mind that these equations may be truncated for the purpose of this example and minor differences
may be seen in the ending decimal places due to the use of calculators or spreadsheets.
SOP 4 Page 14 of 18
December 2003
⎛ 0.001 2 ⎞
9.999 904 1 ⎜ 1 - ⎟
CM x = ⎝ 7.84 ⎠
0.999 850
Cx = CM x – Nx
Cx = 9.999 873 51 g – 10 g
U = uc * 2
2 2 2
uc = u s + s p + uo
The uncertainty for the standard, U, must be divided by the k factor to determine the us.
Uncertainty Statement
The uncertainty reported is the root sum square of the standard uncertainty of the standard, the
standard deviation of the process, and an uncorrected systematic error for lack of buoyancy
corrections, multiplied by a coverage factor of 2 (k=2) for an approximate 95 percent confidence
interval. Factors not considered in the evaluation: magnetism (weights are considered to meet
magnetism specifications unless measurement aberrations are noted), balance eccentricity and
linearity (these factors are considered as a part of the measurement process when obtaining the
standard deviation of the process).
SOP 4 Page 15 of 18
December 2003
Compliance Evaluation
We have to evaluate the correction with its expanded uncertainty to determine if the weight is in
tolerance or not. The magnitude of the expanded uncertainty has to be less than 1/3 of the
tolerance to be able to perform that evaluation, according to ASTM E617-97 and OIML R111.
Load = 10 g
ASTM E617 OIML R111
Class Tolerance (mg) Class Tolerance (mg)
0 0.025 E1 0.020
1 0.050 E2 0.060
2 0.054 F1 0.20
If we look at three times the uncertainty: 0.011 mg x 3 = 0.033 mg, we realize that the
uncertainty complies with the 1/3 rule for ASTM classes 1, 2, and OIML classes E2, F1.
Next, we look at the correction with the uncertainty: - 0.126 mg ± 0.011 mg.
We can see that the absolute value of the correction is within: 0.115 ≤ Cx ≤ 0.137, therefore, it
only complies with OIML class F1.
Reporting
The conventional mass correction and uncertainty would be reported as follows:
CX = - 0.126 mg ± 0.011 mg
SOP 4 Page 16 of 18
December 2003
Observations:
Observation No. Weights Balance Observations, Units mg
Time: 9:00 AM
1 (O1) X + tx 20 93
2 (O2) S + ts 17 21
3 (O3) S + ts+ sw 67 08
4 (O4) X + tx + sw 70 81
Time: 9:05 AM
SOP 4 Page 17 of 18
December 2003
Use equation 3.1.2 for optional sequence B (XSSX) with NO buoyancy corrections4:
⎡ ( O1 - O2 ) + ( O4 - O3 ) ⎤ ⎡ CM sw ⎤
C x = C s + CM t s - CM t x + ⎢ ⎥ ⎢ O - O ⎥ + Ns - Nx
⎣ 2 ⎦ ⎣ 3 2 ⎦
Cx = 1.018 236 mg
U = uc * 2
2 2 2
uc = u s + s p + uo
The uncertainty for the standard, U, must be divided by the k factor for the standard and the tare
weight to determine each us. The additional uncertainty for not performing the air buoyancy
correction can be determined using the magnitude of the air buoyancy correction from SOP 2.
uc = 0.019 617 15 mg
Cx = 1.018 mg ± 0.039 mg
OR
4 Keep in mind that these equations may be truncated for the purpose of this example and minor differences
may be seen in the ending decimal places due to the use of calculators or spreadsheets.
SOP 4 Page 18 of 18
December 2003
SOP No. 7
1.0 Introduction
1.1. Purpose
1.2. Prerequisites
1.2.1. Verify that valid calibration certificates are available for the standards
used in the test.
1.2.2. Verify that the standards to be used have sufficiently small standard
uncertainties for the intended level of calibration. Primary standards
should not be used at this level.
1.2.3. Verify that the balance that is used is in good operating condition with
sufficiently small process standard deviation as verified by a valid control
chart or preliminary experiments to ascertain its performance quality when
a new balance is put into service.
1.2.5. Verify that the laboratory facilities meet the following minimum
conditions to meet the expected uncertainty possible with this procedure.
SOP 7 Page 1 of 15
December 2003
2. Methodology
2.2. Summary
The single substitution procedure is the same for all of the balances mentioned
above, but the adjustment of the balance to prepare for the intercomparison and
the selection of the sensitivity weight differ slightly depending upon the balance
used. When steps specific to a particular balance are required, they are given in
subsections of the procedure identified by a, b, and c along with the balance type.
2.3.1. Precision balance with sufficient capacity and sensitivity for the
calibrations planned.
2.3.3. Calibrated small standard weights with recent calibration values that are
traceable to NIST to be used as tare weights.
2.3.5. Forceps to handle the weights, or gloves to be worn if the weights are
moved by hand.
SOP 7 Page 2 of 15
December 2003
2.3.6. Stop watch or other timing device to observe the time of each
measurement.
2.4. Symbols
1 The barometer, thermometer, and hygrometer are used to determine the air density at the time of
the measurement. The air density is used to make an air buoyancy correction. The accuracies
specified are recommended for high precision calibration. Less accurate equipment can be used
with only a small degradation in the overall accuracy of the measurement.
SOP 7 Page 3 of 15
December 2003
2.5. Procedure
2.5.1.1. Place the test weight and standards in the balance chamber or
near the balance overnight to permit the weights and the balance
to attain thermal equilibrium.
1
a. Single-pan mechanical balance - less than the range of
10
the optical scale.
1
c. Combination balance - less than the range of the digital
10
indications.
SOP 7 Page 4 of 15
December 2003
1 S + ts O1
2 X + tx O2
3 X + tx + sw O3
SOP 7 Page 5 of 15
December 2003
1 X + tx O1
2 S + ts O2
3 S + ts + sw O3
SOP 7 Page 6 of 15
December 2003
3. Calculations
3.1. No air buoyancy correction. Calculate the conventional mass correction, Cx, for
the test weight as follows, according to the optional sequence used. In each case,
the conventional mass corrections for the standard weight(s), Cs, the conventional
mass of the tare weights, CM t s and CM t x , and the conventional mass of the
sensitivity weight, CMsw, are included. The symbols Ns and Nx refer to the
nominal values of S and X, respectively. If no tare weights, and equal nominal
values are used, those terms may all be deleted from the equations.
.
3.1.1. Optional Sequence A (SXX)
⎡ CM sw ⎤
C x = Cs + CM t s - CM t x + ( O2 - O1 ) ⎢ + Ns - N x
⎣ ( O3 - O2 ) ⎥⎦
⎡ CM sw ⎤
C x = Cs + CM t s - CM t x + ( O1 - O2 ) ⎢ + Ns - N x
⎣ ( O3 - O2 ) ⎥⎦
3.2. Air Buoyancy Correction
3.2.1. Calculate the air density, ρa, as described in the Appendix to SOP No. 2.
3.2.2. Calculate the mass of the test weight, Mx, and its mass correction Cx using
the mass of the standard weight(s), the tare weights and the sensitivity
weights according to the optional sequence used.
⎡ ⎛ ρa ⎞⎤
⎢ M sw ⎜⎜ 1 - ⎟⎟ ⎥
⎛ ρ ⎞ ⎛ ρ ⎞ ⎛ ρ ⎞ ⎝ ρsw ⎠⎥
M s ⎜⎜ 1 - a ⎟⎟ + M t s ⎜⎜ 1 - a ⎟⎟ - M t x ⎜⎜ 1 - a ⎟⎟ + ( O2 - O1 ) ⎢
⎝ ρs ⎠ ⎝ ρt s ⎠ ⎝ ρt x ⎠ ⎢ ( O3 - O2 ) ⎥
⎢ ⎥
Mx = ⎣⎢ ⎦⎥
⎛ ρ ⎞
⎜⎜ 1 - a ⎟⎟
⎝ ρx ⎠
SOP 7 Page 7 of 15
December 2003
⎡ ⎛ ρa ⎞ ⎤
⎢ M sw ⎜⎜ 1 - ⎟⎟ ⎥
⎛ ρ ⎞ ⎛ ρ ⎞ ⎛ ρ ⎞ ⎝ ρ sw ⎠ ⎥
M s ⎜⎜ 1 - a ⎟⎟ + M ⎜⎜ 1 - a ⎟⎟ - M ⎜⎜ 1 - a ⎟⎟ + ( O1 - O2 ) ⎢
( O3 - O2 )
ts tx
⎝ ρs ⎠ ⎝ ρt s ⎠ ⎝ ρt x ⎠ ⎢ ⎥
⎢ ⎥
Mx = ⎣⎢ ⎦⎥
⎛ ρ ⎞
⎜⎜ 1 - a ⎟⎟
⎝ ρx ⎠
Cx = Mx - Nx
where Nx is the nominal value for X.
