1) INTENDED USE
HBsAg Elisa is an enzyme immunoassay diagnostic kit for in-vitro qualitative detection of hepatitis B surface antigen
(HBsAg) in human serum or plasma (heparin, citrate or EDTA).
3) TEST DESCRIPTION
HBsAg Elisa is a solid-phase enzyme immunoassay (ELISA= enzyme-linked immuno-sorbent assay ) based on the
"sandwich principle".
The solid phase of the microtiter plate is made of polystyrene wells coated with mouse monoclonal antibodies specific
for HBsAg; whereas guinea pig polyclonal antibody purified by affinity chromatography is used to prepare the anti-HBs•
peroxidase (horseradish) conjugate in the liquid-phase.
When a serum or plasma specimen containing HBsAg is added to the anti-HBs antibody-coated wells together with the
peroxidase conjugated anti-HBs antibody and incubated, an antibody-HBsAg-antibody-peroxidase complex will form on
the wells.
After washing the microtiter plate to remove unbound material, a solution of TMB substrate is added to the wells and
incubated. A color develops in proportion to the amount of HBsAg boud to Anti-HBs. The peroxidase-TMB reaction is
stopped by addition of sulfuric acid. The optical density of developed color is read with a suitable photometer at 450nm
with a selected reference wavelength within 620 to 690nm*1.
Stop Solution
1 bottle
(8) 1N sulfuric acid
STOP SOLN 12 ml
2. The evaluation of the HBsAg Sensitivity Panel PHA807 (BBI, USA) yielded the following results:
Panel No. of panel members Results
BBI HBsAg Sensitivity Panel Equivalent to or better than
PHA807-01~21
PHA807 the competitors.
Catalogue Nr: KAPG4SGE3 PI number : 1701153 Revision Nr : 110627/1
3. The evaluation of the HBsAg Mixed Titer Performance Panel PHA205 (BBI, USA) yielded the following results:
Panel No. of panel members Results
BBI HBsAg Mixed Titer Equivalent to or better than the
PHA205-01~25
Performance Panel PHA205 competitors.
4. The HBsAg Mixed Titer Performance Panel VHA601 (BBI, USA) evaluation yielded the following results:
No. of panel
Panel Results
members
BBI HBsAg Mixed Titer Performance Met BBI’s expected
VHA601-01~06
Panel VHA601 results.
5. The evaluation of the HBsAg Mixed Titer Performance Panel QHA711 (BBI, USA) yielded the following results:
No. of panel
Panel Results
members
BBI HBsAg Mixed Titer Performance Met BBI’s expected
QHA711-01~06
Panel QHA711 results.
6. INTS HBsAg Subtype Panel Test Result:
The results of the evaluation demonstrate that all known HBsAg serotypes (HBsAg subtypes) available in the
INTS HBsAg subtype panel can be detected by both HBsAg Elisa and the reference assay up to 104x ~ 106x
dilutions.
7. Summary table of the evaluation of all tested seroconversion panels:
HBsAg Positive Result From Date of First Blood Collection
Reference HBsAg Difference in Bleed Difference in
Assay HBsAg Elisa Days Bleed #s
PPanel ID (B in days) (panel member
(A in days) (A-B) No.)
PHM907(ay) 50 50 0 0
PHM916(ay) 62 65 -3 -1
PHM920(ad) 26 26 0 0
PHM921(ad) 0 0 0 0
PHM930(ad) 3 3 0 0
PHM933(ad) 7 9 -2 -1
PHM934(ad) 0 0 0 0
PHM935A
9 21 -12 -3
(ad)
PHM935B
231 217 14 -1
(ad)
NabiSB0411 6 6 0 0
PHM903(ad) 6 10 -4 -1
PHM904(ad) 7 18 -11 -1
PHM906(ad) 137 0 137 1
PHM909(ad) 9 9 0 0
PHM910(ad) 35 18 17 1
PHM912(ad) 42 0 42 7
PHM914(ad) 0 146 -146 -1
PHM915(ind) 12 28 -16 -5
PHM917(ind) 36 >43 ->7 ->2
PHM918 (ad) 7 7 0 0
PHM923 (ay) 15 21 -6 -1
PHM924 (ad) 29 35 6 1
PHM925(ind) 8 14 -6 -1
PHM926(ind) 13 15 -2 -1
PHM927(ind) 4 7 -3 -1
PHM928 (ad) 9 9 0 0
PHM929 (ad) 14 18 -2 -1
PHM931(ind) 19 21 -2 -1
PHM932 (ad) 61 61 0 0
PHM935A
21 28 -7 -2
(ad)
----- ----- Sum = -13 -----
Over all
Average -13/30 = 0.4 days -----
Catalogue Nr: KAPG4SGE3 PI number : 1701153 Revision Nr : 110627/1
Summary of the evaluation of all tested seroconversion panels:
The results have shown that the HBsAg Elisa is nearly equivalent to the reference assay. On average the HBsAg
Elisa test is only 0.4 days later in the 30 tested seroconversion panels in comparison to the reference assay.
5.11.3) Mutants
1. Mutant Panel
A Hepatitis B Pre Core Mutant panel of Teragenix, USA was tested. HBsAg Elisa can detect the results while
panel’s dilution ratio is 1:2 or 1:5.
