Xr-Doc Aa83678-20-En Arl Genx (Service Manual) 0520

This document provides certification and training documentation for repairing and maintaining a GenX generator. Only those who have completed specialized training at a specific location are authorized to access the service manual. The document is assigned to an individual who has completed the training and obtained certification, making them responsible for any issues that arise if untrained people attempt repairs.

Uploaded by

MGRenJc
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (2 votes)
822 views549 pages

Xr-Doc Aa83678-20-En Arl Genx (Service Manual) 0520

This document provides certification and training documentation for repairing and maintaining a GenX generator. Only those who have completed specialized training at a specific location are authorized to access the service manual. The document is assigned to an individual who has completed the training and obtained certification, making them responsible for any issues that arise if untrained people attempt repairs.

Uploaded by

MGRenJc
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 549

This documentation can be obtained ONLY AFTER a specific training in Ecublens.

GenX generator Service Manual

This document AA83678 is dedicated personally to :

Family name : ………………………………..

First name : ………………………………..

Working for : ………………………………..

Country : ………………………………..

Remark : ………………………………..

The above mentioned person has followed the special training course for the GenX
generator and has successfully obtained the certificate allowing him to repair and to
maintain the GenX generator. Consequently the person is aware of the risk of the
electrocution.

For the Thermo Training Institute :

Date and place : Signature :

I confirm having followed the special training course for the GenX generator and having
successfully obtained the certificate allowing to repair and to maintain the GenX generator

Date and place : Signature :

This AA83678 document must be used ONLY by the person who has
followed the special training of GenX generator and that has
successfully obtained the certificate to work on it.

Thermo Fisher Scientific is not responsible for any trouble occurring to


the generator or any injury if untrained personal tries to maintain or
repair the GenX.

No AA83678-20
1 2
H
Hydrogen
He
Helium

3 4 5 6 7 8 9 10
Li
Lithium
Be
Beryllium
B
Boron
C
Carbon
N
Nitrogen
O
Oxygen
F
Fluorine
Ne
Neon

11 12 13 14 15 16 17 18
Na Mg
Sodium Magnesium
Al
Aluminum
Si
Silicon
P
Phosphorus
S
Sulfur
Cl
Chlorine
Ar
Argon

19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36
K
Potassium
Ca Sc
Calcium Scandium
Ti
Titanium
V
Vanadium
Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br
Chromium Manganese Iron Cobalt Nickel Copper Zinc Gallium Germanium Arsenic Selenium Bromine
Kr
Krypton

37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54
Rb Sr
Rubidium Strontium
Y
Yttrium
Zr Nb Mo Tc Ru Rh Pd Ag Cd
Zirconium Niobium Molybdenum Technetium Ruthenium Rhodium Palladium Silver Cadmium
In
Indium
Sn Sb Te
Tin Antimony Tellurium
I
Iodine
Xe
Xenon

55 56 57
* 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86
Cs Ba La
Cesium Barium Lanthanum
Hf
Hafnium
Ta
Tantalum
W
Wolfram
Re Os
Rhenium Osmium
Ir
Iridium
Pt
Platinum
Au Hg
Gold Mercury
Tl
Thallium
Pb
Lead
Bi
Bismuth
Po
Polonium
At Rn
Astatine Radon

87 88 89 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118
**
Fr
Francium
Ra Ac Rf Db Sg Bh Hs Mt Ds Rg Cn Nh
Radium Actinium Rutherfordium Dubnium Seaborgium Bohrium Hassium Meitnerium Darmstadtium Roentgenium Copernicium Nihonium
Fl
Flerovium
Mc Lv Ts Og
Moscovium Livermorium Tennessine Oganesson

58 59 60 61 62 63 64 65 66 67 68 69 70 71
* Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
Cerium Praseodymium Neodymium Promethium Samarium Europium Gadolinium Terbium Dysprosium Holmium Erbium Thulium Ytterbium Lutetium

90 91 92 93 94 95 96 97 98 99 100 101 102 103


**
Th Pa
Thorium Protactinium
U
Uranium
Np Pu Am Cm Bk Cf
Neptunium Plutonium Americium Curium Berkelium Californium
Es Fm Md No Lr
Einsteinium Fermium Mendelevium Nobelium Lawrencium

Thermo Scientific ARL GEN'X

Service Manual
AA83678 - 20 • May 2020
SERVICE MANUAL

ARL GEN'X

Report MANUAL

No AA83678-20

Language E

Date May 2020

Department Doc

Author MMa

AA83678-20
The information in this document is subject to change without
notice.
We assume no responsibility for any errors that may appear
in this document.

Die Angaben in diesem Dokument können ohne vorherige


Information geändert werden.
Wir übernehmen keine Verantwortung für eventuelle Mängel
in diesem Dokument.

Les informations contenues dans ce document sont sujettes


à changement sans préavis.
Nous n’assumons aucune responsabilité pour toutes erreurs
éventuelles contenues dans ce document.

Ci riserviamo il diritto di modificare le informazioni contenute


nel presente manuale senza preavviso e non ne assumiamo
nessuna responsabilità quanto ad eventuali errori che
potrebbero esservi accidentalmente introdotti.

Las informaciones contenidas en este documento pueden


ser modificadas sin previo aviso.
Nosotros declinamos toda responsabilidad sobre los
eventuales errores contenidos en este documento.

AA83678-20
PAGE DE MODIFICATION

ÄNDERUNGSANGABEN

REVISION CONTROL SHEET

Date Modification Description


Datum Änderung Beschreibung
Date Revision Description

2005
April 01 General updating
November 02 OM3865 - S112248-0 added
S112072 becomes 05
2006
April 03 General updating

2007
November 04 Chapters 1, 4, 10 updated.
Chapter 11 added
Schematic S010801 added in chapter 8.

2008
June 05 Chapters 2, 4, 10, 11 updated

2009
January 06 Chapters 9 and 11 updated
S112073 becomes 07
April 07 Chapter 9 – Catalogue updated
July 08 Chapters 1, 6, 7, 9, 10 updated (OM4112 – 4128)
S112072 and S112066 updated
October 09 Chapters 1 and 4 updated

2010
January 10 Chapters 2 and 4 updated
S112067 becomes 11
S112073 becomes 09
November 11 S112064 becomes 11
S112065 becomes 8
S112067 becomes 12 (version -9 added)
S112073 becomes 10 (version -7 added)
Chapter 4 updated

2011
April 12 Chapter 9 – Catalogue updated / KXXGENX130 – 140 added
July 13 S112247 becomes 04
S112248 becomes 02
S112067 becomes 13
S112073 becomes 11
August 14 Chapter 9 – Catalogue updated / KXXGENX150 added
Pages 4-20 and 10-25 updated (alarm 8282)
November 15 Chapter 9 – Catalogue updated / KXXGENX151 added

AA83678-20
PAGE DE MODIFICATION

ÄNDERUNGSANGABEN

REVISION CONTROL SHEET

Date Modification Description


Datum Änderung Beschreibung
Date Revision Description

2012

February 16 Chapter 9 – Catalogue updated / KXXGENX152 - 153 added

2014
August 17 Chapter 9 – Catalogue updated / KXXGENX199: Fuses added
for XPSI and XPSP boards

2018 18 Chapter 8 : drawings S112306 and S112307 added


July Chapter 9 : new PNs for XPSI and XPSP boards in chapters
KXXGENX40, 80, 90, 120 and 151

2020
May 19 SDR25503: chapter 9 updated with info about ARL 9900 from
serial number 9953964 including 9953165, 9953396,
9953835, 9953933, 9953941, 9953943.

20 SDR25593: current specifications updated in chapter 9.

AA83678-20
TABLE OF CONTENTS
Table of Contents

1 GENX X-RAY GENERATOR SYSTEM OVERVIEW ............................... 1-1


NON-GENX, HIGH POWER SYSTEM OVERVIEW ................................................. 1-1

NON-GENX, LOW POWER SYSTEM OVERVIEW ................................................. 1-2

GENX, LOW AND HIGH POWER SYSTEM OVERVIEW ........................................ 1-3

GENX INTERCONNECTION ................................................................................... 1-6

GENX SUB-ASSEMBLY PARTS ............................................................................. 1-8


XPSI assembly ..................................................................................................... 1-9
XPSR board ......................................................................................................... 1-9
XPSC board ......................................................................................................... 1-9
XPSP assembly ................................................................................................. 1-10
Multiplier assembly ............................................................................................. 1-10
XPSB backplane board ...................................................................................... 1-10

GENX SUB-ASSEMBLY FOR NEW INSTRUMENT .............................................. 1-11

GENX VERSION COMPATIBLE WITH XPS BOARD FOR RETROFIT ................. 1-12

2 GENX DESCRIPTION .............................................................................. 2-3


X-RAY POWER SUPPLY, INPUT BOARD (XPSI) ................................................... 2-4

DESCRIPTION ......................................................................................................... 2-5


The other functions controlled by this board are .................................................. 2-7
X-ray ON Lamps .............................................................................................. 2-7
Low voltages and temperatures ....................................................................... 2-8
The neutral to earth potential monitoring ......................................................... 2-8

X-RAY POWER SUPPLY, POWER BOARD (XPSP) (REDESIGNED) .................... 2-9

X-RAY POWER SUPPLY, MULTIPLIER ASSEMBLY (XPSM) .............................. 2-14

COMMAND BOARD, X-RAY POWER SUPPLY CONTROL (XPSC) ..................... 2-17


Functions of the µ C 80C196KR ........................................................................ 2-18
Functions of the 80c196kr µ Controller ports ..................................................... 2-19
Functions of the PPI ports .................................................................................. 2-21
AD converters .................................................................................................... 2-23
DA Converter ..................................................................................................... 2-24
X-Ray On Lamps 24 VDC 1200 W Generator .................................................... 2-24

AA83678 Service Manual ARL GenX I


Table of Contents

Water control circuit ........................................................................................... 2-25


Water parameter requests ................................................................................. 2-25
Deionised water control circuit ....................................................................... 2-26
KV, mA Monitoring Calibration ........................................................................... 2-26
HV servo-loop control ......................................................................................... 2-27
Filament current servo-loop control .................................................................... 2-27
Emission current servo-loop control ................................................................... 2-28
Power Factor Correction control......................................................................... 2-28
Switches and jumpers .................................................................................... 2-30

DISPLAY INFORMATION TERMINAL COMMAND ............................................... 2-31

STATUS CHANNELS ............................................................................................. 2-33

2048 - 2559: XPS FIRMWARE STATUS ............................................................... 2-34

ALARMS LIST FOR FIRMWARE VERSION 3.2.8 AND LOWER .......................... 2-35

ALARMS LIST FOR FIRMWARE VERSION XPSCSXRD 3.0.0 ............................ 2-44

3 INSTALLATION AND REMOVING PROCEDURE .................................. 3-1


GENERAL CONTENT .............................................................................................. 3-1

INTRODUCTION ...................................................................................................... 3-1

HOW TO REMOVE A GENX GENERATOR FROM THE XRF INSTRUMENT ........ 3-2
Removal procedure .............................................................................................. 3-2

HOW TO INSTALL A GENX GENERATOR INTO THE XRF INSTRUMENT ........... 3-4
Installation procedure ........................................................................................... 3-4

HOW TO REMOVE AND INSTALL BOARDS AND MULTIPLIER ASSEMBLY


IN THE GENX GENERATOR ................................................................................... 3-6
Introduction .......................................................................................................... 3-6
Removal procedure ............................................................................................ 3-11
Installation procedure ......................................................................................... 3-13
Installation of the high voltage cable .................................................................. 3-15

4 TROUBLESHOOTING ............................................................................. 4-1


GEN’X DIAGNOSTIC FLOW CHART ...................................................................... 4-1

DETAILED DIAGNOSTIC FLOW CHART ................................................................ 4-4

ALARMS LIST ........................................................................................................ 4-13

II Service Manual ARL GenX AA83678


Table of Contents

Fatal alarms ....................................................................................................... 4-13


How to diagnostic insulation problems ............................................................... 4-16
HV MULTIPLIERS AND GENERATOR HV CONNECTOR INSULATION
problems ............................................................................................................ 4-16
X-Ray tube insulation problems ......................................................................... 4-17
HV cable insulation problems ............................................................................. 4-17
Warning alarms .................................................................................................. 4-29

5 FREQUENTLY ASKED QUESTIONS ..................................................... 5-1

6 GENX 600 W ............................................................................................ 6-1


600 W XPSI BOARD ................................................................................................ 6-1

600 W XPSP BOARD ............................................................................................... 6-2


X-ray On Lamps 24 VDC 600 W Generator ......................................................... 6-2
XPSC 24 VDC X-ray On Lamps control circuit ..................................................... 6-2

7 GENX RETROFIT .................................................................................... 7-1


GENX, RETROFIT SYSTEM OVERVIEW ............................................................... 7-1
Previous parts ................................................................................................................... 7-1
New parts ....................................................................................................................... 7-1

INTERCONNECTIONS ............................................................................................ 7-3


Interconnection bloc diagram ............................................................................... 7-3

XPSR BOARD .......................................................................................................... 7-4


XPSR bloc diagram .............................................................................................. 7-4
XPSR IS DIVIDED IN THREE MAIN PARTS ....................................................... 7-4

PART 1 LOW VOLTAGE POWER SUPPLIES..................................................... 7-5


Low voltage power supply circuit.......................................................................... 7-5
+ 24 VDC Safety circuit ........................................................................................ 7-7
Jumper W2 ........................................................................................................... 7-7

PART 2 XPS GENERATOR EMULATION ........................................................... 7-8


Serial link ............................................................................................................. 7-8
Jumper settings .................................................................................................... 7-9

PART 3 DISPLAY .............................................................................................. 7-10

GENX RETROFIT TROUBLE SHOOTING PROCEDURE..................................... 7-11


KV regulation error ............................................................................................. 7-11

AA83678 Service Manual ARL GenX III


Table of Contents

MA regulation error ............................................................................................ 7-11


Over voltage safety ............................................................................................ 7-11
Over emission current safety.............................................................................. 7-11
Over Load .......................................................................................................... 7-12
Filament current limit .......................................................................................... 7-12
Open circuit on filament ..................................................................................... 7-12
X-ray lamps defect ............................................................................................. 7-12
Low KV limit ....................................................................................................... 7-12
Leakage current limit .......................................................................................... 7-13
HV Inverter error (HVFlt) .................................................................................... 7-13
Fil Inverter error (FilFlt) ...................................................................................... 7-14
Arcs detection .................................................................................................... 7-14
Interlock closed .................................................................................................. 7-14
Interlock opened................................................................................................. 7-15
Power supply miscellaneous faults .................................................................... 7-15

GENX INSTALLATION PROCEDURE ................................................................... 7-17


Installation procedures for boards installed in the GenX generator .................... 7-17

9 CATALOGUE........................................................................................... 9-1
GENX I GENERATOR 3.6KW 60KV/120MA GENX I GENERATEUR 3.6KW
60KV/120MA ............................................................................................................ 9-5

GENX II GENERATOR 3.6KW 70KV/120MA GENX II GENERATEUR 3.6KW


70KV/120MA .......................................................................................................... 9-10

GENX III GENERATOR 4.2KW 60KV/120MA GENX III GENERATEUR 4.2KW


60KV/120MA .......................................................................................................... 9-15

GENX IV GENERATOR 4.2KW 70KV/140MA GENX IV GENERATEUR 4.2KW


70KV/140MA .......................................................................................................... 9-20

GENX V GENERATOR 1.2KW 60KV/60MA GENX V GENERATEUR 1.2KW


60KV/60MA ............................................................................................................ 9-25

GENX RET GENERATOR 3.6KW 70KV/120MA GENX RET GENERATEUR


3.6KW 70KV/120MA .............................................................................................. 9-29

GENX RET GENERATOR 4.2KW 70KV/140MA GENX RET GENERATEUR


4.2KW 70KV/140MA .............................................................................................. 9-34

GENX VI GENERATOR 2.5KW 60KV/100MA GENX VI GENERATEUR 2.5KW


60KV/100MA .......................................................................................................... 9-39

IV Service Manual ARL GenX AA83678


Table of Contents

GENXRD GENERATOR GENXRD GENERATEUR .............................................. 9-44

GENXRD RET GENERATOR GENXRD RET GENERATEUR .............................. 9-48

GENXRDF18 GENERATOR GENXRDF18 GENERATEUR .................................. 9-52

GENXRDV18 GENERATOR GENXRDV18 GENERATEUR.................................. 9-57

GENX VII GENERATOR 600 W 60KV/30MA GENX VII GENERATEUR 600 W


60KV/30MA ............................................................................................................ 9-62

GENX VIII GENERATOR PERFORM’X 4.2KW 70KV/140MA GENX VIII


GENERATEUR PERFORM’X 4.2KW 70KV/140MA ............................................... 9-66

GENX IX GENERATOR PERFORM’X 2.5KW 60KV/100MA GENX IX


GENERATEUR PERFORM’X 2.5KW 60KV/100MA ............................................... 9-71

GENX X GENERATOR PERFORM’X 4.2KW 60KV/120MA GENX X


GENERATEUR PERFORM’X 4.2KW 60KV/120MA ............................................... 9-76

GENX XI GENERATOR PERFORM’X 1.5KW 60KV/70MA GENX XI


GENERATEUR PERFORM’X 1.5KW 60KV/70MA................................................. 9-81

GENX XII GENERATOR PERFORM’X 2.5KW 60KV/100MA FOR X-RAY


TUBE UCCO30 GENX XII GENERATEUR PERFORM’X 2.5KW 60KV/100MA
POUR TUBE RAYON-X UCCO30.......................................................................... 9-85

GENX XIII GENERATOR PERFORM’X 4.0KW 60KV/140MA FOR X-RAY


TUBE UCCO30 GENX XIII GENERATEUR PERFORM’X 4.0KW 60KV/140MA
POUR TUBE RAYON-X UCCO30.......................................................................... 9-90

GENX X GENERATOR ARL 9900 VERSION 2020 1.5KW 60KV/100MA GENX


X GENERATEUR ARL 9900 VERSION 2020 1.5KW 60KV/100MA ...................... 9-95

GENX X GENERATOR ARL 9900 VERSION 2020 2.5KW 60KV/140MA GENX


X GENERATEUR ARL 9900 VERSION 2020 2.5KW 60KV/140MA .................... 9-100

GENX X GENERATOR ARL 9900 VERSION 2020 4.2KW 60KV/120MA GENX


X GENERATEUR ARL 9900 VERSION 2020 4.2KW 60KV/120MA .................... 9-105

KXXGENX199 TOOLKIT AND MISCELLANEOUS OUTILLAGE ET


DIVERS WERKZEUGE UND DIVERSES ....................................................... 9-110

10 GENXRD X'TRA ................................................................................. 10-1


DESCRIPTION OF THE MAIN FUNCTIONS ......................................................... 10-1
Working principle ........................................................................................... 10-2
XPSI Air (X-Ray Power Supply Input) ............................................................ 10-2
XPSRD (X-Ray Power Supply Retrofit XRD) ................................................. 10-2
XPSCD (X-Ray Power Supply Control XRD) ................................................. 10-2
XPSP (X-Ray Power Supply Power) .............................................................. 10-2
XPSMD .......................................................................................................... 10-3

AA83678 Service Manual ARL GenX V


Table of Contents

XPSB (X-Ray Power Supply Backplane) ........................................................... 10-3

XPSI ....................................................................................................................... 10-4


The other functions controlled by this board are ................................................ 10-6
Low voltages and temperatures ..................................................................... 10-6
The neutral to earth potential monitoring ....................................................... 10-6

XPSP ...................................................................................................................... 10-7

XPSMD ................................................................................................................ 10-10

XPSCD ................................................................................................................. 10-11


HV servo-loop control ....................................................................................... 10-12
Filament current servo-loop control .................................................................. 10-12
Emission current servo-loop control ................................................................. 10-12
Power Factor Correction control....................................................................... 10-13
Switches and jumpers ...................................................................................... 10-13

XPSRD ................................................................................................................. 10-15


XPSRD bloc diagram ....................................................................................... 10-16
XPSR IS DIVIDED IN TWO MAIN PARTS ....................................................... 10-16
Part 1 Low voltage power supplies ........................................................... 10-16
Low voltage power supply circuit ................................................................. 10-17
+ 24 VDC Safety circuit .................................................................................... 10-18

PART 2 XPS GENERATOR EMULATION ....................................................... 10-19


Serial link ......................................................................................................... 10-20
Jumper settings ................................................................................................ 10-20

DISPLAY .............................................................................................................. 10-21

GENXRD TROUBLE SHOOTING PROCEDURE ................................................ 10-22


Alarms list and description ............................................................................... 10-22
Master Processor Settings ............................................................................... 10-31
Power limit setting ........................................................................................ 10-31
Filament current limit .................................................................................... 10-31
Specifications ................................................................................................... 10-32

11 GENXRD WORKSTATION ................................................................ 11-1


DESCRIPTION OF THE MAIN FUNCTIONS ......................................................... 11-1
The interconnection board GenXRD-GenXRF allows controlling 2
generators by 4 distinct control categories ..................................................... 11-2
One generator configured .................................................................................. 11-5
Two generators configured ................................................................................ 11-5

VI Service Manual ARL GenX AA83678


Table of Contents

ECO Mode ......................................................................................................... 11-6


New X-R command ............................................................................................ 11-7
XC: X-RAY generator calibration........................................................................ 11-8
Generator selection .......................................................................................... 11-10
Working principle.............................................................................................. 11-12
XPSI (X-Ray Power Supply Input) Water cooled ......................................... 11-12
XPSCD (X-Ray Power Supply Control XRD) ............................................... 11-12
XPSP (X-Ray Power Supply Power) Water cooled ...................................... 11-12
XPSMD (X-Ray Power Supply Multiplier XRD) Water cooled ...................... 11-12
XPSB (X-Ray Power Supply Backplane) ......................................................... 11-13

XPSI ..................................................................................................................... 11-14


The other functions controlled by this board are .............................................. 11-16
Low voltages and temperatures ................................................................... 11-16
The neutral to earth potential monitoring ..................................................... 11-16

XPSP .................................................................................................................... 11-17

XPSMD ................................................................................................................ 11-21

XPSCD ................................................................................................................. 11-22


Functions of the µ C 80C196KR ...................................................................... 11-25
Functions of the 80c196kr µ Controller ports ................................................... 11-26
Functions of the PPI ports ................................................................................ 11-28
AD converters .................................................................................................. 11-30
DA Converter ................................................................................................... 11-31
X-Ray On Lamps 24 VDC 1200 W Generator .................................................. 11-31
Water control circuit ......................................................................................... 11-32
Water parameter requests ............................................................................... 11-32
Deionised water control circuit ..................................................................... 11-33
KV, mA Monitoring Calibration ......................................................................... 11-33
Switches and jumpers ...................................................................................... 11-34
NODE XPSCMXRD ..................................................................................... 11-34

WORKSTATION GENXRD TROUBLE SHOOTING PROCEDURE ..................... 11-35


GENXRD WORKSTATION ALARMS AND DIAGNOSTIC ............................... 11-35
HV Multipliers and generator HV connector Insulation problems ..................... 11-37
X-Ray tube insulation problems ....................................................................... 11-38
HV cable or tube insulation problems ............................................................... 11-38

INSTALLATION PROCEDURE OF GENXRD CHASSIS FOR ARL 9900


SERIES WORKSTATION..................................................................................... 11-48
Introduction ...................................................................................................... 11-48

AA83678 Service Manual ARL GenX VII


Table of Contents

Safety ............................................................................................................... 11-48


Installation procedure ....................................................................................... 11-49
Material list, tools and documentation .............................................................. 11-53
Special packaging for returned material ........................................................... 11-53

SPECIFICATIONS ............................................................................................... 11-54

VIII Service Manual ARL GenX AA83678


1

GENX X-RAY GENERATOR


SYSTEM OVERVIEW
GenX X-Ray Generator System Overview Chapter 1

1 GENX X-RAY GENERATOR SYSTEM OVERVIEW


Non-GenX, high power system overview
The non-GenX high power system uses 2 parts. An interface board called “XPS board” and a 3.6/4.2 kW
19’’ rack.

Generator HV
KVMonFb
Divider
Master control
HV HV HV HV
KVRef Corrector Inverter Multiplier

Power
Monitoring
PowerRef C
Water-cooling control Power
Interface I_FilRef Control
A
Board C
Xray Tube
K
MaRef Emission
Current Filament
Corrector Inverter
Safety interlock MaMonFb
Emission
Safety Current
Interlock Reading

XrayOn
R
Control

LVPS

Figure 1.1

AA83678 Service Manual ARL GenX 1-1


Chapter 1 GenX X-Ray Generator System Overview

Non-GenX, low power system overview


The non-GenX low power system uses 2 parts. An interface board called “XPSLP board” and a 600 W 19’’
rack.

Generator
KV Monitor HV
Divider

HV HV HV
Voltage Reference Corrector Inverter Multiplier

Power
M onitoring Power A

C Control X-Ray
Power Reference
Tube
Anode mA Monitor
Filament

XPSLP Board
Current
Emission Filament Sensor
Filament Current Reference
Current Driver
mA Reference
Corrector

C
Safety Filament Current

Cathode mA Monitor

Figure 1.2

1-2 Service Manual ARL GenX AA83678


GenX X-Ray Generator System Overview Chapter 1

GenX, low and high power system overview


List of GenX X-ray power Supplies that are available:

GenX I 3.6kW 60kV/120mA


GenX II 3.6kW 70kV/120mA
GenX III 4.2kW 60kV/120mA
GenX IV 4.2kW 70kV/140mA
GenX V 1.2kW 60kV/60mA
GenX R 3.6kW 70kV/120mA
GenX R 4.2kW 70kV/140mA
GenX VI 2.5kW 60kV/100mA
GenX VII 600W 60kV/30mA
GenXRD 3.6 kW 45kV/44mA
GenXRD RET 3.6 kW 45kV/44mA
GenXRDF18, 1.8 kW
GenXRDV18, 1.8 kW
The GenX generator does supply both the anode voltage potential and the cathode heating current to the X-
Ray tube.

An extended deionised water circuit does sink the rack temperature.

The interface board called XPSC replacing XPS (or XPSLP) is integrated into the 19” GenX rack. XPS (or
XPSLP) are absent from the instrument Electronic Rack.

Generator

HighVoltage HV
Divider
Master control
HvPwm Control HV HV HV
Inverter Multiplier

EmissionCurrent+
Cooling control
Control
A
Board
Xray Tube
K
FiamentPwm Control Filament
Safety interlock
Inverter

EmissionCurrent-

Instrument LVPS R

Figure 1.3

AA83678 Service Manual ARL GenX 1-3


Chapter 1 GenX X-Ray Generator System Overview

Figure 1.4

Figure 1.5 Top view

1-4 Service Manual ARL GenX AA83678


GenX X-Ray Generator System Overview Chapter 1

Figure 1.6 Back view

Figure 1.7 Front view

AA83678 Service Manual ARL GenX 1-5


Chapter 1 GenX X-Ray Generator System Overview

GenX interconnection
Interconnection diagrams:

S112163 Actual ADVANT’X +, XP, XP+ interconnection schematic (with j103 connector)
S112157 Future ADVANT’X +, XP, XP+ interconnection schematic (without j103 connector)
S112162 Actual 9800 XP, XP+ interconnection schematic (with j103 connector)
S112156 Future 9800 XP, XP+ interconnection schematic (without j103 connector)
S112287 9900 LP 1200W, GenX Low Power Interconnection
S112156 9900 XP/XP+, GenX XP/XP+ Interconnection
S010801 GenXRD - GenXRF

The interface with the instrument is made with a 96 wires flat cable (J104). This connection corresponds to
the upper instrument electronic rack backplane bus (J100 to J109). By this way the GenX receives the low
power supply voltages (+5V, +24V, +12V and –12V), the XSN communication signals, the display signal,
pressostat interlock and the master to slave control signals (XpsReset and XpsReady).

The “water cooling system” and the “X-ray safety interlock controls” are connected via M8 connector to
XPSC.

The mains voltage (230 VAC) is connected XPSI board via J2 connector.

The J3 on the XPSI board is used to connect the 230 V “X-ray lamp ON”. 24 VDC Lamps are connected to
J407 of XPSC.

The X-ray tube filament is connected to J1 of XPSM board.

For the X-ray tube high voltage connection, a “federal standard” connector is plugged in the J402 of the
XPSM assembly.

1-6 Service Manual ARL GenX AA83678


GenX X-Ray Generator System Overview Chapter 1

Input power connector230 VAC. Data connector 2x12 with X-ray tube Filament connector
WAGO 6x pins. J2 red key lock. M8 WAGO 5x pins. J1

Instrument ground cable Data connector 32x3 Water fast fitting connector
with border key lock. J104 male

Water fast fitting connector Connector X-ray On Lamps X-ray tube HV connector.
male for version 24VDC JP1
WAGO 7 pins. J407

Connector X-ray On Lamps


for version 230VAC. J3

Figure 1.8

AA83678 Service Manual ARL GenX 1-7


Chapter 1 GenX X-Ray Generator System Overview

GenX sub-assembly parts


The GenX is divided in 5 parts:

XPSI assembly X-Ray Power Supply Input


XPSR board X-Ray Power Supply Retrofit
XPSC board X-Ray Power Supply Control
XPSP assembly X-Ray Power Supply Power
XPSM assembly X-Ray Power Supply Multiplier

XPSB board X-Ray Power Supply Backplane

The mains XPSI


XrayPowerSupplyInput
Board

XPSR
Retrofit XrayPowerSupplyRetrofit
Control Board

Safety XPSC
Control XrayPowerSupplyControl
Communication Board XPSB
LVPS Xray
Power
XPSP Supply
XrayPowerSupplyPower Backplane
Board Board

Filament XPSM
XrayPowerSupplyMultiplier
Board

Multiplier
Multiplier And
HighVoltage Box Filament
Assembly

Figure 1.9

1-8 Service Manual ARL GenX AA83678


GenX X-Ray Generator System Overview Chapter 1

XPSI Board XPSC Board XPSP Board Multiplier Assembly

Figure 1.10

XPSI assembly
The XPSI assembly is made with the XPSI board and the cooling system for the power components. The
XPSI board converts the 230 VAC from the mains voltage into a DC voltage from 360 V to 440 V. Pay
attention to that the DC voltage is referenced to the mains voltage.

XPSR board
This board will be used to emulate the non-GenX system for field retrofit capability. With this board
installed, the instrument uses the XPS interface.

Remark: this board is not yet available. It is still in development.

XPSC board
This board is the interface with the XQM master board in the instrument. It controls all functions in the
GenX like the different servo-loops for the kV, mA and filament PS and few watchdog safety systems. It
controls too the different external parts like the safety interlock and the water cooler system.

AA83678 Service Manual ARL GenX 1-9


Chapter 1 GenX X-Ray Generator System Overview

XPSP assembly
The XPSP assembly is made with the XPSP board and the cooling system for the power component. The
XPSP board converts the intermediate DC voltage (360 V- 440 V) into AC voltage sources, which supplies
current to the filament transformer and the HV transformer.

Pay attention to that the DC voltage is referenced to the mains.

Multiplier assembly
The multiplier assembly is made with the following parts:

XPSM board
Filament transformer
HV transformer
Multiplier box
Cooling system

The XPSM board has three main functions: the filament power supply (rectifier/voltage and current
measurement/transformer and X-ray tube connections), the HV measurement and the emission current
measurement.

The filament transformer via the XPSM board is used to supply up to 120W to the X-ray tube filament.

The HV transformer via the multiplier is used to supply up to 4200W to the X-ray tube anode.

XPSB backplane board


The backplane or motherboard print is a bus supporting 5 slots made of 5 x 32 pins female internal
connectors (J902-J906) and 2 5 x 32 pins female external connectors J901 and J907, of Europe format.

Each board has its specific position in the rack. There are no keys to prevent the connection of a board to the
wrong slot.

1-10 Service Manual ARL GenX AA83678


GenX X-Ray Generator System Overview Chapter 1

GenX sub-assembly for new instrument


The configuration for new instruments (ADVANT’X XP and XP+) does not use the XPSR board. The LVPS
are supplied via the instrument electronic rack and flat cable from the instrument LVPS. In this case the XPS
board in the instrument rack is not used.

T he m ains X PSI
X ra yP o w erS up p lyInp u t
B o ard

S afety X PSC
C o ntro l X ra yP o w erS up p lyC o ntro l
C o m m u nica tio n B o ard X PSB
LVPS X ra y
P o w er
X PSP S up p ly
X ra yP o w erS up p lyP o w er B ackp lane
B o ard B o ard

F ila m e nt X PSM
X ra yP o w erS up p lyM ultip lier
B o ard

M ultip lier
M ultip lier A nd
H ig hV o lta ge B ox F ila m e nt
A sse m b ly

Figure 1.11

AA83678 Service Manual ARL GenX 1-11


Chapter 1 GenX X-Ray Generator System Overview

GenX version compatible with XPS board for retrofit


(Still in development, not yet available)

The retrofit configuration uses one more board in the GenX called XPSR. The XPSR board emulates the
non-GenX generator.

This version is compatible with the current non-GenX generators in the field.

The mains XPSI


XrayPowerSupplyInput
Board

XPSR
Retrofit XrayPowerSupplyRetrofit
Control Board

XPSC
XrayPowerSupplyControl
Board XPSB
Xray
Power
XPSP Supply
XrayPowerSupplyPower Backplane
Board Board

Filament XPSM
XrayPowerSupplyMultiplier
Board

Multiplier
Multiplier And
HighVoltage Box Filament
Assembly

Figure 1.12

1-12 Service Manual ARL GenX AA83678


2

GENX DESCRIPTION
GenX Description Chapter 2

2 GENX DESCRIPTION
Two processors are used to control all functions. One is a micro-controller (µC) and the other is a “digital
signal processing” (DSP) processor.

The µC does dialog with the instrument master board XQM of the electronic rack and controls the DSP, the
filament power supply and some background safety checks.

Bloc diagram

L D 400Vrtm
L
HighVoltage
i Multiplier
230Vac u u 200Vrtm

N 0Vrtm

DSP 1000 M Ohm

Instrument
Master
Board
uC 1000 M Ohm
XQM

i
400Vrtm

+200Vrtm
PI PWM
0Vrtm

Figure 2.1

AA83678 Service Manual ARL GenX 2-3


Chapter 2 GenX Description

X-ray Power Supply, Input board (XPSI)

Figure 2.2

2-4 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Description
The main functions controlled by the XPSI board are the mains (230VAC) power factor correction for the
current, the mains current limitation and the X-ray lamps control.

The mains voltage is used via a filter and a relay. When the generator switches on, the relay is energized to
close both contacts. After the rectifier diodes, the FET1 is used to open the circuit if the current is above a
current limit. When the FET1 is open, the current flow through the integrated diode D1of the FET circuit,
due to the inductor L. When the current is below a current limit the FET1 will drive.

The current is controlled by two devices:

1. The DSP, that does evaluate the current every 50 microseconds


2. A hardware fast comparator referenced to 5 V. 0.2 volts/A. In both cases if the current is excessive the
FET1 is set OFF.

FET1 L D2 +400Vrtm
L
C1
IL FET2 +12kV
230Vac U1 D1 U2 C2

N 0Vrtm

uC DSP

Figure 2.3

The FET2 and the integrated diode D2 of the FET circuit are used to step up the voltage to 400 V +/- 10 %
this voltage is adapted according to the GenX output voltage. 360 V for 0 to 30 kV. 360 V to 440 V
proportionally from 30 kV to 70 kV). Each 50 µs the current “IL”, the voltages “U1” and “U2” are evaluated
in order to define the FET2 OFF/ON conduction time. The maximum OFF/ON time is 25 us.

The current must be relative to the input voltage

AA83678 Service Manual ARL GenX 2-5


Chapter 2 GenX Description

V U1

IL

UD1
U2
FET2
OFF
U1

IL

FET2
ON

25 us

Figure 2.4

During the OFF time the current flows through D2 to load the capacitors C1 and C2. The C1/C2 voltage is
measured by the DSP and compared to the set-point value. The On time is regulated relative to this
comparison.

The proportional 400V working voltage will be used to supply:

 The IGBTs on the XPSP power board, generating the anode high voltage, via the transformers and the
multiplier.
 The IGBTs on the XPSP power board, generating the cathode (filament) heating current.

If the current is excessive, the driver IC 4 is shut off:

 By the DSP with the CurrentLimit signal dropping to 0 V,


 By the hardware comparator made of IC 8B and transistor Q2; If the current sampling voltage, 0.2 V/A,
does exceed the V Ref, Q2 is locked.

2-6 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Proportional 400 V

U2 400V Monitoring
440 V

360 V

HV
30 kV 70 kV
Figure 2.5

The other functions controlled by this board are


The 230VAC X-ray ON lamps current monitoring,
The low voltage power supply voltages monitoring,
The power component temperatures monitoring,
The neutral to earth potential monitoring.
X-ray ON Lamps
XPSC does monitor the X-ray ON lamps current.

CASE ADC LIMIT ALARM N° ALARM DESIGNATION SEVERITY


Both lamps O.k. 70..199
Lamps shorted 200..255 8377 X-Ray Lamps shorted Fatal
One lamp defect 20..69 8864 X-ray warning lamp broken Warning
Both lamps defect 0..19 8219 X-ray ON lamps broken Fatal

ADC value can displayed using following terminal mode command:

Terminal Mode Cmd Designation Remark and typical value


DY 2,3,11,3,4 XPSI, LampCurrent XPSI AD CHAN4:100 (ON case)
XPSI AD CHAN4:000 (OFF case)

AA83678 Service Manual ARL GenX 2-7


Chapter 2 GenX Description

Low voltages and temperatures


The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.

Use the following terminal mode commands to display these values:

Terminal Mode Cmd Designation Remark and typical value


DY 2,3,11,3,0 XPSI, +5VM XPSI AD CHAN0: 204
DY 2,3,11,3,1 XPSI, +12V XPSI AD CHAN1: 204
DY 2,3,11,3,2 XPSI, +24V XPSI AD CHAN2: 204
DY 2,3,11,3,3 XPSI, -12V XPSI AD CHAN3: 051
DY 2,3,11,3,4 XPSI, LampCurrent XPSI AD CHAN4: 100 (ON case)
DY 2,3,11,3,6 XPSI, InputTemp1 XPSI AD CHAN6: 81(at 25°C)
DY 2,3,11,3,7 XPSI, InputTemp2 XPSI AD CHAN7: 81 (at 25°C)

The neutral to earth potential monitoring


The neutral potential does cross a resistor divider before to be applied to the neutral to ground comparator
made of IC 6.

This device allows the DSP to detect a floating neutral or a neutral-phase inversion.

To monitor the difference between both potentials use the following terminal mode command:

DY 2,3,11,3,5 XPSI, NtoGndVoltage XPSI AD CHAN5: 001

The value expressed in Digits should approach 0.

Remark : The two power MOS-FETs are cooled:


By water, if high power application (2.5 kW, 3.6 kW, 4.2 kW)
By air, for low power application (1200 W).

2-8 Service Manual ARL GenX AA83678


GenX Description Chapter 2

X-ray Power Supply, Power board (XPSP) (Redesigned)

Figure 2.6

This power board contains the switching components (IGBTs) for the high voltage transformer and for the
filament power supplies.

Part of XPSP HV
Transformer
FET1 L IGBT1 C3
230Vac D +400V
HighVoltage
i C1
u
FET2
u +200Vrtm
Multiplier
C2 IGBT2
0Vrtm

DSP 1000 M Ohm

External
Master
Board
uC 1000 M Ohm

Part of XPSP

IGBT3
i
400Vrtm

PI PWM +200Vrtm

0Vrtm Filament
IGBT4 Transformer

Figure 2.7

The HV switching works at 20 kHz. The transformer primary is connected to the C1/C2 point which
corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V). The control board
commands alternately the IGBT1 and IGBT2. The On time duration depends of the output power needed. By
this way, the transformer primary receives, via C3, plus and minus 200 V referenced to the fix 200 VRTM.

AA83678 Service Manual ARL GenX 2-9


Chapter 2 GenX Description

Chronogram for low power case:

UC3-0Vrtm

+400V

+200V

0 t

IGBT1
Command

IGBT2
Command

Figure 2.8

Chronogram for high power case:

UC3-0Vrtm

+400V

+200V

0 t

IGBT1
Command

IGBT2
Command

Figure 2.9

The transformer inductor and C3 capacitor make a resonant circuit. By this way the current look like
sinusoidal:

2-10 Service Manual ARL GenX AA83678


GenX Description Chapter 2

UC3-0Vrtm

+400V

+200V

0 t
Transformer
Primary
current

Figure 2.10

The filament power supply switching works at 40 kHz. The transformer primary is connected to the C1/C2
point which corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V). The
control board commands alternately the IGBT3 and IGBT4. The On time duration depends of the output
power needed (from 80 W to 120 W). By this way, the transformer primary receives positive and negative
200 V referenced to the fix 200 VRTM.

See after the voltage and the current at the primary of the filament transformer:
Filament
transformer

+200V

-200V t

Filament
Transformer Primary
current

0 t

Figure 2.11

The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.

AA83678 Service Manual ARL GenX 2-11


Chapter 2 GenX Description

HV Over Current Safety:


The current applied to the HV transformer located on XPSM is monitored by the IC10.
The current sampling voltage is applied to a dual window comparator.
The Positive current limit reference is 4.3V and 2.65V corresponding to a positive current of 140 and 5A.
The Negative Current limit reference is 2.36V and 0.71V corresponding to a negative current of -140 and -
5A.

U (V)
Uref 4.3 V

Uref 2.65 V

Uref 2.36 V

Uref 0.71 V

I (A)
-I Max -5A 5A I Max
-140A 140A
Trip Switch On Level Trip

Figure 2.12

If the current is out of limit the system does generate a /PwmTrip to the DSP of XPSC.
The PwmTrip does disable the HvPwmH and HvPwmL driving the gates of the HV IGBT Inverter Stages.
To the ather hand the system does generate /Limit I Hv. This signal does with the transistors M6, M9 and
Galvanic insulated drivers IC5 and IC8 disable the HvPwmH and HvPwmL driving the gates of the HV
IGBT Inverter Stages.

Filament Over Current safety:


The current applied to the Filament transformer located on XPSM is monitored by the IC11.
The current sampling voltage is applied to a dual window comparator.
The Positive current limit reference is 3V and 2.57V corresponding to a positive current of 15 and 3A.
The Negative current limit reference is 2.35V and 1.87V corresponding to a negative current of -15 and -3A.

U (V)
Uref 3.0 V

Uref 2.5 V

Uref 2.35 V

Uref 1.87 V

I (A)
-I Max -3A 3A I Max
-15A 15A
Trip Switch On Level Trip

Figure 2.13

2-12 Service Manual ARL GenX AA83678


GenX Description Chapter 2

If the current is out of limit the system does generate /Limit I Fil. This signal does with the transistors M3,
M12 and Galvanic insulated drivers IC3 and IC13 disable the FilPwmH and FilPwmL driving the gates of
the Filament IGBT Inverter Stages.

+24V Safety:
If < 20V or if > 23V the IC7 does generate Lim_24V.
If active the HV and Filament IGBT Gate control signals are disabled.

Low voltages:
Gate low voltages: +VGate 1, +VGate L1, +VGate 3, +VGate L3 -VGate 1, -VGate L1, -VGate 3, -VGate
L3 are generated from the +24V via galvanic insulated voltage generators IC1, IC4, IC6, IC9.
Emitor low Voltages: EmH1, EmH2, EmL1, EmL2, are generated from the +24V via galvanic insulated
voltage generators IC1, IC4, IC6, IC9.
+5VA is generated from +24V via low voltage generator IC2.
Designation Remark and typical value
Terminal Mode Cmd
DY 2,3,11,4,0 XPSP, +5VM XPSP AD CHAN0: 204
DY 2,3,11,4,1 XPSP, +24V( XPSP AD CHAN1: 204
DY 2,3,11,4,2 XPSP, -12VRTM (True or False) Note 1 XPSP AD CHAN2: >128 (True)
DY 2,3,11,4,3 XPSP, +12VRTM (True or False) Note 2 XPSP AD CHAN3: >128 (True)
DY 2,3,11,4,4 XPSP, NC XPSP AD CHAN4:
DY 2,3,11,4,5 XPSP, NC XPSP AD CHAN5:
DY 2,3,11,4,6 XPSP, PowerTemp XPSP AD CHAN6: 212
DY 2,3,11,4,7 XPSP, XpspTemp XPSP AD CHAN7: 81(at 25°C)
Note 1: Since XPSP Schematic S112 067 Mode Level 10 the +12 VRTM generator has been suppressed.
Since XPSP Schematic S112 067 Mode Level 11 the + 12 VRTM generator has been suppressed.
A + 5 VA_PTB has been introduced to check the presence of the External Voltage Presence.

Note 2: Since XPSP Schematic S112 067 Mode Level 10 the -12 VRTM generator has been suppressed.
Since XPSP Schematic S112 067 Mode Level 11 the - 12 VRTM generator has been suppressed.
A + 5 VA_PTB has been introduced to check the presence of the External Voltage Presence.

The six-pack IGBT is cooled by water, for high power application, and by air, for low power application.

AA83678 Service Manual ARL GenX 2-13


Chapter 2 GenX Description

X-ray Power Supply, Multiplier ASSEMBLY (XPSM)

Figure 2.14

The XPSM assembly has three main functions. First is the HV supply, second is the emission current
measurement and third is the filament current supply.

Part of XPSM assembly


230Vac FET1 L D HighVoltage
+400V
i FET2 Multiplier
u u C1 +200Vrtm
C2
0Vrtm

DSP 1000 M Ohm

External
Master
Board
uP 1000 M Ohm

Part of XPSM
assembly
i
400Vrtm

+200Vrtm
PI PWM
0Vrtm
Filament
Transformer

Figure 2.15

2-14 Service Manual ARL GenX AA83678


GenX Description Chapter 2

The 400V from the IGBTs on the XPSP board, via XPSM board goes to the HV transformer. At the
transformer secondary the voltage changes according to the IGBT1 and IBGT2 ON time. The maximum
voltage is near 12000 V. The full wave multiplier increases six times this voltage to reach at maximum
70000 V. Two dividers are used to monitor the high voltage. The command board uses one of them to
control the HV via the servo loop function according to asking high voltage (called “kV” in the WinXRF).
The second is used to control the first via a subtraction function.

Use the status channels to retrieve the HV current values.

Terminal Mode Designation Remark and typical value


RS 2048 X-ray kilovolts [kV] Value according to kV set point
RS 2097 XPSM HV divider difference [VADC] (VADC-2V) correspond to 1887VHV

The emission current is measured at two positions. The first position is the multiplier entrance current, which
represents the high voltage current. The second position is the current that return from the filament
(Kirchoff’s current law). The control board uses the first to control via the servo loop function, the asked
emission current (called “mA” in the WinXRF). The control board, to calculate the leakage current, uses
both measurements by subtraction function. The leakage current corresponds to the “multiplier absorbed
current” minus the “filament return current”. When the leakage current is not coherent, the control board
prevents to switch power on or if the power is on, switches it off. See table after for details.

Case before power switch on:

CASE ALARM ALARM DESIGNATION TIME SEVERITY


Leakage < -5 mA 9770 X-ray negative leakage current 50 ms Debug
error
Leakage < -5 mA 8389 X-ray negative leakage current 500 ms Fatal
error
Leakage > 5 mA 9769 X-ray positive leakage current 50 ms Debug
error
Leakage > 5 mA 8390 X-ray positive leakage current 500 ms Fatal
error

Remark : I leakage = I multiplier absorbed – I filament return

Case during power on:

CASE ALARM ALARM DESIGNATION TIME SEVERITY


Leakage < -5 mA 9770 X-ray negative leakage current 50 ms Debug
error
Leakage < -5 mA 8389 X-ray negative leakage current 500 ms Fatal
error
Leakage > 5 mA 9769 X-ray positive leakage current 50 ms Debug
error
Leakage > 5 mA 8390 X-ray positive leakage current 500 ms Fatal
error
[Leakage] >1 mA 9760 X-ray leakage current error 50 ms Debug
[Leakage] >1 mA 8865 X-ray leakage current error 10 s Warning

[Leakage] >1.5 mA 9759 X-ray leakage current error 50 ms Debug


[Leakage] >1.5 mA 8304 X-ray leakage current error 10 s Warning

Remark : [Leakage] correspond to the absolute value of the leakage current

AA83678 Service Manual ARL GenX 2-15


Chapter 2 GenX Description

You can retrieve the leakage value by using the terminal mode command explained after:

Terminal Mode Cmd Designation Remark and typical value


RS 2069 X-ray tube leakage current [mA] -0.3 mA > Leakage +0.3 mA

The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.

Terminal Mode Cmd Designation Remark and typical value


DY 2,3,11,5,0 XPSM, +5VM XPSM AD CHAN0: 204
DY 2,3,11,5,1 XPSM, +12V XPSM AD CHAN1: 204
DY 2,3,11,5,2 XPSM, -12V XPSM AD CHAN2: 051
DY 2,3,11,5,3 XPSM, Arc XPSM AD CHAN3: 000
DY 2,3,11,5,4 XPSM, HvDividerDifference XPSM AD CHAN4: 127
DY 2,3,11,5,5 XPSM, FilamentVoltage XPSM AD CHAN5: 130 (ON)
DY 2,3,11,5,6 XPSM, XpsmFilamentTemp XPSM AD CHAN6:
DY 2,3,11,5,7 XPSM, XpsmHvTemp XPSM AD CHAN7: 255

The multiplier box is cooled by water (Only in high power application). Two fans cool the rest of the
multiplier assembly.

2-16 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Command board, X-ray Power Supply Control (XPSC)

Figure 2.16

This board is the interface with the XQM master board in the instrument. It controls all functions in the
GenX, like the different servo-loops for the HV, emission current and filament power supply and few
watchdog safety systems. It controls too the different external parts like interlock safety circuit and water
cooler system.

HV
Transformer
230Vac FET1 L D +400Vrtm
HighVoltage
C1
u
i FET2
+200Vrtm
Multiplier
u
C2
0Vrtm

DSP 1000 M Ohm

External
Master
Board
uP 1000 M Ohm

XPSC i
400Vrtm
Part
+200Vrtm
PI PWM
0Vrtm Filament
Transformer

Figure 2.17

AA83678 Service Manual ARL GenX 2-17


Chapter 2 GenX Description

Functions of the µ C 80C196KR

Figure 2.18

2-18 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Functions of the 80c196kr µ Controller ports


Port P0 used as A/D converter for the control of the water status:

Port State Designation Remark


P0 0 analogic primary water temperature at the input From 5°C to 40°C
P0 1 analogic primary water temperature at the output From 5°C to 45°C
P0 2 analogic deionised water temperature at the input From 5°C to 50°C
P0 3 analogic deionised water temperature at the output From 5°C to 53°C
P0 4 analogic Deionised water level. Used in ON/OFF mode
P0 5 analogic deionised water resistivity driver 3.3 V (>3 < 4.5)
P0 6 analogic deionised water resistivity From 550 K to Infinite
P0 7 analogic 2.5 Volt Reference Voltage Not used
Port Address: 1FDAH

Port State Designation Remark


P1 0 I
P1 1 I /HvArc Arcs detection from XPSM high voltage divider
P1 2 I /CuArc Arcs detection from XPSM negative emission current -
P1 3 I /MaxFilCur Filament heating current limitation max. If >14 A
P1 4 I FilCurSD Filament heating current Shut Down. If >14.5 A
P1 5 O WdToggle Watch Dog activity supervision.
P1 6 I /AdcDor0 ADC Data Output Ready
P1 7 I AdcEoc0 ADC End Of Conversion
Port Address: IN:1FCFH OUT:1FCDH

Port State Designation Remark


P2 0 O TXD XSN serial data transmission
P2 1 I RXD XSN serial data reception
P2 2 I Adcsars ADC in conversion state
P2 3 O XpsReady XPS presence acknowledge message to XQM
P2 4 O AdcConv0 Start Conversion
P2 5 O XpsOk XPS µ C OK state (LED2 GREEN ON)
P2 6 O XpsNotOk XPS µ C failure NOT OK state (LED1 RED ON)
P2 7 O ClkOut EPLD 610 clock

Port Address: IN:1FD6H OUT:1FD4H

Remarks: P3.0 to P3.7 AD0 to AD7 bus


P4.0 to P4.7 AD8 to AD15 bus

AA83678 Service Manual ARL GenX 2-19


Chapter 2 GenX Description

Port State Designation Remark


P5 0 O ALE Address Latch Enable
P5 1 O EnWd Enable Watch Dog
P5 2 O /WR Write
P5 3 O /RD Read
P5 4
P5 5 O SsioTxSync Serial link communication transmit synchronisation
P5 6 Ready Wait State Signal
P5 7 I EPROM’s Serial EEPROM Chip Selection Signal IC12
Port Address: IN:1FF7H OUT:1FF5H

Port State Designation Remark


P6 0 O /XsnTxDEn XSN data transmission enable for serial port and display
P6 1 O Adc0Cs0 Miscellaneous 8×12 bits ADC12138 Chip Select IC 3
P6 2 I Adc0Cs1 Miscellaneous 8×8 bits ADCO838 Chip Select IC 7. First AD
P6 3 O Adc0Cs2 Miscellaneous 8×8 bits ADCO838 Chip Select IC 8. Sec. AD
P6 4 O SClk ADCO 838 and ADC12138 Clock
P6 5 O SdatIn ADCO 838 and ADC12138 Data out
P6 6 I SClk ADCO 838 and ADC12138 Clock
P6 7 I SdatOut ADCO 838 and ADC12138 Data in
Port Address: IN:1FD7H OUT:1FD5H

2-20 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Functions of the PPI ports

Figure 2.19

PPI 1 82C55 IC 10
Port State Designation Remark
PPI1 PA 0 I /Int1 Pressostat Active if pressure < 25 KPA
PPI1 PA 1 I /Int2 Electronic rack safety switch
PPI1 PA 2 I /Int3 Tank cover + X-ray tube position switches
PPI1 PA 3 I /Int4 Magazine + load lift cover switches
PPI1 PA 4 I /Int5 X-ray generator rear switch
PPI1 PA 5 I /IntRet 60°C secondary water thermostat
PPI1 PA 6 I /24VPOK +24V instrument OK
PPI1 PA 7 I /WaCoOn Water cooler ON
Port Address: FFC0H

Port State Designation Remark


PPI1 PB 0 I ExRelayPresent Power Distribution Board Main Contactor 24 V Coil Return
PPI1 PB 1 I
PPI1 PB 2 I
PPI1 PB 3 I
PPI1 PB 4 I PrWaFlow Primary Water Flow Reed Contact
PPI1 PB 5 I DeWaFlow Deionised Water Flow Reed Contact
PPI1 PB 6 I
PPI1 PB 7 I
Port Address: FFC1H

Port State Designation Remark


PPI1 PC 0 O BuzzerOn Buzzer On Command
PPI1 PC 1 O PrCoolerCtrl Logic Primary Water Valve Command to L293D IC23
PPI1 PC 2 O XrayLampOn 24V X-Ray On Lamp command. Only used on 1.2 kW
PPI1 PC 3 O /DspRst DSP Reset command
PPI1 PC 4 O ExternaRelay Power Distribution Board Main Contactor coil command
PPI1 PC 5 O WaPump Deionised Water Pump command
PPI1 PC 6 O WaCooler External Water Cooler ON command
PPI1 PC 7 O /EnFilCu Filament Heating PWM Generator Enable. IC 20
Port Address: FFC2H

AA83678 Service Manual ARL GenX 2-21


Chapter 2 GenX Description

Figure 2.20

PPI 2 82C55 IC 11:

Port State Designation Remark


PPI2 PA 0 I SW1.1 From XPSC board switch 1
PPI2 PA 1 I SW1.2 From XPSC board switch 2
PPI2 PA 2 I SW1.3 From XPSC board switch 3
PPI2 PA 3 I SW1.4 From XPSC board switch 4
PPI2 PA 4 I JumperW2.1 From XPSC board W2 jumper 1
PPI2 PA 5 I JumperW2.2 From XPSC board W2 jumper 2
PPI2 PA 6 I jumperW2.3 From XPSC board W2 jumper 3
PPI2 PA 7 I JumperW2.4 From XPSC board W2 jumper 4
Port Address: FF80H

Port State Designation Remark


PPI2 PB 0 O PowerOn XPSI Contactor K1 coil Command
PPI2 PB 1 O EnPower IC26 Latch Input to Output Validation
PPI2 PB 2 O
PPI2 PB 3 O
PPI2 PB 4 O
PPI2 PB 5 O /DspPCe DSP Code EEPROM
PPI2 PB 6 O EnDisRd2 En XPSC Display data Transfer
PPI2 PB 7 O EnXpscDis Enable DSP Data Display Not used
Port Address: FF81H

Port State Designation Remark


PPI2 PC 0 O /DspCs DSP Chip Selection
PPI2 PC 1 O /AdcCs1Ext XPSI 8×8 bits status ADC chip selection IC 3
PPI2 PC 2 O /AdcCs2Ext XPSP 8×8 bits status ADC chip selection IC 1
PPI2 PC 3 O /AdcCs3Ext XPSM 8×8 bits status ADC chip selection IC 2
PPI2 PC 4 O
PPI2 PC 5 O TfsSync DSP Transmit synchronisation
PPI2 PC 6 O SsioRxSync Serial link Communication Receive synchronisation
PPI2 PC 7 O RfsSync DSP Receive synchronisation
Port Address: FF82H

2-22 Service Manual ARL GenX AA83678


GenX Description Chapter 2

AD converters
First AD Miscellaneous 8×8 bits AD converter. To display the different AD input values, use the
terminal mode commands DY
See display information. AD display
commands.doc

Max Filament Current:


14 Amps = 3724 mV = 190 Digits

Safety Filament Current:


14.5 Amps = 3857 mV = 197 Digits

Lamp1:
24V DC for GEN’X 1200 W

Lamp2:
24V DC for GEN’X 1200 W

Filament Current Monitoring:


0.266 V/A
5000 mV = 255 digits
1digit = 19.6 mV

Filament Current Reference Monitoring:


0.266V/A
5000 mV = 255 digits
1digit = 19.6 mV

Second AD Miscellaneous 8×8 bits AD converter.


Low voltages:

Pos. 5V = 204 digits


Pos. 12V = 204 digits
Pos. 24V = 204 digits
Neg. 5V = 51 digits
Neg. 12V = 51 digits

HV Arc Threshold
55 Digits = 1080mV

Current Arc Threshold


0 Digits = 0V

AA83678 Service Manual ARL GenX 2-23


Chapter 2 GenX Description

AD Miscellaneous 8×12 bits AD converter.


Neg. Emission Current -: From XPSM

3500 mV for 140 mA 25 mV/ mA

Pos. Emission Current: From XPSM


3500 mV for 140 mA 25 mV/ mA

High Voltage: From XPSM


4000 mV for 75 kV 53.333 mV/mA

XPSC temp:
Should not exceed 65°C

DA Converter

Filament Heating Current Reference:


266 mV/A

High Voltage Arc Threshold


55 Digits = 1080 mV

Current Arc Threshold


0 Digits = 0 mV

Max Filament Current:


14 Amps = 3724 mV = 190 Digits

Safety Filament Current:


14.5 Amps = 3857 mV = 197 Digits

X-Ray On Lamps 24 VDC 1200 W Generator

This circuit allows detecting the X-Ray On Lamps


state.
Both OK.
Lamp1 defect. W Alarm 8829
Lamp2 defect. W Alarm 8830
Both defect. F Alarm 8281
Short-Circuit. F Alarm 8281

2-24 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Water control circuit


The parameters monitored by the µ controller are:

 The tap or primary water flow. An alarm is immediately issued if the flow drops. However, the default
will only shut down the generator after a delay of 20 seconds.
 The tap or primary water inlet and outlet temperatures.
 The deionised water flow.
 The deionised water inlet and outlet temperatures.
 The deionised water resistivity (560k**cm).
 The deionised water level
The change of state of cooling safety causes the immediate shut down of the generator by acting on the
generator interlock input. The parameter responsible for the shut down will generate an alarm reported to the
ACS software.

Water parameter requests


Tap or primary water

 Flow 2.5 to 4.5 l/min


 Pressure <4 bars
 Inlet temperature 5°C < Inlet <18°C

Deionised water

 Flow 2 l/min
 Pressure <3.5 bars
 Outlet temperature <60°C

 Regulation temperature 36°C (can be modified by user)

 Resistivity >550 K

AA83678 Service Manual ARL GenX 2-25


Chapter 2 GenX Description

Deionised water control circuit


MULTIVIBRATOR ASTABLE
+14V
45Hz VPP = 28V
-14V

22ms
+ DEIONIZED WATER RESI. DRIVER IN
-
+14V
0V
3.3V
tol 3<>4.5V
A/D
PORT 0
1M
DEIONIZED
WATER RESIST.
+ IN
resistivity gauge RX
-

Figure 2.21

To monitor the deionised water resistivity select the status RS 2054

RX VIN VPP Remark


 4.5V 28VPP OPEN CIRCUIT
1 M 2.5V 7VPP
560 K 1.8V 5VPP
0 0V 0V SHORT CIRCUIT.

KV, mA Monitoring Calibration


To perform the calibration type in the terminal mode command menu:
XC 1
The system will automatically perform the calibration.

2-26 Service Manual ARL GenX AA83678


GenX Description Chapter 2

HV servo-loop control
The control board receives the asked HV from the PC software application via the COM port and the
instrument master board. The asked HV is compared with the real HV to generate the HV error, which is
used through the PID corrector to PWM duty-cycle. The PWM signals control the power stage on XPSP
board to supply the right voltage according to the HV error. The XPSM assembly transformer and multiplier
increase this voltage to generate the real HV. The DSP evaluates the values each 50 us.

uC DSP
HV
HV Multiplier
set point Power
+ PID PWM Box
- Stage

HV
D HighVoltage 18762
A

Part of Part of
Part of XPSP board XPSM assembly
XPSC board

Figure 2.22

Filament current servo-loop control


The control board receives the desire current from the emission current servo-loop control part. This value is
converted, via a digital-to-analogic converter, into analogic voltage. This voltage corresponds to the needed
current (0.266 V/A).

The rest of the bloc diagram works like a controlled current source. I.e., the filament current changes
according to the control voltage. 1 volt corresponds to 3.76 amperes. The set-point voltage is compared with
the current sensor voltage to control, via the proportional corrector, the PWM duty-cycle at 40 kHz. The
power board contains the power stage, which generates the power for the transformer. The transformer is
used to increase the current at the right level (up to 12 A). The filament receives a DC current through four
rectifier diodes.

uC
Current D + P PWM + -
set point A - X-ray
Tube
Filament

Part of Part of Part of


XPSC board XPSP board XPSM assembly

Figure 2.23

AA83678 Service Manual ARL GenX 2-27


Chapter 2 GenX Description

Emission current servo-loop control


The control board receives from the instrument master board the desired emission current. This value is
compared, via an analogic-to-digital converter, to the inverted value, which corresponds to the current
absorbed in the multiplier. The comparison result is used through the PID corrector to control the filament
current power supply part. The µC evaluates the values each 50 ms.

uC
Emission
current
set point
+
-
PID Filament current power supply

HV
Power Multiplier
Stage

D -1
A
R
Part of Part of
Part of XPSC board XPSP board XPSI assembly

Figure 2.24

Power Factor Correction control


The DSP receives from the µC the desired intermediate DC working voltage according to the asked high
voltage (HV). During power off state this value stays to zero. When the switches on, the value changes to
360 V and then changes according to the asked output voltage (kV). This value is compared to the real
intermediate voltage to calculate the servo-loop error.

The rectified input voltage is weighted by this voltage error to become the current reference for a second
servo-loop. The real current, to generate the current error, is compared to this reference. The GPFC corrector
uses the current reference, the real current and the current error values to control the PWM duty-cycle, which
controls the Power MOS FET (FET2).

The FET1 is controlled by three ways. The first one is the intermediate voltage via a P corrector. The PWM
duty-cycle changes according to intermediate voltage. The second one is made on the real current value and
a comparator with a current threshold. At switching power on, this current threshold corresponds to 2
amperes to limit the inrush current. When after 10 seconds, the intermediate voltage exceeds 360 V, the
current threshold changes to 50 Amperes. This corresponds to the maximum acceptable current during power
on state. The last way controls the FET1 by hardware. The real current is compared with a reference value.
When the real current exceeds 50 amperes the FET1 is switched off.

2-28 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Part of XPSI board


P
FET1

FET2

PWM

A
D
P

-
A +
D

P P
DSP
PI

GPFC PWM

Reference D
+ -
Voltage A

Part of XPSC board

Figure 2.25

The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogic to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.

Terminal Mode Cmd Designation Remark and typical value


DY 2,3,11,5,0 XPSM, +5VM XPSM AD CHAN0: 204
DY 2,3,11,5,1 XPSM, +12V XPSM AD CHAN1: 204
DY 2,3,11,5,2 XPSM, -12V XPSM AD CHAN2: 051
DY 2,3,11,5,3 XPSM, Arc XPSM AD CHAN3: 000
DY 2,3,11,5,4 XPSM, HvDividerDifference XPSM AD CHAN4: 127
DY 2,3,11,5,5 XPSM, FilamentVoltage XPSM AD CHAN5: 130 (ON)
DY 2,3,11,5,6 XPSM, XpsmFilamentTemp XPSM AD CHAN6:
DY 2,3,11,5,7 XPSM, XpsmHvTemp XPSM AD CHAN7: 255

AA83678 Service Manual ARL GenX 2-29


Chapter 2 GenX Description

Switches and jumpers


ON Secondary water cooling regulation inactive
1
OFF Secondary water cooling regulation active
S1 2 OFF Reserved
3 OFF Reserved
4 OFF Reserved
W2 1,2,3,4 Generator type
W2.4 W2.3 W2.2 W2.1
removed removed removed removed Unknown generator type
installed removed installed removed 2500 W 60 kV/ 100 mA
removed removed removed installed 3600 W, 60 kV/ 120 mA
removed removed installed removed 3600 W, 70 kV/ 120 mA
removed removed installed installed 4200 W, 60 kV/ 120 mA
removed installed removed removed 4200 W, 70 kV/ 140 mA
removed installed removed installed 1200 W, 60 kV/ 60 mA
removed installed installed removed Retrofit 3600 W, 60 kV/ 120 mA
removed installed installed installed Retrofit 3600 W, 70 kV/ 120 mA
installed removed removed removed Retrofit 4200 W, 60 kV/ 120 mA
installed removed removed installed Retrofit 4200 W, 70 kV/ 140 mA

2-30 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Display information terminal command


INFO OPTION DESCRIPTION
1 2
DY 2,3,2 2 - Secondary water level SW level : full
5 - Secondary water flow SW flow : ON
DY 2,3,3 0 - Current X-ray state (default display) 50kV 12mA
1 0 Uncalibrated feedback kilovolts [kV] KV back: 50.00kV
3 Calibrated feedback kilovolts [kV] KV back cal:50.00 kV
4 Calibrated and filtered feedback kilovolts KV back fil:50.00 kV
[kV]
3 0 Uncalibrated feedback milliamps [mA] MA back: 12.00mA
3 Calibrated feedback milliamps[mA] MA back cal:12.00 mA
4 Calibrated and filtered feedback MA back fil:12.00 mA
milliamps[mA]
5 0 Reference I filament [A] Fil I Ref: 9.54A
DY 2,3,4 0 - X-ray generator control state SU:NA EX:10 CAL: 0
1 Primary and secondary water control state
WA: 7 SW: 2
2 - Generator state GenTsk:STAND-BY
DY 2,3,7 0 - Secondary water cooling SW PWM: 25% 44.70C
DY 2,3,10 - - All safety switches closed Interlock: closed
Spectrometer tank pressostat open Interlock: open
Interlock: Pressostat.
Spectrometer tank cover safety switch Interlock: open
open Interlock: Intlk2
Primary chamber or primary beam Interlock: open
filter or elliptic mask safety switch Interlock: Intlk3
open
X-ray tube position safety switch open Interlock: open
Interlock: Intlk4
High-voltage protection or Interlock: open
X-ray tube connection safety switch open Interlock: Intlk5
X-ray secondary water over-temperature Interlock: open
sensor switch open Interlock: IntlkRtn
DY 2,3,11 1 0 XPSC, First AD, MaxFilCur XPSC AD1 CHAN0: 255
1 XPSC, First AD, SafFilCur XPSC AD1 CHAN1: 255
2 XPSC, First AD, Lamp1 XPSC AD1 CHAN2: 255
3 XPSC, First AD, Lamp2 XPSC AD1 CHAN3: 255
4 XPSC, First AD, Nothing XPSC AD1 CHAN4: 255
5 XPSC, First AD, FilCurMon XPSC AD1 CHAN5: 255
6 XPSC, First AD, FilCurRefMon XPSC AD1 CHAN6: 255
7 XPSC, First AD, Nothing XPSC AD1 CHAN7: 255
2 0 XPSC, Second AD, Nothing XPSC AD2 CHAN0: 255
1 XPSC, Second AD, +5V XPSC AD2 CHAN1: 255
2 XPSC, Second AD, +12V XPSC AD2 CHAN2: 255
3 XPSC, Second AD, +24V XPSC AD2 CHAN3: 255
4 XPSC, Second AD, -12V XPSC AD2 CHAN4: 255
5 XPSC, Second AD, -5V XPSC AD2 CHAN5: 255
6 XPSC, Second AD, HvArcTh XPSC AD2 CHAN6: 255
7 XPSC, Second AD, CuArcTh XPSC AD2 CHAN7: 255
3 0 XPSI, +5VM XPSI AD CHAN0: 255
1 XPSI, +12V XPSI AD CHAN1: 255

AA83678 Service Manual ARL GenX 2-31


Chapter 2 GenX Description

2 XPSI, +24V XPSI AD CHAN2: 255


3 XPSI, -12V XPSI AD CHAN3: 255
4 XPSI, LampCurrent XPSI AD CHAN4: 255
5 XPSI, NtoGndVoltage XPSI AD CHAN5: 255
6 XPSI, InputTemp1 XPSI AD CHAN6: 255
7 XPSI, InputTemp2 XPSI AD CHAN7: 255
4 0 XPSP, +5VM XPSP AD CHAN0: 255
1 XPSP, +24V XPSP AD CHAN1: 255
2 XPSP, -12VRTM (True or False) XPSP AD CHAN2: 255
3 XPSP, +12VRTM (True or False) XPSP AD CHAN3: 255
4 XPSP, Nothing XPSP AD CHAN4: 255
5 XPSP, Nothing XPSP AD CHAN5: 255
6 XPSP, PowerTemp XPSP AD CHAN6: 255
7 XPSP, XpspTemp XPSP AD CHAN7: 255
5 0 XPSM, +5VM XPSM AD CHAN0: 255
1 XPSM, +12V XPSM AD CHAN1: 255
2 XPSM, -12V XPSM AD CHAN2: 255
3 XPSM, Arc XPSM AD CHAN3: 255
4 XPSM, HvDividerDifference XPSM AD CHAN4: 255
5 XPSM, FilamentVoltage XPSM AD CHAN5: 255
6 XPSM, XpsmFilamentTemp XPSM AD CHAN6: 255
7 XPSM, XpsmHvTemp XPSM AD CHAN7: 255

Displaying the content of a memory location DY 2,3,100,0,Address,Data Type,Base


Example: displaying the content of port A of PPI 1 DY 2,3,100,0,FFC0H,0,2
Reading the content of a memory location RB 2,3,0,Address
Example: reading the content of port A of PPI 1 RB 2,3,0,FFC0H

2-32 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Status channels
Terminal Mode Cmd Designation Remarks
RS 2048 X-ray kilovolts [kV] Available in all versions
RS 2049 X-ray milliamps [mA] Available in all versions
RS 2050 X-ray cooling primary water input High power version only
temperature [0C]
RS 2051 X-ray cooling primary water output High power version only
temperature [0C]
RS 2052 X-ray cooling secondary water input Available in all versions
temperature [0C]
RS 2053 X-ray cooling secondary water output Available in all versions
temperature [0C]
RS 2054 X-ray cooling secondary water Available in all versions
resistivity. [k.cm]
RS 2055 X-ray tube filament current [A] Available in all versions
RS 2056 XPS +15Volts [V] Non-GenX or GenX retrofit version
RS 2057 XPS -15Volts [V] Non-GenX or GenX retrofit version
RS 2058 XPS +24Volts [V] Available in all versions
RS 2061 XPS +5Volts [V] Available in all versions
RS 2063 X-ray activity time [hour] Available in all versions
RS 2064 X-ray tube arcs count [] Available in all versions
RS 2067 Laboratory ambient temperature [0C] Low power version only
RS 2068 X-ray generator output temperature Low power version only
[0C]
RS 2069 X-ray tube leakage current [mA] Available in all versions
0
RS 2070 XPSC board temperature [ C] Available in all versions
RS 2071 XPS +12Volts [V] Available in all versions
RS 2072 XPS -12Volts [V] Available in all versions
RS 2073 XPS +2.5Volts reference [V] Low power version only
RS 2080 XPSC -5Volts [V] Available in all GenX versions
RS 2081 XPSI +5Volts [V] Available in all GenX versions
RS 2082 XPSI +12Volts [V] Available in all GenX versions
RS 2083 XPSI +24Volts [V] Available in all GenX versions
RS 2084 XPSI –12Volts [V] Available in all GenX versions
RS 2085 XPSI Neutral to Gnd [V] Available in all GenX versions
RS 2086 XPSI Mosfet M1 temperature [0C] Available in all GenX versions
RS 2087 XPSI Mosfet M2 temperature [0C] Available in all GenX versions
RS 2088 XPSP +5Vm [V] Available in all GenX versions
RS 2089 XPSP +24Volts [V] Available in all GenX versions
RS 2090 XPSP -10Vrtm [V] Available in all GenX versions
RS 2091 XPSP +15Vrtm [V] Available in all GenX versions
RS 2092 XPSP IGBT temperature [0C] Available in all GenX versions
RS 2093 XPSP board temperature [0C] Available in all GenX versions
RS 2094 XPSM +5Vm [V] Available in all GenX versions
RS 2095 XPSM +12Volts [V] Available in all GenX versions
RS 2096 XPSM -12Volts [V] Available in all GenX versions
RS 2097 XPSM Hv divider difference [kV] Available in all GenX versions
RS 2098 X-ray tube filament voltage [V] Available in all GenX versions
0
RS 2099 XPSM filament temperature [ C] Available in all GenX versions
RS 2100 XPSM Hv temperature [0C] Available in all GenX versions
RS 2101 Monitoring 400Volts [V] Available in all GenX versions

AA83678 Service Manual ARL GenX 2-33


Chapter 2 GenX Description

2048 - 2559: XPS Firmware status

2048 XPS ALL kV/mA measurement calibration


Read only : NO Sent after initialisation : NO
Parameters: Type: Default firm: Limit firm: Default soft:
KV reading offset [kV] FLOAT 0.0 - 0.0
KV reading slope [] FLOAT 1.0 - 1.0
MA reading offset [mA] FLOAT 0.0 - 0.0
MA reading slope [] FLOAT 1.0 - 1.0
2049 XPS ADVANT’X Low power temperature offset
Read only : NO Sent after initialisation : NO
Parameters: Type: Default firm: Limit firm: Default soft:
State (A5A5H = defined) STRING FFFFH - FFFFH
Laboratory temperature offset [oC] FLOAT 0.0 - 0.0
X-ray power supply temperature offset [oC] FLOAT 0.0 - 0.0
Secondary water input temperature offset [oC] FLOAT 0.0 - 0.0
Secondary water output temperature offset [oC] FLOAT 0.0 - 0.0
XPS power stage temperature offset [oC] FLOAT 0.0 - 0.0
2051 XPS ADVANT’XP X-ray arc occurrences
ADVANT’XP GenX
Read only : YES Sent after initialisation : NO
Parameters: Type: Default firm: Limit firm: Default soft:
Record 1 (most recent) [] STRING - 0
“XPS time tic” “kV” “mA” “state” “arc”
Record 2 [] STRING - 0
“XPS time tic” “kV” “mA” “state” “arc”
“ “ “ “ “ “
Record 20 (less recent) [] STRING - 0
“XPS time tic” “kV” “mA” “state” “arc”
2052 XPSC ADVANT’XP Gen'X XPSC option switches
Read only : NO Sent after initialisation : YES
Parameters: Type: Default firm: Limit firm: Default soft:
State (A5A5 = defined) STRING FFFFH - 0
Secondary water cooling regulation inactive BYTE 0 - 0
Not used BYTE 0 - 0
Not used BYTE 0 - 0
Not used BYTE 0 - 0
Note: If “state <> defined”, these value are determined during the initialisation of XPSC.

