Xr-Doc Aa83678-20-En Arl Genx (Service Manual) 0520
Xr-Doc Aa83678-20-En Arl Genx (Service Manual) 0520
Country : ………………………………..
Remark : ………………………………..
The above mentioned person has followed the special training course for the GenX
generator and has successfully obtained the certificate allowing him to repair and to
maintain the GenX generator. Consequently the person is aware of the risk of the
electrocution.
I confirm having followed the special training course for the GenX generator and having
successfully obtained the certificate allowing to repair and to maintain the GenX generator
This AA83678 document must be used ONLY by the person who has
followed the special training of GenX generator and that has
successfully obtained the certificate to work on it.
No AA83678-20
1 2
H
Hydrogen
He
Helium
3 4 5 6 7 8 9 10
Li
Lithium
Be
Beryllium
B
Boron
C
Carbon
N
Nitrogen
O
Oxygen
F
Fluorine
Ne
Neon
11 12 13 14 15 16 17 18
Na Mg
Sodium Magnesium
Al
Aluminum
Si
Silicon
P
Phosphorus
S
Sulfur
Cl
Chlorine
Ar
Argon
19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36
K
Potassium
Ca Sc
Calcium Scandium
Ti
Titanium
V
Vanadium
Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br
Chromium Manganese Iron Cobalt Nickel Copper Zinc Gallium Germanium Arsenic Selenium Bromine
Kr
Krypton
37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54
Rb Sr
Rubidium Strontium
Y
Yttrium
Zr Nb Mo Tc Ru Rh Pd Ag Cd
Zirconium Niobium Molybdenum Technetium Ruthenium Rhodium Palladium Silver Cadmium
In
Indium
Sn Sb Te
Tin Antimony Tellurium
I
Iodine
Xe
Xenon
55 56 57
* 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86
Cs Ba La
Cesium Barium Lanthanum
Hf
Hafnium
Ta
Tantalum
W
Wolfram
Re Os
Rhenium Osmium
Ir
Iridium
Pt
Platinum
Au Hg
Gold Mercury
Tl
Thallium
Pb
Lead
Bi
Bismuth
Po
Polonium
At Rn
Astatine Radon
87 88 89 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118
**
Fr
Francium
Ra Ac Rf Db Sg Bh Hs Mt Ds Rg Cn Nh
Radium Actinium Rutherfordium Dubnium Seaborgium Bohrium Hassium Meitnerium Darmstadtium Roentgenium Copernicium Nihonium
Fl
Flerovium
Mc Lv Ts Og
Moscovium Livermorium Tennessine Oganesson
58 59 60 61 62 63 64 65 66 67 68 69 70 71
* Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Cerium Praseodymium Neodymium Promethium Samarium Europium Gadolinium Terbium Dysprosium Holmium Erbium Thulium Ytterbium Lutetium
Service Manual
AA83678 - 20 • May 2020
SERVICE MANUAL
ARL GEN'X
Report MANUAL
No AA83678-20
Language E
Department Doc
Author MMa
AA83678-20
The information in this document is subject to change without
notice.
We assume no responsibility for any errors that may appear
in this document.
AA83678-20
PAGE DE MODIFICATION
ÄNDERUNGSANGABEN
2005
April 01 General updating
November 02 OM3865 - S112248-0 added
S112072 becomes 05
2006
April 03 General updating
2007
November 04 Chapters 1, 4, 10 updated.
Chapter 11 added
Schematic S010801 added in chapter 8.
2008
June 05 Chapters 2, 4, 10, 11 updated
2009
January 06 Chapters 9 and 11 updated
S112073 becomes 07
April 07 Chapter 9 – Catalogue updated
July 08 Chapters 1, 6, 7, 9, 10 updated (OM4112 – 4128)
S112072 and S112066 updated
October 09 Chapters 1 and 4 updated
2010
January 10 Chapters 2 and 4 updated
S112067 becomes 11
S112073 becomes 09
November 11 S112064 becomes 11
S112065 becomes 8
S112067 becomes 12 (version -9 added)
S112073 becomes 10 (version -7 added)
Chapter 4 updated
2011
April 12 Chapter 9 – Catalogue updated / KXXGENX130 – 140 added
July 13 S112247 becomes 04
S112248 becomes 02
S112067 becomes 13
S112073 becomes 11
August 14 Chapter 9 – Catalogue updated / KXXGENX150 added
Pages 4-20 and 10-25 updated (alarm 8282)
November 15 Chapter 9 – Catalogue updated / KXXGENX151 added
AA83678-20
PAGE DE MODIFICATION
ÄNDERUNGSANGABEN
2012
2014
August 17 Chapter 9 – Catalogue updated / KXXGENX199: Fuses added
for XPSI and XPSP boards
2020
May 19 SDR25503: chapter 9 updated with info about ARL 9900 from
serial number 9953964 including 9953165, 9953396,
9953835, 9953933, 9953941, 9953943.
AA83678-20
TABLE OF CONTENTS
Table of Contents
GENX VERSION COMPATIBLE WITH XPS BOARD FOR RETROFIT ................. 1-12
ALARMS LIST FOR FIRMWARE VERSION 3.2.8 AND LOWER .......................... 2-35
HOW TO REMOVE A GENX GENERATOR FROM THE XRF INSTRUMENT ........ 3-2
Removal procedure .............................................................................................. 3-2
HOW TO INSTALL A GENX GENERATOR INTO THE XRF INSTRUMENT ........... 3-4
Installation procedure ........................................................................................... 3-4
9 CATALOGUE........................................................................................... 9-1
GENX I GENERATOR 3.6KW 60KV/120MA GENX I GENERATEUR 3.6KW
60KV/120MA ............................................................................................................ 9-5
Generator HV
KVMonFb
Divider
Master control
HV HV HV HV
KVRef Corrector Inverter Multiplier
Power
Monitoring
PowerRef C
Water-cooling control Power
Interface I_FilRef Control
A
Board C
Xray Tube
K
MaRef Emission
Current Filament
Corrector Inverter
Safety interlock MaMonFb
Emission
Safety Current
Interlock Reading
XrayOn
R
Control
LVPS
Figure 1.1
Generator
KV Monitor HV
Divider
HV HV HV
Voltage Reference Corrector Inverter Multiplier
Power
M onitoring Power A
C Control X-Ray
Power Reference
Tube
Anode mA Monitor
Filament
XPSLP Board
Current
Emission Filament Sensor
Filament Current Reference
Current Driver
mA Reference
Corrector
C
Safety Filament Current
Cathode mA Monitor
Figure 1.2
The interface board called XPSC replacing XPS (or XPSLP) is integrated into the 19” GenX rack. XPS (or
XPSLP) are absent from the instrument Electronic Rack.
Generator
HighVoltage HV
Divider
Master control
HvPwm Control HV HV HV
Inverter Multiplier
EmissionCurrent+
Cooling control
Control
A
Board
Xray Tube
K
FiamentPwm Control Filament
Safety interlock
Inverter
EmissionCurrent-
Instrument LVPS R
Figure 1.3
Figure 1.4
GenX interconnection
Interconnection diagrams:
S112163 Actual ADVANT’X +, XP, XP+ interconnection schematic (with j103 connector)
S112157 Future ADVANT’X +, XP, XP+ interconnection schematic (without j103 connector)
S112162 Actual 9800 XP, XP+ interconnection schematic (with j103 connector)
S112156 Future 9800 XP, XP+ interconnection schematic (without j103 connector)
S112287 9900 LP 1200W, GenX Low Power Interconnection
S112156 9900 XP/XP+, GenX XP/XP+ Interconnection
S010801 GenXRD - GenXRF
The interface with the instrument is made with a 96 wires flat cable (J104). This connection corresponds to
the upper instrument electronic rack backplane bus (J100 to J109). By this way the GenX receives the low
power supply voltages (+5V, +24V, +12V and –12V), the XSN communication signals, the display signal,
pressostat interlock and the master to slave control signals (XpsReset and XpsReady).
The “water cooling system” and the “X-ray safety interlock controls” are connected via M8 connector to
XPSC.
The mains voltage (230 VAC) is connected XPSI board via J2 connector.
The J3 on the XPSI board is used to connect the 230 V “X-ray lamp ON”. 24 VDC Lamps are connected to
J407 of XPSC.
For the X-ray tube high voltage connection, a “federal standard” connector is plugged in the J402 of the
XPSM assembly.
Input power connector230 VAC. Data connector 2x12 with X-ray tube Filament connector
WAGO 6x pins. J2 red key lock. M8 WAGO 5x pins. J1
Instrument ground cable Data connector 32x3 Water fast fitting connector
with border key lock. J104 male
Water fast fitting connector Connector X-ray On Lamps X-ray tube HV connector.
male for version 24VDC JP1
WAGO 7 pins. J407
Figure 1.8
XPSR
Retrofit XrayPowerSupplyRetrofit
Control Board
Safety XPSC
Control XrayPowerSupplyControl
Communication Board XPSB
LVPS Xray
Power
XPSP Supply
XrayPowerSupplyPower Backplane
Board Board
Filament XPSM
XrayPowerSupplyMultiplier
Board
Multiplier
Multiplier And
HighVoltage Box Filament
Assembly
Figure 1.9
Figure 1.10
XPSI assembly
The XPSI assembly is made with the XPSI board and the cooling system for the power components. The
XPSI board converts the 230 VAC from the mains voltage into a DC voltage from 360 V to 440 V. Pay
attention to that the DC voltage is referenced to the mains voltage.
XPSR board
This board will be used to emulate the non-GenX system for field retrofit capability. With this board
installed, the instrument uses the XPS interface.
XPSC board
This board is the interface with the XQM master board in the instrument. It controls all functions in the
GenX like the different servo-loops for the kV, mA and filament PS and few watchdog safety systems. It
controls too the different external parts like the safety interlock and the water cooler system.
XPSP assembly
The XPSP assembly is made with the XPSP board and the cooling system for the power component. The
XPSP board converts the intermediate DC voltage (360 V- 440 V) into AC voltage sources, which supplies
current to the filament transformer and the HV transformer.
Multiplier assembly
The multiplier assembly is made with the following parts:
XPSM board
Filament transformer
HV transformer
Multiplier box
Cooling system
The XPSM board has three main functions: the filament power supply (rectifier/voltage and current
measurement/transformer and X-ray tube connections), the HV measurement and the emission current
measurement.
The filament transformer via the XPSM board is used to supply up to 120W to the X-ray tube filament.
The HV transformer via the multiplier is used to supply up to 4200W to the X-ray tube anode.
Each board has its specific position in the rack. There are no keys to prevent the connection of a board to the
wrong slot.
T he m ains X PSI
X ra yP o w erS up p lyInp u t
B o ard
S afety X PSC
C o ntro l X ra yP o w erS up p lyC o ntro l
C o m m u nica tio n B o ard X PSB
LVPS X ra y
P o w er
X PSP S up p ly
X ra yP o w erS up p lyP o w er B ackp lane
B o ard B o ard
F ila m e nt X PSM
X ra yP o w erS up p lyM ultip lier
B o ard
M ultip lier
M ultip lier A nd
H ig hV o lta ge B ox F ila m e nt
A sse m b ly
Figure 1.11
The retrofit configuration uses one more board in the GenX called XPSR. The XPSR board emulates the
non-GenX generator.
This version is compatible with the current non-GenX generators in the field.
XPSR
Retrofit XrayPowerSupplyRetrofit
Control Board
XPSC
XrayPowerSupplyControl
Board XPSB
Xray
Power
XPSP Supply
XrayPowerSupplyPower Backplane
Board Board
Filament XPSM
XrayPowerSupplyMultiplier
Board
Multiplier
Multiplier And
HighVoltage Box Filament
Assembly
Figure 1.12
GENX DESCRIPTION
GenX Description Chapter 2
2 GENX DESCRIPTION
Two processors are used to control all functions. One is a micro-controller (µC) and the other is a “digital
signal processing” (DSP) processor.
The µC does dialog with the instrument master board XQM of the electronic rack and controls the DSP, the
filament power supply and some background safety checks.
Bloc diagram
L D 400Vrtm
L
HighVoltage
i Multiplier
230Vac u u 200Vrtm
N 0Vrtm
Instrument
Master
Board
uC 1000 M Ohm
XQM
i
400Vrtm
+200Vrtm
PI PWM
0Vrtm
Figure 2.1
Figure 2.2
Description
The main functions controlled by the XPSI board are the mains (230VAC) power factor correction for the
current, the mains current limitation and the X-ray lamps control.
The mains voltage is used via a filter and a relay. When the generator switches on, the relay is energized to
close both contacts. After the rectifier diodes, the FET1 is used to open the circuit if the current is above a
current limit. When the FET1 is open, the current flow through the integrated diode D1of the FET circuit,
due to the inductor L. When the current is below a current limit the FET1 will drive.
FET1 L D2 +400Vrtm
L
C1
IL FET2 +12kV
230Vac U1 D1 U2 C2
N 0Vrtm
uC DSP
Figure 2.3
The FET2 and the integrated diode D2 of the FET circuit are used to step up the voltage to 400 V +/- 10 %
this voltage is adapted according to the GenX output voltage. 360 V for 0 to 30 kV. 360 V to 440 V
proportionally from 30 kV to 70 kV). Each 50 µs the current “IL”, the voltages “U1” and “U2” are evaluated
in order to define the FET2 OFF/ON conduction time. The maximum OFF/ON time is 25 us.
V U1
IL
UD1
U2
FET2
OFF
U1
IL
FET2
ON
25 us
Figure 2.4
During the OFF time the current flows through D2 to load the capacitors C1 and C2. The C1/C2 voltage is
measured by the DSP and compared to the set-point value. The On time is regulated relative to this
comparison.
The IGBTs on the XPSP power board, generating the anode high voltage, via the transformers and the
multiplier.
The IGBTs on the XPSP power board, generating the cathode (filament) heating current.
Proportional 400 V
U2 400V Monitoring
440 V
360 V
HV
30 kV 70 kV
Figure 2.5
This device allows the DSP to detect a floating neutral or a neutral-phase inversion.
To monitor the difference between both potentials use the following terminal mode command:
Figure 2.6
This power board contains the switching components (IGBTs) for the high voltage transformer and for the
filament power supplies.
Part of XPSP HV
Transformer
FET1 L IGBT1 C3
230Vac D +400V
HighVoltage
i C1
u
FET2
u +200Vrtm
Multiplier
C2 IGBT2
0Vrtm
External
Master
Board
uC 1000 M Ohm
Part of XPSP
IGBT3
i
400Vrtm
PI PWM +200Vrtm
0Vrtm Filament
IGBT4 Transformer
Figure 2.7
The HV switching works at 20 kHz. The transformer primary is connected to the C1/C2 point which
corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V). The control board
commands alternately the IGBT1 and IGBT2. The On time duration depends of the output power needed. By
this way, the transformer primary receives, via C3, plus and minus 200 V referenced to the fix 200 VRTM.
UC3-0Vrtm
+400V
+200V
0 t
IGBT1
Command
IGBT2
Command
Figure 2.8
UC3-0Vrtm
+400V
+200V
0 t
IGBT1
Command
IGBT2
Command
Figure 2.9
The transformer inductor and C3 capacitor make a resonant circuit. By this way the current look like
sinusoidal:
UC3-0Vrtm
+400V
+200V
0 t
Transformer
Primary
current
Figure 2.10
The filament power supply switching works at 40 kHz. The transformer primary is connected to the C1/C2
point which corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V). The
control board commands alternately the IGBT3 and IGBT4. The On time duration depends of the output
power needed (from 80 W to 120 W). By this way, the transformer primary receives positive and negative
200 V referenced to the fix 200 VRTM.
See after the voltage and the current at the primary of the filament transformer:
Filament
transformer
+200V
-200V t
Filament
Transformer Primary
current
0 t
Figure 2.11
The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.
U (V)
Uref 4.3 V
Uref 2.65 V
Uref 2.36 V
Uref 0.71 V
I (A)
-I Max -5A 5A I Max
-140A 140A
Trip Switch On Level Trip
Figure 2.12
If the current is out of limit the system does generate a /PwmTrip to the DSP of XPSC.
The PwmTrip does disable the HvPwmH and HvPwmL driving the gates of the HV IGBT Inverter Stages.
To the ather hand the system does generate /Limit I Hv. This signal does with the transistors M6, M9 and
Galvanic insulated drivers IC5 and IC8 disable the HvPwmH and HvPwmL driving the gates of the HV
IGBT Inverter Stages.
U (V)
Uref 3.0 V
Uref 2.5 V
Uref 2.35 V
Uref 1.87 V
I (A)
-I Max -3A 3A I Max
-15A 15A
Trip Switch On Level Trip
Figure 2.13
If the current is out of limit the system does generate /Limit I Fil. This signal does with the transistors M3,
M12 and Galvanic insulated drivers IC3 and IC13 disable the FilPwmH and FilPwmL driving the gates of
the Filament IGBT Inverter Stages.
+24V Safety:
If < 20V or if > 23V the IC7 does generate Lim_24V.
If active the HV and Filament IGBT Gate control signals are disabled.
Low voltages:
Gate low voltages: +VGate 1, +VGate L1, +VGate 3, +VGate L3 -VGate 1, -VGate L1, -VGate 3, -VGate
L3 are generated from the +24V via galvanic insulated voltage generators IC1, IC4, IC6, IC9.
Emitor low Voltages: EmH1, EmH2, EmL1, EmL2, are generated from the +24V via galvanic insulated
voltage generators IC1, IC4, IC6, IC9.
+5VA is generated from +24V via low voltage generator IC2.
Designation Remark and typical value
Terminal Mode Cmd
DY 2,3,11,4,0 XPSP, +5VM XPSP AD CHAN0: 204
DY 2,3,11,4,1 XPSP, +24V( XPSP AD CHAN1: 204
DY 2,3,11,4,2 XPSP, -12VRTM (True or False) Note 1 XPSP AD CHAN2: >128 (True)
DY 2,3,11,4,3 XPSP, +12VRTM (True or False) Note 2 XPSP AD CHAN3: >128 (True)
DY 2,3,11,4,4 XPSP, NC XPSP AD CHAN4:
DY 2,3,11,4,5 XPSP, NC XPSP AD CHAN5:
DY 2,3,11,4,6 XPSP, PowerTemp XPSP AD CHAN6: 212
DY 2,3,11,4,7 XPSP, XpspTemp XPSP AD CHAN7: 81(at 25°C)
Note 1: Since XPSP Schematic S112 067 Mode Level 10 the +12 VRTM generator has been suppressed.
Since XPSP Schematic S112 067 Mode Level 11 the + 12 VRTM generator has been suppressed.
A + 5 VA_PTB has been introduced to check the presence of the External Voltage Presence.
Note 2: Since XPSP Schematic S112 067 Mode Level 10 the -12 VRTM generator has been suppressed.
Since XPSP Schematic S112 067 Mode Level 11 the - 12 VRTM generator has been suppressed.
A + 5 VA_PTB has been introduced to check the presence of the External Voltage Presence.
The six-pack IGBT is cooled by water, for high power application, and by air, for low power application.
Figure 2.14
The XPSM assembly has three main functions. First is the HV supply, second is the emission current
measurement and third is the filament current supply.
External
Master
Board
uP 1000 M Ohm
Part of XPSM
assembly
i
400Vrtm
+200Vrtm
PI PWM
0Vrtm
Filament
Transformer
Figure 2.15
The 400V from the IGBTs on the XPSP board, via XPSM board goes to the HV transformer. At the
transformer secondary the voltage changes according to the IGBT1 and IBGT2 ON time. The maximum
voltage is near 12000 V. The full wave multiplier increases six times this voltage to reach at maximum
70000 V. Two dividers are used to monitor the high voltage. The command board uses one of them to
control the HV via the servo loop function according to asking high voltage (called “kV” in the WinXRF).
The second is used to control the first via a subtraction function.
The emission current is measured at two positions. The first position is the multiplier entrance current, which
represents the high voltage current. The second position is the current that return from the filament
(Kirchoff’s current law). The control board uses the first to control via the servo loop function, the asked
emission current (called “mA” in the WinXRF). The control board, to calculate the leakage current, uses
both measurements by subtraction function. The leakage current corresponds to the “multiplier absorbed
current” minus the “filament return current”. When the leakage current is not coherent, the control board
prevents to switch power on or if the power is on, switches it off. See table after for details.
You can retrieve the leakage value by using the terminal mode command explained after:
The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.
The multiplier box is cooled by water (Only in high power application). Two fans cool the rest of the
multiplier assembly.
Figure 2.16
This board is the interface with the XQM master board in the instrument. It controls all functions in the
GenX, like the different servo-loops for the HV, emission current and filament power supply and few
watchdog safety systems. It controls too the different external parts like interlock safety circuit and water
cooler system.
HV
Transformer
230Vac FET1 L D +400Vrtm
HighVoltage
C1
u
i FET2
+200Vrtm
Multiplier
u
C2
0Vrtm
External
Master
Board
uP 1000 M Ohm
XPSC i
400Vrtm
Part
+200Vrtm
PI PWM
0Vrtm Filament
Transformer
Figure 2.17
Figure 2.18
Figure 2.19
PPI 1 82C55 IC 10
Port State Designation Remark
PPI1 PA 0 I /Int1 Pressostat Active if pressure < 25 KPA
PPI1 PA 1 I /Int2 Electronic rack safety switch
PPI1 PA 2 I /Int3 Tank cover + X-ray tube position switches
PPI1 PA 3 I /Int4 Magazine + load lift cover switches
PPI1 PA 4 I /Int5 X-ray generator rear switch
PPI1 PA 5 I /IntRet 60°C secondary water thermostat
PPI1 PA 6 I /24VPOK +24V instrument OK
PPI1 PA 7 I /WaCoOn Water cooler ON
Port Address: FFC0H
Figure 2.20
AD converters
First AD Miscellaneous 8×8 bits AD converter. To display the different AD input values, use the
terminal mode commands DY
See display information. AD display
commands.doc
Lamp1:
24V DC for GEN’X 1200 W
Lamp2:
24V DC for GEN’X 1200 W
HV Arc Threshold
55 Digits = 1080mV
XPSC temp:
Should not exceed 65°C
DA Converter
The tap or primary water flow. An alarm is immediately issued if the flow drops. However, the default
will only shut down the generator after a delay of 20 seconds.
The tap or primary water inlet and outlet temperatures.
The deionised water flow.
The deionised water inlet and outlet temperatures.
The deionised water resistivity (560k**cm).
The deionised water level
The change of state of cooling safety causes the immediate shut down of the generator by acting on the
generator interlock input. The parameter responsible for the shut down will generate an alarm reported to the
ACS software.
Deionised water
Flow 2 l/min
Pressure <3.5 bars
Outlet temperature <60°C
Resistivity >550 K
22ms
+ DEIONIZED WATER RESI. DRIVER IN
-
+14V
0V
3.3V
tol 3<>4.5V
A/D
PORT 0
1M
DEIONIZED
WATER RESIST.
+ IN
resistivity gauge RX
-
Figure 2.21
HV servo-loop control
The control board receives the asked HV from the PC software application via the COM port and the
instrument master board. The asked HV is compared with the real HV to generate the HV error, which is
used through the PID corrector to PWM duty-cycle. The PWM signals control the power stage on XPSP
board to supply the right voltage according to the HV error. The XPSM assembly transformer and multiplier
increase this voltage to generate the real HV. The DSP evaluates the values each 50 us.
uC DSP
HV
HV Multiplier
set point Power
+ PID PWM Box
- Stage
HV
D HighVoltage 18762
A
Part of Part of
Part of XPSP board XPSM assembly
XPSC board
Figure 2.22
The rest of the bloc diagram works like a controlled current source. I.e., the filament current changes
according to the control voltage. 1 volt corresponds to 3.76 amperes. The set-point voltage is compared with
the current sensor voltage to control, via the proportional corrector, the PWM duty-cycle at 40 kHz. The
power board contains the power stage, which generates the power for the transformer. The transformer is
used to increase the current at the right level (up to 12 A). The filament receives a DC current through four
rectifier diodes.
uC
Current D + P PWM + -
set point A - X-ray
Tube
Filament
Figure 2.23
uC
Emission
current
set point
+
-
PID Filament current power supply
HV
Power Multiplier
Stage
D -1
A
R
Part of Part of
Part of XPSC board XPSP board XPSI assembly
Figure 2.24
The rectified input voltage is weighted by this voltage error to become the current reference for a second
servo-loop. The real current, to generate the current error, is compared to this reference. The GPFC corrector
uses the current reference, the real current and the current error values to control the PWM duty-cycle, which
controls the Power MOS FET (FET2).
The FET1 is controlled by three ways. The first one is the intermediate voltage via a P corrector. The PWM
duty-cycle changes according to intermediate voltage. The second one is made on the real current value and
a comparator with a current threshold. At switching power on, this current threshold corresponds to 2
amperes to limit the inrush current. When after 10 seconds, the intermediate voltage exceeds 360 V, the
current threshold changes to 50 Amperes. This corresponds to the maximum acceptable current during power
on state. The last way controls the FET1 by hardware. The real current is compared with a reference value.
When the real current exceeds 50 amperes the FET1 is switched off.
FET2
PWM
A
D
P
-
A +
D
P P
DSP
PI
GPFC PWM
Reference D
+ -
Voltage A
Figure 2.25
The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogic to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.
Status channels
Terminal Mode Cmd Designation Remarks
RS 2048 X-ray kilovolts [kV] Available in all versions
RS 2049 X-ray milliamps [mA] Available in all versions
RS 2050 X-ray cooling primary water input High power version only
temperature [0C]
RS 2051 X-ray cooling primary water output High power version only
temperature [0C]
RS 2052 X-ray cooling secondary water input Available in all versions
temperature [0C]
RS 2053 X-ray cooling secondary water output Available in all versions
temperature [0C]
RS 2054 X-ray cooling secondary water Available in all versions
resistivity. [k.cm]
RS 2055 X-ray tube filament current [A] Available in all versions
RS 2056 XPS +15Volts [V] Non-GenX or GenX retrofit version
RS 2057 XPS -15Volts [V] Non-GenX or GenX retrofit version
RS 2058 XPS +24Volts [V] Available in all versions
RS 2061 XPS +5Volts [V] Available in all versions
RS 2063 X-ray activity time [hour] Available in all versions
RS 2064 X-ray tube arcs count [] Available in all versions
RS 2067 Laboratory ambient temperature [0C] Low power version only
RS 2068 X-ray generator output temperature Low power version only
[0C]
RS 2069 X-ray tube leakage current [mA] Available in all versions
0
RS 2070 XPSC board temperature [ C] Available in all versions
RS 2071 XPS +12Volts [V] Available in all versions
RS 2072 XPS -12Volts [V] Available in all versions
RS 2073 XPS +2.5Volts reference [V] Low power version only
RS 2080 XPSC -5Volts [V] Available in all GenX versions
RS 2081 XPSI +5Volts [V] Available in all GenX versions
RS 2082 XPSI +12Volts [V] Available in all GenX versions
RS 2083 XPSI +24Volts [V] Available in all GenX versions
RS 2084 XPSI –12Volts [V] Available in all GenX versions
RS 2085 XPSI Neutral to Gnd [V] Available in all GenX versions
RS 2086 XPSI Mosfet M1 temperature [0C] Available in all GenX versions
RS 2087 XPSI Mosfet M2 temperature [0C] Available in all GenX versions
RS 2088 XPSP +5Vm [V] Available in all GenX versions
RS 2089 XPSP +24Volts [V] Available in all GenX versions
RS 2090 XPSP -10Vrtm [V] Available in all GenX versions
RS 2091 XPSP +15Vrtm [V] Available in all GenX versions
RS 2092 XPSP IGBT temperature [0C] Available in all GenX versions
RS 2093 XPSP board temperature [0C] Available in all GenX versions
RS 2094 XPSM +5Vm [V] Available in all GenX versions
RS 2095 XPSM +12Volts [V] Available in all GenX versions
RS 2096 XPSM -12Volts [V] Available in all GenX versions
RS 2097 XPSM Hv divider difference [kV] Available in all GenX versions
RS 2098 X-ray tube filament voltage [V] Available in all GenX versions
0
RS 2099 XPSM filament temperature [ C] Available in all GenX versions
RS 2100 XPSM Hv temperature [0C] Available in all GenX versions
RS 2101 Monitoring 400Volts [V] Available in all GenX versions
Introduction
Basically each GenX generator is equipped with 4 electronic boards:
All above boards and assembly multiplier are connected to a unique motherboard.
The secondary deionised water of the XRF instrument is used to cool XPSI, XPSP boards and Multiplier
assembly.
Note: XPS board, which was present in the instrument electronic cabinet, has been moved and
integrated inside the GenX generator.
- Wait for 1 minutes to allow the completion of the secondary deionised water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing down the
two main breakers 16A/32A
- Disconnect the GenX generator ground cable located on the top left side on the rear GenX panel
- Unscrew fixing screws located on the front panel of GenX generator. Slide “partially” the GenX from
its location without disconnect the water tubes and electrical cables and taking care to avoid destroying
the electrical connectors and water tubes.
- Disconnect the two external water tubes using the water fast fitting connector. See below picture
- Disconnect all the external electrical connectors and X-Ray High Voltage cable connector as shown in
the picture below.
Input power connector230 VAC. Data connector 2x12 with X-ray tube Filament connector
WAGO 6x pins. J2 red key lock. M8 WAGO 5x pins. J1
Instrument ground cable Data connector 32x3 Water fast fitting connector
with border key lock. J104 male
Water fast fitting connector Connector X-ray On Lamps X-ray tube HV connector.
male for version 24VDC JP1
WAGO 7 pins. J407
Figure 3.1
IMPORTANT: In order to avoid damage during shipment, use the packaging built
specially for GenX generator to absorb shocks during shipment.
- Install the GenX generator into its previous location in the XRF instrument.
- Connect all the external electrical connectors as shown in the picture. Below.
Take care about the connector lock key, especially for the data connector 2x12 with red key lock (Ref.
M8).
- Connect the two external water tubes using the water fast fitting connectors. See picture below.
- Install the GenX generator completely in its location and fix it with screws located on the front panel of
the GenX generator.
- Connect the GenX generator ground cable located on the top left side on the rear GenX panel.
- Switch ON the XRF instrument by pulling up the two main breakers 16A/32A and releasing the red
emergency button.
- Wait until all XRF instrument starting sequences have been completed and then send the instrument
configuration.
Input power connector230 VAC. Data connector 2x12 with X-ray tube Filament connector
WAGO 6x pins. J2 red key lock. M8 WAGO 5x pins. J1
Instrument ground cable Data connector 32x3 Water fast fitting connector
with border key lock. J104 male
Water fast fitting connector Connector X-ray On Lamps X-ray tube HV connector.
male for version 24VDC JP1
WAGO 7 pins. J407
Figure 3.2
IMPORTANT:
Before to have access inside the GenX generator, you have to read carefully the following
caution notes in order to avoid any trouble occurring to the generator or any injury.
Warning
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk
DANGER: LETHAL VOLTAGES are present in the GenX generator
whenever the generator is connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX
generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after
power disconnection.
The figure hereafter shows the location of the boards t in the GenX generator. It’s very important to respect
board’s location.
Figure 3.3
The picture hereafter shows the position of the four screws holding the GenX rear panel:
Screw 2
Screw 1
Screw 3 Screw 4
Figure 3.8
The picture hereafter shows the position of the three screws holding the Multiplier assembly on the bottom
side:
Screw 1
Screw 3
Screw 2
Figure 3.9
The picture hereafter shows the position of the four screws holding the GenX top cover:
Screw 1 Screw 2
Screw 3 Screw 4
Figure 3.10
Removal procedure
This procedure starts from the step where the GenX generator is removed from the XRF instrument. Note:
To remove the GenX generator from the instrument, consult the previous chapter “ HOW TO REMOVE A
GENX GENERATOR FROM THE XRF INSTRUMENT”.
VERY IMPORTANT:
Electrocution risk
DANGER: LETHAL VOLTAGES are present in the GenX generator
whenever the generator is connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX
generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after power
disconnection.
VERY IMPORTANT:
- Disconnect the two remaining water fast fitting connectors: 1x connexion at the multiplier assembly and
1x connexion between XPSI and XPSP boards.
- Remove the rear panel of the GenX by unscrewing 4x screws. See the screws location at the previous
chapter “INTRODUCTION”.
- At this step, you are able to remove all the boards from the rear side of the GenX.
Note: All boards are connected to a unique mother board located near to the GenX front panel.
- Hereafter is the detail of the removal of each boards and multiplier assembly
VERY IMPORTANT : Take care, when you remove the Multiplier assembly from its location, do not
destroy electronic components located on the way during the removal of the XPSM board => XPSP
board electronic components and XPSM board electronic components.
Have a look to the picture showing the Multiplier assembly at the previous chapter “INTRODUCTION”
Installation procedure
This procedure starts from the step where the GenX generator is removed from the XRF instrument. Note:
To remove the GenX generator from the instrument, consult the previous chapter “HOW TO REMOVE A
GENX GENERATOR FROM THE XRF INSTRUMENT”.
VERY IMPORTANT:
Electrocution risk
DANGER: LETHAL VOLTAGES are present in the GenX generator
whenever the generator is connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX
generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after power
disconnection.