This panel consists of 3 samples Core Promoter (A1762/G1764) wild type; PreCore codon 28 mutant/wild type
infection, of 4 samples Core Promotor (A1762/G1764) wild type; PreCore codon 28 mutant, of 2 samples Core
Promotor (T1762/A1754) MUTANT/(A1762/G1754) wild type mixed infection.
2. Mutant samples
The detection rate of HBsAg Elisa is 17/21 which is better than reference kit’s (14/21).
Reference Reference
Mutation HbsAg Elisa Mutation HbsAg Elisa
kit kit
Int Int Int Int
113A114 positive positive C137W positive negative
P120L positive positive C139Y positive positive
R122I negative negative K141E positive positive
R122T positive positive P142L/G145R negative negative
E122I positive positive P142L positive positive
T123N positive positive D144G positive negative
122RA123 negative negative G145K positive negative
C124R negative negative C147T/C148T positive positive
T131I positive positive E164D positive positive
T131P positive positive S174N positive positive
Y134S positive positive
Detection Detection
8/11 8/11 9/10 6/11
Rate Rate
5.11.6) Precision
1. Intra-assay reproducibility
2. Inter-assay reproducibility
The precision evaluation experiments were performed over 10 operating days is using five serum samples (with
borderline positive and clearly above cutoff value HBsAg levels).
Standard Coefficient of
Sample ID N Mean
Deviation Variation
F1 10 0.0609 0.0185 30.41
F2 10 0.0770 0.0189 24.56
F3 10 0.1000 0.0245 24.51
F4 10 0.1786 0.0462 25.87
F5 10 0.6107 0.1412 23.12
The calculated CV´s ranges between 30.4 for an HBsAg negative sample and 23.1 % for an HBsAg low positive
sample (= acceptable values for the inter-assay imprecision of an immunoassay in microtiter plate format).
5.11.8) Traceability
The DIAsource HBsAg Master Calibrator has been calibrated against the British Working Standard for HBsAg
(NIBSC-Code: 01/476-006) using the HBsAg Elisa assay. The relative potency (ratio) of the British Working
Standard for HBsAg versus the DIAsource HBsAg Master Calibrator is 4.081 (3.853-4.325 95% CI). The
concentration of the Positive Control of HBsAg Elisa assay has been determined against the DIAsource HBsAg
Master Calibrator and was established with 42 IU/ml ± 20%.
↓
Add 50 µl of Anti-HBs‧ Peroxidase Solution into each
reaction well, except one blank.
↓
Incubate the plate at +37±1°C for 80 minutes.
↓
Wash the plate.
(Choose one of the following two methods for enzymatic
reaction with color development)
↙ ↘
Mix equal volumes of Add 50 µl of Chromogenic
Chromogenic TMB TMB concentrate to wells
concentrate and Substrate and then add 50 µl of
buffer. Add 100 µl of the Substrate buffer. Mix well,
mixed solution to wells. gently.
↘ ↙
Incubate at R.T. for 30 minutes.
↓
Add 100µl of 1N sulfuric acid into each well.
↓
Determine absorbance using 450 nm as reading wavelength
with 620-690nm reference wavelength
6) NOTES
*1 The reference wavelength of the photometer to be used can be 620 nm to 690 nm. However, the user should validate
the photometer in combination with HBsAg Elisa before use.
*2 Incomplete inactivation of hepatitis B virus after heat treatment at +60°C for 10 hours, J. Infect. Dis. 138:242-244.
7) BIBLIOGRAPHY
1. Blumberg BS, Alter HJ, Visnich S. A „new“ antigen in leukemia sera. JAMA, 1965;191:101- 106.
2. Dane DS, Cameron CH, Briggs M. Virus-like particles in serum of patients with Australia-antigen-associated
hepatitis. Lancet 1970; 1: 695 - 698.
3. Aach RD, Grisham JW, Paker CW. Detection of Australia antigen by radioimmunoassay. Proc Natl Acad Sci. USA
1971;68:1056-1060.
4. Kim CY, Tikes JG. Purification and biophysical characterization of hepatitis antigen. J Clin Invest. 1973; 52:1176-
1186.
5. Wolters G. Kuijpers LP, Kacaki J, Schuurs AH, Enzyme linked immunosorbent assay for hepatitis B surface antigen.
J Infect. Dis 1977;136:Suppl 311-377.
6. Shih JW, Cote PJ Jr, Dapolito GM, Gerin JL. Production of monoclonal antibodies against hepatitis B surface antigen
(HBsAg)by somatic cell hybrids. J Virol Methods. 1980;1:257-273.
7. Hoofnagle JH, Di Bisceglie AM. Serologic diagnosis of acute and chronic viral hepatitis. Semin Liver Dis.
1991;11:73-83
1) BUT DU DOSAGE
L’HBsAg Elisa est un kit de diagnostic d'essai immuno-enzymatique pour la détection qualitative in vitro de l’antigène
de surface de l’hépatite B (HBsAg) dans le sérum ou le plasma humain (hépariné, citraté ou sur EDTA).
