Prothermo NMT532 is an intrinsically safe multi-signal converter and average temperature sensor for inventory control applications. It consists of precision multi-spot Pt100 temperature elements and provides average temperature measurement and data transmission via local HART communication. It is best suited for accurate inventory measurement when connected to a Tank Side Monitor NRF590 with a Micropilot radar gauge or a Proservo NMS5. The device has high accuracy, is intrinsically safe, compatible with FieldCare, compact, simple to install, and maintenance free.

Uploaded by

dtt000001
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
21 views26 pages

Ti00049gen18 15

Prothermo NMT532 is an intrinsically safe multi-signal converter and average temperature sensor for inventory control applications. It consists of precision multi-spot Pt100 temperature elements and provides average temperature measurement and data transmission via local HART communication. It is best suited for accurate inventory measurement when connected to a Tank Side Monitor NRF590 with a Micropilot radar gauge or a Proservo NMS5. The device has high accuracy, is intrinsically safe, compatible with FieldCare, compact, simple to install, and maintenance free.

Uploaded by

dtt000001
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 26

TI00049G/08/EN/18.

15 Products Solutions Services


71282287

Technical Information
Prothermo NMT532
Intrinsically Safe Multi-signal Converter with
Precision Average Temperature Sensor for In-
ventory Control

Application
Prothermo NMT532 consists of an intelligent local HART signal converter and average
temperature sensor.
For average temperature measurement, it consists of
precision multi-spot Pt100 (Max.6) elements which have fixed interval (2m or 3m).
NMT532 is a highly capable solution for a variety of tank gauging applications and pro-
vides constant average temperature data via local HART communication.
For accurate inventory measurement, it is best suited
connected to Tank Side Monitor NRF590 with Micropilot radar tank gauge or Proservo
NMS5.

Features and Benefits

• High Accuracy

• Intrinsically Safe Device Allowing for Safest Electrical Configuration Possible

• Compatible with FieldCare

• Simple and Economical

• Compact Size and Weight

• High-reliability and Easy Installation

• Maintenance Free
Prothermo NMT532

Table of Contents

Important Document Information . . . . . . . . . . . . . . . . . . . . .3 Operating Condition: Installation . . . . . . . . . . . . . . . . . . . . 11


Notes on Safety Conventions and Symbols . . . . . . . . . . . . . . . . 3 Process Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Recommended Installation Height . . . . . . . . . . . . . . . . . . . . . . 11
Function and System Design . . . . . . . . . . . . . . . . . . . . . . . . . .4 Recommended Stilling Well Installation . . . . . . . . . . . . . . . . . 12
Measuring System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Installation Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Operation Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Anchor Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
System Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Wire Hook+ Top Anchor & Stilling Well . . . . . . . . . . . . . . . . . 13
Connection with NRF590 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Mounting on Pressurized Tank . . . . . . . . . . . . . . . . . . . . . . . . . 15
NMT532 + FMR53x + NRF560 . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Connection with NMS5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 Operating Condition: Terminal Connection . . . . . . . . . . . 16
NMT532 + NMS5+ NRF560 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 NMT532 Terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
NMS5 Terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Input and Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 NRF590 Terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Measured Variables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Number of Elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Mechanical Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 NMT532 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Alarm Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Output Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Human Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Auxiliary Energy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8 Operation Using FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Load HART . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Overvoltage Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8 Certificates and Approvals . . . . . . . . . . . . . . . . . . . . . . . . . 20
Cable Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 CE Mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Supply Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Ex Approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Power Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 External Standards and Guidelines . . . . . . . . . . . . . . . . . . . . . . 20
Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Order Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Performance Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . 9 NMT532 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Temperature Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Temperature Measuring Range . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Reference Operating Conditions . . . . . . . . . . . . . . . . . . . . . . . . . 9 Anchor Weight (High Profile, D120) . . . . . . . . . . . . . . . . . . . . 22
Maximum Measured Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Anchor Weight (Low Profile, Hexagon H41) . . . . . . . . . . . . . . 22
New Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Wire Hook and Top Anchor . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Operating Condition: Environment and Process . . . . . . . 10 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24


Ambient Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Technical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Storage Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Climate Class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Degree of Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Electromagnetic Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Stainless Steel Conversion Table . . . . . . . . . . . . . . . . . . . . . . . . 25
Process Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Process Pressure Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Data Transmission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2 Endress+Hauser
Prothermo NMT532

Important Document Information


Notes on Safety Conventions Symbols for Safety Conventions
and Symbols
Symbol Meaning

DANGER DANGER!
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
A0011189-EN
serious or fatal injury.