VERY IMPORTANT:
- Disconnect the two remaining water fast fitting connectors: 1x connexion at the multiplier assembly and
1x connexion between XPSI and XPSP boards.
- Remove the rear panel of the GenX by unscrewing 4x screws. . See the screws location at the previous
chapter “INTRODUCTION”.
- At this step, you are able to install all the boards from the rear side of the GenX.
Note : All boards are connected to a unique mother board located near to the GenX front panel
- Hereafter is the detail of the installation of each boards and multiplier assembly
pass through two “click” distinct positions when the XPSI board is pushing to the final position. Only
one “click” inform that XPSI board isn’t in the final position.
Have a look to the picture showing the XPSI board at the previous chapter“INTRODUCTION”
- At this step, all boards and multiplier assembly are installed and locked to the final mechanical position
to the unique mother board.
- Install the rear panel of the GenX by screwing 4x screws. See the screws location at the previous chapter
“INTRODUCTION”.
- VERY IMPORTANT: Connect the two water fast fitting connectors: 1x connexion at the multiplier
assembly and 1x connexion between XPSI and XPSP boards. Have a look at the chapter “HOW TO
INSTALL A GENX GENERATOR INTO THE XRF INSTRUMENT” showing the rear side of the
GenX with water fast fitting connector installed.
- At this step, the GenX generator is ready to be installed into the XRF instrument. Have a look at the
chapter “HOW TO INSTALL A GENX GENERATOR INTO THE XRF INSTRUMENT”.
Cable
Screw (2x)
Grub screw
(Screw out before tightening)
KXx9499D001800
Silicone Oil
Figure 3.11 X-ray tube and generator connector parts
1. Make sure that the termination plug and the mounting accessories are clean.
2. Screw out the grub screw and slip the ring nut over the sleeve.
3. Place the divided flange onto the ring flange.
4. Wet the silicone washer (both sides) and the rubber gasket with silicone oil.
5. Fit the rubber gasket onto the plug, against the ring flange.
6. Push the silicone washer over the contact pins.
7. Insert the termination plug into the receptacle socket (check the nipple on the plug to ensure correct
positioning of the contact pins).
8. Hand tightens the ring nut.
9. TIGHTEN THE GRUB SCREW.
IMPORTANT:
Always replaces the silicone washer by a new one in case the ring nut has been loosened.
TROUBLESHOOTING
Troubleshooting Chapter 4
4 TROUBLESHOOTING
Gen’X diagnostic flow chart
Switch ON Gen'X
No
XRAY OFF WAIT ON XRAY
GetStatusGenerator GetAllStatus
GetStatusInputCurrent GetStatusGenerator
XR Cmd = ON
^
InputVoltage > 280 [V] ?
No INTERLOCK
GetStatusGenerator Yes
GetStatusInputCurrent
Buzzer On
RECTIFIER No
Interlock Ok ?
INPUTVOLTAGE
GetAllStatus
GetStatusGenerator
Yes
No WATERS ON
GetStatusGenerator Monitoring 400V > 250 [V] ?
GetStatusInputCurrent
GetInterlockStatus
Yes
No
Waters Ok ?
POWER ON XRAY
GetAllStatus
GetStatusGenerator
Yes
CHANGE KV MA
GetAllStatus
GetStatusGenerator
Resistivity Ok?
No
WAIT XRAY IN TOL
Yes GetAllStatus
GetStatusGenerator
SWITCH ON XRAY
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus
GetCoolantStatus Kv Ma InTolerance ?
GetStatusADMon400V
Yes
Power On Relay
XRAY ON
GetAllStatus
Xray Lamp On GetStatusGenerator
CheckKvMaRegulation
Figure 4.1
Details
No No
GetStatusGenerator
GetAlarmDSP GetStatusADMon400V
CheckStatusTemp
CheckStatusVoltage
No
No GetAllStatus
GetInterlockStatus
GetCoolantStatus
GetStatusInputCurrent
GetFilamentStatus
CheckWarningLamp
No
No
GetStatusVoltage
GetCoolantStatus
Primary water
&
Yes
Secondary water
== Fault ?
No
CheckStatusTemp
FAULT GENERATOR
Figure 4.2
XR Cmd = OFF
SWITCH OFF KV MA No
GetAllStatus
GetStatusGenerator
Kv & Ma == 0 ?
Yes
Cooling Off ?
Yes
XRAY OFF
GetStatusGenerator
GetStatusInputCurrent
Figure 4.3
Switch ON Gen'X
XRAY OFF No
WAIT ON XRAY
GetStatusGenerator GetAllStatus
GetStatusInputCurrent GetStatusGenerator
XR Cmd = ON Step 4
^
InputVoltage > 280 [V] ?
No INTERLOCK
GetStatusGenerator Yes
GetStatusInputCurrent
Buzzer On
Step 1
RECTIFIER No
Interlock Ok ?
INPUTVOLTAGE
GetAllStatus
GetStatusGenerator
Yes Step 5
No WATERS ON
GetStatusGenerator Monitoring 400V > 250 [V] ?
GetStatusInputCurrent
GetInterlockStatus
Step 2 Yes
No
Waters Ok ?
POWER ON XRAY
GetAllStatus
GetStatusGenerator
Yes
Step 6
Monitoring 400V > 340 [V] ?
SW RESISTIVITY
No
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus
GetCoolantStatus Yes
Step 3
CHANGE KV MA
GetAllStatus
GetStatusGenerator
Resistivity Ok?
No
WAIT XRAY IN TOL
Yes GetAllStatus
GetStatusGenerator
Step 7
SWITCH ON XRAY
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus
GetCoolantStatus Kv Ma InTolerance ?
GetStatusADMon400V
XRAY ON
GetAllStatus
Xray Lamp On GetStatusGenerator
CheckKvMaRegulation
Figure 4.4
Details
No No
GetStatusGenerator
GetAlarmDSP GetStatusADMon400V
CheckStatusTemp
CheckStatusVoltage
No
No GetAllStatus
GetInterlockStatus
GetCoolantStatus
GetStatusInputCurrent
GetFilamentStatus
CheckWarningLamp
No
No
GetStatusVoltage
GetCoolantStatus
Primary water
&
Yes
Secondary water
== Fault ?
No
CheckStatusTemp
FAULT GENERATOR
Figure 4.5
XR Cmd = OFF
SWITCH OFF KV MA No
GetAllStatus
GetStatusGenerator
Kv & Ma == 0 ?
Yes
Cooling Off ?
Yes
XRAY OFF
GetStatusGenerator
GetStatusInputCurrent
Figure 4.6
Step 1
XR Cmd = ON Check Instruments Interlocks.
Alarms: 8297-8301.
Check the board is properly plugged on the mother board.
INTERLOCK
NO Alarms: Fatal alarms from 8204 to 8208 and 8248 to 8253
GetStatusGenerator 8204 to 8208 and 8248 to 8252 are interlock µ
GetStausInputCurrent
switches contacts
8253 is the water bloc thermal 60°C safety switch
Figure 4.7
Step 2
Checks instrument Water Parameters.
Figure 4.8
Step 3
Check resistivity
Resistivity Ok
Figure 4.9
Step 4
SWITCH ON
Check XPSI board
X-RAY
Input Voltage:
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus Alarms: 8378-8381. XPSI Board Fuse F3 is defect
GetCoolantstatus Alarms: 8292. No Input Voltage. XPSI Board Fuse F1 is
GetStatusADMon400V defect.
Contactor K1 is not driving.
Check Power On signal:
Display T. M. Cmd. DY 2,3,100,0,FF81H,0,2
Power On Relay
X-Ray lamps:
Figure 4.10
Step 5
Alarm: 8293. Time-out to reach 250 V
Yes
GetAllStatus
GetStatusGenerator
M2 rectifier diodes
Figure 4.11
Step 6
Alarm: 8294. The voltage does not step up to 360 V.
M2 Rectifier and Diode D19 have to be checked
POWER ON
X-RAY
Use the same methods as in Step 5
No
GetAllStatus
GetStatusGenerator
Monitoring 400 V
> 340 V Interrupt resistor R67 link (unsolder one leg), remove driver IC4,
and measure with a DVM Drain-Gate capacity between pin 7 and
pin 8 PowerMosfet M2.
It should show a value between 5 to 20nF,
The value does depend upon the DVM.
Interrupt resistor R66 link (unsolder one leg), remove driver IC1,
and measure with a DVM Drain-Gate capacity between pin 7 and
pin 8 PowerMosfet M1.
It should show a value between 5 to 20nF,
The value does depend upon the DVM.
Status: RS 2101
Figure 4.12
Step 7
Alarms: 8730 (kV mA in tolerance timeout.
CHANGE kV mA 8196-8199, 8273-8274.
Make sure the XC X-Ray calibration has been performed.
GetAllStatus At this level one can suspect the inverter stages of XPSP
GetStatusGenerator The following procedure is valid for:
XPSP Schematics S112067 No. Mod. 9 and lower
Measure XPSP IGBTs
With an Ohmmeter measure between pin 13 and pin 16 (Value
measured = ).
WAIT X-RAY IN Idem avec pin 13 et pin 15.
No
TOLERANCE Idem avec pin 13 et pin 14.
GetAllStatus Idem avec pin 17 et pin 16.
GetStatusGenerator Idem avec pin 17 et pin 15.
Idem avec pin 17 et pin 14.
kV mA in tolerance
Figure 4.13
DVM Z1 HV Z2 HV Z3 FIL.
+ C (2) Ω ∞ ∞ ∞
Com E(3) ≈2V ≈ 2V ≈ 2V
DVM Z4 HV Z5 HV Z6 FIL.
+ C (2) Ω ∞ ∞ ∞
Com E(3) ≈2V ≈ 2V ≈ 2V
DVM Z1 HV Z2 HV Z3 FIL.
+ E (3) Ω v530 ≈ 530 ≈ 560
Com C(2) 0.37 V 0.37 V 0.37 V
DVM Z4 HV Z5 HV Z6 FIL.
+ E (3) Ω ≈ 530 ≈ 530 ≈ 560
Com C(2) 0.37 V 0.37 V 0.37 V
Figure 4.14
Alarms list
Fatal alarms
Alarm Issued Severity Description
number by
8192 XPS F Unknown command received from XQM96
S: Reset ICS
8228 XPS F Primary water inlet temperature gauge fault Not used
8229 XPS F Primary water outlet temperature gauge fault Not used
8230 XPS F Secondary water inlet temperature gauge fault
8231 XPS F Secondary water outlet temperature gauge fault
S: XC command has to be executed. Corrupted data in the EEPROM of XPSC. Check firmware Status.
IMPORTANT:
Before to have access inside the GenX generator, you have to read carefully the following
caution notes in order to avoid any trouble occurring to the generator or any injury.
Warning
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk
- Switch OFF the X-ray power supply generator by typing the terminal mode command: #XR
0,0,3,3
- Wait for 1 minute to allow the completion of the secondary deionised water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing
down the two main breakers 16A/32A
- Remove all the panels around the GenX generator.
- Remove the HV cable connector on the back side of the generator. Inspect the connector. No
spark over traces should be visible.
Make sure no wire or any other conductor is in the vicinity of the HV connector.
- Switch OFF the X-ray power supply generator by typing the terminal mode command:
- #XR 0,0,3,3
- Wait for 1 minute to allow the completion of the secondary deionized water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing
down the two main breakers 16A/32A
- Remove all the panels around the GenX generator.
- Remove the HV cable connector on the back side of the generator and from the X-Ray tube.
Inspect the connectors; on the cable, on the generator and on the X-Ray tube. No spark-over
traces should be visible.
- With a Meger check the HV cable insulation.
- The resistivity must be infinite.
- If it is not change the HV cable.
Note: The HV cable connectors must be cleaned yearly. Both connectors’ silicone washers must be
replaced and lubricated yearly.
8268 XPS F X-ray generator identified is different of 4.2 kW type. Not used.
8269 XPS F X-ray generator identified is of 4.2 kW type. Not used
kV Regulation
kV > 0 No No test
Yes
Yes kV > 5 kV No
Yes
Yes No
Figure 4.15
mA Regulation
mA > 0 No No test
Yes
Yes mA > 10 mA No
Yes
Yes No
Figure 4.16
8275 XPSLP F X-ray tube current anode regulation error. Not used
8276 XPSLP F X-ray tube filament current fatal limit reached.
8278 XPSLP F X-ray power supply interlock signal defective. Not used
8284 XPS F X-ray generator switched off without reason Not used
8288 XPSLP F Incoherence between the anode current and the cathode current.
Not used
8305 XPSC F Power off, issued by DSP; 400V too low (<350V)
Too fast X-Ray power supply transition. 400 V voltage drop. May happen when changing X-Ray tube
conditions in terminal mode, in “On Fast mode” and “Operational Mode”.
See Step 6
8307 XPSC F Power off, issued by DSP; Current too high (>35A)
8355 XPSC F Status : Mon 400V too low Suppressed after V 3.2-0
8356 XPSC F Status : Mon 400V too high Suppressed after V 3.2-0
K1 i
L
IL C1 +12kV
230Vac U1 U2
C2
N
15Vrtm
uC DSP
Figure 4.17
At power up if the 32 Amp breaker is on, the voltage leaks from T1 through the drivers to the capacitors C.
The 400 V measuring device does read about 15 V. If not, the 32 Amp breaker is down or the IGBTs of
XPSP are in short circuit.
See Step 4
PTB-TÜV
SwitchOn XRAY SwitchOff XRAY
External Voltage =
Yes
Present ?
internalRelay = OFF
8385 F_EXTERNAL_RELAY_INCOHERENCY
No -10Vrtm and +15Vrtm = ON
externalRelay = ON Yes
8386 F_POWER_ON_RELAY_INCOHERENCY
No Internal Voltage 320V = ON
No
External Voltage =
No
Present ?
externalRelay = OFF
8387 F_NO_EXTERNAL_RELAY_FEEDBACK
-10Vrtm and +15Vrtm = Off
Yes 24 V ExRelayPresent = OFF
internalRelay = ON
External Voltage =
Present ?
Internal Voltage =
No
Present ?
8388 F_NO_POWER_ON_RELAY_FEEDBACK No
400Vrtm (350V) = OFF
Yes
internalRelay = OFF
Figure 4.18
A0 = -1.93370
A1 = 1.04972
Figure 4.19
Multiplier
mA + Assembly
a)
i
i b)
A
C
mA -
Figure 4.20
Warning alarms
8704 XPS W Initial primary water flow insufficient
8705 XPS W Primary water cooler OFF
8706 XPS W Spectrometer pressostat open
8707 XPS W Spectrometer tank cover safety open
8708 XPS W Primary chamber or PBF safety open
8709 XPS W X-ray tube position safety open
8710 XPS W X-ray tube connection safety open
8711 XPS W Illegal interlock
8712 XPS W Primary water inlet temperature gauge fault
8713 XPS W Primary water outlet temperature gauge fault
8714 XPS W Secondary water inlet temperature gauge fault
8715 XPS W Secondary water outlet temperature gauge fault
8716 XPS W Secondary water flow OFF
8717 XPS W Primary water flow insufficient
8718 XPS W Secondary water level gauge fault
8719 XPS W Secondary water level too low
8720 XPS W Secondary water resistivity driver fault
8721 XPS W Secondary water resistivity gauge fault
8722 XPS W Secondary water resistivity too low
8723 XPS W XPS A/D converter time-out
8725 XPS W X-ray kV mA measurement not calibrated
8726 XPS W X-ray kV calibration slope error
8727 XPS W X-ray kV calibration offset error
8728 XPS W X-ray mA calibration slope error
8729 XPS W X-ray mA calibration offset error
8730 XPS W X-ray kV mA in tolerance time-out
8731 XPS W Status: X-ray arc rate too big
8733 XPS W Low kV limit fault
8734 XPS W Leak current fault
8735 XPS W Secondary water temperature safety open
8738 XPS W De-ionised water tank almost empty.
8739 XPS W Unknown XPS status number
8740 XPS W XPS status sampling period too big
8741 XPS W Status: X-ray kV too low
8742 XPS W Status: X-ray kV too high
8743 XPS W Status: X-ray mA too low
8744 XPS W Status: X-ray mA too high
8745 XPS W Status: primary water input temperature too low
8746 XPS W Status: primary water input temperature too high
8747 XPS W Status: primary water output temperature too low
8748 XPS W Status: primary water output temperature too high
8749 XPS W Status: secondary water input temperature too low
8750 XPS W Status: 2nd water input temp too high
8751 XPS W Status: 2nd water output temp too low
8752 XPS W Status: 2nd water output temp too high
8753 XPS W Status: secondary water resistivity too low
8754 XPS W Status: 2nd water resistivity too high
8755 XPS W Status: X-tube filament current too low
8756 XPS W Status: X-tube filament current too high
8757 XPS W Status: XPS +15V too low
FREQUENTLY ASKED
QUESTIONS
Frequently asked Questions Chapter 5
12
10
20°C or 60 kV
Year
6 25°C or 40 kV
30°C or 30 kV
4
0
2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 4.2
Pow er
Figure 5.1
Remark: The time scale corresponds to a continuous working at constant power, 24 hours per day, and
365 days per year. Using WinXRF “Eco Mode” to decrease the X-ray tube power dissipation
between analyses will drastically increase the lifetime.
I receive an alarm « Mains voltage below 195 Vac (after 20 ms)” or “Mains voltage above 253 Vac
(after 20 ms)”, what is the problem?
Check the mains voltage 230Vac.
If often add a voltage stabilizer.
I receive an alarm « Filament current sensor incoherency”, what is the problem?
Change the XPSM board (or change the Sens1 component on XPSM)
Change the XPSC board.
I receive an alarm « Filament voltage sensor incoherency”, what is the problem?
Change the XPSM board (or change the IC1 and the IC3 components on XPSM).
Change the XPSC board.
I receive an alarm « XPSI input current sensor error”, what is the problem?
Change the XPSM board (or change the Sens2 component on XPSI).
Change the XPSC board.
GENX 600 W
GENX 600 W Chapter 6
6 GENX 600 W
The GEN’X 600 W working principle is the same as for the high power model.
The 600 W model is air cooled by additive fans installed on XPSI and XPSP boards. Therefore no water
connections are needed for XPSI, XPSP and Multiplier assembly. Nevertheless the water fast fittings are left
on the multiplier assembly.
XPSC Firmware is dedicated to 600 W power. The maximum voltage is 60kV while the emission current is
limited to 30 mA.
Figure 6.1
Figure 6.2
Figure 6.3
Figure 6.4
GENX RETROFIT
GenX Retrofit Chapter 7
7 GENX RETROFIT
The GenX RETROFIT generator is dedicated to replace previous solid state
generators for field retrofit capability, concerning ARL XRF instruments of the
series 9400 XP /XP+ Series, ADVANTI’X XP / XP+ Series, 9800 XP / XP+ Series.
Retrofit XPSR
XPS X-rayPowerSupplyRetrofit
Control Board
XPSC
X-rayPowerSupplyControl
Board XPSB
X-ray
Power
XPSP Supply
X-rayPowerSupplyPower Backplane
Board Board
Filament XPSM
X-rayPowerSupplyMultiplier
Board
HighVoltage Multiplier
Multiplier And
Box Filament
Assembly
Figure 7.1
An extra board called XPSR is installed inside the generator. This board will be used as emulator between
the GenX generator and the XPS or XRPS CONTROL BOARD plugged into the instrument electronic rack.
The secondary water circuit has to be adapted to supply the GenX power stages heat exchangers.
Figure 7.2
Interconnections
Instrument Type Schematic name Schematic N°
9400 / ADVANT’X Series XPS C.B. INTERCONNECTION (9400) S214071
9800 SERIES XPS C.B. INTERCONNECTION (9800) S214433
Remark: These schematics are part of the: 9400 / ADVANT’X schematic file AA83484
9800 schematic file AA83400
3 x 21 P CABLES
M5
S U B-D S UB-D
50 P J1 54 M6
50 P J4 01
M7
E L E C TR O NIC
R AC K
XP S C . B . B A CK P L A NE XP SR
XP SC R E T DISPL A Y
XP SI
IN S TR U M E N T G EN 'X R
X PSP
EL EC T R ON IC
XP SM
RACK
Figure 7.3
Remark:
XPSR board
Figure 7.4
RetGenX
Safety
And
Water
interface
Figure 7.5
Figure 7.6
Figure 7.7
The regulated 24 VDC is applied to five low voltage regulators used to generate respectively
+15 VDC
-15 VDC
+5 VDC
+12 VDC
- 12 VDC
+24Vout does supply the 24 V to all the boards plugged on the GenX mother board.
Jumper W2
In normal operation the jumper W2 must be removed.
Installing the jumper leads to human harm due to X-ray emission and/or electrocution.
EMULATOR XPSCRET
ADUC812
Outgoing
Signals To/From
Bus XPSCRET
Serial Data Link
Ingoing
Signals
Bus
Figure 7.8
To do so:
• The various alarms generated by XPSC are linked to the ADCU812 processor through a serial link
and converted, to be recognised by XPS.
• The start up synchronisation and safety control signals generated by XPS are converted to serial
Data and transmitted to XPSC.
• The analogic Ref signals are converted to serial Data by the ADUC812 built in ADC
• To be transmitted to XPSC.
• The Monitoring signals are converted from serial Data by the ADUC812 built in DAC to analogic
voltages to be supervised by XPS.
Serial link
Serial link is of the same type as XSN (X-ray Services Network) used in the 9800 / ADVANT’X series
instruments electronic racks.
The jumper W1 must be placed between contacts 1-2 internal, for proper communication.
Jumper settings
JUMPER SETTING
W1 1-2 Internal
W2 Removed
JP1 EnProg Removed
Figure 7.9
Part 3 Display
Schematic and layout: S215199
Figure 7.10
The 1x 20 characters display is mounted and plugged on the display front panel.
Bellow find the list of the ACS XPS alarms and the corresponding GenX alarms that can be monitored on the
GenX display.
KV regulation error
8273 XPS F X-ray kV out of regulation
Number Description
8273 F_KV_OUT_OF_REGULATION
MA regulation error
8274 XPS F X-ray mA out of regulation
Number Description
8274 F_MA_OUT_OF_REGULATION
Indicate the "KV" too high, above 60kV or above 70kV, depending of the GenX type.
KV too high supervised by the DSP for a fast control.
Displayed alarm issued by GenX:
Number Description
8197 F_KV_TOO_HIGH
8306 F_HV_TOO_HIGH_DSP
Indicate the "Ma" too high, > 120mA or > 140mA, depending of the GenX type.
Displayed alarms issued by GenX:
Number Description
8199 F_MA_TOO_HIGH
8306 F_MA_TOO_LOW
Over Load
8213 XPS F Over power on X-ray power supply
Number Description
8276 F_SAFETY_FIL_CUR_LIMIT_REACHED
8283 F_SHORTED_FILAMENT
Number Description
8215 F_OPEN_FILAMENT
Number Description
8281 F_WARNING_LAMP
8377 F_SHORTED_LAMP
At start-up, if the X-ray On lamps are defect, the buzzer does beep three times to indicate
the error.
Low KV limit
8220 XPS F Low kilovolts limit
Number Description
8196 F_KV_TOO_LOW
Power Curve:
150
3.0 kW
140
130 3.6 kW
120 4.2 kW
110
100
90
80
mA
70
60
50
40
30
20
10
0
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75
kV
Figure 7.12
Indicate the current leakage is above 5mA or indicate the absolute current leakage is above 1,5mA with retry.
Displayed alarms issued by GenX:
Number Description
8304 F_X_RAY_LEAK_CURRENT_TOO_HIGH
8389 F_XRAY_NEG_LEAK_CURRENT_ERROR
8390 F_XRAY_POS_LEAK_CURRENT_ERROR
Number Description
8295 F_TEMP_IGBT_TOO_HIGH
8300 F_XPSP_AD8BIT_CONV
8314 F_TEMP_XPSP_TOO_HIGH
8316 F_TEMP_HV_TOO_HIGH
8329 F_PLUS_5_VM_XPSP_TOO_LOW
8330 F_PLUS_5_VM_XPSP_TOO_HIGH
8349 F_PLUS_24_VOLT_XPSP_TOO_LOW
8350 F_PLUS_24_VOLT_XPSP_TOO_HIGH
8368 F_TEMP1_XPSP_OVERFLOW
8369 F_TEMP1_XPSP_UNDERFLOW
8370 F_TEMP2_XPSP_OVERFLOW
8371 F_TEMP2_XPSP_UNDERFLOW
8378 F_MINUS_10_VRTM_TOO_LOW
8379 F_MINUS_10_VRTM_TOO_HIGH
8380 F_PLUS_15_VRTM_TOO_LOW
8381 F_PLUS_15_VRTM_TOO_HIGH
Number Description
8282 F_FILAMENT_POWER_SUPPLY_FAULT
8301 F_XPSM_AD8BIT_CONV
8315 F_TEMP_FILAMENT_TOO_HIGH
8323 F_FILAMENT_CURR_SENSOR_INCOHERENCY
8324 F_FILAMENT_VOLT_SENSOR_INCOHERENCY
8331 F_PLUS_5_VM_XPSM_TOO_LOW
8332 F_PLUS_5_VM_XPSM_TOO_HIGH
8337 F_PLUS_12_VOLT_XPSM_TOO_LOW
8338 F_PLUS_12_VOLT_XPSM_TOO_HIGH
8343 F_MINUS_12_VOLT_XPSM_TOO_LOW
8344 F_MINUS_12_VOLT_XPSM_TOO_HIGH
8372 F_TEMP1_XPSM_OVERFLOW
8373 F_TEMP1_XPSM_UNDERFLOW
8374 F_TEMP2_XPSM_OVERFLOW
8375 F_TEMP2_XPSM_UNDERFLOW
Arcs detection
8244 XPS F Status: X-ray arc rate too big
Indicates that an arc has occurred within the previous second (If more than 3 arcs are detected in 10 seconds,
XPS does shutdown the GenX). No alarms are displayed.
Interlock closed
Indicates the EXTERNAL INTERLOCK connections are in the closed position. This also indicates closure
of internal power supply interlocks. The generator turns OFF if one of the contacts is opened or if the
spectrometer pressure rises above 25kPa.
XPS Interlock related alarms are uploaded to the ACS depending of the instrument type.
Interlock opened
Indicates the EXTERNAL INTERLOCK connections are in the open position. This also indicates opening of
internal power supply interlocks. The generator turns OFF if one of the contacts is opened or if the
spectrometer pressure rises above 25kPa.
If all contacts are closed and if the spectrometer pressure goes below 25kPa, it is possible to start the
generator.
XPS Interlock related alarms are uploaded to the ACS depending of the instrument type.
Power supply miscellaneous faults
8214 XPS F X-ray power supply fault
8324 F_FILAMENT_VOLT_SENSOR_INCOHERENCY
8325 F_PLUS_5_VOLT_XPSI_TOO_LOW
8326 F_PLUS_5_VOLT_XPSI_TOO_HIGH
8327 F_PLUS_5_VOLT_XPSC_TOO_LOW
8328 F_PLUS_5_VOLT_XPSC_TOO_HIGH
8329 F_PLUS_5_VM_XPSP_TOO_LOW
8330 F_PLUS_5_VM_XPSP_TOO_HIGH
8331 F_PLUS_5_VM_XPSM_TOO_LOW
8332 F_PLUS_5_VM_XPSM_TOO_HIGH
8333 F_PLUS_12_VOLT_XPSI_TOO_LOW
8334 F_PLUS_12_VOLT_XPSI_TOO_HIGH
8335 F_PLUS_12_VOLT_XPSC_TOO_LOW
8336 F_PLUS_12_VOLT_XPSC_TOO_HIGH
8337 F_PLUS_12_VOLT_XPSM_TOO_LOW
8338 F_PLUS_12_VOLT_XPSM_TOO_HIGH
8339 F_MINUS_12_VOLT_XPSI_TOO_LOW
8340 F_MINUS_12_VOLT_XPSI_TOO_HIGH
8341 F_MINUS_12_VOLT_XPSC_TOO_LOW
8342 F_MINUS_12_VOLT_XPSC_TOO_HIGH
8343 F_MINUS_12_VOLT_XPSM_TOO_LOW
8344 F_MINUS_12_VOLT_XPSM_TOO_HIGH
8345 F_PLUS_24_VOLT_XPSI_TOO_LOW
8346 F_PLUS_24_VOLT_XPSI_TOO_HIGH
8347 F_PLUS_24_VOLT_XPSC_TOO_LOW
8348 F_PLUS_24_VOLT_XPSC_TOO_HIGH
8349 F_PLUS_24_VOLT_XPSP_TOO_LOW
8350 F_PLUS_24_VOLT_XPSP_TOO_HIGH
8353 F_MINUS_5_VOLT_XPSC_TOO_LOW
8354 F_MINUS_5_VOLT_XPSC_TOO_HIGH
8355 F_MON_400_VOLT_TOO_LOW
8356 F_MON_400_VOLT_TOO_HIGH
8360 F_XPSI_AD_MON400V_ERROR
8362 F_XPSC_TEMP_OVERFLOW
8363 F_XPSC_TEMP_UNDERFLOW
8364 F_TEMP1_XPSI_OVERFLOW
8365 F_TEMP1_XPSI_UNDERFLOW
8366 F_TEMP2_XPSI_OVERFLOW
8367 F_TEMP2_XPSI_UNDERFLOW
8368 F_TEMP1_XPSP_OVERFLOW
8369 F_TEMP1_XPSP_UNDERFLOW
8370 F_TEMP2_XPSP_OVERFLOW
8371 F_TEMP2_XPSP_UNDERFLOW
8372 F_TEMP1_XPSM_OVERFLOW
8373 F_TEMP1_XPSM_UNDERFLOW
8374 F_TEMP2_XPSM_OVERFLOW
8375 F_TEMP2_XPSM_UNDERFLOW
8378 F_MINUS_10_VRTM_TOO_LOW
8379 F_MINUS_10_VRTM_TOO_HIGH
8380 F_PLUS_15_VRTM_TOO_LOW
8381 F_PLUS_15_VRTM_TOO_HIGH
8382 F_NEUTRAL_TO_GND_TOO_LOW
8383 F_NEUTRAL_TO_GND_TOO_HIGH
8384 F_XPSI_INPUT_CURRENT_ERROR
Instrument : XRF
N° Assembly : S703123T
File : S430414-1.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1. Change air direction of fan 2 6.07.05 JMa
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 1 / 6
Document: Installation procedure XPSI board S703123T
1 Introduction:
2 Board configuration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 2 / 6
Document: Installation procedure XPSI board S703123T
3 Firmware compatibility
No firmware is installed on the XPSI board.
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution
must be read carefully in order to avoid any trouble occurring to the
generator and any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 3 / 6
Document: Installation procedure XPSI board S703123T
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 4 / 6
Document: Installation procedure XPSI board S703123T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- At this step, you are able to remove the XPSI board from the rear side of the
GenX.
XPSI board is connected to a unique backplane board XPSB located near the
GenX front panel.
- Use the water tube connectors, located at the end of the board, to extract the
XPSI board from its location.
Pull the XPSI board very slightly without a big force to avoid to damage the
fixing of the water cooling block on the XPSI board.
Take care, as mentioned above, to avoid to damage electronic components
located on the way during removal of the XPSI board.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 5 / 6
Document: Installation procedure XPSI board S703123T
- Install the new XPSI board in the correct location, see chapter 4.1 of this
document.
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 6 / 6
Document: Installation procedure XPSI board Air Cooling S703778T
Instrument : XRF
N° Assembly : S703778T
File : S430427-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 1 / 6
Document: Installation procedure XPSI board Air Cooling S703778T
1 Introduction:
2 Board configuration:
Connector P1.
Connector J2. Connector J902.
Solder side
Components side Components side
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 2 / 6
Document: Installation procedure XPSI board Air Cooling S703778T
3 Firmware compatibility
No firmware is installed on the XPSI board.
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution
must be read carefully in order to avoid any trouble occurring to the
generator and any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 3 / 6
Document: Installation procedure XPSI board Air Cooling S703778T
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 4 / 6
Document: Installation procedure XPSI board Air Cooling S703778T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- At this step, you are able to remove the XPSI board from the rear side of the
GenX.
XPSI board is connected to a unique backplane board XPSB located near the
GenX front panel.
- Use your left hand (Put your thumb on the J3 connector and your others fingers
on the solder XPSI board side) to extract the the XPSI board from its location.
Pull the XPSI board very slightly without a big force to avoid to damage the J3
connector.
Take care, as mentioned above, to avoid to damage electronic components
located on the way during removal of the XPSI board.
- Install the new XPSI board in the correct location, see chapter 4.1 of this
document.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 5 / 6
Document: Installation procedure XPSI board Air Cooling S703778T
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 6 / 6
Document: Installation procedure XPSP board S703126T
Instrument : XRF
N° Assembly : S703126T
File : S430415-1.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1. Change air direction of fan 2 6.07.05 JMa
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 1 / 6
Document: Installation procedure XPSP board S703126T
1 Introduction:
2 Board configuration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 2 / 6
Document: Installation procedure XPSP board S703126T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 3 / 6
Document: Installation procedure XPSP board S703126T
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 4 / 6
Document: Installation procedure XPSP board S703126T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- At this step, you are able to remove the XPSP board from the rear side of the
GenX.
