SKZ2050说明书
SKZ2050说明书
1
Product manual Product manual
data, and upload the data via USB interface to the computer for data analysis or
I. Product overview of portable gas detector printing.
★ High temperature gas detection (optional)
portable gas detector can be used in: rapid detection of a variety of gas concentration The optional high-temperature sampling and cooling filter handle or high-
and measurement of environmental temperature and humidity, if the measurement temperature high-humidity pretreatment system will be possible to detect the
exceeded the limit value, the alarm will emit sound & light vibration as alarm. smoke gas at the temperature of 400 ℃. The detection for the gases at higher
adopted 2.31-inch high-definition color screen for real-time display, adopted the temperatures can be customized.
gas sensors of industrially well-known brands, and the main detection principles ★ It is possible to switch among three display modes
include: electrochemical, infrared, catalytic combustion, thermal conductivity, PID Simultaneously display the concentrations of four types of gases, repeatedly
photoionization, and so on. Advanced electric circuit design, mature kernel algorithm display the concentration of single-channel gases in large font, and display the real-
processing, unique external appearance design, had achieved a number of patents time curve.
about the software works, external appearance, and so on. is suitable for detecting ★ Graphical display, and use the form of curve to reflect the trend of gas
the concentrations of gases in the atmosphere environment in the pipeline or confined concentration changes in a certain period of time.
space; and the purity of the gas leakage and various high concentration single gases ● Data recovery function; in case of mis-operation, it is possible to select partial
with the background gas as nitrogen gas or oxygen gas. The types of gases for or full recovery.
detection are more than 500 types. ● It is possible to set whether to display the maximum value, minimum value,
average value or not.
● Standard USB charging port, with charging protection function, USB hot-plug
supported, and the charging status instrument can work properly.
II. Product characteristics of portable gas detector:
● 4600mA high capacity rechargeable polymer battery is adopted, so it is
possible to work for a long period of time.
★ It is possible to simultaneously detect 1 ~ 6 types of gases, and the units can be
● High-precision temperature and humidity measurements (optional
switched freely
configuration).
Units that can be selected: PPM, mg/m3, Vol%, LEL%, PPHM, ppb, mg/l.
● Real-time detection or timing detection is supported; when no detection will
★ Built-in pump-suction type measurement, built-in water vapor, dust filter, so as
be done, it is possible to shut down the pump so as to extend the startup time.
to prevent the damage of sensors and instruments due to the water vapor and dust,
● Multiple alarm modes, and multi-directional and three-dimensional indication
it can be used in the high humidity, and high dust environment, and the response is
alarm states.
fast; special gas route design, the sampling distance is more than 10 meters, and it
It includes sound & light alarm, vibration alarm, display screen visual alarm.
is possible to directly detect the gas of negative or positive pressure at -0.5 ~ 2 kg.
Alarm types include: concentration alarm, under-voltage alarm, and failure alarm.
★ Rich human-machine interface
● Multiple alarm mode settings: low alarm, high alarm, interval alarm, and
2.31-inch high-definition color screen, display the information such as real-time
weighted average alarm.
concentration, alarm, time, temperature, humidity, storage, and so on; the menu
● Mis-operation recognition function: automatic recognition and prevention
interface adopted high definition simulation graphics to display the function name
of the concentration calibration mis-operation, which is possible to avoid the
of each menu.
poor performance that may be caused by human factors.
★ Function of large-capacity data storage (capacity can be customized), multiple
● Zero-point automatic tracking; long-term use will not be affected by the zero
storage modes are supported, and RS485 communication is an optional
drift.
configuration
● Multi-level calibration for target points, can ensure the linearity and accuracy
Standard configuration is the data storage capacity for 100, 000 entries; real-
of the measurements.
time storage and fixed time storage are supported, or only the data of the alarm
● Chinese and English interfaces can be selected.
concentration will be stored; it is supported to use the local host to view, and delete
2 3
Product manual Product manual
● Broad operating temperatures: -40~+70℃, and temperature compensation is -40 ℃ ~ + 70 ℃, and the optional configuration of high-temperature
Sample gas
supported. sampling and cooling filter handle will be possible to detect the smoke
temperature
● Log recording gas at the temperature of 400 ℃ or higher.
Record the calibration logs, repair logs, failure countermeasure solutions, Temperature
Optional configuration: temperature at -40 ℃ ~ + 70 ℃, accuracy level
sensor service life due reminder, and function of reminder for the time of next and humidity
at 0.5 ℃; humidity at 0 ~ 100% RH, precision level at 3 % RH
measurement
concentration calibration.
