0% found this document useful (0 votes)
1K views540 pages

Proteus XR F Generator Service Manual SM 5764770-1EN 4

This document is the revision 4 manual for the GE Healthcare Proteus XR/f medical x-ray generator. It contains safety and operational instructions, as well as information on maintenance, repairs, and replacement parts. Revisions from previous versions include updated safety symbols, replacement of hardware components, and renewed parts numbers. The manual advises on proper handling and environmental disposal of the device.

Uploaded by

fortroni
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
1K views540 pages

Proteus XR F Generator Service Manual SM 5764770-1EN 4

This document is the revision 4 manual for the GE Healthcare Proteus XR/f medical x-ray generator. It contains safety and operational instructions, as well as information on maintenance, repairs, and replacement parts. Revisions from previous versions include updated safety symbols, replacement of hardware components, and renewed parts numbers. The manual advises on proper handling and environmental disposal of the device.

Uploaded by

fortroni
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Technical

Publications

5764770--1EN
Revision 4

Proteus XR/f

Generator Manual
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

2
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

IMPORTANT PRECAUTIONS
LANGUAGE

3
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

4
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

5
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

6
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

7
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation “damage
in shipment” written on all copies of the freight or express bill before delivery is accepted or “signed for” by a
General Electric representative or a hospital receiving agent. Whether noted or concealed, damage MUST
be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim for
damage if an inspection is not requested within this 14 day period.
To file a report:
G Call 1--800--548--3366 and use option 8.
G Fill out a report on http://us44hdd21/sctq/InstallFulfill/InstalFulfillment.htm
G Contact your local service coordinator for more information on this process.

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. In addition, electrical feeds into the Power
Distribution Unit shall be performed by licensed electrical contractors. Other connections between pieces of
electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel. The
products involved (and the accompanying electrical installations) are highly sophisticated, and special
engineering competence is required. In performing all electrical work on these products, GE will use its own
specially trained field engineers. All of GE‘s electrical work on these products will comply with the requirements
of the applicable electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE‘s field engineers, personnel of
third--party service companies with equivalent training, or licensed electricians) to perform electrical servicing
on the equipment.

IMPORTANT...X--RAY PROTECTION
X--ray equipment if not properly used may cause injury. Accordingly, the instructions herein contained should
be thoroughly read and understood by everyone who will use the equipment before you attempt to place this
equipment in operation. The General Electric Company, Healthcare Group, will be glad to assist and cooperate
in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x--radiation other than the useful
beam, no practical design of equipment can provide complete protection. Nor can any practical design compel
the operator to take adequate precautions to prevent the possibility of any persons carelessly exposing
themselves or others to radiation.
It is important that anyone having anything to do with x--radiation be properly trained and fully acquainted with
the recommendations of the National Council on Radiation Protection and Measurements as published in
NCRP Reports available from NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda,
Maryland 20814, and of the International Commission on Radiation Protection, and take adequate steps to
protect against injury.
The equipment is sold with the understanding that the General Electric Company, Healthcare Group, its
agents, and representatives have no responsibility for injury or damage which may result from improper use
of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.

8
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEHC iTrak (issue tracking) process to report all omissions, errors, and defects
in this publication.

ENVIRONMENTAL STATEMENT
ON THE LIFE CYCLE OF THE EQUIPMENT OR SYSTEM
This equipment or system contains environmentally dangerous components and materials (such as PCB‘s,
electronic components, used dielectric oil, lead, batteries etc.) which, once the life-cycle of the equipment or
system comes to an end, becomes dangerous and need to be considered as harmful waste according to the
international, domestic and local regulations.
The manufacturer recommends to contact an authorized representative of the manufacturer or an authorized
waste management company once the life-cycle of the equipment or system comes to an end to remove this
equipment or system.

9
GE Healthcare Proteus XR/f Generator
REV 4 5764770--1EN

10
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

REVISION HISTORY

REV DATE REASON FOR CHANGE

1 JAN 26, 2017 First edition.

2 JAN 04, 2018 Update of: Extra Fuse 50 Amp. for two--phase Generators,
Workstation configuration, AEC Calibration procedure, AEC
Checking in Maintenance, Aluminium Filter and tools, FRU part
numbers for LV-DRAC Software, FRU part numbers review for
Proteus XR/c and XR/f, Schematics

3 FEB 27, 2018 Tools and Test equipment for AEC Calibration, AEC Calibration
procedure update, Job Cards chapter.

4 APR 16, 2020 Safety Symbols, removed old information for XR/f AT System,
new X-ray Generator Control (PC Interface Box), Generator
Cover divided in two parts, replacement of HT Controller Board
A3000-33 by A3000--80/81, updated renewal parts and
schematics.

This Document is the English original version, edited and supplied by the manufacturer.

11
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ADVISORY SYMBOLS

The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.

DANGERS ADVISE OF CONDITIONS OR SITUATIONS THAT


IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS
PERSONAL INJURY OR DEATH.

ADVISE OF CONDITIONS OR SITUATIONS THAT IF NOT


HEEDED OR AVOIDED COULD CAUSE SERIOUS PERSONAL
INJURY, OR CATASTROPHIC DAMAGE OF EQUIPMENT OR
DATA.

Advise of conditions or situations that if not heeded or


avoided could cause personal injury or damage to equipment
or data.

Note . Alert readers to pertinent facts and conditions. Notes represent


information that is important to know but which do not necessarily
relate to possible injury or damage to equipment.

12
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SAFETY SYMBOLS
The following safety symbols may appear in the equipment.

Their meaning are described below.

Caution. Consult accompanying documents.

Safety Symbol. Follow instructions for use, especially those


instructions identified with Advisory Symbols to avoid any
risk for the Patient or Operator.
(Only applies to Standard IEC 60601--1:2005 and
IEC 60601-1:2005+AMD1:2012)

Manufacturer.

Date of Manufacture.

Medical Device.

Catalogue Number (Model reference).

Serial Number.

Model Configuration.

13
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

General Mandatory action.

Type B applied part.

IPX0 Protection against harmful ingress of water or particulate matter.


IP Classification: Ordinary.

Ionizing radiation.

Non-ionizing electromagnetic radiation.

Radiation of Laser apparatus.


Do not stare into beam.
(Only applicable to equipment with Laser Pointer)

Dangerous voltage.

General warning, caution, risk of danger.

Warning: Ionizing radiation.

14
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Warning: Non-ionizing radiation.

Warning: Laser beam.

Warning: Electricity.

Warning: Do not place fingers between mobile and fixed parts of


the equipment, it may cause serious injuries to patient or operator.
As well, make sure the patient extremities are correctly positioned
into limit areas during operation, movement of parts may cause
serious damages to patient.

Electrostatic sensitive devices.

No pushing.

No sitting.

No stepping on surface.

Do not handle.

15
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Emergency stop.

“Stand-by” power.
(Only applies to IEC 60601-1:2005 and IEC 60601-1:2005+AMD1:2012)

“ON” power.

“OFF” power.

“ON” / “OFF” (push-push).


Each position, “ON” or “OFF”, is a stable position.

Alternating current.

Three-phase alternating current.

Three-phase alternating current with neutral conductor.

N Connection point for the neutral conductor on Permanently


Installed equipment.

16
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Direct current.

Both direct and alternating current.

Protective Earth (Ground).

Earth (Ground).

This symbol according to the European Directive indicates that the


Waste of Electrical and Electronic Equipment (WEEE) must not be
disposed of as unsorted municipal waste and must be collected
separately. Please contact an authorized representative of the
manufacturer or an authorized waste management company for
information concerning the decommissioning of your equipment.

This separate collection symbol is affixed to a battery or its


packing, to advise that the battery must be recycled or disposed of
in accordance with local or country laws. The letters below the
symbol indicate whether certain elements (Li=Lithium, PB=Lead,
CD=Cadmium, Hg=Mercury) are contained in the battery. All
batteries removed from the equipment must be properly recycled
Li/Pb/Cd/Hg or disposed. Please contact an authorized representative of the
manufacturer or an authorized waste management company for
information concerning the decommissioning of your equipment.

Pollution Control. (Only applicable to People’s Republic of China (PRC)).


This symbol indicates the product contains hazardous materials in
excess of the limits established by the Chinese Standards. It must
not be disposed of as unsorted municipal waste and must be
collected separately. Please contact an authorized representative
of the manufacturer or an authorized waste management company
for information concerning the decommissioning of your
equipment.

17
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

18
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

TABLE OF CONTENTS
Section Page

CHAPTER 1 -- SAFETY AND SERVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

1 LOCK-OUT / TAG-OUT (LOTO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

1.1 System Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

1.2 LOTO Procedure for System Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

2 CONTROL SWITCHES AND BUTTONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7

2.1 X-Ray Generator Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7

2.2 Power ON / OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8

3 APPLYING SYSTEM POWER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9

4 EQUIPMENT SAFETY -- ELECTROSTATIC DISCHARGE (ESD) . . . . . . . . . . . . . . 1-11

4.1 Generating Static . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11

4.2 Personal Grounding Methods and Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-12

4.3 Grounding the Work Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13

4.4 Recommended Materials and Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-13

CHAPTER 2 - INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

1.1 Tools and Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

1.2 Pre-installation Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

1.3 General Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

2 UNPACKING, CABINET, PC INTERFACE BOX INSTALLATION,


AND POWER LINE CONNECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7

2.1 PC Interface Box Control Panel Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12

19
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Section Page

3 CABLE CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17

3.1 Cable Routing inside Generator Cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17


3.1.1 General Cable Routing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
3.2 High Voltage Cables Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20

3.3 X-ray Tube Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22


3.3.1 Stator Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
3.4 Interconnection Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
3.4.1 Virtual Console (PC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
3.4.2 Door Interlock Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
3.4.3 Warning Light Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
3.4.4 Collimator Lamp and System Locks . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
3.4.5 230 VAC Power Supply for Two External Devices (optional) . . . . . . . . 2-28
3.4.6 Ion Chambers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28

4 FINAL INSTALLATION AND CHECKS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31

4.1 Cable Fastening and Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-31

5 SYSTEM INTERCONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33

5.1 System Interconnection Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-33

5.2 System Interconnection Maps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-35

CHAPTER 3 -- CONFIGURATION AND CALIBRATION . . . . . . . . . . . . . . . . . . . 3-1

1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3

1.1 Tools and Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4

1.2 Generator Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4


1.2.1 Minimum Current Time Product (mAs) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
1.2.2 Exposure Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
1.2.3 HV Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
1.2.4 Duty Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5

20
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Section Page

2 CONFIGURATION AND TEST SWITCHES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7

2.1 3024SW1 - ATP Console CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8

2.2 3024SW2 - ATP Console CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8

2.3 3024SW3 - ATP Console CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9

2.4 3024SW4 - ATP Console CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9

2.5 3000SW2 - HT Controller Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9

2.6 Basic Configuration of Generator Boards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10

3 GENERATOR SERVICE MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11

3.1 Exposure Counters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19

3.2 Workstations Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20

4 EXTENDED MEMORY SETTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25

4.1 Extended Memory Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25

4.2 How to Enter and Store Data in the Extended Memory . . . . . . . . . . . . . . . . . . . 3-26

4.3 Limit of Maximum kW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28

5 X-RAY TUBE SELECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29

5.1 X-ray Tube Insert Protection Curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29

5.2 Generators with LF-RAC (Low Speed Starter) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30


5.2.1 Stator Voltage and Capacitor Selection . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
5.2.1.1 Configuration for One Tube with Standard Stator . . . . . . . . . . 3-30
5.2.1.2 Configuration for One Tube with the Starting Voltage
at 110 VAC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
5.2.2 Programming of Rotor Acceleration Time and Rad Filament
Setting Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
5.3 Generators with LV-DRAC (High Speed Starter) . . . . . . . . . . . . . . . . . . . . . . . . 3-33
5.3.1 Minimum Time for Ready (Delay) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
5.3.2 Tube Type Selection for Tube-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
5.3.3 Self-running Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
5.3.4 Protections -- Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
5.3.5 Anode Stator Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-37
5.3.6 Programming of Rad Filament Setting Time . . . . . . . . . . . . . . . . . . . . . 3-37

21
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Section Page
5.4 Anode Rotation Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-39

5.5 Focal Spots Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-40

6 CALIBRATION PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-43

6.1 Filament Stand-by Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-45

6.2 Exposure Time Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-45

6.3 kV Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-47

6.4 Digital mA Loop Closed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-52

6.5 Digital mA Loop Open (X-ray Tube Calibration) . . . . . . . . . . . . . . . . . . . . . . . . . 3-57


6.5.1 Auto-calibration of Digital mA Loop Open . . . . . . . . . . . . . . . . . . . . . . . . 3-57
6.5.2 Manual Calibration of Digital mA Loop Open . . . . . . . . . . . . . . . . . . . . . 3-62

6.6 AEC Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-67


6.6.1 Settings before the AEC Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-70
6.6.1.1 Preparation of the CS7 PC Workstation . . . . . . . . . . . . . . . . . 3-70
6.6.1.2 Preparation of the X-ray System . . . . . . . . . . . . . . . . . . . . . . . 3-77
6.6.1.3 Setup the X-ray System Test Conditions
(Table or Wall Stand) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-80
6.6.2 Procedure of AEC Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-80
6.6.2.1 Tube Warm-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-80
6.6.2.2 Calibrate the AEC Pre-Amp Master Gain Potentiometer . . . 3-81
6.6.2.3 Calibrate the kV Tracking Compensation . . . . . . . . . . . . . . . . 3-87
6.6.3 Verify AEC Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-90
6.6.3.1 Kv Tracking Compensation Test . . . . . . . . . . . . . . . . . . . . . . . . 3-90
6.6.3.2 Ion Chamber Cell Balance Test . . . . . . . . . . . . . . . . . . . . . . . . 3-94
6.6.4 AEC Density Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-98
6.7 Final Checks after Generator Calibration . . . . . . . . . . . . . . . . . . . . . . . . 3-99

APPENDIX A -- DATA BOOK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-101

CHAPTER 4 -- TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3

1.1 Tools and Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5

1.2 General Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6

22
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Section Page

2 GENERAL PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7

2.1 Low DC Voltage Power Supply Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7

2.2 Microprocessors and General Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9

2.3 Replacement of Memory in ATP Console CPU Board and/or


HT Controller Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9

3 ERROR CODES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11

CHAPTER 5 -- JOB CARDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1

1 JOB CARDS FOR ADJUSTMENT / REPLACEMENT PROCEDURES . . . . . . . . . . 5-3

1-001 ATP Console CPU Board (A3024-221) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7

1-002 EPROM-U24 on the ATP Console CPU Board (A3024-221) . . . . . . . . . . . . . . 5-10

1-003 AEC Control Board (A3012-25) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12

1-004 HT Controller Board (A3000-80/81) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14

1-005 Microcontroller+ROM U5 on the HT Controller Board (A3000-80/81) . . . . . . . 5-18

1-006 Filament Driver Board (A3004-10) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-20

1-007 Interface Control Board (A3009-11) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22

1-008 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24

1-009 Inverter Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26

1-010 Input Rectifier (A3255-03) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34

1-011 Rectifiers on the Inverter Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-37

1-012 Main Storage Capacitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-40

1-013 Charge/Discharge Monitor Board (A3212-01) . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-43

1-014 IPM Driver Board (A3063-06) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-46

1-015 IGBT’s in the Inverter Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-49

1-016 Low Speed -- LF--RAC Board (A3096-02) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-56

1-017 High Speed -- Interface DRAC--HF (A3240-05) . . . . . . . . . . . . . . . . . . . . . . . . . . 5-58

1-018 High Speed -- Interface DRAC--HF (A3240-05) . . . . . . . . . . . . . . . . . . . . . . . . . . 5-60

1-019 High Speed -- Clamping Board (A3109-01) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-62

23
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Section Page
1-020 High Voltage Tank -- One Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-64

1-021 AEC Adaptation Board (A3263-03) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-66

1-022 Locks Board (A3214-02) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-69

2-001 AEC Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-71

CHAPTER 6 -- PREVENTIVE MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1

1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3

2 PERIODIC MAINTENANCE PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5

2.1 Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5

2.2 General Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6

2.3 General Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7


2.3.1 External Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
2.3.2 Internal Cabinet Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7

2.4 Cable Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8


2.4.1 Ground Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
2.4.2 AC Power Supply in X-ray Room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8

2.5 Console Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9

2.6 HV Transformer Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10

2.7 X-Ray Tube Condition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10

2.8 Radiographic Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11


2.8.1 Test for kV Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
2.8.2 Test for Digital mA Loop Open . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
2.8.3 Test for Digital mA Loop Closed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13

2.9 AEC Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13

CHAPTER 7 -- RENEWAL PARTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

CHAPTER 8 - SCHEMATICS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1

24
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

CHAPTER 1
SAFETY AND SERVICE

1-1
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

1-2
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 1 LOCK-OUT / TAG-OUT (LOTO)

1.1 SYSTEM POWER

Name of Equipment: Proteus XR/f Systems (Workstation, X-Ray Generator


and Positioners).

Number of Locks: One per person working on the system.

Title(s) of Employees Authorized to Perform LOTO: Those trained in


Lock-out / Tag-out (LOTO).

Title(s) of Affected Employees and How to Notify: Hospital Personnel;


notified by verbal communication.

Energy Source Yes No Location of Energy Isolating Means Magnitude of Energy

Electrical X System’s Main “A1” Power Disconnect 208 -- 240 VAC, 1 Phase
380 -- 480 VAC, 3 Phase

Pneumatic X
Hydraulic X
Gas/Water/Steam X
Chemical X
Mechanical Motion X
Gravity X
Springs X
Thermal X
Stored Energy X
Air Under Pressure X
Oil Under Pressure X
Water Under Pressure X
Gas Under Pressure X
Steam X
Other X

Type(s) of Equipment and/or Method(s) Selected to Dissipate or Isolate


Stored Energy: Allow 3 minutes for stored energy to dissipate.

Type(s) of Equipment and/or Method(s) Used to Ensure Disconnection‘s:


One Lock & Tag for Main “A1” Disconnect.

1-3
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1.2 LOTO PROCEDURE FOR SYSTEM POWER

1. Prepare for shutdown of equipment by notifying affected personnel


working in the area that Lock-Out / Tag-Out is being performed.

2. Bring system software down.

3. Locate the Main Power Supply (Service) Disconnect (A1) that feeds the
Generator cabinet. Turn off the mains supply Circuit Breaker.

4. Lock Out / Tag Out the Circuit Breaker of the Main Power Supply
Disconnect (A1).

5. When servicing and/or maintenance procedures are related to the


Generator cabinet:

a. Remove the Generator cover and with an insulated voltmeter


measure the incoming power connection to the Generator at the
main line power fuses (F3 and F4 for single-phase; F3, F4 and F5
for three-phase) located at the right side of the Cabinet.

Input Transf. 6T2

GND Studs
L1 Input Line

L2 Input Line

L3 Input Line

Neutral Stud
(optional)

1-4
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

b. If no voltage is present at the main line power fuses of the


Generator, check that the Main Storage Capacitors of the High
Voltage Inverter have bled off their residual charge by observing
that the LEDs on the Charge/Discharge Monitor Boards are
extinguished. This takes approximately three (3) minutes after the
power is shut off.

LED DS1

CHARGE/DISCHARGE
MONITOR BOARDS

c. If the LEDs on the Charge-Discharge Monitor Boards are not lit,


proceed to service the equipment as required.

d. Verify LOTO by ensuring that the system can not switch on. Then
operate the switches on the equipment submodules to confirm
there is not connection. Return the switches/breakers to the OFF
position.

6. When servicing and/or maintenance procedures are related to the Table


of Proteus XR/f ET:

a. Turn off the Circuit Breaker SW1 of the Table located at the right
back side of the Table Base.

Note . If the Generator is on, there is a 220 volt AC applied to relay K2


(remote Start) through J1 Pins 4 and 5.

b. Remove the Table covers and with an insulated voltmeter


measure the incoming power connection to the Table at the
terminals 1 and 3 of the Circuit Breaker SW1.

Table Circuit Breaker SW1

1-5
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

c. If no voltage is present at Circuit Breaker SW1 of the Table,


proceed to service the equipment as required.

d. Verify LOTO by ensuring that the system can not switch on.

7. When servicing and/or maintenance procedures are related to the Floor


Mounted Tube Stand and/or Wall Stand of Proteus XR/f ET:

a. Turn off the Circuit Breaker SW1 of the Table located at the right
back side of the Table Base and the ON/OFF Emergency Switch
of the Table.

b. Proceed to service the equipment as required.

c. Verify LOTO by ensuring that the system can not switch on.

8. When servicing and/or maintenance procedures are related to the Floor


Mounted Tube Stand, Table or Wall Stand of Proteus XR/f ST:

a. In this case, the Floor Mounted Tube Stand, Table and Wall Stand
are powered through the Generator, then proceed as described in
step 5.

b. Proceed to service the equipment as required.

9. After service has been performed, notify the affected personnel that the
equipment will be re-energized and LOTO devices are being removed.

10. Assure the safety of the area and that all safety devices and guards have
been replaced.

11. Return system functions to a preset state.

12. Remove Lock-Out / Tag-Out equipment from the main disconnect.

13. Locate the Main Power Service Disconnect feeding the system cabinet
and re--apply the power.

14. Locate the Main Power Supply (Service) Disconnect (A1) that feeds the
Generator cabinet and reapply the power by turning on the mains supply
Circuit Breaker.

15. Perform system start-up.

1-6
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 2 CONTROL SWITCHES AND BUTTONS

2.1 X-RAY GENERATOR CONTROL

The Generator parameters are controlled through the Console of the Image
Acquisition Workstation. The Generator is equipped with a PC Interface Box
that comprises the following controls:

1. Generator Power ON.

2. Generator Power OFF.

3. “Generator Power ON” Indicator, it lights when the Generator is ON.

4. Handswitch, with the controls for radiographic “Preparation” and


“Exposure”.

4
1

1-7
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.2 POWER ON / OFF

ON

OFF

X-Ray
Generator
Control

X-Ray
Generator

ON: The Generator is turned ON by pressing this push-button (Pilot light is ON).

OFF: The Generator and System is turned OFF by switching off the user
interface application in the user Workstation. It turns off firstly the user
Workstation and then the Generator.

The Generator is also turned OFF by pressing the OFF push-button of the X-ray
Generator Control (Pilot light is OFF).

For Generators equipped with “High Speed Rotor Controller”, if the X-ray Tube
is rotating when pressing the “OFF” push-button, the Generator will stop the
anode immediately and then the unit will be turned off (approx. 3 seconds). The
equipment only turns off if “Preparation” is not activated.

IN THE EVENT OF AN EMERGENCY FORCIBLY DEPRESS


THE X-RAY ROOM “EMERGENCY OFF SWITCH” (USUALLY
A RED MUSHROOM-SHAPED SWITCH).

THIS SWITCH SHOULD BE LOCATED ON OR NEAR THE


X-RAY ROOM ELECTRICAL CABINET, USUALLY PLACED
NEAR THE GENERATOR CONTROL CONSOLE. MORE THAN
ONE OF THESE SWITCHES MAY BE PLACED AROUND THE
ROOM FOR GREATER ACCESSIBILITY.

TO ISOLATE THE EQUIPMENT FROM MAINS, TURN OFF THE


SWITCH LOCATED AT THE ROOM ELECTRICAL CABINET.

1-8
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 3 APPLYING SYSTEM POWER

POTENTIAL ELECTRICAL SHOCK. PERFORM LOTO


BEFORE PERFORMING SERVICE OF ANY KIND (REFER TO
SECTION 1, “LOCK-OUT / TAG-OUT (LOTO)”).

Disengage all facility (i. e., hospital) E-Stops to ensure that your system powers
up safely, properly and reliably.

The System should be powered by the same Main Power Supply (Service)
Disconnect (A1) (Room Electrical Cabinet) where the X-ray Generator is
connected.

To turn the System ON:

1. Turn ON Room Electrical Cabinet Switch. The Emergency OFF Switch


must not be activated.

IN THE EVENT OF AN EMERGENCY FORCIBLY DEPRESS


THE X-RAY ROOM “EMERGENCY OFF SWITCH” (USUALLY
A RED MUSHROOM-SHAPED SWITCH).

THIS SWITCH SHOULD BE LOCATED ON OR NEAR THE


X-RAY ROOM ELECTRICAL CABINET, USUALLY PLACED
NEAR THE GENERATOR CONTROL CONSOLE. MORE THAN
ONE OF THESE SWITCHES MAY BE PLACED AROUND THE
ROOM FOR GREATER ACCESSIBILITY.

TO ISOLATE THE EQUIPMENT FROM MAINS, TURN OFF THE


SWITCH LOCATED AT THE ROOM ELECTRICAL CABINET.

1-9
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2. Press the X-Ray Generator Control ON button, the Green Lamp turns
ON.

3. Check that the Table, Floor Mounted Tube Stand and Wall Stand is turned
ON once the Generator is ON. If required, switch the Table ON using it
owner Circuit Breaker or ON/OFF Emergency Switch.

Note . Also refer to the corresponding Operation or Service Manual of


each Proteus XR/f ET or ST system.

4. Turn ON the Workstation.

To turn the System OFF:

1. Turn OFF the Workstation.

2. Press the X-Ray Generator Control OFF button, the Green Lamp turns
OFF.

3. Turn OFF the Room Electrical Cabinet Switch (with the Emergency
Switch in OFF position).

Note . Also refer to the corresponding Operation or Service Manual of


each Proteus XR/f ET or ST system.

1-10
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 4 EQUIPMENT SAFETY -- ELECTROSTATIC


DISCHARGE (ESD)

A sudden discharge of static electricity from your finger or other conductor can
destroy static-sensitive devices or microcircuits. Often the spark is neither felt
nor heard, but damage occurs. An electronic device exposed to electrostatic
discharge (ESD) might not appear to be affected at all and can work perfectly
throughout a normal cycle. The device can function normally for a while, but it
has been degraded in the internal layers, reducing its life expectancy.

Networks built into many integrated circuits provide some protection, but in
many cases the discharge contains enough power to alter device parameters
or melt silicon junctions.

4.1 GENERATING STATIC

The following table shows that:

 Different activities generate different amounts of static electricity.

 Static electricity increases as humidity decreases.

Relative Humidity

Event 55% 40% 10%

Walking across carpet 7,500 V 15,000 V 35,000 V

Motions of bench worker 400 V 800 V 6,000 V

Removing bubble pack from PCB 7,000 V 20,000 V 26,500 V

Packing PCBs in foam--lined box 5,000 V 11,000 V 21,000 V

Many electronic components are sensitive to ESD. Circuitry design and


structure determine the degree of sensitivity. The following packaging and
grounding precautions are necessary to prevent damage to electric
components and accessories.

 Transport products in static-safe containers, such as tubes, bags, or


boxes to avoid hand contact.

 Protect all electrostatic parts and assemblies with conductive or


approved containers or packaging.

 Keep electrostatic sensitive parts in their containers until they arrive at


static-free stations.

1-11
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

 Place items on a grounded surface before removing them from their


container.

 When handling or touching a sensitive component or assembly, ground


yourself by touching the chassis.

 Avoid contact with pins, leads, or circuitry.

 Place reusable electrostatic-sensitive parts from assemblies in


protective packaging or conductive foam.

4.2 PERSONAL GROUNDING METHODS AND EQUIPMENT

Use the following equipment to prevent static electricity damage to equipment:

 Wrist straps are flexible straps with a maximum of one--megohm ±10%


resistance in the ground cords. To provide a proper ground, wear the
strap against bare skin. The ground cord must be connected and fit
snugly into the banana plug connector on the grounding mat or
workstation.

 Heel straps, toe straps, and boot straps can be used at standing
workstations and are compatible with most types of shoes or boots. On
conductive floors or dissipative floor mats, use them on both feet with a
maximum of one-megohm ±10% resistance between the operator and
ground.

Static Shielding Protection Levels

Method Voltage

Antistatic plastic 1,500

Carbon--loaded plastic 7,500

Metallized laminate 15,000

1-12
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4.3 GROUNDING THE WORK AREA

To prevent static damage of components and parts in the work area:

 Cover the work surface with approved static--dissipative material.


Provide a wrist strap connected to the work surface and properly
grounded tools and equipment.

 Use static-dissipative mats, foot straps, or air ionizers to give added


protection.

 Handle electrostatic sensitive components, parts, and assemblies by the


case or PCB laminate.

 Handle components only at static--free work areas.

 Turn off power and input signals before inserting and removing
connectors or test equipment.

 Use fixtures made of static-safe materials when fixtures must directly


contact dissipative surfaces.

 Keep work area free of non-conductive materials, such as ordinary


plastic assembly aids and Styrofoam.

4.4 RECOMMENDED MATERIALS AND EQUIPMENT

Materials and equipment that are recommended for use in preventing static
electricity include:

 Antistatic tape, smocks, aprons, or sleeve protectors.

 Conductive bins, foam, tabletop workstations with ground cord of


one--megohm (±10%) resistance, and other assembly or soldering aids.

 Static--dissipating table or floor mats with hard tie to ground.

 ESD Field service kits.

 Static awareness labels.

 Wrist straps and footwear straps providing one-megohm (±10%)


resistance.

 Material handling packages:

G Conductive plastic bags, plastic tubes, and tote boxes.

G Opaque shielding bags.

G Transparent metallized shielding bags and shielding tubes.

1-13
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

1-14
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

CHAPTER 2
INSTALLATION

2-1
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

2-2
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 1 INTRODUCTION

The Installation process depends on the Generator and System configuration.


Installation must be performed in the order indicated along this document.
Perform only the sections required to install this Generator.

1.1 TOOLS AND TEST EQUIPMENT

The following hand tools and products are required for the Installation:

 Standard service engineers tool kit including Allen and Torx key sets.

 Electric drill motor and assorted bits including ∅ 8 mm and ∅ 10 mm bits.

 Silicone Insulating Grease (proofing compound).

 Alcohol cleaning agent.

The following test equipment is required for Configuration and Calibration:

 Calculator.

 Digital Multimeter.

 Non-invasive kVp Meter.

 Digital mAs Meter.

 Only for AEC purposes:

G Homogeneous Phantom of Aluminium with a purity of not less that


99% and thickness of 21 mm for the Collimator Filter Holder
(Part no. 2281087-2).

G Homogeneous Phantoms of PMMA (Polymethylmethacrylate)


with a surface of 43 x 43 cm and different thickness (cm) to
achieve all the different thickness used in the AEC verification
tests: 8, 10, 15, 20, 25 and 28 cm.

G (Optional) Dosimeter, Unfors RaySafe Xi (or equivalent).

G A Trimmer for adjustment of the Ion Chamber Pre-Amplifier (AEC


tiny potentiometer) and Adhesive Tape.

2-3
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1.2 PRE-INSTALLATION CHECK

Prior to beginning installation, it is recommended to inspect the site and verify


that the X-ray room complies with Pre-installation requirements, such as:

 Incoming Line.

 Main Switch and Safety Devices.

 Conduits.

 Space Requirements.

(Refer to the “Pre-Installation” document.)

1.3 GENERAL CAUTIONS

OPERATOR AND SERVICE MANUALS SHOULD BE


CAREFULLY READ AND UNDERSTOOD BY SERVICE
PERSONNEL BEFORE USING AND SERVICING THE
EQUIPMENT, ESPECIALLY THE INSTRUCTIONS
CONCERNING SAFETY, REGULATORY, DOSAGE AND
RADIATION PROTECTION. KEEP THE MANUALS WITH THE
EQUIPMENT AT ALL TIMES AND PERIODICALLY REVIEW
THE OPERATING AND SAFETY INSTRUCTIONS.

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE-DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF.

ALWAYS HAVE THE “IPM DRIVER BOARD” CONNECTED IN


THE GENERATOR PREVIOUS TO MAINS POWER IS
ACTIVATED IN IT. IF THE “IPM DRIVER BOARD” IS NOT
CONNECTED, PERMANENT DAMAGE WILL OCCUR TO
IGBTS.

2-4
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

TO AVOID THE RISK OF ELECTRIC SHOCK, THIS


EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY
MAINS WITH PROTECTIVE EARTH. DO NOT TOUCH ANY
HEATSINK OF THE CIRCUIT BOARDS EVEN WITH THE
GENERATOR TURNED OFF. PREVIOUS TO DISASSEMBLE
ANY BOARD, REMOVE ALL CONNECTORS PLUGGED TO IT.

THIS GENERATOR IS PERMANENTLY CONNECTED TO THE


POWER LINE, AND POWERED ON UNLESS THE SAFETY
SWITCH INSTALLED IN THE ROOM ELECTRICAL CABINET
IS OFF. WHEN THE GENERATOR IS POWERED, THE NEON
LAMP (GREEN) LOCATED ON THE TRANSFORMER 6T2
(GENERATOR CABINET) IS ON.

INTERNAL PARTS OF THE GENERATOR (ALL FUSES, LINE


CONTACTOR (6K5), INPUT TRANSFORMER (6T2), ON/OFF
RELAY (3K3) AND LF-RAC MODULE) ARE PERMANENTLY
POWERED ON THROUGH POWER LINE ALTHOUGH THE
CONTROL CONSOLE IS OFF. BE SURE THAT THE SAFETY
SWITCH IS OFF BEFORE HANDLING ANY INTERNAL PART
OF THE EQUIPMENT.

2-5
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

2-6
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 2 UNPACKING, CABINET, PC INTERFACE BOX


INSTALLATION, AND POWER LINE
CONNECTION

The Generator is shipped in one box to facilitate transport and installation.

Upon receipt of the X-ray unit and associated equipment, inspect all shipping
containers for signs of damage. If damage is found, immediately notify the
carrier or their respective agent.

1. Open the shipping box. Take out the Control Console, Interconnection
Cables and other furnished parts. Do not discard any packing material
such as envelopes, boxes or bags until all parts are accounted for as
listed on the packing list.

2. Remove the packing material from the pallet.

3. Remove the Generator Covers and then remove the Generator Cabinet
from the shipping pallet, placing it near its chosen room position. This
operation requires at least two people.

4. When the equipment is unpacked, verify that all items on the customer
order are present, and the hardware and internal wiring is secure.

5. Check the part numbers / serial numbers of each component with its
identification labels, and inspect all pieces for visible damage. If any
damaged parts are found, repair or order replacements to prevent
unnecessary delay in installation.

Illustration 2-1
Compact Generator

2-7
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6. Usually, the Generator Cabinet is freestanding. Place the Cabinet near


its chosen place in the room. Level it using the Adjustable Leveling Legs
at the bottom of the Cabinet. Keep the Base at the maximum distance
from the floor.

Seismic areas and other conditions require the Generator to be secured


to the floor. There are mounting holes on the bottom of the Cabinet. Keep
the four Leveling Legs at the same height (refer to Illustration 2-2).

Illustration 2-2
Adjustable Leveling Legs

Adjustable Leveling Legs

7. Leave enough working area around the equipment to allow unhindered


access until its final assembly.

8. Two metallic plates, washers and nuts are supplied in a bag with the
Generator for mounting them at both sides of the Generator frame in
order to cover the shipping holes. Secure both plates using the bolts in
the Generator frame.

Illustration 2-3
Installation of Metallic Plates

Metallic Plates (2)

2-8
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

POWER LINE CONNECTION

KEEP IN MIND THE GENERAL CAUTIONS FOR LINE


POWERED GENERATORS INDICATED IN SECTION 1.3.

DO NOT POWER ON THE GENERATOR UNTIL SPECIFICALLY


INSTRUCTED IN THIS SERVICE MANUAL.

9. Verify that the power supply line is “OFF” in the Room Electrical Cabinet.
Verify that the power line to the Generator is cut when the Emergency
Switch(es) is(are) activated.

10. The power supply line should conform with the Generator model defined
in the “Pre-Installation” document. Wire sizes indicated in this document
are relative to the power supply line and wire length. Verify that the power
line voltage and phase of the Generator coincides with the one for Room
Electrical Cabinet.

11. Cut the cables to the appropriate length and remove insulation from both
ends of the power and ground wires. Connect them to the respective
terminals in the Room Electrical Cabinet.

12. Route the Power Line Cables to the Ground Terminal and Input Line
Fuses. These cables can be secured to the Fastening Bar of the Cabinet
and routed internally along the rear side of the Cabinet; or they can be
routed through the Round Cable Outlet on the Rear Cover of the Cabinet
(always apply Local Codes for cable routing). (Refer to Illustration 2-4.)

Illustration 2-4
Cable Routing in the Line Powered Generator

Cable Entrance

Fastening Bar

Power Line Cables

OR
Input Transformer 6T2

Ground Terminals
Round Cable Entrance
(at the Rear Cover)

Input Line Fuses Power Line Cables

2-9
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

13. For Single Phase Generators, connect the Power wires L1 and N (L2) to
the Fuse Holders of F3 and F4 (right side of the Cabinet), and the Ground
wire to the Ground stud in the Cabinet Frame (above these fuses or close
to the right side of the HV Transformer).

SINGLE PHASE GENERATORS ARE FACTORY DELIVERED


TO OPERATE ON PHASE AND NEUTRAL. IN CASE OF
CONNECTING THE EQUIPMENT TO A TWO-PHASE LINE,
REPLACE THE NEUTRAL CARTRIDGE WITH THE 50 A FUSE
SUPPLIED IN THE BAG CONTAINING THE GENERATOR
INTERCONECTION CABLES AND THE PC INTERFACE BOX
PROVIDED WITH THE GENERATOR.

14. For Three Phase Generators, connect the Power wires L1, L2 and L3 to
the Fuse Holders of F3, F4 and F5 (right side of the Cabinet), and the
Ground wire to the Ground Studs in the Cabinet Frame (located above
these fuses or close to the right side of the HV Transformer).

Three Phase Generators do not need Neutral (N) wire connection from
the Line. If the unit is provided with the optional Fuse Module, connect
the Neutral (N) wire from the Line to the Neutral Stud below F3 Fuse
Holder.

Illustration 2-5
Power Line connections

50A Extra Fuse for Two-Phase Line

Input Transf. 6T2

GND Studs
GND Studs
L1 Input Line

L2 Input Line

L3 Input Line

Neutral Stud
(optional)

2-10
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

15. The whole System (Tables, Wall Stand, etc.) can be switched ON/OFF
when the Generator is switched ON/OFF. For this, power the System
through the output terminals of the Line Contactor 6K5 (upper contactor
close to the Input Transformer). These terminals are located underneath
the Line Contactor 6K5.

Illustration 2-6
Line Contactor 6K5

Line Contactor 6K5

Input Transformer 6T2

Output Terminals

16. According to the nominal voltage of the line, verify or connect the
wire “:” to the indicated terminal (TB) of Transformer 6T2. This wire is
factory connected to 230 V~ (for 1-Phase), 400 V~ or 480 V~ (for
3-Phase). (Refer to Schematic 543020XX).

Note . For 220 V~ power line, connect the wire “:” to the 230 V~
terminals. For 380 V~ power line, connect the wire “:” to the 400
V~ terminals.

Illustration 2-7
Connections on Transformer 6T2

6T2 Terminal Strip

6T2

17. After connecting the Power Line Cables, secure them to the Fastening
Bar using cable ties if they are routed over the Fastening Bar, or using
a suitable clamp if they are routed through the Round Cable Outlet on the
Rear Cover of the Cabinet (always apply Local Codes).

2-11
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.1 PC INTERFACE BOX CONTROL PANEL INSTALLATION

The Handswitch support and the optional sloping support are shipped in a
separate box.

Note . There are two types of PC Interface Box, the pictures and process
below describes the new model. Both models are completely
compatible.

Illustration 2-8
PC Interface Box and Handswitch support

Sloping Support for


horizontal mounting

PC Interface Box Handswitch Support

The PC Interface Box can be either wall mounted or freestanding on a table. For
the wall mounting option the Handswitch support is mandatory, otherwise it is
optional, but recommended.

For wall mounting, it is not recommended to install the sloping support in order
to maintain the Handswitch in vertical position.

On the contrary, if the PC Interface Box is going to be used freestanding on a


table, it is recommended to install also the sloping support in order to place the
Handswitch with the necessary inclination to avoid it from slipping out of the
support.

2-12
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Handswitch support installation:

a. Unscrew both screws from the bottom cover and put it aside.

Illustration 2-9
Bottom cover

b. Drill two holes for the Handswitch support nuts and screws using
a 3.5 mm (9/64“) drill bit.

Illustration 2-10
Drilling points

Drill

2-13
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

G To install only the Handswitch support, place the two nuts in the
inner hexagonal spaces and screw the Handswitch support from
the outer part of the PC Interface Box.

Illustration 2-11
Handswitch support

G To install the Handswitch support with the sloping support, first


screw the sloping support from the inner part of the PC Interface
Box and, then, screw the Handswitch support to it.

Illustration 2-12
Handswitch support with sloping support

2. PC Interface Box installation.

G Freestanding: Reinstall the bottom cover using the previously


removed screws and place the PC Interface Box in its final
position.

Illustration 2-13
Freestanding installation

2-14
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

G Wall mounting:

a. For wall mounting, the bottom cover is discarded, so it is possible


to switch the cables position from the top cable outlet to the bottom
cable outlet. In that case, cut the tie-wrap that hold the cables,
route them through the opposite outlet and hold them together and
to the PC Interface Box with a new tie-wrap.

Illustration 2-14
Cables outlet

b. Fix the Wall support to the wall using 4 anchor screws (not
supplied).

c. Slide the PC Interface Box into the Wall support, taking care that
it fits in both tracks, and secure it to the support with a screw on
the top.

Illustration 2-15
Wall support

2-15
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Illustration 2-16
Wall mounting installation

3. Proceed with the cable connections (refer to Section 3.4.1 and


Section 5.2 -- Map 6070064 for cable connections).

2-16
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 3 CABLE CONNECTIONS

This section provides the information necessary to connect the Generator


Cables with the system and options.

Note . For more information about electrical requirements and cable


connections, refer to the “Pre-Installation” document and
Section 5 “System Interconnections” at the end of this document.

Note . Identification of some terminal connections (TB, TS), boards, etc...


along with this document (text and schematics) may have a prefix
number which indicates the module number in the equipment.
(a.e. TS2 as 10TS2 or 11TS2).

Some safety devices such as the Safety Switch / Emergency Switch, Warning
Light, and Door Interlock Switch are supplied and installed by the customer.
Verify that safety devices have been properly installed and routed during the
Pre-Installation procedure.

3.1 CABLE ROUTING INSIDE GENERATOR CABINET

3.1.1 GENERAL CABLE ROUTING

1. Before connecting the Interconnection cables within the Generator


Cabinet, cables must be first connected to each Device (Tables, Wall
Stands, etc.) and routed through the raceways. Remove the ferrite blocks
of the cables (factory clamped) when it is required to carry out a correct
routing, then re-install the ferrite blocks where they originally were
around cables.

2. Inside the Generator Cabinet, all Interconnection cables must be routed


over the Fastening Bar (upper rear bar) of the Cabinet Frame minding the
upper Cable Outlet at the rear side of the Cabinet Cover. (Refer to
Illustration 2-17).

2-17
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Illustration 2-17
Cable Routing in Compact Generator

Stator & Interconnections


Cables Outlet (rear side)

Cable Ties

Stator & Interconnections


Cables Entrance
Cable Ties

Fastening Bar

Input Transformer
Power Line Cables

Ground Terminals HV Cables

Input Line Fuses Stator & Interconnections Cables

Power Line
Cable Entrance

Input Transformer

Stator & Interconnections


Cable Entrance

Power Line Cable Entrance

Input Line Fuses

2-18
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. For Generators with the Low or High Speed Starter located on a shelf
(Module 10 or 11), Stator and Interconnections Cables have to be routed
internally through the Cabinet close to the HV Transformer.

In order to avoid signal interferences, it is strongly


recommended to fold and fasten close to the Generator
Cabinet the portion of cables not routed (see picture below).
Never wrap in circles.

YES NO

4. Connect all cables as indicated in Section 3 “Cable Connections”.

5. Secure all cables to the Fastening Bar using cable ties after all
cable / wire connections are complete.

2-19
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3.2 HIGH VOLTAGE CABLES CONNECTION

Connect the HV Cables in the HV Transformer (Generator).

These cables must enter into the Generator through the cable outlet on the
upper side of the Cabinet and then attached to the fastening bar (upper rear bar)
of the Cabinet frame minding the upper cable outlet at the rear side of the
Cabinet cover.

The Terminal Pins of the High Voltage cables are extremely


delicate and easily damaged. Therefore they must be handled
carefully. Make sure that they are straight and that the splits
in the pins are open (parallel to sides).

Anode and Cathode cables are furnished according to the room layout (length
of the cables).

1. The mounting accessories of each termination plug are factory


assembled. For extended information refer to the HV Cable
manufacturer’s instructions located inside the HV Cable package.

Do not install the Silicone washer supplied with the HV


Cables.

2. Unscrew the grub screw of the ring nut. (Refer to the illustration below.)

Contact Pins Nipple Ring Nut

Termination Plug Grub Screw

Contact

2-20
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. Prepare the High Voltage terminals that will be installed in the HV


Transformer.

G Put approximately 1 cm (0.5”) of HV Silicone Oil in the HV


Transformer receptacles (included in the HV Cables package).

1 cm (0.5”)
HV Cables Package HV Silicone Oil HV Transformer Receptacle

G If HV Silicone Oil is not available, fill the receptacles using silicone


paste provided with the X-ray Tube.

HV Cable Terminal with Silicone Paste

HV Silicone Paste

4. Carefully insert the Anode and Cathode termination plug into the
respective receptacle socket (watch the nipple on the plug to ensure
correct positioning of the contact pins).

5. Hand tighten the ring nut. It must be secure. Tighten the grub screw.

2-21
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6. Prepare the High Voltage terminals that will be installed in the X-ray Tube
receptacles as previously described. Apply Silicone Paste over the entire
surface of the Plug including the Pins.

7. Carefully connect the Anode and Cathode cables from the HV


Transformer to the X-ray Tube receptacles. Ensure that all connections
are made correctly, maintaining correct Anode and Cathode orientation.

8. Hand tighten the ring nut. It must be secure. Tighten the grub screw.

3.3 X-RAY TUBE CONNECTION

3.3.1 STATOR CABLE

Route the Stator cable together with the HV cables to the Generator Cabinet.
Connect the Stator cable terminals to the indicated Terminal Block TS2:

TERMINAL TS2
STATOR WIRES
TUBE-1

MAIN TS2-1

AUX (Shift) TS2-2

COMMON TS2-3

Note . Terminal Block TS2 may be marked as 10TS2 or 11TS2


depending on the Generator model.

MAKE SURE THAT STATOR WIRES ARE PROPERLY


CONNECTED. BEFORE MAKING ANY EXPOSURE, CHECK
THAT THE ANODE ROTATES CORRECTLY.

2-22
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

A LV-DRAC (Low Voltage - Digital Rotatory Anode Controller) is required for


High Speed X-ray Tubes.

IN GENERATORS EQUIPPED WITH LV-DRAC: THE LV-DRAC


OUTPUT CAN BE AS HIGH AS 1000 VRMS. FOR SAFETY
REASONS (TO AVOID ELECTRIC SHOCKS), THE STATOR
CABLE MUST BE SHIELDED AND BOTH ENDS OF THE
SHIELD MUST BE CONNECTED TO GROUND.

DUE TO ELECTROMAGNETIC INTERFERENCE (EMC)


PROBLEMS, THE IGBT’S HEATSINK IS NOT GROUNDED. IT
IS CONNECTED TO THE NEGATIVE TERMINAL OF THE
INPUT RECTIFIER. TO AVOID ELECTRIC SHOCK, BE SURE
THAT THE INPUT LINE IS DISCONNECTED AND THE
CAPACITOR BANK IS PROPERLY DISCHARGED BEFORE
MANIPULATING THE LV-DRAC.

FANS

Wires from fans should be routed with the Stator Cables, and connected to the
indicated terminal of the Generator Cabinet. Depending on the model of X-ray
Tube, the fans are powered at 115 V~ or 220 V~. Make the following
connections to select the fan voltage.

For Compact Generators with the Low Speed Starter LF-RAC located on a shelf
at the bottom of the Generator (module-10), connect wires from fans to:

GENERATOR WITH LOW SPEED STARTER


TUBE
CONNECTION WIRES FROM FANS 115 V~ 220 V~

10TS2-6 and 10TS2-7 TB4-T1 with TB1-22 or TB1-23 TB4-T1 with TB1-25 or TB1-26
AS TUBE-1
on the Generator Cabinet on the LF-RAC Board on the LF-RAC Board

For Compact Generators with the High Speed Starter LV-DRAC located on a
shelf at the bottom of the Generator (module-11), connect wires from fans to:

GENERATOR WITH HIGH SPEED STARTER


TUBE
CONNECTION WIRES TO FANS 115 V~ 220 V~

11TS2-6 and 11TS2-7 Jump 11TS2-6 with 11TS2-17 Jump 11TS2-6 with 11TS2-18
AS TUBE-1
on the LV-DRAC Module on the LV-DRAC Module on the LV-DRAC Module

2-23
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

THERMOSTAT OR PRESSURE SWITCH SIGNAL

If the X-ray Tube is provided with a Safety Thermostat (approx. 65oC) or


Pressure Switch (must be NC Contact), the two wires should be routed to the
Terminal Block TS2 in the Generator Cabinet and connected to the following
Terminals.

In case that the X-ray Tube is provided with a Safety Thermostat (approx. 65oC)
and a Pressure Switch (both must be NC Contacts), connect them in series
before routing, connecting both wire-ends to their respective Terminals in TS2.

THERMOSTAT WIRES TUBE-1

THERMOSTAT SIGNAL TS2-4

THERMOSTAT COMMON TS2-5

Note . Terminal Block TS2 may be marked as 10TS2 or 11TS2


depending on the Generator model.
If an X-ray Tube is not provided with Thermostat signal, jump both connections
in the Terminal Block TS2 (refer to above table).

GND AND/OR SHIELD

The connection of the GND and/or Shield wire of the Stator cables depend on
the Generator model.

GENERATOR MODEL
CONNECTION OF
Starter type and Location GND and/or SHIELD WIRE

LF-RAC (Low Speed)


10TS2--8 or 10TS2--16
Lower Cabinet Shelf (Module 10)

LV-DRAC (High Speed) 11TS2--8 (for Tube-1)


Lower Cabinet Shelf (Module 11) 11TS2--16 (for Tube-2)

2-24
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3.4 INTERCONNECTION CABLES

This section identifies the cables and runs needed for Generator and System
Interconnection. Route and connect the interconnection cables from each
component installed in the system to the Generator Cabinet as indicated in
Illustration 2-18 and Section 5.2 -- Interconnection Maps.

Interconnection cables should not be routed into the same


conduit or cable raceway as the Power or High Voltage
cables.

Illustration 2-18
Interconnection Cables

GENERATOR CABINET

J5 SERIAL COMMUNICATION
(factory supplied)

ROOM DOOR CABLE


(customer supplied)
TS1
ROOM LIGHT CABLE
(customer supplied)

ATP CONSOLE CPU

J15 HANDSWITCH (optional)


(factory supplied)

CABLES FROM BUCKY 1 & 2


TS1 (ONLY FOR SYSTEMS WITH 1 OR 2 BUCKY)
(field supplied)

TS2 STATOR, FAN & THERMOSTAT FROM TUBES


(field supplied)

TS3

COMPATIBILITIES ION CHAMBERS


ADAPTATIONS (customer supplied)

POWER LINE & GND FOR


L1 / L2 / L3 / N / GND LINE POWERED GENERATORS
(customer supplied)

2-25
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3.4.1 VIRTUAL CONSOLE (PC)

Systems using a Virtual Console running on a PC have a PC Interface Box


installed between the PC and Generator. (Refer to Section 5.2 --
Map 6070064).

Perform the following connections:

1. Remove the Back Cover of the PC Interface Box.

2. Connect the Serial Interconnection Cable (A3352--xx) to J5 of the


J5 Generator Cabinet from J1 of the PC Interface Box.

3. Connect the Handswitch Cable to J2 of the PC Interface Box.

4. Connect the Interface Box--Computer Cable (A21089-XX) to J3 and J4


of the PC Interface Box and the other end (2 connectors) to COM1
(Communication) and J1 (Auto ON/OFF) of the Computer connectors.

Note . J1 connector is only available in Computers provided with Auto


ON/OFF Board inside (factory installed).

5. Re-install the Back Cover of the PC Interface Box.

6. Check to set proper Line Voltage on PC. Plug the Power Line cable for
the Computer to a110 V~ or 230 V~ socket.

GENERATOR PC INTERFACE PC
CABINET BOX (COMPUTER)

SERIAL CABLE (A3352--xx) PC BOX / PC CABLE


J5 J1 J1
(A21089-xx)

J2 COM1
HANDSWITCH
J3

J4 POWER LINE CABLE


TO 110 / 230 VAC SOCKET

3.4.2 DOOR INTERLOCK SIGNAL

Connect two wires from the Room Door Interlock Switch(es) to Terminal Block
3TS1-22 (Door signal) and 3TS1-23 (Door Rtn - GND). If the X-ray Room is not
provided with a Door signal, place a jumper between both connections in
Terminal Block 3TS1.

2-26
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3.4.3 WARNING LIGHT SIGNAL

Room Warning Lamp(s) for “X-ray ON” indication can be externally powered,
or internally by the Terminal Block 3TS1. Room Lamp(s) must be connected
through the Terminal Block 3TS1-47 and 3TS1-48 (internal relay on Interface
Control Board) to enable the Generator switches On/Off the Room Warning
Lamps. (Refer to Section 5.2 -- I/F-008).

3.4.4 COLLIMATOR LAMP AND SYSTEM LOCKS

The Generator can supply power to the Collimator Lamp and System Locks
(Table, Wall Stand, etc.)

Connect wires from the Collimator Lamp to Terminal Block TB7-3 (24 V~) and
TB7-4 (0 V~) of the Lock Board.

Connect wires from the Locks to Terminal Block TB7-5 (+24 V=) and TB7-6 (0
V=) of the Lock Board.

Lock Board

Note . When required, voltages (V~ and V =) on TB7 can be changed by


connecting their respective wires to the other available terminals
on the Input Transformer 6T2. (Refer to Schematics 543020XX).

At this point, proceed to perform the complete Configuration


and the Calibration procedures except for AEC procedures.
Once Configuration and Calibration tasks have been
performed, proceed with the rest of the Installation and
Calibration tasks.

2-27
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3.4.5 230 V~ POWER SUPPLY FOR TWO EXTERNAL DEVICES (OPTIONAL)

Optionally the Generator can be provided with a Terminal Strip 5TS1 at the top
of the cabinet in order to make easy the connection to supply 230 V~ for two
external devices such as two Detector Power Supplies.

5TS1

3.4.6 ION CHAMBERS

Note . The Generator is only compatible with Ion Chambers that output
a positive ramp.

1. Connect each Ion Chamber cable to J1 (IC1) and J2 (IC2) of the “AEC
Adaptation Board” (A3263--03). The reference of the Ion Chamber cable
supplied by the Generator manufacturer is A3253--02.

Illustration 2-19
Four Ion Chamber Connection

* Connections free of voltage IC Adaptation


GENERATOR Cables
AEC ADAPTATION BOARD
(A3263--03)
ION CHAMBER Cable
J1 IC 1 PREAMPLIFIER
POWER INTERFACE Cable
TS1 TB1
ION CHAMBER Cable
J2 IC 2 PREAMPLIFIER

ATP CONSOLE BOARD


J5
J3

AEC INTERFACE Cable


J4 J5

2-28
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2. Configure Jumpers from JP1 to JP8 and from JP13 to JP16 in “B”
position.

3. Claymount Ion Chambers requires a GND (TB1-10) reference voltage for


“Area Selections” and “AEC Reset” signals. The reference voltage is
provided from: TB1-7 “Relay 1”.

Connect a wire-jumper between terminal TB1-10 (GND) and TB1--7


(Relay 1) of the “AEC Adaptation Board” (A3263--03).

DO NOT TURN ON THE GENERATOR UNTIL THE INPUT


SIGNALS OF ALL THE ION CHAMBERS ARE CONNECTED TO
THE REQUIRED VOLTAGE. OTHER VOLTAGE MIGHT
DAMAGE THE ION CHAMBERS.

4. Ion Chambers output must be 0 V= when there is no-radiation (No-Offset


adjustment). Configure Jumpers from JP9 to JP12 in “A” position.

2-29
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

2-30
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 4 FINAL INSTALLATION AND CHECKS

4.1 CABLE FASTENING AND COVERS

Note . Before re-installing cabinet covers, perform the rest of the required
Calibration procedures (e.g. AEC).

Check that all electrical connections are firm and secure. Cables should be
correctly routed. (Refer to Section 3.1)

In order to avoid signal interferences, it is strongly


recommended to fold and fasten close to the Generator
Cabinet the portion of cables not routed (see picture below).
Never wrap in circles.

YES NO

2-31
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Assemble both parts of the Cabinet Cover using the screws. Position the Cover
over the Generator Cabinet connecting the internal ground wires to the Cover.
Finally secure the Cover to the Generator Cabinet with the screws on both
sides.

The Power Line, High Voltage and Interconnections cables must pass through
the corresponding cable outlets and be fixed with their fastening plates.

Cover Assembly

Cable Outlets and Fastening Plates

2-32
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 5 SYSTEM INTERCONNECTIONS

5.1 SYSTEM INTERCONNECTION SIGNALS

All input signals are active low. This means the inputs must be pulled to ground
(chassis ground of the Generator) thru relay contacts, by a transistor or other
switching device. The current requirement of the switch is less than 10 mA.

Do not apply 115 / 220 V~ logic signals to any of the logic


inputs. If 115 / 220 V~ logic signals are used in the X-ray table
(i.e. fluoro command), these signals must be converted to a
contact closure by a relay.

The outputs signals from the Generator to the subsystem devices are usually
active low (switched to chassis ground of the Generator). The outputs are open
collector transistor drivers with a maximum current of 0.5 Amperes.

Table 2-1
System Interconnection Signals

SIGNAL NAME SIGNAL DESCRIPTION

AUX BUCKY SPLY External voltage supply required for the Bucky motion, when this voltage is not +24 V=.

--BUCKY 1 DR CMD A low signal to the Interface Control Board as a command to output a Bucky-1 (normally the Table Bucky) drive signal.

--BUCKY 1 MOTION This low going signal from Bucky-1 indicates Bucky-1 motion, and therefore the exposure is enabled.

BUCKY 1 DR This signal is originated from the Bucky supply of the Power Module when an exposure order. It starts the Bucky.

--BUCKY 2 DR CMD A low signal to the Interface Control Board as a command to output a Bucky-2 (normally the Vertical Bucky Stand) drive signal.

--BUCKY 2 MOTION This low going signal from Bucky-2 indicates Bucky-2 in motion, and therefore the exposure is enabled.

BUCKY 2 DR This signal is originated from the Bucky supply of the Power Module when an exposure order. It starts the Bucky.

--BUCKY EXP This low going (0 volts) signal starts the Bucky exposure. The signal originates on the Interface Board

BUCKY SPLY Voltage supply required for the Bucky drive command.

C--HT CLK Serial data clock to the HT Control Board. This clock synchronizes the C--HT DATA signal.

C--HT DATA Serial data to the HT Control Board. This data is synchronous with the C--HT CLK signal.

This active low signal indicates that NO EXPOSURE HOLD condition exists at the Collimator. This input is read only when the
--COLLIMATOR
Radiographic Tube is selected.

--DOOR This low signal is the interlock for the Door of the X-ray room.

Low going Expose signal to the HT Control Board. If --PREP is low then a Spot Film or RAD exposure is made, else a Fluoro
--EXP
exposure is made.

--FLD1 DR A low signal to select the right field in the Ion Chamber.

--FLD2 DR A low signal to select the left field in the Ion Chamber.

--FLD3 DR A low signal to select the center field in the Ion Chamber.

2-33
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Table 2-1 (cont.)


System Interconnection Signals

SIGNAL NAME SIGNAL DESCRIPTION

HT--C CLK Serial data clock from the HT Control Board. This clock synchronizes the HT--C DATA signal.

HT--C DAT Serial data from the HT Control Board. This data is synchronous with the HT--C CLK signal.

This signal is low when the switch in the high voltage transformer is in the RAD position. This is a safety interlock which
--HT INL
prevents an exposure if the high voltage switch (in the HV Transformer) is in the wrong position.

This analogic signal (originates in the optional AEC Control Board) controls the output of the HV Power Supply on the Interface
HV PT CRL
Control Board. +5 volts programs the output to be 0 volts, and 0 volts programs the output to approximately --1200 volts.

IC GND GND for the IC SPLY.

IC1 INPUT This input is the output of the Bucky 1 Ion Chamber (normally the Table Ion Chamber).

IC2 INPUT This input is the output of the Bucky 2 Ion Chamber (normally the Vertical Bucky Stand Ion Chamber).

IC3 INPUT This input is the output of the Spot Film Ion Chamber.

IC SPLY Power supply for the Ion Chamber. This output should be within the range of 500 to 800 volts.

--kV DWN A low signal is a command for the HT Control Board to drive the Fluoro kVp DOWN during a Fluoro exposure in ABC mode.

--kV UP A low signal is a command for the HT Control Board to drive the Fluoro kVp UP during a Fluoro exposure in ABC mode.

--LINE CONT A low signal energizes the main line contactor K5 in the Power Module.

Signal synchronous with the AC line. This signal originates in the Interface Board and is used to synchronize Fluoro exposures
LINE SYNC
with the AC line.

Commands to the HT Control Board to boost X-ray Tube Filament to the value of mA selected and to start the X-ray Tube Rotor
--PREP
is RAD Tube is selected.

This low going signal indicates the system is ready to make an exposure (Prep cycle complete). This signal is used to interface
--READY
to certain peripheral devices such as Film Changers, etc.

This low going signal indicates the X-ray preparation or exposure. This signal is used to interface to the Room X-ray warning
--ROOM LIGHT
light.

--STRT DR A low signal to indicate the start of an exposure to the Ion Chamber.

--THERMOSTAT-1 This signal from X-ray Tube indicates the overheat of the Tube-1.

2-34
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.2 SYSTEM INTERCONNECTION MAPS

Refer to the following maps for details of the wire connections.

Note . Please, keep in mind the following clarifying notes about the
information in the following maps:

*1) XIF and GIPS Interfaces are not applicable to the Proteus XR/f
systems.

*2) Millennium name corresponds to Proteus XR/f ET system.


Multirad name corresponds to Proteus XR/f ST system.

SYSTEM INTERCONNECTION

 RS-232/422/485 Serial Communication . . . . . . . . . . . . . . . A6188--03

 PC--CTSC--ON/OFF BOX--SHF . . . . . . . . . . . . . . . . . . . . . . 6070064

 Earthing Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I/F-103

 ISID -- Image System Interconnection Diagram


Table-WallStand Proteus XR/f ET or ST --
Konica-Minolta AeroDR -- Generator . . . . . . . . . . . . . . . . . . 54303172 (*1)

ROOM LAMPS

 Room Warning Light Interface . . . . . . . . . . . . . . . . . . . . . . . I/F-008

AEC -- ION CHAMBERS

 AEC Compatibility with AID / VACUTEC / COMET /


CLAYMOUNT for more than one Ion Chamber . . . . . . . . . I/M-015

2-35
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

2-36
ATP CONSOLE BD. ( Multilayer )
J5 HARNESS
J14 J5

J2 J1 1 14 PREP/EXP COMM
POWER CABLE PREP/EXP COMM
2 16
1 1 PREP ORDER PREP ORDER
GND GND 3 15
2 2 EXP ORDER EXP ORDER
+12V UNR +12V UNR 4 18
3 3 POWER OFF POWER OFF
+12V +12V 5 19 POWER ON
4 4 POWER ON
OFF OFF 6 17
5 5 POWER COMM POWER COMM
ON ON
6 6
GND GND
7 7 TS1
GND GND
8 8
+12V UNR +12V UNR 15 10
+12 VDC
9 9
--12V --12V 52 11 -- ALOE
12 PWR GND
COMMUNICATION 18 13
J3 CABLE J3 GND PWR GND
17 24 +24 VDC UNR
+24V UNR
--EXP 1 1 --EXP 25 +24 VDC UNR
C--HT DAT 2 2 C--HT DAT 22 22
DOOR
HT--C DAT 3 3 HT--C DAT RS--xxx 23 23 DOOR RTN
--PREP 4 4 --PREP J8 51 21
5 5 PT INPUT
C--HT CLK C--HT CLK
1 6
6 6 ACT EXP ACT EXP
HT--C CLK HT--C CLK 2
RXD+ 2 RXD+ ( RXD )
HV PT CRL 7 7 HV PT CRL 3 3 See Note
8 8 TXD+ TXD+ ( TXD )
--KV DWN --KV DWN 4 20
9 9 +5V
LINE SYNC LINE SYNC 5 7 LOGIC GND
10 10 GND
--LINE CONT --LINE CONT
6 5
11 11 --BUCKY 2 DR CMD
--BUCKY 2 DR CMD 7 4
12 12 TXD-- TXD--
--BUCKY EXP --BUCKY EXP 8 1
13 13 RXD-- RXD--
--HT INL --HT INL 9 8
14 14 CHASSIS GND
--BUCKY 1 DR CMD --BUCKY 1 DR CMD
9
15 15 CHASSIS GND
--KV UP --KV UP
/////
RS--232
J7
Note.-- Signals between ( ) when RS--232 Serial Communication
1
ACT EXP
2
RXD
3
TXD
4
DTR
5
GND
6
7
8
9

NAME DATE SHEET / OF


DRAWING F. GARCIA 04/04/04
DWG:
A6188--03
1/1
REVISED A. DIAZ 04/04/04 A REV

A CN 04/148 F. GARCIA 09/09/04 SEDECAL RS--232/422/485 SERIAL COMMUNICATION


REV DESCRIPTION ISSUED BY DATE
8 7 6 5 4 3 2 1

D D

X-Ray Generator
Cabinet (SHF)

Hand
Switch

SHF INTERFACE
PANEL
PC INTERFACE
BOX
C (A16296-01) PC or CTSC C

Interface Box Cable AUTO ON/OFF


(A3352-01) Handswitch (A6805-16) (A3179-XX)
J5 J2
J3
J1/J2

Interface Box-Computer
Cable (A21089-XX)
J4 COM1

J1

B B

A
SYSTEM INTERCONNECTION A

NAME DATE SHEET / OF


PBA: PCB:
DRAWING M.González 2019-07-08
1/3
REVISED Sonia Perez 2019-07-08 A REV

6070064 DI
Confidentiality
n and privacy notice:All rights reserved. Unless otherwise specified and without written permission from SEDECAL, no part of this documentation may be reproduced or utilized in any form or by any means, electronic
or mechanical, including photocopying and microfilm, and may not be distributed or transmitted to third parties. A Frist release M.González 2019-07-08
SEDECAL CONFIDENTIAL
PC-CTSC - ON-OFF BOX - SHF
REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1

D D

GENERATOR CABINET

PC INTERFACE BOX
(A16296-01)

PBA INTERFACE BOX


(A40042-XX)

R1 DS1
+24V UNR

Interface Box
PWR GND
Cable
24V_UNR (A3352-01)
J5 J3 Handswitch (A6805-16)

25 25
Handswitch
C C
13 13
TS3 D4 COLL LIGHT SWITCH
+12V_ISO V_ON SW_Light
24 20 20
COLL LIGHT SW J2
22
TxD J14
1
GND PREP/EXP COMM 8
GND 15 V ISO
1 14 14 5
PREP
3 EXP ORDER 3 EXP ORDER
EXP EXP EXT2
2 15 15 6
K2A 5 PREP ORDER
PREP ORDER
power ON PREP EXT2
K1A 4 16 16
power OFF 6 PC

GND OFF
GND AUTO
ON/OFF
Board
ON (A3179-XX)
Interface
U9 Box-Computer Cable
1 6 (A21089-XX)
C1+ V- J1/J2
2 J1 J2
V+
3
C1- PWR ON
4 16 ON BUTTON
C2+ VCC 19 19 1 4 +12V
PWR OFF 1
OFF BUTTON
18 18 3 1
5 15
C2- GND PWR COM GND
GND 15 V ISO
17 17 2 6 2
11 14

12 13 COM1
J4
10 7
J8 TxD
2 2 1 3
9 8 RxD
2 RxD
B TxD
3 3 3 4 2 B
5 GND 15 V ISO
MAX3232EID
7 7 6 5

GND SW_Light_RTN

5 5
10 10
11 11
21 21
22 22
8 8

9 9

GND_CHASSIS GND

A A

PC WIRING INTERCONNECTION

NAME DATE SHEET / OF


PBA: PCB:
DRAWING M.González2019-07-08
2/3
REVISED Sonia Perez2019-07-08 A REV

6070064 DI
Confidentiality
n and privacy notice: All rights reserved. Unless otherwise specified and without written permission from SEDECAL, no part of this documentation may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, and may not be distributed or transmitted to third parties. A Frist release M.González2019-07-08 SEDECAL CONFIDENTIAL
PC-CTSC - ON-OFF BOX - SHF
REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1

D D

GENERATOR CABINET

PC INTERFACE BOX
(A16296-01)

PBA INTERFACE BOX


(A40042-XX)

R1 DS1
+24V UNR

Interface Box
PWR GND
Cable
24V_UNR (A3352-01)
J5 J3 Handswitch (A6805-16)

25 25
Handswitch
C C
13 13
TS3 D4 COLL LIGHT SWITCH
+12V_ISO V_ON SW_Light
24 20 20
COLL LIGHT SW J2
22
TxD J14
1
GND PREP/EXP COMM 8
GND 15 V ISO
1 14 14 5
PREP
3 EXP ORDER 3 EXP ORDER
EXP EXP EXT2
2 15 15 6
K2A 5 PREP ORDER
PREP ORDER
power ON PREP EXT2
K1A 4 16 16
power OFF 6 CTSC

GND OFF
GND AUTO
ON/OFF
Board
ON (A3179-XX)
Interface
U9 Box-Computer Cable
1 6 (A21089-XX)
C1+ V- J1/J2
2 J1 J2
V+
3
C1- PWR ON
4 16 ON BUTTON
C2+ VCC 19 19 1 4 +12V
PWR OFF 1
OFF BUTTON
18 18 3 1
5 15
C2- GND PWR COM GND
GND 15 V ISO
17 17 2 6 2
11 14

12 13 COM1
J4
10 7
J8 TxD
2 2 1 3
9 8 RxD
2 RxD
B TxD
3 3 3 4 2 B
5 GND 15 V ISO
MAX3232EID
7 7 6 5

GND SW_Light_RTN

5 5
10 10
11 11
21 21
22 22
8 8
J?