⎛ ρ ⎞
M x ⎜⎜ 1 - n ⎟⎟
⎝ ρx ⎠
CM x =
⎛ ρ ⎞
⎜1 - n ⎟
⎝ 8.0 ⎠
3.2.5. If requested, the apparent mass versus the reference density of brass may
be calculated. This value should only be used when calibrating
mechanical balances that have been adjusted to this reference density.
2 Conventional Mass: “The conventional value of the result of weighing a body in air is equal to the mass of a
standard, of conventionally chosen density, at a conventionally chosen temperature, which balances this body at this
reference temperature in air of conventionally chosen density.” The conventions are: reference density 8.0 g/cm3;
reference temperature 20 °C; normal air density 0.0012 g/cm3. Conventional mass was formerly called “Apparent
Mass versus 8.0 g/cm3” in the United States. See OIML IR 33 (1973, 1979), under revision.
SOP 7 Page 8 of 15
December 2003
⎛ ρ ⎞
M x ⎜⎜ 1 - n ⎟⎟
⎝ ρx ⎠
AM x vs brass =
⎛ ρn ⎞
⎜1 - ⎟
⎝ 8.3909 ⎠
4. Measurement Assurance
4.1. Duplicate the process with a suitable check standard (See GLP 1, SOP 9, SOP 30,
and Sec. 7.4)
4.2. Plot the check standard value and verify that it is within established limits; a t-test
may be incorporated to check observed value against accepted value.
4.3. The mean of the check standard is used to evaluate bias and drift over time.
4.4. Check standard observations are used to calculate the standard deviation of the
measurement process, sp.
5. Assignment of Uncertainty
SOP 7 Page 9 of 15
December 2003
6. Report
SOP 7 Page 10 of 15
December 2003
Appendix
Single Substitution Data Sheet
(Optional Sequence A)
SXX
Laboratory data and conditions:
Operator
Date Temperature
Balance Pressure
Load Relative Humidity
Standard deviation of the process, from control chart, sp
Mass Correction = True Mass if using buoyancy correction. Mass Correction = Conventional Mass if
NOT using buoyancy correction. Density is used only with buoyancy corrections.
Observations:
Measurement # Weights Balance Observations, Units _________
Time:
1 (O1) S + ts
2 (O2) X + tx
3 (O3) X + tx + sw
Time:
SOP 7 Page 11 of 15
December 2003
Appendix
Single Substitution Data Sheet
(Optional Sequence B)
XSS
Laboratory data and conditions:
Operator
Date Temperature
Balance Pressure
Load Relative Humidity
Standard deviation of the process, from control chart, sp
Mass Correction = True Mass if using buoyancy correction. Mass Correction = Conventional Mass if
NOT using buoyancy correction. Density is used only with buoyancy corrections.
Observations:
Measurement # Weights Balance Observations, Units _________
Time:
1 (O1) X + tx
2 (O2) S + ts
3 (O3) S + tS + sw
Time:
SOP 7 Page 12 of 15
December 2003
Mass Correction = True Mass if using buoyancy correction. Mass Correction = Conventional Mass if
NOT using buoyancy correction. Density is used only with buoyancy corrections.
Observations:
Measurement # Weights Balance Observations, Units _____mg____
Time: 3:40 PM
1 (O1) S + ts 12 62
2 (O2) X + tx 12 51
3 (O3) X + tx + sw 62 37
Time: 3:45 PM
SOP 7 Page 13 of 15
December 2003
Since no tare weights were used and equal nominal values were used, the equation may be
simplifies as follows:
⎡ CM sw ⎤
Cx = Cs + ( O2 - O1 ) ⎢ ⎥
⎣ ( O3 - O2 ) ⎦
⎡ 49.916 mg ⎤
C x = 0.255 mg + ( 12.51 - 12.62 ) ⎢ ⎥
⎣ ( 62.37 - 12.51 ) ⎦
Cx = 0.144 876 mg
Calculation of Uncertainty
U = uc * 2
uc = u s + s p + uo
2 2 2
The uncertainty for the standard from the calibration report (or data sheet) must be divided by
the k factor to determine the us. Refer to SOP 29 for the use of multiple standards. An additional
uncertainty for not performing the air buoyancy correction can be determined using the
magnitude of the air buoyancy correction from SOP 2.
uc = 0.021 330 7 mg
U = 0.021 330 7 * 2 = 0.042 661 5 mg
Uncertainty Statement
The uncertainty reported is the root sum square of the standard uncertainty of the standard, the
standard deviation of the process, and an uncorrected systematic error for lack of buoyancy
corrections, multiplied by a coverage factor of 2 (k=2) for an approximate 95 % confidence
interval. Factors not considered in the evaluation: magnetism (weights are considered to meet
magnetism specifications unless measurement aberrations are noted), balance eccentricity and
3 Keep in mind that these equations may be truncated for the purpose of this example and minor differences
may be seen in the ending decimal places due to the use of calculators or spreadsheets.
SOP 7 Page 14 of 15
December 2003
linearity (these factors are considered as a part of the measurement process when obtaining the
standard deviation of the process).
Compliance Evaluation
You may need to evaluate the conventional mass correction and its uncertainty to determine if a
weight is or is not within specified tolerances. The magnitude of the expanded uncertainty must
be less than 1/3 of the tolerance to be able to perform that evaluation, according to ASTM
E 617-97 and OIML R111 documentary standards.
Load = 50 g
ASTM E 617 OIML R111
Class Tolerance (mg) Class Tolerance (mg)
1 0.12 E2 0.10
2 0.25 F1 0.30
If we look at three times the calculated expanded uncertainty: 0.043 mg x 3 = 0.129 mg, we
observe that the uncertainty complies with the 1/3 rule for ASTM Class 2, and OIML Class F1
but not ASTM Class 1 or OIML Class E2.
Next, the weight value is considered to be within tolerance when the absolute value of its error
plus its uncertainty, do not exceed the tolerance established for the particular class of weight.
We can see that the correction of 0.145 is within these limits: 0.102 ≤ Cx ≤ 0.188, therefore, the
value also complies with ASTM Class 2 and OIML Class F1, and a compliance statement may be
included on the calibration report.
Reporting
Cx = 0.145 mg ± 0.043 mg
SOP 7 Page 15 of 15
December 2003
SOP No. 8
1. Introduction
1.1. This SOP describes procedures to be followed for determining whether or not
mass standards are within the tolerances specified for a particular class of
standards (e.g., NIST Class F, ASTM Class 5, 6, 7 or OIML Classes M1, M2, and
M3) where the uncertainty is much smaller than the tolerance application. The
described procedures permit the metrologist to report that the weights under test
were compared against a reference standard with the results reported on the
laboratory report form. The comparison is important because the built-in weights
of a balance do not represent laboratory standards unless they have been formally
calibrated. If a State law or other regulation requires that field weights be
compared against the State (or reference) standards, this procedure can be used to
fulfill this requirement. It is suitable for calibration when moderate accuracy is
required, and does not eliminate errors due to drift. The procedure does not
incorporate measurement control steps to ensure the validity of the standards and
the measurement process; additional precautions must be taken. The expanded
uncertainty with this procedure must be ≤ 1/3 of the tolerance per ASTM and
OIML standards. If uncertainty to tolerance ratios are greater than recommended,
SOP 7 (single substitution) or SOP 4 (double substitution) is preferred.
Note: If you use SOP 8, you are most likely using working standards that are
equivalent to your customer’s weights. If this is the case, the working standards
should be at a higher level of calibration than the weights being calibrated with
them. Therefore, you will need a better balance and procedure to calibrate your
working standards than the one used to calibrate your customer’s weights.
1.2. Prerequisites
1.2.1. Verify that valid calibration values are available for the standards used in
the test.
1.2.2. Verify that the working standards to be used have sufficiently small
standard uncertainties for the intended level of calibration. Primary
standards should not be used at this level.
1.2.3. Verify that the balance that used is in good operating condition with
sufficiently small process standard deviation as verified by a valid control
chart or preliminary experiments to ascertain the performance quality
when a new balance is put into service.
1.2.5. Verify that the laboratory facilities meet the following minimum
conditions to meet the expected uncertainty possible with this procedure.