WARNING WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
A0011190-EN
serious or fatal injury.

CAUTION CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
A0011191-EN
minor or medium injury.
NOTICE!
NOTICE
This symbol contains information on procedures and other facts which do not result in
A0011192-EN
personal injury.

Symbols for Certain Types of Information


Symbol Meaning
Allowed
Indicates procedures, processes or actions that are allowed.
A0011182

Recommendation
Indicates procedures, processes or actions that are recommended.
A0011183

Forbidden
Indicates procedures, processes or actions that are forbidden.
A0011184

Tip
Indicates additional information.
A0011193

Endress+Hauser 3
Prothermo NMT532

Function and System Design


Measuring System NMT532 is compact and economical. The average temperature sensor consists of six Pt100 elements
at an interval of 2 or 3 meters. Temperature data is transmitted to NRF560 or NMS5 via
intrinsically safe (i.s.) 2-wire local HART signal.

Operation Principle

Local HART Noise Filter


Communication

Power Supply Module

Converter Housing

CPU Module

Flange

Gas Phase

Pt100 Multi- spot


Elements Up to 6 Points

Liquid Phase

Figure 1: Function and System Design

System Design
Data Management Field Process
Temperature

Level Gauge
System

Host Application

Field Interface
NXA820
Tankvision NXA83
DCS Liquid /Gas Temperature
PLC Level
Others

Pressure Gauge

Figure 2: System Layout

Endress+Hauser offers a wide range of solutions to integrate field data into your process management
requirement.
The following diagrams describe some individual solutions according to various Ex concepts. For addi-
tional application requirements, contact local Endress+Hauser representatives.

4 Endress+Hauser
Prothermo NMT532

Connection with NRF590

Ex i HART Loop
(Data Transmission)

FMR Power
(DC, Ex i)
NMT532

Tankvision
sion
FMR 540

Liquid Level

Liquid & Gas


Temperature
Fieldbus Protocol
Power (AC/DC)

Water Interface
(WB)
NRF590

Figure 3: Connection with NRF590

NMT532 + FMR53x + Temperature and level measurement with data collection and calculations via NRF590 allows for opti-
NRF590 mal inventory control. Basic functionality of NMT532 is displayed and configured on NRF590. Detailed
NMT532 functionality and data access can be performed by FieldCare.

NMT532 receives radar level data from NRF590 and then calculates liquid and gas phase average tem-
perature. Calculated and standard data including temperature element raw data and device status are
transmitted to NRF590.

All gathered data in the interface unit is sent to inventory management software, such as
Endress+Hauser tank vision, Tank computer or directly sent to the customer's specific DCS or PLC.

Endress+Hauser 5
Prothermo NMT532

Connection with NMS5

Ex i HART Loop
(Data Transmission)

NMT532
Power (AC/DC)
Ex d HART Loop
(Data Transmission
& Remote Control)

NMS5

Fieldbus Protocol Liquid & Gas


Tankvision Temperature

Power (AC/DC)
Liquid Level

NRF560

Figure 4: Connection with NMS5

NMT532 + NMS5+ NRF560 NMT532 is used most effectively with NMS5 to provide average temperature, level, water
interface, and density measurement.

All the necessary configuration and parameter settings for NMT532 are performed via either NMS5 or
FieldCare.

NMT532 receives liquid level data from NMS5, then calculates liquid and gas phase average
temperature. Calculated data and basic information including raw data for each temperature element
and device status are transmitted to NMS5.

