This document summarizes the specifications of the MIC-2510 insulation resistance meter. It can measure insulation resistance up to 2TΩ with selectable test voltages from 50-2500V. It also measures capacity, temperature, DC/AC voltage, and continuity. Measurement results and conditions are stored in 990 memory cells that can wirelessly transmit data to a PC. The meter meets the IEC 61557 standard and includes standard test leads and accessories.

Uploaded by

Mohamed Rizk
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
51 views2 pages

Datasheet Mic-2510 en v3

This document summarizes the specifications of the MIC-2510 insulation resistance meter. It can measure insulation resistance up to 2TΩ with selectable test voltages from 50-2500V. It also measures capacity, temperature, DC/AC voltage, and continuity. Measurement results and conditions are stored in 990 memory cells that can wirelessly transmit data to a PC. The meter meets the IEC 61557 standard and includes standard test leads and accessories.

Uploaded by

Mohamed Rizk
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

Index: WMGBMIC2510

INSULATION RESISTANCE METER


MIC-2510
INSULATION
RESISTANCE
MEASUREMENT
UP TO

2TΩ
IN ACCORDANCE
2
WITH IEC 61557- LL CAPTURE
°C FU
AND RECORDING
T
OF MEASUREMEN
CONDITIO NS

Capability of wireless transmission


of the data from memory to a PC
-ADDITIONALLY ORDER THE WIRELESS USB INTERFACE

! Insulation resistance measurement:


!
- selectable measurement voltage 100, 250, 500, 1000 V or any in the range of 50…2500 V at 10 V resolution,
- continuous indication of measured insulation resistance or leakage current,
- automatic discharge of measured object capacitive voltage after the completion of insulation resistance measurement,
- acoustic signaling of 5 seconds intervals to facilitate capturing time characteristics,
- metered T1, T2 and T3 test times for measuring one or two absorption coefficients from the range of 1…600 s,
- capability of automatic measurement of multi-wire cables with the additional AutoISO-2500 adapter,
- indication of actual test voltage during measurement,
- protection against measuring live objects.

! Continuity measurement of protective connections and equipotential bonding in accordance with PN-EN 61557-4
with current > 200 mA.

! Circuit continuity and resistance measurement at low voltage.


!
- circuit resistance measurement ( < 999 Ω) with current < 15 mA,
- quick acoustic signaling for a circuit of resistance lower than 10 Ω
Sonel S.A. - compensation (automatic zero) of test lead resistance.
ul. Wokulskiego 11 ! Measurement of leakage current during insulation resistance testing.
58-100 Świdnica !
! Measurement of capacity during the measurement of RISO.
tel. +48 74 85 83 860 !
! Continuous measurement of surrounding temperature with the capability of recording results in memory.
fax +48 74 85 83 809 !
! DC and AC voltage measurement in the range of 0…600 V.
!
! 990 cells of memory (11880 records) with the capability of wireless data transmission to a PC (with the USB-OR-1
[email protected] !
!
adapter) or through a USB cable.
Power supply from battery packs.
www.sonel.pl !
! The instruments meet the requirements of the PN-EN 61557 standard.
nr 214551 QM/UM
ISO 9001
ISO 14001
nr 214551 OH PL
PN-N-18001

MIC-2510
Insulation resistance measurement (two-lead)
Measurement range acc. to IEC 61557-2 for RISOmin=UISOnom/IISOmax...2TΩ (IISOmax=1mA) Capacity measurement

Range Resolution Accuracy Display range Resolution Accuracy

1...999nF 1nF
0.0...999.9kΩ 0.1kΩ ±(5% w.m. + 10 digits)
1.00...9.99µF 0.01µF
1.000...9.999MΩ 0.001MΩ
10.00...99.99MΩ 0.01MΩ · Capacity measurement result is displayed after the RISO measurement

100.0..999.9MΩ 0.1MΩ
±(3% w.m. + 20 digits) Temperature measurement
1.000...9.999GΩ 0.001GΩ
10.00...99.99GΩ 0.01GΩ Range Resolution Accuracy