SOP 8 Page 1 of 12
December 2003
2. Methodology
This method is applicable to all lower echelon mass calibration (tolerance testing)
provided that the uncertainty requirements can be met. The achievable precision
using this procedure is appropriate, provided the expanded uncertainty of the
measurement is no more than one-third of the permissible tolerance of the mass
standard tested. The accuracy achievable with this procedure depends on the
accuracy of the calibration of the working standards and the precision of the
intercomparison.
2.2. Summary
2.3. Apparatus/Equipment
2.4.1. Select a reference standard of the same nominal value as the weight under
test. Place the standard on the balance pan. Adjust the optical scale
reading (See GMP No. 4) to approximately midscale using uncalibrated
SOP 8 Page 2 of 12
December 2003
tare weights and the balance’s coarse and fine Zero controls. This setting
must not be altered during a measurement sequence. Record the reading
as O1.
2.4.3. Calculate the value of a scale division using the equation in 3.2. If it is
within ± 2 % of nominal value (usual case) the nominal value of a division
can be used for tolerance testing.
2.4.4. Remove the sensitivity weight and adjust the optical scale to account for
corrected value of standard used.
Example: Suppose that the nominal range of the optical scale is 100
mg and that the reference standard has a correction of -2.5 mg. The optical
scale is adjusted to read 47.5 mg when the standard is on the pan. Under
this condition, the reading 50.0 mg represents the nominal mass of the
reference standard.
2.4.6. Place the weight to be tested on the balance pan, read the optical scale and
record reading as Xn. The error in the weight is the amount by which the
indication deviates from the mid-scale reading. If the weight indication is
more than the mid-scale value, the weight is heavy by the indicated
difference; if the indication is less than the mid-scale value, the weight is
light.
2.4.7. After several weights have been tested (no more than 10), put the standard
on the balance pan and record the reading. The difference between this
indication and the previous one for the standard indicates a balance drift.
This drift will normally be very small. If the drift exceeds 10 % of the
tolerance applicable to the weights under test or affects a measurement
result to the extent that a weight may be out of tolerance, the measurement
should be repeated and more frequent checks of the standard should be
made or a more appropriate procedure should be used.
2.4.8. Readjust the optical scale at any time that a significant difference is
observed when rechecking a standard.
2.4.9. Calculate the mass correction for the unknown weights using the
appropriate equation in Section 3.
2.5.1. Select a reference standard of the same nominal value as the weight under
test. Place the standard on the balance pan. Adjust the optical scale
reading (See GMP No. 4) to midscale using uncalibrated tare weights and
the balance’s coarse and fine Zero controls. This setting must not be
altered during a measurement sequence. Record the reading as O1.
SOP 8 Page 3 of 12
December 2003
2.5.3. Calculate the value of a scale division using the equation in 3.2. If the
sensitivity is within ± 2 % of nominal value (usual case) of the scale
division, the nominal value of a division may be used.
2.5.4. Remove the sensitivity weight and re-adjust the optical scale to obtain a
midscale indication, if the indication has changed from that set in 2.5.1.
2.5.6. Place the weight to be tested on the balance pan, read the optical scale and
record the indication as Xn. The error in the weight is the amount by
which the indication deviates from the mid-scale reading. If the weight
indication is more than the mid-scale value, the weight is heavier than the
standard by the indicated difference; if the indication is less than the mid-
scale value, the weight is lighter than the standard.
2.5.7. After several weights have been tested (no more than 10 without
rechecking the standard) put the standard on the balance pan and record
the reading. The difference between this indication and the previous one
for the standard indicates a balance drift. This drift will normally be very
small. If the drift exceeds 10 % of the tolerance applicable to the weights
under test or affects a measurement result to the extent that a weight may
be out of tolerance, the measurement should be repeated and more
frequent checks of the standard should be made or a more appropriate
procedure should be used.
2.5.8. Readjust the optical scale at any time that a significant difference is
observed when rechecking a standard.
Cx = Cs + (Xn-O1).
2.6.1. Select a reference standard of the same nominal value as the weight under
test. Place the standard on the pan. Zero the balance and record reading
as O1.
2.6.2. Add a calibrated sensitivity weight (sw ≥ 2 times the tolerance but not
exceeding 1 % of the balance capacity) and record the reading as O2.
Verify whether the nominal scale division is within ± 2 % of nominal
value of the scale division using the equation in 3.2. If so, the nominal
value of the scale division may be used.
2.6.3. Remove sensitivity weight and zero the balance so weight differences, d,
can be read directly from the balance indications.
SOP 8 Page 4 of 12
December 2003
2.6.5. Place the weight to be tested on the balance pan. Record the reading as
Xn.
2.6.6. After several weights have been tested (no more than 10 without
rechecking the standard) recheck the zero as in 2.5.3 and record the
reading. The difference between this indication and the previous one for
the standard indicates a balance drift. This drift will normally be small. If
the drift exceeds 10 % of the tolerance applicable to the weights under test
or affects a measurement result to the extent that a weight may be out of
tolerance, the measurement should be repeated and more frequent checks
of the standard should be made or a more appropriate procedure should be
used.
2.6.7. Readjust the zero at any time that a significant difference is observed
when rechecking a standard.
2.6.8. Calculate the mass correction for each weight using the equation
2.7.1. Select a reference standard of the same nominal value as the weight under
test. Place the standard on the left balance pan together with small,
calibrated weights equal to the correction required for the standard,
provided it is light. If (and only if) the standard is heavy, do nothing
further at this point but follow instructions in 2.6.4. Add sufficient
counterweights to the right pan to obtain a sum of turning points of
approximately twice midscale value. If necessary, number the graduated
scale such that adding weights to the left pan will increase the balance
reading. Record the sum of the turning points as 01.
2.7.2. Add an appropriate calibrated sensitivity weight to the left pan and record
the sum of the turning points as 02. Calculate the sensitivity,
CM sw
sensitivity =
( O2 - O1 )
where CMsw is the conventional mass of the sensitivity weight.
2.7.4. Place weight to be tested on the left pan. If the standard used in 2.6.1 was
heavy, add small correction weights to the left pan, equivalent to the
correction required for the standard. Add small, calibrated tare weights as
required to left or right pan to obtain an approximate balance and record
the sum of the turning points as Xn.
SOP 8 Page 5 of 12
December 2003
2.7.5.1. If added tare weights are placed on the left pan (tx).
⎡ CM sw ⎤
C x = ( X n - O1 ) ⎢ ⎥ - CM t x
⎣ ( O2 - O1 ) ⎦
2.7.5.2. If added tare weights are placed on the right pan (ts)
⎡ CM sw ⎤
C x = ( X n - O1 ) ⎢ ⎥ + CM t s
⎣ ( O2 - O1 ) ⎦
2.7.6. After several weights have been tested (no more than 10 without
rechecking the standard), recheck the turning point 01, as described in
2.6.1. Only a small difference should be observed. If the difference
exceeds 2 % of the sum of the turning points in O1, the measurement
should be repeated and more frequent checks of the standard should be
made or a more appropriate procedure should be used.
2.8.1. Compare the correction plus the expanded uncertainty of the weight tested
with the tolerance for the class of weights to which it belongs. If the
absolute value of the correction plus the expanded uncertainty is numeri-
cally smaller than the tolerance, the weight is considered to be within
tolerance. If the correction is larger than permissible, the weight is
considered to be outside the tolerance and appropriate action should be
taken. It is recommended that weights whose absolute value of the
correction exceeds 75 % of the tolerance limit be adjusted.
SOP 8 Page 6 of 12
December 2003
3. Calculations
3.1. Air buoyancy corrections are generally not made with the modified substitution.
Calculate the conventional mass correction, Cx, for the test weight as follows,
according to the optional sequence used. In each case, the conventional mass
corrections for the standard weight(s) are included.
CM sw
sensitivity =
( O2 - O1 )
If the sensitivity error is less than 2 % of the nominal value of a division on the
optical scale or the electronic range of operation, proceed with the modified
substitution. If the sensitivity error is greater, SOP 7, Single Substitution may be
acceptable.
3.3. Calculate the mass correction of each unknown weight as follows if the correction
for the standard IS NOT used in setting a reference point on the balance:
Cx = Cs + d
Cx = Cs + ( Xn – O1 )
SOP 8 Page 7 of 12
December 2003
Note: If an electronic balance is used and zeroed with the standard on the
balance, O1 is “0" and d becomes the Xn balance reading.