2-34 Service Manual ARL GenX AA83678


GenX Description Chapter 2

Alarms list for firmware version 3.2.8 and lower

Alarm Issued Severity Description


number by
8192 XPS F Unknown command received from XQM96
8193 XPS F Primary water flow sensor error
8194 XPS F Secondary water flow sensor error
8195 XPS F Secondary water pump error
8196 XPS F Desired voltage (kV) too low
8197 XPS F Desired voltage (kV) too high
8198 XPS F Desired current (mA) too low
8199 XPS F Desired current (mA) too high
8200 XPS F Desired power too high
8201 XPS F Illegal power change mode
8202 XPS F Illegal power change speed
8203 XPS F Illegal anode
8204 XPS F Spectrometer pressostat open
8205 XPS F Spectrometer tank cover safety open
8206 XPS F Primary chamber or PBF safety open
8207 XPS F X-ray tube position safety open
8208 XPS F X-ray tube connection safety open
8209 XPS F Illegal interlock
8210 XPS F Time-out on X-ray ON ack. or X-ray lamps fault
8211 XPS F Over kilovolts on X-ray power supply
8212 XPS F Over milliamps on X-ray power supply
8213 XPS F Over power on X-ray power supply
8214 XPS F X-ray power supply fault
8215 XPS F X-ray tube filament open
8216 XPS F X-ray tube filament over current
8217 XPS F Kilovolts regulation fault
8218 XPS F Milliamps regulation fault
8219 XPS F X-ray ON lamp broken
8220 XPS F Low kilovolts limit
8221 XPS F Leak current too high
8222 XPS F Secondary water flow OFF
8223 XPS F Secondary water temperature too high
8224 XPS F 'PA' or 'CO' command not allowed now
8225 XPS F Secondary water resistivity driver fault
8226 XPS F Secondary water resistivity gauge fault
8227 XPS F Secondary water resistivity too low
8228 XPS F Primary water inlet temperature gauge fault
8229 XPS F Primary water outlet temperature gauge fault
8230 XPS F Secondary water inlet temperature gauge fault
8231 XPS F Secondary water outlet temperature gauge fault
8232 XPS F Primary water cooler OFF
8233 XPS F Primary water OFF
8234 XPS F Secondary water level gauge fault
8235 XPS F Secondary water level too low
8236 XPS F Water control command not allowed now
8237 XPS F Bad water condition
8239 XPS F Unknown status command

AA83678 Service Manual ARL GenX 2-35


Chapter 2 GenX Description

8240 XPS F X-ray kV calibration slope error


8241 XPS F X-ray kV calibration offset error
8242 XPS F X-ray mA calibration slope error
8243 XPS F X-ray mA calibration offset error
8244 XPS F Status: X-ray arc rate too big
8245 XPS F XPS 12 bits A/D converter time-out
8246 XPS F Water cooler flow time-out
8247 XPS F Secondary water temperature safety open
8248 XPS F Spectrometer pressostat interlock signal defective
8249 XPS F Spectrometer tank cover interlock signal defective
8250 XPS F Primary chamber or PBF interlock signal defective
8251 XPS F X-ray tube position interlock signal defective
8252 XPS F X-ray tube connection interlock signal defective
8253 XPS F Secondary water temperature interlock signal defective
8256 XPS F XPS EEPROM defective
8257 XPS F Unknown X-ray ray power-supply type.
8268 XPS F X-ray generator identified is different of 4.2 kW type.
8269 XPS F X-ray generator identified is of 4.2 kW type.
8270 XPSLP F SSIO communication timeout.
8271 XPSLP F Serial AD12 conversion timeout.
8272 XPSLP F Serial AD12 auto-calibration timeout.
8273 XPS F X-ray “kV” out of regulation
8274 XPS F X-ray “mA” out of regulation
8275 XPSLP F X-ray tube current anode regulation error.
8276 XPSLP F X-ray tube filament current fatal limit reached.
8277 XPSLP F X-ray power supply interlock open.
8278 XPSLP F X-ray power supply interlock signal defective.
8279 XPSLP F X-ray power supply off.
8280 XPSLP F Unknown X-ray power supply fault.
8281 XPSLP F X-ray warning lamps broken.
8282 XPSLP F X-ray tube filament power supply fault.
8283 XPSLP F X-ray tube filament shorted.
8284 XPS F X-ray generator switched off without reason
8285 XPSLP F Secondary water resistivity too high
8286 XPSLP F X-ray power supply current limitation.
8287 XPSLP F X-ray power supply power limitation.
8288 XPSLP F Incoherence between the anode current and the cathode current.
8291 XPS F Invalid X-ray generator calibration value
8292 XPSC F Generators mains voltage not present, time-out
8293 XPSC F Generator 400V not > 250V, time-out
8294 XPSC F Generator 400V not > 380V, time-out
8295 XPSC F Generator, temperature IGBT too high
8296 XPSLP F Bad filament current
8297 XPSC F XPSC First 8 bits A/D converter error
8298 XPSC F XPSC Second 8 bits A/D converter error
8299 XPSC F XPSI 8 bits A/D converter error
8300 XPSC F XPSP 8 bits A/D converter error
8301 XPSC F XPSM 8 bits A/D converter error
8302 XPSC F Generator, temperature Mosfet M1 too high
8303 XPSC F Generator, temperature Mosfet M2 too high
8304 XPSC F Leakage current too high
8305 XPSC F Power off, issued by DSP; 400V too low (<350V)

2-36 Service Manual ARL GenX AA83678


GenX Description Chapter 2

8306 XPSC F Power off, issued by DSP; HV too high (>71kV)


8307 XPSC F Power off, issued by DSP; Current too high (>35A)
8308 XPSC F Mains voltage below 195 Vac (after 20 ms)
8309 XPSC F Mains voltage above 253 Vac (after 20 ms)
8310 XPSC F Unknown reply from DSP, issued by XPSC
8311 XPSC F Too many parameters in SSIO, issued by DSP
8312 XPSC F Unknown DSP Cmd, issued by DSP
8313 XPSC F PWMTRIP -> Power Off, issued by DSP
8314 XPSC F Generator, temperature XPSP too high
8315 XPSC F Generator, temperature filament too high
8316 XPSC F Generator, temperature HIV too high
8317 XPSC F Generator, temperature XPSC too high
8318 XPSC F Resistivity circuit open
8323 XPSC F Filament current sensor incoherency
8324 XPSC F Filament voltage sensor incoherency
8325 XPSC F Status : XPSI +5V too low
8326 XPSC F Status : XPSI +5V too high
8327 XPSC F Status : XPSC +5V too low
8328 XPSC F Status : XPSC +5V too high
8329 XPSC F Status : XPSP +5Vm too low
8330 XPSC F Status : XPSP +5Vm too high
8331 XPSC F Status : XPSM +5Vm too low
8332 XPSC F Status : XPSM +5Vm too high
8333 XPSC F Status : XPSI +12V too low
8334 XPSC F Status : XPSI +12V too high
8335 XPSC F Status : XPSC +12V too low
8336 XPSC F Status : XPSC +12V too high
8337 XPSC F Status : XPSM +12V too low
8338 XPSC F Status : XPSM +12V too high
8339 XPSC F Status : XPSI -12V too low
8340 XPSC F Status : XPSI -12V too high
8341 XPSC F Status : XPSC –12V too low
8342 XPSC F Status : XPSC -12V too high
8343 XPSC F Status : XPSM -12V too low
8344 XPSC F Status : XPSM -12V too high
8345 XPSC F Status : XPSI +24V too low
8346 XPSC F Status : XPSI +24V too high
8347 XPSC F Status : XPSC +24V too low
8348 XPSC F Status : XPSC +24V too high
8349 XPSC F Status : XPSP +24V too low
8350 XPSC F Status : XPSP +24V too high
8351 XPSC F Status : XPSM +2V5ref too low
8352 XPSC F Status : XPSM +2V5ref too high
8353 XPSC F Status : XPSC –5V too low
8354 XPSC F Status : XPSC –5V too high
8355 XPSC F Status : Mon 400V too low
8356 XPSC F Status : Mon 400V too high
8359 XPSC F Leakage Current Error
8360 XPSC F XPSI A/D Mon400V Error
8362 XPSC F XPSC temperature Board overflow
8363 XPSC F XPSC temperature Board underflow
8364 XPSC F XPSI temp. PowerMosfet M1 overflow

AA83678 Service Manual ARL GenX 2-37


Chapter 2 GenX Description

8365 XPSC F XPSI temp. PowerMosfet M1 underflow


8366 XPSC F XPSI temp. PowerMosfet M2 overflow
8367 XPSC F XPSI temp. PowerMosfet M2 underflow
8368 XPSC F XPSP temperature IGBT overflow
8369 XPSC F XPSP temperature IGBT underflow
8370 XPSC F XPSP temperature Board overflow
8371 XPSC F XPSP temperature Board underflow
8372 XPSC F XPSM temp. Filament overflow
8373 XPSC F XPSM temp. Filament underflow
8374 XPSC F XPSM temp. HIV overflow
8375 XPSC F XPSM temp. HIV underflow
8377 XPSC F X-Ray Lamp shorted
8378 XPSC F Status : XPSP –10Vrtm too low
8379 XPSC F Status : XPSP –10Vrtm too high
8380 XPSC F Status : XPSP +15Vrtm too low
8381 XPSC F Status : XPSP +15Vrtm too high
8382 XPSC F Neutral to Gnd too low
8383 XPSC F Neutral to Gnd too high
8384 XPSC F XPSI input current sensor error
8385 XPSC F External relay incoherency
8386 XPSC F Power On relay incoherency
8387 XPSC F No external relay feedback
8388 XPSC F No power relay feedback
8389 XPSC F X-ray negative leakage current error
8390 XPSC F X-ray positive leakage current error
8391 XPSC F Primary water already on
8393 XPSC F Mains voltage missing
8394 XPSC F High-Voltage Incoherency
8704 XPS W Initial primary water flow insufficient
8705 XPS W Primary water cooler OFF
8706 XPS W Spectrometer pressostat open
8707 XPS W Spectrometer tank cover safety open
8708 XPS W Primary chamber or PBF safety open
8709 XPS W X-ray tube position safety open
8710 XPS W X-ray tube connection safety open
8711 XPS W Illegal interlock
8712 XPS W Primary water inlet temperature gauge fault
8713 XPS W Primary water outlet temperature gauge fault
8714 XPS W Secondary water inlet temperature gauge fault
8715 XPS W Secondary water outlet temperature gauge fault
8716 XPS W Secondary water flow OFF
8717 XPS W Primary water flow insufficient
8718 XPS W Secondary water level gauge fault
8719 XPS W Secondary water level too low
8720 XPS W Secondary water resistivity driver fault
8721 XPS W Secondary water resistivity gauge fault
8722 XPS W Secondary water resistivity too low
8723 XPS W XPS A/D converter time-out
8725 XPS W X-ray kV mA measurement not calibrated
8726 XPS W X-ray kV calibration slope error
8727 XPS W X-ray kV calibration offset error
8728 XPS W X-ray mA calibration slope error

2-38 Service Manual ARL GenX AA83678


GenX Description Chapter 2

8729 XPS W X-ray mA calibration offset error


8730 XPS W X-ray kV mA in tolerance time-out
8731 XPS W Status: X-ray arc rate too big
8733 XPS W Low kV limit fault
8734 XPS W Leak current fault
8735 XPS W Secondary water temperature safety open
8738 XPS W De-ionised water tank almost empty.
8739 XPS W Unknown XPS status number
8740 XPS W XPS status sampling period too big
8741 XPS W Status: X-ray kV too low
8742 XPS W Status: X-ray kV too high
8743 XPS W Status: X-ray mA too low
8744 XPS W Status: X-ray mA too high
8745 XPS W Status: primary water input temperature too low
8746 XPS W Status: primary water input temperature too high
8747 XPS W Status: primary water output temperature too low
8748 XPS W Status: primary water output temperature too high
8749 XPS W Status: secondary water input temperature too low
8750 XPS W Status: 2nd water input temp too high
8751 XPS W Status: 2nd water output temp too low
8752 XPS W Status: 2nd water output temp too high
8753 XPS W Status: secondary water resistivity too low
8754 XPS W Status: 2nd water resistivity too high
8755 XPS W Status: X-tube filament current too low
8756 XPS W Status: X-tube filament current too high
8757 XPS W Status: XPS +15V too low
8758 XPS W Status: XPS +15V too high
8759 XPS W Status: XPS -15V too low
8760 XPS W Status: XPS -15V too high
8761 XPS W Status: XPS +24V too low
8762 XPS W Status: XPS +24V too high
8763 XPS W Status: XPS spare 4 too low
8764 XPS W Status: XPS spare 4 too high
8765 XPS W Status: XPS spare 5 too low
8766 XPS W Status: XPS spare 5 too high
8767 XPS W Status: XPS +5V too low
8768 XPS W Status: XPS +5V too high
8769 XPS W Status: XPS spare 7 too low
8770 XPS W Status: XPS spare 7 too high
8771 XPS W Status: X-ray total activity time too low
8772 XPS W Status: X-ray total activity time too high
8773 XPS W Status: arc rate parameter error
8774 XPS W Status: X-ray tube total arc occurrences too low
8775 XPS W Status: X-ray tube total arc occurrences too high
8776 XPS W 60 kV to 70 kV delay too low
8777 XPS W 60 kV to 70 kV delay too high
8800 XPS W SSIO Tx interrupt time-out
8801 XPS W XPS serial A/D converter EOC time-out
8802 XPS W SSIO first Rx interrupt time-out
8803 XPS W SSIO second Rx interrupt time-out
8804 XPS W XPS serial A/D converter channel incoherence
8805 XPS W Secondary water temperatures incoherence

AA83678 Service Manual ARL GenX 2-39


Chapter 2 GenX Description

8806 XPS W Desired 2nd water output temp too low


8807 XPS W Desired 2nd water output temp too high
8808 XPS W Water alarms inhibited
8817 XPSLP W Status: laboratory temperature too low
8818 XPSLP W Status: laboratory temperature too high
8819 XPSLP W X-ray tube temperature interlock
8820 XPSLP W Status: X-ray power supply air temperature too low.
8821 XPSLP W Status: X-ray power supply air temperature too high.
8822 XPS W De-ionised water temperature below tolerance
8823 XPS W De-ionised water temperature above tolerance
8824 XPSLP W Eeprom write failure.
8825 XPSLP W X-ray power supply interlock open.
8826 XPSLP W Default temperature offset used.
8827 XPSLP W Unknown firmware parameter command.
8828 XPSLP W Secondary water OFF time-out
8829 XPSLP W X-ray 1st. warning lamp broken.
8830 XPSLP W X-ray 2nd. warning lamp broken.
8831 XPSLP W Status: X-ray emission leakage current too low.
8832 XPSLP W Status: X-ray emission leakage current too high.
8833 XPSLP W X-ray power supply current limitation.
8834 XPSLP W X-ray power supply power limitation.
8835 XPSLP W Status: XPS power stage temperature too low.
8836 XPSLP W Status: XPS power stage temperature too high.
8837 XPSLP W XPS power stage temperature overflow.
8838 XPSLP W XPS power stage temperature underflow.
8839 XPSLP W Status: XPS +12V too low.
8840 XPSLP W Status: XPS +12V too high.
8841 XPSLP W Status: XPS -12V too low.
8842 XPSLP W Status: XPS -12V too high.
8843 XPSLP W Status: XPS +2.5Vref too low.
8844 XPSLP W Status: XPS +2.5Vref too high.
8845 XPSLP W Secondary water resistivity too high
8846 XPSLP W Incoherence between the anode current and the cathode current.
8849 XPSC W XPSC temperature overflow
8850 XPSC W XPSC temperature underflow
8851 XPSC W XPSI temperature1 overflow
8852 XPSC W XPSI temperature1 underflow
8853 XPSC W XPSI temperature2 overflow
8854 XPSC W XPSI temperature2 underflow
8855 XPSC W XPSP temperature1 overflow
8856 XPSC W XPSP temperature1 underflow
8857 XPSC W XPSP temperature2 overflow
8858 XPSC W XPSP temperature2 underflow
8859 XPSC W XPSM temperature1 overflow
8860 XPSC W XPSM temperature1 underflow
8861 XPSC W XPSM temperature2 overflow
8862 XPSC W XPSM temperature2 underflow
8863 XPSSP W Filament current limit
8864 XPSC W X-ray warning lamp broken
8865 XPSC W Leakage current > 500[µA]
8866 XPSSP W Cooling off time-out
8867 XPSC W Unknown reply from DSP

2-40 Service Manual ARL GenX AA83678


GenX Description Chapter 2

8885 XPSC W Status : XPSC -5V too low


8886 XPSC W Status : XPSC -5V too high
8887 XPSC W Status : XPSI +5V too low
8888 XPSC W Status : XPSI +5V too high
8889 XPSC W Status : XPSI +12V too low
8890 XPSC W Status : XPSI +12V too high
8891 XPSC W Status : XPSI +24V too low
8892 XPSC W Status : XPSI +24V too high
8893 XPSC W Status : XPSI -12V too low
8894 XPSC W Status : XPSI -12V too high
8895 XPSC W Status : XPSI Neutral to GND too low
8896 XPSC W Status : XPSI Neutral to GND too high
8897 XPSC W Status : XPSI Temp1(M1) too low
8898 XPSC W Status : XPSI Temp1(M1) too high
8899 XPSC W Status : XPSI Temp2(M2) too low
8900 XPSC W Status : XPSI Temp2(M2) too high
8901 XPSC W Status : XPSP +5Vm too low
8902 XPSC W Status : XPSP +5Vm too high
8903 XPSC W Status : XPSP +24V too low
8904 XPSC W Status : XPSP +24V too high
8905 XPSC W Status : XPSP -12Vrtm too low
8906 XPSC W Status : XPSP -12Vrtm too high
8907 XPSC W Status : XPSP +12Vrtm too low
8908 XPSC W Status : XPSP +12Vrtm too high
8909 XPSC W Status : XPSP Power temp.(IGBT) too low
8910 XPSC W Status : XPSP Power temp.(IGBT) too high
8911 XPSC W Status: XPSP Board temp. too low
8912 XPSC W Status: XPSP Board temp. too high
8913 XPSC W Status : XPSM +5Vm too low
8914 XPSC W Status : XPSM +5Vm too high
8915 XPSC W Status : XPSM +12V too low
8916 XPSC W Status : XPSM +12V too high
8917 XPSC W Status : XPSM -12V too low
8918 XPSC W Status : XPSM -12V too high
8919 XPSC W Status : XPSM HV divider diff. too low
8920 XPSC W Status : XPSM HV divider diff. too high
8921 XPSC W Status : XPSM Filament voltage too low
8922 XPSC W Status : XPSM Filament voltage too high
8923 XPSC W Status: XPSM Filament temp. too low
8924 XPSC W Status: XPSM Filament temp. too high
8925 XPSC W Status: XPSM HV temp. too low
8926 XPSC W Status: XPSM HV temp. too high
8927 XPSC W Status : Monitoring 400V too low
8928 XPSC W Status : Monitoring 400V too high
8929 XPSC W General status type
8930 XPSC W General status too low
8931 XPSC W General status too high
8932 XPSC W Default switches XPSC used
8933 XPSC W Primary water flow sensor error
8934 XPSC W Primary water already on
9472 XPS I Secondary water temperature regulation ON
9473 XPS I Secondary water temperature regulation OFF

AA83678 Service Manual ARL GenX 2-41


Chapter 2 GenX Description

9474 XPS I Water cooler present


9475 XPS I Water cooler absent
9476 XPS I X-ray generator selected is of 3.0 kW type.
9477 XPS I X-ray generator selected is of 3.6 kW type.
9478 XPS I X-ray generator selected is of 4.2 kW type.
9479 XPSLP I Eeprom write successful.
9480 XPSLP I X-ray generator selected is of 600W type.
9481 XPSLP I X-ray generator selected is of 1.2 kW type.
9484 XPSC I X-ray generator selected is of 3.6kW 60kV/120mA type.
9485 XPSC I X-ray generator selected is of 3.6kW 70kV/120mA type.
9486 XPSC I X-ray generator selected is of 4.2kW 60kV/120mA type.
9487 XPSC I X-ray generator selected is of 4.2kW 70kV/140mA type.
9488 XPSC I X-ray generator selected is of 1.2kW 60kV/60mA type.
9489 XPSC I Error X-ray generator selected type.
9728 XPS D No access to XSN network
9729 XPS D Tx interrupt time-out on XSN network
9730 XPS D Time-out on command size from XQM96
9731 XPS D Time-out on command data string from XQM96
9732 XPS D Time-out on command checksum from XQM96
9733 XPS D Time-out on broadcast command checksum
9734 XPS D String reception retry on checksum error
9735 XPS D Command Rx retry on checksum error
9736 XPS D String emission retry on checksum error
9737 XPS D Reply emission retry on checksum error
9738 XPS D Time-out on reply acknowledge
9739 XPS D No access to display
9740 XPS D Tx interrupt time-out on display
9741 XPS D Power-up on XPS
9742 XPS D X-ray tube spark
9743 XPS D Message size too large on command from XQM96
9744 XPS D XPS 12 bits A/D converter time-out
9745 XPSLP D Serial AD12 bits converter SSIO access denied.
9746 XPSLP D Serial AD8 bits converter SSIO access denied.
9747 XPSLP D Serial DA converter SSIO access denied.
9748 XPSLP D Serial Eeprom SSIO access denied.
9749 XPSLP D Retry on Eeprom write.
9750 XPSLP D X-ray tube filament current warning limit reached.
9751 XPSLP D X-ray tube anode current overflow.
9752 XPSLP D Retry on X-ray tube filament power supply fault.
9753 XPSLP D Retry on X-ray tube filament open.
9754 XPSLP D Retry on X-ray tube filament shorted.
9755 XPSLP D Retry on X-ray tube filament current fatal limit reached.
9756 XPSLP D Retry on secondary water flow sensor error.
9757 XPSLP D X-ray tube filament current regulation in limit.
9758 XPSSP D X-ray PS power limitation
9759 XPSC D X-ray leakage current too high (>1[mA])
9760 XPSC D X-ray leakage current (>500[µA])
9761 XPSC D XPSC First 8 bits A/D converter error
9762 XPSC D XPSC Second 8 bits A/D converter error
9763 XPSC D XPSI 8 bits A/D converter error
9764 XPSC D XPSP 8 bits A/D converter error
9765 XPSC D XPSM 8 bits A/D converter error

2-42 Service Manual ARL GenX AA83678


GenX Description Chapter 2

9766 XPSC D XPSC DSP Power-up


9767 XPSC D X-ray leakage current error
9768 XPSC D DSP ssio access denied
9769 XPSC D X-ray positive leakage current error
9770 XPSC D X-ray negative leakage current error
9771 XPSC D Primary water flow sensor error

AA83678 Service Manual ARL GenX 2-43


Chapter 2 GenX Description

Alarms list for firmware version XPSCSXRD 3.0.0


8192 XPSC GenXRF F Unknown command received from XSM
8193 XPSC GenXRF F Primary water flow sensor error
8194 XPSC GenXRF F Secondary water flow sensor error
8195 XPSC GenXRF F Secondary water pump error
8196 XPSC GenXRF F Desired voltage (kV) too low
8197 XPSC GenXRF F Desired voltage (kV) too high
8198 XPSC GenXRF F Desired current (mA) too low
8199 XPSC GenXRF F Desired current (mA) too high
8200 XPSC GenXRF F Desired power too high
8201 XPSC GenXRF F Illegal power change mode
8202 XPSC GenXRF F Illegal power change speed
8203 XPSC GenXRF F Illegal anode
8204 XPSC GenXRF F Spectrometer pressostat open
8205 XPSC GenXRF F Electronic rack safety switch open
8206 XPSC GenXRF F Spectro tank cover or x-ray tube position safety open
8207 XPSC GenXRF F Load lift cover or cassette magazine safety open
8208 XPSC GenXRF F X-ray ps high voltage connection safety open
8209 XPSC GenXRF F Illegal interlock
8210 XPSC GenXRF F Time-out on X-ray ON ack
8215 XPSC GenXRF F X-ray tube filament open
8222 XPSC GenXRF F Secondary water flow OFF
8223 XPSC GenXRF F Secondary water temperature too high
8225 XPSC GenXRF F Secondary water resistivity driver fault
8226 XPSC GenXRF F Secondary water resistivity gauge fault
8227 XPSC GenXRF F Secondary water resistivity too low
8230 XPSC GenXRF F Secondary water inlet temperature gauge fault
8231 XPSC GenXRF F Secondary water outlet temperature gauge fault
8235 XPSC GenXRF F Secondary water level too low
8236 XPSC GenXRF F Water control command not allowed now
8237 XPSC GenXRF F Bad water condition
8239 XPSC GenXRF F Unknown status command
8240 XPSC GenXRF F X-ray kV calibration slope error
8241 XPSC GenXRF F X-ray kV calibration offset error
8242 XPSC GenXRF F X-ray mA calibration slope error
8243 XPSC GenXRF F X-ray mA calibration offset error
8244 XPSC GenXRF F Status: X-ray arc rate too big
8247 XPSC GenXRF F Secondary water temperature safety open
8248 XPSC GenXRF F Spectrometer pressostat interlock signal defective
8249 XPSC GenXRF F Electronic rack interlock signal defective
8250 XPSC GenXRF F Spectrometer tank cover or x-ray tube interlock signal defective
8251 XPSC genXRF F Load lift cover or cassette magazine interlock signal defective
8252 XPSC GenXRF F X-ray power supply high voltage connection interlock signal defective
8253 XPSC GenXRF F Secondary water temperature interlock signal defective
8257 XPSC GenXRF F Unknown X-ray PS type
8270 XPSC GenXRF F SSIO communication timeout
8271 XPSC GenXRF F Serial AD12 conversion timeout
8272 XPSC GenXRF F Serial AD12 auto-calibration timeout
8273 XPSC GenXRF F X-ray 'kV' out of regulation
8274 XPSC GenXRF F X-ray 'mA' out of regulation
8276 XPSC GenXRF F X-ray tube filament current fatal limit reached
8277 XPSC GenXRF F X-ray power supply interlock open
8281 XPSC GenXRF F X-ray warning lamps broken

2-44 Service Manual ARL GenX AA83678


GenX Description Chapter 2

8282 XPSC GenXRF F X-ray tube filament power supply fault


8283 XPSC GenXRF F X-ray tube filament shorted
8285 XPSC GenXRF F Secondary water resistivity too high
8291 XPSC GenXRF F Invalid X-ray generator calibration value
8292 XPSC GenXRF F Generator mains voltage not present
8293 XPSC GenXRF F Generator 400V not > 250V
8294 XPSC GenXRF F Generator 400V not > 320V
8295 XPSC GenXRF F Generator, Temperature IGBT too high
8297 XPSC GenXRF F XPSC First 8 bits A/D converter error
8298 XPSC GenXRF F XPSC Second 8 bits A/D converter error
8299 XPSC GenXRF F XPSI 8 bits A/D converter error
8300 XPSC GenXRF F XPSP 8 bits A/D converter error
8301 XPSC GenXRF F XPSM 8 bits A/D converter error
8302 XPSC GenXRF F Generator temperature MOSFET M1 too high
8303 XPSC GenXRF F Generator temperature MOSFET M2 too high
8304 XPSC GenXRF F Leakage current too high
8305 XPSC GenXRF F Power Off issued by DSP; 400V too low (<320V)
8306 XPSC GenXRF F Power Off issued by DSP; HV too high (>71kV)
8307 XPSC GenXRF F Power Off issued by DSP; Current too high (>40A)
8308 XPSC GenXRF F Mains voltage below 195 Vac (after 20 ms)
8309 XPSC GenXRF F Mains voltage above 253 Vac (after 20 ms)
8310 XPSC GenXRF F Unknown reply from DSP
8311 XPSC GenXRF F Too many parameters in SSIO
8312 XPSC GenXRF F Unknown DSP Cmd
8313 XPSC GenXRF F PWMTRIP -> Power Off
8314 XPSC GenXRF F Generator temperature XPSP too high
8315 XPSC GenXRF F Generator temperature Filament too high
8316 XPSC GenXRF F Generator temperature Hv too high
8317 XPSC GenXRF F Generator temperature XPSC too high
8318 XPSC GenXRF F Resistivity circuit Open
8323 XPSC GenXRF F Filament current sensor incoherency
8324 XPSC GenXRF F Filament voltage sensor incoherency
8325 XPSC GenXRF F Status: XPSI +5V too low
8326 XPSC GenXRF F Status: XPSI +5V too high
8327 XPSC GenXRF F Status: XPSC +5V too low
8328 XPSC GenXRF F Status: XPSC +5V too high
8329 XPSC GenXRF F Status: XPSP +5Vm too low
8330 XPSC GenXRF F Status: XPSP +5Vm too high
8331 XPSC GenXRF F Status: XPSM +5Vm too low
8332 XPSC GenXRF F Status: XPSM +5Vm too high
8333 XPSC GenXRF F Status: XPSI +12V too low
8334 XPSC GenXRF F Status: XPSI +12V too high
8335 XPSC GenXRF F Status: XPSC +12V too low
8336 XPSC GenXRF F Status: XPSC +12V too high
8337 XPSC GenXRF F Status: XPSM +12V too low
8338 XPSC GenXRF F Status: XPSM +12V too high
8339 XPSC GenXRF F Status: XPSI -12V too low
8340 XPSC GenXRF F Status: XPSI -12V too high
8341 XPSC GenXRF F Status: XPSC -12V too low
8342 XPSC GenXRF F Status: XPSC -12V too high
8343 XPSC GenXRF F Status: XPSM -12V too low
8344 XPSC GenXRF F Status: XPSM -12V too high
8345 XPSC GenXRF F Status: XPSI +24V too low
8346 XPSC GenXRF F Status: XPSI +24V too high

AA83678 Service Manual ARL GenX 2-45


Chapter 2 GenX Description

8347 XPSC GenXRF F Status: XPSC +24V too low


8348 XPSC GenXRF F Status: XPSC +24V too high
8349 XPSC GenXRF F Status: XPSP +24V too low
8350 XPSC GenXRF F Status: XPSP +24V too high
8353 XPSC GenXRF F Status: XPSC -5V too low
8354 XPSC GenXRF F Status: XPSC -5V too high
8360 XPSC GenXRF F XPSI AD Mon400V Error
8362 XPSC GenXRF F XPSC temperature Board overflow
8363 XPSC GenXRF F XPSC temperature Board underflow
8364 XPSC GenXRF F XPSI temp. PowerMosfet M1 overflow
8365 XPSC GenXRF F XPSI temp. PowerMosfet M1 underflow
8366 XPSC GenXRF F XPSI temp. PowerMosfet M2 overflow
8367 XPSC GenXRF F XPSI temp. PowerMosfet M2 underflow
8368 XPSC GenXRF F XPSP temperature IGBT overflow
8369 XPSC GenXRF F XPSP temperature IGBT underflow
8370 XPSC GenXRF F XPSP temperature Board overflow
8371 XPSC GenXRF F XPSP temperature Board underflow
8372 XPSC GenXRF F XPSM temp. Filament overflow
8373 XPSC GenXRF F XPSM temp. Filament underflow
8374 XPSC GenXRF F XPSM temp. Hv overflow
8375 XPSC GenXRF F XPSM temp. Hv underflow
8377 XPSC GenXRF F X-Ray Lamp Shorted
8378 XPSC GenXRF F Status: XPSP -10Vrtm too low
8379 XPSC GenXRF F Status: XPSP -10Vrtm too high
8380 XPSC GenXRF F Status: XPSP +15Vrtm too low
8381 XPSC GenXRF F Status: XPSP +15Vrtm too high
8382 XPSC GenXRF F Neutral to Gnd too low
8383 XPSC GenXRF F Neutral to Gnd too high
8384 XPSC GenXRF F XPSI input current sensor error
8385 XPSC GenXRF F External relay incoherency
8386 XPSC GenXRF F Power On relay incoherency
8387 XPSC GenXRF F No external relay feedback
8388 XPSC GenXRF F No Power relay feedback
8389 XPSC GenXRF F X-ray negative leakeage current error
8390 XPSC GenXRF F X-ray positive leakeage current error
8391 XPSC GenXRF F Primary water already on
8393 XPSC GenXRF F Mains voltage missing
8394 XPSC GenXRF F High-Voltage Incoherency
8448 XPSC GenXRD F Unknown command received from XSM
8449 XPSC GenXRD F Primary water flow sensor error
8450 XPSC GenXRD F Secondary water flow sensor error
8451 XPSC GenXRD F Secondary water pump error
8452 XPSC GenXRD F Desired voltage (kV) too low
8453 XPSC GenXRD F Desired voltage (kV) too high
8454 XPSC GenXRD F Desired current (mA) too low
8455 XPSC GenXRD F Desired current (mA) too high
8456 XPSC GenXRD F Desired power too high
8457 XPSC GenXRD F Illegal power change mode
8458 XPSC GenXRD F Illegal power change speed
8459 XPSC GenXRD F Illegal anode
8460 XPSC GenXRD F Spectrometer pressostat open
8461 XPSC GenXRD F Electronic rack safety switch open
8462 XPSC GenXRD F Spectro tank cover or xray tube safety open
8463 XPSC GenXRD F Load lift cover or cassette magazine safety open

2-46 Service Manual ARL GenX AA83678


GenX Description Chapter 2

8464 XPSC GenXRD F X-ray ps high voltage connection safety open


8465 XPSC GenXRD F Illegal interlock
8466 XPSC GenXRD F Time-out on X-ray ON ack
8467 XPSC GenXRD F X-ray tube filament open
8468 XPSC GenXRD F Secondary water flow OFF
8469 XPSC GenXRD F Secondary water temperature too high
8470 XPSC GenXRD F Secondary water resistivity driver fault
8471 XPSC GenXRD F Secondary water resistivity gauge fault
8472 XPSC GenXRD F Secondary water resistivity too low
8473 XPSC GenXRD F Secondary water inlet temperature gauge fault
8474 XPSC GenXRD F Secondary water outlet temperature gauge fault
8475 XPSC GenXRD F Secondary water level too low
8476 XPSC GenXRD F Water control command not allowed now
8477 XPSC GenXRD F Bad water condition
8478 XPSC GenXRD F Unknown status command
8479 XPSC GenXRD F X-ray kV calibration slope error
8480 XPSC GenXRD F X-ray kV calibration slope error
8481 XPSC GenXRD F X-ray mA calibration slope error
8482 XPSC GenXRD F X-ray mA calibration offset error
8483 XPSC GenXRD F Status: X-ray arc rate too big
8484 XPSC GenXRD F Secondary water temperature safety open
8485 XPSC GenXRD F Spectro pressostat interlock signal defective
8486 XPSC GenXRD F Electronic rack interlock signal defective
8487 XPSC GenXRD F Interlock signal defective
8488 XPSC GenXRD F Interlock signal defective
8489 XPSC GenXRD F Interlock signal defective
8490 XPSC GenXRD F Interlock signal defective
8491 XPSC GenXRD F Unknown X-ray PS type
8492 XPSC GenXRD F SSIO communication timeout
8493 XPSC GenXRD F Serial AD12 conversion timeout
8494 XPSC GenXRD F Serial AD12 auto-calibration timeout
8495 XPSC GenXRD F X-ray 'kV' out of regulation
8496 XPSC GenXRD F X-ray 'mA' out of regulation
8497 XPSC GenXRD F X-ray power supply interlock open
8498 XPSC GenXRD F X-ray warning lamps broken
8499 XPSC GenXRD F X-Ray Lamp Shorted
8500 XPSC GenXRD F X-ray tube filament power supply fault
8501 XPSC GenXRD F Invalid X-ray generator calibration value
8502 XPSC GenXRD F Generator mains voltage not present
8503 XPSC GenXRD F Generator 400V not > 250V
8504 XPSC GenXRD F Generator 400V not > 320V
8505 XPSC GenXRD F Generator temperature IGBT too high
8506 XPSC GenXRD F XPSC First 8 bits A/D converter error
8507 XPSC GenXRD F XPSC Second 8 bits A/D converter error
8508 XPSC GenXRD F XPSI 8 bits A/D converter error
8509 XPSC GenXRD F XPSP 8 bits A/D converter error
8510 XPSC GenXRD F XPSM 8 bits A/D converter error
8511 XPSC GenXRD F Generator temperature MOSFET M1 too high
8512 XPSC GenXRD F Generator temperature MOSFET M2 too high
8513 XPSC GenXRD F Power Off issued by DSP; 400V too low (<320V)
8514 XPSC GenXRD F Power Off issued by DSP; HV too high (>71kV)
8515 XPSC GenXRD F Power Off issued by DSP; Current too high (>40A)
8516 XPSC GenXRD F Mains voltage below 195 Vac (after 20 ms)
8517 XPSC GenXRD F Mains voltage above 253 Vac (after 20 ms)

AA83678 Service Manual ARL GenX 2-47


Chapter 2 GenX Description

8518 XPSC GenXRD F Unknown reply from DSP


8519 XPSC GenXRD F Too many parameters in SSIO
8520 XPSC GenXRD F Unknown DSP Cmd
8521 XPSC GenXRD F PWMTRIP -> Power Off
8522 XPSC GenXRD F Generator temperature XPSP too high
8523 XPSC GenXRD F Generator temperature Filament too high
8524 XPSC GenXRD F Generator temperature Hv too high
8525 XPSC GenXRD F Generator temperature XPSC too high
8526 XPSC GenXRD F Resistivity circuit Open
8527 XPSC GenXRD F Filament current sensor incoherency
8528 XPSC GenXRD F Filament voltage sensor incoherency
8529 XPSC GenXRD F Status: XPSI +5V too low
8530 XPSC GenXRD F Status: XPSI +5V too high
8531 XPSC GenXRD F Status: XPSC +5V too low
8532 XPSC GenXRD F Status: XPSC +5V too high
8533 XPSC GenXRD F Status: XPSP +5Vm too low
8534 XPSC GenXRD F Status: XPSP +5Vm too high
8535 XPSC GenXRD F Status: XPSM +5Vm too low
8536 XPSC GenXRD F Status: XPSM +5Vm too high
8537 XPSC GenXRD F Status: XPSI +12V too low
8538 XPSC GenXRD F Status: XPSI +12V too high
8539 XPSC GenXRD F Status: XPSC +12V too low
8540 XPSC GenXRD F Status: XPSC +12V too high
8541 XPSC GenXRD F Status: XPSM +12V too low
8542 XPSC GenXRD F Status: XPSM +12V too high
8543 XPSC GenXRD F Status: XPSI -12V too low
8544 XPSC GenXRD F Status: XPSI -12V too high
8545 XPSC GenXRD F Status: XPSC -12V too low
8546 XPSC GenXRD F Status: XPSC -12V too high
8547 XPSC GenXRD F Status: XPSM -12V too low
8548 XPSC GenXRD F Status: XPSM -12V too high
8549 XPSC GenXRD F Status: XPSI +24V too low
8550 XPSC GenXRD F Status: XPSI +24V too high
8551 XPSC GenXRD F Status: XPSC +24V too low
8552 XPSC GenXRD F Status: XPSC +24V too high
8553 XPSC GenXRD F Status: XPSP +24V too low
8554 XPSC GenXRD F Status: XPSP +24V too high
8555 XPSC GenXRD F Status: XPSC -5V too low
8556 XPSC GenXRD F Status: XPSC -5V too high
8557 XPSC GenXRD F XPSI AD Mon400V Error
8558 XPSC GenXRD F XPSC temperature Board overflow
8559 XPSC GenXRD F XPSC temperature Board underflow
8560 XPSC GenXRD F XPSI temp. PowerMosfet M1 overflow
8561 XPSC GenXRD F XPSI temp. PowerMosfet M1 underflow
8562 XPSC GenXRD F XPSI temp. PowerMosfet M2 overflow
8563 XPSC GenXRD F XPSI temp. PowerMosfet M2 underflow
8564 XPSC GenXRD F XPSP temperature IGBT overflow
8565 XPSC GenXRD F XPSP temperature IGBT underflow
8566 XPSC GenXRD F XPSP temperature Board overflow
8567 XPSC GenXRD F XPSP temperature Board underflow
8568 XPSC GenXRD F XPSM temp. Filament overflow
8569 XPSC GenXRD F XPSM temp. Filament underflow
8570 XPSC GenXRD F XPSM temp. Hv overflow
8571 XPSC GenXRD F XPSM temp. Hv underflow

2-48 Service Manual ARL GenX AA83678


GenX Description Chapter 2

8572 XPSC GenXRD F Status: XPSP -10Vrtm too low


8573 XPSC GenXRD F Status: XPSP -10Vrtm too high
8574 XPSC GenXRD F Status: XPSP +15Vrtm too low
8575 XPSC GenXRD F Status: XPSP +15Vrtm too high
8576 XPSC GenXRD F Neutral to Gnd too low
8577 XPSC GenXRD F Neutral to Gnd too high
8578 XPSC GenXRD F XPSI input current sensor error
8579 XPSC GenXRD F External relay incoherency
8580 XPSC GenXRD F Power On relay incoherency
8581 XPSC GenXRD F No external relay feedback
8582 XPSC GenXRD F No Power relay feedback
8583 XPSC GenXRD F Primary water already on
8584 XPSC GenXRD F Max filament current limit reached
8585 XPSC GenXRD F Mains voltage missing
8586 XPSC GenXRD F High-Voltage Incoherency
8704 XPSC GenXRF W Initial primary water flow insufficient
8706 XPSC GenXRF W Spectrometer pressostat open
8707 XPSC GenXRF W Electronic rack safety switch open
8708 XPSC GenXRF W Spectrometer tank cover or x-ray tube position safety open
8709 XPSC GenXRF W Load lift cover or cassette magazine safety open
8710 XPSC GenXRF W X-ray power supply high voltage connection safety open
8711 XPSC GenXRF W Secondary water temperature safety open
8714 XPSC GenXRF W Secondary water inlet temperature gauge fault
8715 XPSC GenXRF W Secondary water outlet temperature gauge fault
8716 XPSC GenXRF W Secondary water flow OFF
8719 XPSC GenXRF W Secondary water level too low
8720 XPSC GenXRF W Secondary water resistivity driver fault
8721 XPSC GenXRF W Secondary water resistivity gauge fault
8722 XPSC GenXRF W Secondary water resistivity too low
8723 XPSC GenXRF W XPSC A/D converter time-out
8725 XPSC GenXRF W X-ray kV mA measurement not calibrated
8726 XPSC GenXRF W X-ray kV calibration slope error
8727 XPSC GenXRF W X-ray kV calibration offset error
8728 XPSC GenXRF W X-ray mA calibration slope error
8729 XPSC GenXRF W X-ray mA calibration offset error
8730 XPSC GenXRF W X-ray kV mA in tolerance time-out
8731 XPSC GenXRF W Status: X-ray arc rate too big
8735 XPSC GenXRF W Secondary water temperature safety open
8738 XPSC GenXRF W De-ionized water tank almost empty
8739 XPSC GenXRF W Unknown XPSC status number
8740 XPSC GenXRF W XPSC status sampling period too big
8741 XPSC GenXRF W Status: X-ray kV too low
8742 XPSC GenXRF W Status: X-ray kV too high
8743 XPSC GenXRF W Status: X-ray mA too low
8744 XPSC GenXRF W Status: X-ray mA too high
8745 XPSC GenXRF W Status: primary water input temperature too low
8746 XPSC GenXRF W Status: primary water input temperature too high
8747 XPSC GenXRF W Status: primary water output temperature too low
8748 XPSC GenXRF W Status: primary water output temperature too high
8749 XPSC GenXRF W Status: secondary water input temperature too low
8750 XPSC GenXRF W Status: 2nd water input temp too high
8751 XPSC GenXRF W Status: 2nd water output temp too low
8752 XPSC GenXRF W Status: 2nd water output temp too high
8753 XPSC GenXRF W Status: secondary water resistivity too low

AA83678 Service Manual ARL GenX 2-49


Chapter 2 GenX Description

8754 XPSC GenXRF W Status: 2nd water resistivity too high


8755 XPSC GenXRF W Status: X-tube filament current too low
8756 XPSC GenXRF W Status: X-tube filament current too high
8761 XPSC GenXRF W Status: XPSC +24V too low
8762 XPSC GenXRF W Status: XPSC +24V too high
8767 XPSC GenXRF W Status: XPSC +5V too low
8768 XPSC GenXRF W Status: XPSC +5V too high
8771 XPSC GenXRF W Status: X-ray total activity time too low
8772 XPSC GenXRF W Status: X-ray total activity time too high
8773 XPSC GenXRF W Status: arc rate parameter error
8774 XPSC GenXRF W Status: X-ray tube total arc occurrences too low
8775 XPSC GenXRF W Status: X-ray tube total arc occurrences too high
8776 XPSC GenXRF W 60 kV to 70 kV delay too low
8777 XPSC GenXRF W 60 kV to 70 kV delay too high
8805 XPSC GenXRF W Secondary water temperatures incoherency
8806 XPSC GenXRF W Desired 2nd water output temp too low
8807 XPSC GenXRF W Desired 2nd water output temp too high
8817 XPSC GenXRF W Laboratory temperature too low
8818 XPSC GenXRF W Laboratory temperature too high
8820 XPSC GenXRF W Status: X-ray power supply air temperature too low
8821 XPSC GenXRF W Status: X-ray power supply air temperature too high
8822 XPSC GenXRF W De-ionised water temp. too low
8823 XPSC GenXRF W De-ionised water temp. too high
8824 XPSC GenXRF W Eeprom write failure
8825 XPSC GenXRF W X-ray power supply breaker off
8827 XPSC GenXRF W Unknown firmware parameter command
8828 XPSC GenXRF W Secondary water OFF time-out
8829 XPSC GenXRF W X-ray 1st. warning lamp broken
8830 XPSC GenXRF W X-ray 2nd. warning lamp broken
8831 XPSC GenXRF W Status: X-ray emission leakage current too low
8832 XPSC GenXRF W Status: X-ray emission leakage current too high
8835 XPSC GenXRF W Status: XPSC power stage temperature too low
8836 XPSC GenXRF W Status: XPSC power stage temperature too high
8839 XPSC GenXRF W Status: XPSC +12V too low
8840 XPSC GenXRF W Status: XPSC +12V too high
8841 XPSC GenXRF W Status: XPSC -12V too low
8842 XPSC GenXRF W Status: XPSC -12V too high
8843 XPSC GenXRF W Status: XPSC +2.5Vref too low
8844 XPSC GenXRF W Status: XPSC +2.5Vref too high
8864 XPSC GenXRF W X-ray warning lamp broken
8865 XPSC GenXRF W Leakage Current > 1000[uA]
8867 XPSC GenXRF W Unknown reply from DSP
8885 XPSC GenXRF W Status: XPSC -5V too low
8886 XPSC GenXRF W Status: XPSC -5V too high
8887 XPSC GenXRF W Status: XPSI +5V too low
8888 XPSC GenXRF W Status: XPSI +5V too high
8889 XPSC GenXRF W Status: XPSI +12V too low
8890 XPSC GenXRF W Status: XPSI +12V too high
8891 XPSC GenXRF W Status: XPSI +24V too low
8892 XPSC GenXRF W Status: XPSI +24V too high
8893 XPSC GenXRF W Status: XPSI -12V too low
8894 XPSC GenXRF W Status: XPSI -12V too high
8895 XPSC GenXRF W Status: XPSI Neutral to GND too low
8896 XPSC GenXRF W Status: XPSI Neutral to GND too high

2-50 Service Manual ARL GenX AA83678


GenX Description Chapter 2

8897 XPSC GenXRF W Status: XPSI Temp1(M1) too low


8898 XPSC GenXRF W Status: XPSI Temp1(M1) too high
8899 XPSC GenXRF W Status: XPSI Temp2(M2) too low
8900 XPSC GenXRF W Status: XPSI Temp2(M2) too high
8901 XPSC GenXRF W Status: XPSP +5Vm too low
8902 XPSC GenXRF W Status: XPSP +5Vm too high
8903 XPSC GenXRF W Status: XPSP +24V too low
8904 XPSC GenXRF W Status: XPSP +24V too high
8905 XPSC GenXRF W Status: XPSP -12Vrtm too low
8906 XPSC GenXRF W Status: XPSP -12Vrtm too high
8907 XPSC GenXRF W Status: XPSP +12Vrtm too low
8908 XPSC GenXRF W Status: XPSP +12Vrtm too high
8909 XPSC GenXRF W Status: XPSP Power temp.(IGBT) too low
8910 XPSC GenXRF W Status: XPSP Power temp.(IGBT) too high
8911 XPSC GenXRF W Status: XPSP Board temp. too low
8912 XPSC GenXRF W Status: XPSP Board temp. too high
8913 XPSC GenXRF W Status: XPSM +5Vm too low
8914 XPSC GenXRF W Status: XPSM +5Vm too high
8915 XPSC GenXRF W Status: XPSM +12V too low
8916 XPSC GenXRF W Status: XPSM +12V too high
8917 XPSC GenXRF W Status: XPSM -12V too low
8918 XPSC GenXRF W Status: XPSM -12V too high
8919 XPSC GenXRF W Status: XPSM Hv divider diff. too low
8920 XPSC GenXRF W Status: XPSM Hv divider diff. too high
8921 XPSC GenXRF W Status: XPSM Filament voltage too low
8922 XPSC GenXRF W Status: XPSM Filament voltage too high
8923 XPSC GenXRF W Status: XPSM Filament temp. too low
8924 XPSC GenXRF W Status: XPSM Filament temp. too high
8925 XPSC GenXRF W Status: XPSM Hv temp. too low
8926 XPSC GenXRF W Status: XPSM Hv temp. too high
8927 XPSC GenXRF W Status: Monitoring 400V too low
8928 XPSC GenXRF W Status: Monitoring 400V too high
8933 XPSC GenXRF W Primary water flow sensor error
8934 XPSC GenXRF W Primary water already on
9088 XPSC GenXRD W Initial primary water flow insufficient
9089 XPSC GenXRD W Spectrometer pressostat open
9090 XPSC GenXRD W Electronic rack safety switch open
9091 XPSC GenXRD W Spectro tank cover or x-ray tube safety open
9092 XPSC GenXRD W Load lift cover or cassette magazine safety open
9093 XPSC GenXRD W X-ray ps high voltage connection safety open
9094 XPSC GenXRD W Secondary water temperature safety open
9095 XPSC GenXRD W Secondary water inlet temperature gauge fault
9096 XPSC GenXRD W Secondary water outlet temperature gauge fault
9097 XPSC GenXRD W Secondary water flow OFF
9098 XPSC GenXRD W Secondary water level too low
9099 XPSC GenXRD W Secondary water resistivity driver fault
9100 XPSC GenXRD W Secondary water resistivity gauge fault
9101 XPSC GenXRD W Secondary water resistivity too low
9102 XPSC GenXRD W XPSC A/D converter time-out
9103 XPSC GenXRD W X-ray kV mA measurement not calibrated
9104 XPSC GenXRD W X-ray kV calibration slope error
9105 XPSC GenXRD W X-ray kV calibration offset error
9106 XPSC GenXRD W X-ray mA calibration slope error
9107 XPSC GenXRD W X-ray mA calibration offset error

AA83678 Service Manual ARL GenX 2-51


Chapter 2 GenX Description

9108 XPSC GenXRD W X-ray kV mA in tolerance time-out


9109 XPSC GenXRD W Status: X-ray arc rate too big
9110 XPSC GenXRD W Secondary water temperature safety open
9111 XPSC GenXRD W De-ionized water tank almost empty
9112 XPSC GenXRD W Unknown XPSC status number
9113 XPSC GenXRD W XPSC status sampling period too big
9114 XPSC GenXRD W Status: X-ray kV too low
9115 XPSC GenXRD W Status: X-ray kV too high
9116 XPSC GenXRD W Status: X-ray mA too low
9117 XPSC GenXRD W Status: X-ray mA too high
9118 XPSC GenXRD W Status: primary water input temperature too low
9119 XPSC GenXRD W Status: primary water input temperature too high
9120 XPSC GenXRD W Status: primary water output temperature too low
9121 XPSC GenXRD W Status: primary water output temperature too high
9122 XPSC GenXRD W Status: secondary water input temperature too low
9123 XPSC GenXRD W Status: 2nd water input temp too high
9124 XPSC GenXRD W Status: 2nd water output temp too low
9125 XPSC GenXRD W Status: 2nd water output temp too high
9126 XPSC GenXRD W Status: secondary water resistivity too low
9127 XPSC GenXRD W Status: 2nd water resistivity too high
9128 XPSC GenXRD W Status: X-tube filament current too low
9129 XPSC GenXRD W Status: X-tube filament current too high
9130 XPSC GenXRD W Status: XPSC +24V too low
9131 XPSC GenXRD W Status: XPSC +24V too high
9132 XPSC GenXRD W Status: XPSC +5V too low
9133 XPSC GenXRD W Status: XPSC +5V too high
9134 XPSC GenXRD W Status: X-ray total activity time too low
9135 XPSC GenXRD W Status: X-ray total activity time too high
9136 XPSC GenXRD W Status: arc rate parameter error
9137 XPSC GenXRD W Status: X-ray tube total arc occurrences too low
9138 XPSC GenXRD W Status: X-ray tube total arc occurrences too high
9139 XPSC GenXRD W 60 kV to 70 kV delay too low
9140 XPSC GenXRD W 60 kV to 70 kV delay too high
9141 XPSC GenXRD W Secondary water temperatures incoherency
9142 XPSC GenXRD W Desired 2nd water output temp too low
9143 XPSC GenXRD W Desired 2nd water output temp too high
9144 XPSC GenXRD W De-ionised water temp. too low
9145 XPSC GenXRD W De-ionised water temp. too high
9146 XPSC GenXRD W Eeprom write failure
9147 XPSC GenXRD W X-ray power supply breaker off
9148 XPSC GenXRD W Unknown firmware parameter command
9149 XPSC GenXRD W Secondary water OFF time-out
9150 XPSC GenXRD W X-ray 1st. warning lamp broken
9151 XPSC GenXRD W X-ray 2nd. warning lamp broken
9152 XPSC GenXRD W Status: X-ray emission leakage current too low
9153 XPSC GenXRD W Status: X-ray emission leakage current too high
9154 XPSC GenXRD W Status: XPSC power stage temperature too low
9155 XPSC GenXRD W Status: XPSC power stage temperature too high
9156 XPSC GenXRD W Status: XPSC +12V too low
9157 XPSC GenXRD W Status: XPSC +12V too high
9158 XPSC GenXRD W Status: XPSC -12V too low
9159 XPSC GenXRD W Status: XPSC -12V too high
9160 XPSC GenXRD W Status: XPSC +2.5Vref too low
9161 XPSC GenXRD W Status: XPSC +2.5Vref too high

2-52 Service Manual ARL GenX AA83678


GenX Description Chapter 2

9162 XPSC GenXRD W Leakage Current > 1000[uA]


9163 XPSC GenXRD W Unknown reply from DSP
9164 XPSC GenXRD W Status: XPSC -5V too low
9165 XPSC GenXRD W Status: XPSC -5V too high
9166 XPSC GenXRD W Status: XPSI +5V too low
9167 XPSC GenXRD W Status: XPSI +5V too high
9168 XPSC GenXRD W Status: XPSI +12V too low
9169 XPSC GenXRD W Status: XPSI +12V too high
9170 XPSC GenXRD W Status: XPSI +24V too low
9171 XPSC GenXRD W Status: XPSI +24V too high
9172 XPSC GenXRD W Status: XPSI -12V too low
9173 XPSC GenXRD W Status: XPSI -12V too high
9174 XPSC GenXRD W Status: XPSI Neutral to GND too low
9175 XPSC GenXRD W Status: XPSI Neutral to GND too high
9176 XPSC GenXRD W Status: XPSI Temp1(M1) too low
9177 XPSC GenXRD W Status: XPSI Temp1(M1) too high
9178 XPSC GenXRD W Status: XPSI Temp2(M2) too low
9179 XPSC GenXRD W Status: XPSI Temp2(M2) too high
9180 XPSC GenXRD W Status: XPSP +5Vm too low
9181 XPSC GenXRD W Status: XPSP +5Vm too high
9182 XPSC GenXRD W Status: XPSP +24V too low
9183 XPSC GenXRD W Status: XPSP +24V too high
9184 XPSC GenXRD W Status: XPSP -12Vrtm too low
9185 XPSC GenXRD W Status: XPSP -12Vrtm too high
9186 XPSC GenXRD W Status: XPSP +12Vrtm too low
9187 XPSC GenXRD W Status: XPSP +12Vrtm too high
9188 XPSC GenXRD W Status: XPSP Power temp.(IGBT) too low
9189 XPSC GenXRD W Status: XPSP Power temp.(IGBT) too high
9190 XPSC GenXRD W Status: XPSP Board temp. too low
9191 XPSC GenXRD W Status: XPSP Board temp. too high
9192 XPSC GenXRD W Status: XPSM +5Vm too low
9193 XPSC GenXRD W Status: XPSM +5Vm too high
9194 XPSC GenXRD W Status: XPSM +12V too low
9195 XPSC GenXRD W Status: XPSM +12V too high
9196 XPSC GenXRD W Status: XPSM -12V too low
9197 XPSC GenXRD W Status: XPSM -12V too high
9198 XPSC GenXRD W Status: XPSM Hv divider diff. too low
9199 XPSC GenXRD W Status: XPSM Hv divider diff. too high
9200 XPSC GenXRD W Status: XPSM Filament voltage too low
9201 XPSC GenXRD W Status: XPSM Filament voltage too high
9202 XPSC GenXRD W Status: XPSM Filament temp. too low
9203 XPSC GenXRD W Status: XPSM Filament temp. too high
9204 XPSC GenXRD W Status: XPSM Hv temp. too low
9205 XPSC GenXRD W Status: XPSM Hv temp. too high
9206 XPSC GenXRD W Status: Monitoring 400V too low
9207 XPSC GenXRD W Status: Monitoring 400V too high
9208 XPSC GenXRD W Primary water flow sensor error
9209 XPSC GenXRD W Primary water already on
9474 XPSC GenXRF I Water cooler present
9475 XPSC GenXRF I Water cooler absent
9479 XPSC GenXRF I Eeprom write successful
9484 XPSC GenXRF I X-ray generator selected type 3.6kW 60kV/120mA
9485 XPSC GenXRF I X-ray generator selected type 3.6kW 70kV/120mA
9486 XPSC GenXRF I X-ray generator selected type 4.2kW 60kV/120mA

AA83678 Service Manual ARL GenX 2-53


Chapter 2 GenX Description

9487 XPSC GenXRF I X-ray generator selected type 4.2kW 70kV/140mA


9488 XPSC GenXRF I X-ray generator selected type 1.2kW 60kV/60mA
9489 XPSC GenXRF I Error X-ray generator selected type.
9494 XPSC GenXRF I Default switch XPSC used
9495 XPSC GenXRF I X-ray generator selected type 2.5kW 60kV/100mA
9600 XPSC GenXRD I X-ray generator selected type 2.2kW 60kV/75mA
9601 XPSC GenXRD I Water cooler present
9602 XPSC GenXRD I Water cooler absent
9603 XPSC GenXRD I Eeprom write successful
9604 XPSC GenXRD I Error X-ray generator selected type.
9728 XPSC GenXRF D No access to XSN network
9729 XPSC GenXRF D Tx interrupt time-out on XSN network
9730 XPSC GenXRF D Time-out on command size from XSM
9731 XPSC GenXRF D Time-out on command data string from XSM
9732 XPSC GenXRF D Time-out on command checksum from XSM
9733 XPSC GenXRF D Time-out on broadcast command checksum
9734 XPSC GenXRF D String reception retry on checksum error
9735 XPSC GenXRF D Command Rx retry on checksum error
9736 XPSC GenXRF D String emission retry on checksum error
9737 XPSC GenXRF D Reply emission retry on checksum error
9738 XPSC GenXRF D Time-out on reply acknowledge
9739 XPSC GenXRF D No access to display
9740 XPSC GenXRF D Tx interrupt time-out on display
9741 XPSC GenXRF D Power-up on XPSC
9742 XPSC GenXRF D X-ray tube spark
9743 XPSC GenXRF D Message size too large on command from XSM
9745 XPSC GenXRF D Serial AD12 bits converter SSIO access denied
9746 XPSC GenXRF D Serial AD8 bits converter SSIO access denied
9748 XPSC GenXRF D Serial Eeprom SSIO access denied
9749 XPSC GenXRF D Retry on Eeprom write
9750 XPSC GenXRF D X-ray tube filament current warning limit reached
9752 XPSC GenXRF D Retry on X-ray tube filament power supply fault
9753 XPSC GenXRF D Retry on X-ray tube filament open
9754 XPSC GenXRF D Retry on X-ray tube filament shorted
9755 XPSC GenXRF D Retry on X-ray tube filament current fatal limit reached
9756 XPSC GenXRF D Retry on secondary water flow sensor error
9757 XPSC GenXRF D X-ray tube filament current regulation in limit
9759 XPSC GenXRF D X-ray leakage current too high (>1[mA])
9760 XPSC GenXRF D X-ray leakage current (>1000[uA])
9761 XPSC GenXRF D XPSC First 8 bits A/D converter error
9762 XPSC GenXRF D XPSC Second 8 bits A/D converter error
9763 XPSC GenXRF D XPSI 8 bits A/D converter error
9764 XPSC GenXRF D XPSP 8 bits A/D converter error
9765 XPSC GenXRF D XPSM 8 bits A/D converter error
9766 XPSC GenXRF D XPSC DSP Power-up
9768 XPSC GenXRF D DSP SSIO access denied
9769 XPSC GenXRF D X-ray positive leakage current error
9770 XPSC GenXRF D X-ray negative leakage current error
9771 XPSC GenXRF D Primary water flow sensor error
9984 XPSC GenXRD D No access to XSN network
9985 XPSC GenXRD D Tx interrupt time-out on XSN network
9986 XPSC GenXRD D Time-out on command size from XSM
9987 XPSC GenXRD D Time-out on command data string from XSM
9988 XPSC GenXRD D Time-out on command checksum from XSM

2-54 Service Manual ARL GenX AA83678


GenX Description Chapter 2

9989 XPSC GenXRD D Time-out on broadcast command checksum


9990 XPSC GenXRD D String reception retry on checksum error
9991 XPSC GenXRD D Command Rx retry on checksum error
9992 XPSC GenXRD D String emission retry on checksum error
9993 XPSC GenXRD D Reply emission retry on checksum error
9994 XPSC GenXRD D Time-out on reply acknowledge
9995 XPSC GenXRD D No access to display
9996 XPSC GenXRD D Tx interrupt time-out on display
9997 XPSC GenXRD D Power-up on XPSC
9998 XPSC GenXRD D X-ray tube spark
9999 XPSC GenXRD D Message size too large on command from XSM
10000 XPSC GenXRD D Serial AD12 bits converter SSIO access denied
10001 XPSC GenXRD D Serial AD8 bits converter SSIO access denied
10002 XPSC GenXRD D Serial Eeprom SSIO access denied
10003 XPSC GenXRD D Retry on Eeprom write
10004 XPSC GenXRD D Retry on X-ray tube filament power supply fault
10005 XPSC GenXRD D Retry on X-ray tube filament open
10006 XPSC GenXRD D Retry on X-ray tube filament current fatal limit
10007 XPSC GenXRD D Retry on secondary water flow sensor error
10008 XPSC GenXRD D X-ray tube filament current regulation in limit
10009 XPSC GenXRD D X-ray leakage current (>1000[uA])
10010 XPSC GenXRD D XPSC First 8 bits A/D converter error
10011 XPSC GenXRD D XPSC Second 8 bits A/D converter error
10012 XPSC GenXRD D XPSI 8 bits A/D converter error
10013 XPSC GenXRD D XPSP 8 bits A/D converter error
10014 XPSC GenXRD D XPSM 8 bits A/D converter error
10015 XPSC GenXRD D XPSC DSP Power-up
10016 XPSC GenXRD D DSP SSIO access denied
10017 XPSC GenXRD D Primary water flow sensor error

AA83678 Service Manual ARL GenX 2-55


3

INSTALLATION AND REMOVING


PROCEDURE
Installation and Removing Procedure Chapter 3

3 INSTALLATION AND REMOVING PROCEDURE


General Content
This troubleshooting procedure is divided into three parts:

− How to remove a GenX generator from the XRF instrument


− How to install a GenX generator into the XRF instrument
− How to remove and install boards and multiplier assembly in the GenX generator

Introduction
Basically each GenX generator is equipped with 4 electronic boards:

X-ray Power Supply Input board called XPSI


X-ray Power Supply Control board called XPSC
X-ray Power Supply Power board called XPSP
X-ray Power Supply Multiplier assembly called XPSM

Note: Each board and assembly can be replaced separately

All above boards and assembly multiplier are connected to a unique motherboard.

Dimension of each board is 480x110mm.

The secondary deionised water of the XRF instrument is used to cool XPSI, XPSP boards and Multiplier
assembly.

Note: XPS board, which was present in the instrument electronic cabinet, has been moved and
integrated inside the GenX generator.

AA83678 Service Manual ARL GenX 3-1


Chapter 3 Installation and Removing Procedure

How to remove a GenX generator from the XRF instrument


Removal procedure
- Switch OFF the X-ray power supply generator by typing the terminal mode command : #XR 0,0,3,3

- Wait for 1 minutes to allow the completion of the secondary deionised water cooling cycle

- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing down the
two main breakers 16A/32A

- Remove all the panels around the GenX generator.

- Disconnect the GenX generator ground cable located on the top left side on the rear GenX panel

- Unscrew fixing screws located on the front panel of GenX generator. Slide “partially” the GenX from
its location without disconnect the water tubes and electrical cables and taking care to avoid destroying
the electrical connectors and water tubes.

- Disconnect the two external water tubes using the water fast fitting connector. See below picture

- Disconnect all the external electrical connectors and X-Ray High Voltage cable connector as shown in
the picture below.

3-2 Service Manual ARL GenX AA83678


Installation and Removing Procedure Chapter 3

Input power connector230 VAC. Data connector 2x12 with X-ray tube Filament connector
WAGO 6x pins. J2 red key lock. M8 WAGO 5x pins. J1

Instrument ground cable Data connector 32x3 Water fast fitting connector
with border key lock. J104 male

Water fast fitting connector Connector X-ray On Lamps X-ray tube HV connector.
male for version 24VDC JP1
WAGO 7 pins. J407

Connector X-ray On Lamps


for version 230VAC. J3

Figure 3.1

- Remove the GenX generator from its location.


- The GenX generator is ready to be shipped.

IMPORTANT: In order to avoid damage during shipment, use the packaging built
specially for GenX generator to absorb shocks during shipment.

AA83678 Service Manual ARL GenX 3-3


Chapter 3 Installation and Removing Procedure

How to install a GenX generator into the XRF instrument


Installation procedure
- Be sure the XRF instrument is completely switched OFF. Red emergency button must be pressed and
the two main breakers 16A/32A must be in down position.

- Remove all the panels around the GenX generator.

- Install the GenX generator into its previous location in the XRF instrument.

- Connect all the external electrical connectors as shown in the picture. Below.
Take care about the connector lock key, especially for the data connector 2x12 with red key lock (Ref.
M8).

- Connect the X-Ray High Voltage cable connector.


Important: Before the connexion of the connector, clean the connector surface, change the silicone
washer (A037323) and add silicone oil (A037324). Note : the silicone washer and oil are supplied with
the GenX generator.

- Connect the two external water tubes using the water fast fitting connectors. See picture below.

- Install the GenX generator completely in its location and fix it with screws located on the front panel of
the GenX generator.

- Connect the GenX generator ground cable located on the top left side on the rear GenX panel.

- Install the panels around the GenX generator.

- Switch ON the XRF instrument by pulling up the two main breakers 16A/32A and releasing the red
emergency button.

- Wait until all XRF instrument starting sequences have been completed and then send the instrument
configuration.

- At this step, the instrument is ready to be used.

3-4 Service Manual ARL GenX AA83678


Installation and Removing Procedure Chapter 3

Input power connector230 VAC. Data connector 2x12 with X-ray tube Filament connector
WAGO 6x pins. J2 red key lock. M8 WAGO 5x pins. J1

Instrument ground cable Data connector 32x3 Water fast fitting connector
with border key lock. J104 male

Water fast fitting connector Connector X-ray On Lamps X-ray tube HV connector.
male for version 24VDC JP1
WAGO 7 pins. J407

Connector X-ray On Lamps


for version 230VAC. J3

Figure 3.2

AA83678 Service Manual ARL GenX 3-5


Chapter 3 Installation and Removing Procedure

How to remove and install boards and multiplier assembly in


the GenX generator
Introduction

IMPORTANT:
Before to have access inside the GenX generator, you have to read carefully the following
caution notes in order to avoid any trouble occurring to the generator or any injury.

Warning
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk
DANGER: LETHAL VOLTAGES are present in the GenX generator
whenever the generator is connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX
generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after
power disconnection.

Accessibility inside the GenX generator


The “opening” of the GenX generator, which is necessary for the
maintenance and repair, must be STRICTLY done by a certified
service engineer who has followed a special training course, see
paragraph “Training”.
Thermo Electron Ecublens Switzerland will be not responsible for
any injury occurred on body person in case of no respect of the
special training accreditation given for GenX generator.

3-6 Service Manual ARL GenX AA83678


Installation and Removing Procedure Chapter 3

The figure hereafter shows the location of the boards t in the GenX generator. It’s very important to respect
board’s location.

XPSI Board XPSC Board XPSP Board Multiplier Assembly

Figure 3.3

AA83678 Service Manual ARL GenX 3-7


Chapter 3 Installation and Removing Procedure

Figure 3.4 XPSI board

Figure 3.5 XPSC board

3-8 Service Manual ARL GenX AA83678


Installation and Removing Procedure Chapter 3

Figure 3.6 XPSP board

Figure 3.7 Multiplier assembly

AA83678 Service Manual ARL GenX 3-9


Chapter 3 Installation and Removing Procedure

The picture hereafter shows the position of the four screws holding the GenX rear panel:

Screw 2
Screw 1

Screw 3 Screw 4

Figure 3.8

The picture hereafter shows the position of the three screws holding the Multiplier assembly on the bottom
side:

Screw 1

Screw 3

Screw 2

Figure 3.9

3-10 Service Manual ARL GenX AA83678


Installation and Removing Procedure Chapter 3

The picture hereafter shows the position of the four screws holding the GenX top cover:

Screw 1 Screw 2

Screw 3 Screw 4

Figure 3.10

Removal procedure
This procedure starts from the step where the GenX generator is removed from the XRF instrument. Note:
To remove the GenX generator from the instrument, consult the previous chapter “ HOW TO REMOVE A
GENX GENERATOR FROM THE XRF INSTRUMENT”.

- Put the GenX generator on a clean table.

VERY IMPORTANT:

Electrocution risk
DANGER: LETHAL VOLTAGES are present in the GenX generator
whenever the generator is connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX
generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after power
disconnection.

AA83678 Service Manual ARL GenX 3-11


Chapter 3 Installation and Removing Procedure

VERY IMPORTANT:

A) WAIT 15 MINUTES TO BE SURE THAT ALL STORED ENERGY IS DISCHARGED


AFTER POWER DISCONNECTION.

B) UNPLUG AND PLUG ACTION OF ALL ELECTRONIC BOARDS MUST BE DELAYED


BY AT LEAST 15 MINUTES. NO RESPECT OF THIS CAUTION NOTE CAN OCCUR
SERIOUS DAMAGE ON THE MOTHERBOARD CONNECTORS DUE TO
CAPACITOR STORED ENERGY.

- Disconnect the two remaining water fast fitting connectors: 1x connexion at the multiplier assembly and
1x connexion between XPSI and XPSP boards.

- Remove the rear panel of the GenX by unscrewing 4x screws. See the screws location at the previous
chapter “INTRODUCTION”.

- At this step, you are able to remove all the boards from the rear side of the GenX.
Note: All boards are connected to a unique mother board located near to the GenX front panel.

- Hereafter is the detail of the removal of each boards and multiplier assembly

- Removal of XPSI board:


The water tube connectors, located at the end of the board, can be used to extract the XPSI board. Pull
the XPSI board very slightly without a big force to avoid destroying the fixing of the water cooling
block on the XPSI board.
VERY IMPORTANT : Take care, when you remove the XPSI board from its location, do not destroy
electronic components located on the way during the removal of the XPSI board => XPSI board
electronic components and XPSC board electronic components.
Have a look to the picture showing the XPSI board at the previous chapter “INTRODUCTION”

- Removal of XPSC board:


The XPSC board can be extracted by using your fingers gripped at the end of the XPSC board. Pull the
XPSC board very slightly without a big force.
VERY IMPORTANT : Take care, when you remove the XPSC board from its location, do not destroy
electronic components located on the way during the removal of the XPSC board => XPSI board
electronic components, XPSC board electronic components and XPSP board electronic components.
Have a look to the picture showing the XPSC board at the previous chapter “INTRODUCTION”

- Removal of XPSP board:


The water tube connectors, located at the end of the boards, can be used to extract the XPSP board. Pull
the XPSP board very slightly without a big force to avoid destroying the fixing of the water cooling
block on the XPSP board.
VERY IMPORTANT : Take care, when you remove the XPSP board from its location, do not destroy
electronic components located on the way during the removal of the XPSP board => XPSC board
electronic components, XPSP board electronic components and XPSM board electronic components.
Have a look to the picture showing the XPSP board at the previous chapter “INTRODUCTION”

- Removal of Multiplier assembly:


The multiplier assembly is connected to the unique mother board through XPSM board.
The Multiplier assembly is secured mechanically by four screws located at the bottom panel of the
GenX generator. See the screws location at the previous chapter “INTRODUCTION”. Unscrew those
three fixing screws. Extract the multiplier assembly by pulling with a slightly force. If the multiplier
assembly can not be removed completely, the top cover of the GenX must be opened to allow
facilitating the way of the High Voltage transformer located on the multiplier assembly. See the screws
location at the previous chapter “INTRODUCTION”.

3-12 Service Manual ARL GenX AA83678


Installation and Removing Procedure Chapter 3

VERY IMPORTANT : Take care, when you remove the Multiplier assembly from its location, do not
destroy electronic components located on the way during the removal of the XPSM board => XPSP
board electronic components and XPSM board electronic components.
Have a look to the picture showing the Multiplier assembly at the previous chapter “INTRODUCTION”

Installation procedure
This procedure starts from the step where the GenX generator is removed from the XRF instrument. Note:
To remove the GenX generator from the instrument, consult the previous chapter “HOW TO REMOVE A
GENX GENERATOR FROM THE XRF INSTRUMENT”.

- Put the GenX generator on a clean table.

VERY IMPORTANT:

Electrocution risk
DANGER: LETHAL VOLTAGES are present in the GenX generator
whenever the generator is connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX
generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after power
disconnection.

VERY IMPORTANT:

A) WAIT 15 MINUTES TO BE SURE THAT ALL STORED ENERGY IS


DISCHARGED AFTER POWER DISCONNECTION.

B) UNPLUG AND PLUG ACTION OF ALL ELECTRONIC BOARDS MUST BE


DELAYED BY AT LEAST 15 MINUTES. NO RESPECT OF THIS CAUTION
NOTE CAN OCCUR SERIOUS DAMAGE ON THE MOTHERBOARD
CONNECTORS DUE TO CAPACITOR STORED ENERGY.

- Disconnect the two remaining water fast fitting connectors: 1x connexion at the multiplier assembly and
1x connexion between XPSI and XPSP boards.

- Remove the rear panel of the GenX by unscrewing 4x screws. . See the screws location at the previous
chapter “INTRODUCTION”.

- At this step, you are able to install all the boards from the rear side of the GenX.
Note : All boards are connected to a unique mother board located near to the GenX front panel

- Hereafter is the detail of the installation of each boards and multiplier assembly

- Installation of XPSI board:


Introduce the XPSI board in the GenX generator. Have a look in the previous chapter
“INTRODUCTION” for the correct location of the XPSI board.
VERY IMPORTANT : Take care, when you install the XPSI board to its location, do not destroy
electronic components located on the way during the installation of the XPSI board => XPSI board
electronic components and XPSC board electronic components.
VERY IMPORTANT : To be sure that XPSI is in the final locked mechanical position, you have to

AA83678 Service Manual ARL GenX 3-13


Chapter 3 Installation and Removing Procedure

pass through two “click” distinct positions when the XPSI board is pushing to the final position. Only
one “click” inform that XPSI board isn’t in the final position.
Have a look to the picture showing the XPSI board at the previous chapter“INTRODUCTION”

- Installation of XPSC board:


Introduce the XPSC board in the GenX generator. Have a look in the previous chapter
“INTRODUCTION” for the correct location of the XPSC board.
VERY IMPORTANT : Take care, when you install the XPSC board to its location, do not destroy
electronic components located on the way during the installation of the XPSC board => XPSI board
electronic components, XPSC board electronic components and XPSP board electronic components.
VERY IMPORTANT : To be sure that XPSC is in the final locked mechanical position, you have to pass
through two “click” distinct positions when the XPSC board is pushing to the final position. Only one
“click” inform that XPSC board isn’t in the final position.
Have a look to the picture showing the XPSC board at the previous chapter “INTRODUCTION”

- Installation of XPSP board:


Introduce the XPSP board in the GenX generator. Have a look in the previous chapter
“INTRODUCTION” for the correct location of the XPSC board.
VERY IMPORTANT : Take care, when you install the XPSP board to its location, do not destroy
electronic components located on the way during the installation of the XPSP board => XPSC board
electronic components, XPSP board electronic components and XPSM board electronic components.
VERY IMPORTANT : To be sure that XPSP is in the final locked mechanical position, you have to
pass through two “click” distinct positions when the XPSP board is pushing to the final position. Only
one “click” inform that XPSP board isn’t in the final position.
Have a look to the picture showing the XPSP board at the previous chapter “INTRODUCTION”

- Installation of Multiplier assembly:


Introduce the Multiplier assembly in the GenX generator. Have a look in the previous chapter
“INTRODUCTION” for the correct location of the multiplier assembly.
VERY IMPORTANT : Take care, when you install the Multiplier assembly to its location, do not
destroy electronic components located on the way during the installation of the XPSM board => XPSP
board electronic components and XPSM board electronic components.
If the Multiplier assembly is stopped on its way to the mother board, the top cover of the GenX must be
open to allow facilitating the way of the High Voltage transformer located on the Multiplier assembly.
See the screws location at the previous chapter “INTRODUCTION”.
VERY IMPORTANT : To be sure that Multiplier assembly is in the final locked mechanical position,
you have to pass through two “click” distinct positions when the XPSM board is pushing to the final
position. Only one “click” inform that XPSM board isn’t in the final position.
Install the three screws located at the bottom panel of the GenX generator to secure mechanically the
Multiplier assembly. See the screws location at the previous chapter “INTRODUCTION”.
Have a look to the picture showing the Multiplier assembly at the previous chapter “INTRODUCTION”

- At this step, all boards and multiplier assembly are installed and locked to the final mechanical position
to the unique mother board.