Since NMS5 is a multi-functional device (measurement and data transmission), a Promonitor NRF560
can act as a tank side remote data indicator and controller for NMS5.

All gathered data in the interface unit is sent to inventory management software, such as
Endress+Hauser’s Tankvision, Fuelsmanager, Tank computer or directly sent to the customer's specific
DCS or PLC.

6 Endress+Hauser
Prothermo NMT532

Input and Output


Measured Variables Liquid and gas temperature range: -20 to +100 °C (-4 to +212 °F)
Probe length: 18.5m (60.6 feet) or less

Number of Elements Maximum of 6 (2 m or 3 m interval)

Communication 2 wire, Endress+Hauser local HART protocol to host commanding gauge


• NRF590
• NMS5

Alarm Signal Error information via the following interface and transmission digital protocol. Refer to Operating
Instructions of each device.
• NMS5: BA00401G
• NRF590: BA00256F, BA00257F

Output Signal Temperature data via 2 wire intrinsically safe local HART protocol.

Connection • NMS5
• NRF590

Endress+Hauser 7
Prothermo NMT532

Auxiliary Energy
Load HART Minimum loading for local HART circuit: 250Ω

Overvoltage Protection NMT532 has internal surge arrester which complies with EN/IEC 61000-4-5 (Line to Line 1.0kV).
Connect the metallic housing of the NMT532 to the tank wall or screen directly with an electrically
conductive lead to ensure reliable potential matching.

Cable Entry Wiring of NMT 532 must meet intrinsically safe requirements.
The following cable entries are available:
• Thread NPT 1/2
• Thread M 20

Supply Voltage 16 to 30V: Ex ia


Only for connection to a certified intrinsically safe circuit with the following maximum values

Ui = 30 V
Ii = 120mA Internal capacitance Ci = 7.9 nF
Pi = 1 W Internal inductance Li = 48 H

Power Consumption 6mA

Grounding NMT532 must be grounded to the tank potential before connection to host gauge. All ground connec-
tions must be compliant with local and company regulations, and checked before the equipment is
commissioned.

8 Endress+Hauser
Prothermo NMT532

Performance Characteristics
Temperature Accuracy ± 0.1°C or better (at reference condition)

*Reference
Accuracy of RTD - Temperature Conversion

Temperature Measuring -20 to +100°C (-4 to +212°F)


Range

Reference Operating Condi- • Temperature = +25°C(77°F) ±5 (9°F)


tions • Pressure = 1013mbar abs. ±20mbar abs. (1013hPa abs. ±20hPa abs.,14.7 psi abs. ±0.3 psi abs.)
• Relative humidity (air) = 65% ±20%

Maximum Measured Error Typical statements for reference conditions, include linearity, repeatability, and hysteresis:
• Linearity:
– Temperature: ±0.15°C (0.27°F) + element deviation (based on IEC 60751/DIN EN 60751class A
standard)

New Module NMT 532 employs a completely new electronic module compared to the previous NMT 535.

NMT 532 NMT 535

CPU Performance 16 bit 8 bit

Clock Speed 2.7648 MHz 0.9216 MHz

Memory Capacity (RAM) 20K bytes 176 bytes

EEPROM 2K bytes 256 bytes

Flash Memory 256K bytes 16K bytes

Total # of Print Boards 4 5


Current Consumption
6mA@16VDCEx ia 10mA@16VDC
(Converter + Temp. Probe)

Endress+Hauser 9
Prothermo NMT532

Operating Condition: Environment and Process


Ambient Temperature -40 C° to +85C°(-40°F to +185°F)
Range

Storage Temperature -40 C° to +85C°(-40°F to +185°F)

Climate Class DIN EN 60068-2-38 (test Z/AD)

Degree of Protection • Housing: IP65, NEMA 4X (Converter only, open housing: IP20)
• Probe: IP68

Electromagnetic Compatibil- • Interference Emission: EN 61326, Electrical Equipment Class B


ity • Interference Immunity: EN 61326, Annex A (Industrial)

Interference Emission and immunity apply to each standard shown above when installing probes in
metal and concrete tanks and when using coax probes.