3.4. Calculate the mass correction of each unknown weight as follows if the correction
for the standard IS used in setting a nominal reference point on the balance:
3.5. If tare weights and unequal nominal values are used, use the following equation
for modifying section 3.3:
Cx = Cs + CMts – CMtx + ( Xn – O1 ) + Ns – Nx
4. Measurement Assurance
4.1. Duplicate the process with a suitable check standard (See GLP 1, SOP 9, SOP 30,
and Sec. 7.4)
4.2. Plot the check standard value and verify that it is within established limits; a t-test
may be incorporated to check observed value against accepted value.
4.3. The mean of the check standard is used to evaluate bias and drift over time.
4.4. Check standard observations are used to calculate the standard deviation of the
measurement process, sp.
5. Assignment of Uncertainty
5.1. The expanded uncertainty for the standard, U, is obtained from the calibration
report. The combined standard uncertainty, uc, is used and not the expanded
uncertainty, U, therefore the reported uncertainty for the standard will usually
need to be divided by the coverage factor k.
5.2. Standard deviation of the measurement process from control chart performance
(See SOP No. 9.) The value for sp is obtained from the control chart data for
check standards using modified substitution measurements.
5.3. Other standard uncertainties usually included at this calibration level include
uncertainties associated with lack of performing buoyancy corrections and can be
determined using the equation given in SOP 2 for the magnitude of the air
buoyancy correction. Buoyancy corrections are not generally needed at the
uncertainty level this procedure is designed for.
SOP 8 Page 8 of 12
December 2003
5.4. The expanded uncertainty, U, must be ≤ 1/3 of the tolerance applicable as per
ASTM E617-97 and OIML R111.
6. Report
SOP 8 Page 9 of 12
December 2003
Appendix
Modified Substitution Data Sheet
Laboratory data and conditions:
Operator
Date Temperature
Balance Pressure
sw
tx
ts
*Mass Correction = True Mass if using buoyancy correction. Mass Correction = Conventional Mass if NOT using
buoyancy correction. Density is used only with buoyancy corrections.
Observations:
Observation No. Weights Balance Observations, Units
Time:
1 (O1) S
2 (O2) S + sw
3 (O3) S
4 (O4) X1
5 (O5) X2
6 (O6) X3
7 (O7) Sc
8 (O8) S
SOP 8 Page 10 of 12
December 2003
Appendix
Modified Substitution Data Sheet
Laboratory data and conditions:
Operator GH
X 5 lb 7.85
sw 5g 0.000 002 g
tx None
ts None
Observations:
Observation No. Weights Balance Observations, Units 0.001 g
Time: 11:00 am
1 (O1) S 0000
2 (O2) S + sw 4999
3 (O3) S 0000
4 (O4) X1 0205
5 (O5) X2 0104
6 (O6) X3 - 0089
7 (O7) Sc - 0005
8 (O8) S 0003
SOP 8 Page 11 of 12
December 2003
Sensitivity guidelines indicate a sensitivity weight, greater than or equal to 2 times the tolerance
(460 mg) but not exceeding 1 % of the balance capacity (23 g). For a 5 lb load, the Class F
tolerance is 230 mg so a 5 g weight was selected.
A 2 % sensitivity error in a 5 gram range would allow readings between 4.900 g and 5.100 g in
the sensitivity test. Calculate the sensitivity:
5.000 002 g
sensitivity = = 1.000 200 4 g/div
4.999 div
The drift observed was 3 mg and is less than 1/10 of the tolerance and 23 mg could be allowed,
so the error for each nominal weight may be calculated as follows.
C x1 = 7.5 mg + ( 205 mg - 0 ) = 212.5 mg
C x 2 = 7.5 mg + ( 104 mg - 0 ) = 111.5 mg
C x 3 = 7.5 mg + ( - 89 mg - 0 ) = - 81.5 mg
C Sc = 7.5 mg + ( - 5 mg - 0) = 2.5 mg
uc = u s + s p + uo
2 2 2
The uncertainty for the standard must be divided by the k factor for the standard and the tare
weight to determine each us. The additional uncertainty for not performing the air buoyancy
correction can be determined using the magnitude of the air buoyancy correction from SOP 2.
Since the working standard was previously calibrated with buoyancy corrections and is the same
density as the unknown weights, the uncertainty for the buoyancy correction drops out of the
equation.
uc = ( 0.167 )2 + ( 0.85 )2
uc = 0.866 19 mg
U = 0.866 19 * 2 = 1.732 38 mg
The expanded uncertainty is less than 1/3 of the tolerance (76 mg) so the unknown weights can
either be reported as “within tolerance” or the values and uncertainties can be reported. The
value plus the uncertainty for the first unknown X1 is within 25 % of the Class F tolerance,
therefore it is desirable to adjust the weight closer to nominal value, although based on the
tolerances and uncertainties, and it can be claimed as “within tolerance.”
SOP 8 Page 12 of 12
December 2003
SOP No. 9
1. Introduction
1.1. This SOP describes procedures for the development of control charts and their use
for demonstrating attainment of statistical control of a mass calibration process.
1.2. Prerequisites
The use of this SOP requires that appropriate apparatus, methodology, and standards
are available, and that the laboratory thoroughly understands the basic principles of
the measurement process used and has had sufficient experience to perform the
necessary operations required for the measurements of concern.
2. Summary
An appropriate check standard (or control standard) is incorporated into the measurement
process and weighed at established intervals, the results are plotted on an x (X-bar) chart.
The abscissa (x) represents the sequence of measurements and the ordinate (y) the measured
values. A central line is drawn, indicating the mean ( x ) of the measured values and control
limits are indicated within which the results of measurements are expected to be randomly
distributed, based on statistical considerations. The system is considered to be in statistical
control when the individual values are within the designated limits. The system is
considered to be out of control if an excessive number of values are present outside
established limits, unusual trends are observed, or if the mean exceeds the control limits.
The statistical information on which the control limits are based can be used to calculate
confidence limits for measurements made while the system is demonstrated to be stable and
in a state of statistical control.
3. Procedure
The monitored system is considered to consist of the balance, the standard operating
procedure, the laboratory environment, the check standard or control standard, the
operator, and any other sources that contribute to the variance or bias of the
measurement data. Any of the above that can be considered to be constant or
negligible contributors to the variance may be consolidated and monitored by a
single control chart. Any that cannot be so considered (for example: different
standard, different balance, different SOP) may require separate control charts.
SOP 9 Page 1 of 5
December 2003
The variability of balance precision that is load dependent must be considered. For
many balances, precision is a function of load, and a distinct control chart is required
(in principle) for every load tested. This is not always feasible, except in the case of
SOP 5 or 28 where check standards are incorporated into the measurement process.
Hence, control charts used for measurement assurance and evaluation of
measurement uncertainty are generally satisfactory if developed using data from
check standards at two or three intervals for each balance appropriately spaced
within the range of balance use, or at least with one check standard for each decade.
On balances where few nominal values (loads) are tested, a control chart should be
established for each load.
3.3.1. The control chart parameters consist of the central line, the best estimate of
the mean of measurements of the check standard, and control (or “action”)
SOP 9 Page 2 of 5
December 2003
and warning limits that represent probabilistic limits for the distribution of
results around the central line. These parameters are evaluated on the basis
of a reasonable number of initial measurements and updated as additional
measurement data are accumulated.
Calculate the mean, x and the estimate of the standard deviation, s in the
conventional manner.
Central Line = x
Upper Control/Action Limit = x +3s
Upper Warning Limit = x +2s
Lower Warning Limit = x -2s
Lower Control/Action Limit = x -3s
Control chart parameters for Echelon III (Class F or other) may be completed
as follows to track practical limits:
Central Line = x
Upper Control/Action Limit = x + 1/4 tolerance
Upper Warning Limit = x + 1/10 tolerance
Lower Warning Limit = x - 1/10 tolerance
Lower Control/Action Limit = x - 1/4 tolerance
SOP 9 Page 3 of 5
December 2003
Note: Ordinarily, upgrading is merited when the amount of new data is equal
to that already used to establish the parameters in use, or when at least seven
additional data points have been recorded.
Calculate x and s for the new set of data and examine for significant
differences from the former using the t-test and F-test, respectively. If the
tests fail and results are significantly different, determine the reason for the
difference, if possible, and decide whether corrective action is required. If
data does not agree within statistical limits, establish new parameters using
the most recent data and note the reasons for not using previous data. If no
significant differences between the data sets are found, pool all data and
calculate new control chart parameters based on all existing data.
The check standard should be measured and plotted with sufficient frequency to
minimize the risk of loss of data during the period from last-known-in to first-
known-out of control condition. It is good practice to measure the check standard at
least once during each period when a set of test measurements is made. For critical
calibrations or those of highest accuracy, it is desirable to alternate measurements of
test items and check standards, but for real-time evaluation it is preferable to
incorporate the check standard in the calibration design as in SOP 5 or SOP 28.