Important: Due to the very closed proximity of the XPSC, XPSR (if installed)
and XPSP boards, those 2x (3x if XPSR present) boards must be removed
following the below order to avoid damage on the electronic components part:
-> First two boards to be removed are -> XPSR (if installed) together with
XPSC board.
-> Second board to be removed is -> XPSP board.
XPSP board is connected to a unique backplane board XPSB located near the
GenX front panel.
- Use the water tube connectors, located at the end of the board, to extract the
XPSP board from its location.
Pull the XPSP board very slightly without a big force to avoid to damage the
fixing of the water cooling block on the XPSP board.
Take care, as mentioned above, to avoid to damage electronic components
located on the way during removal of the XPSP board.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 5 / 6
Document: Installation procedure XPSP board S703126T
- Install the new XPSP board in the correct location, see chapter 4.1 of this
document.
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 6 / 6
Document: Installation procedure XPSP board Air Cooling S703779T
Instrument : XRF
N° Assembly : S703779T
File : S430428-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 1 / 6
Document: Installation procedure XPSP board Air Cooling S703779T
1 Introduction:
2 Board configuration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 2 / 6
Document: Installation procedure XPSP board Air Cooling S703779T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 3 / 6
Document: Installation procedure XPSP board Air Cooling S703779T
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 4 / 6
Document: Installation procedure XPSP board Air Cooling S703779T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- At this step, you are able to remove the XPSP board from the rear side of the
GenX.
Important: Due to the very closed proximity of the XPSC, XPSR (if installed)
and XPSP boards, those 2x (3x if XPSR present) boards must be removed
following the below order to avoid damage on the electronic components part:
-> First two boards to be removed are -> XPSR (if installed) together with
XPSC board.
-> Second board to be removed is -> XPSP board.
XPSP board is connected to a unique backplane board XPSB located near the
GenX front panel.
- Use the Transformer T1 (located on the XPSP board) to extract, with your
fingers, the XPSP board from its location.
Pull the XPSP board very slightly without a big force to avoid to damage the
transformer T1.
Take care, as mentioned above, to avoid to damage electronic components
located on the way during removal of the XPSP board.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 5 / 6
Document: Installation procedure XPSP board Air Cooling S703779T
- Install the new XPSP board in the correct location, see chapter 4.1 of this
document.
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 6 / 6
Document: Installation procedure XPSR36 board S703125T
Instrument : XRF
N° Assembly : S703125T
File : S430418-1.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1. Change location of bridges W1 / W2 1.03.06 JMa
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 1 / 6
Document: Installation procedure XPSR36 board S703125T
1 Introduction:
IMPORTANT:
XPSR36 board with P/N: S703125T must be installed ONLY in the GenX RET
generator 3.6kW with P/N: S703685T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
2 Board configuration:
1 1
2
2
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 2 / 6
Document: Installation procedure XPSR36 board S703125T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 3 / 6
Document: Installation procedure XPSR36 board S703125T
P 108
J 401
Rear side
TOP side Right side
Left
side
Front side
BOTTOM side
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 4 / 6
Document: Installation procedure XPSR36 board S703125T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- At this step, you are able to remove the XPSR36 board together with the
XPSCRET board due to the closed proximity of those two boards to avoid
damage on the electronic components part.
XPSR36 board is connected to a unique backplane board XPSB located near
the GenX front panel.
- To extract the XPSR36 board from its location, let plugged the male J401
connector and use it to pull the board out.
To extract the XPSCRET board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 5 / 6
Document: Installation procedure XPSR36 board S703125T
- Install the new XPSR36 board, together with the XPSCRET, in the correct
location, see chapter 4.1 of this document.
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 6 / 6
Document: Installation procedure XPSR42 board S703785T
Instrument : XRF
N° Assembly : S703785T
File : S430419-1.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1. Change location of bridges W1 / W2 1.03.06 JMa
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 1 / 6
Document: Installation procedure XPSR42 board S703785T
1 Introduction:
IMPORTANT:
XPSR42 board with P/N: S703785T must be installed ONLY in the GenX RET
generator 4.2kW with P/N: S703687T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
2 Board configuration:
1 1
2
2
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 2 / 6
Document: Installation procedure XPSR42 board S703785T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 3 / 6
Document: Installation procedure XPSR42 board S703785T
P 108
J 401
Rear side
TOP side Right side
Left
side
Front side
BOTTOM side
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 4 / 6
Document: Installation procedure XPSR42 board S703785T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- At this step, you are able to remove the XPSR42 board together with the
XPSCRET board due to the closed proximity of those two boards to avoid
damage on the electronic components part.
XPSR42 board is connected to a unique backplane board XPSB located near
the GenX front panel.
- To extract the XPSR42 board from its location, let plugged the male J401
connector and use it to pull the board out.
To extract the XPSCRET board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 5 / 6
Document: Installation procedure XPSR42 board S703785T
- Install the new XPSR42 board, together with the XPSCRET, in the correct
location, see chapter 4.1 of this document.
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 6 / 6
Document: Installation procedure XPSCRET board S703784T
Instrument : XRF
N° Assembly : S703784T
File : S430426-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 1 / 7
Document: Installation procedure XPSCRET board S703784T
1 Introduction:
IMPORTANT:
XPSCRET board with P/N: S703784T must be installed ONLY in the GenX
generators “R” 3.6kW with P/N: S703685T et 4.2kW with P/N: S703687T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
Note:
kV/mA measurement calibration parameters are stored in the XPS board
installed in the electronic rack of the XRF instrument and in the WinXRF
software.
2 Board configuration:
Bridge
W3. Firmware XPSCR.
Location IC15.
Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED
Connector J104. 1 4
Components side. 1
1 4 1 2
2
Connector M8.
Solder side.
Connector J407.
Solder side. Bridge
W2.
Components view
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 2 / 7
Document: Installation procedure XPSCRET board S703784T
INSTALLED
Horizontally
to the C53
W3-2 XPSCRET setting
component
direction
Notes:
1) The factory setting is in bold.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 3 / 7
Document: Installation procedure XPSCRET board S703784T
3 Firmware compatibility
2x Firmwares XPSCRET (Location IC15) and XPSCS (Location IC13) are installed
on the XPSCRET board.
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 4 / 7
Document: Installation procedure XPSCRET board S703784T
Green LED
Red LED
M8
J 104
J 407
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 5 / 7
Document: Installation procedure XPSCRET board S703784T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- At this step, you are able to remove the XPSCRET board together with the
XPSR board due to the closed proximity to avoid damage on the electronic
components part.
XPSCRET board is connected to a unique backplane board XPSB located near
the GenX front panel.
- To extract the XPSCRET board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
To extract the XPSR board from its location, let plugged the male J401
connector and use it to pull the board out.
- Install the new XPSCRET board, together with the XPSR, in the correct location,
see chapter 4.1 of this document.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 6 / 7
Document: Installation procedure XPSCRET board S703784T
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
No kV/mA calibration is required when XPSCRET is replaced.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 7 / 7
Document: Installation procedure XPSC1260 board S703726T
Instrument : XRF
N° Assembly : S703726T
File : S430424-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 1 / 8
Document: Installation procedure XPSC1260 board S703726T
1 Introduction:
IMPORTANT:
XPSC1260 board with P/N: S703726T must be installed ONLY in the GenX
generator 1.2kW (60kV/60mA) with P/N: S703683T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC1260 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC1260 board stored in the WinXRF software.
2 Board configuration:
Bridge
W3. Firmware XPSCM1260.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED
Connector J104. 1 4
Components side. 1
1 4 1 2
2
Connector M8.
Solder side.
Connector J407.
Solder side. Bridge
W2.
Components view
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 2 / 8
Document: Installation procedure XPSC1260 board S703726T
INSTALLED
Horizontally
to the C53
W3-2 XPSC1260 setting
component
direction
Notes:
1) The factory setting is in bold.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 3 / 8
Document: Installation procedure XPSC1260 board S703726T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 4 / 8
Document: Installation procedure XPSC1260 board S703726T
Green LED
Red LED
M8
J 104
J 407
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 5 / 8
Document: Installation procedure XPSC1260 board S703726T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- To extact the XPSC1260 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
- Install the new XPSC1260 board in the correct location, see chapter 4.1 of this
document.
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 6 / 8
Document: Installation procedure XPSC1260 board S703726T
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC1260 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.
Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.
IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:
At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 7 / 8
Document: Installation procedure XPSC2560 board S704026T
Instrument : XRF
N° Assembly : S704026T
File : S430661-0.doc
Mod. Description Date Name
0. Document creation 28.03.06 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 1 / 8
Document: Installation procedure XPSC2560 board S704026T
1 Introduction:
IMPORTANT:
XPSC2560 board with P/N: S704026T must be installed ONLY in the GenX
generator 2.5kW (60kV/100mA) with P/N: S704025T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC2560 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC2560 board stored in the WinXRF software.
2 Board configuration:
Bridge
W3. Firmware XPSCM2560.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13 Connector J904
Switch
S1. W1. Components side
Red LED
Connector J104. 1 4
Components side. 1
1 4 1 2
2
Connector M8.
Solder side.
Connector J407.
Solder side. Bridge
W2.
Components view
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 2 / 8
Document: Installation procedure XPSC2560 board S704026T
INSTALLED
Horizontally
to the C53
W3-2 XPSC2560 setting
component
direction
Notes:
1) The factory setting is in bold.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 3 / 8
Document: Installation procedure XPSC2560 board S704026T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 4 / 8
Document: Installation procedure XPSC2560 board S704026T
Green LED
Red LED
M8
J 104
J 407
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 5 / 8
Document: Installation procedure XPSC2560 board S704026T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- To extact the XPSC2560 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
- Install the new XPSC2560 board in the correct location, see chapter 4.1 of this
document.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 6 / 8
Document: Installation procedure XPSC2560 board S704026T
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC2560 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.
Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.
IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 7 / 8
Document: Installation procedure XPSC2560 board S704026T
At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 8 / 8
Document: Installation procedure XPSC1260 board S703726T
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 8 / 8
Document: Installation procedure XPSC3660 board S703722T
Instrument : XRF
N° Assembly : S703722T
File : S430416-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 1 / 8
Document: Installation procedure XPSC3660 board S703722T
1 Introduction:
IMPORTANT:
XPSC3660 board with P/N: S703722T must be installed ONLY in the GenX
generator 3.6kW (60kV/120mA) with P/N: S703690T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC3660 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC3660 board stored in the WinXRF software.
2 Board configuration:
Bridge
W3. Firmware XPSCM3660.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED
Connector J104. 1 4
Components side. 1
1 4 1 2
2
Connector M8.
Solder side.
Connector J407.
Solder side. Bridge
W2.
Components view
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 2 / 8
Document: Installation procedure XPSC3660 board S703722T
INSTALLED
Horizontally
to the C53
W3-2 XPSC3660 setting
component
direction
Notes:
1) The factory setting is in bold.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 3 / 8
Document: Installation procedure XPSC3660 board S703722T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 4 / 8
Document: Installation procedure XPSC3660 board S703722T
Green LED
Red LED
M8
J 104
J 407
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 5 / 8
Document: Installation procedure XPSC3660 board S703722T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- To extact the XPSC3660 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
- Install the new XPSC3660 board in the correct location, see chapter 4.1 of this
document.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 6 / 8
Document: Installation procedure XPSC3660 board S703722T
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC3660 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.
Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.
IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 7 / 8
Document: Installation procedure XPSC3660 board S703722T
At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 8 / 8
Document: Installation procedure XPSC3670 board S703723T
Instrument : XRF
N° Assembly : S703723T
File : S430421-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 1 / 8
Document: Installation procedure XPSC3670 board S703723T
1 Introduction:
IMPORTANT:
XPSC3670 board with P/N: S703723T must be installed ONLY in the GenX
generator 3.6kW (70kV/120mA) with P/N: S703680T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC3670 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC3670 board stored in the WinXRF software.
2 Board configuration:
Bridge
W3. Firmware XPSCM3670.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED
Connector J104. 1 4
Components side. 1
1 4 1 2
2
Connector M8.
Solder side.
Connector J407.
Solder side. Bridge
W2.
Components view
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 2 / 8
Document: Installation procedure XPSC3670 board S703723T
INSTALLED
Horizontally
to the C53
W3-2 XPSC3670 setting
component
direction
Notes:
1) The factory setting is in bold.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 3 / 8
Document: Installation procedure XPSC3670 board S703723T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 4 / 8
Document: Installation procedure XPSC3670 board S703723T
Green LED
Red LED
M8
J 104
J 407
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 5 / 8
Document: Installation procedure XPSC3670 board S703723T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- To extact the XPSC3670 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
- Install the new XPSC3670 board in the correct location, see chapter 4.1 of this
document.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 6 / 8
Document: Installation procedure XPSC3670 board S703723T
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC3670 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.
Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.
IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 7 / 8
Document: Installation procedure XPSC3670 board S703723T
At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 8 / 8
Document: Installation procedure XPSC4260 board S703724T
Instrument : XRF
N° Assembly : S703724T
File : S430422-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 1 / 8
Document: Installation procedure XPSC4260 board S703724T
1 Introduction:
IMPORTANT:
XPSC4260 board with P/N: S703724T must be installed ONLY in the GenX
generator 4.2kW (60kV/140mA) with P/N: S703681T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC4260 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC4260 board stored in the WinXRF software.
2 Board configuration:
Bridge
W3. Firmware XPSCM4260.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED
Connector J104. 1 4
Components side. 1
1 4 1 2
2
Connector M8.
Solder side.
Connector J407.
Solder side. Bridge
W2.
Components view
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 2 / 8
Document: Installation procedure XPSC4260 board S703724T
INSTALLED
Horizontally
to the C53
W3-2 XPSC4260 setting
component
direction
Notes:
1) The factory setting is in bold.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 3 / 8
Document: Installation procedure XPSC4260 board S703724T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 4 / 8
Document: Installation procedure XPSC4260 board S703724T
Green LED
Red LED
M8
J 104
J 407
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 5 / 8
Document: Installation procedure XPSC4260 board S703724T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- To extact the XPSC4260 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
- Install the new XPSC4260 board in the correct location, see chapter 4.1 of this
document.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 6 / 8
Document: Installation procedure XPSC4260 board S703724T
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC4260 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.
Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.
IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 7 / 8
Document: Installation procedure XPSC4260 board S703724T
At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 8 / 8
Document: Installation procedure XPSC4270 board S703725T
Instrument : XRF
N° Assembly : S703725T
File : S430423-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 1 / 8
Document: Installation procedure XPSC4270 board S703725T
1 Introduction:
IMPORTANT:
XPSC4270 board with P/N: S703725T must be installed ONLY in the GenX
generator 4.2kW (70kV/140mA) with P/N: S703682T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.
Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC4270 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC4270 board stored in the WinXRF software.
2 Board configuration:
Bridge
W3. Firmware XPSCM4270.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED
Connector J104. 1 4
Components side. 1
1 4 1 2
2
Connector M8.
Solder side.
Connector J407.
Solder side. Bridge
W2.
Components view
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 2 / 8
Document: Installation procedure XPSC4270 board S703725T
INSTALLED
Horizontally
to the C53
W3-2 XPSC4270 setting
component
direction
Notes:
1) The factory setting is in bold.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 3 / 8
Document: Installation procedure XPSC4270 board S703725T
3 Firmware compatibility
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 4 / 8
Document: Installation procedure XPSC4270 board S703725T
Green LED
Red LED
M8
J 104
J 407
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 5 / 8
Document: Installation procedure XPSC4270 board S703725T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- To extact the XPSC4270 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
- Install the new XPSC4270 board in the correct location, see chapter 4.1 of this
document.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 6 / 8
Document: Installation procedure XPSC4270 board S703725T
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC4270 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.
Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.
IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 7 / 8
Document: Installation procedure XPSC4270 board S703725T
At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 8 / 8
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument, XPSC board and GenX.
New kV/mA measurement calibration parameters are stored in the new XPSC board. Those new
kV/mA parameters will replace the ones from the original “old” XPSC board stored in the OXSAS
software.
Note No fan and no water block are installed on XPSC board.
CAUTION!It is important to verify that you have the correct XPSC board. Incorrect XPSC board will
damage the X-ray tube permanently.
The picture below indicates the XPSC board including places to check:
● Check the XPSC Part Number - 1226732,
● Check Jumpers,
● The IC15 must have the blue label as well as the instrument's reset button.
.
Page 1 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
2 Safety
Note Do not perform any operation without proper safety training specific to the hazard(s).
CAUTION!
A Caution message indicates a potentially hazardous situation that, if not avoided, may
result in moderate or minor injury. It may also be used to alert against unsafe practices.
WARNING!
A Warning message indicates a potentially hazardous situation that, if not avoided, may
result in death or severe injury.
DANGER!
A Danger message indicates an imminently hazardous situation that, if not avoided, will
result in death or severe injury.
.
Page 2 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
WARNING!
Radiation danger – Indicates that X-Ray hazard may be present. Take necessary
precautions.
SERVICE!
A Service message indicates that the described procedure should be performed by
qualified personnel only.
3 Prerequisites
SERVICE! This document must be used ONLY by the person who followed the special training of
GenX generator and successfully obtained the certificate to service GenX generators.
CAUTION!Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the GenX generator.
4 Preparation
.
Page 3 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
Note 2) The secondary water cooling regulation can be set using the firmware status N° 2052.
.
Page 4 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
5 Procedure
.
Page 5 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
WARNING! Before working with the new GenX generator, please read carefully the following
information and cautions.
DANGER! LETHAL VOLTAGES are present in the GenX generator whenever the generator is
connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.
.
Page 6 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
Note For more information concerning the removal of the generator from the XRF instrument, consult the
GenX generator service manual P/N: AA83678, delivered during the certification training course given
in Ecublens by TTI department.
CAUTION! During the removal and installation of the GenX boards and Multiplier Assembly, always
put down the GenX generator with bottom side in contact with the surface table, see the
picture below. DO NOT turn the GenX upside down!
2. WAIT 15 MINUTES to be sure that all stored energy is discharged after power disconnection.
DANGER! DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.
3. Unplug and plug action of all electronic boards must be delayed by at least 15 MINUTES.
CAUTION!Not respecting of this caution note can cause serious damage on the GenX backplane
board XPSB and body’s injury due to energy stored in the capacitor.
4. Remove the 4x screws fixing the rear panel of the GenX generator.
5. Remove the rear panel of the GenX generator.
.
Page 7 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.
6. At this step, you are able to remove the XPSC board: XPSC board is connected to a unique
backplane board XPSB located near the GenX front panel.
CAUTION!IMPORTANT GENERAL NOTE: If you want to remove all the GenX boards located inside
the GenX, you must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed between them):
a) First board to be removed is -> XPSI board
b) Second two boards to be removed are -> XPSR (if installed) together with XPSC board.
c) Third board to be removed is -> XPSP board
d) Fourth “board” to be removed is -> Multiplier assembly (don’t forget the 3x locking
screws located at the bottom panel of the GenX generator).
7. To extract the XPSC board from its location, use your fingers to pull together the front connectors
J104, J407 and M8.
8. Install the new XPSC board in the correct location, see Location of the XPSC Board Inside the
GenX Generator on page 6. Avoid to damage electronic components located on the way during
installation of the XPSC board!
CAUTION!IMPORTATN NOTE: The XPSC board must be installed at the final locked mechanical
position, by passing through two “distinct clicks positions” when XPSC board is pushed
to the final position.
9. Check the correct installation of all the GenX boards and multiplier assembly (correct location
and final mechanical position) to the backplane XPSB board.
10. Install the rear panel of the GenX by screwing totally 4x screws.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.
.
Page 8 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
CAUTION! Caution during the M8 connector plug in: M8 female connector must be introduced in
parallel with the M8 male connector respecting the key lock.
5.4 Calibration
1. Store the new kV/mA measurement calibration parameters of the new XPSC board to the
OXSAS software. To store those new kV/mA parameters, read the firmware status 2048.
Note Only if alarms related to the precision of the kV/mA measurement appear, then perform GenX
calibration using the terminal mode command #XC 1.
CAUTION! Before to perform a GenX calibration kV/mA, the X-ray tube must be already
WARMED UP.
Use the following terminal mode command to calibrate the GenX generator:
Command Description
XR 10,10,1,2 This command is not needed if the X-ray power is already ON.
"Wait for tolerance" wording must be not present on the instrument display to continue to type
the following command.
CAUTION! For shipment of XPSC board, it must always be packed in a special packaging.
.
Page 9 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback
Revision Table
FEI Company, part of Thermo Fisher Scientific, work instructions are proprietary information and confidential. This procedure is property of Thermo
Fisher and for Thermo Fisher internal use only and must not be duplicated or disseminated for any third party without the express consent of hermo
Fisher. Printed or electronic copies of this procedure are uncontrolled and intended only for immediate use. The electronic files are the controlled
versions and are to be used as the master copies. Incomplete printed copies are not to be used and must be discarded. The Government's rights to use,
modify, reproduce, release, perform, display, or disclose these technical data are restricted to those rights specified in DFARS 252.227-7015(b)(2), FAR
52.227-14(g)(2)(Alternate II) and FAR 12.211. Any reproduction of technical data or portions thereof marked with this legend must also reproduce the
markings. Any person, other than the Government, who has been provided access to such data, must promptly notify Thermo Fisher.
.
Page 10 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument, XPSC board and GenX.
New kV/mA measurement calibration parameters are stored in the new XPSC board. Those new
kV/mA parameters will replace the ones from the original “old” XPSC board stored in the OXSAS
software.
Note No fan and no water block are installed on XPSC board.
CAUTION!It is important to verify that you have the correct XPSC board. Incorrect XPSC board will
damage the X-ray tube permanently.
The picture below indicates the XPSC board including places to check:
● Check the XPSC part Number - 1226741,
● Check Jumpers,
● The IC15 must have the blue label as well as the instrument's reset button.
.
Page 1 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
2 Safety
Note Do not perform any operation without proper safety training specific to the hazard(s).
CAUTION!
A Caution message indicates a potentially hazardous situation that, if not avoided, may
result in moderate or minor injury. It may also be used to alert against unsafe practices.
WARNING!
A Warning message indicates a potentially hazardous situation that, if not avoided, may
result in death or severe injury.
.
Page 2 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
DANGER!
A Danger message indicates an imminently hazardous situation that, if not avoided, will
result in death or severe injury.
WARNING!
Radiation danger – Indicates that X-Ray hazard may be present. Take necessary
precautions.
SERVICE!
A Service message indicates that the described procedure should be performed by
qualified personnel only.
3 Prerequisites
SERVICE! This document must be used ONLY by the person who followed the special training of
GenX generator and successfully obtained the certificate to service GenX generators.
CAUTION!Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the GenX generator.
.
Page 3 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
4 Preparation
Note 2) The secondary water cooling regulation can be set using the firmware status N° 2052.
.
Page 4 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
.
Page 5 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
5 Procedure
WARNING! Before working with the new GenX generator, please read carefully the following
information and cautions.
DANGER! LETHAL VOLTAGES are present in the GenX generator whenever the generator is
connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.
.
Page 6 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
Note For more information concerning the removal of the generator from the XRF instrument, consult the
GenX generator service manual P/N: AA83678, delivered during the certification training course given
in Ecublens by TTI department.
CAUTION! During the removal and installation of the GenX boards and Multiplier Assembly, always
put down the GenX generator with bottom side in contact with the surface table, see the
picture below. DO NOT turn the GenX upside down!
.
Page 7 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
2. WAIT 15 MINUTES to be sure that all stored energy is discharged after power disconnection.
DANGER! DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.
3. Unplug and plug action of all electronic boards must be delayed by at least 15 MINUTES.
CAUTION!Not respecting of this caution note can cause serious damage on the GenX backplane
board XPSB and body’s injury due to energy stored in the capacitor.
4. Remove the 4x screws fixing the rear panel of the GenX generator.
5. Remove the rear panel of the GenX generator.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.
6. At this step, you are able to remove the XPSC board: XPSC board is connected to a unique
backplane board XPSB located near the GenX front panel.
.
Page 8 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
CAUTION!IMPORTANT GENERAL NOTE: If you want to remove all the GenX boards located inside
the GenX, you must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed between them):
a) First board to be removed is -> XPSI board
b) Second two boards to be removed are -> XPSR (if installed) together with XPSC board.
c) Third board to be removed is -> XPSP board
d) Fourth “board” to be removed is -> Multiplier assembly (don’t forget the 3x locking
screws located at the bottom panel of the GenX generator).
7. To extract the XPSC board from its location, use your fingers to pull together the front connectors
J104, J407 and M8.
8. Install the new XPSC board in the correct location, see Location of the XPSC Board Inside the
GenX Generator on page 7. Avoid to damage electronic components located on the way during
installation of the XPSC board!
CAUTION!IMPORTATN NOTE: The XPSC board must be installed at the final locked mechanical
position, by passing through two “distinct clicks positions” when XPSC board is pushed
to the final position.
9. Check the correct installation of all the GenX boards and multiplier assembly (correct location
and final mechanical position) to the backplane XPSB board.
10. Install the rear panel of the GenX by screwing totally 4x screws.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.
CAUTION! Caution during the M8 connector plug in: M8 female connector must be introduced in
parallel with the M8 male connector respecting the key lock.
.
Page 9 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
5.4 Calibration
1. Store the new kV/mA measurement calibration parameters of the new XPSC board to the
OXSAS software. To store those new kV/mA parameters, read the firmware status 2048.
Note Only if alarms related to the precision of the kV/mA measurement appear, then perform GenX
calibration using the terminal mode command #XC 1.
CAUTION! Before to perform a GenX calibration kV/mA, the X-ray tube must be already
WARMED UP.
Use the following terminal mode command to calibrate the GenX generator:
Command Description
XR 10,10,1,2 This command is not needed if the X-ray power is already ON.
"Wait for tolerance" wording must be not present on the instrument display to continue to type
the following command.
CAUTION! For shipment of XPSC board, it must always be packed in a special packaging.
Revision Table
.
Page 10 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback
FEI Company, part of Thermo Fisher Scientific, work instructions are proprietary information and confidential. This procedure is property of Thermo
Fisher and for Thermo Fisher internal use only and must not be duplicated or disseminated for any third party without the express consent of hermo
Fisher. Printed or electronic copies of this procedure are uncontrolled and intended only for immediate use. The electronic files are the controlled
versions and are to be used as the master copies. Incomplete printed copies are not to be used and must be discarded. The Government's rights to use,
modify, reproduce, release, perform, display, or disclose these technical data are restricted to those rights specified in DFARS 252.227-7015(b)(2), FAR
52.227-14(g)(2)(Alternate II) and FAR 12.211. Any reproduction of technical data or portions thereof marked with this legend must also reproduce the
markings. Any person, other than the Government, who has been provided access to such data, must promptly notify Thermo Fisher.
.
Page 11 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument, XPSC board and GenX.
New kV/mA measurement calibration parameters are stored in the new XPSC board. Those new
kV/mA parameters will replace the ones from the original “old” XPSC board stored in the OXSAS
software.
Note No fan and no water block are installed on XPSC board.
CAUTION!It is important to verify that you have the correct XPSC board. Incorrect XPSC board will
damage the X-ray tube permanently.
The picture below indicates the XPSC board including places to check:
● Check the XPSC part Number - 1226755,
● Check Jumpers,
● The IC15 must have the blue label as well as the instrument's reset button.
.
Page 1 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
2 Safety
Note Do not perform any operation without proper safety training specific to the hazard(s).
CAUTION!
A Caution message indicates a potentially hazardous situation that, if not avoided, may
result in moderate or minor injury. It may also be used to alert against unsafe practices.
WARNING!
A Warning message indicates a potentially hazardous situation that, if not avoided, may
result in death or severe injury.
.
Page 2 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
DANGER!
A Danger message indicates an imminently hazardous situation that, if not avoided, will
result in death or severe injury.
WARNING!
Radiation danger – Indicates that X-Ray hazard may be present. Take necessary
precautions.
SERVICE!
A Service message indicates that the described procedure should be performed by
qualified personnel only.
3 Prerequisites
SERVICE! This document must be used ONLY by the person who followed the special training of
GenX generator and successfully obtained the certificate to service GenX generators.
CAUTION!Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the GenX generator.
.
Page 3 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
4 Preparation
Note 2) The secondary water cooling regulation can be set using the firmware status N° 2052.
.
Page 4 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
.
Page 5 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
5 Procedure
WARNING! Before working with the new GenX generator, please read carefully the following
information and cautions.
DANGER! LETHAL VOLTAGES are present in the GenX generator whenever the generator is
connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.
.
Page 6 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
Note For more information concerning the removal of the generator from the XRF instrument, consult the
GenX generator service manual P/N: AA83678, delivered during the certification training course given
in Ecublens by TTI department.
CAUTION! During the removal and installation of the GenX boards and Multiplier Assembly, always
put down the GenX generator with bottom side in contact with the surface table, see the
picture below. DO NOT turn the GenX upside down!
.
Page 7 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
2. WAIT 15 MINUTES to be sure that all stored energy is discharged after power disconnection.
DANGER! DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.
3. Unplug and plug action of all electronic boards must be delayed by at least 15 MINUTES.
CAUTION!Not respecting of this caution note can cause serious damage on the GenX backplane
board XPSB and body’s injury due to energy stored in the capacitor.
4. Remove the 4x screws fixing the rear panel of the GenX generator.
5. Remove the rear panel of the GenX generator.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.
6. At this step, you are able to remove the XPSC board: XPSC board is connected to a unique
backplane board XPSB located near the GenX front panel.
.
Page 8 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
CAUTION!IMPORTANT GENERAL NOTE: If you want to remove all the GenX boards located inside
the GenX, you must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed between them):
a) First board to be removed is -> XPSI board
b) Second two boards to be removed are -> XPSR (if installed) together with XPSC board.
c) Third board to be removed is -> XPSP board
d) Fourth “board” to be removed is -> Multiplier assembly (don’t forget the 3x locking
screws located at the bottom panel of the GenX generator).
7. To extract the XPSC board from its location, use your fingers to pull together the front connectors
J104, J407 and M8.
8. Install the new XPSC board in the correct location, see Location of the XPSC Board Inside the
GenX Generator on page 7. Avoid to damage electronic components located on the way during
installation of the XPSC board!
CAUTION!IMPORTATN NOTE: The XPSC board must be installed at the final locked mechanical
position, by passing through two “distinct clicks positions” when XPSC board is pushed
to the final position.
9. Check the correct installation of all the GenX boards and multiplier assembly (correct location
and final mechanical position) to the backplane XPSB board.
10. Install the rear panel of the GenX by screwing totally 4x screws.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.
CAUTION! Caution during the M8 connector plug in: M8 female connector must be introduced in
parallel with the M8 male connector respecting the key lock.
.
Page 9 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
5.4 Calibration
1. Store the new kV/mA measurement calibration parameters of the new XPSC board to the
OXSAS software. To store those new kV/mA parameters, read the firmware status 2048.
Note Only if alarms related to the precision of the kV/mA measurement appear, then perform GenX
calibration using the terminal mode command #XC 1.
CAUTION! Before to perform a GenX calibration kV/mA, the X-ray tube must be already
WARMED UP.
Use the following terminal mode command to calibrate the GenX generator:
Command Description
XR 10,10,1,2 This command is not needed if the X-ray power is already ON.
"Wait for tolerance" wording must be not present on the instrument display to continue to type
the following command.
CAUTION! For shipment of XPSC board, it must always be packed in a special packaging.
Revision Table
.
Page 10 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback
FEI Company, part of Thermo Fisher Scientific, work instructions are proprietary information and confidential. This procedure is property of Thermo
Fisher and for Thermo Fisher internal use only and must not be duplicated or disseminated for any third party without the express consent of hermo
Fisher. Printed or electronic copies of this procedure are uncontrolled and intended only for immediate use. The electronic files are the controlled
versions and are to be used as the master copies. Incomplete printed copies are not to be used and must be discarded. The Government's rights to use,
modify, reproduce, release, perform, display, or disclose these technical data are restricted to those rights specified in DFARS 252.227-7015(b)(2), FAR
52.227-14(g)(2)(Alternate II) and FAR 12.211. Any reproduction of technical data or portions thereof marked with this legend must also reproduce the
markings. Any person, other than the Government, who has been provided access to such data, must promptly notify Thermo Fisher.
.
Page 11 of 11 Confidential and proprietary information
Document: Installation procedure Multiplier Assembly S703758T
Multiplier Assembly
Instrument : XRF
N° Assembly : S703758T
File : S430420-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 1 / 10
Document: Installation procedure Multiplier Assembly S703758T
1 Introduction:
The Multiplier Assembly includes two different parts:
A) XPSM board
B) Multiplier box
Note: XPSM board and Multiplier box can not be delivered separately.
Those two parts must always remain connected together.
Don’t disconnect cables between those two parts.