3) DESCRIPTION DU TEST
L’HBsAg Elisa est un essai immuno-enzymatique en phase solide (ELISA = enzyme-linked immuno-sorbent assay)
basé sur le « principe du sandwich ».
La phase solide de la plaque de microtitration est composée de puits en polystyrène tapissés d’anticorps monoclonaux de
souris spécifiques de l’HBsAg ; par ailleurs, un anticorps polyclonal de cochon d’Inde purifié par chromatographie
d’affinité est utilisé pour préparer le conjugué anti-HBs• peroxydase (de raifort) dans la phase liquide.
Lorsqu’un échantillon de sérum ou de plasma contenant des HBsAg est ajouté avec l’anticorps anti-HBs conjugué
contenant de la peroxydase aux puits tapissés d’anticorps anti-HBs et incubé, un complexe anticorps-HBsAg-anticorps-
peroxydase se forme sur les puits.
Après lavage de la plaque de microtitration pour éliminer le matériel non lié, une solution du substrat TMB est ajoutée
aux puits et la plaque est incubée. Une couleur apparaît, proportionnelle à la quantité d’HBsAg liés aux anti-HBs. La
réaction peroxydase-TMB est stoppée par ajout d’acide sulfurique. La densité optique de la couleur développée est lue à
450 nm à l’aide d’un photomètre adapté avec une longueur de référence comprise entre 620 et 690 nm*1.
Solution de lavage
conc. (20X)
Tampon phosphate concentré avec 1 flacon
(7)
WASH SOLN CONC du Tween-20 58 ml
Solution d'arrêt
1 flacon
(8) Acide sulfurique 1N
12 ml
Temp. de
Kit/composants État Stabilité
stockage
Original 18 mois
Kit HBsAg Elisa +2 à +8 °C
Après ouverture 1 mois
Original 18 mois
Contrôle HBsAg positif +2 à +8 °C
Après ouverture 1 mois
Original 18 mois
Contrôle HB négatif +2 à +8 °C
Après ouverture 1 mois
Original 24 mois
Plaque anti-HBs +2 à +8 °C
Après ouverture 1 mois
Original 18 mois
Solution du conjugué Anti-HBs‧HRPO +2 à +8 °C
Après ouverture 1 mois
Temp. Original 24 mois
Solution de lavage concentrée (20X)
ambiante Après ouverture 1 mois
Temp.
Diluée 2 jours
Solution de lavage diluée 20X ambiante
+2 à +8 °C Diluée 1 semaine
Original 24 mois
Chromogène TMB concentré +2 à +8 °C
Après ouverture 1 mois
Initial 24 mois
Tampon du substrat +2 à +8 °C
Après ouverture 1 mois
Temp.
Mélange substrat TMB Mélange 6 heures
ambiante
Temp. Original 24 mois
Acide sulfurique 1N
ambiante Après ouverture 1 mois
5) Mode d’emploi
5.1) Avertissements :
5.1.1) Ce kit de réactifs est uniquement à usage professionnel.
5.1.2) Ce kit de réactifs est uniquement destiné à un diagnostic in vitro.
5.1.3) Placer l’ensemble des réactifs du kit et des échantillons à température ambiante (+20 à +30 °C) et mélanger
soigneusement avant utilisation.
5.1.4) Ne pas utiliser les réactifs après la date d’expiration.
5.1.5) Ne pas échanger les réactifs entre différents lots.
5.1.6) Ne pas porter la pipette à la bouche.
5.1.7) Ne pas fumer ou manger dans les zones où des échantillons ou des réactifs sont manipulés.
5.1.8) Tous les composants du kit et les échantillons doivent être considérés comme potentiellement dangereux pour la
santé. Ils doivent être utilisés et éliminés conformément aux procédures de sécurité de votre laboratoire. Ces
procédures de sécurité incluent probablement le port de gants de protection et éviter de produire des aérosols.
5.1.9) Les échantillons potentiellement infectieux et les déversements ou fuites non acides doivent être
minutieusement nettoyés à l’aide d’hypochlorite de sodium à 5 % ou traités selon votre procédure de contrôle
des risques biologiques potentiels.
5.1.10) Avant d’éliminer les échantillons et les kits de réactifs usagés dans les ordures ménagères, ils doivent être
traités conformément à la procédure locale de traitement des déchets biologiques potentiellement dangereux
ou comme suit :
Les déchets liquides et solides doivent être stérilisés par autoclave à +121 °C pendant au moins 30 minutes.
5.11) Performance
5.11.1) Spécificité diagnostique
Résultats de l’évaluation européenne de la performance pour l’HBsAg Elisa de DIAsource – Réactivité des
échantillons « Donneur » et « Clinique » HBV négatifs.
2. L’évaluation de l’HBsAg Sensitivity Panel PHA807 (BBI, États-Unis) a donné les résultats suivants :
Nbre de membres du
Panel Résultats
panel
Panel de sensibilité HBsAg Équivalents ou meilleurs que
PHA807-01~21
PHA807 (BBI) les concurrents.
3. L’évaluation de l’HBsAg Mixed Titer Performance Panel PHA205 (BBI, États-Unis) a donné les résultats
suivants :
Nbre de membres du
Panel Résultats
panel
Panel de performance d’un
Équivalents ou meilleurs que les
mélange de titres HBsAg PHA205 PHA205-01~25
concurrents.