Whenever there has been a long period of inactivity, it is good practice to make a
series of measurements of the check standard and to plot the results on a control chart
to demonstrate attainment of statistical control prior to resuming measurements with
that specific calibration system.
4.1.1. If plotted points are stable and randomly distributed within the warning
limits, decide that the system is in control.
SOP 9 Page 4 of 5
December 2003
4.1.2. If a plotted point is outside the warning limits but within the control limits,
investigate the presence of calculation errors. If none were made, re-measure
the check standard. The re-measured value must be within the warning limits
to merit the decision of "in control". If the results are not within limits,
consider the measurement process "out of control". Reject all data obtained
since last "in-control" measurement and take corrective action (hence
“action” limit). Accept no further data until the system is demonstrated to be
in-control as indicated by at least two successive measurements of the check
standard within the warning limits.
If a plotted point is outside the control limits and arithmetically correct, the
system is considered to be out of control. Data are rejected, corrective
actions must be taken and re-attainment of statistical control demonstrated, as
above, before data may be accepted.
4.2.1. Absence of a significant difference between the central line and the accepted
value for the check standard may be considered as evidence of insignificant
bias at the level of confidence of the statistical test used. This conclusion is
valid, as long as the system remains in control. On occasion, small
differences (less than 1 s) from unknown sources will be obvious over time
and the value observed for the bias should be incorporated into the
uncertainty per SOP 29.
4.2.2. The estimate of the standard deviation of the process, sp, used to establish the
control limits may be used to calculate confidence intervals for all pertinent
measurements made while the system is in control. However, see SOP 29 for
calculation of measurement uncertainty using the process variability, sp. The
value of the test weight is said to be within the limits
y ±U
where y represents the mean of the measurements on the test weight and
U = k * u S2 + s P2 + uO2 , with the value of k determined by the confidence level
required for the interval.
SOP 9 Page 5 of 5
March 2003
SOP No. 29
1. Introduction
1.1. Purpose
Laboratories performing calibrations that meet ISO/IEC Guide 25, ISO/IEC
17025, or ANSI/NCSL Z 540-1-1994 must report uncertainties in conformance
with the 1993 ISO Guide to the Expression of Uncertainty in Measurement
(hereafter called the GUM). This SOP provides instruction for the laboratory to
meet this requirement.
1.2. Prerequisites
1.2.1. Calibration certificates with valid uncertainties must be available for all
standards.
1.2.3. Knowledge of the technical basis for the measurement is critical for
completeness in uncertainty evaluation. This can be obtained through
brainstorming, experimentation, interlaboratory comparisons, cause and
effect diagrams and the like. Flow charts at the end of this SOP show a
number of common factors affecting measurements in the areas of mass,
length, and volume.
2. Methodology
SOP 29 Page 1 of 11
March 2003
2.2. Summary
Clearly specify the measurement process in question, including the item being measured
and the input quantities upon which it depends. This will usually require a quantitative
expression related to the process. Where possible, you may reference an SOP or other
method description along with the specific standards and measurement assurance
process that is used to adequately complete this step.
What follows are the most common uncertainties associated with metrological
measurements. Keep in mind that this list is exhaustive. Each item listed below is
identified as a standard uncertainty u, when determined using Type B methods of
evaluation and a standard uncertainty s, when determined with Type A methods of
evaluation (statistical methods). Each standard uncertainty is further defined by an
arbitrary subscript related to the source for ease in remembering that source.
3.2.1. Standard uncertainty from the measurement process, sp, (Type A evaluation).
SOP 29 Page 2 of 11
March 2003
The value for sp is obtained from the control chart limits and the current
knowledge that the measurements are in a state of statistical control.
This will have to be ascertained by measuring at least one check standard
during the course of the current measurements.
The information for the standards comes from the calibration report,
generally reported as an expanded uncertainty with its coverage factor
(k). The expanded uncertainty is simply divided by k to obtain the
combined uncertainty for the standard, uc, which represents the us when
used in your laboratory.
SOP 29 Page 3 of 11
March 2003
These are factors related to the measurement equation, but distinct from the
standard uncertainties associated with the process and the standards. These items
are often much smaller in a well-controlled process than the standard uncertainties
associated with the process and the standards. Examples are given in the flow
charts.
The laboratory must carefully consider any other components of uncertainty that
might contribute to the uncertainty of the measurement. In mass measurements,
these might include effects from the magnitude of air buoyancy corrections as
discussed in SOP 2, when these corrections are not made, or the uncertainty from
the calculation of air density when buoyancy corrections are made (also discussed
in SOP 2).
Additionally, the laboratory should include any other components that are not
negligible, such as the uncertainty associated with the density of the standards, the
density of the unknown test items, effects due to lack of thermal equilibrium
during the test, or known and quantified effects due to magnetized mass standards.
Each component that is considered should be included as an additional standard
uncertainty uo1, uo2, uo3, etc., and included in the RSS equation when data shows
these factors to be significant. Note that components of uncertainty are not
limited to those mentioned. Documentation of each component evaluated should
be maintained to complete the documentation required by ISO/IEC 17025.
3.2.4. Standard uncertainty due to factors unrelated to the measurement process, uu.
These are factors that may be related to characteristics of the items being testing
or of the standard and are usually minimized in well-known and controlled
measurement processes.
3.2.5. Special uncertainties from other sources (Type B evaluations). Includes bias or
unidentified errors.
SOP 29 Page 4 of 11
March 2003
3.2.5.2.2. Evaluate whether or not the bias is less than two times the
combined uncertainty (i.e., bias < 2 u s2 + s2p + u o2 ).
3.2.5.2.3. If the bias is less than twice the combined uncertainty, it may
be included in the uncertainty using one of several
approaches that must clearly be communicated in the report.
3.2.5.2.4. If the bias is larger than twice the combined uncertainty, the
error must be investigated further prior to providing
calibration data.
SOP 29 Page 5 of 11
March 2003
3.2.5.3. Adding the bias to the expanded uncertainty (e.g., used in PMAP
software). In this case, the bias is simply added to the expanded
U + bias = (uc * k ) + bias
uncertainty and is reported as such.
3.2.5.4. Example when uncertainties for the laboratory data and the reference
data are considered equivalent (e.g., laboratory data is compared to
another laboratory’s data).
bias 1
u = or more simply : 0.29 d, where d is the bias
d 2 3
3.2.5.5. Example when uncertainties for the laboratory data are considered
secondary to a reference value (e.g., the difference between the
laboratory data and NIST data). This method was originally published in
1994.
bias
u = or more simply : 0.577 d, where d is the bias
d 3
All uncertainty estimates identified in the previous step must be quantified in units that
represent the measured values. Type A methods of evaluation usually provide
quantified estimates in the units of interest.
Type B methods of evaluation may be conducted with spreadsheets using the basic
expression identified in the SOP or identified when the process was specified. Scenario
testing can be done to determine the impact and quantify specific variables on the final
measured quantity. The knowledge gained in this step often proves useful in identifying
potential areas of improvement.
SOP 29 Page 6 of 11
March 2003
In those cases where the uncertainty factors were determined statistically (Type A
methods), the standard deviation is used to represent the standard uncertainty. In other
cases, estimates must be made to ensure that the quantified uncertainties represent “one-
standard-deviation” values or a k=1 coverage level. The appropriate distribution factor
must be used when converting values.
The combined standard uncertainty, uc, includes the standard uncertainty reported for
the standards used, us, the standard uncertainty of the measurement process, sp, the
standard uncertainty from other sources, uo, which includes all other factors the
laboratory considers significant, the standard uncertainty due to factors related to the
measured item but unrelated to the measurement process, uu, and finally, the standard
uncertainty due to bias or differences, ud, when ud is included. The standard
uncertainties are combined using the root-sum-of-the-squares (RSS) method as follows:
2
u c = s p + us + uo + uu + ud
2 2 2 2
SOP 29 Page 7 of 11
March 2003
Once the uncertainty has been calculated, a statement such as the following is reported:
The combined standard uncertainty includes the standard uncertainty reported for the
standard, for the measurement process, and for any observed deviations from reference
(e.g., NIST) values, which are less than surveillance limits. The combined standard
uncertainty is multiplied by k, a coverage factor of (2, 3) to give the expanded
uncertainty (which defines an interval with an approximate (95, 99) % level of
confidence).