2 Assembly configuration:
Air direction
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 2 / 10
Document: Installation procedure XPSC3660 board S703722T
COMPONENTS VIEW
Multiplier Assembly has 2x fans located on the side panel and connected to the XPSM board:
Fan identification: Connected to:
Fan 1
-> P1 connector
Installed on the side
Red color of the fan cable is connected to: +24VDC
panel of the Multiplier
Black color of the fan cable is connected to: PGND
Assembly
Fan 2 -> P1 connector
Installed on the side
Red color of the fan cable is connected to: +24VDC
panel of the Multiplier
Black color of the fan cable is connected to: PGND
Assembly
Note:
Air direction of the fans 1 and 2 are oriented to the XPSM board side.
3 Firmware compatibility
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 3 / 10
Document: Installation procedure XPSC3660 board S703722T
4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.
Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 4 / 10
Document: Installation procedure Multiplier Assembly S703758T
Connector
JP 1
J1
Water male
fast fitting
connectors
Rear side
Left
side
Front side
BOTTOM side
GENX Generator VIEW
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 5 / 10
Document: Installation procedure Multiplier Assembly S703758T
VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.
- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- Remove the top cover of the GenX generator by unscrewing totally 4x screws.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- Remove XPSI, XPSR (if installed), XPSC and XPSP boards by following the
below IMPORTANT General Note.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 6 / 10
Document: Installation procedure XPSC3660 board S703722T
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
Use the water fast fitting connector of the multiplier block to extract the Multiplier
Assembly from its location.
Pull the Multiplier Assembly slightly without a big force.
Use the opening of the top side generator to facilitate, with the help of your
fingers, the way out of the High Voltage transformer located on the Multiplier
Assembly.
Notes:
1) Multiplier Assembly (through XPSM boad) is connected to a unique
backplane board XPSB located near the GenX front panel.
2) Multiplier Assembly is mechanically secured by three screws located at
the bottom panel of the GenX generator.
- Install the new Multiplier Assembly in the correct location, see chapter 4.1 of this
document.
For only the first installation of the Multiplier Assembly in the GenX
generator, the two connectors J906, male side on the XPSM board and female
side on the GenX backplane board XPSB, must be well mechanically aligned.
See below procedure.
A) Insert the new Multiplier Assembly in the correct location, see chapter 4.1
of this document, at about 5cm before the final board location.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 7 / 10
Document: Installation procedure Multiplier Assembly S703758T
B) Unscrew of about ¼ turn the 5x nuts located on the XPSM board solder side.
See below picture:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 8 / 10
Document: Installation procedure Multiplier Assembly S703758T
E) Fix the XPSM board to the Multiplier block by tightening (not too strong)
of ¼ turn the 5x nuts located on the XPSM board solder side.
Follow the nuts order 1->2->3->4->5 as specified in the below picture:
1
3
5
4 2
- Install all the GenX generator boards respecting the following order:
-> First board to be installed is -> XPSP board
-> Second two boards to be installed are -> XPSR (if installed) together with
XPSC board
-> Third board to be installed is -> XPSI board
Take care to avoid to damage electronic components located on the way
during installation of the GenX generator board.
- Install the top cover of the GenX generator by screwing totally 4x screws.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 9 / 10
Document: Installation procedure Multiplier Assembly S703758T
- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.
4.3 Calibration:
N°. N° Mod
Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 10 / 10
Document : DOC,Update Firmware generator X-Ray
Instrument :
N° Ensemble :
Fichier : P:\Docu\ZAUGG\S428625-0.doc
INTRODUCTION
You have received the Firmwares kit S703xxx which includes all the firmwares necessary for the well function of
a X-Ray generator 3.6 kW limited at 3 kW for the instruments 9400 with S/N < 200 and 9800 with S/N < 100.
In this Firmware kit S703xxx, there are firmwares which are not used, depending of your instruments type and
configuration.
Note : the following firmware isn’t used in your 9400 instrument, you can throw it out :
XSH V3.x-x
Note : the following firmware isn’t used in your 9400 instrument, you can throw it out :
XCM V3.x-x
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 2 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 3 / 21 0
Document : DOC, Update Firmware generator X-Ray
AUTHOR : J. B. Masset
TABLE OF CONTENTS :
Subject description :
1. Introduction
2. Switches configuration on the electronic boards for 9400 instruments
3. Switches configuration on the electronic boards for 9800 instruments
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 4 / 21 0
Document : DOC, Update Firmware generator X-Ray
Subject number : 1
Introduction :
The purpose of this Technical News is to inform you of the switches configuration on the different
9400/9800 electronic boards.
This Technical News will be edited periodically in order to have always the latest update.
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 5 / 21 0
Document : DOC, Update Firmware generator X-Ray
Subject number : 2
Note : If you replace the display by a new type, on the 9400 instruments, you must upgrade all the firmwares to
the version V3.x-x.
Hereafter is the list of the last firmware version dated January 2000 :
XQM88 V3.0-0 S702906 ! This version needs 1 more RAM 32K x 8 ( A033453 )
XQM96 V3.0-0 S702907
XGT V3.1-0 S702908
*XCM V3.1-0 S702909 *Option : Magazine 12 positions
*XSH V3.0-2 S702990 *Option : Magazine XY
*XPS60RT V3.1-1 S703022 *Installed in 9400 instr. with S/N < 200
*XPS60 V3.1-1 S703020 *Installed in 9400 instr. with S/N >= 200
*XPS70RT V3.1-1 S703023 *Installed in 9400 instr. with S/N < 200. Option 70 kV
*XPS70 V3.1-1 S703021 *Installed in 9400 instr. with S/N >= 200. Option 70 kV
XGI V3.1-1 S703019 ! This version needs 1 more RAM 32K x 8 ( A033453 )
XGDI V3.1-0 S702911
XDI V3.1-0 S702912
XGDI
XCM
XGT
XSH
XGI
XDI
XQM88 V 3.0 a a a a a a a a
MASTER
NODE
XQM96 V 3.0 a a a a
XGI V 3.1 a a
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 6 / 21 0
Document : DOC, Update Firmware generator X-Ray
Note : In case of installation of a new displays kit S702928 ( P/N for old display type is : A035481 and new type is
S702777 ), put the switch S1-6 to ON position.
For information : the new display has been introduced since the instrument 9400 >= S/N 223.
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 7 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 8 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 9 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 10 / 21 0
Document : DOC, Update Firmware generator X-Ray
This firmware is installed on the XGI board S702005T or S702735T “ MMB-II part “.
XGI chip’s location : Two EPROMS : EPROM 1/2 in IC7 socket and 2/2 in IC11 socket.
Three RAM ( A033453 ) 32k x 8 in IC10 / IC15 / IC16 sockets.
XGI’s switches and jumpers function located on the XGI board “ Adaptation part “ :
This firmware is installed on the XGI board S702005T or S702735T “ Adaptation part “.
XGDI chip’s location : One EPROM in IC1 socket on the XGI board “ Adaptation part “.
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 11 / 21 0
Document : DOC, Update Firmware generator X-Ray
For your information, find hereafter the firmware compatibility table of the V2.x-x :
S LAV E N O DE S
V 2.0
V 2.1
V 2.2
V 2.3
V 2.4
V 2.0
V 2.0
V 2.0
V 2.1
V 2.2
XPS60/70(RT)V 2.3
V 2.1
V 2.0
V 2.1
V 2.2
V 2.0
V 2.0
V 2.1
XPS60/70
XPS60/70
XPS60/70
MASTER
XQM96
XQM96
XQM96
XQM96
XQM96
NODES
XGDI
XCM
XGT
XSH
XGI
XGI
XGI
XDI
XDI
XQ M 88 V 2 .0
XQ M 88 V 2 .1
XQ M 88 V 2 .2
XQ M 88 V 2 .3
XQ M 96 V 2 .0
XQ M 96 V 2 .1
XQ M 96 V 2 .2
XQ M 96 V 2 .3
XQ M 96 V 2 .4
XG I V 2 .0
XG I V 2 .1
XG I V 2 .2
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 12 / 21 0
Document : DOC, Update Firmware generator X-Ray
Subject number : 3
The purpose of this subject is to inform you of the switches configuration on the 9800 instruments
electronic boards.
Hereafter is the list of the last firmware version dated January 2000 :
XQI V3.3-1 S702950
XSM V3.0-1 S702879
XGT V3.1-0 S702908
XST V3.1-1 S702965
XSP V3.0-1 S702793
XMIR V3.0-0 S702643 *Option : Mono
*XCM V3.1-0 S702909 *Option : Magazine 12 positions
*XHI V3.2-2 S702959 *Option : Magazine XY
*XPS60RT V3.1-1 S703022 *Installed in 9800 instr. with S/N < 100
*XPS60 V3.1-1 S703020 *Installed in 9800 instr. with S/N >= 100
*XPS70RT V3.1-1 S703023 *Installed in 9800 instr. with S/N < 100. Option 70 kV
*XPS70 V3.1-1 S703021 *Installed in 9800 instr. with S/N >= 100. Option 70 kV
*XGI V3.1-1 S703019 *Option : Gonio
*XGDI V3.1-0 S702911 *Option : Gonio
*XDD V3.1-1 S702996 *Option : XRD
*XDDD V3.0-1 S702949 *Option : XRD
*XDI V3.1-0 S702912 *Option : Gonio or XRD
XDDD
XMIR
XGDI
XGDI
XCM
XCM
XSM
XDD
XDD
XGT
XST
XST
XSP
XHI
XHI
XGI
XGI
XDI
XDI
XQI V 3.0 a a r a r a a r r a r a a a r a r a r a a r
XQI V 3.1 a a r a r a a a r a r a a a r a r a r a a r
MASTER NODES
XQI V 3.2 a a r a r a a r a r a a a a r a r a r a a r
XQI V 3.3 a a a a a a a r a r a a a a a a a a a a a a
XSM V 3.0 a a a a a a a a a a a a
XGI V 3.0 a a a a
XGI V 3.1 a a a a
XDD V 3.0 a a a
XDD V 3.1 a a a
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 13 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 14 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 15 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 16 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 17 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 18 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 19 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 20 / 21 0
Document : DOC, Update Firmware generator X-Ray
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 21 / 21 0
8
DRAWINGS
Drawings Chapter 8
8 DRAWINGS
R22 +5V
2K7
+5V
C16 C88
C92 +12V 100nF J902
100nF
100nF R18 +5VRTM AGND
4
10K IC2 AGND R29 +24V 1 1Z
AGND R25 3 + IC5A R24 2 1A
18K 1 8 1 InputVoltage GND 3
AGND 10k C62 R32 2 - U1 4
1B
2 7 10nF 18k OP484 150 P1 +24Vin 5
1C
C51 1D
C93 R21 100nF 1 + 6
11
2Z
+24V C54 C60 3 6 2 7 2A
100nF R27 AGND AGND AGND - PGND 8
10K 100nF 10nF 4 5 18k AGND +12Vin 9
2B
R26 B WAGO 2P 2C
AGND C48 10
R30 FAN 24V 11
2D
10k -12Vin AGND 3Z
+5VM 47K AGND HCPL7800 10nF 12 3A
R28 PGND 13 3B
C49 270 14 3C
Low voltage 8x8bit -12V AGND +5Vin 15
100nF C94 3D
AD status converter DGND 16 4Z
100nF R34 17
IC3 /PwmTrip 4A
10K 18 4B
AGND 20 VCC CH0 1 AGND R38 19 4C
19 2 33K (Usually not intalled) 20
/AdcCs1 V+ CH1 4D
18 CS CH2 3 AGND D18 21 5Z
SdataOut 17 4 +15VRTM2 /PwmTrip 22
Sclk DI CH3 LampCurrent 5A
16 CLK CH4 5 C95 23 5B
AdcSars 15 6 NtoGndVoltage R16 IC4 BAT85 24
SdataIn SARS CH5 InputTemp1 100nF R41 5C
14 DO CH6 7 240K C50 R35 25 5D
13 8 InputTemp2 10K PTC3 8 CurrentLimit 26
Vref+5V R44 +5V SE CH7 Vref+5V 100nF Line1 6Z
12 VREF COM 9 AGND +15VRTM 330 27 6A
11 10 R3 R17 7 2 R69 28
10 AGND DGND
5.6/25°C +5VRTM 150K 1K 29
6B
D8 R51 6C
ADC0838 R20 -5.6VRTM2 6 3 R57 30 6D
C85 AGND DGND 0 AGND 31
C84 10 7Z
10uF 5 32 7A
1nF 1N4007 220 100K +5V 33
T1 D9 C89 7B
1 6 C57 34 7C
AGND AGND D20 C56 C44 D10 100uF/35V 1N5352B R67 HCPL3101 100nF 35
C97 7D
2200uF/25V 100uF/35V 1N5339B 10K 6 36 8Z
7 C35 +5VM 470nF AGND Line2 37
8
1N4007 8A
9 4 IC8B R54 38 8B
R58 5 Vref+5V 39
C99 1nF Q2 + 8C
1nF M2 7 10K 40 8D
5 10 C66 C67 500V 500V R23 MPSA05 6 41
2200uF/25V D13 8 VUM 33-05 10K 2K2 -
42
9Z
2x15V 100uF/35V OP284 9A
D12 PTC4 1N5339B MOV3 7 43
4
InputTemp2 9B
-5.6VRTM 2 44 9C
45 9D
1N4007
+
Line1 5.6/25°C S20K250 C47 DGND AGND C91 R59 R55 R56 46 10Z
RV2 100nF 1nF 470 1K 10K 47
Line2 D19 1 3 48
10A
BYT30 4K7/25° 10B
PTC5 49 10C
+15VRTM2 AGND AGND AGND AGND 50 10D
AGND 51
5 -
+400VRTM 11Z
D6 5.6/25°C 52 11A
1 T2 6 I2 C53 D7 53
SENS2 11B
2200uF/25V 1N5352B +12V 54 11C
1N4007 55
C43 100nF 11D
7 5 1 56 12Z
9 57
SENSOR
C42 100nF 12A
Wires (2 x 0.2mm) 2 58 12B
U/I
D14 D11 59
C69 only for water R15 0.08V/A InputCurrent 12C
0.16V/A
0V OUT
5 10 1N5339B 3 -12V AGND 60 12D
2x15V PTC6 2200uF/25V cooled version 61
C37 1N4007 470 13Z
-5.6VRTM2 6 4 C41 62 13A
R60 D21 63
5.6/25°C AGND 1nF 470 BZX85C5V1 64
13B
13C
22nF SCD20PBR 65 13D
66
3
14Z
AGND AGND AGND 67 14A
FI2 +15VRTM 68 14B
L6 69 14C
FPP 2-45 250uH IC1 70 14D
C46 R31 71 15Z
8 CurrentControl +200VRTM 72
100nF 15A
330 73 15B
7 2 R68 +5V 74 15C
PGND +24V R10 1K 75 15D
-5.6VRTM 6 3 76
4
5
2
C19 16Z
10 77 16A
2 5 100nF 78
1 D16 +5VM DGND 79
16B
P2 R66 16C
1.5K220A R36 80 16D
GND B WAGO 2P 10K HCPL3101 +5VRTM IC7 AGND 5 + IC5B 81
R52 Mon400V 17Z
7 82 17A
Not mounted R4 1 8 10k C70 R39 6 - 83
150 17B
+
OP484 R49 84
-
2
7
8
4
6
6 32A 19D
+10% 5
10nF
I1
0 1 C13 C21 C15 C14 C25 R5 MOV5
96
97
20Z
230V -15% 4
3
2
Earth
GND
GND C23
C2
2.2uF
C1
2.2uF
C45 C68
+24V 5 - +
1000uF 1000uF 1000uF 1000uF 1000uF
250V 250V 250V 250V 250V
10K/5W S20K250 98
99
20A
20B
20C
2.2uF 2.2uF +200VRTM 100
1 Neutral 10nF 20D
101 21Z
WAGO D1 102 21A
MOV2 1N4007 C28 C34 C31 C30 C29 R7 MOV6 103
M1
3
21B
S20K320 1000uF 1000uF 1000uF 1000uF 1000uF 10K/5W S20K250 104 21C
VUM 33-05 250V 250V 250V 250V 250V 105
0VRTM 21D
106 22Z
GND 107 22A
108 22B
109 22C
110 22D
R6 111 23Z
D3 Q1 PowerOn 112 23A
1N4007 MPSA05 +5V 113 23B
2K2 114 23C
115 23D
C18 116 24Z
PGND 100nF 117 24A
+5VRTM 118 24B
IC6 R37 119
IC5C 24C
AGND 10 + R50 120 24D
R11 1 8 8 NtoGndVoltage 121
10k 25Z
R13 3M9 C63 R33 9 - 122 25A
2 7 10nF 18k OP484 150 123
C80 25B
124 25C
3M9 3 6 100nF 125
C59 C40 C58 C39 C74 AGND 126
25D
D4 R14 R12 26Z
1N4007 10nF 10nF 100nF 4 5 R43 AGND 127 26A
2K7 2K7 Vref+2V 128 26B
18k 129 26C
PTC2 GND 1000uF/6.3V 1000uF/6.3V HCPL7800 R40 130 26D
250V/0.1A C71 131 27Z
R53 10k 132 27A
AGND 10nF 133 27B
F2 100K 134 27C
0.8A C96 135 27D
1nF 136
SENS1 28Z
+12V 137 28A
D15 138 28B
5 1 C32 100nF 139
FI1 28C
140
SENSOR
J3 R8 142
Neutral IC5D /AdcCs1 R65 29A
D5
0V OUT
DIN41612
XPSI, X-ray Power Supply Input Board Echelle Dessiné MBz 03.10.00 Remplacé par
Contrôlé
Remplace
Format No
ThermoFisher S112064
Mod. No
S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 12
+5V
+5V
C16 C88
100nF J902
100nF
+5VRTM AGND
4
IC2 AGND R29 +24V 1 1Z
3 + IC5A R24 2 1A
1 8 1 InputVoltage GND 3
10k C62 R32 2 - 4
1B
2 7 10nF 18k OP484 150 P1 +24Vin 5
1C
C51
FAN 1
1D
R21 100nF 1 3 + 6
11
2Z
C54 C60 3 6 2 4 7 2A
AGND AGND AGND - PGND 8
100nF 10nF 4 5 18k AGND B WAGO 2P +12Vin 9
2B
R26 C48 2C
10 2D
11 3Z
HCPL7800 10k 10nF -12Vin AGND 12 3A
R28 PGND 13 3B
270 14 3C
AGND +5Vin 15 3D
DGND 16 4Z
17 4A
CurrentLimit /PwmTrip 18 4B
19 4C
R78 20 4D
21 5Z
+15VRTM2 22 5A
100K +5V 23 5B
R16 C89 24
IC4 5C
R70 240K R35 100nF 25 5D
+5VRTM 8 26 6Z
C50 AGND Line1 27
8
R69 330 6A
220 R17 R51 100nF 7 2 IC8A 28 6B
150K 1K + 3 29 6C
C102 D22 6 3 R79 R57 1 R77 30
47uF/35V 10 2 31
6D
1N5339B +5V - 7Z
IC9 PTC3 5 2K2 LM393N 32 7A
+15VRTM 33K 10K 33
4
7B
3 ON/OFF n.c. 5 ACPL-312T 34 7C
5.6/25°C C101 -15VRTM2 R75 R76 35 7D
L7 6 47uF/35V C106 C97 11K 10K 36
MF5 +VOUT 1uF Line2 8Z
2 +VIN +5VM 470nF AGND 37 8A
Com 7 IC8B R54 38 8B
MF-R110 C103 R58 5 Vref+5V 39
Q2 + 8C
220uF/35V 1 -VIN -VOUT 8 C104 7 40 8D
47uF/35V C105 R23 MPSA05 6 3K9 41
D23 W08G TMR3-2423WI PTC4 1uF 10K 2K2 LM393N
-
42
9Z
+
9A
-15VRTM 43 9B
InputTemp2 C107 44
C91 R59 R55 R56 9C
5.6/25°C 100nF 45
2
9D
Placer les condos 1uF Q3 DGND 100pF 1K5 1K5 2K7 46 10Z
C105, C106, C110, C111 C108 RV2 C47 47 10A
1 4K7/25° 100nF 48
-
3
10D
+15VRTM2 500V AGND AGND 51 11Z
3 5 R67 R71 +400VRTM 52
ON/OFF n.c. 11A
5.6/25°C C109 10K 1k +12V 53 11B
6 47uF/35V C110 SENS2 54
+VOUT 1uF 11C
11
2 55
8
+VIN 11D
Com 7 5 1 C43 100nF 56 12Z
57
SENSOR
12A
T1 1 -VIN -VOUT 8 C111 Wires (2 x 0.2mm) 2 C42 100nF 58 12B
C112
U/I
47uF/35V only for water R15 59 12C
1uF 0.16V/A InputCurrent
0V OUT
230V-15V TMR3-2423WI PTC6 D19 3 0.08V/A 60 12D
STTH6006W cooled version 61
-15VRTM2 470 13Z
6 4 62
1
15C
FPP 2-45 P2 S20K320 4.7 R10 1K 75 15D
6 3 76 16Z
B WAGO 2P MOV7 4.7 77 16A
5 C19 78
4
3
16B
Not mounted DGND 100nF 79 16C
250uH
+
80
-
Mon400V 17Z
MOV4 7 82 17A
1 8 AGND 10k 6 - 83 17B
GND OP484 150 84
R4 17C
S20K320 2 7 C70 R39 85
10K C83 86
17D
C79 10nF 18k R49 18Z
C35 R48 3 6 100nF 87
InputTemp1 C75 18A
R19 10nF 270 18k 88 18B
GND F3 K1 270K 1uF 4 5 89 18C
0.8A 24V 1nF 500V R42 AGND AGND 90 18D
5 RV1 C3 C76 AGND 91
MOV1 0VRTM 19Z
13 6 4K7/25° 100nF HCPL7800 10nF 92 19A
S20K320 F1 4 14 10k 93
J2 Line 12 3 R47 AGND 94
19B
19C
6 C8 32A 11 AGND AGND 200K 95 19D
+10% 5
10nF 1 8
D24
STTH6006W
D26
STTH6006W D27
96
97
20Z
230V -15% 4
3
2
Earth
GND
GND C23
C2
2.2uF
C1
2.2uF
C45 C68 9 16
+24V
+400VRTM 98
99
20A
20B
20C
1 2.2uF 2.2uF D1 STTH6006W 100 20D
Neutral 10nF I1 D28 D29 101
MOV5 21Z
WAGO STTH6006W STTH6006W C13 C21 C15 C14 C25 R5 102 21A
MOV2 1N4007 D16 1000uF 1000uF 1000uF 1000uF 1000uF 10K/5W S20K250 103
2
21B
S20K320 1.5K220A Q5 Q4 250V 250V 250V 250V 250V 104 21C
C115 C114 105 21D
1 1 +200VRTM 106 22Z
GND 107 22A
IKW75N60T
IKW75N60T
1nF 1nF 108
3
22B
109 22C
D17 R73 R74 R66 R72 C28 C34 C31 C30 C29 R7 MOV6 110 22D
R6 1.5K250A 10K 1k 10K 1k 1000uF 1000uF 1000uF 1000uF 1000uF 10K/5W S20K250 111 23Z
D3 Q1 PowerOn 250V 250V 250V 250V 250V 112 23A
1N4007 MPSA05 113 23B
2K2 0VRTM 114 23C
115 23D
116 24Z
DGND 117 24A
+5V 118 24B
119 24C
120 24D
121 25Z
C18 122 25A
+5VRTM 100nF 123 25B
IC6 R37 124
IC5C 25C
AGND 10 + R50 125 25D
R11 1 8 8 NtoGndVoltage 126 26Z
R13 3M9 10k 9 - 127 26A
2 7 OP484 150 128 26B
C63 R33 C80 129 26C
3M9 3 6 10nF 18k R43 100nF 130
C58 C39 C74 26D
D4 131 27Z
1N4007 R14 C59 C40 R12 10nF 4 5 132
2K7 100nF 27A
PTC2 2K7 1000uF/6.3V 10nF AGND 18k 133 27B
250V/0.1A AGND 134 27C
GND 1000uF/6.3V HCPL7800 R40 C71 135 27D
136 28Z
137 28A
R53 AGND 10k 10nF 138 28B
F2 139 28C
0.8A Vref+2V 140
100K 141
28D
C96 1nF 5x100 29Z
+12V 142 29A
SENS1 +5VM /AdcCs1 R65 143
SdataOut R64 29B
D15 144 29C
5 1 C32 100nF Sclk R63 145
FI1 AdcSars R62 29D
146
SENSOR
3 5 3 13 - D5 149 30C
Earth 14 LampCurrent PowerOn 150
X-ray ON Lamp 2
1
Line GND 2 1 6 4 20K 12 +
C92 +12V 151
30D
31Z
-12V OP484 1N4148 152 31A
IEC FPP 2-45 C20 C100 R9 C38 100nF R18 153
Vref+5V 31B
SCD01PBR 10nF 10nF 100K 10uF/6.3V 10K 154 31C
AGND R25 Vref+2V 155
InputCurrent 31D
18K 156 32Z
AGND InputVoltage 157
Mon400V 32A
AGND AGND AGND AGND AGND AGND AGND 158 32B
C93 159 32C
+24V GND 160
100nF 32D
R27
10K
AGND
R30
+5VM 47K AGND DIN41612
Format No
Hole Hole
ThermoFisher S112064
Mod. No
S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 15
J104 J904
1A 1 +5V
1B 2 GND R89
3 0.266V/A FilamentCurrent W3 1 1Z
1C R10 IC1A EnWd
2A 4 /XgtReset 15K R87 GND 2 1A
5 /XcmReset D3 1K 1 C113 FilCurMon Test 3 1B
2B IC2A
6 /XpsReset * 3 D1 C24 R15 IC18B +5V 4 1C
2C
7 /XshReset 2 2 1 WdToggle 47nF 8 MaxFilCur 1K C84 LD3 +24Vin 5 1D
3A BAT85 R104 R7
3B 8 /Xgi1Reset 1 R70 100uF 6 2Z
9 /Xgi2Reset 7403 BAT85 1uF 1K +12V 7 Pressostat/24V 7 2A
3C R5 D2 470K 35V 390 PGND
4A 10 /XdiReset 74HC14 R6 C6 DspOk 8 2B
11 /XxxReset 1K 1M 1N4148 IC20 CMP401 1K AGND /DspCs Green +12Vin 9 2C
4B 470nF
4C 12 /XxxReady C111 C107 10 2D
13 MonosReady 1 -EA 16 C88 10nF CurrentLimit DGND +5V SsioRxSync 11
5A VREF5.1V 10nF -12Vin AGND 3Z
14 MmbTxd DGND DGND DGND DGND 12 3A
5B
15 MmbRxd /Reset FilCurRef R75 R82
2 +EA 15 10nF CurrentControl C57 SsioTxSync 13
5C DisplayRd2 VIN +5V DGND 3B
6A 16 10K 47K 3 +5V AGND AGND 14 3C
17 +5V C8 14 AGND FilPwmH DspWdToggle +5Vin 15
6B
18 XgnR+ 47nF IC2B C95 SYNC OUTPUT B
+5V +5V 10nF C65 3D
16 4Z
6C Reset DGND
7A 19 XgnR- 3 4 10nF 4 OSC OUT VC 13 C93 C85 17 4A
20 XgnT+ DGND AGND /PwmTrip 18 4B
7B 10nF 10nF
21 XgnT- 5 CT 12 10nF 19
4
7C 74HC14 GND 4C
22 /SpAdcInt AGND 20 4D
108
107
106
105
104
103
102
101
100
8A XpsReady FilPwmL
98
91
90
99
97
96
95
94
93
92
89
88
87
86
85
84
83
81
80
79
78
77
76
75
74
72
70
69
68
67
66
65
64
62
61
60
59
58
71
82
73
63
8B 23 /Cs0 6 RT OUTPUT A 11 AGND IC18A AGND R67 21 5Z
24 /Cs1 +24V R148 D13 6 HvPwmH 22 5A
8C XsnR1+ +5V DGND
25 7 10 2 23
EIA
GND
GND
VDD
VDD
GND
GND
GND
EIB
EIS
EIZ
ETU0
ETU1
AUX0
AUX1
P0
P1
P2
P3
P4
P5
P6
P7
P8
P9
P10
P11
D0
D1