Process Temperature Range Temperature probe: -20 to +100°C (-4 to 212°F)

Process Pressure Limits 1bar (100kPa, 14.5psi)

WARNING

When the pressure inside the tank exceeds 1 bar (100KPa, 14.5psi), stilling well which does not have
any holes or slits must be installed.

Data Transmission 2.5mm coaxial cable & common ground

10 Endress+Hauser
Prothermo NMT532

Operating Condition: Installation


Process Connection The following flange sizes are available:
• NPS 2" Cl.150 RF, 304 flange ASME B16.5
• DN50 PN10 B1, 304 flange EN1092-1 (DIN2527 C)

Recommended Installation
Height
NMT532

Ordered Installation Height:


Below Flange to End of Temp. Probe

Recommended Bottom Clearance:


with Mounting Attachment A,B,C,D : 400mm
from Tank Floor

Figure 5: Recommended Installation

NOTICE
The required bottom clearance of both the temperature probe and WB sensor varies depending on the
anchoring method. Consider the required bottom clearance when ordering NMT 532. See the recom-
mended bottom clearance in the above illustration and/or consult Endress+Hauser representatives for
further information.

The standard location of the lowest temperature element should be set at 500 mm (20") from the bot-
tom of the tank regardless of probe type.

Endress+Hauser 11
Prothermo NMT532

Recommended Stilling Well CAUTION


Installation Datum plate should be mounted on the bottom of the tank below the slotted stilling well (see B ) or
located 300mm (12 inches) or more below the slotted stilling well (see A ).
If the anchor weight is not used when installing stilling well, the water should fill the tank up to the
bottom from the end of the stilling well, enough to allow liquid to enter/exit the pipe.

WARNING

Avoid turbulence in the water so that WB (Water Bottom) sensor will not be damaged.

Slotted Stilling Pipe

Datum Plate B

Slotted

300mm or more
Stilling Pipe

Max. 150mm

Max. 150mm

Datum Plate A
Slot-hole
300mm or more

Ø25mm Slotted
Stilling Pipe

Datum Plate
B
A

Figure 6: Stilling Well

Installation Equipment Contents of Anchoring Hardware: Based on The Choice of "100: Mounting Attachment"

C: D: F:
A: B:
Anchor Weight Tension Wire + Tension Wire +
No Installation Anchor Weight
(Low Profile, Wire Hook + Wire Hook +
Material (High Profile, D120)
Hexagon H41) NPT1 Top Anchor R1 Top Anchor
Bottom hook Bottom hook Bottom hook Bottom hook Bottom hook
Anchor weight Anchor weight Base plate Base plate
Converter + Temp. Probe Sling wire Sling wire Wire hook Wire hook
NPT1 top anchor R1 top anchor
Tension wire Tension wire

12 Endress+Hauser
Prothermo NMT532

Anchor Weight

High Profile Anchor Weight Low Profile Anchor Weight

Element Position #1
(Bottom Element)
Bottom Hook
Clearance below
Temp. Sensor

500mm
400mm
Tank Floor
Figure 7: Anchor Weight
"High profile anchor weight" is the anchor method designed for converter + temperature probe.
"Low profile anchor weight" is the anchor method designed for small tank nozzle [max. 2 inches (50A)].
Temperature probes with anchor weights methods have a recommended clearance 400mm (16").

Wire Hook+ Top Anchor &


Stilling Well

Stilling Well
(Min.Ø2")

Element Position #1
(Bottom Element)

Ø26mm
Clearance below
Bottom Hook
500mm

400mm

Tank Floor

Figure 8: Wire Hook, Top Anchor, Stilling Well

Temperature probes with "wire hook + top anchor" and "stilling well" methods have a recommended
clearance 400mm (16").

*Refer to "Accessories" for details of anchor weight, wire hook, and top anchor.