Power supply 3.7V DC, 4600mA high capacity rechargeable polymer battery
● Meet the requirements on the design of the intrinsically safe circuits, such as
anti-static, anti-electromagnetic interference. Display mode 2.31-inch high-definition color screen
● Protection level is up to IP65, and it is possible to prevent the splash water, Built-in pump-suction type measurement, and flow rate at 500 ml / min.
dust, and shock. Calibration flow rate should be greater than 500 ml / min, three-way
Detection mode
pipe should be connected, so as to ensure that the excessive gas will be
discharged from the bypass
It is possible to set as sound & light alarm, vibration alarm, sound &
Alarm mode
light + vibration alarm, turning off the alarm.
III. Technical parameters of portable gas detector:
Communication USB (charging and communication), optional: RS485 communication
interface interface
Arbitrary combination of 1 ~ 4 types of gases such as toxic gases, Standard configuration is the data storage capacity for 100, 000 entries
Data storage
oxygen gas, carbon dioxide, inflammable and explosive gases, TVOC,
Gases to be detected Protection level IP65
and so on.
Optional configuration: temperature and humidity measurements. Explosion-proof
Intrinsically safe type
All the cases which requires the portable rapid detection of gas type
Application concentrations such as petroleum, chemical, pharmaceutical, Explosion-proof
scenarios environmental protection, combustion gas distribution, warehousing, Ex ia IIC T4 Ga
mark
smoke gas analysis, air governance, and so on.
Exterior
0 ~ 1, 10, 100, 1000, 5000, 50000, 100000ppm, 200 mg/L, 100%LEL, 178×67×40mm (L×W×H)
dimensions
Detection range 20%, 50%, 100%Vol, can be selected; and other ranges can be
customized. Weight 300g
0.01ppm or 0.001ppm (0 ~ 10 ppm);0.01ppm (0 ~ 100 ppm), Manual, qualification certificate, warranty card, USB charger (including
Standard
Resolution 0.1ppm (0 ~ 1000 ppm), 1ppm (0 ~ 10000 ppm or more), the data cable), high-grade aluminum instrument case, belt clip,
accessories
0.01 mg/l (0 ~ 200 mg/l), 0.1% LEL, 0.01%Vol, 0.001% Vol humidity dust filter
Electrochemical, catalytic combustion, infrared, thermal conductivity, Temperature and humidity measurement function, 1.2 m retractable
Detection principle PID photoionization, and so on, depending on the type of gas, range, sampling handle (1-10 meter hose, and the standard length is 1
field environment and user demand. meter), 0.4 meter stainless steel sampling handle (with dust filter, not
Optional items retractable), high temperature sampling and cooling filter handle, high
Electrochemical principle: 2 ~ 3 years; Oxygen gas: 2 years or 6 years temperature humidity pretreatment system, multiple humidity dust
Sensor service life can be selected; infrared principle: 5 ~ 10 years; catalytic combustion: 3 filter, hanging rope, CD-ROM (upper level computer communication
years; thermal conductivity: 5 years; PID photoionization: 2 ~ 3 years. software)
Maximum
≤±3% F.S (higher precision levels can be customized)
allowable error
Linearity ≤±2% Repeatability ≤±2% Uncertainty ≤±2%
Response time T90≤30 seconds Recovery time ≤30 seconds
Working Temperature: -40 ℃ ~ + 70 ℃, humidity: ≤10 ~ 95% RH, and built-in
environment filters can be used in high humidity or high dust environment.