- Install the rear panel of the GenX by screwing 4x screws. See the screws location at the previous chapter
“INTRODUCTION”.

- VERY IMPORTANT: Connect the two water fast fitting connectors: 1x connexion at the multiplier
assembly and 1x connexion between XPSI and XPSP boards. Have a look at the chapter “HOW TO
INSTALL A GENX GENERATOR INTO THE XRF INSTRUMENT” showing the rear side of the
GenX with water fast fitting connector installed.

- At this step, the GenX generator is ready to be installed into the XRF instrument. Have a look at the
chapter “HOW TO INSTALL A GENX GENERATOR INTO THE XRF INSTRUMENT”.

3-14 Service Manual ARL GenX AA83678


Installation and Removing Procedure Chapter 3

Installation of the high voltage cable


This procedure is valid for the connection between the high voltage cable and the X-ray tube and also
between the high voltage cable and the power supply generator. The following picture shows the plug with
the mounting accessories.

Contact pin Plug Nipple Ring Flange Sleeve

Cable

Screw (2x)

Grub screw
(Screw out before tightening)

Rubber Gasket Ring Nut


see note 1
Divided Flange
Silicone Washer
see note 2

KXx9499D001800
Silicone Oil
Figure 3.11 X-ray tube and generator connector parts

1. Make sure that the termination plug and the mounting accessories are clean.
2. Screw out the grub screw and slip the ring nut over the sleeve.
3. Place the divided flange onto the ring flange.
4. Wet the silicone washer (both sides) and the rubber gasket with silicone oil.
5. Fit the rubber gasket onto the plug, against the ring flange.
6. Push the silicone washer over the contact pins.
7. Insert the termination plug into the receptacle socket (check the nipple on the plug to ensure correct
positioning of the contact pins).
8. Hand tightens the ring nut.
9. TIGHTEN THE GRUB SCREW.

IMPORTANT:

Always replaces the silicone washer by a new one in case the ring nut has been loosened.

AA83678 Service Manual ARL GenX 3-15


4

TROUBLESHOOTING
Troubleshooting Chapter 4

4 TROUBLESHOOTING
Gen’X diagnostic flow chart
Switch ON Gen'X
No
XRAY OFF WAIT ON XRAY
GetStatusGenerator GetAllStatus
GetStatusInputCurrent GetStatusGenerator

XR Cmd = ON
^
InputVoltage > 280 [V] ?

No INTERLOCK
GetStatusGenerator Yes
GetStatusInputCurrent
Buzzer On

RECTIFIER No
Interlock Ok ?
INPUTVOLTAGE
GetAllStatus
GetStatusGenerator

Yes

No WATERS ON
GetStatusGenerator Monitoring 400V > 250 [V] ?
GetStatusInputCurrent
GetInterlockStatus

Yes

No
Waters Ok ?
POWER ON XRAY
GetAllStatus
GetStatusGenerator

Yes

Monitoring 400V > 340 [V] ?


SW RESISTIVITY
No
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus
GetCoolantStatus Yes

CHANGE KV MA
GetAllStatus
GetStatusGenerator
Resistivity Ok?

No
WAIT XRAY IN TOL
Yes GetAllStatus
GetStatusGenerator
SWITCH ON XRAY
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus
GetCoolantStatus Kv Ma InTolerance ?
GetStatusADMon400V

Yes
Power On Relay

XRAY ON
GetAllStatus
Xray Lamp On GetStatusGenerator
CheckKvMaRegulation

Figure 4.1

AA83678 Service Manual ARL GenX 4-1


Chapter 4 Troubleshooting

Details

No No
GetStatusGenerator
GetAlarmDSP GetStatusADMon400V
CheckStatusTemp
CheckStatusVoltage

Monitoring 400V < 10 [V]


GetStatusGenerator == Fault ? Yes Yes or
Monitoring 400V > 350 [V]

No
No GetAllStatus
GetInterlockStatus
GetCoolantStatus
GetStatusInputCurrent
GetFilamentStatus
CheckWarningLamp

Yes InputCurrent > 2 [A]

GetAllStatus == Fault ? Yes

No

No
GetStatusVoltage

GetCoolantStatus

Yes Status voltage == Fault ?

Primary water
&
Yes
Secondary water
== Fault ?

No

CheckStatusTemp

Yes Temperature == Fault ?

FAULT GENERATOR

Figure 4.2

4-2 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

Switch OFF Gen'X


XRAY ON
GetAllStatus
GetStatusGenerator
CheckKvMaRegulation

XR Cmd = OFF

SWITCH OFF KV MA No
GetAllStatus
GetStatusGenerator

Kv & Ma == 0 ?

Yes

POWER OFF XRAY


GetStatusGenerator
GetCoolantStatus

Power Off Relay

Xray Lamp Off

WAIT COOLING OFF No


GetStatusGenerator
GetCoolantStatus

Cooling Off ?

Yes

XRAY OFF
GetStatusGenerator
GetStatusInputCurrent

Figure 4.3

AA83678 Service Manual ARL GenX 4-3


Chapter 4 Troubleshooting

Detailed diagnostic flow chart

Switch ON Gen'X
XRAY OFF No
WAIT ON XRAY
GetStatusGenerator GetAllStatus
GetStatusInputCurrent GetStatusGenerator

XR Cmd = ON Step 4
^
InputVoltage > 280 [V] ?

No INTERLOCK
GetStatusGenerator Yes
GetStatusInputCurrent
Buzzer On

Step 1
RECTIFIER No
Interlock Ok ?
INPUTVOLTAGE
GetAllStatus
GetStatusGenerator

Yes Step 5

No WATERS ON
GetStatusGenerator Monitoring 400V > 250 [V] ?
GetStatusInputCurrent
GetInterlockStatus

Step 2 Yes

No
Waters Ok ?
POWER ON XRAY
GetAllStatus
GetStatusGenerator

Yes
Step 6
Monitoring 400V > 340 [V] ?
SW RESISTIVITY
No
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus
GetCoolantStatus Yes
Step 3
CHANGE KV MA
GetAllStatus
GetStatusGenerator
Resistivity Ok?

No
WAIT XRAY IN TOL
Yes GetAllStatus
GetStatusGenerator
Step 7
SWITCH ON XRAY
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus
GetCoolantStatus Kv Ma InTolerance ?
GetStatusADMon400V

Step 4 Power On Relay


Yes

XRAY ON
GetAllStatus
Xray Lamp On GetStatusGenerator
CheckKvMaRegulation

Figure 4.4

4-4 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

Details

No No
GetStatusGenerator
GetAlarmDSP GetStatusADMon400V
CheckStatusTemp
CheckStatusVoltage

Monitoring 400V < 10 [V]


GetStatusGenerator == Fault ? Yes Yes or
Monitoring 400V > 350 [V]

No
No GetAllStatus
GetInterlockStatus
GetCoolantStatus
GetStatusInputCurrent
GetFilamentStatus
CheckWarningLamp

Yes InputCurrent > 2 [A]

GetAllStatus == Fault ? Yes

No

No
GetStatusVoltage

GetCoolantStatus

Yes Status voltage == Fault ?

Primary water
&
Yes
Secondary water
== Fault ?

No

CheckStatusTemp

Yes Temperature == Fault ?

FAULT GENERATOR

Figure 4.5

AA83678 Service Manual ARL GenX 4-5


Chapter 4 Troubleshooting

Switch OFF Gen'X


XRAY ON
GetAllStatus
GetStatusGenerator
CheckKvMaRegulation

XR Cmd = OFF

SWITCH OFF KV MA No
GetAllStatus
GetStatusGenerator

Kv & Ma == 0 ?

Yes

POWER OFF XRAY


GetStatusGenerator
GetCoolantStatus

Power Off Relay

Xray Lamp Off

WAIT COOLING OFF No


GetStatusGenerator
GetCoolantStatus

Cooling Off ?

Yes

XRAY OFF
GetStatusGenerator
GetStatusInputCurrent

Figure 4.6

4-6 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

Step 1
XR Cmd = ON Check Instruments Interlocks.

Alarms: 8297-8301.
Check the board is properly plugged on the mother board.
INTERLOCK
NO Alarms: Fatal alarms from 8204 to 8208 and 8248 to 8253
GetStatusGenerator 8204 to 8208 and 8248 to 8252 are interlock µ
GetStausInputCurrent
switches contacts
8253 is the water bloc thermal 60°C safety switch

Display: To get a dynamic monitoring of the contacts use the


following T.M. Cmd
DY 2,3,10
INTERLOCK OK
YES

Figure 4.7

Step 2
Checks instrument Water Parameters.

Alarms: 8193, 8194. No transition from the read relays.


WATERS ON Check the magnets or Read relay positions.
NO
GetStatusGenerator Display Primary Water flow state (8193, 8232, 8233)
GetStausInputCurrent Use the following T.M. Cmd
GetInterlockStatus
PW 1 Prim. Water Valve Open→Flow ON
PW 0 Prim. Water Valve Closed→Flow OFF
Secondary water flow state (8194, 8222)
Use the following T.M. Cmd
DY 2,3,2,5
SW 1→Sec. Water ON
SW 0 →Sec. Water OFF
WATERS OK
Alarms: 8234, 8235. Check Reservoir water level check
gauge connection.
Alarms: 8223. Bad thermal exchange. Check, clean heat
exchanger.
Alarms: 8230, 8231. NTC defect or bad connection.
Alarms: 8247, 8285. Check 60°C safety switch.

Figure 4.8

AA83678 Service Manual ARL GenX 4-7


Chapter 4 Troubleshooting

Step 3
Check resistivity

SW Alarms: 8225. Check XPSC Resistivity Driver.


RESISTIVITY Alarms: 8226, 8285 Resitivity gauge open circuit or short
No GetStatusGenerator circuit.
GetStausInputCurrent Alarms: 8227. Bad water or bad ion exchanger.
GetInterlockStatus Status RS 2054
Getcoolantstatus

Resistivity Ok

Figure 4.9

Step 4
SWITCH ON
Check XPSI board
X-RAY
Input Voltage:
GetStatusGenerator
GetStatusInputCurrent
GetInterlockStatus Alarms: 8378-8381. XPSI Board Fuse F3 is defect
GetCoolantstatus Alarms: 8292. No Input Voltage. XPSI Board Fuse F1 is
GetStatusADMon400V defect.
Contactor K1 is not driving.
Check Power On signal:
Display T. M. Cmd. DY 2,3,100,0,FF81H,0,2
Power On Relay

X-Ray lamps:

X-RayLamp On Alarms: 8219, 8377 Check Bulbs.


Check XPSI Board Fuse F2.
Check if lamps connection cable is correctly
plugged.
Display T. M. Cmd. DY 2,3,11,3,4
WAIT ON X-RAY
2 Lamps open → 0-18 Digits 8219
No 2 Lamps short → 20 -255 Digit 8377
GetAllStatus
2 Lamps OK → 100 Digits
GetStatusGenerator

Alarms: 8308, 8309. Input voltage <195V or > 253V


Or
Alarms: 8516, 8517. Input voltage <195V or > 253V

InputVoltage > 280 V

Figure 4.10

4-8 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

Step 5
Alarm: 8293. Time-out to reach 250 V
Yes

Buzzer On PowerMosfet M1 Input rectifier.

Using a DVM, selector on Diodes, measure the diodes between


pin 5 (anode) and pin 1 (cathode). Value to be obtained  500mV.
Do the same but invert the polarity on the DVM. Value to be
RECTIFIER obtained   .
INPUT VOLTAGE
No

GetAllStatus
GetStatusGenerator

In the same way measure between pin 5 (anode) and pin 3


(cathode). Value to be obtained  500mV. . Do the same, but
invert the polarity on the DVM. Value to be obtained   .
Monitoring 400 V Idem between pin 1 (anode) and pin 2 (cathode).
> 250 V
Idem between pin 3 (anode) and pin 2 (cathode).

M2 rectifier diodes

Using a DVM, selector on Diodes, measure diodes between


(anode) pin 1 and (cathode) pin 2. Value to be obtained  500mV.
Idem with (anode) pin 3and (cathode) pin 2.

Figure 4.11

AA83678 Service Manual ARL GenX 4-9


Chapter 4 Troubleshooting

Step 6
Alarm: 8294. The voltage does not step up to 360 V.
M2 Rectifier and Diode D19 have to be checked
POWER ON
X-RAY
Use the same methods as in Step 5
No

GetAllStatus
GetStatusGenerator

Measure PowerMosfet M2: Drain-Gate capacity

Monitoring 400 V
> 340 V Interrupt resistor R67 link (unsolder one leg), remove driver IC4,
and measure with a DVM Drain-Gate capacity between pin 7 and
pin 8 PowerMosfet M2.
It should show a value between 5 to 20nF,
The value does depend upon the DVM.

Interrupt resistor R66 link (unsolder one leg), remove driver IC1,
and measure with a DVM Drain-Gate capacity between pin 7 and
pin 8 PowerMosfet M1.
It should show a value between 5 to 20nF,
The value does depend upon the DVM.

Status: RS 2101

M1 Diode has to be checked

PowerMosfet M1 400V Diode:

Apply the same diode measurement method as above


Measure between pin 4 (anode) and pin 6 (cathode).

If XPSI is showing problems,


Change the board.

(It is possible to replace some components like drivers HCPL3101


IC1 and IC4).

Figure 4.12

4-10 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

Step 7
Alarms: 8730 (kV mA in tolerance timeout.
CHANGE kV mA 8196-8199, 8273-8274.
Make sure the XC X-Ray calibration has been performed.
GetAllStatus At this level one can suspect the inverter stages of XPSP
GetStatusGenerator The following procedure is valid for:
XPSP Schematics S112067 No. Mod. 9 and lower
Measure XPSP IGBTs
With an Ohmmeter measure between pin 13 and pin 16 (Value
measured =  ).
WAIT X-RAY IN Idem avec pin 13 et pin 15.
No
TOLERANCE Idem avec pin 13 et pin 14.
GetAllStatus Idem avec pin 17 et pin 16.
GetStatusGenerator Idem avec pin 17 et pin 15.
Idem avec pin 17 et pin 14.

kV mA in tolerance

Measure of IGBTs Drain-Gate capacity:


Interrupt R55, R56, R57, R58 resistors conductivity by unsoldering
1 leg of each.
Remove drivers IC2, IC3, IC4, IC5, IC6, IC7.
Measure Drain-Gate capacity between :
pin 1 et 2
pin 3 et 4
pin 5 et 6
pin 7 et 8
pin 9 et 10
pin 11 et 12
The value to be obtained should be between 5 and 20nF,
It depends upon measurement device.

Figure 4.13

AA83678 Service Manual ARL GenX 4-11


Chapter 4 Troubleshooting

The following procedure is valid for:


XPSP Schematics S112067 No. Mod. 11 and higher.

DVM Z1 HV Z2 HV Z3 FIL.
+ C (2) Ω ∞ ∞ ∞
Com E(3) ≈2V ≈ 2V ≈ 2V
DVM Z4 HV Z5 HV Z6 FIL.
+ C (2) Ω ∞ ∞ ∞
Com E(3) ≈2V ≈ 2V ≈ 2V
DVM Z1 HV Z2 HV Z3 FIL.
+ E (3) Ω v530 ≈ 530 ≈ 560
Com C(2) 0.37 V 0.37 V 0.37 V
DVM Z4 HV Z5 HV Z6 FIL.
+ E (3) Ω ≈ 530 ≈ 530 ≈ 560
Com C(2) 0.37 V 0.37 V 0.37 V
Figure 4.14

4-12 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

Alarms list
Fatal alarms
Alarm Issued Severity Description
number by
8192 XPS F Unknown command received from XQM96
S: Reset ICS

8193 XPS F Primary water flow sensor error


8194 XPS F Secondary water flow sensor error
8195 XPS F Secondary water pump error
S: See Step 2

8196 XPS F Desired voltage (kV) too low


Asked kV conditions not allowed by firmware.
S: Select coherent conditions

8197 XPS F Desired voltage (kV) too high


Firmware and XPSC jumpers setting not compatible.
S: Select coherent conditions

8198 XPS F Desired current (mA) too low


Asked mA conditions not allowed by firmware.
S: Select coherent conditions

8199 XPS F Desired current (mA) too high


Firmware and XPSC jumpers setting not compatible.
S: Select coherent conditions

8200 XPS F Desired power too high


Asked power (kV × mA) not allowed by firmware.
S: Select coherent conditions

8201 XPS F Illegal power change mode


Erroneous instructions from XQM
S: Reset ICS

8202 XPS F Illegal power change speed


KV / mA Out of power change speed out of range. < 1W / min. or > 3000 W / min.
S: Set coherent conditions.

8203 XPS F Illegal anode Not Used


8204 XPS F Spectrometer pressostat open
8205 XPS F Spectrometer tank cover safety open
8206 XPS F Primary chamber or PBF safety open
8207 XPS F X-ray tube position safety open
8208 XPS F X-ray tube connection safety open
8209 XPS F Illegal interlock
See Step 1

8210 XPS F Time-out on X-ray ON ack.

AA83678 Service Manual ARL GenX 4-13


Chapter 4 Troubleshooting

Synchronisation error in 80C196KR


S: Reset ICS

8211 XPS F Over kilovolts on X-ray power supply


Used on Retrofit
8212 XPS F Over milliamps on X-ray power supply
Used on Retrofit
8213 XPS F Over power on X-ray power supply
Used on Retrofit
8214 XPS F X-ray power supply fault
Used on Retrofit

8215 XPS F X-ray tube filament open

Filament Curent < 1 A Filament Voltage > 18V


Bad filament connection.
Disconnect filament and measure filament resistivity with a DVM: It should be 1 Ohm.
If not filament is defect.
S: Change X-Ray tube.

8216 XPS F X-ray tube filament over current


Used on Retrofit
8217 XPS F Kilovolts regulation fault
Used on Retrofit
8218 XPS F Milliamps regulation fault
Used on Retrofit
8219 XPS F X-ray ON lamp broken
Used on Retrofit
8220 XPS F Low kilovolts limit
Used on Retrofit
8221 XPS F Leak current too high
Used on Retrofit

8222 XPS F Secondary water flow OFF


8223 XPS F Secondary water temperature too high If > 50° C
See Step 2

8224 XPS F ‘PA’ or ‘CO’ command not allowed now


Used on Retrofit

8225 XPS F Secondary water resistivity driver fault If > 5 V or < 2V


8226 XPS F Secondary water resistivity gauge fault Gauge not connected
8227 XPS F Secondary water resistivity too low If < 300 k Ώ / ccm
See Step 3

8228 XPS F Primary water inlet temperature gauge fault Not used
8229 XPS F Primary water outlet temperature gauge fault Not used
8230 XPS F Secondary water inlet temperature gauge fault
8231 XPS F Secondary water outlet temperature gauge fault

4-14 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

If value = 0, NTC defect or not connected.


S: Change NTC
8232 XPS F Primary water cooler OFF not used
8233 XPS F Primary water OFF not used
8234 XPS F Secondary water level gauge fault not used

8235 XPS F Secondary water level too low


S: Top up deionised water reservoir

8236 XPS F Water control command not allowed now


In normal operation primitive water control commands not allowed.
S: Don’t use primitive commands in that situation.

8237 XPS F Bad water condition


If any water error
See Step 2

8239 XPS F Unknown status command


Status list in the software not updated.

8240 XPS F X-ray kV calibration slope error


If slope < 0.8 or > 1.2  Out of limits
S: XC

8241 XPS F X-ray kV calibration offset error


If offset < -1000V or >1000 V  Out of limits
S: XC

8242 XPS F X-ray mA calibration slope error


If slope < 0.8 or > 1.2  Out of limits
S: XC

8243 XPS F X-ray mA calibration offset error


If offset < -1.0 mA or >1.0 mA  Out of limits

S: XC command has to be executed. Corrupted data in the EEPROM of XPSC. Check firmware Status.

8244 XPS F Status: X-ray arc rate too big


The number of arcs does exceed the high value limit defined in the status channel 2064.

AA83678 Service Manual ARL GenX 4-15


Chapter 4 Troubleshooting

How to diagnostic insulation problems


8244 XPS F Status: X-ray arc rate too big
The number of arcs does exceed the high value limit defined in the status channel 2064.

IMPORTANT:
Before to have access inside the GenX generator, you have to read carefully the following
caution notes in order to avoid any trouble occurring to the generator or any injury.

Warning

Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk

DANGER : LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER : LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after power
disconnection.

Accessibility inside the GenX generator

The “opening” of the GenX generator, which is necessary for the


maintenance and repair, must be STRICTLY done by a certified
service engineer who has followed a special training course, see
paragraph “Training”.

Thermo Fisher Scientific Switzerland will be not responsible for any


injury occurred on body person in case of no respect of the special
training accreditation given for GenX generator.

HV MULTIPLIERS AND GENERATOR HV CONNECTOR INSULATION


problems
When increasing the high voltage, a noise near the multiplier can be heard, and an alarm “PWMTRIP”
Power Off, issued by DSP” is generated, the multiplier assembly can be suspected.

Change the multiplier assembly.

4-16 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

X-Ray tube insulation problems

- Switch OFF the X-ray power supply generator by typing the terminal mode command: #XR
0,0,3,3
- Wait for 1 minute to allow the completion of the secondary deionised water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing
down the two main breakers 16A/32A
- Remove all the panels around the GenX generator.
- Remove the HV cable connector on the back side of the generator. Inspect the connector. No
spark over traces should be visible.

Make sure no wire or any other conductor is in the vicinity of the HV connector.

- Put back the cover of the instrument generator cabinet.


- Inhibit any approach from external people.
- Switch 32 A and 16 A breaker ON.
- Use terminal mode command:
- XR 50,0,1,1,1,1.
- Display the high voltage with the terminal mode command:
- DY 2,3,3,1,3
- For this test don’t exceed 5 minutes.
- If the voltage value is stable and if there are no particular noise or alarms concerning insulation,
The X-Ray tube or HV cable is faulty.

HV cable insulation problems

- Switch OFF the X-ray power supply generator by typing the terminal mode command:
- #XR 0,0,3,3
- Wait for 1 minute to allow the completion of the secondary deionized water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing
down the two main breakers 16A/32A
- Remove all the panels around the GenX generator.
- Remove the HV cable connector on the back side of the generator and from the X-Ray tube.
Inspect the connectors; on the cable, on the generator and on the X-Ray tube. No spark-over
traces should be visible.
- With a Meger check the HV cable insulation.
- The resistivity must be infinite.
- If it is not change the HV cable.

Note: The HV cable connectors must be cleaned yearly. Both connectors’ silicone washers must be
replaced and lubricated yearly.

AA83678 Service Manual ARL GenX 4-17


Chapter 4 Troubleshooting

It is recommended to systematically exchange the HV cable every 5 years.

8245 XPS F XPS 12 bits A/D converter time-out Not used

8246 XPS F Water cooler flow time-out Not used

8247 XPS F Secondary water temperature safety open


8248 XPS F Spectrometer pressostat interlock signal defective
8249 XPS F Spectrometer tank cover interlock signal defective
8250 XPS F Primary chamber or PBF interlock signal defective
8251 XPS F X-ray tube position interlock signal defective
8252 XPS F X-ray tube connection interlock signal defective
8253 XPS F Secondary water temperature interlock signal defective

See step 2 and Step 1

8256 XPS F XPS EEPROM defective Not used

8257 XPS F Unknown X-ray ray power-supply type.


XPSC Incoherent Jumpers settings.
S: Set jumpers according to technical description.

8268 XPS F X-ray generator identified is different of 4.2 kW type. Not used.
8269 XPS F X-ray generator identified is of 4.2 kW type. Not used

8270 XPSLP F SSIO communication timeout.


All serial links on XPSC
S: Reset

8271 XPSLP F Serial AD12 conversion timeout.


No serial ADC communication.
S: Reset

8272 XPSLP F Serial AD12 auto-calibration timeout.


Auto calibration not successful
S: Reset

8273 XPS F X-ray “kV” out of regulation

4-18 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

kV Regulation

kV > 0 No No test

Yes

Yes kV > 5 kV No

< -20%of value


> 5% of value
> 5% of value

Yes
Yes No

Fatal Alarm 8273 No OK

Figure 4.15

8274 XPS F X-ray “mA” out of regulation

mA Regulation

mA > 0 No No test

Yes

Yes mA > 10 mA No

< -20%of value


> 10% of value
> 10% of value

Yes
Yes No

Fatal Alarm 8274 No OK

Figure 4.16

8275 XPSLP F X-ray tube current anode regulation error. Not used
8276 XPSLP F X-ray tube filament current fatal limit reached.

AA83678 Service Manual ARL GenX 4-19


Chapter 4 Troubleshooting

Filament current does exceed 14.5 A


Filament insulation error. Short circuit to the ground.
Disconnect filament connector and measure insulation between filament leads and X-Ray tube body and/or
water pipes. Resistivity with a DVM should show ∞. If not the filament insulation is bad.
Change X-Ray tube.

8277 XPSLP F X-ray power supply interlock open.


Pressostat open.
S: Check tank pressure. Check Pressostat.

8278 XPSLP F X-ray power supply interlock signal defective. Not used

8279 XPSLP F X-ray power supply off. Not used

8280 XPSLP F Unknown X-ray power supply fault. Not used

8281 XPSLP F X-ray warning lamps broken.


See Step 4

8282 XPSLP F X-ray tube filament power supply fault.


Heating current is < 1 A and heating voltage is < 2 volts
Fuse F2 2A (orange soldered fuse) of XPSP board
SG3525AN IC 20 of XPSC.
Filament heating power IGBTs on XPSP.
The filament power supply part of the XPSM board (part of the Multiplier assembly) can be suspected.

8283 XPSLP F X-ray tube filament shorted.


Heating current is > 1 A and heating voltage is < 3.5 volts
Disconnect filament leads and with a DVM measure Filament conductivity.
If the value is close to 0 Ω, the X-Ray tube is dead.
S: Change X-ray tube.

8284 XPS F X-ray generator switched off without reason Not used

8285 XPSLP F Secondary water resistivity too high Not used

8286 XPSLP F X-ray power supply current limitation. Not used

8287 XPSLP F X-ray power supply power limitation. Not used

8288 XPSLP F Incoherence between the anode current and the cathode current.
Not used

8291 XPS F Invalid X-ray generator calibration value


Manual XC kV or / and mA inputs.
If low kV is 20, the high kV must almost be low + 1kV.
If low mA is 20, the high mA must almost be low + 1mA.
S: Enter the coherent kV / mA calibration values, or use default values.

8292 XPSC F Generators mains voltage not present, time-out


If main is < 280VAC. In theory the voltage is 230 × 2
See Step 4

4-20 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

8293 XPSC F Generator 400V not > 250V, time-out


Time out = 5 sec.
See Step 5

8294 XPSC F Generator 400V not > 380V, time-out


Time out = 5 sec.
See Step 6

8295 XPSC F Generator, temperature IGBT too high


3.6 and 4.2 kW → 65°C Water too hot, environmental temperature too high.
1.2 kW → 120°C Bad venting, fans defect, environmental temperature too high.

8296 XPSLP F Bad filament current Not used

8297 XPSC F XPSC First 8 bits A/D converter error. IC 7


8298 XPSC F XPSC Second 8 bits A/D converter error. IC 8
Communication error.
S: Reset ICS

8299 XPSC F XPSI 8 bits A/D converter error


8300 XPSC F XPSP 8 bits A/D converter error
Communication error.
XPSI Board not properly plugged.
S: Shut down power, plug board properly
S: Reset ICS

8301 XPSC F XPSM 8 bits A/D converter error


Communication error.
XPSM Board not properly plugged.
S: Shut down power, plug board properly
S: Reset ICS

8302 XPSC F Generator, temperature Mosfet M1 too high


8303 XPSC F Generator, temperature Mosfet M2 too high
If > 65°C
Water too hot.

8304 XPSC F Leakage current too high


If (absolute anode I – absolute cathode I) > 1.5 mA.
Check X-Ray tube insulation between body and HV connector.
Check X-Ray tube insulation between body and filament connector.
If not OK X-Ray tube is faulty.
S: Change X-ray tube.

8305 XPSC F Power off, issued by DSP; 400V too low (<350V)
Too fast X-Ray power supply transition. 400 V voltage drop. May happen when changing X-Ray tube
conditions in terminal mode, in “On Fast mode” and “Operational Mode”.
See Step 6

8306 XPSC F Power off, issued by DSP; HV too high (>71kV)


KV Regulation error. KV Regulation loop fault.

8307 XPSC F Power off, issued by DSP; Current too high (>35A)

AA83678 Service Manual ARL GenX 4-21


Chapter 4 Troubleshooting

After a power up retry the current can rise above 35 A.

8308 XPSC F Mains voltage below 195 Vac (after 20 ms)


8309 XPSC F Mains voltage above 253 Vac (after 20 ms)
Main power supply fault. The main voltage is checked by the DSP. If out of tolerance the DSP does
immediately shut down the power.
See Step 4

8310 XPSC F Unknown reply from DSP, issued by XPSC


80C196KR no dialog with DSP.
S: Reset ICS. GV 2,3,1.

8311 XPSC F Too many parameters in SSIO, issued by DSP


S: Reset ICS.

8312 XPSC F Unknown DSP Cmd, issued by DSP


S: Reset ICS.

8313 XPSC F PWMTRIP -> Power Off, issued by DSP


Issued by XPSP due to repetitive power stage problems, where the current is > 100A
Check the IGBT.
Issued by XPSM, due to arcs or spark over on the multiplier assembly of XPSM.
If > 15 kV (> 2V at comparator input) at the secondary of the HV transformer of XPSM. Normal 12kV for
70 kV

8314 XPSC F Generator, temperature XPSP too high


If > 75 ° C
Bad cooling. Check fans. Laboratory temperature too high.

8315 XPSC F Generator, temperature filament too high


If XPSM filament heating diodes D1 to D4 > 100 ° C.
Check fans. Laboratory temperature too high.

8316 XPSC F Generator, temperature HV too high


If XPSM HV conversion unit > 75 ° C
Check fans. Laboratory temperature too high.

8317 XPSC F Generator, temperature XPSC too high


If XPSC > 65 °C
Check fans. Laboratory temperature too high.

8318 XPSC F Resistivity circuit open


If > 4490 mV on XPSC 196 P0.6 AD IN.
Check Input Amplifier IC 19 A.
Resistivity gauge disconnected or bad contact.

4-22 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

8323 XPSC F Filament current sensor incoherency


8324 XPSC F Filament voltage sensor incoherency
If before start the sensor delivers a value > 1 A → XPSM Filament current sensor defect. RS 2055
If before start the sensor delivers a value > 10 V → XPSM Filament voltage sensor defect. RS 2098
Measured values out of range, not coherent.

8325 XPSC F Status : XPSI +5V too low <4V


8326 XPSC F Status : XPSI +5V too high >6V
8327 XPSC F Status : XPSC +5V too low <4V
8328 XPSC F Status : XPSC +5V too high >6V
8329 XPSC F Status : XPSP +5Vm too low <4V
8330 XPSC F Status : XPSP +5Vm too high >6V
8331 XPSC F Status : XPSM +5Vm too low <4V
8332 XPSC F Status : XPSM +5Vm too high >6V
8333 XPSC F Status : XPSI +12V too low < 8.5 V
8334 XPSC F Status : XPSI +12V too high > 15 V
8335 XPSC F Status : XPSC +12V too low < 8.5 V
8336 XPSC F Status : XPSC +12V too high > 15 V
8337 XPSC F Status : XPSM +12V too low < 8.5 V
8338 XPSC F Status : XPSM +12V too high > 15 V
8339 XPSC F Status : XPSI -12V too low > -15 V
8340 XPSC F Status : XPSI -12V too high > -8.5 V
8341 XPSC F Status : XPSC –12V too low > -15 V
8342 XPSC F Status : XPSC -12V too high > -8.5 V
8343 XPSC F Status : XPSM -12V too low > -15 V
8344 XPSC F Status : XPSM -12V too high > -8.5 V
8345 XPSC F Status : XPSI +24V too low < 19 V
8346 XPSC F Status : XPSI +24V too high > 27 V
8347 XPSC F Status : XPSC +24V too low < 19 V
8348 XPSC F Status : XPSC +24V too high > 27 V
8349 XPSC F Status : XPSP +24V too low < 19 V
8350 XPSC F Status : XPSP +24V too high > 27 V
8351 XPSC F Status : XPSM +2V5ref too low NOT USED
8352 XPSC F Status : XPSM +2V5ref too high NOT USED
8353 XPSC F Status : XPSC –5V too low > -6 V
8354 XPSC F Status : XPSC –5V too high < -4 V
Low voltages out of tolerance.
Check instrument low voltage power supplies.

8355 XPSC F Status : Mon 400V too low Suppressed after V 3.2-0
8356 XPSC F Status : Mon 400V too high Suppressed after V 3.2-0

8359 XPSC F Leakage Current Error Suppressed after V 3.2-0


Replaced by 8389, 8390.

8360 XPSC F XPSI A/D Mon400V Error


Previous to start, if U < 10V or > 350 V.
If > 350 V → XPSI voltage monitoring IC 7 Defect.

AA83678 Service Manual ARL GenX 4-23


Chapter 4 Troubleshooting

K1 i
L
IL C1 +12kV
230Vac U1 U2
C2
N
15Vrtm

uC DSP

Figure 4.17

At power up if the 32 Amp breaker is on, the voltage leaks from T1 through the drivers to the capacitors C.
The 400 V measuring device does read about 15 V. If not, the 32 Amp breaker is down or the IGBTs of
XPSP are in short circuit.

8362 XPSC F XPSC temperature Board overflow


If > 109 ° C. Out of conversion limits. Mainly indicating NTC is defect.

8363 XPSC F XPSC temperature Board underflow


If < 9.9 ° C. Out of conversion limits. Mainly indicating NTC is defect.

8364 XPSC F XPSI temp. PowerMosfet M1 overflow


8365 XPSC F XPSI temp. PowerMosfet M1 underflow
8366 XPSC F XPSI temp. PowerMosfet M2 overflow
8367 XPSC F XPSI temp. PowerMosfet M2 underflow
All same as above

8368 XPSC F XPSP temperature IGBT overflow


8369 XPSC F XPSP temperature IGBT underflow
All same as above

8370 XPSC F XPSP temperature Board overflow


8371 XPSC F XPSP temperature Board underflow
All same as above

8372 XPSC F XPSM temp. Filament overflow


8373 XPSC F XPSM temp. Filament underflow
All same as above

8374 XPSC F XPSM temp. HV overflow


8375 XPSC F XPSM temp. HV underflow
All same as above
8377 XPSC F X-Ray Lamp shorted
Lamps shorted If ADC converter Digits are between 200..255

8378 XPSC F Status : XPSP –10Vrtm too low


8379 XPSC F Status : XPSP –10Vrtm too high
8380 XPSC F Status : XPSP +15Vrtm too low
8381 XPSC F Status : XPSP +15Vrtm too high
The -10 Vrtm is measure in ON/OFF mode. If present → + 5V. If not present → 0 V.
The +15 Vrtm is measure in ON/OFF mode. If present → + 5V. If not present → 0 V.

4-24 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

See Step 4

8382 XPSC F Neutral to Gnd too low → Not really useful


DY 2,3,11,3,5
RS 2085

8383 XPSC F Neutral to Gnd too high


DY 2,3,11,3,5
RS 2085
If potential is > 200VAC.
The neutral has a floating potential.
Check the main power supply parameters.
Install an insulation transformer if necessary with the 0 volt connected to the ground.
Phase / Neutral are inverted.

8384 XPSC F XPSI input current sensor error


If previous to start the sensor does measure a current > 2 A.
Sensor is defect. If the value is > it means → no current limitation.

8385 XPSC F External relay incoherency


8386 XPSC F Power On relay incoherency
8387 XPSC F No external relay feedback
8388 XPSC F No power relay feedback

AA83678 Service Manual ARL GenX 4-25


Chapter 4 Troubleshooting

PTB-TÜV
SwitchOn XRAY SwitchOff XRAY

External Voltage =
Yes
Present ?

internalRelay = OFF
8385 F_EXTERNAL_RELAY_INCOHERENCY
No -10Vrtm and +15Vrtm = ON

externalRelay = ON Yes

Internal Voltage = Internal Voltage =


Yes Yes
Present ? Present ?

8386 F_POWER_ON_RELAY_INCOHERENCY
No Internal Voltage 320V = ON
No

External Voltage =
No
Present ?
externalRelay = OFF

8387 F_NO_EXTERNAL_RELAY_FEEDBACK
-10Vrtm and +15Vrtm = Off
Yes 24 V ExRelayPresent = OFF

internalRelay = ON

External Voltage =
Present ?

Internal Voltage =
No
Present ?

8388 F_NO_POWER_ON_RELAY_FEEDBACK No
400Vrtm (350V) = OFF
Yes

Continuation of the Continuation of the


process process
externalRelay = OFF

internalRelay = OFF

Figure 4.18

4-26 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

8389 XPSC F X-ray negative leakage current error


IF (Anode i – Cathode i) does exceed - 5mA (- 6mA, -7 mA, etc). RS 2069
a) Bad contacts. Check connections between XPSM and XPSC.
b) After an XC process, if the system is not able to reach the upper mA value due to X-ray tube filament
defect, the calculated A0 coefficient is negative. See example hereafter.

A0 = -1.93370
A1 = 1.04972

Figure 4.19

AA83678 Service Manual ARL GenX 4-27


Chapter 4 Troubleshooting

8390 XPSC F X-ray positive leakage current error


IF (Anode i – Cathode i) does exceed + 5mA (+ 6 mA, + 7 mA, etc). RS 2069
a) XPSM leakage. Check leakage with HV connector disconnected. Set X-Ray conditions to 50kV 0 mA.
Monitor emission current leakage. It should be very close to zero.
S: If not OK, change Multiplier assembly.
b) X-ray tube leakage. Measure the X-Ray tube insulation.
S: If not OK, change tube.

Multiplier
mA + Assembly
a)
i
i b)
A
C

mA -

Figure 4.20

8391 XPSC F Primary water already on


Condition: the duty cycle must be between 5% and 95%
If before start the primary water on signal is at 1 logic.
S: Check read contact. Clean the water bloc.
See Step 1

4-28 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

Warning alarms
8704 XPS W Initial primary water flow insufficient
8705 XPS W Primary water cooler OFF
8706 XPS W Spectrometer pressostat open
8707 XPS W Spectrometer tank cover safety open
8708 XPS W Primary chamber or PBF safety open
8709 XPS W X-ray tube position safety open
8710 XPS W X-ray tube connection safety open
8711 XPS W Illegal interlock
8712 XPS W Primary water inlet temperature gauge fault
8713 XPS W Primary water outlet temperature gauge fault
8714 XPS W Secondary water inlet temperature gauge fault
8715 XPS W Secondary water outlet temperature gauge fault
8716 XPS W Secondary water flow OFF
8717 XPS W Primary water flow insufficient
8718 XPS W Secondary water level gauge fault
8719 XPS W Secondary water level too low
8720 XPS W Secondary water resistivity driver fault
8721 XPS W Secondary water resistivity gauge fault
8722 XPS W Secondary water resistivity too low
8723 XPS W XPS A/D converter time-out
8725 XPS W X-ray kV mA measurement not calibrated
8726 XPS W X-ray kV calibration slope error
8727 XPS W X-ray kV calibration offset error
8728 XPS W X-ray mA calibration slope error
8729 XPS W X-ray mA calibration offset error
8730 XPS W X-ray kV mA in tolerance time-out
8731 XPS W Status: X-ray arc rate too big
8733 XPS W Low kV limit fault
8734 XPS W Leak current fault
8735 XPS W Secondary water temperature safety open
8738 XPS W De-ionised water tank almost empty.
8739 XPS W Unknown XPS status number
8740 XPS W XPS status sampling period too big
8741 XPS W Status: X-ray kV too low
8742 XPS W Status: X-ray kV too high
8743 XPS W Status: X-ray mA too low
8744 XPS W Status: X-ray mA too high
8745 XPS W Status: primary water input temperature too low
8746 XPS W Status: primary water input temperature too high
8747 XPS W Status: primary water output temperature too low
8748 XPS W Status: primary water output temperature too high
8749 XPS W Status: secondary water input temperature too low
8750 XPS W Status: 2nd water input temp too high
8751 XPS W Status: 2nd water output temp too low
8752 XPS W Status: 2nd water output temp too high
8753 XPS W Status: secondary water resistivity too low
8754 XPS W Status: 2nd water resistivity too high
8755 XPS W Status: X-tube filament current too low
8756 XPS W Status: X-tube filament current too high
8757 XPS W Status: XPS +15V too low

AA83678 Service Manual ARL GenX 4-29


Chapter 4 Troubleshooting

8758 XPS W Status: XPS +15V too high


8759 XPS W Status: XPS -15V too low
8760 XPS W Status: XPS -15V too high
8761 XPS W Status: XPS +24V too low
8762 XPS W Status: XPS +24V too high
8763 XPS W Status: XPS spare 4 too low
8764 XPS W Status: XPS spare 4 too high
8765 XPS W Status: XPS spare 5 too low
8766 XPS W Status: XPS spare 5 too high
8767 XPS W Status: XPS +5V too low
8768 XPS W Status: XPS +5V too high
8769 XPS W Status: XPS spare 7 too low
8770 XPS W Status: XPS spare 7 too high
8771 XPS W Status: X-ray total activity time too low
8772 XPS W Status: X-ray total activity time too high
8773 XPS W Status: arc rate parameter error
8774 XPS W Status: X-ray tube total arc occurrences too low
8775 XPS W Status: X-ray tube total arc occurrences too high
8776 XPS W 60 kV to 70 kV delay too low
8777 XPS W 60 kV to 70 kV delay too high
8800 XPS W SSIO Tx interrupt time-out
8801 XPS W XPS serial A/D converter EOC time-out
8802 XPS W SSIO first Rx interrupt time-out
8803 XPS W SSIO second Rx interrupt time-out
8804 XPS W XPS serial A/D converter channel incoherence
8805 XPS W Secondary water temperatures incoherence
8806 XPS W Desired 2nd water output temp too low
8807 XPS W Desired 2nd water output temp too high
8808 XPS W Water alarms inhibited
8817 XPSLP W Status: laboratory temperature too low
8818 XPSLP W Status: laboratory temperature too high
8819 XPSLP W X-ray tube temperature interlock
8820 XPSLP W Status: X-ray power supply air temperature too low.
8821 XPSLP W Status: X-ray power supply air temperature too high.
8822 XPS W De-ionised water temperature below tolerance
8823 XPS W De-ionised water temperature above tolerance
8824 XPSLP W Eeprom write failure.
8825 XPSLP W X-ray power supply interlock open.
8826 XPSLP W Default temperature offset used.
8827 XPSLP W Unknown firmware parameter command.
8828 XPSLP W Secondary water OFF time-out
8829 XPSLP W X-ray 1st. warning lamp broken.
8830 XPSLP W X-ray 2nd. warning lamp broken.
8831 XPSLP W Status: X-ray emission leakage current too low.
8832 XPSLP W Status: X-ray emission leakage current too high.
8833 XPSLP W X-ray power supply current limitation.
8834 XPSLP W X-ray power supply power limitation.
8835 XPSLP W Status: XPS power stage temperature too low.
8836 XPSLP W Status: XPS power stage temperature too high.
8837 XPSLP W XPS power stage temperature overflow.
8838 XPSLP W XPS power stage temperature underflow.
8839 XPSLP W Status: XPS +12V too low.

4-30 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

8840 XPSLP W Status: XPS +12V too high.


8841 XPSLP W Status: XPS -12V too low.
8842 XPSLP W Status: XPS -12V too high.
8843 XPSLP W Status: XPS +2.5Vref too low.
8844 XPSLP W Status: XPS +2.5Vref too high.
8845 XPSLP W Secondary water resistivity too high
8846 XPSLP W Incoherence between the anode current and the cathode current.
8849 XPSC W XPSC temperature overflow
8850 XPSC W XPSC temperature underflow
8851 XPSC W XPSI temperature1 overflow
8852 XPSC W XPSI temperature1 underflow
8853 XPSC W XPSI temperature2 overflow
8854 XPSC W XPSI temperature2 underflow
8855 XPSC W XPSP temperature1 overflow
8856 XPSC W XPSP temperature1 underflow
8857 XPSC W XPSP temperature2 overflow
8858 XPSC W XPSP temperature2 underflow
8859 XPSC W XPSM temperature1 overflow
8860 XPSC W XPSM temperature1 underflow
8861 XPSC W XPSM temperature2 overflow
8862 XPSC W XPSM temperature2 underflow
8863 XPSSP W Filament current limit
8864 XPSC W X-ray warning lamp broken
8865 XPSC W Leakage current > 500[uA]
8866 XPSSP W Cooling off time-out
8867 XPSC W Unknown reply from DSP
8885 XPSC W Status : XPSC -5V too low
8886 XPSC W Status : XPSC -5V too high
8887 XPSC W Status : XPSI +5V too low
8888 XPSC W Status : XPSI +5V too high
8889 XPSC W Status : XPSI +12V too low
8890 XPSC W Status : XPSI +12V too high
8891 XPSC W Status : XPSI +24V too low
8892 XPSC W Status : XPSI +24V too high
8893 XPSC W Status : XPSI -12V too low
8894 XPSC W Status : XPSI -12V too high
8895 XPSC W Status : XPSI Neutral to GND too low
8896 XPSC W Status : XPSI Neutral to GND too high
8897 XPSC W Status : XPSI Temp1(M1) too low
8898 XPSC W Status : XPSI Temp1(M1) too high
8899 XPSC W Status : XPSI Temp2(M2) too low
8900 XPSC W Status : XPSI Temp2(M2) too high
8901 XPSC W Status : XPSP +5Vm too low
8902 XPSC W Status : XPSP +5Vm too high
8903 XPSC W Status : XPSP +24V too low
8904 XPSC W Status : XPSP +24V too high
8905 XPSC W Status : XPSP -12Vrtm too low
8906 XPSC W Status : XPSP -12Vrtm too high
8907 XPSC W Status : XPSP +12Vrtm too low
8908 XPSC W Status : XPSP +12Vrtm too high
8909 XPSC W Status : XPSP Power temp.(IGBT) too low
8910 XPSC W Status : XPSP Power temp.(IGBT) too high

AA83678 Service Manual ARL GenX 4-31


Chapter 4 Troubleshooting

8911 XPSC W Status : XPSP Board temp. too low


8912 XPSC W Status : XPSP Board temp. too high
8913 XPSC W Status : XPSM +5Vm too low
8914 XPSC W Status : XPSM +5Vm too high
8915 XPSC W Status : XPSM +12V too low
8916 XPSC W Status : XPSM +12V too high
8917 XPSC W Status : XPSM -12V too low
8918 XPSC W Status : XPSM -12V too high

8919 XPSC W Status : XPSM HV divider diff. too low


If < 2V
RS 2097
DY 2,3,11,5,4
Divider assembly defect

8920 XPSC W Status : XPSM HV divider diff. too high


If > 2V
RS 2097
DY 2,3,11,5,4
Divider assembly defect

8921 XPSC W Status : XPSM Filament voltage too low


8922 XPSC W Status : XPSM Filament voltage too high
8923 XPSC W Status: XPSM Filament temp. too low
8924 XPSC W Status: XPSM Filament temp. too high
8925 XPSC W Status : XPSM HV temp. too low
8926 XPSC W Status : XPSM Hv temp. too high
8927 XPSC W Status : Monitoring 400V too low
8928 XPSC W Status : Monitoring 400V too high
8929 XPSC W General status type
8930 XPSC W General status too low
8931 XPSC W General status too high
8932 XPSC W Default switches XPSC used
8933 XPSC W Primary water flow sensor error
8934 XPSC W Primary water already on
9472 XPS I Secondary water temperature regulation ON
9473 XPS I Secondary water temperature regulation OFF
9474 XPS I Water cooler present
9475 XPS I Water cooler absent
9476 XPS I X-ray generator selected is of 3.0 kW type.
9477 XPS I X-ray generator selected is of 3.6 kW type.
9478 XPS I X-ray generator selected is of 4.2 kW type.
9479 XPSLP I Eeprom write successful.
9480 XPSLP I X-ray generator selected is of 600W type.
9481 XPSLP I X-ray generator selected is of 1.2 kW type.
9484 XPSC I X-ray generator selected is of 3.6kW 60kV/120mA type.
9485 XPSC I X-ray generator selected is of 3.6kW 70kV/120mA type.
9486 XPSC I X-ray generator selected is of 4.2kW 60kV/120mA type.
9487 XPSC I X-ray generator selected is of 4.2kW 70kV/140mA type.
9488 XPSC I X-ray generator selected is of 1.2kW 60kV/60mA type.
9489 XPSC I Error X-ray generator selected type.
9728 XPS D No access to XSN network

4-32 Service Manual ARL GenX AA83678


Troubleshooting Chapter 4

9729 XPS D Tx interrupt time-out on XSN network


9730 XPS D Time-out on command size from XQM96
9731 XPS D Time-out on command data string from XQM96
9732 XPS D Time-out on command checksum from XQM96
9733 XPS D Time-out on broadcast command checksum
9734 XPS D String reception retry on checksum error
9735 XPS D Command Rx retry on checksum error
9736 XPS D String emission retry on checksum error
9737 XPS D Reply emission retry on checksum error
9738 XPS D Time-out on reply acknowledge
9739 XPS D No access to display
9740 XPS D Tx interrupt time-out on display
9741 XPS D Power-up on XPS
9742 XPS D X-ray tube spark
9743 XPS D Message size too large on command from XQM96
9744 XPS D XPS 12 bits A/D converter time-out
9745 XPSLP D Serial AD12 bits converter SSIO access denied.
9746 XPSLP D Serial AD8 bits converter SSIO access denied.
9747 XPSLP D Serial DA converter SSIO access denied.
9748 XPSLP D Serial Eeprom SSIO access denied.
9749 XPSLP D Retry on Eeprom write.
9750 XPSLP D X-ray tube filament current warning limit reached.
9751 XPSLP D X-ray tube anode current overflow.
9752 XPSLP D Retry on X-ray tube filament power supply fault.
9753 XPSLP D Retry on X-ray tube filament open.
9754 XPSLP D Retry on X-ray tube filament shorted.
9755 XPSLP D Retry on X-ray tube filament current fatal limit reached.
9756 XPSLP D Retry on secondary water flow sensor error.
9757 XPSLP D X-ray tube filament current regulation in limit.
9758 XPSSP D X-ray PS power limitation
9759 XPSC D X-ray leakage current too high (>1[mA])
9760 XPSC D X-ray leakage current (>500[uA])
9761 XPSC D XPSC First 8 bits A/D converter error
9762 XPSC D XPSC Second 8 bits A/D converter error
9763 XPSC D XPSI 8 bits A/D converter error
9764 XPSC D XPSP 8 bits A/D converter error
9765 XPSC D XPSM 8 bits A/D converter error
9766 XPSC D XPSC DSP Power-up
9767 XPSC D X-ray leakage current error
9768 XPSC D DSP ssio access denied
9769 XPSC D X-ray positive leakage current error
9770 XPSC D X-ray negative leakage current error
9771 XPSC D Primary water flow sensor error

AA83678 Service Manual ARL GenX 4-33


5

FREQUENTLY ASKED
QUESTIONS
Frequently asked Questions Chapter 5

5 FREQUENTLY ASKED QUESTIONS


How do I do to know if the failure is due to the X-ray tube or to the GenX?
Act according to the alarm messages.
Use the terminal command XR 0,0,1,1,1,1 to check the 400V intermediate voltage.
Use the terminal command XR 60,0,1,1,1,1 to check the high voltage.
Unplug the high voltage connector in the GenX and try again the XR 60,0,1,1,1,1.
How do I do to determinate which part is defective?
Act according to the alarm messages.
Measure the power components on the XPSI and XPSP boards.
I cannot switch the GenX on, what is the problem?
Act according to the alarm messages.
After few minutes the GenX switch off itself and generate a thermal alarm, what is the problem?
The LAB could be too hot (above 30 °C)
A fan could be defective.
Too match dust in the GenX.
When I increase the high voltage, I hear a noise near the multiplier, an alarm “PWMTRIP -> Power
Off, issued by DSP” is generate, what is the problem?
Too match dust in the GenX.
Change the multiplier assembly.
When I increase the power, an alarm “DSP trip” is generate, what is the problem?
First try to change the XPSP board (or change the six 1000uF/250V capacitors)
After try to change the XPSI board (or change the ten 1000uF/250V capacitors)
Then try to change the multiplier.
I permanently receive a fatal alarm “XPS communication not possible”, what is the problem?
First try to switch the instrument off few seconds and try again.
Then change the XPSC board and check the power components on the XPSI and XPSP boards.
I receive an alarm « XPSP –10Vrtm too high” or “XPSP +15Vrtm too low”, what is the problem?
Check the breaker 32A on the instrument.
Check the mains voltage 230Vac.
Check the fuse F3 on XPSI board.
Check the fuse F1 on XPSP board.
I receive an alarm « Generators mains voltage not present, time-out”, what is the problem?
First check the breaker 32A and the mains voltage.
After change the XPSI boards and check the XPSP power components.
I receive an alarm « Generator 400V not > 250V, time-out”, what is the problem?
Change the XPSI boards and check the XPSP power components.
I receive an alarm « Generator 400V not > 380V, time-out”, what is the problem?
Change the XPSI boards and check the XPSP power components.
I receive an alarm « Power off, issued by DSP; 400V too low (<350V)”, what is the problem?
Try to switch on again (could be perturbations from the mains voltage).
Change the XPSI boards and check the XPSP power components.

AA83678 Service Manual ARL GenX 5-1


Chapter 5 Frequently asked Questions

What is the life of the GenX?


The lifetime depends directly of the components temperature and especially electrolytic capacitors
temperature. See after the estimate MTBF graphic, which indicate the lifetime according to the lab
temperature and the X ray tube dissipation power.

MTBF (Pow er,kV,Lab. Tem p.)

12

10

20°C or 60 kV
Year

6 25°C or 40 kV
30°C or 30 kV
4

0
2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 4.2
Pow er

Figure 5.1

Remark: The time scale corresponds to a continuous working at constant power, 24 hours per day, and
365 days per year. Using WinXRF “Eco Mode” to decrease the X-ray tube power dissipation
between analyses will drastically increase the lifetime.

I receive an alarm « Mains voltage below 195 Vac (after 20 ms)” or “Mains voltage above 253 Vac
(after 20 ms)”, what is the problem?
Check the mains voltage 230Vac.
If often add a voltage stabilizer.
I receive an alarm « Filament current sensor incoherency”, what is the problem?
Change the XPSM board (or change the Sens1 component on XPSM)
Change the XPSC board.
I receive an alarm « Filament voltage sensor incoherency”, what is the problem?
Change the XPSM board (or change the IC1 and the IC3 components on XPSM).
Change the XPSC board.
I receive an alarm « XPSI input current sensor error”, what is the problem?
Change the XPSM board (or change the Sens2 component on XPSI).
Change the XPSC board.

5-2 Service Manual ARL GenX AA83678


6

GENX 600 W
GENX 600 W Chapter 6

6 GENX 600 W
The GEN’X 600 W working principle is the same as for the high power model.

The 600 W model is air cooled by additive fans installed on XPSI and XPSP boards. Therefore no water
connections are needed for XPSI, XPSP and Multiplier assembly. Nevertheless the water fast fittings are left
on the multiplier assembly.

XPSC Firmware is dedicated to 600 W power. The maximum voltage is 60kV while the emission current is
limited to 30 mA.

J407 for X-RAY On Lamps

Figure 6.1

600 W XPSI Board


The figure below shows the XPSI board thermal exchanger mounted on the solder side. Beside the thermal
exchanger, the board is the same as for a GEN’X 3.6 or 4.2 KW type.

Figure 6.2

AA83678 Service Manual ARL GenX 6-1


Chapter 6 GENX 600 W

600 W XPSP Board


The figure below shows the XPSP board thermal exchanger mounted on the solder side. Beside the thermal
exchanger, the board is the same as for a GEN’X 3.6 or 4.2 KW type.

Figure 6.3

X-ray On Lamps 24 VDC 600 W Generator


Gen’X does support 2 types of X-ray On lamps: 24 VDC or 230VAC

If one lamp of any type is connected, the generator does start.

Nevertheless the standard configuration is described in the next table.

Instrument type X-ray On lamps type Connector


ADVANT’X+ 1200W 24 VDC XPSC J407 WAGO
ADVANT’X XP/XP 230 VAC XPSI J3 WAGO
9800/9800+ XP/XP+ 230 VAC XPSI J3 WAGO
9900 LP 1200W 24 VDC XPSC J407 WAGO
9900 XP/XP+ 24 VDC XPSC J407 WAGO

XPSC 24 VDC X-ray On Lamps control circuit


This circuit allows detecting the X-Ray On Lamps
state.
Both OK.
Lamp1 defect. W Alarm 8829
Lamp2 defect. W Alarm 8830
Both defect. F Alarm 8281
Short-Circuit. F Alarm 8281

Figure 6.4

6-2 Service Manual ARL GenX AA83678


7

GENX RETROFIT
GenX Retrofit Chapter 7

7 GENX RETROFIT
The GenX RETROFIT generator is dedicated to replace previous solid state
generators for field retrofit capability, concerning ARL XRF instruments of the
series 9400 XP /XP+ Series, ADVANTI’X XP / XP+ Series, 9800 XP / XP+ Series.

GenX, Retrofit system overview

The mains XPSI


X-rayPowerSupplyInput
Board

Retrofit XPSR
XPS X-rayPowerSupplyRetrofit
Control Board

XPSC
X-rayPowerSupplyControl
Board XPSB
X-ray
Power
XPSP Supply
X-rayPowerSupplyPower Backplane
Board Board

Filament XPSM
X-rayPowerSupplyMultiplier
Board

HighVoltage Multiplier
Multiplier And
Box Filament
Assembly

Figure 7.1

Two specific generators are available; 3.6 KW 4.2 KW


There part numbers are: S703685T S703687T

Previous parts New parts

DESIGNATION Part N° DESIGNATION Part N°


GENERATOR,X-RAY,70KV,3.6KW,PR S702825 RetGenX3.6kW70kV S703685T
GENERATOR,XRAY,3.6KW,Refurbished S702825R RetGenX3.6kW70kV,Refurbished S703685R
GENERATOR,X-RAY,70KV,140MA,PR S703247 RetGenX4.2kW70kV S703687T
GENERATOR,X-RAY,4.2KW,Refurbished S703247R RetGenX4.2kW70KV,Refurbished S703687R

AA83678 Service Manual ARL GenX 7-1


Chapter 7 GenX Retrofit

An extra board called XPSR is installed inside the generator. This board will be used as emulator between
the GenX generator and the XPS or XRPS CONTROL BOARD plugged into the instrument electronic rack.

RetGenX does emulate non GenX 3.6 / 4.2 kW generators.

The secondary water circuit has to be adapted to supply the GenX power stages heat exchangers.

XPSR BOARD XPS INTERCONNECTION CONNECTOR

Figure 7.2

7-2 Service Manual ARL GenX AA83678


GenX Retrofit Chapter 7

Interconnections
Instrument Type Schematic name Schematic N°
9400 / ADVANT’X Series XPS C.B. INTERCONNECTION (9400) S214071
9800 SERIES XPS C.B. INTERCONNECTION (9800) S214433

Remark: These schematics are part of the: 9400 / ADVANT’X schematic file AA83484
9800 schematic file AA83400

Interconnection bloc diagram

3 x 21 P CABLES
M5
S U B-D S UB-D
50 P J1 54 M6
50 P J4 01
M7

E L E C TR O NIC
R AC K
XP S C . B . B A CK P L A NE XP SR

XP SC R E T DISPL A Y
XP SI
IN S TR U M E N T G EN 'X R
X PSP
EL EC T R ON IC
XP SM
RACK

Figure 7.3

Remark:

On GenX RETROFIT, XPSC becomes: XPSCRET Part N° S703784T

AA83678 Service Manual ARL GenX 7-3


Chapter 7 GenX Retrofit

XPSR board

Figure 7.4

Two specific XPSR Boards are available; 3.6 KW 4.2 KW


The Part numbers are: S703125T S703785T

XPSR bloc diagram

RetGenX

XPS 50P Cable Connector uC


Board SUB-D 50P ADUC812
XPSC
Display
Board
XPSR
LVPS Board

Safety
And
Water
interface

Figure 7.5

XPSR IS DIVIDED IN THREE MAIN PARTS


1. Low Voltage power Supply
2. XPS Generator Emulation
3. Display

7-4 Service Manual ARL GenX AA83678


GenX Retrofit Chapter 7

Part 1 Low voltage power supplies


XPS (plugged into the instrument electronic rack)
+ 5 VoltsOut regulated
Low voltage power supplies for XPS +15 VoltsOut regulated
-15 VoltsOut regulated
+ 24 Volts
XPSR (plugged into GEN’X rack)
+ 5 Volts regulated
Low voltage power supplies for XPSR +12 Volts regulated
-12 Volts regulated
+ 24 Volts
GenX backplane low Voltage power supplies
+ 5 VoltsOut regulated
Low voltage power supplies for mother board +12 VoltsOut regulated
-12 VoltsOut regulated
+ 24 VoltsOut From Interlock RelayRY1

Low voltage power supply circuit


The low voltage power supply generator is supplied by the 230VAC Line1 and Line2 via XPSI Board.

The 230VAC is converted to 24VDC relative to the following bloc diagram

Figure 7.6

AA83678 Service Manual ARL GenX 7-5


Chapter 7 GenX Retrofit

Voltage regulator bloc diagram TOP 247

Figure 7.7

The regulated 24 VDC is applied to five low voltage regulators used to generate respectively

+15 VDC
-15 VDC
+5 VDC
+12 VDC
- 12 VDC

All low voltages are protected by Multi-fuses.

7-6 Service Manual ARL GenX AA83678


GenX Retrofit Chapter 7

+ 24 VDC Safety circuit


The + 24 V is connected to the safety relay RY1.

If the signal “/InterlockOk” is true, the potential is 0 V.

The relay RY1 drives the + 24V its contacts to + 24Vout.

+24Vout does supply the 24 V to all the boards plugged on the GenX mother board.

If the interlock is not OK the +24Vout is not supplied to the boards;

1. Following alarms are issued by the concerned boards.

8345 XPSC F Status : XPSI +24V too low


8346 XPSC F Status : XPSI +24V too high
8347 XPSC F Status : XPSC +24V too low
8348 XPSC F Status : XPSC +24V too high
8349 XPSC F Status : XPSP +24V too low
8350 XPSC F Status : XPSP +24V too high

2. XPSI main contactor K1 does not drive.

Following alarms are issued:

8293 XPSC F Generator 400V not > 250V, time-out


8294 XPSC F Generator 400V not > 380V, time-out
8305 XPSC F Power off, issued by DSP; 400V too low (<350V)
8308 XPSC F Mains voltage below 195 VAC (after 20 ms)

In any situation described above, the generator is shut down.

Jumper W2
In normal operation the jumper W2 must be removed.
Installing the jumper leads to human harm due to X-ray emission and/or electrocution.

AA83678 Service Manual ARL GenX 7-7


Chapter 7 GenX Retrofit

Part 2 XPS Generator Emulation

XPS XPSR GEN'X

EMULATOR XPSCRET
ADUC812

Outgoing
Signals To/From
Bus XPSCRET
Serial Data Link
Ingoing
Signals
Bus

Figure 7.8

The task of the µ Controller ADUC812 is to simulate a non GenX generator.

To do so:
• The various alarms generated by XPSC are linked to the ADCU812 processor through a serial link
and converted, to be recognised by XPS.
• The start up synchronisation and safety control signals generated by XPS are converted to serial
Data and transmitted to XPSC.
• The analogic Ref signals are converted to serial Data by the ADUC812 built in ADC
• To be transmitted to XPSC.
• The Monitoring signals are converted from serial Data by the ADUC812 built in DAC to analogic
voltages to be supervised by XPS.

Serial link
Serial link is of the same type as XSN (X-ray Services Network) used in the 9800 / ADVANT’X series
instruments electronic racks.

The jumper W1 must be placed between contacts 1-2 internal, for proper communication.

The serial communication is watchdog controlled. In case of communication timeout or if a communication


error is detected, XPSC does shut down the generator.

The following alarm is displayed.

8270 XPSC F SSIO communication timeout.

7-8 Service Manual ARL GenX AA83678


GenX Retrofit Chapter 7

Jumper settings
JUMPER SETTING
W1 1-2 Internal
W2 Removed
JP1 EnProg Removed

Figure 7.9

AA83678 Service Manual ARL GenX 7-9


Chapter 7 GenX Retrofit

Part 3 Display
Schematic and layout: S215199

Display Display panel

Figure 7.10

The 1x 20 characters display is mounted and plugged on the display front panel.

The display front panel is plugged on both external side connectors.

Display Display front panel

Figure 7.11 Display

The display data are controlled by XPSCRET.

7-10 Service Manual ARL GenX AA83678


GenX Retrofit Chapter 7

GenX Retrofit Trouble Shooting Procedure


The causes of an untimely GenX breakdown can be determined from the ACS software alarms or by the
XPSCRET internal alarms posted on the GenX front panel display.

Bellow find the list of the ACS XPS alarms and the corresponding GenX alarms that can be monitored on the
GenX display.

KV regulation error
8273 XPS F X-ray kV out of regulation

Indicate the "kV" regulation error.


The KV regulation is out of tolerance. -20% or +5% of the set point.
Displayed alarm issued by GenX:

Number Description
8273 F_KV_OUT_OF_REGULATION

MA regulation error
8274 XPS F X-ray mA out of regulation

Indicate the "mA" regulation error.


The mA regulation is out of tolerance. -20% or +10% of the set point.
Displayed alarm issued by GenX:

Number Description
8274 F_MA_OUT_OF_REGULATION

Over voltage safety


8211 XPS F Over kilovolts on X-ray power supply

Indicate the "KV" too high, above 60kV or above 70kV, depending of the GenX type.
KV too high supervised by the DSP for a fast control.
Displayed alarm issued by GenX:

Number Description
8197 F_KV_TOO_HIGH
8306 F_HV_TOO_HIGH_DSP

Over emission current safety


8212 XPS F Over milliamps on X-ray power supply

Indicate the "Ma" too high, > 120mA or > 140mA, depending of the GenX type.
Displayed alarms issued by GenX:

Number Description
8199 F_MA_TOO_HIGH
8306 F_MA_TOO_LOW

AA83678 Service Manual ARL GenX 7-11


Chapter 7 GenX Retrofit

Over Load
8213 XPS F Over power on X-ray power supply

Indicate the consign power has exceeded the power reference.


The alarm Over Load is generated by the card XPSR.

Filament current limit


8216 XPS F X-ray tube filament over current

Indicate the filament current is > 12A.


Displayed alarms issue by GenX:

Number Description
8276 F_SAFETY_FIL_CUR_LIMIT_REACHED
8283 F_SHORTED_FILAMENT

Open circuit on filament


8215 XPS F X-ray tube filament open

Indicate the filament is open circuit.


Displayed alarm issued by GenX:

Number Description
8215 F_OPEN_FILAMENT

X-ray lamps defect


8219 XPS F X-ray ON lamp broken

Indicate the lamps are broken or shorted.


Displayed alarms issued by GenX:

Number Description
8281 F_WARNING_LAMP
8377 F_SHORTED_LAMP

At start-up, if the X-ray On lamps are defect, the buzzer does beep three times to indicate
the error.

Low KV limit
8220 XPS F Low kilovolts limit

Indicate the condition KV is prohibited compared to the power curve.


Displayed alarm issued by GenX:

Number Description
8196 F_KV_TOO_LOW

7-12 Service Manual ARL GenX AA83678


GenX Retrofit Chapter 7

Power Curve:

150
3.0 kW
140
130 3.6 kW

120 4.2 kW
110
100
90
80
mA

70
60
50
40
30
20
10
0
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75

kV

Figure 7.12

Leakage current limit


8221 XPS F Leak current too high

Indicate the current leakage is above 5mA or indicate the absolute current leakage is above 1,5mA with retry.
Displayed alarms issued by GenX:

Number Description
8304 F_X_RAY_LEAK_CURRENT_TOO_HIGH
8389 F_XRAY_NEG_LEAK_CURRENT_ERROR
8390 F_XRAY_POS_LEAK_CURRENT_ERROR

HV Inverter error (HVFlt)

8217 XPS F Kilovolts regulation fault

Indicate an error with the XPSP sub-assembly.


Displayed alarms issued by GenX:

Number Description
8295 F_TEMP_IGBT_TOO_HIGH
8300 F_XPSP_AD8BIT_CONV
8314 F_TEMP_XPSP_TOO_HIGH
8316 F_TEMP_HV_TOO_HIGH
8329 F_PLUS_5_VM_XPSP_TOO_LOW
8330 F_PLUS_5_VM_XPSP_TOO_HIGH

AA83678 Service Manual ARL GenX 7-13


Chapter 7 GenX Retrofit

8349 F_PLUS_24_VOLT_XPSP_TOO_LOW
8350 F_PLUS_24_VOLT_XPSP_TOO_HIGH
8368 F_TEMP1_XPSP_OVERFLOW
8369 F_TEMP1_XPSP_UNDERFLOW
8370 F_TEMP2_XPSP_OVERFLOW
8371 F_TEMP2_XPSP_UNDERFLOW
8378 F_MINUS_10_VRTM_TOO_LOW
8379 F_MINUS_10_VRTM_TOO_HIGH
8380 F_PLUS_15_VRTM_TOO_LOW
8381 F_PLUS_15_VRTM_TOO_HIGH

Fil Inverter error (FilFlt)


8218 XPS F Milliamps regulation fault

Indicate an error with the XPSM sub-assembly.


Displayed alarms issued by GenX:

Number Description
8282 F_FILAMENT_POWER_SUPPLY_FAULT
8301 F_XPSM_AD8BIT_CONV
8315 F_TEMP_FILAMENT_TOO_HIGH
8323 F_FILAMENT_CURR_SENSOR_INCOHERENCY
8324 F_FILAMENT_VOLT_SENSOR_INCOHERENCY
8331 F_PLUS_5_VM_XPSM_TOO_LOW
8332 F_PLUS_5_VM_XPSM_TOO_HIGH
8337 F_PLUS_12_VOLT_XPSM_TOO_LOW
8338 F_PLUS_12_VOLT_XPSM_TOO_HIGH
8343 F_MINUS_12_VOLT_XPSM_TOO_LOW
8344 F_MINUS_12_VOLT_XPSM_TOO_HIGH
8372 F_TEMP1_XPSM_OVERFLOW
8373 F_TEMP1_XPSM_UNDERFLOW
8374 F_TEMP2_XPSM_OVERFLOW
8375 F_TEMP2_XPSM_UNDERFLOW

Arcs detection
8244 XPS F Status: X-ray arc rate too big

Indicates that an arc has occurred within the previous second (If more than 3 arcs are detected in 10 seconds,
XPS does shutdown the GenX). No alarms are displayed.

Interlock closed
Indicates the EXTERNAL INTERLOCK connections are in the closed position. This also indicates closure
of internal power supply interlocks. The generator turns OFF if one of the contacts is opened or if the
spectrometer pressure rises above 25kPa.

XPS Interlock related alarms are uploaded to the ACS depending of the instrument type.

7-14 Service Manual ARL GenX AA83678


GenX Retrofit Chapter 7

Interlock opened
Indicates the EXTERNAL INTERLOCK connections are in the open position. This also indicates opening of
internal power supply interlocks. The generator turns OFF if one of the contacts is opened or if the
spectrometer pressure rises above 25kPa.

If all contacts are closed and if the spectrometer pressure goes below 25kPa, it is possible to start the
generator.

XPS Interlock related alarms are uploaded to the ACS depending of the instrument type.
Power supply miscellaneous faults
8214 XPS F X-ray power supply fault

It Indicates a GenX error.


Displayed alarms issued by GenX:
Number Description
8192 UNKNOWN_CMD_ERROR
8199 F_POWER_TOO_HIGH
8201 F_ILLEGAL_POWER_CHANGE_MODE
8202 F_ILLEGAL_POWER_CHANGE_SPEED
8203 F_ILLEGAL_ANODE
8210 F_XRAY_ON_ACK_TIMEOUT
8244 F_XRAY_ARC_RATE
8257 F_UNKNOWN_XRAY_PS_TYPE
8270 F_SSIO_TIMEOUT
8271 F_AD12_CONV_TIMEOUT
8272 F_AD12_AUTOCAL_TIMEOUT
8282 F_FILAMENT_POWER_SUPPLY_FAULT
8292 F_POWER_ON_TIMEOUT
8293 F_WAIT_RECTIFIER_INPUTVOLTAGE
8294 F_WAIT_POWER_ON_TIMEOUT
8295 F_TEMP_IGBT_TOO_HIGH
8297 F_XPSC_FIRST_AD8BIT_CONV
8298 F_XPSC_SECOND_AD8BIT_CONV
8299 F_XPSI_AD8BIT_CONV
8300 F_XPSP_AD8BIT_CONV
8301 F_XPSM_AD8BIT_CONV
8302 F_TEMP_MOSFET_M1_TOO_HIGH
8303 F_TEMP_MOSFET_M2_TOO_HIGH
8305 F_MON400V_DSP
8307 F_CURRENT_TOO_HIGH_DSP
8308 F_INPUTVOLTAGE_TOO_LOW_DSP
8309 F_INPUTVOLTAGE_TOO_HIGH_DSP
8310 F_UNKNOWN_REPLY_FROM_DSP
8311 F_TOO_MANY_PARAM_IN_SSIO_CMD
8312 F_UNKNOWN_DSP_CMD
8313 F_PWMTRIP_DSP
8314 F_TEMP_XPSP_TOO_HIGH
8315 F_TEMP_FILAMENT_TOO_HIGH
8316 F_TEMP_HV_TOO_HIGH
8317 F_TEMP_XPSC_TOO_HIGH
8323 F_FILAMENT_CURR_SENSOR_INCOHERENCY

AA83678 Service Manual ARL GenX 7-15


Chapter 7 GenX Retrofit

8324 F_FILAMENT_VOLT_SENSOR_INCOHERENCY
8325 F_PLUS_5_VOLT_XPSI_TOO_LOW
8326 F_PLUS_5_VOLT_XPSI_TOO_HIGH
8327 F_PLUS_5_VOLT_XPSC_TOO_LOW
8328 F_PLUS_5_VOLT_XPSC_TOO_HIGH
8329 F_PLUS_5_VM_XPSP_TOO_LOW
8330 F_PLUS_5_VM_XPSP_TOO_HIGH
8331 F_PLUS_5_VM_XPSM_TOO_LOW
8332 F_PLUS_5_VM_XPSM_TOO_HIGH
8333 F_PLUS_12_VOLT_XPSI_TOO_LOW
8334 F_PLUS_12_VOLT_XPSI_TOO_HIGH
8335 F_PLUS_12_VOLT_XPSC_TOO_LOW
8336 F_PLUS_12_VOLT_XPSC_TOO_HIGH
8337 F_PLUS_12_VOLT_XPSM_TOO_LOW
8338 F_PLUS_12_VOLT_XPSM_TOO_HIGH
8339 F_MINUS_12_VOLT_XPSI_TOO_LOW
8340 F_MINUS_12_VOLT_XPSI_TOO_HIGH
8341 F_MINUS_12_VOLT_XPSC_TOO_LOW
8342 F_MINUS_12_VOLT_XPSC_TOO_HIGH
8343 F_MINUS_12_VOLT_XPSM_TOO_LOW
8344 F_MINUS_12_VOLT_XPSM_TOO_HIGH
8345 F_PLUS_24_VOLT_XPSI_TOO_LOW
8346 F_PLUS_24_VOLT_XPSI_TOO_HIGH
8347 F_PLUS_24_VOLT_XPSC_TOO_LOW
8348 F_PLUS_24_VOLT_XPSC_TOO_HIGH
8349 F_PLUS_24_VOLT_XPSP_TOO_LOW
8350 F_PLUS_24_VOLT_XPSP_TOO_HIGH
8353 F_MINUS_5_VOLT_XPSC_TOO_LOW
8354 F_MINUS_5_VOLT_XPSC_TOO_HIGH
8355 F_MON_400_VOLT_TOO_LOW
8356 F_MON_400_VOLT_TOO_HIGH
8360 F_XPSI_AD_MON400V_ERROR
8362 F_XPSC_TEMP_OVERFLOW
8363 F_XPSC_TEMP_UNDERFLOW
8364 F_TEMP1_XPSI_OVERFLOW
8365 F_TEMP1_XPSI_UNDERFLOW
8366 F_TEMP2_XPSI_OVERFLOW
8367 F_TEMP2_XPSI_UNDERFLOW
8368 F_TEMP1_XPSP_OVERFLOW
8369 F_TEMP1_XPSP_UNDERFLOW
8370 F_TEMP2_XPSP_OVERFLOW
8371 F_TEMP2_XPSP_UNDERFLOW
8372 F_TEMP1_XPSM_OVERFLOW
8373 F_TEMP1_XPSM_UNDERFLOW
8374 F_TEMP2_XPSM_OVERFLOW
8375 F_TEMP2_XPSM_UNDERFLOW
8378 F_MINUS_10_VRTM_TOO_LOW
8379 F_MINUS_10_VRTM_TOO_HIGH
8380 F_PLUS_15_VRTM_TOO_LOW
8381 F_PLUS_15_VRTM_TOO_HIGH
8382 F_NEUTRAL_TO_GND_TOO_LOW
8383 F_NEUTRAL_TO_GND_TOO_HIGH
8384 F_XPSI_INPUT_CURRENT_ERROR

7-16 Service Manual ARL GenX AA83678


GenX Retrofit Chapter 7

GenX Installation procedure

Installation procedures for boards installed in the GenX generator

P/N Description Procedure N°


S703123T XPSI (Water cooling version) S430414
S703778T XPSI (Air cooling version) S430427
S703126T XPSP (Water cooling version) S430415
S703779T XPSP (Air cooling version) S430428
S703125T XPSR36 S430418
S703785T XPSR42 S430419
S703784T XPSCRET S430426
S703726T XPSC1260 S430424
S704026T XPSC2560 S430661
S703722T XPSC3660 S430416
S703723T XPSC3670 S430421
S703724T XPSC4260 S430422
S703725T XPSC4270 S430423
S703758T Multiplier Assembly S430420
S703721T XPSCD Board S430824
S703876T XPSRD Board S430825
S703953T XRD Multiplier Assembly S430828
Update Firmware Generator X-Ray S438625

AA83678 Service Manual ARL GenX 7-17


Document: Installation procedure XPSI board S703123T

X-ray Power Supply Input XPSI Board

Water Cooling Version

Instrument : XRF
N° Assembly : S703123T
File : S430414-1.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1. Change air direction of fan 2 6.07.05 JMa
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 1 / 6
Document: Installation procedure XPSI board S703123T

1 Introduction:

XPSI board is cooled with water.