Endress+Hauser 13
Prothermo NMT532

NMT532 #1 Element Position

NMT532 All Types (Length Below Flange)

NMT532 - + + + + D/F
NMT532 - + + + + A

NMT532 - + + + + B
NMT532 - + + + + C
100mm
400mm

Figure 9: #1 Element Position


Mounting and Element Position of NMT532 Anchor Weight Method

Electrical Compartment

Flange

Uppermost
Element

Flexible Tube

Element Position #1

Clearance: Bottom of
Tank to Bottom Hook Ø26mm(1.02”)
500mm

400mm

Anchor Weight
Anchor Weight (High Profile)
(Low Profile)

Figure 10: Anchor Weight Method

14 Endress+Hauser
Prothermo NMT532

Mounting on Pressurized Pressurized tank is required to install a stilling well to protect the probe from the pressure.
Tank
NMS5

NMT

Ball Valve

Reducer

Measuring Wire

Displacer

Stilling Well

Figure 11: Stilling Well for Pressurized Tank

CAUTION
• When the pressure inside the tank exceeds 1 bar (100KPa, 14.5psi), a stilling well which does not
have any holes or slits must be installed.
• NMT532 is installed in the stilling well from the top of the tank nozzle.
• Cover the bottom of the stilling well and weld it to protect the probe from the pressure.

Welding Part
Figure 12: Welding Part of Stilling Well

Endress+Hauser 15
Prothermo NMT532

Operating Condition: Terminal Connection


NMT532 Terminal NOTICE
NMT532 allows an intrinsically safe HART connection only. Refer to the IS regulation for establishing
wiring & field device layout.

Ground Terminal

FG
Screened Twisted Pair
Z1 AR1
or Steel armored Wire
Z2 Z3

H2- H2+ H1- H1+


to NRF 590
+ Terminal 24, 26 or 28
- Terminal 25, 27 or 29
to NMS5/NMS7
Blind Plug + Terminal 24
CAUTION - Terminal 25
Metal Cable Gland Only:
Shield of HART communication
H2- H2+ H1- H1+
line must be grounded.
Temp. Data
Intrinsically Safe HART on NMT532
+: H2+ Terminal + : H1 + Terminal
- : H2 - Terminal - : H1 - Terminal

Figure 13: NMT532 Terminal

NMS5 Terminal Since NMT 532 is an intrinsically safe instrument, the terminal connection to
Ex i side on local HART connection is allowed on the NMS terminal housing.

+ + COM +
1 L 4 Non IS HART 10 16 Operation 22 4 ... 20 mA
Power Supply Alarm Contact
L _ to NRF or Others _ HOIST _ Channel 2
ARS AC 85 ... 264V 50/60Hz 5 11 17 Contact 23
or Input
2 N DC 20 ... 62V AC20 ... 55V 6 + Digital Output 12
+ 18 STOP 24 +
Alarm Contact Ex i HART
G _ Rack Bus RS485, _ _ HART
ARS 7 13 19 25
Serial Pulse,or HART
+
3 G 8 14 + 20 + 4 ... 20 mA 26
N Alarm Contact
ARS 9
_ _ Alarm Contact 21
_ Channel 1
15
from NMT539

Figure 14: NMS5 Terminal

CAUTION
Do not connect NMT532 HART communication on terminals 4 & 5 on NMS5. These terminals are
designed to connect Ex d HART communication.

16 Endress+Hauser
Prothermo NMT532

NRF590 Terminal

RTD
i.s. Module Wiring

mA

24V
Internaly
HART Interconnected
Sensor as One HART
Fieldbus Loop

For Micropilot
S-series Only!

Figure 15: NRF590 Terminal

NOTICE
NRF590 has three sets of IS HART terminals. These three pairs are looped internally.

CAUTION
Do not connect signal HART lines from NMT532 to terminals 30 & 31. These terminals are designed
to supply drive power of FMR 53x series only.Mechanical Construction

Endress+Hauser 17
Prothermo NMT532

Mechanical Construction
NMT532 Dimensions

142.5mm

215mm

SUS304

SUS316

SUS316L

The specifications of temperature probe


vary depending on the tank height.