4 5
Product manual Product manual
At the time of startup, the information about the company, the countdown time which
is needed for the startup self-check and sensor preheating will be displayed, as shown
in Fig. 1 and Fig. 2:
Fig. 1 Fig. 2
At the time of startup, the address number and software version of this unit of
equipment will be displayed, as shown in Fig. 3:
6 7
Product manual Product manual
By pressing the Confirmation key “ ⓛ key” , it will be possible to switch between Select the channels to be calibrated, press “ ⓛ” key to carry out the zero-point
different displaying modes, and press “》》” key to switch the channels. There is calibration; after the calibration is successful, it will display: the statement of“
figure illustration for each mode (There will be deviation between the figure and Calibration is successful”, as shown in Fig. 11, Fig. 12,and Fig. 13:
the actual product, so it is recommended to use the figure by referring to the actual
product), as shown in ,Fig.5、6、7、8:
Fig. 5 Fig. 6 After the calibration of CO2 in the air is successful, it will be displayed as about
400ppm on the screen, as shown in Fig. 14:
Fig. 7 Fig. 8
Fig. 14
VII. Menu operation instructions
If the calibration is not successful, it will display: the statement of “ Calibration failed”
7.1 Zero-point calibration operation instructions ; if O2 or N2 was in the air, the zero-point calibration will fail, as shown in Fig. 15:
Only when the zero drift of the sensor is too large, or it is required to accurately detect
very low concentrations, the zero-point calibration will be carried out. In the test
interface, press the“ MENU” key for 3 seconds to enter the password interface, enter
11111, move the cursor to the "Confirm" menu, and press OK to enter the menu, use
“《《” key and “》》” key to select the "zero-point calibration" menu, and presss“
ⓛ” to enter the interface of "zero-point calibration", as shown in Figs. 9 and 10:
Fig. 15
Fig. 9 Fig. 10
8 9
Product manual Product manual
7.2 Concentration calibration operation instructions Introduce the standard gas or the gas to be measured, move the cursor to the “
The deviation of the concentration which was detected out by the instrument from calibration” position, and then press “ ⓛ” key for calibration, if successful, it will
the standard gas concentration can be corrected by calibrating the concentration display the statement of “ Success”, if unsuccessful, it will display the statement of “
or changing the display coefficients.In the test interface, press the “ MENU” key Failure”, as shown in Fig. 19,Fig. 20 and Fig. 21:
for 3 seconds to enter the menu, use “《《” key and “》》” key to select the“
Concentration calibration” menu, as shown in Fig. 16:
Level I concentration value < Level II concentration value < Level III concentration
value
Fig. 17
By default, the cursor will appear at the position of Level I concentration, press “ ⓛ” Concentration calibration sequence for three levels:
key to enter the setting of the concentration values, and use “《《” key and “》》” Level I concentration calibration →Level II concentration calibration →Level III
key to modify the numerical values, after finished the modification, press “ ⓛ” key to concentration calibration
save (Note: the modified values can become valid only after the successful calibration, If the concentration of the introduced standard gas is too low, the calibration
otherwise, at the next time of entering the menu, the value will return to the previous may not succeed.
concentration value), as shown in Fig. 18:
Fig. 18
10 11
Product manual Product manual
7.3 Operation instructions to restore factory settings 7.4 Data recording operation instructions
In the test interface, press the “ MENU” key for 3 seconds to enter the menu, use In the test interface, press the“ MENU” key for 3 seconds to enter the menu, use “《《”
“《《” key and “》》” key to select the menu of “ Restore the factory setting”, and key and “》》” key to select the menu of “ data recording”, and press “ ⓛ” key to
press“ ⓛ” to enter the interface of “ Restore the factory setting”, as shown in Fig. 22 enter the interface of “ data recording”, as shown in Fig. 27:
and Fig. 23:
Fig. 31
12 13
Product manual Product manual
If there is no record for the query, it will be as shown in Fig. 32: 7.5 Alarm setting operation instructions
In the test interface, press the “ MENU” key for 3 seconds to enter the menu, use
“《《” key and “》》” key to select the menu of “ Alarm setting”, and press “ ⓛ”
key to enter the interface of“ Alarm setting”; use “《《” key and “》》” key to
select the “ Alarm numerical value”,“ Alarm parameters”,“ Alarm mode”, and press“
ⓛ” key to enter, as shown in Fig. 39 and Fig. 40:
Fig. 32
Please go to “ System setting” → “ System setting” to check whether the storage
function is enabled or not. Select the“ timing storage” that is the real time storage
function, and set the storage cycle (time interval).
As shown in the figure, follow the following steps for startup, as shown in Fig. 33,
Fig. 34, Fig. 35, Fig. 36, Fig. 37,Fig. 38 : Fig. 39 Fig. 40
Fig. 44
14 15
Product manual Product manual
Use “《《” key and “》》” key to switch the cursor, and use “ ⓛ” key to select the Alarm mode: generally no setting is needed, because it had been set by default already.