Mains MEDICAL GRADE


2
PS, CLASS II 3
1
9 9
DC Plug

GND

GND_CHASSIS

A A

CTSC WIRING INTERCONNECTION

NAME DATE SHEET / OF


PBA: PCB:
DRAWING M.González2019-07-08
3/3
REVISED Sonia Perez2019-07-08 A REV

6070064 DI
Confidentiality
n and privacy notice: All rights reserved. Unless otherwise specified and without written permission from SEDECAL, no part of this documentation may be reproduced or utilized in any form or by any means, electronic or mechanical, including photocopying and microfilm, and may not be distributed or transmitted to third parties. A Frist release M.González2019-07-08 SEDECAL CONFIDENTIAL
PC-CTSC - ON-OFF BOX - SHF
REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
GENERATOR CABINET

(1) Central Ground


(5) (2) GND Stud
(6) (4)
(3) Cabinet Cover GND

ROOM (4) Back Panel GND


ELECTRICAL (7) (3)
CABINET (5) Front Panel GND

(6) Filter LF1 Cover GND

(7) Adaptations Panel GND

(8) Console GND (Bottom Panel)

(9) Console Support GND

(1) (10) Pedestal Tube GND


GND cable
(2)
( yellow/green, AWG #10 ) (11) Pedestal Cover GND

(12) GND STUD

(12) Note.-- (9) applicable only


GND for metallical Box Console
OPERATOR CONSOLE

HV TANK
GND cable
( yellow/green, AWG #10 )
See Note (8)
(9)

PEDESTAL
(Option)

(10)
(11)

NAME DATE SHEET / OF


DRAWING F. GARCIA 30/06/04
I/F--103
1/1
REVISED A. DIAZ 30/06/04 REV

SEDECAL EARTHING DIAGRAM


REV DESCRIPTION ISSUED BY DATE
8 7 6 5 4 3 2 1

WS

serial cable (**) DETECTOR ASSEMBLY


Ethernet
PC
Detector (*)
D D

ON-OFF BOX
(**)

Hamdswitch cable (**)


230 VAC
Line 230 Vac & GND
cable (**)

COM2 serial cable (**)

(**) See DI 6070064 for interconnection details.

C C

AERODR INTERFACE UNIT (*)

TABLE
no POE

HUB
DETECTOR ASSEMBLY
POE POE POE

Detector Cable (*) Detector (*)

Interface

XGIFF
Panel
J5

Detector Cable (*)


Room Interface Cable (*)
J4

230 VAC & GND


B B

Line 230 Vac


Ethernet cable
(S0025659-02)

MAIN FUSES AND GND

ACCESS POINT
(It depends or
the country
code)

(*) Supplied by Konica


Ground Cable (A6383-02)

Generator
Cabinet Ground Cable (A6383-02)

Room Power
Connection ROOM CONFIGURATION
LINE
A CABLE E-Stop A
Line Voltage

NAME DATE SHEET / OF


230 VAC MAINS PBA: PCB:
DRAWING L.M.Pascual 06/09/2019
1/ 3
REVISED S.Pérez 06/09/2019 D C REV

54303172
D NC 19_0029 L.M.Pascual 06/09/2019
Confidentiality
n and privacy notice: All rights reserved. Unless otherwise specified and without written permission from SEDECAL, no part of this documentation may be reproduced or utilized in any form or by any means, electronic
or mechanical, including photocopying and microfilm, and may not be distributed or transmitted to third parties. C NC 17/0504 M.A.Gonzalez 10/01/2018
SEDECAL CONFIDENTIAL
ISID-R TABLE-WS MAN - AeroDR-CS7 - SHF
REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1

J4 6 J4
AERODR Interface Unit (*)
6

12

Room Interface XIF


AERODR Interface Unit (*) 12 CN2
13 Cable (note 1)

GENERATOR CABINET 13 1

D 10 XGIF Assy D
KM-READY
Room Interface 10 2
Generator Interface Cable 11 Cable (note 1) CN4
Konica AeroDR - SHF 1K2
cable (A15800-01) 11
ATP-CONSOLE
Board 5 1 3
(A3024-XX) 1
PREPARE
J4 5
RP3-22 ohm PREP 7 PREPARE
2 CN3
J13-9 6 1K2 7

U7-14 3 1
EXPOSE
not connected
12

GND 8 4 2
4
not connected
13 1K2 EXPOSE
8
RP3-22 ohm ALOE 5 3
9 MAINGATE
J2-24 TS1-52 10
9
U7-14 12 Vdc 6 4
not connected
11 2
MAINGATE
2 5
TS1-15 5 3
GND
C C
CN3 3
7 4 6

Interface Control 3 4
Board
(A3009-XX) KM-READY
2 1
K5
BUCKY1 DR 1
J4-8 TS1-4 8 4
14

BUCKYSPLY (note 1) Supplied by 14


J4-9 TS1-2 9 15 Konica

15
XGIF interface (note 1) Supplied by
TS1-6 2 Konica
XIF interface

24 Vdc TS1-7 not connected


3 AERODR Interface Unit (*)

TS1-13 not connected


4
K3
3
K5
GND
BUCKY1 MOT GIPS
J4-11 TS1-5 1
B J4 B

24 Vdc TS1-14 8
not connected
14

GND
TS1-17 15 9 Room Interface
Cable (note 1) XCN2
BUCKYREADY

2 1

3 2

X See time diagram on page 3 4 3

1 BUCKYSTART
5

14

15
A A
(note 1) Supplied by DETECTOR WIRING
Konica
GIPS interface NAME DATE SHEET / OF
PBA: PCB:
DRAWING L.M.Pascual 06/09/2019
2/3
REVISED S.Pérez 06/09/2019 D C REV

54303172
D NC 19_0029 L.M.Pascual 06/09/2019
Confidentiality
n and privacy notice: All rights reserved. Unless otherwise specified and without written permission from SEDECAL, no part of this documentation may be reproduced or utilized in any form or by any means, electronic
or mechanical, including photocopying and microfilm, and may not be distributed or transmitted to third parties. C NC 17/0504 M.A.Gonzalez 10/01/2018
SEDECAL CONFIDENTIAL
ISID-R TABLE-WS MAN - AeroDR-CS7 - SHF
REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1

D D

WS Configuration:
bucky 1 selected

Enabled

Disabled Disabled
X-Ray enable (SW message)

12 Vdc_ISO
PREP (handswitch first step)
atp-Console K1

12 Vdc_ISO
EXP (handswitch second step)
atp-Console K2
C C

12 Vdc

1 PREP - PREPARE input Generator starts the rotor


and heats the filament
AND

generator READY:
filament and rotor ON

12 Vdc

2 BUCKY 1 DR - EXPOSE input

24 Vdc

3 BUCKY 1 MOT - KM READY output detector accumulation time

B 12 Vdc B

4 ALOE - MAINGATE input exposure time

A
INTERFACE TIME DIAGRAM A

NAME DATE SHEET / OF


PBA: PCB:
DRAWING L.M.Pascual 06/09/2019
3/3
REVISED S.Pérez 06/09/2019 D C REV

54303172
D NC 19_0029 L.M.Pascual 06/09/2019
Confidentiality
n and privacy notice: All rights reserved. Unless otherwise specified and without written permission from SEDECAL, no part of this documentation may be reproduced or utilized in any form or by any means, electronic
or mechanical, including photocopying and microfilm, and may not be distributed or transmitted to third parties. C NC 17/0504 M.A.Gonzalez 10/01/2018
SEDECAL CONFIDENTIAL
ISID-R TABLE-WS MAN - AeroDR-CS7 - SHF
REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
Interconnection 1.-- For Generator Interface with control relay and externally powered

POWER TS1 X--RAY ROOM


MODULE
47
}
ROOM LIGHT SUP
115 / 220 VAC

48
ROOM LIGHT SW
ROOM
WARNING
LIGHT

Interconnection 2.-- For Generator Interface with control relay and internally powered

POWER TS1
MODULE X--RAY ROOM
115 or 220 VAC SW
*
add jumper
ROOM
47 WARNING
ROOM LIGHT SUP LIGHT
48
ROOM LIGHT SW

0 VAC
(in TS1-3 or TS1-10) *

* Select the power supply on TS1 according to the lamp voltage.


Add jumper to TS1-26 for 220 VAC or to TS1-27 for 115 VAC.

Interconnection 3,-- Additional option to meet some Local Electrical Codes

POWER
MODULE TS1
ROOM LIGHT SUP 47

Contact 7A 1/6HP
125, 250 V AC
3A 30 V DC
48
ROOM LIGHT SW

NAME DATE SHEET / OF


E I/F update F. GARCIA 20/06/04 DRAWING F. GARCIA 04/19/95
I/F--008
1/1
D I/F update F. GARCIA 02/02/02 REVISED A. DIAZ 04/19/95 E D C B A REV
C I/F changing F. GARCIA 05/05/99
B ROOM WARNING LIGHT INTERFACE
SEDECAL
TS1 changing F. GARCIA 24/05/96
A EMC F. GARCIA 29/03/96 INTERFAZ LAMPARA RAYOS--X SALA
REV DESCRIPTION ISSUED BY DATE
GENERATOR
POWER
MODULE
TB1
AEC ADAPTATION BD. ( A3263--03 )
TS1 POWER INTERFACE Cable 9 ION CHAMBER #1
J1
42 1 1 ION CHAMBER Cable Preamplifier
IC GND 1

16 --12 VDC 2 7 --12 VDC 7

15 +12 VDC 3 8 +12 VDC 8


17 +24 VDC 4 9 GND 9

18 GND 8 5 IC1 OUTPUT 5


IC1 OUT
5 +12 VDC 3 CENTER AREA 3
FLD2--IC1
6 C1 C3 2 LEFT AREA 2
-- 4 0.1 uF 100uF FLD1--IC1
7 U1 35V 6 RIGHT AREA 6
TP1 RP4 +12 VDC TP2 RP4 +12 VDC FLD3--IC1
jumper IC1 10K IC2 10K LF347
10 TP6 RP1 +12 VDC R12 R7 4 RESET 4
10K 5 6 RP3 3K01 1 2 RP3 3K01 + STRT--IC1
IC4 R6 11
10K 10K
5 6 RP2 3K01 R10 -- 9 R11 R9 -- 13 R8 C2 C4 Screen
10K 1K U1 6 5 1K U1 2 1 1K --12 VDC 100uF
R4 -- 6 R5
LF347 RP3
1K
RP3
0.1 uF
35V
CONSOLE 1K U1 6 5 1K 8 LF347
AEC INTERFACE Cable 10K 14 10K
J5 J4 LF347 RP2 C10 + 10
C9 + 12
+12 VDC TP5
7 10K 7 7 8 3 3 4 TP4 RP1
0.1 uF 0.1 uF
5 GND 5 C8 + 5 RP4 RP4
IC3 10K R1 GND
ION CHAMBER #2
0.1 uF 7 7 8 8 10K RP2 3K01 J2
4 10K 1 2
RP1 JP9 JP10 10K ION CHAMBER Cable Preamplifier
7 IC4 INPUT 7 8 10K B A
IC1 OUT
B A
IC2 OUT R3 -- 2 R2 1 1
JP12 1K U1 2 1 1K
IC4 OUT --12 VDC
1 IC1 INPUT 1 B A LF347 RP2 7 7
1 10K --12 VDC
6 6 C7 + 3 +12 VDC
IC2 INPUT 0.1 uF 3 3 4 8 +12 VDC 8
RP1
2 IC3 INPUT 2 4 10K 9 9
JP11 GND
IC3 OUT
8 --FLD 2 DR 8 B A 5 IC1 OUTPUT 5
IC2 OUT
3 --FLD 1 DR 3 3 CENTER AREA 3
FLD2--IC2
4 --FLD 3 DR 4 2 LEFT AREA 2
FLD1--IC2
9 --STRT DR 9 6 RIGHT AREA 6
FLD3--IC2
Shield 4 RESET 4
K10 K6 K2 STRT--IC2
JP8 JP6 JP14 K15
A A Screen
C6 6 1 6 1 6 1 A 6 1
100uF B B B
35V CR2
1N4148 13 8 13 8 13 8 13
FLD2--IC1 FLD2--IC2 FLD2--IC3 8
J3 ION CHAMBER #3
ION CHAMBER Cable Preamplifier
C5 1 1
K11 K7 K3 K14
0.1 uF JP4 JP2 JP16
A A 7 --12 VDC 7
6 1 6 1 6 1 A 6 1
B B B 8 +12 VDC 8
CR3
1N4148 13 8 13 8 13 8 13 8 9
FLD1--IC1 FLD1--IC2 FLD1--IC3 9 GND
5 IC1 OUTPUT 5
IC3 OUT
K12 K8 K4 K13 3 CENTER AREA 3
JP7 JP5 JP15 FLD2--IC3
A A A 2
6 1 6 1 6 1 6 1 LEFT AREA 2
B B B FLD1--IC3
CR4 6 RIGHT AREA 6
1N4148 13 8 13 8 13 8 FLD3--IC3
FLD3--IC2 FLD3--IC3 13 8
FLD3--IC1 4 RESET 4
STRT--IC3
Screen
TP3 K9 K5 K1 K16
JP3 JP1 JP13
ST
A A 6 1 A
6 1 6 1 6 1
R13 B B B
10K CR1
1N4148 13 8 13 8 13 8 J5 ION CHAMBER #4
STRT--IC1 STRT--IC2 STRT--IC3 13 8
1 ION CHAMBER Cable Preamplifier
1

7 --12 VDC 7

8 +12 VDC 8
9 GND 9

5 IC1 OUTPUT 5
IC4 OUT
JUMPERS JP9, JP10, JP11, JP12 Ion Chamber Jumpers Jumpers Jumpers 3 CENTER AREA 3
Type JP3,JP4,JP7,JP8 JP1,JP2,JP5,JP6 JP13, JP14, JP15, JP16 2 LEFT AREA 2
POS. A -- NO OFFSET ADJUSTMENT
IC1=IC2=IC3=IC4 B B B 6 6
POS. B -- OFFSET ADJUSTMENT RIGHT AREA
4 RESET 4
IC1=IC2=IC3 B B A
Screen
IC1=IC2 B A A
/ / /
IC1=IC2=IC3=IC4 A A A

NAME DATE SHEET / OF


Note.-- Compatible interface with preamplifier for Ion Chamber types : DRAWING F. GARCIA 07/08/96
DWG:
IM--015
1/1
-- Expos--AID REVISED A. DIAZ 10/10/96 B A REV
-- Vacutec 70145/70151
-- Comet Ion Chambers with PA--021 Preamplifier B new interface F. García 30/09/10
AEC COMPATIBILITY with
A Version 03 F. GARCIA 10/10/99 SEDECAL
-- Claymount with preamplifier SSMC1006 AID / VACUTEC / COMET / CLAYMOUNT
REV DESCRIPTION ISSUED BY DATE
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

CHAPTER 3
CONFIGURATION AND CALIBRATION

3-1
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

3-2
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 1 INTRODUCTION

Configuration provides the initial settings for extended memory and checkout
procedures that must be carried out before making X-ray exposures. When the
initial setup and checkout has been completed the generator will be ready for
calibration.

Calibration information and procedures to perform all the adjustments required


to establish optimal performance of this generator are contained in this chapter.

Some configuration and calibration data are entered in digital form and stored
in a non-volatile memory chip (U3-EEPROM) located on the HT Controller
Board, thus no battery back-up is required.

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES (LEDS) ON THE CHARGE-DISCHARGE MONITOR
BOARDS ARE OFF, APPROXIMATELY THREE (3) MINUTES
AFTER THE UNIT IS TURNED OFF.

DO NOT SUPPLY THE MAIN POWER FROM THE ELECTRICAL


ROOM CABINET (MAIN DISCONNECT) OR TURN ON THE
SYSTEM UNTIL SPECIFICALLY INSTRUCTED IN THIS
DOCUMENT.

Note . Configuration and Calibration procedures must be performed in


the order listed in this document.

Note . Record all the configuration settings and calibration data in the
Data Book (refer to Appendix A, “Data Book,” at the end of this
Chapter).

3-3
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1.1 TOOLS AND TEST EQUIPMENT

The following test equipment is required for Configuration and Calibration:

 Standard service engineers tool kit including Allen and Torx key sets, and
a Trimmer for Ion Chamber Pre-Amplifier (AEC tiny potentiometer).

 Calculator.

 Digital Multimeter.

 Non-invasive kVp Meter.

 Digital mAs Meter.

 Only for AEC purposes:

G Homogeneous Phantom of Aluminium with a purity of not less that


99% and thickness of 21 mm for the Collimator Filter Holder
(Part no. 2281087-2).

G Homogeneous Phantoms of PMMA (Polymethylmethacrylate)


with a surface of 43 x 43 cm and different thickness (cm) to
achieve all the different thickness used in the AEC verification
tests: 8, 10, 15, 20, 25 and 28 cm.

G (Optional) Dosimeter, Unfors RaySafe Xi (or equivalent).

G Adhesive Tape.

1.2 GENERATOR SPECIFICATIONS

1.2.1 MINIMUM CURRENT TIME PRODUCT (mAs)

 Minimum Current Time Product obtained at 0.1 s is 1 mAs.

 Minimum Current Time Product within the specified ranges of


compliance for linearity and constancy is 0.1 mAs.

3-4
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1.2.2 EXPOSURE ACCURACY

Note . Specified accuracy does not include test equipment accuracy.

ACCURACY
PARAMETERS
(with 12 BITS HT Controller)

kVp ± 3% (±1 kVp)

RAD mA ± 4% (±1 mA)

Exposure Time ± 2% (±0.1 ms)

1.2.3 HV FREQUENCY

The operating HV Frequency of this generator is 25 kHz / 30 kHz.

1.2.4 DUTY CYCLE

The Generator duty cycle is continuous, but limits should be set during
installation depending on the capacity of the X-ray Tube.

3-5
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

3-6
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 2 CONFIGURATION AND TEST SWITCHES

Specific versions of EPROM-U24 on the ATP Console CPU Board and


Microcontroller-U5 on the HT Controller Board are based on the generator
configuration. (Refer to Illustration 3-1.)
Illustration 3-1
EPROM and Switch locations

INTERFACE CONTROL
-- A3009
J8 3000SW2
3024SW4
U5

J7

HT CONTROLLER
--A3000

U24

POWER SUPPLY

FILAMENT
DRIVER
A3004
3024SW2

3024SW3
POWER CABINET -- FRONT PANEL

3024SW1

ATP CONSOLE CPU BOARD -- A3024

ATP Console DIP Switches must be configured with the generator turned OFF,
and they are read when the generator is turned ON again.

3-7
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.1 3024SW1 - ATP CONSOLE CPU BOARD

Set DIP switch 3024SW1 in accordance with Table 3-1.

Table 3-1
System Configuration DIP Switch 3024SW1 on the ATP Console CPU Board

3024SW1 POSITION OPEN (OFF) CLOSED (ON)

1 60 Hz *1) 50 Hz *1)

Normal -- Application mode.


2 -- Exposures are inhibited when
Door Interlock Switch is opened.

3 Not used. Set in “OFF” position. Not used.

4 Not used. Set in “OFF” position. Not used.

Note.-- *1) This switch is related with the frequency of the Rotor Controller. For Generators with High Speed Starter
(LV-DRAC) set always SW1-1 to 60 Hz, in the rest of Generators set SW1-1 in accordance with the Power Line
Frequency.

2.2 3024SW2 - ATP CONSOLE CPU BOARD

Set DIP switch 3024SW2 verifying that each position is set as per Table 3-2.

Table 3-2
Test DIP Switch 3024SW2 on the ATP Console CPU Board

3024SW2 POSITION OPEN (OFF) CLOSED (ON)

Skips reception with the HT Controller. Use only


for display purposes, troubleshooting or Demo
1 Consoles when there is no Power Module. Be Normal -- Application mode.
sure that J3 connector is not plugged to the ATP
Console CPU Board.

Tick Sound (button / command acknowledge) is Tick Sound (button / command acknowledge) is
2
not emitted by the ATP Console CPU Board. emitted by the ATP Console CPU Board.

3 Normal -- Application mode. Service Mode .

kV Log (Renard) Scale Mode for kV variation kV Lineal Scale Mode for kV variation (Normal
4
(kV changes in logarithmic steps) (if available). mode) (kV changes one by one).

3-8
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.3 3024SW3 - ATP CONSOLE CPU BOARD

DIP switch 3024SW3 is not used for configuration but all the switches must be
set in the OFF position.

2.4 3024SW4 - ATP CONSOLE CPU BOARD

DIP Switch 3024SW4 is not used for configuration but all the switches must be
set in the OFF position.

2.5 3000SW2 - HT CONTROLLER BOARD

HT Controller DIP switches can be configured while the generator is ON except


DIP switch 3000SW2-1.

Set DIP switch 3000SW2 as indicated in Table 3-3.

Table 3-3
Test DIP Switch 3000SW2 on the HT Controller Board in the Power Module

3000SW2
OPEN (OFF) CLOSED (ON)
POSITION

Programming the Rotor Boost Time and


1 Normal.
Filament Setting Time *1) *2)

2 Normal. Bypasses: Filament, Rotor Ready, Error E11. *1) *3)

3 Normal -- Not used. Not used.

4 Normal -- Digital mA Loop Closed Digital mA Loop Open / Filament Current Constant *1)

5 125 kV *4) 150 kV *4)

6 Normal -- Generator with Tube-1 Not used.

7 Normal -- Filament Boosting for Tube--1 No Filament Boosting for Tube--1 *4)

8 Normal -- Not used. Not used.

Notes.-- *1) Set in Closed (ON) position only for Calibration and Testing procedures when indicated in the Service Manual.

*2) Note that SW2-1 in Closed (ON) position is only set to program the Rotor Acceleration Time and the Rad Filament
Settling Time, therefore it changes the functions of Switches SW2-2, SW2-7, and SW2-8. Refer to Section 3 of this
chapter.

*3) This turns the filaments off so no radiation will be produced during the exposure.

WARNING: THE kV OUTPUT OF THE HV TRANSFORMER WILL BE WHATEVER IS SET BY THE CONSOLE. IF
THE X-RAY TUBE HV CABLES ARE NOT CONNECTED INTO THE HV TRANSFORMER, FILL COMPLETELY BOTH
HV RECEPTACLES WITH HV OIL.

*4) Set SW2-5 according to the Generator kV rating (refer to the Generator model and/or specifications).

*4) Set to “No Filament Boosting” when using X-ray Tubes with a Small Focal Spot smaller than 0.6.

3-9
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.6 BASIC CONFIGURATION OF GENERATOR BOARDS

The following Jumpers are factory set or removed to configure the Generator
Boards according to the customer order. Check the jumper positions in the
Generator Boards.

GENERATOR BOARDS JUMPERS POSITION

JP1, JP2, JP3, JP5 and JP6 in “2” (soldered)


HT CONTROLLER BOARD
(A3000--80/81)
JP4 in “1” (soldered)

FILAMENT CONTROL BOARD


W1 in “A” (installed)
(A3004--10)

W1 in “2--3” (installed)
INTERFACE CONTROL BOARD
W2 in “1--2” (installed)
(A3009--11)
W3 to W8 removed

JP1, JP2 and JP3 in “B” (soldered)

JP4 in “A” (installed)

JP5 in “B” (installed)

JP6 in “A” (soldered)

Connector J8 configured for RS232 so:


JP9, JP10 and JP11 in “A” (soldered)
ATP CONSOLE CPU BOARD
JP7, JP8, JP21 and JP22 do not matter jumpers position (in “A” soldered)
(A3024--221)
JP12 removed

JP13 (not installed)

JP14 (soldered)

JP15, JP16, JP17 and JP18 removed

JP19 in “A” (soldered)

JP1 in “C” (installed)


AEC CONTROL BOARD
JP2 and JP3 in “A” (installed)
(A3012--25)
JP4 in “B” (installed)

JP1, JP2, JP3, JP4, JP5, JP6, JP7, JP8, JP13, JP14, JP15 and JP16 in “B” (installed)
AEC ADAPTATION BOARD
(A3263--03)
JP9, JP10, JP11, JP12 in “A” (installed)

GCA BOARD
W1 in “ON”
(A3548--01)

3-10
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 3 GENERATOR SERVICE MODE

The Generator Service Mode (GSM) program allows the access to the service
screens for Reading the Exposure Counters, Configuration and Calibration
procedures.

Note . The TechService Sedecal Application must be installed on the


CS7 PC Workstation.

The Generator PC Interface Box should be connected to COM 1 Port at the CS7
PC Workstation (refer to Section 3.4.1 of the Installation chapter). Check the
proper configuration of the COM Ports in the PC and modify it if needed.

If, for any reason, it is necessary to check or modify the COM Port where the
Generator PC Interface Box is connected to the CS7 PC Workstation, proceed
as follow:

1. Go to the Windows Desktop:

a. Power ON the Proteus XR/f.


If prompted log on as:
user: maintenance
password: kmmg5678

b. Select:

1 2 3

“Utility” button is located


at the bottom of the Enter the password (kmmg5678) and press “OK”
“System Monitor” screen 4

Press “Close”

Continues in next page

3-11
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Press “Shutdown” and confirm with “OK”

Click “Windows Desktop”

2. Externally in the PC, verify which is the COM Port where the Generator
PC Interface Box is connected (for example COM 3). Then check or
modify the COM Port configuration in the PC and in the TechService
Sedecal Application, as described in the next steps 3. and 4.

3-12
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. Configuration of COM Port in the PC.

Open a folder in the Windows Desktop and select


“This PC” --> “Manage” --> “Device Manager” --> “Ports COM & LPT)” Continues in next page

3-13
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Select the corresponding “Communication Port” and “Properties”


Select the “Port Settings” tab --> “Advanced” --> the chosen “COM Port number”
Save the selection by clicking “OK” --> “OK”

3-14
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Configuration of COM Port in the TechService Application.

From the Windows Desktop, look up the “TechService” configuration settings file, the complete path is:
“This PC \ OS(C:) \ Program Files (x86) \ Sedecal TechService \ TechService”

Edit this file with “Notepad”, modify the number of the chosen “COM Port” and save the file.

3-15
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

To start the GSM program perform the following sequence:

1. From the CS7 Desktop, launch the TechService application on the CS7
PC Workstation as follows:

a. With the Generator power off, verify that the Generator is in


Service Mode, that is, dip switch SW2-3 on the ATP Console Board
is in ON position.

b. Power ON the Proteus XR/f.


If prompted log on as:
user: maintenance
password: kmmg5678

c. Select:

1 2 3

“Utility” button is located


at the bottom of the Enter the password (kmmg5678) and press “OK”
“System Monitor” screen 4

Press “Close”

Press “Shutdown” and confirm with “OK”

3-16
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

d. Click “Windows Desktop” and launch the “TechService” shortcut.

After accessing to the GSM program, a black Information Area appears at the
lower right corner of the screen to show some messages related with the
process (e.g. “Power Up the Generator”).

If after pressing the “Configuration”, “Calibration” or Auto Calibration” buttons,


the GSM program prompts an error message: “Please check calibration dip
switch and toggle with power off”; it means that these functions are disabled
because position of dip switch 3024SW2-3 on the “ATP Console CPU Board”
is not in “Service Mode Allowed”. Turn OFF the generator, change the dip
switch 3024SW2-3 to Closed (ON) position, turn ON the generator and start the
GSM program again.

Illustration 3-2
Example of GSM Menu

3-17
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

The screens to “Read Exposure Counters”, to enter in “Configuration”, “Manual


Calibration” or “Auto Calibration” are displayed after selecting the respective
button on the right side. Press again the selected button (in yellow on the right
side) to return to the GSM menu.

The “Exit” button can also be used to return to the GSM menu, specially if the
other buttons are disabled.

Note . Whenever the “Configuration” menu is closed (by pressing any of


the “Configuration” or “Exit” buttons) a double-beep will sound
confirming the storage of the values set for each workstation.

During operation, the color of the selected buttons changes to yellow when they
are selected.

Press the “Close” button to exit the GSM program.

Exposure status indicators for “Ready” and “X-ray On” are located on the upper
right area of the GSM screens. The “Information Area” displays data related to
the service mode, remaining heat units, working mode, errors indicators, etc.

When an error code or message is displayed on the GSM program press the
“Reset Error” button to reset the error indication.

Illustration 3-3
Status and Error Indicators

Workstation Selected Exposure Indicators

Exit Button

Button of Menu Selected

Reset Error Button

Error Indicator

Information Areas

3-18
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3.1 EXPOSURE COUNTERS

The Exposure Counters display the number of Rad exposures, in this case read
the exposures for Tube-1 RAD.

1. Enter in the GSM program and select the “Read Exposure Counters”
button.

2. This screen shows the Exposure Counters.

3. Exit from this screen by pressing the “Read Exposure Counters” button
or the “Exit” button again.

Illustration 3-4
Exposure Counters

3-19
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3.2 WORKSTATIONS CONFIGURATION

This screen is used to view the configuration of the Workstations (Tube, WM,
Ion Chamber, etc.) and the AEC Areas selected by default for each Workstation
in the system (optional).

The workstations can be configured according to the customer preferences or


default. If a workstation is configured with the value “Tube -- 0”, its button can
not be selected during operation.

CUSTOMIZED CONFIGURATION

Note . This procedure has to be performed always that “ATP Console


CPU Board” is replaced by a new one.

1. Enter in the GSM program and select the “Configuration” button.

2. Once in Configuration mode, it indicates “RAD”.

Illustration 3-5
Configuration screen

3-20
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. Select the first workstation to be configured, by pressing the respective


button (only the icon of the selected workstation has different color). The
following values can be shown on the “Configuration” screen:

DISPLAY FUNCTION VALUE DESCRIPTION

0 No-configured workstation (Not selectable by the operator)

TUBE Tubes 1 Tube-1

2 Tube-2

0 -- Direct Direct

First Receptor, usually in the Table: CR-Film inserted in the


Tray/Bucky or DR (Direct DR or DR inserted in the Receptor
1 -- Bucky 1 cabinet).

* Konica-Minolta Detectors use only the signal as Bucky 1.

Second Receptor, usually in the Wall Stand: CR-Film inserted


Devices - 2 -- Bucky 2 in the Tray/Bucky or DR (Direct DR or DR inserted in the
WM Receptor cabinet).
Working
g Modes

3 -- STD Tomo Standard Tomo

4 -- STD RF Standard RF (Spot Film Device)

5 -- DSI Digital RAD and Fluoro

6 -- Cine Cine

7 -- DSA DSA

0 No AEC

1 Ion Chamber-1 (IC-1)

2 Ion Chamber-2 (IC-2)


AEC IC AEC Ion Chambers
3 Ion Chamber-3 (IC-3)

4 Ion Chamber-4 (IC-4)

5 Photomultiplier (PT-INPUT)

WS LOCK Not used

kV
Not used
TRACKING

RAD/RF Generator Type RAD Generator for RAD application

Notes: Some of listed values are not configurable depending on the Positioners / Generator model .

3-21
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Set the new value by pressing the corresponding “Increase” or


“Decrease” buttons. Set also the selected AEC Areas by default for each
Workstation in the system (optional).

5. Select the next workstations to be configured and set the respective


values of each one.

6. Exit from configuration mode by pressing the “Configuration” button, then


a double-bip will sound confirming the process.

The next table shows the recommended configuration of Workstations in a


Proteus XR/f System.

WS1 = DR1 Direct mode on the Table Top


without AEC Ion Chamber
nor Alignment Interlocks / Receptor Detection

WS2 = DR1 in the Table Receptor Cabinet


with AEC Ion Chamber
and Alignment Interlocks / Receptor Detection

WS3 = DR2 in the Wall Stand Receptor Cabinet


with AEC Ion Chamber and
and Alignment Interlocks / Receptor Detection

WS4 to WS7 = Not used.

WS8 = Direct mode without AEC Ion Chamber


nor Alignment Interlocks / Receptor Detection

Direct on Table Wall Stand Service


N/A N/A N/A N/A
Table Top Receptor Receptor Use
WORKSTATION
WS1 WS2 WS3 WS4 WS5 WS6 WS7 WS8

TUBE 1 1 1 0 0 0 0 1

WM 1 1 1 0 0 0 0 0

AEC IC 0 1 2 0 0 0 0 0

WS LOCK 0 0 0 0 0 0 0 0

kV TRACKING N/A N/A N/A N/A N/A N/A N/A N/A

RAD/RF RAD RAD RAD RAD RAD RAD RAD RAD

3-22
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

DEFAULT CONFIGURATION

Default configuration sets some default values to each workstation. It only


should be used to re-initialize the workstation configuration when the complete
configuration has been lost or it is not possible to select any workstation.

1. Enter in the GSM program and select the “Configuration” button.

2. Press the “Default Configuration” button.

3. Exit from configuration mode by pressing the “Configuration” button, then


a double-bip will sound confirming the process.

4. It is recommended to perform a proper configuration of each workstation


in the system after a default configuration.

3-23
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

3-24
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 4 EXTENDED MEMORY SETTING

4.1 EXTENDED MEMORY LOCATIONS

Miscellaneous configuration and calibration data are stored in the Extended


Memory Locations. It is recommended to note the values factory stored in each
Memory Location. (Refer to Table 3-4)

Note . For generators with only one Radiographic X-ray Tube, this tube
has to be configured, calibrated and used as Tube-1.

Table 3-4
Extended Memory Locations

MEMORY LOCATION FUNCTION VALUE

E01 TUBE-1 -- RAD filament stand-by (Autocalibrated. Not field changeable)

E02 TUBE-1 -- RAD tube type

E03 Low Digital mA Loop Closed (from 10 mA to 80 mA)

E04 AEC-1 calibration

E05 High Digital mA Loop Closed (from 100 mA)

E06 kV Loop

E07 Maximum kW (Factory set. Only field changeable to lower value)

E08 AEC-1 tracking

E09 AEC-2 calibration

E10 AEC-2 tracking

E12 AEC Density Scale

E13 TUBE-1 -- Exposure Time adjustment - Delay

E15 TUBE-1 -- Exposure Time adjustment - Ceq kV

Note.-- Memory Locations not specified are not used.

3-25
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4.2 HOW TO ENTER AND STORE DATA IN THE EXTENDED MEMORY

The Extended Memory data are entered from the Console when the unit is in
service mode. Access to memory locations as indicated below:

1. Turn the Generator OFF and set the Test dip switch 3024SW2-3 on the
ATP Console CPU Board in “On” position to permit the service mode.

2. Power ON the System and select the “Manual Calibration” button on the
GSM program.

3. Select the needed Workstation (WS) using the “Up” and “Down” buttons,
then press the “OK” button to enter in calibration mode.

Note . In calibration mode, only the kV and mA parameters can be


modified, values for Time and mAs are factory programmed.

4. Increase the mA value beyond the maximum mA position. The mA


Display will show the first Extended Memory location (E01), they will
continue sequentially as the “Increase mA” button is pressed.

The values stored in each location are shown on the “Calibration Value”
Display after pressing the “Read” button or after pressing the “Increase”
or “Decrease” buttons of this panel. Since these buttons are also used
to increase or decrease the stored values, one number should be added
or subtracted from the reading, to obtain the current stored value.

5. Select the new value by pressing the “Increase” and “Decrease” buttons.
Each time these buttons are pressed the value displayed on the
calibration panel is increased or decreased one step.

6. Store the new value by pressing the “Store” button.

Note . If the “Store” button is not pressed after a new value is selected,
no modified data will be retained.

7. Exit from calibration mode by pressing the “Manual Calibration” button


again.

8. Turn the Generator OFF and set dip switch 3024SW2-3 on the ATP
Console CPU Board in “Off” position to place the Generator in normal
mode.

3-26
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Illustration 3-6
Calibration

Calibration Value Panel

3-27
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4.3 LIMIT OF MAXIMUM kW

The Maximum kW of the Generator is factory set according to the Generator


performance. Generator kW can be limited to a lower value.

Note . This limit can be set to a lower value to match the maximum
Generator power to the Line power, due to a high line impedance
(refer to Pre-installation document).

1. Enter in calibration mode by pressing the “Manual Calibration” button on


the GSM program. Select any workstation (WS) and press the “OK”
button.
2. Select the E07 Memory Location (memory location is shown on the mA
Display).
3. Set the new limit of Maximum kW by pressing the “Increase” or
“Decrease” calibration buttons and store the value by pressing the
“Store” button.
4. Exit from calibration mode.

Note . Record configuration data for E07 in the Data Book (refer to
Appendix A, “Data Book,” at the end of this Chapter).

3-28
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 5 X-RAY TUBE SELECTION

5.1 X-RAY TUBE INSERT PROTECTION CURVES

In order to properly select the X-ray tube insert protection curves for the X-ray
Tube connected to the Generator, perform the following procedure:

1. Enter in calibration mode by pressing the “Manual Calibration” button on


the GSM program. Select one of the workstation (WS) related to the
X-ray Tube to be configured. Then press the “OK” button.

2. Select the memory location E02 for Tube-1 (memory location is shown
on the mA Display).

3. Identify in Table 3-5 the X-ray tube that is being installed and note its tube
type number.

Table 3-5
X-ray Tube Numbers

POWER RATINGS
TUBE TUBE CODE FOCAL
MODEL KHU
NUMBER (ID) SPOT LS (kW) HS (kW)

005 412 CANON E7254FX 0.6 / 1.2 23 / 60 40 / 102 400

006 407 CANON E7884X 0.6 / 1.2 21 / 52 -- 300

Note .-- Power Ratings are for 60 Hz. To calculate Power Ratings for 50 Hz multiply the values by 0.91

4. Set the tube number by pressing the “Increase” or “Decrease” calibration


buttons until the correct number is showed on the “Calibration Value”
panel.

5. Store the value by pressing the “Store” button.

6. Verify that the tube code (ID) shown in the mAs Display is the same of
the tube code listed in Table 3-5. The tube code (ID) can be only read for
the selected X-ray Tube after pressing the “Store” button.

7. Exit from calibration mode.

Note . Record configuration data in the Data Book (refer to Appendix A,


“Data Book,” at the end of this Chapter).

3-29
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.2 GENERATORS WITH LF-RAC (LOW SPEED STARTER)

5.2.1 STATOR VOLTAGE AND CAPACITOR SELECTION

Check that the capacitor value of the Low Speed Starter


corresponds to the value recommended by the X-ray Tube
manufacturer. If needed replace the capacitor. Also, the Rotor
speed must be indicated by the manufacturer.

The DC Brake of the Low Speed Starter (LF-RAC) can be removed by


desoldering CR6 on the LF-RAC Board (refer to schematic 543020xx). In this
case, the Tube will remain coasting after releasing the “Prep” order.

5.2.1.1 CONFIGURATION FOR ONE TUBE WITH STANDARD STATOR

Voltage and capacitor is factory set to 220 VAC, 30 μF. In all cases, refer to
X-ray Tube Product Data.

5.2.1.2 CONFIGURATION FOR ONE TUBE WITH THE STARTING VOLTAGE AT 110 VAC

When the stator requires a starting voltage of 110 VAC (e.g. X-ray Tube Canon
E7252) perform the following modifications:

 If the Power Input Transformer 6T2 is for using with power lines up to
240 VAC (part number 50509030), remove the wire labelled as “4” that
is connected to Terminal 4 (230 VAC RTR) and connect it to Terminal 3
or 8 (110 VAC).

 If the Power Input Transformer 6T2 is for using with power lines up to 530
VAC (part number 50509029), remove the wire labelled as “4” that is
connected to Terminal 4 (230 VAC RTR) and connect it to Terminal 40
(110 VAC).

 For X-ray Tube Canon E7252 replace also the Fuse F1 (6A) on the
LF-RAC Board by another fuse of 10 A.

3-30
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.2.2 PROGRAMMING OF ROTOR ACCELERATION TIME AND RAD FILAMENT SETTING TIME

Rotor Acceleration Time is determined by the X-ray Tube and


Rotor characteristics and it must be considered when the
Generator is about to be configured. X-ray Tube could be
permanently damaged unless the required RPM are reached
before an exposure. (Refer to technical information of the
X-ray Tube).

Dip Switch 3000SW2 on the HT Controller Board is used to program:

 Rotor Acceleration Time. That depends on Stator voltage, Stator


frequency, Stator type, quality of X-ray tube bearings, and X-ray tube
anode size. A reed tachometer or a stroboscope can be used to
determine the anode RPM. Be sure that the Rotor Acceleration Times
meet all requirements for anticipated customer applications.

This value is programmable from 0.8 to 2.7 seconds. After this time the
Rotor is hold running in maintaining mode as long as “Prep” is active.

 Rad Filament Setting Time. This parameter has the same configuration
value than the Rotor Acceleration Time. Sometimes, if it is required to
increase the Rad Filament Setting Time to the next value, configure the
respective switches again. This adjustment avoids Error-12.

Note . The Rotor Acceleration Time and Rad Filament Setting Time is
factory set to 1.8 seconds. Maintain this value when it is unknown
or not provided with the X-ray Tube documentation.

1. Turn the Generator OFF and note current settings of the dip switch
3000SW2 on the HT Controller Board.

Note . Configuration of these times are only allowed when dip switch
3000SW2-1 is in “Closed” (On) position after power the Generator
OFF and back ON again.

3-31
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2. Set dip switches 3000SW2-1 and 3000SW2-2 as indicated below, in


order to enable the selection of times with the Low Speed Starter. Dip
switch3000SW2-1 has to be switched ONLY with the Generator
powered OFF.

3000SW2-1 (selection enable) 3000SW2-2 (Low Speed Starter)

ON OFF

3. Configure the Rotor and Filament Times by setting the dip switches
3000SW2-7 and 3000SW2-8 per Table 3-6.

Table 3-6
Low Speed: Configuration of Rotor and Filament Times

TUBE-1 3000SW2-7 3000SW2-8


ROTOR ACCELERATION TIME
AND FILAMENT SETTING TIME OPEN (OFF) CLOSED (ON) OPEN (OFF) CLOSED (ON)

0.8 seconds - -

1.2 seconds - -

1.8 seconds - -

2.7 seconds - -

Note . The default value of the Rotor Acceleration Time and Rad
Filament Setting Time for Tubes Canon E7254FX and E7884X is
set to 1.8 seconds.

Note . Record configuration data in the Data Book (refer to Appendix A,


“Data Book,” at the end of this Chapter).

4. To validate previous configuration, turn the Generator ON, wait until


Error--01 (E01) appears on the Console and turn the Generator OFF.

5. Set dip switch 3000SW2 to the original settings as noted in step-1. (Refer
to Section 2.5 for the normal settings of Dip Switch 3000SW2).

3-32
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.3 GENERATORS WITH LV-DRAC (HIGH SPEED STARTER)

The LV-DRAC is a High Speed Rotor Controller located on the Generator


Cabinet. Its configuration is made through dip switches 3243SW1 to 3243SW4
on the Control DRAC Board.

BEFORE MANIPULATING THE LV-DRAC, MAKE SURE THAT


THE INPUT LINE IS DISCONNECTED AND THE CAPACITOR
BANK IS PROPERLY DISCHARGED. WAIT UNTIL LEDS DL7
AND DL8 ON THE CONTROL DRAC BOARD ARE OFF.

Illustration 3-7
LV-DRAC Status Diagram

CHANGE
OF TUBE

ACCELERATION STAND-BY ACCELERATION


UP TO 3300 RPM UP TO 10000 RPM

RUNNING RUNNING
AT 3300 RPM AT 10000 RPM

BRAKING SELF-RUNNING MODE BRAKING


FROM 3300 RPM FROM 10000 RPM

The Function of these switches is the following:

3243SW1-x FUNCTION REMARKS

1
For further information about these switches
2 MINIMUM TIME FOR READY ((DELAY))
refer to Section 5.3.1
3

4 Set dip
p switches in accordance with the tables in
S i 5.3.2
Section 32
5
Also refer to position of dip switches:
6 TUBE TYPE SELECTION FOR TUBE-1
3243SW3-5, 3243SW3-7, 3243SW3-8
7
For further information about these switches
8 refer to Section 5.3.2

3-33
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3243SW2-x FUNCTION REMARKS

1 -- 8 NOT USED*

* Note: Set switches not used in “ON” position.

3243SW3-x FUNCTION REMARKS

ON = INHIBITED
LOW SPEED ROTATION BRAKE MODE
1
(3300 rpm to 0 rpm)
OFF = ACTIVATED

2 -- 4 NOT USED*

ON = WHEN THE TUBE CONFIGURED AS


TUBE-1 IS LISTED ON TUBE TABLE-1

TUBE-1 SELECTION DEPENDING ON OFF = WHEN THE TUBE CONFIGURED AS


5
THE TUBE LIST TUBE-1 IS LISTED ON TUBE TABLE-2

For further information about these switches


refer to Section 5.3.2

OFF= ALWAYS SET IN “OFF“.


6 PORT FLUORO MODE
DISABLE PORT FLUORO.

7 ON = ALWAYS SET IN “ON”


TUBE TYPE SELECTION
For further information about these switches
8 refer to Section 5.3.2

* Note: Set switches not used in “ON” position.

3243SW4-x FUNCTION REMARKS

1 -- 5 NOT USED*

ON = ENABLED SELF-RUNNING

OFF = DISABLED SELF-RUNNING


6 SELF RUNNING MODE
For further information about these switches
refer to Section 5.3.3

ON = INHIBITED

OFF = ACTIVATED
7 PROTECTIONS -- ERRORS (NORMAL OPERATION MODE)

For further information about these switches


refer to Section 5.3.4

ON = ACTIVATED
DC BRAKE MODE
8
(FOR LOW AND HIGH SPEED)
OFF = INHIBITED

* Note: Set switches not used in “ON” position.

3-34
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.3.1 MINIMUM TIME FOR READY (DELAY)

3243SW1-x
MINIMUM TIME
(SECONDS) 1 2 3

0* ON ON ON

0.6 OFF ON ON

0.8 ON OFF ON

1.0 OFF OFF ON

1.2 ON ON OFF

1.5 OFF ON OFF

2.0 ON OFF OFF

3.0 OFF OFF OFF

* Note: It is recommended to set these switches at “0 seconds”.

5.3.2 TUBE TYPE SELECTION FOR TUBE-1

TUBE-1 for CONVENTIONAL USE in TUBE TABLE-1

SWITCH 3243SWx-x
TUBE FAMILY
P
Pos
(Stator -  Anode) SW1-4 SW1-5 SW1-6 SW1-7 SW1-8 SW3-5 SW3-7 SW3-8

CANON 7254X
4 ON ON OFF ON ON ON ON ON
(4” Anode, XS-RB Stator (20/38 ohm), 400 KHU)

NOTE.-- Position not specified are nor used.

3-35
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.3.3 SELF-RUNNING MODE

The continuous starting and braking of the anode produces an overheating in


the X-ray Tube Stator (e.g. during calibration procedure). The “Self-running
mode” avoids this overheating.

In the “Self-running mode” (switch 3243SW4-6 in the “ON” position), the anode
remains running for approximately 1 minute when rotates at 3300 rpm or
10000 rpm.

The LV-DRAC enters in the “Self-running mode” when “Preparation” is selected


three consecutive times only from the Console buttons or Handswitch, and
the time between two of the consecutive accelerations is shorter than
30 seconds (refer to illustration below).

PREPARATION SIGNALS
t < 30 s t < 30 s t < 30 s t > 30 s t < 30 s t < 30 s t < tsr

ANODE ROTATION

tsr tsr
Self-running mode Self-running mode

tsr (time for self-running mode) = approx. 1 minute at 3300 rpm or 10000 rpm.

5.3.4 PROTECTIONS -- ERRORS

PROTECTIONS -- ERRORS 3243SW4-7

INHIBITED
ON
(LED DL1 on the Control DRAC Board is always illuminated)

ACTIVATED (Normal Operation mode)


(LED DL1 on the Control DRAC Board is non-illuminated OFF
indicating that the DRAC is working properly).

Note: The Error Codes are usually shown on the Console preceded by the letter “E”
(e.g. E51) (For Error Codes refer to the Troubleshooting document).

WHEN SWITCH 3243SW4-7 IS IN THE “ON” POSITION, THE


ERROR PROTECTIONS OF THE LV-DRAC ARE INHIBITED.

3-36
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.3.5 ANODE STATOR SELECTION

For Generators with High Speed Starter, the X-ray Tube Family (Anode Stator
+ Insert) is configured by setting the respective dip switches 3243SW1 (pos. 4
to 8) on the Control DRAC Board, as previously performed.

Configuration of these dip switches automatically determines the appropriate


Starting and Running Stator Voltage and Rotor Acceleration Time of the
selected Tube Family.

5.3.6 PROGRAMMING OF RAD FILAMENT SETTING TIME

Dip Switch 3000SW2 on the HT Controller Board is used to program:

 Rad Filament Setting Time. This value is programmable from 0.8 to 2.7
seconds. It can be initially set as the same value assigned for the Rotor
Acceleration Time (refer to technical information of the X-ray Tube)

Note . The Rad Filament Setting Time is factory set to 1.8 seconds.
Maintain this value when it is unknown or not provided with the
X-ray Tube documentation.

If it is required to increase the Rad Filament Setting Time to the next


value, configure the respective switches again. This adjustment avoids
Error-12.

1. Turn the Generator OFF and note current settings of the dip switch
3000SW2 on the HT Controller Board.

Note . Configuration of these times are only allowed when dip switch
3000SW2-1 is in “Closed” (On) position after powering the
Generator OFF and back ON again.

2. Set dip switches 3000SW2-1 and 3000SW2-2 as indicated below, in


order to enable the selection of times with the High Speed Starter. Dip
switch 3000SW2-1 has to be switched ONLY with the Generator
powered OFF.

3000SW2-1 (selection enable) 3000SW2-2 (High Speed Starter)

ON ON

3-37
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. Configure the Filament Setting Times by setting the dip switches


3000SW2-7 and 3000SW2-8 per Table 3-7. The Filament Setting Time
should be configured in accordance to Rotor Acceleration Time of the
X-ray Tube.

Table 3-7
High Speed: Configuration of Filament Setting Time

3000SW2-7 3000SW2-8
TUBE-1
FILAMENT SETTING TIME OPEN (OFF) CLOSED (ON) OPEN (OFF) CLOSED (ON)

0.8 seconds - -

1.2 seconds - -

1.8 seconds - -

2.7 seconds - -

Note . Record configuration data in the Data Book (refer to Appendix A,


“Data Book,” at the end of this Chapter).

4. To validate previous configuration, turn the Generator ON, wait until


Error--01 (E01) appears on the Console and turn the Generator OFF.

5. Set dip switch 3000SW2 to the original settings as noted in step-1. (Refer
to Section 2.5 for the normal settings of Dip Switch 3000SW2).

3-38
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.4 ANODE ROTATION TEST

Perform the following tests for checking the low and high speed when it is
required.

NEVER MAKE EXPOSURES DURING THE TESTS, THE


PERSON CLOSE TO THE X-RAY TUBE WILL BE EXPOSED.

Note . Two (2) people are needed for this test, one (1) at the Console and
the service engineer looking at the Anode of the X-ray tube. These
tests also can be done by hearing the sound of the anode rotating.

Note . For this test it is recommended to turn off the Tube Filaments
(switch 3000SW2-2 on the HT Controller in the “On” position) and
use a stroboscope to measure the anode speed.

1. With the switch 3024SW2-3 on the ATP Console CPU Board in “On”
position (service mode), turn the Console ON and select the
corresponding X-ray Tube.

2. Select a low value for kVp and mAs for checking the Anode Rotation at
Low Speed.

3. Press the “Prep” push-button and visually check that the Tube anode
rotates in the proper way. (Refer to the X-ray Tube documentation).

4. Hold pressed the “Prep” push-button and check that the rotation speed
of the Tube anode is in compliance with the X-ray Tube specifications.

5. Release the “Prep” push-button.

6. For Generators with LV-DRAC, select a high value for kVp and mAs for
checking the Anode Rotation at High Speed. Repeat steps 3, 4 and 5.

3-39
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5.5 FOCAL SPOTS CONFIGURATION

This configuration determines which mA station will be the smallest mA station


for the Large Focal Spot. It is possible to configure all the mA stations for the
Small Focal Spot or for the Large Focal Spot.

The smallest mA station for the Large Focal Spot must be selected according
to the Tube ratings for the Small Filament and the customer preference.

IF THE mA STATION FOR FOCAL SPOT CHANGE IS NOT


CONFIGURED ACCORDING TO THE X-RAY TUBE RATINGS,
THE TUBE FILAMENTS MAY BE PERMANENTLY DAMAGED.

1. With the generator OFF, set dip switch 3024SW2-3 on the ATP Console
CPU Board in “On” position to permit the service mode.

2. Power ON the System. Enter in calibration mode by pressing the “Manual


Calibration” button on the GSM program. Select any workstation (WS)
and press the “OK” button.

3. Select the smallest mA station for the Large Focal Spot using the
“Increase” or “Decrease” mA buttons. When is required to configure all
mA stations for the Small Focal Spot, select “E01” Memory Location.

Note . Default value is factory set at 200 mA except when using X-ray
Tubes with Small Focal Spot smaller than 0.6 .

4. Press the “Toggle” button to store the select mA station and then press
the “Confirm” or “Dismiss” buttons to confirm or cancel the process.
When it is confirmed, the ATP Console CPU Board emits a
“double-beep”.

IF THE FOCAL SPOT SWITCH-OVER POINT IS CHANGED


AFTER mA CALIBRATION, THE mA STATIONS AFFECTED
MUST BE RE-CALIBRATED.

5. Exit from calibration mode.

3-40
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Perform the following test (it is not mandatory).

1. Select the highest mA station for the Small Focal Spot. Verify that
effectively the Small Filament is ON (lighted) and the Large Filament is
OFF. Observe filaments through the X-ray tube window.

2. Select the lowest mA station for the Large Focal Spot. Verify that
effectively the Large Filament is ON (lighted) and the Small Filament is
OFF. Observe filaments through the X-ray tube window.

Illustration 3-8
Focal Spots Configuration

3-41
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

3-42
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 6 CALIBRATION PROCEDURES

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES (LEDS) ON THE CHARGE-DISCHARGE MONITOR
BOARDS ARE OFF, APPROXIMATELY THREE (3) MINUTES
AFTER THE UNIT IS TURNED OFF.

DO NOT SUPPLY THE MAIN POWER FROM THE ELECTRICAL


ROOM CABINET (MAIN DISCONNECT) OR TURN ON THE
SYSTEM UNTIL SPECIFICALLY INSTRUCTED IN THIS
DOCUMENT.

Calibrate the Generator immediately after Configuration is


completed.

Note . Enter in GSM Program for Calibration procedures as described in


the “Configuration” document.

Enter and store calibration data in the Extended Memory


Locations as described in Section 4.2.

Record all the calibration data in the Data Book (refer to Appendix
A, “Data Book,” at the end of this Chapter).

3-43
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Before calibration, bear in mind that:

 For calibration and measurement a kVp a non-invasive kVp meter is


needed.

 For calibration and measurement an mA or mAs Meter is needed


plugged to the banana connections on the HV Transformer (Digital mA
Loops calibration).

Note . Test points on the HT Controller Board can also be used to monitor
the kVp and mA readings but should not be used to calibrate the
unit. These test points must be checked with scope:

- mA test point is TP-5 and the scale factor is:


- from 10 to 80 mA, 1 volt = 10 mA
- greater than 100 mA, 1 volt = 100 mA

- kVp test point is TP-7 and the scale factor is 1 volt = 33.3 kVp
(0.3 volt = 10 kVp)

 Verify that DIP switch 3024SW2-3 on the ATP Console CPU Board is in
the ON position to permit the service mode.

 Verify the position of the DIP switches on the HT Controller Board during
every calibration procedure:

DIP SWITCH OPEN (OFF) CLOSED (ON)

Disables Filament and Rotor Interlocks


Position during operation -- Enables Filament and
3000SW2-2 (this turns off the filament so no radiation will
Rotor Interlocks
be produced during the exposure).

3000SW2-4 Position during operation -- Digital mA Loop Closed Digital mA Loop Open / Filament Current Constant

Note . Only for Generators with Low Speed Starter (LF-RAC) (it does not
apply to Generators with High Speed Starter - LV-DRAC):

-- When the Digital mA Loop is open (dip switch 3000SW2-4 in


“On”), the rotor runs for two minutes after releasing the handswitch
button from “Preparation” position. After this time the rotor will
brake (unless DC Brake is removed).

-- When the Digital mA Loop is closed (dip switch 3000SW2-4 in


“Off”), the rotor will brake after releasing the handswitch button
from “Preparation” position (unless DC Brake is removed).

3-44
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.1 FILAMENT STAND-BY CURRENT

Note . The Filament Stand-by value is auto-calibrated by the generator


and automatically stored into the E01 Memory Location. Filament
stand-by values are not field changeable.

6.2 EXPOSURE TIME ADJUSTMENT

The values stored in these Extended Memory Locations only affect Exposure
Times for techniques below 20 ms. The Memory Locations which affect short
exposure times are:

MEMORY LOCATION FUNCTION

E13 EXPOSURE TIME ADJUSTMENT - DELAY

E15 EXPOSURE TIME ADJUSTMENT - Ceq kVp

The generator has been optimized at the factory to produce correct exposures
at the lower times (<20 ms.) Therefore do not change the value factory set for
these Memory Locations (E13 and E15).

The Exposure Time adjustment is calibrated by performing the following steps:

1. Enter in calibration mode by pressing the “Manual Calibration” button on


the GSM Program. Select an available workstation (WS) and press the
“OK” button.

2. Select the E13 Memory Location. The value in this Memory Location
adjusts the time delay of the exposure. It is factory set for a value of 18.

3-45
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. Select the E15 Memory Location. The value in this Memory Location is
set in relation to the length in meters of one of the HV Cables
(1 ft = 0.3048 m). Verify the HV Cable length in meters and set the
following value:

HV CABLE LENGTH VALUE TO SET IN MEMORY

4m 27

6m 31

9m 38

12 m 45

14 m 49

16 m 54

For another HV Cable length value = (2.2711 x cables length) + 17.744

4. Store the value of each Memory Location by pressing the “Store” button
of the calibration panel.

5. Exit from calibration mode and record the new values in the Data Book.

Note . Record the configuration data for E13 and E15 in the Data Book.
(Refer to Appendix A, “Data Book,” at the end of this Chapter).

3-46
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.3 KV LOOP

Extended Memory Location E06 contains the calibration factor for kV Loop.
Each number above or below of the indicated in the E06 memory location
increases or decreases respectively the kV gain value.

Note . Value in E06 Memory Location is only related to the Generator


performance (it is not related to the X-ray Tube or another
components installed), so value in this Memory Location is
factory adjusted. Only perform this procedure if the HT Controller
Board and/or HV Transformer have been replaced in the unit.

The kV Gain for kV Loop can be manually calibrated with a


Non-Invasive kV Meter (recommended procedure) or
Auto-calibrated with HV Bleeder.

Manual Calibration of E06 Memory Location

1. With the Generator power OFF:

G Set Dip switch 3000SW2-2 on the HT Controller Board in “Off”


position (enables Filament and Rotor Interlocks).

G Set Dip switch 3000SW2-4 on the HT Controller Board in “On”


position (Digital mA Loop Open / Filament Current Constant).

G Remove the link between the banana plug connections on the HV


Transformer. Connect the mAs Meter to the banana plug
connections to measure mA or mAs.

G Place and center a Non-Invasive kVp Meter on the X-ray Tube


output at the required SID (refer to the Non-Invasive kVp Meter
documentation).

2. Enter in calibration mode by pressing the “Manual Calibration” button on


the GSM Program. Select an available workstation (WS) and press the
“OK” button.

3. Select the E06 Memory Location and read the calibration value by
pressing the “Read” button. Enter the value “200” and store it by pressing
the “Store” button.

3-47
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Calibrate manually the Filament Current Number for 80 kV / 200 mA


combination, as indicated in the following steps (if it has not been
previously calibrated).

In calibration mode, Filament Current Numbers are shown on the


calibration panel by pressing the “Read” button after selecting the
respective kV / mA combination. They can be changed by pressing the
“Increase” and “Decrease” buttons and stored by pressing the “Store”
button of the calibration panel.

Select 80 kV, 200 mA, Large Focus. Enter the value “600” as Filament
Current Number (calibration value) and press the “Store” button. Make
an exposure with these parameters. The mAs read on the mAs Meter
must be the same mAs displayed on the calibration screen with a
tolerance of ±0.1 mAs (tolerance of the parameter and mAs Meter).

If the mAs is low, increase the filament number. If the mA is high (or
“Generator Overload” Error is shown), decrease the filament number.
Press the “Store” button before making a new exposure. Repeat until the
mAs read is correct and the mA station is calibrated.

3-48
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Enter in user mode inside calibration mode by pressing the “Go to User
Mode” button.

6. Select: RAD Menu, 80 kV, 200 mA, 100 ms and Large Focus. Make an
exposure and note the kV value at the end of the exposure.

7. If calibration of the kV Loop is correct (80 ±1 kV), record value “200” in


the Data Book (refer to Appendix A, “Data Book,” at the end of this
Chapter).

8. If calibration of the kV Loop is not correct:

a. Exit from the “User Mode” screen by pressing the “Manual


Calibration” button. Select the E06 Memory Location. Press the
“Read” button to read the value stored.

b. Increase or decrease the value to increase or decrease the kV


respectively. Enter the new value and store it by pressing the
“Store” button.

c. Exit calibration mode and repeat the exposure (steps 5 and 6) to


determine if the new value has had the proper effect, if not repeat
step-8.

d. When it is correct, record the new value for E06 Memory Location
in the Data Book (refer to Appendix A, “Data Book,” at the end of
this Chapter).

9. After calibration of E06 Memory Location, remove the Non-Invasive kVp


Meter.

3-49
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Auto-Calibration of E06 Memory Location

1. With the Generator power OFF:

G Set Dip switch 3000SW2-2 on the HT Controller Board in “On”


position (disables Filament and Rotor Interlocks).

G Remove the HV Cables from the X-ray Tube and connect them to
the HV Bleeder, then connect a short couple of HV Cables from the
HV Bleeder to the X-ray Tube.

2. Enter in Auto-calibration mode by pressing the “Auto Calibration” button


on the GSM menu. Select an available workstation (WS) and press the
“OK” button. Then, select the E06 Memory Location and press the “OK”
button.

Confirm or leave the Auto-calibration by pressing the respective button


(“Yes” or “No”) on the new screen.

3-50
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Confirm or leave the Auto-calibration (second confirmation) by pressing


the respective button (“Confirm” or “Dismiss”) on the calibration screen.

Auto-calibration is activated when the “Auto Calibration” button is lighted


and the “Press Prep and Expose” message appears on the screen.

3. Make an exposure (technique parameters are pre-programmed at


100 kV, 200 mA and 32 ms and they can be shown when pressing the
“Prep” button).

4. Read the kVp measured with the HV Bleeder and enter this value on the
kV Display by pressing the “kV Increase” or “kV Decrease” buttons.

5. Exit from Auto-calibration mode pressing the “Auto Calibration” button.


At this moment, the Generator will calculate and store the new value in
E06 Memory Location. Auto-calibration is deactivated and the process
is finished when the screen shows the GSM menu.

6. Press the “Manual Calibration” button. Select an available workstation


(WS) and press the “OK” button. Then, select the E06 Memory Location
and read its new value on the Calibration Display. Record this value in
the Data Book (refer to Appendix A, “Data Book,” at the end of this
Chapter).

7. Exit calibration mode.

8. After calibration of E06 Memory Location:

G Switch the Generator power OFF.

G Remove the HV Bleeder and connect the HV Cables from the


Generator directly to the X-ray Tube.

3-51
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.4 DIGITAL mA LOOP CLOSED

Extended Memory Locations E03 and E05 contain the calibration factor for
Digital mA Loop Closed. Each number above or below the indicated in the
Memory Locations increases or decreases respectively the mA gain value.

Values in E03 and E05 Memory Locations are only related to


the Generator performance (they are not related to the X-ray
Tube installed), so values in these Memory Locations are
factory adjusted. Only perform this procedure if the HT
Controller Board and/or HV Transformer have been replaced
in the unit.

The mA Gain for Digital mA Loop Closed is calibrated by performing the


following steps:

1. With the Generator power OFF, set:

G Dip switch 3000SW2-2 on the HT Controller Board in “Off” position


(enables Filament and Rotor Interlocks).

G Dip switch 3000SW2-4 on the HT Controller Board in “On” position


(Digital mA Loop Open / Filament Current Constant).

G Remove the link between the banana plug connections on the HV


Transformer. Connect the mAs Meter to the banana plug
connections to measure mA or mAs.

2. Calibration of E03 and E05 Memory Locations is performed by means of


the “Auto Calibration” menu.

EACH TIME THAT AUTO-CALIBRATION IS ACTIVATED IN


ONE OF THESE MEMORY LOCATIONS, CALIBRATION DATA
STORED FOR THIS MEMORY LOCATION IS DELETED AND A
NEW CALIBRATION FOR IT WILL BE REQUIRED.

3-52
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. Enter in Auto-calibration mode by pressing the “Auto Calibration” button


on the GSM menu. Select an available workstation (WS) and press the
“OK” button. Then, select the respective Memory Location (E03 or E05)
and press the “OK” button.

Confirm or leave the Auto-calibration by pressing the respective button


(“Yes” or “No”) on the new screen.

Confirm or leave the Auto-calibration (second confirmation) by pressing


the respective button (“Confirm” or “Dismiss”) on the calibration screen.

3-53
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Auto-calibration is activated when the “Auto Calibration” button is lighted


and the “Press Prep and Expose” message appears on the screen.

4. Calibrate E03 and E05 Memory Locations as described in the following


pages.

3-54
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

For Low mA stations (from 10 mA to 80 mA) (E03 Memory Location):

1. Enter in Auto-calibration mode by pressing the “Auto Calibration” button


on the GSM menu and select the E03 Memory Location. Confirm all the
following screens.

2. When Auto-calibration is activated, “Auto Calibration” button is lighted


and the mAs value is shown on the screen (3.5 mAs).

3. Perform the following steps:

a. Make an exposure (technique parameters are pre-programmed at


80 kV, 32 mA and 100 ms and they can be shown when pressing
the “Prep” button).

b. Read the mAs measured on the mAs Meter.

c. Read the Filament Current Number shown on the calibration


panel.
d. Increase or decrease the Filament Current Number (by pressing
the “Increase” or “Decrease” buttons) to determine the correction
needed to obtain a value between 3.0 and 4.0 mAs in the mAs
Meter after making a new exposure.

e. Repeat the above steps until a proper mAs value is obtained in the
mAs Meter.

4. Enter the mAs value read in the mAs Meter (step 3.) in the mAs Display
pressing the “mAs Increase” or “mAs Decrease” buttons.

5. Exit from Auto-calibration mode pressing the “Auto Calibration” button.


At this moment, the Generator will calculate and store the new value in
E03 Memory Location. Auto-calibration is deactivated and the process
is finished when the screen shows the GSM menu.

6. Press the “Manual Calibration” button. Select an available workstation


(WS) and press the “OK” button. Then, select the E03 Memory Location
and read its new value on the Calibration Display. Record this value in
the Data Book (refer to Appendix A, “Data Book,” at the end of this
Chapter).

7. Exit calibration mode.

3-55
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

For High mA stations (from 100 mA) (E05 Memory Location):

1. Enter in Auto-calibration mode by pressing the “Auto Calibration” button


on the GSM menu and select the E05 Memory Location. Confirm all the
following screens.

2. When Auto-calibration is activated, “Auto Calibration” button is lighted


and the mAs value is shown on the screen (7.75 mAs).

3. Perform the following steps:

a. Make an exposure (technique parameters are pre-programmed at


80 kV, 125 mA and 65 ms and they can be shown when pressing
the “Prep” button).

b. Read the mAs measured on the mAs Meter.

c. Read the Filament Current Number shown on the calibration


panel.
d. Increase or decrease the Filament Current Number (by pressing
the “Increase” or “Decrease” buttons) to determine the correction
needed to obtain a value between 7.00 and 8.50 mAs in the mAs
Meter after making a new exposure.

e. Repeat the above steps until a proper mAs value is obtained in the
mAs Meter.