(BBI)
4. L’évaluation du HBsAg Mixed Titer Performance Panel VHA601 (BBI, États-Unis) a donné les résultats suivants :
Nbre de membres du
Panel Résultats
panel
Panel de performance d’un mélange de Rencontre les résultats du
VHA601-01~06
titres HBsAg VHA601 (BBI) BBI escomptés.
5. L’évaluation du HBsAg Mixed Titer Performance Panel QHA711 (BBI, États-Unis) a donné les résultats suivants :
Nbre de membres du
Panel Résultats
panel
Panel de performance d’un mélange de Rencontre les résultats du
QHA711-01~06
titres HBsAg QHA711 (BBI) BBI escomptés.
6. Résultats du test avec l’INTS HBsAg Subtype Panel :
Les résultats de l’évaluation démontrent que l’ensemble des sérotypes HBsAg connus (sous-types HBsAg)
disponibles dans le panel des sous-types HBsAg INTS peuvent être détectés aussi bien par l’HBsAg Elisa que par
l’essai de référence à des dilutions pouvant aller jusqu’à 10 4x ~ 106x.
7. Tableau de synthèse de l’évaluation de l’ensemble des panels de séroconversion testés :
5.11.3) Mutants
1. Panel mutant
Un panel de mutants dans le PreCore de l’hépatite B de Teragenix, États-Unis a été testé. L’HBsAg Elisa peut
détecter les résultats pour un rapport de dilution du panel de 1:2 ou 1:5.
Ce panel contient 3 échantillons de Promoteur Core (A1762/G1764) de type sauvage ; mutant dans le codon 28
de la région PreCore/infection de type sauvage, 4 échantillons de Promoteur Core (A1762/G1764) de type sauvage ;
mutant dans le codon 28 de la région PreCore, 2 échantillons de Promoteur Core (T1762/A1754)
MUTANT/(A1762/G1754) infection mélangée de type sauvage.
Nº de catalogue : KAPG4SGE3 Nº de PI : 1701153 Nº de révision : 110627/1
2. Échantillons mutants
Le taux de détection de l’HBsAg Elisa est 17/21, soit un meilleur résultat que le kit de référence (14/21).
Kit de Kit de
Mutation HBsAg Elisa Mutation HBsAg Elisa
référence référence
Int Int Int Int
113A114 positif positif C137W positif négatif
P120L positif positif C139Y positif positif
R122I négatif négatif K141E positif positif
R122T positif positif P142L/G145R négatif négatif
E122I positif positif P142L positif positif
T123N positif positif D144G positif négatif
122RA123 négatif négatif G145K positif négatif
C124R négatif négatif C147T/C148T positif positif
T131I positif positif E164D positif positif
T131P positif positif S174N positif positif
Y134S positif positif
Taux de Taux de
8/11 8/11 9/10 6/11
détection détection
5.11.6) Précision
1. Reproductibilité intra-essai
La précision intra-essai a été déterminée à l’aide d’un échantillon de contrôle positif et de deux échantillons de
sérum de patient de différentes concentrations en HBsAg (taux légèrement supérieur au niveau du seuil de
positivité et taux moyen) qui ont été analysés 20 fois en un seul run sur 3 jours.
ID Coefficient de
Série Moyenne Écart type
échantillon variation
1 PC 1 :32 0,7541 0,0857 11,36
1 PS 1 :32 2,2766 0,1571 6,90
1 PS 1 :64 1,3159 0,1168 8,88
2 PC 1 :32 0,9671 0,0358 3,70
2 PS 1 :32 2,7325 0,1025 3,75
2 PS 1 :64 1,5822 0,0888 5,61
3 PC 1 :32 0,9669 0,0909 9,40
3 PS 1 :32 2,6088 0,2367 9,07
3 PS 1 :64 1,6502 0,1499 9,08
Le CV calculé variait entre 3,7 % et 11,36 %. (= valeur acceptable pour un essai immuno-enzymatique sous forme
de plaque de microtitration).
2. Reproductibilité inter-essai
Le CV calculé variait entre 30,4 % pour un échantillon HBsAg négatif et 23,1 % pour un échantillon HBsAg
faiblement positif (= valeurs acceptables pour l’imprécision inter-essai d’un essai immuno-enzymatique sous
forme de plaque de microtitration).
5.11.8) Traçabilité
L’HBsAg Master Calibrator de DIAsource a été calibré par rapport à l’étalon de travail HBsAg britannique
(British Working Standard for HBsAg ) (NIBSC-Code : 01/476-006) à l'aide de l'essai HBsAg Elisa. L’efficacité
relative (rapport) de l’étalon de travail HBsAg britannique par rapport à l’HBsAg Master Calibrator de DIAsource
est de 4,081 (3,853-4,325 IC à 95 %). La concentration du contrôle positif de l’essai HBsAg Elisa a été déterminée
par rapport à l’HBsAg Master Calibrator de DIAsource et a été établie à 42 UI/ml ± 20 %.