Bottom Hook
SUS316
Ø10

Ø26

Figure 16: NMT532 Dimensions

Weight Approx. 8kg


Condition: 6 elements
Temp. probe: 11.5m
Flange: NPS 2" Cl.150 RF, 304 flange ASME B16.5

Material Elements: Class A Pt100, IEC 60751/DIN EN 60751/ JISC 1604


Housing: Aluminum die-casting
Temp. probe: SUS316, SUS316L flexible tube (refer to "Dimension")

18 Endress+Hauser
Prothermo NMT532

Human Interface
Operation Using FieldCare NMT532 can also be operated via FieldCare. These programs support
commissioning, securing of data, signal analysis, and documentation of the instruments.

FieldCare supports the following functions:


• Online configuration of transmitters
• Loading and saving of instrument data (Upload/Download)
• Documentation of measuring points

Endress+Hauser 19
Prothermo NMT532

Certificates and Approvals


CE Mark By attaching the CE mark, Endress+Hauser confirms that the instruments pass the required tests.

Ex Approval
Ex Approval Class

ATEX II 1/2 G Ex ia IIB T4-T6 Ga/Gb


IEC Ex ia IIB T4-T6 Ga/Gb
FM IS Class I, Div. 1, Gp. C, D, T6, T4, Class I, Zone 0, AEx ia IIB, T6, T4
CSA Ex ia Class I, Div.1, Gp. C, D, T6, T5, T4, Ex ia IIB T6, T5, T4
NPESI Ex ia IIB T4-T6

External Standards and IEC 61326 Appendix: A, Immunity according to table A-1
Guidelines
EN 60529
Protection class of housing (IP-code)

EN 61326
Emissions (equipment class B), compatibility (appendix A - industrial area)

20 Endress+Hauser
Prothermo NMT532

Order Information
NMT532
010 Approval:
7 FM IS Cl.I Div.1 Gr. C-D
8 CSA IS Cl.I Div.1 Gr. C-D
B ATEX Ex ia IIB T4 - T6
F IEC Ex ia IIB T4 - T6
G NEPSI Ex ia IIB T2-T6

020 Cable Entry:


B Thread NPT1/2
D Thread M20

030 Process Connection:


1 NPS 2" Cl.150 RF, 304 flange ASME B16.5
2 DN50 PN10 B1, 304 flange EN1092-1 (DIN2527 C)
9 Special version, TSP-no. to be spec.

040 Probe Length; Element; Interval:


022 ...mm; 2x Pt100; 2 m
032 ...mm; 3x Pt100; 2 m
042 ...mm; 4x Pt100; 2 m
052 ...mm; 5x Pt100; 2 m
062 ...mm; 6x Pt100; 2 m
023 ...mm; 2x Pt100; 3 m
033 ...mm; 3x Pt100; 3 m
043 ...mm; 4x Pt100; 3 m
053 ...mm; 5x Pt100; 3 m
063 ...mm; 6x Pt100; 3 m

050 Additional Option:


A Not selected
B Anchor weight, High profile
C Anchor weight, low profile
D Tensioning wire, wire hook, NPT1 top anchor
F Tensioning wire, wire hook, R1 top anchor

NMT532 Complete product designation

Endress+Hauser 21
Prothermo NMT532

Accessories
Anchor Weight (High Pro- Mounting Attachment Option: B
file, D120) This high profile anchor type is designed for Converter + Temperature Probe Versions.

CAUTION
The lowest measuring element (lowest temperature measurement position) must be set at a position
of approximately 400mm (16") from the bottom of the tank.
Ensure that the opening size of the nozzle is 6 inches (150A) or more when installing the high profile
anchor weight from a nozzle at the top of the tank.

Material:
Weight : JIS SS400 (Mild Carbon Steel)
Ring: JIS SS400 (Mild Carbon Steel)

Weight

42mm (1.65")
Approx. 16kg

222mm (8.74")
180mm (7.09")

120mm (4.72")

Figure 17: Anchor Weight for High Profile

Different dimensions, weight, and material for the anchor weight are also available.