100.0...999.9GΩ 0.1GΩ -40.0...99.9°C 0.1°C ±(3% w.m. + 8 digits)


1.000...2.000TΩ 0.001TΩ -40.0...221.8°F 0.1°F ±(3% w.m. + 16 digits)
Values of measured resistance depending on measurement voltage ! measurement with the use of an external probe

Voltage UISO Measurement range


DC and AC voltage measurement
50V 50GΩ Range Resolution Accuracy
100V 100GΩ
0...600V 1V ±(3% w.m. + 2 digits)
250V 250GΩ
! Frequency range : 45...65Hz
500V 500GΩ
1000V 1.00TΩ
2500V 2.00TΩ

Measurement of insulation resistance with the AutoISO-2500 adapter

Range Resolution Accuracy

0.0...999.9kΩ 0.1kΩ
1.000...9.999MΩ 0.001MΩ
10.00...99.99MΩ 0.01MΩ ±(3% w.m. + 20 digits)
of the meter + max 1%
100.0..999.9MΩ 0.1MΩ of AutoISO-2500
uncertainty
1.000...9.999GΩ 0.001GΩ
10.00...99.99GΩ 0.01GΩ
±(3% w.m. + 20 digits)
100.0...400.0GΩ 0.1GΩ of the meter + max 5%
of AutoISO-2500 uncertainty
Values of measured resistance depending on measurement voltage

Voltage UISO Measurement range Standard accessories of the meter:


100V 100GΩ - test lead banana plug; 1,8m; 5kV; red WAPRZ1X8REBB
- test lead banana plug; 1,8m; 5kV; blue WAPRZ1X8BUBB
250V 250GΩ - test lead banana plug; 1,8m; 5kV; black WAPRZ1X8BLBB
- USB cable WAPRZUSB
500V, 1000V, 2500V 400GΩ
- "crocodile" clip K02 5kV; black WAKROBL20K04
- "crocodile" clip K05 5kV; red WAKRORE20K05
Continuity measurement of protective connections and equipotential bonding - "crocodile" clip K05 5kV; blue WAKROBU20K05
with 200 mA current - pin probe 5kV with banana connector; red WASONREOGB2
Measurement range acc. to PN-EN 61557-4: 0.10...999Ω - pin probe 5kV with banana connector; black WASONBLOGB2
- carrying case L4 WAFUTL4
- power supply adaptor Z7 WAZASZ7
Range Resolution Accuracy - battery pack
- meter harness
0.00...19.99Ω 0.01Ω - DVD with software, including “SONEL Reader” (reading data from memory)
±(2% w.m. + 3 digits)
20.0...199.9Ω 0.1Ω - calibration certificate
- warranty card
200...999Ω 1Ω ±(4% w.m. + 3 digits)
· Voltage on open terminals: 8...16V
· Output current at R <2Ω: ISC >200mA
· Compensation of test lead resistance Electrical safety:
· Current flowing in both directions, mean value of resistance is displayed - type of insulation double, in acc. with PN-EN 61010-1 and IEC 61557
- measurement category IV 600V (III 1000V) in acc. with PN-EN 61010-1
Resistance measurement with low current - case protection rating in acc. with PN-EN 60529 IP54
Range Resolution Accuracy Other technical specifications:
- power supply of the meter SONEL L-1 NiMH 9.6V battery pack
0.0...199.9Ω 0.1Ω ±(2% w.m. + 3 digits) - weight of the meter approx. 1,3kg
200...999Ω 1Ω ±(4% w.m. + 4 digits) - dimensions 260 x 190 x 60 mm
- display LCD segment display
· Voltage on open terminals: 8...16V - measurement results memory 990 cells, 11880 records
· Output current > 10mA - transmission of measurement results USB or wireless interface
· Acoustic signal for measured < 10Ω ± 10%
· Compensation of test lead resistance
The acronym “m.v.” stands for a “measured reference value”.

v.3

You might also like