4. Enter the mAs value read in the mAs Meter (step 3.) in the mAs Display
pressing the “mAs Increase” or “mAs Decrease” buttons.

5. Exit from Auto-calibration mode pressing the “Auto Calibration” button.


At this moment, the Generator will calculate and store the new value in
E05 Memory Location. Auto-calibration is deactivated and the process
is finished when the screen shows the GSM menu.

6. Press the “Manual Calibration” button. Select an available workstation


(WS) and press the “OK” button. Then, select the E05 Memory Location
and read its new value on the Calibration Display. Record this value in
the Data Book (refer to Appendix A, “Data Book,” at the end of this
Chapter).

7. Exit calibration mode.

8. After calibration of E03 and E05 Memory Locations:

G Switch the Generator power OFF.

G Disconnect the mAs Meter to the banana plug connections.

G Re-install the link between the banana plug connections on the HV


Transformer.

3-56
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.5 DIGITAL mA LOOP OPEN (X-RAY TUBE CALIBRATION)

To achieve the most accurate calibration, this procedure has to be


automatically performed by the Generator (Auto-calibration). Calibration
procedure will be manually performed by the field engineer only if
Auto-calibration is not possible.

Two different methods are described in this section: Auto-calibration and


Manual Calibration.

Digital mA Loop Open is calibrated by performing the following steps:

1. With the Generator power OFF, set:

G Dip switch 3000SW2-2 on the HT Controller Board in “Off” position


(enables Filament and Rotor Interlocks).

G Dip switch 3000SW2-4 on the HT Controller Board in “On” position


(Digital mA Loop Open / Filament Current Constant).

2. Turn the Generator ON.

3. Perform the Auto-calibration procedure as described in Section 6.5.1 for


each X-ray Tube in the system.

6.5.1 AUTO-CALIBRATION OF DIGITAL mA LOOP OPEN

Auto-calibration of the Filament Current Numbers is divided in two separated


procedures related to the mA stations configured for the Small or Large Focal
Spots.

It is recommended to start with the Small Focal Spot (first group) and continue
with the Large Focal Spot (second group).

EACH TIME THAT AUTO-CALIBRATION IS ACTIVATED IN A


mA STATION (OR IN “E01” MEMORY LOCATION), ALL THE
FILAMENT CURRENT NUMBERS OF THE SELECTED FOCAL
SPOT ARE AUTOMATICALLY SET TO “344”. SO A NEW
COMPLETE CALIBRATION OF THE FILAMENT CURRENT
NUMBERS FOR THIS FOCAL SPOT WILL BE REQUIRED.

3-57
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Auto-calibration starts with the minimum available mA station for the selected
Focal Spot at 50 kV and follows with the other combinations of mA stations for
the selected Focal Spot at 80 kV, 120 kV and 40 kV.

1. Enter in Auto-calibration mode by pressing the “Auto Calibration” button


on the GSM menu.

2. Select an available workstation (WS) of the X-ray Tube to be calibrated


and press the “OK” button. This workstation has to be one of the
previously configured as “Direct”.

3. Select the Small Focal Spot by pressing the “Small” button and then
press the “OK” button.

Confirm or leave the Auto-calibration by pressing the respective button


(“Yes” or “No”) on the new screen.

3-58
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Confirm or leave the Auto-calibration (second confirmation) by pressing


the respective button (“Confirm” or “Dismiss”) on the calibration screen.

5. Auto-calibration is activated when the “Auto Calibration” button is lighted


and the “Press Prep and Expose” message appears on the screen. At
this moment, the Generator has checked the mA stations available for
the Small Focal Spot.

3-59
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6. Check that the Heat Units capacity available for the X-ray Tube is 100%
or nearly.

BEFORE MAKING ANY EXPOSURE IN AUTO-CALIBRATION,


VERIFY THAT THE LINK BETWEEN THE BANANA PLUG
CONNECTIONS ON THE HV TRANSFORMER IS INSTALLED.

7. Keep fully pressed the Handswitch button to perform continuous


exposures.

Note . In Auto-calibration mode, all technique parameters are factory


pre-programmed and they can not be changed.

Auto-calibration can be paused momentarily releasing the


Handswitch button, whenever there is not an exposure in
process. Do not exit from Auto-calibration before the
procedure has been completed.

Note . Auto-calibration can be cancelled by pressing the “Auto


Calibration” button. A message on the screen informs that
Auto-calibration has been cancelled and after a while the screen
shows the GSM menu.

Note . The mA filament number will populate on the bottom of the screen
when that mA station is calibrated before moving to the next
station. This number can be updated for the filament current
numbers chart for the backup.

If during Auto-calibration process, any error indication is shown


momentarily (such as “Tube Overload”, etc.), it means that Generator
can not calibrate in this moment the selected kV / mA combination
(because anode overheated, space charge, Generator power limit, etc.).
In this case, the Generator will continue with Auto-calibration of the
following available kV / mA combinations for the selected Focal Spot. At
the end of the process it will try to calibrate or calculate the combinations
previously uncalibrated.

If the Heat Units capacity available is less than 40%, exposures are
inhibited momentarily and message “Tube too hot” is shown on the
screen. This message will disappear and exposures can be made again
when the X-ray Tube begins to cool and recovers the Heat Units capacity.

At this point, it is recommended to wait until the Heat Units capacity


available is close to the 80% before making any exposure.

3-60
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Generator tries to calibrate each kV / mA combination for ten (10)


attempts (maximum). If calibration is cancelled (after ten attempts),
message “Auto calibration failure” is shown on the screen until press the
“Auto Calibration” button to exit from Auto-calibration mode and go back
to the GSM menu. Calibration can be also cancelled due to space charge
during calibration of the lowest kV at the highest mA stations for the Focal
Spot selected, so message “Auto calibration failure” is also shown on the
screen until press the “Auto Calibration” button to exit from
Auto-calibration mode and go back to the GSM menu.

ONLY IF AUTO-CALIBRATION IS CANCELLED DUE TO


“AUTO CALIBRATION FAILURE”, CONTINUE THE
AUTO-CALIBRATION PROCEDURE FOR THE OTHER FOCAL
SPOT. CHECK AT THE END OF THE AUTO-CALIBRATION
PROCEDURE WHICH kV / mA COMBINATIONS HAVE NOT
BEEN AUTO-CALIBRATED FOR EACH FOCAL SPOT (THESE
COMBINATIONS HAVE THE FILAMENT CURRENT NUMBER
SET TO “344”). MANUALLY CALIBRATE THESE kV / mA
COMBINATIONS AS EXPLAIN IN SECTION 6.5.2.

When Auto-calibration is successfully performed, message “Auto


Calibration OK” is shown on the screen until press the “Auto Calibration”
button to exit from Auto-calibration mode and go back to the GSM menu.

8. Repeat the same procedure for the Large Focal Spot.

Enter in Auto-calibration mode by pressing the “Auto Calibration” button


on the GSM menu.

Select an available workstation (WS) of the X-ray Tube to be calibrated


and press the “OK” button. This workstation has to be one of the
previously configured as “Direct”.

Select the Large Focal Spot by pressing the “Large” button and then
press the “OK” button.

Confirm or leave the Auto-calibration by pressing the respective button


on each screen.

Auto-calibration is activated when the “Auto Calibration” button is lighted.


At this moment, the Generator has checked the mA stations available for
the Large Focal Spot.

Before starting the exposures, it is recommended to wait until the Heat


Units capacity available is close to the 80%.

3-61
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

9. After performing both procedures (for Small and Large Focal Spots),
enter in “Manual Calibration” mode and select each combination of the
available mA stations for each Focal Spot at the kV break points (40, 50,
80 and 120 kV). Press the “Read” button to read on the calibration panel
the new value of the Filament Current Number stored for each
combination and record the new values in the Data Book (refer to
Appendix A, “Data Book,” at the end of this Chapter).

Note that the highest mA station for Small Focal Spot may have numbers
larger than the lowest mA station for Large Focal Spot. This is normal.

10. Exit from calibration mode.

11. Turn the Generator power OFF and set Dip Switch 3000SW2-4 on the
HT Controller Board in “Off” position (Digital mA Loop Closed).

6.5.2 MANUAL CALIBRATION OF DIGITAL mA LOOP OPEN

This procedure describes the Manual calibration of all the Filament Current
Numbers. It has to be also used to calibrate the kV / mA combinations not
performed during Auto-calibration procedure because it has not been
completed. These combinations have the Filament Current Number set to
“344”, so only these combinations have to be manually calibrated as described
in this procedure. If Auto-calibration for one of the Focal Spots has been
successful, it is only required to perform the manual calibration of the mA station
do not calibrate for the other Focal Spot.

Manual Calibration is initiated at the 80 kV break point by entering the


appropriate Filament Current Number for the proper tube current at each
selectable mA. Calibration at the other kV break points (40, 50, 80 and 120 kV)
are obtained by adding or subtracting values as indicated in Table 3-8.

Although the suggested values (Table 3-8) could change depending on the
X-ray tube used, entering those values will approximate accurate calibration
without making excessive exposures.

In “Manual Calibration” mode, the Filament Current Numbers are shown on the
calibration panel by pressing the “Read” button after selecting the respective
kV / mA combination. The value can be changed by pressing the “Increase” or
“Decrease” buttons of the calibration panel and stored by pressing the “Store”
button.

Note that in calibration mode, only the mA stations and kV (at the break points)
can be selected.

3-62
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Table 3-8
mA Calibration Numbers Change

FILAMENT CURRENT NUMBERS AT kV BREAK POINT


mA
A STATION
40 50 80 120

10 A1+7 A1+6 A1 A1 --5

12.5 A2+7 A2+6 A2 A2 --5

16 A3+7 A3+6 A3 A3 --5

20 A4+7 A4+6 A4 A4 --5

25 A5+7 A5+6 A5 A5 --5

32 A6+7 A6+6 A6 A6 --5

40 A7+7 A7+6 A7 A7 --5

50 A8+7 A8+6 A8 A8 --5

65 A9+7 A9+6 A9 A9 --5

80 A10+7 A10+6 A10 A10 --5

100 A11+10 A11+8 A11 A11 --7

125 A12+10 A12+8 A12 A12 --7

160 A13+10 A13+8 A13 A13 --7

200 A14+10 A14+8 A14 A14 --7

250 A15+10 A15+8 A15 A15 --7

320 A16+14 A16+11 A16 A16 --9

400 A17+14 A17+11 A17 A17 --9

500 A18+14 A18+11 A18 A18 --9

650 A19+14 A19+11 A19 A19 --9

800 A20+14 A20+11 A20 A20 --9

Note.-- The mA station values depends on the Generator model. Some models do not contain all the mA stations listed above.

3-63
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. With the Generator power OFF, set:

G Dip switch 3000SW2-2 on the HT Controller Board in “Off” position


(enables Filament and Rotor Interlocks).

G Dip switch 3000SW2-4 on the HT Controller Board in “On” position


(Digital mA Loop Open / Filament Current Constant).

G Remove the link between the banana plug connections on the HV


Transformer. Connect the mAs Meter to the banana plug
connections to measure mA or mAs.

2. Turn the Generator ON and enter in calibration mode by pressing the


“Manual Calibration” button on the GSM menu.

3. Select an available workstation (WS) of the X-ray Tube to be calibrated


and press the “OK” button. This workstation has to be one of the
previously configured as “Direct”.

4. Check that the Heat Units capacity available for the X-ray Tube is 100%
or nearly.

5. Accordingly to X-ray Tube ratings or maximum Generator power, check


which kV / mA combinations in Table 3-8 are allowed.

If error “Tube Overload” is shown after selection of a kV / mA combination,


it means this combination is not allowed for the selected X-ray Tube.

In calibration mode, if Generator power is exceeded by a kV / mA


combination selection, error “E-16” is shown after “Preparation”. Reset
this error by pressing the “Reset Error” button.

Note which combinations in Table 3-8 can not be calibrated by making


exposures (combinations not allowed due to Tube rating, maximum
Generator power, space charge, etc.) and the Exposure Time assigned
to these combinations in calibration mode.

6. Enter in user mode inside calibration mode by pressing the “Go to User
Mode” button. Select the mA station and Exposure Time of each kV / mA
combination not allowed. Increase or decrease the kV value as required
to determine the kV value allowed nearest to the kV value of the
combination. Note the kV value allowed in the respective cell of
Table 3-8.

7. Exit from the “User Mode” screen by pressing the “Manual Calibration”
button. Select 80 kV and lowest mA station available (first combination
available). Enter a Filament Number of “500” for this combination.

3-64
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

8. Make an exposure. The mAs read on the mAs Meter must be the same
mAs displayed on the calibration screen with a tolerance of ± 0.1 mAs
(tolerance of the parameter and mAs Meter). If the mAs read is close to
zero, increase the filament number in big steps (e.g. increase values in
40). As the mAs read is close to the mAs displayed on the Console,
increase (or reduce) the filament number in smaller steps (e.g. increase
value in 30, 20, 10, ...).

If the mAs is low, increase the filament number. If the mA is high,


decrease the filament number. Press the “Store” button before making
a new exposure. Repeat until the mA station is calibrated.

Note . Press the “Store” button to store the new data (filament number)
before selecting the next kV or mA stations.

Calibration data (presently in memory) may or may not be


close to your requirements. If it is not close, the potential
exists to damage the X-ray tube (i.e. too much mA). Thus, as
you start the mA calibration procedure note how close or how
far off the mA break points are. If a large adjustment (more
than 40 points) is required at the low mA stations, make
estimated adjustments to the high mA stations before those
exposures are made.

9. Select the next mA station at 80 kV. Before making any exposure, enter
as filament number the value calibrated for the previous mA station
increased in 10.

If the mAs is low, increase the filament number. If the mA is high,


decrease the filament number. Press the “Store” button before making
a new exposure. Repeat until the mA station is calibrated.

10. Complete the calibration process for all mA stations at 80 kV as


described before. When select the first mA station for the Large Focal
Spot, enter as a filament number the value calibrated for the first mA
station for the Small Focal Spot. Note that the highest mA station for
Small Focal Spot may have numbers larger than the lowest mA station
for Large Focal Spot. This is normal.

Press the “Read” button to read on the calibration panel the new value
of the Filament Current Number stored for each mA station at 80 kV.
Record the new values in the Data Book (refer to Appendix A, “Data
Book,” at the end of this Chapter).

3-65
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Complete the calibration process for the remaining kV / mA combinations


using Table 3-8 as a guide. It is not necessary to make exposures to do
this. Compute the value for all the kV break points of each available mA
station although the Generator power can not reach all the kV / mA
combinations. Select the corresponding kV / mA combination and enter
the computed value.

12. Check calibration at all break points (making exposures) and correct any
calibration points as needed.

Note . If “Tube Overload” error is shown directly after selection of an


allowed combination (refer to step-5.), wait until the X-ray tube
anode cools down to permit the calibration of the mA station.

13. Recalculate the values of the non-allowed combinations in accordance


to the new values obtained by exposures. (Refer to Table 3-8).

14. Enter in user mode inside calibration mode by pressing the “Go to User
Mode” button. Select the mA station, Exposure Time and kV value noted
for each kV / mA combination not allowed (refer to step-6.). Check
calibration at these kV / mA combinations by making exposures. If
needed, enter in calibration mode and correct the Filament Current
Number of the respective kV / mA combination not allowed.

15. Exit from the “User Mode” screen by pressing the “Manual Calibration”
button. In calibration mode, select each combination of the available mA
stations at the kV break points (40, 50, 80 and 120 kV). Press the “Read”
button to read on the kV Display the new value of the Filament Current
Number stored for each combination. Record the new values in the Data
Book (refer to Appendix A, “Data Book,” at the end of this Chapter).

Note . Note that the highest mA station for Small Focal Spot may have
numbers larger than the lowest mA station for Large Focal Spot.
This is normal.

16. Exit calibration mode.

17. After calibration of Filament Current Numbers:

G Switch the Generator power OFF.

G Disconnect the mAs Meter to the banana plug connections.

G Re-install the link between the banana plug connections on the HV


Transformer.

G Set Dip Switch 3000SW2-4 on the HT Controller Board in


“Off” position (Digital mA Loop Closed).

3-66
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6 AEC CALIBRATION

This section provides the calibration steps to achieve proper AEC performance
for Proteus XR/f Radiographic System, in order to achieve the correct image at
the lowest possible dose.

The term “Optimum Gray Level” refers to some measurement criteria and may
be either:

 a measured exposure in μGray, or

 a system calculation called EI (Exposure Index), or

 a system calculation called S-Value.

This procedure will use the EI measurement method because of variations in


back-scatter characteristics, detector conversion factor, and detector
sensitivity. The EI value is calculated within a 4 cm x 4 cm ROI (Region of
Interest), and should be checked using a raw image. Raw images must be
acquired with the “S-Value” exam tag (in the QC group). An exposure of 1 μGray
will produce an equivalent of 102 to 110 EI.

AEC performance is controlled by a hardware gain potentiometer in the


Pre-Amplifier of the Ion Chamber and several software parameters stored in the
Generator (Extended Memory Locations).

The following table indicates the Extended Memory Locations related to AEC
Calibration.

FUNCTION MEMORY LOCATION

AEC-1 CALIBRATION E04

AEC-1 TRACKING E08

AEC-2 CALIBRATION E09

AEC-2 TRACKING E10

AEC DENSITY SCALE E12

Previous to AEC Calibration, identify and note which Memory


Locations are assigned / related to each Ion Chamber in the
system regarding to the configured Workstation / Devices.

3-67
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

AEC Calibration is carried out using a specified test setup and a Homogeneous
Aluminum Phantom (21 mm thickness, 99% purity) installed in the Collimator
guides, to produce a testing exposure time between 50 ms and 300 ms.

Illustration 3-9
Homogeneous Aluminum Phantom

The radiation signal detected by the sensors (also known as Ion Chambers) is
stepped up by an adjustable gain amplifier. Its output is passed to an AEC
control circuit with a comparator stage, which activates the AEC Stop command
when the signal exceeds the reference.

Therefore, the AEC Stop depends on the Signal (kV, mA, SID, Phantom, and
Gain Pot) compared to the Software Parameters.

Radiation Signal
Pre-Amp
Sensors AEC STOP
Gain Pot
L. C. R.
SW Parameters Reference
DAC

AEC Control Board

EXP ORDER

STRT DR

DAC Vout (Reference)


(calibration data setting)

Ion Chamber Output


EXP Start
AEC STOP
(AEC Control Board)
EXP Stop
AEC Function Waveform

3-68
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

The AEC calibration procedure is carried out in 3 parts:

 The 1st part is to adjust the Optimum Gray Level at 70 kV.

 The 2nd part is to adjust kV-Tracking for the same Optimum Gray Level
at 55 kV and 90 kV (¦ 20%).

 The 3rd part is to fine-tune and validate the AEC performance using the
imaging sub-system.

SAFETY PRECAUTIONS

This procedure produces ionizing radiation. Refer to the


product documentation for additional safety advice.

PREREQUISITES

 The X-Ray Generator and Imaging System are functional, i.e. exposures
can be taking through CS7 PC Workstation.

 The X-Ray Tube is calibrated and the System alignment meets


specifications.

Before starting with the AEC Calibration, it is necessary that


the Alignment of X-Ray Beam and the Alignment of Light Field
with X-Ray Field should be performed.

Note . If the X-Ray beam energy (kV, mA) is mis-calibrated or mis-aligned


(SID), the AEC accuracy will be affected.

 The Service Engineer has been trained on the Proteus XR/f and is
familiar with operating the CS7 and TechService applications.

3-69
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6.1 SETTINGS BEFORE THE AEC CALIBRATION

6.6.1.1 PREPARATION OF THE CS7 PC WORKSTATION

1. Verify that Exposure Index (EI) appears in the annotations on the “CS7
Utility” screen.

If EI value appears on the screen, skip to step 2.

If not, enable it in the “CS7 Utility” screen as follows:

1 2 3

“Utility” button is located


at the bottom of the Enter the password (kmmg5678) and press “OK”
“System Monitor” screen
4

Select “User Tool” tab in the upper left


corner of the “Service Tool” screen

Continues in next page

3-70
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Click “Screen Setting”


6

Click “Overlay Setting on Exposed Images”

Tick on “Text Annotation” and “Exposure Index” boxes, then


press “OK” --> “Save (YES)” --> “OK” --> “Close” --> “Close”

Note . If the customer does not want to see the Exposure Index data on
the images, deselect the Exposure Index overlay setting after the
AEC calibration. Otherwise it is recommended to leave it for further
Service checks and if Biomed engineers are familiar with EI
values.

3-71
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2. Enable “Exam Tag Copying” in the “CS7 Utility” screen as follows:

1 2 3

“Utility” button is located


at the bottom of the Enter the password (kmmg5678) and press “OK”
“System Monitor” screen
4

Select “Service Tool” tab in the upper


left corner of the “Service Tool” screen

5 Click on the “Center Icon”

Continues in next page

3-72
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6
Click “Setting”

Click “Exposed Image Copying / Exchan...”

Tick on “Exam Tag Copying” Box, then press


OK” --> “Save (YES)” --> “OK” --> “Close” --> “Close” --> “Close”

3-73
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. Verify that the S-Value exam tag (in the QA/QC group) is configured for
Table and Wall Stand exams.

If S-Value exam tag is configured, skip to step 4.

If not, modify it in the Edit Menu as follows:

1 2 3

“Utility” button is located


Enter the password (kmmg5678) and press “OK”
at the bottom of the
“System Monitor” screen
4

Select “User Tool” tab in the upper left


corner of the “Service Tool” screen

Click “Edit Exam Tag”

Continues in next page

3-74
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Select “Property List” tab and then find


the row “QC / S-Value” on the list

(1) Select ”Property List” tab.

6 (2) Filter using Free Word “QC”.

(3) In the S Value row, click the “Priority Modality” field.

(4) Open the droplist and select “Upright- >Table- >Cassette” ,


that means “Wall Stand --> Table --> Direct Detector”

Then press “OK” --> “OK” --> “Close” --> “Close”

3-75
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Exit the CS7 Application from the “Patient” screen as follows:

1 2 3

“Utility” button is located


at the bottom of the Enter the password (kmmg5678) and press “OK”
“System Monitor” screen
4

Press “Close”

Press “Shutdown” and confirm with “OK”

3-76
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6.1.2 PREPARATION OF THE X-RAY SYSTEM

1. From the CS7 Desktop, launch the TechService application on the CS7
PC Workstation as follows:

a. With the Generator power off, verify that:


-- the Generator is in Service Mode, that is, dip switch SW2-3 on
the ATP Console Board is in ON position.
-- Jumper JP3 on the AEC Control Board A3012-25 is in “A”
position.

b. Power ON the Proteus XR/f.


If prompted log on as:
user: maintenance
password: kmmg5678

c. Select:

1 2 3

“Utility” button is located


at the bottom of the Enter the password (kmmg5678) and press “OK”
“System Monitor” screen 4

Press “Close”

Press “Shutdown” and confirm with “OK”

3-77
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

d. Click “Windows Desktop” and launch the “TechService” shortcut.

2. In the TechService application, select the “Manual Calibration” mode and


enter in the corresponding Extended Memory Locations the default
values related to the AEC Calibration.

a. Click on “Manual Calibration” and select a Workstation (e.g. WS1),


click on “OK”.

Click on “Manual Calibration” and select a Workstation (e.g. WS1), click on “OK”.

3-78
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

b. Select the Extended Memory Location to be checked (it is shown


on the “mA Display”) and display the value stored (it is shown on
the “Calibration Value” Display after pressing the “Read” button).
If needed, select the correct value by pressing the “Increase” and
“Decrease” buttons and store it by pressing the “Store” button.

Calibration Value Panel

MEMORY LOCATION FUNCTION DEFAULT VALUE

E04 AEC-1 calibration Table Ion Chamber 70

E08 AEC-1 tracking (kVp Compensation) Table Ion Chamber 4

E09 AEC-2 calibration Wall Stand Ion Chamber 70

E10 AEC-2 tracking (kVp Compensation) Wall Stand Ion Chamber 4

E12 AEC Density Scale All Ion Chambers 25

c. Select 100 mA or any mA value to exit the Extended Memories


selection. Exit from “Manual Calibration” by pressing the “Exit”
button.

d. Close the TechService application.

3-79
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6.1.3 SETUP THE X-RAY SYSTEM TEST CONDITIONS (TABLE OR WALL STAND)

1. Remove the Detector from the Cassette Tray.

2. Adjust the SID to 150 cm (130 cm for ST Table).

3. Remove the Anti-scatter Grid.

4. Install the Aluminum Phantom in the Collimator rails.

6.6.2 PROCEDURE OF AEC CALIBRATION

6.6.2.1 TUBE WARM-UP

1. Reboot the entire system from Sys Loader. It activates the custom
changes made before, re-establishes the Gen/PC communication and
restart the applications.

2. Login as maintenance.

3. Open a test patient and select exam tag S-Value with the following
parameters: 70 kV, 100 mAs, 200 mA, 500 ms.

4. Fully close the Collimator blades.

5. Select the Detector (1417 or 1717 as applicable).

6. Locate the “Copy Exam Tag” button at the bottom of the “Exam List”.
Click it repeatedly to create 2 duplicates of the settings above.

7. Perform a Tube Warm-up procedure taking 3 exposures, 15 seconds


apart. Be sure that no one will be exposed.

3-80
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6.2.2 CALIBRATE THE AEC PRE-AMP MASTER GAIN POTENTIOMETER

1. Install a Detector in the Tray in the portrait mode and start a new Exam
for S-Value.

2. Adjust / confirm the AEC Calibration acquisition settings are correct:

(1) Exam tag (Wall Stand v. Table)

(2) Detector selection

(3) Bucky (Wall Stand v. Table)

(4) Technique: 70 kV, 50 mA, 400 ms

(5) AEC: Center / Medium Film-Screen combination

1 4
2

3. On the Collimator, fully open the blades.

4. Locate the “Copy Exam Tag” button at the bottom of the “Exam List”. Click
it repeatedly to create 6 duplicates of the settings above. Create more
copies if needed.

3-81
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Make an exposure on the newly copied exams and note the EI value and
no error messages are displayed. The target Exposure Index (EI) for
optimized image quality depends on the Digital Detector type.

Target Exposure Index (EI)


AeroDR Detector Type
for optimized image quality

HQ 305 ¦ 6

S 415 ¦ 8

G If the measured EI value is reasonably close to the target EI,


small adjustments in the AEC Calibration parameter (E04 or E09)
may be done instead of adjusting the Gain Potentiometer of the Ion
Chamber Pre-Amplifier (no need to remove covers).

In this case, select the Extended Memory Location to be checked


(E04 or E09) and correct value by pressing the “Increase” and
“Decrease” buttons and store it by pressing the “Store” button
(refer to Section 6.6.1.2).

MEMORY LOCATION FUNCTION DEFAULT VALUE

E04 AEC-1 calibration Table Ion Chamber 70

E09 AEC-2 calibration Wall Stand Ion Chamber 70

Note . The Image Gray Level / Dose increases / decreases when the
calibration number is increased / decreased.

The AEC accuracy is better in all the useful range as the AEC
calibration number is closer to 70.

Record the values for the Memory Locations E04 and E09 in the
Data Book (refer to Appendix A, “Data Book,” at the end of this
Chapter).

Go to next Section 6.6.2.3.

3-82
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

G If the measured EI value is incorrect:

-- Access the Ion Chamber Pre-Amplifier Box.

-- Tweak the tiny Gain Potentiometer.


(CCW decreases the dose)

-- Make an exposure and read the EI value.

-- Repeat these steps as needed until within the


specifications.

In this case, go to next step-6. (Table) or step-7. (Wall Stand).

6. Access to the Gain Potentiometer of the IC Pre-Amplifier in the


Table (XR/f ET / ST)

a. Doubletap on the Tabletop movement pedal and push the Tabletop


to the front as much as possible to gain access to the Tracking Bar.

b. Cut the tie wrap next to the Tracking Bar connector.

c. Unscrew the cable tie clamp.

d. Disconnect the Tracking Bar connector.

Cable Tie Clamp

Tracking Bar Cable Connector

3-83
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

e. Unscrew the Tracking Bar Plate (4+2 screws).

f. Carefully leave the Tracking Bar with its Plate resting beside the
Detector Assembly.

Tracking Bar with its Plate

g. Unscrew the two (2) torx upper screws of the Pre-amplifier plate.

3-84
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

h. Perform the Gain Potentiometer adjustment with a trimmer


adjustment tool (tiny screwdriver).

7. Access to the Gain Potentiometer of the IC Pre-Amplifier in the Wall


Stand (XR/f ET / ST)

Note . The following procedure shows a left hand load Wall Stand. For
right hand load Wall Stand, proceed in the same way adapting
processes from the right hand load point of view.

a. Position the Receptor Assembly at a comfortable position to


remove the Receptor Assembly Cover.

b. Take out the grid.

c. Unscrew the eight (8) screws that attach the Receptor Assembly
Cover at the back of the assembly and carefully take out the Cover
towards the handle of the Detector Tray.

3-85
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

d. Unscrew the two (2) torx upper screws of the Pre-Amplifier Plate.

e. Perform the Gain Potentiometer adjustment with a trimmer


adjustment tool (tiny screwdriver).

3-86
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6.2.3 CALIBRATE THE KV TRACKING COMPENSATION

To maintain constant Image Gray Level regardless of the kV at which a Detector


exposed, the Generator provides kV compensation. (Refer to Illustration 3-10).

Illustration 3-10
AEC kV Tracking Curve

250

200
DAC OUTPUT in byte (8 Bit) (U3-2 of AEC Control Board)

150

100

70 10
8

50 6
4
2
0
--2
--4
0
--6
--8
--10

--50
40 50 60 70 80 90 100 110 120 130 140 150
kV

3-87
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Make an exposure at 55 kV and other exposure at 90 kV.

2. Check and note the EI value at 55 kV and 90 kV, compared to 70 kV.


Variation shall be less than ¦ 15% with pre-set +4 value.

TABLE WALL STAND

55 kV 70 kV 90 kV 55 kV 70 kV 90 kV

EI
(Exposure Index)

Fine-tune
Value in Memory Location E08 (Table) = __ Value in Memory Location E10 (Wall Stand) = __
kV Tracking

MEMORY LOCATION FUNCTION DEFAULT VALUE

E08 AEC-1 tracking (kVp Compensation) Table Ion Chamber 4 = (pre-set +4)

E10 AEC-2 tracking (kVp Compensation) Wall Stand Ion Chamber 4 = (pre-set +4)

Note . Values for AEC Tracking range is from --10 to +10.


(Refer to Illustration 3-10).

AEC-kV
TRACKING +10 +9 +8 +7 +6 +5 +4 +3 +2 +1 0 --1 --2 --3 --4 --5 --6 --7 --8 --9 --10
CURVE

VALUE TO ENTER
IN THE MEMORY 10 9 8 7 6 5 4 3 2 1 0 255 254 253 252 251 250 249 248 247 246
LOCATION

3. If the variation percentage is incorrect, change the value to +3.

In this case, select the Extended Memory Location to be modified (E08


or E10) and correct value by pressing the “Increase” and “Decrease”
buttons and store it by pressing the “Store” button (refer to Section
6.6.1.2).

a. Repeat the exposures at 55 kV and 90 kV and compare the EI


value again.

3-88
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

b. If the variation value is not ±15% of the Image Gray Level,


calculate the new value for the AEC Tracking number in each
Memory Location as follows:

-- If the Image Gray Level at 55 kV has to be decreased and


the Image Gray Level at 90 kV has to be increased then increase
the Tracking value in one.

-- If the Image Gray Level at 55 kV has to be increased and


the Image Gray Level at 90 kV has to be decreased then
decrease the Tracking value in one.

Note . A Tracking value of 10 will have no effect on AEC kV


compensation.

c. Repeat these steps as needed until within the specifications.

4. Repeat the above steps for all the Ion Chambers installed with the
Generator.

5. Record the values for the Memory Locations E08 and E10 in the Data
Book (refer to Appendix A, “Data Book,” at the end of this Chapter).

Note . Once the calibration procedures of AEC Pre-Amplifier Gain


Potentiometer and kV Tracking Compensation are finished,
remove from the Collimator guides the Phantom of Aluminum.

3-89
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6.3 VERIFY AEC PERFORMANCE

6.6.3.1 KV TRACKING COMPENSATION TEST

Note . This checking is carried out using homogeneous Phantom of


PMMA (Polymethylmethacrylate) with a surface of 43 x 43 cm and
different thickness (cm) according to the exposure test to be
performed.

TESTING CONDITIONS

1 Testing conditions on Table (XR/f ET / ST):

2 ¡ Align Center-On-Center:
Lateral-Horizontal Arm, Table Center Detent.
Long-Laser to Cassette Tray Handle.
3
© Remove Aluminium Phantom and DAP.
Set Field Size = full open the Collimator blades.
4
¢ SID = 150 cm (130 cm for ST Table)
5
£ PMMA Phantom = 43 cm x 43 cm x thickness
according to the exposure test to be performed.

¤ Flat Panel Detector (FPD) installed.


Anti-scatter Grid removed.

3-90
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Testing conditions on Wall Stand (XR/f ET / ST):

¡ Align Center-On-Center:
Lateral-Vertical Arm, Wall Stand Receptor Center.
5
4 Long-Laser to Cassette Tray Handle.
3 2
© Remove Aluminium Phantom and DAP.
Set Field Size = full open the Collimator blades.
1
¢ SID = 150 cm

£ PMMA Phantom= 43 cm x 43 cm x thickness


6
according to the exposure test to be performed.

¤ Flat Panel Detector (FPD) installed.


Anti-scatter Grid removed.

¥ Cardboard Boxes or Polystyrene at a sufficient


height to support the PPMA Phantom stack
centered with the Tube X-Ray Beam, and with the
Tube located at a height above the Tabletop
of the Table.

Example of PMMA positioning on Wall Stand

3-91
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Install a Detector in the Tray in the portrait mode and start a new Exam
for S-Value.

2. Select the following parameters and settings:

(1) Exam tag (Wall Stand v. Table)

(2) Detector selection

(3) Bucky (Wall Stand v. Table)

(4) Technique:

-- kV, mA and PMMA thickness as indicated in the next table


(refer to step-4.), mA values and PMMA thickness values
are given for indication in order to have an exposure time
more than 50 ms and less than 300 ms.

-- 1 s back-up Time.

(5) AEC: Center / Medium Film-Screen combination

1 4
2

3. Locate the “Copy Exam Tag” button at the bottom of the “Exam List”. Click
it repeatedly to create 5 duplicates of the settings above. Create more
copies if needed.

3-92
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Make the exposures as indicated in the next table and note the EI value.
Check that no error messages are displayed. The target Exposure
Index (EI) for optimized image quality depends on the Digital Detector
type.

Target Exposure Index (EI)


AeroDR Detector Type
for optimized image quality

HQ 305 ¦ 6

S 415 ¦ 8

EI value
PMMA Thickness
kV mA
A Ti
Time ((ms))
(cm)
Table Wall Stand

55 25 50 < x < 300 8

70 10 50 < x < 300 10

80 10 50 < x < 300 15

100 10 50 < x < 300 20

125 10 50 < x < 300 25

130 10 50 < x < 300 28

5. Verify the performance accuracy, ± 20%.


Note the values in the table above.

EI max -- EI min
x 100% =
EI max

6. If the test fails, check the following items:

G Requested system and exposure conditions described early in the


procedure.

G Tube calibration.

G Selected Film Screen combination.

G Tube warm-up.

3-93
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6.3.2 ION CHAMBER CELL BALANCE TEST

Note . This test is carried out using homogeneous PMMA (Polymethyl


methacrylate), 43 x 43 x 15 cm, so that all 3 cells receive the same
amount of scatter radiation. As an alternative to PMMA, the Al
phantom in the collimator rails may be used (not recommended).
This procedure is used only to confirm that the balance of the 3
cells is correctly adjusted from factory.

TESTING CONDITIONS

1 Testing conditions on Table (XR/f ET / ST):

2 ¡ Align Center-On-Center:
Lateral-Horizontal Arm, Table Center Detent.
Long-Laser to Cassette Tray Handle.
3
© Remove Aluminium Phantom and DAP.
Set Field Size = base of PMMA Phantom stack.
4
¢ SID = Grid under test (100 / 150 cm)
5
£ PMMA Phantom = 43 x 43 x 15 cm

¤ Flat Panel Detector (FPD) installed.


Anti-scatter Grid in or out according to testing
procedure.

Testing conditions on Wall Stand (XR/f ET / ST):

¡ Align Center-On-Center:
Lateral-Vertical Arm, Wall Stand Receptor Center.
5
4 Long-Laser to Cassette Tray Handle.
3 2
© Remove Aluminium Phantom and DAP.
Set Field Size = base of PMMA Phantom stack.
1
¢ SID = Grid under test (100 / 150 or 180 cm)

£ PMMA Phantom = 43 x 43 x 15 cm
6
¤ Flat Panel Detector (FPD) installed.
Anti-scatter Grid in or out according to testing
procedure.

¥ Cardboard Boxes or Polystyrene at a sufficient


height to support the PPMA Phantom stack
centered with the Tube X-Ray Beam, and with the
Tube located at a height above the Tabletop
of the Table.

3-94
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Install a Detector in the Tray in the portrait mode and start a new Exam
for S-Value.

2. Select the following parameters and settings:

(1) Exam tag (Wall Stand v. Table)

(2) Detector selection

(3) Bucky (Wall Stand v. Table)

(4) Technique: 70 kV, 50 mA, 1 s back-up Time.

(5) AEC: Left, Center or Right according to the exposure test


to be performed / Medium Film-Screen combination.

1 4
2

3. Locate the “Copy Exam Tag” button at the bottom of the “Exam List”. Click
it repeatedly to create as many duplicates of the settings above as
exposures will be done in next steps. Create more copies if needed.

4. Make a text exposure and note the exposure time displayed on the
application, it should be a time between 50 ms and 300 ms. If not,
increase or decrease mA value (to reduce the time increase the mA
value, and viceversa) and make the exposure again.

Note the final mA value to be used in the next exposures.

3-95
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Make the following series of exposures and record the exposure time.

EXPOSURES NO GRID EXPOSURES WITH GRID


TABLE at 70 kV and mA at 70 kV and mA
(SID 100 cm))
LEFT CENTER RIGHT LEFT CENTER RIGHT

Exposure #1 tL1 = tC1 = tR1 = tL1 = tC1 = tR1 =

Exposure #2 tL2 = tC2 = tR2 = tL2 = tC2 = tR2 =

Exposure #3 tL3 = tC3 = tR3 = tL3 = tC3 = tR3 =

Average value of the Exposure Time tL = tC = tR = tL = tC = tR =

Left vs center right vs center left vs center right vs center

% Difference

EXPOSURES NO GRID EXPOSURES WITH GRID


WALL STAND at 70 kV and mA at 70 kV and mA
(SID 100 cm))
LEFT CENTER RIGHT LEFT CENTER RIGHT

Exposure #1 tL1 = tC1 = tR1 = tL1 = tC1 = tR1 =

Exposure #2 tL2 = tC2 = tR2 = tL2 = tC2 = tR2 =

Exposure #3 tL3 = tC3 = tR3 = tL3 = tC3 = tR3 =

Average value of the Exposure Time tL = tC = tR = tL = tC = tR =

Left vs center right vs center left vs center right vs center

% Difference

3-96
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

EXPOSURES NO GRID EXPOSURES WITH GRID


WALL STAND at 70 kV and mA at 70 kV and mA
(SID 150 cm or 180 cm))
LEFT CENTER RIGHT LEFT CENTER RIGHT

Exposure #1 tL1 = tC1 = tR1 = tL1 = tC1 = tR1 =

Exposure #2 tL2 = tC2 = tR2 = tL2 = tC2 = tR2 =

Exposure #3 tL3 = tC3 = tR3 = tL3 = tC3 = tR3 =

Average value of the Exposure Time tL = tC = tR = tL = tC = tR =

Left vs center right vs center left vs center right vs center

% Difference

6. Calculate the average of the 3 exposure times for each cell. Verify
performance accuracy by comparing the left and right averages against
the center average. Tolerance ± ¦20%.

tL -- tC
% of difference between “Center” and “Left” Fields = x 100 = %
tC

tR -- tC
% of difference between “Center” and “Right” Fields = x 100 = %
tC

3-97
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.6.4 AEC DENSITY SCALE

AEC is calibrated with “Density 0” selected (Normal Image Gray level). The
Density can be increased or decreased in several steps.

The variation percentage of the density scale is factory set at 25%. This
variation can be set according to the customer preferences by changing the
value stored in E12 Memory Location. This value applies to both tubes.

The range of the scale is “from --4 to +4”.

The following table shows some examples for the variation percentage of the
density scale with reference to the value stored in E12 Memory Location.

DENSITY SCALE FROM --4 TO +4 (OPTIONAL) . VARIATION OVER NORMAL OPTICAL DENSITY (N)
E12
VALUE
--4 --3 --2 --1 0 +1 +2 +3 +4

5 N x 0.82 N x 0.86 N x 0.91 N x 0.95 N N x 1.05 N x 1.10 N x 1.16 N x 1.22

10 N x 0.68 N x 0.75 N x 0.83 N x 0.91 N N x 1.10 N x 1.21 N x 1.33 N x 1.46

25 N x 0.41 N x 0.51 N x 0.64 N x 0.80 N N x 1.25 N x 1.56 N x 1.95 N x 2.44

NOTE: With scale from “--4 to +4” the useful range for the value stored in E12 Memory Location is from 1 to 25 .

Note . In most cases the factory default of 25 for E12 Value is the best
setting.

Record the value for E12 Memory Location in the Data Book (refer to Appendix
A, “Data Book,” at the end of this Chapter).

3-98
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6.7 FINAL CHECKS AFTER GENERATOR CALIBRATION

Verify that all Configuration and Calibration data have been properly stored in
memory.

1. Enter in calibration mode and check that the values noted for the
“Filament Current Numbers” and “Extended Memory Locations” tables
of the Data Book (refer to Appendix A, “Data Book,” at the end of this
Chapter) are the same that the values displayed and stored in memory.
Press the “Read” button to read the stored values.

2. Exit from calibration mode and Service mode.

3. Turn the Generator OFF and verify position of dip switches on the HT
Controller Board are:

G Dip switch 3000SW2-2 in “Off” position (enables Filament and


Rotor Interlocks).

G Dip switch 3000SW2-4 in “Off” position (Digital mA Loop Closed).

4. Set the Test dip switch 3024SW2-3 on the ATP Console CPU Board in
“Off” position to place the Generator in normal operating mode.

5. Check the HV Transformer and verify the mAs meter is removed and the
banana plug jumper is installed.

3-99
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

3-100
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

APPENDIX A DATA BOOK

This Data Book is the register of the Configuration and Calibration data of the
Generator.

Note . Enter the data with a pencil in order to modify them later due to
future changes.

Note . Verify that the “Configuration Control Sheet” and “Final Test
Results” pages from factory have been included with the
equipment.

If the HT Controller Board or the ATP Console CPU Board are


replaced, check specially that Extended Memory data have
not been lost or modified with the Board change. Compare
Extended Memory data displayed on the Console with the
values noted in this document.

Also, make some exposures using different techniques and


Focal Spot and check that mA stations are calibrated
correctly, if not perform Calibration procedures.

3-101
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Table A-1
3024SW1 - ATP Console CPU Board

3024SW1 POSITION OPEN (OFF) CLOSED (ON)

Table A-2
3024SW2 - ATP Console CPU Board

3024SW2 POSITION OPEN (OFF) CLOSED (ON)

Table A-3
3024SW3 and 3024SW4 - ATP Console CPU Board

Note . DIP switches 3024SW3 and 3024SW4 are not used for
configuration but all their switches must be set in the OFF position.

Table A-4
3000SW2 - HT Controller Board

3000SW2 POSITION OPEN (OFF) CLOSED (ON)

3-102
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Table A-5
Basic Configuration in Generator Boards

GENERATOR BOARDS JUMPERS POSITION

JP1, JP2, JP3, JP5 and JP6 in “2” (soldered)


HT CONTROLLER BOARD
(A3000--80/81)
JP4 in “1” (soldered)

FILAMENT CONTROL BOARD


W1 in “A” (installed)
(A3004--10)

W1 in “2--3” (installed)

INTERFACE CONTROL BOARD


W2 in “1--2” (installed)
(A3009--11)

W3 to W8 removed

JP1, JP2 and JP3 in “B” (soldered)

JP4 in “A” (installed)

JP5 in “B” (installed)

JP6 in “A” (soldered)

Connector J8 configured for RS232 so:


JP9, JP10 and JP11 in “A” (soldered)
ATP CONSOLE CPU BOARD JP7, JP8, JP21 and JP22 do not matter
((A3024--221)) jumpers position (in “A” soldered)

JP12 removed

JP13 (not installed)

JP14 (soldered)

JP15, JP16, JP17 and JP18 removed

JP19 in “A” (soldered)

JP1 in “C” (installed)

AEC CONTROL BOARD


JP2 and JP3 in “A” (installed)
(A3012--25)

JP4 in “B” (installed)

JP1, JP2, JP3, JP4, JP5, JP6, JP7, JP8, JP13,


AEC ADAPTATION BOARD JP14, JP15 and JP16 in “B” (installed)
(A3263 03)
(A3263--03)
JP9, JP10, JP11, JP12 in “A” (installed)

GCA BOARD
W1 in “ON”
(A3548--01)

3-103
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Table A-6
Workstations

WORKSTATION WS1 WS2 WS3 WS4 WS5 WS6 WS7 WS8

TUBE

WM

AEC IC

WS LOCK

kV TRACKING

RAD/RF

Table A-7
Extended Memory Locations

MEMORY LOCATION FUNCTION VALUE

E01 TUBE-1 -- RAD filament stand-by (Autocalibrated. Not field changeable)

E02 TUBE-1 -- RAD tube type

E03 Low Digital mA Loop Closed (from 10 mA to 80 mA)

E04 AEC-1 calibration

E05 High Digital mA Loop Closed (from 100 mA)

E06 kV Loop

E07 Maximum kW (Factory set. Only field changeable to lower value)

E08 AEC-1 tracking

E09 AEC-2 calibration

E10 AEC-2 tracking

E12 AEC Density Scale

E13 TUBE-1 -- Exposure Time adjustment - Delay

E15 TUBE-1 -- Exposure Time adjustment - Ceq kV

Note.-- Memory Locations not specified are not used.

3-104
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Table A-8
Rotor Acceleration Time Configuration

3000SW2-2
OPERATION MODE
OPEN (OFF) CLOSED (ON)

Rotor Speed Low Speed High Speed

TUBE-1 3000SW2-7 3000SW2-8


ROTOR ACCELERATION TIME
AND FILAMENT SETTING TIME OPEN (OFF) CLOSED (ON) OPEN (OFF) CLOSED (ON)

_____ seconds

Table A-9
DRAC Control Configuration

3243SW1 3243SW2 3243SW3 3243SW4

1 1 1 1

2 2 2 2

3 3 3 3

4 4 4 4

5 5 5 5

6 6 6 6

7 7 7 7

8 8 8 8

3-105
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Table A-10
mA Calibration Numbers

TUBE-1

FILAMENT CURRENT NUMBERS AT kVp BREAK POINT


mA
A STATION
40 50 80 120

10

12.5

16

20

25

32

40

50

65

80

100

125

160

200

250

320

400

500

650

800

Note.-- The mA station values depend on the Generator model. Some models do not contain all the mA stations listed above.

3-106
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

CHAPTER 4
TROUBLESHOOTING

4-1
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

4-2
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 1 INTRODUCTION

The Generator contains many self-diagnostic routines which greatly facilitate


troubleshooting. Self-diagnostic functions require a proper performance of all
microprocessors. Each microprocessor contains LEDs that indicate correct
operation.

As a general rule, the first step in any troubleshooting procedure is to verify


correct Power Supply Voltages and perform a visual inspection of all Boards and
Cable connections.

Upon arrival at the X-ray Room, the Service Engineer should carry out the
following operations:

 If the Generator cannot be Powered up, run the troubleshooting for Error
Code E01.

 If the Generator can be Powered up:

G Check which Error Code is displayed on the Console and run the
troubleshooting routines for the last Error Code displayed.

G If there is no error code displayed, try to reproduce the failure in


the conditions reported by the Operator. The Error Code displayed
might be different from the one reported. In all cases, run the
troubleshooting routines for the last Error Code displayed.

4-3
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Whenever the HV Transformer, X-ray Tube, ATP Console CPU


Board, HT Controller Board, or Filament Control Board is
replaced in the Generator, the respective Configuration and
Calibration procedures must be performed.

If the HT Controller Board or the ATP Console CPU Board are


replaced, check that the Extended Memory data have not
been lost or modified with the Board change. Compare
Extended Memory data displayed on the Console with the
values noted in the Data Book.

Revision of Boards should be checked, the new Board must


be compatible with the replaced one, that is, the same o
higher revision than the replaced Board. The EEPROM
information has to match the Generator options.

If the HT Controller Board is replaced in the Power Module,


transfer U3-EEPROM from the old Board to the new Board.
U3-EEPROM contains calibration data.

If U3-EEPROM is not transferred, a complete Calibration must


be performed.

Make various exposures, using different techniques and


Focal Spots, to verify that mA stations are calibrated
correctly. If they are not, perform Calibration procedures.

Update and record in the Data Book any new data entered in
the Extended Memory Locations.

4-4
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1.1 TOOLS AND TEST EQUIPMENT

The following hand tools and products are required for the Troubleshooting:

 Standard service engineers tool kit including Allen and Torx key sets, and
a reversible ratchet with socket set.

 Silicone Insulating Grease (proofing compound) (included in the X-ray


Tube package).

 HV Oil (included in the HV Cables package).

 Alcohol cleaning agent.

The following test equipment is required for some Configuration and Calibration
procedures carried out during Troubleshooting:

 Calculator.

 Anti-static Kit.

 Digital Multimeter.

 Non-invasive kVp Meter.

 Digital mAs Meter.

 Oscilloscopy.

 Only for AEC purposes:

G Homogeneous Phantom of Aluminium with a purity of not less that


99% and thickness of 21 mm for the Collimator Filter Holder
(Part no. 2281087-2).

G Homogeneous Phantoms of PMMA (Polymethylmethacrylate)


with a surface of 43 x 43 cm and different thickness (cm) to
achieve all the different thickness used in the AEC verification
tests: 8, 10, 15, 20, 25 and 28 cm.

G (Optional) Dosimeter, Unfors RaySafe Xi (or equivalent).

G A Trimmer for adjustment of the Ion Chamber Pre-Amplifier (AEC


tiny potentiometer) and Adhesive Tape.

4-5
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1.2 GENERAL CAUTIONS

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES (LEDS) ON THE CHARGE-DISCHARGE MONITOR
BOARDS ARE OFF (APPROXIMATELY THREE (3) MINUTES
AFTER THE UNIT IS TURNED OFF).

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY


HEATSINK OF THE CIRCUIT BOARDS EVEN WITH THE
GENERATOR TURNED OFF. PREVIOUS TO DISASSEMBLE
ANY BOARD, REMOVE ALL CONNECTORS PLUGGED TO IT.

DO NOT SUPPLY THE MAIN POWER FROM THE ELECTRICAL


ROOM CABINET (MAIN DISCONNECT) OR TURN ON THE
SYSTEM UNTIL SPECIFICALLY INSTRUCTED IN THIS
DOCUMENT.

ALWAYS HAVE THE “IPM DRIVER BOARD” CONNECTED IN


THE GENERATOR PREVIOUS TO ACTIVATING MAINS
POWER. IF THE “IPM DRIVER BOARD” IS NOT CONNECTED,
IRREVERSIBLE DAMAGE WILL OCCUR TO THE IGBTS.

THIS GENERATOR IS PERMANENTLY CONNECTED TO THE


POWER LINE AND POWERED ON UNLESS THE SAFETY
SWITCH INSTALLED IN THE ROOM ELECTRICAL CABINET
IS OFF. WHEN THE GENERATOR IS POWERED ON, THE
NEON LAMP (GREEN) LOCATED ON THE TRANSFORMER
6T2 (GENERATOR CABINET) IS ON.

INTERNAL PARTS OF THE GENERATOR (ALL FUSES, LINE


CONTACTOR (6K5), INPUT TRANSFORMER (6T2), ON/OFF
RELAY (3K3) AND LF-RAC MODULE) RECEIVE
CONTINUOUS POWER THROUGH THE POWER LINE
ALTHOUGH THE CONTROL CONSOLE IS OFF. BE SURE
THAT THE SAFETY SWITCH IS OFF BEFORE HANDLING ANY
INTERNAL PART OF THE EQUIPMENT.

4-6
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 2 GENERAL PROCEDURES

2.1 LOW DC VOLTAGE POWER SUPPLY TEST

The Generator operates from a Low DC Voltage Power Supply located in the
Front Panel (MOD. 3) of the Generator Cabinet. (Refer to Illustration 4-1).

Turn the Generator ON and with a Digital Multimeter measure between:

 P2-3 (+) and P2-4 on the HT Controller Board. Check that the voltage
at this point is +5 (±0.2 VDC). If required, adjust voltage with the
+5 VDC Adjustment Potentiometer on the Power Supply Board.

 P2-2 (+) and P2-4 on the HT Controller Board. Check that the voltage
at this point is +12 (±0.1 VDC). If required, adjust voltage with the
+12 VDC Adjustment Potentiometer on the Power Supply Board.

 P2-1 (--) and P2-4 on the HT Controller Board. Check that the voltage
at this point is --12 (±0.1 VDC). If required, adjust voltage with the
--12 VDC Adjustment Potentiometer on the Power Supply Board.

Illustration 4-1
Power Supply and HT Controller Board in the Front Panel

HT CONTROLLER BOARD

+12V adjust --12V adjust +5V adjust

POWER SUPPLY

4-7
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

The ATP Console CPU Board operates from a 12 VDC Unregulated Supply
located in the Generator Cabinet that supplies to a 5 VDC switching regulator
located in the ATP Console CPU Board.

Turn the generator ON and with a Digital Multimeter measure between:

 J1-2 or J1-8 (12V UNR) with J1-6 or J1-1 (GND) on the ATP Console CPU
Board; likewise, TP11 (GND UNR) with TP9 (12V UNR) on the ATP
Console CPU Board. Check that the voltage at this point
is +12 (±1.5 VDC).

If the measured voltage is not +12 (±1.5 VDC), check the 12 VDC
Unregulated Supply of the generator, that is, between terminals 11 and
12 of the Input Transformer 6T2, Fuse 3F7 (3A, 250 V), Contactor 3K3,
Rectifier 3BR4, and Capacitor 3C7. (Refer to schematic 543020xx).

 TP1 GND and TP3 +5 V on the ATP Console CPU Board. Check that the
voltage at this point is +5 (±0.2 VDC). If not, replace the ATP Console
CPU Board.

Note . No adjustments are required for both voltage measurements.

Illustration 4-2
ATP Console CPU Board in the Back Panel

ATP CONSOLE CPU BOARD

TP1, TP3, TP9, TP11

4-8
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.2 MICROPROCESSORS AND GENERAL OPERATION

The following LEDs indicate the proper operation of each Microprocessor in the
Generator:

 The LED DS1 located on the HT Controller Board blinks fast during
power up, then slows to a steady blink of about two (2) per second,
indicating that the Microprocessor U5 is operating normally.

 The LED DS2 located on the ATP Console CPU Board normally blinks
at the same rate as LED DS1 on the HT Controller Board, indicating that
communication between both Boards is correct.

Also observe the following LEDs to facilitate general troubleshooting:

 When LED DS1 located on the Interface Control Board is ON (lit), it


indicates that the Generator is ON.

 The LED DS1 located on the ATP Console CPU Board is normally ON
(lit), indicating that the Watch-Dog Timer of the Console is operating and
insuring the correct timing of data communications with the HT Controller
Board.

2.3 REPLACEMENT OF MEMORY IN ATP CONSOLE CPU BOARD AND/OR


HT CONTROLLER BOARD

Before starting up the system, you have to set ON the switch A3024SW2--3 at
the ATP Console CPU Board to enable Service Mode. A few seconds after
pressing the ON button on the Console, you will see E10 (this is shown because
the EPROM U24 has been replaced).

Reset the error indication by pressing the respective button on the Console and
keep it pressed until Error 10 disappears.

After this, the normal start up will take place.

It is recommended to set this switch back to OFF once the installation has been
completed, so the operator will not have access to Configuration or Calibration
Modes (Service Modes).

4-9
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

4-10
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 3 ERROR CODES

ERROR CODE LIST


ERROR CODE DESCRIPTION

“ -- -- -- -- -- -- ” System failure.

Failure in power up routine.


E01
No communication link between HT Controller Board and ATP Console CPU Board.

Failure in power up routine. RAD Generator configured as R&F Generator.


E02
No communication link between ATP Console CPU Board and Fluoro CPU Board.

E03 Failure in power up routine. Workstations are not configured.

E04 “Prep” signal received without Console order.

E06 “Prep” and/or “Exposure” orders activated during power-up routine.

E08 Wrong data for X-ray Tube-1.

E09 Generator overload. Arcing or IGBT fault.

EEPROM corrupted or not initialized on ATP Console CPU Board or on HT Controller Board.
E10
Wrong data calibration.

E11 No voltage in the Main Storage Capacitors of the Generator (Inverter Module).

E12 No mA during exposure or mA out of tolerance. Wrong filament current.

E13 No kVp during exposure, kVp out of tolerance (±33%) or insufficient time parameters during AEC operation.

E14 Exposure signal without X-ray Exposure Console command.

E15 Filament Driver Circuit Open, No current on Filament. Wrong selection of Focal Spot detected during “Prep”.

E16 Invalid value of : kVp, mA or kW.

E17 Communication error among ATP Console, CPU Board and HT Controller Board.

E18 Rotor error or Rotor running without order (only with Low Speed Starter).

E19 mA detected without “Exp” command.

E20 kVp detected without “Exp” command.

E21 Wrong Tube-1 selection.

E23 Calibration data not stored

E24 Bucky Movement Signal has not been detected or DR Device is not ready for exposure.

E27 Failure in EPROM U24 of the ATP Console CPU Board. Bad checksum.

E33 No communication link between the Generator and the Operator Console or PC Unit.

E34 Technique error.

Door Open.
E35 For Generators with HV Transformer equipped with a Security Pressure Switch, the HV Transformer
surpasses the maximum internal pressure capacity.

E36 Safety Thermostat Open. Overheating.

E37 Tube Overload.

4-11
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE LIST


ERROR CODE DESCRIPTION

E41 Dosimeter failure. Communication failure between Tube-1 Dosimeter and Generator.

E42 Dosimeter failure. Autotest error on Tube-1 Dosimeter.

E43 Dosimeter failure. Tube-1 DAP Ion Chamber status check error.

E48 Collimator Error.

E50 Interrupted Exposure.

E51 LV-DRAC: Checksum failure or Microcontroller U17 corrupted in Control DRAC Board.

E52 LV-DRAC: Microcontroller RAM failure.

E53 LV--DRAC: Insufficient DC BUS voltage at low level voltage (220 VAC).

E54 LV--DRAC: Insufficient DC BUS voltage at high level voltage (400 / 480 VAC).

E55 LV--DRAC: Excessive DC BUS voltage at 480 VAC.

E58 LV--DRAC: Excessive current in the main winding during acceleration up to 3300 RPM

E59 LV--DRAC: Excessive current in the auxiliary winding during acceleration up to 3300 RPM

E60 LV--DRAC: Insufficient current in the auxiliary winding during acceleration up to 3300 RPM

E61 LV--DRAC: Insufficient current in the main winding during acceleration up to 3300 RPM

E62 LV--DRAC: Excessive current in the main winding during acceleration up to 10000 RPM

E63 LV--DRAC: Excessive current in the auxiliary winding during acceleration up to 10000 RPM

E64 LV--DRAC: Insufficient current in the auxiliary winding during acceleration up to 10000 RPM

E65 LV--DRAC: Insufficient current in the main winding during acceleration up to 10000 RPM

E66 LV--DRAC: Excessive current in the main winding running at 3300 RPM

E67 LV--DRAC: Excessive current in the auxiliary winding running at 3300 RPM

E68 LV--DRAC: Insufficient current in the auxiliary winding running at 3300 RPM

E69 LV--DRAC: Insufficient current in the main winding running at 3300 RPM

E70 LV--DRAC: Excessive current in the main winding running at 10000 RPM

E71 LV--DRAC: Excessive current in the auxiliary winding running at 10000 RPM

E72 LV--DRAC: Insufficient current in the auxiliary winding running at 10000 RPM

E73 LV--DRAC: Insufficient current in the main winding running at 10000 RPM

E74 LV--DRAC: Excessive current in the main winding braking at 3300 RPM

E75 LV--DRAC: Excessive current in the auxiliary winding braking at 3300 RPM

E76 LV--DRAC: Insufficient current in the auxiliary winding braking at 3300 RPM

E77 LV--DRAC: Insufficient current in the main winding braking at 3300 RPM

E78 LV--DRAC: Excessive current in the main winding braking at 10000 RPM

E79 LV--DRAC: Excessive current in the auxiliary winding braking at 10000 RPM

E80 LV--DRAC: Insufficient current in the auxiliary winding braking at 10000 RPM

4-12
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE LIST


ERROR CODE DESCRIPTION

E81 LV--DRAC: Insufficient current in the main winding braking at 10000 RPM

E82 LV-DRAC: Wrong X-ray Tube selection.

E83 LV--DRAC: Excessive current in DC Brake.

E87 LV--DRAC: Insufficient current in common wire during acceleration up to 3300 RPM

E88 LV--DRAC: Insufficient current in common wire running at 3300 RPM.

E89 LV--DRAC: Insufficient current in common wire during acceleration up to 10000 RPM.

E90 LV--DRAC: Insufficient current in common wire running at 10000 RPM.

E91 LV--DRAC: Incorrect signal measure in IPRINC (CH2).

E92 LV--DRAC: Incorrect signal measure in IAUX (CH3).

E93 LV--DRAC: Incorrect signal measure in ICOM (CH4).

E95 Rapid Termination

E98 DIP Switch 3024SW2-3 on ATP Console Board set for Configuration and Calibration Mode Active.

4-13
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : “-- -- -- -- -- --” -- SYSTEM FAILURE

DESCRIPTION : System failure.


ERROR TYPE : Fatal error. Generator must be switched off.
APPLICABLE TO : All Generators
APPEARS WHEN : This indication may appear at any time together with another Error Code on the Console.
INFORMATION / SYMPTOM : This Error Code requires the Generator to be turned OFF/ON before it can be fixed.

ACTIONS
1. Turn the Generator OFF, wait a few seconds, and turn it ON.

4-14
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E01

DESCRIPTION : Failure in power up routine.


No communication link between HT Controller Board and ATP Console CPU Board.
ERROR TYPE : Fatal error. Generator can not continue with power up.
APPLICABLE TO : All Generators
APPEARS WHEN : Only during initialization phase. If it appears during normal functioning of equipment, it means
that a problem has caused a power off in the unit.
INFORMATION / SYMPTOM : This Error Code requires that the Generator be turned OFF/ON before it can be fixed.

POSSIBLE CAUSES
Dipswitch 3000SW2--1 is in “ON” position at the HT Controller Board.
It has not been possible to establish a correct communication link between the ATP Console CPU
Board and the HT Controller Board during power ON.
The communication link between the ATP Console Board and the HT Controller Board is not reliable.
The Microprocessor U5 on the HT Controller Board has not started-up.
The Generator has been affected by a loss of the main power.
Faulty DC power supply.

ACTIONS

1. With the Generator switched OFF, check that Dipswitch 3000SW2--1 is in “OFF” position at the
HT Controller Board.

This switch is set in “ON” position only for programming the Rotor Acceleration Time, Rad
Filament Setting Time, Fluoro Rotor and Filament Hold-over Time, therefore it changes the
functions of Switches SW2-2 and SW2-4 to SW2-8. For normal operation it must set in “OFF”
position.

2. Turn OFF/ON the Generator to verify that the initialization procedure has been established
correctly. Check that connections of the ATP Console CPU Board and HT Controller Board are
secure.

3. If the error appears just after EPROM U24 on the ATP Console Board (A3024-xx) or
Microprocessor U5 on the HT Controller Board (A3000-xx) have been replaced or updated,
then check the software version compatibility. If necessary, replace or update again .

4. Visually check the performance of DS1 on the HT Controller Board during start-up.

G If DS1 remains illuminated (not blinking), replace the HT Controller Board (A3000-xx).

G If DS1 blinks, go to step-6.

G If DS1 remains OFF, go to step-5.

4-15
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. The DC supply may be faulty either at the source or at the Boards themselves.

G On the HT Controller Board, check for DC at TP3 (+12V), TP4 (-12V) and TP2 (+5V) with
GND connection at either TP1, TP9 & TP10.

G On the ATP Console Board, measure between TP9 (+12V UNREG) with TP11 (GND
UNREG); TP10 (+12V ISO), TP3 (+5V), TP7 (+12V) and TP8 (-12V) with TP1 (GND).

-- If DC supply is not present, there might be a faulty fuse: shutdown the Generator and
test Fuse 3F9 on the Generator with an Multimeter (replacing if necessary). Fuse 3F9
is located below the Power Supply 3PS1 in the Generator Front Panel.

-- If DC supply is present (or error persists with correct DC), proceed to the next step.

6. Check continuity between J3 terminals 2, 3, 5 and 6 on the ATP Console CPU Board with
P1 terminals 4, 11, 5 and 10 on the HT Controller Board respectively. Connector 6J3 terminals
2, 3, 5 and 6 on the Generator should also be checked. Repair or replace if necessary.

If there is continuity in these connections, proceed to the next step.

7. Verify with an oscilloscope the signal presence (pulses) for C-HT CLK (J3-5) and C-HT DAT
(J3-2) from the ATP Console Board to the HT Controller Board (P1-5 and P1-4 respectively).

G If any signal is not present, replace the ATP Console CPU Board.

G If the signals are present, proceed to the next step.

8. Verify with an oscilloscope the signal presence (pulses) for HT-C CLK (P1-10) and HT-C DAT
(P1-11) from the HT Controller Board to the ATP Console CPU Board (J3-6 and J3-3
respectively). If any signal is not present, replace the HT Controller Board. If the problem
persists, replace Microprocessor U5 on the HT Controller Board.

4-16
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E02

DESCRIPTION : Failure in power up routine. RAD Generator configured as R&F Generator.


No communication link between ATP Console CPU Board and Fluoro CPU Board.
ERROR TYPE : Fatal error. Generator can not continue with power up.
APPLICABLE TO : R&F Generators or Generators configured by mistake as R&F.
APPEARS WHEN : Only during initialization phase. If it appears during normal functioning of equipment, it means
that a problem has caused a power--off in the unit.
INFORMATION / SYMPTOM : This Error Code requires that the Generator be turned OFF/ON before it can be fixed.
An incorrect or corrupt communication link occurred between the ATP Console and the Fluoro CPU
Boards during power ON and the Generator is not able to make exposures.

POSSIBLE CAUSES
A RAD Generator has been configured by mistake as a R&F Generator. See Configuration Section.
The communication link between ATP Console CPU Board and Fluoro CPU Board is not reliable.
The Fluoro CPU Board does not work properly.
The Microprocessor U9 on the Fluoro CPU Board did not start-up.
An incorrect communication between the ATP Console and Fluoro CPU Board has been
established during power ON.
After changing the ATP Console CPU Board or U18 / U23. The error is removed by entering and
then exiting from Workstation configuration.

ACTIONS
1. Turn OFF/ON the Generator to verify that the initialization procedure has been established
correctly. Check that connections of the ATP Console CPU Board and Fluoro CPU Board are
secure.

2. If the error appears just after EPROM U24 on the ATP Console Board (A3024-xx) or
Microprocessor U9 on the Fluoro CPU Board (A3213-xx) have been replaced or updated, then
check the software version compatibility. If necessary, replace or update again .

3. Visually check the performance of DS1 on the Fluoro CPU Board during startup.

G If DS1 remains illuminated (not blinking), replace Microcontroller U9 on the Fluoro CPU
Board.

G If DS1 blinks, go to step-6.

G If DS1 remains OFF, go to step-5.

4. The DC supply may be faulty either at the source or at the Boards themselves. On the Fluoro
CPU Board, check for DC at TP5 (+12V), TP3 (+5V) with GND connection at TP4. If DC supply
is present (or error persists with correct DC), proceed to the next step.

5. Check continuity between J4 terminals 1, 7, 3 and 4 on the ATP Console CPU Board with J4
terminals 1, 4, 5 and 7 on the Fluoro CPU Board respectively. Repair or replace if necessary.

If there is continuity in these connections, proceed to the next step.

4-17
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6. Verify signal presence for C-FL CLK (J4-7) and C-FL DAT (J4-1) from the ATP Console CPU
Board to the Fluoro CPU Board (J4-4 & J4-1 respectively).

G If any signal is not present, replace the ATP Console CPU Board.

G If the signals are present, proceed to the next step.

7. Verify signal presence for FL-C CLK (J4-7) and FL-C DAT (J4-5) from the Fluoro CPU Board
to the ATP Console CPU Board (J4-4 and J4-3 respectively). If any signal is not present,
replace the Fluoro CPU Board . If the problem persists, replace Microcontroller U9 on the
Fluoro CPU Board.