SOP 29 Page 8 of 11
March 2003
Uncertaint
Flow Chart for
Expanded Calibration
u c x k (usually =
uc
Combined Standard Uncertainty of
ud
Standard Uncertainty Due
us
sp Minimal Systematic
Standard Uncertainty uo
Standard Uncertainty of
with Calibration, Check Standard Uncertainty uo
and with Factors Unrelated to
- standard deviation of - drift of the
measuremen - deviation between
- restraint (calibration - equipment uncertainty - Round robin data
- between-time standard values in lab and reported
- temperature, pressure, reproducibilit
- equipment on calibration
- thermal - thermal - "Handbook" values on
calibration, linearity, - discrepancies observed
- - air (e.g. coefficient of
weight interlaboratory
- surface - expansion, air density
-
- - pressure
- NOTE: If values do not
-
criteria given, these factors
be
SOP 29 Page 9 of 11
March 2003
Uncertaint
Flow Chart for Length
Expanded Calibration
uc x k (usually =
uc
Combined Standard Uncertainty of
us
ud
Standard
Standard Uncertainty Due
Associated with
Minimal Systematic
Check Standards and
sp
Unknowns
Standard Uncertainty uo uo
Process Standard - drift of the Factors Unrelated to
- restraint (calibration report)
Associated with - deviation between Measuremen
- uncertainty of optical scale
- standard deviation of the values in lab and reported
graph paper
measurement (typically - equipment uncertainty values on calibration reports - Round robin data showing
- uncertainty of applied
range temperatur - discrepancies observed on reproducibilit
- thermal
- lighting - thermal equilibrium interlaboratory - "Handbook" values on
-
- correction - bias due to reading the standards (e.g. coefficient
- surface
- microscope or graph thermal
-
- hysteresis of
NOTE: If values do not
- readability of
criteria given, these
-condition of tape
must be
SOP 29 Page 10 of 11
March 2003
Uncertaint
Flow Chart for
Expanded Calibration
uc x k (usually =
uc
Combined Standard Uncertainty of
ud
Standard Uncertainty Due
Minimal Systematic
us
sp uo
Standard Uncertainty uo
Standard Uncertainty of Standard Uncertainty
with Calibration, Check Factors Unrelated to
with - drift of the
and
- standard deviation of - deviation between
- Round robin data
measurement (from range or - equipment uncertainty values in lab and reported
- restraint (calibration reproducibilit
standard temperature, pressure, on calibration
- thermal - "Handbook" values on
- between-time standard - thermal - discrepancies observed
- (e.g. coefficient of
(from check - air interlaboratory
- surface expansion, air density
- - - bias due to reading the
- water density
- pressure
NOTE: If values do not
criteria given, these factors
be
SOP 29 Page 11 of 11
December 2003
SOP 30
1. Introduction
1.1. Purpose
The Process Measurement Assurance Program (PMAP) is used for the control
and surveillance of measurement performance using check standards in
measurement and calibration procedures. Incorporation of these measurement
control steps ensures the validity of the measurement process and the standards
used. The variables used in calculation and assignment of uncertainty can be
determined and controlled by the use of this Standard Operating Procedure
(SOP).
1.2. Prerequisites
1.2.1. A Standard Operating Procedure that describes and provides rigor and
consistency in the calibration process.
1.2.2. Check standards that represent the items to be calibrated. These check
standards should be stable and their values should be established with
accuracy, since they will be used to control the uncertainty in the
calibration process. Check standard values are to be determined by
NIST or inside the laboratory with a procedure one level higher than
the calibration process to be controlled.
1.2.3. The operator must be experienced in the calibration process and the
standard operating procedure for the calibration to which this method
is applied. The operator should also have had specific training on
SOP 29 (uncertainty identification and calculations).
SOP 30 Page 1 of 9
December 2003
2. Methodology
2.2. Summary
The check standard is selected to evaluate the standard deviation of the process
(sp), other process uncertainties (uo) such as material density or air density
accuracy, and possible bias (ud) of the process (see SOP 29). Reference
measurements for the check standard are performed after calibrating the
calibration standards and after the servicing of the measurement instrument.
Statistical tools, ‘F’ and ‘t’ tests, determine if there are process changes in
variability and bias (systematic error). These tests are used when process
performance is questionable; when current data is evaluated to establish new
reference control limits; and when evaluating uncertainty statements.
SOP 30 Page 2 of 9
December 2003
2.3.3. A check standard that has been selected to evaluate the established
objectives of: the 1) measurement process uncertainty; 2) the
calibration standards; or 3) both the process and the standards.
2.5. Procedure
SOP 30 Page 3 of 9
December 2003
SOP 30 Page 4 of 9
December 2003
2.5.2.3. Record the measured value (Sc) and plot it on the control chart
and evaluate it with reference to the accepted value of the
check standard. Record date, time, and information tags with
data.
SOP 30 Page 5 of 9
December 2003
u c = ( s p ) 2 + (u s ) 2 + (u o ) 2 + (u d ) 2
U = k * uc
SOP 30 Page 6 of 9
December 2003
SOP 30 Page 7 of 9
December 2003
( s p small ) 2
n new n old
SOP 30 Page 8 of 9
December 2003
These changes can require repeating the PMAP procedure (from 2.5
through 2.5.6).
SOP 30 Page 9 of 9
March 2003
8. STATISTICAL TECHNIQUES
Statistics are used in metrology to summarize experimental data, to provide the basis for
assessing its quality, and to provide a basis for making probabilistic decisions in its use. The
essential basic statistical information for describing a simple data set is:
If the set is a random sample of the population from which it was derived, if the measurement
process is in statistical control, and if all of the observations are independent of one another, then
s is an estimate of the population standard deviation, σ , and x is an unbiased estimate of the
mean, µ.
The population consists of all possible measurements that could have been made under the test
conditions for a stable test sample. In this regard, the metrologist must be aware that any
changes in the measurement system (known or unknown) could possibly result in significant
changes in its operational characteristics, and, hence the values of the mean and standard
deviation. Whenever there is doubt, statistical tests should be made to determine the significance
of any apparent differences before statistics are combined.
The following discussion reviews some useful statistical techniques for interpreting
measurement data. In presenting this information, it is assumed that the reader is already
familiar with basic statistical concepts. For a detailed discussion of the following techniques and
others not presented here, it is recommended that the reader consult NIST Handbook 91 -
Experimental Statistics, by Mary G. Natrella [19]. That handbook also contains comprehensive
statistical tables from which the tables contained in Chapter 9 of this publication were taken.
Mean, x =
( x1 + x 2 + x 3 + ... + x n )
n
∑( xn − x )
2
Standard deviation estimate, s =
n −1
The estimate, s, is based on n-1 degrees of freedom
Given k differences of duplicate measurements, d1, d2, d3, …, dk, a useful formula for estimating
the standard deviation is:
∑ d i2
sd = where sd is based on k degrees of freedom.
2k
Note that d 1 = xi′ − x i′′ , for example.
The values d1, d2 etc., may be differences of duplicate measurements of the same sample (or
object) at various times, or they may be the differences of duplicate measurements of several
similar samples (or objects).
The range, R, is defined as the difference between the largest and smallest values in a set of
measurements.
Mean, R =
(R1 + R2 + R3 + ... + Rk )
k
R
Standard deviation can be estimated by the formula, s R =
d 2*
The value of d 2* will depend on the number of sets of measurements used to calculate s R , and on
the number of measurements in each set, i.e., 2 for duplicates, 3 for triplicates, etc. Consult a
table such as Table 9.1 for the appropriate value of d 2* to use. The effective number of degrees
of freedom for s R is in the table.
Estimates of the standard deviation obtained at several times may be combined (pooled) to
obtain a better estimate based upon more degrees of freedom. The following equation may be
used for this purpose:
sp =
(n1 − 1)s12 + (n2 − 1)s22 + (n3 − 1)s32 + ... + (nk − 1)sk2 where
(n1 − 1) + (n2 − 1) + (n3 − 1) + ... + (nk − 1)
Estimation of the within-series, sw, and between-series, sb standard deviation, (also referred to as
short-term and long-term standard deviations in the applications described here) is an important
way to characterize a measurement process. The former provides guidance as to how many
repetitions of a measurement are required to obtain a result on a single occasion with a given
precision, while the latter is a better estimate of the precision of replication (reproducibility) of a
result on various occasions and is a more realistic evaluation of measurement variability.
To estimate these standard deviations, sets of measurements may be made on several occasions.
To simplify the calculations, each set should consist of the same number of measurements. For
most measurements, it is recommended that duplicate measurements be made on at least 12
separate occasions when estimating sw and sb.
Given k sets of duplicate measurements made on k occasions the following table and calculation
can be made.