Water block is installed on the XPSI board solder side.
2x water tubes with fast fitting male and female connectors are connected
to the water block.
2x fans are installed on the XPSI board

2 Board configuration:

Below is the layout of the XPSI board:


Fan 2 installed on Connector P2. Connector P1.
Connector J2. the I1 filter. Connector J902.
Components side Solder side
Components side Components side Components side

Air direction <-----


from the fan 1. < - - - - -
Solder side <-----

Connector J3. COMPONENTS VIEW Fan 1 installed on


Components side the solder side

XPSI board has no bridge or no switch to be configured.

XPSI board has 3x electrical external connectors:


Connector identification: Connected to:
J2 -> Input Power connector 230VAC Wago 6x
J3 -> X-Ray ON lamps with version 230VAC
J902 -> GenX backplane board XPSB

XPSI board has 2x fans:


Fan identification: Connected to:
-> P1 connector
Fan 1
Red color of the fan cable is connected to: +24VDC
Installed on the solder side
Black color of the fan cable is connected to: PGND
Fan 2 -> P2 connector
Installed on the components Red color of the fan cable is connected to: +24VDC
side Black color of the fan cable is connected to: PGND
Note:
1) Air direction of the fan 1 is oriented as shown as the above layout drawing
2) Fan 2 is sucking air from the XPSI I1 filter component.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 2 / 6
Document: Installation procedure XPSI board S703123T

3 Firmware compatibility
No firmware is installed on the XPSI board.

4 Board installation

IMPORTANT:
Before to have access inside the GenX generator, the following caution
must be read carefully in order to avoid any trouble occurring to the
generator and any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 3 / 6
Document: Installation procedure XPSI board S703123T

4.1 Location of the XPSI board inside the GenX generator:

XPSI XPSR XPSC XPSP Multiplier


Board Board Board Board assembly
J2
J3

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 4 / 6
Document: Installation procedure XPSI board S703123T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSI board from the rear side of the
GenX.
XPSI board is connected to a unique backplane board XPSB located near the
GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- Use the water tube connectors, located at the end of the board, to extract the
XPSI board from its location.
Pull the XPSI board very slightly without a big force to avoid to damage the
fixing of the water cooling block on the XPSI board.
Take care, as mentioned above, to avoid to damage electronic components
located on the way during removal of the XPSI board.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 5 / 6
Document: Installation procedure XPSI board S703123T

- Install the new XPSI board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSI board.
VERY IMPORTANT: The XPSI board must be installed at the final locked
mechanical position, by passing through two “distinct clicks positions”
when XPSI board is pushing to the final position.

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

4.3 Calibration:

No kV/mA calibration is required when XPSI board is replaced.

5 Special packaging for returned material:


The XPSI board requires special packaging for transportation to avoid damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430414 1
Page 6 / 6
Document: Installation procedure XPSI board Air Cooling S703778T

X-ray Power Supply Input XPSI Board

Air Cooling Version

Instrument : XRF
N° Assembly : S703778T
File : S430427-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 1 / 6
Document: Installation procedure XPSI board Air Cooling S703778T

1 Introduction:

XPSI board with P/N: S703778T is cooled with air flow.


Air exchanger radiator is installed along the XPSI board on solder side.

1x fans is installed on the XPSI board

2 Board configuration:

Below is the layout of the XPSI board Air cooling version:

Connector P1.
Connector J2. Connector J902.
Solder side
Components side Components side

Air direction <-----


from the fan 1. < - - - - -
Solder side <-----

Connector J3. COMPONENTS VIEW Fan 1 installed on


Components side the solder side

XPSI board has no bridge or no switch to be configured.

XPSI board has 3x electrical external connectors:


Connector identification: Connected to:
J2 -> Input Power connector 230VAC Wago 6x
J3 -> X-Ray ON lamps with version 230VAC
J902 -> GenX backplane board XPSB

XPSI board has 2x fans:


Fan identification: Connected to:
-> P1 connector
Fan 1
Red color of the fan cable is connected to: +24VDC
Installed on the solder side
Black color of the fan cable is connected to: PGND
Note:
1) Air direction of the fan 1 is oriented as shown as the above layout drawing

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 2 / 6
Document: Installation procedure XPSI board Air Cooling S703778T

3 Firmware compatibility
No firmware is installed on the XPSI board.

4 Board installation

IMPORTANT:
Before to have access inside the GenX generator, the following caution
must be read carefully in order to avoid any trouble occurring to the
generator and any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 3 / 6
Document: Installation procedure XPSI board Air Cooling S703778T

4.1 Location of the XPSI board inside the GenX generator:

XPSI Board XPSP Board


Air cooling XPSR XPSC Multiplier
Air cooling
version Board Board assembly
version
J2
J3

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 4 / 6
Document: Installation procedure XPSI board Air Cooling S703778T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSI board from the rear side of the
GenX.
XPSI board is connected to a unique backplane board XPSB located near the
GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- Use your left hand (Put your thumb on the J3 connector and your others fingers
on the solder XPSI board side) to extract the the XPSI board from its location.
Pull the XPSI board very slightly without a big force to avoid to damage the J3
connector.
Take care, as mentioned above, to avoid to damage electronic components
located on the way during removal of the XPSI board.

- Install the new XPSI board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSI board.
VERY IMPORTANT: The XPSI board must be installed at the final locked
mechanical position, by passing through two “distinct clicks positions”
when XPSI board is pushing to the final position.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 5 / 6
Document: Installation procedure XPSI board Air Cooling S703778T

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

4.3 Calibration:

No kV/mA calibration is required when XPSI board is replaced.

5 Special packaging for returned material:


The XPSI board requires special packaging for transportation to avoid damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430427 0
Page 6 / 6
Document: Installation procedure XPSP board S703126T

X-ray Power Supply Power XPSP Board

Water Cooling Version

Instrument : XRF
N° Assembly : S703126T
File : S430415-1.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1. Change air direction of fan 2 6.07.05 JMa
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 1 / 6
Document: Installation procedure XPSP board S703126T

1 Introduction:

XPSP board is cooled with water.


Water block is installed on the XPSP board solder side.
2x water tubes with fast fitting male and female connectors are connected
to the water block.
2x fans are installed on the XPSP board.

2 Board configuration:

Below is the layout of the XPSP board:

Fan 2 installed on Connector P2 on Connector P1 on Connector J905.


the component side component side solder side Components side

Air direction < - - - -


from the fan 1. < - - - -
Solder side <----

COMPONENTS VIEW Fan 1 installed on


the solder side

XPSP board has no bridge or no switch to be configured.

XPSP board has 1x electrical external connector:


Connector identification: Connected to:
J905 -> GenX backplane board XPSB

XPSP board has 2x fans:


Fan identification: Connected to:
-> P1 connector
Fan 1
Red color of the fan cable is connected to: +24VDC
Installed on the solder side
Black color of the fan cable is connected to: PGND
Fan 2 -> P2 connector
Installed on the components Red color of the fan cable is connected to: +24VDC
side Black color of the fan cable is connected to: PGND
Note:
1) Air direction of the fan 1 is oriented as shown as the above layout drawing.
2) Fan 2 is sucking air from the XPSP board.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 2 / 6
Document: Installation procedure XPSP board S703126T

3 Firmware compatibility

No firmware is installed on the XPSP board.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:

Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 3 / 6
Document: Installation procedure XPSP board S703126T

4.1 Location of the XPSP board inside the GenX generator:

XPSI XPSR XPSC XPSP Multiplier


Board Board Board Board assembly

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 4 / 6
Document: Installation procedure XPSP board S703126T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSP board from the rear side of the
GenX.
Important: Due to the very closed proximity of the XPSC, XPSR (if installed)
and XPSP boards, those 2x (3x if XPSR present) boards must be removed
following the below order to avoid damage on the electronic components part:
-> First two boards to be removed are -> XPSR (if installed) together with
XPSC board.
-> Second board to be removed is -> XPSP board.
XPSP board is connected to a unique backplane board XPSB located near the
GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- Use the water tube connectors, located at the end of the board, to extract the
XPSP board from its location.
Pull the XPSP board very slightly without a big force to avoid to damage the
fixing of the water cooling block on the XPSP board.
Take care, as mentioned above, to avoid to damage electronic components
located on the way during removal of the XPSP board.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 5 / 6
Document: Installation procedure XPSP board S703126T

- Install the new XPSP board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during removal of the XPSP board.
VERY IMPORTANT: The XPSP board must be installed at the final locked
mechanical position, by passing through two “distinct clicks positions”
when XPSP board is pushing to the final position.

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

4.3 Calibration:

No kV/mA calibration is required when XPSP board is replaced.

5 Special packaging for returned material:


The XPSP board requires special packaging for transportation to avoid damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430415 1
Page 6 / 6
Document: Installation procedure XPSP board Air Cooling S703779T

X-ray Power Supply Power XPSP Board

Air Cooling Version

Instrument : XRF
N° Assembly : S703779T
File : S430428-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 1 / 6
Document: Installation procedure XPSP board Air Cooling S703779T

1 Introduction:

XPSP board with P/N: S703779T is cooled with air flow.


Air exchanger radiator is installed along the XPSP board on solder side.

1x fans are installed on the XPSP board.

2 Board configuration:

Below is the layout of the XPSP board Air cooling version:

Connector P1 on Connector J905.


solder side Components side

Air direction < - - - -


from the fan 1. < - - - -
Solder side <----

COMPONENTS VIEW Fan 1 installed on


the solder side

XPSP board has no bridge or no switch to be configured.

XPSP board has 1x electrical external connector:


Connector identification: Connected to:
J905 -> GenX backplane board XPSB

XPSP board has 2x fans:


Fan identification: Connected to:
-> P1 connector
Fan 1
Red color of the fan cable is connected to: +24VDC
Installed on the solder side
Black color of the fan cable is connected to: PGND
Note:
Air direction of the fan 1 is oriented as shown as the above layout drawing.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 2 / 6
Document: Installation procedure XPSP board Air Cooling S703779T

3 Firmware compatibility

No firmware is installed on the XPSP board.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:

Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 3 / 6
Document: Installation procedure XPSP board Air Cooling S703779T

4.1 Location of the XPSP board inside the GenX generator:

XPSI Board XPSP Board


Air cooling XPSR XPSC Multiplier
Air cooling
version Board Board assembly
version

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 4 / 6
Document: Installation procedure XPSP board Air Cooling S703779T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSP board from the rear side of the
GenX.
Important: Due to the very closed proximity of the XPSC, XPSR (if installed)
and XPSP boards, those 2x (3x if XPSR present) boards must be removed
following the below order to avoid damage on the electronic components part:
-> First two boards to be removed are -> XPSR (if installed) together with
XPSC board.
-> Second board to be removed is -> XPSP board.
XPSP board is connected to a unique backplane board XPSB located near the
GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- Use the Transformer T1 (located on the XPSP board) to extract, with your
fingers, the XPSP board from its location.
Pull the XPSP board very slightly without a big force to avoid to damage the
transformer T1.
Take care, as mentioned above, to avoid to damage electronic components
located on the way during removal of the XPSP board.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 5 / 6
Document: Installation procedure XPSP board Air Cooling S703779T

- Install the new XPSP board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during removal of the XPSP board.
VERY IMPORTANT: The XPSP board must be installed at the final locked
mechanical position, by passing through two “distinct clicks positions”
when XPSP board is pushing to the final position.

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

4.3 Calibration:

No kV/mA calibration is required when XPSP board is replaced.

5 Special packaging for returned material:


The XPSP board requires special packaging for transportation to avoid damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430428 0
Page 6 / 6
Document: Installation procedure XPSR36 board S703125T

X-ray Power Supply Retrofit XPSR36 Board

Instrument : XRF
N° Assembly : S703125T
File : S430418-1.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1. Change location of bridges W1 / W2 1.03.06 JMa
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 1 / 6
Document: Installation procedure XPSR36 board S703125T

1 Introduction:
IMPORTANT:
XPSR36 board with P/N: S703125T must be installed ONLY in the GenX RET
generator 3.6kW with P/N: S703685T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note: No fan and no water block are installed on XPSR36 board.

2 Board configuration:

Below is the layout of the XPSR36 board:

Connector P108. Bridge Bridge Connector J903.


Components side. W1. W2. Components side.

1 1
2
2

Connector J401. Bridge Components view


Componants side. JP1.

XPSR36 board bridges configuration:


Bridge State Description
W1-1 INSTALLED XPSR36 setting
W1-2 REMOVED XPSR36 setting

Bridge State Description


W2-1 REMOVED XPSR36 setting
W2-2 REMOVED XPSR36 setting

Bridge State Description


JP1 REMOVED XPSR36 setting
Note:
The factory setting is in bold.

XPSR36 board has 3x electrical external connectors:


Connector identification: Connected to:
P108 -> Not used
J401 -> XRF instrument electronic rack
J903 -> GenX backplane board XPSB

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 2 / 6
Document: Installation procedure XPSR36 board S703125T

3 Firmware compatibility

Note: Firmware XPSR36 is downloaded in factory.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 3 / 6
Document: Installation procedure XPSR36 board S703125T

4.1 Location of the XPSR36 board inside the GenX generator:

XPSI XPSR36 XPSCRET XPSP Multiplier


Board Board Board Board assembly

P 108
J 401

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side
TOP side Right side
Left
side

Front side
BOTTOM side

RET GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 4 / 6
Document: Installation procedure XPSR36 board S703125T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSR36 board together with the
XPSCRET board due to the closed proximity of those two boards to avoid
damage on the electronic components part.
XPSR36 board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extract the XPSR36 board from its location, let plugged the male J401
connector and use it to pull the board out.
To extract the XPSCRET board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 5 / 6
Document: Installation procedure XPSR36 board S703125T

- Install the new XPSR36 board, together with the XPSCRET, in the correct
location, see chapter 4.1 of this document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSR36 and XPSCRET boards.
VERY IMPORTANT: The XPSR36 and XPSCRET boards must be installed
at the final locked mechanical position, by passing through two “distinct
clicks positions” when XPSR36 and XPSCRET boards are pushing to the
final position.

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

4.3 Calibration:

No kV/mA calibration is required when XPSR36 is replaced.

5 Special packaging for returned material:


The XPSR36 board requires special packaging for transportation to avoid damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430418 1
Page 6 / 6
Document: Installation procedure XPSR42 board S703785T

X-ray Power Supply Retrofit XPSR42 Board

Instrument : XRF
N° Assembly : S703785T
File : S430419-1.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1. Change location of bridges W1 / W2 1.03.06 JMa
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 1 / 6
Document: Installation procedure XPSR42 board S703785T

1 Introduction:
IMPORTANT:
XPSR42 board with P/N: S703785T must be installed ONLY in the GenX RET
generator 4.2kW with P/N: S703687T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note: No fan and no water block are installed on XPSR42 board.

2 Board configuration:

Below is the layout of the XPSR42 board:

Connector P108. Bridge Bridge Connector J903.


Components side. W1. W2. Components side.

1 1
2
2

Connector J401. Bridge Components view


Componants side. JP1.

XPSR42 board bridges configuration:


Bridge State Description
W1-1 INSTALLED XPSR42 setting
W1-2 REMOVED XPSR42 setting

Bridge State Description


W2-1 REMOVED XPSR42 setting
W2-2 REMOVED XPSR42 setting

Bridge State Description


JP1 REMOVED XPSR42 setting
Note:
The factory setting is in bold.

XPSR42 board has 3x electrical external connectors:


Connector identification: Connected to:
P108 -> Not used
J401 -> XRF instrument electronic rack
J903 -> GenX backplane board XPSB

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 2 / 6
Document: Installation procedure XPSR42 board S703785T

3 Firmware compatibility

Note: Firmware XPSR42 is downloaded in factory.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 3 / 6
Document: Installation procedure XPSR42 board S703785T

4.1 Location of the XPSR42 board inside the GenX generator:

XPSI XPSR42 XPSCRET XPSP Multiplier


Board Board Board Board assembly

P 108
J 401

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side
TOP side Right side
Left
side

Front side
BOTTOM side

RET GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 4 / 6
Document: Installation procedure XPSR42 board S703785T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSR42 board together with the
XPSCRET board due to the closed proximity of those two boards to avoid
damage on the electronic components part.
XPSR42 board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extract the XPSR42 board from its location, let plugged the male J401
connector and use it to pull the board out.
To extract the XPSCRET board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 5 / 6
Document: Installation procedure XPSR42 board S703785T

- Install the new XPSR42 board, together with the XPSCRET, in the correct
location, see chapter 4.1 of this document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSR42 and XPSCRET boards.
VERY IMPORTANT: The XPSR42 and XPSCRET boards must be installed
at the final locked mechanical position, by passing through two “distinct
clicks positions” when XPSR42 and XPSCRET boards are pushing to the
final position.

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

4.3 Calibration:

No kV/mA calibration is required when XPSR42 is replaced.

5 Special packaging for returned material:


The XPSR42 board requires special packaging for transportation to avoid damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430419 1
Page 6 / 6
Document: Installation procedure XPSCRET board S703784T

X-ray Power Supply Control XPSCRET Board

Instrument : XRF
N° Assembly : S703784T
File : S430426-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 1 / 7
Document: Installation procedure XPSCRET board S703784T

1 Introduction:
IMPORTANT:
XPSCRET board with P/N: S703784T must be installed ONLY in the GenX
generators “R” 3.6kW with P/N: S703685T et 4.2kW with P/N: S703687T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note:
kV/mA measurement calibration parameters are stored in the XPS board
installed in the electronic rack of the XRF instrument and in the WinXRF
software.

Note: No fan and no water block are installed on XPSCRET board.

2 Board configuration:

Below is the layout of the XPSCRET board:

Bridge
W3. Firmware XPSCR.
Location IC15.
Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED

Connector J104. 1 4
Components side. 1
1 4 1 2
2

Connector M8.
Solder side.

Connector J407.
Solder side. Bridge
W2.
Components view

XPSCRET board switches configuration:


Switch State Description
S1-1 OFF Spare
S1-2 OFF Spare
S1-3 OFF Spare
S1-4 OFF Spare
Notes:
The factory setting is in bold.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 2 / 7
Document: Installation procedure XPSCRET board S703784T

XPSCRET board bridges configuration:


Bridge State Description
W1-1 REMOVED XPSCRET setting
W1-2 REMOVED XPSCRET setting

Bridge State Description


W2-1 INSTALLED XPSCRET setting
W2-2 REMOVED XPSCRET setting
W2-3 REMOVED XPSCRET setting
W2-4 INSTALLED XPSCRET setting

Bridge State Description


INSTALLED
Horizontally
to the C53
W3-1 XPSCRET setting
component
direction

INSTALLED
Horizontally
to the C53
W3-2 XPSCRET setting
component
direction

Notes:
1) The factory setting is in bold.

XPSCRET board has 4x electrical external connectors:


Connector identification: Connected to:
M8 -> Not used
J104 -> Not used
J407 -> X-Ray ON lamps with version 24VDC + PTB-TUV
J904 -> GenX backplane board XPSB

XPSCRET board has 2 LEDS (1x green and 1x red):


LED State Description
Blinking XPSCRET is in correct working condition
Green OFF XPSCRET isn’t in correct working condition
ON XPSCRET isn’t in correct working condition
OFF XPSCRET is in correct working condition
Red
ON XPSCRET isn’t in correct working condition

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 3 / 7
Document: Installation procedure XPSCRET board S703784T

3 Firmware compatibility

2x Firmwares XPSCRET (Location IC15) and XPSCS (Location IC13) are installed
on the XPSCRET board.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 4 / 7
Document: Installation procedure XPSCRET board S703784T

4.1 Location of the XPSCRET board inside the GenX


generator:

XPSI XPSR XPSCRET XPSP Multiplier


Board Board Board Board assembly

Green LED
Red LED

M8

J 104
J 407

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 5 / 7
Document: Installation procedure XPSCRET board S703784T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSCRET board together with the
XPSR board due to the closed proximity to avoid damage on the electronic
components part.
XPSCRET board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extract the XPSCRET board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.
To extract the XPSR board from its location, let plugged the male J401
connector and use it to pull the board out.

- Install the new XPSCRET board, together with the XPSR, in the correct location,
see chapter 4.1 of this document.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 6 / 7
Document: Installation procedure XPSCRET board S703784T

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSCRET and XPSR boards.
VERY IMPORTANT: The XPSCRET and XPSR boards must be installed at
the final locked mechanical position, by passing through two “distinct
clicks positions” when XPSCRET and XPSR boards are pushing to the
final position.

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

4.3 Calibration:
No kV/mA calibration is required when XPSCRET is replaced.

5 Special packaging for returned material:


The XPSCRET board requires special packaging for transportation to avoid
damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430426 0
Page 7 / 7
Document: Installation procedure XPSC1260 board S703726T

X-ray Power Supply Control XPSC1260 Board

Instrument : XRF
N° Assembly : S703726T
File : S430424-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 1 / 8
Document: Installation procedure XPSC1260 board S703726T

1 Introduction:
IMPORTANT:
XPSC1260 board with P/N: S703726T must be installed ONLY in the GenX
generator 1.2kW (60kV/60mA) with P/N: S703683T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC1260 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC1260 board stored in the WinXRF software.

Note: No fan and no water block are installed on XPSC1260 board.

2 Board configuration:

Below is the layout of the XPSC1260 board:

Bridge
W3. Firmware XPSCM1260.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED

Connector J104. 1 4
Components side. 1
1 4 1 2
2

Connector M8.
Solder side.

Connector J407.
Solder side. Bridge
W2.
Components view

XPSC1260 board switches configuration:


Switch State Description
ON: Disable the secondary water cooling regulation
S1-1 OFF
OFF: Enable the secondary water cooling regulation
S1-2 OFF Spare
S1-3 OFF Spare
S1-4 OFF Spare
Notes:
1) The factory setting is in bold.
2) The secondary water cooling regulation can be setted using the firmware status N° 2052.
This firmware status 2052 is available from firmwares XPSCM1260 and XPSCS
version >= V3.2-1.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 2 / 8
Document: Installation procedure XPSC1260 board S703726T

XPSC1260 board bridges configuration:


Bridge State Description
W1-1 REMOVED XPSC1260 setting
W1-2 REMOVED XPSC1260 setting

Bridge State Description


W2-1 INSTALLED XPSC1260 setting
W2-2 REMOVED XPSC1260 setting
W2-3 INSTALLED XPSC1260 setting
W2-4 REMOVED XPSC1260 setting

Bridge State Description


INSTALLED
Horizontally
to the C53
W3-1 XPSC1260 setting
component
direction

INSTALLED
Horizontally
to the C53
W3-2 XPSC1260 setting
component
direction

Notes:
1) The factory setting is in bold.

XPSC1260 board has 4x electrical external connectors:


Connector identification: Connected to:
M8 -> XRF instrument interlocks and water control
J104 -> XRF instrument electronic rack (J100 -> J109)
J407 -> X-Ray ON lamps with version 24VDC + PTB-TUV
J904 -> GenX backplane board XPSB

XPSC1260 board has 2 LEDS (1x green and 1x red):


LED State Description
Blinking XPSC1260 is in correct working condition
Green OFF XPSC1260 isn’t in correct working condition
ON XPSC1260 isn’t in correct working condition
OFF XPSC1260 is in correct working condition
Red
ON XPSC1260 isn’t in correct working condition

XPSC1260 board with the factory kV/mA measurement calibration parameters:


Parameter: Value:
kV reading offset
kV reading slope
mA reading offset
mA reading slope
Note:
Those kV/mA parameters values are fulled in factory.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 3 / 8
Document: Installation procedure XPSC1260 board S703726T

3 Firmware compatibility

2x Firmwares XPSCM1260 (Location IC15) and XPSCS (Location IC13) are


installed on the XPSC1260 board.
XPSCM1260 and XPSCS firmwares must always have the same version level.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 4 / 8
Document: Installation procedure XPSC1260 board S703726T

4.1 Location of the XPSC1260 board inside the GenX


generator:

XPSI Board XPSP Board


Air cooling XPSR XPSC1260 Multiplier
Air cooling
version Board Board assembly
version

Green LED
Red LED

M8

J 104
J 407

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 5 / 8
Document: Installation procedure XPSC1260 board S703726T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSC1260 board.


XPSC1260 board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extact the XPSC1260 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.

- Install the new XPSC1260 board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSC1260 board.
VERY IMPORTANT: The XPSC1260 board must be installed at the final
locked mechanical position, by passing through two “distinct clicks
positions” when XPSC1260 board is pushing to the final position.

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 6 / 8
Document: Installation procedure XPSC1260 board S703726T

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

- Caution during the M8 connector plug in:


M8 female connector must be introduced in parallel with the M8 male connector
respecting the key lock.

4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC1260 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.

Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.

IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:

XR 10,10,1,2 This command isn’t needed if X-ray power is already on.

“Wait for tolerance” wording must be not present on the


instrument display to continue to type the following
command

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

The GenX calibration takes few minutes.

At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 7 / 8
Document: Installation procedure XPSC2560 board S704026T

X-ray Power Supply Control XPSC2560 Board

Instrument : XRF
N° Assembly : S704026T
File : S430661-0.doc
Mod. Description Date Name
0. Document creation 28.03.06 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 1 / 8
Document: Installation procedure XPSC2560 board S704026T

1 Introduction:
IMPORTANT:
XPSC2560 board with P/N: S704026T must be installed ONLY in the GenX
generator 2.5kW (60kV/100mA) with P/N: S704025T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC2560 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC2560 board stored in the WinXRF software.

Note: No fan and no water block are installed on XPSC2560 board.

2 Board configuration:

Below is the layout of the XPSC2560 board:

Bridge
W3. Firmware XPSCM2560.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13 Connector J904
Switch
S1. W1. Components side
Red LED

Connector J104. 1 4
Components side. 1
1 4 1 2
2

Connector M8.
Solder side.

Connector J407.
Solder side. Bridge
W2.
Components view

XPSC2560 board switches configuration:


Switch State Description
S1-1 OFF XPSC2560 setting
S1-2 OFF Spare
S1-3 OFF Spare
S1-4 OFF Spare
Notes:
1) The factory setting is in bold.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 2 / 8
Document: Installation procedure XPSC2560 board S704026T

XPSC2560 board bridges configuration:


Bridge State Description
W1-1 REMOVED XPSC2560 setting
W1-2 REMOVED XPSC2560 setting

Bridge State Description


W2-1 REMOVED XPSC2560 setting
W2-2 INSTALLED XPSC2560 setting
W2-3 REMOVED XPSC2560 setting
W2-4 INSTALLED XPSC2560 setting

Bridge State Description


INSTALLED
Horizontally
to the C53
W3-1 XPSC2560 setting
component
direction

INSTALLED
Horizontally
to the C53
W3-2 XPSC2560 setting
component
direction

Notes:
1) The factory setting is in bold.

XPSC2560 board has 4x electrical external connectors:


Connector identification: Connected to:
M8 -> XRF instrument interlocks and water control
J104 -> XRF instrument electronic rack (J100 -> J109)
J407 -> X-Ray ON lamps with version 24VDC + PTB-TUV
J904 -> GenX backplane board XPSB

XPSC2560 board has 2 LEDS (1x green and 1x red):


LED State Description
Blinking XPSC2560 is in correct working condition
Green OFF XPSC2560 isn’t in correct working condition
ON XPSC2560 isn’t in correct working condition
OFF XPSC2560 is in correct working condition
Red
ON XPSC2560 isn’t in correct working condition

XPSC2560 board with the factory kV/mA measurement calibration parameters:


Parameter: Value:
kV reading offset
kV reading slope
mA reading offset
mA reading slope
Note:
Those kV/mA parameters values are fulled in factory.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 3 / 8
Document: Installation procedure XPSC2560 board S704026T

3 Firmware compatibility

2x Firmwares: XPSCM2560 (Location IC15) and XPSCS (Location IC13) are


installed on the XPSC2560 board.
XPSCM2560 and XPSCS firmwares must always have the same version level.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 4 / 8
Document: Installation procedure XPSC2560 board S704026T

4.1 Location of the XPSC2560 board inside the GenX


generator:

XPSI XPSC2560 XPSP Multiplier


Board Board Board assembly

Green LED
Red LED

M8

J 104
J 407

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 5 / 8
Document: Installation procedure XPSC2560 board S704026T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSC2560 board.


XPSC2560 board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extact the XPSC2560 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.

- Install the new XPSC2560 board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSC2560 board.
VERY IMPORTANT: The XPSC2560 board must be installed at the final
locked mechanical position, by passing through two “distinct clicks
positions” when XPSC2560 board is pushing to the final position.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 6 / 8
Document: Installation procedure XPSC2560 board S704026T

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

- Caution during the M8 connector plug in:


M8 female connector must be introduced in parallel with the M8 male connector
respecting the key lock.

4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC2560 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.

Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.

IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:

XR 10,10,1,2 This command isn’t needed if X-ray power is already on.

“Wait for tolerance” wording must be not present on the


instrument display to continue to type the following
command

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 7 / 8
Document: Installation procedure XPSC2560 board S704026T

The GenX calibration takes few minutes.

At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.

5 Special packaging for returned material:


The XPSC2560 board requires special packaging for transportation to avoid
damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430661 0
Page 8 / 8
Document: Installation procedure XPSC1260 board S703726T

5 Special packaging for returned material:


The XPSC1260 board requires special packaging for transportation to avoid
damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430424 0
Page 8 / 8
Document: Installation procedure XPSC3660 board S703722T

X-ray Power Supply Control XPSC3660 Board

Instrument : XRF
N° Assembly : S703722T
File : S430416-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 1 / 8
Document: Installation procedure XPSC3660 board S703722T

1 Introduction:
IMPORTANT:
XPSC3660 board with P/N: S703722T must be installed ONLY in the GenX
generator 3.6kW (60kV/120mA) with P/N: S703690T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC3660 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC3660 board stored in the WinXRF software.

Note: No fan and no water block are installed on XPSC3660 board.

2 Board configuration:

Below is the layout of the XPSC3660 board:

Bridge
W3. Firmware XPSCM3660.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED

Connector J104. 1 4
Components side. 1
1 4 1 2
2

Connector M8.
Solder side.

Connector J407.
Solder side. Bridge
W2.
Components view

XPSC3660 board switches configuration:


Switch State Description
ON: Disable the secondary water cooling regulation
S1-1 OFF
OFF: Enable the secondary water cooling regulation
S1-2 OFF Spare
S1-3 OFF Spare
S1-4 OFF Spare
Notes:
1) The factory setting is in bold.
2) The secondary water cooling regulation can be setted using the firmware status N° 2052.
This firmware status 2052 is available from firmwares XPSCM3660 and XPSCS
version >= V3.2-1.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 2 / 8
Document: Installation procedure XPSC3660 board S703722T

XPSC3660 board bridges configuration:


Bridge State Description
W1-1 REMOVED XPSC3660 setting
W1-2 REMOVED XPSC3660 setting

Bridge State Description


W2-1 INSTALLED XPSC3660 setting
W2-2 REMOVED XPSC3660 setting
W2-3 REMOVED XPSC3660 setting
W2-4 REMOVED XPSC3660 setting

Bridge State Description


INSTALLED
Horizontally
to the C53
W3-1 XPSC3660 setting
component
direction

INSTALLED
Horizontally
to the C53
W3-2 XPSC3660 setting
component
direction

Notes:
1) The factory setting is in bold.

XPSC3660 board has 4x electrical external connectors:


Connector identification: Connected to:
M8 -> XRF instrument interlocks and water control
J104 -> XRF instrument electronic rack (J100 -> J109)
J407 -> X-Ray ON lamps with version 24VDC + PTB-TUV
J904 -> GenX backplane board XPSB

XPSC3660 board has 2 LEDS (1x green and 1x red):


LED State Description
Blinking XPSC3660 is in correct working condition
Green OFF XPSC3660 isn’t in correct working condition
ON XPSC3660 isn’t in correct working condition
OFF XPSC3660 is in correct working condition
Red
ON XPSC3660 isn’t in correct working condition

XPSC3660 board with the factory kV/mA measurement calibration parameters:


Parameter: Value:
kV reading offset
kV reading slope
mA reading offset
mA reading slope
Note:
Those kV/mA parameters values are fulled in factory.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 3 / 8
Document: Installation procedure XPSC3660 board S703722T

3 Firmware compatibility

2x Firmwares XPSCM3660 (Location IC15) and XPSCS (Location IC13) are


installed on the XPSC3660 board.
XPSCM3660 and XPSCS firmwares must always have the same version level.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 4 / 8
Document: Installation procedure XPSC3660 board S703722T

4.1 Location of the XPSC3660 board inside the GenX


generator:

XPSI XPSR XPSC3660 XPSP Multiplier


Board Board Board Board assembly

Green LED
Red LED

M8

J 104
J 407

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 5 / 8
Document: Installation procedure XPSC3660 board S703722T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSC3660 board.


XPSC3660 board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extact the XPSC3660 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.

- Install the new XPSC3660 board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSC3660 board.
VERY IMPORTANT: The XPSC3660 board must be installed at the final
locked mechanical position, by passing through two “distinct clicks
positions” when XPSC3660 board is pushing to the final position.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 6 / 8
Document: Installation procedure XPSC3660 board S703722T

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

- Caution during the M8 connector plug in:


M8 female connector must be introduced in parallel with the M8 male connector
respecting the key lock.

4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC3660 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.

Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.

IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:

XR 10,10,1,2 This command isn’t needed if X-ray power is already on.

“Wait for tolerance” wording must be not present on the


instrument display to continue to type the following
command

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 7 / 8
Document: Installation procedure XPSC3660 board S703722T

The GenX calibration takes few minutes.

At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.

5 Special packaging for returned material:


The XPSC3660 board requires special packaging for transportation to avoid
damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430416 0
Page 8 / 8
Document: Installation procedure XPSC3670 board S703723T

X-ray Power Supply Control XPSC3670 Board

Instrument : XRF
N° Assembly : S703723T
File : S430421-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 1 / 8
Document: Installation procedure XPSC3670 board S703723T

1 Introduction:
IMPORTANT:
XPSC3670 board with P/N: S703723T must be installed ONLY in the GenX
generator 3.6kW (70kV/120mA) with P/N: S703680T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC3670 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC3670 board stored in the WinXRF software.

Note: No fan and no water block are installed on XPSC3670 board.

2 Board configuration:

Below is the layout of the XPSC3670 board:

Bridge
W3. Firmware XPSCM3670.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED

Connector J104. 1 4
Components side. 1
1 4 1 2
2

Connector M8.
Solder side.

Connector J407.
Solder side. Bridge
W2.
Components view

XPSC3670 board switches configuration:


Switch State Description
ON: Disable the secondary water cooling regulation
S1-1 OFF
OFF: Enable the secondary water cooling regulation
S1-2 OFF Spare
S1-3 OFF Spare
S1-4 OFF Spare
Notes:
1) The factory setting is in bold.
2) The secondary water cooling regulation can be setted using the firmware status N° 2052.
This firmware status 2052 is available from firmwares XPSCM3670 and XPSCS
version >= V3.2-1.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 2 / 8
Document: Installation procedure XPSC3670 board S703723T

XPSC3670 board bridges configuration:


Bridge State Description
W1-1 REMOVED XPSC3670 setting
W1-2 REMOVED XPSC3670 setting

Bridge State Description


W2-1 REMOVED XPSC3670 setting
W2-2 INSTALLED XPSC3670 setting
W2-3 REMOVED XPSC3670 setting
W2-4 REMOVED XPSC3670 setting

Bridge State Description


INSTALLED
Horizontally
to the C53
W3-1 XPSC3670 setting
component
direction

INSTALLED
Horizontally
to the C53
W3-2 XPSC3670 setting
component
direction

Notes:
1) The factory setting is in bold.

XPSC3670 board has 4x electrical external connectors:


Connector identification: Connected to:
M8 -> XRF instrument interlocks and water control
J104 -> XRF instrument electronic rack (J100 -> J109)
J407 -> X-Ray ON lamps with version 24VDC + PTB-TUV
J904 -> GenX backplane board XPSB

XPSC3670 board has 2 LEDS (1x green and 1x red):


LED State Description
Blinking XPSC3670 is in correct working condition
Green OFF XPSC3670 isn’t in correct working condition
ON XPSC3670 isn’t in correct working condition
OFF XPSC3670 is in correct working condition
Red
ON XPSC3670 isn’t in correct working condition

XPSC3670 board with the factory kV/mA measurement calibration parameters:


Parameter: Value:
kV reading offset
kV reading slope
mA reading offset
mA reading slope
Note:
Those kV/mA parameters values are fulled in factory.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 3 / 8
Document: Installation procedure XPSC3670 board S703723T

3 Firmware compatibility

2x Firmwares XPSCM3670 (Location IC15) and XPSCS (Location IC13) are


installed on the XPSC3670 board.
XPSCM3670 and XPSCS firmwares must always have the same version level.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 4 / 8
Document: Installation procedure XPSC3670 board S703723T

4.1 Location of the XPSC3670 board inside the GenX


generator:

XPSI XPSR XPSC3670 XPSP Multiplier


Board Board Board Board assembly

Green LED
Red LED

M8

J 104
J 407

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 5 / 8
Document: Installation procedure XPSC3670 board S703723T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSC3670 board.


XPSC3670 board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extact the XPSC3670 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.

- Install the new XPSC3670 board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSC3670 board.
VERY IMPORTANT: The XPSC3670 board must be installed at the final
locked mechanical position, by passing through two “distinct clicks
positions” when XPSC3670 board is pushing to the final position.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 6 / 8
Document: Installation procedure XPSC3670 board S703723T

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

- Caution during the M8 connector plug in:


M8 female connector must be introduced in parallel with the M8 male connector
respecting the key lock.

4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC3670 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.

Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.

IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:

XR 10,10,1,2 This command isn’t needed if X-ray power is already on.

“Wait for tolerance” wording must be not present on the


instrument display to continue to type the following
command

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 7 / 8
Document: Installation procedure XPSC3670 board S703723T

The GenX calibration takes few minutes.

At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.

5 Special packaging for returned material:


The XPSC3670 board requires special packaging for transportation to avoid
damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430421 0
Page 8 / 8
Document: Installation procedure XPSC4260 board S703724T

X-ray Power Supply Control XPSC4260 Board

Instrument : XRF
N° Assembly : S703724T
File : S430422-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 1 / 8
Document: Installation procedure XPSC4260 board S703724T

1 Introduction:
IMPORTANT:
XPSC4260 board with P/N: S703724T must be installed ONLY in the GenX
generator 4.2kW (60kV/140mA) with P/N: S703681T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC4260 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC4260 board stored in the WinXRF software.

Note: No fan and no water block are installed on XPSC4260 board.

2 Board configuration:

Below is the layout of the XPSC4260 board:

Bridge
W3. Firmware XPSCM4260.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED

Connector J104. 1 4
Components side. 1
1 4 1 2
2

Connector M8.
Solder side.

Connector J407.
Solder side. Bridge
W2.
Components view

XPSC4260 board switches configuration:


Switch State Description
ON: Disable the secondary water cooling regulation
S1-1 OFF
OFF: Enable the secondary water cooling regulation
S1-2 OFF Spare
S1-3 OFF Spare
S1-4 OFF Spare
Notes:
1) The factory setting is in bold.
2) The secondary water cooling regulation can be setted using the firmware status N° 2052.
This firmware status 2052 is available from firmwares XPSCM4260 and XPSCS
version >= V3.2-1.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 2 / 8
Document: Installation procedure XPSC4260 board S703724T

XPSC4260 board bridges configuration:


Bridge State Description
W1-1 REMOVED XPSC4260 setting
W1-2 REMOVED XPSC4260 setting

Bridge State Description


W2-1 INSTALLED XPSC4260 setting
W2-2 INSTALLED XPSC4260 setting
W2-3 REMOVED XPSC4260 setting
W2-4 REMOVED XPSC4260 setting

Bridge State Description


INSTALLED
Horizontally
to the C53
W3-1 XPSC4260 setting
component
direction

INSTALLED
Horizontally
to the C53
W3-2 XPSC4260 setting
component
direction

Notes:
1) The factory setting is in bold.

XPSC4260 board has 4x electrical external connectors:


Connector identification: Connected to:
M8 -> XRF instrument interlocks and water control
J104 -> XRF instrument electronic rack (J100 -> J109)
J407 -> X-Ray ON lamps with version 24VDC + PTB-TUV
J904 -> GenX backplane board XPSB

XPSC4260 board has 2 LEDS (1x green and 1x red):


LED State Description
Blinking XPSC4260 is in correct working condition
Green OFF XPSC4260 isn’t in correct working condition
ON XPSC4260 isn’t in correct working condition
OFF XPSC4260 is in correct working condition
Red
ON XPSC4260 isn’t in correct working condition

XPSC4260 board with the factory kV/mA measurement calibration parameters:


Parameter: Value:
kV reading offset
kV reading slope
mA reading offset
mA reading slope
Note:
Those kV/mA parameters values are fulled in factory.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 3 / 8
Document: Installation procedure XPSC4260 board S703724T

3 Firmware compatibility

2x Firmwares XPSCM4260 (Location IC15) and XPSCS (Location IC13) are


installed on the XPSC4260 board.
XPSCM4260 and XPSCS firmwares must always have the same version level.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 4 / 8
Document: Installation procedure XPSC4260 board S703724T

4.1 Location of the XPSC4260 board inside the GenX


generator:

XPSI XPSR XPSC4260 XPSP Multiplier


Board Board Board Board assembly

Green LED
Red LED

M8

J 104
J 407

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 5 / 8
Document: Installation procedure XPSC4260 board S703724T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSC4260 board.


XPSC4260 board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extact the XPSC4260 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.

- Install the new XPSC4260 board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSC4260 board.
VERY IMPORTANT: The XPSC4260 board must be installed at the final
locked mechanical position, by passing through two “distinct clicks
positions” when XPSC4260 board is pushing to the final position.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 6 / 8
Document: Installation procedure XPSC4260 board S703724T

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

- Caution during the M8 connector plug in:


M8 female connector must be introduced in parallel with the M8 male connector
respecting the key lock.

4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC4260 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.

Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.

IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:

XR 10,10,1,2 This command isn’t needed if X-ray power is already on.

“Wait for tolerance” wording must be not present on the


instrument display to continue to type the following
command

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 7 / 8
Document: Installation procedure XPSC4260 board S703724T

The GenX calibration takes few minutes.

At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.

5 Special packaging for returned material:


The XPSC4260 board requires special packaging for transportation to avoid
damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430422 0
Page 8 / 8
Document: Installation procedure XPSC4270 board S703725T

X-ray Power Supply Control XPSC4270 Board

Instrument : XRF
N° Assembly : S703725T
File : S430423-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 1 / 8
Document: Installation procedure XPSC4270 board S703725T

1 Introduction:
IMPORTANT:
XPSC4270 board with P/N: S703725T must be installed ONLY in the GenX
generator 4.2kW (70kV/140mA) with P/N: S703682T.
In case of no respect of this above caution, Thermo Electron Ecublens will
be not responsible for any damage of the Genx generator and for any body’s
injury.

Note:
New kV/mA measurement calibration parameters are stored in the new
XPSC4270 board.
Those new kV/mA parameters will replace the ones from the original “old”
XPSC4270 board stored in the WinXRF software.

Note: No fan and no water block are installed on XPSC4270 board.

2 Board configuration:

Below is the layout of the XPSC4270 board:

Bridge
W3. Firmware XPSCM4270.
Location IC15. Firmware XPSCS.
Green LED Bridge Location IC13
Switch Connector J904
S1. W1. Components side
Red LED

Connector J104. 1 4
Components side. 1
1 4 1 2
2

Connector M8.
Solder side.

Connector J407.
Solder side. Bridge
W2.
Components view

XPSC4270 board switches configuration:


Switch State Description
ON: Disable the secondary water cooling regulation
S1-1 OFF
OFF: Enable the secondary water cooling regulation
S1-2 OFF Spare
S1-3 OFF Spare
S1-4 OFF Spare
Notes:
1) The factory setting is in bold.
2) The secondary water cooling regulation can be setted using the firmware status N° 2052.
This firmware status 2052 is available from firmwares XPSCM4270 and XPSCS
version >= V3.2-1.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 2 / 8
Document: Installation procedure XPSC4270 board S703725T

XPSC4270 board bridges configuration:


Bridge State Description
W1-1 REMOVED XPSC4270 setting
W1-2 REMOVED XPSC4270 setting

Bridge State Description


W2-1 REMOVED XPSC4270 setting
W2-2 REMOVED XPSC4270 setting
W2-3 INSTALLED XPSC4270 setting
W2-4 REMOVED XPSC4270 setting

Bridge State Description


INSTALLED
Horizontally
to the C53
W3-1 XPSC4270 setting
component
direction

INSTALLED
Horizontally
to the C53
W3-2 XPSC4270 setting
component
direction

Notes:
1) The factory setting is in bold.

XPSC4270 board has 4x electrical external connectors:


Connector identification: Connected to:
M8 -> XRF instrument interlocks and water control
J104 -> XRF instrument electronic rack (J100 -> J109)
J407 -> X-Ray ON lamps with version 24VDC + PTB-TUV
J904 -> GenX backplane board XPSB

XPSC4270 board has 2 LEDS (1x green and 1x red):


LED State Description
Blinking XPSC4270 is in correct working condition
Green OFF XPSC4270 isn’t in correct working condition
ON XPSC4270 isn’t in correct working condition
OFF XPSC4270 is in correct working condition
Red
ON XPSC4270 isn’t in correct working condition

XPSC4270 board with the factory kV/mA measurement calibration parameters:


Parameter: Value:
kV reading offset
kV reading slope
mA reading offset
mA reading slope
Note:
Those kV/mA parameters values are fulled in factory.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 3 / 8
Document: Installation procedure XPSC4270 board S703725T

3 Firmware compatibility

2x Firmwares XPSCM4270 (Location IC15) and XPSCS (Location IC13) are


installed on the XPSC4270 board.
XPSCM4270 and XPSCS firmwares must always have the same version level.

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 4 / 8
Document: Installation procedure XPSC4270 board S703725T

4.1 Location of the XPSC4270 board inside the GenX


generator:

XPSI XPSR XPSC4270 XPSP Multiplier


Board Board Board Board assembly

Green LED
Red LED

M8

J 104
J 407

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 5 / 8
Document: Installation procedure XPSC4270 board S703725T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- At this step, you are able to remove the XPSC4270 board.


XPSC4270 board is connected to a unique backplane board XPSB located near
the GenX front panel.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC board.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

- To extact the XPSC4270 board from its location, use your fingers to pull
together the front connectors J104, J407 and M8.

- Install the new XPSC4270 board in the correct location, see chapter 4.1 of this
document.

Take care, as mentioned above, to avoid to damage electronic components


located on the way during installation of the XPSC4270 board.
VERY IMPORTANT: The XPSC4270 board must be installed at the final
locked mechanical position, by passing through two “distinct clicks
positions” when XPSC4270 board is pushing to the final position.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 6 / 8
Document: Installation procedure XPSC4270 board S703725T

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

- Caution during the M8 connector plug in:


M8 female connector must be introduced in parallel with the M8 male connector
respecting the key lock.

4.3 Calibration:
- Store the new kV/mA measurement calibration parameters of the new
XPSC4270 board to the WinXRF software.
To store those new kV/mA parameters, read the firmware status 2048.

Note:
Only if alarms related to the precision of the kV/mA measurement appear, then
perform GenX calibration using the terminal mode command #XC 1.

IMPORTANT:
Before to perform a GenX calibration kV/mA, the X-ray tube must be already
“warmed up”.
Use the following terminal mode command to calibrate the GenX generator:

XR 10,10,1,2 This command isn’t needed if X-ray power is already on.

“Wait for tolerance” wording must be not present on the


instrument display to continue to type the following
command

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 7 / 8
Document: Installation procedure XPSC4270 board S703725T

The GenX calibration takes few minutes.

At the end of the GenX calibration, store the new kV/mA parameters by reading
the firmware status 2048.

5 Special packaging for returned material:


The XPSC4270 board requires special packaging for transportation to avoid
damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430423 0
Page 8 / 8
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Install XPSC Board in Gen'X 1.5 kW


1 Introduction
CAUTION! IMPORTANT INFORMATION:
The XPSC board with the part number: 1226732 must be installed ONLY in the GenX
generator 1.5 kW (60 kV/100 mA) with the part number: S604811K/1226736. If not
followed, it will result in damaging of the X-ray tube!

Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument, XPSC board and GenX.

New kV/mA measurement calibration parameters are stored in the new XPSC board. Those new
kV/mA parameters will replace the ones from the original “old” XPSC board stored in the OXSAS
software.
Note No fan and no water block are installed on XPSC board.

1.1 Verify that You Have Correct XPSC Board

CAUTION!It is important to verify that you have the correct XPSC board. Incorrect XPSC board will
damage the X-ray tube permanently.

The picture below indicates the XPSC board including places to check:
● Check the XPSC Part Number - 1226732,
● Check Jumpers,
● The IC15 must have the blue label as well as the instrument's reset button.

.
Page 1 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

2 Safety
Note Do not perform any operation without proper safety training specific to the hazard(s).

2.1 Hazard Statements and Symbols


The following messages are used throughout the manuals to highlight information.
Note A Note message indicates the information in it requires special attention.

CAUTION!

A Caution message indicates a potentially hazardous situation that, if not avoided, may
result in moderate or minor injury. It may also be used to alert against unsafe practices.

WARNING!

A Warning message indicates a potentially hazardous situation that, if not avoided, may
result in death or severe injury.

DANGER!
A Danger message indicates an imminently hazardous situation that, if not avoided, will
result in death or severe injury.

.
Page 2 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

WARNING!
Radiation danger – Indicates that X-Ray hazard may be present. Take necessary
precautions.

SERVICE!
A Service message indicates that the described procedure should be performed by
qualified personnel only.

3 Prerequisites
SERVICE! This document must be used ONLY by the person who followed the special training of
GenX generator and successfully obtained the certificate to service GenX generators.

CAUTION!Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the GenX generator.

4 Preparation

4.1 Board Configuration

.
Page 3 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

4.1.1 XPSC1560 Board Switches Configuration

Switch State Description

S1-1 OFF ON: Disable the secondary water cooling regulation


OFF: Enable the secondary water cooling regulation

S1-2 OFF Spare

S1-3 OFF Spare

S1-4 OFF Spare

Note 1) The factory setting is in bold.

Note 2) The secondary water cooling regulation can be set using the firmware status N° 2052.

4.1.2 XPSC1560 Board Bridges Configuration

Bridge State Description

W1-1 REMOVED XPSC setting

W1-2 REMOVED XPSC setting

Bridge State Description

W2-1 INSTALLED XPSC setting

W2-2 REMOVED XPSC setting

W2-3 INSTALLED XPSC setting

W2-4 INSTALLED XPSC setting

Bridge State Description

W3-1 INSTALLED -Horizontally to the C53 component direction XPSC setting

W3-2 INSTALLED - Horizontally to the C53 component direction XPSC setting

.
Page 4 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Note 1) The factory setting is in bold.

4.1.3 XPSC Board Has 4x Electrical External Connectors

Connector identification: Connected to:

M8 XRF instrument interlocks and water control

J104 XRF instrument electronic rack (J100 -> J109)

J407 X-Ray ON lamps with version 24VDC + PTB-TUV

J904 GenX backplane board XPSB

4.1.4 XPSC Board Has 2 LEDS (1x Green and 1x Red)

LED State Description

Green Blinking XPSC is in correct working condition


OFF XPSC isn’t in correct working condition
ON XPSC isn’t in correct working condition

Red OFF XPSC is in correct working condition


ON XPSC isn’t in correct working condition

4.2 Firmware Compatibility


2x Firmwares XPSCM (Location IC15) and XPSCS (Location IC13) are installed on the XPSC
board.
XPSCM and XPSCS firmwares must always have the same version level.

5 Procedure

5.1 Board Installation


Before working on the GenX generator, the following caution must be read carefully in order to avoid
any trouble occurring to the generator and any body’s injury:

.
Page 5 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

WARNING! Before working with the new GenX generator, please read carefully the following
information and cautions.

DANGER! LETHAL VOLTAGES are present in the GenX generator whenever the generator is
connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.

CAUTION! Accessibility inside the GenX generator:


The "opening" of the GenX generator, which is necessary for the maintenance and repair,
must be STRICTLY done by a certified service engineer who followed a special GenX
training course.Thermo Fisher Scientific will be not responsible for any injury occurred
on body person in case of not respecting of the special training accreditation given for
GenX generator.
Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument.

5.2 Location of the XPSC Board Inside the GenX Generator


The picture below indicates the location of the XPSC board in the GenX generator.

5.3 Installation Procedure


Note This procedure starts from the step when the GenX generator is already removed from the XRF
instrument.

.
Page 6 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Note For more information concerning the removal of the generator from the XRF instrument, consult the
GenX generator service manual P/N: AA83678, delivered during the certification training course given
in Ecublens by TTI department.

1. Put the GenX generator on a clean table.

CAUTION! During the removal and installation of the GenX boards and Multiplier Assembly, always
put down the GenX generator with bottom side in contact with the surface table, see the
picture below. DO NOT turn the GenX upside down!

2. WAIT 15 MINUTES to be sure that all stored energy is discharged after power disconnection.

DANGER! DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.

3. Unplug and plug action of all electronic boards must be delayed by at least 15 MINUTES.

CAUTION!Not respecting of this caution note can cause serious damage on the GenX backplane
board XPSB and body’s injury due to energy stored in the capacitor.

4. Remove the 4x screws fixing the rear panel of the GenX generator.
5. Remove the rear panel of the GenX generator.

.
Page 7 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.

6. At this step, you are able to remove the XPSC board: XPSC board is connected to a unique
backplane board XPSB located near the GenX front panel.

CAUTION!IMPORTANT GENERAL NOTE: If you want to remove all the GenX boards located inside
the GenX, you must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed between them):
a) First board to be removed is -> XPSI board
b) Second two boards to be removed are -> XPSR (if installed) together with XPSC board.
c) Third board to be removed is -> XPSP board
d) Fourth “board” to be removed is -> Multiplier assembly (don’t forget the 3x locking
screws located at the bottom panel of the GenX generator).
7. To extract the XPSC board from its location, use your fingers to pull together the front connectors
J104, J407 and M8.
8. Install the new XPSC board in the correct location, see Location of the XPSC Board Inside the
GenX Generator on page 6. Avoid to damage electronic components located on the way during
installation of the XPSC board!

CAUTION! Protect ESD sensitive devices from electrostatic discharge.


Always work with:
• An ESD wristband.
• An ESD table mat.
• ESD-safe gloves.

CAUTION!IMPORTATN NOTE: The XPSC board must be installed at the final locked mechanical
position, by passing through two “distinct clicks positions” when XPSC board is pushed
to the final position.

9. Check the correct installation of all the GenX boards and multiplier assembly (correct location
and final mechanical position) to the backplane XPSB board.
10. Install the rear panel of the GenX by screwing totally 4x screws.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.

11. Connect the two water fast fitting connectors:


1x connection between XPSI and XPSP boards,
1x connection at the multiplier assembly.
12. At this step, the GenX generator is ready to be installed into the XRF instrument.
13. Consult the GenX generator service manual P/N: AA83678, for the installation of the GenX
generator into the XRF instrument.

.
Page 8 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

CAUTION! Caution during the M8 connector plug in: M8 female connector must be introduced in
parallel with the M8 male connector respecting the key lock.

5.4 Calibration
1. Store the new kV/mA measurement calibration parameters of the new XPSC board to the
OXSAS software. To store those new kV/mA parameters, read the firmware status 2048.
Note Only if alarms related to the precision of the kV/mA measurement appear, then perform GenX
calibration using the terminal mode command #XC 1.

CAUTION! Before to perform a GenX calibration kV/mA, the X-ray tube must be already
WARMED UP.

Use the following terminal mode command to calibrate the GenX generator:

Command Description

XR 10,10,1,2 This command is not needed if the X-ray power is already ON.
"Wait for tolerance" wording must be not present on the instrument display to continue to type
the following command.

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

2. The GenX calibration takes a few minutes.


3. At the end of the GenX calibration, store the new kV/mA parameters by reading the firmware
status 2048.

5.5 Special Packaging for Returned Material


The XPSC board requires special shipment packaging to avoid damage.

CAUTION! For shipment of XPSC board, it must always be packed in a special packaging.

[End of Work Instruction]

.
Page 9 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 1.5 kW Doc Number: Provide Field
307444 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Revision Table

Revision Date ECO number Description of Changes

A 02-APR-2020 SDR21079 Initial release

FEI Company, part of Thermo Fisher Scientific, work instructions are proprietary information and confidential. This procedure is property of Thermo
Fisher and for Thermo Fisher internal use only and must not be duplicated or disseminated for any third party without the express consent of hermo
Fisher. Printed or electronic copies of this procedure are uncontrolled and intended only for immediate use. The electronic files are the controlled
versions and are to be used as the master copies. Incomplete printed copies are not to be used and must be discarded. The Government's rights to use,
modify, reproduce, release, perform, display, or disclose these technical data are restricted to those rights specified in DFARS 252.227-7015(b)(2), FAR
52.227-14(g)(2)(Alternate II) and FAR 12.211. Any reproduction of technical data or portions thereof marked with this legend must also reproduce the
markings. Any person, other than the Government, who has been provided access to such data, must promptly notify Thermo Fisher.

.
Page 10 of 10 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Install XPSC Board in Gen'X 2.5 kW


1 Introduction
CAUTION! IMPORTANT INFORMATION:
The XPSC board with the part number: 1226741 must be installed ONLY in the GenX
generator 2.5 kW (60 kV/140 mA) with the part number: S604812K/1226748. If not
followed, it will result in damaging of the X-ray tube!

Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument, XPSC board and GenX.

New kV/mA measurement calibration parameters are stored in the new XPSC board. Those new
kV/mA parameters will replace the ones from the original “old” XPSC board stored in the OXSAS
software.
Note No fan and no water block are installed on XPSC board.

1.1 Verify that You Have Correct XPSC Board

CAUTION!It is important to verify that you have the correct XPSC board. Incorrect XPSC board will
damage the X-ray tube permanently.

The picture below indicates the XPSC board including places to check:
● Check the XPSC part Number - 1226741,
● Check Jumpers,
● The IC15 must have the blue label as well as the instrument's reset button.

.
Page 1 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

2 Safety
Note Do not perform any operation without proper safety training specific to the hazard(s).

2.1 Hazard Statements and Symbols


The following messages are used throughout the manuals to highlight information.
Note A Note message indicates the information in it requires special attention.

CAUTION!

A Caution message indicates a potentially hazardous situation that, if not avoided, may
result in moderate or minor injury. It may also be used to alert against unsafe practices.

WARNING!

A Warning message indicates a potentially hazardous situation that, if not avoided, may
result in death or severe injury.

.
Page 2 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

DANGER!
A Danger message indicates an imminently hazardous situation that, if not avoided, will
result in death or severe injury.

WARNING!
Radiation danger – Indicates that X-Ray hazard may be present. Take necessary
precautions.

SERVICE!
A Service message indicates that the described procedure should be performed by
qualified personnel only.

3 Prerequisites
SERVICE! This document must be used ONLY by the person who followed the special training of
GenX generator and successfully obtained the certificate to service GenX generators.

CAUTION!Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the GenX generator.

.
Page 3 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

4 Preparation

4.1 Board Configuration

4.1.1 XPSC2560 Board Switches Configuration

Switch State Description

S1-1 OFF ON: Disable the secondary water cooling regulation


OFF: Enable the secondary water cooling regulation

S1-2 OFF Spare

S1-3 OFF Spare

S1-4 OFF Spare

Note 1) The factory setting is in bold.

Note 2) The secondary water cooling regulation can be set using the firmware status N° 2052.

4.1.2 XPSC2560 Board Bridges Configuration

Bridge State Description

W1-1 REMOVED XPSC setting

W1-2 REMOVED XPSC setting

.
Page 4 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Bridge State Description

W2-1 INSTALLED XPSC setting

W2-2 REMOVED XPSC setting

W2-3 INSTALLED XPSC setting

W2-4 INSTALLED XPSC setting

Bridge State Description

W3-1 INSTALLED XPSC setting


Horizontally to the C53 component direction

W3-2 INSTALLED XPSC setting


Horizontally to the C53 component direction

Note 1) The factory setting is in bold.

4.1.3 XPSC Board Has 4x Electrical External Connectors

Connector identification: Connected to:

M8 XRF instrument interlocks and water control

J104 XRF instrument electronic rack (J100 -> J109)

J407 X-Ray ON lamps with version 24VDC + PTB-TUV

J904 GenX backplane board XPSB

.
Page 5 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

4.1.4 XPSC Board Has 2 LEDS (1x Green and 1x Red)

LED State Description

Green Blinking XPSC is in correct working condition


OFF XPSC isn’t in correct working condition
ON XPSC isn’t in correct working condition

Red OFF XPSC is in correct working condition


ON XPSC isn’t in correct working condition

4.2 Firmware Compatibility


2x Firmwares XPSCM (Location IC15) and XPSCS (Location IC13) are installed on the XPSC
board.
XPSCM and XPSCS firmwares must always have the same version level.

5 Procedure

5.1 Board Installation


Before working on the GenX generator, the following caution must be read carefully in order to avoid
any trouble occurring to the generator and any body’s injury:

WARNING! Before working with the new GenX generator, please read carefully the following
information and cautions.

DANGER! LETHAL VOLTAGES are present in the GenX generator whenever the generator is
connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.

.
Page 6 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

CAUTION! Accessibility inside the GenX generator:


The "opening" of the GenX generator, which is necessary for the maintenance and repair,
must be STRICTLY done by a certified service engineer who followed a special GenX
training course.Thermo Fisher Scientific will be not responsible for any injury occurred
on body person in case of not respecting of the special training accreditation given for
GenX generator.
Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument.

5.2 Location of the XPSC Board Inside the GenX Generator


The picture below indicates the location of the XPSC board in the GenX generator.

5.3 Installation Procedure


Note This procedure starts from the step when the GenX generator is already removed from the XRF
instrument.

Note For more information concerning the removal of the generator from the XRF instrument, consult the
GenX generator service manual P/N: AA83678, delivered during the certification training course given
in Ecublens by TTI department.

1. Put the GenX generator on a clean table.

CAUTION! During the removal and installation of the GenX boards and Multiplier Assembly, always
put down the GenX generator with bottom side in contact with the surface table, see the
picture below. DO NOT turn the GenX upside down!

.
Page 7 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

2. WAIT 15 MINUTES to be sure that all stored energy is discharged after power disconnection.

DANGER! DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.

3. Unplug and plug action of all electronic boards must be delayed by at least 15 MINUTES.

CAUTION!Not respecting of this caution note can cause serious damage on the GenX backplane
board XPSB and body’s injury due to energy stored in the capacitor.

4. Remove the 4x screws fixing the rear panel of the GenX generator.
5. Remove the rear panel of the GenX generator.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.

6. At this step, you are able to remove the XPSC board: XPSC board is connected to a unique
backplane board XPSB located near the GenX front panel.

.
Page 8 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

CAUTION!IMPORTANT GENERAL NOTE: If you want to remove all the GenX boards located inside
the GenX, you must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed between them):
a) First board to be removed is -> XPSI board
b) Second two boards to be removed are -> XPSR (if installed) together with XPSC board.
c) Third board to be removed is -> XPSP board
d) Fourth “board” to be removed is -> Multiplier assembly (don’t forget the 3x locking
screws located at the bottom panel of the GenX generator).
7. To extract the XPSC board from its location, use your fingers to pull together the front connectors
J104, J407 and M8.
8. Install the new XPSC board in the correct location, see Location of the XPSC Board Inside the
GenX Generator on page 7. Avoid to damage electronic components located on the way during
installation of the XPSC board!

CAUTION! Protect ESD sensitive devices from electrostatic discharge.


Always work with:
• An ESD wristband.
• An ESD table mat.
• ESD-safe gloves.

CAUTION!IMPORTATN NOTE: The XPSC board must be installed at the final locked mechanical
position, by passing through two “distinct clicks positions” when XPSC board is pushed
to the final position.

9. Check the correct installation of all the GenX boards and multiplier assembly (correct location
and final mechanical position) to the backplane XPSB board.
10. Install the rear panel of the GenX by screwing totally 4x screws.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.

11. Connect the two water fast fitting connectors:


1x connection between XPSI and XPSP boards,
1x connection at the multiplier assembly.
12. At this step, the GenX generator is ready to be installed into the XRF instrument.
13. Consult the GenX generator service manual P/N: AA83678, for the installation of the GenX
generator into the XRF instrument.

CAUTION! Caution during the M8 connector plug in: M8 female connector must be introduced in
parallel with the M8 male connector respecting the key lock.

.
Page 9 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

5.4 Calibration
1. Store the new kV/mA measurement calibration parameters of the new XPSC board to the
OXSAS software. To store those new kV/mA parameters, read the firmware status 2048.
Note Only if alarms related to the precision of the kV/mA measurement appear, then perform GenX
calibration using the terminal mode command #XC 1.

CAUTION! Before to perform a GenX calibration kV/mA, the X-ray tube must be already
WARMED UP.

Use the following terminal mode command to calibrate the GenX generator:

Command Description

XR 10,10,1,2 This command is not needed if the X-ray power is already ON.
"Wait for tolerance" wording must be not present on the instrument display to continue to type
the following command.

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

2. The GenX calibration takes a few minutes.


3. At the end of the GenX calibration, store the new kV/mA parameters by reading the firmware
status 2048.

5.5 Special Packaging for Returned Material


The XPSC board requires special shipment packaging to avoid damage.

CAUTION! For shipment of XPSC board, it must always be packed in a special packaging.

[End of Work Instruction]

Revision Table

Revision Date ECO number Description of Changes

A 02-APR-2020 SDR21079 Initial rlease

.
Page 10 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 2.5 kW Doc Number: Provide Field
307445 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

FEI Company, part of Thermo Fisher Scientific, work instructions are proprietary information and confidential. This procedure is property of Thermo
Fisher and for Thermo Fisher internal use only and must not be duplicated or disseminated for any third party without the express consent of hermo
Fisher. Printed or electronic copies of this procedure are uncontrolled and intended only for immediate use. The electronic files are the controlled
versions and are to be used as the master copies. Incomplete printed copies are not to be used and must be discarded. The Government's rights to use,
modify, reproduce, release, perform, display, or disclose these technical data are restricted to those rights specified in DFARS 252.227-7015(b)(2), FAR
52.227-14(g)(2)(Alternate II) and FAR 12.211. Any reproduction of technical data or portions thereof marked with this legend must also reproduce the
markings. Any person, other than the Government, who has been provided access to such data, must promptly notify Thermo Fisher.

.
Page 11 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Install XPSC Board in Gen'X 4.2 kW


1 Introduction
CAUTION! IMPORTANT INFORMATION:
The XPSC board with the part number: 1226755 must be installed ONLY in the GenX
generator 4.2 kW (60 kV/120 mA) with the part number: S604807K/1226757. If not
followed, it will result in damaging of the X-ray tube!

Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument, XPSC board and GenX.

New kV/mA measurement calibration parameters are stored in the new XPSC board. Those new
kV/mA parameters will replace the ones from the original “old” XPSC board stored in the OXSAS
software.
Note No fan and no water block are installed on XPSC board.

1.1 Verify that You Have Correct XPSC Board

CAUTION!It is important to verify that you have the correct XPSC board. Incorrect XPSC board will
damage the X-ray tube permanently.

The picture below indicates the XPSC board including places to check:
● Check the XPSC part Number - 1226755,
● Check Jumpers,
● The IC15 must have the blue label as well as the instrument's reset button.

.
Page 1 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

2 Safety
Note Do not perform any operation without proper safety training specific to the hazard(s).

2.1 Hazard Statements and Symbols


The following messages are used throughout the manuals to highlight information.
Note A Note message indicates the information in it requires special attention.

CAUTION!

A Caution message indicates a potentially hazardous situation that, if not avoided, may
result in moderate or minor injury. It may also be used to alert against unsafe practices.

WARNING!

A Warning message indicates a potentially hazardous situation that, if not avoided, may
result in death or severe injury.

.
Page 2 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

DANGER!
A Danger message indicates an imminently hazardous situation that, if not avoided, will
result in death or severe injury.

WARNING!
Radiation danger – Indicates that X-Ray hazard may be present. Take necessary
precautions.

SERVICE!
A Service message indicates that the described procedure should be performed by
qualified personnel only.

3 Prerequisites
SERVICE! This document must be used ONLY by the person who followed the special training of
GenX generator and successfully obtained the certificate to service GenX generators.

CAUTION!Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the GenX generator.

.
Page 3 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

4 Preparation

4.1 Board Configuration

4.1.1 XPSC4260 Board Switches Configuration

Switch State Description

S1-1 OFF ON: Disable the secondary water cooling regulation


OFF: Enable the secondary water cooling regulation

S1-2 OFF Spare

S1-3 OFF Spare

S1-4 OFF Spare

Note 1) The factory setting is in bold.

Note 2) The secondary water cooling regulation can be set using the firmware status N° 2052.

4.1.2 XPSC4260 Board Bridges Configuration

Bridge State Description

W1-1 REMOVED XPSC setting

W1-2 REMOVED XPSC setting

.
Page 4 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

Bridge State Description

W2-1 INSTALLED XPSC setting

W2-2 REMOVED XPSC setting

W2-3 INSTALLED XPSC setting

W2-4 INSTALLED XPSC setting

Bridge State Description

W3-1 INSTALLED XPSC setting


Horizontally to the C53 component direction

W3-2 INSTALLED XPSC setting


Horizontally to the C53 component direction

Note 1) The factory setting is in bold.

4.1.3 XPSC Board Has 4x Electrical External Connectors

Connector identification: Connected to:

M8 XRF instrument interlocks and water control

J104 XRF instrument electronic rack (J100 -> J109)

J407 X-Ray ON lamps with version 24VDC + PTB-TUV

J904 GenX backplane board XPSB

.
Page 5 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

4.1.4 XPSC Board Has 2 LEDS (1x Green and 1x Red)

LED State Description

Green Blinking XPSC is in correct working condition


OFF XPSC isn’t in correct working condition
ON XPSC isn’t in correct working condition

Red OFF XPSC is in correct working condition


ON XPSC isn’t in correct working condition

4.2 Firmware Compatibility


2x Firmwares XPSCM (Location IC15) and XPSCS (Location IC13) are installed on the XPSC
board.
XPSCM and XPSCS firmwares must always have the same version level.

5 Procedure

5.1 Board Installation


Before working on the GenX generator, the following caution must be read carefully in order to avoid
any trouble occurring to the generator and any body’s injury:

WARNING! Before working with the new GenX generator, please read carefully the following
information and cautions.

DANGER! LETHAL VOLTAGES are present in the GenX generator whenever the generator is
connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.

.
Page 6 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

CAUTION! Accessibility inside the GenX generator:


The "opening" of the GenX generator, which is necessary for the maintenance and repair,
must be STRICTLY done by a certified service engineer who followed a special GenX
training course.Thermo Fisher Scientific will be not responsible for any injury occurred
on body person in case of not respecting of the special training accreditation given for
GenX generator.
Thermo Fisher Scientific declines all responsibilities for improper use by unauthorized
personnel or non-authorized operation of the instrument.

5.2 Location of the XPSC Board Inside the GenX Generator


The picture below indicates the location of the XPSC board in the GenX generator.

5.3 Installation Procedure


Note This procedure starts from the step when the GenX generator is already removed from the XRF
instrument.

Note For more information concerning the removal of the generator from the XRF instrument, consult the
GenX generator service manual P/N: AA83678, delivered during the certification training course given
in Ecublens by TTI department.

1. Put the GenX generator on a clean table.

CAUTION! During the removal and installation of the GenX boards and Multiplier Assembly, always
put down the GenX generator with bottom side in contact with the surface table, see the
picture below. DO NOT turn the GenX upside down!

.
Page 7 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

2. WAIT 15 MINUTES to be sure that all stored energy is discharged after power disconnection.

DANGER! DANGER: LETHAL STORED ENERGY is still present in the GenX generator after
disconnection of the power line source. Wait 15 minutes to be sure that all stored energy
is discharged after power disconnection.

3. Unplug and plug action of all electronic boards must be delayed by at least 15 MINUTES.

CAUTION!Not respecting of this caution note can cause serious damage on the GenX backplane
board XPSB and body’s injury due to energy stored in the capacitor.

4. Remove the 4x screws fixing the rear panel of the GenX generator.
5. Remove the rear panel of the GenX generator.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.

6. At this step, you are able to remove the XPSC board: XPSC board is connected to a unique
backplane board XPSB located near the GenX front panel.

.
Page 8 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

CAUTION!IMPORTANT GENERAL NOTE: If you want to remove all the GenX boards located inside
the GenX, you must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed between them):
a) First board to be removed is -> XPSI board
b) Second two boards to be removed are -> XPSR (if installed) together with XPSC board.
c) Third board to be removed is -> XPSP board
d) Fourth “board” to be removed is -> Multiplier assembly (don’t forget the 3x locking
screws located at the bottom panel of the GenX generator).
7. To extract the XPSC board from its location, use your fingers to pull together the front connectors
J104, J407 and M8.
8. Install the new XPSC board in the correct location, see Location of the XPSC Board Inside the
GenX Generator on page 7. Avoid to damage electronic components located on the way during
installation of the XPSC board!

CAUTION! Protect ESD sensitive devices from electrostatic discharge.


Always work with:
• An ESD wristband.
• An ESD table mat.
• ESD-safe gloves.

CAUTION!IMPORTATN NOTE: The XPSC board must be installed at the final locked mechanical
position, by passing through two “distinct clicks positions” when XPSC board is pushed
to the final position.

9. Check the correct installation of all the GenX boards and multiplier assembly (correct location
and final mechanical position) to the backplane XPSB board.
10. Install the rear panel of the GenX by screwing totally 4x screws.
Note Only in case of 1x longer screw and 3x shorter, the longer screw is located on the top of the GenX
generator and on the multiplier assembly side. Consult the GenX generator service manual P/N:
AA83678, to see the screws location.

11. Connect the two water fast fitting connectors:


1x connection between XPSI and XPSP boards,
1x connection at the multiplier assembly.
12. At this step, the GenX generator is ready to be installed into the XRF instrument.
13. Consult the GenX generator service manual P/N: AA83678, for the installation of the GenX
generator into the XRF instrument.

CAUTION! Caution during the M8 connector plug in: M8 female connector must be introduced in
parallel with the M8 male connector respecting the key lock.

.
Page 9 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

5.4 Calibration
1. Store the new kV/mA measurement calibration parameters of the new XPSC board to the
OXSAS software. To store those new kV/mA parameters, read the firmware status 2048.
Note Only if alarms related to the precision of the kV/mA measurement appear, then perform GenX
calibration using the terminal mode command #XC 1.

CAUTION! Before to perform a GenX calibration kV/mA, the X-ray tube must be already
WARMED UP.

Use the following terminal mode command to calibrate the GenX generator:

Command Description

XR 10,10,1,2 This command is not needed if the X-ray power is already ON.
"Wait for tolerance" wording must be not present on the instrument display to continue to type
the following command.

XC 3 Uncalibrated kV/mA parameters with default values

XC 1 Kv and mA will be setted at different conditions

2. The GenX calibration takes a few minutes.


3. At the end of the GenX calibration, store the new kV/mA parameters by reading the firmware
status 2048.

5.5 Special Packaging for Returned Material


The XPSC board requires special shipment packaging to avoid damage.

CAUTION! For shipment of XPSC board, it must always be packed in a special packaging.

[End of Work Instruction]

Revision Table

Revision Date ECO number Description of Changes

A 02-APR-2020 SDR21079 Initial release

.
Page 10 of 11 Confidential and proprietary information
Install XPSC Board in Gen'X 4.2 kW Doc Number: Provide Field
307446 Feedback

TC: WARNING: paper copies are uncontrolled Date: 4/3/2020 Rev: A

FEI Company, part of Thermo Fisher Scientific, work instructions are proprietary information and confidential. This procedure is property of Thermo
Fisher and for Thermo Fisher internal use only and must not be duplicated or disseminated for any third party without the express consent of hermo
Fisher. Printed or electronic copies of this procedure are uncontrolled and intended only for immediate use. The electronic files are the controlled
versions and are to be used as the master copies. Incomplete printed copies are not to be used and must be discarded. The Government's rights to use,
modify, reproduce, release, perform, display, or disclose these technical data are restricted to those rights specified in DFARS 252.227-7015(b)(2), FAR
52.227-14(g)(2)(Alternate II) and FAR 12.211. Any reproduction of technical data or portions thereof marked with this legend must also reproduce the
markings. Any person, other than the Government, who has been provided access to such data, must promptly notify Thermo Fisher.