4-18
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E03

DESCRIPTION : Failure in power up routine. Workstations are not configured.


ERROR TYPE : Fatal error. Generator can not continue with power up.
APPLICABLE TO : All Generators.
APPEARS WHEN : Only during initialization phase. If it appears during normal functioning of equipment, it means
that a problem has caused a power off in the unit.
INFORMATION / SYMPTOM : This Error Code requires that the Generator be turned OFF/ON before it can be fixed.
It has not been possible to establish a correct communication link between the ATP Console CPU
Board and the HT Controller Board during power ON and the Generator is not able to make
Exposures.

POSSIBLE CAUSES
All workstations have been configured as Tube = 0.
The EEPROM (U18) on the ATP Console CPU Board is defective.

ACTIONS
1. Configure workstations according to the Service Manual.

2. If not fixed after the previous steps, replace the ATP Console CPU Board and configure
workstations again.

4-19
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E04

DESCRIPTION : “Prep” signal received without Console order.


ERROR TYPE : Fatal error. Generator re-start automatically once error is solved.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment once initialization phase is over.
INFORMATION / SYMPTOM : This Error Code requires that the Generator be turned OFF/ON before it can be fixed.
An incorrect or corrupt communication link has been established between the ATP Console CPU
Board and the HT Controller Board.

POSSIBLE CAUSES
The Generator receives the “Prep” signal without a Console command.

ACTIONS
1. Check continuity between P1-3 on HT Controller Board and J3-4 on ATP Console CPU Board.
Check with special care connector 6J3 of the Generator (6J3-4 of the Generator connects with
P2-3 on the HT Control Board).

2. If all is correct, check signal from P1-3 to U5 on HT Controller Board.

3. If the signal is OK, replace the HT Controller Board .

4-20
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E06

DESCRIPTION : “Prep” and/or “Exposure” orders activated during power-up routine.


ERROR TYPE : Informative.
APPLICABLE TO : All Generators
APPEARS WHEN : At any moment once initialization phase is over.
INFORMATION / SYMPTOM : This Error Code requires that the Generator be turned OFF/ON before it can be fixed.

POSSIBLE CAUSES
The Generator has detected “Prep” or “Exposure” signals during initialization.

ACTIONS
1. In Generators equipped with “Prep” and “Exposure” keys on the Console, check that no object
has accidentally activated the function. Also check for possible damages on keys. Check for
damages on flat cables to connector J9 on the ATP Console CPU Board.

2. In Generators with external Handswitch, check contacts, cable and connectors.

3. If the error persists, check for continuity between TS1-37 and J2-17 on the ATP Console CPU
Board and TS1-36 and J2-4 also on the ATP Console CPU Board.

4. If the error persists, replace the ATP Console CPU Board.

4-21
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E08

DESCRIPTION : Wrong data for X-ray Tube-1.


ERROR TYPE : Fatal error. Generator can not continue with power up.
APPLICABLE TO : All Generators
APPEARS WHEN : Only during initialization phase.
INFORMATION / SYMPTOM : This Error Code requires that the Generator be turned OFF/ON before it can be fixed.
An incorrect or corrupt communication link has been established between the ATP Console CPU
Board and the HT Controller Board during power ON.

POSSIBLE CAUSES
Data on Extended Memory Location “E02” are larger than the maximum allowed.

ACTIONS
1. Replace the HT Controller Board.

4-22
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E09 -- GENERATOR OVERLOAD

DESCRIPTION : Generator overload. Arcing or IGBT fault.


ERROR TYPE : Informative.
APPLICABLE TO : All Generators
APPEARS WHEN : During the exposure and In Stand-by,
INFORMATION / SYMPTOM : This error may appear at the Console as “E09” Error Code or as a “Generator Overload”
indication.
During exposure an over current on the IGBTs of the HV Inverter Module has been detected. This
may be produced by an arc or malfunction on the HV Circuitry.
In stand-by, the Console is continuously displaying “E09” or “Generator Overload” due to a defective
or overheated IGBT Module.

POSSIBLE CAUSES
Symptom-1:
G Defective X-ray Tube.
G Defective HV Transformer or HV Cable.
G Defective Inverter Module.
G Defective HT Controller Board.
Symptom-2:
G Extremely high Duty Cycle on Rad and Fluoro operation.

ACTIONS

A. PRELIMINARY

1. Select minimum kVp , minimum mA, and 80 ms (for example 40 kVp, 10mA, and 80 ms). Make
preparation and check that anode rotates in the X-ray Tube. If the anode is not rotating
correctly, check the starter and the Stator connections.

2. In case the anode is rotating correctly, make an exposure:


G If “09” or “Generator Overload” appears, follow the procedure in paragraph B, “Inverter
Module Test.”
G If not, follow step 3.

3. Increment kVp in 10 kVp steps, select the same mA and time. Make an exposure:
G If “E09” or “Generator Overload” appears, or the exposure is cut before 80 ms, then follow
Procedure C, “HV Transformer Test.”
G If not, keep on increasing the kVp in 10 kVp steps (60, 70, 80, 90, 100, 110, 120, and 125
kVp for 125 kVp HV Transformers; and 130, 140, 150 kVp for 150 kVp HV Transformers)
making Exposures at each kVp selected.
-- If “E09” or “Generator Overload” appears or the exposure is cut before 80 ms at any
kVp selected, follow Procedure C, “HV Transformer Test.”
-- If “E09” or “Generator Overload” has not appeared at the maximum kVp or the
exposure was not cut before 80 ms, it means that the HV Transformer and the HV
Cables are OK. Then follow Procedure F, “RANDOM “E09” or “Generator Overload”.”
In this case high mA causes “E09” or “Generator Overload”.

4-23
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

B. INVERTER MODULE TEST

Illustration 4-3 shows a Flowchart for Procedure “B.”

Illustration 4-3
Flowchart for B: Inverter Module Test

TURN THE GENERATOR OFF


WAIT UNTIL THE INVERTER IS DISCHARGED
MAKE A JUMPER DC--BUS+ AND DC--BUS--
MEASURE AND CHECK

NO
V¶ 0,3V REPLACE IGBT

YES

REMOVE JUMPER DC--BUS


REMOVE SHIELDED CABLES THAT CONNECT
INVERTER TO THE HV TRANSFORMER
SET 3000SW2-2=ON
TURN ON GENERATOR & MAKE EXPOSURE

“E09” OR NO
RE--CONNECT H V TRANSFORMER AND INVERTER
EXPOSITION CUT?
PERFORM PROCEDURE C, “HV TRANSFORMER TEST”

YES

REPLACE INVERTER

4-24
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Turn the Generator and Electrical Room Cabinet (Main Disconnect) OFF.

2. Remove the cover from the Generator Cabinet.

3. Wait for the DC Bus of the Inverter to be fully discharged. When it is fully discharged the LEDs
on the Charge-Discharge Monitor Board will be completely turned off.

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF THE HIGH


VOLTAGE INVERTER DO NOT CONTAIN ANY RESIDUAL CHARGE. WAIT
UNTIL THE LIGHT EMITTING DIODES ON THE CHARGE- DISCHARGE
MONITOR BOARDS ARE OFF, APPROXIMATELY THREE (3) MINUTES
AFTER THE UNIT IS TURNED OFF.

4. When LEDs are off make a jumper between DC-BUS+ and DC--BUS-- (use a wire AWG 22
(0.5 mm2) or higher). Make sure that there is less than 10 VDC across the BUS. (Refer to
Illustration 4-4 to see where to make the jumper.)

5. Measure with a Multimeter in Diode (or ohms) between C2E1 (positive polarity) and E2 or C1
(negative polarity) in both IGBTs (refer to Illustration 4-4 for more details). Repeat the measure
with different polarity between C2E1 (negative polarity) and E2 or C1 (positive polarity).

Voltage should be around 0.3 V (or the resistance must be a high impedance) for the IGBT to
be OK. Normally when an IGBT is broken the voltage is = 0 volts (or the resistance is zero [0]
Ohms) or very close.

4-25
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Illustration 4-4
Jumper and Measurement Points

Jumper between DC-BUS+ and DC-BUS-- E2 C2E1

3--PHASE LINE POWERED GENERATOR

Jumper between DC-BUS+ and DC-BUS-- E2 C2E1

1--PHASE LINE POWERED GENERATOR

4-26
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6. Repeat the measurements done in step 5 for the other IGBT.

DON’T FORGET TO REMOVE THE JUMPER ACROSS THE DC BUS AFTER


ALL MEASUREMENTS ARE MADE, OTHERWISE THE INVERTER WILL
SUFFER SERIOUS DAMAGE.

7. If any of the IGBTs are short-circuited, replace the IGBT.

8. If both IGBTs are OK, remove both shielded cables that connect the Inverter to the HV
Transformer: P1, P3, and SHLD (P2). Isolate the three (3) wires completely from each other
and from the metal sheet or ground.

MAKE SURE THE WIRES ARE PERFECTLY ISOLATED AND THAT NO


SHORT-CIRCUIT IS MADE, OTHERWISE SERIOUS DAMAGES COULD
RESULT.

9. Set DIP switch 3000SW2--2 in the ON position at the HT Controller Board.

10. Turn the Room Electrical Cabinet (Mains Disconnect) and Generator ON.

11. Make an exposure:


G If “E09” or “Generator Overload” does not appear and the exposure has not been cut
before 80 ms, re--connect both shielded cables to the HV Transformer (P1, P2, and P3).
Follow Procedure C, “HV Transformer Test.”
G If “E09” or “Generator Overload” appears or the exposure has been cut before 80 ms,
replace the whole Inverter Module.

Note . At the end of an Exposure and right after releasing the Handswitch, error “E13” is
shown on the Console. This is normal. Reset and continue.

12. Turn the Electrical Room Cabinet (Main Disconnect) and generator OFF. Wait three (3) minutes
for the Main Storage Capacitors to discharge.

13. Set DIP switch 3000SW2-2 in the OFF position at the HT Controller Board. Re-connect both
shielded cables that connect the Inverter to the HV Transformer P1, P3, and SHLD (P2).

4-27
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

C. HV TRANSFORMER TEST

1. Make these connections with the Oscilloscope:


G CH.1 on + mA (TP14) on the HT Controller Board.
G CH.2 on - mA (TP13) on the HT Controller Board.
G Base Time in 10 ms per division and 1 V per division

2. Turn the mains and Generator ON. Set DIP switch 3000SW2-4 to the ON position at the HV
Controller Board.

3. Select 40 kVp, 10mA, and 80 ms. Make an exposure and check that both waveforms are almost
symmetric (a difference of 12% is normal).
G If it is OK, follow the procedure in step 4.
G If it is not OK, check:
-- that in the mA Test Point of the HV Transformer the jumper is correctly placed and
secure.
-- that connector J1 is correctly placed and secure in the HV Transformer.
-- continuity between J1-D and P4-7, J1-E and P4-6, J1-B and P4-2, and J1-C and
P4-1. Check that they are correctly connected and secure.
-- if after these actions the waveform is not OK, replace the HV Transformer.

4. Turn the Generator and Electrical Room Cabinet (Main Disconnect) OFF. Wait three (3)
minutes for the Main Storage Capacitors to discharge.

5. Remove the HV Cables from the HV Transformer (anode and cathode) and fill the HV
Receptacles with oil.

6. On the HT Controller Board, make a jumper FIL (TP8) and + 5 V (TP2).

7. Turn the Electrical Room Cabinet (Main Disconnect) and Generator ON.

8. Make these connections with the Oscilloscope:


G CH.1 on + kV (TP11) on the HT Controller Board.
G CH.2 on --kV (TP12) on the HT Controller Board.
G Base time in 10 msec. per division and 2 V per division.

9. Select 40 kVp, 10 mA, and 80 ms. Make an Exposure and check that both waveforms are
symmetric. A difference of 12% is normal.

Note . If error code “E15” appears after pressing PREP, reset the Error and select the
lowest mA for Large Focus. Repeat the exposure.

G If it is OK, follow procedure in step 10.


G If it is not OK, follow Procedure D, “Asymmetry on the kVp Loop.”

4-28
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

10. Make exposures from 40 kVp to 150 kVp with the same Exposure Time and check that all kVp
waveforms are symmetric and the values are similar according the table below.

Select TP11 & TP12 on the HT Controller

50 kVp 2.1 V

70 kVp 2.9 V

90 kVp 3.7 V

110 kVp 4.5 V

130 kVp 5.3 V

150 kVp 6.1 V

11. If the waveforms are not symmetric (within 12% at any point), follow Procedure D, “Asymmetry
on the kVp Loop.”

12. If everything is OK, it means that HV Transformer is OK and the problem could be in the X--ray
Tube or in the HV Cables. In order to determine exactly what part must be replaced (the X--ray
Tube or the HV Cables) it is necessary to replace one of both components and perform the
procedure again. In most of the cases the part to be replaced usually is the X--ray Tube. To know
when the Tube begins to arc follow Procedure E, “X--ray Tube Test.”

13. Turn the Generator and Electrical Room Cabinet (Main Disconnect) OFF and wait three (3)
minutes for the Main Storage Capacitors to discharge. Set DIP switch 3000SW2--4 in the OFF
position at the HV Controller Board and remove the jumper between FIL (TP8) and + 5 V (TP2).

D. ASYMMETRY ON THE kVp LOOP

This procedure determines what part should be replaced due to asymmetry on the kVp Loop:
the High Voltage Transformer or the HT Controller Board.

1. Turn the generator and Electrical Room Cabinet (Main Disconnect) OFF and wait three (3)
minutes for the Main Storage Capacitors to discharge.

2. Set DIP Switch 3000SW2--2 to the “ON” position at the HV Controller Board.

3. Turn the Room Electrical Cabinet (Main Disconnect) and generator ON.

4. Make these connections with the Oscilloscope:


G CH.1 on +kV (TP11) on the HT Controller Board.
G CH.2 on kVp (TP16) on the HT Controller Board.

5. Apply +12VDC (P2--2) to +kV (P4--2) for 500 ms. This operation may cause Error “20” if the
voltage is applied for more than 500 ms; in this case turn the generator OFF and ON to reset
the error.

4-29
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

6. Check the waveform’s result:

G If it is OK, follow the procedure in step 7.


G If it is not OK, replace the HT Controller Board. It is recommended to perform Procedure
C, “HV Transformer Test,” again.

7. Make these connections with the Oscilloscope:


G CH.1 on --kV (TP12) on HT the Controller Board.
G CH.2 on kVp (TP16) on HT the Controller Board.

8. Apply --12VDC (P2--1) to --kV (P4--1) for 500 ms. This operation may cause Error “20” if the
voltage is applied for more than 500 ms. In this case turn the generator OFF and ON to reset
the error.

9. Check the waveform’s result:

G If it is OK, replace the HV Transformer.


G If it is not OK, replace the HT Controller Board. It is recommended to perform Procedure
C, “HV Transformer Test,” again.

10. Turn the Generator and Electrical Room Cabinet (Main Disconnect) OFF and wait three (3)
minutes for the Main Storage Capacitors to discharge.

11. Set DIP Switch 3000SW2--2 in the OFF position at the HV Controller Board.

4-30
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

E. X-RAY TUBE TEST

After the performance of the above referred test everything is found OK, the Service (Field) Engineer
may want to know the actual status of the Tube. Perform the following procedure in order to
determine the point in which the X-ray Tube begins to arc.

It is strongly recommended to replace the Tube as soon as possible to


prevent potential damage to the Generator.

1. Select 40kVp, 10mA, and 100 ms.

2. Make an exposure.

3. Increment kVp in 10 kVp steps, select same mA and time. Make an exposure.
G If “E09” or “Generator Overload” appears follow Procedure C, “HV Transformer Test,” if
it has not been performed before. If Procedure C has been already performed follow the
procedure in step 4.
G If “E09” or “Generator Overload” does not appear, keep on incrementing the kVp in steps
of 10 kVp (up to the maximum kVp), making exposures at each kVp selected.
-- If “E09” or “Generator Overload” appears at any kVp selected it means that the Tube
has dielectric problems above the selected kVp.
-- If “E09” or “Generator Overload“ does not appear (up to maximum kVp), it means that
arcing may be due to mA or kW. Follow the procedure in step 4.

4. Select 40 kVp and 10 mA. Increment the mA one station and make an Exposure. Keep on
incrementing the mA station (making exposures) until “E09” or “Generator Overload” appears.
This will give an idea of the maximum mA allowed by the Tube without arcing. If “E09” or
“Generator Overload” does not appear follow step 5.

5. If the Tube still does not arc, the problem is related to kW = KV*mA. Make selections on the
Console at 100 ms incrementing kV and mA. A point will be reached in which “E09” or
“Generator Overload” will appear. This will give an idea on the approximate value of kVp and
mA that can be handled by the X--ray Tube. Also, this value may change when the Tube heats
up.

F. RANDOM “E09” OR “GENERATOR OVERLOAD”

If everything is OK and random “E09” or “Generator Overload” appears, check:


1. That the signal IGBT FAULT on Pin 3 of P5 on the HT Controller Board is not low (logic 0) in
stand--by and during the exposure. If there is noise, check for a loose connection between Pin
3 of P5 on HT Controller Board and Pin 4 of J2 in both IPM Driver Boards

2. If IGBT FAULT is active during an exposure, try to isolate when it occurs. It may be due to noise
coming from any device outside the Generator. Or it may occur when selecting a high power
Exposure and the voltage of the main line goes down more than 10%.

4-31
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E10

DESCRIPTION : EEPROM corrupted or not initialized on ATP Console CPU Board or on HT Controller Board.
Wrong data calibration.
ERROR TYPE : Fatal Error during power up (when EEPROM U3 on HT Controller Board is corrupted or not
initialized).

Informative error during power up (when EEPROM U18 on ATP Console CPU Board is
corrupted or not initialized or when EPROM U24 on ATP Console CPU Board has been
changed). In both cases, “E10” appears together with “E34” Error Code or “Technique Error”
indication.
APPLICABLE TO : All Generators.
APPEARS WHEN : Only during initialization phase.
INFORMATION / SYMPTOM : Generator does not continue with start up.

POSSIBLE CAUSES
EEPROM U3 on the HT Controller Board corrupted or not initialized.
EEPROM U18 on the ATP Console CPU Board corrupted or not initialized.
EPROM U24 on the ATP Console CPU Board has been changed.

ACTIONS
1. If EPROM U24 in the ATP Console CPU Board has been replaced, reset the Error Code to
acknowledge that the NVRAM has been initialized.
2. If the error does not reset, turn the Generator OFF and set Dip-switch A3024SW2-3 in the ON
position to allow Service Mode.
3. Turn the Generator ON and enter into Workstations Configuration and follow the checks as
indicated in the Service Manual (do not forget to exit from Configuration mode to store the
Workstation Configuration).
4. If the problem persists and “E10” appears together with the “E34” Error Code or “Technique
Error” indication, replace the ATP Console CPU Board.
5. If the problem persists and only “E10” appears on the Console, replace the HT Controller
Board.

4-32
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E11

DESCRIPTION : No voltage in the Main Storage Capacitors of the Generator (Inverter Module).
ERROR TYPE : Informative. “Prep” is not allowed.
APPLICABLE TO : All Line Powered Generators
APPEARS WHEN : Only during initialization phase or when pressing “PREP”.
INFORMATION / SYMPTOM : --

POSSIBLE CAUSES
No voltage in the Main Storage Capacitors (5C1, 5C2, 5C3, 5C4) of the Generator (Inverter Module).
Defective Charge/Discharge Monitor Board.
Precharge 6K5 contactor located inside the Generator Cabinet is not energized.
Main line fuses F3 and/or F4 are blown.
“--CHRG” signal of Pin 7 on connector P1 on the HT Controller Board is not present.
Cables disconnected accidentally or damaged connectors.

ACTIONS
A. IF LED DS1 IN CHARGE/DISCHARGE MONITOR IS ILLUMINATED
Check the voltage at the Main Storage Capacitors (5C1, 5C2, 5C3, 5C4) (Inverter Module)
measuring the voltage between DC BUS (+) and DC BUS (--) as indicated in the illustration below.

CHARGE/DIS-
CHARGE
MONITOR BOARD

DC BUS (+)
1--PHASE LINE
POWERED GENERATOR

DC BUS (--)

CHARGE/DIS-
CHARGE
MONITOR BOARDS

3--PHASE LINE
POWERED GENERATOR

DC BUS (+) DC BUS (--)

4-33
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Voltage at the Main Storage Capacitors should be the voltage of the Input Power Line multiply by
1.4142 for Line Powered Generators (not in “Prep”) (±10%). See the following examples:

Input Power Line Voltage at the Main Storage Capacitors

Line Powered Generator at 230 VAC 230 x 1.4142 = 325 VDC (±10%)

Line Powered Generator at 400 VAC 400 x 1.4142 = 566 VDC (±10%)

Line Powered Generator at 480 VAC 480 x 1.4142 = 679 VDC (±10%)

 If the voltage is OK, check that P1-7 on the HT Controller is at logic level “0”. (0.75 V for
Generators with 1 Charge/ Discharge Monitor Board and 1.5 V for Generators with 2
Charge/Discharge Monitor Boards).
G If logic level is OK (logic level 0), replace the HT Controller Board.
G If logic level is not OK, check the links between P1-7 on the HT Controller Board and P1-2
on Charge/Discharge Monitor Board #1 and P1-1 on Charge/Discharge Monitor Board
#1 and P1-2 on Charge/Discharge Monitor Board #2,
-- If these links are OK, replace defective Charge/Discharge Monitor Board.
-- If they are not OK, repair connection.

 If the voltage in capacitors is not OK, check VAC on AC1, AC2 and AC3 at the Input Rectifier
Board.
G If VAC is not OK, verify input connections and input fuses.
G If VAC is OK, disconnect BUS+ and BUS-- at the Input Rectifier Board and check the VDC.
-- If VDC is not OK, replace any defective components (CR1, CR2, CR3) at the Input
Rectifier Board.
-- If VDC is OK, check for and replace any defective components at the Generator
Cabinet.

B. IF LED DS1 IS NOT ILLUMINATED


 Check input VAC on 6LF1,
G If VAC is OK, replace any defective components at the Input Rectifier Board (CR2, CR3
or CR1).
G If VAC on 6LF1 is not OK, check that contactor 6K5 is ON.
-- If 6K5 is ON, check VAC on T1,T2 and T3 (in 6K5).
-- If voltage is OK, replace the defective 6R1.
-- If voltage is not OK, replace the defective fuse (F3, F4 or F5).

 If Contactor 6K5 is not ON, check if signal +24VPSU is OK.


G If the signal is not OK, check and/or replace 3F6 and 6T2.
G If +24VPSU is OK, check that signal --LINE CONT is at 0 VDC,
-- if --LINE CONT is OK (0 VDC), replace 6K5.
-- If --LINE CONT is not OK, check continuity in J3-10 at the ATP Console CPU Board.
-- If --LINE CONT is OK, repair the connection between J3-10 and “A” in 6K5, at the
ATP Console CPU Board.
-- If --LINE CONT is not OK at the ATP Console CPU Board, replace the ATP
Console CPU Board.

4-34
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E12

DESCRIPTION : No mA during exposure or mA out of tolerance. Wrong filament current.


ERROR TYPE : Informative.
APPLICABLE TO : All Generators
APPEARS WHEN : After exposition.
INFORMATION / SYMPTOM : Error 12 appears after the Exposure to alert the operator that the mA at the start of the exposure has
not been correct. During the first 10 ms the Generator applies constant filament current to the Tube.
This current is proportional to the current already calibrated for that mA station at the kVp selected
for that Exposure (filament numbers). Near the end of these 10 ms, the HT Controller Board reads
the mA and if they are found to be 30% under or over what has been selected, it sends error “12” to
the Console.

POSSIBLE CAUSES
Calibration data for kVp and mA is not correct causing error E12.
The mA jumper on the HV Transformer is open, or it is not making good contact. The mA read at
the beginning of the exposure is 50% of the correct value (because one branch is open).
There is a problem on the reading of the mA.
No correct heating prior to the Exposure. The filament has not reached its correct temperature and
the mA at the start of the exposure is low. It usually occurs when the “Prep” and “Exp” buttons are
pressed down at the same time.
Making an exposure immediately after getting out of Calibration mode in Extended Memory.
+5 VDC , +12 VDC or --12 VDC Power Supplies of HT Control Board (measured at TP2, TP3 and
TP4 of this Board) have excessive ripple or VDC measured is not correct.

ACTIONS
1. Connect a Oscilloscope to the following Test Points on the HT Controller Board and check that
the voltage is correct. Adjust an incorrect voltage with the respective Potentiometer in the
Power Supply Board (refer to Section 2.1 -- Low DC Voltage Power Supply Test):
G TP2 (+5 VDC) on HT Control Board is adjusted with R12 Pot. in the Power Supply Board.
G TP3 (+12 VDC) on HT Control Board is adjusted with R26 Pot. in the Power Supply Board.
G TP4 (--12 VDC) on HT Control Board is adjusted with R25 Pot. in the Power Supply Board.

2. Check calibration data for the mA Open Loop (filament numbers) as described in the Service
Manual for all combinations of kVp and mA when this error appears.

With a Oscilloscope connected to Test Point TP5 (mA) on the HT Controller Board, check that
the mA reading is within a ratio of 1V=100 mA (± 5%) for software V2 and V3. For later versions
(V4 and up), the ratio is 1V=10mA from the min. mA to 80 mA and 1V=100mA from 100 mA
to maximum rating.

If it is not, the cause could be that the mA second test is not measuring correctly, or a wrong
measurement performed in the Generator. (Refer to step 3).

3. Check that the jumper in the mA Test Point of the HV Transformer is correctly placed and
secure.

4-35
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With a Oscilloscope check that during the entire Exposure signals on Test Points TP13 (--mA)
and TP14 (+mA) on the HT Control Board connectors are symmetrical (12%).
G If one is found missing or not symmetrical, measure on Pin 6 and 7 of J4 on the HT Control
Board.
G If they are symmetrical on both Points, the problem could be on the HT Control Board.
G If they are not correct, check that the connections made on the HV Transformer in J1-E,
D, K terminals are correctly connected and secure. Also check that the GND wire is
connected to GND stud.
G If connections are correct the problem is in the HV Transformer.

5. Connect an Oscilloscope to Test Point TP5 (mA) on the HT Controller Board.

Note . For the following test with a High Speed Generator, the self-running mode has not
been active. DIP switch 3243SW4--6 on the Control DRAC Board must be in the
“OFF” position (only High Speed Generator).

Check that:
G when the exposure is made by pressing first the PREP control and then the EXP control,
the mA reading at the beginning of exposure is correct.
G and when an exposure is made by pressing at the same time the PREP and EXP controls,
the mA reading at the beginning of exposure is low.

6. For Low Speed Generators:


G Reprogram the “Rotor Acceleration and Filament Setting Time” as stated in the Service
Manual (Configuration chapter) : one (1) step over the previously configured time (e.g.:
if it was 1.2 seconds, reprogram for 1.8 seconds) and check if boosting is correctly
configured.

If it is not correct, reprogram the “Rotor Acceleration and Filament Setting Time” one (1)
step over the previously configured time and test again.

7. For High Speed Generators:


G Reprogram the “RAD Filament Setting Time” as stated in the Service Manual
(Configuration chapter) : one (1) step over the previously configured time (e.g.: if the
Filament Setting Time programmed was 1.8 seconds, reprogram it for 2.7 seconds).

In order to verify that the problem has disappeared:


G select the highest mA station for Small Focus and the lowest kVp allowed for this mA
station. Make an Exposure by pressing at the same time the PREP and EXP controls;
check that the mA reading at the beginning of the Exposure is correct.
G select the highest mA station for Large Focus and the lowest kVp allowed for this mA
station. Make an Exposure by pressing at the same time the PREP and EXP controls;
check that the mA reading at the beginning of the Exposure is correct.

If it is not correct, reprogram the “RAD Filament Setting Time” one (1) step over the
previously configured time and test again.

4-36
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E13

DESCRIPTION : No kVp during exposure, kVp out of tolerance (±33%) or insufficient time parameters during
AEC operation.
ERROR TYPE : Informative. May abort exposition.
APPLICABLE TO : All Generators
APPEARS WHEN : During and after exposures.
INFORMATION / SYMPTOM : No kVp during exposure.

POSSIBLE CAUSES

Note . This error assumes that the fault is not due to arcing in the X-ray Tube.

Connection between the ATP Console CPU Board and HT Controller Board (--EXP signal)
Connection on the IPM Driver Boards
Faulty HT Controller
Faulty IPM Driver Board
Faulty IGBT
Open Serial Capacitor (5C9)
5V OUT during exposure
Power to IPM Drivers is not supplied from the Interface Control Board

ACTIONS
1. With an Oscilloscope, measure between TP-12 & TP-11 on the HT Controller. These signals
should be incomplete and/or irregular.

2. Assure that the power supply to the HT Controller Board and the two IPM Drivers are correct.

HT Controller: Schematic A3000-xx -- Check for 5V, --12VAC and +12VAC.

IPM Driver: Schematic A3063-06 -- Check for +24 VDC ±10 % (ROHS Generator) and check
for 5 V (both voltages are supplied from the Interface Control Board). 5V should be constant
(not presenting any variation), including during the entire Exposure.

3. Measure the pulses at P3-1 (KV DR1) and P3-2 (KV DR2) on the HT Controller. If these pulses
are not present during the Exposure, change the HT Controller. If they are present, proceed
to the next step.

The following procedures involve the IPM Driver Boards which are
connected to the IGBTs. This area has dangerous HIGH VOLTAGE and must
be treated with great care.

4-37
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Ensure that KV DR1 and KV DR2 reach the IPM Drivers at J1-1 & J1-2. If these signals are
not present at the connection points, revise the connection or cables. If these signals are
present, proceed to the next step.

5. Turn OFF the equipment. Dismount one of the IPM Drivers from the IGBT disconnecting the
terminals to IGBT side but keeping J1 & J2 connected.

Turn ON the equipment and ensure the power supply by measuring at J3-1 (+15 V OUTPUT
1) & J3-4 (GND 1), and by measuring at J4-1 (+15 V OUTPUT 2) & J4-4 (GND 2).

Turn OFF the equipment. If +15 V OUTPUT is not present, repair or replace if necessary.
Return the first IPM Driver to its original position connected to the IGBT. Then dismount the
second IPM Driver and repeat this step.

If +15 V OUTPUT is present in both IPM Drivers. proceed to the next step.

6. Turn OFF the equipment and wait a few moments for the equipment to fully discharge.
Remove the rear cover of the Generator Cabinet for accessing to Capacitor 5C9 located at the
back side of the HV Inverter. Check the Capacitor 5C9 as follows:
G by using a meter (e.g. Fluke 87) in “capacitor” mode
-- If Capacitor 5C9 is in good condition, the meter shows: “O.L μf (out of limit)” .
-- If Capacitor 5C9 is defective, the meter shows an abnormally low value (e.g: 0.19 μf).

G or, by using a meter in “ohms” mode


-- If Capacitor 5C9 is in good condition, the meter shows the resistance changing. If the
polarity of the probes is reversed, the meter shows a negative resistance.
-- If Capacitor 5C9 is defective, the meter shows a very high Impedance (1 MOhms)
or Open circuit.

If Capacitor 5C9 is defective, replace it. If not, proceed to the next step.

Capacitor 5C9

7. Refer to Schematic 543020xx for the connections between the HV Transformer and the IGBTs.
Measure the resistance (continuity) of the following cables from the IGBTs to the HV
Transformer, replacing or repairing where necessary:
G Capacitor 5C9 (Transformer side terminal) and P1 of the HV Transformer.
G C2E1 and P2

8. Disconnect P1 & P2 on the HV Transformer. Check that the resistance at P1 & P2 is low
(approximately 0.2 ). If the resistance is too high, replace the HV Transformer. If the
resistance is correct, proceed to the next step.

4-38
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

9. On the HT Controller Board, place SW2-2 to ON. Disconnect A2(B) +24V PSU (Faston
Terminal) at the 6K5 contactor (some are simply marked as K5).

10. Turn ON the equipment and verify that 6K5 does not activate and that there is NO TENSION
on the DC Bus (±300 VDC), Charge / Discharge Screws or Capacitor Bars.

Note . The following action requires the Inverter Module (L2), to which the
Charge/Discharge Boards are mounted, to be dismounted in order to gain access
to the IPM Driver Boards. Do not disconnect the Charge/Discharge Boards.

11. With an Oscilloscope at one of the IPM Driver boards, measure pulses between J3-4 (GND 1)
& J3-3 (CP1 KVDR1) and pulses between J4-4 (GND 2) & J4-3 (CN1 KVDR2). Repeat the
test on the other IPM Driver board. If the pulses are not present, replace the respective IPM
Driver Board.

4-39
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E14

DESCRIPTION : Exposure signal without X-ray Exposure Console command.


ERROR TYPE : Informative.
APPLICABLE TO : All Generators
APPEARS WHEN : During and after exposition.
INFORMATION / SYMPTOM : Exposure signal without X-ray Exposure Console command.

POSSIBLE CAUSES
The “Exp” signal is active on the HT Controller Board.

ACTIONS
1. Remove the connector J1 on HT Controller Board and check grounding of Pin 6 of P1.

2. If Pin 6 of P1 is grounded, replace the HT Controller Board.

3. If Pin 6 of P1 on the HT Controller Board is not grounded, then check grounding of Pin 6 of
Connector J1.

4. If it is grounded, remove the connector J3 on the ATP Console CPU Board and check grounding
of Pin 6 of Connector J1 again. If it is grounded, repair the wire short-circuit.

5. If it is not grounded, replace the ATP Console CPU Board.

Note . The resistance value between Pin 6 of P1 on HT Controller Board and Ground (with
all cables connected) should be around 900 Ω.

4-40
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E15

DESCRIPTION : Filament Driver Circuit Open, No current on Filament. Wrong selection of Focal Spot detected
during “Prep”.
ERROR TYPE : Informative. Does not allow “Prep”. Requires to be reset twice.
APPLICABLE TO : All Generators
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : No current detected on Focal Spot (Filaments Off). After resetting the Error Code, the filament
driver will be shut off.

POSSIBLE CAUSES
Absence of --12 V Power Supply at the HT Controller
A There is no power supplied to the Filament Board
Poor connection at the Cathode HV Cable
Defective Filament Transformer inside the HV Transformer
B Wrong signal --FIL1 ACK

ACTIONS

Note . “E15” may be provoked by an absence of --12 V (power supply) to the HT Controller
at P2-1 or from the Low Voltage Power Supply (LVDC Power Supply 3PS1) at
3TS1-16. See schematics 543020xx and HT Controller 3000-xx for further
information.

Before Troubleshooting for the Possible Causes listed above, proceed with the
following steps.

1. Verify -12V from the Low Voltage Power Supply to the HT Controller, identifying and correcting
any possible short circuit. If -12V is present, Troubleshoot for the causes listed in the Possible
Causes section above (Causes A and B). It -12V is not present, proceed to the next step.

2. There is a possibility of a short circuit at the Ion Chamber connections. If the system has AEC
and doesn’t have an AEC Adaptation Board, proceed to the next step. If the system has an AEC
Adaptation Board, proceed to step 4.

3. It is likely that there is a short circuit in 3TS1, most probably at 3TS1-16.

4. Disconnect J1, J2, J3 & J5 at the AEC Adaptation Board that go to the Ion Chambers and test
for a short circuit in one of these aforementioned connections.

4-41
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

A. THERE IS NO SUPPLY TO THE FILAMENT BOARD, A POOR CONNECTION ON THE


CATHODE HV CABLE, A DEFECTIVE FILAMENT TRANSFORMER

1. Visual Check for Unplugged Connectors:


With the Generator turned OFF, first of all take a visual inspection at all the connectors in the
Front Panel of the Generator (HT Controller Board, Filament Driver Board, Interface Control
Board and the High Voltage Switch Control Board). It could have happened that by an accident
some connector has been unplugged when removing or placing the generator cover. Take a
close inspection because sometimes a connector is disconnected a little bit from one of the
sides but not totally disconnected from the connector of the board. Please take a close look
with a flash light.

2. X--Ray Tube Filaments Resistance Test:


With the Generator turned OFF, measure the Filaments of each Tube by disconnecting the
Cathode High Voltage Cable from the HV Transformer (Generator side)

There are three pins located at the male end of the Plug: “C” (Common), “L” (Large) and “S”
(Small). With an Ohm meter, measure between “C” and “S”. If the filaments are OK (not opened),
the reading should be close to zero Ohms. Also measure between “C” and “L” and between “L”
and “S”.

If there is an open circuit in any of the Tube filaments, measure the resistance on the Tube side.
Remove the Cable from the Tube and test the ohms in the female plug (Tube side) with an Ohm
meter that has sufficiently long probes.

Nipple Ring Nut

Contact Pins “C”, “L”, “S” Plug


Cable

Nipple

C
L S

Termination Plug Grub Screw


Plug

3. Primary Filament Resistance Test (1):


With the Generator turned off, measure the resistance of the primary of the Filament
Transformer for the Small Focus by removing connector J1 on the HV Transformer and
measuring on the Connector of the HV Transformer between the pin marked G (FIL SUP) and
H (FIL 2 RTN). Also measure between G (FIL SUP) and F (FIL 1 RTN). These readings should
be about 5 ohms each primary.

4-42
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Primary Filament Resistance Test (2):


Reconnect J1 to the HV Transformer and ensure that it is securely connected and “locked” into
position by twisting the outside cover until a “click” is heard. Measure the Ohms between the
following points (Each one should read about 5 Ohms):
-- FIL 1 RTN (Large Focus) P3-4 on the Filament Driver Board and P4-15 on the
Interface Control Board.
-- FIL 2 RTN (Small Focus) P3-4 on the Filament Driver Board and P4-16 on the
Interface Control Board.
Move around the J1 cable that goes into the HV Transformer looking for a possible wrong
connection.

5. High Voltage Switch Connection:


a. If the primary resistance tests do not reveal the source of the error, remove the Cathode
High Voltage Cable from the HV Transformer.
b. Set dipswitch “SW2--2” of the HT Controller Board in “ON” position to turn OFF the
Filament Circuit (No Error 15) .
c. Turn the Generator “ON” and select “Tube 1” (or “Tube 2”).
d. Measure with an Ohm Meter the pins inside the HV Transformer receptacle for “Tube 1”
. For this, use long probes to be able to access to the pins of the “Female” High Voltage
Connector of the HV Transformer. Measure between Common (this is the pin in line with

the “Nipple”) and Small (it is the pin to the left). Measure also between Common and Large
(it is the pin to the right) . The resistance should be around 0.5 Ohms for each Filament.
e. Select “Tube 2”, if it is present, and repeat the measurement done in step-d. for “Tube 2”.

6. Once the test is finished, reconnect the Cathode HV cable for “Tube 1” in its position of the HV
Transformer (it is marked in the Cable and in the HV Transformer). Clean the cable with a piece
of clean dry paper before placing it inside the HV Transformer. Make sure you tighten the big
washer of the HV Cable into the HV Transformer.

7. Repeat the previous step for “Tube 2”.

8. Set dipswitch “SW2--2” of the HT Controller Board in “OFF” position to turn ON the Filament
Circuit.

4-43
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

B. WRONG SIGNAL --FIL1 ACK


Perform the following diagnosis if after pressing “Prep” the Error Code “E15” appears on the
Console.

Set a Techinque that select FIL1


( FIL1 is Small Focus in Tube 1 and it is Large Focus in Tube 2)

Check --FIL1 ACK Press “Prep”


Check and REPAIR NO at Interface Board in YES and check --FIL1 ACK at
link P1-8 HT Controller P1-8 at HT Controller. Interface Board (P4-22).
& P4-22 Interface Board Is it ok? Is it Low Level?

NO
YES

Check if K7
is active when
Replace YES selecting FIL1*.
HT Controller Board
Press Prep?
Is it active?
NO

YES Check signal


Replace FIL1 SLC at P4-19
YES
Interface Board Interface Board.
Is it at Low Level when
selecting FIL1*?
NO

Check and fix


link P1-13 in Check Signal --FIL1
YES
HT Controller SCL in P1-13 at HT
with P4--19 in Controll. Is it OK?
Interface Board
NO

NOTES
1) FIL1 is Small Focus in Tube1 and it is Large Focus in Tube2.
2) FIL1 SLC is activated when selecting FIL1 during Stand by for Replace
RAD Tubes. For R&F Tubes it is activated at “Prep”. HT Controller Board

4-44
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E16

DESCRIPTION : Invalid value of : kVp, mA or kW.


ERROR TYPE : Informative. Does not allow “Prep” or “Exp”.
APPLICABLE TO : All Generators
APPEARS WHEN : In “Prep”
INFORMATION / SYMPTOM : Selected mA or kVp are not correct.

POSSIBLE CAUSES
Wrong maximum kVp configuration.
Wrong software compatibility on the Generator Cabinet and Console.
When pressing “Prep” during the calibration of “Digital mA Loop Open” with a technique that
overpasses the Generator power kVp/mA.

ACTIONS
1. Check dipswitch--5 of SW2 on the HT Controller Board , it must be set in accordance with the
Generator configuration (maximum kVp of Generator) (dipswitch--5 must be in “open” position
for 125 kVp Generator, / dipswitch--5 must be in “closed” position for 150 kVp Generator).

2. When “Digital mA Loop Open” is manually calibrated (in calibration mode), this error may
appears if the Generator power is exceeded by a kVp / mA combination selection. Reset the
error indication and enter manually the Filament Current number following the instructions
described in the Calibration chapter without make any exposure for that combination (kVp /
mA).

4-45
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E17

DESCRIPTION : Communication error among ATP Console CPU Board and HT Controller Board.
ERROR TYPE : Fatal Error. Generator opens line contactor.
APPLICABLE TO : All Generators
APPEARS WHEN : Once initialization phase is over at any moment.
INFORMATION / SYMPTOM : No communication link between HT Controller Board and ATP Console CPU Board.
This Error Code requires that the Generator be turned OFF/ON before it can be fixed.

POSSIBLE CAUSES
The communication link between the ATP Console Board and the HT Controller Board is not reliable.
The Generator has been affected by a loss of the main power.
Faulty DC power supply.
Noise on the bucky circuitry.
Defective HT Controller Board or defective ATP Console CPU Board.

ACTIONS
1. Turn Generator OFF/ON.

2. If “E01” appears, follow the same procedure as for “E01”.

3. If “E01” does not appear, error “E17” can be due:


G a loss of the main power.
G an intermittent error produced by an external device, install a R--C filter in the power supply
and at Bucky start circuitry.

4-46
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E18 OR ROTOR ERROR

DESCRIPTION : Rotor error or Rotor running without order (only with Low Speed Starter).
ERROR TYPE : Fatal Error when the Rotor is running without order. Exposure is not allowed.
Informative Error when the Rotor is not running while “Prep” is active. Exposure is not allowed.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : The X-ray Tube is not detected to be rotating while “Prep” is active, then exposure is inhibited.
For Generators with Low Speed Starter, it can be due to the X-ray Tube anode is rotating
without a command from the Console.

POSSIBLE CAUSES
For all Generators:
“--PREP” signal is not going low in the HT Controller Board when “Prep” is pressed.

Only for Generators with High Speed Starter:


HT Controller Board is not receiving the “RTR Ready” signal back from the Control DRAC Board.

Only for Generators with Low Speed Starter:


“–RTR” signal on HT Controller Board is active.
Defective relay K1 (solid State) on the low speed module.
12 VDC power supply missing on the HT Controller Board.
Defective Low Speed Control or HT Controller Boards.

ACTIONS

A. PREVIOUS CHECK FOR ALL GENERATORS (LOW OR HIGH SPEED STARTER)


Check that the “--PREP” signal is going low in the HT Controller Board when “Prep” is active.
 “--PREP” signal can be measured in connector P1--3 of the HT Controller Board or by looking
at the LED DS2 in the HT Controller Board (it must be ON when “Prep” is active).
 If not, check the communication cable (J3) between the ATP console CPU Board and the
Generator connector 6J3 and then to the HT Controller Board (P1--3). Make sure it is properly
connected and that all the pins are well inserted on both sides (especially 6J3--4 on both sides).
G If there is not signal continuity between the ATP Console CPU Board (J3--4) and the HT
Controller Board (P1--3), repair or replace the communication cable.
G If the “--PREP” signal is going low and there is signal continuity between the ATP Console
CPU Board (J3--4) and the HT Controller Board (P1--3) and LED DS2 in the HT Controller
Board is OFF when “Prep” is active, replace the ATP Console CPU Board.
G If the “--PREP” signal is going low and there is signal continuity between the ATP Console
CPU Board (J3--4) and the HT Controller Board (P1--3) and LED DS2 in the HT Controller
Board is ON when “Prep” is active, perform procedure B for Generators with High Speed
Starter or procedure C for Generators with Low Speed Starter.

4-47
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

B. ONLY FOR GENERATORS WITH HIGH SPEED STARTER (LV-DRAC)


In “Prep” (the Rotor is not running while “Prep” is active)
“Rotor Error” means that the ATP Console CPU Board is not receiving the “Ready” communication
signal back from the HT Controller Board. This could be also because the HT Controller Board is
not receiving the “Ready” (--RTR RDY) signal from the High Speed Starter (LV-DRAC).
1. Perform the Programming of the High Speed Starter and RAD Filament Setting Time as
explained in the Configuration Chapter of the Generator Service Manual. Perform the following
sequence:
G Turn OFF the Generator and take note of the setting of Dipswitch 3000SW2 on the HT
Controller Board.
G Set all Dipswitch positions to OFF.
G Set Dipswitch of positions 1, 2, 5 and 7 to ON (position 2 to ON informs to the HT Controller
Board that it has a LV-DRAC connected).
G Turn ON the Generator and wait of E01. If E01 is not displayed leave the Generator for
1 minute and then proceed to turn it off.
G Set all Dipswitch positions as they were originally.
2. If after this, the “Rotor Error” is still present check that the “--RTR RUN” signal is going low in
the Control DRAC Board by measuring J4--20 in the Control DRAC Board when “Prep” is
active. Normally, if this signal is received by the Control DRAC Board a noise can be heard on
the LV-DRAC Transformers because the starter is trying to start the Tube.
G If there is no signal, check the signal continuity between the HT Controller Board (P5--5),
connector 6J1 of the Generator Cabinet (6J1--1) and Control DRAC Board (J4--20). If the
continuity is correct but there is no signal in the Control DRAC Board (J4--20), replace the
HT Controller Board.
G If this signal is arriving to the Control DRAC Board then most probably is that the LV-DRAC
(High Speed Starter) is accelerating the Tube and therefore the “--RTR RDY” signal is not
being received in the HT Controller Board.
-- Measure the signal “--RTR RDY” (active--low) in J4--4 of the Control DRAC Board.
Also the LED DL3 must be ON on the Control DRAC Board. If either is not correct:
-- Set Dipswitches for “Minimum Time for Ready (Delay)” on the Control DRAC
Board are set for “0” seconds (Dipswitches 3243SW1--1, SW1--2 and SW1--3 in ON
position).
-- Check the voltage at the following Test Points on the Control DRAC Board:
+5 VDC at TP14, +15 VDC at TP15, --15 VDC at TP37, using for GND reference the
following Test Points on the Control DRAC Board: TP2, TP4, TP12, TP13, or TP19.
If they are correct, replace the Control DRAC Board.
-- If voltages at the above points are not correct, turn the generator OFF and
wait three (3) minutes for the Main Storage Capacitors to discharge. Remove the
Ribbon Cable from Connector J3 at the Control DRAC Board. Turn the generator ON.

Check the voltage at the following points on the Interface DRAC Board:
+5 VDC at Anode of diode CR6,
+15 VDC at Anode of diode CR5 or at Cathode of diode CR6,
--15 VDC at Cathode of diode CR7.
For GND reference use the Terminal J2--3 on the Interface DRAC Board.

If voltages at the above points are correct, replace the Control DRAC Board.

If any voltage at the above points are not correct, verify at the Interface DRAC
Board that there is 220 VAC between J3--1 and J3--2, and fuse F1 and F2
are correct, if 220 VAC is present and the fuses are correct replace the
Interface DRAC Board.
-- If the signal “--RTR RDY” (active--low) in J4--4 of the Control DRAC Board is correct
and the LED DL3 is ON on the Control DRAC Board, check the signal continuity
between the Control DRAC Board (J4--4), connector 6J1 of the Generator Cabinet
(6J1--5) and HT Controller Board (P5--4). If the continuity is correct but there is no
signal in the HT Controller Board (P5--4), replace the HT Controller Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-48
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

C. ONLY FOR GENERATORS WITH LOW SPEED STARTER (LF-RAC)


In Stand-by (the Rotor is running without order)
1. Check if the signal “RTR I” Test Point (TP6 on HT Controller ) is 0 VDC.
G If it is 0 VDC, replace the HT Controller Board.
G If it is not 0 VDC, remove wire P4--3 “RTR I2” on the HT Controller Board and check again
the voltage at TP6.
-- If it is not 0 VDC, replace the HT Controller Board.
-- If it is 0 VDC, connect wire P4--3 “RTR I2” on the HT Controller Board and then ensure
that Solid Relay K1 (4K1 or 10K1) in the Low Speed Starter Module works properly.
For that, check the voltage between terminals 3 and 4 of the Solid Relay K1 is 0 VDC.
2. If the voltage between terminals 3 and 4 of the Solid Relay K1 it is not 0 VDC, replace the HT
Controller Board.
3. If the voltage between terminals 3 and 4 of the Solid Relay K1 is 0 VDC, check the voltage
between terminal 1 and 2 of the Solid Relay K1 is 60 VAC.
G If it is 60 VAC replace the LF-RAC Board.
G If it is not 60 VAC replace the Solid Relay K1 (4K1 or 10K1).

In “Prep” (the Rotor is not running while “Prep” is active)


Refer to the following Flowchart for procedure in “Prep”.

4-49
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

IF “E18” OR ROTOR ERROR APPEARS IN “PREP” (ONLY FOR LF-RAC)

VERIFY THAT:
LOW SPEED STARTER IS CONFIGURED TO
VISUALLY CHECK CORRECT TUBE
THAT ANODE IS YES AND CAPACITOR INSTALLED.
TURNING WHEN +
PRESSING “PREP”. VERIFY AT SIGNAL DELAYED +24V THAT LINE
DOES IT TURN? VOLTAGE IS TOO LOW. THAT MAY CAUSE
ROTOR ERROR OR E18.
+
CHECK WITH THE OSCILLOSCOPE SIGNAL RTRI
AT TP6 HT CONTROLLER BOARD.
SEE ILLUSTRATION BELOW.
NO
2.2 V
PREP

CHECK THAT
6K6 CONTACTOR 0.8
NO IS ENERGIZED WHEN
REPLACE 6K6 CONTACTOR GENERATOR IS POWER ON
(IT SUPPLIES “0 VAC RTR” running
TO LF-RAC BOARD).
IS IT OK?
CHECK TUBE STATOR

YES t
YES

tprep

CHECK LINK YES CHECK VAC ON TS2


FROM TS2 TO TUBE CONNECTIONS STATOR
STATOR. WHEN PRESSING “PREP”.
IS IT OK? IS IT OK?

NO NO

REPAIR CONNECTIONS

YES CHECK SIGNAL


REPLACE FUSE 6F6 +24 DELAYED.
IS IT OK?

NO

CHECK IF
YES --START AND --ACC NO REPEAT CONFIGURATION OF ROTOR
REPLACE K1 OR LF--RAC BOARD ARE AT LOW LOGIC LEVEL
ACCELERATION TIME FOR LOW SPEED
WHEN PRESSING “PREP”.
IS IT OK?

4-50
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E19

DESCRIPTION : mA detected without “Exp” command.


ERROR TYPE : Fatal Error. Generator has opened a line contactor that remains in an endless loop. It is
necessary to turn off the equipment.
APPLICABLE TO : All Generators
APPEARS WHEN : In stand-by or during initialization.
INFORMATION / SYMPTOM : Current in tube without ”Prep” command.

POSSIBLE CAUSES
 12 VDC power supplies missing on the HT Controller Board.
Defective HT Controller.
mA signal on H T Controller is active.

ACTIONS
1. Check  12 VDC power supplies.

2. In stand-by mode (no exposure), check for 0 VDC in TP5, TP13, TP14 as well as in connector
P4-6 and P4-7 of the HT Controller Board.

3. If voltage V ¸ 0 VDC in P4-6 and P4-7, check connections between P4-6 / P4-7 and J1 of the
HV Transformer.

4. If it is OK, replace HT Controller Board.

Note . As reference values:

-- the resistance measured between P4-6 or P4-7 on the HT Controller Board and
Ground (with all cables connected) should be around 900 Ω.

-- the resistance measured between P4-6 or P4-7 on the HT Controller Board and
Ground (with cable of J4 connector removed from the HT Controller Board) should
be around 200 KΩ.

4-51
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E20

DESCRIPTION : kVp detected without “Exp” command.


ERROR TYPE : Fatal Error. Generator has opened a line contactor that remains in an endless loop.
APPLICABLE TO : All Generators.
APPEARS WHEN : In stand-by or during initialization.
INFORMATION / SYMPTOM : kVp detected without “Exp” command.
“E20” on the Console indicates that the error must be solved with the equipment OFF.

POSSIBLE CAUSES
--12 VDC power supply missing.
Defective HT Controller Board.

ACTIONS
1. Check --12 VDC power supply.

2. In stand-by mode (no exposure), check for 0 VDC in TP7, TP11, TP12, as well as in connector
P4-1 and P4-2 of the HT Controller Board.

3. If voltage V ¸ 0 VDC in P4-1 and P4-2, check connections between P4-1 / P4-2 and J1 of the
HV Transformer.

4. If it is OK, replace HT Controller Board.

Note . As reference values:

-- the resistance measured between P4-1 on the HT Controller Board and Ground
(with all cables connected) should be around 18.5 KΩ.

-- the resistance measured between P4-2 on the HT Controller Board and Ground
(with all cables connected) should be around 17.5 KΩ.

-- the resistance measured between P4-1 or P4-2 on the HT Controller Board and
Ground (with cable of J4 connector removed from the HT Controller Board) should
be around 20 KΩ.

4-52
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E21

DESCRIPTION : Wrong Tube-1 selection.


ERROR TYPE : Informative. Does not allow operation until the error is solved or another Tube is selected.
APPLICABLE TO : All Generators.
APPEARS WHEN : Principally after initialization, at any moment or when changing Tube selection.
INFORMATION / SYMPTOM : Wrong Tube 1 selection.

POSSIBLE CAUSES
The –HT INTLK is not grounding (Tube 1 selection).
The --HT INTLK is missing.

ACTIONS

Note . Before performing any action, check the Workstation configuration for the Tube.

1. Check the –HT INTLK in 6J3-13 (Generator Cabinet), it should be 0 VDC with Tube-1 selected.

2. If it is not 0 VDC, check the Jumper between 6J3-13 and 6J3-16 (GND).

3. If 6J3-13 = 0 VDC, check J3--13 on the ATP Console CPU Board.

4. If not 0 VDC in J3-13, check link between 6J3--13 of the Generator Cabinet and J3-13 on the
ATP Console CPU Board.

5. If voltage is 0 VDC and “E21” appears, replace the ATP Console CPU Board.

4-53
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E23

DESCRIPTION : Calibration data not stored


ERROR TYPE : Indicative although it is almost impossible to find if not provoked.
APPLICABLE TO : All Generators.
APPEARS WHEN : When trying to save calibration data.
INFORMATION / SYMPTOM : Calibration data not stored. Calibration value intended to be stored has not been properly
recorded.

POSSIBLE CAUSES
This problem may be shown randomly.
This problem is continuos or occurs frequently and the communication link between the Generator
Cabinet and the Console is too noisy.

ACTIONS
1. Check calibration data and enter them manually if required.

2. Check communication cable (J3) between ATP Console CPU Board and HT Controller Board.

3. Route communication cables in a different way.

4-54
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E24

DESCRIPTION : Bucky Movement Signal has not been detected or DR Device is not ready for exposure.
ERROR TYPE : Indicative. Does not allow exposition.
APPLICABLE TO : All Generators.
APPEARS WHEN : At the beginning of the exposition.
INFORMATION / SYMPTOM : Wrong Exposure signal from Bucky or DR Devices.

POSSIBLE CAUSES
Wrong Workstation configuration or selection.
Wrong wiring of the Bucky.
Malfunction of the Bucky / DR Device / Standard Tomo Device.

ACTIONS
1. Check the configuration of the selected Workstation corresponds to the device (Bucky / DR /
Standard Tomo ) connected to the Generator (refer to the Installation / Configuration chapters
of the Generator Service Manual).

2. If none of these devices (Bucky / DR / Standard Tomo) have to be connected to the Generator
and “E24” Error Code is present, check if the Workstation configuration is correct, that is,
“Device value” ¸ 1 (Bucky-1), or ¸ 2 (Bucky-2), or ¸ 3 (Standard Tomo) (refer to the
Configuration chapter of the Generator Service Manual).

3. If the error is present selecting a Workstation configured as Bucky-1 (1) or Standard Tomo (3),
proceed with step 4.

If the error is present selecting a Workstation configured as Bucky-2 (2), proceed with step 5.

4. The error is present selecting a Workstation configured as Bucky-1 (1) or Standard


Tomo (3), make a provisional jumper between 3TS1--5 and 3TS1--6 of the Generator. Then,
make an exposure.

a. If the exposure is finished and “E24” Error Code is not present, check cable connection
between 3TS1 of the Generator and the respective Device (Bucky or DR). (Refer to the
Interconnection Map at the end of the Installation chapter of the Generator Service
Manual).

b. If “E24” Error Code is present, check if Relay K5 of the Interface Control Board (A3009--xx)
is activated.

c. If Relay K5 is not activated, check if signal BUCKY1 DRCMD is active (“Low Level”) in
P4--12 of the Interface Control Board when making an exposure.
-- If signal BUCKY1 DRCMD is active, replace the Interface Control Board (A3009--xx).
-- If it is not active, check this signal (BUCKY1 DRCMD) is active (“Low Level”) in J3--14
of the ATP Console CPU Board (A3024--xx) when making an exposure.

If it is active, check connection between J3--14 of the ATP Console CPU Board
(A3024--xx) and P4--12 of the Interface Control Board.

If it is not active, replace the ATP Console CPU Board (A3024--xx).

4-55
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

d. If Relay K5 is activated, check if Relay K3 is activated when making an exposure.


-- If Relay K3 is activated, check if signal BUCKY EXP is active (“Low Level”) in P4--6
of the Interface Control Board when making an exposure.
-- If signal BUCKY EXP is not active, replace the Interface Control Board (A3009--xx).
-- If signal BUCKY EXP is active, check for this signal active in J3--12 of the ATP
Console CPU Board (A3024--xx) when making an exposure.

If it is active, replace the ATP Console CPU Board (A3024--xx).

If it is not active, check connection between J3--12 of the ATP Console CPU Board
(A3024--xx) and P4--6 of the Interface Control Board.

5. The error is present selecting a Workstation configured as Bucky-2, Make a provisional


jumper between 3TS1--12 and 3TS1--13 of the Generator. Then make an exposure.

a. If the exposure is finished and “E24” Error Code is not present, check cable connection
between 3TS1 of the Generator and the respective Device (Bucky or DR). (Refer to the
Interconnection Map at the end of the Installation chapter of the Generator Service
Manual).

b. If “E24” Error Code is present, check if Relay K4 of the Interface Control Board (A3009--xx)
is activated.

c. If Relay K4 is not activated, check if signal BUCKY2 DRCMD is active (“Low Level”) in
P4--13 of the Interface Control Board when making an exposure.
-- If signal BUCKY2 DRCMD is active, replace the Interface Control Board (A3009--xx).
-- If it is not active, check this signal (BUCKY2 DRCMD) is active (“Low Level”) in J3--11
of the ATP Console CPU Board (A3024--xx) when making an exposure.

If it is active, check connection between J3--11 of the ATP Console CPU Board
(A3024--xx) and P4--13 of the Interface Control Board.

If it is not active, replace the ATP Console CPU Board (A3024--xx).

d. If Relay K4 is activated, check if Relay K3 is activated when making an exposure.


-- If Relay K3 is activated, check if signal BUCKY EXP is active (“Low Level”) in P4--6
of the Interface Control Board when making an exposure.
-- If signal BUCKY EXP is not active, replace the Interface Control Board (A3009--xx).
-- If signal BUCKY EXP is active, check for this signal active in J3--12 of the ATP
Console CPU Board (A3024--xx) when making an exposure.

If it is active, replace the ATP Console CPU Board (A3024--xx).

If it is not active, check connection between J3--12 of the ATP Console CPU Board
(A3024--xx) and P4--6 of the Interface Control Board.

6. Finally, remove the provisional jumpers in 3TS1 of the Generator.

4-56
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E27

DESCRIPTION : Failure in EPROM U24 of the ATP Console CPU Board. Bad checksum.
ERROR TYPE : Fatal Error. Generator does not start.
APPLICABLE TO : All Generators
APPEARS WHEN : During Generator initialization.
INFORMATION / SYMPTOM : The ATP Console CPU Board performs a checksum procedure of EPROM U24 when it has been
configured to a non-allowed calibration and it has found this value to be incorrect.

POSSIBLE CAUSES
The ATP Console CPU Board has been configured in a non communication mode.
The non volatile RAM does not calculate and compare the ATP Console CPU Board checksum.
The EPROM U24 has been corrupted.

ACTIONS
1. Check if dip-switch 3024SW2 on ATP Console CPU Board is correctly configured (refer to
Configuration Chapter -- “Configuration and Test Switches” of the Service Manual).

2. If error remains, replace the ATP Console CPU Board installing the EPROM U24 from the
previous board that has been removed.

3. If error remains, replace the EPROM U24 on ATP Console CPU Board.

4-57
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E33

DESCRIPTION : No communication link between the Generator and the Operator Console or PC Unit.
ERROR TYPE : Indicative.
APPLICABLE TO : All Generators with Operator Console.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Serial Communication Error.

POSSIBLE CAUSES
Connecting cable between Console and Generator is loose or damaged.
ATP Console CPU Board, or Operator Console, or Serial Port of the PC is damaged.
The Generator is turned OFF and PC is ON.

ACTIONS
This error means that there is NO serial communication (RS232) between the ATP Console CPU
Board and the Operator Console or PC (depends on the installation).

Note . The ATP Console CPU Board (for serial communication) is located at the back side
of the Front Panel the Generator cabinet.

1. Check the connection cable between connector J8 (or J7 depends on the Generator
configuration) from the ATP Console CPU Board to connector J5 of the Generator cabinet.
Then check the serial communication cable and connections from J5 to the Operator Console
or to the ON/OFF Interface Box (when a PC is used). (Refer to schematics 54301052 and
A6188--03 of the Installation chapter in the Service Manual).

2. In order to isolate the problem, turn the Generator ON and measure if any of the following
voltages are present in order to know if the Generator is really turned ON.
G 240 VAC SW: between 3TS1--26 and 3TS1--10.
G 115 VAC SW: between 3TS1--27 and 3TS1--10.
G Relay 3K3 (located on the front door of the generator) is energized.
G LED DS1 in the Interface Control Board is ON.

3. If the Generator is turned ON, check the following points:


a. 12 VDC UNR: In the ATP Console CPU Board between TP9 and TP11.

If 12 VDC UNR is not present in the ATP Console CPU Board, check Fuse 3F7. If Fuse
3F7 is OK, check the presence of 11 VAC at the input of Rectifier 3BR4.
-- If 11VAC is not present, check for 11 VAC between terminals 11 and 12 of
Transformer 6T2 and check Relay 3K3 works properly.
-- If 11 VAC is present at the input of Rectifier 3BR4, check for 12 VDC UNR measuring
in Capacitor 3C7. If 12 VDC UNR is not present, replace the Rectifier 3BR4. If
12 VDC UNR is present, check the wire connections.

4-58
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

b. If 12 VDC UNR is present between TP9 and TP11 of the ATP Console CPU Board, check
for:
-- 12 VDC ISO: In the ATP Console CPU Board between TP10 and TP1.
-- 5 VDC of the ATP: In the ATP Console CPU Board between TP3 and TP1.
-- DS1 of the ATP Console CPU Board (normally ON).

If any of the voltages is not present or correct, or if LED DS1 is OFF, replace the ATP
Console CPU Board.

c. If previous steps are correct, the problem is isolated in the RS232 connection between
the ATP Console CPU Board and the Operator Console or PC (depends on the
installation).

Check the Serial Port on the PC and if possible the RxD and TxD signal between the ATP
Console CPU Board and the PC. (Refer to schematics 54301052 and A6188--03 of the
Installation chapter in the Service Manual). Replace any defective part found.

To check the Serial Communication using an Oscilloscope measure the following signal:
-- On the ATP Console CPU Board, measure and check that there is a wave form on
the connector J8--3, if some RS232 signal is present that means the ATP Console
CPU Board is sending some information to the Operator Console, in case that no
signal is present on this pin replace the ATP Console CPU Board.
-- On the ATP Console CPU Board, measure and check that there is a wave form on
the connector J8--2, if some RS232 signal is present that means the ATP Console
CPU Board is receiving some information from the Operator Console, in case that
no signal is present check the following points:

-- For Push-button Console, check that all VDC are present on the Operator Console
and all internal connections are secure inside the Console.

-- For Touch Screen Console, check that all internal connections are secure inside
the Console.

-- If after the previous checking the wave form is present on the connector J8--2 but
the “E33” Error Code persists, then replace the ATP Console CPU Board.

4-59
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E34 -- TECHNIQUE ERROR

DESCRIPTION : Technique error.

ERROR TYPE : Informative without acoustic alarm or


Fatal error after exposition. It does not allow exposure.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment after parameter selection or after exposure.
INFORMATION / SYMPTOM : Warning Indication: “Technique error” on Consoles with written indicators. For other Consoles,
“E34” appears on the display.

POSSIBLE CAUSES
Calibration for specific parameters is incorrect.

ACTIONS
1. If this error occurs with time parameters close to 1ms -- The cable capacity is excessive for a
short exposure. Also the calibration for that parameters is wrong. Check the Calibration section
of the Service Manual: “Exposure Time Adjustment”.

2. DANGER: If it happens after an exposition, it indicates a failure in the exposition timer;


backup timer has been cut. Reset APR and reconfigure values.

4-60
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E35 -- DOOR OPEN

DESCRIPTION : Door Open.


For Generators with HV Transformer equipped with a Security Pressure Switch, the HV
Transformer surpasses the maximum internal pressure capacity.
ERROR TYPE : Informative. It may inhibit Exposition depending on the configuration of Dipswitch 3024SW1--2
on the ATP Console CPU Board.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Indicated with warning “Door Open” or error code “E35”. Does not allow “Prep” and “Exp”.

POSSIBLE CAUSES
“Door Open” Circuit is open (X-Ray Room Door open, cable disconnected or jumper cable not
installed).
When the HV Transformer is equipped with a Pressure Switch: Could indicate a need to replace the
HV Transformer because it has surpassed its maximum internal pressure capacity and is close to
deforming.
Fault at the ATP Console CPU Board.

ACTIONS
1. If the X-Ray Door signal is setup, ensure continuity in the Door Interlock Signal from the X-Ray
Room Door through 3TS1 (3TS1-22 & 3TS1-23) and J2 (J2-19 & J2-18) on the ATP Console
CPU Board (see the Installation Chapter of the Service Manual: “Door Interlock Signal”),
correcting continuity where necessary.

If the X-Ray Door signal is not setup, ensure the same continuity as above from the Jumper
at 3TS1 (3TS1-22 & 3TS1-23).

2. If the error persists and the HV Transformer is equipped with a Security Pressure Switch, this
error may appear when the HV Transformer surpasses its maximum internal pressure capacity
and is close to deforming. In this case, the HV Transformer must be replaced.

3. If the error persists and the HV Transformer has been replaced or is not equipped with a
Security Pressure Switch, replace the ATP Console CPU Board.

4-61
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E36 -- HEAT UNITS

DESCRIPTION : Safety Thermostat Open. Overheating.


ERROR TYPE : Informative without acoustic alarm. Does not allow expositions.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Indicated with “Heat Unit” warning in Consoles with written indicators. For the rest of Consoles
“E36” appears in display. Does not allow “Prep” and “Exp”.

POSSIBLE CAUSES
The thermostat of the selected Tube is open due to an overheated Tube Housing
(temperature > 60oC), a defective thermostat connection, or a faulty thermostat.
Discontinuity problem inside the wires or at connection points.
If the Tube Housing is cool (temperature < 60oC) and there is a closed circuit (connections are not
faulty and Voltmeter reading for Tube-1/2 is 0 VDC), there is a damaged ATP.
It may also appear during calibration due to the high volume of exposures (starting, braking and
Anode heating)

ACTIONS
1. If the error is due to overheating, wait until the temperature is lower than 60oC, the thermostat
will be closed.
2. When there is no thermostat on the X-ray Tubes, the respective thermostat signals have to be
connected to ground (For Tube-1: TS2-4 & TS2-5)
3. Check for a correct installation of the Thermostat/Presostat (connections for Tube-1: TS2-4 &
TS2-5).
G If the Tube does not have this signal, make the Jumpers as indicated above.
G If the Tube has this signal, check for a faulty ATP Console CPU Board.
START

Is the Tube Housing


Is there a Thermostat?
YES YES Wait for Tube Does the NO Normal
Warm?
error persist? Operation
(Temp <60oC) to cool
YES
NO

NO

Install Thermostat
or
Make Jumper following
Action 2 above Check
Continuity YES Damaged ATP Console CPU Board
Action 2 Replace ATP Console CPU Board
(0 VDC)
NO

Discontinuity Problem
Faulty Thermostat YES Check for break in circuitry inside
NO Is the Thermostat
Replace wires and at connection points
Closed?
Thermostat Replace necessary parts

4-62
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E37 -- TUBE OVERLOAD

DESCRIPTION : Tube Overload.