Table 8.1
Occasion Measured Values Range Mean
x1′ x1′′
(x1′ + x1′′) = x
1 R1 1
2
x2′ x′2′
(x2′ + x2′′ ) = x
2 R2 2
2
x3′ x3′′
(x3′ + x3′′ ) = x
3 R3 3
2
· · · · ·
· · · · ·
· · · · ·
x′k x′k′
(xk′ + xk′′ ) = x
k Rk k
2
1. Calculate R=
(R1 + R2 + R3 + ... + Rk )
k
R
2. Calculate sw = s R −
d 2*
Note: One may use the procedure of 8.2 to calculate sw if preferred.
3. Calculate sx as follows:
x =
( x1 + x 2 + x 3 + ... + x k )
k
∑ ( xk − x )
2
sx =
k −1
s w2
sb = s x2 −
2
Note that sb is an estimate of the long term component of the standard deviation of a single
measurement. The long term standard deviation of the mean of n measurements taken at a single
occasion is estimated by:
s w2
sb ( x n ) = sb2 −
n
Important note: Do not use this approach for handling within-series and between-time standard
deviations with the Mass Code. Separate formulas are available for that application.
The estimation of the confidence interval for the mean of n measurements is one of the most
frequently used statistical calculations. The formula used will depend on whether the population
standard deviation,σ, is known or whether it is estimated on the basis of measurements of a
sample(s) of the population.
In the usual situation, s is known, based on υ degrees of freedom and the formula for use is:
ts
x ±
n
Variable Description
x sample mean
s estimate of standard deviation
n number of measurements on which the mean is based
t Student's t value, based on the confidence level desired and
the υ degrees of freedom associated with s (see Table 9.3).
Note that t → z as n →∞ . For many practical purposes, the standard deviation may be
considered as known when estimated by at least 30 degrees of freedom.
The interval is asymmetrical because the standard deviation is ordinarily underestimated when
small numbers of measurements are involved due to the fact that large deviations occur
infrequently in a limited measurement process. Indeed, it is the general experience of
metrologists that a few measurements appear to be more precise than they really are.
The basic information required to compute the interval is an estimate of the standard deviation, s,
and the number of degrees of freedom on which the estimate is based. The relationships to use
are:
The values for BL and BU depend upon the confidence level and degrees of freedom associated
with s. Values for use in calculating the confidence level are given in Table 9.7. A more
extensive table (Table A-20) is available in NIST Handbook 91 [19].
Statistical tolerance intervals define the bounds within which a percentage of the population is
expected to lie with a given level of confidence. For example, one may wish to define the limits
within which 95 % of measurements would be expected to lie with a 95 % confidence of being
correct. The interval is symmetrical and is computed using the expression
x ± ks
Variable Description
p the proportion or percentage of the individual measurements to be
included
γ the confidence coefficient to be associated with the interval
n the number of measurements on which the estimate, s, is based
Table 9.6 may be used to obtain values for k for frequently desired values of γ and p. A more
extensive table is Table A-6 found in NIST Handbook 91 [19].
The F-test may be used to decide whether there is sufficient reason to believe that two estimates
of a standard deviation differ significantly. The ratio of the variances (square of the standard
deviation) is calculated and compared with tabulated values. Unless the computed ratio is larger
than the tabulated value, there is no reason to believe that the observed standard deviations are
significantly different.
where sL is the numerically larger value and sS is the smaller value of the two estimates under
consideration.
The critical value of F depends on the significance level chosen for the decision (test) and the
number of degrees of freedom associated with sL and sS, respectively.
Table 9.4 contains critical values for F at the 95 % level of confidence. The tabulated values of
F are not expected to be exceeded with 95 % confidence on the basis of chance alone. As an
example, if both the numerator and the denominator values for s were each based on 9 degrees of
freedom, an F value no larger than 4.03 is expected with 95 % confidence, due to the
uncertainties of the s values, themselves. Table A-5 of NIST Handbook 91 [19] contains values
for F for other confidence levels.
The F-test is useful for comparing the precision of methods, equipment, laboratories, or
metrologists, for example. An inspection of Table 9.4 shows that when either of the values of s
is based on a small number of degrees of freedom, the F value is large. Consequently, the
significance of decisions based on small changes in precision can be supported statistically only
by a relatively large number of measurements. If such changes are suspected, but the data
requirement is difficult to meet, the decision may need to be made on the basis of information
about the measurement process itself.
The F-test is also useful for deciding whether estimates of the standard deviation made at various
times differ significantly. Such questions need to be answered when deciding on whether to
revise control limits of a control chart, for example.
The question may arise as to whether a measured value agrees or significantly disagrees with a
stated value for the measured object. The evaluation can be based on whether or not the
confidence interval for the measured value encompasses the stated value. The confidence
interval is calculated using the expression
ts
x ±
n
as previously described in Section 8.6. In using this expression, n represents the number of
measurements used to calculate the mean, x , and t depends on the degrees of freedom, υ ,
associated with s and the confidence level needed when making the decision. Note that one may
use historical data for estimating s, such as a control chart for example, in which case υ will
represent the degrees of freedom associated with establishment of the control limits and may be
considerably larger than n − 1 .
This discussion is concerned with deciding whether the means of two measured values, A and B,
are in agreement. The data sets used for this purpose may consist of the following:
xA xB
sA sB
nA nB
Case I
Confirming (or assuming) that s A and s B are not significantly different, they are pooled, as
already described (but repeated here for convenience) and used to calculate a confidence interval
for the difference of the means. If this is larger than the observed difference, there is no reason
to believe that the means differ. The steps to follow when making the calculation described
above are:
sp =
(n A − 1)s A2 + (nB − 1)s B2
(n A − 1) + (nB − 1)
sp will be estimated with nA + nB - 2 degrees of freedom
Case II
Confirming (or assuming) that s A and s B are significantly different, their individual values are
used to calculate U ∆ as outlined below.
Conducting operations in random sequences can avert problems of bias that might stem from a
particular order of operations. For example, in the measurement of a series of items, it might be
difficult to determine whether systematic trends in the measured values were due to differences
in the items or to measurement system drift unless the items were measured in random order.
ignore it and proceed until a valid match is encountered. This becomes the first item in the
sequence. Continuing in the same manner, items are selected in the sequence in which their
serial numbers are encountered ignoring the repetition of previously identified items. The
procedure is continued until all items have been randomly selected.
Specimen No.
08 03 09 05 06 02 07 10 04 01
Order
1 2 3 4 5 6 7 8 9 10
9. Reference Tables
Table 9.10 Coefficient of Linear and Cubical Expansion for Materials Commonly Used
in Laboratory Apparatus and Standards
R / d 2* → σ
*
Intermediate values for d 2 and υ may be obtained by interpolation, or from the reference from which this
table was adapted.
Example: If 10 sets of measurements were made and each set consisted of two measurements (duplicates),
*
the value for d 2 is 1.16; if 15 sets of measurements were made and each set consisted of three measurements
*
(triplicates) the value for d 2 is 1.71.