parameters. The default setting of O2 and N2 is <AL&>AH alarm; the default setting of other
Alarm method: press “ ⓛ” key to select sound & light, vibration, sound & light + gases is>AL&>AH alarm, and it is possible to modify or set to“ Turn off alarm” (Note,
vibration, no alarm, as shown in Fig. 45: if“ Turn off alarm” is selected, then, when the measured concentration value is higher
than the alarm value, no alarm will be generated), as shown in Fig. 48 and Fig. 49:
Alarm sound: 3 kinds of sound can be selected, as shown in Fig. 46: 7.6 System setting operation instructions
In the test interface, press the “ MENU” key for 3 seconds to enter the menu, use
“《《” key and “》》” key to select the menu of “ System setting”, and press “ ⓛ”
key to enter the interface of system setting, as shown in Fig. 50:
Fig. 46
Alarm hysteresis: that is the alarm return difference, it is possible to set the alarm
return difference for each channel; if it is set as 0, as long as the value reaches the set
Fig. 50
value of alarm, the alarm will be activated immediately; if it is not desired to have
frequent occurrence of alarming at the critical point of the set value of alarm, then, the Setting of range: it will display the maximum detection range of each gas, and has
alarm hysteresis should be set to a certain value, and use “ ⓛ” key and “《《” key been set up before leaving the manufacturer, if modification is needed, please call the
and “》》” key to modify, as shown in Fig. 47: manufacturer to confirm whether it is possible or not to modify; when some sensors
are measuring under overload conditions, the sensor may be damaged.
As shown in Fig. 51 and Fig. 52:
Fig. 47
Fig. 51 Fig. 52
16 17
Product manual Product manual
Setting of time: modify date and time, after pressed “ ⓛ” key, use “《《” key and Storage method: it is possible to set two modes which are the timing storage, and
“》》” key to select and modify, and then, press “ ⓛ” key to save, as shown in Fig. alarm trigger;“ timing storage” is to store the data for one time during each period of
53 and Fig. 54: storage cycle, as shown in Fig. 59:
Fig. 60
Storage cycle: the time interval for storing the data, it is possible to set as 1-9999
Fig. 55 Fig. 56 Fig. 57 seconds, and it is valid only for the timing storage.
Setting of storage: move the cursor to“ Setting of storage”, press “ ⓛ” key to enter As shown in Fig. 61:
the setting interface.as shown in Fig. 58:
Fig. 61
Fig. 58
18 19
Product manual Product manual
Storage function: select Turn on or Turn off the data storage function, as shown in Timing measurement: if the value was set,after startup, once reached the set time, the
Fig. 62: measurement will stop automatically; when needed to measure again, just press the“
PUMP” key as the switch of the pump to start the measurement again.as shown in Fig.
65:
Fig. 62
Setting of communication: when there is no connection with the computer or data
transmission, there will be no need for setting, as shown in Fig. 63: Fig. 65
The default measuring time is 0 seconds; if this value is not modified, this function
will not work; when it is modified to a certain value,
this function will start working; if it is required to use the timing measurement
function, in generalcases, it is necessary to modify the value to be 30-60 seconds,
press the pump switch of“ PUMP” key for one time to start the measurement; once
reached the set time, stop the measurement.
Interface of Timing Measurement: the set time and the remaining time will be
displayed at the bottom part of the display screen,as shown in Fig. 66,Fig.67:
Fig. 63
Equipment address: address number of the local host, and can be set as 1-255
Communication interface: the mode of automatic recognition communication
Baud rate: communication speed, generally select 9600.
Fig. 66 Fig. 67
Fig. 64
20 21
Product manual Product manual
Preheating time: it is the time for preheating and stabilizing which is needed by the
7.7 Unit setting operation instructions sensor, and it is not recommended for modification, as shown in Fig. 72:
In the test interface, press the“ MENU” key for 3 seconds to enter the menu, use
“《《” key and “》》” key to select the menu of second page of“ Unit setting”,
press“ ⓛ” key to enter the interface of setting, as shown in Fig. 68:
Fig. 72
Language options: default is Chinese, English can be selected, as shown in Fig. 73:
Fig. 68
Automatic conversion between various units is possible to switch between the
following units, as shown in Fig. 69 and Fig. 70:
PPM, Vol%, LEL%, ppb, PPHM, mg/m3, mg/l, g/m3
Fig. 73
Name of the gas: the molecular formula will be displayed by default, as shown in Fig.
74:
Fig. 69 Fig. 70
7.8 Other setting operation instructions
In the test interface, press the“ MENU” key for 3 seconds to enter the menu, use
“《《” key and “》》” key to select the menu of second page of“ Other setting”,
press“ ⓛ” key to enter the interface of setting, as shown in Fig. 71:
Fig. 74
Display the maximum value: select Open, then the maximum value of the
measurement of this time and the maximum value of the last measurement in the
single-channel interface will be displayed.