.
Page 11 of 11 Confidential and proprietary information
Document: Installation procedure Multiplier Assembly S703758T

Multiplier Assembly

Instrument : XRF
N° Assembly : S703758T
File : S430420-0.doc
Mod. Description Date Name
0. Document creation 9.05.05 JMa
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 1 / 10
Document: Installation procedure Multiplier Assembly S703758T

1 Introduction:
The Multiplier Assembly includes two different parts:
A) XPSM board
B) Multiplier box

Note: XPSM board and Multiplier box can not be delivered separately.
Those two parts must always remain connected together.
Don’t disconnect cables between those two parts.

Multiplier Assembly is cooled with water.


2x male fast fittings connectors are installed on the Multiplier Assembly.

2x fans are installed on the side panel of the Multiplier Assembly.

2 Assembly configuration:

2.1 Multiplier Assembly:


Below is the picture of the Multiplier Assembly:

Fan 1 installed on Fan 2 installed on


the side panel. HV connector JP1
the side panel.

Air direction

High Voltage Multiplier box


transformer

Connector J906 XPSM board Connector J1 2x Water male fast


fitting connectors
Multiplier Assembly top view

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 2 / 10
Document: Installation procedure XPSC3660 board S703722T

2.2 XPSM board configuration:


Below is the layout of the XPSM board:

Connector P1 on Connector J906.


Connector J1. components side Components side
Components side

COMPONENTS VIEW

XPSM board has no bridge and no switch to be configured.

XPSM board has 2x electrical external connectors:


Connector identification: Connected to:
J1 -> Filament X-ray tube cable Wago 5x
J906 -> GenX backplane board XPSB

Multiplier Assembly has 2x fans located on the side panel and connected to the XPSM board:
Fan identification: Connected to:
Fan 1
-> P1 connector
Installed on the side
Red color of the fan cable is connected to: +24VDC
panel of the Multiplier
Black color of the fan cable is connected to: PGND
Assembly
Fan 2 -> P1 connector
Installed on the side
Red color of the fan cable is connected to: +24VDC
panel of the Multiplier
Black color of the fan cable is connected to: PGND
Assembly
Note:
Air direction of the fans 1 and 2 are oriented to the XPSM board side.

3 Firmware compatibility

No firmware is installed on the Multiplier assembly.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 3 / 10
Document: Installation procedure XPSC3660 board S703722T

4 Board installation
IMPORTANT:
Before to have access inside the GenX generator, the following caution must
be read carefully in order to avoid any trouble occurring to the generator and
any body’s injury.

Warning:
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk:

DANGER: LETHAL VOLTAGES are present in the GenX generator


whenever the generator is connected to the power line source.

DANGER: LETHAL STORED ENERGY is still present in the GenX


generator after disconnection of the power line source.
Wait 15 minutes to be sure that all stored energy is discharged
after power disconnection.

Accessibility inside the GenX generator:

The “opening” of the GenX generator, which is necessary for


the maintenance and repair, must be STRICTLY done by a
certified service enfineer who has followed a special GenX
training course.

Thermo Electron Ecublens Switzerland will be not responsible for


any injury occurred on body person in case of no respect of
the special training accreditation given for GenX generator.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 4 / 10
Document: Installation procedure Multiplier Assembly S703758T

4.1 Location of the Multiplier assembly inside the GenX


generator:

XPSI XPSR XPSC XPSP Multiplier


Board Board Board Board Ground Assembly
connection

Connector
JP 1

J1
Water male
fast fitting
connectors

Rear view of the GenX generator

4.2 Installation procedure


This procedure starts from the step when the GenX generator is already removed
from the XRF instrument.
Note: For more information concerning the removal of the generator from the XRF
instrument, consult the GenX generator service manual P/N: AA83678, delivered
during the certification training course given in Ecublens by TTI department.

- Put the GenX generator on a clean table.


Important:
During the removal and installation of the GenX boards and Multiplier
Assembly, always put down the GenX generator with bottom side in
contact with the surface table, see below picture. Don’t turn the GenX
upside down.

Rear side

TOP side Right side

Left
side

Front side
BOTTOM side
GENX Generator VIEW

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 5 / 10
Document: Installation procedure Multiplier Assembly S703758T

VERY IMPORTANT:
A) WAIT 15 MINUTES to be sure that all stored energy is discharged after
power disconnection.
B) Unplug and plug action of all electronic boards must be delayed by at
least 15 MINUTES.
No respect of this caution note can cause serious damage on the GenX
backplane board XPSB and any body’s injury due to the capacitor stored
energy.

- Disconnect the two remaining water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- Remove the rear panel of the GenX generator by unscrewing totally 4x screws.
Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Remove the top cover of the GenX generator by unscrewing totally 4x screws.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Remove XPSI, XPSR (if installed), XPSC and XPSP boards by following the
below IMPORTANT General Note.

IMPORTANT General Note:


If you want to remove all the GenX boards located inside the GenX, you
must follow a certain removal order boards as described below to avoid
damage on the electronic components part (some boards are very closed
between them):
-> First board to be removed is -> XPSI board
-> Second two boards to be removed are -> XPSR (if installed) together
with XPSC boards.
-> Third board to be removed is -> XPSP board
-> Fourth “board” to be removed is -> Multiplier assembly (don’t forget the
3x locking screws located at the bottom panel of the GenX generator).

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 6 / 10
Document: Installation procedure XPSC3660 board S703722T

- Remove the Multiplier Assembly by unscrewing the 3x locking screws located at


the bottom panel of the GenX generator.
See below picture:

Consult the GenX generator service manual P/N: AA83678, to see the screws
location.
Use the water fast fitting connector of the multiplier block to extract the Multiplier
Assembly from its location.
Pull the Multiplier Assembly slightly without a big force.
Use the opening of the top side generator to facilitate, with the help of your
fingers, the way out of the High Voltage transformer located on the Multiplier
Assembly.
Notes:
1) Multiplier Assembly (through XPSM boad) is connected to a unique
backplane board XPSB located near the GenX front panel.
2) Multiplier Assembly is mechanically secured by three screws located at
the bottom panel of the GenX generator.

- Install the new Multiplier Assembly in the correct location, see chapter 4.1 of this
document.
For only the first installation of the Multiplier Assembly in the GenX
generator, the two connectors J906, male side on the XPSM board and female
side on the GenX backplane board XPSB, must be well mechanically aligned.
See below procedure.

Important: Once the Multiplier Assembly is mechanically aligned, no need to


repeat this connector J906 backplane alignment.

Hereafter is the mechanical alignment procedure of the Multiplier Assembly for


the first installation only:

A) Insert the new Multiplier Assembly in the correct location, see chapter 4.1
of this document, at about 5cm before the final board location.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 7 / 10
Document: Installation procedure Multiplier Assembly S703758T

B) Unscrew of about ¼ turn the 5x nuts located on the XPSM board solder side.
See below picture:

5x nuts of the XPSM board


to be unscrewed of ¼ turn

C) Install the Multiplier Assembly to the J906 backplane connector by taking


care of the following points:
- Insert the Multiplier Assembly in the J906 backplane connector by
checking visually the good parallelism alignment between the 2x J906
connectors. See below picture:

Check visually the good


parallelism alignment
between the 2x J906
connectors

- Install the Multiplier Assembly to the final locked mechanical position


by passing through two “distinct clicks positions”.
Use your 2x hands to press the Multiplier Assembly to the final position
as shown in the below picture:

Use your right hand to


hold back the
backplane board XPSB

Use your left hand to


press the XPSM board
to the final position.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 8 / 10
Document: Installation procedure Multiplier Assembly S703758T

D) Fix the Multiplier Assembly by screwing the 3x locking screws located at


the bottom panel of the GenX generator:
See below picture:

E) Fix the XPSM board to the Multiplier block by tightening (not too strong)
of ¼ turn the 5x nuts located on the XPSM board solder side.
Follow the nuts order 1->2->3->4->5 as specified in the below picture:

1
3

5
4 2

Follow the nuts order 1->2->3->4->5.


Screw of ¼ turn the 5x nuts.

- Install all the GenX generator boards respecting the following order:
-> First board to be installed is -> XPSP board
-> Second two boards to be installed are -> XPSR (if installed) together with
XPSC board
-> Third board to be installed is -> XPSI board
Take care to avoid to damage electronic components located on the way
during installation of the GenX generator board.

- Install the top cover of the GenX generator by screwing totally 4x screws.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

- Check the correct installation of all the GenX boards and multiplier assembly
(correct location and final mechanical position) to the backplane XPSB board.

- Install the rear panel of the GenX by screwing totally 4x screws.


Note: Only in case of 1x longer screw and 3x shorter, the longer screw is located
on the top of the GenX generator and on the multiplier assembly side.
Consult the GenX generator service manual P/N: AA83678, to see the screws
location.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 9 / 10
Document: Installation procedure Multiplier Assembly S703758T

- Connect the two water fast fitting connectors:


1x connexion between XPSI and XPSP boards
1x connexion at the multiplier assembly

- At this step, the GenX generator is ready to be installed into the XRF instrument.
Consult the GenX generator service manual P/N: AA83678, for the installation of
the GenX generator into the XRF instrument.

4.3 Calibration:

No kV/mA calibration is required when Multiplier assembly is replaced.

5 Special packaging for returned material:


The Multiplier Assembly requires special packaging for transportation to avoid
damage.
Important: This special packaging must always be used for transportation purposes.

N°. N° Mod

Thermo Electron SA
En Vallaire Ouest C
Case postale
CH- Switzerland 1024
Ecublens
S430420 0
Page 10 / 10
Document : DOC,Update Firmware generator X-Ray

Update Firmware generator X-Ray

Instrument :

N° Ensemble :

Fichier : P:\Docu\ZAUGG\S428625-0.doc

Mod. Description Date Nom


0 Création du document 14.03.00 JMa
1
2
3
4
N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 1 / 21 0
Document : DOC, Update Firmware generator X-Ray

INTRODUCTION
You have received the Firmwares kit S703xxx which includes all the firmwares necessary for the well function of
a X-Ray generator 3.6 kW limited at 3 kW for the instruments 9400 with S/N < 200 and 9800 with S/N < 100.
In this Firmware kit S703xxx, there are firmwares which are not used, depending of your instruments type and
configuration.

A) If you have a 9400 instrument :

1. With a 12 position magazine, install the following firmwares :


XQM88 V3.x-x ! This version needs 1 more RAM 32K x 8
XQM96 V3.x-x
XGT V3.x-x
XCM V3.x-x
XPS60RT V3.x-x
XGI V3.x-x ! This version needs 1 more RAM 32K x 8
XGDI V3.x-x
XDI V3.x-x

Note : the following firmware isn’t used in your 9400 instrument, you can throw it out :

XSH V3.x-x

2. With a XY magazine, install the following firmwares :


XQM88 V3.x-x ! This version needs 1 more RAM 32K x 8
XQM96 V3.x-x
XGT V3.x-x
XSH V3.x-x
XPS60RT V3.x-x
XGI V3.x-x ! This version needs 1 more RAM 32K x 8
XGDI V3.x-x
XDI V3.x-x

Note : the following firmware isn’t used in your 9400 instrument, you can throw it out :

XCM V3.x-x

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 2 / 21 0
Document : DOC, Update Firmware generator X-Ray

B) If you have a 9800 instrument :

Install the following firmware :


XPS60RT V3.x-x
Note : the following firmwares aren’t used in your 9800 instrument, you can throw them out :
XQM88 V3.x-x
XQM96 V3.x-x
XGT V3.x-x
XCM V3.x-x
XSH V3.x-x
XGI V3.x-x
XGDI V3.x-x
XDI V3.x-x
For your information :
The switches configuration is described in the following X-Ray Technical News No. 62 document.

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 3 / 21 0
Document : DOC, Update Firmware generator X-Ray

X-RAY TECHNICAL NEWS : 62


ARL Applied Research Laboratories S.A. Tel. X-ray TSG: (++41)(21)694 7619
Technical Support Group E-Mail X-Ray TSG : [email protected]
En Vallaire Ouest C, case postale Tel. Head Office: (++41)(21)694 7111
CH-1024 Ecublens Fax Head Office: (++41)(21)694 7112
Switzerland

X-RAY TECHNICAL SUPPORT

INSTRUMENTS CONCERNED : 9400 and 9800 Instruments

SUBJECT : Switches configuration on the electronic boards

AUTHOR : J. B. Masset

DATE : 10th January 2000

TABLE OF CONTENTS :

Subject description :
1. Introduction
2. Switches configuration on the electronic boards for 9400 instruments
3. Switches configuration on the electronic boards for 9800 instruments

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 4 / 21 0
Document : DOC, Update Firmware generator X-Ray

Subject number : 1

Introduction :
The purpose of this Technical News is to inform you of the switches configuration on the different
9400/9800 electronic boards.
This Technical News will be edited periodically in order to have always the latest update.

Action in the field :


When you change an electronic board in the field, verify the swithes configuration according to this Technical
News No. 62.
Personal notes :

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 5 / 21 0
Document : DOC, Update Firmware generator X-Ray

Subject number : 2

Switches configuration on the electronic boards for 9400 instruments :


The purpose of this subject is to inform you of the switches configuration on the 9400 instruments electronic
boards.

Note : If you replace the display by a new type, on the 9400 instruments, you must upgrade all the firmwares to
the version V3.x-x.

Hereafter is the list of the last firmware version dated January 2000 :
XQM88 V3.0-0 S702906 ! This version needs 1 more RAM 32K x 8 ( A033453 )
XQM96 V3.0-0 S702907
XGT V3.1-0 S702908
*XCM V3.1-0 S702909 *Option : Magazine 12 positions
*XSH V3.0-2 S702990 *Option : Magazine XY
*XPS60RT V3.1-1 S703022 *Installed in 9400 instr. with S/N < 200
*XPS60 V3.1-1 S703020 *Installed in 9400 instr. with S/N >= 200
*XPS70RT V3.1-1 S703023 *Installed in 9400 instr. with S/N < 200. Option 70 kV
*XPS70 V3.1-1 S703021 *Installed in 9400 instr. with S/N >= 200. Option 70 kV
XGI V3.1-1 S703019 ! This version needs 1 more RAM 32K x 8 ( A033453 )
XGDI V3.1-0 S702911
XDI V3.1-0 S702912

Note :These above firmwares are the latest version at present.

Hereafter is the 9400 firmware compatibility table :


SLAVE NODES
V 3.0
V 3.1
V 3.1
V 3.1
V 3.0
V 3.1
V 3.1
V 3.1
XPS60/70/RT
XQM96

XGDI
XCM
XGT

XSH
XGI

XDI

XQM88 V 3.0 a a a a a a a a
MASTER
NODE
XQM96 V 3.0 a a a a
XGI V 3.1 a a

a Compatible r Incompatible Not applicable

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 6 / 21 0
Document : DOC, Update Firmware generator X-Ray

Switches function on the different boards for the 9400 instruments :


1) XQM88 V3.0-0 S702906 :
This firmware is installed on the XQM board S702002T “ MMB-II part “.
XQM88 chip’s location : Two EPROMS : EPROM 1/2 in IC7 socket and EPROM 2/2 in IC11 socket.
Three RAM ( A033453 ) 32k x 8 in IC10 / IC15 / IC16 sockets.
XQM88’s switches and jumpers function located on the XQM board “ Adaptation part “ :

1 ON Return global alarm numbers


OFF Return normal alarm numbers
2 ON Return diagnostic alarms
OFF Do not return diagnostic alarms
3 ON Return information alarms
OFF Do not return information alarms
S1 4 OFF reserved
5 OFF reserved
6 ON Display type SAMSUNG or ISE (new)
OFF Display type NEC (old)
7 OFF reserved
8 OFF reserved
W1 2 INSTALLED No detector on open position of primary chamber shutter
REMOVED Detector on open position of primary chamber shutter
3 INSTALLED Issue a fatal alarm if no cassette is loaded
REMOVED Issue a warning alarm if no cassette is loaded

Note : In case of installation of a new displays kit S702928 ( P/N for old display type is : A035481 and new type is
S702777 ), put the switch S1-6 to ON position.

For information : the new display has been introduced since the instrument 9400 >= S/N 223.

2) XQM96 V3.0-0 S702907 :


This firmware is installed on the XQM board S702002T “ Adaptation part “.
XQM96 chip’s location : One EPROM in IC4 socket on the XQM board “ Adaptation part “.

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 7 / 21 0
Document : DOC, Update Firmware generator X-Ray

3) XGT V3.1-0 S702908 :


This firmware is installed on the XGT board S702003T or S702733T.
XGT chip’s location : One EPROM in IC6 socket.

Switches and jumpers function on the XGT board :


1 ON Allow HVPS independently of FPC gas regulation
OFF Allow HVPS only if FPC gas in regulation
S1 2 ON Disable FPC gas regulation
OFF Enable FPC gas regulation
3 ON Enable HVPS under AIR environment
OFF Disable HVPS under AIR environment
W1 1,2 Instrument type
W1.2 W1.1
Removed removed ARL-9400
removed installed Reserved
installed removed ARL-9800
installed installed Reserved
3 REMOVED Reserved
4 REMOVED Reserved

4) XCM V3.1-0 S702909 :

This firmware is installed on the XCM board S702004T or S702734T.


XCM chip’s location : One EPROM in IC4 socket.
Switches and jumpers function on the XCM board :
1 ON Reserved
2 ON Reserved
S1 3 ON Reserved
4 ON Instrument type ARL-9400
OFF Instrument type ARL-9800

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 8 / 21 0
Document : DOC, Update Firmware generator X-Ray

5) XSH V3.0-2 S702990 :

This firmware is installed on the XSH board S702579T.


XSH chip’s location : One EPROM in IC5 socket.
Switches and jumpers function on the XSH board :
1,2,3 S1.3 S1.2 S1.1
OFF OFF ON Cassette trays version
OFF ON OFF Sample and cassette trays version
ON OFF OFF Sample trays version
S1 6,7 S1.7 S1.6
OFF X Show disabled
ON OFF Show active without load
ON ON Show active with load
8 ON Covers safety inactive
OFF Covers safety active

6) XPS RT V3.1-1 S703022/23 for instrument 9400 < 200 :

This firmware is installed on the XPS board S702007T.


XPS chip’s location : One EPROM in IC4 socket.
Switches and jumpers function on the XPS board :
1 ON Reserved
2 ON Secondary water output temperature regulation inactive
OFF Secondary water output temperature regulation active
3,4 X-ray generator type
SW1.4 SW1.3
ON ON 3.0 kW generator
SW1 ON OFF 3.6 kW generator
OFF ON 4.2 kW generator
OFF OFF reserved
5 ON
Switch off generator on specific events:
-Leak current lasts for more than 2 seconds
OFF Do not switch off generator on specific events
6 ON Reserved
7 ON Reserved
8 ON Reserved

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 9 / 21 0
Document : DOC, Update Firmware generator X-Ray

7) XPS V3.1-1 S703020/21 for instrument 9400 >= 200 :

This firmware is installed on the XPS board S702847T.


XPS chip’s location : One EPROM in IC4 socket.
Switches and jumpers function on the XPS board :
1 ON Reserved
2 ON Secondary water output temperature regulation inactive
OFF Secondary water output temperature regulation active
3,4 X-ray generator type
SW1.4 SW1.3
ON ON 3.0 kW generator
SW1 ON OFF 3.6 kW generator
OFF ON 4.2 kW generator
OFF OFF reserved
5 ON
Switch off generator on specific events:
-Leak current lasts for more than 2 seconds
OFF Do not switch off generator on specific events
6 ON Reserved
7 ON Reserved
8 ON Reserved

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 10 / 21 0
Document : DOC, Update Firmware generator X-Ray

8) XGI V3.1-1 S703019 :

This firmware is installed on the XGI board S702005T or S702735T “ MMB-II part “.
XGI chip’s location : Two EPROMS : EPROM 1/2 in IC7 socket and 2/2 in IC11 socket.
Three RAM ( A033453 ) 32k x 8 in IC10 / IC15 / IC16 sockets.
XGI’s switches and jumpers function located on the XGI board “ Adaptation part “ :

1 ON SHOW starts at power-up


OFF Normal goniometer functions at power-up
ON XDI board simulated
2
OFF Normal dialogue between XGI and XDI
ON Use coefficients in non-volatile memory for crystal/detector position
S1
3 regulation
OFF Use default coefficients for crystal/detector position regulation
4-6 ON/OFF Goniometer identity number (S1.4 to S1.6 are OFF on 9400-XRF)
7 not used
8 not used
1 INSTALLED Goniometer without crystal changer
REMOVED Goniometer with crystal changer
INSTALLED Goniometer without collimator changer
2
REMOVED Goniometer with collimator changer
3 not used
INSTALLED During SHOW, alarms summary displayed continuously
4
W1 REMOVED During SHOW, alarms summary displayed every 10 passes
W1.6 W1.5
Rem rem Default coefficients for crystal/detector position regulation
5,6 Rem inst 50% of default coefficients for crystal/detector position regulation
Inst rem 150% of default coefficients for crystal/detector position regulation
Inst inst not used
7 not used
8 not used

9) XGDI V3.1-0 S702911 :

This firmware is installed on the XGI board S702005T or S702735T “ Adaptation part “.

XGDI chip’s location : One EPROM in IC1 socket on the XGI board “ Adaptation part “.

10) XDI V3.1-0 S702912 :

This firmware is installed on the XDI board S702006T or S702736T.


XDI chip’s location : One EPROM in IC2 socket.
Switches and jumpers function on the XDI board :
S1 1 ON Test pulse generator ON
OFF Test pulse generator OFF

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 11 / 21 0
Document : DOC, Update Firmware generator X-Ray

For your information, find hereafter the firmware compatibility table of the V2.x-x :

S LAV E N O DE S

V 2.0
V 2.1
V 2.2
V 2.3
V 2.4
V 2.0
V 2.0
V 2.0
V 2.1
V 2.2
XPS60/70(RT)V 2.3
V 2.1
V 2.0
V 2.1
V 2.2
V 2.0
V 2.0
V 2.1
XPS60/70
XPS60/70
XPS60/70
MASTER
XQM96
XQM96
XQM96
XQM96
XQM96
NODES

XGDI
XCM
XGT

XSH
XGI
XGI
XGI

XDI
XDI
XQ M 88 V 2 .0 ‰ ‰ ‰ ‰ ‰ ‰ ‰ ‰ ‰    ‰   ‰ ‰ ‰
XQ M 88 V 2 .1  ‰ ‰ ‰ ‰ ‰ ‰ ‰ ‰   ‰ ‰   ‰ ‰ ‰
XQ M 88 V 2 .2  ‰ ‰ ‰ ‰ ‰ ‰ ‰ ‰   ‰  ‰ ‰ ‰ ‰ ‰
XQ M 88 V 2 .3     ‰ ‰ ‰   ‰ ‰ ‰  ‰ ‰ ‰ ‰ ‰
XQ M 96 V 2 .0 ‰ ‰ ‰ ‰ ‰ ‰ 
XQ M 96 V 2 .1 ‰ ‰ ‰ ‰ ‰ ‰ ‰
XQ M 96 V 2 .2 ‰ ‰ ‰ ‰ ‰ ‰ ‰
XQ M 96 V 2 .3 ‰ ‰ ‰ ‰ ‰ ‰ ‰
XQ M 96 V 2 .4 ‰ ‰ ‰ ‰ ‰ ‰ ‰
XG I V 2 .0 ‰ ‰ 
XG I V 2 .1 ‰ ‰ 
XG I V 2 .2 ‰  ‰

‰ c o m p a tib le  in c o m p a tib le n o t a p p lic a b le

Action in the field :


When you change an electronic board in the field, verify the switches configuration according to this Technical
News No. 62.
Personal notes :

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 12 / 21 0
Document : DOC, Update Firmware generator X-Ray

Subject number : 3

Switches configuration on the electronic boards for 9800 instruments :

The purpose of this subject is to inform you of the switches configuration on the 9800 instruments
electronic boards.
Hereafter is the list of the last firmware version dated January 2000 :
XQI V3.3-1 S702950
XSM V3.0-1 S702879
XGT V3.1-0 S702908
XST V3.1-1 S702965
XSP V3.0-1 S702793
XMIR V3.0-0 S702643 *Option : Mono
*XCM V3.1-0 S702909 *Option : Magazine 12 positions
*XHI V3.2-2 S702959 *Option : Magazine XY
*XPS60RT V3.1-1 S703022 *Installed in 9800 instr. with S/N < 100
*XPS60 V3.1-1 S703020 *Installed in 9800 instr. with S/N >= 100
*XPS70RT V3.1-1 S703023 *Installed in 9800 instr. with S/N < 100. Option 70 kV
*XPS70 V3.1-1 S703021 *Installed in 9800 instr. with S/N >= 100. Option 70 kV
*XGI V3.1-1 S703019 *Option : Gonio
*XGDI V3.1-0 S702911 *Option : Gonio
*XDD V3.1-1 S702996 *Option : XRD
*XDDD V3.0-1 S702949 *Option : XRD
*XDI V3.1-0 S702912 *Option : Gonio or XRD

Note : These above firmwares are the latest version at present.

Hereafter is the 9800 firmware compatibility table :


SLAVE NODES
V 3.0
V 3.0
V 3.1
V 3.0
V 3.1
V 3.0
V 3.1
V 3.1
V 3.2
V 3.0
V 3.1
V 3.0
V 3.0
V 3.0
V 3.1
V 3.0
V 3.1
V 3.0
V 3.1
V 3.0
V 3.0
V 3.1
XPS60/70/RT
XPS60/70
XGT98

XDDD
XMIR

XGDI
XGDI
XCM
XCM
XSM

XDD
XDD
XGT

XST
XST

XSP
XHI
XHI

XGI
XGI

XDI
XDI

XQI V 3.0 a a r a r a a r r a r a a a r a r a r a a r
XQI V 3.1 a a r a r a a a r a r a a a r a r a r a a r
MASTER NODES

XQI V 3.2 a a r a r a a r a r a a a a r a r a r a a r
XQI V 3.3 a a a a a a a r a r a a a a a a a a a a a a
XSM V 3.0 a a a a a a a a a a a a
XGI V 3.0 a a a a
XGI V 3.1 a a a a
XDD V 3.0 a a a
XDD V 3.1 a a a

a Compatible r Incompatible Not applicable

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 13 / 21 0
Document : DOC, Update Firmware generator X-Ray

Switches function on the different boards for the 9800 instruments :


1) XQI V3.3-1 S702950 :
This firmware is installed on the XQM board S702417T “ MMB-II part “.
XQI chip’s location : Two EPROMS : EPROM 1/2 in IC7 socket and EPROM 2/2 in IC11 socket.
Three RAM ( A033453 ) 32k x 8 in IC10 / IC15 / IC16 sockets.
XQI’s switches and jumpers function located on the XQM board “ Adaptation part “ :

1 ON Return global alarm numbers


OFF Return normal alarm numbers
2 ON Return diagnostic alarms
OFF Do not return diagnostic alarms
3 ON Return information alarms
OFF Do not return information alarms
4 ON Stop cassette rotation at origin
OFF Stop cassette rotation immediately
S1 5 ON During a quantitative analysis, the counting on a monochromator is stopped
only when the counting on all the goniometers and all the monochromators
is over.
OFF The counting time on each monochromator is the one defined in the
monochromator lines library
6 OFF Reserved
7 OFF Reserved
8 ON NS and PA commands processes are simulated
OFF NS and PA commands processes are executed normally
1 Removed Reserved
2 Installed Load of unexpected sample is allowed
Removed Load of unexpected sample is not allowed
3 Removed Reserved
W1 4 Removed Reserved
5 Removed Reserved
6 Removed Reserved
7 Removed Reserved
8 Removed Reserved

2) XSM V3.0-1 S702879 :


This firmware is installed on the XQM board S702417T “ Adaptation part “.
XSM chip’s location : One EPROM in IC4 socket on XQM board “ Adaptation part “.

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 14 / 21 0
Document : DOC, Update Firmware generator X-Ray

3) XGT V3.1-0 S702908 :

This firmware is installed on the XGT board S702733T.


XGT chip’s location : One EPROM in IC6 socket.
Switches and jumpers function on the XGT board :
1 ON Allow HVPS independently of FPC gas regulation
OFF Allow HVPS only if FPC gas in regulation
S1 2 ON Disable FPC gas regulation
OFF Enable FPC gas regulation
3 ON Enable HVPS under AIR environment
OFF Disable HVPS under AIR environment
4 OFF Reserved
1,2 Instrument type
W1.2 W1.1
removed removed ARL-9400
removed installed Reserved
W1 installed removed ARL-9800
installed installed Reserved
3 Removed Reserved
4 Removed Reserved

4) XST V3.1-1 S702965 :

This firmware is installed on the XST board S702576T.


XST chip’s location : One EPROM in IC5 socket.
Switches and jumpers function on the XST board :
1,2 S1.2 S1.1
OFF OFF Sample height detection enabled
OFF ON Sample height detection disabled.
Sample height forced for cassette of 50 mm
ON OFF Sample height detection disabled.
Sample height forced for cassette of 36 mm
S1 ON ON Sample height detection disabled.
Sample height determined automatically by the analysis lift
3 ON Analysis lift moves at full speed
OFF Analysis lift in low speed approach mode
4 ON Analysis lift close completely the sample stage
OFF Analysis lift close partially the sample stage
5 ON Load Lift driver security enabled
OFF Load Lift driver security disabled
6 OFF Reserved
7 OFF Reserved
8 OFF Reserved

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 15 / 21 0
Document : DOC, Update Firmware generator X-Ray

5) XSP V3.0-1 S702793 :

This firmware is installed on the XSP board S702594T.


XSP chip’s location : One EPROM in IC5 socket.
Switches and jumpers function on the XSP board :
1 OFF Reserved
2 OFF Reserved
S1 3 OFF Reserved
4 OFF Reserved

6) XMIR V3.0-0 S702643 :

This firmware is installed on the XMI board “Digital part” S702456.


XMIR chip’s location : One micro-controller 87C196KR in IC6 socket.
Switches and jumpers function on the XMI “Digital part” board :
SW1 0-9 Unit digit coding of the node identity number
SW2 0-3 Decimal digit coding of the node identity number
1 Installed Micro-controller watchdog function disabled (read at power-up)
Removed Micro-controller watchdog function enabled (read at power-up)
BR 2 Installed Monochromator intensity filter position sensors disabled
Removed Monochromator intensity filter position sensors enabled
3 Removed Reserved
4 Removed Reserved

7) XCM V3.1-0 S702909 :

This firmware is installed on the XCM board S702734T.


XCM chip’s location : One EPROM in IC4 socket.
Switches and jumpers function on the XCM board :
1 ON Reserved
2 ON Reserved
S1 3 ON Reserved
4 ON Instrument type ARL-9400
OFF Instrument type ARL-9800

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 16 / 21 0
Document : DOC, Update Firmware generator X-Ray

8) XHI V3.2-2 S702959 :

This firmware is installed on the XHI board S702577T.


XHI chip’s location : One EPROM in IC5 socket.
Switches and jumpers function on the XHI board :
1,2,3 S1.3 S1.2 S1.1
OFF OFF ON Cassette version
OFF ON OFF Sample version
ON OFF OFF OEM version
4 OFF Reserved
S1 5 OFF Reserved
6 OFF Reserved
7 ON Show active
OFF Show disabled
8 ON Covers safety inactive
OFF Covers safety active

9) XPS RT V3.1-1 S703022/23 for instrument 9800 < 100 :

This firmware is installed on the XPS board S702737T.


XPS chip’s location : One EPROM in IC4 socket.
Switches and jumpers function on the XPS board :
1 ON Reserved
2 ON Secondary water output temperature regulation inactive
OFF Secondary water output temperature regulation active
3,4 X-ray generator type
SW1.4 SW1.3
ON ON 3.0 kW generator
SW1 ON OFF 3.6 kW generator
OFF ON 4.2 kW generator
OFF OFF reserved
5 ON Switch off generator on specific events:
-Leak current lasts for more than 2 seconds
OFF Do not switch off generator on specific events
6 ON Reserved
7 ON Reserved
8 ON Reserved

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 17 / 21 0
Document : DOC, Update Firmware generator X-Ray

10) XPS V3.1-1 S703020/21 for instrument 9400 >= 100 :

This firmware is installed on the XPS board S702869T.


XPS chip’s location : One EPROM in IC4 socket.
Switches and jumpers function on the XPS board :
1 ON Reserved
2 ON Secondary water output temperature regulation inactive
OFF Secondary water output temperature regulation active
3,4 X-ray generator type
SW1.4 SW1.3
ON ON 3.0 kW generator
SW1 ON OFF 3.6 kW generator
OFF ON 4.2 kW generator
OFF OFF reserved
5 ON
Switch off generator on specific events:
-Leak current lasts for more than 2 seconds
OFF Do not switch off generator on specific events
6 ON Reserved
7 ON Reserved
8 ON Reserved

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 18 / 21 0
Document : DOC, Update Firmware generator X-Ray

11) XGI V3.1-1 S703019 :

This firmware is installed on the XGI board S702735T “ MMB-II part “.


XGI chip’s location : Two EPROMS : EPROM 1/2 in IC7 socket and 2/2 in IC11 socket.
Three RAM ( A033453 ) 32k x 8 in IC10 / IC15 / IC16 sockets.
XGI’s switches and jumpers function located on the XGI board “ Adaptation part “ :

1 ON SHOW starts at power-up


OFF Normal goniometer functions at power-up
ON XDI board simulated
2
OFF Normal dialogue between XGI and XDI
ON Use coefficients in non-volatile memory for crystal/detector position
S1
3 regulation
OFF Use default coefficients for crystal/detector position regulation
4-6 ON/OFF Goniometer identity number (S1.4 to S1.6 are OFF on 9400-XRF)
7 not used
8 not used
1 INSTALLED Goniometer without crystal changer
REMOVED Goniometer with crystal changer
INSTALLED Goniometer without collimator changer
2
REMOVED Goniometer with collimator changer
3 not used
INSTALLED During SHOW, alarms summary displayed continuously
4
W1 REMOVED During SHOW, alarms summary displayed every 10 passes
W1.6 W1.5
Rem rem Default coefficients for crystal/detector position regulation
5,6 Rem inst 50% of default coefficients for crystal/detector position regulation
Inst rem 150% of default coefficients for crystal/detector position regulation
Inst inst not used
7 not used
8 not used

12) XGDI V3.1-0 S702911 :

This firmware is installed on the XGI board S702735T “ Adaptation part “.


XGDI chip’s location : One EPROM in IC1 socket on the XGI board “ Adaptation part “.

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 19 / 21 0
Document : DOC, Update Firmware generator X-Ray

13) XDD V3.1-1 S702996 :

This firmware is installed on the XRDI board S702419T “ MMB-II part “.


XDD chip’s location : Two EPROMS : EPROM 1/2 in IC7 socket and 2/2 in IC11 socket.
Three RAM ( A033453 ) 32k x 8 in IC10 / IC15 / IC16 sockets.
XDD’s switches and jumpers function located on the XRDI board “ Adaptation part “ :
1 ON SHOW starts at power-up
OFF Normal XRD-system functions at power-up
2 ON XDI board simulated
OFF Normal dialogue between XDD and XDI
S1 3 ON Use coefficients in non-volatile memory for detector position regulation
OFF Use default coefficients for detector position regulation
4-6 XRD-system identity number
S1.6 S1.5 S1.4
OFF ON ON XRD-system 1
ON OFF OFF XRD-system 2
7 OFF Reserved
8 OFF Reserved
1 Removed Reserved
2 Removed Reserved
3 Removed Reserved
4 Installed During SHOW, alarms summary displayed continuously
W1 Removed During SHOW, alarms summary displayed every 10 passes
5,6 W1.6 W1.5
removed removed Default coefficients for crystal/detector position regulation
removed installed 50% of default coefficients for crystal/detector position regulation
installed removed 150% of default coefficients for crystal/detector position regulation
installed installed not used
7 Removed Reserved
8 Removed Reserved

14) XDDD V3.0-1 S702949 :

This firmware is installed on the XRDI board S702419T “ Adaptation part “.


XDDD chip’s location : One EPROM in IC1 socket on the XRDI board “ Adaptation part “.

15) XDI V3.1-0 S702912 :

This firmware is installed on the XDI board S702736T


XDI chip’s location : One EPROM in IC2 socket.
Switches and jumpers function on the XDI board :
S1 1 ON Test pulse generator ON
OFF Test pulse generator OFF

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 20 / 21 0
Document : DOC, Update Firmware generator X-Ray

Action in the field :


When you will change an electronic board in the field, verify the swithes configuration according to this Technical
News No. 62.
Personal notes :

N°. N°
Mod
S428625
En Vallaire Ouest C CH-1024 Ecublens
Page 21 / 21 0
8

DRAWINGS
Drawings Chapter 8

8 DRAWINGS

Drawings list for the GenX training document :

S112163 GenX Adaptation Interconnection


S112157 GenX Interconnection
S112162 GenX Adaptation Interconnection
S112156 GenX Interconnection
S112287 GenX Interconnection 9900 LP
S010801 GenXRD – GenXRF Interconnection
S112069 XPSB, X-Ray Power Supply Mother Board
S214773 XPSB, X-Ray Power Supply Mother Board
S112064 XPSI, X-Ray Power Supply Input Board (old and new card)
S112070 XPSI, X-Ray Power Supply Input Board (old and new card)
S112065 XPSC, X-Ray Power Supply Control Board
S112071 XPSC, X-Ray Power Supply Control Board
S112306 XPSC V2 Electrical schematic
S112307 XPSC V2 Layout
S112067 XPSP, X-Ray Power Supply Power Board (old and new card)
S112073 XPSP, X-Ray Power Supply Power Board (old and new card)
S112068 XPSM, X-Ray Power Supply Multiplier Assembly
S112074 XPBM, X-Ray Power Supply Multiplier Board
S112066 XPSR, X-Ray Power Supply Retrofit (old and new card)
S112072 XPSR, X-Ray Power Supply Retrofit (old and new card)
S215199 GenX Display Assembly
S112247 X-Ray Power Supply Multiplier Assembly XRD Version (N.H.V.)
S112248 X-Ray Power Supply Multiplier Assembly XRD Version

AA83678 Service Manual ARL GenX 8-1


+5VM

R22 +5V
2K7
+5V
C16 C88
C92 +12V 100nF J902
100nF
100nF R18 +5VRTM AGND

4
10K IC2 AGND R29 +24V 1 1Z
AGND R25 3 + IC5A R24 2 1A
18K 1 8 1 InputVoltage GND 3
AGND 10k C62 R32 2 - U1 4
1B
2 7 10nF 18k OP484 150 P1 +24Vin 5
1C
C51 1D
C93 R21 100nF 1 + 6

11
2Z
+24V C54 C60 3 6 2 7 2A
100nF R27 AGND AGND AGND - PGND 8
10K 100nF 10nF 4 5 18k AGND +12Vin 9
2B
R26 B WAGO 2P 2C
AGND C48 10
R30 FAN 24V 11
2D
10k -12Vin AGND 3Z
+5VM 47K AGND HCPL7800 10nF 12 3A
R28 PGND 13 3B
C49 270 14 3C
Low voltage 8x8bit -12V AGND +5Vin 15
100nF C94 3D
AD status converter DGND 16 4Z
100nF R34 17
IC3 /PwmTrip 4A
10K 18 4B
AGND 20 VCC CH0 1 AGND R38 19 4C
19 2 33K (Usually not intalled) 20
/AdcCs1 V+ CH1 4D
18 CS CH2 3 AGND D18 21 5Z
SdataOut 17 4 +15VRTM2 /PwmTrip 22
Sclk DI CH3 LampCurrent 5A
16 CLK CH4 5 C95 23 5B
AdcSars 15 6 NtoGndVoltage R16 IC4 BAT85 24
SdataIn SARS CH5 InputTemp1 100nF R41 5C
14 DO CH6 7 240K C50 R35 25 5D
13 8 InputTemp2 10K PTC3 8 CurrentLimit 26
Vref+5V R44 +5V SE CH7 Vref+5V 100nF Line1 6Z
12 VREF COM 9 AGND +15VRTM 330 27 6A
11 10 R3 R17 7 2 R69 28
10 AGND DGND
5.6/25°C +5VRTM 150K 1K 29
6B
D8 R51 6C
ADC0838 R20 -5.6VRTM2 6 3 R57 30 6D
C85 AGND DGND 0 AGND 31
C84 10 7Z
10uF 5 32 7A
1nF 1N4007 220 100K +5V 33
T1 D9 C89 7B
1 6 C57 34 7C
AGND AGND D20 C56 C44 D10 100uF/35V 1N5352B R67 HCPL3101 100nF 35
C97 7D
2200uF/25V 100uF/35V 1N5339B 10K 6 36 8Z
7 C35 +5VM 470nF AGND Line2 37

8
1N4007 8A
9 4 IC8B R54 38 8B
R58 5 Vref+5V 39
C99 1nF Q2 + 8C
1nF M2 7 10K 40 8D
5 10 C66 C67 500V 500V R23 MPSA05 6 41
2200uF/25V D13 8 VUM 33-05 10K 2K2 -
42
9Z
2x15V 100uF/35V OP284 9A
D12 PTC4 1N5339B MOV3 7 43

4
InputTemp2 9B
-5.6VRTM 2 44 9C
45 9D
1N4007

+
Line1 5.6/25°C S20K250 C47 DGND AGND C91 R59 R55 R56 46 10Z
RV2 100nF 1nF 470 1K 10K 47
Line2 D19 1 3 48
10A
BYT30 4K7/25° 10B
PTC5 49 10C
+15VRTM2 AGND AGND AGND AGND 50 10D
AGND 51

5 -
+400VRTM 11Z
D6 5.6/25°C 52 11A
1 T2 6 I2 C53 D7 53
SENS2 11B
2200uF/25V 1N5352B +12V 54 11C
1N4007 55
C43 100nF 11D
7 5 1 56 12Z
9 57

SENSOR
C42 100nF 12A
Wires (2 x 0.2mm) 2 58 12B

U/I
D14 D11 59
C69 only for water R15 0.08V/A InputCurrent 12C
0.16V/A

0V OUT
5 10 1N5339B 3 -12V AGND 60 12D
2x15V PTC6 2200uF/25V cooled version 61
C37 1N4007 470 13Z
-5.6VRTM2 6 4 C41 62 13A
R60 D21 63
5.6/25°C AGND 1nF 470 BZX85C5V1 64
13B
13C
22nF SCD20PBR 65 13D
66
3

14Z
AGND AGND AGND 67 14A
FI2 +15VRTM 68 14B
L6 69 14C
FPP 2-45 250uH IC1 70 14D
C46 R31 71 15Z
8 CurrentControl +200VRTM 72
100nF 15A
330 73 15B
7 2 R68 +5V 74 15C
PGND +24V R10 1K 75 15D
-5.6VRTM 6 3 76
4
5
2

C19 16Z
10 77 16A
2 5 100nF 78
1 D16 +5VM DGND 79
16B
P2 R66 16C
1.5K220A R36 80 16D
GND B WAGO 2P 10K HCPL3101 +5VRTM IC7 AGND 5 + IC5B 81
R52 Mon400V 17Z
7 82 17A
Not mounted R4 1 8 10k C70 R39 6 - 83
150 17B
+

OP484 R49 84
-

on Low Power version D17 10K 10nF 18k C83 17C


1.5K250A 2 7 85 17D
InputTemp1 AGND 100nF 86
U2 R19 C79 R48 3 6 AGND 18k 87
18Z
C75 R42 C76 18A
K1 MOV7 RV1 C3 270K 10nF 270 AGND 88 18B
GND C98 100nF 1uF 4 5 89
F3 24V 4K7/25 10k 10nF 18C
0.8A MOV8 1nF 90 18D
FAN 24V 4 500V R47 91
MOV1 0VRTM 19Z
2 MOV4 AGND AGND 200K HCPL7800 92 19A
S20K320 F1 8 3xS20K320 93
J2 Line 6 +400VRTM 94
19B
19C
C8 AGND 95

2
7
8

4
6
6 32A 19D
+10% 5
10nF
I1
0 1 C13 C21 C15 C14 C25 R5 MOV5
96
97
20Z

230V -15% 4
3
2
Earth
GND
GND C23
C2
2.2uF
C1
2.2uF
C45 C68
+24V 5 - +
1000uF 1000uF 1000uF 1000uF 1000uF
250V 250V 250V 250V 250V
10K/5W S20K250 98
99
20A
20B
20C
2.2uF 2.2uF +200VRTM 100
1 Neutral 10nF 20D
101 21Z
WAGO D1 102 21A
MOV2 1N4007 C28 C34 C31 C30 C29 R7 MOV6 103
M1

3
21B
S20K320 1000uF 1000uF 1000uF 1000uF 1000uF 10K/5W S20K250 104 21C
VUM 33-05 250V 250V 250V 250V 250V 105
0VRTM 21D
106 22Z
GND 107 22A
108 22B
109 22C
110 22D
R6 111 23Z
D3 Q1 PowerOn 112 23A
1N4007 MPSA05 +5V 113 23B
2K2 114 23C
115 23D
C18 116 24Z
PGND 100nF 117 24A
+5VRTM 118 24B
IC6 R37 119
IC5C 24C
AGND 10 + R50 120 24D
R11 1 8 8 NtoGndVoltage 121
10k 25Z
R13 3M9 C63 R33 9 - 122 25A
2 7 10nF 18k OP484 150 123
C80 25B
124 25C
3M9 3 6 100nF 125
C59 C40 C58 C39 C74 AGND 126
25D
D4 R14 R12 26Z
1N4007 10nF 10nF 100nF 4 5 R43 AGND 127 26A
2K7 2K7 Vref+2V 128 26B
18k 129 26C
PTC2 GND 1000uF/6.3V 1000uF/6.3V HCPL7800 R40 130 26D
250V/0.1A C71 131 27Z
R53 10k 132 27A
AGND 10nF 133 27B
F2 100K 134 27C
0.8A C96 135 27D
1nF 136
SENS1 28Z
+12V 137 28A
D15 138 28B
5 1 C32 100nF 139
FI1 28C
140
SENSOR

C33 100nF 28D


4 3 2 BZX85C6V2 5x100 141 29Z
U/I

J3 R8 142
Neutral IC5D /AdcCs1 R65 29A
D5
0V OUT

3 5 3 -12V AGND 13 - 143 29B


Earth C100 14 LampCurrent SdataOut R64 144
X-ray ON Lamp 2
1
Line GND 2 1 6 4 C20 20K
10nF 12 +
1N4148
Sclk
AdcSars
R63
R62
145
29C
29D
10nF OP484 146 30Z
IEC FPP 2-45 R9 C38 SdataIn R61 147
CurrentControl 30A
SCD01PBR AGND AGND AGND AGND 100K 10uF/6.3V 148 30B
CurrentLimit 149
PowerOn 30C
150 30D
151 31Z
AGND AGND 152 31A
153 31B
Vref+5V 154
Vref+2V 31C
155 31D
InputCurrent 156
InputVoltage 32Z
157 32A
Mon400V 158 32B
159 32C
GND 160 32D

DIN41612

TP1 TP3 TP7


+12V +5V +5VM
MF1 L1 +24V L2 L4 +5V L5
+24Vin +12Vin MF2 +5Vin MF4
+24V +12V
C17
MF-R250 MF-R110 C11 MF-R110
C7 C12
C6
220uF 100nF
TP4 TP10
C4 C9 1uF Remark: VRTM = Voltage Referenced To the Mains
100nF TP2 AGND AGND TP8 TP9 100nF TP6 TP11
220uF 220uF AGND
AGND GND GND DGND DGND
R1 PGND R2
PGND C5 C10 TP5 Mod: 4 Version série 10.04.03 PPa 7 OM3834 28.10.04 PPa 10 OM4000 03.01.07 PPa
0 L3 220uF 100nF -12V 0
MF3 DGND 5 OM3753 & OM3758 17.12.03 PPa 8 OM3884 21.02.05 PPa 11 Addendum OM4000 16.03.07 PPa
GND -12Vin GND
-12V
6 OM3815 14.06.04 PPa 9 OM3964 06.04.06 PPa 12 OM4194 26.19.11 PPa
MF-R110

XPSI, X-ray Power Supply Input Board Echelle Dessiné MBz 03.10.00 Remplacé par

Contrôlé
Remplace

Liste S703123 - S703778 Schémas S112064 Vu


Ensemble ---------- Circuit S920074 Groupe Technique
Appareil XRF Implantation S112070 Ingénieur M.Bonzon 9850

Format No
ThermoFisher S112064
Mod. No

S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 12
+5V
+5V
C16 C88
100nF J902
100nF
+5VRTM AGND

4
IC2 AGND R29 +24V 1 1Z
3 + IC5A R24 2 1A
1 8 1 InputVoltage GND 3
10k C62 R32 2 - 4
1B
2 7 10nF 18k OP484 150 P1 +24Vin 5
1C
C51

FAN 1
1D
R21 100nF 1 3 + 6

11
2Z
C54 C60 3 6 2 4 7 2A
AGND AGND AGND - PGND 8
100nF 10nF 4 5 18k AGND B WAGO 2P +12Vin 9
2B
R26 C48 2C
10 2D
11 3Z
HCPL7800 10k 10nF -12Vin AGND 12 3A
R28 PGND 13 3B
270 14 3C
AGND +5Vin 15 3D
DGND 16 4Z
17 4A
CurrentLimit /PwmTrip 18 4B
19 4C
R78 20 4D
21 5Z
+15VRTM2 22 5A
100K +5V 23 5B
R16 C89 24
IC4 5C
R70 240K R35 100nF 25 5D
+5VRTM 8 26 6Z
C50 AGND Line1 27

8
R69 330 6A
220 R17 R51 100nF 7 2 IC8A 28 6B
150K 1K + 3 29 6C
C102 D22 6 3 R79 R57 1 R77 30
47uF/35V 10 2 31
6D
1N5339B +5V - 7Z
IC9 PTC3 5 2K2 LM393N 32 7A
+15VRTM 33K 10K 33

4
7B
3 ON/OFF n.c. 5 ACPL-312T 34 7C
5.6/25°C C101 -15VRTM2 R75 R76 35 7D
L7 6 47uF/35V C106 C97 11K 10K 36
MF5 +VOUT 1uF Line2 8Z
2 +VIN +5VM 470nF AGND 37 8A
Com 7 IC8B R54 38 8B
MF-R110 C103 R58 5 Vref+5V 39
Q2 + 8C
220uF/35V 1 -VIN -VOUT 8 C104 7 40 8D
47uF/35V C105 R23 MPSA05 6 3K9 41
D23 W08G TMR3-2423WI PTC4 1uF 10K 2K2 LM393N
-
42
9Z
+

9A
-15VRTM 43 9B
InputTemp2 C107 44
C91 R59 R55 R56 9C
5.6/25°C 100nF 45

2
9D
Placer les condos 1uF Q3 DGND 100pF 1K5 1K5 2K7 46 10Z
C105, C106, C110, C111 C108 RV2 C47 47 10A
1 4K7/25° 100nF 48
-

près des drivers ACPL-312T MOV3 C99 49


10B
10C
IC10 PTC5 S20K250 1nF IKW75N60T 1nF AGND AGND AGND AGND AGND 50

3
10D
+15VRTM2 500V AGND AGND 51 11Z
3 5 R67 R71 +400VRTM 52
ON/OFF n.c. 11A
5.6/25°C C109 10K 1k +12V 53 11B
6 47uF/35V C110 SENS2 54
+VOUT 1uF 11C
11

2 55
8

+VIN 11D
Com 7 5 1 C43 100nF 56 12Z
57

SENSOR
12A
T1 1 -VIN -VOUT 8 C111 Wires (2 x 0.2mm) 2 C42 100nF 58 12B
C112

U/I
47uF/35V only for water R15 59 12C
1uF 0.16V/A InputCurrent

0V OUT
230V-15V TMR3-2423WI PTC6 D19 3 0.08V/A 60 12D
STTH6006W cooled version 61
-15VRTM2 470 13Z
6 4 62
1

R60 D21 13A


5.6/25°C -12V D2 C41 63 13B
Line1 1N4148 1nF 470 BZX85C5V1 64 13C
SCD20PBR AGND AGND 65 13D
C37 C98 66
Line2 14Z
+15VRTM AGND AGND AGND AGND 67 14A
68 14B
10nF 1nF 500V 69 14C
70 14D
+24V IC1 R31 71
PGND
3

CurrentControl +200VRTM 15Z


MOV8 C46 8 72 15A
FI2 R80 100nF 330 73 15B
7 2 R68 +5V 74
2
1

15C
FPP 2-45 P2 S20K320 4.7 R10 1K 75 15D
6 3 76 16Z
B WAGO 2P MOV7 4.7 77 16A
5 C19 78
4
3

16B
Not mounted DGND 100nF 79 16C
250uH
+

80
-

on Low Power version S20K320 ACPL-312T R36 16D


L6 +5VM -15VRTM +5VRTM IC7 5 + IC5B 81
FAN 2 R52
4
5
2

Mon400V 17Z
MOV4 7 82 17A
1 8 AGND 10k 6 - 83 17B
GND OP484 150 84
R4 17C
S20K320 2 7 C70 R39 85
10K C83 86
17D
C79 10nF 18k R49 18Z
C35 R48 3 6 100nF 87
InputTemp1 C75 18A
R19 10nF 270 18k 88 18B
GND F3 K1 270K 1uF 4 5 89 18C
0.8A 24V 1nF 500V R42 AGND AGND 90 18D
5 RV1 C3 C76 AGND 91
MOV1 0VRTM 19Z
13 6 4K7/25° 100nF HCPL7800 10nF 92 19A
S20K320 F1 4 14 10k 93
J2 Line 12 3 R47 AGND 94
19B
19C
6 C8 32A 11 AGND AGND 200K 95 19D
+10% 5
10nF 1 8
D24
STTH6006W
D26
STTH6006W D27
96
97
20Z

230V -15% 4
3
2
Earth
GND
GND C23
C2
2.2uF
C1
2.2uF
C45 C68 9 16
+24V
+400VRTM 98
99
20A
20B
20C
1 2.2uF 2.2uF D1 STTH6006W 100 20D
Neutral 10nF I1 D28 D29 101
MOV5 21Z
WAGO STTH6006W STTH6006W C13 C21 C15 C14 C25 R5 102 21A
MOV2 1N4007 D16 1000uF 1000uF 1000uF 1000uF 1000uF 10K/5W S20K250 103

2
21B
S20K320 1.5K220A Q5 Q4 250V 250V 250V 250V 250V 104 21C
C115 C114 105 21D
1 1 +200VRTM 106 22Z
GND 107 22A

IKW75N60T

IKW75N60T
1nF 1nF 108

3
22B
109 22C
D17 R73 R74 R66 R72 C28 C34 C31 C30 C29 R7 MOV6 110 22D
R6 1.5K250A 10K 1k 10K 1k 1000uF 1000uF 1000uF 1000uF 1000uF 10K/5W S20K250 111 23Z
D3 Q1 PowerOn 250V 250V 250V 250V 250V 112 23A
1N4007 MPSA05 113 23B
2K2 0VRTM 114 23C
115 23D
116 24Z
DGND 117 24A
+5V 118 24B
119 24C
120 24D
121 25Z
C18 122 25A
+5VRTM 100nF 123 25B
IC6 R37 124
IC5C 25C
AGND 10 + R50 125 25D
R11 1 8 8 NtoGndVoltage 126 26Z
R13 3M9 10k 9 - 127 26A
2 7 OP484 150 128 26B
C63 R33 C80 129 26C
3M9 3 6 10nF 18k R43 100nF 130
C58 C39 C74 26D
D4 131 27Z
1N4007 R14 C59 C40 R12 10nF 4 5 132
2K7 100nF 27A
PTC2 2K7 1000uF/6.3V 10nF AGND 18k 133 27B
250V/0.1A AGND 134 27C
GND 1000uF/6.3V HCPL7800 R40 C71 135 27D
136 28Z
137 28A
R53 AGND 10k 10nF 138 28B
F2 139 28C
0.8A Vref+2V 140
100K 141
28D
C96 1nF 5x100 29Z
+12V 142 29A
SENS1 +5VM /AdcCs1 R65 143
SdataOut R64 29B
D15 144 29C
5 1 C32 100nF Sclk R63 145
FI1 AdcSars R62 29D
146
SENSOR

SdataIn R61 30Z


4 3 2 C33 100nF BZX85C6V2 R22 147 30A
U/I

J3 R8 2K7 CurrentControl 148


Neutral IC5D CurrentLimit 30B
0V OUT

3 5 3 13 - D5 149 30C
Earth 14 LampCurrent PowerOn 150
X-ray ON Lamp 2
1
Line GND 2 1 6 4 20K 12 +
C92 +12V 151
30D
31Z
-12V OP484 1N4148 152 31A
IEC FPP 2-45 C20 C100 R9 C38 100nF R18 153
Vref+5V 31B
SCD01PBR 10nF 10nF 100K 10uF/6.3V 10K 154 31C
AGND R25 Vref+2V 155
InputCurrent 31D
18K 156 32Z
AGND InputVoltage 157
Mon400V 32A
AGND AGND AGND AGND AGND AGND AGND 158 32B
C93 159 32C
+24V GND 160
100nF 32D
R27
10K
AGND
R30
+5VM 47K AGND DIN41612

C49 Low voltage 8x8bit


TP1 TP3 TP7 C94 -12V
+5V +5VM 100nF AD status converter
L1 L2 +12V L4 +5V L5 100nF R34
MF1 +24V MF2 MF4
+24Vin +12Vin +5Vin IC3 10K
+24V +12V
C17 AGND 20 VCC CH0 1 AGND R38
MF-R250 MF-R110 C11 MF-R110 19 2 33K
C7 C12
C6
220uF 100nF
TP4 TP10
C4 C9 1uF /AdcCs1 18
V+
CS
CH1
CH2 3 AGND Remark: VRTM = Voltage Referenced To the Mains
100nF TP2 AGND AGND TP8 TP9 100nF TP6 TP11 SdataOut 17 4
220uF 220uF Sclk DI CH3 LampCurrent
AGND GND GND DGND DGND AGND 16 CLK CH4 5 C95
PGND R2 AdcSars 15 6 NtoGndVoltage
R1 SdataIn SARS CH5 InputTemp1 100nF R41
PGND C5 C10 TP5 14 7
13
DO CH6
8 InputTemp2 10K Mod: 7 OM3834 28.10.04 PPa 10 OM4000 03.01.07 PPa 13 Addendum OM4245 25.11.11 JSo-PPa
0 L3 220uF 100nF -12V 0 Vref+5V R44 +5V SE CH7 Vref+5V
MF3 DGND 12 VREF COM 9 AGND 8 OM3884 21.02.05 PPa 11 Addendum OM4000 16.03.07 PPa 14 Finalisation OM4245 05.06.12 JSo
GND -12Vin GND 11 10 R3
-12V 10 AGND DGND
MF-R110 9 OM3964 06.04.06 PPa 12 OM4194 26.09.11 PPa 15 Finalisation bis OM4245 03.07.12 JSo
ADC0838
C84 C85 AGND DGND 0 AGND
10uF/6.3V 1nF
XPSI, X-ray Power Supply Input Board Echelle Dessiné MBz 03.10.00 Remplacé par
Accessories
AGND AGND Contrôlé
Remplace
Hole Hole Liste S703123 - S703778 Schémas S112064 Vu
Ensemble ---------- Circuit S920074 Groupe Technique
Hole GND Appareil XRF Implantation S112070 Ingénieur M.Bonzon 9850

Format No
Hole Hole
ThermoFisher S112064
Mod. No

S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 15
J104 J904
1A 1 +5V
1B 2 GND R89
3 0.266V/A FilamentCurrent W3 1 1Z
1C R10 IC1A EnWd
2A 4 /XgtReset 15K R87 GND 2 1A
5 /XcmReset D3 1K 1 C113 FilCurMon Test 3 1B
2B IC2A
6 /XpsReset * 3 D1 C24 R15 IC18B +5V 4 1C
2C
7 /XshReset 2 2 1 WdToggle 47nF 8 MaxFilCur 1K C84 LD3 +24Vin 5 1D
3A BAT85 R104 R7
3B 8 /Xgi1Reset 1 R70 100uF 6 2Z
9 /Xgi2Reset 7403 BAT85 1uF 1K +12V 7 Pressostat/24V 7 2A
3C R5 D2 470K 35V 390 PGND
4A 10 /XdiReset 74HC14 R6 C6 DspOk 8 2B
11 /XxxReset 1K 1M 1N4148 IC20 CMP401 1K AGND /DspCs Green +12Vin 9 2C
4B 470nF
4C 12 /XxxReady C111 C107 10 2D
13 MonosReady 1 -EA 16 C88 10nF CurrentLimit DGND +5V SsioRxSync 11
5A VREF5.1V 10nF -12Vin AGND 3Z
14 MmbTxd DGND DGND DGND DGND 12 3A
5B
15 MmbRxd /Reset FilCurRef R75 R82
2 +EA 15 10nF CurrentControl C57 SsioTxSync 13
5C DisplayRd2 VIN +5V DGND 3B
6A 16 10K 47K 3 +5V AGND AGND 14 3C
17 +5V C8 14 AGND FilPwmH DspWdToggle +5Vin 15
6B
18 XgnR+ 47nF IC2B C95 SYNC OUTPUT B
+5V +5V 10nF C65 3D
16 4Z
6C Reset DGND
7A 19 XgnR- 3 4 10nF 4 OSC OUT VC 13 C93 C85 17 4A
20 XgnT+ DGND AGND /PwmTrip 18 4B
7B 10nF 10nF
21 XgnT- 5 CT 12 10nF 19

4
7C 74HC14 GND 4C
22 /SpAdcInt AGND 20 4D

108
107
106
105
104
103
102
101

100
8A XpsReady FilPwmL

98
91
90

99
97
96
95
94
93
92
89
88
87
86

85
84
83
81
80
79
78
77
76
75
74
72
70
69
68
67
66
65
64
62
61
60
59
58

71
82
73
63
8B 23 /Cs0 6 RT OUTPUT A 11 AGND IC18A AGND R67 21 5Z
24 /Cs1 +24V R148 D13 6 HvPwmH 22 5A
8C XsnR1+ +5V DGND
25 7 10 2 23

EIA

GND
GND

VDD

VDD

GND
GND
GND
EIB

EIS
EIZ

ETU0
ETU1

AUX0
AUX1

P0

P1

P2

P3

P4

P5

P6

P7

P8

P9

P10

P11

D0
D1
D2
D3
D4
D5
D6
D7
D8
D9
D10
D11
D12
D13
D14
D15
D16
D17
D18
D19
D20
D21
D22
D23
9A XsnR1- XrayLampC Q2 DISCH. SHUTDOWN SafFilCur 10K HvPwmL C151 5B
26 5 24 5C
9B
27 TGND 180 8 SOFT-START 9 1N3595 D17 C45 56 10nF 25
9C XsnT1+ R107 COMP. C99 AH R68 FilPwmH 5D
10A 28 IRF9Z24 CMP401 C104 AL 55 0 26 6Z
29 XsnT1- 10K 1N3595 10nF 1uF 109 54 Not mounted 27
C108

14

13
10B R112 R97 C101 CONVST BH 6A
30 TGND SG3525AN 10nF 110 GND BL 53
R64 0 FilPwmL 28 6B
10C
31 XpsIntOut C156 C155 C154 C153 20K C83 10nF AGND AGND +5VM 111 IC5 Not mounted DGND 29
11A Pressostat/24V 10K Q3 R110 XrayLampOn VDD 6C
11B 32 1nF 100uF 112 GND GND 52 30 6D
11C 33 C168 1nF 1nF 1nF 1nF MPSA05 10K AGND AGND 113 AVDD ADMC401 VDD 51 C72 R136 31 7Z
34 10nF AGND AGND AGND R98 /MaxFilCur C114 114 AVSS 100 32 7A
12A
35 D11 10nF AGND 115 VIN7 +5V DGND
12B XdiHvpsIlk 10K CH 50 10nF
33 7B
12C 36 Synchro DGND +5V AGND 1N3595 116 VREF CL 49 34 7C
37 Dvm+ PGND FilCurSD -5V AGND C7 C15 Hv 117 VIN6 48 10K 35 7D
13A XrayLamp1 R109 10K Lamp1 PWMSYNC R61
13B 38 Dvm- AGND 118 REFCOM PWMTRIP 47 +5V R49 36 8Z
39 XgtReady R115 XrayLamp2 Lamp2 /EnFilCu 10uF 100nF 119 VIN5
13C C9 AGND PWMPOL 28 +5V 37 8A
14A 40 XcmReady 3K3 AGND 120 CAPT PWMSR 22 +5V 10K 38 8B
41 XpsReady R114 121 VIN4 R41 10K 39 8C
14B 10K D19 D18 100nF C28 C16 R1322
14C 42 XshReady 122 BSHAN GND 46 40 8D
43 C152 2xBZX55C4V7 123 ASHAN
15A Xgi1Ready
1nF
PTC2 PTC1 VDD 45 C71 41 9Z
15B 44 Xgi2Ready C965 C10 100nF 10uF R9 22 124 VIN0 +5V DGND 42 9A
45 XdiReady C965 125 44 43 9B
15C
46 Sd0 DGND AGND CAPB
126 VIN1
SCLK0 10nF DspDisp
16A 100nF C26 TFS0 43 44 9C
16B 47 Sd1 PGND DGND +5V AGND 127 CML RFS0 42 R56 45 9D
16C 48 Sc IC23 100nF 128 VIN2 DT0 41 DGND 46 10Z
17A 49 EN1 1 129 GAIN DR0 40 10K 47 10A
50 +12V 130 VIN3
17B +12VP
+24Vin C21 C27 SCLK1 39 TfsSync
48 10B
17C 51 +12VP GND 13 PGND 131 SENSE TFS1/IRQ1 38 49 10C
18A 52 +12VP PrWaValve 3 OUT1 C70 IC6A 132 AVSS RFS1/IRQ0/SROM 37 Rfs1 50 10D
53 +12VP 12 100nF 100nF 133 36 DspDt1 51 11Z
MF5 R37 OP495

4
18B GND +24VIntOn AVDD DT1/F0 PROMorEEPROM
54 +12VP XpsIntOut C31 35 DspDr1B 52 11A

PWDACK
18C DR1B/FI

CLKOUT
AGND 10nF DspSclk1

BMODE
55 4 + 3 AGND 34 1 DATA 2 53 11B

RESET
19A GND 10K DR1A/F1 CLK

MMAP

CLKIN
56 WaterPump 6 OUT2 MF-R110 Vref+2V 1 C148 C149 C150 C147 6 CEO RST/OE 3 54 11C

XTAL
PWD
GND

GND

GND

GND
GND
GND

GND
DMS

BGH

POR
BMS
PMS

VDD

VDD

VDD
19B C54

A13
A12
A11
A10
RN2A 100nF

WR
57 100nF 100nF R52 /DspPCe
GND 5 - 2 CE 4 55 11D

BG
RD

NC
NC
Pressostat 100nF 100nF

BR

A9
A8
A7
A6
A5
A4
A3
A2
A1
A0
19C 2K7 100nF
20A 58 PGND Via XGT C36 C20 10K VPP 7 56 12Z
59 PrCoolCtrl
IN1 2 R42 5 GND VCC 8 57 12A

11

134
135
136
138
137
140
139
15
57
141
142
143
144
1
2
3
4
6
7
8
10
11
12

5
9

13
14
16
17

18
19

20
21

23
24

25
26
27

29

30

31

32
33
20B WaterCooler Pressostat/24V Int1 C43 R21 +5V
20C 60 11 OUT3 DGND DGND 58 12B
61 7 WaPump 10nF AGND AGND AGND AGND 100uF IC13 37LV65 1K 59 12C
21A
62 -12VP
IN2 10K AGND 10nF C44 C56 C64 DGND 60 12D
21B RN2B C58 orXC1765D
63 -12VP 10 WaCooler -12V R121 R122 R123 R120 AGND +5VM DGND R46 orAT17C65 C52 61 13Z
12.96MHz

FROM INSTRUMENT INTERLOCK


21C IN3 2K7
64 -12VP Buzzer 14 OUT4 100nF R24 100 100 100 100 R29 DGND 62 13A
22A
65 BuzzerOn 10nF 10nF 10nF 10nF
22B -12VP IN4 15 Intlk2
R47 Int2
AGND
FilCurMon +5V R32 1K Y1 63 13B
22C 66 -12VP C94 C102 C98 C96 DGND DGND 10K 0.266V/A +5V 64 13C
67 L10 10K C91 C90
23A +24Vin EN2 9 +5V 10K RN2C +5V DGND 10K +5V DGND +5V DGND DGND 65 13D
23B 68 1nF 1nF 1nF 1nF Mon400V
3.3V/400V 66 14Z
69 8 2K7 C59 20pF 20pF 67 14A
23C Not connected VS +24V C17 InputVoltage
R25 R36 /DspRst
24A 70 GND GND GND GND 100nF +12V 10pF 3.95V/400V +5V +5V 68 14B
71 Vref+5V
24B VSS 16 +5V Intlk3
R50 Int3 InputCurrent 10K 10K 69 14C
24C 72 DGND DGND R81 0.1V/A DGND 70 14D
73 L9 R34 L293D D6 71 15Z
25A +12Vin IC6B C117 10K 1N3595
25B 74 10K/0.1% 6 -
RN2D 100K IC18C 72 15A
75 7 Vref+2V5 10nF C63 C92 R86 HvArcTh 9 73 15B
25C
76 5 + 2K7 100nF HighVoltage 13 - 16 /HvArc 74 15C
26A OP495 R74
26B 77 PGND R53 10K 14 10 +5VM 75 15D
78 L8 AGND R38 +5V Intlk4 DGND DGND Int4 1uF 12 + 76 16Z
26C -12Vin C51 D9 IC19D 10K
27A 79 10K/0.1% BZ1 +5V 10K D4 CMP401 77 16A
80 1nF 3 1 RN2E 1N3595 AGND LF347N BZX55C5V1 R33 78 16B
27B L7 C66
27C 81 2K7 10K 79 16C
82 +5Vin AGND AGND +5V 100nF XpscTemp 80 16D
28A
28B 83 R57 AGND 81 17Z
84 4 2 R69 Intlk5 DGND DGND Int5 AGND C55 82 17A
28C C29 RV1
29A 85 4K7 10K +12V JTAG 100nF 83 17B
86 R113 R108 R111 R116 DGND BUZZER 10pF 84 17C
29B RN2F C68 4K7/25
29C 87 4K7 4K7 4K7 4K7 R96 J1 +5V +12V +24V 85 17D
88 DGND 2K7 100nF D14 AGND AGND 86 18Z
30A PrWaTIn
89 R62 1N3595 IC18D +5V +5V 87 18A

8
7
6
5
4
3
2
1
30B
90 DGND PrWaTOut IntlkRtn DGND DGND IntRet C105 R92 100K CuArcTh 11 R28 R35 R40 88 18B
30C WaterTIn EmissionCurrent- /CuArc
31A 91 10K RN2G 9 - R103 15 R1544K7 2K7 18K 47K 89 18C
92 WaterTOut 8 12 +5VM 90 18D
31B
93 2K7 C73 1uF 10K 10 + 91 19Z
31C
94 C136 C124 C130 C142 100nF D12 IC19C 10K CMP401 4K7 AD status converter R39 R45 92 19A
32A D16 R31 C159 C160 C161
95 1N3595 AGND LF347N BZX55C5V1 C38 Low voltage 8x8bit 10K 10K 10K 100nF 93 19B
100nF

Tms
R155 100nF

Tclk
32B
96 GND 1uF 1uF 1uF 1uF DGND DGND 94 19C

Tdo
Tdi
32C
DGND 10nF IC8 95 19D
DIN41612 AGND 20 VCC CH0 1 AGND AGND AGND AGND AGND AGND 96 20Z
AGND TP8 AGND 19 V+ 2 97 20A
FilamentCurrentRef /AdcCs2 CH1
+12V AGND 3X 1N4148 18 CS CH2 3 98 20B
Vref+5V DGND 0.266V/A SdataOut 17 DI 4 99 20C
Sclk CH3
16 CLK CH4 5 -12V 100 20D
FilCurRef AdcSars 15 SARS
IC6C D10 D8 D7 CH5 6 101 21Z
2

IC25 9 R90 Vref+5V +12V SdataIn 14 DO HvArcTh


C157 CH6 7 102 21A
-
REF02 8 10K IC4 C80 CuArcTh
10nF Tdo C110 C74 10nF Vref+5V R60 +5V 13 SE CH7 8 R44 103 21B
6 10 +
TDO 35 Tdi 12 VREF COM 9 33K 104 21C
OP495 WaLevel IC2C ClkOut 11 10 11 10 105 21D
AGND R43 DEIONIZED WATER LEVEL +5V 3 4
CLK0/I0 TDI
13 Tclk 1nF 10nF AGND 10 AGND DGND
-5V 106 22Z
C62 SW1 /Wr TCLK Tms R48 C162

11
10K R101 D15 R100 R12 5 6 1 2 43 I/O TMS 32 AGND AGND C82 C77 ADC0838 107 22A

1
4

100nF IC16 R66 AGND DGND AGND 10K 100nF 108 22B
AD15 A7 IC15 10uF 1nF R51
W1 2 I/O I/O 14 A6 1K 109 22C

VDD
VREF
1K 1N4148 C118 18K R93 10K 74HC14 AD14 A0 AD0 AD7 FilCurRefMon
AGND R16 3 I/O I/O 15 /DataCs 12 A0 O0 13 AD1 13 DB7 VOUT8 2 AGND 15K 110 22D
100uF C40 AD13 4 I/O A1 AD6
C115 100K C13
AD12 I/O 39 /CodeCs A2
11 A1 O1 14 AD2 AD5
14 DB6 VOUT7 3 AGND AGND 111 23Z
AGND R95 1uF 67 XTAL1 P1.0/EPA0/T2CLK 57 5 I/O I/O 38 /WrDac0 10 A2 O2 15 AD3 15 DB5 VOUT6 4 112 23A
AGND 100uF 15K 10pF X1 56 /HvArc AD11 6 37 A3 9 17 AD4 16 5 R55 C163 113 23B
P1.1/EPA1 /CuArc AD10 I/O I/O /PPI2Cs A4 A3 O3 AD4 AD3 DB4 VOUT5 HvArcTh 100nF
10K DGND DGND 16MHz P1.2/EPA2/T2DIR 55 7 I/O I/O 36 8 A4 O4 18 17 DB3
FF40H VOUT4 6 10K 114 23C
R91 AGND AGND C41 66 XTAL2 54 /MaxFilCur AD9 8 I/O 17 4MHz A5 7 A5 19 AD5 AD2 18 DB2 - 7 CuArcTh +5VM Vref+5V 115 23D
C100 R94 LF347N P1.3/EPA3 FilCurSD AD8 I/O O5 VOUT3
+ 5 10pF P1.4/EPA4 53 9 I/O I/O 31 /Ppi1Cs A6 6 A6 O6 20 AD6 AD1 19 DB1 FF48H VOUT2 8 MaxFilCur AGND 116 24Z
WaResist 7 WdToggle /Rd A7 AD7 AD0 SafFilCur
10K R84 P1.5/EPA5 52 /AdcDor0 /Ale I/O 30 Ready AD8
5 A7 O7 21 20 DB0 VOUT1 9 Miscellaneous 8x8bit 117 24A
5M6 IC19B - 6 28 NMI P1.6/EPA6 51 AdcEoc0
16 I/O I/O 29 ClkOut AD9
27 A8 C37 AD converter 118 24B
1uF AGND P1.7/EPA7 50 CLK1/I1 33 26 A9 Miscellaneous 119 24C

GND
10nF

VSS
-WR
C112 /Reset DGND 0 27 AD10 23 8x12bit IC7 120 24D

A2
A1
A0
R102 RESET AD0 AD11 A10 MaxFilCur
AGND P3.0/AD0 26 M4A5-64/32 XPSC01L 25 A11 DA converter 20 VCC CH0 1 121 25Z
8x100 25 AD1 AD12 4 A12 AD7228 AGND 19 V+ 2 SafFilCur 122 25A

21
22
23
24
12
10
R88 +12V 1M P3.1/AD1 AD2 AD13 /WrDac0 /AdcCs1 CH1
220nF PrimaryWaterTemperatureInput R140 40 P0.0/ACH0 P3.2/AD2 24 28 A13 18 CS CH2 3 Lamp1 123 25B
C116 PrimaryWaterTemperatureOutput R141 41 P0.1/ACH1 AD3 AD14 A2 SdataOut 4 Lamp2
1M P3.3/AD3 23 AD4 IC17 AD15
29 A14
A1 -5V Sclk
17 DI CH3 124 25C
AGND DeionizedWaterTemperatureInput R142 42 P0.2/ACH2 P3.4/AD4 22 3 A15 16 CLK CH4 5 FilCurMon 125 25D
R143 43 P0.3/ACH3 AD5 AD0 A0 A0 C106 AdcSars XsnR1+
DeionizedWaterTemperatureOutput P3.5/AD5 21 3 D1 Q1 2 DGND 2 A16 AGND 15 SARS CH5 6 FilCurRefMon 126 26Z
4

100nF DEIONIZED WATER RESISTIVITY DeionizedWaterLevel R144 44 P0.4/ACH4 AD6 AD1 A1 SdataIn XsnR1-
P3.6/AD6 20 AD7 AD2
4 D2 Q2 5 A2 /CodeCs
1nF 14 DO CH6 7 127 26A
3 + R72 D5 R73 WaterResistivityDriver R145 45 P0.5/ACH5 P3.7/AD7 19 7 D3 Q3 6 22 CE 0000H R59 +5V 13 SE CH7 8 128 26B
1 DeionizedWaterResistivity R146 46 P0.6/ACH6 AD3 8 D4 9 A3 /Rd 24 - AGND Vref+5V 12 9 XsnT1+ 129 26C
Vref+2V5 AD8 AD4 Q4 OE VREF COM
2 - 15K R147 47 P0.7/ACH7 P4.0/AD8 18 13 D5 Q5 12 A4 1 VPP DFFFH 10 11 AGND DGND 10 XsnT1- 130 26D
M8 IC19A 1K 1N4148 R79 17 AD9 AD5 14 D6 15 A5 31 PGM 131 27Z
PrWaTIn LF347N C97 C158 XsnTxD P4.1/AD9 AD10 AD6 Q6 A6 +5V
1 15K 32 P2.0/TXD P4.2/AD10 16 17 D7 Q7 16 A7 ADC0838 +5V 132 27A
11