D2
D3
D4
D5
D6
D7
D8
D9
D10
D11
D12
D13
D14
D15
D16
D17
D18
D19
D20
D21
D22
D23
9A XsnR1- XrayLampC Q2 DISCH. SHUTDOWN SafFilCur 10K HvPwmL C151 5B
26 5 24 5C
9B
27 TGND 180 8 SOFT-START 9 1N3595 D17 C45 56 10nF 25
9C XsnT1+ R107 COMP. C99 AH R68 FilPwmH 5D
10A 28 IRF9Z24 CMP401 C104 AL 55 0 26 6Z
29 XsnT1- 10K 1N3595 10nF 1uF 109 54 Not mounted 27
C108
14
13
10B R112 R97 C101 CONVST BH 6A
30 TGND SG3525AN 10nF 110 GND BL 53
R64 0 FilPwmL 28 6B
10C
31 XpsIntOut C156 C155 C154 C153 20K C83 10nF AGND AGND +5VM 111 IC5 Not mounted DGND 29
11A Pressostat/24V 10K Q3 R110 XrayLampOn VDD 6C
11B 32 1nF 100uF 112 GND GND 52 30 6D
11C 33 C168 1nF 1nF 1nF 1nF MPSA05 10K AGND AGND 113 AVDD ADMC401 VDD 51 C72 R136 31 7Z
34 10nF AGND AGND AGND R98 /MaxFilCur C114 114 AVSS 100 32 7A
12A
35 D11 10nF AGND 115 VIN7 +5V DGND
12B XdiHvpsIlk 10K CH 50 10nF
33 7B
12C 36 Synchro DGND +5V AGND 1N3595 116 VREF CL 49 34 7C
37 Dvm+ PGND FilCurSD -5V AGND C7 C15 Hv 117 VIN6 48 10K 35 7D
13A XrayLamp1 R109 10K Lamp1 PWMSYNC R61
13B 38 Dvm- AGND 118 REFCOM PWMTRIP 47 +5V R49 36 8Z
39 XgtReady R115 XrayLamp2 Lamp2 /EnFilCu 10uF 100nF 119 VIN5
13C C9 AGND PWMPOL 28 +5V 37 8A
14A 40 XcmReady 3K3 AGND 120 CAPT PWMSR 22 +5V 10K 38 8B
41 XpsReady R114 121 VIN4 R41 10K 39 8C
14B 10K D19 D18 100nF C28 C16 R1322
14C 42 XshReady 122 BSHAN GND 46 40 8D
43 C152 2xBZX55C4V7 123 ASHAN
15A Xgi1Ready
1nF
PTC2 PTC1 VDD 45 C71 41 9Z
15B 44 Xgi2Ready C965 C10 100nF 10uF R9 22 124 VIN0 +5V DGND 42 9A
45 XdiReady C965 125 44 43 9B
15C
46 Sd0 DGND AGND CAPB
126 VIN1
SCLK0 10nF DspDisp
16A 100nF C26 TFS0 43 44 9C
16B 47 Sd1 PGND DGND +5V AGND 127 CML RFS0 42 R56 45 9D
16C 48 Sc IC23 100nF 128 VIN2 DT0 41 DGND 46 10Z
17A 49 EN1 1 129 GAIN DR0 40 10K 47 10A
50 +12V 130 VIN3
17B +12VP
+24Vin C21 C27 SCLK1 39 TfsSync
48 10B
17C 51 +12VP GND 13 PGND 131 SENSE TFS1/IRQ1 38 49 10C
18A 52 +12VP PrWaValve 3 OUT1 C70 IC6A 132 AVSS RFS1/IRQ0/SROM 37 Rfs1 50 10D
53 +12VP 12 100nF 100nF 133 36 DspDt1 51 11Z
MF5 R37 OP495
4
18B GND +24VIntOn AVDD DT1/F0 PROMorEEPROM
54 +12VP XpsIntOut C31 35 DspDr1B 52 11A
PWDACK
18C DR1B/FI
CLKOUT
AGND 10nF DspSclk1
BMODE
55 4 + 3 AGND 34 1 DATA 2 53 11B
RESET
19A GND 10K DR1A/F1 CLK
MMAP
CLKIN
56 WaterPump 6 OUT2 MF-R110 Vref+2V 1 C148 C149 C150 C147 6 CEO RST/OE 3 54 11C
XTAL
PWD
GND
GND
GND
GND
GND
GND
GND
DMS
BGH
POR
BMS
PMS
VDD
VDD
VDD
19B C54
A13
A12
A11
A10
RN2A 100nF
WR
57 100nF 100nF R52 /DspPCe
GND 5 - 2 CE 4 55 11D
BG
RD
NC
NC
Pressostat 100nF 100nF
BR
A9
A8
A7
A6
A5
A4
A3
A2
A1
A0
19C 2K7 100nF
20A 58 PGND Via XGT C36 C20 10K VPP 7 56 12Z
59 PrCoolCtrl
IN1 2 R42 5 GND VCC 8 57 12A
11
134
135
136
138
137
140
139
15
57
141
142
143
144
1
2
3
4
6
7
8
10
11
12
5
9
13
14
16
17
18
19
20
21
23
24
25
26
27
29
30
31
32
33
20B WaterCooler Pressostat/24V Int1 C43 R21 +5V
20C 60 11 OUT3 DGND DGND 58 12B
61 7 WaPump 10nF AGND AGND AGND AGND 100uF IC13 37LV65 1K 59 12C
21A
62 -12VP
IN2 10K AGND 10nF C44 C56 C64 DGND 60 12D
21B RN2B C58 orXC1765D
63 -12VP 10 WaCooler -12V R121 R122 R123 R120 AGND +5VM DGND R46 orAT17C65 C52 61 13Z
12.96MHz
8
7
6
5
4
3
2
1
30B
90 DGND PrWaTOut IntlkRtn DGND DGND IntRet C105 R92 100K CuArcTh 11 R28 R35 R40 88 18B
30C WaterTIn EmissionCurrent- /CuArc
31A 91 10K RN2G 9 - R103 15 R1544K7 2K7 18K 47K 89 18C
92 WaterTOut 8 12 +5VM 90 18D
31B
93 2K7 C73 1uF 10K 10 + 91 19Z
31C
94 C136 C124 C130 C142 100nF D12 IC19C 10K CMP401 4K7 AD status converter R39 R45 92 19A
32A D16 R31 C159 C160 C161
95 1N3595 AGND LF347N BZX55C5V1 C38 Low voltage 8x8bit 10K 10K 10K 100nF 93 19B
100nF
Tms
R155 100nF
Tclk
32B
96 GND 1uF 1uF 1uF 1uF DGND DGND 94 19C
Tdo
Tdi
32C
DGND 10nF IC8 95 19D
DIN41612 AGND 20 VCC CH0 1 AGND AGND AGND AGND AGND AGND 96 20Z
AGND TP8 AGND 19 V+ 2 97 20A
FilamentCurrentRef /AdcCs2 CH1
+12V AGND 3X 1N4148 18 CS CH2 3 98 20B
Vref+5V DGND 0.266V/A SdataOut 17 DI 4 99 20C
Sclk CH3
16 CLK CH4 5 -12V 100 20D
FilCurRef AdcSars 15 SARS
IC6C D10 D8 D7 CH5 6 101 21Z
2
11
10K R101 D15 R100 R12 5 6 1 2 43 I/O TMS 32 AGND AGND C82 C77 ADC0838 107 22A
1
4
100nF IC16 R66 AGND DGND AGND 10K 100nF 108 22B
AD15 A7 IC15 10uF 1nF R51
W1 2 I/O I/O 14 A6 1K 109 22C
VDD
VREF
1K 1N4148 C118 18K R93 10K 74HC14 AD14 A0 AD0 AD7 FilCurRefMon
AGND R16 3 I/O I/O 15 /DataCs 12 A0 O0 13 AD1 13 DB7 VOUT8 2 AGND 15K 110 22D
100uF C40 AD13 4 I/O A1 AD6
C115 100K C13
AD12 I/O 39 /CodeCs A2
11 A1 O1 14 AD2 AD5
14 DB6 VOUT7 3 AGND AGND 111 23Z
AGND R95 1uF 67 XTAL1 P1.0/EPA0/T2CLK 57 5 I/O I/O 38 /WrDac0 10 A2 O2 15 AD3 15 DB5 VOUT6 4 112 23A
AGND 100uF 15K 10pF X1 56 /HvArc AD11 6 37 A3 9 17 AD4 16 5 R55 C163 113 23B
P1.1/EPA1 /CuArc AD10 I/O I/O /PPI2Cs A4 A3 O3 AD4 AD3 DB4 VOUT5 HvArcTh 100nF
10K DGND DGND 16MHz P1.2/EPA2/T2DIR 55 7 I/O I/O 36 8 A4 O4 18 17 DB3
FF40H VOUT4 6 10K 114 23C
R91 AGND AGND C41 66 XTAL2 54 /MaxFilCur AD9 8 I/O 17 4MHz A5 7 A5 19 AD5 AD2 18 DB2 - 7 CuArcTh +5VM Vref+5V 115 23D
C100 R94 LF347N P1.3/EPA3 FilCurSD AD8 I/O O5 VOUT3
+ 5 10pF P1.4/EPA4 53 9 I/O I/O 31 /Ppi1Cs A6 6 A6 O6 20 AD6 AD1 19 DB1 FF48H VOUT2 8 MaxFilCur AGND 116 24Z
WaResist 7 WdToggle /Rd A7 AD7 AD0 SafFilCur
10K R84 P1.5/EPA5 52 /AdcDor0 /Ale I/O 30 Ready AD8
5 A7 O7 21 20 DB0 VOUT1 9 Miscellaneous 8x8bit 117 24A
5M6 IC19B - 6 28 NMI P1.6/EPA6 51 AdcEoc0
16 I/O I/O 29 ClkOut AD9
27 A8 C37 AD converter 118 24B
1uF AGND P1.7/EPA7 50 CLK1/I1 33 26 A9 Miscellaneous 119 24C
GND
10nF
VSS
-WR
C112 /Reset DGND 0 27 AD10 23 8x12bit IC7 120 24D
A2
A1
A0
R102 RESET AD0 AD11 A10 MaxFilCur
AGND P3.0/AD0 26 M4A5-64/32 XPSC01L 25 A11 DA converter 20 VCC CH0 1 121 25Z
8x100 25 AD1 AD12 4 A12 AD7228 AGND 19 V+ 2 SafFilCur 122 25A
21
22
23
24
12
10
R88 +12V 1M P3.1/AD1 AD2 AD13 /WrDac0 /AdcCs1 CH1
220nF PrimaryWaterTemperatureInput R140 40 P0.0/ACH0 P3.2/AD2 24 28 A13 18 CS CH2 3 Lamp1 123 25B
C116 PrimaryWaterTemperatureOutput R141 41 P0.1/ACH1 AD3 AD14 A2 SdataOut 4 Lamp2
1M P3.3/AD3 23 AD4 IC17 AD15
29 A14
A1 -5V Sclk
17 DI CH3 124 25C
AGND DeionizedWaterTemperatureInput R142 42 P0.2/ACH2 P3.4/AD4 22 3 A15 16 CLK CH4 5 FilCurMon 125 25D
R143 43 P0.3/ACH3 AD5 AD0 A0 A0 C106 AdcSars XsnR1+
DeionizedWaterTemperatureOutput P3.5/AD5 21 3 D1 Q1 2 DGND 2 A16 AGND 15 SARS CH5 6 FilCurRefMon 126 26Z
4
100nF DEIONIZED WATER RESISTIVITY DeionizedWaterLevel R144 44 P0.4/ACH4 AD6 AD1 A1 SdataIn XsnR1-
P3.6/AD6 20 AD7 AD2
4 D2 Q2 5 A2 /CodeCs
1nF 14 DO CH6 7 127 26A
3 + R72 D5 R73 WaterResistivityDriver R145 45 P0.5/ACH5 P3.7/AD7 19 7 D3 Q3 6 22 CE 0000H R59 +5V 13 SE CH7 8 128 26B
1 DeionizedWaterResistivity R146 46 P0.6/ACH6 AD3 8 D4 9 A3 /Rd 24 - AGND Vref+5V 12 9 XsnT1+ 129 26C
Vref+2V5 AD8 AD4 Q4 OE VREF COM
2 - 15K R147 47 P0.7/ACH7 P4.0/AD8 18 13 D5 Q5 12 A4 1 VPP DFFFH 10 11 AGND DGND 10 XsnT1- 130 26D
M8 IC19A 1K 1N4148 R79 17 AD9 AD5 14 D6 15 A5 31 PGM 131 27Z
PrWaTIn LF347N C97 C158 XsnTxD P4.1/AD9 AD10 AD6 Q6 A6 +5V
1 15K 32 P2.0/TXD P4.2/AD10 16 17 D7 Q7 16 A7 ADC0838 +5V 132 27A
11
1
2
3
4
22K R65 C11 8922A AD7 27 D7 37 /WaCoOn AD7 27 37 22K C167
MM74HC423A
2
4
6
8
1nF R118 PA7 D7 PA7 10nF
7K5 100nF T=100ms 6 C
6
8
7
6
5
1
3
5
7
ExternalRelay 35 RESET /PrWaFlow SdataOut DspDr1B
Q4
/XsnTxDEn
XpscNotOk/Red
/Ppi1Cs PB4 22 /DeWaFlow /Ppi2Cs
35 RESET PB4 22 /DspPCe
10
RfsSync
4 A2 Y2 16 Rfs1
MM74HC423A
MPSA05 2K2 5 74HC14 6 CS PB5 23 6 CS PB5 23 9 6 A3 Y3 14 T=100us
24 Reset 24 EnDisRd2 8 DspDt1 8 12 Sdatain
XsnR1+ 11 PB6 PB6 EnXpscDis A4 Y4 ExSclk
PB7 25 PB7 25 7 +5V 11 A5 Y5 9
7 6 13 A6 ExSdataOut
PGND
XsnR1- 12
Address
BuzzerOn /DspCs Y6 7
Ppi1Cs FFC0...FFFF FFC0H PC0 14 FF80H PC0 14 5 15 A7 Y7 5
PrCoolCtrl /AdcCs1Ext R119
Ppi2Cs FF80...FFBF - PC1 15 - PC1 15 4 17 A8 Y8 3
8922A WrDac0 FF40...FF7F FFFFH O PC2 16
XrayLampOn
/DspRst
FFBFH O PC2 16
/AdcCs2Ext
/AdcCs3Ext
3
/DspCs
10K
IC1B
DataCs E000...FF3F PC3 17 ExternalRelay PC3 17 2 1 1OE
EnDisRd2 IC1C
CodeCs 0000...DFFF PC4 13 WaPump PC4 13 TfsSync
1 19 2OE
5 12 12
DisplayRd2 R19
6
4 8
10 O PC5
PC6 11
10
WaCooler
/EnFilCu
O PC5
PC6 11
10
SsioRxSync
RfsSync
9*10K
DGND DGND
SN74ACT244 Mod: 1 Version pré-série 07.08.01 PPa 4 Version série 10.03.04 PPa 7 MOD23765 03.03.05 PPa
C34 100 9 XsnTxD PC7 PC7
1nF 2 Version pré-série (2) 13.11.01 PPa 5 OM3753 17.12.03 PPa 8 OM4012 04.05.07 PPa
7403 82C55 82C55
+5V 7403 Reset 3 Version "Phase-in" 09.09.02 PPa 6 OM3834 28.10.04 PPa 9 OM4049 10.04.08 PPa
R17
DGND 10K
Dessiné 15.11.00 PPa Remplacé par:
R20
10K
IC1D
DGND
+5V
TP5 XPSC, X-ray Power Supply Control Board Echelle
2
4
6
R89 FilamentCurrent
0.266V/A W3 1 1Z
R10 IC1A EnWd Test 2 1A
D3 1K C113 15K R87 FilCurMon GND
1 IC2A 3 1B
1
3
5
/GenxReset 3 D1 C24 R15 IC18B +5V 4 1C
2 2 1 WdToggle 47nF 8 MaxFilCur 1K C84 LD3 +24Vin 5 1D
BAT85 R104 R7
1 R70 100uF 6 2Z
7403 BAT85 1uF 1K +12V 7 Pressostat/24V 7 2A
R5 D2 470K 35V 390 PGND
74HC14 R6 C6 DspOk 8 2B
1K 1M 1N4148 IC20 CMP401 1K AGND /DspCs Green +12Vin 9 2C
470nF
C111 C107 10 2D
1 -EA 16 C88 10nF 10nF CurrentLimit DGND +5V SsioRxSync 11
VREF5.1V -12Vin AGND 3Z
DGND DGND DGND DGND 12
/Reset FilCurRef R75 R82
2 +EA 15 10nF CurrentControl C57 SsioTxSync 13
3A
VIN +5V DGND 3B
10K 47K 3 +5V AGND AGND 14 3C
C8 FilPwmH DspWdToggle +5Vin
SYNC OUTPUT B 14 AGND
10nF C65
15 3D
47nF +5V IC2B C95 +5V +5V 16 4Z
3 4 Reset 10nF 4 OSC OUT DGND
VC 13 C93 C85 /PwmTrip
17 4A
DGND AGND 18 4B
R115 5 CT 10nF 10nF
74HC14 GND 12 10nF 19 4C
4
3K3 AGND 20 4D
108
107
106
105
104
103
102
101
100
GenxReady XpsReady FilPwmL
98
91
90
99
97
96
95
94
93
92
89
88
87
86
85
84
83
81
80
79
78
77
76
75
74
72
70
69
68
67
66
65
64
62
61
60
59
58
71
82
73
63
6 RT OUTPUT A 11 AGND IC18A AGND R67 21 5Z
+24V R148 D13 6 HvPwmH 22 5A
7 DISCH. +5V DGND
SHUTDOWN 10 2 23 5B
EIA
EIB
EIZ
EIS
ETU0
ETU1
AUX0
AUX1
GND
GND
VDD
P0
P1
P2
P3
P4
P5
P6
P7
P8
P9
P10
P11
D0
D1
D2
D3
D4
D5
D6
D7
D8
D9
D10
D11
D12
D13
D14
D15
D16
D17
D18
D19
D20
D21
D22
D23
VDD
GND
GND
GND
XrayLampC Q2 SafFilCur 10K HvPwmL
C152 5 24 5C
1nF 180 8 9 1N3595 D17 C45 56 C151 25 5D
R107 SOFT-START COMP. C99 AH R68 FilPwmH
IRF9Z24 CMP401 C104 AL 55 0 26 6Z
1N3595 10nF 109 CONVST
DGND 10K C108 1uF BH 54 Not mounted 27 6A
14
13
R112 110 GND R64 0 FilPwmL
R97 C101
C83
SG3525AN 10nF +5VM
10nF BL 53 10nF 28 6B
R110 20K 1nF AGND AGND 111 VDD IC5 Not mounted DGND 29 6C
10K Q3 XrayLampOn 100uF 112 GND 52 30 6D
GND
MPSA05 10K AGND AGND 113 AVDD ADMC401 VDD 51 C72 R136 31 7Z
AGND AGND AGND R98 /MaxFilCur C114 114 AVSS 32 7A
10nF AGND +5V DGND 100
10K D11 115 VIN7 CH 50 33 7B
1N3595 116 VREF 10nF
AGND
FilCurSD C7 Hv CL 49 34 7C
-5V AGND C15 117 48 35 7D
XrayLamp1 R114
R109 10K Lamp1 AGND
VIN6
118 REFCOM
PWMSYNC
47 R6110K 36 8Z
XrayLamp2 Lamp2 /EnFilCu PWMTRIP +5V R49
C9 AGND 10uF 100nF 119 VIN5 PWMPOL 28 +5V 37 8A
AGND 120 CAPT PWMSR 22 +5V 10K 38 8B
10K D19 D18 R1322 121 VIN4 R4110K 39 8C
100nF C28 C16 122 BSHAN 46 40 8D
GND
PTC2 PTC1 2xBZX55C4V7 123 ASHAN VDD 45 C71 41 9Z
C965 C965 100nF 10uF R9 22 124 VIN0 +5V DGND 42 9A
C10 125 CAPB
AGND SCLK0 44 10nF DspDisp
43 9B
100nF C26 126 VIN1 TFS0 43 44 9C
PGND DGND +5V AGND 127 CML RFS0 42 R56 45 9D
IC23 100nF 128 VIN2 DT0 41 DGND 46 10Z
EN1 1 129 GAIN DR0 40 10K 47 10A
130 VIN3 SCLK1 39 48 10B
+24Vin +12V C21 C27 TfsSync
PrWaValve GND 13 PGND 131 SENSE TFS1/IRQ1 38 Rfs1
49 10C
3 OUT1 132 AVSS RFS1/IRQ0/SROM 37 50 10D
12 MF5 R37 C70 IC6A 100nF 100nF 133 36 DspDt1 51 11Z
4
GND XpsIntOut +24VIntOn AVDD DT1/F0 DspDr1B PROMorEEPROM
C31 DR1B/FI 35 52 11A
PWDACK
CLKOUT
10nF DspSclk1
GND 4 3 DR1A/F1 34 1 DATA CLK 2 53 11B
BMODE
AGND + AGND
RESET
10K
MMAP
CLKIN
WaterPump 6 OUT2 MF-R110 Vref+2V 1 C148 C149 C150 C147 6 CEO RST/OE 3 54 11C
XTAL
PWD
DMS
GND
GND
GND
GND
GND
GND
GND
BMS
PMS
BGH
POR
VDD
VDD
VDD
C54
A13
A12
A11
A10
RN2A 100nF
WR
5 2 R52 4 /DspPCe 55 11D
RD
BG
NC
NC
Environment 100nF 100nF 100nF 100nF
BR
A9
A8
A7
A6
A5
A4
A3
A2
A1
A0
-
GND 2K7 100nF CE
Coherency OP495 C20 10K VPP 7 56 12Z
PrCoolCtrl
IN1 2 R42 C36 5 GND VCC 8 57 12A
11
134
135
136
138
137
140
139
15
57
141
142
143
144
1
2
3
4
6
7
8
10
11
12
5
9
13
14
16
17
18
19
20
21
23
24
25
26
27
29
30
31
32
33
WaterCooler 11 OUT3 Pressostat/24V DGND DGND Int1 C43 R21 +5V 58 12B
WaPump 10nF 100uF
IN2 7 10K AGND
10nF
AGND AGND AGND AGND IC13 1K 59 12C
R24 -12V C44 C56 C64 DGND 60 12D
10 WaCooler RN2B C58 R121 R122 R123 R120 AGND +5VM DGND AT17LV256-10PU 61 13Z
R46 12.96MHz C52
8
7
6
5
4
3
2
1
DGND 2K7 100nF AGND 86 18Z
PrWaTIn R62 AGND R1544K7 87 18A
PrWaTOut IntlkRtn DGND DGND IntRet +12V C29 10pF R28 R35 R40 88 18B
WaterTIn 2K7 18K 47K 89 18C
WaterTOut 10K RN2G 4K7 +5VM 90 18D
C73 D14 R96
2K7 91 19Z
Tms
R155
Tclk
Tdo
C136 C124 C130 C142 100nF 1N3595 IC18D AD status converter C159 R39 C160 R45 C161 92 19A
Tdi
100K R31
+5V 3X C105 R92 CuArcTh 11 DGND Low voltage 8x8bit 10K 100nF 10K 100nF 10K 100nF 93 19B
1uF 1uF 1uF 1uF 1N4148 DGND DGND EmissionCurrent- 9 - R103 15 /CuArc C38 94 19C
8 12 10nF IC8 95 19D
1uF 10K 10 + 10K 20 VCC CH0 1 AGND AGND AGND AGND AGND AGND 96 20Z
D12 IC19C TP8 19 V+
AGND
R90
D10 D8 D7
LF347N
D16 CMP401 DGND
FilamentCurrentRef /AdcCs2
AGND CH1 2 97 20A
+12V 10K 1N3595 AGND BZX55C5V1 18 CS CH2 3 98 20B
Vref+5V SdataOut 17 DI
0.266V/A
Sclk CH3 4 99 20C
16 CLK CH4 5 -12V 100 20D
WaLevel AGND FilCurRef AdcSars 15 SARS 6 101
2
11
10K R101 D15 R100 R12 5 6 1 2 TDO 35 Tdi AGND AGND C82 C77 ADC0838 100nF 107 22A
1
4
VREF
VDD
1K 1N4148 C118 18K 10K 74HC14 /Wr A0 AD0 AD7 13 FilCurRefMon
AGND R16 C40 43 I/O TMS 32 12 A0 O0 13 2 AGND 15K 110 22D
100uF R93 100K C13 A1 11 A1 14 AD1 AD6 14 DB7 VOUT8
3 AGND AGND 111 23Z
C115 AD15 2 A7 A2 O1 AD2 AD5 15 DB6 VOUT7
AGND R95 15K 1uF 10pF 67 XTAL1 P1.0/EPA0/T2CLK 57 I/O I/O 14 A6 10 A2 O2 15 AD3 DB5 VOUT6 4 C163 112 23A
100uF X1 /HvArc AD14 A3 AD4 R55
AGND P1.1/EPA1 56 /CuArc AD13
3 I/O I/O 15 /DataCs A4
9 A3 O3 17 AD4 AD3
16 DB4 VOUT5 5 HvArcTh
100nF 113 23B
DGND DGND C41 16MHz P1.2/EPA2/T2DIR 55 4 I/O I/O 39 /CodeCs 8 A4 O4 18 AD5 17 DB3 FF40H VOUT4 6 10K 114 23C
10K AGND AGND 10pF 66 54 /MaxFilCur AD12 5 38 A5 7 19 AD2 18 - 7 CuArcTh +5VM Vref+5V 115 23D
C100 R94 LF347N R91 XTAL2 P1.3/EPA3 FilCurSD AD11 6 I/O I/O /WrDac0 A6 A5 O5 AD6 AD1 19 DB2 VOUT3 MaxFilCur
+ 5 P1.4/EPA4 53 I/O I/O 37 6 A6 O6 20 DB1 FF48H VOUT2 8 AGND 116 24Z
WaResist 7 WdToggle AD10 7 /PPI2Cs A7 AD7 AD0 20 SafFilCur
10K R84 P1.5/EPA5 52 /AdcDor0 AD9 I/O I/O 36 4MHz AD8
5 A7 O7 21 DB0 VOUT1 9 Miscellaneous 8x8bit 117 24A
5M6 IC19B - 6 28 NMI P1.6/EPA6 51 AdcEoc0 AD8
8 I/O I/O 17 /Ppi1Cs AD9
27 A8 AD converter 118 24B
1uF AGND P1.7/EPA7 50 9 I/O I/O 31 /Rd 26 A9 Miscellaneous C37 119 24C
GND
-WR
VSS
C112 /Reset DGND 0 27 30 AD10 23 8x12bit 10nF IC7 120 24D
A2
A1
A0
R102 RESET AD0 /Ale I/O Ready AD11 A10 MaxFilCur
AGND P3.0/AD0 26 16 I/O I/O 29 25 A11 DA converter 20 VCC CH0 1 121 25Z
AD1 ClkOut AD12 SafFilCur
8x100 P3.1/AD1 25 CLK1/I1 33 4 A12 AD7228 AGND 19 V+ CH1 2 Lamp1 122 25A
21
22
23
24
12
10
R88 +12V 1M 220nF R140 40 P0.0/ACH0 AD2 AD13 /WrDac0 /AdcCs1
PrimaryWaterTemperatureInput P3.2/AD2 24 AD3
IC4
AD14
28 A13
A2 SdataOut
18 CS CH2 3 Lamp2 123 25B
1M C116 PrimaryWaterTemperatureOutput R141 41 P0.1/ACH1 P3.3/AD3 23 M4A5-64/32 XPSC01L 29 A14 -5V 17 DI CH3 4 124 25C
DeionizedWaterTemperatureInput R142 42 22 AD4 IC17 AD15 3 A1 Sclk 16 5 125 25D
AGND P0.2/ACH2 P3.4/AD4 AD5 AD0 A0 A15 A0 C106 AdcSars CLK CH4 FilCurMon XsnR1+
DeionizedWaterTemperatureOutput R143 43 P0.3/ACH3 21 3 2 DGND 2 AGND 15 6 126 26Z
4
J450 100nF 100nF +5V +5V 48 ANGND P6.0/EPA8 58 /XsnTxDEn 3 6 10nF 4 23 6 C42 C19 C30 148 R134
29 EA /AdcCs0 SdataIn 4 DI NC AD13 A12 /AdcCs0 DO CH5 XpscTemp 30B
SUD-D 50PM P6.1/EPA9 59 /AdcCs1 DO GND 5
AD14
28 A13
/AdcConv0
22 CS CH6 7 149 30C R135
P6.2/T1CLK 60 3 A14 21 CONV CH7 8 1uF 1uF 1uF 150 30D R149
34 1 L7 DGND +5V R26 R78 AGND 61 /AdcCs2 EEPROM AD15 31 AdcEoc0 20 9 EmissionCurrent+ 151
DGND 18 1 2 +5Vin 1M 10K P6.3/T1DIR
62 DGND 2
A15
19
EOC COM
10 HighVoltage 13x 100 Ohm 152
31Z
P6.4/SC0 93C66 DGND A16 DGND PD MUXOUT1 31A
35 2 SdataIn EmissionCurrent-
L8
-12Vin R71 P6.5/SD0 63 Sclk
or
/DataCs 22 Vref+5V AGND
18 AGND A/DIN1 11 AGND
Vref+5V
153 31B
19 1 2 P6.6/SC1 64 +5V CE1 E000H 17 VREF+ MUXOUT2 12 154 31C
36 3 C35 R77 R63 SdataOut IC12A Vref+2V
AGND L9
+12Vin /WaterCoolOn R76 +24V 10K
/WaCoOn P6.7/SD1 65 EepromCs 1 +5V /Wr 30 CE2 - R58
10
16 VREF- A/DIN2 13 InputCurrent
155 31D
20 1 2 0 0 R27 CS VCC 8 29 WE FF3FH 15 VA+ DGND 14 156 32Z
37 4 Sclk 2 SK /Rd 24 InputVoltage
L10
+24Vin
C173
4K7 1uF IC9 80C196KR 180 SdataOut NC 7 OE Mon400V
157 32A
21 1 2 IC22 3 DI NC 6
C67 C75 ADC12138 158 32B
38 5 1uF DGND DGND ClkOut SdataIn 4 DO GND 5 RAM128K 10uF 10nF DGND AGND 159 32C
PGND PGND PC817 /AdcConv0 C78
22 160 32D
WaResist 39 6 XpsIntOut PGND AdcSars EEPROM DGND 10nF IC26
23 Pressostat/24V /XsnTxDEn 93C66/SO AGND AGND DspWdToggle FilPwmH 2 A1 Y1 18
WaLevel 40 7 Intlk2 FilPwmL 4 A2
Intlk3 DGND
+5V DGND +5V +5VM AGND
WdToggle HvPwmH Y2 16
24 6 A3 Y3 14 GND DIN41612
WaterTOut 41 8 Intlk4 HvPwmL 8 A4 Y4 12
3
1
2
3
4
WaterPump AD7 PA6 /WaCoOn AD7 D6 PA6 22K C167 B
30 DGND 13 27 D7 37 27 37 MM74HC423A
2
4
6
8
XrayLamp1 47 XrayLampC R118 PA7 D7 PA7 10nF
14 DGND LD2 T=100ms 6 C
31 XrayLamp2 8922AM 1K R4 /Rd 5 RD 18 ExRelayPresent /Rd 5 PowerOn
XsnT1- XsnT1+
IC2F
/Wr PB0 /Wr RD PB0 18 EnPower S1 W2
7 R/C
48 15 330 36 WR PB1 19 36 WR PB1 19
6
8
7
6
5
1
3
5
7
33 9 DGND XpscNotOk/Red 35 RESET 22 /PrWaFlow 35 22 10 SdataOut 4 16 DspDr1B MM74HC423A
DGND /Ppi1Cs PB4 /DeWaFlow /Ppi2Cs RESET PB4 /DspPCe RfsSync A2 Y2 Rfs1
50 17 8 74HC14 6 CS PB5 23 6 CS PB5 23 9 6 A3 Y3 14 T=100us
10 24 Reset 24 EnDisRd2 8 DspDt1 8 A4 12 Sdatain Accessories
/XsnTxDEn PB6 PB6 EnXpscDis Y4 ExSclk
PB7 25 PB7 25 7 +5V 11 A5 Y5 9 ExSdataOut
C153 5 Address 6 13 A6 7
52
DGND
C79 C49 C89 C4 C3 C60 C47 C25 C61 C48 C146 C164 C165
-12Vin
AGND AGND
0
PGND
-12Vin
MF2
1
L3
2
C18
220uF
C12
100nF
-12V
ThermoFisher S112306
No Mod.
13
10
35
9
SPM1 7D
DGND DGND 2.2uF 1nF/500V 1nF/500V 36 8Z
8 Line2 37
C37 38
8A
100nF 0VRTM2 8B
2 7 39 8C
R11 40 8D
R63 3 6 R55 10K 1 41 9Z
1K 10 42 9A
5 0VRTM2 2 43 9B
+5V 44 9C
R57 10K 5 45 9D
HCPL3101 46 10Z
R23 0VRTM2 6 +400VRTM 47 10A
1K 48 10B
+15VRTM PowerTemp 18 16 ToHvTransfo 49
IC5 10C
C3 C4 C5 R3 50 10D
R25 C31 15 1000uF 1000uF 1000uF 10K/5W 51 11Z
HvPwmL 8 100nF F2 250V 250V 250V +400VRTM 52
C38 NTC
14 ToFilamentTransfo +200VRTM 53
11A
120 R64 2 7 100nF 19 54
11B
4
AGND 10A 11C
1K R12 AGND 55 11D
3 6 -10VRTM R58 10K 7 C17 C18 C19 R4 56
C68 12Z
10 1000uF 1000uF 1000uF 10K/5W 10nF/1kV 57 12A
5 0VRT 8 T2 T3 250V 250V 250V 58
HvPwmH R32 Q5 0VRTM 12B
59 12C
MPSA05 60 12D
10K R31 HCPL3101 R56 10K 3 C70 61
2
13Z
4K7 AGND AGND 62 13A
+15VRTM 0VRT 4 1nF/500V 63
IC3 D10 D9 D14 D13 13B
-
64 13C
DGND DGND 65 13D
8 MWI75-06A7T 66
17
11
12
C36 14Z
67 14A
R62 2 7 100nF D12 D11 68
D16 D15 14B
1K R10 69
+
0VRT 14C
3 6 -10VRTM 4xBYV26 4xBYV26 70 14D
10 71 15Z
5 +200VRTM 72
R60 10K 15A
73 15B
74 15C
R27 HCPL3101 75 15D
FilPwmL C67 76
FAN 24V 16Z
680 1nF/500V 77 16A
U2 Not mounted 78 16B
on Low Power version 79
FilPwmH R68 Q6 80
16C
16D
MPSA05 +15VRTM 81
+
-
10K IC7 17Z
R67 82 17A
4K7 P2 +5VM 83 17B
8 0VRT 84
C40 B WAGO 2P 17C
85 17D
DGND DGND R66 2 7 100nF 86
1
2
IC8 18Z
1K R14 87 18A
3 6 -10VRTM PC817 88 18B
10 +24V PGND 89 18C
5 90 18D
91 19Z
R20 R16 0VRTM 92
D19 19A
HCPL3101 33K 1K 93 19B
94 19C
R42 BAT85 +5VM +24V 95 19D
AGND 96 20Z
D21 C7 97 20A
100K +5V 98
BZX55C7V5 10nF/1kV 20B
R17 R19 99 20C
2K7 47K 100 20D
C62 -10VRTM 101 21Z
C65 470nF 10nF 102 21A
R40 +5VM 103
8
21B
10K C63 C64 +15VRTM GND 104
AGND IC10A + Vref+5V 21C
Q1 R44 3 8x8bit ADC 10nF R15 10nF R18 105 21D
MPSA05 1 C26 Status AD converter 10K 10K 106 22Z
2 HvIgbtCurrent /AdcCs2 R51 100 107
2K2 - 10nF IC1 D22 22A
OP284 108 22B
R38 SdataOut R50 100 20 1 AGND AGND AGND AGND C72 109
C58 C56 BZX55C11V
4
10 TP1 +24V
BYT11
MF1 L1 +24V DIN41612
+15VRTM +15VRTM2 +24Vin
D4 PTC1 R21 D5
MF-R250 C2 C9 TP2
C46 D8
1N4007 5.6/25°C 10 BYT11 220uF 100nF PGND
2200uF/25V 1N5355B R2
C45
Line1 1 T1 12 C23 C27 D3 D6
2200uF/25V 0
2200uF/25V 2200uF/25V 1N5352B 1N5355B PGND
GND
11
8 0VRT 0VRTM2 0VRTM3
F1
Line2 6 7
230V/2x15V/10VA +5V
0.5A C16 C28 D2 TP3
2200uF/25V 2200uF/25V 1N5347B +5V +5VM
D1 PTC2 L2 +5V L3
+5Vin MF2 R29
10K
1N4007 5.6/25°C MF-R110
C10 XpspTemp Mod: 1 Version pré-série 23.08.01 PPa 4 Version série 11.04.03 PPa 7 Addendum OM3834 30.11.04 PPa
C1 C8 TP5 TP6
-10VRTM TP4 1uF 2 Version pré-série (2) 07.11.01 PPa 5 OM3753 & OM3758 15.01.04 PPa 8 OM3884 21.02.05 PPa
220uF 100nF AGND AGND C32
DGND RV1
R1 4K7/25 100nF 3 Version "phase in" 03.09.02 PPa 6 OM3834 28.10.04 PPa 9 OM4033 30.07.07 PPa
0 02.12.00 MBo Remplacé par:
GND DGND AGND AGND AGND
XPSP X-ray Power Supply Power Board Echelle Dessiné
Contrôlé
Remplace:
ThermoFisher
Groupe No No Mod.