Display the minimum value: select Open, then the minimum value of the measurement
of this time and the minimum value of the last measurement in the single-channel
interface will be displayed.
Fig. 71
22 23
Product manual Product manual
Sensor information: it is possible to view the service life and the used years of the
sensor on this equipment, so as to avoid the extended use which will affect the
As shown in Fig. 75:
accuracy and reliability, as shown in Fig. 79 and Fig. 80:
Fig. 75
7.9 Explanation about local host information
Fig. 79 Fig. 80
In the test interface, press the“ MENU” key for 3 seconds to enter the menu,
use “《《” key and “》》” key to select the menu of second page of“ Local
information”, press“ ⓛ” key to enter the interface of view, as shown in Fig. 76: Repair records: record the repair information of this equipment, as shown in Fig.
81and Fig. 82:
Fig. 81 Fig. 82
Fig. 76
Guidance about the failures: it will introduce the simple troubleshooting methods for
Equipment information: it is possible to view the relevant information about this
the failures; at present, it could not be viewed temporarily, as shown in Fig. 83:
equipment, as shown in Fig. 77 and Fig. 78:
Fig. 83
Fig. 77 Fig. 78
24 25
Product manual Product manual
VIII. Common failures and countermeasure solutions temperature and humidity may cause the great fluctuations of the numerical values
of the measurement by the instrument in a short period of time; after the air the
Failure 1: it is not possible to detect when the concentration is low temperature and humidity became constant, the numerical values will become
Countermeasure solutions: relatively stable too.
1. Check whether the pump is working or not; when the pump is working properly, 2. Verify whether the zero-point calibration or target point calibration was carried out
there is slight vibration, and if the finger is used to block the gas inlet for 2 seconds, for the instrument or not; if the zero-point calibration was carried out in the case
it is possible to feel the obvious suction force. And then, check whether the gas inlet when the gas to be measured existed, it might be impossible to detect out the gases
of the filter is clogged or not, or whether the gas leakage and no suction force were at low concentrations; if the target point calibration was carried out in the case when
caused by the poor sealing of the connection joint or not. the gas to be measured existed, but the concentration value after calibrated did
2. Introduce the nitrogen gas for the zero-point calibration or carry out the zero- not match the actual concentration value, it might cause that the fluctuation of the
point calibration in the clean air, and after the calibration is finished, carry out the numerical values of the instrument is very great, or the numerical values which are
detection immediately. measured is relatively small; for these 2 cases, it is possible to be solved through the
3. If it was still impossible to detect out the gas to be measured after the zero-point operation of Restoring the factory setting.
calibration, then it is necessary to conduct the operation of Restoring the factory 3. If the problem still could not be solved, it will be necessary to verify whether the
setting. gases of high concentrations had been introduced or the gases of high concentrations
4. If it is still impossible to detect out after restored the factory setting, then it will had impacted the sensor; if the sensor had been impacted, after 24 hours of aging
be necessary to introduce the nitrogen gas again, or carry out the zero-point through connecting the power supply to the instrument, the numerical values are till
calibration in clean air again, after the calibration is finished, carry out the detection instable or the numerical values still could not decrease further, the sensor may be
immediately. damaged due to the impact, and it will be necessary to replace the sensor.
5. Check whether the connection cable of the sensor has been damaged by human Failure 3: detection result is inaccurate
factors or there was poor contact. Countermeasure solutions:
6. If it is still not possible to detect out after the above mentioned four steps are done, 1. Check whether the gas concentration on the spot is accurate or not; sometimes
then, it will be necessary to check whether the gas to be measured exist or not on the difference between the theoretical value and the actual value is great, the
the spot, or whether the concentration of the detected gas is very low indeed; if it best method is to verify the accuracy of the instrument by the introduction of the
is lower than the minimum detection limit value of the instrument, it will not be standard gases, or send to a third party metrology authority for checking.
possible for the detection. 2. If the use time of the sensor was too long, there might be some errors of the
measurement values, and it will be necessary to check with the manufacturer
Failure 2: there is no gas to be detected in the air, but the value fluctuates greatly whether it is still possible or not to continue using the sensor; if the sensor itself has
or bounces wildly almost been close to the service life, even at the moment after the calibration, it can
Countermeasure solutions: be used normally, soon, it will become not possible to work properly again, so, it is
1. In general, it belongs to the normal range if the short-term fluctuation range of recommended to replace the sensor.
zero point is less than 1% of the maximum range; under the situation without
the presence of the gas to be measured, it belongs to the normal range if the Failure 4: the alarm is activated even when the numerical value was 0 or the
long-term drift is less than 2% of the maximum range; if exceeded this range, alarm value in the air had not been reached
it is necessary to check whether the gas to be measured exists or not on the Countermeasure solutions:
spot, or the great fluctuations of air temperature and humidity resulted in the 1. Check whether the alarm has been modified or not.
instability of the numerical values, and in general cases, the great fluctuations in 2. Check whether the alarm method, alarm mode had been modified or not.