2 PrWaTOut 1uF XsnRxD 33 P2.1/RXD AD11 AD7 C81 XpscDisplay


WaterTIn
1000uF P4.3/AD11 15 AD12
18 D8 Q8 19 ROM128K AGND DGND AGND
DisplayRd2
133 27B
3 C87 AdcSars 34 P2.2/EXINT P4.4/AD12 14 C76 134 27C
4 WaterTOut XpsReady 35 P2.3/BREQ AD13 /ALE IC14
AGND AGND AGND P4.5/AD13 13 AD14
11 CLK
A0 AD0
10uF 1nF DGND 135 27D
5 AGND 100nF -12V R80 /AdcConv0 36 P2.4/INTOUT P4.6/AD14 12 1 CLR 12 A0 D0 13 +5VM 136 28Z R124
6 WaLevel +5V XpscOk 37 P2.5/HOLD 11 AD15 IC2D A1 11 14 AD1 AGND AGND 137 28A
PGND XpscNotOk P4.7/AD15 A2 A1 D1 AD2 R125
7 AGND 38 P2.6/HLDA Ale 9 8 +5V 74HC273 10 A2 D2 15 138 28B R126
8 WaResist 1K ClkOut 39 P2.7/CLKOUT 4 A3 9 17 AD3 Miscellaneous 8x12bit 139
PGND R85 P5.0/ALE/ADV EnWd A4 A3 D3 AD4 C14 FilamentCurrent 28C R127
9 P5.1/INST 3 74HC14 8 A4 D4 18 AD converter 140 28D R128
10 PrWaValve 100K R105 R99 AGND 9 /Wr A5 7 19 AD5 1nF /AdcCs3Ext 141
P5.2/WR/WRL /Rd A6 A5 D5 IC3 R18 100 EmissionCurrent- 29Z R129
11 /PrWaFlow 4K7 4K7 P5.3/RD 7 R117 6 A6 D6 20 AD6 3.5V/140mA Ik /AdcCs2Ext 142 29A R130
12 /DeWaFlow 1 A7 5 21 AD7 AGND 28 1 /AdcCs1Ext 143
Intlk2 /PrWaFlow C69 P5.4/SLPINT SsioTxSync +5V AD8 A7 D7 /AdcDor0 VD+ CH0 Ia R23 100 EmissionCurrent+ ExSdataOut 29B R131
13 AGND +5V P5.5/BHE/WRH 8 +5V 27 A8 27 DOR CH1 2 3.5V/140mA 144 29C
14 Intlk3 /DeWaFlow 1nF 2 Ready 1K IC12 AD9 26 4MHz 26 3 ExSclk 145
Intlk4 P5.6/READY EepromCs AD10 A9 Sclk CCLK CH2 Hv R30 100 HighVoltage AdcSars 29D
15 49 VREF P5.7/BUSWIDTH 10 EepromCs 1 CS VCC 8 23 A10 25 SCLK CH3 4 4V/75KV 146 30Z R132
16 Intlk5 +5V C109 C103 6 VPP Sclk 2 SK 7 C50 AD11 25 SdataOut 24 5 SdataIn 147
IntlkRtn SdataOut 3 NC AD12 A11 SdataIn DI CH4
C42 C19 C30 30A R133
17 100nF 100nF +5V +5V 48 ANGND P6.0/EPA8 58 /XsnTxDEn 6 10nF 4 23 6 148 R134
18 WaterCoolOn 29 EA 59 /AdcCs0 SdataIn 4 DI NC
5 AD13 28
A12 /AdcCs0 22
DO CH5
7 XpscTemp 149
30B
PGND R71 P6.1/EPA9 /AdcCs1 DO GND AD14 A13 /AdcConv0 CS CH6 30C R135
19 P6.2/T1CLK 60 3 A14 21 CONV CH7 8 1uF 1uF 1uF 150 30D R149
20 WaterPump +24V 10K DGND R26 R78 AGND 61 /AdcCs2 EEPROM AD15 31 A15 AdcEoc0 20 EOC 9 EmissionCurrent+ 151 31Z
/WaCoOn P6.3/T1DIR COM HighVoltage
21 WaterCooler 1M 10K P6.4/SC0 62 93C66 DGND DGND 2 A16 DGND 19 PD MUXOUT1 10 13x 100 Ohm 152 31A
22 Pressostat/24V SdataIn EmissionCurrent-
AGND P6.5/SD0 63 Sclk or /DataCs Vref+5V AGND 18 AGND A/DIN1 11 AGND
Vref+5V
153 31B
23 IC22 P6.6/SC1 64 +5V 22 CE1 E000H 17 VREF+ MUXOUT2 12 154 31C
24 C35 R77 R63 SdataOut IC12a Vref+2V
PC817 P6.7/SD1 65 EepromCs +5V /Wr 30 CE2 - R58
10
16 VREF- A/DIN2 13 InputCurrent
155 31D
R76 0 0 R27 1 CS VCC 8 29 FF3FH 15 14 156
Sclk 2 /Rd 24 WE VA+ DGND InputVoltage 32Z
1uF IC9 80C196KR 180 SdataOut 3 SK NC 7 OE Mon400V
157 32A
DI NC 6 C67 C75 ADC12138 158 32B
4K7 DGND DGND DGND ClkOut SdataIn 4 DGND AGND
/AdcConv0 DO GND 5 RAM128K 10uF 10nF 159 32C
C78 160 32D
+5V AdcSars EEPROM DGND 10nF IC26
/XsnTxDEn AGND AGND DspWdToggle FilPwmH 2 A1
J407
Buzzer
93C66/SO
FilPwmL Y1 18
+5VM AGND 4 A2 Y2 16
3

1 XrayLamp1 WdToggle HvPwmH


2 6 A3 Y3 14
GND DIN41612
XrayLampC HvPwmL 8 A4
3 XrayLamp2 XsnT1+ IC21A CurrentControl Y4 12
4 16 +5V +5V +5V IC27A 11 A5 Y5 9
1 XsnRxD CurrentLimit 13 A6
5 R83 15 IC10 IC11 PowerOn Y6 7
6 +24V PGND LD1 1 A Q 13 15 A7 Y7 5
R3 AD0 34 D0 4 Int1 AD0 34 D0 4 R150 2 B 4 17 A8
1K IC2E Y8 3
WAGO
7 C169
1nF
XsnT1- 4 DGND XpscOk 11 10
330 AD1
AD2
33 D1
32 D2
PPI 1 PA0
PA1 3
2
Int2
Int3
AD1
AD2
33 D1
32
PPI 2 PA0
PA1 3
2
220K C166
2.2uF 14
Q
R106
10K 1
R152
10K
AD3 PA2 Int4 AD3 D2 PA2 C 1OE
14 31 D3 1 31 1 +5V 15 +5V 19
PGND R54
ExRelayPresent 13
2
+5V
XpscOk/Green AD4
AD5
30 D4
29 D5
I PA3
PA4 40
39
Int5
IntRet
AD4
AD5
30 D4
29 D5
D3
I PA3
PA4 40
39
+5V +5V
3 CLR
R/C
9 A
IC27B
5 DGND
2OE
DGND
74HC14 PA5 PA5 R151 Q SN74ACT244
AD6 28 D6 +24VIntOn AD6
C170 +5V PA6 38 28 D6 PA6 38 10 B Q 12

1
2
3
4
22K R65 C11 8922A AD7 27 D7 37 /WaCoOn AD7 27 37 22K C167
MM74HC423A

2
4
6
8
1nF R118 PA7 D7 PA7 10nF
7K5 100nF T=100ms 6 C
6

1K LD2 R4 /Rd 5 18 ExRelayPresent /Rd 5 18 PowerOn 7 R/C


IC21B IC2F RD PB0 RD PB0 S1
PGND 9 /Wr 36 WR /Wr EnPower W2
XpscNotOk
330
A0 PB1 19 A0
36 WR PB1 19 IC24 EnPower
DGND DGND 8 DGND 13 12 9 A0 PB2 20 9 A0 PB2 20 11 CLR
10 A1 8 A1 A1 RN1 Sclk DspSclk1
PB3 21 8 A1 PB3 21 2 A1 Y1 18
R153
I O

8
7
6
5

1
3
5
7
ExternalRelay 35 RESET /PrWaFlow SdataOut DspDr1B
Q4
/XsnTxDEn
XpscNotOk/Red
/Ppi1Cs PB4 22 /DeWaFlow /Ppi2Cs
35 RESET PB4 22 /DspPCe
10
RfsSync
4 A2 Y2 16 Rfs1
MM74HC423A
MPSA05 2K2 5 74HC14 6 CS PB5 23 6 CS PB5 23 9 6 A3 Y3 14 T=100us
24 Reset 24 EnDisRd2 8 DspDt1 8 12 Sdatain
XsnR1+ 11 PB6 PB6 EnXpscDis A4 Y4 ExSclk
PB7 25 PB7 25 7 +5V 11 A5 Y5 9
7 6 13 A6 ExSdataOut
PGND
XsnR1- 12
Address
BuzzerOn /DspCs Y6 7
Ppi1Cs FFC0...FFFF FFC0H PC0 14 FF80H PC0 14 5 15 A7 Y7 5
PrCoolCtrl /AdcCs1Ext R119
Ppi2Cs FF80...FFBF - PC1 15 - PC1 15 4 17 A8 Y8 3
8922A WrDac0 FF40...FF7F FFFFH O PC2 16
XrayLampOn
/DspRst
FFBFH O PC2 16
/AdcCs2Ext
/AdcCs3Ext
3
/DspCs
10K
IC1B
DataCs E000...FF3F PC3 17 ExternalRelay PC3 17 2 1 1OE
EnDisRd2 IC1C
CodeCs 0000...DFFF PC4 13 WaPump PC4 13 TfsSync
1 19 2OE
5 12 12
DisplayRd2 R19
6
4 8
10 O PC5
PC6 11
10
WaCooler
/EnFilCu
O PC5
PC6 11
10
SsioRxSync
RfsSync
9*10K
DGND DGND
SN74ACT244 Mod: 1 Version pré-série 07.08.01 PPa 4 Version série 10.03.04 PPa 7 MOD23765 03.03.05 PPa
C34 100 9 XsnTxD PC7 PC7
1nF 2 Version pré-série (2) 13.11.01 PPa 5 OM3753 17.12.03 PPa 8 OM4012 04.05.07 PPa
7403 82C55 82C55
+5V 7403 Reset 3 Version "Phase-in" 09.09.02 PPa 6 OM3834 28.10.04 PPa 9 OM4049 10.04.08 PPa
R17
DGND 10K
Dessiné 15.11.00 PPa Remplacé par:
R20
10K
IC1D
DGND
+5V
TP5 XPSC, X-ray Power Supply Control Board Echelle

EnXpscDis R11 L4 +12V


R14 13 MF3 Contrôlé
XpscDisplay 11 10K R22 TP2 TP7 +12Vin Remplace:
DspDisp +12V
C23 12 +5V +5VM +24V
100 BZX85C5V1 L5 MF1 L2 +5V L1 MF4 L6 MF-R110 C33 C22 TP4 Vu
1nF 470 +5Vin +24Vin Liste S703124 Schémas S112065
7403 +5V AGND -5V +24V AGND
R8 C2 R1 220uF 100nF Ensemble ---------- Circuit S920076 Groupe Technique:
DGND D20 C39 MF-R250 C5 C1 MF-R250 C53 C46 TP6 XRF Ingénieur M.Bonzon
10K R138 R139 R137 1uF AGND Appareil Implantation S112071 9850
C79 C49 C89 C4 C3 C60 C47 C25 C61 C48 C146 C164 C165 1uF TP1 PGND 0
120 120 120 220uF 100nF 220uF 100nF GND C18 C12 TP3
DGND R2 Groupe No
DGND
DGND
-12Vin
AGND AGND
0
PGND
-12Vin
MF2 L3 220uF 100nF
-12V
ThermoFisher S112065
No Mod.

DECOUPLING CAPACITORS = 10nF


DGND GND
MF-R110
-12V S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 9
J904
+5V

2
4
6
R89 FilamentCurrent
0.266V/A W3 1 1Z
R10 IC1A EnWd Test 2 1A
D3 1K C113 15K R87 FilCurMon GND
1 IC2A 3 1B

1
3
5
/GenxReset 3 D1 C24 R15 IC18B +5V 4 1C
2 2 1 WdToggle 47nF 8 MaxFilCur 1K C84 LD3 +24Vin 5 1D
BAT85 R104 R7
1 R70 100uF 6 2Z
7403 BAT85 1uF 1K +12V 7 Pressostat/24V 7 2A
R5 D2 470K 35V 390 PGND
74HC14 R6 C6 DspOk 8 2B
1K 1M 1N4148 IC20 CMP401 1K AGND /DspCs Green +12Vin 9 2C
470nF
C111 C107 10 2D
1 -EA 16 C88 10nF 10nF CurrentLimit DGND +5V SsioRxSync 11
VREF5.1V -12Vin AGND 3Z
DGND DGND DGND DGND 12
/Reset FilCurRef R75 R82
2 +EA 15 10nF CurrentControl C57 SsioTxSync 13
3A
VIN +5V DGND 3B
10K 47K 3 +5V AGND AGND 14 3C
C8 FilPwmH DspWdToggle +5Vin
SYNC OUTPUT B 14 AGND
10nF C65
15 3D
47nF +5V IC2B C95 +5V +5V 16 4Z
3 4 Reset 10nF 4 OSC OUT DGND
VC 13 C93 C85 /PwmTrip
17 4A
DGND AGND 18 4B
R115 5 CT 10nF 10nF
74HC14 GND 12 10nF 19 4C

4
3K3 AGND 20 4D

108
107
106
105
104
103
102
101

100
GenxReady XpsReady FilPwmL

98
91
90

99
97
96
95
94
93
92
89
88
87
86

85
84
83
81
80
79
78
77
76
75
74
72
70
69
68
67
66
65
64
62
61
60
59
58

71
82
73
63
6 RT OUTPUT A 11 AGND IC18A AGND R67 21 5Z
+24V R148 D13 6 HvPwmH 22 5A
7 DISCH. +5V DGND
SHUTDOWN 10 2 23 5B

EIA
EIB
EIZ
EIS

ETU0
ETU1

AUX0
AUX1

GND
GND

VDD

P0

P1

P2

P3

P4

P5

P6

P7

P8

P9

P10

P11

D0
D1
D2
D3
D4
D5
D6
D7
D8
D9
D10
D11
D12
D13
D14
D15
D16
D17
D18
D19
D20
D21
D22
D23

VDD

GND
GND
GND
XrayLampC Q2 SafFilCur 10K HvPwmL
C152 5 24 5C
1nF 180 8 9 1N3595 D17 C45 56 C151 25 5D
R107 SOFT-START COMP. C99 AH R68 FilPwmH
IRF9Z24 CMP401 C104 AL 55 0 26 6Z
1N3595 10nF 109 CONVST
DGND 10K C108 1uF BH 54 Not mounted 27 6A

14

13
R112 110 GND R64 0 FilPwmL
R97 C101
C83
SG3525AN 10nF +5VM
10nF BL 53 10nF 28 6B
R110 20K 1nF AGND AGND 111 VDD IC5 Not mounted DGND 29 6C
10K Q3 XrayLampOn 100uF 112 GND 52 30 6D
GND
MPSA05 10K AGND AGND 113 AVDD ADMC401 VDD 51 C72 R136 31 7Z
AGND AGND AGND R98 /MaxFilCur C114 114 AVSS 32 7A
10nF AGND +5V DGND 100
10K D11 115 VIN7 CH 50 33 7B
1N3595 116 VREF 10nF
AGND
FilCurSD C7 Hv CL 49 34 7C
-5V AGND C15 117 48 35 7D
XrayLamp1 R114
R109 10K Lamp1 AGND
VIN6
118 REFCOM
PWMSYNC
47 R6110K 36 8Z
XrayLamp2 Lamp2 /EnFilCu PWMTRIP +5V R49
C9 AGND 10uF 100nF 119 VIN5 PWMPOL 28 +5V 37 8A
AGND 120 CAPT PWMSR 22 +5V 10K 38 8B
10K D19 D18 R1322 121 VIN4 R4110K 39 8C
100nF C28 C16 122 BSHAN 46 40 8D
GND
PTC2 PTC1 2xBZX55C4V7 123 ASHAN VDD 45 C71 41 9Z
C965 C965 100nF 10uF R9 22 124 VIN0 +5V DGND 42 9A
C10 125 CAPB
AGND SCLK0 44 10nF DspDisp
43 9B
100nF C26 126 VIN1 TFS0 43 44 9C
PGND DGND +5V AGND 127 CML RFS0 42 R56 45 9D
IC23 100nF 128 VIN2 DT0 41 DGND 46 10Z
EN1 1 129 GAIN DR0 40 10K 47 10A
130 VIN3 SCLK1 39 48 10B
+24Vin +12V C21 C27 TfsSync
PrWaValve GND 13 PGND 131 SENSE TFS1/IRQ1 38 Rfs1
49 10C
3 OUT1 132 AVSS RFS1/IRQ0/SROM 37 50 10D
12 MF5 R37 C70 IC6A 100nF 100nF 133 36 DspDt1 51 11Z

4
GND XpsIntOut +24VIntOn AVDD DT1/F0 DspDr1B PROMorEEPROM
C31 DR1B/FI 35 52 11A

PWDACK

CLKOUT
10nF DspSclk1
GND 4 3 DR1A/F1 34 1 DATA CLK 2 53 11B

BMODE
AGND + AGND

RESET
10K

MMAP

CLKIN
WaterPump 6 OUT2 MF-R110 Vref+2V 1 C148 C149 C150 C147 6 CEO RST/OE 3 54 11C

XTAL
PWD
DMS

GND

GND

GND

GND
GND
GND

GND
BMS
PMS

BGH

POR
VDD

VDD

VDD
C54

A13
A12
A11
A10
RN2A 100nF

WR
5 2 R52 4 /DspPCe 55 11D

RD

BG

NC
NC
Environment 100nF 100nF 100nF 100nF

BR

A9
A8
A7
A6
A5
A4
A3
A2
A1
A0
-
GND 2K7 100nF CE
Coherency OP495 C20 10K VPP 7 56 12Z
PrCoolCtrl
IN1 2 R42 C36 5 GND VCC 8 57 12A

11

134
135
136
138
137
140
139
15
57
141
142
143
144
1
2
3
4
6
7
8
10
11
12

5
9

13
14
16
17

18
19

20
21

23
24

25
26
27

29

30

31

32
33
WaterCooler 11 OUT3 Pressostat/24V DGND DGND Int1 C43 R21 +5V 58 12B
WaPump 10nF 100uF
IN2 7 10K AGND
10nF
AGND AGND AGND AGND IC13 1K 59 12C
R24 -12V C44 C56 C64 DGND 60 12D
10 WaCooler RN2B C58 R121 R122 R123 R120 AGND +5VM DGND AT17LV256-10PU 61 13Z
R46 12.96MHz C52

FROM INSTRUMENT INTERLOCK


Buzzer IN3 2K7
14 OUT4 100nF 100 100 100 100 R29 DGND 62 13A
BuzzerOn 10K 10nF 10nF 10nF 10nF
IN4 15 Intlk2 R47 Int2
AGND
FilCurMon +5V R32 1K Y1 63 13B
C94 C102 C98 C96 DGND DGND 0.266V/A +5V 64 13C
9 10K +5V DGND +5V DGND C91 C90 +5V DGND DGND 65 13D
EN2 +5V 10K RN2C Mon400V 10K
1nF 1nF 1nF 1nF 3.3V/400V +5VM 66 14Z
2K7 C59 20pF 20pF
VS 8 +24V InputVoltage R25 R36 /DspRst
67 14A
GND GND GND GND 100nF 3.95V/400V +5V +5V 68 14B
R33
VSS 16 +5V Intlk3 R50 Int3 InputCurrent 0.1V/A 10K 10K 69 14C
DGND DGND 10K DGND 70 14D
IC6B L293D XpscTemp 71 15Z
C117 10K
R34 6 - RN2D 72 15A
7 Vref+2V5 10nF C63 +12V C17 C55 73 15B
2K7 RV1
5 + 100nF 100nF 74 15C
OP495 PGND R53 R81 4K7/25 75 15D
10K/0.1% R38 +5V Intlk4 DGND DGND Int4 D6 10pF 76 16Z
C51 BZ1 +5V 1N3595 IC18C AGND AGND 77 16A
10K/0.1% 10K 100K
1nF 3 1 RN2E C92 R86 HvArcTh 9 78 16B
2K7 C66 HighVoltage 13 - 16 /HvArc 79 16C
100nF R74
AGND AGND +5V 10K 14 10 80 16D
R57 1uF 12 + JTAG 81 17Z
4 2 R69 Intlk5 DGND DGND Int5 D9 IC19D 10K D4 CMP401 82 17A
4K7 1N3595 AGND LF347N BZX55C5V1 J1 83 17B
R113 R108 R111 R116 DGND 10K 84 17C
BUZZER RN2F
4K7 4K7 4K7 4K7 C68 +5V +5V +5V +12V +24V 85 17D

8
7
6
5
4
3
2
1
DGND 2K7 100nF AGND 86 18Z
PrWaTIn R62 AGND R1544K7 87 18A
PrWaTOut IntlkRtn DGND DGND IntRet +12V C29 10pF R28 R35 R40 88 18B
WaterTIn 2K7 18K 47K 89 18C
WaterTOut 10K RN2G 4K7 +5VM 90 18D
C73 D14 R96
2K7 91 19Z

Tms
R155

Tclk
Tdo
C136 C124 C130 C142 100nF 1N3595 IC18D AD status converter C159 R39 C160 R45 C161 92 19A

Tdi
100K R31
+5V 3X C105 R92 CuArcTh 11 DGND Low voltage 8x8bit 10K 100nF 10K 100nF 10K 100nF 93 19B
1uF 1uF 1uF 1uF 1N4148 DGND DGND EmissionCurrent- 9 - R103 15 /CuArc C38 94 19C
8 12 10nF IC8 95 19D
1uF 10K 10 + 10K 20 VCC CH0 1 AGND AGND AGND AGND AGND AGND 96 20Z
D12 IC19C TP8 19 V+
AGND
R90
D10 D8 D7
LF347N
D16 CMP401 DGND
FilamentCurrentRef /AdcCs2
AGND CH1 2 97 20A
+12V 10K 1N3595 AGND BZX55C5V1 18 CS CH2 3 98 20B
Vref+5V SdataOut 17 DI
0.266V/A
Sclk CH3 4 99 20C
16 CLK CH4 5 -12V 100 20D
WaLevel AGND FilCurRef AdcSars 15 SARS 6 101
2

IC25 Vref+5V SdataIn CH5 HvArcTh 21Z


C157 9 - IC6C C178 AGND +12V 14 DO CH6 7 102 21A
REF02 8 C80 13 8 CuArcTh R44 103
10nF C110 C74 10nF Vref+5V R60 +5V SE CH7 21B
6 10 + 1uF 12 VREF COM 9 33K 104 21C
OP495 IC2C 10 11 AGND DGND 10 105 21D
AGND R43 DEIONIZED WATER LEVEL AGND +5V 3 4 1nF 10nF AGND C162 -5V 106 22Z
C62 SW1 Tdo R48

11
10K R101 D15 R100 R12 5 6 1 2 TDO 35 Tdi AGND AGND C82 C77 ADC0838 100nF 107 22A

1
4

100nF ClkOut 11 CLK0/I0 IC16 R66 10K


TDI 10 Tclk IC15 10uF 1nF
AGND DGND AGND 108 22B
W1 TCLK 13 Tms 1K R51 109 22C

VREF

VDD
1K 1N4148 C118 18K 10K 74HC14 /Wr A0 AD0 AD7 13 FilCurRefMon
AGND R16 C40 43 I/O TMS 32 12 A0 O0 13 2 AGND 15K 110 22D
100uF R93 100K C13 A1 11 A1 14 AD1 AD6 14 DB7 VOUT8
3 AGND AGND 111 23Z
C115 AD15 2 A7 A2 O1 AD2 AD5 15 DB6 VOUT7
AGND R95 15K 1uF 10pF 67 XTAL1 P1.0/EPA0/T2CLK 57 I/O I/O 14 A6 10 A2 O2 15 AD3 DB5 VOUT6 4 C163 112 23A
100uF X1 /HvArc AD14 A3 AD4 R55
AGND P1.1/EPA1 56 /CuArc AD13
3 I/O I/O 15 /DataCs A4
9 A3 O3 17 AD4 AD3
16 DB4 VOUT5 5 HvArcTh
100nF 113 23B
DGND DGND C41 16MHz P1.2/EPA2/T2DIR 55 4 I/O I/O 39 /CodeCs 8 A4 O4 18 AD5 17 DB3 FF40H VOUT4 6 10K 114 23C
10K AGND AGND 10pF 66 54 /MaxFilCur AD12 5 38 A5 7 19 AD2 18 - 7 CuArcTh +5VM Vref+5V 115 23D
C100 R94 LF347N R91 XTAL2 P1.3/EPA3 FilCurSD AD11 6 I/O I/O /WrDac0 A6 A5 O5 AD6 AD1 19 DB2 VOUT3 MaxFilCur
+ 5 P1.4/EPA4 53 I/O I/O 37 6 A6 O6 20 DB1 FF48H VOUT2 8 AGND 116 24Z
WaResist 7 WdToggle AD10 7 /PPI2Cs A7 AD7 AD0 20 SafFilCur
10K R84 P1.5/EPA5 52 /AdcDor0 AD9 I/O I/O 36 4MHz AD8
5 A7 O7 21 DB0 VOUT1 9 Miscellaneous 8x8bit 117 24A
5M6 IC19B - 6 28 NMI P1.6/EPA6 51 AdcEoc0 AD8
8 I/O I/O 17 /Ppi1Cs AD9
27 A8 AD converter 118 24B
1uF AGND P1.7/EPA7 50 9 I/O I/O 31 /Rd 26 A9 Miscellaneous C37 119 24C

GND
-WR

VSS
C112 /Reset DGND 0 27 30 AD10 23 8x12bit 10nF IC7 120 24D

A2
A1
A0
R102 RESET AD0 /Ale I/O Ready AD11 A10 MaxFilCur
AGND P3.0/AD0 26 16 I/O I/O 29 25 A11 DA converter 20 VCC CH0 1 121 25Z
AD1 ClkOut AD12 SafFilCur
8x100 P3.1/AD1 25 CLK1/I1 33 4 A12 AD7228 AGND 19 V+ CH1 2 Lamp1 122 25A

21
22
23
24
12
10
R88 +12V 1M 220nF R140 40 P0.0/ACH0 AD2 AD13 /WrDac0 /AdcCs1
PrimaryWaterTemperatureInput P3.2/AD2 24 AD3
IC4
AD14
28 A13
A2 SdataOut
18 CS CH2 3 Lamp2 123 25B
1M C116 PrimaryWaterTemperatureOutput R141 41 P0.1/ACH1 P3.3/AD3 23 M4A5-64/32 XPSC01L 29 A14 -5V 17 DI CH3 4 124 25C
DeionizedWaterTemperatureInput R142 42 22 AD4 IC17 AD15 3 A1 Sclk 16 5 125 25D
AGND P0.2/ACH2 P3.4/AD4 AD5 AD0 A0 A15 A0 C106 AdcSars CLK CH4 FilCurMon XsnR1+
DeionizedWaterTemperatureOutput R143 43 P0.3/ACH3 21 3 2 DGND 2 AGND 15 6 126 26Z
4

P3.5/AD5 AD6 AD1 D1 Q1 A1 A16 SdataIn SARS CH5 FilCurRefMon XsnR1-


100nF DEIONIZED WATER RESISTIVITY DeionizedWaterLevel R144 44 P0.4/ACH4 P3.6/AD6 20 4 D2 Q2 5 1nF 14 DO CH6 7 127 26A
3 + R145 45 P0.5/ACH5 AD7 AD2 A2 /CodeCs
R72 D5 R73 WaterResistivityDriver P3.7/AD7 19 AD3
7 D3 Q3 6 A3 /Rd
22 CE 0000H
Vref+5V R59 +5V 13 SE CH7 8 XsnT1+
128 26B
1 DeionizedWaterResistivity R146 46 P0.6/ACH6 8 D4 Q4 9 24 OE - AGND 12 VREF COM 9 129 26C
2 - Vref+2V5 R147 47 18 AD8 AD4 13 12 A4 1 DFFFH 11 10 XsnT1- 130 26D
IC19A 1K 1N4148 15K R79 C158 P0.7/ACH7 P4.0/AD8
17 AD9 AD5 14
D5 Q5
15 A5 31
VPP 10 AGND DGND
131 27Z
LF347N C97 XsnTxD P4.1/AD9 AD10 AD6 D6 Q6 A6 +5V PGM
15K 1uF 32 P2.0/TXD P4.2/AD10 16 17 D7 Q7 16 A7 ADC0838 +5V 132 27A
11

XsnRxD 33 P2.1/RXD AD11 AD7 C81 XpscDisplay


1000uF P4.3/AD11 15 AD12
18 D8 Q8 19 AT27C010-70PU AGND DGND AGND
DisplayRd2
133 27B
AdcSars 34 P2.2/EXINT P4.4/AD12 14 C76 134 27C
C87 AGND AGND AGND XpsReady 35 P2.3/BREQ 13 AD13 /ALE 11 IC14 10uF 1nF 135 27D
-12V 100nF /AdcConv0 36 P2.4/INTOUT
P4.5/AD13
12 AD14 1
CLK A0 12 13 AD0 +5VM DGND 136 28Z
XpscOk P4.6/AD14 AD15 CLR A1 A0 D0 AD1 R124
+5V 37 P2.5/HOLD P4.7/AD15 11 IC2D 11 A1 D1 14 AD2 AGND AGND 137 28A R125
AGND R80 XpscNotOk 38 P2.6/HLDA Ale 9 8 +5V 74HC273 A2 10 A2 15 138
A3 D2 28B R126
ClkOut 39 P2.7/CLKOUT P5.0/ALE/ADV 4 9 A3 D3 17 AD3 C14 139 28C R127
3 EnWd 74HC14 A4 8 A4 18 AD4 FilamentCurrent 140 28D
1K R105 R99 P5.1/INST /Wr A5 D4 AD5 1nF /AdcCs3Ext R128
AGND P5.2/WR/WRL 9 7 A5 D5 19 Miscellaneous 8x12bit 141 29Z R129
R85 4K7 4K7 7 /Rd A6 6 20 AD6 IC3 AD converter 3.5V/140mA Ik R18 100 EmissionCurrent- /AdcCs2Ext 142
P5.3/RD R117 A7 A6 D6 AD7 /AdcCs1Ext 29A R130
100K C69 P5.4/SLPINT 1 +5V 5 A7 D7 21 AGND 28 VD+ CH0 1 143 29B R131
/PrWaFlow +5V 8 SsioTxSync +5V AD8 27 A8 /AdcDor0 27 DOR 2 3.5V/140mA Ia R23 100 EmissionCurrent+ ExSdataOut 144 29C
/DeWaFlow 1nF P5.5/BHE/WRH Ready 1K IC12 AD9 4MHz CH1 ExSclk
P5.6/READY 2 26 A9 26 CCLK CH2 3 145 29D
GND 49 VREF EepromCs AD10 Sclk Hv R30 100 HighVoltage AdcSars
AGND
C109 C103 P5.7/BUSWIDTH 10 Sclk
1 CS VCC 8 AD11
23 A10
SdataOut
25 SCLK CH3 4 4V/75KV
SdataIn
146 30Z R132
6 VPP 2 SK NC 7 C50 25 A11 24 DI CH4 5 147 30A R133
SdataOut AD12 SdataIn
51

J450 100nF 100nF +5V +5V 48 ANGND P6.0/EPA8 58 /XsnTxDEn 3 6 10nF 4 23 6 C42 C19 C30 148 R134
29 EA /AdcCs0 SdataIn 4 DI NC AD13 A12 /AdcCs0 DO CH5 XpscTemp 30B
SUD-D 50PM P6.1/EPA9 59 /AdcCs1 DO GND 5
AD14
28 A13
/AdcConv0
22 CS CH6 7 149 30C R135
P6.2/T1CLK 60 3 A14 21 CONV CH7 8 1uF 1uF 1uF 150 30D R149
34 1 L7 DGND +5V R26 R78 AGND 61 /AdcCs2 EEPROM AD15 31 AdcEoc0 20 9 EmissionCurrent+ 151
DGND 18 1 2 +5Vin 1M 10K P6.3/T1DIR
62 DGND 2
A15
19
EOC COM
10 HighVoltage 13x 100 Ohm 152
31Z
P6.4/SC0 93C66 DGND A16 DGND PD MUXOUT1 31A
35 2 SdataIn EmissionCurrent-
L8
-12Vin R71 P6.5/SD0 63 Sclk
or
/DataCs 22 Vref+5V AGND
18 AGND A/DIN1 11 AGND
Vref+5V
153 31B
19 1 2 P6.6/SC1 64 +5V CE1 E000H 17 VREF+ MUXOUT2 12 154 31C
36 3 C35 R77 R63 SdataOut IC12A Vref+2V
AGND L9
+12Vin /WaterCoolOn R76 +24V 10K
/WaCoOn P6.7/SD1 65 EepromCs 1 +5V /Wr 30 CE2 - R58
10
16 VREF- A/DIN2 13 InputCurrent
155 31D
20 1 2 0 0 R27 CS VCC 8 29 WE FF3FH 15 VA+ DGND 14 156 32Z
37 4 Sclk 2 SK /Rd 24 InputVoltage
L10
+24Vin
C173
4K7 1uF IC9 80C196KR 180 SdataOut NC 7 OE Mon400V
157 32A
21 1 2 IC22 3 DI NC 6
C67 C75 ADC12138 158 32B
38 5 1uF DGND DGND ClkOut SdataIn 4 DO GND 5 RAM128K 10uF 10nF DGND AGND 159 32C
PGND PGND PC817 /AdcConv0 C78
22 160 32D
WaResist 39 6 XpsIntOut PGND AdcSars EEPROM DGND 10nF IC26
23 Pressostat/24V /XsnTxDEn 93C66/SO AGND AGND DspWdToggle FilPwmH 2 A1 Y1 18
WaLevel 40 7 Intlk2 FilPwmL 4 A2
Intlk3 DGND
+5V DGND +5V +5VM AGND
WdToggle HvPwmH Y2 16
24 6 A3 Y3 14 GND DIN41612
WaterTOut 41 8 Intlk4 HvPwmL 8 A4 Y4 12
3

25 Intlk5 C155 R156 +5V +5V +5V CurrentControl 11 A5 9


PrWaTIn 42 IntlkRtn CurrentLimit Y5
9 1nF 10K IC21A IC27A 13 A6 Y6 7
26 PrWaValve XsnT1+ 16 IC10 IC11 1 13 PowerOn 15
WaterTIn 43 /DeWaFlow XsnRxD LD1 R3 AD0 Int1 AD0 R150 A Q A7 Y7 5
10 1 IC2E 34 D0 PA0 4 34 D0 PA0 4 2 B Q 4 17 A8 Y8 3
PrWaTOut 44
27
11
/PrWaFlow
/GenxReset
R83
1K
15
XpscOk 11 10
330 AD1
AD2
33 D1
32 D2
PPI 1 PA1 3
2
Int2
Int3
AD1
AD2
33
32
PPI 2
D1 PA1 3
2
220K C166
2.2uF 14
R106
10K 1
R152
10K
GenxReady XsnT1- AD3 PA2 Int4 AD3 D2 PA2 C 1OE
28 R157 4 31 D3 PA3 1 31 D3 PA3 1 +5V AA22451 15 R/C +5V 19 2OE
RelayOff 45 12
29
/WaterCoolOn
WaterCooler
10K
14
DGND
XpscOk/Green AD4
AD5
30 D4
29 D5
I PA4 40
Int5
IntRet
AD4
AD5
30 D4 I PA4 40 +5V +5V IC27B
DisplayRd2 46 Buzzer
C156 +5V 74HC14
AD6 PA5 39 +24VIntOn AD6
29 D5 PA5 39 3 CLR
R151
9 A Q 5 DGND SN74ACT244 DGND
13 1nF 2 +5V 28 D6 38 28 38 10 Q 12

1
2
3
4
WaterPump AD7 PA6 /WaCoOn AD7 D6 PA6 22K C167 B
30 DGND 13 27 D7 37 27 37 MM74HC423A

2
4
6
8
XrayLamp1 47 XrayLampC R118 PA7 D7 PA7 10nF
14 DGND LD2 T=100ms 6 C
31 XrayLamp2 8922AM 1K R4 /Rd 5 RD 18 ExRelayPresent /Rd 5 PowerOn
XsnT1- XsnT1+
IC2F
/Wr PB0 /Wr RD PB0 18 EnPower S1 W2
7 R/C
48 15 330 36 WR PB1 19 36 WR PB1 19
6

32 XsnR1+ XpscNotOk 13 12 A0 9 A0 20 A0 9 20 IC24 EnPower 11


XsnR1- A1 PB2 A1 A0 PB2 RN1 Sclk DspSclk1 CLR
49 16 8 A1 PB3 21 8 A1 PB3 21 2 A1 Y1 18
IC21B
I O

8
7
6
5

1
3
5
7
33 9 DGND XpscNotOk/Red 35 RESET 22 /PrWaFlow 35 22 10 SdataOut 4 16 DspDr1B MM74HC423A
DGND /Ppi1Cs PB4 /DeWaFlow /Ppi2Cs RESET PB4 /DspPCe RfsSync A2 Y2 Rfs1
50 17 8 74HC14 6 CS PB5 23 6 CS PB5 23 9 6 A3 Y3 14 T=100us
10 24 Reset 24 EnDisRd2 8 DspDt1 8 A4 12 Sdatain Accessories
/XsnTxDEn PB6 PB6 EnXpscDis Y4 ExSclk
PB7 25 PB7 25 7 +5V 11 A5 Y5 9 ExSdataOut
C153 5 Address 6 13 A6 7
52

XsnR1+ BuzzerOn /DspCs Y6


1nF Ppi1Cs FFC0...FFFF FFC0H PC0 14 PrCoolCtrl
FF80H PC0 14 /AdcCs1Ext
5 15 A7 Y7 5 Fid Fid
11 Ppi2Cs FF80...FFBF - 15 - 15 4 R119 17 Y8 3
GND XsnR1-
12
7 WrDac0 FF40...FF7F FFFFH O PC1
PC2 16
XrayLampOn
/DspRst
FFBFH O PC1
PC2 16
/AdcCs2Ext
/AdcCs3Ext
3
/DspCs
10K
A8
DataCs E000...FF3F PC3 17 ExternalRelay PC3 17 2 1 1OE Fid Fid
C154 CodeCs 0000...DFFF PC4 13 PC4 13 1 19 2OE
8922AM WaPump TfsSync
1nF R54 PC5 12 PC5 12
IC1B
5 EnDisRd2 DGND IC1C
RelayOff ExRelayPresent O PC6 11
10
WaCooler
/EnFilCu
O PC6 11
10
SsioRxSync 9*10K
RfsSync DGND DGND
SN74ACT244
Mod: 1 OM4012 04.05.07 PPa
DisplayRd2 R19 C11 PC7 PC7
6 10 22K R65 2 Serie Version 07.12.10 PPa
C34 4 8 1nF C170 7K5 100nF 82C55 82C55
100 9 XsnTxD Reset
1nF 3 OM4266 08.01.13 GZb-PPa
7403 PGND
R17 7403 DGND DGND
Dessiné 31.10.05 MBz Remplacé par:
DGND +5V 10K
Q4
R153
ExternalRelay TP5 XPSC2, X-ray Power Supply Control Board Echelle
+5V MPSA05 +12V
R20
10K
DGND
IC1D
EnXpscDis
2K2
TP2 TP7 +12Vin MF31
L4
2 +12V
Version 2 Contrôlé
Remplace:
R14 13 R11 PGND +5V +5VM +24V
XpscDisplay 11 10K BZX85C5V1 L5 MF1 L2 +5V L1 MF4 L6 MF-R110 TP4 Vu
R22 DspDisp +5Vin +24Vin Liste S704011 Schémas S112306
C23 12 AGND 1 2 -5V 1 2 1 2 1 2 +24V C33 C22 AGND
100 C2 100nF Ensemble ---------- Circuit S920090 Groupe Technique:
470 R1 220uF
1nF 7403 +5V D20 C39 MF-R250 MF-R250 TP6 Neoseq Ingénieur G. Zbinden
1uF AGND Appareil Implantation S112307 9850
R8 R138 R139 R137 1uF C5 C1 TP1 C53 C46 PGND
DGND 120 120 120 100nF GND 0
10K 220uF DGND R2 220uF 100nF TP3 Groupe No

DGND
C79 C49 C89 C4 C3 C60 C47 C25 C61 C48 C146 C164 C165
-12Vin
AGND AGND
0
PGND
-12Vin
MF2
1
L3
2
C18
220uF
C12
100nF
-12V
ThermoFisher S112306
No Mod.

DECOUPLING CAPACITORS = 10nF DGND


DGND GND
MF-R110
-12V S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland Page: 1/1 --- 3
FilPwmH R26 J905
680
R6 1 1Z
FilPwmL Q3 2
GND 1A
MPSA05 +15VRTM3 3 1B
10K R5 IC6 4
+24Vin 1C
4K7 5 1D
8 +24V 6
C39 2Z
PGND 7 2A
DGND DGND R65 2 7 100nF 0VRTM3 8 2B
1K R13 U1 +12V 9
P1 2C
3 6 10 2D
10 1 + AGND 11 3Z
5 2 12 3A
- -12V 13 3B
B WAGO 2P 14 3C
HCPL3101 +5Vin 15
FAN 24V 16
3D
DGND 4Z
17 4A
PGND /PwmTrip 18 4B
19 4C
+15VRTM2 20 4D
IC2 21 5Z
R24 22 5A
HvPwmH 8 23
C35 24
5B
120 R61 2 7 100nF 0VRTM2 25
5C
5D
1K R9 26 6Z
3 6 Line1 27 6A
10 28 6B
5 R59 10K 29
HvPwmL R8 Q4 6C
30 6D
MPSA05 0VRTM3 31
10K 7Z
R7 HCPL3101 32 7A
4K7 33 7B
+15VRTM2 C6 C69 C71 34
IC4 7C

13

10
35

9
SPM1 7D
DGND DGND 2.2uF 1nF/500V 1nF/500V 36 8Z
8 Line2 37
C37 38
8A
100nF 0VRTM2 8B
2 7 39 8C
R11 40 8D
R63 3 6 R55 10K 1 41 9Z
1K 10 42 9A
5 0VRTM2 2 43 9B
+5V 44 9C
R57 10K 5 45 9D
HCPL3101 46 10Z
R23 0VRTM2 6 +400VRTM 47 10A
1K 48 10B
+15VRTM PowerTemp 18 16 ToHvTransfo 49
IC5 10C
C3 C4 C5 R3 50 10D
R25 C31 15 1000uF 1000uF 1000uF 10K/5W 51 11Z
HvPwmL 8 100nF F2 250V 250V 250V +400VRTM 52
C38 NTC
14 ToFilamentTransfo +200VRTM 53
11A
120 R64 2 7 100nF 19 54
11B

4
AGND 10A 11C
1K R12 AGND 55 11D
3 6 -10VRTM R58 10K 7 C17 C18 C19 R4 56
C68 12Z
10 1000uF 1000uF 1000uF 10K/5W 10nF/1kV 57 12A
5 0VRT 8 T2 T3 250V 250V 250V 58
HvPwmH R32 Q5 0VRTM 12B
59 12C
MPSA05 60 12D
10K R31 HCPL3101 R56 10K 3 C70 61

2
13Z
4K7 AGND AGND 62 13A
+15VRTM 0VRT 4 1nF/500V 63
IC3 D10 D9 D14 D13 13B

-
64 13C
DGND DGND 65 13D
8 MWI75-06A7T 66

17

11
12
C36 14Z
67 14A
R62 2 7 100nF D12 D11 68
D16 D15 14B
1K R10 69

+
0VRT 14C
3 6 -10VRTM 4xBYV26 4xBYV26 70 14D
10 71 15Z
5 +200VRTM 72
R60 10K 15A
73 15B
74 15C
R27 HCPL3101 75 15D
FilPwmL C67 76
FAN 24V 16Z
680 1nF/500V 77 16A
U2 Not mounted 78 16B
on Low Power version 79
FilPwmH R68 Q6 80
16C
16D
MPSA05 +15VRTM 81

+
-
10K IC7 17Z
R67 82 17A
4K7 P2 +5VM 83 17B
8 0VRT 84
C40 B WAGO 2P 17C
85 17D
DGND DGND R66 2 7 100nF 86

1
2
IC8 18Z
1K R14 87 18A
3 6 -10VRTM PC817 88 18B
10 +24V PGND 89 18C
5 90 18D
91 19Z
R20 R16 0VRTM 92
D19 19A
HCPL3101 33K 1K 93 19B
94 19C
R42 BAT85 +5VM +24V 95 19D
AGND 96 20Z
D21 C7 97 20A
100K +5V 98
BZX55C7V5 10nF/1kV 20B
R17 R19 99 20C
2K7 47K 100 20D
C62 -10VRTM 101 21Z
C65 470nF 10nF 102 21A
R40 +5VM 103
8

21B
10K C63 C64 +15VRTM GND 104
AGND IC10A + Vref+5V 21C
Q1 R44 3 8x8bit ADC 10nF R15 10nF R18 105 21D
MPSA05 1 C26 Status AD converter 10K 10K 106 22Z
2 HvIgbtCurrent /AdcCs2 R51 100 107
2K2 - 10nF IC1 D22 22A
OP284 108 22B
R38 SdataOut R50 100 20 1 AGND AGND AGND AGND C72 109
C58 C56 BZX55C11V
4

C60 100 VCC CH0 22C


R36 100nF 10nF R34 R53 R54 AGND 19 V+ CH1 2 110 22D
1nF 15 15 12 Sclk R49 100 18 3 ToHvTransfo C73 111
D17 10K CS CH2 23Z
DGND 17 DI CH3 4 C29 +5VM R28 112 23A
BAT85 AdcSars R48 100 16 5 1K C74 113
AGND AGND AGND AGND AGND AGND AGND AGND 15
CLK CH4
6 10nF 114
23B
SdataIn R47 100 SARS CH5 PowerTemp 23C
14 DO CH6 7 115 23D
13 8 XpspTemp IC9 3x 2u2/250V 116
/PwmTrip Vref+5V R30 +5V SE CH7 24Z
12 VREF COM 9 AGND PC817 117 24A
10 11 AGND DGND 10 118 24B
C11 R46 119 24C
D18 ADC0838 0VRT 120
470pF R33 24D
(Not mounted) BAT85 C53 C41 121 25Z
C55 AGND 0 33K 122
1nF 10nF 25A
C66 470nF 10uF DGND AGND 123 25B
D20 AGND 124 25C
AGND AGND AGND AGND 125 25D
Q2 R45 R43 BAT85 126 26Z
127 26A
MPSA05 ToFilamentTransfo 128
2K2 100K 26B
R41 129 26C
IC10B 10K 130 26D
+ 5 Vref+5V 131 27Z
7 132 27A
DGND - 6 FilamentIgbtCurrent 133 27B
OP284 134 27C
R39 135
C61 100 C59 C57 FilPwmH 27D
R37 100nF 10nF R35 136 28Z
10K 1nF 180 FilPwmL 137
HvPwmH 28A
138 28B
HvPwmL 139 28C
AGND AGND AGND AGND AGND FilamentCurrent 140 28D
/AdcCs3 141 29Z
/AdcCs2 142 29A
/AdcCs1 143 29B
SdataOut SdataOut 144
Sclk 29C
Sclk 145 29D
AdcSars AdcSars 146
SdataIn 30Z
SdataIn 147 30A
CurrentControl 148 30B
CurrentLimit 149 30C
PowerOn 150 30D
EmissionCurrent+ 151 31Z
HighVoltage 152 31A
EmissionCurrent- 153 31B
Vref+5V 154 31C
Vref+2V 155 31D
InputCurrent 156 32Z
InputVoltage 157 32A
Mon400V 158 32B
159 32C
+15VRTM3 GND 160 32D
R22 D7

10 TP1 +24V
BYT11
MF1 L1 +24V DIN41612
+15VRTM +15VRTM2 +24Vin
D4 PTC1 R21 D5
MF-R250 C2 C9 TP2
C46 D8
1N4007 5.6/25°C 10 BYT11 220uF 100nF PGND
2200uF/25V 1N5355B R2
C45
Line1 1 T1 12 C23 C27 D3 D6
2200uF/25V 0
2200uF/25V 2200uF/25V 1N5352B 1N5355B PGND
GND
11
8 0VRT 0VRTM2 0VRTM3

F1
Line2 6 7
230V/2x15V/10VA +5V
0.5A C16 C28 D2 TP3
2200uF/25V 2200uF/25V 1N5347B +5V +5VM
D1 PTC2 L2 +5V L3
+5Vin MF2 R29
10K
1N4007 5.6/25°C MF-R110
C10 XpspTemp Mod: 1 Version pré-série 23.08.01 PPa 4 Version série 11.04.03 PPa 7 Addendum OM3834 30.11.04 PPa
C1 C8 TP5 TP6
-10VRTM TP4 1uF 2 Version pré-série (2) 07.11.01 PPa 5 OM3753 & OM3758 15.01.04 PPa 8 OM3884 21.02.05 PPa
220uF 100nF AGND AGND C32
DGND RV1
R1 4K7/25 100nF 3 Version "phase in" 03.09.02 PPa 6 OM3834 28.10.04 PPa 9 OM4033 30.07.07 PPa
0 02.12.00 MBo Remplacé par:
GND DGND AGND AGND AGND
XPSP X-ray Power Supply Power Board Echelle Dessiné

Contrôlé
Remplace:

Liste S703126 - S703779 Schémas S112067 Vu


Ensemble ---------- Circuit S920078 Groupe Technique:
Appareil XRF Implantation S112073 Ingénieur M.Bonzon 9850

ThermoFisher
Groupe No No Mod.

S C I E N T I F I C
S112067
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 9
+5VA
R78 20k
R77
10k +5V
C69 J905
R88 D2 C52 R83 +5V 100n
+VGateL_3 100n IC16 +24V_Vent
39k
LMV7235 1

5
1Z
270 BYG20J DGND 2 1A
AGND 3 + GND 3 1B
R84 1 +5V P2 4 1C
IC10 4 - 2 4 + +24Vin 5 1D
1 IN OUT 4 1 3 6 2Z
IKW40T120_L2 IKW40T120_L2 IKW40T120_L2 2 5 ToHvTransfo 10k C53 R42 - 7
PGND

2
IN OUT 2A
Z1 Z2 Z3 3 IN OUT 6 1n 1k B WAGO 2P 8 2B
+5VM C22 D1 +12V 9
GH1 GH2 GH3 2C
1 1 1 R44 0 FAN 24V 10 2D
7 9 Mes_I_Transfo Limit_I_Transfo 1 2 /PwmTrip 11
C48
VCC OUT
PGND AGND 12
3Z
1n

3
EmH1 EmH2 EmH3 3A
100n 8 GND REF 10 AGND DGND -12V 13 3B
+400VRTM C33 BAT54T1 14
+5Vin 3C
LAS50-TP 10n 15 3D
AGND 16 4Z
R8 C13 C12 C11 C21 +24V_Vent DGND 17
/PwmTrip 4A
10k/5W 1000uF 1000uF 1000uF DGND 18 4B
4.7uF 19 4C
+200VRTM R63 R54 24k 20 4D
P1 21 5Z
IKW40T120_L2 IKW40T120_L2 IKW40T120_L2 2 4 + 22

2
C20 5A
R25 C23 C26 C25 Z4 Z5 Z6 10k +5V 1 3 23 5B
10k/5W 1000uF 1000uF 1000uF - 24 5C
4.7uF GL1 1 GL2 1 GL3 1 +5V C66 B WAGO 2P 25
0VRTM +5VA 5D
IC14 100n 26 6Z
LMV7235 FAN 24V 27

5
3

3
EmL1 EmL2 EmL3 +5VA_PTB 6A
IC19 D3 28 6B
R53 3 + DGND PGND TR1 29 6C
10k 1 2 VOUT VIN 3 1 2 30 6D
4 - 31 7Z
Limit courant pos Not mounted 4 VOUT ADJ 1 BAT54T1 32 7A
C44 R62 C45 on Low Power version C73 33

2
7B
100n 5.1k 1n Limit courant neg comp+ = 3V C71 34 7C
comp- = 2.57V 1u LM317-SOT223 47u 35 7D
DGND Comp+ = 2.36V +5VA 36 8Z
AGND AGND Comp- = 0.71V 37 8A
R80 33k AGND 38 8B
R79 240 R90 R89 VB0.35/1/6 39 8C
10k +5V 715 40 8D
DGND 41 9Z
C70 42 9A
C54 R85 +5V 100n AGND 43 9B
100n 30k IC17 44 9C
LMV7235 45

5
9D
DGND 46 10Z
AGND 3 + 47 10A
R86 1 +5V 48 10B
4 - 49 10C
50 10D
10k C55 R69 51

2
+400VRTM 11Z
+5VM 1n 1k 52 11A
53 11B
DGND R75 0 54 11C
C49 IC11 AGND /Limit_I_Fil 55
Mes_I_Fil Not mounted 11D
100n 7 VCC_5V Uout 9 C32 56 12Z
8 GND Ref 10 10nF/1kV 57 12A
C43 58 12B
AGND Isol. galvanique 10n 59
LEM6 12C
1 I_mes1 I_mes6 6 R65 R56 47k 60 12D
LEM6 2 5 LEM5 61
LEM5 I_mes2 I_mes5 13Z
3 I_mes3 I_mes4 4 DGND 62 13A
10k +5V 63 13B
64 13C
LEM_LTSR ToFilamentTransfo +5V C67 65
+5VA 13D
IC15 100n 66 14Z
LMV7235 67

5
14A
68 14B
R55 3 + DGND 69 14C
10k 1 70 14D
4 - 71 15Z
+200VRTM 72 15A
C46 R64 C47 73

2
15B
100n 5.1k 1n Limit courant neg 74 15C
75
Isol Galv DGND Comp+ = 2.35V C19 76
15D
+VGate_1 16Z
AGND AGND 1nF/500V 77
Isol Galv +VGate_3 Not mounted Comp- = 1.87V 78
16A
+24V 16B
79 16C
IC4 R12 +24V 80
C14 16D
0 IC1 81 17Z
3 12 10u C3 82
+VIN +VOUT 17A
R14 1 3 +VIN +VOUT 12 10u 83 17B
C58 1 18 C15 EmH1 84
-VIN NC C6 17C
1u 10u C57 1 -VIN NC 18 85 17D
Com 7 1u 10u 86 18Z
7 EmH3 87
EmH2 Com 18A
-VOUT 9 88 18B
PGND C16 9 89
-VOUT C7 +5VA 18C
10u R13 1 PGND 90 18D
TES 1-2423 10u 91 19Z
-VGate_1 TES 1-2423 0VRTM 92
-VGate_3 19A
93 19B
R37 94 19C
10k R30 100k 95 19D
+5V 96 20Z
C39 97 20A
+VGate_1 R38 C35 98
+24V 30k 1u R31 10nF/1kV 99
20B
Isol Galv +5V 1k 100
20C
Q1 Isol Galv +24V 101
20D

5
+5VA 21Z
MJD44H11 IC2 102 21A
R40 AGND 3 + 103
IC5 Lim_24V 21B
R23 270 3 VIN VOUT 2 47k 1 GND 104 21C
HvPwmH 1 8 R15 R16 4.7 R9 270 IC3 4 - 105
IN VCC GH1 FilPwmH +VGate_3 C17 21D
R21 0 1 IN VCC 8 C8 1 ADJ VOUT 4 106 22Z
7 R7 1u C10 R39 C36 IC7 C31 107

2
Out GH3 22A
R19 1k Out 7 10k 1u LMV7235 10n 108 22B
HvPwmL 2 10 R6 1k LM317-SOT223 1u DGND 109
Cat R20 GH2 FilPwmL 10 22C
3 Cat Ref 6 2 Cat 110 22D
10k C18 R22 3 6 R10 DGND +5V DGND ToHvTransfo C30 111
Cat Ref AGND 23Z
M4 4 NC Ref 5 10n 0 R18 4.7 10k DGND 112 23A
FDV301N M1 4 NC Ref 5 R4 R3 240 C60 113 23B
DGND ACPL_H312 FDV301N 715 100n 114 23C
MJD45H11 DGND ACPL_H312 3x 2u2/250V 115
Q2 -VGate_3 23D
R17 1k 116 24Z
Lim_24V R5 1k AGND DGND C38 117
-VGate_1 Lim_24V 24A
118 24B
119 24C
M5 120 24D
FDV301N M2 +24V 121 25Z
DGND FDV301N
+24V_Vent TP1 +24V 122 25A
DGND 123 25B
MF1 124 25C
R24 +24Vin 125
Limit_I_Transfo 25D
R11 126 26Z
/Limit_I_Fil MF-R250 C1 C4 TP3 127
ToFilamentTransfo 26A
1k 220u 128 26B
M6 1k 100n PGND 129
R2 26C
FDV301N M3 130 26D
DGND FDV301N 131 27Z
DGND 0 PGND 132 27A
GND 133 27B
134 27C
135 27D
FilPwmH 136
FilPwmL 28Z
137 28A
HvPwmH 138
HvPwmL 28B
139 28C
FilamentCurrent 140 28D
/AdcCs3 141 29Z
TP2 /AdcCs2 142
+5VA 29A
+5V +5VM /AdcCs1 143 29B
L3 +5V L2 SdataOut SdataOut 144
+5Vin MF2 Sclk 29C
Sclk 145 29D
R72 AdcSars AdcSars 146
MF-R110 SdataIn 30Z
1K SdataIn 147 30A
C9 PowerTemp CurrentControl 148
Isol Galv +VGateL_1 Isol Galv C2 C5 1u
TP5
CurrentLimit 149
30B
+VGateL_3 220u 100n AGND C68 PowerOn 150
30C
+24V R73 30D
R1 100n EmissionCurrent+ 151 31Z
IC6 +24V 6.2k HighVoltage 152
C27 31A
IC9 0 EmissionCurrent- 153 31B
3 12 10u C37 GND DGND AGND Vref+5V 154
+VIN +VOUT 31C
R26 1 3 +VIN +VOUT 12 10u AGND Vref+2V 155 31D
C59 1 18 C28 EmL1 InputCurrent 156
-VIN NC C40 32Z
1u 10u C61 1 -VIN NC 18 InputVoltage 157 32A
7 ComGateL_1 1u 10u Hole Hole Mon400V 158
Com EmL3 32B
Com 7 R57 159 32C
9 EmL2 TP7 TP8 TP6 TP4 +5VA 160
-VOUT C29 32D
PGND -VOUT 9 DGND DGND DGND DGND 0
10u R27 1 PGND C42
TES 1-2423 10u
-VGateL_1 TES 1-2423 R82 DGND GND
-VGateL_3 10K DIN41612
DGND DGND DGND DGND
XpspTemp

RV1 C50
+VGateL_1 100n
4K7/25
Isol Galv
Q3 Isol Galv AGND AGND
MJD44H11
R43 270 IC8 +5VA +24V
HvPwmL 1 8 R28 R33 4.7 R70 270 IC13
IN VCC GL1 FilPwmL +VGateL_3
R36 0 1 IN VCC 8
Out 7 R61
R34 1k 7 GL3 R46 R49
HvPwmH 10 Out
2 Cat R60 1k 2K7 47K
3 6 R35 GL2 FilPwmH 2 10
Cat Ref R41 Cat R71
10k C34 3 Cat Ref 6
M7 4 5 10n 0 R29 4.7 10k +5VA
NC Ref
FDV301N M10 4 NC Ref 5 C62 C63
DGND ACPL_H312 FDV301N 8x8bit ADC 10n R47 10n R50
MJD45H11 DGND ACPL_H312 C41 Status AD converter 10K 10K +5VA_PTB
R32 1k Q4 -VGateL_3 /AdcCs2 R48 100
Lim_24V 10n IC12 Mod: 10 New Design 25.05.09 PPa 13 Add. OM4180 + OM4193 03.03.11 PPa
R52 1k
-VGateL_1 Lim_24V SdataOut R51 100 20 1 AGND AGND AGND AGND
VCC CH0 11 OM4130 17.07.09 PPa
AGND 19 V+ CH1 2
M8 Sclk R59 100 18 3
CS CH2 12 OM4180 30.08.10 PPa
FDV301N M11 17 DI CH3 4
DGND FDV301N AdcSars R68 100 16 5
CLK CH4 Remplacé par:
DGND
SdataIn R74 100
15
14
SARS
DO
CH5
CH6
6
7 PowerTemp
XpspTemp
+5VA_PTB Accesories
XPSP X-ray Power Supply Power Board Echelle Dessiné 15.05.09 TMa
R45 R81 +5V 13 SE CH7 8
Limit_I_Transfo R76 Vref+5V 12 9 Hole Hole Contrôlé
/Limit_I_Fil VREF COM Remplace:
10 11 AGND DGND 10
1k R87
M9 1k ADC0838 Liste S703126 - S703779 Schémas S112067 Vu
FDV301N M12 C51 Fid Fid
DGND FDV301N C56 AGND 0 Ensemble ---------- Circuit S920078 Groupe Technique:
DGND 10u 1n DGND AGND Fid Fid Ingénieur T. Maulaz
Appareil XRF Implantation S112073 9850
AGND AGND Fid Fid Groupe No
Fid Fid ThermoFisher S112067
No Mod.

S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 13
+24V
P1
1 5 +12V
2 6 + +200VRTM
3 7
4 8 - C64
AGND +5V
10nF

4
WAGO 4P ToHvTransfo
J906 FAN 24V IC5A 3
IC5C R65 +
PGND R79 10 1
D20 IC5D - 13 8
+
R67 2 R18
R76 10K -
1 + /PwmTrip 14 9 OP484 10K
1Z
2 + 12 10K OP484
-
R66 C66 T1A T1B

11
1A 1K XpsmHvTemp
1B 3 GND - BAT85 OP484 R78 10K 10K 47pF
4 MF6 D21 Vref+2V C67
1C BAT85 R68
1D 5 +24Vin FAN 24V 10nF C35 RV2
6 1K AGND -12V AGND 100nF 4K7/25

4
2Z C65 10K
2A 7 MF-R110 AGND
8 PGND DGND 10nF C77 T1
2B +12Vin R64
2C 9 R80 +12V AGND AGND
2D 10 AGND R69 HvTransfo
11 220pF 1K 30K
3Z AGND -12Vin 1K
3A 12 C72 Not on the Board
13 IC5B C63

2
3B
3C 14 6 - 10nF
15 +5Vin 7 3n3
3D
16 5 + AGND

4
4Z DGND IC4B
4A 17 OP484 5 R41 R63
18 7
+ AGND IC4A 3
4B R56 +
4C 19 6 1
20 AGND OP484
-
2 1K 20K
4D 10k -
5Z 21 Positive HV OP484 h18 R28
22 D11 R52 R59

11
5A Arc C44
5B 23
24 C40 1nF 200M
5C
25 1N4148 100 10K 10nF 20x 4n7/3kV
5D C59
6Z 26 -12V
6A 27 Line1 R21 C34 1nF AGND GAP6
28 Line1 C33 100K D12 R48 (Not mounted) R35 M51-C90
6B
29 10uF BZX85C5V6 AGND C1 C2 C3 C4 C5
6C 10nF
6D 30 5x 20nF/12kV
31 100 1K
7Z Negative HV
7A 32 AGND GND
7B 33 AGND AGND AGND
7C 34 D1 D2 D3 D4 D5 D6 D7 D8 D9 D10 D11 D12
35 +5V IC4C 10 MF5
7D + AGND R42 R13 h1
36 8 R57 IC4D + 12
8Z
37 Line2 9 14 24x Z100UFG
8A
38 Line2 D22 OP484
-
13 1K MF-R050 15/3W GAP1
8B
39 BAT85 Positive HV 10k -
A81-C230 D1' D2' D3' D4' D5' D6' D7' D8' D9' D10' D11' D12'
8C OP484
40 R26 C45 Wired resistors JP1
8D EmissionCurrent+ R53 R60
9Z 41 1uF GAP3
42 M51-C90 GND C6 C7 C8 C9 C10 C11 6x 20nF/12kV R1 R2 R3 R4 R5 R6 R7 R8
9A
43 100 D13 100 10k 1
9B C60 2
9C 44 BAT85 AGND 3
9D 45 1nF 8x 100/9W
10Z 46 R49 (Not mounted) R36 GND D13' D14' D15' D16' D17' D18' D19' D20' D21' D22' D23' D24'
10A 47 D10 Fed.Standard
48 AGND 1N5347 C32
10B
49 100 1K 24x Z100UFG
10C Negative HV 100uF
10D 50 R40 R34 R30 R25
11Z 51 D13 D14 D15 D16 D17 D18 D19 D20 D21 D22 D23 D24
11A 52 HP:4x100
11B 53 +400VRTM
54 LP:1x100
11C
11D 55 C12 C13 C14 C15 C16
12Z 56 5x 20nF/12kV
12A 57 AGND
12B 58 HT Output board S940188
12C 59 +12V
12D 60 R19 R14 h4
61 IC3B 5 6x Multiplier board S930191
13Z +
13A 62 +5V C62 7 1K 15/3W
13B 63 - 6 GAP2 Ra1
64 10nF OP484 A81-C230
13C R58
13D 65 D14
66 AGND 1K R55

4
14Z BAT85
14A 67 GAP4 GND GND
68 IC3A 3 C36 D8 R16 C30
14B HvDividerDifference R38 R32 + 1K M51-C90
14C 69 1 10nF 1N5347 53K3 10uF 5x 200M/15kV 5x 470pF/15kV Divisor board S940189
70 2 R61 Wired resistors
14D
71 1K 1K OP484
-
10K
15Z +200VRTM Vref+2V R17 R16 R15 R14 R13
72 D16 R12 R11 R10 R9

11
15A
15B 73 C43 BAT85 C39
15C 74 10nF C74 AGND AGND AGND AGND GND
75 10nF 10nF 4x 1K/9W
15D
16Z 76 -12V C21 C20 C19 C18 C17
16A 77 AGND
16B 78 AGND R28 AGND R23
79 AGND R27 C27
16C
16D 80 10K 1K 33M/9kV 470pF/15kV
17Z 81 +5V
17A 82 Interface board
17B 83 D19
17C 84 S930192 R26 R25 R24 R23 R22
85 BAT85 R21 R20 R19 R18
17D IC3C R20 R15 h5
18Z 86 R62 R51 + 10
87 HighVoltage 8
18A
88 9 1K 15/3W GAP3 Ra2 4x 1K/9W
18B 100 100 -
18C 89 OP484 A81-C230 C26 C25 C24 C23 C22
18D 90 D18 Wired resistors
19Z 91 BAT85 R47 5x 200M/15kV 5x 470pF/15kV Divisor board S940189
19A 92 GAP5 GND GND Multiplier Box S703757 (not on the board)
93 0VRTM C37 D9 R17 C31
19B
94 1K 10nF 1N5347 53K3 M51-C90
19C 10uF
19D 95 AGND
20Z 96
20A 97
20B 98
20C 99 AGND AGND AGND AGND GND
20D 100
21Z 101 GND
21A 102
21B 103
21C 104
21D 105
22Z 106
22A 107
22B 108
22C 109
22D 110 GND GND GND
23Z 111 h57 H55
112 ToHvTransfo ToFilamentTransfo 1 T2
23A
23B 113 C13 D6
114 3 D2 D1 GAP2
23C 1nF/500V 1.5KE68A M51-C90
23D 115
24Z 116
24A 117 4 I1 J1
24B 118 h58 H56 1 2 C1 1
119 +200VRTM 2 D3 D4
24C 47nF/100V SENS1 C18 2
24D 120 FilamentTransfo 3 4 1nF/500V 3
121 4xMUR1520 C16 C22 SCD15PBR
25Z 4
25A 122 1000uF/65V 1000uF/65V 5
123 Not on the Board

6
25B GAP1
25C 124 MF4 C20 D7 WAGO
25D 125 M51-C90
126 MF-R050 1nF/500V 1.5KE68A
26Z
26A 127
128 ToFilamentTransfo
26B
26C 129 0VI U/I GND GND GND GND
26D 130 R11
131 +5V EmissionCurrent- SENSOR
27Z
27A 132 100
27B 133 C21
134 100uF/35V OUT 0V
27C R4 C25
135

1
2
3
4
27D 10nF
28Z 136 FilPwmH 10K R12 R9 R8 R10
28A 137 FilPwmL
138 HvPwmH XpsmFilamentTemp
28B HP:4x100
28C 139 HvPwmL AGND LP:1x100 +12V AGND
140 FilamentCurrent C17 RV1
28D /AdcCs3
29Z 141 +5V +12V -12V 100nF 4K7/25
29A 142 /AdcCs2 AGND
29B 143 /AdcCs1 AGND
144 SdataOut AGND
29C Sclk C11 C10 -12V
29D 145 R31 R46 R43 +5V +5VI
146 AdcSars 2K7 18K 33K 10nF 10nF
30Z SdataIn
30A 147
30B 148 CurrentControl +5V C69 C70 C71 R22 AGND AGND IC6
149 CurrentLimit 8x8bit ADC R27 100nF R37 100nF R50 100nF FilamentCurrent 0.266V/A R7 R3 2 4 C78
30C 10K +VIN +VO
150 PowerOn Status AD converter 10K 10K 330K 330K 1 3 10uF
30D
151 EmissionCurrent+ /AdcCs3 R70 100 100 -VIN -VO
31Z HighVoltage C38 IC2
31A 152 IK0505SA
153 EmissionCurrent- SdataOut R71 100 10nF 20 1 AGND AGND AGND AGND AGND
31B Vref+5V VCC CH0 C41
31C 154 19 V+ CH1 2 R33 +5V +5VI
155 Vref+2V Sclk R72 100 AGND 18 3 Vref+5V 1uF +5V C76 AGND 0VI
31D CS CH2 Arc
32Z 156 InputCurrent 17 DI CH3 4 C75
157 InputVoltage AdcSars R73 100 16 5 HvDividerDifference AGND 10K IC1
32A CLK CH4 FilamentVoltage
32B 158 Mon400V 15 SARS CH5 6 D17 10nF
159 SdataIn R74 100 14 7 XpsmFilamentTemp AGND 8 1 10nF
32C DO CH6 XpsmHvTemp C68 BAT85
32D 160 +5V 13 SE CH7 8 R5
GND Vref+5V R54 12 9 10nF OP484 - 13 R44 1K 7 2
VREF COM FilamentVoltage 0.242V/Vfil R45 R24
10 11 AGND DGND 10 R75 14 R6 1K
+ 12 R39 1K 6 3
C61 C53 1K 1K IC3D
ADC0838 C56 1K
DIN41612 10uF 1nF AGND 0 D15 5 4
DGND AGND 100uF/35V R29 C19 C15
BAT85
10K 100nF 100nF
AGND AGND AGND HCPL7800

AGND AGND AGND 0VI

Accessories

Hole Hole Hole Hole

MF1 L1 MF3 L3 Hole Hole Hole Hole


+12Vin +5Vin Mod: 1 Version "Phase-in" 10.09.02 PPa 4 OM4212 13.09.11 PPa 7 Change value C72 - 23.07.12 PPa
+12V +5V Hole Hole Hole Hole 2 Version série 14.04.03 PPa 5 OM4217 + OM4256 28.03.12 PPa
MF-R110 C4 C7 MF-R110 C2 C5 Hole Hole Hole Hole
220uF 100nF 220uF 100nF 3 OM3753 & OM3758 14.01.04 PPa 6 Maj. OM4217 + OM4256 28.03.12 PPa
R1
AGND Hole Hole Hole Hole
GND 0 C3 C6 Hole Hole Hole XPSM Echelle Dessiné 23.08.01 PPa Remplacé par

220uF 100nF
-12Vin MF2 L2
-12V XPSM X-ray Power Supply Multiplier Board
Hole Hole Hole Hole
X-ray Power Supply Multiplier Assembly Contrôlé
Remplace
MF-R110 Vu
Liste S703758 Schémas S112068
Hole Hole Hole Hole
Assembly : S703127 Ensemble ---------- Circuit ---------- Groupe Technique
Appareil XRF Implantation ---------- Ingénieur M.Bonzon 9850
GND
PCB : S920079 Format No

Layout : S112074 RAD RAD RAD RAD ThermoFisher S112068


Mod. No

Fid Fid Fid Fid S C I E N T I F I C


Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 7
2 D7
5
T1 BYV28 J903
8
11
1 1Z
GND 2

13

14
1A
3 1B
GND GND 4 1C
+24Vout 5 1D
2 D12 6 2Z
5 PGND 7 2A
T3 8 2B
BYV28 +12Vout 9 2C
8 10 2D
11 AGND 11 3Z
-12Vout 12 3A
13

13

14
3B
D23 14 3C
GND GND +5Vout 15 3D
1.5KE120A 16 4Z
DGND 17 4A
18 4B
2 D20 19 4C
D25 5 20 4D
D24 BYV26E T2 21 5Z
BYV28 22
C87 1.5KE120A 8 23
5A
10n +24V 5B
11 24 5C
+12Vout +12V 25 5D
+15Vout 26

13

14
Line1 6Z
27 6A
GND GND MF3 MF6 MF4 C38 28
IC12 IC7 6B
MF-R250 R80 R82 R54 MF-R050 MF-R020 29 6C
2 D29 L5 4 MF8 4 1nF/500V GND 30
FB FB 6D
5 1 IN LM2576 10K L8 L10 MF-R050 1 IN LM2576 9K1 L3 L6 31 7Z
2 1K5 2 32
T4 50uH OUT +15V OUT 7A
R104 HS1 BYV28 33

GND

GND
Line1 R105 C57 C78 470uH 50uH C74 470uH 50uH 7B
C33

ON

ON
2.2M 8 C53 470uF/50V C37 34 7C
HS-580 11 R78 D13 C83 C76 R50 D10 C48 35
4.7/4W 470uF/50V C59 7D
R95 470nF 100nF 1K BYV28 2200uF/16V 100uF/16V 470nF 100nF 1K BYV28 100uF/16V 36

5
2200uF/16V Line2 8Z
37

13

14
8A
- + 2 L R48 PGND C40 38 8B
7 GND GND PGND 0 R86 AGND AGND AGND AGND R85 AGND AGND AGND AGND 39
D 5K6/5W 8C
3 X PGND GND 47K 47K 1nF/500V GND 40 8D
R106 W08G 5 F 41 9Z
Line2 4 42
D2 1
S Z1 AGND AGND AGND AGND 43
9A
4.7/4W C R94 R91 9B
44 9C
5K6/5W 45 9D
IC10 150 BZX85C24V 46
R35 10Z
TOP249_TO220_7P 47 10A
R103 R13 24K IC13 IC8 48
IC3D R79 R83 R55 10B
6.8 R56 + 12 FB 4 FB 4 49 10C
D8 14 1 IN LM2576 L9 1 IN LM2576 L4 50 10D
5 1 13 10K 2 1K5 10K 2 9K1 51
OUT OUT 11Z
IC14 270
-
1N5352B OP484 R34 52

GND

GND
C72 C80 470uH C30 C85 470uH 11A
H11D1Z C21

ON

ON
2K7 C84 AGND C77 AGND 53 11B
C4 10nF 10nF R81 D14 2200uF/16V C60 C49 54
100uF/16V R51 D11 11C
C91 4 2 470nF 100nF 1K BYV28 470nF 100nF 2200uF/16V 100uF/16V 55

5
BYV28 11D
47uF/35V C5 R9 AGND L11 1K 56
C39 C50 C90 AGND 57
12Z
L7 12A
1uF AGND R87 R88 58
10K 50uH 12B
1uF 47K MF9 47K 50uH 59 12C
2 x 10uF/450V D19 R107 R4 MF-R050 60
MF5 MF7 12D
10K 10K 61 13Z
1N5346B MF-R050 MF-R020 62 13A
Vref+2V5 -15Vout 63 13B
64 13C
AGND C6 +5Vout +5V 65 13D
+24V +24Vout 66
10nF IC4 -12Vout -12V 67
14Z
RN1 W2 R33 MF2 MF1 14A
+5V FB 4 68 14B
+5V 1 2 1 LM2576 L1 L2 MF-R110 MF-R050 69
IN 3K3 14C
9*10K 3 4 AGND OUT 2 70 14D
71

GND
150uH
1
2
3
4
5
6
7
8
9
10

R89 50uH 15Z


C23

ON
R96 72 15A
470nF 47K R32 D1 C27 73
1K C36 15B
1K BYV28 74
100uF/16V

5
15C
2200uF/16V 75 15D
4 5 76 16Z
3 6 AGND AGND AGND AGND 77 16A
2 7 RY1 78 16B
AGND D35 R108 79
C46

C58
C63
C67
C44

C54

G2RL-2 +5V 16C


1 8 TR2 +15V 80 16D
R36 C41 MPSA05 81 17Z
6x1nF 1K 10nF BZX55C12 47K 82 17A
D22 83 17B
D30 C94 D36 84 17C
GND DGND GND BAT85 85 17D
J401 1N4007 86
VccCtl 18Z
1 VccCtl 1K R37 2.2uF BAT85 87 18A
1K R40 /CPSon AGND 88
2 /InterlockOk 18B
3 /XrReady=/Interlock 10K R43 89 18C
1K R46 /XrayOn 90
4 XrayOffOR/XrayOff 18D
5 1K R52 91 19Z
+12V 92
6 19A
7 93 19B
1K R60 /XrReset C2 VccCtl 94
8 /ExtCtrl 19C
9 1K R64 95 19D
1uF /InterlockOk KVRefMon 96

4
10 +24V 20Z
11 R14 SGND 97 20A
+5Vout 47 R66 /XonAcq KVRef 3 + IC2A 98
12 R6 R97 20B
13
47 R67 OverKV 1 99 20C
47 R69 LowKV 1K 2 - +5V +12V 470 100
14 C11 OP484 3K3/0.1% 20D
47 R71 OverLoad C3 101
15
47 R75 I_FilLim 10nF R10 D15 C15 102
21Z
R1 R98
11

16 BAT85 21A
17
47 R76 OverMA 1K/0.1% 10nF 103 21B
47 R39 /LkILim SGND S1 R38 1uF 104
18 1K C34 Reset 0 21C
19
47 R41 PSFaultOR/PSFault 1M AGND 105 21D
AGND AGND AGND BAT85 106

4
20 AGND 22Z
47 R49 HVInvFlt -12V 1uF R15 107
21
SGND C43 3 + IC3A 108
22A
22 R30 D33 R8 22B
47 R58 FilInvFlt MARef 5 + IC2B DGND 100nF 1 D34 KVMonFb 109
23 R21 10K 22C
+24V 7 C12 2 - 110

1
24 1K C31 6 - DGND OP484 BAT85 100 C19 111
22D
25 GND R90 3K3/0.1% 100pF 23Z
KVRef 10nF OP484 11.0592MHz R3 10nF 112
D16

11
26 C16 C65 IC9 23A
KVCommon R57 R16 IC6A 113
27 SGND MARef 1K XrLamp 1K/0.1% BAT85 10nF DGND 32 43 /InterlockOk SGND 114
23B
28 R26 74HC14 10pF XTAL1 P0.0/AD0 33K 23C
29
MACommon TR1 SGND X1 P0.1/AD1 44 VccCtl R12 SGND 115 23D
SGND 1K 33 45 FilInvFlt 10K 116
MPSA05

2
30 AGND 1K C66 10pF XTAL2 P0.2/AD2 HVInvFlt C35 24Z
31 R73 AGND AGND AGND DGND P0.3/AD3 46 117 24A
AGND AGND KVRegFlt 15 49 LowKV -12V 1uF 118
32 PowerRef RESET P0.4/AD4 Arcs 24B
33 47 R20 P0.5/AD5 50 SGND 119 24C
PowerCom PowerRef 10 + IC2C 1 51 /XonAcq AGND 120
34 SGND I_FilRef
R22 P1.0/ADC0/T2 P0.6/AD6 /LkILim 24D
35 8 2 P1.1/ADC1/T2EX P0.7/AD7 52 R23 121 25Z
I_FilCom 1K 9 - 3 5 + IC3B 122
36 SGND R44 C25 3K3/0.1% P1.2/ADC2 R24 25A
MARegFlt OP484 4 28 C26 7 I_FilMon 123
37 KVMonFb R17 D17 C17 /CPSon 11
P1.3/ADC3 P2.0/A08/A16
29 OverKV 100K 6 - 124
25B
38 47 10nF BAT85 P1.4/ADC4 P2.1/A09/A17 1uF 100 25C
MAMonFb
R27 1K/0.1% 10nF /XrReset 12 30 OverLoad OP484 125
39 /ExtCtrl P1.5/ADC5/SS P2.2/A10/A18 OverMA XsnR1+ 25D
40 R62 SGND 13 P1.6/ADC6 P2.3/A11/A19 31 R28 C20 126 26Z
+5Vout Arcs 14 36 I_FilLim SGND XsnR1- 127
41
KVMonRef KVRef 1K AGND AGND AGND
P1.7/ADC7 P2.4/A12/A20
37 OpenFil 10nF 128
26A
42
MAMonRef MARef 47 16
P2.5/A13/A21
38 XrLamp R18 1K5 XsnT1+ 129
26B
43 I_FilRef
R5 IC2D P3.0/RXD P2.6/A14/A22 PSFaultOR/PSFault XsnT1- 26C
44 12 + R7 17 P3.1/TXD P2.7/A15/A23 39 470 SGND 130 26D
+5Vout 14 /XrayOn 18 131
45 +15Vout I_FilMon 1K XrayOffOR/XrayOff P3.2/INT0 27Z
46 13 - 19 P3.3/INT1/MISO ALE 42 R59 +5V +5V 132 27A
MonCom C8 OP484 3K3/0.1% 22 40 SGND XpscDisplay 133
47 SGND 10nF R11 D18 C18 23
P3.4/T0 EA/VPP
DisplayRd2 134
27B
48 -15Vout BAT85 P3.5/T1/CONVST 10K 27C
R77 R2 1K/0.1% 10nF MARegFlt 24 135
49 GND OpenFil SGND KVRegFlt 25
P3.6/WR
BAT85 FilPwmH DGND 136
27D
50 P3.7/RD 28Z
47 DAC0 9 R31 FilPwmL 137 28A
SUB-D 50PF 1K AGND AGND AGND 26 10 10 + IC3C D32 R25 HvPwmH 138
SCLOCK DAC1 D31 MAMonFb 28B
27 SDATA/MOSI 8 HvPwmL 139 28C
R100 +5V +5V 10K C32 9 - FilamentCurrent 140
41 OP484 BAT85 100 /AdcCs3Ext 141
28D
MARefMon PSEN 100pF C22 29Z
0 JP1 C93 R29 R102 10nF /AdcCs2Ext 142 29A
EnProg C52 5 AVDD 100nF /AdcCs1Ext 143 29B
R99 100nF 6 AGND SGND 0 SdataOut 144 29C
7 33K SGND Sclk 145
KVRefMon 8
CREF
DGND R19 AdcSars 146
29D
0 VREF 30Z
AGND 10K SdataIn 147
C82

30A
C81
C42
C45
C55
C61
C62
C75
C51
C64

R101
C73
C79
C68
C69
C71

ADUC812 CurrentControl 148 30B


R109 +12V 470 CurrentLimit 149 30C
10K SGND PowerOn 150 30D
15x1nF R53 C47 EmissionCurrent+ 151
R47 MARefMon 31Z
1K 100nF Vref+2V5 HighVoltage 152 31A
1K EmissionCurrent- 153 31B
R92 Vref+5V 154 31C
GND R61 R63 10K/0.1% Vref+2V 155 31D
C92 C56 InputCurrent 156 32Z
0 1uF 100nF InputVoltage 157 32A
DGND 0 TGND Z2 R93 Mon400V 158
LM385 10K/0.1% 32B
R84 R65 159 32C
160 32D
0 0 SGND AGND AGND AGND AGND
GND
AGND
GND DIN41612

+5V +5V
+5V
R45 R42
+5V 1K 1K
D6 +5V R68
D9 NotOk/Red 470
C1 C9 Ok/Green IC6B

3
100nF IC11A XsnR1+
16

1uF XRF Configuration:


IC1 4 3
TGND 1 W1 16 R70 Remove R99 & R100
C7
2 1uF 74HC14 1 2 Internal 1 100
DC/DC 3 3 4 15 XsnR1- XRD Configuration:
C29 CONV. 4 IC6C External Remove R98 & R102
6 C14 4
5 6 5 DGND R72
TGND 1uF 1uF 14 470
74HC14 2
13
P108 14 11 DGND Mod: 4 OM3926 16.08.05 PPa 7 OM3953 21.02.06 PPa 10 OM4303 26.11.12 JSo - PPa
5 5 OM3865 08.11.05 PPa 8
9 XsnT1+ OM3991 28.11.06 PPa
8922A
6

4 TGND 7 10 Accessories 6 Addendum OM3865 06.12.05 PPa 9 OM4128 06.08.09 PPa


8 R74
3 DGND 1K
XsnT1- Remplacé par:
7
2
13 12
XPSR, X-ray Power Supply Retrofit Echelle Dessiné 28.05.01 PPa
6 Heatsink Heatsink Heatsink
1 8 9 Contrôlé
Remplace:
COM1 C24 C28 C13 C10 Heatsink Heatsink Heatsink Liste S703125 Schémas S112066 Vu
1nF 1nF 1nF 1nF DS14C232CN
15

Ensemble ---------- Circuit S920077 Groupe Technique:


Appareil XRF / XRD Implantation S112072 Ingénieur T.Maulaz 9850
TGND TGND TGND TGND Heatsink Heatsink Heatsink
TGND Groupe No
ThermoFisher S112066
No Mod.