S C I E N T I F I C
S112067
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 9
+5VA
R78 20k
R77
10k +5V
C69 J905
R88 D2 C52 R83 +5V 100n
+VGateL_3 100n IC16 +24V_Vent
39k
LMV7235 1
5
1Z
270 BYG20J DGND 2 1A
AGND 3 + GND 3 1B
R84 1 +5V P2 4 1C
IC10 4 - 2 4 + +24Vin 5 1D
1 IN OUT 4 1 3 6 2Z
IKW40T120_L2 IKW40T120_L2 IKW40T120_L2 2 5 ToHvTransfo 10k C53 R42 - 7
PGND
2
IN OUT 2A
Z1 Z2 Z3 3 IN OUT 6 1n 1k B WAGO 2P 8 2B
+5VM C22 D1 +12V 9
GH1 GH2 GH3 2C
1 1 1 R44 0 FAN 24V 10 2D
7 9 Mes_I_Transfo Limit_I_Transfo 1 2 /PwmTrip 11
C48
VCC OUT
PGND AGND 12
3Z
1n
3
EmH1 EmH2 EmH3 3A
100n 8 GND REF 10 AGND DGND -12V 13 3B
+400VRTM C33 BAT54T1 14
+5Vin 3C
LAS50-TP 10n 15 3D
AGND 16 4Z
R8 C13 C12 C11 C21 +24V_Vent DGND 17
/PwmTrip 4A
10k/5W 1000uF 1000uF 1000uF DGND 18 4B
4.7uF 19 4C
+200VRTM R63 R54 24k 20 4D
P1 21 5Z
IKW40T120_L2 IKW40T120_L2 IKW40T120_L2 2 4 + 22
2
C20 5A
R25 C23 C26 C25 Z4 Z5 Z6 10k +5V 1 3 23 5B
10k/5W 1000uF 1000uF 1000uF - 24 5C
4.7uF GL1 1 GL2 1 GL3 1 +5V C66 B WAGO 2P 25
0VRTM +5VA 5D
IC14 100n 26 6Z
LMV7235 FAN 24V 27
5
3
3
EmL1 EmL2 EmL3 +5VA_PTB 6A
IC19 D3 28 6B
R53 3 + DGND PGND TR1 29 6C
10k 1 2 VOUT VIN 3 1 2 30 6D
4 - 31 7Z
Limit courant pos Not mounted 4 VOUT ADJ 1 BAT54T1 32 7A
C44 R62 C45 on Low Power version C73 33
2
7B
100n 5.1k 1n Limit courant neg comp+ = 3V C71 34 7C
comp- = 2.57V 1u LM317-SOT223 47u 35 7D
DGND Comp+ = 2.36V +5VA 36 8Z
AGND AGND Comp- = 0.71V 37 8A
R80 33k AGND 38 8B
R79 240 R90 R89 VB0.35/1/6 39 8C
10k +5V 715 40 8D
DGND 41 9Z
C70 42 9A
C54 R85 +5V 100n AGND 43 9B
100n 30k IC17 44 9C
LMV7235 45
5
9D
DGND 46 10Z
AGND 3 + 47 10A
R86 1 +5V 48 10B
4 - 49 10C
50 10D
10k C55 R69 51
2
+400VRTM 11Z
+5VM 1n 1k 52 11A
53 11B
DGND R75 0 54 11C
C49 IC11 AGND /Limit_I_Fil 55
Mes_I_Fil Not mounted 11D
100n 7 VCC_5V Uout 9 C32 56 12Z
8 GND Ref 10 10nF/1kV 57 12A
C43 58 12B
AGND Isol. galvanique 10n 59
LEM6 12C
1 I_mes1 I_mes6 6 R65 R56 47k 60 12D
LEM6 2 5 LEM5 61
LEM5 I_mes2 I_mes5 13Z
3 I_mes3 I_mes4 4 DGND 62 13A
10k +5V 63 13B
64 13C
LEM_LTSR ToFilamentTransfo +5V C67 65
+5VA 13D
IC15 100n 66 14Z
LMV7235 67
5
14A
68 14B
R55 3 + DGND 69 14C
10k 1 70 14D
4 - 71 15Z
+200VRTM 72 15A
C46 R64 C47 73
2
15B
100n 5.1k 1n Limit courant neg 74 15C
75
Isol Galv DGND Comp+ = 2.35V C19 76
15D
+VGate_1 16Z
AGND AGND 1nF/500V 77
Isol Galv +VGate_3 Not mounted Comp- = 1.87V 78
16A
+24V 16B
79 16C
IC4 R12 +24V 80
C14 16D
0 IC1 81 17Z
3 12 10u C3 82
+VIN +VOUT 17A
R14 1 3 +VIN +VOUT 12 10u 83 17B
C58 1 18 C15 EmH1 84
-VIN NC C6 17C
1u 10u C57 1 -VIN NC 18 85 17D
Com 7 1u 10u 86 18Z
7 EmH3 87
EmH2 Com 18A
-VOUT 9 88 18B
PGND C16 9 89
-VOUT C7 +5VA 18C
10u R13 1 PGND 90 18D
TES 1-2423 10u 91 19Z
-VGate_1 TES 1-2423 0VRTM 92
-VGate_3 19A
93 19B
R37 94 19C
10k R30 100k 95 19D
+5V 96 20Z
C39 97 20A
+VGate_1 R38 C35 98
+24V 30k 1u R31 10nF/1kV 99
20B
Isol Galv +5V 1k 100
20C
Q1 Isol Galv +24V 101
20D
5
+5VA 21Z
MJD44H11 IC2 102 21A
R40 AGND 3 + 103
IC5 Lim_24V 21B
R23 270 3 VIN VOUT 2 47k 1 GND 104 21C
HvPwmH 1 8 R15 R16 4.7 R9 270 IC3 4 - 105
IN VCC GH1 FilPwmH +VGate_3 C17 21D
R21 0 1 IN VCC 8 C8 1 ADJ VOUT 4 106 22Z
7 R7 1u C10 R39 C36 IC7 C31 107
2
Out GH3 22A
R19 1k Out 7 10k 1u LMV7235 10n 108 22B
HvPwmL 2 10 R6 1k LM317-SOT223 1u DGND 109
Cat R20 GH2 FilPwmL 10 22C
3 Cat Ref 6 2 Cat 110 22D
10k C18 R22 3 6 R10 DGND +5V DGND ToHvTransfo C30 111
Cat Ref AGND 23Z
M4 4 NC Ref 5 10n 0 R18 4.7 10k DGND 112 23A
FDV301N M1 4 NC Ref 5 R4 R3 240 C60 113 23B
DGND ACPL_H312 FDV301N 715 100n 114 23C
MJD45H11 DGND ACPL_H312 3x 2u2/250V 115
Q2 -VGate_3 23D
R17 1k 116 24Z
Lim_24V R5 1k AGND DGND C38 117
-VGate_1 Lim_24V 24A
118 24B
119 24C
M5 120 24D
FDV301N M2 +24V 121 25Z
DGND FDV301N
+24V_Vent TP1 +24V 122 25A
DGND 123 25B
MF1 124 25C
R24 +24Vin 125
Limit_I_Transfo 25D
R11 126 26Z
/Limit_I_Fil MF-R250 C1 C4 TP3 127
ToFilamentTransfo 26A
1k 220u 128 26B
M6 1k 100n PGND 129
R2 26C
FDV301N M3 130 26D
DGND FDV301N 131 27Z
DGND 0 PGND 132 27A
GND 133 27B
134 27C
135 27D
FilPwmH 136
FilPwmL 28Z
137 28A
HvPwmH 138
HvPwmL 28B
139 28C
FilamentCurrent 140 28D
/AdcCs3 141 29Z
TP2 /AdcCs2 142
+5VA 29A
+5V +5VM /AdcCs1 143 29B
L3 +5V L2 SdataOut SdataOut 144
+5Vin MF2 Sclk 29C
Sclk 145 29D
R72 AdcSars AdcSars 146
MF-R110 SdataIn 30Z
1K SdataIn 147 30A
C9 PowerTemp CurrentControl 148
Isol Galv +VGateL_1 Isol Galv C2 C5 1u
TP5
CurrentLimit 149
30B
+VGateL_3 220u 100n AGND C68 PowerOn 150
30C
+24V R73 30D
R1 100n EmissionCurrent+ 151 31Z
IC6 +24V 6.2k HighVoltage 152
C27 31A
IC9 0 EmissionCurrent- 153 31B
3 12 10u C37 GND DGND AGND Vref+5V 154
+VIN +VOUT 31C
R26 1 3 +VIN +VOUT 12 10u AGND Vref+2V 155 31D
C59 1 18 C28 EmL1 InputCurrent 156
-VIN NC C40 32Z
1u 10u C61 1 -VIN NC 18 InputVoltage 157 32A
7 ComGateL_1 1u 10u Hole Hole Mon400V 158
Com EmL3 32B
Com 7 R57 159 32C
9 EmL2 TP7 TP8 TP6 TP4 +5VA 160
-VOUT C29 32D
PGND -VOUT 9 DGND DGND DGND DGND 0
10u R27 1 PGND C42
TES 1-2423 10u
-VGateL_1 TES 1-2423 R82 DGND GND
-VGateL_3 10K DIN41612
DGND DGND DGND DGND
XpspTemp
RV1 C50
+VGateL_1 100n
4K7/25
Isol Galv
Q3 Isol Galv AGND AGND
MJD44H11
R43 270 IC8 +5VA +24V
HvPwmL 1 8 R28 R33 4.7 R70 270 IC13
IN VCC GL1 FilPwmL +VGateL_3
R36 0 1 IN VCC 8
Out 7 R61
R34 1k 7 GL3 R46 R49
HvPwmH 10 Out
2 Cat R60 1k 2K7 47K
3 6 R35 GL2 FilPwmH 2 10
Cat Ref R41 Cat R71
10k C34 3 Cat Ref 6
M7 4 5 10n 0 R29 4.7 10k +5VA
NC Ref
FDV301N M10 4 NC Ref 5 C62 C63
DGND ACPL_H312 FDV301N 8x8bit ADC 10n R47 10n R50
MJD45H11 DGND ACPL_H312 C41 Status AD converter 10K 10K +5VA_PTB
R32 1k Q4 -VGateL_3 /AdcCs2 R48 100
Lim_24V 10n IC12 Mod: 10 New Design 25.05.09 PPa 13 Add. OM4180 + OM4193 03.03.11 PPa
R52 1k
-VGateL_1 Lim_24V SdataOut R51 100 20 1 AGND AGND AGND AGND
VCC CH0 11 OM4130 17.07.09 PPa
AGND 19 V+ CH1 2
M8 Sclk R59 100 18 3
CS CH2 12 OM4180 30.08.10 PPa
FDV301N M11 17 DI CH3 4
DGND FDV301N AdcSars R68 100 16 5
CLK CH4 Remplacé par:
DGND
SdataIn R74 100
15
14
SARS
DO
CH5
CH6
6
7 PowerTemp
XpspTemp
+5VA_PTB Accesories
XPSP X-ray Power Supply Power Board Echelle Dessiné 15.05.09 TMa
R45 R81 +5V 13 SE CH7 8
Limit_I_Transfo R76 Vref+5V 12 9 Hole Hole Contrôlé
/Limit_I_Fil VREF COM Remplace:
10 11 AGND DGND 10
1k R87
M9 1k ADC0838 Liste S703126 - S703779 Schémas S112067 Vu
FDV301N M12 C51 Fid Fid
DGND FDV301N C56 AGND 0 Ensemble ---------- Circuit S920078 Groupe Technique:
DGND 10u 1n DGND AGND Fid Fid Ingénieur T. Maulaz
Appareil XRF Implantation S112073 9850
AGND AGND Fid Fid Groupe No
Fid Fid ThermoFisher S112067
No Mod.
S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 13
+24V
P1
1 5 +12V
2 6 + +200VRTM
3 7
4 8 - C64
AGND +5V
10nF
4
WAGO 4P ToHvTransfo
J906 FAN 24V IC5A 3
IC5C R65 +
PGND R79 10 1
D20 IC5D - 13 8
+
R67 2 R18
R76 10K -
1 + /PwmTrip 14 9 OP484 10K
1Z
2 + 12 10K OP484
-
R66 C66 T1A T1B
11
1A 1K XpsmHvTemp
1B 3 GND - BAT85 OP484 R78 10K 10K 47pF
4 MF6 D21 Vref+2V C67
1C BAT85 R68
1D 5 +24Vin FAN 24V 10nF C35 RV2
6 1K AGND -12V AGND 100nF 4K7/25
4
2Z C65 10K
2A 7 MF-R110 AGND
8 PGND DGND 10nF C77 T1
2B +12Vin R64
2C 9 R80 +12V AGND AGND
2D 10 AGND R69 HvTransfo
11 220pF 1K 30K
3Z AGND -12Vin 1K
3A 12 C72 Not on the Board
13 IC5B C63
2
3B
3C 14 6 - 10nF
15 +5Vin 7 3n3
3D
16 5 + AGND
4
4Z DGND IC4B
4A 17 OP484 5 R41 R63
18 7
+ AGND IC4A 3
4B R56 +
4C 19 6 1
20 AGND OP484
-
2 1K 20K
4D 10k -
5Z 21 Positive HV OP484 h18 R28
22 D11 R52 R59
11
5A Arc C44
5B 23
24 C40 1nF 200M
5C
25 1N4148 100 10K 10nF 20x 4n7/3kV
5D C59
6Z 26 -12V
6A 27 Line1 R21 C34 1nF AGND GAP6
28 Line1 C33 100K D12 R48 (Not mounted) R35 M51-C90
6B
29 10uF BZX85C5V6 AGND C1 C2 C3 C4 C5
6C 10nF
6D 30 5x 20nF/12kV
31 100 1K
7Z Negative HV
7A 32 AGND GND
7B 33 AGND AGND AGND
7C 34 D1 D2 D3 D4 D5 D6 D7 D8 D9 D10 D11 D12
35 +5V IC4C 10 MF5
7D + AGND R42 R13 h1
36 8 R57 IC4D + 12
8Z
37 Line2 9 14 24x Z100UFG
8A
38 Line2 D22 OP484
-
13 1K MF-R050 15/3W GAP1
8B
39 BAT85 Positive HV 10k -
A81-C230 D1' D2' D3' D4' D5' D6' D7' D8' D9' D10' D11' D12'
8C OP484
40 R26 C45 Wired resistors JP1
8D EmissionCurrent+ R53 R60
9Z 41 1uF GAP3
42 M51-C90 GND C6 C7 C8 C9 C10 C11 6x 20nF/12kV R1 R2 R3 R4 R5 R6 R7 R8
9A
43 100 D13 100 10k 1
9B C60 2
9C 44 BAT85 AGND 3
9D 45 1nF 8x 100/9W
10Z 46 R49 (Not mounted) R36 GND D13' D14' D15' D16' D17' D18' D19' D20' D21' D22' D23' D24'
10A 47 D10 Fed.Standard
48 AGND 1N5347 C32
10B
49 100 1K 24x Z100UFG
10C Negative HV 100uF
10D 50 R40 R34 R30 R25
11Z 51 D13 D14 D15 D16 D17 D18 D19 D20 D21 D22 D23 D24
11A 52 HP:4x100
11B 53 +400VRTM
54 LP:1x100
11C
11D 55 C12 C13 C14 C15 C16
12Z 56 5x 20nF/12kV
12A 57 AGND
12B 58 HT Output board S940188
12C 59 +12V
12D 60 R19 R14 h4
61 IC3B 5 6x Multiplier board S930191
13Z +
13A 62 +5V C62 7 1K 15/3W
13B 63 - 6 GAP2 Ra1
64 10nF OP484 A81-C230
13C R58
13D 65 D14
66 AGND 1K R55
4
14Z BAT85
14A 67 GAP4 GND GND
68 IC3A 3 C36 D8 R16 C30
14B HvDividerDifference R38 R32 + 1K M51-C90
14C 69 1 10nF 1N5347 53K3 10uF 5x 200M/15kV 5x 470pF/15kV Divisor board S940189
70 2 R61 Wired resistors
14D
71 1K 1K OP484
-
10K
15Z +200VRTM Vref+2V R17 R16 R15 R14 R13
72 D16 R12 R11 R10 R9
11
15A
15B 73 C43 BAT85 C39
15C 74 10nF C74 AGND AGND AGND AGND GND
75 10nF 10nF 4x 1K/9W
15D
16Z 76 -12V C21 C20 C19 C18 C17
16A 77 AGND
16B 78 AGND R28 AGND R23
79 AGND R27 C27
16C
16D 80 10K 1K 33M/9kV 470pF/15kV
17Z 81 +5V
17A 82 Interface board
17B 83 D19
17C 84 S930192 R26 R25 R24 R23 R22
85 BAT85 R21 R20 R19 R18
17D IC3C R20 R15 h5
18Z 86 R62 R51 + 10
87 HighVoltage 8
18A
88 9 1K 15/3W GAP3 Ra2 4x 1K/9W
18B 100 100 -
18C 89 OP484 A81-C230 C26 C25 C24 C23 C22
18D 90 D18 Wired resistors
19Z 91 BAT85 R47 5x 200M/15kV 5x 470pF/15kV Divisor board S940189
19A 92 GAP5 GND GND Multiplier Box S703757 (not on the board)
93 0VRTM C37 D9 R17 C31
19B
94 1K 10nF 1N5347 53K3 M51-C90
19C 10uF
19D 95 AGND
20Z 96
20A 97
20B 98
20C 99 AGND AGND AGND AGND GND
20D 100
21Z 101 GND
21A 102
21B 103
21C 104
21D 105
22Z 106
22A 107
22B 108
22C 109
22D 110 GND GND GND
23Z 111 h57 H55
112 ToHvTransfo ToFilamentTransfo 1 T2
23A
23B 113 C13 D6
114 3 D2 D1 GAP2
23C 1nF/500V 1.5KE68A M51-C90
23D 115
24Z 116
24A 117 4 I1 J1
24B 118 h58 H56 1 2 C1 1
119 +200VRTM 2 D3 D4
24C 47nF/100V SENS1 C18 2
24D 120 FilamentTransfo 3 4 1nF/500V 3
121 4xMUR1520 C16 C22 SCD15PBR
25Z 4
25A 122 1000uF/65V 1000uF/65V 5
123 Not on the Board
6
25B GAP1
25C 124 MF4 C20 D7 WAGO
25D 125 M51-C90
126 MF-R050 1nF/500V 1.5KE68A
26Z
26A 127
128 ToFilamentTransfo
26B
26C 129 0VI U/I GND GND GND GND
26D 130 R11
131 +5V EmissionCurrent- SENSOR
27Z
27A 132 100
27B 133 C21
134 100uF/35V OUT 0V
27C R4 C25
135
1
2
3
4
27D 10nF
28Z 136 FilPwmH 10K R12 R9 R8 R10
28A 137 FilPwmL
138 HvPwmH XpsmFilamentTemp
28B HP:4x100
28C 139 HvPwmL AGND LP:1x100 +12V AGND
140 FilamentCurrent C17 RV1
28D /AdcCs3
29Z 141 +5V +12V -12V 100nF 4K7/25
29A 142 /AdcCs2 AGND
29B 143 /AdcCs1 AGND
144 SdataOut AGND
29C Sclk C11 C10 -12V
29D 145 R31 R46 R43 +5V +5VI
146 AdcSars 2K7 18K 33K 10nF 10nF
30Z SdataIn
30A 147
30B 148 CurrentControl +5V C69 C70 C71 R22 AGND AGND IC6
149 CurrentLimit 8x8bit ADC R27 100nF R37 100nF R50 100nF FilamentCurrent 0.266V/A R7 R3 2 4 C78
30C 10K +VIN +VO
150 PowerOn Status AD converter 10K 10K 330K 330K 1 3 10uF
30D
151 EmissionCurrent+ /AdcCs3 R70 100 100 -VIN -VO
31Z HighVoltage C38 IC2
31A 152 IK0505SA
153 EmissionCurrent- SdataOut R71 100 10nF 20 1 AGND AGND AGND AGND AGND
31B Vref+5V VCC CH0 C41
31C 154 19 V+ CH1 2 R33 +5V +5VI
155 Vref+2V Sclk R72 100 AGND 18 3 Vref+5V 1uF +5V C76 AGND 0VI
31D CS CH2 Arc
32Z 156 InputCurrent 17 DI CH3 4 C75
157 InputVoltage AdcSars R73 100 16 5 HvDividerDifference AGND 10K IC1
32A CLK CH4 FilamentVoltage
32B 158 Mon400V 15 SARS CH5 6 D17 10nF
159 SdataIn R74 100 14 7 XpsmFilamentTemp AGND 8 1 10nF
32C DO CH6 XpsmHvTemp C68 BAT85
32D 160 +5V 13 SE CH7 8 R5
GND Vref+5V R54 12 9 10nF OP484 - 13 R44 1K 7 2
VREF COM FilamentVoltage 0.242V/Vfil R45 R24
10 11 AGND DGND 10 R75 14 R6 1K
+ 12 R39 1K 6 3
C61 C53 1K 1K IC3D
ADC0838 C56 1K
DIN41612 10uF 1nF AGND 0 D15 5 4
DGND AGND 100uF/35V R29 C19 C15
BAT85
10K 100nF 100nF
AGND AGND AGND HCPL7800
Accessories
220uF 100nF
-12Vin MF2 L2
-12V XPSM X-ray Power Supply Multiplier Board
Hole Hole Hole Hole
X-ray Power Supply Multiplier Assembly Contrôlé
Remplace
MF-R110 Vu
Liste S703758 Schémas S112068
Hole Hole Hole Hole
Assembly : S703127 Ensemble ---------- Circuit ---------- Groupe Technique
Appareil XRF Implantation ---------- Ingénieur M.Bonzon 9850
GND
PCB : S920079 Format No
13
14
1A
3 1B
GND GND 4 1C
+24Vout 5 1D
2 D12 6 2Z
5 PGND 7 2A
T3 8 2B
BYV28 +12Vout 9 2C
8 10 2D
11 AGND 11 3Z
-12Vout 12 3A
13
13
14
3B
D23 14 3C
GND GND +5Vout 15 3D
1.5KE120A 16 4Z
DGND 17 4A
18 4B
2 D20 19 4C
D25 5 20 4D
D24 BYV26E T2 21 5Z
BYV28 22
C87 1.5KE120A 8 23
5A
10n +24V 5B
11 24 5C
+12Vout +12V 25 5D
+15Vout 26
13
14
Line1 6Z
27 6A
GND GND MF3 MF6 MF4 C38 28
IC12 IC7 6B
MF-R250 R80 R82 R54 MF-R050 MF-R020 29 6C
2 D29 L5 4 MF8 4 1nF/500V GND 30
FB FB 6D
5 1 IN LM2576 10K L8 L10 MF-R050 1 IN LM2576 9K1 L3 L6 31 7Z
2 1K5 2 32
T4 50uH OUT +15V OUT 7A
R104 HS1 BYV28 33
GND
GND
Line1 R105 C57 C78 470uH 50uH C74 470uH 50uH 7B
C33
ON
ON
2.2M 8 C53 470uF/50V C37 34 7C
HS-580 11 R78 D13 C83 C76 R50 D10 C48 35
4.7/4W 470uF/50V C59 7D
R95 470nF 100nF 1K BYV28 2200uF/16V 100uF/16V 470nF 100nF 1K BYV28 100uF/16V 36
5
2200uF/16V Line2 8Z
37
13
14
8A
- + 2 L R48 PGND C40 38 8B
7 GND GND PGND 0 R86 AGND AGND AGND AGND R85 AGND AGND AGND AGND 39
D 5K6/5W 8C
3 X PGND GND 47K 47K 1nF/500V GND 40 8D
R106 W08G 5 F 41 9Z
Line2 4 42
D2 1
S Z1 AGND AGND AGND AGND 43
9A
4.7/4W C R94 R91 9B
44 9C
5K6/5W 45 9D
IC10 150 BZX85C24V 46
R35 10Z
TOP249_TO220_7P 47 10A
R103 R13 24K IC13 IC8 48
IC3D R79 R83 R55 10B
6.8 R56 + 12 FB 4 FB 4 49 10C
D8 14 1 IN LM2576 L9 1 IN LM2576 L4 50 10D
5 1 13 10K 2 1K5 10K 2 9K1 51
OUT OUT 11Z
IC14 270
-
1N5352B OP484 R34 52
GND
GND
C72 C80 470uH C30 C85 470uH 11A
H11D1Z C21
ON
ON
2K7 C84 AGND C77 AGND 53 11B
C4 10nF 10nF R81 D14 2200uF/16V C60 C49 54
100uF/16V R51 D11 11C
C91 4 2 470nF 100nF 1K BYV28 470nF 100nF 2200uF/16V 100uF/16V 55
5
BYV28 11D
47uF/35V C5 R9 AGND L11 1K 56
C39 C50 C90 AGND 57
12Z
L7 12A
1uF AGND R87 R88 58
10K 50uH 12B
1uF 47K MF9 47K 50uH 59 12C
2 x 10uF/450V D19 R107 R4 MF-R050 60
MF5 MF7 12D
10K 10K 61 13Z
1N5346B MF-R050 MF-R020 62 13A
Vref+2V5 -15Vout 63 13B
64 13C
AGND C6 +5Vout +5V 65 13D
+24V +24Vout 66
10nF IC4 -12Vout -12V 67
14Z
RN1 W2 R33 MF2 MF1 14A
+5V FB 4 68 14B
+5V 1 2 1 LM2576 L1 L2 MF-R110 MF-R050 69
IN 3K3 14C
9*10K 3 4 AGND OUT 2 70 14D
71
GND
150uH
1
2
3
4
5
6
7
8
9
10
ON
R96 72 15A
470nF 47K R32 D1 C27 73
1K C36 15B
1K BYV28 74
100uF/16V
5
15C
2200uF/16V 75 15D
4 5 76 16Z
3 6 AGND AGND AGND AGND 77 16A
2 7 RY1 78 16B
AGND D35 R108 79
C46
C58
C63
C67
C44
C54
4
10 +24V 20Z
11 R14 SGND 97 20A
+5Vout 47 R66 /XonAcq KVRef 3 + IC2A 98
12 R6 R97 20B
13
47 R67 OverKV 1 99 20C
47 R69 LowKV 1K 2 - +5V +12V 470 100
14 C11 OP484 3K3/0.1% 20D
47 R71 OverLoad C3 101
15
47 R75 I_FilLim 10nF R10 D15 C15 102
21Z
R1 R98
11
16 BAT85 21A
17
47 R76 OverMA 1K/0.1% 10nF 103 21B
47 R39 /LkILim SGND S1 R38 1uF 104
18 1K C34 Reset 0 21C
19
47 R41 PSFaultOR/PSFault 1M AGND 105 21D
AGND AGND AGND BAT85 106
4
20 AGND 22Z
47 R49 HVInvFlt -12V 1uF R15 107
21
SGND C43 3 + IC3A 108
22A
22 R30 D33 R8 22B
47 R58 FilInvFlt MARef 5 + IC2B DGND 100nF 1 D34 KVMonFb 109
23 R21 10K 22C
+24V 7 C12 2 - 110
1
24 1K C31 6 - DGND OP484 BAT85 100 C19 111
22D
25 GND R90 3K3/0.1% 100pF 23Z
KVRef 10nF OP484 11.0592MHz R3 10nF 112
D16
11
26 C16 C65 IC9 23A
KVCommon R57 R16 IC6A 113
27 SGND MARef 1K XrLamp 1K/0.1% BAT85 10nF DGND 32 43 /InterlockOk SGND 114
23B
28 R26 74HC14 10pF XTAL1 P0.0/AD0 33K 23C
29
MACommon TR1 SGND X1 P0.1/AD1 44 VccCtl R12 SGND 115 23D
SGND 1K 33 45 FilInvFlt 10K 116
MPSA05
2
30 AGND 1K C66 10pF XTAL2 P0.2/AD2 HVInvFlt C35 24Z
31 R73 AGND AGND AGND DGND P0.3/AD3 46 117 24A
AGND AGND KVRegFlt 15 49 LowKV -12V 1uF 118
32 PowerRef RESET P0.4/AD4 Arcs 24B
33 47 R20 P0.5/AD5 50 SGND 119 24C
PowerCom PowerRef 10 + IC2C 1 51 /XonAcq AGND 120
34 SGND I_FilRef
R22 P1.0/ADC0/T2 P0.6/AD6 /LkILim 24D
35 8 2 P1.1/ADC1/T2EX P0.7/AD7 52 R23 121 25Z
I_FilCom 1K 9 - 3 5 + IC3B 122
36 SGND R44 C25 3K3/0.1% P1.2/ADC2 R24 25A
MARegFlt OP484 4 28 C26 7 I_FilMon 123
37 KVMonFb R17 D17 C17 /CPSon 11
P1.3/ADC3 P2.0/A08/A16
29 OverKV 100K 6 - 124
25B
38 47 10nF BAT85 P1.4/ADC4 P2.1/A09/A17 1uF 100 25C
MAMonFb
R27 1K/0.1% 10nF /XrReset 12 30 OverLoad OP484 125
39 /ExtCtrl P1.5/ADC5/SS P2.2/A10/A18 OverMA XsnR1+ 25D
40 R62 SGND 13 P1.6/ADC6 P2.3/A11/A19 31 R28 C20 126 26Z
+5Vout Arcs 14 36 I_FilLim SGND XsnR1- 127
41
KVMonRef KVRef 1K AGND AGND AGND
P1.7/ADC7 P2.4/A12/A20
37 OpenFil 10nF 128
26A
42
MAMonRef MARef 47 16
P2.5/A13/A21
38 XrLamp R18 1K5 XsnT1+ 129
26B
43 I_FilRef
R5 IC2D P3.0/RXD P2.6/A14/A22 PSFaultOR/PSFault XsnT1- 26C
44 12 + R7 17 P3.1/TXD P2.7/A15/A23 39 470 SGND 130 26D
+5Vout 14 /XrayOn 18 131
45 +15Vout I_FilMon 1K XrayOffOR/XrayOff P3.2/INT0 27Z
46 13 - 19 P3.3/INT1/MISO ALE 42 R59 +5V +5V 132 27A
MonCom C8 OP484 3K3/0.1% 22 40 SGND XpscDisplay 133
47 SGND 10nF R11 D18 C18 23
P3.4/T0 EA/VPP
DisplayRd2 134
27B
48 -15Vout BAT85 P3.5/T1/CONVST 10K 27C
R77 R2 1K/0.1% 10nF MARegFlt 24 135
49 GND OpenFil SGND KVRegFlt 25
P3.6/WR
BAT85 FilPwmH DGND 136
27D
50 P3.7/RD 28Z
47 DAC0 9 R31 FilPwmL 137 28A
SUB-D 50PF 1K AGND AGND AGND 26 10 10 + IC3C D32 R25 HvPwmH 138
SCLOCK DAC1 D31 MAMonFb 28B
27 SDATA/MOSI 8 HvPwmL 139 28C
R100 +5V +5V 10K C32 9 - FilamentCurrent 140
41 OP484 BAT85 100 /AdcCs3Ext 141
28D
MARefMon PSEN 100pF C22 29Z
0 JP1 C93 R29 R102 10nF /AdcCs2Ext 142 29A
EnProg C52 5 AVDD 100nF /AdcCs1Ext 143 29B
R99 100nF 6 AGND SGND 0 SdataOut 144 29C
7 33K SGND Sclk 145
KVRefMon 8
CREF
DGND R19 AdcSars 146
29D
0 VREF 30Z
AGND 10K SdataIn 147
C82
30A
C81
C42
C45
C55
C61
C62
C75
C51
C64
R101
C73
C79
C68
C69
C71
+5V +5V
+5V
R45 R42
+5V 1K 1K
D6 +5V R68
D9 NotOk/Red 470
C1 C9 Ok/Green IC6B
3
100nF IC11A XsnR1+
16
Heatsink
S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 10
+24V C12
P1
+12V
1 3 220pF +200VRTM
2 4 FAN R68 C64
WAGO 2P AGND +5V
10nF
4
10K ToHvTransfo
J906 PGND R67 IC5A 3
R7 10K OP484 - 9 1
+
D20 IC5D - 13 8 2 R18
R76 -
1 /PwmTrip 14 R78
+ 10 10K OP484 10K
1Z Vref+2V
2 + 12 IC5C C66 T1A T1B
11
1A 1K XpsmHvTemp
1B 3 GND BAT85 OP484 47pF
4 MF6 D21 1K C67
1C BAT85 C65
1D 5 +24Vin C11 10nF C35 RV2
6 10nF 10nF R66 -12V AGND 100nF 4K7/25
4
2Z
2A 7 MF-R110 10K AGND
8 PGND DGND AGND T1
2B +12Vin R64
2C 9 +12V AGND AGND
2D 10 AGND R69 HvTransfo
11 R65 1K 30K
3Z AGND -12Vin IC5B
3A 12 5 10K C72 Not on the Board
13 7
+ AGND IC4B 5 C63
2
3B + AGND
3C 14 - 6 7 10nF
15 +5Vin OP484 6 2.7nF
3D -
16 OP484 AGND
4
4Z DGND
4A 17 R8 R41 R63
18 IC4A 3
4B R59 +
4C 19 1K 1
20 2 1K 20K
4D
21 1K OP484
-
5Z HV2 R28
22 D11 R48
11
5A Arc C44
5B 23
24 C40 1nF 200M
5C 100 18x 4n7/3kV
5D 25 1N4148 10nF
6Z 26 -12V
6A 27 Line1 R21 C34 AGND GAP6
28 Line1 C33 100K D12 R35 M51-C90
6B
29 10uF BZX85C5V6 AGND C1 C2 C3 C4
6C 10nF
6D 30 4x 20nF/12kV
31 IC4C 10 1K
7Z + AGND
7A 32 AGND 8 GND
7B 33 AGND AGND AGND - 9
7C 34 OP484 D1 D2 D3 D4 D5 D6 D7 D8 D9 D10
35 MF5 R13
7D IC4D + R42 HV1
8Z 36 +5V R60 12
8A 37 Line2 14 1K 20x Z100UFG
38 Line2 13 MF-R050 15/3W GAP1
8B
39 D22 1K -
A81-C230 D1' D2' D3' D4' D5' D7' D8' D9' D10'
8C BAT85 OP484
40 C45 Wired resistors
8D JP1
9Z 41 R26 R49 R36 1uF GAP3
42 EmissionCurrent+ M51-C90 GND C6 C7 C8 C9 C10 5x 20nF/12kV R1 R2 R3 R4 R5 R6 R7 R8
9A S
9B 43 100 1K L
44 100 D13 AGND
9C C
9D 45 BAT85 8x 100/9W
10Z 46 R11 GND D13' D14' D15' D16' D17' D18' D19' D20' D21' D22' Fed.Standard
47 EmissionCurrent- D10
10A
10B 48 100 1N5347
10C 49 AGND C32 20x Z100UFG
10D 50 C25 R40 R34 R30 R25 100uF/35V
11Z 51 10nF C9 D13 D14 D15 D16 D17 D18 D19 D20 D21 D22
11A 52 HP:4x100
11B 53 +400VRTM 1nF
54 AGND LP:1x100
11C
11D 55 R55 C12 C13 C14 C15
12Z 56 4x 20nF/12kV
12A 57 AGND
58 53k3 HT Output board S940188
12B
12C 59 +12V
12D 60 R14 DV1 5x Multiplier board S930191
61 OP484 - 6
13Z
13A 62 +5V C62 7 56K
13B 63 + 5 GAP2 Ra1
64 10nF IC3B A81-C230
13C R58
13D 65 D14
66 AGND 1K
4
14Z BAT85
14A 67 AGND GAP4 GND GND
68 IC3A 3 D8 C1
14B HvDividerDifference R38 R32 + M51-C90
14C 69 1 1N5347 10uF 5x 200M/15kV 5x 470pF/15kV Divisor board S940189
70 2 R61 Wired resistors
14D
71 1K 1K OP484 -
10K
15Z +200VRTM Vref+2V R17 R16 R15 R14 R13
72 D16 R12 R11 R10 R9
11
15A
15B 73 C43 BAT85 C39
15C 74 10nF C74 C10 AGND GND AGND
75 10nF 10nF 4x 1K/9W
15D
16Z 76 -12V 1nF C21 C20 C19 C18 C17
16A 77 AGND
16B 78 AGND R28 AGND R23 R47
79 AGND R27 C27
16C
16D 80 10K 1K 53k3 33M/9kV 470pF/15kV
17Z 81 +5V
17A 82 Interface board
17B 83 D19
17C 84 S930220 R26 R25 R24 R23 R22
85 IC3D 12 BAT85 R21 R20 R19 R18
17D + AGND OP484 R15 DV2
18Z 86 14 R62 R51
- 9
87 13 HighVoltage 8
18A
88 OP484
- + 10 56K GAP3 Ra2 4x 1K/9W
18B 1K 100 IC3C
18C 89 A81-C230 C26 C25 C24 C23 C22
18D 90 R6 D18 Wired resistors
19Z 91 BAT85 5x 200M/15kV 5x 470pF/15kV Divisor board S940189
19A 92 1K AGND GAP5 GND GND
19B 93 0VRTM D9 C8
94 1N5347 M51-C90
19C
95 AGND
10uF Multiplier Box S703952 (not on the board)
19D
20Z 96
20A 97
20B 98
20C 99 AGND GND AGND
20D 100
21Z 101
21A 102
21B 103 GND
21C 104
21D 105
22Z 106
22A 107
22B 108
22C 109
22D 110 +5V
23Z 111
112 ToHvTransfo T2B
23A ToFilamentTransfo T2
23B 113
114 R4
23C
23D 115 10K 2
24Z 116
117 XpsmFilamentTemp 17
24A
24B 118 2
24C 119 C17 RV1 T2A T3
120 100nF 4K7/25 +200VRTM
24D
25Z 121 Not on the Board
25A 122 AGND FilamentTransfo
25B 123 AGND 30
25C 124
25D 125
26Z 126 T2D T2C T3A T3B
26A 127
128 ToFilamentTransfo D2 D1 D6 D23
26B
26C 129
26D 130
27Z 131
27A 132
133 D3 D4 D5 D7
27B
27C 134
27D 135 +5V 4xBYV26 4xBYV26
28Z 136 FilPwmH
28A 137 FilPwmL AGND AGND
28B 138 HvPwmH D17
28C 139 HvPwmL
140 FilamentCurrent BAT85 R2
28D /AdcCs3
29Z 141 +5V +12V -12V R45 R24 1K3 R22 1V/A
142 /AdcCs2 FilamentVoltage 0.242V/Vfil FilamentCurrent 1V/A
29A
29B 143 /AdcCs1 1K 1K
144 SdataOut C41 100 R5
29C Sclk R46 C56
29D 145 R31 R43 100uF/35V D15 39
146 AdcSars 2K7 18K 33K BAT85
30Z SdataIn 1uF
30A 147 R3
30B 148 CurrentControl +5V C69 C70 C71 AGND 430
149 CurrentLimit 8x8bit ADC R27 100nF R37 100nF R50 100nF AGND AGND
30C 10K
30D 150 PowerOn Status AD converter 10K 10K
151 EmissionCurrent+ /AdcCs3 R70 100 AGND AGND
31Z HighVoltage C38 IC2
31A 152
153 EmissionCurrent- SdataOut R71 100 10nF 20 1 AGND AGND AGND AGND AGND
31B Vref+5V VCC CH0
31C 154 19 V+ CH1 2
155 Vref+2V Sclk R72 100 AGND 18 3 Vref+5V
31D CS CH2 Arc
32Z 156 InputCurrent 17 DI CH3 4
157 InputVoltage AdcSars R73 100 16 5 HvDividerDifference
32A CLK CH4 FilamentVoltage
32B 158 Mon400V 15 SARS CH5 6
159 SdataIn R74 100 14 7 XpsmFilamentTemp
32C DO CH6 XpsmHvTemp C68
32D 160 +5V 13 SE CH7 8
GND Vref+5V R54 12 9 10nF
VREF COM
10 11 AGND DGND 10 R75
C61 C53 ADC0838
DIN41612 10uF 1nF AGND 0
DGND AGND
AGND AGND
Accessories
TP3 TP1
L1 +12V L3 +5V
MF1 MF3 Hole Hole Hole Hole
+12Vin +5Vin Mod: 1 MOD23923 16.11.05 PPa 4 OM4204 27.04.11 PPa
+12V +5V Hole Hole Hole Hole 2 Add. OM3865 21.02.06 PPa 5 OM4204 Add.2 06.07.11 PPa
MF-R110 C4 C7 TP4 MF-R110 C2 C5 Hole Hole Hole Hole
220uF 100nF 220uF 100nF 3 OM4187 11.10.10 PPa
R1 AGND
AGND Hole Hole Hole Hole
22.07.02 MBz Remplacé par:
GND 0 C3 C6 TP2 Hole Hole Hole Hole X-ray Power Supply Mutiplier Assembly Echelle Dessiné
Layout : S112248
GND
ThermoFisher S112247
No Mod.
