How To Upgrade CW in DC-3 - 1 0 Version

The document provides installation instructions for adding a CW module to a Mindray DC-3 ultrasound system. It describes disassembling the exterior of the machine and interior components to access the installation location for the CW board, and installing the necessary software.

Uploaded by

Jairo Manzaneda
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
113 views18 pages

How To Upgrade CW in DC-3 - 1 0 Version

The document provides installation instructions for adding a CW module to a Mindray DC-3 ultrasound system. It describes disassembling the exterior of the machine and interior components to access the installation location for the CW board, and installing the necessary software.

Uploaded by

Jairo Manzaneda
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 18

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

CW Module installation guide in DC-3


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

REVISION CONTROL

Document Title: CW Module installation guide in DC-3


Author: Mindray Service Department

Date By Action Pages


June 26, 2008 Klaus Hu New document

Review/Approval History

Date By Action Pages

Last Modified: June 17, 2007 Page 2 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Contents
Page
1 BEFORE DISASSEMBLY ................................................................................................................................4
THINGS REQUIRED:.....................................................................................................................................................4
NOTE1: CERTAIN ESD PROTECTION IS NEEDED. ........................................................................................................4
NOTE2: DO NOT CUT OFF THE CABLE HOOPS ............................................................................................................4
NOTE3: MAKE SURE IT SHOULD BE ESPECIALLY CAREFUL WHEN YOU UNPLUG PART FROM OTHER PART AND DO THE
REVERSE. ................................................................................................................................................................... 4

2 OPEN THE MACHINE .....................................................................................................................................5

3 ADDING CW BOARD ..................................................................................................................................... 17

4 CW SOFTWARE INSTALLATION .............................................................................................................. 18

Last Modified: June 17, 2007 Page 3 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

1 Before disassembly

Things required:
a. CW board
b. CW license key disk

Note1: Certain ESD protection is needed.


Electro Static Discharge is very harmful to the boards inside the system. Please adopt certain
measures to protect the boards from ESD especially in dry environment.

Note2: DO NOT cut off the cable hoops


There are many cable hoops attached to the shield covers. Please DO NOT cut them off, otherwise, it may
influence the cable layout when you assemble the system.

Note3: Make sure it should be especially careful when you unplug part from
other part and do the reverse.

Last Modified: June 17, 2007 Page 4 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

2 Open the machine


Before you can see the inner structure of DC-3, please remove the Rear Cover.
Step 1: Remove all the Silicon Rubber Caps on the Rear Cover

Step 2: Remove all the Screws on the Rear Cover


Note: there are three Screws at the bottom of the Rear Cover.

Last Modified: June 17, 2007 Page 5 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step 3: Pull out the Rear Cover

Last Modified: June 17, 2007 Page 6 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Now you can see I/O interface and AC Power Input Module.

I/O interface

AC Power
Input Module

Step 4: Remove six screws on both side of the module

Last Modified: June 17, 2007 Page 7 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step 5: Rotate the panel in the indicated direction

Step 6: Remove the three screws on the right side of the I/O Interface module.

Last Modified: June 17, 2007 Page 8 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step 7: Rotate the panel in the indicated direction

Step 8: Disconnect the soft cable before you take out the panel
Please remove
the fragile cable
carefully, and
there is a Red
Line on the
upper position.

Last Modified: June 17, 2007 Page 9 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step 9: Then we can list up the module and pull it out

2) Pull it out
The metal sticks are
1.
used to locate the
module.

1) Lift the panel up.

Step 10: Disconnect all cables and connectors

Last Modified: June 17, 2007 Page 10 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Remove the screws on the main box

Last Modified: June 17, 2007 Page 11 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step 11: Remove front bottom cover


Remove three screws

Pull out front bottom cover

Last Modified: June 17, 2007 Page 12 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step 12: Remove front cover


Remove plastic caps and screws on both sides of the front cover

Pull out the front cover

Last Modified: June 17, 2007 Page 13 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step 13: Remove the screws on the front side of the unit

Step 14: pull out the main box

Last Modified: June 17, 2007 Page 14 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step 15: Remove screws and pull out the I/O port Board

Step 16: Remove screws and metal cover

Last Modified: June 17, 2007 Page 15 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Last Modified: June 17, 2007 Page 16 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

3 Adding CW board
Add the CW board to its place like following picture.
Be careful of the slot that they fix themselves.

CW board

Last Modified: June 17, 2007 Page 17 of 18


SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

4 CW software installation
(1) Press [Setup] key  [System] menu  [Option] menu

CW software installed

(2) Select CW
(3) Click ‘Install’ menu
(4) Run the license key file in driver F:
(5) System info ‘ install success’.

PS: the license key file is unique related with Serial number and Mac address.

Congratulations, now you have Continuous Wave Doppler.


In CW mode, kindly use phased array transducer 2P2 to do scanning.

Last Modified: June 17, 2007 Page 18 of 18