Heatsink
S C I E N T I F I C
Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 10
+24V C12
P1
+12V
1 3 220pF +200VRTM
2 4 FAN R68 C64
WAGO 2P AGND +5V
10nF

4
10K ToHvTransfo
J906 PGND R67 IC5A 3
R7 10K OP484 - 9 1
+
D20 IC5D - 13 8 2 R18
R76 -
1 /PwmTrip 14 R78
+ 10 10K OP484 10K
1Z Vref+2V
2 + 12 IC5C C66 T1A T1B

11
1A 1K XpsmHvTemp
1B 3 GND BAT85 OP484 47pF
4 MF6 D21 1K C67
1C BAT85 C65
1D 5 +24Vin C11 10nF C35 RV2
6 10nF 10nF R66 -12V AGND 100nF 4K7/25

4
2Z
2A 7 MF-R110 10K AGND
8 PGND DGND AGND T1
2B +12Vin R64
2C 9 +12V AGND AGND
2D 10 AGND R69 HvTransfo
11 R65 1K 30K
3Z AGND -12Vin IC5B
3A 12 5 10K C72 Not on the Board
13 7
+ AGND IC4B 5 C63

2
3B + AGND
3C 14 - 6 7 10nF
15 +5Vin OP484 6 2.7nF
3D -
16 OP484 AGND

4
4Z DGND
4A 17 R8 R41 R63
18 IC4A 3
4B R59 +
4C 19 1K 1
20 2 1K 20K
4D
21 1K OP484
-
5Z HV2 R28
22 D11 R48

11
5A Arc C44
5B 23
24 C40 1nF 200M
5C 100 18x 4n7/3kV
5D 25 1N4148 10nF
6Z 26 -12V
6A 27 Line1 R21 C34 AGND GAP6
28 Line1 C33 100K D12 R35 M51-C90
6B
29 10uF BZX85C5V6 AGND C1 C2 C3 C4
6C 10nF
6D 30 4x 20nF/12kV
31 IC4C 10 1K
7Z + AGND
7A 32 AGND 8 GND
7B 33 AGND AGND AGND - 9
7C 34 OP484 D1 D2 D3 D4 D5 D6 D7 D8 D9 D10
35 MF5 R13
7D IC4D + R42 HV1
8Z 36 +5V R60 12
8A 37 Line2 14 1K 20x Z100UFG
38 Line2 13 MF-R050 15/3W GAP1
8B
39 D22 1K -
A81-C230 D1' D2' D3' D4' D5' D7' D8' D9' D10'
8C BAT85 OP484
40 C45 Wired resistors
8D JP1
9Z 41 R26 R49 R36 1uF GAP3
42 EmissionCurrent+ M51-C90 GND C6 C7 C8 C9 C10 5x 20nF/12kV R1 R2 R3 R4 R5 R6 R7 R8
9A S
9B 43 100 1K L
44 100 D13 AGND
9C C
9D 45 BAT85 8x 100/9W
10Z 46 R11 GND D13' D14' D15' D16' D17' D18' D19' D20' D21' D22' Fed.Standard
47 EmissionCurrent- D10
10A
10B 48 100 1N5347
10C 49 AGND C32 20x Z100UFG
10D 50 C25 R40 R34 R30 R25 100uF/35V
11Z 51 10nF C9 D13 D14 D15 D16 D17 D18 D19 D20 D21 D22
11A 52 HP:4x100
11B 53 +400VRTM 1nF
54 AGND LP:1x100
11C
11D 55 R55 C12 C13 C14 C15
12Z 56 4x 20nF/12kV
12A 57 AGND
58 53k3 HT Output board S940188
12B
12C 59 +12V
12D 60 R14 DV1 5x Multiplier board S930191
61 OP484 - 6
13Z
13A 62 +5V C62 7 56K
13B 63 + 5 GAP2 Ra1
64 10nF IC3B A81-C230
13C R58
13D 65 D14
66 AGND 1K

4
14Z BAT85
14A 67 AGND GAP4 GND GND
68 IC3A 3 D8 C1
14B HvDividerDifference R38 R32 + M51-C90
14C 69 1 1N5347 10uF 5x 200M/15kV 5x 470pF/15kV Divisor board S940189
70 2 R61 Wired resistors
14D
71 1K 1K OP484 -
10K
15Z +200VRTM Vref+2V R17 R16 R15 R14 R13
72 D16 R12 R11 R10 R9

11
15A
15B 73 C43 BAT85 C39
15C 74 10nF C74 C10 AGND GND AGND
75 10nF 10nF 4x 1K/9W
15D
16Z 76 -12V 1nF C21 C20 C19 C18 C17
16A 77 AGND
16B 78 AGND R28 AGND R23 R47
79 AGND R27 C27
16C
16D 80 10K 1K 53k3 33M/9kV 470pF/15kV
17Z 81 +5V
17A 82 Interface board
17B 83 D19
17C 84 S930220 R26 R25 R24 R23 R22
85 IC3D 12 BAT85 R21 R20 R19 R18
17D + AGND OP484 R15 DV2
18Z 86 14 R62 R51
- 9
87 13 HighVoltage 8
18A
88 OP484
- + 10 56K GAP3 Ra2 4x 1K/9W
18B 1K 100 IC3C
18C 89 A81-C230 C26 C25 C24 C23 C22
18D 90 R6 D18 Wired resistors
19Z 91 BAT85 5x 200M/15kV 5x 470pF/15kV Divisor board S940189
19A 92 1K AGND GAP5 GND GND
19B 93 0VRTM D9 C8
94 1N5347 M51-C90
19C
95 AGND
10uF Multiplier Box S703952 (not on the board)
19D
20Z 96
20A 97
20B 98
20C 99 AGND GND AGND
20D 100
21Z 101
21A 102
21B 103 GND
21C 104
21D 105
22Z 106
22A 107
22B 108
22C 109
22D 110 +5V
23Z 111
112 ToHvTransfo T2B
23A ToFilamentTransfo T2
23B 113
114 R4
23C
23D 115 10K 2
24Z 116
117 XpsmFilamentTemp 17
24A
24B 118 2
24C 119 C17 RV1 T2A T3
120 100nF 4K7/25 +200VRTM
24D
25Z 121 Not on the Board
25A 122 AGND FilamentTransfo
25B 123 AGND 30
25C 124
25D 125
26Z 126 T2D T2C T3A T3B
26A 127
128 ToFilamentTransfo D2 D1 D6 D23
26B
26C 129
26D 130
27Z 131
27A 132
133 D3 D4 D5 D7
27B
27C 134
27D 135 +5V 4xBYV26 4xBYV26
28Z 136 FilPwmH
28A 137 FilPwmL AGND AGND
28B 138 HvPwmH D17
28C 139 HvPwmL
140 FilamentCurrent BAT85 R2
28D /AdcCs3
29Z 141 +5V +12V -12V R45 R24 1K3 R22 1V/A
142 /AdcCs2 FilamentVoltage 0.242V/Vfil FilamentCurrent 1V/A
29A
29B 143 /AdcCs1 1K 1K
144 SdataOut C41 100 R5
29C Sclk R46 C56
29D 145 R31 R43 100uF/35V D15 39
146 AdcSars 2K7 18K 33K BAT85
30Z SdataIn 1uF
30A 147 R3
30B 148 CurrentControl +5V C69 C70 C71 AGND 430
149 CurrentLimit 8x8bit ADC R27 100nF R37 100nF R50 100nF AGND AGND
30C 10K
30D 150 PowerOn Status AD converter 10K 10K
151 EmissionCurrent+ /AdcCs3 R70 100 AGND AGND
31Z HighVoltage C38 IC2
31A 152
153 EmissionCurrent- SdataOut R71 100 10nF 20 1 AGND AGND AGND AGND AGND
31B Vref+5V VCC CH0
31C 154 19 V+ CH1 2
155 Vref+2V Sclk R72 100 AGND 18 3 Vref+5V
31D CS CH2 Arc
32Z 156 InputCurrent 17 DI CH3 4
157 InputVoltage AdcSars R73 100 16 5 HvDividerDifference
32A CLK CH4 FilamentVoltage
32B 158 Mon400V 15 SARS CH5 6
159 SdataIn R74 100 14 7 XpsmFilamentTemp
32C DO CH6 XpsmHvTemp C68
32D 160 +5V 13 SE CH7 8
GND Vref+5V R54 12 9 10nF
VREF COM
10 11 AGND DGND 10 R75
C61 C53 ADC0838
DIN41612 10uF 1nF AGND 0
DGND AGND
AGND AGND

Accessories
TP3 TP1
L1 +12V L3 +5V
MF1 MF3 Hole Hole Hole Hole
+12Vin +5Vin Mod: 1 MOD23923 16.11.05 PPa 4 OM4204 27.04.11 PPa
+12V +5V Hole Hole Hole Hole 2 Add. OM3865 21.02.06 PPa 5 OM4204 Add.2 06.07.11 PPa
MF-R110 C4 C7 TP4 MF-R110 C2 C5 Hole Hole Hole Hole
220uF 100nF 220uF 100nF 3 OM4187 11.10.10 PPa
R1 AGND
AGND Hole Hole Hole Hole
22.07.02 MBz Remplacé par:
GND 0 C3 C6 TP2 Hole Hole Hole Hole X-ray Power Supply Mutiplier Assembly Echelle Dessiné

220uF 100nF -12V


-12Vin MF2 L2
-12V Board XPSMD X-ray Power Supply Multiplier for XRD
Hole Hole Hole
XRD version (Negative High Voltage) Contrôlé
Remplace:
Hole Hole Hole Hole
MF-R110 Vu
Liste S703953 Schémas S112247
Assembly : S703719 Ensemble ---------- Circuit S920087 Groupe Technique:
Hole Hole Hole Hole Appareil Xtra Implantation S112248 Ingénieur T.Maulaz 9850
PCB : S920087 Groupe No

Layout : S112248
GND
ThermoFisher S112247
No Mod.

Fid Fid Fid Fid S C I E N T I F I C


Thermo Fisher Scientific - En Vallaire Ouest C - Case postale -1024 Ecublens - Switzerland --- Page: 1/1 5
9

CATALOGUE
Catalogue Chapter 9

9 CATALOGUE
Foreword
Some spare parts are marked with a blue label containing the following text:
This component was developed and tested by Thermo Fisher Scientific to comply with our
instrument's technical specifications in order to guarantee and maintain analytical performances
as well as safety standards.

For safety reasons, this part must be provided only through Thermo distribution companies. Any
attempt to fit a part provided by another supplier than Thermo Fisher Scientific may result in
damages and loss of performances, and thus render Thermo Fisher Scientific warranty and
responsibility void.

Avant-propos
Certaines pièces de rechange sont marquées par une étiquette bleue comportant le texte suivant:
This component was developed and tested by Thermo Fisher Scientific to comply with our
instrument's technical specifications in order to guarantee and maintain analytical performances
as well as safety standards.

For safety reasons, this part must be provided only through Thermo distribution companies. Any
attempt to fit a part provided by another supplier than Thermo Fisher Scientific may result in
damages and loss of performances, and thus render Thermo Fisher Scientific warranty and
responsibility void.

La traduction française de ce texte est la suivante:


Ce composant a été développé et testé par Thermo Fisher Scientific pour répondre aux
spécifications techniques de nos appareils afin de garantir et conserver les performances
analytiques ainsi que les normes de sécurité.

Pour des raisons de sécurité, cette pièce doit être fournie par le réseau de distribution Thermo
Electron. Toute tentative de montage d'une pièce provenant d'une autre source de distribution que
celle de Thermo Fisher Scientific pourrait engendrer des dommages et une perte de performances,
et par conséquence rendre la responsabilité et la garantie de Thermo Fisher Scientific nulle.

Vorwort
Einige Ersatzteile werden mit einem blauen Aufkleber geliefert, der folgenden Text enthält:
This component was developed and tested by Thermo Fisher Scientific to comply with our
instrument's technical specifications in order to guarantee and maintain analytical performances
as well as safety standards.

For safety reasons, this part must be provided only through Thermo distribution companies. Any
attempt to fit a part provided by another supplier than Thermo Fisher Scientific may result in
damages and loss of performances, and thus render Thermo Fisher Scientific warranty and
responsibility void.

Die deutsche Übersetzung lautet wie folgt:


Dieses Teil wurde von Thermo Fisher Scientific speziell für den Einsatz in unseren Geräten
entwickelt und getestet, um sowohl die analytischen Leistungen zu gewährleisten als auch die
Einhaltung der Sicherheitsvorschriften zu garantieren.

Aus Sicherheitsgründen muss dieses Teil über Thermo-Unternehmen oder –Vertretungen bezogen
werden. Jeglicher Versuch, Teile aus anderen Bezugsquellen einzubauen, kann zu Schäden und
Leistungsverlusten führen und damit die Thermo Fisher Scientific Garantie und Verantwortung
nichtig machen.

AA83678 Service Manual ARL GenX 9-1


Chapter 9 Catalogue

Spare parts list


Liste de pièces de rechange
Ersatzteil-Liste

Description
List
Mod. Description Note
KXx…
Beschreibung

GENX00 1 GenX I 3.6 kW 60kV/120m

GENX10 1 GenX II 3.6 kW 70kV/120mA

GENX20 1 GenX III 4.2kW 60kV/120mA

GENX30 1 GenX IV 4.2 kW 70kV/140mA

GENX40 0 GenX V 1.2 kW 60kV/60mA

GENX50 1 GenX R 3.6 kW 70kV/120mA

GENX60 1 GenX R 4.2 kW 70kV/140mA

GENX70 1 GenX VI 2.5 kW 60kV/100mA

GENX80 2 GenXRD 3.0 kW

GENX90 2 GenXRD RET 3.0 kW

GENX100 3 GenXRDF18, 1.8 kW

GENX110 2 GenXRDV18, 1.8 kW

GENX120 0 GenX VII 600 W 60kV/30mA

GENX130 1 GenX VIII PFX 4.2 kW 70kV/140mA

GENX140 1 GenX IX PFX 2.5 kW 60kV/100mA

GENX150 2 GenX X PFX 4.2 kW 60kV/120mA

GENX151 2 GenX XI PFX 1.5 kW 60kV/70mA

GENX152 0 GenX XII PFX 2.5 kW 60kV/100mA for X-Ray Tube


UCCO30

GENX153 0 GenX XIII PFX 4.0 kW 60kV/140mA for X-Ray Tube


UCCO30

GENX160 0 GenX Generator ARL 9900 Version 2020 1.5 kW


60kV/100mA

9-2 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX161 0 GenX Generator ARL 9900 Version 2020 2.5 kW


60kV/140mA

GENX162 0 GenX Generator ARL 9900 Version 2020 4.2 kW


60kV/120mA
Toolkit and miscellaneous
GENX199 3 Outillage et divers
Werkzeuge und Diverses

Remark :

A Recommended spare part

B Full spare part list

C Consumable spare part

AA83678 Service Manual ARL GenX 9-3


Chapter 9 Catalogue

KXXGENX00-01

LIST GENX00

9-4 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX I GENERATOR 3.6KW 60KV/120MA


GENX I GENERATEUR 3.6KW 60KV/120MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Positions 6 and 8 updated
1 Positions 6 et 8 mises à jour 4112
Positionen 6 und 8 auf den letzten Stand gebracht

Note :

AA83678 Service Manual ARL GenX 9-5


Chapter 9 Catalogue

Figure 9.1

9-6 Service Manual ARL GenX AA83678


Catalogue Chapter 9

N° P/N Qty Description Type Note Maintenance:


Description greyed items
Beschreibung have to be
considered
1 S703690T 1 GenX I 3.6kW 60kV/120mA B
GenX I 3.6kW 60kV/120mA
………..
2 S703722T 1 XPSC3660 I board A
Carte XPSC3660 I
…………..
3 S703123T 1 XPSI board A
Carte XPSI
………..
4 S703126T 1 XPSP board A
Carte XPSP
…………..
5 S703758T 1 Multiplier assembly A
Ensemble Multiplicateur
………….
6 S703838 1 Kit I, fans GENXRFHP C 1 Replace fans after
Kit I ventilateurs GENXRFHP 2 years
……………
7 A034398 2 Fan “small” for XPSI & XPSP C Replace fans after
Ventilateur “petit” pour XPSI 2 years
et XPSP
…………….
8 S704791 2 Fan “big” for XPSI & XPSP C Replace fans after
Ventilateur “grand” pour XPSI 2 years
et XPSP
…………….
9 A039356 2 Fan for Multiplier assembly. C Replace fans after
Ventilateur pour ensemble 2 years
multiplicateur
……………
10 A040026 4 Fast fitting female
Raccord rapide femelle
……………
11 A040027 2 Fast fitting male
Raccord rapide male
……………
12 A040232 2 Fixed Fitting male 2
Raccord fixe male
……………
13 AA22511 8 Collar D9-D16
Collier D9-D16
…………….

AA83678 Service Manual ARL GenX 9-7


Chapter 9 Catalogue

14 A040528 2 Pipe D16/10 3


Tuyau D16/10
…………….

Note :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-8 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX10-01

LIST GENX10

AA83678 Service Manual ARL GenX 9-9


Chapter 9 Catalogue

GENX II GENERATOR 3.6KW 70KV/120MA


GENX II GENERATEUR 3.6KW 70KV/120MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Positions 6 and 8 updated
1 Positions 6 et 8 mises à jour 4112
Positionen 6 und 8 auf den letzten Stand gebracht

Note :

9-10 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.2

AA83678 Service Manual ARL GenX 9-11


Chapter 9 Catalogue

N° P/N Qty Description Type Note Maintenance:


Description greyed items
Beschreibung have to be
considered
1 S703680T 1 GenX II 3.6kW 70kV/120mA B
GenX II 3.6kW 70kV/120mA
………..
2 S703723T 1 XPSC3670 II board A
Carte XPSC3670 II
…………..
3 S703123T 1 XPSI board A
Carte XPSI
………..
4 S703126T 1 XPSP board A
Carte XPSP
…………..
5 S703758T 1 Multiplier assembly A
Ensemble Multiplicateur
………….
6 S703838 1 Kit I, fans, GENXRFHP C Replace fans after
Kit I ventilateurs, GENXRFHP 2 years
……………
7 A034398 2 Fan “small” for XPSI & XPSP C Replace fans after
Ventilateur “petit” pour XPSI 2 years
et XPSP
…………….
2 Fan “big” for XPSI & XPSP C Replace fans after
8 S704791
Ventilateur “grand” pour XPSI 2 years
et XPSP
…………….
9 A039356 2 Fan for Multiplier assembly. C Replace fans after
Ventilateur pour ensemble 2 years
multiplicateur
……………
10 A040026 4 Fast fitting female
Raccord rapide femelle
……………
11 A040027 2 Fast fitting male
Raccord rapide male
……………
12 A040232 2 Fixed Fitting male 2
Raccord fixe male
……………
13 AA22511 8 Collar D9-D16
Collier D9-D16
…………….
14 A040528 2 Pipe D16/10 3
Tuyau D16/10
……………
9-12 Service Manual ARL GenX AA83678
Catalogue Chapter 9

Note :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-13


Chapter 9 Catalogue

KXXGENX20-01

LIST GENX20

9-14 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX III GENERATOR 4.2KW 60KV/120MA


GENX III GENERATEUR 4.2KW 60KV/120MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Positions 6 and 8 updated
1 Positions 6 et 8 mises à jour 4112
Positionen 6 und 8 auf den letzten Stand gebracht

Note :

AA83678 Service Manual ARL GenX 9-15


Chapter 9 Catalogue

Figure 9.3

9-16 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX III 4.2kW 60kV/140mA
1 S703681T 1 GenX III 4.2kW 60kV/140mA B
………..
XPSC4260 III board
2 S703724T 1 Carte XPSC4260 III A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

AA83678 Service Manual ARL GenX 9-17


Chapter 9 Catalogue

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Note :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-18 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX30-01

LIST GENX30

AA83678 Service Manual ARL GenX 9-19


Chapter 9 Catalogue

GENX IV GENERATOR 4.2KW 70KV/140MA


GENX IV GENERATEUR 4.2KW 70KV/140MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Positions 6 and 8 updated
1 Positions 6 et 8 mises à jour 4112
Positionen 6 und 8 auf den letzten Stand gebracht

Note :

9-20 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.4

AA83678 Service Manual ARL GenX 9-21


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX IV 4.2kW 70kV/140mA
1 S703682T 1 GenX IV 4.2kW 70kV/140mA B
………..
XPSC4270 IV board
2 S703725T 1 Carte XPSC4270 IV A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

9-22 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Note :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-23


Chapter 9 Catalogue

KXXGENX40-00

LIST GENX40

9-24 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX V GENERATOR 1.2KW 60KV/60MA


GENX V GENERATEUR 1.2KW 60KV/60MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

Note :

AA83678 Service Manual ARL GenX 9-25


Chapter 9 Catalogue

Figure 9.5

9-26 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N P/N Qty Description Type Note greyed items
° Beschreibung have to be
considered
GenX V 1.2kW 60kV/60mA
1 S703683T 1 GenX V 1.2kW 60kV/60mA B
………..
XPSC1260 V board
2 S703726T 1 Carte XPSC1260 V A
…………..
XPSI board
3 S703778T 1 Carte XPSI A
………..
XPSP board
4 S703779T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit II, fans GENXRFLP Replace fans after
6 S703839 1 Kit II ventilateurs GENXRFLP C 1 2 years
……………
Fan for XPSI & XPSP Replace fans after
7 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly. Replace fans after
8 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………

Note :
1 Positions 7,8 are included in this item.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-27


Chapter 9 Catalogue

KXXGENX50-01

LIST GENX50

9-28 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX RET GENERATOR 3.6KW 70KV/120MA


GENX RET GENERATEUR 3.6KW 70KV/120MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Positions 7 and 9 updated
1 Positions 7 et 9 mises à jour 4112
Positionen 7 und 9 auf den letzten Stand gebracht

Note :

AA83678 Service Manual ARL GenX 9-29


Chapter 9 Catalogue

Figure 9.6

9-30 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX R 3.6kW 70kV/120mA
1 S703685T 1 GenX R 3.6kW 70kV/120mA B
………..
XPSCRET PR board
2 S703784T 1 Carte XPSCRET PR A
…………..
XPSR36 PR board
3 S703125T A
Carte XPSR36 PR
XPSI board
4 S703123T 1 Carte XPSI A
………..
XPSP board
5 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
6 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
7 S703838 1 Kit I ventilateurs GENXRFHP C 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
8 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
S704791 Fan “big” for XPSI & XPSP Replace fans after
9 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
10 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
11 A040026 4 Raccord rapide femelle
……………
Fast fitting male
12 A040027 2 Raccord rapide male
……………
Fixed Fitting male
13 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
14 AA22511 8 Collier D9-D16
………………

AA83678 Service Manual ARL GenX 9-31


Chapter 9 Catalogue

Pipe D16/10
15 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-32 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX60-01

LIST GENX60

AA83678 Service Manual ARL GenX 9-33


Chapter 9 Catalogue

GENX RET GENERATOR 4.2KW 70KV/140MA


GENX RET GENERATEUR 4.2KW 70KV/140MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Positions 7 and 9 updated
1 Positions 7 et 9 mises à jour 4112
Positionen 7 und 9 auf den letzten Stand gebracht

Note :

9-34 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.7

AA83678 Service Manual ARL GenX 9-35


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX R 4.2kW 70kV/140mA
1 S703687T 1 GenX R 4.2kW 70kV/140mA B
………..
XPSCRET board
2 S703784T 1 Carte XPSCRET A
…………..
XPSR42 PR board
3 S703785T A
Carte XPSR42 PR
XPSI board
4 S703123T 1 Carte XPSI A
………..
XPSP board
5 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
6 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
7 S703838 1 Kit I ventilateurs GENXRFHP C 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
8 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
9 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
10 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
11 A040026 4 Raccord rapide femelle
……………
Fast fitting male
12 A040027 2 Raccord rapide male
……………
Fixed Fitting male
13 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
14 AA22511 8 Collier D9-D16
……………

9-36 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Pipe D16/10
15 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-37


Chapter 9 Catalogue

KXXGENX70-01

LIST GENX70

9-38 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX VI GENERATOR 2.5KW 60KV/100MA


GENX VI GENERATEUR 2.5KW 60KV/100MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

0 23.12.05 Introduction GEN’X 2.5 kW 3942

Positions 6 and 8 updated


1 Positions 6 et 8 mises à jour 4112
Positionen 6 und 8 auf den letzten Stand gebracht

Note :

AA83678 Service Manual ARL GenX 9-39


Chapter 9 Catalogue

Figure 9.8

9-40 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX VI 2.5kW 60kV/100mA
1 S704025T 1 GenX VI 2.5kW 60kV/100mA B
………..
XPSC2560 VI board
2 S704026T 1 Carte XPSC2560 VI A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

AA83678 Service Manual ARL GenX 9-41


Chapter 9 Catalogue

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-42 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX80-02

LIST GENX80

AA83678 Service Manual ARL GenX 9-43


Chapter 9 Catalogue

GENXRD GENERATOR
GENXRD GENERATEUR
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

0 Introduction GEN’X for XRD X’TRA

Positions 7 and 9 updated


1 Positions 7 et 9 mises à jour
Positionen 7 und 9 auf den letzten Stand gebracht
Position 1 updated
2 Position 1 mise à jour
Position 1 auf den letzten Stand gebracht

Note :

9-44 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.9

AA83678 Service Manual ARL GenX 9-45


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenXRD 3.0kW
1 S703720T 1 GenXRD 3.0kW B
………..
XPSCD 3060 board
2 S703721T 1 Carte XPSCD 3060 A
…………..
XPSRD3060 board
3 S703876T A
Carte XPSRD3060
XPSI board
4 S703778T 1 Carte XPSI A
………..
XPSP board
5 S703779T 1 Carte XPSP A
…………..
Multiplier assembly XRD 3
6 S703953T 1 kW A
Ensemble Multiplicateur XRD
3kW
………….
Kit II, fans GENXRDAIR Replace fans after
7 S705115 1 Kit II ventilateurs C 1 2 years
GENXRDAIR
……………
Fan for XPSI & XPSP Replace fans after
8 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly Replace fans after
9 A040863 1 XRD 3KW. C 2 years
Ventilateur pour ensemble
multiplicateur XRD 3KW
……………

Note :
1 Positions 7,8,9 are included in this item.

Remark / Remarque / Bemerkung :

9-46 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX90-02

LIST GENX90

AA83678 Service Manual ARL GenX 9-47


Chapter 9 Catalogue

GENXRD RET GENERATOR


GENXRD RET GENERATEUR
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

0 Introduction GEN’X RETROFIT for XRD X’TRA

Positions 7 and 9 updated


1 Positions 7 et 9 mises à jour
Positionen 7 und 9 auf den letzten Stand gebracht
Position 1 updated
2 Position 1 mise à jour
Position 1 auf den letzten Stand gebracht

Note :

9-48 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.10

AA83678 Service Manual ARL GenX 9-49


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenXRD RET 3.0kW
1 S703720T 1 GenXRD RET 3.0kW B
………..
XPSCDR 3060 board
2 S703721T 1 Carte XPSCDR 3060 A
…………..
XPSRD3060 board
3 S703876T A
Carte XPSRD3060
XPSI board
4 S703778T 1 Carte XPSI A
………..
XPSP board
5 S703779T 1 Carte XPSP A
…………..
Multiplier assembly XRD 3 kW
6 S703953T 1 Ensemble Multiplicateur XRD A
3kW
………….
Kit II, fans GENXRDAIR Replace fans after
7 S705115 1 Kit II ventilateurs C 1 2 years
GENXRDAIR
……………
Fan for XPSI & XPSP Replace fans after
8 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly Replace fans after
9 A040863 1 XRD 3KW. C 2 years
Ventilateur pour ensemble
multiplicateur XRD 3KW
……………

Note :
1 Positions 8,9 are included in this item.

Remark / Remarque / Bemerkung :

9-50 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX100-03

LIST GENX100

AA83678 Service Manual ARL GenX 9-51


Chapter 9 Catalogue

GENXRDF18 GENERATOR
GENXRDF18 GENERATEUR
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

0 Introduction GEN’X for NEXRD 4010

Positions 6 and 8 updated


1 Positions 6 et 8 mises à jour 4112
Positionen 6 und 8 auf den letzten Stand gebracht
Positions 6 and 9 updated
2 Positions 6 et 9 mises à jour
Positionen 6 und 9 auf den letzten Stand gebracht
Position 6 updated
3 Position 6 mise à jour
Position 6 auf den letzten Stand gebracht

Note :

9-52 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.11

AA83678 Service Manual ARL GenX 9-53


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenXRDF18, 1.8 kW
1 S704385T 1 GenXRDF18, 1.8 kW B
………..
XPSC XRDF18 board
2 S704387T 1 Carte XPSC XRDF18 A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly XRD 3 kW
5 S703953T 1 Ensemble Multiplicateur XRD A
3kW
………….
Kit I, fans GENXRDWATER Replace fans after
6 S705114 1 Kit I ventilateurs C 1 2 years
GENXRDWATER
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A040863 1 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………

9-54 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7, 8, 9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-55


Chapter 9 Catalogue

KXXGENX110-02

LIST GENX110

9-56 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENXRDV18 GENERATOR
GENXRDV18 GENERATEUR
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

0 Introduction GEN’X for NEXRD 4010

Positions 6 and 8 updated


1 Positions 6 et 8 mises à jour 4112
Positionen 6 und 8 auf den letzten Stand gebracht
Positions 6 and 9 updated
2 Positions 6 et 9 mises à jour
Positionen 6 und 9 auf den letzten Stand gebracht
Position 6 updated
3 Positions 6 mise à jour
Position 6 auf den letzten Stand gebracht

Note :

AA83678 Service Manual ARL GenX 9-57


Chapter 9 Catalogue

Figure 9.12

9-58 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenXRDV18, 1.8kW
1 S704368T 1 GenXRDV18, 1.8kW B
………..
XPSC XRDV18 board
2 S704379T 1 Carte XPSC XRDV18 A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly XRD 3 kW
5 S703953T 1 Ensemble Multiplicateur XRD A
3kW
………….
Kit I, fans GENXRDWATER Replace fans after
6 S705114 1 Kit I ventilateurs C 1 2 years
GENXRDWATER
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A040863 1 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………

AA83678 Service Manual ARL GenX 9-59


Chapter 9 Catalogue

Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….
Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7, 8, 9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-60 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX120-00

LIST GENX120

AA83678 Service Manual ARL GenX 9-61


Chapter 9 Catalogue

GENX VII GENERATOR 600 W 60KV/30MA


GENX VII GENERATEUR 600 W 60KV/30MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

0 July 2009 Introduction Gen‘X 600 W ARL 9900 Series 4119

Note :

9-62 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.13

AA83678 Service Manual ARL GenX 9-63


Chapter 9 Catalogue

Description Maintenance:
N P/N Qty Description Type Note greyed items
° Beschreibung have to be
considered
GenX VII 600 W 60kV/30mA
1 S704818T 1 GenX VII 600 W 60kV/30mA B
………..
XPSC0660 VII board
2 S704819T 1 Carte XPSC0660 VII A
…………..
XPSI board
3 S703778T 1 Carte XPSI A
………..
XPSP board
4 S703779T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit II, fans GENXRFLP Replace fans after
6 S703839 1 Kit II ventilateurs GENXRFLP C 1 2 years
……………
Fan for XPSI & XPSP Replace fans after
7 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly. Replace fans after
8 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………

Note :
1 Positions 7,8 are included in this item.

Remark / Remarque / Bemerkung :

9-64 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX130-01

LIST GENX130

AA83678 Service Manual ARL GenX 9-65


Chapter 9 Catalogue

GENX VIII GENERATOR PERFORM’X 4.2KW 70KV/140MA


GENX VIII GENERATEUR PERFORM’X 4.2KW 70KV/140MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

1 Introduction Gen’X VIII ARL PERFORM’X Series

Note :

9-66 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.14

AA83678 Service Manual ARL GenX 9-67


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX VIII 4.2kW 70kV/140mA
1 S705076 1 GenX VIII 4.2kW 70kV/140mA B
………..
XPSC4270 board VIII
2 S704931T 1 Carte XPSC4270 VIII A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

9-68 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-69


Chapter 9 Catalogue

KXXGENX140-01

LIST GENX140

9-70 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX IX GENERATOR PERFORM’X 2.5KW 60KV/100MA


GENX IX GENERATEUR PERFORM’X 2.5KW 60KV/100MA
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

1 Introduction Gen’X IX ARL PERFORM’X Series

Note :

AA83678 Service Manual ARL GenX 9-71


Chapter 9 Catalogue

Figure 9.15

9-72 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX IX 2.5kW 60kV/100mA
1 S705075 1 GenX IX 2.5kW 60kV/100mA B
………..
XPSC2560LC IX board
2 S704930T 1 Carte XPSC2560LC IX A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

AA83678 Service Manual ARL GenX 9-73


Chapter 9 Catalogue

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-74 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX150-02

LIST GENX150

AA83678 Service Manual ARL GenX 9-75


Chapter 9 Catalogue

GENX X GENERATOR PERFORM’X 4.2KW 60KV/120MA


GENX X GENERATEUR PERFORM’X 4.2KW 60KV/120MA

From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

0 August 2011 Introduction Gen’X X ARL PERFORM‘X


Position 2 updated
1 Position 2 mise à jour
Position 2 auf den letzten Stand gebracht
Position 1 updated
2 Position 1 mise à jour
Position 1 auf den letzten Stand gebracht

Note :

9-76 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.16

AA83678 Service Manual ARL GenX 9-77


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX X 4.2kW 60kV/120mA
1 S705178 1 GenX X 4.2kW 60kV/120mA B
………..
XPSC4260LC board X
2 S705173 1 Carte XPSC4260LC X A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

9-78 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-79


Chapter 9 Catalogue

KXXGENX151-02

LIST GENX151

9-80 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX XI GENERATOR PERFORM’X 1.5KW 60KV/70MA


GENX XI GENERATEUR PERFORM’X 1.5KW 60KV/70MA

From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

0 Nov. 2011 Introduction Gen’X XI ARL PERFORM‘X Series


Position 6 updated
1 Position 6 mise à jour
Position 6 auf den letzten Stand gebracht

2 July 2018 PN numbers of items 3 and 4 updated

Note :

AA83678 Service Manual ARL GenX 9-81


Chapter 9 Catalogue

Figure 9.17

9-82 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX XI 1.5kW 60kV/70mA
1 S705074 1 GenX XI 1.5kW 60kV/70mA B
………..
XPSC1560LC board XI
2 S704929T 1 Carte XPSC1560LC XI A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit II, fans GENXRFHP Replace fans after
6 S703838 1 Kit II ventilateurs GENXRFHP C 1 2 years
……………
Fan for XPSI & XPSP Replace fans after
7 A040411 2 Ventilateur pour XPSI et C 2 years
XPSP
…………….
Fan for Multiplier assembly Replace fans after
8 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………

Note :
1 Positions 7,8 are included in this item.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-83


Chapter 9 Catalogue

KXXGENX152-00

LIST GENX152

9-84 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX XII GENERATOR PERFORM’X 2.5KW 60KV/100MA


for X-Ray Tube UCCO30
GENX XII GENERATEUR PERFORM’X 2.5KW 60KV/100MA
pour Tube Rayon-X UCCO30
From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

Note :

AA83678 Service Manual ARL GenX 9-85


Chapter 9 Catalogue

Figure 9.18

9-86 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX XII 2.5kW 60kV/100mA
1 S705219 1 for X-Ray Tube UCCO30 B
GenX XII 2.5kW 60kV/100mA
pour Tube rayon-X UCCO30
………..
XPSC2560LCU30 XII board
2 S705217 1 Carte XPSC2560LCU30 XII A
…………..
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

AA83678 Service Manual ARL GenX 9-87


Chapter 9 Catalogue

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-88 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX153-00

LIST GENX153

AA83678 Service Manual ARL GenX 9-89


Chapter 9 Catalogue

GENX XIII GENERATOR PERFORM’X 4.0KW 60KV/140MA


for X-Ray Tube UCCO30
GENX XIII GENERATEUR PERFORM’X 4.0KW 60KV/140MA
pour Tube Rayon-X UCCO30

From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung

Note :

9-90 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.19

AA83678 Service Manual ARL GenX 9-91


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX XIII 4.0kW 60kV/140mA
1 S705220 1 for X-Ray Tube UCCO30 B
GenX XIII 4.0kW 60kV/140mA
pour Tube rayon-X UCCO30
………..
XPSC4260LCU30 XIII board
2 S705218 1 Carte XIII XPSC4260LCU30 A
………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

9-92 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-93


Chapter 9 Catalogue

KXXGENX160-00

LIST GENX160

9-94 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX X GENERATOR ARL 9900 VERSION 2020 1.5KW


60KV/60MA
GENX X GENERATEUR ARL 9900 VERSION 2020
1.5KW 60KV/60MA

From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Introduction Gen’X 9900 from serial number 9953964
0 May 2020 including 9953165, 9953396, 9953835, 9953933, 9953941,
9953943.

Note :

AA83678 Service Manual ARL GenX 9-95


Chapter 9 Catalogue

Figure 9.20

9-96 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX 1.5kW 60kV/60mA
1 1226736 1 GenX 1.5kW 60kV/60mA B
………..
CARTE,XPSC,GENX 9900
2 1226732 1 1.5kW SP A

………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

AA83678 Service Manual ARL GenX 9-97


Chapter 9 Catalogue

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-98 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX161-00

LIST GENX161

AA83678 Service Manual ARL GenX 9-99


Chapter 9 Catalogue

GENX X GENERATOR ARL 9900 VERSION 2020 2.5KW


60KV/100MA
GENX X GENERATEUR ARL 9900 VERSION 2020
2.5KW 60KV/100MA

From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Introduction Gen’X 9900 from serial number 9953964
0 May 2020 including 9953165, 9953396, 9953835, 9953933, 9953941,
9953943.

Note :

9-100 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Figure 9.21

AA83678 Service Manual ARL GenX 9-101


Chapter 9 Catalogue

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX 2.5kW 60kV/100mA
1 1226748 1 GenX 2.5kW 60kV/100mA B
………..
CARTE,XPSC,GENX 9900
2 1226741 1 2.5 kW A

………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

9-102 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-103


Chapter 9 Catalogue

KXXGENX162-00

LIST GENX162

9-104 Service Manual ARL GenX AA83678


Catalogue Chapter 9

GENX X GENERATOR ARL 9900 VERSION 2020 4.2KW


60KV/120MA
GENX X GENERATEUR ARL 9900 VERSION 2020 4.2KW
60KV/120MA

From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Introduction Gen’X 9900 from serial number 9953964
0 May 2020 including 9953165, 9953396, 9953835, 9953933, 9953941,
9953943.

Note :

AA83678 Service Manual ARL GenX 9-105


Chapter 9 Catalogue

Figure 9.22

9-106 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
GenX 4.2kW 60kV/120mA
1 1226757 1 GenX 4.2kW 60kV/120mA B
………..
CARTE,XPSC,GENX 9900
2 1226755 1 4.2 kW A

………
XPSI board
3 S703123T 1 Carte XPSI A
………..
XPSP board
4 S703126T 1 Carte XPSP A
…………..
Multiplier assembly
5 S703758T 1 Ensemble Multiplicateur A
………….
Kit I, fans GENXRFHP Replace fans after
6 S703838 1 Kit I ventilateurs GENXRFHP C 1 2 years
……………
Fan “small” for XPSI & XPSP Replace fans after
7 A034398 2 Ventilateur “petit” pour XPSI C 2 years
et XPSP
…………….
Fan “big” for XPSI & XPSP Replace fans after
8 S704791 2 Ventilateur “grand” pour XPSI C 2 years
et XPSP
…………….
Fan for Multiplier assembly. Replace fans after
9 A039356 2 Ventilateur pour ensemble C 2 years
multiplicateur
……………
Fast fitting female
10 A040026 4 Raccord rapide femelle
……………
Fast fitting male
11 A040027 2 Raccord rapide male
……………
Fixed Fitting male
12 A040232 2 Raccord fixe male 2
……………
Collar D9-D16
13 AA22511 8 Collier D9-D16
…………….

AA83678 Service Manual ARL GenX 9-107


Chapter 9 Catalogue

Pipe D16/10
14 A040528 2 Tuyau D16/10 3
……………

Notes :
1 Positions 7,8,9 are included in this item.
2 This fixed fitting male is installed with Loctite 542, see chapter GENX99
3 Mention in your order the length in meter.

Remark / Remarque / Bemerkung :

9-108 Service Manual ARL GenX AA83678


Catalogue Chapter 9

KXXGENX199-03

LIST GENX199

AA83678 Service Manual ARL GenX 9-109


Chapter 9 Catalogue

KXXGENX199 TOOLKIT AND MISCELLANEOUS


OUTILLAGE ET DIVERS
WERKZEUGE UND DIVERSES

From - To Description
Level
De - à Description OM Note
Mod.
Von - Bis Beschreibung
Position 1 updated
1 Position 1 mise à jour 4112
Position 1 auf den letzten Stand gebracht
Positions 4 and 5 added
2 Positions 4 et 5 ajoutées
Positionen 4 und 5 hinzugefügt
Positions 6 to 8 added
3 Positions 6 à 8 ajoutées
Positionen 6 bis 8 hinzugefügt

Note :

9-110 Service Manual ARL GenX AA83678


Catalogue Chapter 9

Description Maintenance:
N° P/N Qty Description Type Note greyed items
Beschreibung have to be
considered
Kit I, fans GENXHP Replace fans
1 S703838 1 Kit I ventilateurs GENXHP C after 2 years
………..
Kit II, fans GENXRFLP Replace fans
2 S703839 1 Kit II ventilateurs GENXRFLP C after 2 years
…………
Loctite 542 to seal water
3 A031375 1 fittings
Loctite 542 étanchéité
……………
Kit I, fans GENXRDWATER
4 S705114 1 Kit I, ventilateurs
GENXRDWATER
……………
Kit II, fans GENXRDAIR
5 S705115 1 Kit II, ventilateurs
GENXRDAIR
……………
Fuse 0.8A (XPSI board)
6 AA20670 2 Fusible 0.8A (Carte XPSI)

Fuse 32A (XPSI board)


7 A039598 1 Fusible 32A (Carte XPSI)

Fuse 0.5A (XPSP board)


8 AA24479 1 Fusible 0.5A (XPSP board)

Note :

Remark / Remarque / Bemerkung :

AA83678 Service Manual ARL GenX 9-111


10

GENXRD X'TRA
GENXRD X'TRA Chapter 10

10 GENXRD X'TRA
Description of the main functions
The GenXRD X-Ray power supply is made of 5 main sub-units:

Mains XPSI
XrayPowerSupplyInput
Board
Master processor XPSRD
Control XrayPowerSupplyRetrofitXRD
Board

XPSCD
XrayPowerSupplyControlXRD XPSB
Board Xray
Power
XPSP Supply
XrayPowerSupplyPowerXRD Backplane
Board Board

XPSMD
XrayPowerSupplyMultiplier
Board

Multiplier
Multiplier And
HighVoltage Box Filament
Assembly

Figure 10.1

AA83678 Service Manual ARL GenX 10-1


Chapter 10 GENXRD X'TRA

Working principle

Part of XPSMD assembly


FET1 L D
230Vac
+400V
HighVoltage S

u
i FET2
C1 +200Vrtm Multiplier L
u
C
C2
0Vrtm

DSP 1000 M Ohm

XPSRD
uP 1000 M Ohm

i
400Vrtm

+200Vrtm
PI PWM
0Vrtm
Filament -HV
Transformer W filament
-e X-Rays
Cu anode

Figure 10.2

XPSI Air (X-Ray Power Supply Input)

XPSI power components (Power-Mosfet) are cooled by a combined radiator-half pipes system.

XPSI does convert the 230VAC main power to 360 - 420 VDC.

XPSI does manage the inrush current, the working current limitation and the power factor correction.

XPSRD (X-Ray Power Supply Retrofit XRD)

XPSRD functions are the emulation of a Spellman X-Ray power supply, the interfacing of the analog to
digital signals and vice versa with XPSCD.

XPSRD is the master in the dialog process.

XPSCD (X-Ray Power Supply Control XRD)

XPSCD does manage the kV regulation, the emission current regulation of the X-Ray tube. It is the brain of
the generator.

XPSCD is the slave in the dialog process.

XPSP (X-Ray Power Supply Power)

XPSP power components (Power-IGBT) are cooled by a combined radiator-half pipes system.

The XPSP board converts the intermediate DC voltage (360 V- 440 V) into AC voltage sources, which
supplies current to the filament transformer and the HV transformer.

10-2 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

XPSMD (X-Ray Power Supply Multiplier XRD)

The multiplier assembly is made with the following parts:

XPSMD board
Filament transformer
HV transformer
Multiplier box with voltage inversion

The XPSMD board has three main functions: the filament power supply, the HV measurement and the
emission current measurement.

The filament transformer via the XPSM board is used to supply up to 50 W to the X-ray tube filament.

The HV transformer via the multiplier is used to supply up to 4000 W to the X-ray tube anode.

XPSB (X-Ray Power Supply Backplane)

The backplane or motherboard print is a bus supporting 5 slots made of 5 x 32 pins female internal
connectors (J902-J906) and 2 5 x 32 pins female external connectors J901 and J907, of Europe format.

Each board has its specific position in the rack. There are no keys to prevent the connection of a board to the
wrong slot.

AA83678 Service Manual ARL GenX 10-3


Chapter 10 GENXRD X'TRA

XPSI Air (X-Ray Power Supply Input)

Figure 10.3

The main functions controlled by the XPSI board are the mains (230VAC) power factor correction for the
current, the mains current limitation and the X-ray lamps control.

The mains voltage is used via a filter and a relay. When the generator switches on, the relay is energized to
close both contacts. After the rectifier diodes, the FET1 is used to open the circuit if the current is above a
current limit. When the FET1 is open, the current flow through the integrated diode D1of the FET circuit,
due to the inductor L. When the current is below a current limit the FET1 will drive.

The current is controlled by two devices:

1. The DSP that does evaluate the current every 50 microseconds


2. A hardware fast comparator referenced to 5 V. 0.2 volts/A. In both cases if the current is excessive the
FET1 is set OFF.

FET1 L D2 +400Vrtm
L
C1
IL FET2 +12kV
230Vac U1 D1 U2 C2

N 0Vrtm

uC DSP

Figure 10.4

The FET2 and the integrated diode D2 of the FET circuit are used to step up the voltage to 400 V +/- 10 %
this voltage is adapted according to the kV output voltage. 360 V for 0 to 30 kV, 360 V to 420 V
proportionally from 30 kV to 60 kV).

Each 50 µs the current “IL”, the voltages “U1” and “U2” are evaluated in order to define the FET2 OFF/ON
conduction time. The maximum OFF/ON time is 25 us.

The current must be relative to the input voltage

10-4 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

V U1

IL

UD1
U2
FET2
OFF
U1

IL

FET2
ON

25 us

Figure 10.5

During the OFF time the current flows through D2 to load the capacitors C1 and C2. The C1/C2 voltage is
measured by the DSP and compared to the set-point value. The On time is regulated relative to this
comparison.

The proportional 400V working voltage will be used to supply:

− The IGBTs on the XPSP power board, generating the anode high voltage, via the transformers and the
multiplier.
− The IGBTs on the XPSP power board, generating the cathode (filament) heating current.

If the current is excessive, the driver IC 4 is shut off:

− By the DSP with the CurrentLimit signal dropping to 0 V,


− By the hardware comparator made of IC 8B and transistor Q2; If the current sampling voltage, 0.2 V/A,
does exceed the V Ref, Q2 is locked.

AA83678 Service Manual ARL GenX 10-5


Chapter 10 GENXRD X'TRA

Proportional 400 V

U2 400V Monitoring
420 V

360 V

HV
30 kV 60 kV
Figure 10.6

The other functions controlled by this board are


♦ The low voltage power supply voltages monitoring,
♦ The power component temperatures monitoring,
♦ The neutral to earth potential monitoring.
Low voltages and temperatures
The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.

The neutral to earth potential monitoring

The neutral potential does cross a resistor divider before to be applied to the neutral to ground comparator IC
6.

This device allows the DSP to detect a floating neutral or a neutral-phase inversion.

Remark: By air, for GenXRD application.

10-6 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

XPSP Air (X-Ray Power Supply Power)

Figure 10.7

This power board contains the switching components (IGBTs) for the high voltage transformer and for the
filament power supplies.

Part of XPSP HV
Transformer
FET1 L IGBT1 C3
230Vac D +400V
HighVoltage

C1
u
i FET2
u +200Vrtm
Multiplier
C2 IGBT2
0Vrtm

DSP 1000 M Ohm

External
Master
Board
uC 1000 M Ohm

Part of XPSP
i
IGBT3
400Vrtm

PI PWM +200Vrtm

0Vrtm Filament
IGBT4 Transformer

Figure 10.8

The HV switching works at 20 kHz. The transformer primary is connected to the C1/C2 point which
corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V).

The control board commands alternately the IGBT1 and IGBT2. The On time duration depends of the output
power needed. By this way, the transformer primary receives, via C3, plus and minus 200 V referenced to
the fix 200 VRTM.

AA83678 Service Manual ARL GenX 10-7


Chapter 10 GENXRD X'TRA

Chronogram for low power case:

UC3-0Vrtm

+400V

+200V

0 t

IGBT1
Command

IGBT2
Command

Figure 10.9

Chronogram for high power case:

UC3-0Vrtm

+400V

+200V

0 t

IGBT1
Command

IGBT2
Command

Figure 10.10

10-8 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

The transformer inductor and C3 capacitor make a resonant circuit. By this way the current look like
sinusoidal:

UC3-0Vrtm

+400V

+200V

0 t
Transformer
Primary
current

Figure 10.11

The filament power supply switching works at 40 kHz. The transformer primary is connected to the C1/C2
point which corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V).

The control board commands alternately the IGBT3 and IGBT4. The On time duration depends of the output
power needed (max. 50 W). By this way, the transformer primary receives positive and negative 200 V
referenced to the fix 200 VRTM.
See after the voltage and the current at the primary of the filament transformer:
Filament
transformer

+200V

-200V t

Filament
Transformer Primary
current

0 t

Figure 10.12

AA83678 Service Manual ARL GenX 10-9


Chapter 10 GENXRD X'TRA

The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.

The six-pack IGBT is cooled by air, for GenXRD application.

XPSMD (X-Ray Power Supply Multiplier XRD)


Part of XPSMD assembly
230Vac FET1 L D HighVoltage
+400V

u
i FET2
C1 +200Vrtm Multiplier
u
C2
0Vrtm

DSP 1000 M Ohm

XPSRD
uP 1000 M Ohm

i
400Vrtm

+200Vrtm
PI PWM
0Vrtm
Filament
Transformer

Figure 10.13

The XPSM assembly has three main functions.

♦ the HV supply
♦ the emission current measurement
♦ the filament current supply
The 400V from the IGBTs on the XPSP board, via XPSM board goes to the HV transformer. At the
transformer secondary the voltage changes according to the IGBT1 and IBGT2 ON time.

The maximum voltage is near 12000 V. The full wave multiplier increases five times this voltage to reach at
maximum - 60000 V. Two dividers are used to monitor the high voltage. The command board uses one of
them to control the HV via the servo loop function accordance with the kV set point. The second is used to
control the first via a subtraction function.

The emission current is measured with the current absorbed by the multiplier. The emission current
measurement allows to indirectly regulating the filament heating current, thus to match with the desired
emission current.

The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.

10-10 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

XPSCD (X-Ray Power Supply Control XRD)

Figure 10.14

XPSCD does dialog with XPSRD via a serial interface.


XPSCD controls the various functions of the X-Ray generator:

♦ the high voltage regulation


♦ the emission current regulation
♦ the filament heating current regulation

HV
Transformer
230Vac FET1 L D +400Vrtm
HighVoltage
C1
u
i FET2
u +200Vrtm
Multiplier
C2
0Vrtm

DSP 1000 M Ohm

XPSRD
uP 1000 M Ohm

XPSCD
400Vrtm
Part
+200Vrtm
PI PWM
0Vrtm Filament
Transformer

Figure 10.15

AA83678 Service Manual ARL GenX 10-11


Chapter 10 GENXRD X'TRA

HV servo-loop control
The control board receives the asked HV from instrument master board via XPSRD. The asked HV is
compared with the real HV to generate the HV error, which is used through the PID corrector to PWM duty-
cycle.
The PWM signals control the power stage on XPSP board to supply the right voltage according to the HV
error. The XPSM assembly transformer and multiplier increase this voltage to generate the real HV. The
DSP evaluates the values each 50 us.

uC DSP
HV
HV Multiplier
Power
set point + PID PWM Box
- Stage

HV
D HighVoltage 18762
A

Part of Part of
Part of XPSP board XPSMD assembly
XPSCD board

Figure 10.16

Filament current servo-loop control


The control board receives the desire current from the emission current servo-loop control part. This value is
converted, via a digital-to-analogic converter, into analogic voltage. This voltage corresponds to the needed
current (0.266V/A).
The rest of the bloc diagram works like a controlled current source. I.e., the filament current changes
according to the control voltage. 1 volt corresponds to 1 ampere. The set-point voltage is compared with the
current sensor voltage to control, via the proportional corrector, the PWM duty-cycle at 40 kHz. The power
board contains the power stage, which generates the power for the transformer. The transformer is used to
increase the current to the desired level (up to 5 A). Both the filament heating current and filament voltages
are monitored.

Emission current servo-loop control


The control board receives from the instrument master board the instruction for the desired emission current.
This value is compared, via an analogic-to-digital converter, to the inverted value, which corresponds to the
current absorbed in the multiplier. The comparison result is used through the PID corrector to control the
filament current power supply part. The µC evaluates the values each 50 ms.

10-12 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

uC
Emission
current
set point
+
-
PID Filament current power supply

HV
Power Multiplier
Stage

D -1
A
R
Part of Part of
Part of XPSCD board XPSP board XPSMD assembly

Figure 10.17

Power Factor Correction control


The DSP receives from the µC the desired intermediate DC working voltage (360 to 420 V) according to the
asked high voltage (HV). During power off state this value stays to zero. When the generator switches on,
the value changes to 360 V and then changes according to the asked output voltage (kV). This value is
compared to the real intermediate voltage to calculate the servo-loop error.

The rectified input voltage is weighted by this voltage error to become the current reference for a second
servo-loop. The real current, to generate the current error, is compared to this reference. The GPFC corrector
uses the current reference, the real current and the current error values to control the PWM duty-cycle, which
controls the Power MOS FET (FET2).
The FET1 is controlled by three ways. The first one is the intermediate voltage via a P corrector. The PWM
duty-cycle changes according to intermediate voltage. The second one is made on the real current value and
a comparator with a current threshold. At switching power on, this current threshold corresponds to 2
amperes to limit the inrush current. When after 10 seconds, the intermediate voltage exceeds 320 V, the
current threshold changes to 40 Amperes. This corresponds to the maximum acceptable current during power
on state. The last way controls the FET1 by hardware. The real current is compared with a reference value.
When the real current exceeds 40 amperes the FET1 is switched off. This circuit allows reacting faster than
the DSP, where its reaction time is related to the sampling period.

Switches and jumpers


1 OFF Reserved
2 OFF Reserved
S1
3 OFF Reserved
4 OFF Reserved
W1 OFF
W2 0,1,2,3 Generator type
W2.3 W2.2 W2.1 W2.0
installed removed removed installed GenXRD

AA83678 Service Manual ARL GenX 10-13


Chapter 10 GENXRD X'TRA

Part of XPSI board


P
FET1

FET2

PWM

A
D
P

-
A +
D

P P
DSP
PI

GPFC PWM

Reference D
+ -
Voltage A

Part of XPSC board

Figure 10.18

The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogic to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC.

10-14 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

XPSRD (X-Ray Power Supply Retrofit XRD)


An extra board called XPSRD is installed inside the generator. This board will be used as emulator between
the GenX generator and the MASTER CONTROL BOARD of the instrument electronic rack.

XPSRD does emulate non GenX generators.

Figure 10.19

50 P CABLES

SUB-D SUB-D
50 P J401
MASTER
CONTROL
BOARD

XPSRD

XPSCD DISPLAY
XPSI
INSTRUMENT GenXRD
XPSP
XPSMD

Figure 10.20

AA83678 Service Manual ARL GenX 10-15


Chapter 10 GENXRD X'TRA

XPSRD bloc diagram

GenXRD

Master 50P Cable Connector uC


Processor SUB-D 50P ADUC812
XPSC
Display
Board
XPSR
LVPS Board

Safety
And
Water
interface

Figure 10.21

XPSR IS DIVIDED IN THREE MAIN PARTS


1. Low Voltage power Supply
2. XPS Generator Emulation
3. Display

Part 1 Low voltage power supplies

XPS (plugged into the instrument electronic rack)


+ 5 VoltsOut regulated
Low voltage power supplies for XPS +15 VoltsOut regulated
-15 VoltsOut regulated
+ 24 Volts
XPSR (plugged into GEN’X rack)
+ 5 Volts regulated
Low voltage power supplies for XPSR +12 Volts regulated
-12 Volts regulated
+ 24 Volts
GenX backplane low Voltage power supplies
+ 5 VoltsOut regulated
Low voltage power supplies for mother board +12 VoltsOut regulated
-12 VoltsOut regulated
+ 24 VoltsOut From Interlock RelayRY1

10-16 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

Low voltage power supply circuit

The low voltage power supply generator is supplied by the 230VAC Line1 and Line2 via XPSI Board.

The 230VAC is converted to 24VDC relative to the following bloc diagram

Figure 10.22

Voltage regulator bloc diagram TOP 247

Figure 10.23

The regulated 24 VDC is applied to five low voltage regulators used to generate respectively

+15 VDC
-15 VDC
+5 VDC
+12 VDC
- 12 VDC

All low voltages are protected by Multi-fuses.

AA83678 Service Manual ARL GenX 10-17


Chapter 10 GENXRD X'TRA

+ 24 VDC Safety circuit


The + 24 V is connected to the safety relay RY1.

If the signal “/InterlockOk” is true, the potential is 0 V.

The relay RY1 drives the + 24V its contacts to + 24Vout.

+24Vout does supply the 24 V to all the boards plugged on the GenX mother board.

If the interlock is not OK the +24Vout is not supplied to the boards;

10-18 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

Part 2 XPS Generator Emulation

XPSRD GEN'X
MASTER
PROCESSOR
BOARD EMULATOR XPSCD
ADUC812

Outgoing
Signals To/From
Bus XPSCD
Serial Data Link
Ingoing
Signals
Bus

Figure 10.24

The task of the µ Controller ADUC812 is to simulate a non GenX generator.

To do so:
• The various alarms generated by XPSCD are linked to the ADCU812 processor through a serial
link and converted, to be recognised by MASTER PROCESSOR BOARD.
• The start up synchronisation and safety control signals generated by MASTER PROCESSOR
BOARD are converted to serial Data and transmitted to XPSCD.
• The analogic Ref signals are converted to serial Data by the ADUC812 built in ADC, to be
transmitted to XPSCD.
• The Monitoring signals are converted from serial Data by the ADUC812 built in DAC to analogic
voltages to be supervised by MASTER PROCESSOR BOARD.

AA83678 Service Manual ARL GenX 10-19


Chapter 10 GENXRD X'TRA

KVMonFb XPSRD

KVRef

PowerRef

Master I_FilRef

Board

MaRef

MaMonFb

Safety
Interlock
XrayOn
Control

Figure 10.25

Serial link
Serial link is of the same type as XSN (X-ray Services Network) used in the 9900 / ADVANT’X series
instruments electronic racks.

The jumper W1 must be placed between contacts 1-2 internal, for proper communication.

The serial communication is watchdog controlled. In case of communication timeout or if a communication


error is detected, XPSCD does shut down the generator.

Jumper settings
JUMPER SETTING
W1 1-2 Internal
W2 Installed
JP1 EnProg Removed
R98, R102 Removed
R99, R100 Installed

10-20 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

Part 3 Display
Schematic and layout: S215199

Display Display panel

Figure 10.26

The 1x 20 characters display is mounted and plugged on the display front panel.

The display front panel is plugged on both external side connectors.

Display Display front panel

Figure 10.27 Display

The display data are controlled by XPSCD.

AA83678 Service Manual ARL GenX 10-21


Chapter 10 GENXRD X'TRA

GenXRD Trouble Shooting Procedure


Possible breakdowns are reported, either to the MASTER PROCESSOR BOARD by mean of LEDs, or to
the GenXRD display by mean of alarms numbers. Up to 10 alarms, running in a loop, can be monitored on
the display.
Alarms list and description
Numéro Description LEDs on Master Processor Board
alarme
8192 Unknown command received from XPSRD POWER SUPPLY FAULT
S Restart to reset Master Processor
8196 Desired voltage (kV) too low KV MIN ERROR
Asked kV conditions not allowed by firmware.
S Select coherent conditions

8197 Desired voltage (kV) too high OVER VOLTAGE


Asked kV conditions not allowed by firmware.
S Select coherent conditions

8198 Desired current (mA) too low mA CURRENT LIMIT


Asked mA conditions not allowed by firmware.
S Select coherent conditions

8199 Desired current (mA) too high mA CURRENT LIMIT


Firmware and XPSC jumpers setting not compatible.
S Select coherent conditions

8200 Desired power too high OVER POWER


Asked power (kV × mA) not allowed by firmware.
S Select coherent conditions

8201 Illegal power change mode POWER SUPPLY FAULT


Erroneous instructions from Master Processor
S Restart to reset Master Processor

8202 Illegal power change speed POWER SUPPLY FAULT


KV / mA Out of power change speed out of range. < 1W / min. or > 3000 W / min.
S Set coherent conditions.

8203 Illegal anode Not used POWER SUPPLY FAULT

8210 Time-out on X-ray ON ack POWER SUPPLY FAULT


Synchronisation error with Master Processor
S Restart to reset Master Processor

8215 X-ray tube filament open N/A

Filament Curent < 1 A Filament Voltage > 16V


Bad filament connection.
Disconnect filament and measure filament resistivity with a DVM: It should be < 1 Ohm.
If not filament is defect.

10-22 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

S Change X-Ray tube

8244 X-ray arc rate too big POWER SUPPLY FAULT


The number of arcs does exceed the high value limit defined in the status channel 2064.

8257 Unknown X-ray ray power-supply type. POWER SUPPLY FAULT


XPSCD Incoherent Jumpers settings.
S Set jumpers according to technical description

8270 SSIO communication timeout. POWER SUPPLY FAULT


All serial links on XPSC
S Restart to reset Master Processor

8271 Serial AD12 conversion timeout. POWER SUPPLY FAULT


No serial ADC communication.
S Restart to reset Master Processor

8272 Serial AD12 auto-calibration timeout. POWER SUPPLY FAULT


Auto calibration not successful
S Restart to reset Master Processor

8273 X-ray “kV” out of regulation N/A

AA83678 Service Manual ARL GenX 10-23


Chapter 10 GENXRD X'TRA

kV Regulation

kV > 0 No No test

Yes

Yes kV > 5 kV No

< -20%of value


> 5% of value
> 5% of value

Yes
Yes No

Fatal Alarm 8273 No OK

Figure 10.28

8274 X-ray “mA” out of regulation N/A

mA Regulation

mA > 0 No No test

Yes

Yes mA > 10 mA No

< -20%of value


> 10% of value
> 10% of value

Yes
Yes No

Fatal Alarm 8274 No OK

Figure 10.29

10-24 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

8280 Unknown identity X-ray power supply fault. POWER SUPPLY FAULT

8281 X-ray warning lockgate broken. N/A

8282 X-ray tube filament power supply fault. POWER SUPPLY FAULT
Heating current is < 1 A and heating voltage is < 2 volts
Fuse F2 2A (orange soldered fuse) of XPSP board
SG3525AN IC 20 of XPSC.
Filament heating power IGBTs on XPSP.
The filament power supply part of the XPSM board (part of the Multiplier assembly) can be suspected.

8292 Generators mains voltage not present, time-out POWER SUPPLY FAULT
If main is < 280VAC. In theory the voltage is 230 × 2 See trouble shooting section 4 step 4

8293 Generator 400V not > 250V, time-out POWER SUPPLY FAULT
Time out = 5 sec. See trouble shooting section 4 step 5

8294 Generator 400V not > 380V, time-out POWER SUPPLY FAULT
Time out = 5 sec. See trouble shooting section 4 step 6

8295 Generator, temperature IGBT too high POWER SUPPLY FAULT


120°C Bad venting, fans defect, environmental temperature too high.

8297 XPSCD First 8 bits A/D converter error POWER SUPPLY FAULT
8298 XPSCD Second 8 bits A/D converter error POWER SUPPLY FAULT
Communication error.
S Power up

8299 XPSI 8 bits A/D converter error POWER SUPPLY FAULT


8300 XPSP 8 bits A/D converter error POWER SUPPLY FAULT
Communication error.
XPSI Board not properly plugged
S Power up

8301 XPSMD 8 bits A/D converter error POWER SUPPLY FAULT


Communication error.
XPSMD Board not properly plugged
S Restart Master Processor

8302 Generator, temperature Mosfet M1 too high POWER SUPPLY FAULT


8303 Generator, temperature Mosfet M2 too high POWER SUPPLY FAULT
If > 65°C
Fan problems or environmental temperature too high

8304 Leakage current too high N/A

8305 Power off, issued by DSP; 400V too low (<350V) POWER SUPPLY FAULT
or 8513
Too fast X-Ray power supply transition. 400 V voltage drop. May happen when changing X-Ray tube
conditions in terminal mode, in “On Fast mode” and “Operational Mode”.
See trouble shooting section 4 step 6

AA83678 Service Manual ARL GenX 10-25


Chapter 10 GENXRD X'TRA

8306 Power off, issued by DSP; HV too high (>71kV) OVER VOLTAGE
or 8514
KV Regulation error. KV Regulation loop fault.

8307 Power off, issued by DSP; Current too high (>35A) POWER SUPPLY FAULT
or 8515
After a power up retry the current can rise above 35 A.

8308 Mains voltage below 195 VAC (after 20 ms) POWER SUPPLY FAULT
or 8516
8309 Mains voltage above 253 VAC (after 20 ms) POWER SUPPLY FAULT
or 8517
Main power supply fault. The main voltage is checked by the DSP. If out of tolerance the DSP does
immediately shut down the power.
See trouble shooting section 4 Step 4

8310 Unknown reply from DSP, issued by XPSCD POWER SUPPLY FAULT
80C196KR no dialog with DSP.
S Power up

8311 Too many parameters in SSIO, issued by DSP POWER SUPPLY FAULT
or 8519
S Power up

8312 Unknown DSP Cmd, issued by DSP POWER SUPPLY FAULT


or 8520
S Power up

8313 PWMTRIP -> Power Off, issued by DSP POWER SUPPLY FAULT
or 8521
Issued by XPSP due to repetitive power stage problems, where the current is > 100A
Check the IGBT.
Issued by XPSM, due to arcs or spark over on the multiplier assembly of XPSM.
If > 15 kV (> 2 V at comparator input) at the secondary of the HV transformer of XPSM. Normal 12kV for
60 kV

8314 Generator, temperature XPSP too high POWER SUPPLY FAULT


If > 75 ° C
Bad cooling. Check fans. Laboratory temperature too high.

8315 Generator, temperature filament too high POWER SUPPLY FAULT


If XPSM filament heating diodes D1 to D4 > 100 ° C.
Check fans. Laboratory temperature too high.

8316 Generator, temperature HV too high POWER SUPPLY FAULT


If XPSMD HV conversion unit > 75 ° C
Check fans. Laboratory temperature too high.

8317 Generator, temperature XPSCD too high POWER SUPPLY FAULT


If XPSCD > 65 °C
Check fans. Laboratory temperature too high.

10-26 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

8323 Filament current sensor incoherency POWER SUPPLY FAULT


8324 Filament voltage sensor incoherency POWER SUPPLY FAULT
If before start the sensor delivers a value > 1 A → XPSMD Filament current sensor defect.
If before start the sensor delivers a value > 10 V → XPSMD Filament voltage sensor defect.
Measured values out of range, not coherent.