ERROR TYPE : Informative without acoustic alarm. Does not allow Exposition.
APPLICABLE TO : All Generators.
APPEARS WHEN : After “Exp” due to Tube overheating. After changing radiographic parameters. In some Consoles
this error may not appear as they are provided with an automatic blocking that disables
parameters above Tube capacity.
INFORMATION / SYMPTOM : Indicated with warning “Tube Overload” in Consoles with written indicators. For the rest of the
Consoles “E37” appears in the display. Exposures are not allowed.

POSSIBLE CAUSES
Parameters selected for a new exposition are above Tube capacity. (Heat Units or Rating).
The Capacity Line frequency is wrong or the Tube selected in the extended position “E02” or “E18”
is not correct.

ACTIONS
1. Wait until X-ray Tube temperature is lowered, so capacity of available Heat Units increase, or
modify Exposition parameters.

2. If Heat Units of the Tube are 100% and “E37” or “Tube Overload” warning appear, check that
Dip-switch 3024SW1--1 on ATP Console CPU Board is properly configured (Dip-switch
3024SW1--1 in “OFF” position for High Speed generators). (Refer to the Configuration Chapter
in the Service Manual).

3. Check the X-Ray Tube configured in the Extended Memory Location “E02” for Tube-1 (refer
to Section“X-ray Tube Selection” of the Configuration Chapter of the Service Manual).

4-63
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E41

DESCRIPTION : Dosimeter failure. Communication failure between the Tube-1 Dosimeter and Generator.
ERROR TYPE : Indicative.
APPLICABLE TO : All Generators with Dosimeter.
APPEARS WHEN : Communication error starting the measuring.
INFORMATION / SYMPTOM : Indicated as Error Code “E41”.

POSSIBLE CAUSES
Wrong Physical connection between Dosimeter and Generator for Tube-1.
Defective Radiation Meter Board (A3170--01) or defective ATP Console CPU Board (A3024--xx).

ACTIONS
1. Turn OFF/ON the Generator to reset the Radiation Measuring System.
2. Check the cable connections between the Dosimeter Ion Chamber for Tube-1 (located under
the Collimator) and the Connector IC-1 located on the Front Panel (Module--3) of the Generator
Cabinet (for more information, refer to Appendix “Adaptation of the Radiation Measuring
System).
3. Check the cable connections between Connector IC-1 and Connector P2 of the Radiation
Meter Board (A3170--01).
4. Check the cable connections between Connector P3 of the Radiation Meter Board (A3170--01)
to Connector J7 of the ATP Console CPU Board (A3024--xx) or to Connector COM4 of the TPC
(if ATP Console CPU Board is not used for the Dosimeter System).
5. Verify +24VDC is present between J1--2 and J1--3 of the Radiation Meter Board (A3170--01).
G If +24VDC is not present, check wire connections between J1 of the Radiation Meter
Board and 3TS1--17/18 in the Generator Cabinet.
G If +24VDC is present, disconnect the JACK connector next to Connector IC1 and check
if verify 23 VDC is present on the JACK connector (Generator side).
-- If 23 VDC is not present, check wire connection between P2 of the Radiation Meter
Board and JACK Connector, and if the connection is correct replace the Radiation
Meter Board (A3170--01).
-- If 23 VDC is present, follow next step.
6. On the Radiation Meter Board, verify between both ends of Diode D1 is 0 VDC present.
G If the voltage in Diode D1 is 24 VDC, check if there is a short-circuit between J1--1 and
GND.
G If the voltage in Diode D1 is 0 VDC, disconnect P2 and P3 of the Radiation Meter Board.
Maintain the Unit switched ON, measure continuity between P3--2 and P2--2, and
between P3--3 and P2--3. If there is not continuity between the mentioned points, replace
the Radiation Meter Board (A3170--01).
7. Only for Radiation Measuring Systems with connections through the ATP Console CPU Board.

Using an Oscilloscope, measure on the ATP Console CPU Board the presence of pulses
between connector J7--2 (reception) and GND, and between J7--3 (transmission) and GND.
G If pulses are present in J7--3 but pulses are not present in J7--2 , replace the DAP Ion
Chamber.
G If pulses are not present in J7--3, replace the ATP Console CPU Board (A3024--xx).
G If pulses are present in both mentioned points, replace the ATP Console CPU Board
(A3024--xx).

4-64
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E42

DESCRIPTION : Dosimeter failure. Autotest error on Tube-1 Dosimeter.


ERROR TYPE : Indicative.
APPLICABLE TO : All Generators with Dosimeter.
APPEARS WHEN : Communication error starting the measuring for Tube-1 or during regular operation.
INFORMATION / SYMPTOM : Indicated as Error Code “E42”. Failure test in Dosimeter.

POSSIBLE CAUSES
Error during electronic checking of DAP Ion Chamber for Tube-1.

ACTIONS
1. Turn OFF/ON the Generator to reset the Radiation Measuring System.

2. If the error persists, replace the DAP Ion Chamber.

4-65
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E43

DESCRIPTION : Dosimeter failure. Tube-1 DAP Ion Chamber status check error.
ERROR TYPE : Indicative.
APPLICABLE TO : All Generators with Dosimeter.
APPEARS WHEN : Communication error starting the measuring for Tube-1 or during regular operation.
INFORMATION / SYMPTOM : Indicated as Error Code “E43”. Wrong DAP Ion Chamber status request.

POSSIBLE CAUSES
Error during DAP Ion Chamber checking. DAP Ion Chamber not operative.

ACTIONS
1. Turn OFF/ON the Generator to reset the Radiation Measuring System.

2. If the error persists, it is necessary to recalibrate the DAP Ion Chamber.

4-66
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E48

DESCRIPTION : Collimator Error.


ERROR TYPE : Informative. Does not allow exposure.
APPLICABLE TO : All Generators.
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Showed as “Technique Error” in Consoles with this light indicator. On other Consoles, “E48” is
displayed.

POSSIBLE CAUSES
Collimator blades closed or in motion during exposure.
Defective Collimator.

ACTIONS
1. Check collimator blades.

2. If they are OK, check that the TS3-20 is at “Low level”


G If it is not at “Low level”: The problem is in the Collimator.
G If the signal is at “Low level”, check J2-6 on ATP Console.
-- if it is at “Low level”, replace ATP Console CPU Board
-- If is at “High level”, check and fix the link between J2-6 and TS3-20.

4-67
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E50

DESCRIPTION : Interrupted Exposure.


ERROR TYPE : Indicative.
APPLICABLE TO : All Generators.
APPEARS WHEN : When operator releases “Exp” control before exposure time has finished.
INFORMATION / SYMPTOM : Indicated as Error Code “E50” or “Exposure Aborted by the Operator”.

POSSIBLE CAUSES
Operator releases “Exp” button before exposure time has finished.

ACTIONS
1. Reset the Error indication.

2. Repeat the exposure.

3. If the error persists, check:


G Handswitch condition and connections, replace if needed.
G “EXP” button on the Console, Console condition and connections, replace the respective
parts if needed.

4-68
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E51

DESCRIPTION : LV-DRAC: Checksum failure or Microcontroller U17 corrupted in Control DRAC Board.
ERROR TYPE : Indicative. System does not allow exposition.
APPLICABLE TO : Generators with High Speed Starter
APPEARS WHEN : After self-test.
INFORMATION / SYMPTOM : At power On, after the Generator autocheck, Error Code “E51 is displayed and it is not possible
to make Exposures.

POSSIBLE CAUSES
DRAC program memory corrupted.

ACTIONS
1. Replace Microcontroller U17 on the Control DRAC Board.

4-69
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E52

DESCRIPTION : LV-DRAC: Microcontroller RAM failure.


ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter
APPEARS WHEN : After self-test.
INFORMATION / SYMPTOM : At power On, after the Generator autocheck, Error Code “E52” is displayed and it is not
possible to make Exposures.

POSSIBLE CAUSES
U17 on the Control DRAC Board is defective.

ACTIONS
1. Replace Microcontroller U17 on Control DRAC Board.

4-70
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E53

DESCRIPTION : LV--DRAC: Insufficient DC BUS voltage at low level voltage (220 VAC).
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter
APPEARS WHEN : At any moment.
INFORMATION / SYMPTOM : Error Code “E53” is displayed and it is not possible to make Prep.

POSSIBLE CAUSES
Low or no DC BUS Voltage on the Control DRAC Board.
Voltage at TP18 on the Control DRAC is <1.1VDC.

ACTIONS
1. Turn the generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV-DRAC MODULE.

2. On the Interface DRAC Board check that the voltage between J2--1 (--) and J2--2 (+) is higher
than 200 VDC.
G If it is not, perform Error Code “E11” checking.
G If it is higher than 200 VDC, go to next step.

3. On the Interface DRAC Board check that there is the same voltage between J2--5 (--) and J2--4
(+) than the voltage measured before between J2--1 (--) and J2--2 (+).
G If VDC is OK, go to step 8.
G If VDC is not OK, go to next step.

4. Turn the generator OFF and wait three (3) minutes for Main Storage Capacitors to discharge.

5. Check Fuses F3 and F4 on the Interface DRAC Board and that the Fuse Holders are making
contact with the Fuses. If necessary, replace the Fuses and close the contacts of the Fuse
Holders.

6. Turn the generator ON.

7. If the error persists, check on the Interface DRAC Board that there is the same voltage between
J2--5 (--) and J2--4 (+) than the VDC measured before between J2--1 (--) and J2--2 (+).
G If VDC is OK, go to step 8.
G If VDC is not OK, turn the generator OFF and wait three (3) minutes for the Main Storage
Capacitors to discharge.
-- If any of the Fuses are blown again, replace the Fuse(s) and replace the Control
DRAC Board.
-- If both Fuses are OK, replace the Interface DRAC Board.

4-71
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

8. On the Control DRAC Board, check that the voltage between J2--1 (--) and J2--2 (+) is the same
than the voltage measured before between J2--1 (--) and J2--2 (+) on the Interface DRAC
Board.
G If VDC is not OK:
-- Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors
to discharge.
-- Check the Faston terminals of the DC BUS Cable on both Boards (Control and
Interface DRAC Boards). If necessary, close the Faston terminals.
-- Check the continuity of the DC BUS Cable. If necessary, replace the DC BUS Cable.
G If VDC is OK, but the error persists, check that the voltage at TP18 on Control DRAC Board
is < 1.1 VDC.

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

-- If VDC is higher than 1.1 VDC, replace the Control DRAC Board.
-- If VDC is lower than 1.1 VDC, check the voltage at the following Test Points on the
Control DRAC Board:
-- +5 VDC at TP14
-- +15 VDC at TP15
-- --15 VDC at TP37

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

-- If voltages at the above Test Points are correct, replace the Control DRAC Board.
-- If any voltage at the above Test Points are not correct, go to next step.

9. Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.

10. Remove the Ribbon Cable from Connector J3 at the Control DRAC Board.

11. Turn the generator ON.

12. Check the voltage at the following points on the Interface DRAC Board:
-- +5 VDC at Anode of diode CR6
-- +15 VDC at Anode of diode CR5 or at Cathode of diode CR6
-- --15 VDC at Cathode of diode CR7

Note . For GND reference use the Terminal J2--3 on the Interface DRAC Board.

-- If voltages at the above points are correct, replace the Control DRAC Board.
-- If any voltage at the above points are not correct, replace the Interface DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-72
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E54

DESCRIPTION : LV--DRAC: Insufficient DC BUS voltage at high level voltage (400 / 480 VAC).
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter
APPEARS WHEN : At any moment
INFORMATION / SYMPTOM : Error Code “E54” is displayed and it is not possible to make Prep.

POSSIBLE CAUSES
Low or no DC BUS Voltage on the Control DRAC Board.
Voltage at TP18 on the Control DRAC is <2.48VDC.

ACTIONS
1. Turn the generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

2. On the Interface DRAC Board check that the voltage between J2--1 (--) and J2--2 (+) is higher
than 447 VDC.
G If it is not, perform Error Code “E11” checking.
G If it is higher than 447 VDC, go to next step.

3. On the Interface DRAC Board, check that there is the same voltage between J2--5 (--) and J2--4
(+) than the voltage measured before between J2--1 (--) and J2--2 (+).
G If VDC is OK, go to step 8.
G If VDC is not OK, go to next step.

4. Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.

5. Check Fuses F3 and F4 on the Interface DRAC Board and that the Fuse Holders are making
contact with the Fuses. If necessary, replace the Fuses and close the contacts of the Fuse
Holders.

6. Turn the generator ON.

7. If the error persists, check on the Interface DRAC Board that there is the same voltage between
J2--5 (--) and J2--4 (+) than the voltage measured before between J2--1 (--) and J2--2 (+).
G If VDC is OK, go to step 8.
G If VDC is not OK, turn the generator OFF and wait three (3) minutes for the Main Storage
Capacitors to discharge.
-- If any of the Fuses are blown again, replace the Fuse(s) and replace the Control
DRAC Board.
-- If both Fuses are OK, replace the Interface DRAC Board.

4-73
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

8. On the Control DRAC Board, check that the voltage between J2--1 (--) and J2--2 (+) is the same
than the voltage measured before between J2--1 (--) and J2--2 (+) on the Interface DRAC
Board.
G If VDC is not OK:
-- Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors
to discharge.
-- Check the Faston terminals of the DC BUS Cable on both Boards (Control and
Interface DRAC Boards). If necessary, close the Faston terminals.
-- Check the continuity of the DC BUS Cable. If necessary, replace the DC BUS Cable.
G If VDC is OK, but the error persists, check that the voltage at TP18 on the Control DRAC
Board is < 2.48 VDC.

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

-- If VDC is higher than 2.48 VDC, replace the Control DRAC Board.
-- If VDC is lower than 2.48 VDC, check the voltage at the following Test Points on the
Control DRAC Board:
-- +5 VDC at TP14
-- +15 VDC at TP15
-- --15 VDC at TP37

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

-- If voltages at the above Test Points are correct, replace the Control DRAC Board.
-- If any voltage at the above Test Points are not correct, go to next step.

9. Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.

10. Remove the Ribbon Cable from Connector J3 at the Control DRAC Board.

11. Turn the generator ON.

12. Check the voltage at the following points on the Interface DRAC Board:
-- +5 VDC at Anode of diode CR6
-- +15 VDC at Anode of diode CR5 or at Cathode of diode CR6
-- --15 VDC at Cathode of diode CR7

Note . For GND reference use the Terminal J2--3 on the Interface DRAC Board.

-- If voltages at the above points are correct, replace the Control DRAC Board.
-- If any voltage at the above points are not correct, replace the Interface DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-74
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E55

DESCRIPTION : LV--DRAC: Excessive DC BUS voltage at 480 VAC.


ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter.
APPEARS WHEN : At any moment
INFORMATION / SYMPTOM : Error Code “E55” is displayed and it is not possible to make Prep.

POSSIBLE CAUSES
DC BUS Voltage out of range on the Control DRAC Board.
Voltage at TP18 on the Control DRAC is > 4.92 VDC.

ACTIONS
1. Turn the generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

2. Check that the input voltage to the Generator (at Input Line Fuses) is according to
specifications (480 VAC, 10%). If it is not, check the Input Power Line.

Note . For a three--phase Generator, the DC BUS voltage should be approximately the
RMS voltage value between phases times 1.5, when the generator is ON and not
making exposures.

G The following steps are a few recommendations to check the Input Power Line. These
steps are not related with the Generator.
G Measure the voltage between all the phases: L1 and L2; L2 and L3; L3 and L1. They may
have a big disequilibrium in one (1) of them.
G If the three (3) measurements are equal, maybe the input voltage in the installation is not
sinusoidal and the peek between any of the phases is too higher.

Use an Oscilloscope with the shield of the probe isolated from ground. Both the probe and
the scope have to be able to handle and insulation voltage of 480 VAC (10%) between
the active of the probe and the shield.

Connect the probe between all the phases: L1 and L2; L2 and L3; L3 and L1. Check the
peek between any of the phases.

4-75
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. If the input voltage is OK, check on the Control DRAC Board that the voltage between J2--1
(--) and J2--2 (+) is < 800 VDC. Take note of the measured VDC.
G If VDC is > 800 VDC, go to step 8.
G If VDC is < 800 VDC, but the error persists, measure on the Control DRAC Board the
voltage between TP18 and a GND Test Point (TP2, TP4, TP12, TP13 or TP19).

Check that VDC measured at TP18 complies with the following scale factor:
181 VDC measured between J2--1 and J2--2 = 1 VDC measured at TP18.
-- If VDC ratio is OK, but the error persists, replace the Control DRAC Board.
-- If VDC ratio is not OK, check the voltage at the following Test Points on the Control
DRAC Board:
-- +5 VDC at TP14
-- +15 VDC at TP15
-- --15 VDC at TP37

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

-- If voltages at the above Test Points are correct, replace the Control DRAC Board.
-- If any voltage at the above Test Points are not correct, go to next step.

4. Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.
5. Remove the Ribbon Cable from Connector J3 at the Control DRAC Board.
6. Turn the generator ON.
7. Check the voltage at the following points on the Interface DRAC Board:
-- +5 VDC at Anode of diode CR6
-- +15 VDC at Anode of diode CR5 or at Cathode of diode CR6
-- --15 VDC at Cathode of diode CR7

Note . For GND reference use the Terminal J2--3 on the Interface DRAC Board.

-- If voltages at the above points are correct, replace the Control DRAC Board.
-- If any voltage at the above points are not correct, replace the Interface DRAC Board.

8. Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.
9. Disconnect the DC BUS Cable from J2--5 (--) and J2--4 (+) on the Interface DRAC Board.
10. Turn the generator ON.
11. Check on the Interface DRAC Board that the voltage between J2--1 (--) and J2--2 (+) is
< 800 VDC.
G If VDC is < 800 VDC, replace the Control DRAC Board.
G If VDC is > 800 VDC, go to the next step.

12. Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.
13. Disconnect the DC BUS Cable from J2--1 (--) and J2--2 (+) on the Interface DRAC Board.

4-76
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

14. Turn the generator ON.


15. At the wire ends (--) and (+) of the DC BUS Cable disconnected before from the Interface DRAC
Board, check that the voltage is < 800 VDC.
G If VDC is < 800 VDC, replace the Interface DRAC Board.
G If VDC is > 800 VDC, go to the next step.

16. Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.
17. Disconnect the wires (--) and (+) of the DC BUS Cable from the Input Rectifier Board.

DC BUS Cable DC BUS Cable


WIRE + WIRE --

DC BUS --

DC BUS +
Input Rectifier Board

18. Turn the generator ON.


19. Check if the voltage is < 800 VDC between BUS+ and BUS-- on the Input Rectifier Board
(Inverter Module).
G If VDC is > 800 VDC, replace the Input Rectifier Board.
G If VDC is < 800 VDC, at least one of the Main Storage Capacitors (Inverter Module) is
defective. Replace the Main Storage Capacitors.

20. Turn the generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-77
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E58

DESCRIPTION : LV--DRAC: Excessive current in the main winding during acceleration up to 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E58” is displayed and it is not possible to make exposures.

POSSIBLE CAUSES
Excessive current in the main winding up to 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-78
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-79
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.

6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Main wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Main wire (11TS2--1 for Tube-1). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. Desolder the wire marked as “12” on the Main Transformer of the LV--DRAC Module. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
10. On the Control DRAC Board, remove wire J1--1 (Main). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
11. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-80
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E59

DESCRIPTION : LV--DRAC: Excessive current in the auxiliary winding during acceleration up to 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E59” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Excessive current in the auxiliary winding up to 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-81
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-82
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.
6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Auxiliary wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Auxiliary wire (11TS2--2 for Tube-1). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. On the Interface DRAC Board, remove J5--1 and J5--4 wires. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Interface DRAC Board.
10. Remove the two (2) wires connected to the Clamping Board. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Clamping Board.
11. Desolder the wire marked as “12” on the auxiliary Transformer of the LV--DRAC Module.
Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
12. On the Control DRAC Board, remove wire J1--4 (Auxiliary). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
13. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-83
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E60

DESCRIPTION : LV--DRAC: Insufficient current in the auxiliary winding during acceleration up to 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E60” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the auxiliary winding up to 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-84
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-85
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-86
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-87
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E61

DESCRIPTION : LV--DRAC: Insufficient current in the main winding during acceleration up to 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E61” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the main winding up to 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-88
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-89
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-90
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-91
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E62

DESCRIPTION : LV--DRAC: Excessive current in the main winding during acceleration up to 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E62” is displayed and it is not possible to make exposures.

POSSIBLE CAUSES
Excessive current in the main winding up to 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-92
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-93
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.

6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Main wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Main wire (11TS2--1 for Tube-1). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. Desolder the wire marked as “12” on the Main Transformer of the LV--DRAC Module. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
10. On the Control DRAC Board, remove wire J1--1 (Main). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
11. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-94
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E63

DESCRIPTION : LV--DRAC: Excessive current in the auxiliary winding during acceleration up to 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E63” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Excessive current in the auxiliary winding up to 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-95
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-96
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists, replace the X--ray
Tube.
6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Auxiliary wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Auxiliary wire (11TS2--2 for Tube-1).
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. On the Interface DRAC Board, remove J5--1 and J5--4 wires. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Interface DRAC Board.
10. Remove the two (2) wires connected to the Clamping Board. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Clamping Board.
11. Desolder the wire marked as “12” on the auxiliary Transformer of the LV--DRAC Module.
Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
12. On the Control DRAC Board, remove wire J1--4 (Auxiliary). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
13. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-97
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E64

DESCRIPTION : LV--DRAC: Insufficient current in the auxiliary winding during acceleration up to 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E64” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the auxiliary winding up to 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-98
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-99
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-100
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-101
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E65

DESCRIPTION : LV--DRAC: Insufficient current in the main winding during acceleration up to 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E65” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the main winding up to 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-102
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-103
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-104
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-105
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E66

DESCRIPTION : LV--DRAC: Excessive current in the main winding running at 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E66” is displayed and it is not possible to make exposures.

POSSIBLE CAUSES
Excessive current in the main winding running at 3300 RPM is detected.
During the acceleration time the Contactor 11K2 remains OFF. Once the acceleration time has been
completed the Contactor 11K2 must be energized to apply to the Stator a lower voltage (running
voltage). This is the step where the errors are appearing (excessive current in running mode). For
any reason, Contactor 11K2 is not energized so that the voltage applied to the Stator is higher than
should be and this is causing the excessive current.
Defective: cable connections, or Contactor 11K2, or Control DRAC Board, or Stator Cable, or Tube
Stator.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-106
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-107
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.

6. Turn the Generator ON.

7. Check that the Contactor 11K2 is energized few moments after the “Prep” order.
G If it is OK, check if Contactor 11K2 is providing a good contact, if needed replace Contactor
11K2.
G If Contactor 11K2 is not energized, check +24 VDC at the coil of 11K2 (between A1 and
A2).

Note . The voltage must be verified when the Contactor is supposed to be ON (after the
acceleration time)

-- If the voltage is OK, check the coil of the Contactor by an ohmmeter. It must be around
160 Ohms. If it is not correct replace the Contactor.
-- If the voltage is not present, check 24 VDC (Delayed) between A1 of Contactor 11K2
and ground (this voltage must be present at any time).
-- If 24 VDC is not present between A1 of Contactor 11K2 and ground, check
continuity between the Interface Control Board (J5--2) and A1 of Contactor
11K2, checking it also in the aerial connector 11J7--1. Repair defective
connections..
-- If 24 VDC is present between A1 of Contactor 11K2 and ground, check the
link between A2 of Contactor 11K2 and J4--2 of the Interface DRAC Board.

If the link is not correct, repair or replace it.

If the link is correct, check low logic level at the cathode of CR13 when
Contactor 11K2 is supposed to be ON. If the low level is present check the
diode CR13, and if diode CR13 is defective replace this diode or the
Interface DRAC Board. Check also the flat cable between Interface DRAC
Board (J1) and Control DRAC Board, if the flat cable is not correct, replace
this cable.

If the flat cable is correct but you have +24 VDC between cathode of CR13
and ground, replace the Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-108
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E67

DESCRIPTION : LV--DRAC: Excessive current in the auxiliary winding running at 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E67” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Excessive current in the auxiliary winding running at 3300 RPM is detected.
During the acceleration time the Contactor 11K2 remains OFF. Once the acceleration time has been
completed the Contactor 11K2 must be energized to apply to the Stator a lower voltage (running
voltage). This is the step where the errors are appearing (excessive current in running mode). For
any reason, Contactor 11K2 is not energized so that the voltage applied to the Stator is higher than
should be and this is causing the excessive current.
Defective: cable connections, or Contactor 11K2, or Control DRAC Board, or Stator Cable, or Tube
Stator.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-109
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-110
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.

6. Turn the Generator ON.

7. Check that the Contactor 11K2 is energized few moments after the “Prep” order.
G If it is OK, check if Contactor 11K2 is providing a good contact, if needed replace Contactor
11K2.
G If Contactor 11K2 is not energized, check +24 VDC at the coil of 11K2 (between A1 and
A2).

Note . The voltage must be verified when the Contactor is supposed to be ON (after the
acceleration time)

-- If the voltage is OK, check the coil of the Contactor by an ohmmeter. It must be around
160 Ohms. If it is not correct replace the Contactor.
-- If the voltage is not present, check 24 VDC (Delayed) between A1 of Contactor 11K2
and ground (this voltage must be present at any time).
-- If 24 VDC is not present between A1 of Contactor 11K2 and ground, check
continuity between the Interface Control Board (J5--2) and A1 of Contactor
11K2, checking it also in the aerial connector 11J7--1. Repair defective
connections..
-- If 24 VDC is present between A1 of Contactor 11K2 and ground, check the
link between A2 of Contactor 11K2 and J4--2 of the Interface DRAC Board.

If the link is not correct, repair or replace it.

If the link is correct, check low logic level at the cathode of CR13 when
Contactor 11K2 is supposed to be ON. If the low level is present check the
diode CR13, and if diode CR13 is defective replace this diode or the
Interface DRAC Board. Check also the flat cable between Interface DRAC
Board (J1) and Control DRAC Board, if the flat cable is not correct, replace
this cable.

If the flat cable is correct but you have +24 VDC between cathode of CR13
and ground, replace the Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-111
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E68

DESCRIPTION : LV--DRAC: Insufficient current in the auxiliary winding running at 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E68” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the auxiliary winding running at 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-112
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-113
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-114
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-115
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E69

DESCRIPTION : LV--DRAC: Insufficient current in the main winding running at 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E69” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the main winding running at 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-116
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-117
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-118
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-119
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E70

DESCRIPTION : LV--DRAC: Excessive current in the main winding running at 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E70” is displayed and it is not possible to make exposures.

POSSIBLE CAUSES
Excessive current in the main winding running at 10000 RPM is detected.
During the acceleration time the Contactor 11K2 remains OFF. Once the acceleration time has been
completed the Contactor 11K2 must be energized to apply to the Stator a lower voltage (running
voltage). This is the step where the errors are appearing (excessive current in running mode). For
any reason, Contactor 11K2 is not energized so that the voltage applied to the Stator is higher than
should be and this is causing the excessive current.
Defective: cable connections, or Contactor 11K2, or Control DRAC Board, or Stator Cable, or Tube
Stator.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-120
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-121
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.

6. Turn the Generator ON.

7. Check that the Contactor 11K2 is energized few moments after the “Prep” order.
G If it is OK, check if Contactor 11K2 is providing a good contact, if needed replace Contactor
11K2.
G If Contactor 11K2 is not energized, check +24 VDC at the coil of 11K2 (between A1 and
A2).

Note . The voltage must be verified when the Contactor is supposed to be ON (after the
acceleration time)

-- If the voltage is OK, check the coil of the Contactor by an ohmmeter. It must be around
160 Ohms. If it is not correct replace the Contactor.
-- If the voltage is not present, check 24 VDC (Delayed) between A1 of Contactor 11K2
and ground (this voltage must be present at any time).
-- If 24 VDC is not present between A1 of Contactor 11K2 and ground, check
continuity between the Interface Control Board (J5--2) and A1 of Contactor
11K2, checking it also in the aerial connector 11J7--1. Repair defective
connections..
-- If 24 VDC is present between A1 of Contactor 11K2 and ground, check the
link between A2 of Contactor 11K2 and J4--2 of the Interface DRAC Board.

If the link is not correct, repair or replace it.

If the link is correct, check low logic level at the cathode of CR13 when
Contactor 11K2 is supposed to be ON. If the low level is present check the
diode CR13, and if diode CR13 is defective replace this diode or the
Interface DRAC Board. Check also the flat cable between Interface DRAC
Board (J1) and Control DRAC Board, if the flat cable is not correct, replace
this cable.

If the flat cable is correct but you have +24 VDC between cathode of CR13
and ground, replace the Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-122
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E71

DESCRIPTION : LV--DRAC: Excessive current in the auxiliary winding running at 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E71” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Excessive current in the auxiliary winding running at 10000 RPM is detected.
During the acceleration time the Contactor 11K2 remains OFF. Once the acceleration time has been
completed the Contactor 11K2 must be energized to apply to the Stator a lower voltage (running
voltage). This is the step where the errors are appearing (excessive current in running mode). For
any reason, Contactor 11K2 is not energized so that the voltage applied to the Stator is higher than
should be and this is causing the excessive current.
Defective: cable connections, or Contactor 11K2, or Control DRAC Board, or Stator Cable, or Tube
Stator.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-123
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-124
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.

6. Turn the Generator ON.

7. Check that the Contactor 11K2 is energized few moments after the “Prep” order.
G If it is OK, check if Contactor 11K2 is providing a good contact, if needed replace Contactor
11K2.
G If Contactor 11K2 is not energized, check +24 VDC at the coil of 11K2 (between A1 and
A2).

Note . The voltage must be verified when the Contactor is supposed to be ON (after the
acceleration time)

-- If the voltage is OK, check the coil of the Contactor by an ohmmeter. It must be around
160 Ohms. If it is not correct replace the Contactor.
-- If the voltage is not present, check 24 VDC (Delayed) between A1 of Contactor 11K2
and ground (this voltage must be present at any time).
-- If 24 VDC is not present between A1 of Contactor 11K2 and ground, check
continuity between the Interface Control Board (J5--2) and A1 of Contactor
11K2, checking it also in the aerial connector 11J7--1. Repair defective
connections..
-- If 24 VDC is present between A1 of Contactor 11K2 and ground, check the
link between A2 of Contactor 11K2 and J4--2 of the Interface DRAC Board.

If the link is not correct, repair or replace it.

If the link is correct, check low logic level at the cathode of CR13 when
Contactor 11K2 is supposed to be ON. If the low level is present check the
diode CR13, and if diode CR13 is defective replace this diode or the
Interface DRAC Board. Check also the flat cable between Interface DRAC
Board (J1) and Control DRAC Board, if the flat cable is not correct, replace
this cable.

If the flat cable is correct but you have +24 VDC between cathode of CR13
and ground, replace the Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-125
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E72

DESCRIPTION : LV--DRAC: Insufficient current in the auxiliary winding running at 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E72” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the auxiliary winding running at 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-126
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-127
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-128
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-129
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E73

DESCRIPTION : LV--DRAC: Insufficient current in the main winding running at 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E73” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the main winding running at 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-130
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-131
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-132
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-133
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E74

DESCRIPTION : LV--DRAC: Excessive current in the main winding braking at 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E74” is displayed and it is not possible to make exposures.

POSSIBLE CAUSES
Excessive current in the main winding braking at 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-134
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-135
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.

6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Main wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Main wire (11TS2--1 for Tube-1). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. Desolder the wire marked as “12” on the Main Transformer of the LV--DRAC Module. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
10. On the Control DRAC Board, remove wire J1--1 (Main). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
11. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-136
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E75

DESCRIPTION : LV--DRAC: Excessive current in the auxiliary winding braking at 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E75” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Excessive current in the auxiliary winding braking at 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-137
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-138
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.
6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Auxiliary wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Auxiliary wire (11TS2--2 for Tube-1). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. On the Interface DRAC Board, remove J5--1 and J5--4 wires. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Interface DRAC Board.
10. Remove the two (2) wires connected to the Clamping Board. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Clamping Board.
11. Desolder the wire marked as “12” on the Auxiliary Transformer of the LV--DRAC Module.
Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
12. On the Control DRAC Board, remove wire J1--4 (Auxiliary). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
13. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-139
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E76

DESCRIPTION : LV--DRAC: Insufficient current in the auxiliary winding braking at 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E76” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the auxiliary winding braking at 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-140
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-141
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-142
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-143
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E77

DESCRIPTION : LV--DRAC: Insufficient current in the main winding braking at 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E77” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the main winding braking at 3300 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-144
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-145
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-146
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-147
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E78

DESCRIPTION : LV--DRAC: Excessive current in the main winding braking at 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E78” is displayed and it is not possible to make exposures.

POSSIBLE CAUSES
Excessive current in the main winding braking at 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-148
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-149
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.

6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Main wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Main wire (11TS2--1 for Tube-1). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. Desolder the wire marked as “12” on the Main Transformer of the LV--DRAC Module. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
10. On the Control DRAC Board, remove wire J1--1 (Main). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
11. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-150
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E79

DESCRIPTION : LV--DRAC: Excessive current in the auxiliary winding braking at 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E79” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Excessive current in the auxiliary winding braking at 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-151
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-152
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.
6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Auxiliary wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Auxiliary wire (11TS2--2 for Tube-1). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. On the Interface DRAC Board, remove J5--1 and J5--4 wires. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Interface DRAC Board.
10. Remove the two (2) wires connected to the Clamping Board. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Clamping Board.
11. Desolder the wire marked as “12” on the Auxiliary Transformer of the LV--DRAC Module.
Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
12. On the Control DRAC Board, remove wire J1--4 (Auxiliary). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
13. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-153
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E80

DESCRIPTION : LV--DRAC: Insufficient current in the auxiliary winding braking at 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E80” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the auxiliary winding braking at 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-154
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-155
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-156
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-157
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E81

DESCRIPTION : LV--DRAC: Insufficient current in the main winding braking at 10000 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E81” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the main winding braking at 10000 RPM is detected.
Defective: Interface DRAC Board, or Control DRAC Board, or Tube Stator, or Stator Cable, or
Clamping Board, or Transformers, or Contactors in LV--DRAC Module.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-158
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-159
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-160
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-161
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E82

DESCRIPTION : LV-DRAC: Wrong X-ray Tube selection.


ERROR TYPE : Indicative. Does not allow exposures.
APPLICABLE TO : All Generators with High Speed Starter (LV-Drac).
APPEARS WHEN : During POWER OFF or in stand by.
INFORMATION / SYMPTOM : E82 appears on display

POSSIBLE CAUSES
Incorrect turning off routine.
Incorrect signals of tube selection in the LV-DRAC.
Defective Delayed Switch Off PCB.
Defective Control DRAC PCB.

ACTIONS
1. With the Generator off, disconnect J12 from A3274 Delayed Switch Off PCB and jumper the
aerial connector J1-1 & 8. Turn the Generator ON, if the error persists, replace A3243 Control
DRAC PCB.
2. If the error dissapears, verify that the generator performs the turning off routines in a correct
way. if ok, replace the Delayed Switch OFF PCB A3274-XX.
3. If the Generator turning off routines are not correct, verify the connections between 6J2 in the
Generator Cabinet and J1-8 in the Delayed Switch OFF PCB as well verify connection between
J1-8 in the Interface Control PCB and J1-1 in the Delayed Switch OFF PCB.
4. If the connections are not ok, repair accordingly, if the connections are ok, verify the
components related to the Turning OFF function of the Generator according to the system
configuration.

4-162
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E83

DESCRIPTION : LV--DRAC: Excessive current in DC Brake.


ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode braking in DC voltage.
INFORMATION / SYMPTOM : Error Code “E83” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Excessive current in the main or auxiliary winding is detected.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-163
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-164
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check for correct isolation between the connections for the Main winding and Auxiliary winding
in 11TS2 on the LV--DRAC Module. Measure the impedance between Main and Ground,
Auxiliary and Ground, Common and Ground; in all cases it must be high impedance
(mega--ohms).
G If the impedance is OK, go to the next step.
G If the impedance is not correct, it can be due to:
-- a wrong connection of the Stator Cable on the Tube side or on the Generator side.
Check the connection of the Stator Cable and connect it properly if necessary.
-- or the Stator Cable is short--circuited, then disconnect the Stator Cable on the Tube
side and measure impedance in the Tube. If the problem persists replace the X--ray
Tube.
6. Turn the Generator ON.

To perform each one of the following checks it is necessary:


- Turn the Generator OFF and wait three (3) minutes to discharge the
Main Storage Capacitors.
- Remove the wire indicated in each step below.
- Turn the Generator ON. Press PREP and note the error code and reset it.

7. On the Tube side, remove the Auxiliary wire. Check:


G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Tube Stator. Replace the
X--ray Tube.
8. On the Generator side, remove the Auxiliary wire (11TS2--2 for Tube-1). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, the problem is in the Stator Cable. Replace the
Stator Cable.
9. On the Interface DRAC Board, remove J5--1 and J5--4 wires. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Interface DRAC Board.
10. Remove the two (2) wires connected to the Clamping Board. Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Clamping Board.
11. Desolder the wire marked as “12” on the Auxiliary Transformer of the LV--DRAC Module.
Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, there is a bad contact from 11TS2 to Terminal
“12” in the Transformer on the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to
Schematic “543020xx” in the Service Manual). Identify faulty connections, clean the
contacts, tighten each wire connection, and/or replace Contactors if necessary.
12. On the Control DRAC Board, remove wire J1--1 (Main) and J1--4 (Auxiliary). Check:
G If an “Excessive Current” code appears, go to the next step.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.
13. On the Control DRAC Board, remove wire J1--3 (Shield). Check:
G If an “Excessive Current” code appears, replace the Control DRAC Board.
G If an “Insufficient Current” code appears, replace the Main Transformer at the LV--DRAC
Module.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-165
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E87

DESCRIPTION : LV--DRAC: Insufficient current in common wire during acceleration up to 3300 RPM
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E87” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the common wire up to 3300 RPM is detected.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-166
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-167
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-168
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-169
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E88

DESCRIPTION : LV--DRAC: Insufficient current in common wire running at 3300 RPM.


ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at Low Speed.
INFORMATION / SYMPTOM : Error Code “E88” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the common wire running at 3300 RPM is detected.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-170
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-171
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-172
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-173
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E89

DESCRIPTION : LV--DRAC: Insufficient current in common wire during acceleration up to 10000 RPM.
ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E89” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in the common wire during acceleration up to 10000 RPM is detected.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-174
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-175
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-176
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-177
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E90

DESCRIPTION : LV--DRAC: Insufficient current in common wire running at 10000 RPM.


ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : During Anode acceleration at High Speed.
INFORMATION / SYMPTOM : Error Code “E90” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
Insufficient current in common wire running at 10000 RPM is detected.

ACTIONS

All the wires / cables removed during the procedure have to be isolated to
avoid contact with any other wire or ground.

1. Verify the Generator is turned OFF.

2. Check that the DIP switches on the Control DRAC Board are set in the correct position. (Refer
to the LV -DRAC Documentation in this Service Manual, especially check that DIP-switches
3243SW3--5 and 3243SW4--5 are in the correct position according to the Tube Family
Selection.)

3. Check each wire connection of the Stator Cable on the Tube side and on 11TS2 on the
LV--DRAC Module are correct and secured (good contact), connect properly, and tighten each
wire if necessary. Check if the error has been corrected. If not, go to the next step.

4-178
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. With the Generator turned OFF, measure the Stator Tube Winding impedances on 11TS2 on
the LV--DRAC Module.

Note . The Contactor 11KT1 on the LV--DRAC Module have to be not energized
(contactor indicator is visible) when measuring the Stator impedances of Tube-1.

Contactor 11KT1 measuring


Stator impedances of Tube-1

G For Tube-1 measure:


-- between Main (11TS2--1) and Common (11TS2--3).
-- between Auxiliary (11TS2--2) and Common (11TS2--3).
-- between Main (11TS2--1) and Auxiliary (11TS2--2), it must be the sum of
impedance measured between Main and Common plus Auxiliary and Common.

Note . Impedance must be as stated in the X--ray Tube Stator specifications. (Refer to the
Manufacturer X--ray Tube Data Sheet.)

-- If the impedance is OK, go to the next step.


-- If the impedance is not correct:
-- Measure the Stator Tube Winding impedances on the Tube side:
-- Between Main and Common.
-- Between Auxiliary and Common.
-- Between Main and Auxiliary, it must be the sum of impedance measured
between Main and Common plus Auxiliary and Common.

-- If the impedance is correct on the Tube, replace the Stator Cable.


-- If the impedance is not correct on the Tube, replace the X--ray Tube.

4-179
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Check that there is a low resistance (ohms) on the primary and secondary side of either the
Main or Auxiliary Transformer (depending if the error is in the Main or Auxiliary winding):
-- For Primary measure between terminals 1--7 of the Transformer.
-- For Secondary measure between terminals 8--10 and 10--12 of the Transformer.
G If the low resistance is OK, go to the next step.
G If the low resistance is not correct, replace the Transformer.

6. On the Control DRAC Board, check that there is a continuity between the following points:
-- For Main measure between J1--1 and J1--3.
-- For Auxiliary measure between J1--3 and J1--4.
G If the continuity is OK, go to the next step.
G If the continuity is not correct, check cable connections in J1 of the Control DRAC Board
and in Primary side of the Transformers.

7. With the Generator turned OFF, check that Contactors 11KT1, 11K2, and 11K3 are not
energized (OFF) on the LV--DRAC Module.
G If it is OK, go to the next step.
G If any Contactor is energized (ON), replace the Contactor.

8. Turn the Generator ON, check on the LV--DRAC Module that ten (10) seconds after power--up
Contactors 11K2 and 11K3 are not energized (OFF) and Contactor 11KT1 is energized (ON)
when Tube-1 is selected.
G If it is OK, go to the next step.
G If Contactor 11K2 or 11K3 is energized (ON), replace the Control DRAC Board.
G If Contactor 11KT1 is not energized (OFF) (when Tube--1 is selected), turn the Generator
OFF and wait three (3) minutes for the Main Storage Capacitors to discharge. Check that
Diode CR14 is OK on the Interface DRAC Board.
-- Diode CR14 is OK, replace the Control DRAC Board.
-- Diode CR14 is not OK, replace the Interface DRAC Board.

9. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

10. Check that there is a continuity between the following points:


-- For Main measure between terminal 12 of the Main Transformer and 11TS2--1 (for
Tube-1).
-- For Auxiliary measure between terminal 12 of the Auxiliary Transformer and
11TS2--2 (for Tube-1).
-- For Common measure between terminal 8 of each Transformer (Main and Auxiliary)
and then between terminal 8 of the Main Transformer and 11TS2--3 (for Tube-1).
G If the continuity is OK, go to the next step.
G If the continuity is not correct, there is a bad contact on the Transformer Terminals or on
the Contactor’s Terminals 11KT1, 11K2, and/or 11K3 (refer to Schematic “543020xx” in
the Service Manual). Identify faulty connections, clean the contacts, and/or tighten each
wire connection.

4-180
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Check the proper operation of the Control DRAC Board. For that:
G Set DIP switch 3243SW4--7 in the “ON” position to inhibit errors.
G Select: 40 kVp, 10 mA and 10 ms (low speed) and Direct Workstation. Check the X--ray
Tube is starting properly by performing an Anode Rotation Test (refer to “Configuration”
chapter in the Service Manual).
-- If the Tube does not start, replace the Control DRAC Board.
-- If the Tube starts, the Control DRAC Board has to be replaced but the unit can work
(for a short period of days) whenever DIP switch 3243SW4--7 is in the “ON” position
to inhibit errors.
-- In both cases, set DIP switch 3243SW4--7 in the “OFF” position after replacing the
Control DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-181
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E91

DESCRIPTION : LV--DRAC: Incorrect signal measure in IPRINC (CH2).


ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : In Stand-by situation.
INFORMATION / SYMPTOM : Error Code “E91” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
The Generator is detecting current through the main wire when it should be zero (0).

ACTIONS
1. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

2. Check the voltage at TP1 and TP17 on the Control DRAC Board. Both voltages should be
0 VDC.

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

G If both voltages at TP1 and TP17 are 0 VDC and the Error Code persists, replace the
Control DRAC Board.
G If any voltage at TP1 and TP17 is different than 0 VDC, go to the next step.

3. Check the voltage at the following Test Points on the Control DRAC Board:
-- +5 VDC at TP14
-- +15 VDC at TP15
-- --15 VDC at TP37

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

G If voltages at the above Test Points are correct, replace the Control DRAC Board.
G If any voltage at the above Test Points are not correct, go to the next step.

4-182
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Turn the Generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.

5. Remove the Ribbon Cable from Connector J3 at the Control DRAC Board.

6. Turn the Generator ON.

7. Check the voltage at the following points on the Interface DRAC Board:
-- +5 VDC at Anode of diode CR6
-- +15 VDC at Anode of diode CR5 or at Cathode of diode CR6
-- --15 VDC at Cathode of diode CR7

Note . For GND reference use the Terminal J2--3 on the Interface DRAC Board.

G If voltages at the above points are correct, replace the Control DRAC Board.
G If any voltage at the above points are not correct, replace the Interface DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-183
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E92

DESCRIPTION : LV--DRAC: Incorrect signal measure in IAUX (CH3).


ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : In Stand-by situation.
INFORMATION / SYMPTOM : Error Code “E92” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
The Generator is detecting current through the auxiliary wire when it should be zero (0).

ACTIONS
1. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

2. Check the voltage at TP3 and TP16 on the Control DRAC Board. Both voltages should be
0 VDC.

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

G If both voltages at TP3 and TP16 are 0 VDC and the Error Code persists, replace the
Control DRAC Board.
G If any voltage at TP3 and TP16 is different than 0 VDC, go to the next step.

3. Check the voltage at the following Test Points on the Control DRAC Board:
-- +5 VDC at TP14
-- +15 VDC at TP15
-- --15 VDC at TP37

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

G If voltages at the above Test Points are correct, replace the Control DRAC Board.
G If any voltage at the above Test Points are not correct, go to the next step.

4-184
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Turn the Generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.

5. Remove the Ribbon Cable from Connector J3 at the Control DRAC Board.

6. Turn the Generator ON.

7. Check the voltage at the following points on the Interface DRAC Board:
-- +5 VDC at Anode of diode CR6
-- +15 VDC at Anode of diode CR5 or at Cathode of diode CR6
-- --15 VDC at Cathode of diode CR7

Note . For GND reference use the Terminal J2--3 on the Interface DRAC Board.

G If voltages at the above points are correct, replace the Control DRAC Board.
G If any voltage at the above points are not correct, replace the Interface DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-185
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E93

DESCRIPTION : LV--DRAC: Incorrect signal measure in ICOM (CH4).


ERROR TYPE : Indicative. Exposure is not allowed.
APPLICABLE TO : All Generators with High Speed Starter.
APPEARS WHEN : In Stand-by situation.
INFORMATION / SYMPTOM : Error Code “E93” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
The Generator is detecting current through the main wire when it should be zero (0).

ACTIONS
1. Turn the Generator ON.

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY HEATSINK OR ANY


COMPONENT IN THE LV- DRAC MODULE.

2. Check the voltage at TP1, TP3, and TP36 on Control DRAC Board. All voltages should be
0 VDC.

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

G If both voltages at TP1 and TP17 are 0 VDC and the Error Code persists, replace the
Control DRAC Board.
G If any voltage at TP1, TP3, or TP36 is different than 0 VDC, go to the next step.

3. Check the voltage at the following Test Points on the Control DRAC Board:
-- +5 VDC at TP14
-- +15 VDC at TP15
-- --15 VDC at TP37

Note . For GND reference use the following Test Points on the Control DRAC Board: TP2,
TP4, TP12, TP13, or TP19.

G If voltages at the above Test Points are correct, replace the Control DRAC Board.
G If any voltage at the above Test Points are not correct, go to the next step.

4-186
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Turn the Generator OFF and wait three (3) minutes for the Main Storage Capacitors to
discharge.

5. Remove the Ribbon Cable from Connector J3 at the Control DRAC Board.

6. Turn the Generator ON.

7. Check the voltage at the following points on the Interface DRAC Board:
-- +5 VDC at Anode of diode CR6
-- +15 VDC at Anode of diode CR5 or at Cathode of diode CR6
-- --15 VDC at Cathode of diode CR7

Note . For GND reference use the Terminal J2--3 on the Interface DRAC Board.

G If voltages at the above points are correct, replace the Control DRAC Board.
G If any voltage at the above points are not correct, replace the Interface DRAC Board.

Connect all the wires / cables removed during the procedure back to the
original connections.

4-187
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E95

DESCRIPTION : Rapid Termination


ERROR TYPE : Indicative. Does not allow exposures.
APPLICABLE TO : Generators with Rapid Termination application Installed.
APPEARS WHEN : While Exposure (after 30% of the exposure back-up time and after 10 ms of exposure, both
conditions have to be fulfilled).
INFORMATION / SYMPTOM : Error Code “E95” is displayed and it is not possible to make Exposures.

POSSIBLE CAUSES
No radiation is detected on the Ion Chamber.
The selected parameters are not appropriate for an exposure with AEC (due to a short backup time
/ mAs Operator selection).
The selected Ion Chamber is not correct.
Dipswitch 3024SW1--3 on the ATP Console CPU Board A3024--XX is in the ON position.

ACTIONS
Rapid Termination is a Safety device that cuts the X-ray exposure in case of an error with the
selected Ion Chamber or the selected parameters (short backup time) are not appropriate for an
exposure with AEC.
AEC Rapid Termination compares the AEC ramp with a 25% of the final value at the 30% of the
Backup Time. It is activated after 30% of the exposure back-up time and after 10 ms of exposure,
both conditions have to be fulfilled.
For a proper operation of the Rapid Termination feature, the operator must select an exposure
back-up time higher or equal to 100 ms whenever the AEC is ON.

1. Press the respective button on the Console to reset the Error indication.

2. Check the Collimator blades are properly open and no object is blocking the X-ray beam.

3. Increase the backup Time / mAs.

4. Select appropriate Ion Chamber, check the Ion Chamber configured for that Workstation is the
same as one physically connected (a.e. IC1 connected to J1 at the AEC Adaptation Board).

5. Check that 3024SW1-3 at the ATP Console CPU Board (A3024--XX) is OFF.

4-188
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ERROR CODE : E98

DESCRIPTION : DIP Switch 3024SW2-3 on ATP Console Board set for Configuration and Calibration Mode
Active.
ERROR TYPE : Informative. It allows normal operation.
APPLICABLE TO : All Generators.
APPEARS WHEN : After turning on the Control Console.
INFORMATION / SYMPTOM : Error Code “E98” is displayed.

POSSIBLE CAUSES
DIP Switch 3024SW2-3 on the ATP Console CPU Board (A3024--xx) set in the ON position.

ACTIONS
1. Reset the error condition by pressing the respective button on the Console.

Note . Keep in mind that this error will appear each time the Generator is turned OFF/ON
during service procedures (configuration, calibration, etc.) whenever Dip-switch
3024SW2-3 on ATP Console CPU Board is in “ON” (closed) position (for Service
Mode allowed).

2. When servicing is finished and the Generator is ready for normal operation, turn the Generator
OFF and set DIP Switch 3024SW2-3 on ATP Console CPU Board in the OFF (open) position
(operation mode).

4-189
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

4-190
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

CHAPTER 5
JOB CARDS

5-1
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

5-2
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 1 JOB CARDS FOR ADJUSTMENT /


REPLACEMENT PROCEDURES

Whenever the HV Transformer, X-ray Tube, ATP Console CPU


Board, HT Controller Board, or Filament Control Board is
replaced in the Generator, the respective Configuration and
Calibration procedures must be performed.

If the HT Controller Board or the ATP Console CPU Board are


replaced, check that the Extended Memory data have not
been lost or modified with the Board change. Compare
Extended Memory data displayed on the Console with the
values noted in the Data Book.

Revision of Boards should be checked, the new Board must


be compatible with the replaced one, that is, the same o
higher revision than the replaced Board. The EEPROM
information has to match the Generator options.

If the HT Controller Board is replaced in the Power Module,


transfer U3-EEPROM from the old Board to the new Board.
U3-EEPROM contains calibration data.

If U3-EEPROM is not transferred, a complete Calibration must


be performed.

Make various exposures, using different techniques and


Focal Spots, to verify that mA stations are calibrated
correctly. If they are not, perform Calibration procedures.

Update and record in the Data Book any new data entered in
the Extended Memory Locations.

Use the Anti-static Protection Device for handling when


replacing any board or electronic part.

5-3
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES (LEDS) ON THE CHARGE-DISCHARGE MONITOR
BOARDS ARE OFF (APPROXIMATELY THREE (3) MINUTES
AFTER THE UNIT IS TURNED OFF).

TO AVOID ELECTRIC SHOCK, DO NOT TOUCH ANY


HEATSINK OF THE CIRCUIT BOARDS EVEN WITH THE
GENERATOR TURNED OFF. PREVIOUS TO DISASSEMBLE
ANY BOARD, REMOVE ALL CONNECTORS PLUGGED TO IT.

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

DO NOT SUPPLY THE MAIN POWER FROM THE ELECTRICAL


ROOM CABINET (MAIN DISCONNECT) OR TURN ON THE
SYSTEM UNTIL SPECIFICALLY INSTRUCTED IN THIS
DOCUMENT.

ALWAYS HAVE THE “IPM DRIVER BOARD” CONNECTED IN


THE GENERATOR PREVIOUS TO ACTIVATING MAINS
POWER. IF THE “IPM DRIVER BOARD” IS NOT CONNECTED,
IRREVERSIBLE DAMAGE WILL OCCUR TO THE IGBTS.

THIS GENERATOR IS PERMANENTLY CONNECTED TO THE


POWER LINE AND POWERED ON UNLESS THE SAFETY
SWITCH INSTALLED IN THE ROOM ELECTRICAL CABINET
IS OFF. WHEN THE GENERATOR IS POWERED ON, THE
NEON LAMP (GREEN) LOCATED ON THE TRANSFORMER
6T2 (GENERATOR CABINET) IS ON.

INTERNAL PARTS OF THE GENERATOR (ALL FUSES, LINE


CONTACTOR (6K5), INPUT TRANSFORMER (6T2), ON/OFF
RELAY (3K3) AND LF-RAC MODULE) RECEIVE
CONTINUOUS POWER THROUGH THE POWER LINE
ALTHOUGH THE CONTROL CONSOLE IS OFF. BE SURE
THAT THE SAFETY SWITCH IS OFF BEFORE HANDLING ANY
INTERNAL PART OF THE EQUIPMENT.

5-4
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD LIST

JOB CARD DESCRIPTION

REPLACEMENTS

1-001 ATP Console CPU Board (A3024-221)

1-002 Memory EPROM-U24 on the ATP Console CPU Board (A3024-221)

1-003 AEC Control Board (A3012-25)

1-004 HT Controller Board (A3000-80/81)

1-005 Microcontroller+ROM U5 on the HT Controller Board (A3000-80/81)

1-006 Filament Driver Board (A3004-10)

1-007 Interface Control Board (A3009-11)

1-008 Power Supply

1-009 Inverter Module

1-010 Input Rectifier (A3255-03)

1-011 Rectifiers on the Inverter Module

1-012 Main Storage Capacitors

1-013 Charge/Discharge Monitor Board (A3212-01)

1-014 IPM Driver Board (A3063-06)

1-015 IGBT’s in the Inverter Module

1-016 Low Speed -- LF--RAC Board (A3096-02)

1-017 High Speed -- Control DRAC Board (A3243-04)

1-018 High Speed -- Interface DRAC--HF (A3240-05)

1-019 High Speed -- Clamping Board (A3109-01)

1-020 High Voltage Tank -- One Tube

1-021 AEC Adaptation Board (A3263-03)

1-022 Locks Board (A3214-02)

ADJUSTMENTS

2-001 AEC Calibration

5-5
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

5-6
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-001: REPLACEMENT OF THE ATP CONSOLE CPU BOARD

SUBASSEMBLY : ATP CONSOLE CPU BOARD A3024-221

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

SW1 SW3 SW2 Board SW4


JP13 Reference
JP4 J6
J13 J1 J2 J3 J5 J7 J8
J4

JP5

JP7, 8, 21, & 22

J9

J14 J10, J11, J16 J12 J15

U18 AEC CONTROL PCB AREA


U23 U24

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-7
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Note the actual setting of all jumpers and switches on the board. We
recommend using the table on the following page for this task. Note the
label on EPROM U24 ( Console Program ).

2. Remove Memories U18, U23 & U24.

3. Remove all board connectors. Do not forget to note their position.

4. Disconnect and remove the AEC Control Board if this option is present.

5. Remove the five Allen screws used to secure the board.

6. Using the anti-static protection device, replace the old board with the new
one. Install the previously removed Memories U18, U23 and U24.

7. Replace the five Allen screws and all previously removed connectors.

8. Set all jumpers and switches to their original positions.

9. Check that RAM U23 and EPROM U18 (APR data) are present. Also
check if EEPROM U24 is on the new board; if not, remove the old one
and put it on the new board (Do not forget to use the anti-static
protection).

10. Power ON the Electrical Cabinet and then the Generator. Workstation
Configuration must be checked for loss or modification. See
Configuration Chapter in the Service Manual.

11. Make a functional check of the Equipment.

12. Note on the Data Book the cause and the date of the replacement and
fix an adhesive label beside the new board indicating the date and name
of the field engineer in order to give maximum information.

Note . If the error persists after having changed the Board, replace the old
components U23 & U18 with the ones supplied with the new
Board. Then reset and reconfigure the Workstation following the
Configuration Chapter in the Service Manual.

5-8
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Note . If Error 10 (E10) appears during startup, follow the steps below:
1. Shut down the system
2. Activate Service Mode by placing SW2-3 to the “ON” position
3. Turn ON the system
4. When E10 appears, reset the error
5. Ensure a correct Workstation Configuration
6. Shut down and exit Service mode by placing SW2-3 to “OFF”

Switch Dipswitch SW1


Position 1 2 3 4
ON
OFF

Switch Dipswitch SW2


Position 1 2 3 4
ON
OFF

Switch Dipswitch SW3


Position 1 2 3 4 5 6 7 8
ON
OFF

Switch Dipswitch SW4


Position 1 2 3 4
ON
OFF

5-9
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-002: REPLACEMENT OF THE MEMORY EPROM-U24 ON THE


ATP CONSOLE CPU BOARD

SUBASSEMBLY : MEMORY EPROM-U24 ON THE ATP CONSOLE CPU BOARD A3024-221

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

U24

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-10
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Check that the new EPROM--U24 has “(U24)” written on its label.

2. Check that the software code written in the EPROM-U24 label is correct.

3. Replace the old Memory EPROM--U24 on the ATP Console CPU Board
with the new one. Insert it in the correct position.

4. Set the DIP switch A3024SW2--3 of the ATP Console CPU Board to the
ON position. This enables the Service Mode.

5. Turn the generator ON. Error Code E10 will be displayed a few seconds
after the unit start-up. This is shown because the EPROM-U24 has been
replaced.

6. Reset the Error Code indication from the Console by pressing and
holding the Error Reset button until Error Code E10 disappears. After this
the normal start-up will be completed.

7. Turn the generator OFF and set DIP switch A3024SW2--3 of the ATP
Console CPU Board to the OFF position. This enables the normal
Application Mode. This will disable the operator from entering the Service
Mode.

5-11
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-003: REPLACEMENT OF THE AEC CONTROL BOARD

SUBASSEMBLY : AEC CONTROL BOARD (A3012-25)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

R10 R13 R12

R11

JP4
R14

JP3
JP2

JP1

J1
see reverse

Version and
Revision state

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-12
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

The AEC Control Board is always mounted directly onto the ATP Console
whether in the Generator Cabinet or the Control Console. For ease of service,
it is recommended to remove the ATP Console.

1. Note the positions of JP1, JP2, JP3 and JP4 on the table below.

JUMPER A B C
JP1
JP2
JP3
JP4

2. Remove the Allen screw used to secure the AEC Control Board to the
ATP CPU Board, holding the nut in place to avoid stripping.

3. Disconnect the flat cable from the AEC Control board. Then disconnect
the pin connector of the Control Board, located at the backside of the
board, by carefully lifting it perpendicular from the ATP Console.

4. Replace the old board with the new one. Connector J1 on the ATP
Console may have more connections than the AEC Control. Ensure that
pin 1 (J1) on the AEC Control lines up perfectly with pin 1 (J12) on the
ATP Console if the size or number of connection points are not the same.

5. Replace the Allen screw, do not forget to replace the spacer

6. Ensure that the jumpers and resistance on the replacement board match
the configuration of the original board (see step 1).

7. Power on the electrical cabinet and then the system.

8. Make a functional check of the AEC by testing the Ion Chambers in the
system. If necessary, reconfigure the AEC system following the
instructions laid out in the Configuration Chapter (AEC Configuration) of
the Service Manual.

9. Make a functional check of the equipment.

10. Note in the Data Book the cause and the date of the replacement and fix
an adhesive label beside the new board indicating the date and name of
field engineer in order to give maximum information.

5-13
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-004: REPLACEMENT OF THE HT CONTROLLER BOARD

SUBASSEMBLY : HT CONTROLLER BOARD A3000--80/81

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

TP2(+5V)
TP3(+12V) TP5 (mA)
R49 (Pot) TP4(--12V) TP7 (kV)
Code and Revision J3 J2 J4

GND GND

Potentiometer
U23-5 R29 (Pot) Adjustment
Data
U23

U19 SW2

U3 J1

U19-5 J5

GND
GND

U5

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-14
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. With the Generator ON and using an Oscilloscope, measure and note the
period of the signals on U19-5.

2. Note the setting for Dipswitch SW2. The next table will help you with this task.
Ensure that the new HT Controller is configured for the system (see note
below).

Switch Dipswitch SW2


Position 1 2 3 4 5 6 7 8
ON
OFF

3. Remove all connectors from the board and note their position.

4. Remove the four Allen screws (one metallic and three nylon) used to
secure the board to the Cabinet.

5. Using the anti-static protection device, replace the old HT Controller


Board with the new one and install it using the Allen screws removed
before.

6. Replace the four Allen screws.

7. Replace all the previously removed connectors. (see Step 3.)

8. Set the switches on SW2 to their original position. (see Step 2.)

DO NOT TURN ON THE SYSTEM UNLESS EXPLICITLY TOLD


TO DO SO IN THESE INSTRUCTIONS.

5-15
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

9. Ensure that EEPROM U3 (Calibration data) and Microprocessor U5


(Cabinet Program) are present on the new board; remove both
components and replace them with the old ones on the new board (do
not forget to use the anti-static protection). U5 is not normally supplied
because it is individual for each system. Make sure that pin 1 on U3 is
kept in place when replacing the EEPROM.

Pin 1 marked on the EEPROM

10. Set SW2--2 to ON and Turn the equipment ON.

Note . R29 and R49 are used to adjust the modulator period for the
Inverter and Filament, U19-5 and U23-5 respectively. There is a
label on the board (shown in photo as “Potentiometer Adjustment
Data”) that displays factory-set inverter periods in seconds for
R29 & R49.

11. Visually check the replaced (old) Board Potentiometer Adjustments,


marked on the label (somewhere on the board). If one or both modulator
period values are missing, refer to the original Final Test Documentation
that came with the system, or refer to next Table-1 as a guideline only.

12. Measuring with an Oscilloscope at U23-5, the period of the signal must
be adjusted with Potentiometer R49 if it does not match the period
detailed in Table-1.

5-16
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

13. Ensure that U19-5 has the same period as measured on the replaced
HT Controller in step 1.; adjusting R29 as necessary. Table-1 details the
possible periods that U19-5 may present.

Table-1
Pot. Adjustment Data
Component -- Pin System Configuration Period
Line Powered Generator
36s
U19-5 (Default Configuration)
Heavy Duty Cycle Inverters (Triple Phase Generators) 45s -- 48s *
Filament Driver Board 3004-xx
U23-5 80s
10/11/12 and higher
-- During the final system test, certain U19 components are set to 45 or 48s. As this depends on the
system and final site, contact the manufacturer in absence of the correct indication for this value.

Note . HT Controllers with revision K and lower cannot be adjusted to an


80s filament period unless Resistor R46 (4K99 ) is installed.

14. Turn OFF the System. Set SW2--2 back to the OFF position.

15. Perform Rotor Programming as detailed in the Configuration Chapter.

16. Power ON the Electrical Cabinet and the System. Check that the
Extended Memory Data have not been lost or modified. Use data noted
in the Data Book during installation. Ensure a proper X--ray Tube
selection. In case of doubts and for additional information on carrying out
Configuration, check the Configuration Chapter in the Service Manual.

17. Carry out the procedures as detailed in the Preventive Maintenance


Chapter of the Service Manual (Test for kV Loop, Test for Digital mA Loop
Open and Test for Digital mA Loop Closed).

18. Note in the Data Book the cause and the date of the replacement and fix
an adhesive label with the Board Potentiometer Adjustments beside the
new board including the date and name of the service engineer.

5-17
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-005: REPLACEMENT OF THE MICROCONTROLLER+ROM U5


ON THE HT CONTROLLER BOARD

SUBASSEMBLY : MICROCONTROLLER+ROM U5 ON THE HT CONTROLLER BOARD


(A3000--80/81)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

U5

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-18
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Check that the new Microcontroller+ROM U5 has “(U5)” written on its


label.

2. Replace the old Microcontroller+ROM U5 on the HT Controller Board


with the new one. Insert it in the correct position.

5-19
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-006: REPLACEMENT OF THE FILAMENT DRIVER BOARD

SUBASSEMBLY : FILAMENT DRIVER BOARD A3004--10

TOOLS : Standard Service Tool Kit, Anti-static Protection Device, Digital mAs Meter
and Oscilloscope.

PROCEDURE

Revision
Information
and W1

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-20
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Note the actual settings of jumper W1 on the board and remove all
connectors of the Board to be replaced and note their position.

2. Remove the four Allen screws used to secure the board to the cabinet.

3. Mount the Replacement Board, securing it with the four Allen screws.

4. Replace all previously removed connectors.

5. Ensure that W1 is in its original position (see Step 1)

6. Power ON the Electrical Cabinet and Systems.

7. Connect the mAs meter on the HV Tank and place a non--invasive meter
on the table under the X-ray beam.

8. To confirm correct parameters, make several exposures with different


technique settings, carrying out a full calibration if not correct.

9. Remove the Test Equipment and note the new memory data.

10. Note in the Data Book the cause and date of replacement and fix an
adhesive label beside the new board indicating the date and name of the
field engineer in order to give maximum info.

5-21
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-007: REPLACEMENT OF THE INTERFACE CONTROL BOARD

SUBASSEMBLY : INTERFACE CONTROL BOARD A3009--11

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

W1
W2
W6

W7
ZONE
A
W5

ZONE
W4 B
W3

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-22
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

The actual components in both zone A and B are detailed in Schematics


A3009-xx in the Service Manual. Ensure that the replacement board is properly
configured for the actual system type before continuing with the replacement.

COMPONENTS COMPONENTS
VERSION SYSTEM TYPE
ZONE A ZONE B
A3009-11 RAD ONLY NOT REQUIRED NOT REQUIRED

1. Note the actual setting of jumpers on the board.

2. Remove all connectors of the board. Note their position.

3. Remove the four Allen screws used to secure the board to the Cabinet.

4. Replace the old board with the new one.

5. Replace the four Allen screws.

6. Replace all connectors previously removed.

7. Set the jumpers in their original positions (see Step 1)

8. Power ON the Electrical Cabinet and Console.

9. Connect the digital Voltmeter on TB1-9 and TB1-1 (GND) and adjust Pot
R29 to obtain the required Voltage for the Ion Chamber in use. Certain
Ion Chambers require 230 VDC while others require 300 VDC (see Ion
Chamber specifications for more information). Do not forget to remove
the Voltmeter when finished.

10. Make several exposures with different AEC technique settings in order
to check if the density is correct compared with previous results.

11. If density values are not correct, perform a full AEC calibration and note
the new memory data.

12. Note in the Data Book the cause and date of the replacement and fix an
adhesive label beside the new board indicating the date and name of the
field engineer in order to give maximum info.

5-23
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-008: REPLACEMENT OF THE POWER SUPPLY

SUBASSEMBLY : POWER SUPPLY

TOOLS : Standard Service Tool Kit, Anti-static Protection Device, and Digital
Multimeter.

PROCEDURE

Soldered
Connections
Plastic Cover Mount

Soldered
Connections

+12V adjust --12V adjust +5V adjust

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-24
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Remove the two (2) screws from the plastic cover mount and take off the
plastic cover.