26 27
Product manual Product manual
3. Check whether the alarm status is the concentration alarm or the failure alarm, the X. Precautions
concentration alarm will display the words of AL or AH, and the red indicator light
may flash, the failure alarm will light yellow. ● For the charger, it is a must to use the original charger; in the case of using the
4. If the alarm is caused by the man-made modifications, it can be resolved by charger which are provided by other manufacturers, the specifications should be
restoring to the factory settings; and the failure alarm requires further checking 4.2V, 2A or 5V, 2A, should be neither greater than 5V, 2A, nor less than 4.2V,
about whether it is caused by the short circuit, open circuit, poor contact, sensor 1A.
failure, and so on or not, or, deliver it back to the manufacturer for checking. ● It is prohibited for the live replacement of the sensor without disconnecting the
power supply.
Failure 5: it is not possible to communicate with the computer ● At the time when the sensor calibration is carried out, and the nitrogen gas or the
Countermeasure solutions: standard gas is introduced, the gas discharged from the gas cylinder should not
1. Check the equipment address, the address which was set in the communication be directly connected to the instrument with a hose, and it must be connected to
software must be same as the equipment address of the instrument. a three-way flow diverter, at the same time of ensuring the ventilation amount
2. At the computer hardware - Equipment Manager - Port, check whether the serial is greater than or equal to 500 ml / min, excessive gas can be discharged away
ports corresponding to RS485 / RS232 converter are connected or not. For the from the three-way bypass, so as to avoid the case that the internal pressure of
equipment of USB interface, it is also necessary to detect whether the serial ports the sensor gas chamber is greater than the normal value.
are connected or not, and whether the serial port numbers are correct or not. ● The operations such as installation, debugging, setting up and so on must be
conducted by professionals.
IX. Equipment maintenance ● The calibration checking of the detector should be carried out regularly; the
sensors which exceeded the valid service life and had failures must be replaced
During the normal use of the detector, the valid service life of the most sensors is 24- timely.
36 months. Within the valid service life, one time of periodic sensor calibration should ● Avoid using the gases higher than the measurement range to impact the sensor.
be performed every 6 months or 1 year, so as to ensure the accuracy and effectiveness ● At the time of occurrence of failure, abnormality, and in accurate detection,
of the gas detection function. The sensors which exceeded the valid service life and please conduct the operation of restoring the factory setting, which can usually
had failures must be replaced. solve the problem; if the problem could not be solved, please contact the
9.1 Sensor replacement manufacturer.
After the sensor failure appears, please deliver the instrument back to the
manufacturer for replacement and recalibration.
9.2 Sensor calibration
Please refer to the detailed description in 7.1 and 7.2 of the Section 7 for details.
Never operate when the standard gas is not ready. At the time of mis-operation,
it is necessary to restore the factory setting through the operations of the
section 7.3.