7) BIBLIOGRAPHIE
1. Blumberg BS, Alter HJ, Visnich S. A „new“ antigen in leukemia sera. JAMA, 1965;191:101- 106.
2. Dane DS, Cameron CH, Briggs M. Virus-like particles in serum of patients with Australia-antigen-associated
hepatitis. Lancet 1970; 1: 695 - 698.
3. Aach RD, Grisham JW, Paker CW. Detection of Australia antigen by radioimmunoassay. Proc Natl Acad Sci. USA
1971;68:1056-1060.
4. Kim CY, Tikes JG. Purification and biophysical characterization of hepatitis antigen. J Clin Invest. 1973; 52:1176-
1186.
5. Wolters G. Kuijpers LP, Kacaki J, Schuurs AH, Enzyme linked immunosorbent assay for hepatitis B surface antigen.
J Infect. Dis 1977;136:Suppl 311-377.
6. Shih JW, Cote PJ Jr, Dapolito GM, Gerin JL. Production of monoclonal antibodies against hepatitis B surface antigen
(HBsAg)by somatic cell hybrids. J Virol Methods. 1980;1:257-273.
7. Hoofnagle JH, Di Bisceglie AM. Serologic diagnosis of acute and chronic viral hepatitis. Semin Liver Dis.
1991;11:73-83
1) USO PREVISTO
El kit de diagnostico HBsAg ELISA es un inmunoensayo enzimático in vitro para la detección cualitativa del antígeno
de superficie de la hepatitis B (HBsAg) en suero o plasma humano (heparina, citrato o EDTA).
El principio del ensayo descrito anteriormente se ilustra también enel siguiente diagrama.
A Muestra que contiene HBsAg:
1. Placa (Anti-HBs) + muestra (HBsAg) + Anti-HBs • peroxidasa
→ HBsAg •Anti-HBs• (Anti-HBs • peroxidasa) complejo de sándwich
2. Complejo de sándwich + solución de sustrato TMB → color de azul a celeste
3. Agregue ácido sulfúrico para detener el desarrollo del color → Lea la DO a 450nm/620-690 nm*8
B Muestra sin HBsAg:
1. Placa (Anti-HBs) + muestra (sin HBsAg) + Anti-HBs • peroxidasa → Anti-HBs (en los pocillos)
2. Anti-HBs (en los pocillos) + solución de sustrato TMB → Incoloro a azul claro
3. Agregue ácido sulfúrico para detener el desarrollo del color → Lea la DO a 450nm/620-690 nm*8
TMB Cromogénico
concentrado 0,6 mg/ml de 3,3',5,5'-tetrametilbenzidina (TMB) 1 vial,
(5) en una base orgánica con DMSO. 12 ml
CHROM TMB CONC
Tampón del
Sustrato 1 vial,
(6) Tampón de ácido cítrico con 0,03% H2O2.
SUB BUF 12 ml
Solución de Lavado
Conc. (20x)
Tampón concentrado de fosfato con Tween-20 1 vial,
(7) WASH SOLN CONC 58 ml
Solución de Parada
1 vial,
(8) 1N ácido sulfúrico
STOP SOLN 12 ml
ITEMS Componentes
(1) Micropipetas y puntas de 50µl y 100 µl
(2) Incubadora o baño maría con control de temperatura a +37 °C.
(3) Equipo para lavado de placas.
(4) Lector de microplacas ELISA:
Longitud de onda dual 450nm con longitud de onda de referencia de 620-690nm, ancho de
banda 10nm*1.
(5) Un analizador EIA de microplacas totalmente automático es opcional. El usuario debe
validar el equipo en conjunto con el kit.
5) INSTRUCCIONES DE USO
5.1) Advertencias:
5.1.1) Este kit de reactivos es sólo para uso profesional.
5.1.2) Este kit de reactivos es sólo para uso de diagnóstico in vitro.
5.1.3) Procure que todos los reactivos del kit y las muestras alcancen temperatura ambiente (+20 to +30 °C) y
mézclelos cuidadosamente antes de usar.
5.1.4) No use reactivos después de su fecha de caducidad.
5.1.5) No intercambie reactivos entre lotes diferentes.
5.1.6) No pipetee con la boca.
5.1.7) No fume o coma en zonas donde se manipulan muestras o reactivos.
5.1.8) Todos los componentes y muestras del kit deben considerarse potencialmente peligrosos para la salud Deben
usarse y eliminarse de acuerdo con los procedimientos de seguridad del laboratorio del usuario. Estos
procedimientos de seguridad probablemente incluirán el uso guantes de protección y evitar la generación de
aerosoles.
5.1.9) Las muestras potencialmente infecciosas y derrames o goteos que no contienen ácidos deben ser limpiados
concienzudamente con hipoclorito de sodio al 5% o tratado de acuerdo con las prácticas del laboratorio para el
control de un potencial riesgo biológico.
5.1.10) Antes de eliminar los desechos de las muestras usadas y reactivos del kit como desechos generales, estos deben
ser tratados de acuerdo con el procedimiento local para desecho con potencial riesgo biológico o tratado como
sigue:
Tanto los desechos sólidos como los líquidos deben ser autoclavados a +121 °C por lo menos por 30 minutos.
El desecho sólido también se puede incinerar.