2. Note and label the positions of the wires soldered to the Board,
Transformer, and Fuse Holder. Unsolder these wires.

3. Remove the Allen screws used to secure the power supply to the cabinet.

4. Replace the old power module with the new one and secure it with the
Allen screws removed before.

5. Solder all the wires previously unsoldered to their respective points.

6. Remove the fuse from the old power supply and insert it into the new
power supply.

7. Turn the system ON.

8. With a Digital Multimeter measure between:

G P2-3 (+) and P2-4 on the HT Controller Board. Check that the
voltage at this point is +5 (±0.2 VDC). If required, adjust the
voltage with the +5 VDC Adjustment Potentiometer on the Power
Supply Board.

G P2-2 (+) and P2-4 on the HT Controller Board. Check that the
voltage at this point is +12 (±0.1 VDC). If required, adjust the
voltage with the +12 VDC Adjustment Potentiometer on the Power
Supply Board.

G P2-1 (--) and P2-4 on the HT Controller Board. Check that the
voltage at this point is --12 (±0.1 VDC). If required, adjust the
voltage with the --12 VDC Adjustment Potentiometer on the Power
Supply Board.

9. Note in the Data Book the cause and date of replacement and fix an
adhesive label beside the new board indicating the date and name of
field engineer in order to give maximum info.

5-25
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-009: REPLACEMENT OF THE INVERTER MODULE

SUBASSEMBLY : INVERTER MODULE

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

Charge / Discharge Boards

IGBTs

Rectifiers
Input Rectifier Board

Seen from the top

GND

Input Rectifier Board Power Cables OUT

Power Cables IN +VDC


--VDC

Seen from the rear

5-26
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN
Note . The code number of the Inverter Module depends on the
Generator Model. Check labels for its identification. images in this
Job Card may vary depending on the Inverter Module model.

5-27
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

1. Remove the Generator Cabinet Cover and place the Generator where
the rear panel can be accessed.

2. Remove Generator Rear Cover and disconnect its GND cable.

3. Remove the two (2) screws of the Generator Front Door and open it.

4. Remove the protective Plastic Safety Covers from the front of the Module
inside the Generator.

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE- DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF. CHECK THAT THE MEASURED VOLTAGE
BETWEEN POSITIVE AND NEGATIVE DC VOLTAGE WIRES
ON THE INPUT RECTIFIER BOARD IS < 10 VDC.

LED on Chard--Discharge Monitor Boards

+VDC --VDC

5-28
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Remove the four (4) screws that secure the LF-RAC or LV-DRAC module
tray to the Generator Chassis and take out the module.

Inverter Module

Front Panel

LV DRAC Module

6. Place a white cloth below the inverter to easily find any screw, nut, spacer
or washer that could drop down to the floor during the replacement.

7. Label and disconnect the following connectors / wires:


(refer next illustration)

If the system is equipped with a High Speed Starter (LV-DRAC):


G Tube Stator wires connected to 11TS2.
G GND wire connected to the LV--DRAC Module.
G Aerial Connector J7.
G Connector J4 at the Control DRAC Board.
G Wires J2--1 and J2--2 at the Interface DRAC Board.

If the system is equipped with a Low Speed Starter (LF-RAC):


G Tube Stator wires connected to 10TS2.
G GND wire connected to the LF--RAC Module.
G Aerial Connectors J7 y J8.

5-29
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

J7

Stops of
Tube Stator wires
LV--DRAC
connected to 11TS2
Module

J2--1 & J2--2

GND Wire J4
LV-DRAC Module

Tube Stator wires connected to 10TS2,


GND Wire
J7 and J8 aerial connectors

Stops of
LF--RAC
Module

LF-RAC Module

5-30
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

8. Remove the two (2) Stops at the rear side of the LV-DRAC Module or
LF-RAC Module and then remove the Module from the Cabinet.

9. Disconnect the Cables from the IPM Drivers (J1 & J2) and the Charge /
Discharge Boards (P1 & P2).

P1 & P2

300 V

J1 & J2

Cable to Starter (1&2)

10. Label and disconnect the Cables from the HV Transformer (P1,
P2--Shield, and P3).

P2--Shield

P1

P3

5-31
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. At the back side of the Cabinet, carefully remove the four (4) Allen screws
at the Back Cover of the Generator, disconnect the GND wire and remove
the Back Cover. This action may be hampered due to the routing of the
Power Line Cables. If this is the case, disconnect the Power Line Cables
from the Input Line Fuses.

12. From the back side of the Cabinet, disconnect the Power cables from the
Input Rectifier Board holding the hexagonal spacer in place while
unscrewing the connection.

Power Cables attached to


the Input Rectifier Board

Ground Cable

13. With a socket wrench, disconnect the GND (yellow and green wires)
attached to the chassis of the Generator from Input Rectifier Board, as
shown in the photo below.

5-32
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

14. If the Inverter Module is equipped with fans, disconnect the supply wires
of the fans from the terminal strip located at the back side of the
Generator. Due to the polarity of the supply wires of the fans, note their
positions in order to restore them properly when re-assembling of the
Inverter Module.

15. The most practical way of dismounting the Inverter Module from the
Generator is by first removing the two (2) lower screws and then
loosening the two (2) upper screws.

Allen Screws securing the Inverter Module

Upper Screws

Lower Screws

16. Hold the module with one (1) hand (to keep the module from falling and
potentially damaging the components underneath) while removing the
already loosened upper screws of the Inverter Module. Remove the
Inverter Module.

This module is heavy and must be removed through the front


of the cabinet.

17. Install the new Inverter Module, attaching the screws with the nuts
removed earlier.

5-33
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

18. Connect all the wires and connectors removed in the previous steps to
the Inverter Module.

MAKE SURE THAT ALL CONNECTIONS ARE NOT ONLY


SECURE BUT ALSO CORRECT.

19. Re-install the Protective Plastic Safety Cover at the front of the Inverter
Module.

20. Remove the white cloth that was placed below the Inverter in step 6.

21. If applicable, re-install the LV-DRAC Module or LF-RAC Module into the
Generator. Install the two (2) Stops at the rear side of the module. Insert
the module and secure it to the Generator Frame. Connect all the wires
and connectors removed in the previous steps in the starter module.

MAKE SURE THAT ALL CONNECTIONS ARE NOT ONLY


SECURE BUT ALSO CORRECT.

22. Close the Front Panel and secure it to the Generator Frame.

23. Switch ON the Generator and there should be no errors upon powering
up. To check that the Inverter Module replacement has been performed
successfully:

G Measure RMS Voltage between 2 AC wires and measure DC


voltage between DC bars and apply the following formula:

V(DC) = Vrms (Line ) x 1.41.

The result of the previous formula should be equal or greater than


the measured voltage between DC bars.

G Make a few exposures and verify that there are X-Rays to ensure
proper functioning.

24. Note in the Data Book the cause and date of the replacement in order to
give maximum info.
25. Re-install the Back Cover and Ground cable, and finally the Generator
Cover.

5-34
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-010: REPLACEMENT OF THE INPUT RECTIFIER

SUBASSEMBLY : INPUT RECTIFIER (A3255-03).

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

Components for
3 Phase Systems

Board Reference

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE- DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF. CHECK THAT THE MEASURED VOLTAGE
BETWEEN POSITIVE AND NEGATIVE DC VOLTAGE WIRES
ON THE INPUT RECTIFIER BOARD IS < 10 VDC.

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-35
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Remove the Generator Cover.

2. At the back side of the Cabinet, carefully remove the four (4) Allen screws
at the Back Cover of the Generator, disconnect the GND wire and remove
the Back Cover. This action may be hampered due to the routing of the
Power Line Cables. If this is the case, disconnect the Power Line Cables
from the Input Line Fuses.

3. From the back side of the Cabinet, disconnect the Power cables from the
Input Rectifier Board holding the hexagonal spacer in place while
unscrewing the connection.

Power Cables attached to


the Input Rectifier Board

Input Rectifier Board

Ground Cable

4. With a socket wrench, disconnect the GND (yellow and green wires)
attached to the chassis of the Generator from Input Rectifier Board, as
shown in the photo below.

5-36
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Replace the old board with the new one.

Before securing the new Input Rectifier Board make sure that
the three (3) washers are installed under the Board (between
the Rectifiers and the Board) at the Power Cable terminal
connections as indicated in the illustration below.

Washers between
Board and Rectifiers

6. Re-install the Ground Cable to the chassis of the Generator from Input
Rectifier Board.

7. Re-install the three (3) hexagon shaped-spacers and its corresponding


Power cables (3) from the Input Rectifier Board with nuts and washers
holding the hexagonal spacer in place while screwing the connection.

8. Note in the Data Book the cause and date of the replacement and fix an
adhesive label in the new board indicating the date and the name of the
field engineer in order to give maximum info.
9. Re-install the Back Cover of the Generator, connecting the GND wire to
the cover and finally the Generator Cover.

5-37
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-011: REPLACEMENT OF THE RECTIFIERS

SUBASSEMBLY : RECTIFIERS ON THE INVERTER MODULE.

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

Bar Codes

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE- DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF. CHECK THAT THE MEASURED VOLTAGE
BETWEEN POSITIVE AND NEGATIVE DC VOLTAGE WIRES
ON THE INPUT RECTIFIER BOARD IS < 10 VDC.

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

5-38
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN
Use the Anti-static Protection Device for handling or
replacing any board or electronic part.

5-39
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Remove the Input Rectifier Board as described in the Job Card 1-010.

2. Label and disconnect the DC BUS Wires (+) and (--) (Nuts and Washers).

3. Remove the four (4) Allen screws that fix the DC BUS Bars (+) and (--)
to the Rectifiers behind.

4. Remove the old Rectifiers and install the new ones. Each Rectifier is fixed
to the Module by four washers and two nuts. Note the orientation of the
Rectifiers.

Remove and replace the Rectifiers respecting the original


orientation (+/- ) in order to maintain the Polarity of the Unit.

Rectifiers

5. Install the DC BUS Bars with its corresponding Allen screws (4).

6. Connect the DC BUS wires (+) and (--) removed before.

5-40
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

7. Re-install the Input Rectifier Board, ensuring that its position matches its
original placement.

Before securing the new Input Rectifier Board make sure that
the three (3) washers are installed under the Board (between
the Rectifiers and the Board) at the Power Cable terminal
connections as indicated in the illustration below.

Washers between
Board and Rectifiers

8. Re-install the three (3) hexagon shaped-spacers and its corresponding


cables with nuts and washers.

9. Note in the Data Book the cause and date of the replacement in order to
give maximum info.

10. Follow with step 6. of the Job Card 1-010.

5-41
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-012: REPLACEMENT OF THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER

SUBASSEMBLY : MAIN STORAGE CAPACITORS.

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

Main Storage Capacitors


of the Inverter Module

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE- DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF. CHECK THAT THE MEASURED VOLTAGE
BETWEEN POSITIVE AND NEGATIVE DC VOLTAGE WIRES
ON THE INPUT RECTIFIER BOARD IS < 10 VDC.

5-42
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

1. Remove the Inverter Module as described in the Job Card 1-009 from
steps 1 to 16.

2. Position the Module face upward in order to access the screws that
secure each Storage Capacitor to the DC BUS. Remove the two (2)
screws that fix the Capacitor to be replaced to the DC BUS. If these
screws serve as connections for wires from the Charge / Discharge
Monitor Board, mark the respective wire(s).
Allen Screws
Mark these wires as needed

Allen Screws Tie Wraps

3. While holding the Capacitor to be replaced, cut the Tie Wrap that holds
the Capacitor to the Module.

5-43
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Place a new Tie Wrap through the holes provided in the Module.

5. Install the new Storage Capacitor, securing it to the DC BUS with the two
(2) Allen screws. Reconnect any wires removed from the Charge /
Discharge Monitor Board, if needed.

MAKE SURE THAT THE DC BUS WIRES (+) AND (- ) ARE


CONNECTED IN THE CORRECT SCREWS AND SECURED.

Allen Screws DC BUS (+) DC BUS (--)

6. Secure the new Storage Capacitor to the module by securing the Tie
Wrap.

7. Re-install the Inverter Module as described in Job Card 1-009 from step
17. to the end of the process.

8. Note in the Data Book the cause and date of the replacement in order to
give maximum info.

5-44
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-013: REPLACEMENT OF THE


CHARGE/DISCHARGE MONITOR BOARD

SUBASSEMBLY : CHARGE/DISCHARGE MONITOR BOARD (A3212-01).

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

DS1

P2

Board Reference P1

DC BUS Connection Wire (+) DC BUS Connection Wire (--)

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE- DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF. CHECK THAT THE MEASURED VOLTAGE
BETWEEN POSITIVE AND NEGATIVE DC VOLTAGE WIRES
ON THE INPUT RECTIFIER BOARD IS < 10 VDC.

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

5-45
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

1. Remove the Generator Cover.

2. Remove the Allen screws that secure the Front Panel to the Generator
Frame and open the Front Panel.

3. Remove the Protective Plastic Safety Cover from the front of the Inverter
Module.

4. In the old Charge / Discharge Monitor Board, label and remove all Board
connectors and note their position.

P1 and P2

DC BUS (+)

DC BUS (--)

5. Label and remove the DC BUS wires (+) and (--) connected to the Board
and note their position.

6. Remove the Allen screws used to secure the Board to the Inverter
Module.

7. Replace the old Board with the new one.

8. Re--install the Allen screws.

5-46
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

9. Re--install the DC BUS wires (+) and (--).

MAKE SURE THAT THE DC BUS WIRES (+) AND (- ) ARE


CONNECTED IN THE CORRECT SCREWS AND SECURED.

10. Turn the Generator ON.

11. Check for correct operation that:

a. LED DS1 is ON.

b. Measure the voltage in Resistor R4. It must be 0 VDC when the


equipment is ON. This should be more than 200 VDC right after
the equipment is Powered OFF.

c. Turn the Generator OFF and wait three (3) minutes for the Main
Storage Capacitors to discharge.

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES (LEDS) ON THE CHARGE- DISCHARGE MONITOR
BOARDS ARE OFF, APPROXIMATELY THREE (3) MINUTES
AFTER THE UNIT IS TURNED OFF.

d. LED DS1 is OFF.

e. Measure the voltage in Resistor R4, it must be 0 VDC or close to


0 VDC (less than 10 VDC).

f. Measure the voltage at DC BUS, it must be 0 VDC or close to 0


VDC (less than 10 VDC).

12. Make a functional check of the equipment.

13. Note in the Data Book the cause and date of the replacement and fix an
adhesive label beside the new board indicating the date and name of
field engineer in order to give maximum info.

14. Re-install the Protective Plastic Safety Cover at the front of the Inverter
Module.

15. Close the Front Panel and secure it to the Generator Frame.

16. Re-install the Generator Cover.

5-47
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-014: REPLACEMENT OF THE IPM DRIVER BOARDS

SUBASSEMBLY : IPM DRIVER BOARDS (A3063-06).

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

Board Reference

IPM DRIVER BOARD (A3063-06)

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE- DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF. CHECK THAT THE MEASURED VOLTAGE
BETWEEN POSITIVE AND NEGATIVE DC VOLTAGE WIRES
ON THE INPUT RECTIFIER BOARD IS < 10 VDC.

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

5-48
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

1. Remove the Generator Cover.

2. Remove the Allen screws that secure the Front Panel to the Generator
Frame and open the Front Panel.

3. Remove the Protective Plastic Safety Cover from the front of the Inverter
Module.

4. Disconnect all main connections to the boards as shown in the photo


below. Do not forget to note their positions.

5. Remove the two Standoffs (hexagonal spacers), used to secure the


Front Panel to the Cabinet, as shown in the photo above.

6. Prop the Inverter Module Front Panel in the position shown in the photo
below.

5-49
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

7. Remove the four screws used to fix the two IPM Boards to the Inverter
Module (2 screws on each board).

IGBT’s Connector Pins

Allen Screws

The IGBT’s Connector Pins inserted on the backside of the


IPM Boards are fragile. Be careful when removing and
installing the replacement board.

8. Remove the IPM Driver Board by pulling them gently and perpendicular
from the Module.

9. Replace the old Board with the new one. When installing the new IPM
Driver Board, ensure that the IGBT’s Connection Pins line up with the
connectors on the Board.

10. Replace all connectors, securing the connection for all affected boards.

11. Re-install the Protective Plastic Safety Cover at the front of the Inverter
Module.

12. Close the Front Panel and secure it to the Generator Frame.

13. Re-install the Generator Cover.

14. Power ON the Electrical Cabinet and then the System.

15. Check for a correct operation.

16. Make a functional check of the equipment.

17. Note in the Data Book the cause and date of the replacement and fix an
adhesive label of the new board indicating the date and name of the field
engineer in order to give maximum info.

5-50
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-015: REPLACEMENT OF THE IGBT

SUBASSEMBLY : IGBT‘S IN THE INVERTER MODULE.

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE- DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF. CHECK THAT THE MEASURED VOLTAGE
BETWEEN POSITIVE AND NEGATIVE DC VOLTAGE WIRES
ON THE INPUT RECTIFIER BOARD IS < 10 VDC.

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

5-51
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN
Use the Anti-static Protection Device for handling or
replacing any board or electronic part.

5-52
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Both IGBTs are located at the middle of the Inverter Module (at the back of the
Coil (L2), IPM Driver Boards, and DC Bus)

Coil (L2)

Charge / Discharge Boards

IPM Driver Boards

TWO (2) IGBTS LOCATED


UNDERNEATH

1. Remove the Generator Cover.

2. Remove the Allen screws that secure the Front Panel to the Generator
Frame and open the Front Panel.

3. Remove the Protective Plastic Safety Cover from the front of the Inverter
Module.

4. Label and disconnect the Cables from the IPM Drivers (J1 & J2) and the
Charge / Discharge Monitor Boards (P1 & P2).

P1 and P2

J1 J2

5-53
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Remove the Allen screws, then label and disconnect the DC BUS wires
from the Charge/Discharge Monitor Boards secured with round terminals
to the Main Storage Capacitors. Take note of the wire positions and mark
the respective wire if necessary.

Allen Screws DC BUS Wires

6. Label and remove the wire connections of the DC BUS / Coil at the upper
corner of the Coil (L2), then re--install the washers and nuts. Take note
of the wire positions and mark the respective wire if necessary.

7. Remove the two (2) Standoffs (hexagonal spacers) that secure the Plate
(with the Coil (L2) and the Charge / Discharge Monitor Boards). Then
remove this assembly (Plate, Coil and Charge / Discharge Monitor
Boards) and the two (2) Standoffs (hexagonal spacers) located under it.

DC BUS -- Coil
Wire Connection

Standoffs

5-54
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

8. Remove the two (2) Allen screws used to secure each IPM Driver Board
to the Inverter Module.

IGBT’s Connector Pins

Allen Screws

The IGBT’s Connector Pins inserted on the backside of the


IPM Boards are fragile. Be careful when removing the
replacement board.

9. Remove both IPM Driver Boards by pulling them gently and


perpendicular from the Module.

10. Remove all the Allen screws that secure the three (3) Capacitors (2 Micro
/ 1000 V) on the DC BUS. Then remove the three (3) Capacitors and the
wires connected to the DC BUS. Take note of the wire positions and mark
the respective wire if necessary.

DC BUS

Capacitors (2 Micro / 1000 V)

5-55
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

11. Remove the rest of the screws and the DC BUS (metallic rail) over both
IGBT.

IGBT

12. Label and disconnect the remaining connections to the IGBT and remove
the screws that fix the IGBT to the Heat Sink. Take note of the IGBT and
wire positions and mark the respective wire if necessary.

13. Install the new IGBT keeping a thin layer of semiconductor silicone paste
between the IGBT and the Heat Sink. Secure the IGBT to the Heat Sink
with the removed Allen Screws and connect to the IGBT the previously
removed wires.

14. Re--install the DC BUS (metallic rail) over both IGBT.

15. Re--install the three (3) Capacitors (2 Micro / 1000 V) and the removed
wires on the DC BUS.

DC BUS

Capacitors (2 Micro / 1000 V)

5-56
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

16. Re--install the IPM Driver Boards. Ensure that the IGBT’s Connection
Pins line up with the connectors on the Board.

The IGBT’s Connector Pins inserted on the backside of the


IPM Boards are fragile. Be careful when installing the Board.

17. Secure the IPM Driver Boards with the two (2) Allen screws.

IGBT’s Connector Pins

Allen Screws

18. Re--install and secure the assembly (Plate, Coil (L2) and Charge /
Discharge Monitor Boards) to the Inverter Module by placing the two (2)
Standoffs (hexagonal spacers) located under it and the two (2) Standoffs
(hexagonal spacers) over it.

19. Re--install the wire connections of the DC BUS / Coil at the upper corner
of the Coil (L2).

DC BUS -- Coil
Wire Connection

Standoffs

5-57
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

20. Connect the DC BUS wires from the Charge / Discharge Monitor Boards
secured with Round Terminals to the Main Storage Capacitors.

Allen Screws DC BUS Wires

21. Connect the Cables to the IPM Drivers (J1 & J2) and the Charge /
Discharge Monitor Boards (P1 & P2).

P1 and P2

J1 J2

22. Re-install the Protective Plastic Safety Cover at the front of the Inverter
Module.

23. Close the Generator Front Panel and re-install the Allen screws that
secure the Front Panel to the Generator Frame.

24. Re-install the Generator Cover.

25. Note in the Data Book the cause and date of the replacement in order to
give maximum info.

5-58
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-016: REPLACEMENT OF THE LOW SPEED BOARD

SUBASSEMBLY : LOW SPEED BOARD / LF RAC (A3096-02)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

Board Reference

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-59
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Remove all connectors and note their positions.

2. Remove the four Allen screws used to secure the board to the Cabinet.

3. Replace the old board with the new one.

4. Replace the four Allen screws.

5. Reconnect all previously removed connectors.

6. Power ON the Electrical Cabinet and the System.

7. Make a functional check for the acceleration, running and braking times,
visually checking the rotation of the anode through the window.

8. Make a functional check of the equipment.

9. Note in the Data Book the cause and date of the replacement and fix an
adhesive label beside the new board indicating the date and name of the
field engineer in order to give maximum info.

5-60
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-017: REPLACEMENT OF THE CONTROL DRAC BOARD

SUBASSEMBLY : CONTROL DRAC BOARD (A3243-04)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

SW1
SW2

J2 SW3

SW4
J1

J3

Board Reference U17

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-61
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Note the actual setting of jumpers on the board.

2. Remove all wires and connectors of the board and note their positions.

3. Remove the four Allen screws used to secure the board to the Cabinet.

4. Replace the old board with the new one.

5. Replace the four Allen screws.

6. Replace all wires and connectors previously removed.

7. Set jumpers to their original position (see Step 1).

8. Place the old U17 onto the new board.

9. Power on the Electrical Cabinet and System.

10. Make a functional check of acceleration, running and braking times


and visually check the rotation of the anode through the window.

11. Make a functional check of the equipment.

12. Note in the Data Book the cause and the date of the replacement and fix
an adhesive label beside the new board indicating the date and name of
the field engineer in order to give the maximum info.

5-62
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-018: REPLACEMENT OF THE INTERFACE DRAC-HT

SUBASSEMBLY : INTERFACE DRAC--HF (A3240-05)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

DS1
Board Reference

High DC Voltage IN

J2

MAKE SURE THAT THE MAIN STORAGE CAPACITORS OF


THE HIGH VOLTAGE INVERTER DO NOT CONTAIN ANY
RESIDUAL CHARGE. WAIT UNTIL THE LIGHT EMITTING
DIODES ON THE CHARGE- DISCHARGE MONITOR BOARDS
ARE OFF, APPROXIMATELY 3 MINUTES AFTER THE UNIT IS
TURNED OFF. CHECK THAT THE MEASURED VOLTAGE
BETWEEN POSITIVE AND NEGATIVE DC VOLTAGE WIRES
ON THE INPUT RECTIFIER BOARD IS < 10 VDC.

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

5-63
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

THE LED DS1 BEGINS TO DIM DURING DISCHARGE AND IS


OFF WHEN COMPLETELY DISCHARGED. NEVERTHELESS,
ENSURE THAT DC VOLTAGE IS NOT PRESENT AT J2-1 & J2-2
BEFORE CONTINUING WITH THE REASSEMBLY.

1. Label and remove all wires and connectors on the board.

2. Remove the four (4) Allen screws that secure the board to the cabinet.

3. Replace the old board with the new one and secure it to the chassis with
the four (4) Allen screws.

4. Replace all previously removed connectors. Ensure that connections line


up properly, the first and the last pin are numbered to aid this task.

5. Turn the electrical cabinet and the generator ON.

6. Make a functional check of the equipment, acceleration, running and


braking times with a visual check of the rotating anode.

7. Note in the Data Book the cause and the date of the replacement and fix
an adhesive label beside the new board indicating the date and name of
the field engineer in order to give the maximum info.

5-64
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-019: REPLACEMENT OF THE CLAMPING BOARD

SUBASSEMBLY : CLAMPING BOARD (A3109-01)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-65
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Label and remove all wires on the board.

2. Remove the four (4) screws that secure the board to the cabinet.

3. Replace the old board with the new board.

4. Replace the four (4) Allen screws.

5. Replace all wires previously removed.

6. Turn the electrical cabinet and the generator ON.

7. Make a functional check of the equipment.

8. Note in the Data Book the cause and date of the replacement and fix an
adhesive label beside the new board indicating the date and name of the
field engineer in order to give maximum info.

5-66
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-020: REPLACEMENT OF THE HIGH VOLTAGE TANK

SUBASSEMBLY : HIGH VOLTAGE TANK (ONE TUBE)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-67
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Label and remove all wires and cables from the HV Transformer. These
wires from the Inverter Module are P1, P3, and “Shield” (SHLD). The
ground wire (GND) is directly connected to the chassis.

2. Replace the old HV Transformer with the new one and replace all
connections previously removed in Step 1.

Note . Do not open the oil cover as the oil level does not need to be
checked in this type of HV Transformer.

3. Power the electrical cabinet and then the system ON.

4. Make a check with a few mA and kV settings and recalibrate the system
if necessary following the calibration procedures for kV Loop, Digital mA
Loop Closed and Digital mA Loop (both Focal Spots) detailed in the
Calibration Chapter of the Service Manual.

5. Return the faulty HV Transformer to the factory.

6. Note in the Data Book the cause and date of the replacement and fix an
adhesive label beside the new board indicating the date and name of the
field engineer in order to give the maximum info.

5-68
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-021: REPLACEMENT OF THE AEC ADAPTATION BOARD

SUBASSEMBLY : AEC ADAPTATION BOARD (A3263-03)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

JP10
JP9
Setup Jumpers

R8

R11
R5

R2

Board Reference

JP11
JP12

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

5-69
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Note the actual setting of the jumpers on the board. Their configuration
depends on the ion chamber setup with the system. The table below will
help you with this task. See Schematic A3263--03 in the Service Manual
for more information.

JUMPERS POSITION

JP3, JP4, JP7, JP8

JP1, JP2, JP5, JP6

JP13, JP14, JP15, JP16

JP9 (IC1)

JP10 (IC2)

JP11 (IC3)

JP12 (IC4)

2. Label and remove all connectors to the board.

3. Remove the four (4) Allen screws that secure the board to the cabinet.

4. Replace the old board with the new board.

5. Replace the four (4) Allen screws.

6. Replace all connectors previously removed.

7. Set jumpers to their original positions (see Step 1). In case jumpers 9 - 12
have come in different positions to those on the previous board, set them
to their necessary positions and ensure a proper ion chamber output
according to the table below. If the jumpers are in the same position,
proceed to the next step.

Ion chambers output has to be 0 VDC when there is no radiation


(no--Offset Adjustment). If an ion chamber output has an offset, it must
be adjusted to 0 VDC with the respective Potentiometer. The table below
will aid with any necessary adjustment.

JUMPERS POSITION
ION CHAMBER OUTPUT
JP9 (IC1) JP10 (IC2) JP11 (IC3) JP12 (IC4)
NO--OFFSET ADJUSTMENT A A A A
OFFSET ADJUSTMENT B B B B
TEST POINT AND POTENTIOMETER
TP1 - R11 TP2 - R8 TP4 - R2 TP12 - R5
(ONLY IF JUMPER IS IN “B” POSITION)

5-70
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

8. Power the electrical cabinet and generator ON.

9. Make several exposures with different AEC technique settings in order


to ensure a correct selection.

10. Make a functional check of the equipment.

11. Note in the Data Book the cause and date of the replacement and fix an
adhesive label beside the new board indicating the date and name of the
field engineer to give maximum info.

5-71
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 1-022: REPLACEMENT OF THE LOCKS BOARD

SUBASSEMBLY : LOCKS BOARD (A3214-02)

TOOLS : Standard Service Tool Kit and Anti-static Protection Device.

PROCEDURE

Board Reference

6A Fuse (F1)
10A Fuse (F2)

BEFORE REPLACING ANY PART BE SURE THAT THE


GENERATOR AND THE ELECTRICAL ROOM CABINET (MAIN
DISCONNECT) ARE POWERED OFF.

PERFORM A LOTO PROCEDURE.

ALSO, FOLLOW THE GENERAL CAUTIONS INDICATED AT


THE BEGINNING OF THIS DOCUMENT.

Use the Anti-static Protection Device for handling or


replacing any board or electronic part.

1. Remove all wires of the board and note their positions.

2. Remove the four screws and nuts with washers used to secure the board
to the Cabinet.

5-72
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN
3. Replace the old board with the new one.

5-73
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

4. Replace the four Allen screws.

5. Replace all wires previously removed.

6. Power ON the Electrical Cabinet and then the System.

7. Make a functional check of the equipment. Check that the Locks,


Collimator light, etc function correctly.

8. Note in the Data Book the cause and date of the replacement and fix an
adhesive label beside the new board indicating the date and name of the
field engineer in order to give maximum info.

5-74
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

JOB CARD 2-001: ADJUSTMENT: AEC CALIBRATION

SUBASSEMBLY : ION CHAMBER ASSEMBLY (TABLE OR WALL STAND)


TOOLS : Standard service engineers tool kit including Allen and Torx key sets, and a
Trimmer for Ion Chamber Pre-Amplifier (AEC tiny potentiometer).

Homogeneous Phantom of Aluminium with a purity of not less that 99% and
thickness of 21 mm for the Collimator Filter Holder (Part no. 2281087-2).

Homogeneous Phantoms of PMMA (Polymethylmethacrylate) with a surface


of 43 x 43 cm and different thickness (cm) to achieve all the different
thickness used in the AEC verification tests: 8, 10, 15, 20, 25 and 28 cm.

(Optional) Dosimeter, Unfors RaySafe Xi (or equivalent).

Adhesive Tape.

Anti-static Protection Device.

PERSONNEL : 1 Service Engineer

PROCEDURE

1 SCOPE / OVERVIEW

This procedure provides the calibration steps to achieve proper AEC


performance for Proteus XR/f Radiographic System, in order to achieve the
correct image at the lowest possible dose.

The term “Optimum Gray Level” refers to some measurement criteria and may
be either:

 a measured exposure in μGray, or

 a system calculation called EI (Exposure Index), or

 a system calculation called S-Value.

This procedure will use the EI measurement method because of variations in


back-scatter characteristics, detector conversion factor, and detector
sensitivity. The EI value is calculated within a 4 cm x 4 cm ROI (Region of
Interest), and should be checked using a raw image. Raw images must be
acquired with the “S-Value” exam tag (in the QC group). An exposure of 1 μGray
will produce an equivalent of 102 to 110 EI.

5-75
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

AEC performance is controlled by a hardware gain potentiometer in the


Pre-Amplifier of the Ion Chamber and several software parameters stored in the
Generator (Extended Memory Locations).

The following table indicates the Extended Memory Locations related to AEC
Calibration.

FUNCTION MEMORY LOCATION

AEC-1 CALIBRATION E04

AEC-1 TRACKING E08

AEC-2 CALIBRATION E09

AEC-2 TRACKING E10

AEC DENSITY SCALE E12

AEC Calibration is carried out using a specified test setup and a Homogeneous
Aluminum Phantom (21 mm thickness, 99% purity) installed in the Collimator
guides, to produce a testing exposure time between 50 ms and 300 ms.

The radiation signal detected by the sensors (also known as Ion Chambers) is
stepped up by an adjustable gain amplifier. Its output is passed to an AEC
control circuit with a comparator stage, which activates the AEC Stop command
when the signal exceeds the reference.

Therefore, the AEC Stop depends on the Signal (kV, mA, SID, Phantom, and
Gain Pot) compared to the Software Parameters.

Radiation Signal
Pre-Amp
Sensors AEC STOP
Gain Pot
L. C. R.
SW Parameters Reference
DAC

AEC Control Board

The AEC calibration procedure is carried out in 3 parts:

 The 1st part is to adjust the Optimum Gray Level at 70 kV.

 The 2nd part is to adjust kV-Tracking for the same Optimum Gray Level
at 55 kV and 90 kV
(¦ 20%).

 The 3rd part is to fine-tune and validate the AEC performance using the
imaging sub-system.

5-76
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

LIST OF AFFECTED PROTEUS XR/F ET OR ST SYSTEMS

CATALOG ITEM DESCRIPTION DETECTOR TYPE

B3900MX (US)
Proteus XRf Package with 14x17 Aero XE HQ
B3926XA (EU)

B3900MY (US)
Proteus XRf Package with 14x17 Aero LT S
B3926HZ (EU)

B3900MZ (US)
Proteus XRf Package with 17x17 Aero HQ HQ
B3926XB (EU)

B1976SP
B1976GF 50kW Single Phase Generator
B1976GA

B1976GD
65kW Three Phase Generator
B1976GG

B1976GH
80kW Three Phase Generator
B1976GE

SAFETY PRECAUTIONS

This procedure produces ionizing radiation. Refer to the


product documentation for additional safety advice.

PREREQUISITES

 The X-Ray Generator and imaging system are functional, i.e. exposures
can be taking through CS7 PC workstation.

 The X-Ray tube is calibrated, and the system alignment meets


specifications.

 Refer to the product documentation for additional calibration and


alignment advice. Note, if the XR beam energy (kV, mA) is mis-calibrated
or mis-aligned (SID), the AEC accuracy will be affected.

 The TechService Sedecal Application must be installed on the CS7


workstation.

 The service provider has been trained on the Proteus XR/f and is familiar
with operating the CS7 and TechService applications

5-77
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2 PROCEDURE

2.1 SETTINGS BEFORE THE AEC CALIBRATION

2.1.1 PREPARATION OF THE CS7 PC WORKSTATION

1. Verify that ”Exposure Index” appears in the annotations. If not, enable it


in the CS7 Utility screen.

Note: this is a user preference setting. Return the setting to its original
state if necessary.

System --> Utility --> User Tool Tab --> Screen Setting --> Overlay Setting
on Exposed Images (click on ”Text Annotation” and ”Exposure Index”
boxes)

Press “OK” --> “Save (YES)” --> “OK” --> “Close” --> “Close”

2. Enable “Exam Tag Copying” in the “CS7 Utility” screen .

System --> Utility --> Center Icon --> Settings --> Exposed Image Copying
/ Exchan...

OK” --> “Save (YES)” --> “OK” --> “Close” --> “Close” --> “Close”

5-78
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3. Verify that the S-Value exam tag (in the QA/QC group) is configured for
Table and Wall Stand exams. If not, modify it in the Edit Menu.

System --> Utility --> User Tool Tab --> Edit Exam Tag.

To do so:

(1) Select ”Property List” tab.


(2) Filter using Free Word “QC”.
(3) In the S Value row, click the “Priority Modality” field.
(4) Open the droplist and select “Upright- >Table- >Cassette” .

“OK” --> “OK” --> “Close” --> “Close”

2.1.2 PREPARATION OF THE X-RAY SYSTEM

1. From the CS7 Desktop, launch the TechService application on the CS7
PC Workstation.

a. Verify that the generator is in Service Mode


(ATP Console SW2--3 = ON).

b. Power ON the Proteus XR/f.


If prompted log on as:
user: maintenance
password: kmmg5678

5-79
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

c. Navigate to the Intermediate (blue bar) Screen:


System --> Utility --> Close --> Shut down --> OK

d. Click “Windows Desktop” and launch the “TechService” shortcut.

2. Enter in Manual Calibration --> Select WS1 --> mA UP --> until Exx
(Extended Memories).

3. Enter the following default values in the corresponding Memory


Locations of the Generator related to the AEC Calibration, Read/Store
the AEC parameters (refer to the Generator Service Manual Chapter 3,
Generator Service Mode for more information)

MEMORY LOCATION FUNCTION DEFAULT VALUE

E04 AEC-1 calibration Table Ion Chamber 70

E08 AEC-1 tracking (kVp Compensation) Table Ion Chamber 4

E09 AEC-2 calibration Wall Stand Ion Chamber 70

E10 AEC-2 tracking (kVp Compensation) Wall Stand Ion Chamber 4

E12 AEC Density Scale All Ion Chambers 25

5-80
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.1.3 SETUP THE X-RAY SYSTEM TEST CONDITIONS (TABLE OR WALL STAND)

1. Remove the Detector from the Cassette Tray.

2. Adjust the SID to 150 cm (130 cm for ST Table).

3. Remove the Anti-scatter Grid.

4. Install the Aluminum Phantom in the Collimator rails.

2.2 PROCEDURE OF AEC CALIBRATION

2.2.1 TUBE WARM-UP

1. Reboot the entire system from Sys Loader. It activates the custom
changes made before, re-establishes the Gen/PC communication and
restart the applications.

2. Login as maintenance.

3. Open a test patient and select exam tag S-Value with the following
parameters: 70 kV, 100 mAs, 200 mA, 500 ms.

4. Fully close the Collimator blades.

5. Select the Detector (1417 or 1717 as applicable).

6. Locate the “Copy Exam Tag” button at the bottom of the “Exam List”.
Click it repeatedly to create 2 duplicates of the settings above.

7. Perform a Tube Warm-up procedure taking 3 exposures, 15 seconds


apart. Be sure that no one will be exposed.

5-81
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.2.2 CALIBRATE THE AEC PRE-AMP MASTER GAIN POTENTIOMETER

1. Install a Detector in the Tray in the portrait mode and start a new Exam
for S-Value.

2. Adjust / confirm the AEC Calibration acquisition settings are correct:

(1) Exam tag (Wall Stand v. Table)

(2) Detector selection

(3) Bucky (Wall Stand v. Table)

(4) Technique: 70 kV, 50 mA, 400 ms

(5) AEC: Center / Medium Film-Screen combination

1 4
2

3. On the Collimator, fully open the blades.

4. Locate the “Copy Exam Tag” button at the bottom of the “Exam List”. Click
it repeatedly to create 6 duplicates of the settings above. Create more
copies if needed.

5-82
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

5. Make an exposure on the newly copied exams and note the EI value and
no error messages are displayed. The target Exposure Index (EI) for
optimized image quality depends on the Digital Detector type.

Target Exposure Index (EI)


AeroDR Detector Type
for optimized image quality

HQ 305 ¦ 6

S 415 ¦ 8

 If the measured EI value is reasonably close to the target EI, small


adjustments in the AEC Calibration parameter (E04 or E09) may be done
instead of adjusting the Gain Potentiometer of the Ion Chamber
Pre-Amplifier (no need to remove covers).

 If the measured EI value is incorrect, Access the Ion Chamber


Pre-Amplifier Box --> Tweak the tiny Gain Potentiometer (CCW
decreases the dose) --> Make an exposure and read the EI value -->
Repeat these steps as needed until within the specifications.

5-83
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.2.3 CALIBRATE THE KV TRACKING COMPENSATION

Calibrate the kV Tracking Parameter (E08 or E10) until within the specifications.

1. Make an exposure at 55 kV and other exposure at 90 kV.

2. Check and note the EI value at 55 kV and 90 kV, compared to 70 kV.


Variation shall be less than ¦ 15% with pre-set +4 value. If not, change
the value to +3.

Suspend the exam and refer to Section 2.1.2 and reopen the suspended
exams --> Perform --> re-check that collimator is open

TABLE WALL STAND

55 kV 70 kV 90 kV 55 kV 70 kV 90 kV

EI
(Exposure Index)

Fine-tune
Value in Memory Location E08 (Table) = __ Value in Memory Location E10 (Wall Stand) = __
kV Tracking

MEMORY LOCATION FUNCTION DEFAULT VALUE

E08 AEC-1 tracking (kVp Compensation) Table Ion Chamber 4 = (pre-set +4)

E10 AEC-2 tracking (kVp Compensation) Wall Stand Ion Chamber 4 = (pre-set +4)

 If panel is not detected, check the power.

 Copy the exams as needed.

5-84
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3 VERIFY AEC PERFORMANCE

3.1 KV TRACKING COMPENSATION TEST

Note . This checking is carried out using homogeneous Phantom of


PMMA (Polymethylmethacrylate) with a surface of 43 x 43 cm and
different thickness (cm) according to the exposure test to be
performed.

TEST CONDITIONS

1 Testing conditions on Table (AT / ET / ST):

2 ¡ Align Center-On-Center:
Lateral-Horizontal Arm, Table Center Detent.
Long-Laser to Cassette Tray Handle.
3
© Remove Aluminium Phantom and DAP.
Set Field Size = full open the Collimator blades.
4
¢ SID = 150 cm (130 cm for ST Table)
5
£ PMMA Phantom = 43 cm x 43 cm x thickness
according to the exposure test to be performed.

¤ Flat Panel Detector (FPD) installed.


Anti-scatter Grid removed.

Testing conditions on Wall Stand (AT / ET / ST):

¡ Align Center-On-Center:
Lateral-Vertical Arm, Wall Stand Receptor Center.
5
4 Long-Laser to Cassette Tray Handle.
3 2
© Remove Aluminium Phantom and DAP.
Set Field Size = full open the Collimator blades.
1
¢ SID = 150 cm

£ PMMA Phantom= 43 cm x 43 cm x thickness


6
according to the exposure test to be performed.

¤ Flat Panel Detector (FPD) installed.


Anti-scatter Grid removed.

¥ Cardboard Boxes or Polystyrene at a sufficient


height to support the PPMA Phantom stack
centered with the Tube X-Ray Beam, and with the
Tube located at a height above the Tabletop
of the Table.

5-85
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Select S-Value with exam tag click button.

2. Set the AEC for Center/Medium.

3. Set SID to 150 cm (130 cm for ST table), open Collimator blades to max
field size and remove the Aluminum flat field phantom.

4. Remove the Grid.

5. Make exposure as indicated and record the EI values.

6. mA values and PMMA thickness values are given for indication in order
to have an exposure time more than 50 ms and less than 300 ms.

EI value
PMMA Thickness
kV mA Time (ms)
(cm) Table Wall Stand

55 25 50 < x < 300 8

70 10 50 < x < 300 10

80 10 50 < x < 300 15

100 10 50 < x < 300 20

125 10 50 < x < 300 25

130 10 50 < x < 300 28

7. Verify the performance accuracy, ± 20%.

EI max -- EI min
x 100% =
EI max

If the test fails, check the following items:

G Requested system and exposure conditions described early in the


procedure.

G Tube calibration.

G Selected Film Screen combination.

G Tube warm-up.

5-86
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

3.2 ION CHAMBER CELL BALANCE TEST

Note . This test is carried out using homogeneous PMMA (Polymethyl


methacrylate), 43 x 43 x 15 cm, so that all 3 cells receive the same
amount of scatter radiation. As an alternative to PMMA, the Al
phantom in the collimator rails may be used (not recommended).
This procedure is used only to confirm that the balance of the 3
cells is correctly adjusted from factory.

TEST CONDITIONS

1 Testing conditions on Table (AT / ET / ST):

2 ¡ Align Center-On-Center:
Lateral-Horizontal Arm, Table Center Detent.
Long-Laser to Cassette Tray Handle.
3
© Remove Aluminium Phantom and DAP.
Set Field Size = base of PMMA Phantom stack.
4
¢ SID = Grid under test (100 / 150 cm)
5
£ PMMA Phantom = 43 x 43 x 15 cm

¤ Flat Panel Detector (FPD) installed.


Anti-scatter Grid in or out according to testing
procedure.

Testing conditions on Wall Stand (AT / ET / ST):

¡ Align Center-On-Center:
Lateral-Vertical Arm, Wall Stand Receptor Center.
5
4 Long-Laser to Cassette Tray Handle.
3 2
© Remove Aluminium Phantom and DAP.
Set Field Size = base of PMMA Phantom stack.
1
¢ SID = Grid under test (100 / 150 or 180 cm)

£ PMMA Phantom = 43 x 43 x 15 cm
6
¤ Flat Panel Detector (FPD) installed.
Anti-scatter Grid in or out according to testing
procedure.

¥ Cardboard Boxes or Polystyrene at a sufficient


height to support the PPMA Phantom stack
centered with the Tube X-Ray Beam, and with the
Tube located at a height above the Tabletop
of the Table.

5-87
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

1. Use the previous CS7 S-value exam tag with the same exposure
parameter: 70 kV, 50 mA, 1 s back-up Time.

2. Fully open the Collimator.

3. Make the following series of exposures and record the exposure time.

EXPOSURES NO GRID EXPOSURES WITH GRID


TABLE
(SID 100 cm) LEFT CENTER RIGHT LEFT CENTER RIGHT

Exposure #1 tL1 = tC1 = tR1 = tL1 = tC1 = tR1 =

Exposure #2 tL2 = tC2 = tR2 = tL2 = tC2 = tR2 =

Exposure #3 tL3 = tC3 = tR3 = tL3 = tC3 = tR3 =

Average value of the Exposure Time tL = tC = tR = tL = tC = tR =

Left vs center right vs center left vs center right vs center

% Difference

EXPOSURES NO GRID EXPOSURES WITH GRID


WALL STAND
(SID 100 cm) LEFT CENTER RIGHT LEFT CENTER RIGHT

Exposure #1 tL1 = tC1 = tR1 = tL1 = tC1 = tR1 =

Exposure #2 tL2 = tC2 = tR2 = tL2 = tC2 = tR2 =

Exposure #3 tL3 = tC3 = tR3 = tL3 = tC3 = tR3 =

Average value of the Exposure Time tL = tC = tR = tL = tC = tR =

Left vs center right vs center left vs center right vs center

% Difference

EXPOSURES NO GRID EXPOSURES WITH GRID


WALL STAND
(SID 150 cm or 180 cm) LEFT CENTER RIGHT LEFT CENTER RIGHT

Exposure #1 tL1 = tC1 = tR1 = tL1 = tC1 = tR1 =

Exposure #2 tL2 = tC2 = tR2 = tL2 = tC2 = tR2 =

Exposure #3 tL3 = tC3 = tR3 = tL3 = tC3 = tR3 =

Average value of the Exposure Time tL = tC = tR = tL = tC = tR =

Left vs center right vs center left vs center right vs center

% Difference

4. Calculate the average of the 3 exposure times for each cell. Verify
performance accuracy by comparing the left and right averages against
the center average. Tolerance ± ¦20%.

tL -- tC
% of difference between “Center” and “Left” Fields = x 100 = %
tC

tR -- tC
% of difference between “Center” and “Right” Fields = x 100 = %
tC

5-88
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

CHAPTER 6
PREVENTIVE MAINTENANCE

6-1
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

6-2
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 1 INTRODUCTION

The purpose of this Periodic Maintenance is to assure continued safe


performance of the X-ray Generator, to increase serviceability, to reduce the
costs (down time, repairs, etc.) and to assure the safety (personal risk).

The following checks and maintenance procedures, together with the


suggested intervals, are the manufacturer’s recommendation for the most
effective Periodic Maintenance schedule for this Generator.

Service tasks here described must be performed exclusively by service


personnel specifically trained on medical X-ray Generators.

The first Periodic Maintenance Service should be performed six (6) months
after installation, and the subsequent services every twelve (12) months.
Periodic Maintenance Service depends on the working load of the Generator
and X-ray Tube.

Note . Take note in the Data Book all the periodic maintenance services
carried out and the data changes made during any maintenance
service.

6-3
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

6-4
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

SECTION 2 PERIODIC MAINTENANCE PROCEDURES

When any major component, such as a X-ray Tube,


HV Transformer or major circuit board, is replaced in the
system, perform the respective Configuration and
Calibration procedures.

Update and take note in the Data Book any new data entered
in memory.

If the HT Control Board or the ATP Console CPU Board is


replaced, check specially that Extended Memory data have
not been lost or modified with the Board change. Compare
Extended Memory data with the values noted in the Data
Book.

Also, make some exposures using different techniques and


Focal Spot and check that mA stations are calibrated
correctly, if not perform Calibration procedures.

Before starting the periodic maintenance procedures, it is recommended to


make a test exposure using the same operating factors and conditions as a
typical exposure.

Perform the X-ray tube warm-up procedure if the tube has not been in use for
approximately one hour. (Refer to Operator Manual of the Console).

2.1 TEST EQUIPMENT

The tools and test equipment required to perform the Periodic Maintenance
Service are the same as specified in Chapter 1 “Installation” and Chapter 3
“Configuration and Calibration”.

6-5
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.2 GENERAL CAUTIONS

MAKE SURE THAT THE MAIN CAPACITORS OF THE HIGH


VOLTAGE INVERTER DO NOT CONTAIN ANY RESIDUAL
CHARGE. WAIT UNTIL THE LIGHT EMITTING DIODES ON
THE CHARGE-DISCHARGE MONITOR BOARDS ARE OFF,
APPROX. 3 MINUTES AFTER THE UNIT IS TURNED OFF.

ALWAYS HAVE THE “IPM DRIVER BOARD” CONNECTED IN


THE GENERATOR PREVIOUS TO MAINS POWER IS
ACTIVATED IN IT. IF THE “IPM DRIVER BOARD” IS NOT
CONNECTED, PERMANENT DAMAGE WILL OCCUR TO
IGBTS.

THIS GENERATOR IS PERMANENTLY CONNECTED TO THE


POWER LINE, AND POWERED ON UNLESS THE SAFETY
SWITCH INSTALLED IN THE ROOM ELECTRICAL CABINET
IS OFF. WHEN THE GENERATOR IS POWERED, THE NEON
LAMP (GREEN) LOCATED ON THE TRANSFORMER 6T2
(GENERATOR CABINET) IS ON.

INTERNAL PARTS OF THE GENERATOR (ALL FUSES, LINE


CONTACTOR (6K5), INPUT TRANSFORMER (6T2), ON/OFF
RELAY (3K3) AND LF-RAC MODULE) ARE PERMANENTLY
POWERED ON THROUGH POWER LINE ALTHOUGH THE
CONTROL CONSOLE IS OFF. BE SURE THAT THE SAFETY
SWITCH IS OFF BEFORE HANDLING ANY INTERNAL PART
OF THE EQUIPMENT.

6-6
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.3 GENERAL CLEANING

NEVER ATTEMPT TO CLEAN OR HANDLE ANY PART OF THE


GENERATOR WHEN IT IS TURNED ON. SWITCH OFF THE
GENERATOR MAIN DISCONNECT BEFORE CLEANING OR
INSPECTING.

DO NOT SUPPLY THE MAIN POWER UNTIL SPECIFICALLY


INSTRUCTED TO DO SO IN THIS DOCUMENT.

2.3.1 EXTERNAL SURFACES

Clean external covers and surfaces frequently, particularly if corroding


chemicals are present and specially parts in contact with the patient, with a soft
cloth moistened in warm water with mild soap solution. Rinse wipe with a soft
cloth moistened in clean water. Do not use cleaners or solvents of any kind.

Clean Console keyboard and displays with a cloth dampened in warm water.
Rinse wipe with a cloth dampened in clean water.

Also check painted surfaces for scratching and touch up as required.

2.3.2 INTERNAL CABINET CLEANING

MAKE SURE THAT THE MAIN CAPACITORS OF THE HIGH


VOLTAGE INVERTER DO NOT RETAIN ANY RESIDUAL
CHARGE. LED ON THE CHARGE-DISCHARGE MONITOR
BOARDS HAVE TO BE OFF.

Remove the external access cover from the Generator Cabinet.

Visually inspect all major components for dust or foreign items. Search carefully
to detect objects which might cause short circuits and for loose connections.

If excess dust is present, clean the interior of the Generator Cabinet using a dry
brush or vacuum cleaner. Make sure that the fans operate properly and the vent
holes of the cabinet are not obstructed.

6-7
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.4 CABLE CHECKS

CAREFULLY HANDLE ALL INTERNAL PARTS OF THE UNIT.

Check that all electrical connections are firm and secure and that all cable
clamps and strain reliefs are in place. Also check that connectors do not have
exposed wire-veins and check cable sheaths (cable cover) for wear and fraying.

Check that all cables are correctly routed.

2.4.1 GROUND CABLE CONNECTIONS

The central reference ground of the X-ray System and Generator is located at
the Generator Cabinet.

Check the ground lead interconnections continuity using a multimeter at its


lowest ohms range.

2.4.2 AC POWER SUPPLY IN X-RAY ROOM

Measure the value of AC power supply between all phases, neutral and ground.
Check that these values comply with the tolerances established at the original
installation.

6-8
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.5 CONSOLE CONDITION

Check the proper connection and condition of the cables connected to the
Console.

1. Turn the Generator ON.

2. Check the Handswitch condition. Verify that the Handswitch cable and
its connection are in good condition.

3. Select a radiographic technique and observe:

G Indicators of the selected workstation and Focal Spot.

G Technique parameters are displayed on the Console. Change


technique parameters and observe that changes are correctly
displayed.

G Select the parameters for an usual exposure. Press “Prep” and


verify that the “Ready” indicator is activated. Release “Prep” and
observe that the “Ready” indicator is deactivated.

G Make the exposure, and verify that radiographic exposure signal


sounds and the “Prep” and “X-ray On” indicators are activated
during the exposure.

4. If AEC is installed, select a technique with AEC and observe that the
indicators of the selected AEC controls are activated.

6-9
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.6 HV TRANSFORMER CONDITION

The HV Transformer contains “Shell Diala AX” oil. Check that there is not oil
leakage.

Make sure that:

 HV oil in the HV Cable terminals is clean and shows no evidence of


arcing.

 HV Cable terminal rings are tight.

2.7 X-RAY TUBE CONDITION

Make sure that:

 All parts are mechanically secure with no oil leaks.

 HV grease on the HV Cable terminals is clean and shows no evidence


of arcing.

 HV Cable terminal rings are tight.

6-10
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.8 RADIOGRAPHIC PARAMETERS

With the generator power OFF, connect:

G Non-invasive kV Meter to measure kV.

G mAs Meter to the banana plug connections on the HV Transformer


to measure mA or mAs (connect the mAs Meter for Digital mA
Loops calibration).

Note . Test points on the HT Controller PCB can also be used to monitor
the kV and mA readings but should not be used to calibrate the
unit. These test points must be checked with scope. (Refer to
Calibration chapter -- Section 2 “Calibration Procedures”, for test
points and scale factors).

2.8.1 TEST FOR kV LOOP

1. Verify that dip switch 3000SW2-2 on the HT Controller Board is in “Off”


position (enables Filament and Rotor Interlocks).

2. Turn the Generator ON and select the “Direct” (No Bucky) workstation in
one of the X-ray Tubes.

3. Select 80 kV, 200 mA (or the first mA station for Large Focus), 100 ms.
Make an exposure and note the kV at the end of the exposure.

4. Check that the kV value read on the kV Meter must be 80 ±1 kV.

If the kV value does not comply with the above value, perform the
respective Calibration procedures.

6-11
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.8.2 TEST FOR DIGITAL mA LOOP OPEN

1. Set the dip switch 3000SW2-4 on the HT Controller Board in “On”


position (Digital mA Loop Open / Filament Current Constant).

Note . Only for Generators with LF-RAC (LSS):

-- When the mA Loop is open (dip switch 3000SW2-4 in “On”), the


rotor runs for two minutes after release the handswitch
push-button from “Preparation” position.

-- When the mA Loop is closed (dip switch 3000SW2-4 in “Off”),


the Tube will brake after release the handswitch push-button from
“Preparation” position.

2. Enter in Manual Calibration selecting the “Direct” (No Bucky) workstation


of the corresponding X-ray Tube.

3. Select 80 kV and the following mA stations. Make an exposure and note


the mAs values read on the mAs Meter.

G Minimum mA for Small Focal Spot.

G Maximum mA for Small Focal Spot.

G Minimum mA for Large Focal Spot.

G Maximum mA for Large Focal Spot.

4. Check that the mAs values read on the mAs Meter must be the same mAs
displayed on the Console with a tolerance of ±6% mAs.

If the mAs values do not comply with the above values, perform the
respective Auto-Calibration procedures.

5. Repeat this test for the second X-ray Tube.

6. Turn the Generator OFF and set the dip switch 3000SW2-4 on the HT
Controller Board in “Off” position (Digital mA Loop Closed).

6-12
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

2.8.3 TEST FOR DIGITAL mA LOOP CLOSED

1. Turn the Generator ON and select the “Direct” (No Bucky) workstation.

2. Select the following parameters, make an exposure and note the mAs
values read on the mAs Meter.

G 80 kV, 100 ms, 50 mA.

G 80 kV, 100 ms, 200 mA.

3. Check that the mAs values read on the mAs Meter must be the same mAs
displayed on the Console with a tolerance of ±4% mAs.

4. If the mAs values do not comply with the above values, perform the
respective Auto-Calibration procedures.

5. Turn the Generator OFF.

2.9 AEC CHECKS

The AEC is calibrated to produce an Optimum Gray Level at 70 kV, and then
AEC kV Tracking Compensation is adjusted to produce the same density at
55 kV and 90 kV.

Verify the AEC performance as described in the Chapter 3 -- Section 6.6.3.

6-13
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

6-14
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

CHAPTER 7
RENEWAL PARTS

7-1
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

7-2
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

COMPACT GENERATOR -- ONE TUBE

ADAPTATION
MODULE

POWER INPUT
MODULE

CONTROL PANEL

HIGH VOLTAGE
TRANSFORMER

ANODE ROTOR STARTER INVERTER MODULE BACK PANEL


(LF-RAC o LV-DRAC)

7-3
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

A0 CONTROL PANEL, MOD. 3

SAT-A3000-80 for 1Phase is compatible


SAT-A3000-80 2343379 ??
with previous SAT-A3000-33P
A1 PCB HT C
PCB, Controller
t ll 1
SAT-A3000-81 for 3Phase is compatible
SAT-A3000-81 2343379 ??
with previous SAT-A3000-33P1

A2 U5 EPROM 1 SAT-SOFT-HT 2343381 See Generator serial number

A3 PCB, Interface Control 1 SAT-A3009-11 SAT-A3009-11

A4 PCB, Filament Driver 1 SAT-A3004-10 5212054

A5 Low Voltage Power Supply 1 SAT-53418001 2197216 PS1

A6 Rectifier Bridge FAGOR / 10 Amp. 3 SAT-53404002 2197217 BR1, BR2 & BR4

A7 Filter Line 1 SAT-50208001 2197218 LF3

A8 ON/OFF Relay 4PDT /110V 1 SAT-51401012 2197298 K3

A9 Capacitor ELE 2400 μF, 50V 2 SAT-54002007 2197300 C6, C7

A10 Fuses F2, F6, F7, F8, F9 A520291-01 A520291-01 See Fuse Kit (item I0)

A1 A3

A2

A10
A4
A8
A5

A6

A6 A10 A7 A9

7-4
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

A1 A2 A3

A4 A5 A6

A7 A8 A9

7-5
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

B0 BACK PANEL, MOD.4

B1 Delayed Switch Off PCB 1 SAT-A3274-01 2347305 Only used with High Speed Generator

B2 CPU

B2.1 PCB, ATP Console CPU 1 SAT-A3024-221 5255289

B2.2 U24 EPROM 1 SAT-SOFT-ATP 2197361 See Generator serial number

B2.3 PCB, AEC Control 1 SAT-A3012-25 5255290

B2.1

B1

B2.3

B2.2

7-6
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

B1

B2.1

B2.2

B2.3

7-7
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

C0 INVERTER MODULE, MOD. 5

Used in 3-Phase Generator


C1 Inverter 3-Phase, 400-480 V, Lo kW 1 SAT-A11700-56 5440952 50 kW and 65 kW at 400-480 V,
and 80 kW at 480 V

Used in 1-Phase
C1 Inverter 1-Phase, 220-240 V, Hi kW 1 SAT-A11700-61 5446682
Generator 50 kW

Inverter 3-Phase, 400-480 V, HI kW Used in 3-Phase Generator


C1 1 SAT-A11700-90 SAT-A11700-90
with Fans 80 kW at 400-440 V

C2 IGBT 300A / 1200V Low Power 2 SAT-53416038 5457602 Used in Inverter A11700-56

C2 IGBT 600A / 600V 2 SAT-53416039 5455277 Used in Inverter A11700-61

C2 IGBT 450A / 1200V 2 SAT-53416048 SAT-53416048 Used in Inverter A11700-90

C3 Capacitor 40 μF, 450 VDC 1 SAT-54001006 2406632 C9, for 220-240 VAC Gen.

C3 Capacitor 10 μF, 1000 VDC 1 SAT-54001011 2197283 C9, for 400-480 VAC Gen.

C4 Capacitor Elec.3700 μF, 450 VDC 4 SAT-54002011 2197284 C1, C2, C3 & C4

C5 Capacitor 2 μF, 1000 V 2 SAT-54008002 2197285 C10, C11

C6 PCB, IPM Driver 2 SAT-A3063-06 5759661

1 PCB for 1-Phase &


C7 PCB, Charge / Discharge Monitor 1/2 SAT-A3212-01 2197287
2 PCB for 3-Phase

C8 Input Rectifier Board 1 SAT-A3255-03 2197288 Includes 2 metallic Plates

CR1, CR2, CR3


C9 Rectifier SKKD 40F10 3/2 SAT-53404019 2197281
(only 2 for 1-Phase)

C10 Equipped Resistor 1 SAT-A3312-01 SAT-A3312-01

C11 Fan Micro Boxer 24 VDC PAPST 3 SAT-53103046 SAT-53103046 Used in Inverter A11700-90

7-8
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

C1

C7

C4

C6

C11

C10

C8
C3
C9
C2

C5

7-9
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

C2 C3 C4

C5 C6 C7

C8 C9

C10 C11

7-10
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

D0 POWER INPUT MODULE, MOD. 6

D1 Contactor 3-Phase / 24V / 50 Amp 2 SAT-51405016 5199153 K5 and K6

D2 Fuses F3, F4, F5, F12, F13 A520291-01 A520291-01 See Fuse Kit (item I0)

D3 Resistor 20 ohms /100 W / 10% 1 SAT-52707021 2197293 R1

D4.1 Transformer T2 1 SAT-50509029 2197294 T2 for 3-Phase 400-480 VAC

D4.2 Transformer T2 1 SAT-50509030 2406637 T2 for 1-Phase 220-240 VAC

D5 Green light Lamp 1 SAT-54203006 2347306 NL1

SAT-A6371-01 2197296 LF1 for 1-Phase or 3-Phase

LF1 for 1-Phase. Compatible with


SAT-A11769-03
D6 Input
p Filter 1 SAT-A6371-01

LF1 for 3-Phase. Compatible with


SAT-A11769-04
SAT-A6371-01

D2

D1
D4.1 / D4.2

D5
D3

D2

D1

D3
D6

D4.1 D4.2 D5 D6

7-11
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

E0 HIGH VOLTAGE TANK - HERMETICAL TANK (ALUMINIUM)

H V Tank (150 kV)/ 1Tube 220V 1 SAT-A6097-16 2406641

H V Tank (150 kV)/ 1Tube 400V 1 SAT-A6097-17 2406642

H V Tank (150 kV)/ 1Tube 480V 1 SAT-A6097-18 2406643

E0

7-12
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

F0 ANODE ROTOR STARTER

F1 LOW SPEED (LF-RAC), MOD. 10

F1.1 PCB, LF-RAC 1 SAT-A3096-02 2197222 This reference is not a renewal part

F1.2 Fuse F1 A520291-01 A520291-01 See Fuse Kit (item I0)

F1.3 Relay Solid State 1 SAT-51404008 2197223 K1

F1.4 Capacitor 30 μF, 370 V 1 SAT-54001002 2197306 C5-2

F1.5 Capacitor 15 μF, 450V 1 SAT-54001012 2197304 C5-1 Optional

F1.6 Filter Line 1 SAT-50208001 2197218 LF2

SAT-A6378-26 2343392 for 400 VAC, 3-Phase


HIGH SPEED STARTER
F2 1 SAT-A6378-27 2406650 for 480 VAC, 3-Phase
(LV-DRAC), MOD. 11
SAT-A6378-28 2406651 for 220 VAC, 1-Phase

F2.1 PCB, Control DRAC 1 SAT-A3243-04 2197309

F2.2 PCB, Interface DRAC with Flat Cable 1 SAT-A3240-05 2197310

F2.3 Fuses F1, F2, F3, F4, F1 Aerial A520291-01 A520291-01 See Fuse Kit (item I0)

F2.4 PCB, Clamping (A3) 1 SAT-A3109-01 2248293

A522001-01 2359601 for 400 VAC

F2.5 LV-DRAC Main & Aux Transformers 2 A522002-01 2406652 for 480 VAC

A522003-01 2406653 for 220 VAC

F2.6 DC BUS cable 1 SAT-A3237-02 2357935

Contactor 24 V / 15 A (2NO/2NC)
F2.7 3 SAT-K00087-01 2303891 ?? K2, K3, KT1
+ Aux

SAT-SOFT-LVDRAC
SAT-SOFT-220
220VAC

SAT-SOFT-LVDRAC
F2.8 U17 EPROM DRAC 1 SAT-SOFT-400
400VAC

SAT-SOFT-LVDRAC
SAT-SOFT-480
480VAC

F2.9 Fan Micro Boxer 24 VDC PAPST 3 SAT-53103046 SAT-53103046

7-13
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

F1.5 F1.6 F1.3 F1.4 F1.1 F1.2

F1.1 F1.3 F1.4

F1.5 F1.6

7-14
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

F2.9

F2.7

F2.1 F2.6 F2.4


F2.5 F2.8 F2.3 F2.3 F2.2

F2.1 F2.2 F2.4

F2.7
F2.5 F2.6

F2.9
F2.8

7-15
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

G0 ADAPTATION MODULE

G1 PCB, Locks 1 SAT-A3214-02 2197295

G1.1 Fuses F1, F2 A520291-01 A520291-01 See Fuse Kit (item I0)

G2 PCB, AEC Adaptation 1 SAT-A3263-03 2197333-2 For 4 Ion Chambers

See Adaptations / Options


(Item H0).
Adaptations / Options - Some Adaptations / Options can be
installed on the upper side of the Adaptation
Module.

G1 G2

G1.1 G1.1

7-16
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

H0 ADAPTATIONS / OPTIONS

H1 Radiation Meter PCB 1 SAT-A3170-01 5309468 Optional for internal Dosimetry

H2 PCB, Auto On /Off 1 SAT-A3179-09 SAT-A3179-02 For installation inside a PC

H3 PC Interface Box (ABS case)

For using with a PC.


H3 PC Interface Box 1 SAT-AS00519
Includes items H3.1 and H3.2

H3.1 PCB, PC Interface Box 1 SAT-A40042--02

H3.2 Hand Switch support 1 SAT-14653-01

Compatible with PC Interface Box


H4 Hand Switch equipped 1 SAT-A6805-16
SAT-AS00519

H5 PC Interface Box (Metal case)

For using with a PC.


H5 Interface Box PC 1 SAT-A6509-40 SAT-A6509-01
Includes items H5.1, H5.2 and H5.3.

H5.1 PCB Interface Box 1 SAT-A9532-01 SAT-A9532-01

H5.2 Push Button 2 SAT-50613050 SAT-50613050

H5.3 Green Lamp 1 SAT-A7262-01 SAT-A7262-01

Compatible with PC Interface Box


H6 Hand Switch equipped - Sub D Conn. 1 SAT-A3223-15 SAT-A3223-15 SAT-A6509-40.
Includes Item H6.1

H6.1 Hand Switch Support 1 SAT-7897-01 SAT-7897-01

7-17
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

H1 H2

H3 H3.1 H3.2 H4

H5.3 H5 H5.1 H6 H6.1

H5.2

7-18
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

The Fuse Kit is common for all the Generator


models and comprises:
Reference Qty
53801002 10
53801003 20
53801008 20
53801011 6
I0 FUSE KIT 1 A520291-01 A520291-01
53801012 10
53801015 10
53801018 6
53801019 10

All fuses installed in the unit are included in


the Fuse Kit.

Fuse 1.5 A /250V , SB SAT-53801012 SAT-53801012 Control Panel: F2

A10 Fuse 3 A / 250V , SB SAT-53801003 SAT-53801003 Control Panel: F6, F7, F8

Fuse 0.4 A / 250V , SB SAT-53801015 SAT-53801015 Power Supply: F9

3-Phase Input Line: F3, F4 and F5


Fuse 50A / 600V SB SAT-53801011 SAT-53801011 2-Phase Input Line: F3 and F4
D2 1-Phase Input Line: F3

Fuse 10A / 250V T SAT-53801008 SAT-53801008 Transformer 2: F12, F13

F1.2 Fuse 6A / 250V SB SAT-53801002 SAT-53801002 LF-RAC: F1

Fuses 0.5A / 250V SAT-53801019 SAT-53801019 Interface DRAC PCB: F1, F2

F2.3 Fuses 15A / 600V FA SAT-53801018 SAT-53801018 Interface DRAC PCB: F3, F4

Fuse 6A / 250V, SB SAT-53801002 SAT-53801002 Interface DRAC PCB: F1 Aerial

Fuse 6A / 250V, SB SAT-53801002 SAT-53801002 Locks Board: F1


G1 1
G1.1
Fuse 10A / 250V, SB SAT-53801008 SAT-53801008 Locks Board: F2

D2
G1.1

D2

A10

A10

7-19
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

F1.2

F2.3 F2.3

7-20
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

J0 INTERCONNECTIONS

J1 AEC Adaptation Cable (J5-J4) 1 A3251-02 A3251-02 0.8 meters

J2 PC Interface Box - Serial Cable 1 A3352-01 A3352-01 15 meters

6 meters.
J3 PC Interface Box - Computer Cable 1 A21089-06 Compatible with PC Interface Box
SAT-AS00519

6 meters.
J4 PC Interface Box - Computer Cable 1 A3363-01 A3363-01 Compatible with PC Interface Box
SAT-A6509-40.