CATALOGUE
Catalogue Chapter 9
9 CATALOGUE
Foreword
Some spare parts are marked with a blue label containing the following text:
This component was developed and tested by Thermo Fisher Scientific to comply with our
instrument's technical specifications in order to guarantee and maintain analytical performances
as well as safety standards.
For safety reasons, this part must be provided only through Thermo distribution companies. Any
attempt to fit a part provided by another supplier than Thermo Fisher Scientific may result in
damages and loss of performances, and thus render Thermo Fisher Scientific warranty and
responsibility void.
Avant-propos
Certaines pièces de rechange sont marquées par une étiquette bleue comportant le texte suivant:
This component was developed and tested by Thermo Fisher Scientific to comply with our
instrument's technical specifications in order to guarantee and maintain analytical performances
as well as safety standards.
For safety reasons, this part must be provided only through Thermo distribution companies. Any
attempt to fit a part provided by another supplier than Thermo Fisher Scientific may result in
damages and loss of performances, and thus render Thermo Fisher Scientific warranty and
responsibility void.
Pour des raisons de sécurité, cette pièce doit être fournie par le réseau de distribution Thermo
Electron. Toute tentative de montage d'une pièce provenant d'une autre source de distribution que
celle de Thermo Fisher Scientific pourrait engendrer des dommages et une perte de performances,
et par conséquence rendre la responsabilité et la garantie de Thermo Fisher Scientific nulle.
Vorwort
Einige Ersatzteile werden mit einem blauen Aufkleber geliefert, der folgenden Text enthält:
This component was developed and tested by Thermo Fisher Scientific to comply with our
instrument's technical specifications in order to guarantee and maintain analytical performances
as well as safety standards.
For safety reasons, this part must be provided only through Thermo distribution companies. Any
attempt to fit a part provided by another supplier than Thermo Fisher Scientific may result in
damages and loss of performances, and thus render Thermo Fisher Scientific warranty and
responsibility void.
Aus Sicherheitsgründen muss dieses Teil über Thermo-Unternehmen oder –Vertretungen bezogen
werden. Jeglicher Versuch, Teile aus anderen Bezugsquellen einzubauen, kann zu Schäden und
Leistungsverlusten führen und damit die Thermo Fisher Scientific Garantie und Verantwortung
nichtig machen.
Description
List
Mod. Description Note
KXx…
Beschreibung
Remark :
KXXGENX00-01
LIST GENX00
Note :
Figure 9.1
Note :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX10-01
LIST GENX10
Note :
Figure 9.2
Note :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX20-01
LIST GENX20
Note :
Figure 9.3
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX III 4.2kW 60kV/140mA
1 S703681T 1 GenX III 4.2kW 60kV/140mA B
………..
XPSC4260 III board
2 S703724T 1 Carte XPSC4260 III A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Note :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX30-01
LIST GENX30
Note :
Figure 9.4
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX IV 4.2kW 70kV/140mA
1 S703682T 1 GenX IV 4.2kW 70kV/140mA B
………..
XPSC4270 IV board
2 S703725T 1 Carte XPSC4270 IV A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Note :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX40-00
LIST GENX40
Note :
Figure 9.5
Description Maintenance:
N P/N Qty Description Type Note greyed items
° Beschreibung have to be
considered
GenX V 1.2kW 60kV/60mA
1 S703683T 1 GenX V 1.2kW 60kV/60mA B
………..
XPSC1260 V board
2 S703726T 1 Carte XPSC1260 V A
…………..
XPSI board
3 S703778T 1 Carte XPSI A
………..
XPSP board
4 S703779T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit II, fans GENXRFLP Replace fans after
6 S703839 1 Kit II ventilateurs GENXRFLP C 1 2 years
……………
Fan for XPSI & XPSP Replace fans after
7 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly. Replace fans after
8 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Note :
1 Positions 7,8 are included in this item.
KXXGENX50-01
LIST GENX50
Note :
Figure 9.6
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX R 3.6kW 70kV/120mA
1 S703685T 1 GenX R 3.6kW 70kV/120mA B
………..
XPSCRET PR board
2 S703784T 1 Carte XPSCRET PR A
…………..
XPSR36 PR board
3 S703125T A
Carte XPSR36 PR
XPSI board
4 S703123T 1 Carte XPSI A
………..
XPSP board
5 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
6 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
7 S703838 1 Kit I ventilateurs GENXRFHP C 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
8 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
S704791 Fan “big” for XPSI & XPSP Replace fans after
9 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
10 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
11 A040026 4 Raccord rapide femelle
……………
Fast fitting male
12 A040027 2 Raccord rapide male
……………
Fixed Fitting male
13 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
14 AA22511 8 Collier D9-D16
………………
Pipe D16/10
15 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX60-01
LIST GENX60
Note :
Figure 9.7
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX R 4.2kW 70kV/140mA
1 S703687T 1 GenX R 4.2kW 70kV/140mA B
………..
XPSCRET board
2 S703784T 1 Carte XPSCRET A
…………..
XPSR42 PR board
3 S703785T A
Carte XPSR42 PR
XPSI board
4 S703123T 1 Carte XPSI A
………..
XPSP board
5 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
6 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
7 S703838 1 Kit I ventilateurs GENXRFHP C 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
8 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
9 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
10 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
11 A040026 4 Raccord rapide femelle
……………
Fast fitting male
12 A040027 2 Raccord rapide male
……………
Fixed Fitting male
13 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
14 AA22511 8 Collier D9-D16
……………
Pipe D16/10
15 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX70-01
LIST GENX70
Note :
Figure 9.8
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX VI 2.5kW 60kV/100mA
1 S704025T 1 GenX VI 2.5kW 60kV/100mA B
………..
XPSC2560 VI board
2 S704026T 1 Carte XPSC2560 VI A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX80-02
LIST GENX80
GENXRD GENERATOR
GENXRD GENERATEUR
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Note :
Figure 9.9
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenXRD 3.0kW
1 S703720T 1 GenXRD 3.0kW B
………..
XPSCD 3060 board
2 S703721T 1 Carte XPSCD 3060 A
…………..
XPSRD3060 board
3 S703876T A
Carte XPSRD3060
XPSI board
4 S703778T 1 Carte XPSI A
………..
XPSP board
5 S703779T 1 Carte XPSP A
…………..
Multiplier assembly XRD 3
6 S703953T 1 kW A
Ensemble Multiplicateur XRD
3kW
………….
Kit II, fans GENXRDAIR Replace fans after
7 S705115 1 Kit II ventilateurs C 1 2 years
GENXRDAIR
……………
Fan for XPSI & XPSP Replace fans after
8 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly Replace fans after
9 A040863 1 XRD 3KW. C 2 years
Ventilateur pour ensemble
multiplicateur XRD 3KW
……………
Note :
1 Positions 7,8,9 are included in this item.
KXXGENX90-02
LIST GENX90
Note :
Figure 9.10
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenXRD RET 3.0kW
1 S703720T 1 GenXRD RET 3.0kW B
………..
XPSCDR 3060 board
2 S703721T 1 Carte XPSCDR 3060 A
…………..
XPSRD3060 board
3 S703876T A
Carte XPSRD3060
XPSI board
4 S703778T 1 Carte XPSI A
………..
XPSP board
5 S703779T 1 Carte XPSP A
…………..
Multiplier assembly XRD 3 kW
6 S703953T 1 Ensemble Multiplicateur XRD A
3kW
………….
Kit II, fans GENXRDAIR Replace fans after
7 S705115 1 Kit II ventilateurs C 1 2 years
GENXRDAIR
……………
Fan for XPSI & XPSP Replace fans after
8 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly Replace fans after
9 A040863 1 XRD 3KW. C 2 years
Ventilateur pour ensemble
multiplicateur XRD 3KW
……………
Note :
1 Positions 8,9 are included in this item.
KXXGENX100-03
LIST GENX100
GENXRDF18 GENERATOR
GENXRDF18 GENERATEUR
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Note :
Figure 9.11
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenXRDF18, 1.8 kW
1 S704385T 1 GenXRDF18, 1.8 kW B
………..
XPSC XRDF18 board
2 S704387T 1 Carte XPSC XRDF18 A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly XRD 3 kW
5 S703953T 1 Ensemble Multiplicateur XRD A
3kW
………….
Kit I, fans GENXRDWATER Replace fans after
6 S705114 1 Kit I ventilateurs C 1 2 years
GENXRDWATER
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A040863 1 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7, 8, 9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX110-02
LIST GENX110
GENXRDV18 GENERATOR
GENXRDV18 GENERATEUR
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Note :
Figure 9.12
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenXRDV18, 1.8kW
1 S704368T 1 GenXRDV18, 1.8kW B
………..
XPSC XRDV18 board
2 S704379T 1 Carte XPSC XRDV18 A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly XRD 3 kW
5 S703953T 1 Ensemble Multiplicateur XRD A
3kW
………….
Kit I, fans GENXRDWATER Replace fans after
6 S705114 1 Kit I ventilateurs C 1 2 years
GENXRDWATER
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A040863 1 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7, 8, 9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX120-00
LIST GENX120
Note :
Figure 9.13
Description Maintenance:
N P/N Qty Description Type Note greyed items
° Beschreibung have to be
considered
GenX VII 600 W 60kV/30mA
1 S704818T 1 GenX VII 600 W 60kV/30mA B
………..
XPSC0660 VII board
2 S704819T 1 Carte XPSC0660 VII A
…………..
XPSI board
3 S703778T 1 Carte XPSI A
………..
XPSP board
4 S703779T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit II, fans GENXRFLP Replace fans after
6 S703839 1 Kit II ventilateurs GENXRFLP C 1 2 years
……………
Fan for XPSI & XPSP Replace fans after
7 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly. Replace fans after
8 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Note :
1 Positions 7,8 are included in this item.
KXXGENX130-01
LIST GENX130
Note :
Figure 9.14
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX VIII 4.2kW 70kV/140mA
1 S705076 1 GenX VIII 4.2kW 70kV/140mA B
………..
XPSC4270 board VIII
2 S704931T 1 Carte XPSC4270 VIII A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX140-01
LIST GENX140
Note :
Figure 9.15
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX IX 2.5kW 60kV/100mA
1 S705075 1 GenX IX 2.5kW 60kV/100mA B
………..
XPSC2560LC IX board
2 S704930T 1 Carte XPSC2560LC IX A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX150-02
LIST GENX150
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Note :
Figure 9.16
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX X 4.2kW 60kV/120mA
1 S705178 1 GenX X 4.2kW 60kV/120mA B
………..
XPSC4260LC board X
2 S705173 1 Carte XPSC4260LC X A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX151-02
LIST GENX151
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Note :
Figure 9.17
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX XI 1.5kW 60kV/70mA
1 S705074 1 GenX XI 1.5kW 60kV/70mA B
………..
XPSC1560LC board XI
2 S704929T 1 Carte XPSC1560LC XI A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit II, fans GENXRFHP Replace fans after
6 S703838 1 Kit II ventilateurs GENXRFHP C 1 2 years
……………
Fan for XPSI & XPSP Replace fans after
7 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly Replace fans after
8 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Note :
1 Positions 7,8 are included in this item.
KXXGENX152-00
LIST GENX152
Note :
Figure 9.18
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX XII 2.5kW 60kV/100mA
1 S705219 1 for X-Ray Tube UCCO30 B
GenX XII 2.5kW 60kV/100mA
pour Tube rayon-X UCCO30
………..
XPSC2560LCU30 XII board
2 S705217 1 Carte XPSC2560LCU30 XII A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX153-00
LIST GENX153
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Note :
Figure 9.19
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX XIII 4.0kW 60kV/140mA
1 S705220 1 for X-Ray Tube UCCO30 B
GenX XIII 4.0kW 60kV/140mA
pour Tube rayon-X UCCO30
………..
XPSC4260LCU30 XIII board
2 S705218 1 Carte XIII XPSC4260LCU30 A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX160-00
LIST GENX160
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Introduction Gen’X 9900 from serial number 9953964
0 May 2020 including 9953165, 9953396, 9953835, 9953933, 9953941,
9953943.
Note :
Figure 9.20
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX 1.5kW 60kV/60mA
1 1226736 1 GenX 1.5kW 60kV/60mA B
………..
CARTE,XPSC,GENX 9900
2 1226732 1 1.5kW SP A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX161-00
LIST GENX161
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Introduction Gen’X 9900 from serial number 9953964
0 May 2020 including 9953165, 9953396, 9953835, 9953933, 9953941,
9953943.
Note :
Figure 9.21
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX 2.5kW 60kV/100mA
1 1226748 1 GenX 2.5kW 60kV/100mA B
………..
CARTE,XPSC,GENX 9900
2 1226741 1 2.5 kW A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX162-00
LIST GENX162
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Introduction Gen’X 9900 from serial number 9953964
0 May 2020 including 9953165, 9953396, 9953835, 9953933, 9953941,
9953943.
Note :
Figure 9.22
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX 4.2kW 60kV/120mA
1 1226757 1 GenX 4.2kW 60kV/120mA B
………..
CARTE,XPSC,GENX 9900
2 1226755 1 4.2 kW A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………
Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.
KXXGENX199-03
LIST GENX199
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Position 1 updated
1 Position 1 mise à jour 4112
Position 1 auf den letzten Stand gebracht
Positions 4 and 5 added
2 Positions 4 et 5 ajoutées
Positionen 4 und 5 hinzugefügt
Positions 6 to 8 added
3 Positions 6 à 8 ajoutées
Positionen 6 bis 8 hinzugefügt
Note :
Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
Kit I, fans GENXHP Replace fans
1 S703838 1 Kit I ventilateurs GENXHP C after 2 years
………..
Kit II, fans GENXRFLP Replace fans
2 S703839 1 Kit II ventilateurs GENXRFLP C after 2 years
…………
Loctite 542 to seal water
3 A031375 1 fittings
Loctite 542 étanchéité
……………
Kit I, fans GENXRDWATER
4 S705114 1 Kit I, ventilateurs
GENXRDWATER
……………
Kit II, fans GENXRDAIR
5 S705115 1 Kit II, ventilateurs
GENXRDAIR
……………
Fuse 0.8A (XPSI board)
6 AA20670 2 Fusible 0.8A (Carte XPSI)
Note :
GENXRD X'TRA
GENXRD X'TRA Chapter 10
10 GENXRD X'TRA
Description of the main functions
The GenXRD X-Ray power supply is made of 5 main sub-units:
Mains XPSI
XrayPowerSupplyInput
Board
Master processor XPSRD
Control XrayPowerSupplyRetrofitXRD
Board
XPSCD
XrayPowerSupplyControlXRD XPSB
Board Xray
Power
XPSP Supply
XrayPowerSupplyPowerXRD Backplane
Board Board
XPSMD
XrayPowerSupplyMultiplier
Board
Multiplier
Multiplier And
HighVoltage Box Filament
Assembly
Figure 10.1
Working principle
u
i FET2
C1 +200Vrtm Multiplier L
u
C
C2
0Vrtm
XPSRD
uP 1000 M Ohm
i
400Vrtm
+200Vrtm
PI PWM
0Vrtm
Filament -HV
Transformer W filament
-e X-Rays
Cu anode
Figure 10.2
XPSI power components (Power-Mosfet) are cooled by a combined radiator-half pipes system.
XPSI does convert the 230VAC main power to 360 - 420 VDC.
XPSI does manage the inrush current, the working current limitation and the power factor correction.
XPSRD functions are the emulation of a Spellman X-Ray power supply, the interfacing of the analog to
digital signals and vice versa with XPSCD.
XPSCD does manage the kV regulation, the emission current regulation of the X-Ray tube. It is the brain of
the generator.
XPSP power components (Power-IGBT) are cooled by a combined radiator-half pipes system.
The XPSP board converts the intermediate DC voltage (360 V- 440 V) into AC voltage sources, which
supplies current to the filament transformer and the HV transformer.
XPSMD board
Filament transformer
HV transformer
Multiplier box with voltage inversion
The XPSMD board has three main functions: the filament power supply, the HV measurement and the
emission current measurement.
The filament transformer via the XPSM board is used to supply up to 50 W to the X-ray tube filament.
The HV transformer via the multiplier is used to supply up to 4000 W to the X-ray tube anode.
The backplane or motherboard print is a bus supporting 5 slots made of 5 x 32 pins female internal
connectors (J902-J906) and 2 5 x 32 pins female external connectors J901 and J907, of Europe format.
Each board has its specific position in the rack. There are no keys to prevent the connection of a board to the
wrong slot.
Figure 10.3
The main functions controlled by the XPSI board are the mains (230VAC) power factor correction for the
current, the mains current limitation and the X-ray lamps control.
The mains voltage is used via a filter and a relay. When the generator switches on, the relay is energized to
close both contacts. After the rectifier diodes, the FET1 is used to open the circuit if the current is above a
current limit. When the FET1 is open, the current flow through the integrated diode D1of the FET circuit,
due to the inductor L. When the current is below a current limit the FET1 will drive.
FET1 L D2 +400Vrtm
L
C1
IL FET2 +12kV
230Vac U1 D1 U2 C2
N 0Vrtm
uC DSP
Figure 10.4
The FET2 and the integrated diode D2 of the FET circuit are used to step up the voltage to 400 V +/- 10 %
this voltage is adapted according to the kV output voltage. 360 V for 0 to 30 kV, 360 V to 420 V
proportionally from 30 kV to 60 kV).
Each 50 µs the current “IL”, the voltages “U1” and “U2” are evaluated in order to define the FET2 OFF/ON
conduction time. The maximum OFF/ON time is 25 us.
V U1
IL
UD1
U2
FET2
OFF
U1
IL
FET2
ON
25 us
Figure 10.5
During the OFF time the current flows through D2 to load the capacitors C1 and C2. The C1/C2 voltage is
measured by the DSP and compared to the set-point value. The On time is regulated relative to this
comparison.
− The IGBTs on the XPSP power board, generating the anode high voltage, via the transformers and the
multiplier.
− The IGBTs on the XPSP power board, generating the cathode (filament) heating current.
Proportional 400 V
U2 400V Monitoring
420 V
360 V
HV
30 kV 60 kV
Figure 10.6
The neutral potential does cross a resistor divider before to be applied to the neutral to ground comparator IC
6.
This device allows the DSP to detect a floating neutral or a neutral-phase inversion.
Figure 10.7
This power board contains the switching components (IGBTs) for the high voltage transformer and for the
filament power supplies.
Part of XPSP HV
Transformer
FET1 L IGBT1 C3
230Vac D +400V
HighVoltage
C1
u
i FET2
u +200Vrtm
Multiplier
C2 IGBT2
0Vrtm
External
Master
Board
uC 1000 M Ohm
Part of XPSP
i
IGBT3
400Vrtm
PI PWM +200Vrtm
0Vrtm Filament
IGBT4 Transformer
Figure 10.8
The HV switching works at 20 kHz. The transformer primary is connected to the C1/C2 point which
corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V).
The control board commands alternately the IGBT1 and IGBT2. The On time duration depends of the output
power needed. By this way, the transformer primary receives, via C3, plus and minus 200 V referenced to
the fix 200 VRTM.
UC3-0Vrtm
+400V
+200V
0 t
IGBT1
Command
IGBT2
Command
Figure 10.9
UC3-0Vrtm
+400V
+200V
0 t
IGBT1
Command
IGBT2
Command
Figure 10.10
The transformer inductor and C3 capacitor make a resonant circuit. By this way the current look like
sinusoidal:
UC3-0Vrtm
+400V
+200V
0 t
Transformer
Primary
current
Figure 10.11
The filament power supply switching works at 40 kHz. The transformer primary is connected to the C1/C2
point which corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V).
The control board commands alternately the IGBT3 and IGBT4. The On time duration depends of the output
power needed (max. 50 W). By this way, the transformer primary receives positive and negative 200 V
referenced to the fix 200 VRTM.
See after the voltage and the current at the primary of the filament transformer:
Filament
transformer
+200V
-200V t
Filament
Transformer Primary
current
0 t
Figure 10.12
The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.
u
i FET2
C1 +200Vrtm Multiplier
u
C2
0Vrtm
XPSRD
uP 1000 M Ohm
i
400Vrtm
+200Vrtm
PI PWM
0Vrtm
Filament
Transformer
Figure 10.13
♦ the HV supply
♦ the emission current measurement
♦ the filament current supply
The 400V from the IGBTs on the XPSP board, via XPSM board goes to the HV transformer. At the
transformer secondary the voltage changes according to the IGBT1 and IBGT2 ON time.
The maximum voltage is near 12000 V. The full wave multiplier increases five times this voltage to reach at
maximum - 60000 V. Two dividers are used to monitor the high voltage. The command board uses one of
them to control the HV via the servo loop function accordance with the kV set point. The second is used to
control the first via a subtraction function.
The emission current is measured with the current absorbed by the multiplier. The emission current
measurement allows to indirectly regulating the filament heating current, thus to match with the desired
emission current.
The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.
Figure 10.14
HV
Transformer
230Vac FET1 L D +400Vrtm
HighVoltage
C1
u
i FET2
u +200Vrtm
Multiplier
C2
0Vrtm
XPSRD
uP 1000 M Ohm
XPSCD
400Vrtm
Part
+200Vrtm
PI PWM
0Vrtm Filament
Transformer
Figure 10.15
HV servo-loop control
The control board receives the asked HV from instrument master board via XPSRD. The asked HV is
compared with the real HV to generate the HV error, which is used through the PID corrector to PWM duty-
cycle.
The PWM signals control the power stage on XPSP board to supply the right voltage according to the HV
error. The XPSM assembly transformer and multiplier increase this voltage to generate the real HV. The
DSP evaluates the values each 50 us.
uC DSP
HV
HV Multiplier
Power
set point + PID PWM Box
- Stage
HV
D HighVoltage 18762
A
Part of Part of
Part of XPSP board XPSMD assembly
XPSCD board
Figure 10.16
uC
Emission
current
set point
+
-
PID Filament current power supply
HV
Power Multiplier
Stage
D -1
A
R
Part of Part of
Part of XPSCD board XPSP board XPSMD assembly
Figure 10.17
The rectified input voltage is weighted by this voltage error to become the current reference for a second
servo-loop. The real current, to generate the current error, is compared to this reference. The GPFC corrector
uses the current reference, the real current and the current error values to control the PWM duty-cycle, which
controls the Power MOS FET (FET2).
The FET1 is controlled by three ways. The first one is the intermediate voltage via a P corrector. The PWM
duty-cycle changes according to intermediate voltage. The second one is made on the real current value and
a comparator with a current threshold. At switching power on, this current threshold corresponds to 2
amperes to limit the inrush current. When after 10 seconds, the intermediate voltage exceeds 320 V, the
current threshold changes to 40 Amperes. This corresponds to the maximum acceptable current during power
on state. The last way controls the FET1 by hardware. The real current is compared with a reference value.
When the real current exceeds 40 amperes the FET1 is switched off. This circuit allows reacting faster than
the DSP, where its reaction time is related to the sampling period.
FET2
PWM
A
D
P
-
A +
D
P P
DSP
PI
GPFC PWM
Reference D
+ -
Voltage A
Figure 10.18
The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogic to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.
Figure 10.19
50 P CABLES
SUB-D SUB-D
50 P J401
MASTER
CONTROL
BOARD
XPSRD
XPSCD DISPLAY
XPSI
INSTRUMENT GenXRD
XPSP
XPSMD
Figure 10.20
GenXRD
Safety
And
Water
interface
Figure 10.21
The low voltage power supply generator is supplied by the 230VAC Line1 and Line2 via XPSI Board.
Figure 10.22
Figure 10.23
The regulated 24 VDC is applied to five low voltage regulators used to generate respectively
+15 VDC
-15 VDC
+5 VDC
+12 VDC
- 12 VDC
+24Vout does supply the 24 V to all the boards plugged on the GenX mother board.
XPSRD GEN'X
MASTER
PROCESSOR
BOARD EMULATOR XPSCD
ADUC812
Outgoing
Signals To/From
Bus XPSCD
Serial Data Link
Ingoing
Signals
Bus
Figure 10.24
To do so:
• The various alarms generated by XPSCD are linked to the ADCU812 processor through a serial
link and converted, to be recognised by MASTER PROCESSOR BOARD.
• The start up synchronisation and safety control signals generated by MASTER PROCESSOR
BOARD are converted to serial Data and transmitted to XPSCD.
• The analogic Ref signals are converted to serial Data by the ADUC812 built in ADC, to be
transmitted to XPSCD.
• The Monitoring signals are converted from serial Data by the ADUC812 built in DAC to analogic
voltages to be supervised by MASTER PROCESSOR BOARD.
KVMonFb XPSRD
KVRef
PowerRef
Master I_FilRef
Board
MaRef
MaMonFb
Safety
Interlock
XrayOn
Control
Figure 10.25
Serial link
Serial link is of the same type as XSN (X-ray Services Network) used in the 9900 / ADVANT’X series
instruments electronic racks.
The jumper W1 must be placed between contacts 1-2 internal, for proper communication.
Jumper settings
JUMPER SETTING
W1 1-2 Internal
W2 Installed
JP1 EnProg Removed
R98, R102 Removed
R99, R100 Installed
Part 3 Display
Schematic and layout: S215199
Figure 10.26
The 1x 20 characters display is mounted and plugged on the display front panel.
kV Regulation
kV > 0 No No test
Yes
Yes kV > 5 kV No
Yes
Yes No
Figure 10.28
mA Regulation
mA > 0 No No test
Yes
Yes mA > 10 mA No
Yes
Yes No
Figure 10.29
8280 Unknown identity X-ray power supply fault. POWER SUPPLY FAULT
8282 X-ray tube filament power supply fault. POWER SUPPLY FAULT
Heating current is < 1 A and heating voltage is < 2 volts
Fuse F2 2A (orange soldered fuse) of XPSP board
SG3525AN IC 20 of XPSC.
