Technical
Information SENCOM 4.0 Platform
Grounding
TI 12A06S01-01EN-P
Introduction....................................................................................................................................................3
1. General.....................................................................................................................................................4
3. Typical installations.................................................................................................................................8
4. Grounding..............................................................................................................................................10
4.2 Grounding when having different ground potentials (Ground loop currents).......................................10
5. Related documents................................................................................................................................13
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 3
INTRODUCTION
NOTE
No part of the Technical Information may be transferred or reproduced without prior written consent from Yokogawa.
Yokogawa reserves the right to make changes in the Technical Information and product at any time without notice or obligation.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 4
1. GENERAL
1.1 What is grounding?
Grounding is a common term used in the electrical industry to mean both “equipment grounding” and “system
grounding”. Equipment grounding means the connection of earth ground to non-current carrying conductive
materials such as conduit, cable trays, junction boxes, enclosures, and motor frames.
Electricity describes the flow of electrons through metal circuits and wires. This electricity is always following
the shortest possible route to ground. Grounding refers to and electrical contact directly to ground. It prevents
undesired potential differences and prevents power surges causing electrical hazards.
Electrical grounding is a backup pathway that provides an alternative route for the current to the ground if there
is a fault in the wiring system.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 5
The SENCOM smart adapter can be operated as a stand-alone analyzer, connecting directly to a DCS, PLC or
data acquisition system.
To guarantee an accurate, reliable and stable measurement, the SA11 SENCOM smart adapter has been
equipped with full galvanically isolation.
Galvanic Isolation is used when two or more electric circuits must communicate, but their grounds may be at dif-
ferent potentials. It is an effective method of breaking ground loops by preventing unwanted current from flowing
between two units sharing a ground conductor.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 6
2. WIRING AND INSTALLATION
2.1 Wired connection
A YOKOGAWA HOST system is connected to the SA11 using the WU11 type S interconnection cable with at
one end wired terminals and on the other end a M9 connector. This cable is specially designed to be installed
in heavy-duty industrial environments. It is specified for reliable transfer of digital signals in IP67 applications,
allowing it to be submerged as a whole. The double shielded cable will protect the connected devices for inter-
ferences from high voltages and currents in the surrounding environment. The WU11 type S cable is available
length up to a maximum of 100 meters (328ft).
When cable lengths other than standard specified by MS-code are required, two different WU11 types (-WP and
-CN) can be combined to any length up to 100 meters.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 7
2.3 Connection concept UM33A-S
NOTE
The intrinsically safe model of the SA11 must be connected to the intrinsically safe model IB100 to guarentee safe installation.
For installation details see user manual IB100.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 8
3. TYPICAL INSTALLATIONS
3.1 Installation SA11 with IB100 in a laboratory environment
Problems with noise causing unstable readings are often caused by unwanted electrostatical discharges in the
environment. When this is observed it is recommended to ground the process liquid. This will for instance reduce
the effect of electrostatic interferences caused by human proximity.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 9
3.3. Typical installations FLXA402 (multiple parameters)
The FLXA402 offers multiple connection possibilities, eliminating the need for multiple analyzers and providing
greater flexibility. This analyzer can connect up to five sensors, realizing interruption-free measurement even
during maintenance. Multiple sensor measurement offers additional functionalities including a variety of calculat-
ed data as well as the option to program the analyzer as a redundant system.
With its modular design, the FLXA402 analyzer offers a range of measurement choices with the respective sen-
sor module, including pH/ORP, resistivity/conductivity (SC), inductive conductivity (ISC), percent concentration,
dissolved oxygen (DO), and 4-20mA input. Dual sensor measurements offer additional functionalities, such as a
calculated data function that can be customized.
Figure 8: Typical installation for FLXA02 with multiple inputs and parameters
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 10
4. GROUNDING
4.1 Grounding with wired HOST system
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 11
• Connecting contact of #82 wire from terminals and metal housing, when SA11 is connected to another
host. Deze zin is volkomen onduidelijk
• Installing the SA11 device using the optional pipe and wall mounting hardware (/UM). Place a ground wire
between the ground terminal of the mounting bracket and the ground reference of the process
(see Figure 13).
An alternate way to prevent ground loop currents is to isolate the SA11 & mounting plate from ground 2 and
keep the wire to terminal 82 connected.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 12
4.4 Grounding with wireless HOST system
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 13
5. RELATED DOCUMENTS
NOTE
When using Yokogawa products please consult the relevant documents for those products.
They are shipped with the product or can be downloaded from the Yokogawa website.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P
Subject to change without notice.
<<Contents>> <<Index>> “SENCOM 4.0 Platform Grounding” 14
YOKOGAWA ELECTRIC CORPORATION YOKOGAWA ELECTRIC ASIA Yokogawa has an extensive sales and
World Headquarters9-32, Nakacho Pte. LTD.5 Bedok South Road distribution network. Please refer to the
2-chome, Musashino-shi Singapore 469270 European website (www.yokogawa.com/eu)
Tokyo 180-8750 Singapore to contact your nearest representative.
Japan www.yokogawa.com/sg
www.yokogawa.com
YOKOGAWA CHINA CO. LTD.
YOKOGAWA CORPORATION OF 3F Tower D Cartelo Crocodile Building
AMERICA
No.568 West Tianshan Road Changing District
2 Dart RoadNewnan GA 30265
Shanghai, China
USA
www.yokogawa.com/cn
www.yokogawa.com/us
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe B.V. TI 12A06S01-01EN-P Jun.24, 2021-00
Subject to change without notice. Printed in The Netherlands