ITEM DESCRIPTION QTY MFT. REFERENCE GEHC REFERENCE REMARKS

K0 MISCELLANEOUS

SAT-6470-08 2357942 Cover assembly in 1 part


K1 Cover Generator 1
Cover assembly in two parts.
SAT-A11740-03
Replaces the Cover SAT-6470-08

K2 High Voltage Cables 1 SAT-6680-12 SAT-6680-12 12 meters

K3 HV Cables Cover 1 SAT-A6855-12 SAT-A6855-12 12 meters

K4 Flat Field Phantom 1 SAT-31528-01 SAT-31528-01

K4

K1

7-21
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

7-22
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

CHAPTER 8
SCHEMATICS

8-1
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

8-2
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

TABLE OF CONTENTS

SCH. No. SCHEMATIC REMARKS

54302010 Compact Generator High or Low Speed Generator

A3096--02 LF-RAC PCB Only used with Low Speed Generator

A3274--01 Delayed Switch-Off PCB

A3243--04 Control DRAC PCB


O l used
Only d with
ith High
Hi h Speed
S d Generator
G t
A3240--05 Interface DRAC PCB

A3109--01 Clamping PCB

A3000--80/81 HT Controller PCB

A3004--10 Filament Control PCB

A3009--11 Interface Control PCB

A3063--06 IPM Driver PCB

A3212--01 Charge / Discharge Monitor PCB

A3024--221 ATP Console CPU PCB

A3012--25 AEC Control PCB

A3263--03 AEC Adaptation PCB

A3170--01 Radiation Meter PCB

A3179--09 Auto ON/OFF PCB

8-3
GE Healthcare Proteus XR/f Generator
REV 4 SM 5764770--1EN

This page intentionally left blank.

8-4
A B C D E

6J2
POWER INPUT TRANSFORMER
6T2 3F2
240 VAC SW LVDC POWER SUPPLY -PWR OFF 4 -PWR OFF
T 1.5A
44 530V 240V 20 9 5 3PS1 +5V +12V -12V (sheets 3 & 4)
+5V -PWR ON 5 -PWR ON
3LF3 1 7 +12V
3F8
115 VAC SW 2 10 -12V 3 +12V
10 480V T 3A 3F9 jumper
3K3 14 1 3 8 GND
115V 18 115 VAC T 0.4A 3
jumper 9 9 -12V
connect 2 4 4
440V (sheets 6
according 9 2
3 & 4) +12V UNR
with the 13
4 0V 17 0 VAC 4
input line 8 +12V UNR
7 415V 3F6
T 3A 6 GND
POWER 21V 15
LAMP 8 400V
7 GND
6NL1
0V 14
230V 1 GND
5 3F7
T 3A
11V 12 11 7
4 +24V-PSU
1 ohm +24V
0V 3K3
11 4w 3TS1
- 3BR4
+ - 3BR2
+ 10 6
28V 35
54 240 VAC SPLY
26V 34 3C7 3C6
40 110V 2400 MF 2400 MF
17 +24V
24V 16
+ 16 -12V
22V 33 - 3BR1
15 +12V
20V 32
240 VAC SW 26 240 VAC SW
3 60V 16V 31 +SUPPLY
(sheets 3 & 4)
115 VAC SW 27 115 VAC SW
14V 30 24 VAC TR -SUPPLY
2 12V 19 (sheets 5)
3
20 VAC TR 3

1 0V 0V 13 12 8
0 VAC TR
Shield
Stud

THIS SHEET ONLY APPLIES TO THREE PHASE GENERATORS


L1 T1
0 VAC RTR
MANUFACTURED FOR INPUT LINES FROM 400 VAC
(sheets 8) Note.- In Three Phase Generators manufactured for Input Lines up to 240 VAC,
60 VAC RTR
Input Transformer 6T2 is as shown on sheet 2.
230 VAC RTR
6F13
T 10A

AC1
6F12
T 10A 6R1
+ BUS +
20 ohm (sheet 5)
100 w C2
1 nF CR1 CR2 CR3
6F3
T 50A 6LF1
L1 T1
L1 C5 C1
6F4 470 nF 470 nF
AC2
T 50A
MAINS

LOAD

L2 T2 L2 T2
L2
2 2
6F5 C4
T 50A 1 nF
L3 T3 L3 T3
L3
C3
470 nF
AC3
GND
- BUS -
C6 (sheet 5)
1 nF INPUT RECTIFIER BOARD
(A3255-03)
NEUTRAL
5A5

12TS4

N
12F14
U +24V-PSU +24V-DELAYED
12F15 T 15A
V
12F16 T 15A
W
T 15A
GND A1 A1
1 1
Note.- 12TS4, 12F14, 12F15,
and 12F16 are optional

A2 6K5 A2 6K6
NAME DATE SHEET / OF
6J3 54302010
K NC16/0390 M.González 30/11/16 DRAWING F.GARCIA 30/09/02
1/8
J NC14/0139 R.Hermosilla 24/04/14 REVISED A.DIAZ 11/11/02 K J I H G REV
-LINE CONT -LINE CONT -CHRG DR
10 I NC13/0015 R.Asenjo 21/01/13
(sheets 3 & (sheets 3 & 4)
R.Asenjo
4) H NC12/0466 08/11/12
SEDECAL COMPACT-LV DRAC/LF RAC
G NC10/0533 F.García 14/04/11
REV DESCRIPTION ISSUED BY DATE
GENERATOR
A B C D E
A B C D E

6J2
POWER INPUT TRANSFORMER
6T2 3F2
240 VAC SW LOW POWER SUPPLY -PWR OFF 4 -PWR OFF
connect T 1.5A
according 7 240V 240V 20 9 5 3PS1 +5V +12V -12V (sheets 3 & 4)
+5V -PWR ON 5 -PWR ON
with the 3LF3 1 7
3F8 +12V
input line 115 VAC SW 2 10 -12V 3 +12V
6 T 3A 3F9 jumper
3K3 14 1 3 8 GND
115V 18 115 VAC T 0.4A 3
jumper 9 9 -12V
POWER 2 4 4
LAMP (sheets 6
230V 3 & 4) 2 +12V UNR
5 13
4 0V 17 0 VAC 4
8 +12V UNR
3F6
6NL1 4 T 3A 6 GND
21V 15
7 GND
21 208V
0V 14 1 GND
3F7
23 120V T 3A
11V 12 11 7
+24V-PSU
3 110V 1 ohm +24V
0V 3K3
11 4w 3TS1
- 3BR4
+ - 3BR2
+ 10 6
8 28V 35
54 240 VAC SPLY
26V 34 3C7 3C6
22 100V 2400 MF 2400 MF
17 +24V
24V 16
+ 16 -12V
22V 33 - 3BR1
15 +12V
20V 32
240 VAC SW 26 240 VAC SW
9 60V 16V 31 +SUPPLY
(sheets 3 & 4)
115 VAC SW 27 115 VAC SW
14V 30 24 VAC TR -SUPPLY
2 12V 19 (sheets 5)
3
20 VAC TR 3

1 0V 0V 13 12 8
0 VAC TR
Shield
Stud

L1 T1
0 VAC RTR THIS SHEET ONLY APPLIES TO SINGLE PHASE GENERATORS
60 VAC RTR
(sheets 8) MANUFACTURED FOR INPUT LINES UP TO 240 VAC
230 VAC RTR

6F13
T 10A

AC2
6F12
T 10A Note.- For Two Phases, replace Neutral
+ BUS +
Cartridge (L2/N) by Fuse 6F4 (sheet 5)
C4
6F3 1 nF CR2 CR3
T 50A 6LF1
L1 T1
L1
6R1
20 ohm
MAINS

LOAD

Neutral Cartridge 100 w


2 or 6F4 2
C5
6F4 470 nF
T 50A
L2 T2 L2 T2
N/L2

AC3

GND
- BUS -
C6 (sheet 5)
1 nF INPUT RECTIFIER BOARD
(A3255-03)
5A5

12TS4

N
12F15
V +24V-PSU +24V-DELAYED
T 15A 12F14
U
T 15A
W
Note.- 12TS4, 12F14, and 12F15,
1 GND are optional A1 A1
1

NAME DATE SHEET / OF


A2 6K5 A2 6K6
6J3 K NC16/0390 M.González 30/11/16 DRAWING F.GARCIA 30/09/02 54302010
2/8
J NC14/0139 R.Hermosilla 24/04/14 REVISED A.DIAZ 11/11/02 K J I H G REV

-LINE CONT -LINE CONT -CHRG DR I NC13/0015 R.Asenjo 21/01/13


10
(sheets 3 & 4) (sheets 3 & 4) R.Asenjo
H NC12/0466 08/11/12
SEDECAL COMPACT-LV DRAC/LF RAC
G NC10/0533 F.García 14/04/11
REV DESCRIPTION ISSUED BY DATE
GENERATOR
A B C D E
A B C D E

THIS SHEET ONLY APPLIES TO GENERATORS WITH HT TRANSFORMERS FOR TWO X-RAY TUBES INTERFACE CONTROL BOARD P5 +5V
3A2
+5V 6 +24V
+24V 8 +24V-PSU
P1
+24V 5
-PWR ON 7 -PWR ON DELAYED
+24V 16 +24V
HT CONTROL BOARD (sheets 1 & 2)
-PWR OFF 8 -PWR OFF
-12V P2 3A1 P1 DELAYED +24V 2
4 PT CRL
4 4
+12V 1 -12V HT-C CLK 10 -LINE CONT 1 -LINE CONT (sheets 1 & 2)
3 ON
+5V 2 +12V HT-C DAT 11 See NOTE 2 +5V OUT 4 +5V OUT (sheet 5)
+12V 2 GND
6J3 3 +5V -TUBE 2 SCL 12 -T2 SCL -TUBE 2 SCL DISCHARGE 1 12 DISCHARGE 1

(sheet 5)
(sheet 8) (sheet 8) 6 +12V
4 GND -FIL 1 SCL 13 DISCHARGE 1 RTN 14 DISCHARGE 1 RTN
PT CRL 7 1 GND
5 GND -START 14 -START DISCHARGE 2 13 DISCHARGE 2
-KV UP 15 (sheet 8)
-PRE CHRG 15 DISCHARGE 2 RTN 10 DISCHARGE 2 RTN 6J3
-KV DWN 8 P1
-CHRG DR 16 -CHRG DR GND 9
HT-C DAT 3 1 -KV UP (sheets 1 & 2) +24V P3 -BUCKY 1 DR CMD 14 -BUCKY 1 DR CMD
-TUBE 3 17
HT-C CLK 6 2 -KV DWN 4 +24V P3 -BUCKY 2 DR CMD 11 -BUCKY 2 DR CMD

-EXP 1 6 -EXP P3 2 GND LINE SYNC 3 9 LINE SYNC

C-HT DAT 2 4 C-HT DAT ERROR DRAC 6 ERROR DRAC (sheet 3) 1 GND 12 -BUCKY EXP
P2
C-HT CLK 5 5 C-HT CLK -KV DR1 1 -KV DR1 16 GND
-ROOM LIGHT
-ROOM LIGHT 2
-KV DR2 (sheet 5) (sheet 5)
-PREP 4 3 -PREP -KV DR2 2
IC SPLY 1 3
9 GND -FIL DR1 4 P4 3TS1
IC SPLY 2 5
7 -CHRG -FIL DR2 5 -BUCKY 1 DR CMD 12 -BUCKY 1 DR CMD
-PREP
3
PT SPLY 1 39 PT SPLY 3
(sheet 3) 8 -FIL 1 ACK GND 3 -BUCKY 2 DR CMD 13 -BUCKY 2 DR CMD
IC GND 7 42 IC GND
-CHRG
9 BUCKY SPLY
(sheet 5) P4 ROOM LIGHT SW ROOM LIGHT SW
P5 4 48
5 AUX BUCKY SPLY
RTR I
3 RTR I -RTR RUN 5 ROOM LIGHT SUP 6 47 ROOM LIGHT SUP
(sheet 8) 11 -BUCKY 1 MOTION
8 FIL I -ACC 7 18 GND
See NOTE 2 10 -BUCKY 2 MOTION P4
4 -BRAKE 6
3TS1 8 BUCKY 1 DR -BUCK EXP 6
1 -KV -HS SCL 9
7 BUCKY 2 DR 115 VAC 2 115 VAC (sheets 1 & 2)
BUCKY SPLY 2 2 +KV -RTR RDY 4
18 -TUBE 2 SCL 0 VAC 1 0 VAC
BUCKY SPLY -IGBT FAULT
9 6 -MA -IGBT FAULT 3
(sheet 5) 19 -FIL 1 SCL 115 VAC SW 3 115 VAC SW
AUX BUCKY SPLY 46 7 +MA -AUTO OFF INH 1
17 FIL PREHEAT FIL 1 RTN 15 NOTE 1:
-BUCKY 1 MOTION 5 5 GND HI MA 10 See NOTE 1 Wires FIL 1 RTN (P4-15) and FIL 2 RTN (P4-16), are
14 FIL RTN FIL 2 RTN 16
-BUCKY 2 MOTION 12 factory inverted on the INTERFACE CONTROL Board
22 -FIL 1 ACK SOL DR 20
BUCKY 1 DR 4 -BRAKE
(sheet 8)
BUCKY 2 DR 11

0 VAC -ACC
3
2
(sheet 8) 6J1 See NOTE 2 2
0 VAC 10 0 VAC

HT TRANSFORMER +24V
BUCKY 1 MOT RTN 6 J1 10 +24V
9HT1 ERROR DRAC
GND 7 K HI MA 8 ERROR DRAC
(sheet 3)
BUCKY 2 MOT RTN 13 P1 P3 A GND 9
DELAYED
GND 14 +24V 5 -FIL DR1 FIL PREHEAT 2 B +KV GND 1 -RTR RUN

SUPPLY (BUCKY 1) 1 +5V 6 -FIL DR2 FIL RTN 1 C -KV 2 -TUBE 1 RTR
Red
SUPPLY (BUCKY 2) 8 3 +5V FIL SUP 4 D +MA 3 -TUBE 2 RTR
mA test posts
8 DELAYED +24V E -MA 4 -HS SCL
Black
7 FIL I F FIL 1 RTN 5 -RTR RDY
HT DR1
1 GND
HT SW CONTROL
H FIL 2 RTN P1 (sheet 5) -SF PREP 6 -SF PREP

(sheet 5)
FILAMENT 10A2 J1 G FIL SUP -FL EXP 7 -FL EXP
P2 DRIVER HT DR2
+24V
J2 1 0 VAC N SW COM P3 (sheet 5) -PREP 12 -PREP
+SUPPLY 4 +SUPPLY BOARD
(sheets 1 & 2) 3A3 4 2 U SW1 11 GND
-SUPPLY 1 -SUPPLY COM
3 3 V SW2 SHLD (sheet 5)
1 2 4 J -HT INTLK 6J3 1

1 5 240 VAC SW 13 -HT INTLK

NOTE 2.- For High Speed Version:


NAME DATE SHEET / OF
-T2 SCL signal is shortcircuited with -TUBE 2 SCL signal. -TUBE 2 SCL K NC16/0390 M.González 30/11/16 DRAWING F.GARCIA 30/09/02 54302010
3/8
(sheet 8) J NC14/0139 R.Hermosilla 24/04/14 REVISED A.DIAZ 11/11/02 K J I H G REV
-ACC and -BRAKE signals for Low Speed Version becomes -TUBE 1 RTR and -TUBE 2 RTR respectively. NC13/0015
I R.Asenjo 21/01/13
R.Asenjo
6J1 connector used only for High Speed version (see LV-DRAC documentation). H NC12/0466 08/11/12
SEDECAL COMPACT-LV DRAC/LF RAC
G NC10/0533 F.García 14/04/11
REV DESCRIPTION ISSUED BY DATE
GENERATOR
A B C D E
A B C D E

THIS SHEET ONLY APPLIES TO GENERATORS WITH HT TRANSFORMER FOR ONLY ONE X-RAY TUBE INTERFACE CONTROL BOARD P5 +5V
3A2
+5V 6 +24V
+24V 8 +24V-PSU
P1
+24V 5
-PWR ON 7 -PWR ON DELAYED
+24V 16 +24V
HT CONTROL BOARD (sheets 1 & 2)
-PWR OFF 8 -PWR OFF
-12V P2 3A1 P1 DELAYED +24V 2
4 PT CRL
4 4
+12V 1 -12V HT-C CLK 10 -LINE CONT 1 -LINE CONT (sheets 1 & 2)
3 ON
+5V 2 +12V HT-C DAT 11 See NOTE 2 +5V OUT 4 +5V OUT (sheet 5)
+12V 2 GND
6J3 3 +5V -TUBE 2 SCL 12 -T2 SCL -TUBE 2 SCL DISCHARGE 1 12 DISCHARGE 1

(sheet 5)
(sheet 8) (sheet 8) 6 +12V
4 GND -FIL 1 SCL 13 DISCHARGE 1 RTN 14 DISCHARGE 1 RTN
PT CRL 7 1 GND
5 GND -START 14 -START DISCHARGE 2 13 DISCHARGE 2
-KV UP 15 (sheet 8)
-PRE CHRG 15 DISCHARGE 2 RTN 10 DISCHARGE 2 RTN 6J3
-KV DWN 8 P1
-CHRG DR 16 -CHRG DR GND 9
HT-C DAT 3 1 -KV UP (sheets 1 & 2) +24V P3 -BUCKY 1 DR CMD 14 -BUCKY 1 DR CMD
-TUBE 3 17
HT-C CLK 6 2 -KV DWN 4 +24V P3 -BUCKY 2 DR CMD 11 -BUCKY 2 DR CMD

-EXP 1 6 -EXP P3 2 GND LINE SYNC 3 9 LINE SYNC

C-HT DAT 2 4 C-HT DAT ERROR DRAC 6 ERROR DRAC (sheet 4) 1 GND 12 -BUCKY EXP
P2
C-HT CLK 5 5 C-HT CLK -KV DR1 1 -KV DR1 16 GND
-ROOM LIGHT
-ROOM LIGHT 2
-KV DR2 (sheet 5) (sheet 5)
-PREP 4 3 -PREP -KV DR2 2
IC SPLY 1 3
9 GND -FIL DR1 4 P4 3TS1
IC SPLY 2 5
7 -CHRG -FIL DR2 5 -BUCKY 1 DR CMD 12 -BUCKY 1 DR CMD
-PREP
3
PT SPLY 1 39 PT SPLY 3
(sheet 4) 8 -FIL 1 ACK GND 3 -BUCKY 2 DR CMD 13 -BUCKY 2 DR CMD
IC GND 7 42 IC GND
-CHRG
9 BUCKY SPLY
(sheet 5) P4 ROOM LIGHT SW ROOM LIGHT SW
P5 4 48
5 AUX BUCKY SPLY
RTR I
3 RTR I -RTR RUN 5 ROOM LIGHT SUP 6 47 ROOM LIGHT SUP
(sheet 8) 11 -BUCKY 1 MOTION
8 FIL I -ACC 7 18 GND
See NOTE 2 10 -BUCKY 2 MOTION P4
4 -BRAKE 6
3TS1 8 BUCKY 1 DR -BUCK EXP 6
1 -KV -HS SCL 9
7 BUCKY 2 DR 115 VAC 2 115 VAC (sheets 1 & 2)
BUCKY SPLY 2 2 +KV -RTR RDY 4
18 -TUBE 2 SCL 0 VAC 1 0 VAC
BUCKY SPLY -IGBT FAULT
9 6 -MA -IGBT FAULT 3
(sheet 5) 19 -FIL 1 SCL 115 VAC SW 3 115 VAC SW
-AUTO OFF INH NOTE 1.- Only for one FL tube:
AUX BUCKY SPLY 46 7 +MA 1 Wires FIL 1 RTN (P4-15) and FIL 2 RTN (P4-16), are
17 FIL PREHEAT FIL 1 RTN 15
-BUCKY 1 MOTION 5 5 GND HI MA 10 factory inverted on the INTERFACE CONTROL Board,
See NOTE 1
14 FIL RTN FIL 2 RTN 16 and 6J3-13 is disconnected and insulated.
-BUCKY 2 MOTION 12
22 -FIL 1 ACK SOL DR 20
BUCKY 1 DR 4 -BRAKE
(sheet 8)
BUCKY 2 DR 11

0 VAC -ACC
3
2
(sheet 8) 6J1 See NOTE 2 2
0 VAC 10 0 VAC

HT TRANSFORMER +24V
BUCKY 1 MOT RTN 6 P1 10 +24V
9HT1 ERROR DRAC
GND 7 K HI MA 8 ERROR DRAC
(sheet 4)
BUCKY 2 MOT RTN 13 P1 P3 A GND 9
DELAYED
GND 14 +24V 5 -FIL DR1 FIL PREHEAT 2 B +KV GND 1 -RTR RUN

SUPPLY (BUCKY 1) 1 +5V 6 -FIL DR2 FIL RTN 1 C -KV 2 -TUBE 1 RTR
Red
SUPPLY (BUCKY 2) 8 3 +5V FIL SUP 4 D +MA 3 -TUBE 2 RTR
mA test posts
8 DELAYED +24V E -MA 4 -HS SCL
Black
7 FIL I F FIL 1 RTN 5 -RTR RDY
HT DR1
1 GND H FIL 2 RTN P1 (sheet 5) -SF PREP 6 -SF PREP

(sheet 5)
FILAMENT G FIL SUP -FL EXP 7 -FL EXP
P2 DRIVER HT DR2
BOARD
N SW COM P3 (sheet 5) -PREP 12 -PREP
+SUPPLY 4 +SUPPLY
(sheets 1 & 2) 3A3 U SW1 11 GND
-SUPPLY 1 -SUPPLY COM
V SW2 SHLD (sheet 5)
1 J -HT INTLK 6J3 1

OPEN - for Fl tube


13 -HT INTLK
GND - for RAD tube

NOTE 2.- For High Speed Version:


NAME DATE SHEET / OF
-T2 SCL signal is shortcircuited with -TUBE 2 SCL signal. K NC16/0390 M.González 30/11/16 DRAWING F.GARCIA 30/09/02 54302010
4/8
J NC14/0139 R.Hermosilla 24/04/14 REVISED A.DIAZ 11/11/02 K J I H G REV
-ACC and -BRAKE signals for Low Speed Version becomes -TUBE 1 RTR and -TUBE 2 RTR respectively. NC13/0015
I R.Asenjo 21/01/13
R.Asenjo
6J1 connector used only for High Speed version (see LV-DRAC documentation). H NC12/0466 08/11/12
SEDECAL COMPACT-LV DRAC/LF RAC
G NC10/0533 F.García 14/04/11
REV DESCRIPTION ISSUED BY DATE
GENERATOR
A B C D E
A B C D E

3TS1
-DOOR
22 J2-19
DOOR RTN
HT DR2 J2-18
23
(sheets 3 & 4) -ROOM LIGHT
24 J2-8
-SF PREP
5C10 5C11 (sheets 3 & 4) J2-4
5R2 36 to J2 connector of the
2 MF 2 MF -FL EXP
4 5L1 1 Mohm, 2w J2-17 ATP CONSOLE board 4
HT DR1 37
PT INPUT
(sheets 3 & 4) 51 J2-13
-ALOE
52 J2-24
-READY

C2E1
C1

C1
E2

C2E1

E2
5C9 53 J2-9
40 MF for Single Phase Generator
10 MF for Three Phase Generator
5IGBT 2 5IGBT 1
4TS3
SPR-+5 REG OUT

SPR-+5 REG OUT


CPI-CONTROL

CPI-CONTROL
-THERM 1

FPO-FAULT

FPO-FAULT
FNO-FAULT

FNO-FAULT
J2-3 1 10/11J7-5

VNC-GND

VNC-GND
VPC-GND

VPC-GND
VPI-VCC

VPI-VCC
-THERM 2

VNI-VCC

VNI-VCC
J2-20 2 10/11J7-6

SNR-

SNR-
(sheets 6, 7, and 8)

CNI-

CNI-
THERM COM
J2-5 3 10/11J7-7
to J2 connector of the
J3 J4 J3 J4 ATP CONSOLE board
2 3 4 1 5 2 3 4 1 5 2 3 4 1 5 2 3 4 1 5 J2-16 7 -SFC

J2-11 18 ABC OUT (only required for DSI option)


IPM DRIVER 2 IPM DRIVER 1 J2-14 19 FL DSI
5A2 5A1

FAULT

FAULT
GND
DR1

GND
J2-15
DR2

DR1
+5V

+5V
CAM SYNC

DR2
20

3 3

J1 J2 J1 J2
1

4
+24 VDC DELAYED For Three Phase Generator

GND
DC+ + P2
+5V OUT

GND 5C3 - 5A3


2 DISCHARGE 1 RTN

DISCHARGE 1
(sheets 3 & 4)
3900 MF 1
5C4
-KV DR1 3900 MF
(sheets 3 & 4) CHARGE/ P1
-KV DR2 DISCHARGE
2 -CHRG
-IGBT FAULT MONITOR #1 (sheet 3)
(sheets 3 & 4)
BOARD 1

BUS+ DC+
To Three Phase Generator
(sheets 1 & 2)
BUS- DC-
To Single Phase Generator
5C1
CHARGE/ P1
3900 MF DISCHARGE
COM 5C2 2
(sheets 3 & 4)
MONITOR #2
3900 MF
2 BOARD 1 2

TS4 5A4 P2
+
2 DISCHARGE 2 RTN
FANS +24VDC (TB7-10 LOCKS BOARD) (sheets 3 & 4)
POWER
SUPPLY
+

0 VDC (TB7-11 LOCKS BOARD)


DC- - 1 DISCHARGE 2
-

LOCKS BOARD
TB7 7A3 TB7
F1 DC+ + P2
T 6A
24 VAC TR 1 3 24 VAC
To Lamp
5C2
3900 MF
5C4
3900 MF - 5A3
2 DISCHARGE 1 RTN

DISCHARGE 1
(sheets 3 & 4)

(Collimator) 5C1 5C3 1


(sheets 1 & 2) 0 VAC TR 2 4 0 VAC 3900 MF 3900 MF
20 VAC TR 9 CHARGE/ P1
F2 DISCHARGE
10 +24 VDC 2 -CHRG
T 10A DC- MONITOR #1 (sheets 3 & 4)
- +
D1 7 +24 VDC BOARD 1
R1 To
1 5 +24 VDC 1
1K5 Locks For Single Phase Generator
C1 - C4
6 0 VDC
2200 MF
8 0 VDC
NAME DATE SHEET / OF
11 0 VDC K NC16/0390 M.González 30/11/16 DRAWING F.GARCIA 30/09/02 54302010
5/8
J NC14/0139 R.Hermosilla 24/04/14 REVISED A.DIAZ 11/11/02 K J I H G REV
I NC13/0015 R.Asenjo 21/01/13
H NC12/0466 R.Asenjo 08/11/12 COMPACT-LV DRAC/LF RAC
G NC10/0533 F.García 14/04/11 SEDECAL
REV DESCRIPTION ISSUED BY DATE
GENERATOR
A B C D E
A B C D E

CONTROL DRAC J3 Ferrite Ferrite J1 INTERFASE DRAC-HF


Core Core
PCB (A3243-03) Ribbon Cable PCB (A3240-05)
Connections ( ) for
+5 VDC 1 1 +5 VDC
11A1 11A2 only one FL tube
+5 VDC 2 2 +5 VDC
11TS2
+5 VDC 3 3 +5 VDC
MAIN T1
+5 VDC 4 4 +5 VDC 1 (9)
AUX T1
GND 5 5 GND 2 (10)
4 J4 4
COM T1
GND 6 6 GND 3 (11)
-THERM 1
1 GND GND 7 7 GND 4 (12)
ERROR DRAC THERM COM
6J1-8 2 -ERROR CODE GND 8 8 GND 5 (13)

+15 VDC See FAN 1


3 NC 9 9 +15 VDC 6 (14)
Note
-RTR RDY 0 VAC
6J1-5 4 RDY 1 +15 VDC 10 10 +15 VDC 7 (15)
GND
5 RDY 1 OUT +15 VDC 11 11 +15 VDC 8 (16)

6 +15 VDC +15 VDC 12 12 +15 VDC


GND 115 VAC SW
6J1-11 7 RDY 1 RTN -15 VDC 13 13 -15 VDC 17
240 VAC SW
8 NC -15 VDC 14 14 -15 VDC 18
Cable to High Speed Starter connection 6J1 (sheets 3 & 4)

9 SPARE 2 V UNR 15 15 V UNR

10 SPARE 2 RTN V UNR 16 16 V UNR


Note:
11 SPARE 1 PRECH IF 17 17 PRECH IF Connect FAN to 11TS2-17
or 11TS2-18 as required.
12 SPARE 1 RTN PRECH IF 18 18 PRECH IF
+24 VDC
6J1-10 13 SF +5 VDC 19 19 +5 VDC
3 -SF PREP 3
6J1-6 14 SF RTN +5 VDC 20 20 +5 VDC

15 FL T3 IF 21 21 T3 IF
-FL EXP
6J1-7 16 FL RTN T3 IF 22 22 T3 IF

17 HS T2 IF 23 23 T2 IF
-HS SCL
6J1-4 18 HS RTN T2 IF 24 24 T2 IF

19 ST T1 IF 25 25 T1 IF
-RTR RUN
6J1-1 20 ST RTN T1 IF 26 26 T1 IF

21 T3
AUXILIARY CLAMPING +24 VDC
22 T3 RTN
J1 TRANSFORMER
23 T2 11K2 11K3
P J5
R2 R1 R2 R1 11K2 A1 11K3 A1
-TUBE 2 RTR 1 12
6J1-3 24 T2 RTN AUX 4
2 2
J4
1 1
10 1
A2 A2
2 K1 1
VR1 F1
7 8 V250LA20A
COM 3 4 2
T 6.25A
2 COM T1 3 K3 3 2

P Note
SHL 2 8 AUX T1
1
VR1
MAIN T1
V250LA20A
10 11K2 J3
4 3
11K3 Note:
7 12 R4 R3 Connections to 11TS2 of
PRIN 1 240 VAC SW 1
R4 R3 the Generator Cabinet to 11TS2
4 3
MAIN 0 VAC 2
TRANSFORMER
J2 Ferrite Ferrite
J2
Core Core
V- 1 5 -DC OUT

V+ 2 4 +DC OUT

3 GND

2 +DC IN
from DC voltage Bus
Note.- 11J7 is an aerial connector 11J7 (sheet 5) 1 -DC IN
+24 VDC
+24V DELAYED 1

1 GND 3 THIS SHEET ONLY APPLIES TO HIGH SPEED STARTER 1

240 VAC SW 4
FOR ONLY ONE X-RAY TUBE (OPTIONAL)
115 VAC SW 2
NAME DATE SHEET / OF
-THERM 1 to 11TS2
4TS3-1 5 K NC16/0390 M.González 30/11/16 DRAWING F.GARCIA 30/09/02 54302010
(sheet 5)
THERM COM 6/8
4TS3-3 7 J NC14/0139 R.Hermosilla 24/04/14 REVISED A.DIAZ 11/11/02 K J I H G REV
I NC13/0015 R.Asenjo 21/01/13
0 VAC 8
R.Asenjo
H NC12/0466 08/11/12
SEDECAL COMPACT-LV DRAC/LF RAC
G NC10/0533 F.García 14/04/11
REV DESCRIPTION ISSUED BY DATE
GENERATOR
A B C D E
A B C D E

CONTROL DRAC J3 Ferrite Ferrite J1 INTERFASE DRAC-HF


Core Core
PCB (A3243-03) Ribbon Cable PCB (A3240-05)
+5 VDC 1 1 +5 VDC
11A1 11A2
+5 VDC 2 2 +5 VDC
11TS2
+5 VDC 3 3 +5 VDC
MAIN T1
+5 VDC 4 4 +5 VDC 1
AUX T1
GND 5 5 GND 2
4 J4 4
COM T1
GND 6 6 GND 3
-THERM 1
1 GND GND 7 7 GND 4
ERROR DRAC THERM COM
6J1-8 2 -ERROR CODE GND 8 8 GND 5

+15 VDC FAN 1


3 NC 9 9 +15 VDC 6
-RTR RDY 0 VAC
6J1-5 4 RDY 1 +15 VDC 10 10 +15 VDC 7
GND
5 RDY 1 OUT +15 VDC 11 11 +15 VDC 8
MAIN T2
6 +15 VDC +15 VDC 12 12 +15 VDC 9
GND AUX T2
6J1-11 7 RDY 1 RTN -15 VDC 13 13 -15 VDC 10
COM T2
8 NC -15 VDC 14 14 -15 VDC 11
Cable to High Speed Starter connection 6J1 (sheets 3 & 4)

-THERM 2
9 SPARE 2 V UNR 15 15 V UNR 12
THERM COM
10 SPARE 2 RTN V UNR 16 16 V UNR 13

PRECH IF FAN 2
11 SPARE 1 PRECH IF 17 17 14

PRECH IF 0 VAC
12 SPARE 1 RTN PRECH IF 18 18 15
+24 VDC GND
6J1-10 13 SF +5 VDC 19 19 +5 VDC 16
3 -SF PREP 3
6J1-6 14 SF RTN +5 VDC 20 20 +5 VDC

15 FL T3 IF 21 21 T3 IF
-FL EXP
6J1-7 16 FL RTN T3 IF 22 22 T3 IF

17 HS T2 IF 23 23 T2 IF
-HS SCL
6J1-4 18 HS RTN T2 IF 24 24 T2 IF

19 ST T1 IF 25 25 T1 IF
-RTR RUN
6J1-1 20 ST RTN T1 IF 26 26 T1 IF

21 T3
Note:
AUXILIARY CLAMPING Connections to 11TS2 of +24 VDC
22 T3 RTN
J1 TRANSFORMER the Generator Cabinet
23 T2 11K2 11K3 11KT1
P J5
R2 R1 R2 R1 R2 R1 11KT1 A1 11K2 A1 11K3 A1
-TUBE 2 RTR 1 12
6J1-3 24 T2 RTN AUX 4 AUX T2

10
2 1 2 1 2 1
AUX T1
} Note
1
J4
A2 A2 A2
2 K1 1
VR1 F1
7 8 V250LA20A
COM 3 4 2
T 6.25A
COM T1 3 K3 3
2

SHL 2
P
1 8 VR1
COM T2
} Note
2

V250LA20A

10 11K2 J3
4 3
11K3 11KT1
7 12 R4 R3 R4 R3
PRIN 1 MAIN T2 240 VAC SW 1

MAIN
R4 R3
4 3 4 3
MAIN T1
} Note
0 VAC 2
to 11TS2

TRANSFORMER
J2 Ferrite Ferrite
J2
Core Core
V- 1 5 -DC OUT

V+ 2 4 +DC OUT

3 GND

DC+ 2 +DC IN
11J7 from DC voltage Bus
Note.- 11J7 is an aerial connector (sheet 5) DC- -DC IN
+24 VDC 1
11TS2
+24V DELAYED 1 11KT1
tube 2 62 61
240 VAC SW 4 18 FAN 2
1
54 53 to 11TS2 1
tube 1
115 VAC SW 2 17 FAN 1 THIS SHEET ONLY APPLIES TO HIGH SPEED STARTER
GND 3
FOR ONE OR TWO X-RAY TUBES (STANDARD)
-THERM 1 11KT1
4TS3-1 5 72 71 NAME DATE SHEET / OF
TUBE 2 SCL
4TS3-2 -THERM 2 for K NC16/0390 M.González 30/11/16 DRAWING F.GARCIA 30/09/02 54302010
(sheet 5) 6 to 11TS2 external TUBE 1 SCL 84 83 7/8
J NC14/0139 R.Hermosilla 24/04/14 REVISED A.DIAZ 11/11/02 K J I H G REV
4TS3-3 THERM COM interface
7 TUBE SCL COMM I NC13/0015 R.Asenjo 21/01/13
R.Asenjo
0 VAC 8 H NC12/0466 08/11/12
SEDECAL COMPACT-LV DRAC/LF RAC
G NC10/0533 F.García 14/04/11
REV DESCRIPTION ISSUED BY DATE
GENERATOR
A B C D E
A B C D E

NOTE.- 10T3, 10C5-1, and


THIS SHEET ONLY APPLIES TO LOW SPEED STARTER 10R4-1 for 330 VAC Stator 10C5-1
for external interface
15 uF

TUBE SCL COM


10C5-2

TUBE 1 SCL

TUBE 2 SCL
Note.- Capacitor 10C5 could be 40 uF

115 VAC SW

240 VAC SW
30 uF
depending on tube used 10R4-1
1 Mohm, 2w NOTE.- Use Tube 2
connections for FL tube

Note.- 10J7 and 10J8 are 10R4-2 10TS2


4 aerial connectors 1 Mohm, 2w 4
30 uF 15 uF
10J8 6 FAN 1

TB1 J1

10

11

12

13

14

15

16

21

22

23

24

25

26

27

28

10

11

12
14 FAN 2

9
230 VAC RTR 1 TB1
3 COM T1
(sheets 1 & 2) 60 VAC RTR KACC
2 K1-1
1 T1/T2 COMM 11 COM T2
0 VAC RTR 3 TB3 TB3 TB4 TB4
2 0 VAC 7 0 VAC
KCTR
F1 T2 T1 T2 T1 K4-2
15 0 VAC
T 6A Tube 2 Tube 1 Tube 2 Tube 1
T2
3 5 THERM COMM
KCTRR
5 KCT KCTRR K5-3
6 THERM COM 13 THERM COMM
10T3 K2-3 K5-1
47 10RC1
2w 7 KBR -THERM 1
-THERM 1 4
330V
7 TB2 Iaux K3-2 TB1
9 -THERM 2
100 nF 12 -THERM 2
CR6 KCTR

+
1 KV 8 T1 1 KBU4J K4-1
10 17 2 AUX T1
Tube 1 KCT
10LF2 10K1 9 K2-1
220V
~ ~ 18 10 AUX T2
2 4 2 1
T1 19 1 MAIN T1
1 3 6 TB2
0V
- 20 9 MAIN T2
3
6 5 KCTR 3
5 T2 K4-4
8 GND
Tube 2 7
4
Imain KBR
16 GND
T1/T2 COMM 9 K3-1
(to 10TS2)
10 1 TP2 J1
CR11 +24V +24V
+24V DELAYED 1N4007
1 +24V DELAYED
10K1 3 10R5 CR10 CR12 C4
475 VM48 1N4007 100 uF
2w +24V Q2 35V
AC + 2N4401
4
KCTR CR4 C3 R8 TP1
K4 1N4007 100nF 1K5 GND
+24V
AC -
CR8 2 GND
KCTRR CR5 1N4007
KCT K5 1N4007 LF-RAC BOARD
K2-2 10A1 CR9 CR7 KCTRR
VM48 1N4007 K5-2
Q1
KCTR AC + 2N4401
K4-3 4
2 2
C2 R7
100nF 1K5
AC - KACC
DS4 K1-2
+24V HLMP1700 3

TP3
KACC KCT KBR
R1 CR1 R2 CR2 R3 CR3 R4 R5 RTRI
10J7 4K75 1N4007 4K75 1N4007 4K75 1N4007 4K99 4K99
K1 K2 K3
240 VAC SW 5
4 240 VAC SW
DS1 DS2 DS3 C1
115 VAC SW HLMP1700 HLMP1700 HLMP1700 R6 22 uF
2 115 VAC SW
475
+24V DELAYED
1 +24V DELAYED
GND GND
3
J1
6

8
9
from HT CONTROL board

-START

-ACC 10

-T2 SCL 11

-BRAKE 12

1 RTRI 13 1

-TUBE 2 SCL 14

0 VAC 8 0 VAC
NAME DATE SHEET / OF
-THERM 1
4TS3-1 5 -THERM 1
K NC16/0390 M.González 30/11/16 DRAWING F.GARCIA 30/09/02 54302010
(sheet 5)

-THERM 2 to 10TS2 8/8


4TS3-2 6 -THERM 2 J NC14/0139 R.Hermosilla 24/04/14 REVISED A.DIAZ 11/11/02 K J I H G REV

THERM COM I NC13/0015 R.Asenjo 21/01/13


4TS3-3 7 THERM COM
R.Asenjo
H NC12/0466 08/11/12
SEDECAL COMPACT-LV DRAC/LF RAC
G NC10/0533 F.García 14/04/11
REV DESCRIPTION ISSUED BY DATE
GENERATOR
A B C D E
TUBE SCL COMM
TUBE 1 SCL

TUBE 2 SCL
FAN 1

FAN 2
COMM 30 MF 15 MF 115 VAC 240 VAC

10 11 12 13 14 15 16 21 22 23 24 25 26 27 28 9 10 11 12
TB1 TB1 J1
1 F1
230 VAC RTR T 6A
TB3 KCT TB3 TB4 KCTRR TB4
2 KACC T2 K2--3 T1 T2 K5--1 T1 KCTR
60 VAC RTR
K1--1 KCTRR K4--2
3
VAC RTR RTN Tube 2 Tube 1 Tube 2 Tube 1 K5--3

T2 KBR
7 TB2 6 5 TB1
K3--2
T1 7
8
330V IAUX CR6 KCTR 17 AUX T1
KBU4J

+
9 K4--1
9 1
Tube 1 KCT 10 18 AUX T2
K2--1
KCTR
19 MAIN T1
T1 K4--4
6 6 5 KBR
TB2 7 20 MAIN T2
_ K3--1
5 T2 IMAIN
230 V 9
4 10 1
Tube 2 +24 VDC J1
CR11 TP2
1N4007 +24V
1 +24 VDC DELAYED
CR10 CR12 C4
+24 VDC VM48 1N4007 100uF
Q2 35V

+
CR4 2N4401
KCTR TP1
1N4007 C3 R8 GND
K4 100nF 1K5 2
+24 VDC _ GND
CR8
KCT CR5 1N4007
KCTRR 1N4007
K2--2 K5 KCTRR
CR9 CR7 K5--2
VM48 1N4007 4
Q1 --TUBE 2 SCL

+
KCTR 2N4401
K4--3 C2 R7
100nF 1K5
_ 3
+24 VDC KACC
DS4
HLMP1700 K1--2
CR1 CR2 CR3
R1 1N4007 R2 1N4007 R3 1N4007
4K75 KACC 4K75 KCT 4K75 KBR TP3
R4 R5 RTRI
K1 K2 K3 4K99 4K99
5 RTR I
DS1 DS2 DS3
HLMP1700 HLMP1700 HLMP1700 R6 C1
475 22uF

J1 6 7 8
--ACC --T2 SCL --BRAKE

Note:
061109
The LF- RAC shown in the Generator schematic
NAME DATE SHEET / OF
(543020XX) is a basic documentation, only for
DRAWING F. GARCIA 21/09/97
DWG:
A3096--02
interconnection purposes. 1/1
Follow this schematic for detailed component REVISED A. DIAZ 21/09/97 B REV
information of the LF- RAC board.
LF -- RAC Board
B CN 03/004 F. García 13/01/03 SEDECAL ( Line Frequency Rotatory Anode Controller )
REV DESCRIPTION ISSUED BY DATE
CONTROL DRAC BOARD +15VDC +15VDC

J1 +15V
J4 C4
R10 TP3
100nF
6 +15 VDC 2 1K5 14
R1 R2 CR1
10K 1M 1N3595
5 RDY1 OUT 3 2
6 OFF OFF DLY
C1 K1 TP1
DLY
470uF 1 6
25V GND CR6 5
1 GND 4
TP5 1N3595 2
21 5
3 7
23 ADD JUMPER U1
CR5 C3 ICM7556
1N3595 3,3uF C2 DELAYED SWITCH--OFF BOARD
22 T3 RTN 6 100nF
( A3274--01 )
CR4 +15VDC
1N3595
24 T2 RTN 7
U1 CR8 R8
+15VDC ICM7556 1N4148 1M
C7
1uF R4 R3 10 CR2
CR3 10K 1M5 1N3595
R7 1N3595 4
C8 30K1 2 K2
8
10uF OFF OFF MANT
GENERATOR Q1
6 MANT TP2
2N4401 12
CABINET 9
CR9
PWR OFF 1N3595 13
R6 CR7
J2 TP6
10K 1N3595
4 PWR OFF 8 R9 C9
11 100nF
150K
C5 C6
OFF COND R5 3,3uF
TP4 100nF
J1 1K5
8 OFF COND 1 14 7--8 1 7--8
K1 K2
INTERFACE CONTROL BOARD

061109
NAME DATE SHEET / OF
DRAWING F. GARCIA 05/05/98
DWG:
A3274--01
1/1
REVISED A. DIAZ 05/05/98 D C B REV
D CN 14/0337 F. GARCIA 12/12/14
C CN 00/183 F. GARCIA 06/09/00
B CN 99/003 F. GARCIA 13/01/99 SEDECAL DELAYED SWITCH--OFF
REV DESCRIPTION ISSUED BY DATE
A B C D E

PTR1
VS1 D19 D20
+5 VDC +5 VDC C15 Q12 BC640
AR1 PD ( 0.......7 ) 3
8x10K C29 100nF TP9
100nF D40 D42 R117 D45 1N5819 BAT49
R29 33.2
BAT49 BAT49 1K D44 1N4148

6
U9 TC4426 1 TF5 6 1N4148 Q5

20
SW1 U13 HCT244 2 7
C11

GNDVDD
PD7 INA OUTA
8 9 2 1Y1 18 5

VCC
1A1 PD6 +5 VDC C65 R116 R30 BC640 D21
7 10 4 1A2 1Y2 16
6 11 6 PD5
1A3 1Y3 14 PD4 AR6
C12 4 INB OUTB 5 3 4
1K 2K21 BAT49
5 12 8 1A4 1Y4 12
4 13 11 PD3 8x4K7 PD0 220nF D41 D43
2Y1 9
4 4
2A1 PD2 PD1
3 14 13 2Y2 7

3
2A2 PD1 PD2 BAT49 BAT49 C59 4
2 15 15 2A3 2Y3 5
1 16 17 PD0 PD3 10uF
2A4 2Y4 3 PD4
1 PD5 VS2 D10 D11

GND
1G PD6 C12 Q15 BC640
CS3 19 2G 9
PD7 100nF TP6
D58 D60 R123 D63 1N5819 BAT49

10
R23 33.2
BAT49 BAT49 1K D62 1N4148

6
U6 TC4426 1 TF2 6 1N4148 Q2
+5 VDC 2 7

GNDVDD
C21 INA OUTA
C47 5
10uF C68 R122 R24 BC640 D12
+5 VDC +5 VDC 4 5 3 4
AR2 C22 INB OUTB
8x10K C30 1K 2K21 BAT49
100nF +5 VDC 220nF D59 D61
U17

3
40
87C54 BAT49 BAT49 C62 10
20
31 39 PD0 10uF

VCC
SW2 U14 HCT244 C43 EA/VPP P0.0(AD0) PD1
P0.1(AD1) 38
8 9 2 PD7 18pF PD2 VS1 D22 D23
1Y1 18 37
VCC

1A1 PD6 P0.2(AD2) PD3 C16 Q11 BC640


7 10 4 1A2 1Y2 16 19 X1 P0.3(AD3) 36 1
6 11 6 PD5 PD4 100nF TP10
1A3 1Y3 14 PD4 P0.4(AD4) 35
PD5 D34 D36 R115 D39 1N5819 BAT49
5 12 8 1A4 1Y4 12 P0.5(AD5) 34
4 13 11 PD3 X1 PD6 R31 33.2
2A1 2Y1 9 PD2 P0.6(AD6) 33
PD7 BAT49 BAT49 1K D38 1N4148
3 14 13 2A2 2Y2 7 P0.7(AD7) 32

6
2 15 15 PD1 C44 12MHz U10 TC4426 TF6 6 1N4148 Q6
2A3 2Y3 5 PD0 18pF
1
3
1 16 17 2Y4 3 2 7 3

GNDVDD
2A4 P11 INA OUTA
18 X2 5
1 C64 R114 R32 BC640 D24
GND

1G
CS2 19 2G P12 4 INB OUTB 5 3 4
1K 2K21 BAT49
9 21 220nF D35 D37
RSTuP RESET P2.0(A8) PWM
10

22

3
P2.1(A9) PWMC 2
29 23 BAT49 BAT49 C58
PSEN P2.2(A10) C1
24 10uF
P2.3(A11) C2
ALE 30 ALE/PROG P2.4(A12) 25 MODP1
26 VS2 D13 D14
P2.5(A13) MODP2
27 C13 Q14 BC640
P2.6(A14) MODA1 7
+5 VDC +5 VDC +5 VDC 28 100nF TP7
AR3 P2.7(A15) MODA2
8x10K C39 R86 D52 D54 R121 D57 1N5819 BAT49
100nF 10K2 R25 33.2
BAT49 BAT49 1K D56 1N4148

6
U7 TC4426 1 TF3 6 1N4148 Q3
2 7
20

GNDVDD
P21 INA OUTA
SW3 U18 HCT244 5
8 9 2 PD7 C67 R120 R26 BC640 D15
1Y1 18 1 10
VCC

1A1 PD6 A0 P1.0(T2) P3.0(RXD) T2


7 10 4 1A2 1Y2 16 A1 2 P1.1(T2EX) P3.1(TXD) 11 T3 P22 4 INB OUTB 5 3 4
6 11 6 PD5 1K 2K21 BAT49
1A3 1Y3 14 PD4 A2 3 P1.2 P3.2(INT0) 12 ST
220nF D53 D55
5 12 8 1A4 1Y4 12 RAW 4 P1.3 P3.3(INT1) 13 HS
4 13 11 PD3
2Y1 9 5 14

3
2A1 PD2 ENuP P1.4 P3.4(T0) FL 8
3 14 13 BAT49 BAT49 C61
2A2 2Y2 7 PD1 AUTO 6 P1.5 P3.5(T1) 15 SF
10uF
2 15 15 2Y3 5 7 16
GND

2A3 PD0 RD P1.6 P3.6(WR) SPARE1


1 16 17 2A4 2Y4 3 WR 8 P1.7 P3.7(RD) 17 SPARE2
VS1 D16 D17
1 C14 Q13 BC640
GND

1G 5
20

2
19 100nF TP8 2
CS1 2G D46 D48 R119 D51 1N5819 BAT49
R27 33.2
10

BAT49 BAT49 1K D50 1N4148

6
U8 TC4426 1 TF4 6 1N4148 Q4
2 7

GNDVDD
A11 INA OUTA
5
C66 R118 R28 BC640 D18
A12 4 INB OUTB 5 3 4
1K 2K21 BAT49
220nF D47 D49
+5 VDC +5 VDC

3
AR4 8x10K C40 BAT49 BAT49 C60 6
100nF 10uF
+15 VDC VS1 +15 VDC VS2
Q10 Q9 VS2 D7 D8
2N4403 2N4403 C11 Q16 BC640
20

SW4 U19 HCT244 100nF TP5 11


8 9 2 PD7 D64 D66 R125 D69 1N5819 BAT49
1Y1 18
VCC

1A1 PD6 R21 33.2


7 10 4 1A2 1Y2 16
6 11 6 PD5 R113 R2 BAT49 BAT49 1K D68 1N4148
1A3 1Y3 14

6
5 12 8 PD4 2K21 2K21 U5 TC4426 TF1 6 1N4148 Q1
1A4 1Y4 12 PD3
1
4 13 11 2Y1 9 A21 2 7

GNDVDD
2A1 PD2 INA OUTA
3 14 13 2A2 2Y2 7 5
2 15 15 PD1 C69 R124 R22 BC640 D9
2A3 2Y3 5 PD0
1 16 17 2A4 2Y4 3 A22 4 INB OUTB 5 3 4
R111 R112 R68 R17 1K 2K21 BAT49
1 220nF D65 D67
GND

1G 475 1K5 475 1K5


19
3

1 CS0 2G 12 1
BAT49 BAT49 C63
10uF
10

DZ3 DZ2
12V 12V
J NC 18/148 J.Mingo 30/10/18 NAME DATE SHEET / OF
I NC 17/0131 Carlos R. 30/03/17 DRAWING F. Díaz 12/06/08
A3243-04
1/3
H NC 14/0356 R. Hermosilla 24/11/15 REVISED A. Díaz 12/06/08 J I H G F E D REV
G NC 14/0356 Carlos R. 16/10/15
F NC 11/185 F. Díaz 30/05/11
E NC 09/204 F. Díaz 14/07/09 SEDECAL CONTROL DRAC
REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

PTR1 VBUS (4V<>724 VDC)


PD ( 0.......7 )
TP18
R63
1K

P+

N-
V
+15 VDC

R66 +5 VDC
22K1 +5 VDC
4x1N4148
D28 D27
C18 R34 R35 R36 R75 R65 D29
J1 470nF 1000V 1M 1M 1M 1M 604K 1N4148 U15

20
4 4
COM U12D D31 R74 A40103-01
3 U12C R67 LF347 1N4148 1K5 PD0
13 DB0 17

VCC
9 LF347 604K 14 3 PD1
CH1 DB1 16 PD2
8 12 DB2 15
SHL R40 R39 R38 R76 R64 10 R73 4 PD3
2 1M 1M 1M 1M 604K 10K2 CH2 DB3 14 PD4
R37 R33 DB4 13 PD5
5 CH3 DB5 12
150K-2w 150K-2w C31 PD6
100nF C36 DB6 11 PD7
6 DB7 9
J2 R72 100nF CH4
V+ D26 D25 8 2
2 VREF CS CS4
DL7 DL8 22K1 +5 VDC 1
RD RD
HLMP1700 HLMP1700 7 19

GND
AGND WR WR
INTR 18
V- C19 C17
1 -15 VDC C27

10
100uF 450V 100uF 450V 100nF

+15 VDC
C5 I.PRINC
PTR1 100nF TP17 TP36
R88
1K R60
U 1V<>8.26A 1K
U2 CSNH151 TP1
3 R8 R7 R10 R4 +5 VDC 3
V+

1-7 12K1 12K1 12K1 12K1 +5 VDC


OP D1
D3 1N4148 D73
J1
V-

8-14 1N4148 U1A D32 R1 R61 1N4148 D72 U1B


U1C U1D R3 2 LF347 1N4148 1K5 1K5 1N4148 LF347 6
PRIN C4 R9 9 LF347 13 LF347 221K 1 7
1 100nF 8 14 3 5
100 10 12 R5 R62
10K2 10K2
D2
R6 1N4148 C1
-15 VDC 12K1 1uF C35 C25
100nF 100nF

+15 VDC
C6 I.AUX
PTR1 100nF TP16
R89
1K
W
U3 CSNH151 TP3
R12 R14 R15 R19 +5 VDC
V+

8-14 12K1 12K1 12K1 12K1


OP D6
D5 1N4148
J1
V-

1-7 1N4148 U4B D33 R18


2
U4D U4C R16 6 LF347 1N4148 1K5 2

AUX C7 R11 13 LF347 9 LF347 221K 7


4 100nF 14 8 5
100 12 10 R20
10K2
D4
R13 1N4148 C10
-15 VDC 12K1 1uF C34
100nF

R53 R55 R57 R58


12K1 12K1 12K1 12K1

P+
PTR1 R54
12K1
D70
1N4148
SKM 40GD 123D
U12A U12B R59
2 LF347 6 LF347 221K
1 7
C28 +15 VDC C3 +15 VDC C8 +15 VDC 3 5
100nF 100nF 100nF
I.PRINC LIMIT
D71
1 3 5 R56 1N4148
12K1 C24
I.AUX LIMIT
1uF
2 4 6
U12D U1B U4A
4

U V W LF347 LF347 LF347


1 1
2
1
3
7 9 11
J NC 18/148 J.Mingo 30/10/18 NAME DATE SHEET / OF
11

11

11

I NC 17/0131 Carlos R. 30/03/17 DRAWING F. Díaz 12/06/08


A3243-04
8 10 12 2/3
C26 C2 C9 H NC 14/0356 R. Hermosilla 24/11/15 REVISED A. Díaz 12/06/08 J I H G F E D REV
100nF 100nF 100nF
G NC 14/0356 Carlos R. 16/10/15
N-
F NC 11/185 F. Díaz 30/05/11
-15 VDC -15 VDC -15 VDC E NC 09/204 F. Díaz 14/07/09 SEDECAL CONTROL DRAC
REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

R50 VUNR +5 VDC +5 VDC


1K5 U11B +5 VDC AR5 8x4K7
RSTuP PD ( 0.......7 )
4 - LM339 C38 C45
I. PRINC
2 C57 DZ1 R71 100nF 100nF
Limit
C22 5 + R94 2K2 J4
100nF 13V 1K5 TP30 OP1A TPL621-4
9 SPARE2
+15 VDC 1nF 1
R90 16 R93

18

52
SPARE2
23K7 U11D R69 +5 VDC
10 - LM339 C41 15 2K2
13 1K5 1nF 12 20 PD0 2 SPARE2 RTN
R126 C32 PD1 10
11 + 21
Q8 22 PD2 R92 2K2
4 10K2 100nF 2N4401 D30 R83 48 PD3 TP27 OP1B TPL621-4 SPARE1 4
R127 1N4148 221K PD4 11
49 3
23K7 U11A R70 R82 50 PD5 14 R91
PD6 SPARE1
6 - LM339 10K2 51
1 1K 56 54 PD7 13 2K2
R45 C33 7 + 4 12 SPARE1 RTN
C46 5 C11
-15 VDC 10K2 100nF I.LIM R98 2K2
+5 VDC TP11 1uF 32 C12
TP31 OP1C TPL621-4
13 SF
U11C R52 R49 Q7 R51 5
R47 8 - LM339 2K21 1K 2N4403 221 44 12 R97
C21 SF
1K5 14 33
9 + 45 11 2K2
I. AUX C22
Limit
R48 C52 6 14 SF RTN
C21 6 P11
NOTE: 4.5 V < >37 Amp. pick 221 10nF R96 2K2
1nF
for I. PRINC or I. AUX 7 P12
TP28 OP1D TPL621-4
15 FL
7
24 8 10 R95
ENuP P21 FL
TP29 TP25 TP21 TP20 23 9 9 2K2
AUTO P22
J3 +5 VDC TP14
U16 8 16 FL RTN
N.U. 68 10 A11
TP24 TP26 TP22 TP23 A3607-02 R102 2K2
+5 VDC 1 ALE 53 11 A12
TP32 OP2A TPL621-4
17 HS
1
+5 VDC 2 PWM 34 2 A21 HS 16 R101
C51
+5 VDC 3 PWMC 26 3 A22 15 2K2
3 470uF 2 18 HS RTN 3

+5 VDC 4 C1 17 30 CS4
R100 2K2
GND 5 C2 19 29 CS3
TP33 OP2B TPL621-4
19 ST
3
GND 6 MODP1 47 28 CS2 ST 14 R99

GND 7
C50
MODP2 59 27 CS1 13 2K2
4 20 ST RTN
GND 8
470uF +15 VDC TP15
MODA1 25 4 CS0
R106 2K2
+15 VDC 9 MODA2 46 TP34 OP2C TPL621-4
21 T3
5
+15 VDC 10 RAW 55 T3 12 R105

+15 VDC 11 A0 14 11 2K2


6 22 T3 RTN
+15 VDC 12
TP37 +5 VDC
R104 2K2
-15 VDC 13
TP35 OP2D TPL621-4
23 T2
7
-15 VDC 14 T2 10 R103
-15 VDC TP2 TP4 TP12 TP13 TP19 R109 R108 R107 R110
C49 9 2K2
2K21 2K21 2K21 2K21 8 24 T2 RTN
470uF
+15 VDC

6 +15 VDC
2
C55 C54 C53 C56 2

+15 VDC
1nF 1nF 1nF 18pF DL3 R43 3K3
VUNR +15 VDC RDY1
4
3 x HLMP1700 15 +5 VDC HLMP1700 RL2
A1
VUNR A2 16 NOTE: Connect RL2-2 pin RDY1 RTN
15 ERROR 7
to RL2-13 pin with a wire at
VUNR 16
DL6 DL5 DL4 C48 22uF DL1
Bottom Layer
HLMP1700 R44 82 PRME15002
5 RDY1 OUT
R85 R81 R79 U20
10

PRECH IF 17
R87 ULN2803 R46 10
22 3K3 3K3 3K3 R77
VCC

PRECH IF 18 11 Q8 I8 8 ERROR CODE 40 8


12 7 RDY1 39 750
Q7 I7 DISCH +5 VDC
13 Q6 I6 6 57
+5 VDC 14 5 PRECH 38 DL2 R41 750
Q5 I5 BRAKE
+5 VDC 19 15 Q4 I4 4 36 37 3
16 3 T3 43
Q3 I3 T2
+5 VDC 20 17 Q2 I2 2 42 HLMP1700
18 1 T1 41 R42 10
GND

Q1 I1
2 ERROR DRAC
+15 VDC C23
35

T3 IF R84 100nF
1

21 22
9

T3 IF 22 1 GND
1 1
3

U11A
T2 IF 23
R80
+
LM339
22
T2 IF 24 - J NC 18/148 J.Mingo 30/10/18 NAME DATE SHEET / OF
I NC 17/0131 Carlos R. 30/03/17 DRAWING F. Díaz 12/06/08
A3243-04
3/3
12

C20 H NC 14/0356 R. Hermosilla 24/11/15 REVISED A. Díaz 12/06/08 J I H G F E D REV


T1 IF 25
R78 100nF
NC 14/0356 Carlos R. 16/10/15
22 G
T1 IF 26 F NC 11/185 F. Díaz 30/05/11
T1 IF: T1 to T2 Commutation
T2 IF: Not Used -15 VDC E NC 09/204 F. Díaz 14/07/09 SEDECAL CONTROL DRAC
T3 IF: Break Control REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

V UNR +15 VDC +5 VDC

CR5 CR6 J1
1N4007 1N4007 +5 VDC
1
+5 VDC
2 +5 VDC

REG1 LM7815 REG2 LM7805 +5 VDC


3
VIN VOUT VIN VOUT

GND

GND
4
4 +5 VDC 4

CR1 CR2

1N4007 1N4007
J3 F1 GND
T 0.5 A
TR1 5
220 VAC SW C1 C3 C4
1 2200uF 10uF 10uF
6 GND
35V 35V 35V

15 VAC GND
7

220 VAC GND


8

15 VAC
F2 +15 VDC
T 0.5 A C2 C5 9
0 VAC 2200uF 10uF +15 VDC
2 35V 35V
10 +15 VDC

CR3 CR4 +15 VDC


R1 11
150K 1N4007 1N4007
2W DS1 +15 VDC

GND
3
12 3

HLMP1700
R2 VIN VOUT
150K REG3 LM7915
J2 2W
13 -15 VDC
F3
T 15 A FILT1
-DC IN 1 5 9 14 -15 VDC
CR7
1N4007
6 10 -15 VDC
CAR-DRAC BUS
V UNR
F4 -15 VDC V UNR 15
T 15 A 2 12
+DC IN V UNR

14
2 3 13 16
1

8
17

18
GND 3 +5 VDC +5 VDC
19

20 +5 VDC
2 2

+DC OUT 4 21 T3 IF

22 T3 IF
-DC OUT 5
T2 IF T1 IF: T1 to T2 Commutation
23 T2 IF: Not Used
T1 IF: Break Control
24 T2 IF
J5
25 T1 IF
1

26 T1 IF

CR9 CR8 CR12 CR13 CR14


STTH1212D STTH1212D
1N4007 1N4007 1N4007
J4
2
1 K1
C6
220nF
VR1 KT2
1.5KE440CA 2

1 1
3 3 K3
PCB: 3240-05D

G NC 09/301 F. Díaz 16/11/09 NAME DATE SHEET / OF


CR11 CR10
STTH1212D STTH1212D F NC 09/178 F. Díaz 09/06/09 DRAWING F. GARCIA 27-09-97 A3240-05
1/1
E CN 01/075 F. GARCIA 03-05-01 REVISED A. DIAZ 27-09-97 H G F E D C B A REV

C, D CN 98/090, 99/94 F. GARCIA 15-02-00


4

H NC 10/252 F. Díaz 28/05/10 SEDECAL INTERFACE DRAC-HF


REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

4 4

R1
680 ALL DIODES CL03-15
15w
IN-1

CR1 CR2 CR3 CR4 CR5 CR6

3 3

R2 C1 R3 C2
150K 10nF 150K 10nF
15w 12.5KV 15w 12.5KV

IN-2

2 2

1 1

NAME DATE SHEET / OF


DRAWING F. GARCIA 12-01-99 A3109-01
1/1
REVISED A. DIAZ 16-03-99 A REV

SEDECAL CLAMPING
A NC16/0089 R. Hermosilla 31/03/16
REV DESCRIPTION ISSUED BY DATE
A B C D E
8 7 6 5 4 3 2 1
+5V +12V +5V
AD[0..15] C96 C34
AD[0..15]
100nF 100nF
AD[0..15]
C60 AD[0..15]
+12V +5V +5V R182 R183 R184 R185 R186 R187 R188 R189
Q4 +5V
R107
MMBT4403 C62 33Pf U5

40
100nF
T89C51RD2 U12 U33A U34

28
6K81 J5B

9
X1 18 39 AD0 10K 10K 10K 10K 10K 10K 10K 10K AD7582 74HCT574 ULN2803
R110 R112 XTAL1 P0.0

VCC
1K C61 33Pf 20MHz 19 38 AD1 AD0 17 2 AD0 2 19 1 18 R146 22.1
4K99 XTAL2 P0.1 DB0 AIN0 mA D1 Q1 I1 Q1 9 -HS SCL

VCC
VDD
37 AD2 AD1 16 AD1 3 18 2 17 * R147 22.1
P0.2 DB1 D2 Q2 I2 Q2 7 -ACC
36 AD3 AD2 15 3 AD2 4 17 3 16 R148 22.1
R1061K R109 P0.3 DB2 AIN1 KVP D3 Q3 I3 Q3 6 -BRAKE
9 35 AD4 AD3 14 AD3 5 16 4 15 R149 22.1
RESET P0.4 DB3 D4 Q4 I4 Q4 5 -RTR RUN
34 AD5 AD4 13 4 AD4 6 15 5 14
D 100K P0.5 DB4 AIN2 FIL I D5 Q5 I5 Q5 D
Q5 31 33 AD6 AD5 12 AD5 7 14 6 13 R150 22.1
EA P0.6 DB5 D6 Q6 I6 Q6
C58 D22 C57 1uF 32 AD7 AD6 11 5 AD6 8 13 7 12
MMBT4401 P0.7 DB6 AIN3 RTR I D7 Q7 I7 Q7 8 -BAT TEST
MMSZ8V2T1 R108 AD7 10 AD7 9 12 8 11
100uF DB7 D8 Q8 I8 Q8 10 HI mA
825 4 24 AD11

GND
-DESEQ P1.3 A0
5 25 AD12 11
-KV SAFETY P1.4 A1 CLK
6 21 AD8 1 +12V R152 +12V
D21 R105 -mA SAFETY P1.5 P2.0 OC
22 AD9 21 23 1K5
825 +5V P2.1 BYSL CLK

9
23 AD10 18
IMBD4448 P2.2 RD
24 AD11 20 6
P2.3 WR VREF Vref R153 1K5

DGND

AGND
25 AD12 19 1
P2.4 CS CAZ

VSS
+5V R28 R154 R30 26 AD13 -12V
P2.5 D33
1K 1K 3K32 27 AD14
P2.6 BAT54S
28 AD15
C90
P2.7

27

7
Q10 100nF
RX R190
7 30 MMBT4403
R16 R18 P1.6 ALE/P 681
8
1K 1K TX P1.7
10
P3.0/RxD
+5V 11 17
P3.1/TxD P3.7/RD C97 C98
12 16 C99
DS2 DS3 U4 P3.2 P3.6/WR D40 100nF 100nF
1 16 13 3 2.2nF J1A
P3.3 P1.2
R20 475 14 2 U2A U20
P3.4 P1.1 MMSZ5231B *
J1B 2 15 15 1 74HCT574 ULN2803 R151 22.1
P3.5 P1.0 17 -TUBE3
3 14 +5V AD0 2 19 1 18 R9 22.1
D1 Q1 I1 Q1 10 HT-C CLK

VSS
R21 475 29 AD1 3 18 2 17 R10 22.1
-KV UP 1 PSEN D2 Q2 I2 Q2 11 HT-C DAT
4 13 AD2 4 17 3 16 R11 22.1
D3 Q3 I3 Q3 12 -TUBE2 SCL
5 12 C101 C100 R191 AD3 5 16 4 15 R12 22.1
-KV DWN 2 PR JP7 D38 D39 D4 Q4 I4 Q4 13 -FIL 1 SCL