AA83678 Service Manual ARL GenX 10-27


Chapter 10 GENXRD X'TRA

8325 XPSI +5V too low <4V POWER SUPPLY FAULT


8326 XPSI +5V too high >6V POWER SUPPLY FAULT
8327 XPSCD +5V too low <4V POWER SUPPLY FAULT
8328 XPSCD +5V too high >6V POWER SUPPLY FAULT
8329 XPSP +5Vm too low <4V POWER SUPPLY FAULT
8330 XPSP +5Vm too high >6V POWER SUPPLY FAULT
8331 XPSMD +5Vm too low <4V POWER SUPPLY FAULT
8332 XPSMD +5Vm too high >6V POWER SUPPLY FAULT
8333 XPSI +12V too low < 8.5 V POWER SUPPLY FAULT
8334 XPSI +12V too high > 15 V POWER SUPPLY FAULT
8335 XPSCD +12V too low < 8.5 V POWER SUPPLY FAULT
8336 XPSCD +12V too high > 15 V POWER SUPPLY FAULT
8337 XPSMD +12V too low < 8.5 V POWER SUPPLY FAULT
8338 XPSMD +12V too high > 15 V POWER SUPPLY FAULT
8339 XPSI -12V too low > -15 V POWER SUPPLY FAULT
8340 XPSI -12V too high > -8.5 V POWER SUPPLY FAULT
8341 XPSCD –12V too low > -15 V POWER SUPPLY FAULT
8342 XPSCD -12V too high > -8.5 V POWER SUPPLY FAULT
8343 XPSMD -12V too low > -15 V POWER SUPPLY FAULT
8344 XPSMD -12V too high > -8.5 V POWER SUPPLY FAULT
8345 XPSI +24V too low < 19 V POWER SUPPLY FAULT
8346 XPSI +24V too high > 27 V POWER SUPPLY FAULT
8347 XPSCD +24V too low < 19 V POWER SUPPLY FAULT
8348 XPSCD +24V too high > 27 V POWER SUPPLY FAULT
8349 XPSP +24V too low < 19 V POWER SUPPLY FAULT
8350 XPSP +24V too high > 27 V POWER SUPPLY FAULT
8353 XPSCD –5V too low > -6 V POWER SUPPLY FAULT
8354 XPSCD –5V too high < -4 V POWER SUPPLY FAULT
Low voltages out of tolerance.
Check instrument low voltage power supplies.

8355 Mon 400V too low POWER SUPPLY FAULT


8356 Mon 400V too high POWER SUPPLY FAULT

8360 XPSI A/D Mon400V Error POWER SUPPLY FAULT


Previous to start, if U < 10V or > 350 V.
If > 350 V → XPSI voltage monitoring IC 7 Defect.

K1 i
L
IL C1 +12kV
230Vac U1 U2
C2
N
15Vrtm

uC DSP

Figure 10.30

At power up if the 32 Amp breaker is on, the voltage leaks from T1 through the drivers to the capacitors C.

10-28 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

The 400 V measuring device does read about 15 V. If not, the 32 Amp breaker is down or the IGBTs of
XPSP are in short circuit.
8362 XPSCD temperature Board overflow POWER SUPPLY FAULT
If > 109 ° C. Out of conversion limits. Mainly indicating NTC is defect.

8363 XPSCD temperature Board underflow POWER SUPPLY FAULT


If < 9.9 ° C. Out of conversion limits. Mainly indicating NTC is defect.

8364 XPSI temp. PowerMosfet M1 overflow POWER SUPPLY FAULT


8365 XPSI temp. PowerMosfet M1 underflow POWER SUPPLY FAULT
8366 XPSI temp. PowerMosfet M2 overflow POWER SUPPLY FAULT
8367 XPSI temp. PowerMosfet M2 underflow POWER SUPPLY FAULT
All same as above

8368 XPSP temperature IGBT overflow POWER SUPPLY FAULT


8369 XPSP temperature IGBT underflow POWER SUPPLY FAULT
All same as above

8370 XPSP temperature Board overflow POWER SUPPLY FAULT


8371 XPSP temperature Board underflow POWER SUPPLY FAULT
All same as above

8372 XPSMD temp. Filament overflow POWER SUPPLY FAULT


8373 XPSMD temp. Filament underflow POWER SUPPLY FAULT
All same as above

8374 XPSMD temp. HV overflow POWER SUPPLY FAULT


8375 XPSMD temp. HV underflow POWER SUPPLY FAULT
8377 X-Ray lockgate shorted N/A
8378 XPSP –10Vrtm too low POWER SUPPLY FAULT
8379 XPSP –10Vrtm too high POWER SUPPLY FAULT
8380 XPSP +15Vrtm too low POWER SUPPLY FAULT
8381 XPSP +15Vrtm too high POWER SUPPLY FAULT
8382 Neutral to Gnd too low POWER SUPPLY FAULT
8383 Neutral to Gnd too high POWER SUPPLY FAULT
If potential is > 200VAC.
The neutral has a floating potential.
Check the main power supply parameters.
Install an insulation transformer if necessary with the 0 volt connected to the ground.
Phase / Neutral are inverted.
8384 XPSI input current sensor error POWER SUPPLY FAULT
If previous to start the sensor does measure a current > 2 A.
Sensor is defect. If the value is > it means →no current limitation.

8389 X-ray negative leakage current error N/A


X-Ray tube leakage.
Measure tube insulation.
S If not OK, change tube.

8390 X-ray positive leakage current error N/A


Multiplier divider leak.
S Change XPSMD.
8392 Filament current limit reached FILAMENT CURRENT LIMIT

AA83678 Service Manual ARL GenX 10-29


Chapter 10 GENXRD X'TRA

An additive Alarm is available and can be monitored on the display. If the INTERLOCK is open,
the next message is displayed.

INTERLOCK OPEN

At the start–up of the GenXRD the firmware versions are displayed in the following manner:

XPSRD V3.0-0

XPSCRD V3.0-0

XPSCS V3.2-6

10-30 Service Manual ARL GenX AA83678


GENXRD X'TRA Chapter 10

Master Processor Settings


Two parameters have to be adjusted on the master processor board:

♦ The power limitation


♦ The filament current limitation

Power limit setting

Connect a DVM on TP2 and GND TP7

Voltage on TP2 = (0.0025*power[W]) + 0.05

Set the potentiometer P2 to define the maximum power limit relative to the next table.

The maximum power is 4 kW that means 10 V on TP2.

Tube Anode
Power

Focus Copper Silver Molybdenum Cobalt Chromium Iron


Cu Ag Mo Co Cr Fe
Broad 2700W 2700W 2700W 2700W 2700W 2200W
Long 2200W 2200W 3000W 1500W 1700W 1100W
Fine
Normal 2000W 2000W 2400W 1800W 1800W 1500W
Fine 1500W 1500W 2000W 1200W 1300W 900W
Ultra 800W 800W 800W 600W 800W 300W
Fine

Filament current limit

Connect a DVM on TP1 and GND TP7

Set P1 to adjust 8V ± 0.01 corresponding to maximum 4 Amperes.

The maximum current is 5 Amperes corresponding to 10 V on TP1.

AA83678 Service Manual ARL GenX 10-31


Chapter 10 GENXRD X'TRA

Specifications
# DESCRIPTION VALUES REMARKS
1 Output power 0 to 4KW
2 Input voltage 230VAC +10% / -15%
50/60Hz 1phase
3 Maximum input current and inrush current 28A at 195VAC
4 Maximum input power 5000W
5 Efficiency > 80%
6 Power factor > 0.9 switched mode load type
7 Output voltage -10 to -60KV Without overshoot.
Short-circuit protected.
8 Output voltage vs Input voltage < 50 ppm Over specified range
9 Output voltage vs Output load < 50 ppm 100 % load changes
10 Output voltage vs Temperature 50 ppm/deg. Over specified range
11 Output voltage accuracy < 0.5 % Related to set-point value
(max offset: 250V)
12 Output voltage ripple 0.03 % rms Below 1KHz
13 Output voltage ripple 0.3 % rms Above 1KHz
14 Emission current 0..100 mA
15 Emission current vs Input voltage < 50 ppm Over specified range
16 Emission current vs Output load < 50 ppm Load change According to voltage
change from -20KV to -60KV
17 Emission current vs Temperature 50 ppm/deg. Over specified range
18 Emission current accuracy < 0.5 % Related to set-point value
(max offset 0.08 mA)
19 Filament power supply voltage 0..14 Vac Over voltage protected
to 14V
20 Filament power supply current 0..5 A short-circuit protected
21 Filament power supply current Output 1 % rms Below 100 Hz
voltage ripple
22 Filament power supply current Output not critical Above 100 Hz
voltage ripple
23 Operating temperature range 0..35deg.C Inlet Air Temperature
24 Cooling system Forced air
30 High voltage connector Type II, US The 3 terminals C, L, S connected
FEDERAL to the high voltage
STANDARD H72,
3 PINS
31 Main connection 6c. WAGO connector
32 X-ray ON valve connector IEC320 Supply: 230V15W
33 Maximum power change ramp 1KW/second Limited by interface
34 Maximum voltage change ramp 100KV/second Limited by interface
35 Maximum filament current change ramp 2A/second Limited by interface
36 X-ray arcs detection. shut off after 5 arcs in 10 seconds
37 Control and command connector SUB-D 50P Female 50x
38 Over temperature protection

10-32 Service Manual ARL GenX AA83678


11

GENXRD WORKSTATION
GENXRD WORKSTATION Chapter 11

11 GENXRD WORKSTATION
Description of the main functions

Figure 11.1

With the ARL 9900 WORKSTATION, 2 generators have to be controlled by XQIX.

The serial part Numbers are:

For GenXRDV18 S704368T This generator is the Valve control version.


It is installed for instruments with a 3.6 or 4.2 kW GenXRF
For GenXRDF18 S704385T This generator is the Fan control version.
It is installed for instruments with a 2.5 kW GenXRF

AA83678 Service Manual ARL GenX 11-1


Chapter 11 GENXRD WORKSTATION

The 2 generators are interconnected via the GenXRD-GenXRF Interconnection board.


Schematic S010801, layout S318795.

Interlocks:
All safety switches closed Interlock: closed
Spectrometer Tank Pressostat open Interlock: open
Interlock: Pressostat.
Electronic Rack Security Switch open Interlock: open
Interlock: Intlk2
Tank Cover or/and Interlock: open
X-Ray Tube Security Switch open Interlock: Intlk3
Load Lift Rack Security Switch open Interlock: open
Interlock: Intlk4
X-Ray Power Supply Rear Rack Security Switch open Interlock: open
Interlock: Intlk5
X-ray Secondary Water Over-temperature Sensor Switch open Interlock: open
(60° C) Interlock: IntlkRtn

The interconnection board GenXRD-GenXRF allows controlling 2 generators by 4


distinct control categories

♦ The command signals


They are interfaced by the opto-couplers IC2, IC3, IC4, and IC5
And by the diodes D1, D2, D3, D4, D6, D7, D8.

WaterPump DeWaterFlow
GenXRF ON Enable
GenXRD OFF Disable
Water Pump DeWaterFlow
GenXRF OFF Disable
GENXRD ON Enable

11-2 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

PrWavalve PrWaFlow
GenXRF ON Enable
GenXRD OFF Disable
PrWavalve PrWaFlow
GenXRF OFF Disable
GENXRD ON Enable

WaterCooler WaterCoolerOn
GenXRF ON Enable
GenXRD ON Enable

♦ The analogue signals


The analogue signals PrWaTin, PrWaTOut, WaTIn, WaTOut, are dispatched to the 2 generators through
the drivers IC1A. The signal WaResist is only and directly linked to the GenXRF, the fact is that the
XRDTube does not need a water resistivity check.
♦ The State signals
These signals like WaLevel, WaCooler On, Intlk2 to IntlkRtn are dispatched to the 2 generators in direct
line.
♦ The Power supply
The Pressostat/24V generated on the XPSC board is applied to the Interlock circuit via the Diodes D5 and
D9.

AA83678 Service Manual ARL GenX 11-3


Chapter 11 GENXRD WORKSTATION

Their respective node addresses are:


For XRF Generator GenXRF 2,3,1
For XRD Generator GenXRD 2,8,1.

CLUSTER 1
DISPLAY 1

ACS XQIX XSM


ID: 1.1 ID: 1.2

CLUSTER 3 (GONIO1) DISPLAY 2

XCMC XDMC CLUSTER 2 CLUSTER 2


ID: 3.4 ID: 3.5 (MONO's) XGT (SERVICES)
ID: 2.1

DISPLAY 3
XMI32 XCM
XGB(I/M) ID: 2.63 ID: 2.2 XRF
ID: 3.1
generator

XPSCM XPSCS
XDI ID: 2.3.0 ID: 2.3.1
ID: 3.3

XHI
CLUSTER 4 (GONIO2) ID: 2.4

CLUSTER 5 (GONIO3)
XSI
CLUSTER 6 (XRD1) ID: 2.5

XCMC XDMC
ID: 6.4 ID: 6.5 XMI02 XSP
ID: 2.33 ID: 2.6 XRD
generator

DISPLAY 6
XMI01 XPSCMXRD XPSCSXRD
XGBX ID: 2.32 ID: 2.8.0 ID: 2.8.1
ID: 6.1

XSN
XDI
ID: 6.3
BASIC CLUSTER BASIC NODE

CLUSTER 7 (XRD2)
OPTIONAL CLUSTER OPTIONAL NODE
XGN
INTEL 386EX INTEL 80C196KR ANALOG ADMC401

Figure 11.2

11-4 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

In normal operation the GenXRF is the master and GenXRD is the slave. The master does control the start-
up sequence (Including PTB safety) and the X-Ray tubes cooling water parameters.
One or the other generator, or both generators, can be disabled by the Firmware status 31.
If GenXRF is disabled, GenXRD becomes automatically the master. In turn it has to control the start-up
sequence (Including PTB safety) and the X-Ray tubes cooling water parameters.
If GenXRD is the Master it has to control the X-Ray tube cooling parameters, with the primary water valve if
GenXRF 3.6 or 4.2 kW versions is installed and by the Water-air exchanger fans, in case of GenXRF 2.5 kW
version is installed.
The XRF generator is still controlled by the XSN node 2.3. The new XRD generator is controlled by the
XSN node 2.8.

The alarms range which was allocated to the XRF generator is now shared by the two generators according
to the following table.

Severity XRF generator XRD generator


FATAL 8192-8447 (256) 8448-8703 (256)
WARNING 8704-9087 (384) 9088-9471 (384)
INFORMATION 9472-9599 (128) 9600-9727 (128)
DIAGNOSTIC 9728-9983 (256) 9984-10239 (256)

The “Status channels” and the “Firmware status” of the new XRD generator are in the range 4096 to 4607.

The firmwares of the 2 generators must support the new IDLE mode. In this mode, the X-ray tube current
emission is not controlled anymore but the filament current is forced to a value defined by a parameter of a
new firmware status (see below).

A new firmware status “X-ray generators IDLE and ECO conditions” (31) allows to enable/disable the
generators. Moreover, it gives IDLE conditions for the XRF generator and the IDLE and ECO conditions for
the XRD generator. The ECO conditions of the XRF generator as well as the entry and exit times in ECO
mode are given by the XG command.
Two parameters have been added to the firmware status “General instrument activity” (8), to report the XRD
generator activity time and the XRD tube sparks.

One generator configured


If one generator only is configured (see firmware status 31), that generator is the “master”; it controls and
monitors the X-ray cooling system.

Two generators configured


If both generators are configured, the XRF generator (2.3) is the “master” generator and the XRD generator
(2.8) becomes the “slave” generator. In “master” mode, a generator controls and monitors the X-ray cooling
system. In “slave” mode a generator monitors the X-ray cooling system only.
At power-up, both generators are set in “master” mode. Before each X-ray conditions command sent to the
XRD generator, XQIX sends a specific command to set the XRD generator in “slave” mode.
When XQIX receives a PA command concerning a generator, that generator is set in ACTIVE mode at the
conditions received with the command. The other generator is set in IDLE mode at the conditions given by
the firmware status 31.

AA83678 Service Manual ARL GenX 11-5


Chapter 11 GENXRD WORKSTATION

31 XQIX ALL X-ray generators IDLE and ECO conditions


Read only : NO Sent after initialisation : NO
Parameters: Type: Default firm: Limit firm: Dflt soft:
State (A5A5H = defined) STRING FFFFH - FFFFH
XRF generator present [] WORD 1 - 1
0: XRF generator absent
>0 : XRF generator present
XRF generator IDLE kV [kV] WORD 20 - 20
XRF generator IDLE filament current [A] FLOAT 5.0 - 5.0
XRF generator IDLE mA [mA] WORD 0 - 0
XRD generator present [] WORD 1 - 1
0: XRD generator absent
>0 : XRD generator present
XRD generator mode [] WORD 0 - 0
0: Set IDLE conditions for ECO mode
>0 : Set ECO conditions for ECO mode
XRD generator IDLE kV [kV] WORD 10 - 10
XRD generator IDLE filament current [A] FLOAT 1.0 - 1.0
XRD generator IDLE mA [mA] WORD 0 - 0
XRD generator ECO kV [kV] WORD 10 - 10
XRD generator ECO mA [mA] WORD 10 - 10

Figure 11.3

ECO Mode

If the ECO mode is set Enabled:


ECO Mode XRF generator GenXRF XRD generator GenXRD
Enter ECO Active: apply parameters defined in the XG Idle: apply parameters of the Firmware
Mode command status 31
Idle: apply parameters defined in the XG Active: apply parameters of the Firmware
command status 31
Exit ECO Active: apply parameters defined in the XG Idle: apply parameters of the Firmware
Mode command status 31
Idle: apply parameters defined in the XG Active: apply parameters of the Firmware
command status 31

If the ECO mode is set Disabled:


It applies the IDLE kV, mA, Filament Current, defined in the firmware status 31, for the generator
that is Idle.

11-6 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

New X-R command

) Command:
XR <X-ray kV>
,<X-ray mA>
,<X-ray generator>
,<X-ray state>
,<X-ray anode>
,<true X-ray kV-mA>

Command parameters definition:


X-ray kV INTEGER
0: 0 kV if “true X-ray kV-mA” = 1.
Same as last XR command if “true X-ray kV-mA” <> 1.
For valid kV value, see LEGAL KV-MA COMBINATIONS graph.

X-ray mA INTEGER
0: 0 mA if “true X-ray kV-mA” = 1.
Same as last XR command if “true X-ray kV-mA” <> 1.
For valid mA value, see LEGAL KV-MA COMBINATIONS graph.

X-ray generator INTEGER


X-ray generator selection
<=3: XRF generator
>3: XRD generator

X-ray state INTEGER


X-ray generator behavior for the current XR command
0: Same as last XR command
1: Slow (to “X-ray kV” – “X-ray mA” at 300 W/min)
2: Fast (to “X-ray kV” – “X-ray mA” at 1500 W/min)
3: Switch OFF X-ray generator
4: Switch X-ray generator in IDLE mode according to firmware status 31.

X-ray anode INTEGER


Always 1

true X-ray kV-mA INTEGER


Interpretation flag of “X-ray kV” and “X-ray mA“ parameters
1: Interprete 0 as “real 0”
<>1: Interprete 0 as same as last XR command
© Returned data:
None

Figure 11.4

Events Commands
X1OFF: XR kV,mA,<=3,3
X2OFF: XR kV,mA,>=4,3
X1ON: XR kV,mA,<=3,1|2 PA 0,kV,mA,0
X2ON: XR kV,mA,>=4,1|2 PA 0,kV,mA,2048
X1IDLE XR kV,mA,<=3,4
X2IDLE XR kV,mA,>=4,4

AA83678 Service Manual ARL GenX 11-7


Chapter 11 GENXRD WORKSTATION

XC: X-RAY generator calibration

) Command:
XC <function>
(,<param1>) Min. 20kV
(,<param2>) Max. 45 kV
(,<param3>) Min. 20 mA
(,<param4>) Max 45 mA
(,<param5>)

Command parameters definition:


function WORD
Desired X-ray generator calibration process function
1: Perform the X-ray generator calibration process.
The X-ray generator is set to different values of kV and mA defined by
the following parameters. New OFFSET and new SLOPE for the kV and
the mA are calculated and stored.
2: Store desired OFFSET and SLOPE for kV and mA given by the
following parameters.
3: Uncalibrate the X-ray generator.
Set OFFSET and SLOPE for kV and mA to their default values.
other: Return current OFFSET and SLOPE parameters for kV and mA.

Param1 REAL
for function = 1
Desired low kV calibration point [kV].
0.0 : default low kV calibration point.

for function = 2
Desired kV OFFSET.

for function = other


Ignored.

Param2 REAL
for function = 1
Desired high kV calibration point [kV].
0.0 : default high kV calibration point.

for function = 2
Desired kV SLOPE.

for function = other


Ignored.

Param3 REAL
for function = 1
Desired low mA calibration point [mA].
0.0 : default low mA calibration point.

for function = 2

11-8 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Desired mA OFFSET.

for function = other


Ignored.

Param4 REAL
for function = 1
Desired high mA calibration point [mA].
0.0 : default high mA calibration point.

for function = 2
Desired mA SLOPE.

for function = other


Ignored.

Param5 WORD
for all functions
X-ray generator selection
0: XRF generator
<>0: XRD generator

© Returned data:
ALARMS STATUS
,<current kV OFFSET>
,<current kV SLOPE>
,<current mA OFFSET>
,<current mA SLOPE>

Returned data definition:


current kV OFFSET REAL

current kV SLOPE REAL

current mA OFFSET REAL

current mA SLOPE REAL

2
The parameters updated by this command are also reachable through the firmware status “kV/mA
measurement calibration” (2048 for XRF generator (0) or 4096 for XRD generator (1)).

Figure 11.5

AA83678 Service Manual ARL GenX 11-9


Chapter 11 GENXRD WORKSTATION

Generator selection
Existing parameters or new parameters of the commands XR, PA and XC allows selecting the
desired generator.
The following state transition diagram summarizes the behavior of XQIX to control both
generators.

start Cmd:X2OFF/
Cmd:X1OFF/ X2OFF
X2OFF; X1OFF

X1OFF
X2OFF

Cmd:X1OFF/
Cmd:X1OFF/
X2OFF; X1OFF
X2OFF; X1OFF
Cmd:X1OFF/
X2OFF; X1OFF Cmd:X2ON/ Cmd:X1ON/
X1IDLE; X2ON X2IDLE; X1ON
Cmd:X2IDLE/
X1IDLE; X2IDLE
Cmd:X1OFF/
X2OFF; X1OFF Cmd:X1OFF/
Cmd:X1IDLE/ X2OFF; X1OFF
X1IDLE Cmd:X1ON/
Cmd:X2OFF/ X1ON; X2IDLE
X2OFF
X1ACTIVE
X2IDLE
X1ACTIVE
X2OFF
Cmd:X2OFF/
X2OFF Cmd:X1IDLE/
X1IDLE Cmd:X1ON/
Cmd:X1ON/
X2IDLE; X1ON
X2IDLE; X1ON

Cmd:X2IDLE/
X1IDLE; X2IDLE Cmd:X1ON/
Cmd:X2ON/
Cmd:X1IDLE/ Cmd:X1OFF/ X1IDLE; X2ON X2IDLE; X1ON
X1IDLE Cmd:X2ON/ X2OFF; X1OFF
X1IDLE; X2ON
Cmd:X1ON/
X2IDLE; X1ON

Cmd:X1IDLE/
X1IDLE Cmd:X1IDLE/
Cmd:X2IDLE/ X1IDLE
X1IDLE; X2IDLE
Cmd:X2ON/
X1IDLE; X2ON

Cmd:X1ON/ Cmd:X2ON/
X2IDLE; X1ON X1IDLE X1IDLE; X2ON X1IDLE Cmd:X2IDLE/ X1IDLE
Cmd:X2OFF/ X2ACTIVE
X2IDLE X1IDLE; X2IDLE X2ECO
X2OFF Cmd:X2IDLE/
exit ECO/
X1IDLE; X2IDLE
X2IDLE
Cmd:X2ON/
X1IDLE; X2ON
Cmd:X2IDLE/
Cmd:X2OFF/ X1IDLE; X2IDLE
Cmd:X1IDLE/
X1IDLE X2OFF
X1IDLE
X2OFF enter ECO/
X2toECO
Cmd:X1IDLE/
X1IDLE
Cmd:X2IDLE/ enter ECO/
Cmd:X2ON/
X1IDLE; X2IDLE X1toECO
Cmd:X1IDLE/ X1IDLE; X2ON
X1IDLE

Cmd:X2OFF/ Cmd:X1ON/
X2OFF X2IDLE; X1ON

exit ECO/ Cmd:X2OFF/


X1fromECO X2OFF

Cmd:X1IDLE/
X1IDLE
Cmd:X2ON/
Cmd:X2OFF/ X1IDLE; X2ON
X2OFF
Cmd:X2IDLE/
X1IDLE; X2IDLE Cmd:X1OFF/
enter ECO/ Cmd:X2ON/
X2OFF; X1OFF
X1toECO X1IDLE; X2ON
Cmd:X2IDLE/
X1IDLE; X2IDLE

X1ECO exit ECO/


X1ECO X1fromECO
X2OFF
X2IDLE
Cmd:X1OFF/
X2OFF; X1OFF
Cmd:X2OFF/
X2OFF

Cmd:X1ON/
X1ON; X2IDLE

Figure 11.6

11-10 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

The GenXRD X-Ray power supply is made of 4 main sub-units:

Mains XPSI
XrayPowerSupplyInput
Board

XPSCD
XQIX
XrayPowerSupplyControlXRD XPSB
Board Xray
Power
XPSP Supply
XrayPowerSupplyPowerXRD Backplane
Board Board

XPSMD
XrayPowerSupplyMultiplier
Board

Multiplier
Multiplier And
HighVoltage Box Filament
Assembly

Figure 11.7

AA83678 Service Manual ARL GenX 11-11


Chapter 11 GENXRD WORKSTATION

Working principle

Part of XPSMD assembly


230Vac FET1 L D HighVoltage
+400V S

u
i FET2
u C1 +200Vrtm Multiplier L
C
C2
0Vrtm

DSP 1000 M Ohm

XQIX
uP 1000 M Ohm

i
400Vrtm

+200Vrtm
PI PWM
0Vrtm
Filament -HV
Transformer W filament
-e X-Rays
Co anode

Figure 11.8

XPSI (X-Ray Power Supply Input) Water cooled

XPSI power components (Power-Mosfet) are cooled by a combined radiator-half pipes system.

XPSI does convert the 230VAC main power to 360 - 400 VDC.

XPSI does manage the inrush current, the working current limitation and the power factor correction.

XPSCD (X-Ray Power Supply Control XRD)

XPSCD does manage the kV regulation, the emission current regulation of the X-Ray tube. It is the brain of
the generator.

XPSCD is the slave in the dialog process.

XPSP (X-Ray Power Supply Power) Water cooled

XPSP power components (Power-IGBT) are cooled by a combined radiator-half pipes system.

The XPSP board converts the intermediate DC voltage (360 V- 400 V) into AC voltage sources, which
supplies current to the filament transformer and the HV transformer.

XPSMD (X-Ray Power Supply Multiplier XRD) Water cooled


The multiplier assembly is made with the following parts:

XPSMD board
Filament transformer

11-12 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

HV transformer
Multiplier box with voltage inversion

The XPSMD board has three main functions: the filament power supply, the HV measurement and the
emission current measurement.

The filament transformer via the XPSM board is used to supply up to 50 W, maximum 4 A to the X-ray tube
filament.

The HV transformer via the multiplier is used to supply up to 1800 W, maximum 50 kV, 60 mA, to the X-
ray tube anode.

XPSB (X-Ray Power Supply Backplane)

The backplane or motherboard print is a bus supporting 5 slots made of 5 x 32 pins female internal
connectors (J902-J906) and 2 5 x 32 pins female external connectors J901 and J907, of Europe format.

Each board has its specific position in the rack. There are no keys to prevent the connection of a board to the
wrong slot.

AA83678 Service Manual ARL GenX 11-13


Chapter 11 GENXRD WORKSTATION

XPSI Water (X-Ray Power Supply Input)

Figure 11.9

The main functions controlled by the XPSI board are the mains (230VAC) power factor correction for the
current, the mains current limitation and the X-ray lamps control.

The mains voltage is used via a filter and a relay. When the generator switches on, the relay is energized to
close both contacts. After the rectifier diodes, the FET1 is used to open the circuit if the current is above a
current limit. When the FET1 is open, the current flow through the integrated diode D1of the FET circuit,
due to the inductor L. When the current is below a current limit the FET1 will drive.

The current is controlled by two devices:

1. The DSP that does evaluate the current every 50 microseconds


2. A hardware fast comparator referenced to 5 V. 0.2 volts/A. In both cases if the current is excessive the
FET1 is set OFF.

FET1 L D2 +400Vrtm
L
C1
IL FET2 +12kV
230Vac U1 D1 U2 C2

N 0Vrtm

uC DSP

Figure 11.10

The FET2 and the integrated diode D2 of the FET circuit are used to step up the voltage to 400 V +/- 10 %
this voltage is adapted according to the kV output voltage. 360 V for 0 to 30 kV, 360 V to 420 V
proportionally from 30 kV to 60 kV).

11-14 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Each 50 µs the current “IL”, the voltages “U1” and “U2” are evaluated in order to define the FET2 OFF/ON
conduction time. The maximum OFF/ON time is 25 us.

The current must be relative to the input voltage

V U1

IL

UD1
U2
FET2
OFF
U1

IL

FET2
ON

25 us

Figure 11.11

During the OFF time the current flows through D2 to load the capacitors C1 and C2. The C1/C2 voltage is
measured by the DSP and compared to the set-point value. The On time is regulated relative to this
comparison.

The proportional 400V working voltage will be used to supply:

− The IGBTs on the XPSP power board, generating the anode high voltage, via the transformers and the
multiplier.
− The IGBTs on the XPSP power board, generating the cathode (filament) heating current.

AA83678 Service Manual ARL GenX 11-15


Chapter 11 GENXRD WORKSTATION

If the current is excessive, the driver IC 4 is shut off:

− By the DSP with the CurrentLimit signal dropping to 0 V,


− By the hardware comparator made of IC 8B and transistor Q2; If the current sampling voltage, 0.2 V/A,
does exceed the V Ref, Q2 is locked.

Proportional 400 V

U2 400V Monitoring
400 V

360 V

HV
30 kV 50 kV
Figure 11.12

The other functions controlled by this board are


♦ The low voltage power supply voltages monitoring,
♦ The power component temperatures monitoring,
♦ The neutral to earth potential monitoring.
Low voltages and temperatures
The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.

The neutral to earth potential monitoring

The neutral potential does cross a resistor divider before to be applied to the neutral to ground comparator IC
6.

This device allows the DSP to detect a floating neutral or a neutral-phase inversion.

Remark: By water, for GenXRD WORKSTATION application.

11-16 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

XPSP Water (X-Ray Power Supply Power)

Figure 11.13

This power board contains the switching components (IGBTs) for the high voltage transformer and for the
filament power supplies.

Part of XPSP HV
Transformer
FET1 L IGBT1 C3
230Vac D +400V
HighVoltage

C1
u
i FET2
u +200Vrtm
Multiplier
C2 IGBT2
0Vrtm

DSP 1000 M Ohm

XQIX
uC 1000 M Ohm

Part of XPSP
i
IGBT3
400Vrtm

PI PWM +200Vrtm

0Vrtm Filament
IGBT4 Transformer

Figure 11.14

AA83678 Service Manual ARL GenX 11-17


Chapter 11 GENXRD WORKSTATION

The HV switching works at 20 kHz. The transformer primary is connected to the C1/C2 point which
corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V).

The control board commands alternately the IGBT1 and IGBT2. The On time duration depends of the output
power needed. By this way, the transformer primary receives, via C3, plus and minus 200 V referenced to
the fix 200 VRTM.

11-18 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Chronogram for low power case:

UC3-0Vrtm

+400V

+200V

0 t

IGBT1
Command

IGBT2
Command

Figure 11.15

Chronogram for high power case:

UC3-0Vrtm

+400V

+200V

0 t

IGBT1
Command

IGBT2
Command

Figure 11.16

AA83678 Service Manual ARL GenX 11-19


Chapter 11 GENXRD WORKSTATION

The transformer inductor and C3 capacitor make a resonant circuit. By this way the current look like
sinusoidal:

UC3-0Vrtm

+400V

+200V

0 t
Transformer
Primary
current

Figure 11.17

The filament power supply switching works at 40 kHz. The transformer primary is connected to the C1/C2
point which corresponds to 200V referenced to the 0 VRTM (referenced to the mains like the 400 V).

The control board commands alternately the IGBT3 and IGBT4. The On time duration depends of the output
power needed (max. 50 W). By this way, the transformer primary receives positive and negative 200 V
referenced to the fix 200 VRTM.
See after the voltage and the current at the primary of the filament transformer:
Filament
transformer

+200V

-200V t

Filament
Transformer Primary
current

0 t

Figure 11.18

11-20 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

The low voltage power supply voltages and the temperatures are check in “watchdog” mode. Periodically, in
background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.

The six-pack IGBT is cooled by water, for GenXRD application.

XPSMD Water (X-Ray Power Supply Multiplier XRD)

Part of XPSMD assembly


FET1 L D
230Vac
+400V
HighVoltage

u
i FET2
u C1 +200Vrtm Multiplier
C2
0Vrtm

DSP 1000 M Ohm

XQIX
uP 1000 M Ohm

i
400Vrtm

+200Vrtm
PI PWM
0Vrtm
Filament
Transformer

Figure 11.19

The XPSM assembly has three main functions.

♦ the HV supply
♦ the emission current measurement
♦ the filament current supply
The 400V from the IGBTs on the XPSP board, via XPSM board goes to the HV transformer. At the
transformer secondary the voltage changes according to the IGBT1 and IBGT2 ON time.

The maximum voltage is near 12000 V. The full wave multiplier increases five times this voltage to reach at
maximum - 60000 V. Two dividers are used to monitor the high voltage. The command board uses one of
them to control the HV via the servo loop function accordance with the kV set point. The second is used to
control the first via a subtraction function.

The emission current is measured with the current absorbed by the multiplier. The emission current
measurement allows to indirectly regulating the filament heating current, thus to match with the desired
emission current.

The low voltage power supply voltages and the temperatures are checked in “watchdog” mode. Periodically,
in background task, the control board accesses to the 8 channels 8 bit analogue to digital converter (ADC). A
fatal alarm will be issued if the values are not coherent or if it is not possible to communicate with the ADC,
the generator is immediately shut down.

AA83678 Service Manual ARL GenX 11-21


Chapter 11 GENXRD WORKSTATION

XPSCD (X-Ray Power Supply Control XRD)

Figure 11.20

XPSCD controls the various functions of the X-Ray generator:

♦ the high voltage regulation


♦ the emission current regulation
♦ the filament heating current regulation

HV
Transformer
230Vac FET1 L D +400Vrtm
HighVoltage
C1
u
i FET2
+200Vrtm
Multiplier
u
C2
0Vrtm

DSP 1000 M Ohm

XQIX
uP 1000 M Ohm

XPSCD
400Vrtm
Part
+200Vrtm
PI PWM
0Vrtm Filament
Transformer

Figure 11.21

11-22 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

HV servo-loop control

The control board receives the asked HV from instrument master board via XQIX. The asked HV is
compared with the real HV to generate the HV error, which is used through the PID corrector to PWM duty-
cycle.
The PWM signals control the power stage on XPSP board to supply the right voltage according to the HV
error. The XPSM assembly transformer and multiplier increase this voltage to generate the real HV. The
DSP evaluates the values each 50 us.

uC DSP
HV
HV Multiplier
Power
set point + PID PWM Box
- Stage

HV
D HighVoltage 18762
A

Part of Part of
Part of XPSP board XPSMD assembly
XPSCD board

Figure 11.22

Filament current servo-loop control


The control board receives the desire current from the emission current servo-loop control part. This value is
converted, via a digital-to-analogic converter, into analogic voltage. This voltage corresponds to the needed
current (0.266V/A).

The rest of the bloc diagram works like a controlled current source. I.e., the filament current changes
according to the control voltage. 1 volt corresponds to 1 ampere. The set-point voltage is compared with the
current sensor voltage to control, via the proportional corrector, the PWM duty-cycle at 40 kHz. The power
board contains the power stage, which generates the power for the transformer. The transformer is used to
increase the current to the desired level (up to 4 A). Both the filament heating current and filament voltages
are monitored.

Emission current servo-loop control


The control board receives from the instrument master board the instruction for the desired emission current.
This value is compared, via an analogic-to-digital converter, to the inverted value, which corresponds to the
current absorbed in the multiplier. The comparison result is used through the PID corrector to control the
filament current power supply part. The µC evaluates the values each 50 ms.

AA83678 Service Manual ARL GenX 11-23


Chapter 11 GENXRD WORKSTATION

uC
Emission
current
set point
+
-
PID Filament current power supply

HV
Power Multiplier
Stage

D -1
A
R
Part of Part of
Part of XPSCD board XPSP board XPSMD assembly

Figure 11.23

Power Factor Correction control


The DSP receives from the µC the desired intermediate DC working voltage (360 to 400 V) according to the
asked high voltage (HV). During power off state this value stays to zero. When the generator switches on,
the value changes to 360 V and then changes according to the asked output voltage (kV). This value is
compared to the real intermediate voltage to calculate the servo-loop error.

The rectified input voltage is weighted by this voltage error to become the current reference for a second
servo-loop. The real current, to generate the current error, is compared to this reference. The GPFC corrector
uses the current reference, the real current and the current error values to control the PWM duty-cycle, which
controls the Power MOS FET (FET2).

The FET1 is controlled by three ways. The first one is the intermediate voltage via a P corrector. The PWM
duty-cycle changes according to intermediate voltage. The second one is made on the real current value and
a comparator with a current threshold. At switching power on, this current threshold corresponds to 2
amperes to limit the inrush current. When after 2 seconds, the intermediate voltage exceeds 320 V, the
current threshold changes to 40 Amperes. This corresponds to the maximum acceptable current during power
on state. The last way controls the FET1 by hardware. The real current is compared with a reference value.
When the real current exceeds 40 amperes the FET1 is switched off. This circuit allows reacting faster than
the DSP, where its reaction time is related to the sampling period. The low voltage power supply voltages
and the temperatures are check in “watchdog” mode. Periodically, in background task, the control board
accesses to the 8 channels 8 bit analogic to digital converter (ADC). A fatal alarm will be issued if the values
are not coherent or if it is not possible to communicate with the ADC.

11-24 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Functions of the µ C 80C196KR

Figure 11.24

AA83678 Service Manual ARL GenX 11-25


Chapter 11 GENXRD WORKSTATION

Functions of the 80c196kr µ Controller ports


Port P0 used as A/D converter for the control of the water status:

Port State Designation Remark


P0 0 analogic primary water temperature at the input From 5°C to 40°C
P0 1 analogic primary water temperature at the output From 5°C to 45°C
P0 2 analogic deionised water temperature at the input From 5°C to 50°C
P0 3 analogic deionised water temperature at the output From 5°C to 53°C
P0 4 analogic Deionised water level. Used in ON/OFF mode
P0 5 analogic deionised water resistivity driver 3.3 V (>3 < 4.5)
P0 6 analogic deionised water resistivity From 550 KΩ to Infinite
P0 7 analogic 2.5 Volt Reference Voltage Not used
Port Address: 1FDAH

Port State Designation Remark


P1 0 I
P1 1 I /HvArc Arcs detection from XPSM high voltage divider
P1 2 I /CuArc Arcs detection from XPSM negative emission current -
P1 3 I /MaxFilCur Filament heating current limitation max. If >14 A
P1 4 I FilCurSD Filament heating current Shut Down. If >14.5 A
P1 5 O WdToggle Watch Dog activity supervision.
P1 6 I /AdcDor0 ADC Data Output Ready
P1 7 I AdcEoc0 ADC End Of Conversion
Port Address: IN:1FCFH OUT:1FCDH

Port State Designation Remark


P2 0 O TXD XSN serial data transmission
P2 1 I RXD XSN serial data reception
P2 2 I Adcsars ADC in conversion state
P2 3 O XpsReady XPS presence acknowledge message to XQM
P2 4 O AdcConv0 Start Conversion
P2 5 O XpsOk XPS µ C OK state (LED2 GREEN ON)
P2 6 O XpsNotOk XPS µ C failure NOT OK state (LED1 RED ON)
P2 7 O ClkOut EPLD 610 clock

Port Address: IN:1FD6H OUT:1FD4H

Remarks: P3.0 to P3.7 AD0 to AD7 bus


P4.0 to P4.7 AD8 to AD15 bus

11-26 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Port State Designation Remark


P5 0 O ALE Address Latch Enable
P5 1 O EnWd Enable Watch Dog
P5 2 O /WR Write
P5 3 O /RD Read
P5 4
P5 5 O SsioTxSync Serial link communication transmit synchronisation
P5 6 Ready Wait State Signal
P5 7 I EPROM’s Serial EEPROM Chip Selection Signal IC12
Port Address: IN:1FF7H OUT:1FF5H

Port State Designation Remark


P6 0 O /XsnTxDEn XSN data transmission enable for serial port and display
P6 1 O Adc0Cs0 Miscellaneous 8×12 bits ADC12138 Chip Select IC 3
P6 2 I Adc0Cs1 Miscellaneous 8×8 bits ADCO838 Chip Select IC 7. First AD
P6 3 O Adc0Cs2 Miscellaneous 8×8 bits ADCO838 Chip Select IC 8. Sec. AD
P6 4 O SClk ADCO 838 and ADC12138 Clock
P6 5 O SdatIn ADCO 838 and ADC12138 Data out
P6 6 I SClk ADCO 838 and ADC12138 Clock
P6 7 I SdatOut ADCO 838 and ADC12138 Data in
Port Address: IN:1FD7H OUT:1FD5H

AA83678 Service Manual ARL GenX 11-27


Chapter 11 GENXRD WORKSTATION

Functions of the PPI ports

Figure 11.25

PPI 1 82C55 IC 10
Port State Designation Remark
PPI1 PA 0 I /Int1 Pressostat Active if pressure < 25 KPA
PPI1 PA 1 I /Int2 Electronic rack safety switch
PPI1 PA 2 I /Int3 Tank cover + X-ray tube position switches
PPI1 PA 3 I /Int4 Magazine + load lift cover switches
PPI1 PA 4 I /Int5 X-ray generator rear switch
PPI1 PA 5 I /IntRet 60°C secondary water thermostat
PPI1 PA 6 I /24VPOK +24V instrument OK
PPI1 PA 7 I /WaCoOn Water cooler ON
Port Address: FFC0H

Port State Designation Remark


PPI1 PB 0 I ExRelayPresent Power Distribution Board Main Contactor 24 V Coil Return
PPI1 PB 1 I
PPI1 PB 2 I
PPI1 PB 3 I
PPI1 PB 4 I PrWaFlow /Primary Water Flow Reed Contact
PPI1 PB 5 I DeWaFlow /Deionised Water Flow Reed Contact
PPI1 PB 6 I
PPI1 PB 7 I
Port Address: FFC1H

Port State Designation Remark


PPI1 PC 0 I BuzzerOn Buzzer On Command
PPI1 PC 1 I PrCoolerCtrl Logic Primary Water Valve Command to L293D IC23
PPI1 PC 2 I XrayLampOn 24V X-Ray On Lamp command. Only used on 1.2 kW
PPI1 PC 3 I /DspRst DSP Reset command
PPI1 PC 4 I ExternaRelay Power Distribution Board Main Contactor coil command
PPI1 PC 5 I WaPump Deionised Water Pump command
PPI1 PC 6 I WaCooler External Water Cooler ON command
PPI1 PC 7 I /EnFilCu Filament Heating PWM Generator Enable. IC 20
Port Address: FFC2H

11-28 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Figure 11.26

PPI 2 82C55 IC 11:

Port State Designation Remark


PPI2 PA 0 I SW1.1 From XPSC board switch 1
PPI2 PA 1 I SW1.2 From XPSC board switch 2
PPI2 PA 2 I SW1.3 From XPSC board switch 3
PPI2 PA 3 I SW1.4 From XPSC board switch 4
PPI2 PA 4 I JumperW2.1 From XPSC board W2 jumper 1
PPI2 PA 5 I JumperW2.2 From XPSC board W2 jumper 2
PPI2 PA 6 I jumperW2.3 From XPSC board W2 jumper 3
PPI2 PA 7 I JumperW2.4 From XPSC board W2 jumper 4
Port Address: FF80H

Port State Designation Remark


PPI2 PB 0 O PowerOn XPSI Contactor K1 coil Command
PPI2 PB 1 O EnPower IC26 Latch Input to Output Validation
PPI2 PB 2 O
PPI2 PB 3 O
PPI2 PB 4 O
PPI2 PB 5 O /DspPCe DSP Code EEPROM
PPI2 PB 6 O EnDisRd2 En XPSC Display data Transfer
PPI2 PB 7 O EnXpscDis Enable DSP Data Display Not used
Port Address: FF81H

Port State Designation Remark


PPI2 PC 0 O /DspCs DSP Chip Selection
PPI2 PC 1 O /AdcCs1Ext XPSI 8×8 bits status ADC chip selection IC 3
PPI2 PC 2 O /AdcCs2Ext XPSP 8×8 bits status ADC chip selection IC 1
PPI2 PC 3 O /AdcCs3Ext XPSM 8×8 bits status ADC chip selection IC 2
PPI2 PC 4 O
PPI2 PC 5 O TfsSync DSP Transmit synchronisation
PPI2 PC 6 O SsioRxSync Serial link Communication Receive synchronisation
PPI2 PC 7 O RfsSync DSP Receive synchronisation
Port Address: FF82H

AA83678 Service Manual ARL GenX 11-29


Chapter 11 GENXRD WORKSTATION

AD converters
First AD Miscellaneous 8×8 bits AD converter. To display the different AD input values, use the
terminal mode commands DY
See display information. AD display
commands.doc

Max Filament Current:


14 Amps = 3724 mV = 190 Digits

Safety Filament Current:


14.5 Amps = 3857 mV = 197 Digits

Lamp1:
24V DC for GEN’X 1200 W

Lamp2:
24V DC for GEN’X 1200 W

Filament Current Monitoring:


0.266 V/A
5000 mV = 255 digits
1digit = 19.6 mV

Filament Current Reference Monitoring:


0.266V/A
5000 mV = 255 digits
1digit = 19.6 mV

Second AD Miscellaneous 8×8 bits AD converter.


Low voltages:

Pos. 5V = 204 digits


Pos. 12V = 204 digits
Pos. 24V = 204 digits
Neg. 5V = 51 digits
Neg. 12V = 51 digits

HV Arc Threshold
55 Digits = 1080mV

Current Arc Threshold


0 Digits = 0V

11-30 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

AA83678 Service Manual ARL GenX 11-31


Chapter 11 GENXRD WORKSTATION

AD Miscellaneous 8×12 bits AD converter.


Neg. Emission Current -: From XPSM

3500 mV for 140 mA 25 mV/ mA

Pos. Emission Current: From XPSM


3500 mV for 140 mA 25 mV/ mA

High Voltage: From XPSM


4000 mV for 75 kV 53.333 mV/mA

XPSC temp:
Should not exceed 65°C

DA Converter

Filament Heating Current Reference:


266 mV/A

High Voltage Arc Threshold


55 Digits = 1080 mV

Current Arc Threshold


0 Digits = 0 mV

Max Filament Current:


14 Amps = 3724 mV = 190 Digits

Safety Filament Current:


14.5 Amps = 3857 mV = 197 Digits

X-Ray On Lamps 24 VDC 1200 W Generator

This circuit allows detecting the X-Ray On Lamps


state.
Both OK.
Lamp1 defect. W Alarm 8829
Lamp2 defect. W Alarm 8830
Both defect. F Alarm 8281
Short-Circuit. F Alarm 8281

11-32 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Water control circuit


The parameters monitored by the µ controller are:

♦ The tap or primary water flow. An alarm is immediately issued if the flow drops. However, the default
will only shut down the generator after a delay of 20 seconds.
♦ The tap or primary water inlet and outlet temperatures.
♦ The deionised water flow.
♦ The deionised water inlet and outlet temperatures.
♦ The deionised water resistivity (560k**cm).
♦ The deionised water level
The change of state of cooling safety causes the immediate shut down of the generator by acting on the
generator interlock input. The parameter responsible for the shut down will generate an alarm reported to the
ACS software.

Water parameter requests


Tap or primary water

♦ Flow 2.5 to 4.5 l/min


♦ Pressure <4 bars
♦ Inlet temperature 5°C < Inlet <18°C

Deionised water

♦ Flow 2 l/min
♦ Pressure <3.5 bars
♦ Outlet temperature <60°C

♦ Regulation temperature 36°C (can be modified by user)

♦ Resistivity >550 KΩ

AA83678 Service Manual ARL GenX 11-33


Chapter 11 GENXRD WORKSTATION

Deionised water control circuit


MULTIVIBRATOR ASTABLE
+14V
45Hz VPP = 28V
-14V

22ms
+ DEIONIZED WATER RESI. DRIVER IN
-
+14V
0V
3.3V
tol 3<>4.5V
A/D
PORT 0
1M
DEIONIZED
WATER RESIST.
+ IN
resistivity gauge RX
-

Figure 11.27

To monitor the deionised water resistivity select the status RS 2054

RX VIN VPP Remark


∞ 4.5V 28VPP OPEN CIRCUIT
1 MΩ 2.5V 7VPP
560 KΩ 1.8V 5VPP
0Ω 0V 0V SHORT CIRCUIT.

KV, mA Monitoring Calibration


To perform the calibration type in the terminal mode command menu:
XC 1
The system will automatically perform the calibration.

11-34 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Switches and jumpers

NODE XPSCMXRD
XPSCMXRD and the sub-node XPSCSXRD, controls:
♦ The XRD high-voltage generator activity.
♦ The XRD X-ray tube filament current regulator.
♦ The XRD X-ray tube cooling circuit only if the XRF generator is absent.
♦ Switches and jumpers function on XPSCM.

ON Secondary water cooling regulation inactive


S1,1
OFF Secondary water cooling regulation active
W2 0,1,2,3 Generator type
W2.3 W2.2 W2.1 W2.0
GenXRDV18 1800 W, 50 kV/60 mA
installed removed installed installed
GenXRDF18 1800 W, 50 kV/60 mA
Figure 11.28

Displaying the content of a memory location DY 2,8,100,0,Address,Data Type,Base


Example: displaying the content of port A of PPI 1 DY 2,8,100,0,FFC0H,0,2
Reading the content of a memory location RB 2,8,0,Address
Example: reading the content of port A of PPI 1 RB 2,8,0,FFC0H

AA83678 Service Manual ARL GenX 11-35


Chapter 11 GENXRD WORKSTATION

WORKSTATION GENXRD Trouble Shooting Procedure

GENXRD WORKSTATION ALARMS AND DIAGNOSTIC


As a general rule, to increase the reliability of breakdown diagnostic, it preferable to disable the suspected
Gen’X generator, using the parameters of the Firmware Status 31. The fact is that a fatal error issued by one
generator may shut down both generators. The alarms issued by such an event can lead to a wrong
diagnostic.

8448 XPSC GenXRD F Unknown command received from XSM


S: Reset ICS

8449 XPSC GenXRD F Primary water flow sensor error


8450 XPSC GenXRD F Secondary water flow sensor error
8451 XPSC GenXRD F Secondary water pump error
S: See Step 2

8452 XPSC GenXRD F Desired voltage (kV) too low


Asked kV conditions not allowed by firmware.
S: Select coherent conditions

8453 XPSC GenXRD F Desired voltage (kV) too high


Firmware and XPSC jumper settings are not compatible.
S: Select coherent conditions

8454 XPSC GenXRD F Desired current (mA) too low


Asked mA conditions not allowed by firmware.
S: Select coherent conditions

8455 XPSC GenXRD F Desired current (mA) too high


Firmware and XPSC jumper settings are not compatible.
S: Select coherent conditions

8456 XPSC GenXRD F Desired power too high


Asked power (kV × mA) not allowed by firmware.
S: Select coherent conditions

8457 XPSC GenXRD F Illegal power change mode


Erroneous instructions from XQIX
S: Reset ICS

8458 XPSC GenXRD F Illegal power change speed


KV / mA Out of power change speed out of range. < 1W / min. or > 3000 W / min.
S: Set coherent conditions.

8459 XPSC GenXRD F Illegal anode. Not Used

8460 XPSC GenXRD F Spectrometer pressostat open


8461 XPSC GenXRD F Electronic rack safety switch open
8462 XPSC GenXRD F Spectro tank cover or X-Ray tube safety open
8463 XPSC GenXRD F Load lift cover or cassette magazine safety open
8464 XPSC GenXRD F X-Ray P.S. high voltage connection safety open

11-36 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

8465 XPSC GenXRD F Illegal interlock


See Step 1

8466 XPSC GenXRD F Time-out on X-ray ON ack


Synchronisation error in 80C196KR
S: Reset ICS

8467 XPSC GenXRD F X-ray tube filament open


Filament Curent < 1 A Filament Voltage > 15V
Bad filament connection.
Disconnect filament and measure filament resistivity with a DVM: It should be 0.3 Ohm.
If not filament is defect.
S: Change X-Ray tube.

8468 XPSC GenXRD F Secondary water flow OFF


8469 XPSC GenXRD F Secondary water temperature too high
See Step 2

8470 XPSC GenXRD F Secondary water resistivity driver fault. If > 5 V or < 2V
8471 XPSC GenXRD F Secondary water resistivity gauge fault. Gauge not connected
8472 XPSC GenXRD F Secondary water resistivity too low. If < 300 k Ώ / cm3
The concerned statuses are disabled in the database for GEXRD for WORKSTATION.

8473 XPSC GenXRD F Secondary water inlet temperature gauge fault.


8474 XPSC GenXRD F Secondary water outlet temperature gauge fault.
If value = 0, NTC defect or not connected.
S: Change NTC

8475 XPSC GenXRD F Secondary water level too low.


S: Top up deionised water reservoir

8476 XPSC GenXRD F Water control command not allowed now


In normal operation primitive water control commands not allowed.
S: Don’t use primitive commands in that situation.

8477 XPSC GenXRD F Bad water condition


If any water error
See Step 2

8478 XPSC GenXRD F Unknown status command


Status list in the software not updated.

8479 XPSC GenXRD F X-ray kV calibration slope error


If slope < 0.8 or > 1.2 Ö Out of limits
S: XC

8480 XPSC GenXRD F X-ray kV calibration Offset error


If offset < -1000V or >1000 V Ö Out of limits
S: XC

8481 XPSC GenXRD F X-ray mA calibration slope error


If slope < 0.8 or > 1.2 Ö Out of limits

AA83678 Service Manual ARL GenX 11-37


Chapter 11 GENXRD WORKSTATION

S: XC

8482 XPSC GenXRD F X-ray mA calibration offset error


If offset < -1.0 mA or >1.0 mA Ö Out of limits

S: XC command has to be executed. Corrupted data in the EEPROM of XPSC. Check firmware Status.

8483 XPSC GenXRD F Status: X-ray arc rate too big


The number of arcs does exceed the high value limit defined in the status channel 4107.

IMPORTANT:
Before to have access inside the GenX generator, you have to read carefully the following
caution notes in order to avoid any trouble occurring to the generator or any injury.

Warning
Before to work with the new GenX generator, please read carefully
the following information and caution.

Electrocution risk
DANGER: LETHAL VOLTAGES are present in the GenX generator
whenever the generator is connected to the power line source.
DANGER: LETHAL STORED ENERGY is still present in the GenX
generator after disconnection of the power line source. Wait 15
minutes to be sure that all stored energy is discharged after power
disconnection.

Accessibility inside the GenX generator


The “opening” of the GenX generator, which is necessary for the
maintenance and repair, must be STRICTLY done by a certified
service engineer who has followed a special training course, see
paragraph “Training”.
Thermo Fisher Scientific Switzerland will be not responsible for any
injury occurred on body person in case of no respect of the special
training accreditation given for GenX generator.

HV Multipliers and generator HV connector Insulation problems

When increasing the high voltage, a noise near the multiplier can be heard, and an alarm “PWMTRIP”
Power Off, issued by DSP” is generated, the multiplier assembly can be suspected.

Change the multiplier assembly.

11-38 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

X-Ray tube insulation problems

- Switch OFF the X-ray power supply generator by typing the terminal mode command: #XR
0,0,4,3
- Wait for 1 minute to allow the completion of the secondary deionised water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing
down the two main breakers 16A/32A
- Remove all the panels around the GenX generator.
- Remove the HV cable connector on the back side of the generator. Inspect the connector. No
spark over traces should be visible.

Make sure no wire or any other conductor is in the vicinity of the HV connector.

- Put back the cover of the instrument generator cabinet.


- Inhibit any approach from external people.
- Switch 32 A and 16 A breaker ON.
- Use terminal mode command:
- XR 40,0,4,1,1,1
- Display the high voltage with the terminal mode command:
- DY 2,8,3,1,3
- For this test don’t exceed 5 minutes.
- If the voltage value is stable and if there are no particular noise or alarms concerning insulation,
The X-Ray tube or HV cable is faulty.

HV cable or tube insulation problems

- Switch OFF the X-ray power supply generator by typing the terminal mode command:
- #XR 0,0,4,3
- Wait for 1 minute to allow the completion of the secondary deionized water cooling cycle
- Switch OFF the complete XRF instrument by pushing the emergency red button and pushing
down the two main breakers 16A/32A
- Remove all the panels around the GenX generator.
- Remove the HV cable connector on the back side of the generator and from the X-Ray tube.
Inspect the connectors; on the cable, on the generator and on the X-Ray tube. No spark-over
traces should be visible.
- With a Meger check the HV cable insulation.
- The resistivity must be infinite.
- If it is not change the HV cable and the tube.
Note: The HV cable connector must be cleaned yearly. The connector silicone washers must be
replaced and lubricated yearly.

AA83678 Service Manual ARL GenX 11-39


Chapter 11 GENXRD WORKSTATION

It is recommended to systematically exchange the HV cable every 5 years.

8484 XPSC GenXRD F Secondary water temperature safety open


8485 XPSC GenXRD F Spectro pressostat interlock signal defective
8486 XPSC GenXRD F Electronic rack interlock signal defective
8487 XPSC GenXRD F Interlock signal defective
8488 XPSC GenXRD F Interlock signal defective
8489 XPSC GenXRD F Interlock signal defective
8490 XPSC GenXRD F Interlock signal defective
See step 2 and Step 1

8491 XPSC GenXRD F Unknown X-ray PS type


XPSC Incoherent Jumpers settings.
S: Set jumpers according to technical description.

8492 XPSC GenXRD F SSIO communication timeout


All serial links on XPSC
S: Reset

8493 XPSC GenXRD F Serial AD12 conversion timeout


No serial ADC communication.
S: Reset

8494 XPSC GenXRD F Serial AD12 auto-calibration timeout


Auto calibration not successful
S: Reset

11-40 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

8495 XPSC GenXRD F X-ray ‘kV' out of regulation

kV Regulation

kV > 0 No No test

Yes

Yes kV > 5 kV No

< -20%of value


> 5% of value
> 5% of value

Yes
Yes No

Fatal Alarm 8273 No OK

Figure 11.29

8496 XPSC GenXRD F X-ray ‘mA’ out of regulation

mA Regulation

mA > 0 No No test

Yes

Yes mA > 10 mA No

< -20%of value


> 10% of value
> 10% of value

Yes
Yes No

Fatal Alarm 8274 No OK

Figure 11.30

This alarm may also arise for the following reason:


The firmware status 31 does define the idle state kV, mA, filament heating current and ECO mode
state. In idle state when the XRF generator is in use the XRD generator is set to the parameters
defined in the Firmware status 31. The emission current in idle state, should be equal to 0 mA. The
filament heating current is set to 1 A. If this last parameter has been increased for any reason, the
XRD tube shows an emission current different from 0 mA, thus generating the alarm 8496, trigging
the shut down of both generators.

AA83678 Service Manual ARL GenX 11-41


Chapter 11 GENXRD WORKSTATION

8497 XPSC GenXRD F X-ray power supply interlock open


Pressostat open.
S: Check tank pressure. Check Pressostat.

8498 XPSC GenXRD F X-ray warning lamps broken


See Step 4

8499 XPSC GenXRD F X-Ray Lamp Shorted


See Step 4
Lamps shorted If ADC converter Digits are between 200..255
8500 XPSC GenXRD F X-ray tube filament power supply fault
Heating current is < 1 A and heating voltage is < 2 volts
Fuse F2 10 A of XPSP board
SG3525AN IC 20 of XPSC.
Filament heating power IGBTs on XPSP.

8501 XPSC GenXRD F Invalid X-ray generator calibration value


Manual XC kV or / and mA inputs.
If low kV is 20, the high kV must almost be low + 1kV.
If low mA is 20, the high mA must almost be low + 1mA.
S: Enter the coherent kV / mA calibration values, or use default values.

8502 XPSC GenXRD F Generator mains voltage not present


If main is < 280VAC. In theory the voltage is 230 × 2
See Step 4

8503 XPSC GenXRD F Generator 400V not > 250V


Time out = 5 sec.
See Step 5

8504 XPSC GenXRD F Generator 400V not > 320V


Time out = 2 sec. (Start after the Generator 400V check above).
See Step 6

8505 XPSC GenXRD F Generator temperature IGBT too high


3.6 and 4.2 kW → 65°C Water too hot, environmental temperature too high.
1.2 kW → 120°C Bad venting, fans defect, environmental temperature too high.
Communication error.
S: Reset ICS

8506 XPSC GenXRD F XPSC First 8 bits A/D converter error IC 7


8507 XPSC GenXRD F XPSC Second 8 bits A/D converter error IC 8
Communication error.
XPSI Board not properly plugged.
S: Shut down power, plug board properly
S: Reset ICS

11-42 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

8508 XPSC GenXRD F XPSI 8 bits A/D converter error


8509 XPSC GenXRD F XPSP 8 bits A/D converter error
Communication error.
XPSM Board not properly plugged.
S: Shut down power, plug board properly
S: Reset ICS

8510 XPSC GenXRD F XPSM 8 bits A/D converter error


Communication error.
XPSM Board not properly plugged.
S: Shut down power, plug board properly
S: Reset ICS

8511 XPSC GenXRD F Generator temperature MOSFET M1 too high


8512 XPSC GenXRD F Generator temperature MOSFET M2 too high
If > 65°C
Water too hot.

8513 XPSC GenXRD F Power Off issued by DSP; 400V too low (<320V)
Too fast X-Ray power supply transition. 400 V voltage drop. May happen when changing X-Ray tube
conditions in terminal mode, in “On Fast mode” and “Operational Mode”.
See Step 6

8514 XPSC GenXRD F Power Off issued by DSP; HV too high (>71kV)
KV Regulation error. KV Regulation loop fault.

8515 XPSC GenXRD F Power Off issued by DSP; Current too high (>40A)
After a power up retry the current can rise above 35 A.

8516 XPSC GenXRD F Mains voltage below 195 VAC (after 20 ms)
8517 XPSC GenXRD F Mains voltage above 253 VAC (after 20 ms)
Main power supply fault. The main voltage is checked by the DSP. If out of tolerance the DSP does
immediately shut down the power.
See Step 4

8518 XPSC GenXRD F Unknown reply from DSP


80C196KR no dialog with DSP.
S Reset ICS. GV 2,8,1

8519 XPSC GenXRD F Too many parameters in SSIO


S: Reset ICS.

8520 XPSC GenXRD F Unknown DSP Cmd


S: Reset ICS.

8521 XPSC GenXRD F PWMTRIP -> Power Off


Issued by XPSP due to repetitive power stage problems, where the current is > 100A
Check the IGBT.
Issued by XPSM, due to arcs or spark over on the multiplier assembly of XPSM.
If > 15 kV (> 2V at comparator input) at the secondary of the HV transformer of XPSM. Normal 12kV for
60 kV.

8522 XPSC GenXRD F Generator temperature XPSP too high

AA83678 Service Manual ARL GenX 11-43


Chapter 11 GENXRD WORKSTATION

If > 75 ° C
Bad cooling. Check fans. Laboratory temperature too high.

8523 XPSC GenXRD F Generator temperature Filament too high


If XPSM filament heating > 100 ° C.
Check fans. Laboratory temperature too high.

8524 XPSC GenXRD F Generator temperature HV too high


If XPSM HV conversion unit > 70 ° C
Check fans. Laboratory temperature too high.

8525 XPSC GenXRD F Generator temperature XPSC too high


If XPSC > 65 °C
Check fans. Laboratory temperature too high.

8527 XPSC GenXRD F Filament current sensor incoherency


8528 XPSC GenXRD F Filament voltage sensor incoherency
If before start the sensor delivers a value > 1 A → XPSM Filament current sensor defect. RS 4103
If before start the sensor delivers a value > 10 V → XPSM Filament voltage sensor defect. RS 4131
Measured values out of range, not coherent.

8529 XPSC GenXRD F Status: XPSI +5V too low <4V


8530 XPSC GenXRD F Status: XPSI +5V too high >6V
8531 XPSC GenXRD F Status: XPSC +5V too low <4V
8532 XPSC GenXRD F Status: XPSC +5V too high >6V
8533 XPSC GenXRD F Status: XPSP +5Vm too low <4V
8534 XPSC GenXRD F Status: XPSP +5Vm too high >6V
8535 XPSC GenXRD F Status: XPSM +5Vm too low <4V
8536 XPSC GenXRD F Status: XPSM +5Vm too high >6V
8537 XPSC GenXRD F Status: XPSI +12V too low < 8.5 V
8538 XPSC GenXRD F Status: XPSI +12V too high > 15 V
8539 XPSC GenXRD F Status: XPSC +12V too low < 8.5 V
8540 XPSC GenXRD F Status: XPSC +12V too high > 15 V
8541 XPSC GenXRD F Status: XPSM +12V too low < 8.5 V
8542 XPSC GenXRD F Status: XPSM +12V too high > 15 V
8543 XPSC GenXRD F Status: XPSI -12V too low > -15 V
8544 XPSC GenXRD F Status: XPSI -12V too high > -8.5 V
8545 XPSC GenXRD F Status: XPSC -12V too low > -15 V
8546 XPSC GenXRD F Status: XPSC -12V too high > -8.5 V
8547 XPSC GenXRD F Status: XPSM -12V too low > -15 V
8548 XPSC GenXRD F Status: XPSM -12V too high > -8.5 V
8549 XPSC GenXRD F Status: XPSI +24V too low < 19 V
8550 XPSC GenXRD F Status: XPSI +24V too high > 27 V
8551 XPSC GenXRD F Status: XPSC +24V too low < 19 V
8552 XPSC GenXRD F Status: XPSC +24V too high > 27 V
8553 XPSC GenXRD F Status: XPSP +24V too low < 19 V
8554 XPSC GenXRD F Status: XPSP +24V too high > 27 V
8555 XPSC GenXRD F Status: XPSC -5V too low > -6 V
8556 XPSC GenXRD F Status: XPSC -5V too high < -4 V

8557 XPSC GenXRD F XPSI AD Mon400V Error

11-44 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Previous to start, if U > 350 V.


If > 350 V → XPSI voltage monitoring IC 7 Defect.

8558 XPSC GenXRD F XPSC temperature Board overflow


If > 109 ° C. Out of conversion limits. Mainly indicating NTC is defect.

8559 XPSC GenXRD F XPSC temperature Board underflow


If < 9.9 ° C. Out of conversion limits. Mainly indicating NTC is defect.
8560 XPSC GenXRD F XPSI temp. PowerMosfet M1 overflow
8561 XPSC GenXRD F XPSI temp. PowerMosfet M1 underflow
8562 XPSC GenXRD F XPSI temp. PowerMosfet M2 overflow
8563 XPSC GenXRD F XPSI temp. PowerMosfet M2 underflow
All same as above

8564 XPSC GenXRD F XPSP temperature IGBT overflow


8565 XPSC GenXRD F XPSP temperature IGBT underflow
All same as above

8566 XPSC GenXRD F XPSP temperature Board overflow


8567 XPSC GenXRD F XPSP temperature Board underflow
All same as above

8568 XPSC GenXRD F XPSM temp. Filament overflow


8569 XPSC GenXRD F XPSM temp. Filament underflow
All same as above

8570 XPSC GenXRD F XPSM temp. HV overflow


8571 XPSC GenXRD F XPSM temp. HV underflow
All same as above

8572 XPSC GenXRD F Status: XPSP -10Vrtm too low


8573 XPSC GenXRD F Status: XPSP -10Vrtm too high
8574 XPSC GenXRD F Status: XPSP +15Vrtm too low
8575 XPSC GenXRD F Status: XPSP +15Vrtm too high
The -10 Vrtm is measure in ON/OFF mode. If present → + 5V. If not present → 0 V.
The +15 Vrtm is measure in ON/OFF mode. If present → + 5V. If not present → 0 V.
See Step 4

8576 XPSC GenXRD F Neutral to Gnd too low


DY 2,8,11,3,5 (For GenXRD)
RS 4118

8577 XPSC GenXRD F Neutral to Gnd too high


DY 2,8,11,3,5
RS 4118
If potential is > 200VAC.
The neutral has a floating potential.
Check the main power supply parameters.
Install an insulation transformer if necessary with the 0 volt connected to the ground.
Phase / Neutral are inverted.

8578 XPSC GenXRD F XPSI input current sensor error

AA83678 Service Manual ARL GenX 11-45


Chapter 11 GENXRD WORKSTATION

If previous to start the sensor does measure a current > 2 A.


Sensor is defect. If the value is > it means → no current limitation.
8579 XPSC GenXRD F External relay incoherency
8580 XPSC GenXRD F Power On relay incoherency
8581 XPSC GenXRD F No external relay feedback
8582 XPSC GenXRD F No Power relay feedback

See bloc diagram and flow chart below:

Main Power Distribution GenX


Phase 1
External Voltage XPSI XPSP
External Contactor 230VAC/-10vrtm
+15vrtm
Internal
N K4 Voltage
PowerOn
External
K1 24VDC
Voltage
Internal Relay

Phase 2
XPSC

External Relay X-R Safety Contactor


µC
RY2
24VDC

Figure 11.31

11-46 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

PTB-TÜV
SwitchOn XRAY SwitchOff XRAY

External Voltage =
Yes
Present ?

internalRelay = OFF
8385 F_EXTERNAL_RELAY_INCOHERENCY
No -10Vrtm and +15Vrtm = ON

externalRelay = ON Yes

Internal Voltage = Internal Voltage =


Yes Yes
Present ? Present ?

8386 F_POWER_ON_RELAY_INCOHERENCY
No Internal Voltage 320V = ON
No

External Voltage =
No
Present ?
externalRelay = OFF

8387 F_NO_EXTERNAL_RELAY_FEEDBACK
-10Vrtm and +15Vrtm = Off
Yes 24 V ExRelayPresent = OFF

internalRelay = ON

External Voltage =
Present ?

Internal Voltage =
No
Present ?

8388 F_NO_POWER_ON_RELAY_FEEDBACK No
400Vrtm (350V) = OFF
Yes

Continuation of the Continuation of the


process process
externalRelay = OFF

internalRelay = OFF

Figure 11.32

8583 XPSC GenXRD F Primary water already on


Condition: the duty cycle must be >5%
If before start the primary water on signal is at 1 logic.
S: Check read contact. Clean the water bloc.
See Step 1

8584 XPSC GenXRD F Max filament current limit reached


Filament current does exceed 4.0 A

AA83678 Service Manual ARL GenX 11-47


Chapter 11 GENXRD WORKSTATION

Filament insulation error. Short circuit to the ground.


Disconnect filament H.V. connector and measure insulation between filament leads and X-Ray tube body
and/or water pipes. Resistivity with a DVM should show ∞. If not the filament insulation is bad.
Change X-Ray tube.

8585 XPSC GenXRD F Mains voltage missing


If both -10Vrtm and +15Vrtm are missing.
No external Voltage: check breaker 32A or/and F3 on XPSI or and Fi on XPSP or/and ADC of XPSP

8586 XPSC GenXRD F High-Voltage Incoherency


If prior to start 2kV are measured.
XPSM Board not properly plugged, or XPSM Board ADC errors.

11-48 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Installation Procedure of GenXRD chassis for ARL 9900 Series


Workstation
Introduction
This following installation procedure explains you how to install a GenXRD chassis in the ARL 9900 Series
Workstation.

Figure 11.33 View of the GenXRD chassis with a 9900 Workstation

Safety
Read carefully those below cautions:

During the instrument’s unloading from the wood palette and the installation’s work,
take care of the cables’s connectors and cables coming from the instrument, DON’T
DAMAGE THEM.
Before starting the installation’s work, be sure that 9900 Workstation instrument isn’t
connected to the Mains 230VAC.

AA83678 Service Manual ARL GenX 11-49


Chapter 11 GENXRD WORKSTATION

Installation procedure
For the installation of the GenXRD chassis, please follow strictly the procedure below:

Step: Description: Illustration / Remarks


1. Unload the instrument from the transport
wood’s palette.

During the unloading operation and


installation’s work, take care of the cables’
connectors and cables coming from the
instrument:
DON’T DAMAGE THEM .

2. Locate the foot of the 9900 Workstation


instrument at the left rear side and on the
water chassis side.
At this instrument’s foot, install the cable tray,
screwing 2x screws.

3. Install all the cables inside the cable tray.

11-50 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

4. Install the rod inside the cable tray using a


screw as shown in the picture right.

5. Install the GenXRD generator inside the


chassis using 4 screws located on the front
panel.

6. Install the GenXRD chassis and align it with


the circular cable tray guide.

Detail of the rotative part between GenXRD


chassis and cable tray.

AA83678 Service Manual ARL GenX 11-51


Chapter 11 GENXRD WORKSTATION

7. Connect the 2x cables in the interlock


switch:
- Remove the plastic cover unscrewing 1x
screw

- Install the 2x cables in the interlock


switch:
Blue colour on the right side and Brown
colour on the left side

- Re-install the plastic cover screwing 1x


screw

8. Connect the following cables at the rear


side of the GenXRD generator:
- HV cable with big O-Ring, silicone
washer and applying silicone oil on the
washer and on the HV connector

- Ground cable to the HV multiplier


assembly using 1 screw

- Flat cable with connector J104

- Cable XRD/XPSC with connector M8

- Cable XRD/XPSC with connector J407

- Mains 230VAC cable and secure it with


a plastic clamp as shown in the picture
right.

Connect the water tubes to:


- XPSI board

- XPSP board

- HV Multiplier Assembly

11-52 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

9. Connect 1x Ground cable between:

- HV multiplier assembly using 1 screw

to
- the flat terminal connector located on
the base of the GenXRD chassis.

10. Connect 1x Ground cable between :


- the “free” fast on terminal located on the
base of the GenXRD chassis

to
- the fast on terminal located inside the
cover of the GenXRD chassis.

11. Install the cover of the GenXRD generator


screwing 2x screws that are located on the top
of the cover.

12. At this step, the installation of the GenXRD


generator with its chassis is finished.

IMPORTANT:

Maintenance and repair on the GenXRD


generator can be performed ONLY with
generator removed from its chassis and stand
in HORIZONTAL POSITION as shown on
the right picture

AA83678 Service Manual ARL GenX 11-53


Chapter 11 GENXRD WORKSTATION

Material list, tools and documentation


Material List supplied with the GenXRD chassis:

Pos.: Qty: Description:


1 1 GenXRD chassis including cover
2 1 GenXRD generator
3 1 Cable tray with 2 screws
4 1 Rod (that will be installed inside the cable tray)

Tools:
Field Service Engineer standard tools

Documentation:
Set of ARL 9900 Series instruments schematics

Special packaging for returned material

IN ORDER TO AVOID DAMAGE DURING SHIPMENT, USE STRICKLY ONLY THE


PACKAGING BUILT SPECIALLY FOR THE GENXRD GENERATOR WHICH WILL
ABSORB SHOCKS DURING SHIPMENT.

IMPORTANT: THE SPECIAL PACKAGING MUST ALWAYS BE USED FOR


TRANSPORTATION PURPOSES.

11-54 Service Manual ARL GenX AA83678


GENXRD WORKSTATION Chapter 11

Specifications
# DESCRIPTION VALUES REMARKS
1 Output power 0 to 1.8 KW
2 Input voltage 230VAC +10% / -15%
50/60Hz 1phase
3 Maximum input current and inrush current 28A at 195VAC
4 Maximum input power 5000W
5 Efficiency > 80%
6 Power factor > 0.9 switched mode load type
7 Output voltage -10 to -50KV Without overshoot.
Short-circuit protected.
8 Output voltage vs Input voltage < 50 ppm Over specified range
9 Output voltage vs Output load < 50 ppm 100 % load changes
10 Output voltage vs Temperature 50 ppm/deg. Over specified range
11 Output voltage accuracy < 0.5 % Related to set-point value
(max offset: 250V)
12 Output voltage ripple 0.03 % rms Below 1KHz
13 Output voltage ripple 0.3 % rms Above 1KHz
14 Emission current 0..60 mA
15 Emission current vs Input voltage < 50 ppm Over specified range
16 Emission current vs Output load < 50 ppm Load change According to voltage
change from -20KV to -60KV
17 Emission current vs Temperature 50 ppm/deg. Over specified range
18 Emission current accuracy < 0.5 % Related to set-point value
(max offset 0.08 mA)
19 Filament power supply voltage 0..14 Vac Over voltage protected
to 14V
20 Filament power supply current 0..4 A short-circuit protected
21 Filament power supply current Output 1 % rms Below 100 Hz
voltage ripple
22 Filament power supply current Output not critical Above 100 Hz
voltage ripple
23 Operating temperature range 0..35deg.C Inlet Air Temperature
24 Cooling system Forced air
30 High voltage connector Type II, US The 3 terminals C, L, S connected
FEDERAL to the high voltage
STANDARD H72,
3 PINS
31 Main connection 6c. WAGO connector
32 X-ray ON Lamps +24 VDC 24VDC 8W
33 Maximum power change ramp 3KW/min Limited by interface
34 Maximum voltage change ramp 100KV/second Limited by interface
35 Maximum filament current change ramp 2A/second Limited by interface
36 X-ray arcs detection. shut off after 3 arcs in 10 seconds
37 Control and command connector 96 pole cable
38 Over temperature protection

AA83678 Service Manual ARL GenX 11-55


Thermo Fisher Scientific (Ecublens) SARL,
Switzerland is ISO certified.

In Switzerland:
Thermo Fisher Scientific +41 (0)21 694 71 11
En Vallaire Ouest C, Case Postale [email protected]
CH-1024 Ecublens, Switzerland

Find out more at thermofisher.com/elemental

© 2020 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher
Scientific and its subsidiaries unless otherwise specified.

You might also like