El desecho líquido no ácido puede tratarse con hipoclorito de sodio diluido a una concentración final de 1%.
Los desechos ácidos deben ser neutralizados antes de tratarlos con hipoclorito de sodio como se mencionó
antes y debe mantenerse por 30 minutos para obtener una desinfección efectiva.
5.1.11) La solución de parada es irritante para los ojos, la piel, vías respiratorias y membranas mucosas. Evite el
contacto de la solución de parada con la piel y membranas mucosas. En caso de contacto, lave con copiosa
cantidad de agua inmediatamente. En caso de inhalación, proporcione aire fresco y busque ayuda médica en
caso de molestias.
5.1.12) El TMB cromogénico concentrado contiene un disolvente orgánico, que es inflamable: peligro de efectos
graves irreversibles por inhalación, contacto con la piel o ingestión. El TMB cromogénico concentrado contiene
dimetil sulfóxido, un irritante para la piel y membranas mucosas.
5.1.13) A pesar de que todo el material de origen humano ha resultado no reactivo para anti-VHC y Anti-VIH y ha sido
inactivado a +56 °C por una hora, el reactivo debe manipularse como material potencialmente infeccioso. *7
Catalogo Nr: KAPG4SGE3 PI Numero : 1701153 Revisión Nr : 110627/1
5.2) Toma de Muestra y Preparación para el Análisis
5.2.1) El paciente no requiere preparación especial antes de la toma de la muestra. La sangre debe ser tomada con
técnicas médicas aprobadas.
5.2.2) Con este kit se puede usar suero o plasma. La muestra de sangre total debe separarse lo antes posible para evitar
hemólisis. Cualquier partícula presente en la muestra (por ej. glóbulos rojos, coágulos de fibrina) deben
removerse antes de usar.
5.2.3) Las muestras deben almacenarse de +2 a +8 °C y evitar inactivación por calor para minimizar el deterioro. Para
almacenar por periodos largos, las muestras se deben congelar por debajo de -20 °C. No se recomienda
almacenarlas en congeladores que se auto descongelan.
5.2.4) Las muestras congeladas deben ser descongelados completamente y mezcladas en forma homogénea antes del
ensayo.
5.2.5) Evite congelar y descongelar en forma sucesiva
5.2.6) ADVERTENCIA
1. La muestra no debe contener compuestos de azida que puedan inhibir la actividad de la peroxidasa en el
conjugado.
2. Muestras de suero coaguladas y muestras con contaminación bacteriana no deben ser usadas
5.3) Almacenaje y estabilidad de los componentes
5.3.1) El kit debe almacenarse entre + 2 y +8 °C. No congelar.
5.3.2) Las tiras de las placas debes usarse dentro de 2 meses después de abrir la bolsa original de aluminio. Las tiras no
usadas deben permanecer en la bolsa de aluminio firmemente sellada.
5.3.3) Almacene los reactivos nuevamente entre +2 y +8 °C inmediatamente después de su uso.
5.3.4) El concentrado (x20) de la solución de lavado es almacenado y transportado entre +2 y +8 °C, lo que puede
causar cristalización. Si hay precipitado de cristales antes del uso, caliente la solución en un baño maría a +37 °C
hasta que los cristales se disuelvan.
NOTA:
a) Use una punta nueva de pipeta para cada muestra para evitar contaminación cruzada
b) Cada placa necesita sus propios controles negativos, positivos y pocillos blancos.
c) No use el valor de corte establecido para otras placas de HBsAg Elisa..
5.5.3) Agregue 50 μl de solución de anti-HBs· Peroxidasa a cada pocillo excepto a el blanco.
NOTA: El TMB cromógencio concentrado debe ser entre incoloro y celeste; de otro modo debe ser eliminado.
La mezcla de TMB cromogénico concentrado con tampón del sustrato debe usarse dentro de 30 minutos
a partir del momento de la mezcla. La mezcla debe protegerse de la luz intensa.
5.5.9) Cubra la placa con la cubierta negra e incube a temperatura ambiente por 30 minutos.
5.5.10) Detenga la reacción agregando 100 µl de solución de parada en cada pocillo incluyendo el blanco.
5.5.11) Determine la absorbancia de los controles y muestras dentro de 30 minutos con un fotómetro de precisión a
450 / 620-690 nm (Longitud de onda de 450 nm para la lectura con 620-690 nm de longitud de onda de
referencia)*2. Use el blanco para blanquear el fotómetro.
NOTA:
El color del blanco debe ser de incoloro a amarillento pálido; de otro modo el ensayo es inválido. En este caso el
ensayo debe ser repetido.
Blanco Sustrato: la absorbancia debe ser menor de 0,100.
2. La evaluación del Panel de Sensibilidad para HBsAg (BBI, USA) produjo los siguientes resultados:
Panel No. De miembros del Resultados
5.11.3) Mutantes
1. Panel de mutantes
Se examinó un panel mutante prenuclear de hepatitis B de Teragenix, EE.UU. HBsAg Elisa puede detectar los
resultados si la relación de la dilución del panel es 1:2 o 1:5.