28 29
Product manual Product manual
Maximum
XI. Table of technical performance and parameters of other Gases to be detected Range allowable error
Minimum Response
optional sensors value
reading time T90
Combustible gas ( EX ) 0-100%Vol <±3%(F.S) 0.1%Vol ≤10 seconds Ozone ( O3 ) 0-2000ppm <±3%(F.S) 0.1ppm ≤30 seconds
Methane ( CH4 ) 0-100%LEL <±3%(F.S) 0.1%LEL ≤10 seconds Ozone ( O3 ) 0-30000ppm <±3%(F.S) 1ppm ≤30 seconds
Methane ( CH4 ) 0-100%Vol <±3%(F.S) 0.1%Vol ≤10 seconds Ozone ( O3 ) 0-20mg/L <±3%(F.S) 0.01mg/L ≤30 seconds
Oxygen gas ( O2 ) 0-30%Vol <±3%(F.S) 0.01%Vol ≤10 seconds Ozone water ( O3 ) 0-20mg/L <±3%(F.S) 0.01mg/L ≤30 seconds
Oxygen gas ( O2 ) 0-100%Vol <±3%(F.S) 0.01%Vol ≤10 seconds Hydrogen sulfide ( H2S ) 0-10ppm <±3%(F.S) 0.001ppm ≤30 seconds
Oxygen gas ( O2 ) 0-5000ppm <±3%(F.S) 1ppm ≤30 seconds Hydrogen sulfide ( H2S ) 0-50ppm <±3%(F.S) 0.01ppm ≤30 seconds
Nitrogen gas ( N2 ) 0-100%Vol <±3%(F.S) 0.01%Vol ≤10 seconds Hydrogen sulfide ( H2S ) 0-100ppm <±3%(F.S) 0.01ppm ≤30 seconds
Carbon monoxide ( CO ) 0-100ppm <±3%(F.S) 0.1ppm ≤25 seconds Hydrogen sulfide ( H2S ) 0-2000ppm <±3%(F.S) 0.1ppm ≤30 seconds
Carbon monoxide ( CO ) 0-1000ppm <±3%(F.S) 0.1ppm ≤25 seconds Hydrogen sulfide ( H2S ) 0-10000ppm <±3%(F.S) 1ppm ≤45 seconds
Carbon monoxide ( CO ) 0-2000ppm <±3%(F.S) 0.1ppm ≤25 seconds Sulfur dioxide ( SO2 ) 0-10ppm <±3%(F.S) 0.001ppm ≤30 seconds
Carbon monoxide ( CO ) 0-20000ppm <±3%(F.S) 1ppm ≤25 seconds Sulfur dioxide ( SO2 ) 0-20ppm <±3%(F.S) 0.01ppm ≤30 seconds
Carbon monoxide ( CO ) 0-100000ppm <±3%(F.S) 1ppm ≤25 seconds Sulfur dioxide ( SO2 ) 0-100ppm <±3%(F.S) 0.01ppm ≤30 seconds
Carbon dioxide ( CO2 ) 0-500ppm <±3%(F.S) 1ppm ≤20 seconds Sulfur dioxide ( SO2 ) 0-500ppm <±3%(F.S) 0.1ppm ≤30 seconds
Carbon dioxide ( CO2 ) 0-2000ppm <±3%(F.S) 1ppm ≤20 seconds Sulfur dioxide ( SO2 ) 0-2000ppm <±3%(F.S) 0.1ppm ≤30 seconds
Carbon dioxide ( CO2 ) 0-5000ppm <±3%(F.S) 1ppm ≤20 seconds Sulfur dioxide ( SO2 ) 0-10000ppm <±3%(F.S) 1ppm ≤30 seconds
Carbon dioxide ( CO2 ) 0-50000ppm <±3%(F.S) 1ppm ≤30 seconds Nitric oxide ( NO ) 0-10ppm <±3%(F.S) 0.001ppm ≤30 seconds
Carbon dioxide ( CO2 ) 0-20%Vol <±3%(F.S) 0.01%Vol ≤30 seconds Nitric oxide ( NO ) 0-100ppm <±3%(F.S) 0.01ppm ≤30 seconds
Carbon dioxide ( CO2 ) 0-100%Vol <±3%(F.S) 0.01%Vol ≤30 seconds Nitric oxide ( NO ) 0-2000ppm <±3%(F.S) 0.1ppm ≤30 seconds
Formaldehyde ( CH2O ) 0-10ppm <±3%(F.S) 0.001ppm ≤30 seconds Nitric oxide ( NO ) 0-5000ppm <±3%(F.S) 1ppm ≤30 seconds
Formaldehyde ( CH2O ) 0-10ppm <±3%(F.S) 0.01ppm ≤30 seconds Nitrogen dioxide ( NO2 ) 0-10ppm <±3%(F.S) 0.001ppm ≤25 seconds
Formaldehyde ( CH2O ) 0-100ppm <±3%(F.S) 0.01ppm ≤30 seconds Nitrogen dioxide ( NO2 ) 0-100ppm <±3%(F.S) 0.01ppm ≤25 seconds