50 0.68
67 1.00
75 1.15
90 1.645
95 1.960
95.28 2.000
99.00 2.575
99.74 3
99.993 4 4
99.999 95 5
100 - 10-9 6
100 - 10-12 7
100 - 10-15 8
100 - 10-18.9 9
100 - 10-23 10
* 80 % 90 % 95 % 98 % 99 % 99.73 %
5 10.01 8.43 7.76 7.39 7.15 6.98 6.85 6.76 6.68 6.62 6.52 6.43 6.33 6.28 6.23 6.18
6 8.81 7.26 6.60 6.23 5.99 5.83 5.70 5.60 5.52 5.46 5.37 5.27 5.17 5.12 5.07 5.01
7 8.07 6.54 5.89 5.52 5.29 5.12 4.99 4.90 4.82 4.76 4.67 4.57 4.47 4.42 4.36 4.31
8 7.57 6.06 5.42 5.05 4.82 4.65 4.53 4.43 4.36 4.30 4.20 4.10 4.00 3.95 3.89 3.84
9 7.21 5.71 5.08 4.72 4.48 4.32 4.20 4.10 4.03 3.96 3.87 3.77 3.67 3.61 3.56 3.51
10 6.94 5.46 4.83 4.47 4.24 4.07 3.95 3.85 3.78 3.72 3.62 3.52 3.42 3.37 3.31 3.26
11 6.72 5.26 4.63 4.28 4.04 3.88 3.76 3.66 3.59 3.53 3.43 3.33 3.23 3.17 3.12 3.06
12 6.55 5.10 4.47 4.12 3.89 3.73 3.61 3.51 3.44 3.37 3.28 3.18 3.07 3.02 2.96 2.91
13 6.41 4.97 4.35 4.00 3.77 3.60 3.48 3.39 3.31 3.25 3.15 3.05 2.95 2.89 2.84 2.78
14 6.30 4.86 4.24 3.89 3.66 3.50 3.38 3.29 3.21 3.15 3.05 2.95 2.84 2.79 2.73 2.67
15 6.20 4.77 4.15 3.80 3.58 3.41 3.29 3.20 3.12 3.06 2.96 2.86 2.76 2.70 2.64 2.59
16 6.12 4.69 4.08 3.73 3.50 3.34 3.22 3.12 3.05 2.99 2.89 2.79 2.68 2.63 2.57 2.51
17 6.04 4.62 4.01 3.66 3.44 3.28 3.16 3.06 2.98 2.92 2.82 2.72 2.62 2.56 2.50 2.44
18 5.98 4.56 3.95 3.61 3.38 3.22 3.10 3.01 2.93 2.87 2.77 2.67 2.56 2.50 2.44 2.38
19 5.92 4.51 3.90 3.56 3.33 3.17 3.05 2.96 2.88 2.82 2.72 2.62 2.51 2.45 2.39 2.33
20 5.87 4.46 3.86 3.51 3.29 3.13 3.01 2.91 2.84 2.77 2.68 2.57 2.46 2.41 2.35 2.29
21 5.83 4.42 3.82 3.48 3.25 3.09 2.97 2.87 2.80 2.73 2.64 2.53 2.42 2.37 2.31 2.25
22 5.79 4.38 3.78 3.44 3.22 3.05 2.93 2.84 2.76 2.70 2.60 2.50 2.39 2.33 2.27 2.21
23 5.75 4.35 3.75 3.41 3.18 3.02 2.90 2.81 2.73 2.67 2.57 2.47 2.36 2.30 2.24 2.18
24 5.72 4.32 3.72 3.38 3.15 2.99 2.87 2.78 2.70 2.64 2.54 2.44 2.33 2.27 2.21 2.15
25 5.69 4.29 3.69 3.35 3.13 2.97 2.85 2.75 2.68 2.61 2.51 2.41 2.30 2.24 2.18 2.12
26 5.66 4.27 3.67 3.33 3.10 2.94 2.82 2.73 2.65 2.59 2.49 2.39 2.28 2.22 2.16 2.09
27 5.63 4.24 3.65 3.31 3.08 2.92 2.80 2.71 2.63 2.57 2.47 2.36 2.25 2.19 2.13 2.07
28 5.61 4.22 3.63 3.29 3.06 2.90 2.78 2.69 2.61 2.55 2.45 2.34 2.23 2.17 2.11 2.05
29 5.59 4.20 3.61 3.27 3.04 2.88 2.76 2.67 2.59 2.53 2.43 2.32 2.21 2.15 2.09 2.03
30 5.57 4.18 3.59 3.25 3.03 2.87 2.75 2.65 2.57 2.51 2.41 2.31 2.20 2.14 2.07 2.01
40 5.42 4.05 3.46 3.13 2.90 2.74 2.62 2.53 2.45 2.39 2.29 2.18 2.07 2.01 1.94 1.88
60 5.29 3.93 3.34 3.01 2.79 2.63 2.51 2.41 2.33 2.27 2.17 2.06 1.94 1.88 1.82 1.74
120 5.15 3.80 3.23 2.89 2.67 2.52 2.39 2.30 2.22 2.16 2.05 1.94 1.82 1.76 1.69 1.61
∞ 5.02 3.69 3.12 2.79 2.57 2.41 2.29 2.19 2.11 2.05 1.94 1.83 1.71 1.64 1.57 1.48
For use for a one-tailed test of equality of standard deviation estimate at 2.5 % level of confidence, or for a
two-tailed test at 5 % level of confidence.
γ = 0.95 γ = 0.99
P 0.75 0.90 0.95 0.99 0.999 0.75 0.90 0.95 0.99 0.999
n
2 22.858 32.019 37.674 48.430 60.573 114.363 160.193 188.491 242.300 303.054
3 5.922 8.380 9.916 12.861 16.208 13.378 18.930 22.401 29.055 36.616
4 3.779 5.369 6.370 8.299 10.502 6.614 9.398 11.150 14.527 18.383
5 3.002 4.275 5.079 6.634 8.415 4.643 6.612 7.855 10.260 13.015
6 2.604 3.712 4.414 5.775 7.337 3.743 5.337 6.345 8.301 10.548
7 2.361 3.369 4.007 5.248 6.676 3.233 4.613 5.488 7.187 9.142
8 2.197 3.136 3.732 4.891 6.226 2.905 4.147 4.936 6.468 8.234
9 2.078 2.967 3.532 4.631 5.899 2.677 3.822 4.550 5.966 7.600
10 1.987 2.839 3.379 4.433 5.649 2.508 3.582 4.265 5.594 7.129
11 1.916 2.737 3.259 4.277 5.452 2.378 3.397 4.045 5.308 6.766
12 1.858 2.655 3.162 4.150 5.291 2.274 3.250 3.870 5.079 6.477
13 1.810 2.587 3.081 4.044 5.158 2.190 3.130 3.727 4.893 6.240
14 1.770 2.529 3.012 3.955 5.045 2.120 3.029 3.608 4.737 6.043
15 1.735 2.480 2.954 3.878 4.949 2.060 2.945 3.507 4.605 5.876
16 1.705 2.437 2.903 3.812 4.865 2.009 2.872 3.421 4.492 5.732
17 1.679 2.400 2.858 3.754 4.791 1.965 2.808 3.345 4.393 5.607
18 1.655 2.366 2.819 3.702 4.725 1.926 2.753 3.279 4.307 5.497
19 1.635 2.337 2.784 3.656 4.667 1.891 2.703 3.221 4.230 5.399
20 1.616 2.310 2.752 3.612 4.614 1.860 2.659 3.168 4.161 5.312
21 1.599 2.286 2.723 3.577 4.567 1.833 2.620 3.121 4.100 5.234
22 1.584 2.264 2.697 3.543 4.523 1.808 2.584 3.078 4.044 5.163
23 1.570 2.244 2.673 3.512 4.484 1.785 2.551 3.040 3.993 5.098
24 1.557 2.225 2.651 3.483 4.447 1.764 2.522 3.004 3.947 5.039
25 1.545 2.208 2.631 3.457 4.413 1.745 2.494 2.972 3.904 4.985
26 1.534 2.193 2.612 3.432 4.382 1.727 2.469 2.941 3.865 4.935
27 1.523 2.178 2.595 3.409 4.353 1.711 2.446 2.914 3.828 4.888
Degrees of
Freedom α = 0.05 α = 0.01 α = 0.001
SU SL SU SL SU SL
Degrees of
α = 0.05 α = 0.01 α = 0.001
Freedom
SU SL SU SL SU SL
*Equations to compute the density of air at any pressure, temperature, and relative humidity are
given in Section 8 of the Appendix to SOP No. 2 and are preferred for all precision mass and
volume calibrations.
7.0
Cast iron 0.000 010 0.000 030
7.2
Copper 0.000 017 0.000 050
Diamond 0.000 001 18 0.000 003 5
Fused silica (quartz) 0.000 000 5 0.000 001 6
Glass, borosilicate (T1CA) 0.000 010
Glass, borosilicate (T1CB) 0.000 015
Glass, soda-lime 0.000 025
Gold 18.0 0.000 043
Invar 0.000 000 4 0.000 001 2
Lead 11.34 0.000 029 0.000 087
Length bench 0.000 010 6
Mercury 0.000 018
Nichrome 8.39 0.000 039
Nichrome V 8.5 0.000 039
Platinum 0.000 011 0.000 027
Polycarbonate plastic 0.000 45
Polypropylene plastic 0.000 24
Polystyrene plastic 0.000 21
Steel, stainless (mass) 7.84, 7.95, 8.0 0.000 045
Steel, stainless (provers) 0.000 018 0.000 047 7 0.000 026 5
Steel, tape, mild 0.000 0116 0.000 033 5
Steel, pressure vessel, low
0.000 016
carbon
Steel, prover, low carbon 0.000 012 0.000 033 5 0.000 018 6
Steel, terne plate 0.000 012 0.000 035 0.000 019 5
Tantalum 16.6 0.000 020
Water (20 °C) 0.000 21
Reference values for cubical coefficient of expansion given in per °F are provided as reference for prover calibrations
used for petroleum products where the reference temperature is 60 °F.