Este panel consiste de 3 muestras de promotor del núcleo basal (A1762/G1764) tipo silvestre; Infección tipo
silvestre, mutante precursor nuclear codón 28, de 4 muestras de promotor del núcleo basal (A1762/G1764) tipo
silvestre; Mutante precursor nuclear codón 28, de 2 muestras de promotor del núcleo basal (T1762/A1754) infección
mixta tipo silvestre MUTANTE (A1762/G1754).
2. Muestras mutantes
La tasa de detección de HBsAg Elisa es de 17/21 que es mejor que la del kit de referencia (14/21).
Kit de Kit de
Mutación HBsAg Elisa Mutación HBsAg Elisa
referencia referencia
Int Int Int Int
113A114 positivo positivo C137W positivo negativo
P120L positivo positivo C139Y positivo positivo
R122I negativo negativo K141E positivo positivo
R122T positivo positivo P142L/G145R negativo negativo
E122I positivo positivo P142L positivo positivo
T123N positivo positivo D144G positivo negativo
122RA123 negativo negativo G145K positivo negativo
C124R negativo negativo C147T/C148T positivo positivo
T131I positivo positivo E164D positivo positivo
T131P positivo positivo S174N positivo positivo
Y134S positivo positivo
Tasa de Tasa de
8/11 8/11 9/10 6/11
detección detección
5.11.6) Precisión
1) Repetibilidad intra-ensayo
La precisión intra-ensayo se determinó usando una muestra de control positivo y dos sueros de pacientes con
distinta concentración de HBsAg (levemente por sobre el nivel de corte y a nivel medio) las que fueron analizadas
en 20 repeticiones en un ensayo único por 3 días.
Desviación Coeficiente de
Ensayo ID de muestra Promedio
Estándar Variación
1 PC 1 :32 0.7541 0.0857 11.36
1 PS 1 :32 2.2766 0.1571 6.90
1 PS 1 :64 1.3159 0.1168 8.88
2 PC 1 :32 0.9671 0.0358 3.70
2 PS 1 :32 2.7325 0.1025 3.75
2 PS 1 :64 1.5822 0.0888 5.61
3 PC 1 :32 0.9669 0.0909 9.40
3 PS 1 :32 2.6088 0.2367 9.07
3 PS 1 :64 1.6502 0.1499 9.08
El CV calculado varió entre 3,7% y 11,36 %. (= valor aceptable para un inmunoensayo que utiliza microplacas).
Los CV calculados varían entre 30,4% para una muestra HBsAg negativa y 23,1 % para una muestra HBsAg
positiva débil (= valores aceptables para la imprecisión entre ensayos de un inmunoensayo que utiliza microplacas).
5.11.8) Seguimiento
El Calibrador Maestro HBsAg de DIAsource ha sido calibrado con el British Working Standard para HBsAg
(Código NIBSC: 01/476-006) usando el ensayo HBsAg ELISA. La potencia relativa (razón) del British Working
Standard para HBsAg versus el Calibrador Maestro HBsAg de DIAsource es 4,081 (3,853-4,325 Intervalo de
Confianza 95%). La concentración del control positivo del ensayo HBsAg ELISA se ha determinado contra el
Calibrador Maestro HBsAg de DIAsource y fue establecida en 42 IU/ml ± 20%.
↘ ↙
Incube a temperatura ambiente por 30 minutos.
↓
Agregue 1N ácido sulfúrico a cada pocillo
↓
Determine la absorbancia usando 450 nm como
longitud de onda para la lectura con un
longitud de onda de referencia de 620-690nm*1
6) NOTAS
*1 La longitud de onda de referencia del espectrofotómetro puede ser de 620nm a 690nm. Sin embargo el usuario debe
validar el foto metro en conjunto con este kit antes de su uso.
*2 Inactivación incompleta del virus de la hepatitis B después de tratamiento con calor a +60°C por 10 horas, J. Infect. Dis.
138:242-244.
7) BIBLIOGRAFIA
1. Blumberg BS, Alter HJ, Visnich S. A „new“ antigen in leukemia sera. JAMA, 1965;191:101- 106.
2. Dane DS, Cameron CH, Briggs M. Virus-like particles in serum of patients with Australia-antigen-associated
hepatitis. Lancet 1970; 1: 695 - 698.
3. Aach RD, Grisham JW, Paker CW. Detection of Australia antigen by radioimmunoassay. Proc Natl Acad Sci. USA
1971;68:1056-1060.
4. Kim CY, Tikes JG. Purification and biophysical characterization of hepatitis antigen. J Clin Invest. 1973; 52:1176-
1186.
5. Wolters G. Kuijpers LP, Kacaki J, Schuurs AH, Enzyme linked immunosorbent assay for hepatitis B surface antigen.
J Infect. Dis 1977;136:Suppl 311-377.
6. Shih JW, Cote PJ Jr, Dapolito GM, Gerin JL. Production of monoclonal antibodies against hepatitis B surface antigen
(HBsAg)by somatic cell hybrids. J Virol Methods. 1980;1:257-273.