20
AD4 6 15 5 14
R17 475 R22 475 A B 510pF 3.3nF 56K2 D5 Q5 I5 Q5
6 11 1MBD4448 1MBD4448 AD5 7 14 6 13
-PREP 3 D6 Q6 I6 Q6
7 10 AD6 8 13 7 12
CH D7 Q7 I7 Q7
C R23 475 AD7 9 12 8 11 C
C-HT CLK 4 D8 Q8 I8 Q8
8 9 +5V R13 22.1

GND
14 -START
11 R14 22.1
-WR DAC CLK 15 -PRE CHRG
1
OC 16 -CHRG DR
C12 +5V
C-HT DAT 5 R32 2K74

9
100nF +5V U13A
U21 74HCT138
-EXP 6 R31 3K32
1 16 15 1 AD8 +5V
-CS DAC Y0 A DS1
14 2 AD9
-CHRG 7 R24 475 -L DAC Y1 B R15 475
2 15 13 3 AD10
C59 Y2 C
3 14 100nF 12
-FIL 1 ACK 8 -EXP Y3 led
U3 11
R19 475 R25 475 Y4

8
4 13 24C16 10 4 AD13
Y5 G2A
5 12 1 6 9 5
A0 SCL Y6 G2B Q1

VCC
2 7 6 +5V
C14 R26 332 A1 Y7 G1 R8 475 IRLZ44NS D3
6 11 3 5 1MBD4448
1nF A2 SDA
7 10
R27 475
7
TEST GND
8 9
C15 C16
C10 R7
4

C13 1uF 100K


100nF 100nF
100nF

+5V +5V
U8 +5V
1 16
J1C R63 475
2 15
BATST 1 18
B 3 14 R174 R175 R176 R177 R178 R179 R180 R181 B
BATST 2 19
BATST 3 20 R64 475
4 13
GND 9
5 12
R166 R167 R168 R169 R170 R171 R172 R173 10K 10K 10K 10K 10K 10K 10K 10K
J3C R65 475
6 11 10K 10K 10K 10K 10K 10K 10K 10K U10A U11A
7 10 74HCT541 74HCT541 SW2
ERR DRAC 6
+5V 2 18 AD0 AD0 18 2 16 1
R66 475 A1 Y1 Y1 A1
8 9 3 17 AD1 AD1 17 3 15 2
A2 Y2 Y2 A2
4 16 AD2 AD2 16 4 14 3
A3 Y3 Y3 A3
+5V 5 15 AD3 AD3 15 5 13 4
R164 U9 A4 Y4 Y4 A4
3K32 6 14 AD4 AD4 14 6 12 5
A5 Y5 Y5 A5
1 16 7 13 AD5 AD5 13 7 11 6
A6 Y6 Y6 A6
J5A R68 475
8
A7 Y7
12 AD6 AD6 12
Y7 A7
8 10 7
2 15 9 11 AD7 AD7 11 9 9 8
-RTR RDY 4 A8 Y8 Y8 A8
3 14
R165
19 19
-IGBT FAULT 3 R69 475 C32 G2 G2
4 13 1 -RD 1
475 C95 100pF G1 G1
5 12
-BATT FAULT 2 100nF
R70 475
6 11
-AUTO OFF INH 1
7 10
R71 475
8 9
FAULT
TP9 TP17
J2 R43 4K02
TP1 TP10 TP18
-12V 1
+12V DS6
+12V 2 TP3
A TP4 GNDGNDGNDGNDGND A
+5V 3 R92 1K +12V R91 1K -12V
+5V TP2 +5V
NAME DATE SHEET / OF
GND 4 R42 PBA: A3000-8X PCB: 90455-02
R41 RHA 15/02/2018
1K62 R93 1K DRAWING
4K02 1 /4
GND 5 C42 C41 15/02/2018
C43 -12V REVISED F. Díaz B A REV
100nF
100nF 100uF
GND 6 DS5 C45 C44
C46 DS4
100nF 100uF 6/9/2017
100uF B NC18/0050 RHA
SEDECAL HT CONTROLLER
A First Edition RHA 30/12/2016
REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1

R44 16K5
U10B +5V U2B +5V U33B +5V
JP4 74HCT541 74HCT574 74HCT574
JP1 JP2 JP4 JP5 JP7 JP8 JP9

A
20 20 20
Vcc Vcc Vcc
HV TANK COMPACT B B B B A A A *

B
R45 36K5 TP16
1V = 33.3KVp
HV TANK VERTICAL A A A A A A A C22 100pF
KVP C31 C11 C89
100nF 100nF 100nF
TP7
+5V
D8 KVPF
10 10 10
R40 10K C24 100nF GND GND GND
D KVP_med D
KV -
IMBD4448 C21
+12V
10nF
KV + TP11
TP12 U7A C25
U6C

4
KV- KV+ LF347 U11B +5V U13B +5V
J4A * LMC6484

4
100nF 9 74HCT541 74HCT138
R1 1K R48 100K
2 D9 8 20 16
-KV 1 R52 10K KVP Vcc Vcc
* R2 1K R133 100K
1 10
3
+KV 2
IMBD4448
C33 C35

11
A B A B A B C27 100nF 100nF

11
C26
10nF
10 8
JP1 JP2 JP5 GND GND
-12V
* 100nF
C23
C5 C6 R46 R47
C3 C4 D1 D2
RTR I1 4 16K5 36K5
1nF 10nF 100pF
1nF 10nF
RTR I2 3
BZV55-C15
R158 34K

BZV55-C15 U18C
LF347
9
R61 8
10
6K04 R155 10K
Vref
* C55

C
JP3 10nF
1 16 C
EA1+ EA2+
STANDARD A +12V
2
3
EA1- EA2-
15
14
+12V
FDB REF U18B
SCAN B 4
5
DTC OC
13
12 6
LF347
TP6 R157 4K99 CT VCC
6 11 7
A RT CQ2
RTR 7 10 5
GND EQ2 C94 C93
D6 R29 C91 8 9
R3 R4 C92 CQ1 EQ1 470nF 100nF
BAT54S 10K
RTR I
JP3 4.7nF R156 100nF U19
10K 10K
10K TL594CD +5V
B
C84 C85
DIF_DEM F_inv_Kv => 36us to 40us (R29) 100nF 100nF U30C
R129
12

8
U27 7
* 22
R73 49K9 1 8 1
CP VCC J3A
KV DEM 2 7 7
D /SD
3 6 2 ULN2803A
/Q /RD 1 -KV DR 1

9
4 5
R62 * R143 4K99 GND Q 2 -KV DR 2
D20 5
49K9 IMBD4448 U30D
74LVC1G74DP 3
+12V R130
6 11
D18 C81
8
C53 100nF BZX384-C5V1 D19 1nF 22
+12V +12V U28

4
R144
+5V 74AHCT2G08
U18A ULN2803A
4

C51 100nF IMBD4448 10K

9
B LF347 B
3 R100
KV - U7C
1 10K R101
LF347
2 4K75 9
11

U17B
8
LM319J
9 + 10
11

7
C54 10
C56 - C30 +5V
R60
100nF 10nF
100nF C52 100K
6

-12V
100nF
R98 R97 R96 R159 R160 R161
R94
1K 3K3 1K
1K65
10K 3K01 10K
-12V D34
+5V
TP15 -KV SAFETY
U18D
LF347 U17A IMBD4448 D35
12
KV + R102 LM319J D37
14 4 -EXP
13
+
12
IMBD4448 J11
10K -DESEQ IMBD4448 D36
5 -
2 IGBT Fault
C48
1
R104 R103 C47 IMBD4448
3

100nF R95 Q11


10K R162
10K 3K01 100nF NDS7002A
R163 FAULT
100K 22

A3000-80 =>According to schematic. -12V


A A
A3000-81 =>D20 unmounted.
A3000-82 =>D20 unmounted, R73 86K6, R147 & R151 0ohm. NAME DATE SHEET / OF
A3000-83 =>D20 unmounted, R45 & R47 69K8, R58 & R56 226K, R59 & R54 25K5, R1 & R2 4K02. 15/02/2018
PBA: A3000-8X PCB: 90455-02
DRAWING RHA
A3000-84 =>D20 unmounted, adjust R29 to have a period of 48us. REVISED F. Díaz 15/02/2018
2 /4
B A REV
A3000-85 =>D20 unmounted, R55 & R57 200K, R38 2K21, D4 & D5 unmounted
A3000-86 =>D20 unmounted, adjust R29 to have a period of 48us, C55 100nF. 6/9/2017
A3000-87 =>D20 unmounted, adjust R29 to have a period of 45us, C55 100nF.
B NC18/0050 RHA
SEDECAL HT CONTROLLER
A First Edition RHA 30/12/2016
A3000-88 =>D20 unmounted, R147 & R151 0ohm. REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1

1V = 10mA (FROM 10mA UP TO 80 mA)


ALL A3000-8X EXCEPT A3000-85

6
+12V C49
1V = 100mA (FROM 100mA UP TO maximum mA) 100nF
U16A
+5V 1V = 20mA (FROM 10mA UP TO 80 mA)
DG300ACWE
AD[0..15] ONLY A3000-85

16
AD[0..15]
* R59 1V = 200mA (FROM 100mA UP TO maximum mA)
4 2
R99
10K 12K7 U16C
TP5
D
AD15 * R58 DG300ACWE D
mA
113K
mA_med
C29 10nF
TP14 TP13 C20 C50
10nF 100nF

10
mA + mA - D11 R39 U6A
J4B 10K LMC6484
R33 1K * R57 100K
2
13 IMBD4448 D10 * 1
-mA 6 R38 10K mA
* 14 3 -12V
R34 1K R55 100K
12
+mA 7
U7D IMBD4448
LF347
R37
TP8
* R56 C28 * 8K87
R54 FIL I
C1
C2
100nF C18 * D4 * C17 113K U16B 12K7
D5 10nF
100nF 100nF U6D
500V 100nF 1SMA5942BT3 1SMA5942BT3 DG300ACWE
500V FIL_med R36 LMC6484
13
13 15 1K 14
R51 10K FIL I
12
U6B
LMC6484
+12V 6
C7 7
100nF 5
U1

13
AD637JR

11
+12V R5
4K75 C78

+VS
C 100nF C
R138 49K9
5 16
CS BUFF OUT
R6 10K U7B
7 11
D7 dB OUTPUT RMS OUT LF347
6
R35 511 BAT54S
15 6 7
FIL I 8 VIN DEN INPUT C9 R137 10K
5
4 1uF
GND 5 OUTPUT OFFSET R140 10K
C19 10
CAV
1
BUFF IN R139 10K
10nF

GND

-VS
-12V 1 16
3 C8 EA1+ EA2+

12
2 15 +12V
EA1- EA2-
3 14
FDB REF
4 13
100nF R134 DTC OC
4K99 5 12
CT VCC
-12V 6 11
RT CQ2
-12V +12V 7 10
C88 GND EQ2 C79 C77
100nF C87 8 9
100nF R49 CQ1 EQ1 470nF 100nF
20K
C86 R136 C75 U23
R135 C74 C76 TL594CD +5V
100nF 205 100nF
7

TP19 TP20 10K 100nF 10nF


AD0 24 3 KV DEM
DB0 VrefA
Vss

Vdd

AD1 23 10 IFIL DEM


DB1 VrefB C82 C83
AD2 22 U31
DB2 U30A
AD3 21 4 8 F_inv_Fil => 79us to 81us (R49) 100nF 100nF
DB3 R131
B AD4 20 R50 R53 18 B
DB4

8
AD5 19 LM336M-5,0 1K 1K U25 1
DB5 22
AD6 18 12 1 8 1
DB6 RfbB CP VCC
AD7 17
DB7 VoutB
9
KV DEM
2
D /SD
7 7 J3B
AD11 16 3 6 2 ULN2803A
A0 /Q /RD 4 -FIL DR 1

9
AD12 15 4 5
A1 R141 GND Q 5 -FIL DR 2
2 5
RfbA 4K99
4 74LVC1G74DP 3
VoutA U30B
6
C80 R132
AD[0..15] 14 17
AD[0..15] LDAC 1nF U26
DGND

13 5 R145 2
WR AGDNA 22

4
1 8 1K54 R142 74AHCT2G08
CS AGNDB
10K
ULN2803A
-L DAC U32
11

9
AD7837
-12V
-WR DAC

-CS DAC

-mA SAFETY

A A

NAME DATE SHEET / OF


15/02/2018
PBA: A3000-8X PCB: 90455-02
DRAWING RHA
3 /4
REVISED F. Díaz 15/02/2018 B A REV

B NC18/0050 RHA 6/9/2017 HT CONTROLLER


SEDECAL
A First Edition RHA 30/12/2016
REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
8 7 6 5 4 3 2 1

+12V +12V

C64 R113
C39 R79
100nF 287
100nF 287
U22A U15A

4
D R115 10K LF347 R81 10K LF347 D
3 3
KVP_med FIL_med Q2
1 Q6 1
2 MMBT4401 2 MMBT4401
D23 D12
C63 IMBD4448 D25 C36 IMBD4448
10nF 10nF D15
BAT54S
R114 R80 BAT54S

1K 1K
R119 R86
R120
10K J8 R87
10K J7
10K 1 10K 1
2 Kvp U15B 2 Filament current
3 LF347 3
6
R121 7
U22B R88
10K 5
LF347 10K
6
7
D26 D16
5
R123 BAT54S R90 10K BAT54S

10K
D27 D17
U22C U15C
IMBD4448 IMBD4448
LF347 U15D LF347
9 9
Q8 LF347 Q3
8 13 8
U22D MMBT4403
10 14 10 MMBT4403
LF347
13 12
14

11

11
C 12 R122 R89 C
287
287

C65 C40
100nF 100nF
-12V -12V

+12V
All components of this page do not mount

C67 R116
100nF 287
U29A
4

+12V
U29B R118 10K LF347
3
LF347 mA_med Q7
6 1
7 2 MMBT4401
5 D24
C66 IMBD4448 C37
10nF D28
R117 BAT54S 100nF
U14D
1K LF347
D13 13
R75 R76 10K
R124 BAT54S 14
U29D R125
10K J9 R83 R82
12
LF347 20K

4
13 U14A
B 14
10K 1
mA J6 1K 1K
2 LF347 B
2
12 1
3 1 DIF_DEM
Input dem. 3
2
R126 3
U14C
10K

11
9 LF347
D14 R78
8
D29 R85 BAT54S R84
10
R128 20K
+5V BAT54S R77 C38
1K 1K
10K 100nF
10K
U29C D30
C73 C70 U24 LF347
9 IMBD4448
100nF 1uF 16 MAX232 Q9
8
2 1 10 MMBT4403
V+ C1+
C69
1uF
11

6 3 R127
V- C1-
4 287 -12V
C72 C2+
C71 U14B
1uF
6 LF347
1uF
5 7
C2- C68 5
100nF
14 11 JP8 -12V
A
13 12 B
J10 CH
7 10
1 TX
2
A 8 9 JP9
D
GND

3 A
A
B
PR
1

NAME DATE SHEET / OF


D31 D32 PBA: A3000-8X PCB: 90455-02
15

PSOT12C PSOT12C RX DRAWING RHA 15/02/2018


4 /4
3 3 REVISED F. Díaz 15/02/2018 B A REV

B NC18/0050 RHA 6/9/2017 HT CONTROLLER


SEDECAL
A First Edition RHA 30/12/2016
2

REV DESCRIPTION ISSUED BY DATE

8 7 6 5 4 3 2 1
5 4 3 2 1

NOTE 1: W1-A for version A3004-10/11/12, W1-B for version A3004-09


NOTE 1: W1-A para versión A3004-10/11/12, W1-B para versión A3004-09

NOTE 2: R7= 2K67 ohms for A3004-09/10


NTC NTC R7= 1K4 ohms for A3004-11
R7= 1K ohms for A3004-12

NOTE 2: R7= 2K67 ohms para A3004-09/10


Tº Tº R7= 1K4 ohms para A3004-11
R18 R19 R7= 1K ohms para A3004-12
D D

R20
+15V 100K 3W

D2
1N4148
+5V
+24V
+5V R5 DS1
TP7 TP6 33
R17 C6 Green
3K74 TP9 C12
100nF 100nF R2
R15 R12 D4 100
R9
P1 8 10K 10K Q2
UF4006

9
GND R16 10K U3 SGP23N60UFD + C10 + C11
1 2K D7
GND 1 7

VDD
2 UF4006 HO 330uF/450V 330uF/450V
+5V 3 +24V
7 10 HIN
+5V 4 2 VB 6
C8
-FIL DR1 5 3
6 12 5 1uF C1 P3
-FIL DR2 6 R14 R13 D6 4
LIN VS 470nF/630Vdc
FIL I 7 3K74 2K UF4006 FIL SUP
DELAY 8 4
VCC 3 3
5
+24V U5 11 SD 2 FIL PREHEAT
HCPL2232
LO 1 1 FIL RTN
T1

VSS
C TP8 2 D1 1 C
R11 COM 1N4148 2
D5 1K IR2113 C7 Pulse. PE-67300. 1 : 300

13
2 5 1uF 4 3

BZX55C15V U4
R10 H11AA1 R1
10K 100
1 4

Q1
SGP23N60UFD

R6
2K43

W1

A
NOTE 1 1V=0.31 Arms

B
TP4

D3
TP3 1N4148
+24V +15V R7
R8 NOTE 2
10 +5V
B
U2 R4 U1 TP1
B
10K LM7805
1 IN OUT 3
+IN +OUT
R3
+ GND + 10K
+ C4 C5 C3 C2
C13 C9 10uF/50V 100nF 2 100nF TP2 TP5
1.5uF/250V-X2 10uF/50V
P2 10uF/50V -IN -OUT
2

+SUPPLY 4
3 EC4BE13
2
1
-SUPPLY
1

L1
C15 C14

10nF/250V-Y2 10nF/250V-Y2

Ground stud
A A

PCB 90068-04

NAME DATE SHEET / OF


F SCH error F. Díaz 18/12/15 DRAWING F. Díaz 26/10/04 A3004-09/10/11/12
1/1
F NC 09/347 F. Díaz 26/11/09 REVISED A. Díaz 26/10/04 F E Db1 Cb1 Bb1 Ab1 b2 b1 REV

E NC 09/162 R. Asenjo 05/06/09


Db1 NC 06/251 F. Díaz 01/12/06
Cb1 NC 05/112 SEDECAL Filament Driver HC
REV DESCRIPTION ISSUED BY DATE
5 4 3 2 1
A B C D E

P6

GND 4
+11V PERM 3
-PWR ON 2
-PWR OFF 1 P3
DS1
3 LINE SYNC
T2 D4
P4 5

2 13 +24VDC
0 VAC 1 R15 R13 +24VDC
4 K1 4
K1 1 6 D16 15 D15 27 D3 1
9 6 1 4 +24 VDC
3 14
115 VAC 5W 3 5 R7 W2 2
1 15K C22
115 VAC 2 4 3 R14 U1 C2 C23 100nF
4 C4 1K5 D ULN2803A 3 10nF 470uF
470uF 11
DB104 Q1 1 GND
8
OFF G MTD3055
R12 R8 2 GND
P1
10 D14 S 3K3
8 K1

9
-PWR OFF 8 12
ON R11 4
-PWR ON 7 C3 3K3
100nF P4
8 K5
12
8 BUCKY 1 DR
P4 4

8 K4
BUCKY SPLY 9 12
7 BUCKY 2 DR
4
W1
AUX BUCKY SPLY 5
1 2 3

+24VDC
3 8 K3 3
L3 12
+24 VDC 4 6 -BUCKY EXP
1000uH K6 1 4
D11 D10 9 K7 4 L1 L2
(4) 13 13 13 13 13 5 12 500uH 500uH
D9 K3 K6 8
K5 K4 K7
(4) 14 14 14 D12 14 D13 14 K7 1
FIL 2 RTN
9 16
-BUCKY 1 DR CMD 12 5
-BUCKY 2 DR CMD 13 15 FIL 1 RTN
K5 1 C6
9 4 1
-BUCKY 1 MOTION 11 22 -FIL 1 ACK
5

4
2.2nF 600Vac R17
K4 1 6 3 499
21 SPARE
9 D17 C5 R16 U2
-BUCKY 2 MOTION 10 5 K9 4 D18 1N4738 2.2uF 1K AQV210EH
12 DB104
+24VDC 8
-TUBE 2 SCL 18 20 SOL DR

6
-FIL 1 SCL 19
FIL PREHEAT 17 13
FIL RTN 14
K9
D19
14
P5
P5
K8 1
2
9 2
DISCHARGE 2 13 5
10 DISCHARGE 2 RTN

K8 4
12
DISCHARGE 1 12 8
14 DISCHARGE 1 RTN
+24 VDC 16
14
R9 K8
D7
13
47
GND 9

K2 1
9 3 SPARE
+24 VDC 5 5
2 DELAYED +24 VDC
+5 VDC OUT 4 4
K2 7 SPARE
+24VDC 12
+24 VDC 8 8
6 +5 VDC
-LINE CONT 1 NOTES:
14 1.- FILAMENT 1 IS SMALL FOR RAD AND LARGE FOR SF
VERSION USED FOR COMPONENTS ON SHEET 2 L1, L2, L3, K6
+24VDC R3 R4 U1 D8 2.- HIGH VOLTAGE SUPPLY ON AEC OR ABS UNITS ONLY
K2 A3009-09 RAD + HV REQUIRED NOT REQUIRED 3.- LI, L2, L3 AND K6 NOT USED ON RAD SYSTEM
3K3 U1 10K ULN2803A
D1 D2 R2 ULN2803A 12 13 A3009-10 R&F NOT REQUIRED REQUIRED 4.- K5 AND D11 NOT PRESENT IN A3009-15
C1 1N4733 10K 16 7
100uF A3009-11 ONLY RAD NOT REQUIRED NOT REQUIRED NOTAS:
3
1 A3009-12 R&F + HV REQUIRED REQUIRED 1.- FILAMENTO 1 ES FINO PARA RAD Y GRUESO PARA SF 1
D6 2.- LA FUENTE DE ALTA TENSION SOLO PARA UNIDADES CON AEC O ABS
A3009-15 MAMO NOT REQUIRED NOT REQUIRED 3.- L1, L2, L3 Y K6 NO SE USAN EN SISTEMAS RAD
U1
D5 R5 ULN2803A 4.- K5 Y D11 NO SE MONTAN EN A3009-15
BZX55C10V 10K 17
NAME DATE SHEET / OF
2
O NC 16/208 MAguado 11/09/16 DRAWING F. Díaz 14/02/06
A3009-09/10/11/12/15
R6 R1 1/2
N NC 14/118 M.González 16/04/14 REVISED A. Díaz 14/02/06 O N M L K Jb1 I b2 I b1 REV
100K 2M2
M NC 13/098 M.González 13/03/13
L NC 09/360 F. Díaz 30/11/09
K NC 08/102 F. Díaz 23/04/08 SEDECAL INTERFACE CONTROL
REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

+24VDC

R10
56
C7 5w W7
100 nF
B A
2 U3 D22
4 T1 RGP02-16E 4
PE6189
Vs
5 In1 Out1 3 1

C10 7 C19 R35


2.2 uF C8 R33 10nF 1M
7 In2 Boot1 4
2 33K 1KV 1w
1 1w W6
Out2 2.2uF
9 Vref 8
C9 63V A B
11 8 2.2 uF 3
En Boot2 C18 R34
P1 R42 1KV 1M
22 GND Sense D21 10nF 1w
L6203 RGP02-16E
+12 VDC 6 6 10

+12 VDC 5
W8 C21 C15 R28
100uF 100nF 6K8
PT CRL 4
B A U5 P2
TL594
C14 6 12
R41 RT V+ 1 PT SPLY
R39 2,2nF 8
6K04 10K C1 +Vref
5 CT C2 11
D23
E1 9
16 +2 E2 10

1N4148 15 14
3 -2 REF 3 3
R40
3 FB DC 13
1K C20 R24 R26
100nF R43 R29 C16 R30 R31 1 4 10K 10K
10K 22K1 100nF 2M2 100K +1 DT
2 -1 GND 7

R23
10K R18 R19
10K 10K 5
-ON 3

R32
R22 W4 47K5
GND 2 GND 10K
A B R25 R27
10K 332
C13
GND 1 100nF
7
C17
100pF

6 ROOM LIGHT SUP


R38 R37 R36
4M75 4M75 4M75 K10 4
A B 12
2
8 2
W5 +Vref
C12 K10 1
100nF +24VDC 9
5
4 ROOM LIGHT SW
A B
13
R21 W3 K10
10K 8 7 D20
14
2
6 SD
2 -ROOM LIGHT
3 R20
U4A 5K
TLV2460
4 C11
100nF

NOTE.- BE SURE THAT THE VALUE OF THE PHOTO TUBE SPLY OUTPUT IN P2-1

PHOTO TUBE SPLY OUTPUT ( P2-1) IS THE REQUIRED BEFORE CONECTING THE SYSTEM INTERFACE

+500 VDC OR +300 VDC : JUMPERS W3 TO W8 IN POS "A", AND ADJUSTING WITH R20 NOTA.- ASEGURESE DE QUE EL VALOR DE LA SALIDA PHOTO TUBE SPLY EN P2-1
VARIABLE NEGATIVE OUTPUT : JUMPERS W3 TO W10 IN POS "B" ES EL REQUERIDO ANTES DE HACER LA INTERCONEXION DEL SISTEMA

1 1
SALIDA PHOTO TUBE SPLY (P2-1)

+500 VDC O +300 VDC : PUENTES W3 A W8 EN POS "A",Y AJUSTAR CON R20
NAME DATE SHEET / OF
SALIDA NEGATIVA VARIABLE : PUENTES W3 A W10 EN POS "B" A3009-09/10/11/12/15
O NC 16/208 MAguado 11/09/16 DRAWING F. Díaz 14/02/06
2/2
N NC 14/118 M.González 16/04/14 REVISED A. Díaz 14/02/06 O N M L K Jb1 I b2 I b1 REV
M NC 13/098 M.González 13/03/13
L NC 09/360 F. Díaz 30/11/09
K NC 08/102 F. Díaz 23/04/08 SEDECAL INTERFACE CONTROL
REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

U2 R2
HCPL4503 20K C7
2 8 100nF

4 CR3 6 4
1N4007
3 5

U5
8 +Vin +Vo 3
C3 9 C1 J3
47 uF +Vin 470 uF
50V 10 50V
-Vin 2 SPR (+5V INPUT 1)
11 -Vin -Vo 1
3 CP1 (-KV DR1)
VLA 106-24151
4 VPC (GND 1)

1 VPI (+15V OUTPUT 1)


J2
11 14
+24Vdc 1 24 C6 U1 5 FPD (-FAULT 1)
1 C5 10 uF 35V HCPL4503
3 10 15 22
470 uF
50V TEL3-2422 C10
8 2
0Vdc 12 13 10 uF 35V 6
2 2 23
5 3
3 3

3
TP5

-FAULT
4

U4 R3
HCPL4503 20K C8
2 8 100nF

CR4 6
1N4007 J4
3 5
2 SNR (+5V INPUT 2)

3 CN1 (-KV DR2)

4 VNC (GND 2)

U6 1 VNI (+15V OUTPUT 2)


8 +Vin +Vo 3 5 FND (-FAULT 2)
C4 9 C2
47 uF +Vin 470 uF
50V 10 50V
-Vin
2 11 -Vin -Vo 1 2

VLA 106-24151

J1
C9
+5V 100nF R5 R6 U3
4 200 200 HCPL4503
TP3
8 2
-DR1 6
1
TP4
5 3
-DR2
2
TP1 TP2

GND
3

1 1

C.I. 90132-06
NAME DATE SHEET / OF
DRAWING F. Díaz 20/06/11
A3063-06
1/1
REVISED A. Díaz 20/06/11 B A REV

B NC 13/001 F. Díaz 04/01/13 SEDECAL IPM DRIVER


REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

4
+ 4

R2 R3 R4 R5 R1 R6
100K 3W 10K 15W 10K 15W 10K 15W 39K 15W 27K 4W

CR2

TO MAIN STORAGE BZX85C56V

CAPACITORS Q1
IRF840
DS1

CR3
BZX85C56V
CR1
3 1N4733A 3

U1 R7
H11B1 3M9
1 6 P1
5 2 CHRG

- 2 4 1 GND

P2
2 DISCHARGE RTN

1 DISCHARGE

2 2

1 1

E CN 98/114 F.GARCIA 10/10/98 ENG F. GARCIA

D CN 98/057 F.GARCIA 29/04/98 CHARGE/DISCHARGE MONITOR CHK A. DIAZ

REV CN99/80 (22/07/99)


C CN 96/033 F.GARCIA 06/03/96
3212-01 REV. F
B CN 95/080 F.GARCIA 10/05/95 DWG
A CN 95/74 F.GARCIA 25/04/95 SEDECAL S.A. REV.

REV.
REV DESCRIPTION ISSUED BY DATE DATE 11/01/95
A B C D E
A B C D E

EXT SYNC K4
from J2-14 14 HT-C CLK AEC STOP
7, 8 +5V
LINE SYNC JP4 1
from J3-9 A B ABC IN U19F C55
K5 TPx FL-C CLK R34 R5 R6 R7 74HCT14 100nF
CAM SYNC +5V U33A 1K 10K 10K 2K55 13 12 U29
from J4-9 1, 14 7, 8 U32A 74HCT32 C59 8259A
+12V ISO 74HCT00 1 100nF 18 11 AD0
U19E IR0 D0 AD1
1 3 D1 10
3 2 RP16 U27 74HCT14 CLK 500 19 9 AD2
2 13 470 2501 IR1 D2 AD3
2 11 10 D3 8
CR15 K4 K5 U33B 2 16 -EXPT 20 7 AD4
1N4148 74HCT32 1 2 K1 C1 IR2 D4 AD5
4
D5 6 4
6 6 4 4 14 U19D 21 5 AD6
-DS PREP 3 4 K2 C2 74HCT14 IR3 D6 AD7
6 22 IR4 D7 4
5 5 6 6 K3 C3 12 9 8 23 IR5
U33C 8 10 TMR 1 24
-INTEN 74HCT32 7 8 K4 C4 C60 IR6
MASTER RESET 9 100nF UART RQ 25 12
+5V IR7 CAS0
8
C51 10 A9 1 13
1nF R14 -MWR CS CAS1
+5V U30 10K U33D -MRD A0 27 15
U28 CP80C88-2Z 74HCT32 A0 CAS2
71084C 21 28 12 -IOWR 2
R9 RESET IO/M -IORD WR
4 RDY1 RDY 5 22 READY WR 29 11 3 RD EN 16
CR7 10K 6 8 R28 330 19 32 13
1N4148 RDY2 CLK CLK RD -IOWR -INTA 26 INTA INT 17
11 17 U32C
R X1 U32B 74HCT00 -IORD INTR
3 XT1 C42 74HCT00 9
C50 AEN1 10MHz 33pF DATA BUS -INTEN
7 AEN2 4 8
10uF 13 16 26 6 10 U20
16V F/C X2 DEN 74HCT245
15 ASYN 5
1 2 C43 19
CSYNPCLK 33pF G
DT/R 27 1 DIR
+5V 33 16 2 18 D0
MN/MX AD0 A0 B0 D1
23 T AD1 15 3 A1 B1 17
U34 14 4 16 D2
74HCT573 AD2 A2 B2 D3
AD3 13 5 A3 B3 15
3 A15 19 2 39 12 6 14 D4 3
A16 Q1 D1 A15 AD4 A4 B4 D5 U19 U20 U21
18 Q2 D2 3 38 A16 AD5 11 7 A5 B5 13
A17 17 4 37 10 8 12 D6 74HCT14 +5V 74HCT245 +5V 8253 +5V
A18 Q3 D3 A17 AD6 A6 B6 D7
16 Q4 D4 5 36 A18 AD7 9 9 A7 B7 11 Vcc 14 Vcc 20 Vcc 24
15 Q5 D5 6 31 HLD
14 7 17 DATA BUS
Q6 D6 NMI C17 C20 C18
13 Q7 D7 8 1 GND
12 9 8 100nF 100nF 100nF
Q8 D8 A8
A9 7 GND 7 GND 10 GND 12
C 11 25 ALE A10 6
1 5 A0..........A7
+5V OE A11 U23
A12 4
INTR 18 3 MK48T08 U22 U23 U24
INTR A13 A0 D0 74HCT573 +5V MK48T08 +5V 27C256 +5V
A14 2 10 A0 D0 11
R18 -INTA 24 35 AD0 A1 9 12 D1 20 28 32
10K INTA A19 AD1 A2 A1 D1 D2 Vcc Vcc Vcc
8 A2 D2 13
AD2 A3 7 15 D3
AD3 A4 A3 D3 D4 C22 C25 C26
6 A4 D4 16
1.665 MHz AD4 A5 5 17 D5 100nF 100nF 100nF
CLK AD5 A6 A5 D5 D6
4 A6 D6 18 GND 10 GND 14 GND 16
AD6 A7 3 19 D7
AD7 A8 A7 D7
25 A8
U26 A9 24
74HC4040 A10 A9 U26 U28 U29
21 A10
10 9 A11 23 74HC4040 +5V 71084C +5V 8259A +5V
CLK Q1 A12 A11
Q2 7 2 A12 Vcc 16 Vcc 18 Vcc 28
11 RST Q3 6
5 U22 20 +5V
Q4 74HCT573 CS C41 C44 C45
2 Q5 3 27 WR E2 26 2
2 2 19 A0 22 100nF 100nF 100nF
Q6 13KHz D1 Q1 A1 OE
Q7 4 3 D2 Q2 18 GND 8 GND 9 GND 14
13 6.5KHz 4 17 A2
Q8 D3 Q3 A3 U24
Q9 12 5 D4 Q4 16
14 6 15 A4 27C512/27C1001/27C4001
Q10 D5 Q5 A5 A0 D0 U30 U31 U32
Q11 15 7 D6 Q6 14 12 A0 D0 13
1 8 13 A6 A1 11 A1 D1 8088 +5V 74HCT573 +5V 74HCT00 +5V
Q12 U21 D7 Q7 A7 A2 D1 14 D2
9 D8 Q8 12 10 A2 D2 15 Vcc 40 Vcc 20 Vcc 14
8253 A3 9 A3 D3
CLK 500 500Hz A0 A4 D3 17 D4
19 A0 11 C 8 A4 D4 18
to U29-19 8 D0 1 A5 7 A5 D5 C46 C48 C27
A1 D0 D1 OE A6 D5 19 D6 100nF 100nF 100nF
20 A1 D1 7 6 A6 D6 20
U38 6 D2 A7 5 A7 D7
74HC4040 -IORD D2 D3 U31 A8 D7 21 GND 20 GND 10 GND 7
22 RD D3 5 27 A8
10 9 4 D4 74HCT573 A9 26 A9 to U34-17 (sheet 1)
CLK Q1 -IOWR D4 D5 A8 A10 A17
Q2 7 23 WR D5 3 2 D1 Q1 19 23 A10 NC 30
11 6 2 D6 3 18 A9 A11 25 A11 U33 U34 U38
RST Q3 A8 D6 D7 D2 Q2 A10 A12 A18 74HCT32 +5V 74HCT573 +5V 74HC4040 +5V
Q4 5 21 CS D7 1 4 D3 Q3 17 4 A12 PGM 31
3 5 16 A11 A13 28 A13 to U34-16 (sheet 1) 14 20 16
Q5 R21 D4 Q4 A12 A14 Vcc Vcc Vcc
Q6 2 18 CLK2 6 D5 Q5 15 29 A14
4 100 15 7 14 A13 3 A15
Q7 CLK1 D6 Q6 A14 A16 +5V C28 C47 C61
Q8 13 9 CLK0 8 D7 Q7 13 2 A16
12 11 10 9 12 24 OE 100nF 100nF 100nF
Q9 C36 G0 OUT0 D8 Q8 Vpp 1
Q10 14 14 G1 OUT1 13 22 CE GND 7 GND 10 GND 8
15 1nF 16 17 11
Q11 G2 OUT2 C
Q12 1 1 OE 12
11
TMR 1 13
1 to U29-24 1
U32D
-EXPT 74HCT00
A15
+5V to J12A-16, J12B-16 and U34-19
D V NC 18/0119 Carlos R. 28/03/19 NAME DATE SHEET / OF
R13 U NC 13/257 R. Hermosilla 19/02/14 DRAWING F. Díaz 09/02/11
A3024-2XX
A1 + 22 G 1/4
T NC 13/417 R. Hermosilla 21/01/14 REVISED A. Díaz 09/02/11 V U T S W REV
SPK CR6 S
1N4148 Q4 S NC 12/278 F. Díaz 25/06/12
MTD3055

W NC19/0078 R. Hermosilla 30/07/19 SEDECAL ATP CONSOLE


REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

J15
2 PREP V ON
3 EXP TP2
V ON 4 R1 CR1
1 COMM 475 1N4007
K3
K1 K2 +12V ISO R29
1, 14 7, 8 1, 14 7, 8 1, 14 7, 8 J14 1K V ON

+12V ISO R4 1 COMM Q1 Q2


330 2 PREP MTD3055 D MTD3055 D
3 EXP C3
13 13 13 4 POWER OFF ON G OFF G 100uF C4 CR2
4 K1 K3 S S 100nF 1N4733 4
K2 5 POWER ON
CR5 CR4 CR3
1N4148 1N4148 1N4148 R3 V ON 6 V ( ON )
6 6 6
330
R2 J1
100
-PWR OFF 4
-AUTO OFF C39
-PWR ON 5
JP1, JP2, AND JP3 ON to J2-21 (page 3) 100nF
GND 7
POSITION "A" FOR -12V 9
NOVEL CONSOLE C38
+12V 3
100nF
A B A B A B +12V UNREG 8 +12V UNREG +12V ISO +5V +12V -12V
+12V UNREG 2 TP9 TP10 TP3 TP7 TP8
GND UNREG 6
JP1 JP2 JP3 U25
GND UNREG 1
COMM PREP EXP A9563-01
D0 12 4 R33 R31 R32 R35 R36
D1 D0 IRQ 1K 1K 1K 1K 1K
13 D1 CN/ST 37
D2 14 36 +12V UNREG REG2 REG1
D3 D2 SHFT U17 DC-DC CONVERTER +12V ISO 78SR105 +5V +12V -12V
15 D3
D4 16 74HCT138
D5 D4 -COL0 +Vin +Vout +Vin +Vout
17 D5 SL0 32 1 A Y0 15
D6 18 33 2 14 -COL1

GND
D7 D6 SL1 B Y1 -COL2
19 D7 SL2 34 3 C Y2 13 -Vin -Vout
35 12 -COL3 C9 C54 C56 C2 C57 C1 C6 C5 C7 C8
V ON SL3 +5V Y3 -COL4 470uF 470uF 470uF 100nF 2200uF 10uF 10uF 100nF 10uF 100nF
Y4 11
-IOWR 11 6 10 -COL5 TP11
WR G1 Y5 GND UNREG
4 9
J9 -IORD 10 5
G2A Y6
7 TP6 PT CRL
RD G2B Y7 TP1 J12A-J12B J1
11-12 BD 23
3 V (ON) -COL0 A12 J11 GND PT CRL 3
1-2 22 CS 1 PT CRL
COL 0 -COL1
COL 1 3-4 -COL2 A0 8 SL2
A0
AEC
21
COL 2 5-6 -COL3 A0 7 SL1 3 A0 CONTROL
COL 3 7-8 5 SL0
COL 4 9-10
-COL4 CLK 3 CLK OA0 27 15 CS1
A1
4 A1
BOARD
-COL5 26
COL 5 13-14 OA1 16 CS2
ROW 0 23-24 38 RL0 OA2 25 13 CS3
-IOWR
5 -IDWR
(A3012-11/12/15/17)
27-28 39 RL1 OA3 24 14 CS4 DATA BUS
ROW 1 -KV UP
ROW 2 29-30 1 RL2 OB0 31 11 CS5 6 -KV UP
25-26 2 RL3 OB1 30 12 CS6
ROW 3 -KV DWN
ROW 4 15-16 5 RL4 OB2 29 9 CS7 19 -KV DWN
ROW 5 17-18 6 RL5 OB3 28 10 CS8 2
7 GND
ROW 6 19-20 RL6
ROW 7 21-22 8 RL7 RESET 9
D0
8 D0
+5V D1
POWER OFF 39-40 D2 7 D1
6 +5V 10 D2
2 9 D3
POWER ON 37-38 D4 9 D3
3 ON 10 OFF +12V ISO
33-34 12 D4
1 13 D5
4 +12V ISO D6 11 D5
35-36 4 11 3 +12V ISO 14 D6
5 12 D7
31-32 2 GND 13 D7
U1A U1B
1 GND
U2E MC14072B MC14072B
AEC STOP 15 AEC STOP
RP1 MC14050B U11 PT INPUT
10K 11 12 8255A (82C55 ) to J2-13 (page 3) A15 16 A15
10 D0 34 DB0.....DB7
D1 D0 to J16 (page 4)
9 33 D1
2 8 U2B D2 32 2
7 MC14050B D3 D2 J10
31 D3
6 5 4 D4 30 4 DB0 DB0 -12V
D5 D4 PA0 DB1 DB1 7 DB0 +12V
5 29 D5 PA1 3 10 DB1 JP5 POSITION: 17 -12V
4 D6 28 2 DB2 DB2 TP5 +5V
D6 PA2 12 DB2 18 +12V
3 U2F D7 27 1 DB3 DB3 A.- VD SELECTION JP5 PULSED FL
+5V 2 MC14050B D7 PA3 DB4 DB4 11 DB3 B.- IC4 SELECTION A B C 20 +5V
PA4 40 14
DB5 DB5 DB4 C.- PT INPUT SELECTION
13 14 PA5 39 13 DB5 24 -PULSED FL
1 1 -IOWR 36 38 DB6 DB6
10
WR PA6 DB7 DB7 16 DB6 21 PT INPUT
PA7 37 15
9 to U37 -IORD DB7 ABC OUT
5 RD 22 ABC OUT
RP15 8 (sheet 3) EN2 to JP19-A (page 3)
10K 7 IN1 A11 6 to J16-7 (page 4)
CS RAPID TERMINATION 23 RAPID TERMINATION
6 IN2 14
5 A0 PC0 6 RS
8 A1 PC1 15 8
4 U2A EN1 J5 J6A J6B J2
PC2 16 9 EN2
3 MC14050B A1 9 17
A0 PC3 5 R/W IC1 1 1 1 1 IC1
2 3 2 13 -CSE2PROM
PC4 SI IC2 6 2 2 2 IC2
18 PB0 PC5 12 IC3 2 3 3 3 IC3
19 11 C/-D +5V
PB1 PC6 SCK IC4/VD 7 4 4 4 IC4/PT INPUT
U2C 20 10
PB2 PC7 4 +5V -FLD1 3 5 5 5 -FLD1
MC14050B 21 PB3 3 VD -FLD2 8 6 6 6 -FLD2
7 6 22 PB4 2 GND -FLD3 4 7 7 7 -FLD3
23 PB5 1 GND -STRT DR 9 8 8 8 -STRT DR
SW1 24 PB6 GND 5 9 9 9 GND
1 8 25 PB7 RESET 35 10 10 10 VD
C62 C / -D
2 7
3 6 to J16-8 (page 4)
4 5 RIBBON CABLE
1 RAPID TERMINATION 33 pF R8 1
to J12A/B-23 1K DS2 +5V
U18 3 4
U1 V ON U2 +5V U11 +5V U17 +5V U25 +5V X25320
MC14072B MC14050B 8255A (82C55) 74HCT138 8279 1 5
CS SI U19B V NC 18/0119 Carlos R. 28/03/19 NAME DATE SHEET / OF
Vcc 14 Vcc 1 Vcc 26 Vcc 16 Vcc 40 A3024-2XX
2 6 74HCT14 U NC 13/257 R. Hermosilla 19/02/14 DRAWING F. Díaz 09/02/11
+5V SD SCK 2/4
T NC 13/417 R. Hermosilla 21/01/14 REVISED A. Díaz 09/02/11 V U T S W REV
C11 C12 C14 C13 C40 3 7 +5V
100nF 100nF 100nF 100nF 100nF WP HLD C49 S NC 12/278 F. Díaz 25/06/12
7 8 7 8 20 4 8 100nF
GND GND GND GND GND GND VCC
W NC19/0078 R. Hermosilla 30/07/19 SEDECAL ATP CONSOLE
REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

NOTE.- ALL DIODES 1N4148 UNLESS SPECIFIED U5A


74HCT02
SET JP13 FOR DIGITAL MODE 2 +5V J3
CR17 1 CR9
JP13 10 -LINE CONT
+5V to U29-18 (page 1) AEC STOP 3 D
R41
2K7 CR8 G Q3 PT CRL
to U21-10 (page 1) -EXPT S MTD3055 to J8-1 and J7-1 (page 4) 7 PT CRL
U4 DS1 U5B -ACT EXP - KV UP
74HCT123 +5V CR18 R40 74HCT02 15 -KV UP
R11 1 10K 5 - KV DWN
CR16 330K 1A R17 R12 C52 8 -KV DWN
2 1B 4
4 3 10K PREP 6 150K 1uF +5V HT-C DAT 4
1CLR RP20 470 3 HT-C DAT
14 1CX
15 R10 +5V LINE SYNC
1RX/CX 1K U5C 1 2 9 LINE SYNC
R20 74HCT02 3 4 -BUCK EXP
9 2A 1Q 13 DATA BUS 5 6 12 -BUCK EXP
C21 10 4 10K 8
1.0 2B 1Q 7 8 HT-C CLK
11 2CLR 10 6
U6 HT-C CLK
6 2CX 2Q 12 9
7 5 8255A RP3 22 -HT INTLK
2RX/CX 2Q D0 U7 13 -HT INTLK
34 D0 PA7 37 7 8
D1 33 38 11 ULN2803
D2 D1 PA6 5 6 1 -EXP
32 D2 PA5 39 13 8 I8 Q8 11 3 4 4
D3 -PREP
31 D3 PA4 40 12 7 I7 Q7 12 1 2
+5V U3/U12/U14/U27/U37 U35/U36 D4 30 U5D 6 13 -PREP
2501 2501 D5 D4 74HCT02 I6 Q6 to J13-9
29 D5 5 I5 Q5 14
1 A1 D6
E7 15 E7 15
D7
28 D6 PA3 1 1 I1 Q1 18 1 2 2 C-HT DAT
27 D7 PA2 2 2 I2 Q2 17 3 4 5
-IOWR C-HT CLK
3 13 13 36 3 3 16

GND
A2 E6 E6 WR PA1 I3 Q3 5 6 11 -BUCKY2 DR CMD
PA0 4 4 I4 Q4 15 7 8 14
-IORD -BUCKY1 DR CMD
5 A3 E5 11 E5 11 5 RD R19 RP4 22 -AUTO OFF
A13 2K74 -DS PREP (PAGE 2, R2) J2

9
7 A4 E4 9 E4 9 6 CS (PAGE 1, K4) RP11 22
A1 U13 25 -ACT EXP
8 A1 1 2 23
ULN2803 ALOE
A0 3 4 21 -AUTO OFF
9 A0 1 I1 Q1 18 5 6 10 EXP OK
PC7 10 2 I2 Q2 17 7 8 8
HT-C DAT RP2 U3 +5V -ROOM LIGHT
PC6 11 3 I3 Q3 16
470 2501 1 12 4 15 -SPARE IN1
3 -BUCK EXP PC5 I4 Q4 22 -SPARE IN1 3
1 2 2 K1 C1 16 8 25 PB7 PC4 13 5 I5 Q5 14 7 8
3 4 4 K2 C2 14 6 24 PB6 PC3 17 8 I8 Q8 11 5 6 9 -READY
FL-C DAT 6 12 23 16 7 12

GND
5 6 K3 C3 4 PB5 PC2 I7 Q7 3 4 24 -ALOE
7 8 8 K4 C4 10 2 22 PB4 PC0 14 6 I6 Q6 13 1 2
-THERMOSTAT 1 21 -DOOR
9 PB3 -DRD RP13 22 19 -DOOR
7 20 PB2 PC1 15
-DOOR RP10 U12 to J16-6 (page 4) -COL

9
5 19 PB1 6 -COL
470 2501 18 35 JP19
-COL 3 PB0 RESET MASTER RESET ABC OUT ABC OUT -LEFT -COMP
7 8 8 K4 C4 10 7 -COMP
6 12 RP9 C10 to J12-22 (page 2) A B
-COMP 5 6 K3 C3 10K 1nF
3 4 4 K2 C2 14 11 ABC OUT/ -LEFT
2 16 -DWR
1 2 K1 C1 DATA BUS
-THERMOSTAT 2 to J16-5 (page 4) -THERMOSTAT 2
RP6 22 20 -THERMOSTAT 2
U15 FS U8
8255A to J16-19 (page 4) ULN2803 7 8 2 -SFD SEL
D0 5 6 -THERMOSTAT 1 1 -GEN OK
JUMPER NORMAL MOBILE D1
34 D0 PA7 37 4 I4 Q4 15 3 4
33 D1 PA6 38 3 I3 Q3 16 1 2 3
D2 C65 -THERMOSTAT 1
JP14 set removed 32 D2 PA5 39 2 I2 Q2 17
D3 31 40 1 18 100nF
D4 D3 PA4 I1 Q1
JP15 removed set 30 D4 PA3 1 5 I5 Q5 14 3 4 12 -DIRECT SEL
D5 29 2 6 13
D6 D5 PA2 I6 Q6 5 6 -SF PREP CR12
JP16 removed set 28 3 7 12

GND
D7 D6 PA1 I7 Q7 1 2 4 -SF PREP
27 D7 PA0 4 8 I8 Q8 11 7 8
JP17 removed set -IDWR 36 CR13
WR +5V RP7 22 -FL EXP
-IDRD 1 SW3 17 -FL EXP
JP18 removed set

9
5 RD
14 1 16 -PT SEL
A14 PC0 2 16 -PT SEL/ -SFC
2 6 CS PC1 15 2 15 2
R41 3 PT INPUT
PC2 16 3 14 13
+5V 475 A1 4 PT INPUT
8 A1 PC3 17 4 13
5 EXT SYNC
PC4 13 5 12 14 EXT SYNC
A0 6
9 A0 PC5 12 7 6 11
100nF 11 7 10 JP23 -SPARE IN2
1K RP12 U14 +5V PC6 8 EXT SYNC 15 -SPARE IN2
PC7 10 8 9
470 2501 1 9 to J2-14
-HT INTLK RP5 J20 5 GND
1 2 2 K1 C1 16 6 25 PB7 18
4 14 24 10K GND
-PT SEL 3 4 K2 C2 7 PB6 2 EXP STOP CR20
5 6 6 K3 C3 12 8 23 PB5 1 EXT REF
8 10 22 1N4148
-FL EXP 7 8 K4 C4 9 PB4 -PREP J4 (to J4 - FLUORO CPU)
3 18 PB0
19
-SF PREP SW2
4
20
PB1 J13 2 -CAM FL EXP
5 PB2 1 C-FL DAT
1 8 10 21 PB3 RESET 35 15 -FS 7 C-FL CLK
2 7 U36 2501 RP18
2
470 10 -TOMO EXP FL-C DAT
3 6 A1 1 11 TOMO ON 3 FL-C DAT
4 5 RP14 16 2
10K C1 K1 2 1 9 -TOMO PREP FL-C CLK
A2 3 1 4
TIME 1 FL-C CLK
14 C2 K2 4 4 3 2 TIME 2
RP19 U37 5 CAM SYNC
470 2501 to RP15 A3 3 TIME 3 9 CAM SYNC
12 C3 K3 6 6 5 4 TIME 4
to J2-15 2 16 (sheet 2) 7 +12V ISO
-SPARE IN2 1 2 K1 C1 IN2 A4 5 -PS (DSI) SEL
3 4 4 K2 C2 14 10 C4 K4 8 8 7 6 5
+5V -CINE (DSA) SEL +12V ISO
5 6 6 K3 C3 12 7 -HCF SEL 6 GND
-SPARE IN1 8 10 IN1
to J2-22 7 8 K4 C4 8 SPARE IN 8 GND
12 EXT REF
14 EXP STOP
1 1
U4 +5V U5 +5V U6 +5V U15 +5V U35 2501 RP17 CR19 13 GND
JP14 JP15 JP16 JP17 JP18
74HCT123 74HCT02 8255A 8255A 1 470 1N4148
A1
Vcc 16 Vcc 14 Vcc 26 Vcc 26 16 C1 K1 2 2 1
A2 3 V NC 18/0119 Carlos R. 28/03/19 NAME DATE SHEET / OF
14 C2 K2 4 4 3 A3024-2XX
C19 C23 C24 C15 5 U NC 13/257 R. Hermosilla 19/02/14 DRAWING F. Díaz 09/02/11
100nF 100nF 100nF 100nF 12
A3
6
3/4
C3 K3 6 5 T NC 13/417 R. Hermosilla 21/01/14 REVISED A. Díaz 09/02/11 V U T S W REV
GND 8 GND 7 GND 7 GND 7 A4 7
10 8 S NC 12/278 F. Díaz 25/06/12
C4 K4 8 7

W NC19/0078 R. Hermosilla 30/07/19 SEDECAL ATP CONSOLE


REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

SERIAL COMMUNICATION REQUIRES THE


COMPONENTS SHOWN IN THIS BOX -ACT EXP
+5V (PAGE 3, J2-25)

4 R15 4
4K7 RS485
JP8 RS485
JP22 JP21 +5V C34 JP7
485 422 485 422 100nF A B
A B A B A B

R16
U10 75

14
MAX489
+5V J8

Vcc
1 NC NC 13 4 +5V
2 R A 12 1 -ACT EXP
3 REB B 11 8 RXD-
4 DE Z 10 7 TXD-
5 D Y 9
+5V C35 6 8 RS232/422

GND
100nF GND NC RS232/422 JP9 9
JP10 232 422
DATA BUS 232 422 6
A B
U16

7
A B 2 RXD+ (RXD)
2681

40
+5V 3 TXD+ (TXD)
D0 5 GND
25 15

Vcc
RS232/422 D1 DO DP7 C31
16 D1 DP6 26
JP11 D2 24 14 10uF
232 422 D3 D2 DP5 16V U9 +5V
17 D3 DP4 27
A B D4 23 13 C33 C32 C30 MAX232
D5 18
D4 DP3
28 10uF 10uF 10uF 2 16 +5V J7
3 D6 D5 DP2 16V 16V 16V V+ Vcc 3
22 D6 DP1 12 1 C1+ 4 DTR
D7 19 29 3 C29 SET JP12 FOR
D7 DP0 C1- 10uF FLUOROSCAN 1 -ACT EXP
4 C2+ 8
-IOWR 8 5 16V
WR C2- +12V UNREG JP12 7
6 V- GND 15
-IORD 9 RD 9
TXDA 30
A10 35 11 11 14
CS TXDB 3 TXD
A0 1 12 13
RS1 2 RXD
A1 6
3 SW4 10 7 GND
RS2 CR24 CR23 5
IP6 37 1 8
A2 5 38 2 7 9 8 1.5KE15CA 1.5KE15CA
RS3 IP5
IP4 39 3 6
A3 6 2 4 5
RS4 IP3 CR22 CR21
IP2 36
31 4 1.5KE15CA 1.5KE15CA
RXDA IP1
10 RXDB IP0 7
32 +5V
X1
IRQ 21 J16
C37 XT2 1 RP8
GND

33pF 3.6864MHz 33 34 10K


X2 D0 2 1 FGND
3 +5V 2 GND
20

4
5
C63 6 3 VDD
33pF
7
R38 4 VEE
2 8 2
10K -DWR
UART RQ 9 Q7 R39 (PAGE 3, U15-3) 5 -WR
2 1
(PAGE 1, U29-25) 10 2N4401 1K -DRD
U19A R37 (PAGE 3, U6-15) 6 -RD
74HCT14 1K EN2
(PAGE 2, J10-9) 7 -CE
C/ -D
(PAGE 2, U11-11) 8 C/ -D
-12V R30 FS
+5V 10K (PAGE 3, U15-2) 19 FS
9
10 -RESET
DB0
+5V C58 DB1 11 DB0
+12V ISO Q5 CR14 1nF DB2 12 DB1
2N4403 1N4148 DB3 13 DB2
J2 14 DB3
R24 DB4
INTEGRATED 1K DB5 15 DB4
R22 R23 DB6 16 DB5
KEYBOARD J4 FLUORO CPU 17 DB6
1K 100K DB7
MASTER RESET 18 DB7
J3 BOARD (PAGE 3, U6-35) TP12
and DB0......DB7 GND
C53 CR10 R27 (PAGE 1, U30-21) (PAGE 2, J10) 20 RV
100uF BZX55C C16 6K81
6V2 1uF

Q6
1 J9 2N4401 1
CR11 R25
ATP J4
1N4148 825 R26
681
CONSOLE DISPLAY V NC 18/0119 Carlos R. 28/03/19 NAME DATE SHEET / OF
BOARD U NC 13/257 R. Hermosilla 19/02/14 F. Díaz 09/02/11
A3024-2XX
J10 BOARD DRAWING
4/4
T NC 13/417 R. Hermosilla 21/01/14 REVISED A. Díaz 09/02/11 V U T S W REV

J11 S NC 12/278 F. Díaz 25/06/12

W NC19/0078 R. Hermosilla 30/07/19 SEDECAL ATP CONSOLE


REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

AEC/ABC SW
J1 U6
U3 +12V U2 U4 ULN2803 J2
D0 7224 A9579-01 74HCT574
D0 8 D1 D0 13 17 D0 3 D0 2 19 1 18
D1 7 D0 RESET A1 D1 Q1 I1 Q1 5 -FLD 1
D2 D1 12 D1 4 D1 3 18 2 17
D2 10 D1 A2 D2 Q2 I2 Q2 6 -FLD 2
D3 +5V U7 D2 11 D2 5 D2 4 17 3 16
D3 9 D2 A3 D3 Q3 I3 Q3 7 -FLD 3
D4 74HCT138 D3 10 D3 6 D3 5 16 4 15
D4 12 D3 A4 D4 Q4 I4 Q4 8 -STRT DR
D5 6 D4 9 D4 6 15 5 14
D5 11 G1 D4 D5 Q5 I5 Q5 9 GND
D6 D5 8 D5 7 14 6 13
D6 14 D5 D6 Q6 I6 Q6 10 VD
D7 5 D6 7 2 D6 8 13 7 12
D7 13 G2B D6 VOUT D7 Q7 I7 Q7
D7 6 D7 9 12 8 11

GND
D7 D8 Q8 I8 Q8
4 VREF 3 7 W 4

-IOWR 5 4 G2A DGND 5 11 CLK OC 1

9
A15 16 3 C Y2 13 15 WR VSS 1

A1 4 2 B Y1 14 LDAC 16
SCL 2
A0 3 1 A Y0 15 CS 14
C25 +12V +5V
SCL 1
100nF
REFERENCE (as calibration number) R21
+12V R19 1K TP6

8
115K LM393 RAPID
2 - U9
R24 1
TP4 CR4
3K32 3 +
RP2 U5 RAMP IN4148 R20 C26
J2 R23 6 -
1K DG501 4K99 1 38K3 100nF
1 2 4 8 7

4
IC 1 1 S1 D + U8A
IC 2 2 3 4 5 S2
5 6 6 1 R16 LM339 J1
IC 3 3 S3 A0 SCL 1
7 8 1M C12 R22 C27
PT INPUT/IC4 4 100nF 22 nF RAPID
16 SCL 2 49K9
C8 C9 C10 A1 23
12 S5 TERMINATION
100nF 100nF 100nF 11 2
S6 VC +12V
10 S7 S4 7 15 AEC STOP
9 15 +5V
S8 A2
3 R8 R15 3
2K2 U2C R4 4 - 100
A9579-01 3K3 2 1 PT CRL
12 S D 11 5 +
C13 +5V
B A R6 C7 U8B 100nF
R10 R7 200 K 1uF LM339
JP4 10K 10K
U2D R13 R12
JP3 A9579-01 10K 10K
R3 U2A 10 9
1K A9579-01 B A +12V S D
FL RAD
16 15 See Note
S D

TP2 TP1 R9
INTEGRATION RESET UP PT DWN PT U8C 100K
LM339 J1
8 -
R1 U2B R14 R11 14 19 -KV DWN
604K A9579-01 10K 10K 9 +
13 14 RP1 +12V
D S 1K
8 7
6 5 +5V
FLUORO ABC TP3 U8D 4 3
ABC IN LM339 2 1
C24 C6 R2 R5 10 -
100nF 1nF 1K 10K 13
2 6 -KV UP 2
11 +
INTEGRATION CR3
C11 R18 IN3595 R17
C
U1 1nF +12V 1K 1K +5V
CA3140
B 2 -
6 B A
J1 JP1 3 +
CR1 CR2 JP2
+5V A 1N4148 1N4148
20 C1 C14
1

+12V 100pF +12V C22 +12V 100nF


18 1uF
-12V +12V +5V

7
-12V U1
17

3
- CA3140 BOARD VERSION JP1 JP2 JP4
6 - 4 -
C5 7 2 +
C3 C2 C4 100nF TP5 5 + 5 3012-21 PHOTOMULTIPLIER B B A
+
10uF 10uF 10uF GND U9 U8B C15
LM393 LM339 100nF 3012-22 FOR TV CAMERA A A A

4
12
GND
2 3012-25 4 ION CHAMBERS C A B
-12V

NOTE.- JUMPER JP3 IN POSITION A FOR ION CHAMBERS


1 U2 +12V C16 U3 +12V U4 +5V U5 +12V C20 U7 +5V WITH HIGH SENSIBILITY (GAIN 2V / mR ) 1
A9579-01 100nF 7224 74HCT574 5108 100nF 74HCT138
+V 8 VCC 18 VCC 20 +V 13 VCC 16

C18 C19 C23


NAME DATE SHEET / OF
C17 100nF 100nF C21 100nF A3012-21/22/25
100nF 100nF DRAWING R. Asenjo 16/04/08
1/1
GND

GND

-V 1 GND 4 GND 10 -V 3 GND 8 D NC 13/228 M.A.González16/09/13 REVISED A. Díaz 16/04/08 D C B A REV


C NC 12/378 M.A.González 19/03/13
14

-12V -12V
2

B NC 08/137 M.A.Gonzalez 21/05/08


A NC 08/105 R. Asenjo 16/04/08 SEDECAL AEC CONTROL
REV DESCRIPTION ISSUED BY DATE
A B C D E
A B C D E

TB1 J1

COMM 1 1 COMM
-12 VDC 2 7 -12 VDC
+12 VDC 3 8 +12 VDC
+24 VDC 4 9 GND
RELAY 4 8 IC1 OUT 5 IC1 OUTPUT
4 RELAY 3 5 FL2-IC1 3 CENTER AREA 4

RELAY 2 6 FL1-IC1 2 LEFT AREA


RELAY 1 7 FL3-IC1 6 RIGHT AREA
+12 VDC +12 VDC +12 VDC +12 VDC -RESET
9 STRT-IC1 4
R6 3K01 R5 1K R12 3K01 R11 1K R7 3K01 R8 1K R1 3K01 R2 1K
GND 10 TP6 TP1 TP2 TP4
RP1 10K RP2 10K RP4 10K RP3 10K RP4 10K RP3 10K RP1 10K RP2 10K +12 VDC
IC4 5 6 6 5 IC1 5 6 6 5 IC2 1 2 2 1 IC3 1 2 2 1

R4 1K 6 R10 1K 9 R9 1K 13 R3 1K 2
7 RP1 10K 8 RP4 10K 14 RP4 10K 1 RP1 10K TP5
5 7 8 10 7 8 12 3 4 3 3 4 GND J2
C8 U1B C10 U1C C9 U1D C7 U1A
J4 100nF LF347 RP2 10K 100nF LF347 RP3 10K 100nF LF347 RP3 10K 100nF LF347 RP2 10K
7 8 7 8 3 4 3 4 COMM
JP12 JP9 JP10 JP11 1
GND 5 7 -12 VDC
B A IC4 OUT B A IC1 OUT B A IC2 OUT B A IC3 OUT
IC4 INPUT 7 8 +12 VDC
-12 VDC
IC1 INPUT 1 9 GND
IC2 OUT
IC2 INPUT 6 5 IC2 OUTPUT
FLD2-IC2
IC3 INPUT 2 3 CENTER AREA
3 3
FLD1-IC2
-FLD 2 DR 8 2 LEFT AREA
FLD3-IC2
-FLD 1 DR 3 6 RIGHT AREA
STRT-IC2
-FLD 3 DR 4 4 -RESET
-STRT DR 9
J3
K10 JP8 K6 JP6 K2 JP14 K15
6 1 6 1 6 1 6 1 1 COMM
A A A
B B B 7 -12 VDC
CR2
1N4148 +12 VDC
FLD2-IC1 FLD2-IC2 FLD2-IC3 8
13 8 13 8 13 8 13 8
C5 C6 GND
9
IC3 OUT
100nF 100uF IC3 OUTPUT
35V 5
FLD2-IC3
K11 JP4 K7 JP2 K3 JP16 K14 CENTER AREA
3
6 1 6 1 6 1 6 1 FLD1-IC3
A A A
2 LEFT AREA
B B B FLD3-IC3
CR3 RIGHT AREA
1N4148 6
FLD1-IC1 FLD1-IC2 FLD1-IC3 STRT-IC3
13 8 13 8 13 8 13 8 4 -RESET
2 2

K12 JP7 K8 JP5 K4 JP15 K13 J5


6 1 6 1 6 1 6 1
A A A
B B B 1 COMM
CR4
1N4148 -12 VDC
FLD3-IC1 FLD3-IC2 FLD3-IC3 7
13 8 13 8 13 8 13 8
TP3 +12 VDC
8
ST GND
9
IC4 OUT
K9 JP3 K5 JP1 K1 JP13 K16 IC4 OUTPUT
R13 5
6 1 6 1 6 1 6 1
A A A
3 CENTER AREA
10K
CR1 B B B
2 LEFT AREA
1N4148
STRT-IC1 STRT-IC2 STRT-IC3
13 8 13 8 13 8 13 8 6 RIGHT AREA

4 -RESET
JUMPERS JP9, JP10, JP11, JP12
+12 VDC
POS. A.- NO OFFSET ADJUSTMENT
POS. B.- OFFSET ADJUSTMENT C1 C3
1 1
4

100nF 100uF
Ion Chamber Jumpers Jumpers Jumpers 35V
U1A
+ Note.- Version 03 as CN 99/45
Type JP3, JP4, JP7, JP8 JP1, JP2, JP5, JP6 JP13, JP14, JP15, JP16
LF347
- NAME DATE SHEET / OF
IC1=IC2=IC3=IC4 B B B DRAWING F. GARCIA 07-07-99 A3263-03
1/1
11

IC1=IC2=IC3 B B A REVISED A. DIAZ 07-07-99 REV

C2 C4
IC1=IC2 B A A
100nF 100uF SEDECAL AEC ADAPTATION
IC1=IC2=IC3=IC4
/ / / A A A -12 VDC 35V
REV DESCRIPTION ISSUED BY DATE
ADAPTACION AEC
A B C D E
A B C D E

+23V
P1
5
1 9 1
4
8
3
7
2 P3
6
1 9 5
13 9
4
11 8
3
7
2
8 6
4 1
6

+24V
+24V K1 J1
2 2
1 16
+23V 1
2
P2 3
5
9
4
8 D1 1N4007
3
7
2
6 U1
1 LM317HV
+23V
VOUT VIN

R1 ADJ

3 C1 221 C2 3
10uF R2 10uF
35V 35V
3K83

4 NAME DATE SHEET / OF 4

DRAWING F. Díaz 05/04/00


A3170-01
1/1
REVISED F. Díaz 05/04/00 A REV

A NC 00/166 F. Díaz 18/07/00 SEDECAL RADIATION METER


REV DESCRIPTION ISSUED BY DATE
A B C D E
5 4 3 2 1

D D

D1 COD1 K1 COK1 C1 COC1


K2 COK2
12VDC PID101 PID102 PIK10
PIC10 PIK20
5PIJ105
1N4007 9PIJ109
ON
2200uF 4PIJ104
25V 8PIJ108
+ 3PIJ103
7PIJ107 J1 COJ1
C J2 COJ2 PID20
D2 COD2
K2 K3 COK3 2PIJ102 C
OMRON G5V-2 6PIJ106
4 PIJ204
3 PIJ203
PID201
1N4007 - PIK20 PIK30 1PIJ101
2 PIJ202 +
PIK10 OFF
1 PIJ201
K1
Hard drive connector OMRON G5V-2
-

J4 COJ4
4
PID301
PIJ404
D3 COD3
-
3 PIJ403
2 PID302
1N4007 PIK30
1
PIJ402
PIJ401
K3
OMRON G5V-2
Floppy disk connector +

B B

A A
PCB: 90241-01
NAME DATE SHEET / OF

DRAWING MC 14/01/10 A3179-09-10-11


1 /1
REVISED F.Diaz 14/01/10 B A REV

B NC 14/127 J.A.Garcia 29/05/14


SEDECAL LOW PROFILE AUTO ON/OFF
A NE 223/10 MC 14/01/10
REV DESCRIPTION ISSUED BY DATE

5 4 3 2 1

You might also like