Formaldehyde ( CH2O ) 0-5000ppm <±3%(F.S) 1ppm ≤50 seconds Nitrogen dioxide ( NO2 ) 0-1000ppm <±3%(F.S) 0.1ppm ≤30 seconds
30 31
Product manual Product manual
Maximum Maximum
Minimum Response Minimum Response
Gases to be detected Range allowable error Gases to be detected Range allowable error
reading time T90 reading time T90
value value
Nitrogen dioxide ( NO2 ) 0-5000ppm <±3%(F.S) 1ppm ≤30 seconds Phosphorus hydride ( PH3 ) 0-25 ppm <±3%(F.S) 0.01ppm ≤30 seconds
Nitrogen oxide ( NOX ) 0-10ppm <±3%(F.S) 0.001ppm ≤30 seconds Phosphorus hydride ( PH3 ) 0-2000 ppm <±3%(F.S) 1ppm ≤30 seconds
Nitrogen oxide ( NOX ) 0-100ppm <±3%(F.S) 0.01ppm ≤30 seconds Chlorine dioxide ( CL O2 ) 0-1ppm <±3%(F.S) 0.001ppm ≤30 seconds
Nitrogen oxide ( NOX ) 0-2000ppm <±3%(F.S) 0.1ppm ≤30 seconds Chlorine dioxide ( CL O2 ) 0-10ppm <±3%(F.S) 0.01ppm ≤30 seconds
Nitrogen oxide ( NOX ) 0-5000ppm <±3%(F.S) 1ppm ≤30 seconds Chlorine dioxide ( CL O2 ) 0-200ppm <±3%(F.S) 0.01ppm ≤30 seconds
Chlorine gas ( CL2 ) 0-10ppm <±3%(F.S) 0.001ppm ≤30 seconds Ethylene oxide ( ETO ) 0-100ppm <±3%(F.S) 0.01ppm ≤30 seconds
Chlorine gas ( CL2 ) 0-20ppm <±3%(F.S) 0.01ppm ≤30 seconds Ethylene oxide ( ETO ) 0-1000ppm <±3%(F.S) 0.1ppm ≤30 seconds
Chlorine gas ( CL2 ) 0-200ppm <±3%(F.S) 0.1ppm ≤30 seconds Ethylene oxide ( ETO ) 0-100%LEL <±3%(F.S) 1%LEL ≤30 seconds
Chlorine gas ( CL2 ) 0-2000ppm <±3%(F.S) 0.1ppm ≤30 seconds Phosgene ( COCL2 ) 0-1ppm <±3%(F.S) 0.001ppm ≤20 seconds
Ammonia gas ( NH3 ) 0-50ppm <±3%(F.S) 0.01ppm ≤30 seconds Phosgene ( COCL2 ) 0-50ppm <±3%(F.S) 0.01ppm ≤20 seconds
Ammonia gas ( NH3 ) 0-100ppm <±3%(F.S) 0.01ppm ≤30 seconds Silane ( SiH4 ) 0-1ppm <±3%(F.S) 0.001ppm ≤30 seconds
Ammonia gas ( NH3 ) 0-1000ppm <±3%(F.S) 0.1ppm ≤30 seconds Silane ( SiH4 ) 0-50ppm <±3%(F.S) 0.01ppm ≤30 seconds
Ammonia gas ( NH3 ) 0-5000ppm <±3%(F.S) 1ppm ≤30 seconds Fluorine gas ( F2 ) 0-1ppm <±3%(F.S) 0.001ppm ≤30 seconds
Ammonia gas ( NH3 ) 0-100%LEL <±3%(F.S) 0.1%LEL ≤10 seconds Fluorine gas ( F2 ) 0-10ppm <±3%(F.S) 0.01ppm ≤30 seconds
Hydrogen gas ( H2 ) 0-100%LEL <±3%(F.S) 0.1%LEL ≤10 seconds Fluorine gas ( F2 ) 0-50ppm <±3%(F.S) 0.01ppm ≤30 seconds
Hydrogen gas ( H2 ) 0-1000ppm <±3%(F.S) 0.1ppm ≤30 seconds Hydrogen fluoride ( HF ) 0-10ppm <±3%(F.S) 0.01ppm ≤30 seconds
Hydrogen gas ( H2 ) 0-20000ppm <±3%(F.S) 1ppm ≤30 seconds Hydrogen fluoride ( HF ) 0-50ppm <±3%(F.S) 0.01ppm ≤30 seconds
Hydrogen gas ( H2 ) 0-40000ppm <±3%(F.S) 1ppm ≤30 seconds Hydrogen bromide ( HBr ) 0-50ppm <±3%(F.S) 0.01ppm ≤30 seconds
Hydrogen gas ( H2 ) 0-100%Vol <±3%(F.S) 0.01%Vol ≤20 seconds Diborane ( B2H6 ) 0-10ppm <±3%(F.S) 0.001ppm ≤30 seconds
34 35