Filament heating power IGBTs on XPSP.
The filament power supply part of the XPSM board (part of the Multiplier assembly) can be suspected.
8292 Generators mains voltage not present, time-out POWER SUPPLY FAULT
If main is < 280VAC. In theory the voltage is 230 × 2 See trouble shooting section 4 step 4
8293 Generator 400V not > 250V, time-out POWER SUPPLY FAULT
Time out = 5 sec. See trouble shooting section 4 step 5
8294 Generator 400V not > 380V, time-out POWER SUPPLY FAULT
Time out = 5 sec. See trouble shooting section 4 step 6
8297 XPSCD First 8 bits A/D converter error POWER SUPPLY FAULT
8298 XPSCD Second 8 bits A/D converter error POWER SUPPLY FAULT
Communication error.
S Power up
8305 Power off, issued by DSP; 400V too low (<350V) POWER SUPPLY FAULT
or 8513
Too fast X-Ray power supply transition. 400 V voltage drop. May happen when changing X-Ray tube
conditions in terminal mode, in “On Fast mode” and “Operational Mode”.
See trouble shooting section 4 step 6
8306 Power off, issued by DSP; HV too high (>71kV) OVER VOLTAGE
or 8514
KV Regulation error. KV Regulation loop fault.
8307 Power off, issued by DSP; Current too high (>35A) POWER SUPPLY FAULT
or 8515
After a power up retry the current can rise above 35 A.
8308 Mains voltage below 195 VAC (after 20 ms) POWER SUPPLY FAULT
or 8516
8309 Mains voltage above 253 VAC (after 20 ms) POWER SUPPLY FAULT
or 8517
Main power supply fault. The main voltage is checked by the DSP. If out of tolerance the DSP does
immediately shut down the power.
See trouble shooting section 4 Step 4
8310 Unknown reply from DSP, issued by XPSCD POWER SUPPLY FAULT
80C196KR no dialog with DSP.
S Power up
8311 Too many parameters in SSIO, issued by DSP POWER SUPPLY FAULT
or 8519
S Power up
8313 PWMTRIP -> Power Off, issued by DSP POWER SUPPLY FAULT
or 8521
Issued by XPSP due to repetitive power stage problems, where the current is > 100A
Check the IGBT.
Issued by XPSM, due to arcs or spark over on the multiplier assembly of XPSM.
If > 15 kV (> 2 V at comparator input) at the secondary of the HV transformer of XPSM. Normal 12kV for
60 kV
K1 i
L
IL C1 +12kV
230Vac U1 U2
C2
N
15Vrtm
uC DSP
Figure 10.30
At power up if the 32 Amp breaker is on, the voltage leaks from T1 through the drivers to the capacitors C.
The 400 V measuring device does read about 15 V. If not, the 32 Amp breaker is down or the IGBTs of
XPSP are in short circuit.
8362 XPSCD temperature Board overflow POWER SUPPLY FAULT
If > 109 ° C. Out of conversion limits. Mainly indicating NTC is defect.
An additive Alarm is available and can be monitored on the display. If the INTERLOCK is open,
the next message is displayed.
INTERLOCK OPEN
At the start–up of the GenXRD the firmware versions are displayed in the following manner:
XPSRD V3.0-0
XPSCRD V3.0-0
XPSCS V3.2-6
Set the potentiometer P2 to define the maximum power limit relative to the next table.
Tube Anode
Power
Specifications
# DESCRIPTION VALUES REMARKS
1 Output power 0 to 4KW
2 Input voltage 230VAC +10% / -15%
50/60Hz 1phase
3 Maximum input current and inrush current 28A at 195VAC
4 Maximum input power 5000W
5 Efficiency > 80%
6 Power factor > 0.9 switched mode load type
7 Output voltage -10 to -60KV Without overshoot.
Short-circuit protected.
8 Output voltage vs Input voltage < 50 ppm Over specified range
9 Output voltage vs Output load < 50 ppm 100 % load changes
10 Output voltage vs Temperature 50 ppm/deg. Over specified range
11 Output voltage accuracy < 0.5 % Related to set-point value
(max offset: 250V)
12 Output voltage ripple 0.03 % rms Below 1KHz
13 Output voltage ripple 0.3 % rms Above 1KHz
14 Emission current 0..100 mA
15 Emission current vs Input voltage < 50 ppm Over specified range
16 Emission current vs Output load < 50 ppm Load change According to voltage
change from -20KV to -60KV
17 Emission current vs Temperature 50 ppm/deg. Over specified range
18 Emission current accuracy < 0.5 % Related to set-point value
(max offset 0.08 mA)
19 Filament power supply voltage 0..14 Vac Over voltage protected
to 14V
20 Filament power supply current 0..5 A short-circuit protected
21 Filament power supply current Output 1 % rms Below 100 Hz
voltage ripple
22 Filament power supply current Output not critical Above 100 Hz
voltage ripple
23 Operating temperature range 0..35deg.C Inlet Air Temperature
24 Cooling system Forced air
30 High voltage connector Type II, US The 3 terminals C, L, S connected
FEDERAL to the high voltage
STANDARD H72,
3 PINS
31 Main connection 6c. WAGO connector
32 X-ray ON valve connector IEC320 Supply: 230V15W
33 Maximum power change ramp 1KW/second Limited by interface
34 Maximum voltage change ramp 100KV/second Limited by interface
35 Maximum filament current change ramp 2A/second Limited by interface
36 X-ray arcs detection. shut off after 5 arcs in 10 seconds
37 Control and command connector SUB-D 50P Female 50x
38 Over temperature protection
GENXRD WORKSTATION
GENXRD WORKSTATION Chapter 11
11 GENXRD WORKSTATION
Description of the main functions
Figure 11.1
Interlocks:
All safety switches closed Interlock: closed
Spectrometer Tank Pressostat open Interlock: open
Interlock: Pressostat.
Electronic Rack Security Switch open Interlock: open
Interlock: Intlk2
Tank Cover or/and Interlock: open
X-Ray Tube Security Switch open Interlock: Intlk3
Load Lift Rack Security Switch open Interlock: open
Interlock: Intlk4
X-Ray Power Supply Rear Rack Security Switch open Interlock: open
Interlock: Intlk5
X-ray Secondary Water Over-temperature Sensor Switch open Interlock: open
(60° C) Interlock: IntlkRtn
WaterPump DeWaterFlow
GenXRF ON Enable
GenXRD OFF Disable
Water Pump DeWaterFlow
GenXRF OFF Disable
GENXRD ON Enable
PrWavalve PrWaFlow
GenXRF ON Enable
GenXRD OFF Disable
PrWavalve PrWaFlow
GenXRF OFF Disable
GENXRD ON Enable
WaterCooler WaterCoolerOn
GenXRF ON Enable
GenXRD ON Enable
CLUSTER 1
DISPLAY 1
DISPLAY 3
XMI32 XCM
XGB(I/M) ID: 2.63 ID: 2.2 XRF
ID: 3.1
generator
XPSCM XPSCS
XDI ID: 2.3.0 ID: 2.3.1
ID: 3.3
XHI
CLUSTER 4 (GONIO2) ID: 2.4
CLUSTER 5 (GONIO3)
XSI
CLUSTER 6 (XRD1) ID: 2.5
XCMC XDMC
ID: 6.4 ID: 6.5 XMI02 XSP
ID: 2.33 ID: 2.6 XRD
generator
DISPLAY 6
XMI01 XPSCMXRD XPSCSXRD
XGBX ID: 2.32 ID: 2.8.0 ID: 2.8.1
ID: 6.1
XSN
XDI
ID: 6.3
BASIC CLUSTER BASIC NODE
CLUSTER 7 (XRD2)
OPTIONAL CLUSTER OPTIONAL NODE
XGN
INTEL 386EX INTEL 80C196KR ANALOG ADMC401
Figure 11.2
In normal operation the GenXRF is the master and GenXRD is the slave. The master does control the start-
up sequence (Including PTB safety) and the X-Ray tubes cooling water parameters.
One or the other generator, or both generators, can be disabled by the Firmware status 31.
If GenXRF is disabled, GenXRD becomes automatically the master. In turn it has to control the start-up
sequence (Including PTB safety) and the X-Ray tubes cooling water parameters.
If GenXRD is the Master it has to control the X-Ray tube cooling parameters, with the primary water valve if
GenXRF 3.6 or 4.2 kW versions is installed and by the Water-air exchanger fans, in case of GenXRF 2.5 kW
version is installed.
The XRF generator is still controlled by the XSN node 2.3. The new XRD generator is controlled by the
XSN node 2.8.
The alarms range which was allocated to the XRF generator is now shared by the two generators according
to the following table.
The “Status channels” and the “Firmware status” of the new XRD generator are in the range 4096 to 4607.
The firmwares of the 2 generators must support the new IDLE mode. In this mode, the X-ray tube current
emission is not controlled anymore but the filament current is forced to a value defined by a parameter of a
new firmware status (see below).
A new firmware status “X-ray generators IDLE and ECO conditions” (31) allows to enable/disable the
generators. Moreover, it gives IDLE conditions for the XRF generator and the IDLE and ECO conditions for
the XRD generator. The ECO conditions of the XRF generator as well as the entry and exit times in ECO
mode are given by the XG command.
Two parameters have been added to the firmware status “General instrument activity” (8), to report the XRD
generator activity time and the XRD tube sparks.
Figure 11.3
ECO Mode
) Command:
XR <X-ray kV>
,<X-ray mA>
,<X-ray generator>
,<X-ray state>
,<X-ray anode>
,<true X-ray kV-mA>
X-ray mA INTEGER
0: 0 mA if “true X-ray kV-mA” = 1.
Same as last XR command if “true X-ray kV-mA” <> 1.
For valid mA value, see LEGAL KV-MA COMBINATIONS graph.
Figure 11.4
Events Commands
X1OFF: XR kV,mA,<=3,3
X2OFF: XR kV,mA,>=4,3
X1ON: XR kV,mA,<=3,1|2 PA 0,kV,mA,0
X2ON: XR kV,mA,>=4,1|2 PA 0,kV,mA,2048
X1IDLE XR kV,mA,<=3,4
X2IDLE XR kV,mA,>=4,4
) Command:
XC <function>
(,<param1>) Min. 20kV
(,<param2>) Max. 45 kV
(,<param3>) Min. 20 mA
(,<param4>) Max 45 mA
(,<param5>)
Param1 REAL
for function = 1
Desired low kV calibration point [kV].
0.0 : default low kV calibration point.
for function = 2
Desired kV OFFSET.
Param2 REAL
for function = 1
Desired high kV calibration point [kV].
0.0 : default high kV calibration point.
for function = 2
Desired kV SLOPE.
Param3 REAL
for function = 1
Desired low mA calibration point [mA].
0.0 : default low mA calibration point.
for function = 2
Desired mA OFFSET.
Param4 REAL
for function = 1
Desired high mA calibration point [mA].
0.0 : default high mA calibration point.
for function = 2
Desired mA SLOPE.
Param5 WORD
for all functions
X-ray generator selection
0: XRF generator
<>0: XRD generator
© Returned data:
ALARMS STATUS
,<current kV OFFSET>
,<current kV SLOPE>
,<current mA OFFSET>
,<current mA SLOPE>
2
The parameters updated by this command are also reachable through the firmware status “kV/mA
measurement calibration” (2048 for XRF generator (0) or 4096 for XRD generator (1)).
Figure 11.5
Generator selection
Existing parameters or new parameters of the commands XR, PA and XC allows selecting the
desired generator.
The following state transition diagram summarizes the behavior of XQIX to control both
generators.
start Cmd:X2OFF/
Cmd:X1OFF/ X2OFF
X2OFF; X1OFF
X1OFF
X2OFF
Cmd:X1OFF/
Cmd:X1OFF/
X2OFF; X1OFF
X2OFF; X1OFF
Cmd:X1OFF/
X2OFF; X1OFF Cmd:X2ON/ Cmd:X1ON/
X1IDLE; X2ON X2IDLE; X1ON
Cmd:X2IDLE/
X1IDLE; X2IDLE
Cmd:X1OFF/
X2OFF; X1OFF Cmd:X1OFF/
Cmd:X1IDLE/ X2OFF; X1OFF
X1IDLE Cmd:X1ON/
Cmd:X2OFF/ X1ON; X2IDLE
X2OFF
X1ACTIVE
X2IDLE
X1ACTIVE
X2OFF
Cmd:X2OFF/
X2OFF Cmd:X1IDLE/
X1IDLE Cmd:X1ON/
Cmd:X1ON/
X2IDLE; X1ON
X2IDLE; X1ON
Cmd:X2IDLE/
X1IDLE; X2IDLE Cmd:X1ON/
Cmd:X2ON/
Cmd:X1IDLE/ Cmd:X1OFF/ X1IDLE; X2ON X2IDLE; X1ON
X1IDLE Cmd:X2ON/ X2OFF; X1OFF
X1IDLE; X2ON
Cmd:X1ON/
X2IDLE; X1ON
Cmd:X1IDLE/
X1IDLE Cmd:X1IDLE/
Cmd:X2IDLE/ X1IDLE
X1IDLE; X2IDLE
Cmd:X2ON/
X1IDLE; X2ON
Cmd:X1ON/ Cmd:X2ON/
X2IDLE; X1ON X1IDLE X1IDLE; X2ON X1IDLE Cmd:X2IDLE/ X1IDLE
Cmd:X2OFF/ X2ACTIVE
X2IDLE X1IDLE; X2IDLE X2ECO
X2OFF Cmd:X2IDLE/
exit ECO/
X1IDLE; X2IDLE
X2IDLE
Cmd:X2ON/
X1IDLE; X2ON
Cmd:X2IDLE/
Cmd:X2OFF/ X1IDLE; X2IDLE
Cmd:X1IDLE/
X1IDLE X2OFF
X1IDLE
X2OFF enter ECO/
X2toECO
Cmd:X1IDLE/
X1IDLE
Cmd:X2IDLE/ enter ECO/
Cmd:X2ON/
X1IDLE; X2IDLE X1toECO
Cmd:X1IDLE/ X1IDLE; X2ON
X1IDLE
Cmd:X2OFF/ Cmd:X1ON/
X2OFF X2IDLE; X1ON
Cmd:X1IDLE/
X1IDLE
Cmd:X2ON/
Cmd:X2OFF/ X1IDLE; X2ON
X2OFF
Cmd:X2IDLE/
X1IDLE; X2IDLE Cmd:X1OFF/
enter ECO/ Cmd:X2ON/
X2OFF; X1OFF
X1toECO X1IDLE; X2ON
Cmd:X2IDLE/
X1IDLE; X2IDLE
Cmd:X1ON/
X1ON; X2IDLE
Figure 11.6
Mains XPSI
XrayPowerSupplyInput
Board
XPSCD
XQIX
XrayPowerSupplyControlXRD XPSB
Board Xray
Power
XPSP Supply
XrayPowerSupplyPowerXRD Backplane
Board Board
XPSMD
XrayPowerSupplyMultiplier
Board
Multiplier
Multiplier And
HighVoltage Box Filament
Assembly
Figure 11.7
Working principle
u
i FET2
u C1 +200Vrtm Multiplier L
C
C2
0Vrtm
XQIX
uP 1000 M Ohm
i
400Vrtm
+200Vrtm
PI PWM
0Vrtm
Filament -HV
Transformer W filament
-e X-Rays
Co anode
Figure 11.8
XPSI power components (Power-Mosfet) are cooled by a combined radiator-half pipes system.
XPSI does convert the 230VAC main power to 360 - 400 VDC.
XPSI does manage the inrush current, the working current limitation and the power factor correction.
XPSCD does manage the kV regulation, the emission current regulation of the X-Ray tube. It is the brain of
the generator.
XPSP power components (Power-IGBT) are cooled by a combined radiator-half pipes system.
The XPSP board converts the intermediate DC voltage (360 V- 400 V) into AC voltage sources, which
supplies current to the filament transformer and the HV transformer.
XPSMD board
Filament transformer
HV transformer
Multiplier box with voltage inversion
The XPSMD board has three main functions: the filament power supply, the HV measurement and the
emission current measurement.
The filament transformer via the XPSM board is used to supply up to 50 W, maximum 4 A to the X-ray tube
filament.
The HV transformer via the multiplier is used to supply up to 1800 W, maximum 50 kV, 60 mA, to the X-
ray tube anode.
The backplane or motherboard print is a bus supporting 5 slots made of 5 x 32 pins female internal
connectors (J902-J906) and 2 5 x 32 pins female external connectors J901 and J907, of Europe format.
Each board has its specific position in the rack. There are no keys to prevent the connection of a board to the
wrong slot.
Figure 11.9
The main functions controlled by the XPSI board are the mains (230VAC) power factor correction for the
current, the mains current limitation and the X-ray lamps control.
The mains voltage is used via a filter and a relay. When the generator switches on, the relay is energized to
close both contacts. After the rectifier diodes, the FET1 is used to open the circuit if the current is above a
current limit. When the FET1 is open, the current flow through the integrated diode D1of the FET circuit,
due to the inductor L. When the current is below a current limit the FET1 will drive.
FET1 L D2 +400Vrtm
L
C1
IL FET2 +12kV
230Vac U1 D1 U2 C2
N 0Vrtm
uC DSP
Figure 11.10
The FET2 and the integrated diode D2 of the FET circuit are used to step up the voltage to 400 V +/- 10 %
this voltage is adapted according to the kV output voltage. 360 V for 0 to 30 kV, 360 V to 420 V
proportionally from 30 kV to 60 kV).
Each 50 µs the current “IL”, the voltages “U1” and “U2” are evaluated in order to define the FET2 OFF/ON
conduction time. The maximum OFF/ON time is 25 us.
V U1
IL
UD1
U2
FET2
OFF
U1
IL
FET2
ON
25 us
Figure 11.11
During the OFF time the current flows through D2 to load the capacitors C1 and C2. The C1/C2 voltage is
measured by the DSP and compared to the set-point value. The On time is regulated relative to this
comparison.
− The IGBTs on the XPSP power board, generating the anode high voltage, via the transformers and the
multiplier.
− The IGBTs on the XPSP power board, generating the cathode (filament) heating current.
Proportional 400 V
U2 400V Monitoring
400 V
360 V
HV
30 kV 50 kV
Figure 11.12
The neutral potential does cross a resistor divider before to be applied to the neutral to ground comparator IC
6.
This device allows the DSP to detect a floating neutral or a neutral-phase inversion.
Figure 11.13
This power board contains the switching components (IGBTs) for the high voltage transformer and for the
filament power supplies.
Part of XPSP HV
Transformer
FET1 L IGBT1 C3
230Vac D +400V
HighVoltage
C1
u
i FET2
u +200Vrtm
Multiplier
C2 IGBT2
0Vrtm
XQIX
uC 1000 M Ohm
Part of XPSP
i
IGBT3
400Vrtm
PI PWM +200Vrtm
0Vrtm Filament
IGBT4 Transformer
Figure 11.14
The HV switching works at 20 kHz. The transformer primary is connected to the C1/C2 point which
corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V).
The control board commands alternately the IGBT1 and IGBT2. The On time duration depends of the output
power needed. By this way, the transformer primary receives, via C3, plus and minus 200 V referenced to
the fix 200 VRTM.
UC3-0Vrtm
+400V
+200V
0 t
IGBT1
Command
IGBT2
Command
Figure 11.15
UC3-0Vrtm
+400V
+200V
0 t
IGBT1
Command
IGBT2
Command
Figure 11.16
The transformer inductor and C3 capacitor make a resonant circuit. By this way the current look like
sinusoidal:
UC3-0Vrtm
+400V
+200V
0 t
Transformer
Primary
current
Figure 11.17
The filament power supply switching works at 40 kHz. The transformer primary is connected to the C1/C2
point which corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V).
The control board commands alternately the IGBT3 and IGBT4. The On time duration depends of the output
power needed (max. 50 W). By this way, the transformer primary receives positive and negative 200 V
referenced to the fix 200 VRTM.
See after the voltage and the current at the primary of the filament transformer:
Filament
transformer
+200V
-200V t
Filament
Transformer Primary
current
0 t
Figure 11.18
The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.
u
i FET2
u C1 +200Vrtm Multiplier
C2
0Vrtm
XQIX
uP 1000 M Ohm
i
400Vrtm
+200Vrtm
PI PWM
0Vrtm
Filament
Transformer
Figure 11.19
♦ the HV supply
♦ the emission current measurement
♦ the filament current supply
The 400V from the IGBTs on the XPSP board, via XPSM board goes to the HV transformer. At the
transformer secondary the voltage changes according to the IGBT1 and IBGT2 ON time.
The maximum voltage is near 12000 V. The full wave multiplier increases five times this voltage to reach at
maximum - 60000 V. Two dividers are used to monitor the high voltage. The command board uses one of
them to control the HV via the servo loop function accordance with the kV set point. The second is used to
control the first via a subtraction function.
The emission current is measured with the current absorbed by the multiplier. The emission current
measurement allows to indirectly regulating the filament heating current, thus to match with the desired
emission current.
The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.
Figure 11.20
HV
Transformer
230Vac FET1 L D +400Vrtm
HighVoltage
C1
u
i FET2
+200Vrtm
Multiplier
u
C2
0Vrtm
XQIX
uP 1000 M Ohm
XPSCD
400Vrtm
Part
+200Vrtm
PI PWM
0Vrtm Filament
Transformer
Figure 11.21
HV servo-loop control
The control board receives the asked HV from instrument master board via XQIX. The asked HV is
compared with the real HV to generate the HV error, which is used through the PID corrector to PWM duty-
cycle.
The PWM signals control the power stage on XPSP board to supply the right voltage according to the HV
error. The XPSM assembly transformer and multiplier increase this voltage to generate the real HV. The
DSP evaluates the values each 50 us.
uC DSP
HV
HV Multiplier
Power
set point + PID PWM Box
- Stage
HV
D HighVoltage 18762
A
Part of Part of
Part of XPSP board XPSMD assembly
XPSCD board
Figure 11.22
The rest of the bloc diagram works like a controlled current source. I.e., the filament current changes
according to the control voltage. 1 volt corresponds to 1 ampere. The set-point voltage is compared with the
current sensor voltage to control, via the proportional corrector, the PWM duty-cycle at 40 kHz. The power
board contains the power stage, which generates the power for the transformer. The transformer is used to
increase the current to the desired level (up to 4 A). Both the filament heating current and filament voltages
are monitored.
uC
Emission
current
set point
+
-
PID Filament current power supply
HV
Power Multiplier
Stage
D -1
A
R
Part of Part of
Part of XPSCD board XPSP board XPSMD assembly
Figure 11.23
The rectified input voltage is weighted by this voltage error to become the current reference for a second
servo-loop. The real current, to generate the current error, is compared to this reference. The GPFC corrector
uses the current reference, the real current and the current error values to control the PWM duty-cycle, which
controls the Power MOS FET (FET2).
The FET1 is controlled by three ways. The first one is the intermediate voltage via a P corrector. The PWM
duty-cycle changes according to intermediate voltage. The second one is made on the real current value and
a comparator with a current threshold. At switching power on, this current threshold corresponds to 2
amperes to limit the inrush current. When after 2 seconds, the intermediate voltage exceeds 320 V, the
current threshold changes to 40 Amperes. This corresponds to the maximum acceptable current during power
on state. The last way controls the FET1 by hardware. The real current is compared with a reference value.
When the real current exceeds 40 amperes the FET1 is switched off. This circuit allows reacting faster than
the DSP, where its reaction time is related to the sampling period. The low voltage power supply voltages
and the temperatures are check in “watchdog” mode. Periodically, in background task, the control board
accesses to the 8 channels 8 bit analogic to digital converter (ADC). A fatal alarm will be issued if the values
are not coherent or if it is not possible to communicate with the ADC.
Figure 11.24
Figure 11.25
PPI 1 82C55 IC 10
Port State Designation Remark
PPI1 PA 0 I /Int1 Pressostat Active if pressure < 25 KPA
PPI1 PA 1 I /Int2 Electronic rack safety switch
PPI1 PA 2 I /Int3 Tank cover + X-ray tube position switches
PPI1 PA 3 I /Int4 Magazine + load lift cover switches
PPI1 PA 4 I /Int5 X-ray generator rear switch
PPI1 PA 5 I /IntRet 60°C secondary water thermostat
PPI1 PA 6 I /24VPOK +24V instrument OK
PPI1 PA 7 I /WaCoOn Water cooler ON
Port Address: FFC0H
Figure 11.26
AD converters
First AD Miscellaneous 8×8 bits AD converter. To display the different AD input values, use the
terminal mode commands DY
See display information. AD display
commands.doc
Lamp1:
24V DC for GEN’X 1200 W
Lamp2:
24V DC for GEN’X 1200 W
HV Arc Threshold
55 Digits = 1080mV
XPSC temp:
Should not exceed 65°C
DA Converter
♦ The tap or primary water flow. An alarm is immediately issued if the flow drops. However, the default
will only shut down the generator after a delay of 20 seconds.
♦ The tap or primary water inlet and outlet temperatures.
♦ The deionised water flow.
♦ The deionised water inlet and outlet temperatures.
♦ The deionised water resistivity (560k**cm).
♦ The deionised water level
The change of state of cooling safety causes the immediate shut down of the generator by acting on the
generator interlock input. The parameter responsible for the shut down will generate an alarm reported to the
ACS software.
Deionised water
♦ Flow 2 l/min
♦ Pressure <3.5 bars
♦ Outlet temperature <60°C
♦ Resistivity >550 KΩ
22ms
+ DEIONIZED WATER RESI. DRIVER IN
-
+14V
0V
3.3V
tol 3<>4.5V
A/D
PORT 0
1M
DEIONIZED
WATER RESIST.
+ IN
resistivity gauge RX
-
Figure 11.27
NODE XPSCMXRD
XPSCMXRD and the sub-node XPSCSXRD, controls:
♦ The XRD high-voltage generator activity.
♦ The XRD X-ray tube filament current regulator.
♦ The XRD X-ray tube cooling circuit only if the XRF generator is absent.
♦ Switches and jumpers function on XPSCM.
8470 XPSC GenXRD F Secondary water resistivity driver fault. If > 5 V or < 2V
8471 XPSC GenXRD F Secondary water resistivity gauge fault. Gauge not connected
8472 XPSC GenXRD F Secondary water resistivity too low. If < 300 k Ώ / cm3
The concerned statuses are disabled in the database for GEXRD for WORKSTATION.
S: XC
S: XC command has to be executed. Corrupted data in the EEPROM of XPSC. Check firmware Status.
IMPORTANT:
Before to have access inside the GenX generator, you have to read carefully the following
caution notes in order to avoid any trouble occurring to the generator or any injury.
Warning
Before to work with the new GenX generator, please read carefully
the following information and caution.
Electrocution risk
DANGER: LETHAL VOLTAGES are present in the GenX generator
whenever the generator is connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX
generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after power
disconnection.
When increasing the high voltage, a noise near the multiplier can be heard, and an alarm “PWMTRIP”
Power Off, issued by DSP” is generated, the multiplier assembly can be suspected.
- Switch OFF the X-ray power supply generator by typing the terminal mode command: #XR
0,0,4,3
- Wait for 1 minute to allow the completion of the secondary deionised water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing
down the two main breakers 16A/32A
- Remove all the panels around the GenX generator.
- Remove the HV cable connector on the back side of the generator. Inspect the connector. No
spark over traces should be visible.
Make sure no wire or any other conductor is in the vicinity of the HV connector.
- Switch OFF the X-ray power supply generator by typing the terminal mode command:
- #XR 0,0,4,3
- Wait for 1 minute to allow the completion of the secondary deionized water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing
down the two main breakers 16A/32A
- Remove all the panels around the GenX generator.
- Remove the HV cable connector on the back side of the generator and from the X-Ray tube.
Inspect the connectors; on the cable, on the generator and on the X-Ray tube. No spark-over
traces should be visible.
- With a Meger check the HV cable insulation.
- The resistivity must be infinite.
- If it is not change the HV cable and the tube.
Note: The HV cable connector must be cleaned yearly. The connector silicone washers must be
replaced and lubricated yearly.
kV Regulation
kV > 0 No No test
Yes
Yes kV > 5 kV No
Yes
Yes No
Figure 11.29
mA Regulation
mA > 0 No No test
Yes
Yes mA > 10 mA No
Yes
Yes No
Figure 11.30
8513 XPSC GenXRD F Power Off issued by DSP; 400V too low (<320V)
Too fast X-Ray power supply transition. 400 V voltage drop. May happen when changing X-Ray tube
conditions in terminal mode, in “On Fast mode” and “Operational Mode”.
See Step 6
8514 XPSC GenXRD F Power Off issued by DSP; HV too high (>71kV)
KV Regulation error. KV Regulation loop fault.
8515 XPSC GenXRD F Power Off issued by DSP; Current too high (>40A)
After a power up retry the current can rise above 35 A.
8516 XPSC GenXRD F Mains voltage below 195 VAC (after 20 ms)
8517 XPSC GenXRD F Mains voltage above 253 VAC (after 20 ms)
Main power supply fault. The main voltage is checked by the DSP. If out of tolerance the DSP does
immediately shut down the power.
See Step 4
If > 75 ° C
Bad cooling. Check fans. Laboratory temperature too high.
Phase 2
XPSC
Figure 11.31
PTB-TÜV
SwitchOn XRAY SwitchOff XRAY
External Voltage =
Yes
Present ?
internalRelay = OFF
8385 F_EXTERNAL_RELAY_INCOHERENCY
No -10Vrtm and +15Vrtm = ON
externalRelay = ON Yes
8386 F_POWER_ON_RELAY_INCOHERENCY
No Internal Voltage 320V = ON
No
External Voltage =
No
Present ?
externalRelay = OFF
8387 F_NO_EXTERNAL_RELAY_FEEDBACK
-10Vrtm and +15Vrtm = Off
Yes 24 V ExRelayPresent = OFF
internalRelay = ON
External Voltage =
Present ?
Internal Voltage =
No
Present ?
8388 F_NO_POWER_ON_RELAY_FEEDBACK No
400Vrtm (350V) = OFF
Yes
internalRelay = OFF
Figure 11.32
Safety
Read carefully those below cautions:
During the instrument’s unloading from the wood palette and the installation’s work,
take care of the cables’s connectors and cables coming from the instrument, DON’T
DAMAGE THEM.
Before starting the installation’s work, be sure that 9900 Workstation instrument isn’t
connected to the Mains 230VAC.
Installation procedure
For the installation of the GenXRD chassis, please follow strictly the procedure below:
- XPSP board
- HV Multiplier Assembly
to
- the flat terminal connector located on
the base of the GenXRD chassis.
to
- the fast on terminal located inside the
cover of the GenXRD chassis.
IMPORTANT:
Tools:
Field Service Engineer standard tools
Documentation:
Set of ARL 9900 Series instruments schematics
Specifications
# DESCRIPTION VALUES REMARKS
1 Output power 0 to 1.8 KW
2 Input voltage 230VAC +10% / -15%
50/60Hz 1phase
3 Maximum input current and inrush current 28A at 195VAC
4 Maximum input power 5000W
5 Efficiency > 80%
6 Power factor > 0.9 switched mode load type
7 Output voltage -10 to -50KV Without overshoot.
Short-circuit protected.
8 Output voltage vs Input voltage < 50 ppm Over specified range
9 Output voltage vs Output load < 50 ppm 100 % load changes
10 Output voltage vs Temperature 50 ppm/deg. Over specified range
11 Output voltage accuracy < 0.5 % Related to set-point value
(max offset: 250V)
12 Output voltage ripple 0.03 % rms Below 1KHz
13 Output voltage ripple 0.3 % rms Above 1KHz
14 Emission current 0..60 mA
15 Emission current vs Input voltage < 50 ppm Over specified range
16 Emission current vs Output load < 50 ppm Load change According to voltage
change from -20KV to -60KV
17 Emission current vs Temperature 50 ppm/deg. Over specified range
18 Emission current accuracy < 0.5 % Related to set-point value
(max offset 0.08 mA)
19 Filament power supply voltage 0..14 Vac Over voltage protected
to 14V
20 Filament power supply current 0..4 A short-circuit protected
21 Filament power supply current Output 1 % rms Below 100 Hz
voltage ripple
22 Filament power supply current Output not critical Above 100 Hz
voltage ripple
23 Operating temperature range 0..35deg.C Inlet Air Temperature
24 Cooling system Forced air
30 High voltage connector Type II, US The 3 terminals C, L, S connected
FEDERAL to the high voltage
STANDARD H72,
3 PINS
31 Main connection 6c. WAGO connector
32 X-ray ON Lamps +24 VDC 24VDC 8W
33 Maximum power change ramp 3KW/min Limited by interface
34 Maximum voltage change ramp 100KV/second Limited by interface
35 Maximum filament current change ramp 2A/second Limited by interface
36 X-ray arcs detection. shut off after 3 arcs in 10 seconds
37 Control and command connector 96 pole cable
38 Over temperature protection
In Switzerland:
Thermo Fisher Scientific +41 (0)21 694 71 11
En Vallaire Ouest C, Case Postale [email protected]
CH-1024 Ecublens, Switzerland
© 2020 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher
Scientific and its subsidiaries unless otherwise specified.