Anchor Weight (Low Profile, Mounting Attachment Option: C


Hexagon H41) The low profile anchor weight is mainly designed to secure WB sensor and to measure a range of the
measurable water level. The anchor weight can also be used as an attachment of converters or tem-
perature probes when installing it to a small nozzle that is being used.

Material:
Weight: JIS SS400 (Mild Carbon Steel)
Ring: JIS SS400 (Mild Carbon Steel)

Weight:
Approx. 12kg

1040mm (40.94")
41mm (1.61")

Figure 18: Anchor Weight for Low Profile

22 Endress+Hauser
Prothermo NMT532

Wire Hook and Top Anchor


Material:
Weight: JIS SS400 (Mild Carbon Steel)
Ring: JIS SS400 (Mild Carbon Steel)

Weight : Approx 1.5kg

21mm(0.83")
9mm(0.35")
150mm(5.91")
70mm (2.76”)
200mm(7.87”)

Figure 19: Wire Hook

Actual tensioning can be completed with SUS316 stranded 3mm diameter tension wire between the
wire hook and the top anchor.

Material
Exterior: ADC (Aluminium)
Internal Parts : SUS316, Carbon Steel
Weight: Approx. 1.2kg
190mm (7.48")

52mm (2.05")

R1 or NPT1( Specified by Order Code)

Figure 20: Top Anchor

NOTICE
The standard process connection of the top anchor is R1 or NPT1 threaded connection.

Endress+Hauser 23
Prothermo NMT532

Documentation
Technical Information TI00452G
Proservo NMS5

TI00462G
Promonitor NRF560

Operating Instructions BA01032G


Prothermo NMT532 (Installation Instructions)

Safety Instructions
NMT532 ATEX IEC NEPSI FM CSA

Average Temperature XA00584G XA00581G XA01260G Ex461-852-1 Ex462-875-1

24 Endress+Hauser
Prothermo NMT532

Appendix
Stainless Steel Conversion The stainless steel material used in products of Endress + Hauser Japan normally have expressions
Table according to Japanese industrial standards, such as JIS or TIIS. Each country or region may have differ-
ent expressions.
The following conversion table contains the expression of equivalent stainless steel material based on
the chemical composition and mechanical properties.

Country Standard Expressions

Japan JIS / TIIS SUS304 SUS304L SUS316 SUS316L

Germany DIN 17006 X5 CrNi 18 10 X2 CrNi 18 11 X5 CrNiMo 17 12 X2 CrNiMo 17 13


X5 CrNi 18 12 2 / 1713 3 2

W.N. 17007 1.4301 1.4303 1.4306 1.4401 / 1.4436 1.4404

France AFNOR Z 6 CN 18-09 Z 2CN 18-10 Z 6 CND 17-11 / Z2 CND 17-12


17 12

Italy UNI X5 CrNi 1810 X2 CrNi 1911 X5 CrNiMo 1712 X2 CrNiMo 1712
/ 1713

U.K. BSI 304S15 / 304S16 304S11 316S31 / 316S33 316S11

U.S.A. AISI 304 304 L 316 316L

E.U. EURONORM X6 CrNi 1810 X3 CrNi 1810 X6 CrNiMo 17 12 X3 CrNiMo 17 12


2 / 17 13 3 2

Spain UNE X6 CrNi 19-10 X2 CrNi 19-10 X6 CrNiMo 17- X2 CrNiMo 17-
12-03 12-03

Russia GOST 08KH18N10 03KH18N11 _ 03KH17N14M2


06KH18N11

- ISO 11 10 20 19

- ASME S30400 S30403 S31600 S31603

NOTICE
Since each standard carries its own mechanical and scientific definition, some expressions may not
have the straight conversion from the Japanese standard. Consult with the local authority or legislature
to ensure the proper comparison with the applied standard prior to determining specification.

Endress+Hauser 25
www.addresses.endress.com

You might also like