7. Hoofnagle JH, Di Bisceglie AM. Serologic diagnosis of acute and chronic viral hepatitis. Semin Liver Dis.
1991;11:73-83
Symboles utilisés
Consulter les instructions d’utilisation
Température de conservation
Utiliser jusque
Numéro de lot
Référence de catalogue
Contrôle
Dispositif médical de diagnostic in vitro
Fabricant
Contenu suffisant pour <n> tests
WASH SOLN CONC Solution de lavage concentrée
CAL 0 0 Calibrateur zéro
CAL N Calibrateur #
CONTROL N
Contrôle #
Ag 125I Traceur
Ab 125I Traceur
Ag 125I CONC Traceur concentré
Ab 125I CONC Traceur concentré
Tubes
INC BUF Tampon d'incubation
ACETONITRILE Acétonitrile
SERUM Sérum
DIL SPE Diluant du spécimen
DIL BUF Tampon de dilution
ANTISERUM Antisérum
IMMUNOADSORBENT Immunoadsorbant
DIL CAL Diluant de calibrateur
REC SOLN Solution de reconstitution
PEG Glycol Polyéthylène
EXTR SOLN Solution d’extraction
ELU SOLN Solution d’elution
GEL Cartouches Bond Elut Silica
PRE SOLN Solution de pré-traitement
NEUTR SOLN Solution de neutralisation
TRACEUR BUF Tampon traceur
Microplaque de titration
Ab HRP HRP Conjugué
Ag HRP HRP Conjugué
Ab HRP CONC HRP Conjugué concentré
Ag HRP CONC HRP Conjugué concentré
CONJ BUF Tampon conjugué
CHROM TMB CONC Chromogène TMB concentré
CHROM TMB Solution chromogène TMB
SUB BUF Tampon substrat
STOP SOLN Solution d’arrêt
INC SER Sérum d'incubation
BUF Tampon
Ab AP AP Conjugué
SUB PNPP Tampon PNPP
BIOT CONJ CONC Biotine conjugué concentré
PREC AGENT Agent de précipitation
AVID HRP CONC Avidine HRP concentré
ASS BUF Tampon de test
Ab BIOT Biotine conjugué
Ab Anticorps spécifique
SAV HRP CONC Concentré streptavidine HRP
NSB Liant non spécifique
2nd Ab Second anticorps
ACID BUF Tampon d'acidification
DIST Distributeur
TRAY Plaque d'incubation
PMSF Solution PMSF
Conserver à l'abri de la lumière
STRIP Bandelette de dots
SUB Substrat
EXTR SOLN CONC Tampon d'extraction concentré
CART Cartouche
SAV HRP Streptavidine-peroxydase de raifort
PIPETTE
PIPETTE Pipette
WASH SOLN Tampon de lavage
P.I. Number : 1701000 Revision nr 140723
Símbolos utilisados
Consultar las instrucciones de uso
Limitación de temperatura
Fecha de caducidad
Codigo de lote
Número de catálogo
Control
Producto sanitario para diagnóstico in vitro
Fabricante
Contenido suficiente para <n> ensayos
WASH SOLN CONC Solución de lavado concentrada
CAL 0 0 Calibrador cero
CAL N Calibrador #
CONTROL N
Control #
Ag 125I Trazador
Ab 125I Trazador
Ag 125I CONC Trazador concentrada
Ab 125I CONC Trazador concentrada
Tubos
INC BUF Tampón de incubación
ACETONITRILE Acetonitrilo
SERUM Suero
DIL SPE Diluyente de Muestra
DIL BUF Tampón de dilución
ANTISERUM Antisuero
IMMUNOADSORBENT Inmunoadsorbente
DIL CAL Diluyente de calibrador
REC SOLN Solución de Reconstitución
PEG Glicol Polietileno
EXTR SOLN Solución de extracción
ELU SOLN Solución de elución
GEL Cartuchos Bond Elut Silica
PRE SOLN Solución de Pre-tratamiento
NEUTR SOLN Solución de Neutralización
TRACEUR BUF Tampón de trazador
Placa de microvaloración
Ab HRP HRP Conjugado
Ag HRP HRP Conjugado
Ab HRP CONC HRP Conjugado concentrada
Ag HRP CONC HRP Conjugado concentrada
CONJ BUF Tampón de Conjugado
CHROM TMB CONC Cromógena TMB concentrada
CHROM TMB Solución Cromógena TMB
SUB BUF Tampón de sustrato
STOP SOLN Solución de Parada
INC SER Suero de Incubación
BUF Tampón
Ab AP AP Conjugado
SUB PNPP Sustrato PNPP
BIOT CONJ CONC Concentrado de conjugado de biotina
PREC AGENT Agente de precipitación
AVID HRP CONC Concentrado avidina-HRP
ASS BUF Tampón de ensayo
Ab BIOT Conjugado de biotina
Ab Anticuerpo específico
SAV HRP CONC Estreptavidina-HRP Concentrado
NSB Unión no específica
2nd Ab Segundo anticuerpo
ACID BUF Tampón de Acidificación
DIST Distribuidor
TRAY Bandejas de incubación
PMSF Solución de PMSF
Proteger de la luz
STRIP Tries Dot
SUB Sustrato
EXTR SOLN CONC Concentrado de tampón de extracción
CART Cartucho
SAV HRP Estreptavidina HRP
PIPETTE Pipeta
WASH SOLN Tampón de lavado