Mammography
1
Outline
Imaging features of breast lesions
Mammography unit
QC tests
Role of other existing and developing modalities
in breast imaging
2
Primary signs of breast pathology
Two of the most common mammographic
abnormalities are:
Mass
Calcifications
3
Mass
4
What to do if there is a mass?
5
Spot Compression View
Compression paddle
- Enables the application of maximal compression to
a small area of breast so that tissue is spread out
24x30cm compression paddle Spot compression paddle
6
Spot Compression View
Compression view
– separate superimposed tissue
in a localized area
=> distinguish a mass from
composite of superimposed
normal glandular tissue
- allow better delineation of
mass margin
7
Is it a mass?
8
Is it a mass?
After spot compression
9
Is it a mass?
10
Is it a mass?
CC, ML and spot-
compression CC
views show a
lobular mass at 6
o’clock.
11
How to tell the location of the
lesion?
MLO view
Lesion at ? o’clock
12
How to tell the location of the
lesion?
Lesion at ? o’clock
13
Is the mass benign or malignant?
14
Mass
Describe in terms of
shape, density and margin
Shape – round, oval,
lobular, irregular
Density- high, equal, low,
fat-containing
15
Mass
Margin – circumscribed,
microlobulated,
obscured, ill-defined,
spiculated
16
Radiographic Characteristics for
Breast Mass
Benign Malignant
Shape Round, oval, lobulated Irregular
Margin Regular, smooth, well- Spiculated, poorly
circumscribed circumscribed
Density Homogeneous Hetergenerous
Surrounding Tissue Displaced Infiltrated, retracted
Calcifications Few, scattered Numerous, confined to
mass area, along duct
17
Mass
Benign – round, oval, regular border
Malignant – spiculated, irregular margin
18
Calcifications
19
Calcifications
Tiny deposits of
calcium
Usually found in
rapid dividing cells,
can be results of
aging and injury
Can be divided into
-Macrocalcifications
-Microcalcifications
20
Calcifications
-Macrocalcifications: Benign
-Microcalcifications: like grains
of salt – 30-50% of breast
cancer that are detected by
mammography are identified by
microcalcifications alone
21
What to do if there are calcifications?
22
Spot Magnification View
Air-gap technique
- No grid should be used to
reduce the radiation dose
as the short SOD
- Scatter radiation is
reduced by the air-gap
Small focal spot
23
Spot Magnification View
Magnification view
- Assist in the evaluation of
microcalcification
- Assist in clarifying the
borders of small masses
(generally under 1cm)
- Used in specimen
radiography
24
Calcifications
Describe in terms distribution and morphology
Distribution
- diffuse
- linear
- segmentary
- regional
- grouped
25
Calcifications
Morphology
Benign calcifications
- skin
- vascular
- round
- punctate (small<0.5mm, round)
- popcorn
- eggshell / rim
Intermediate calcifications
– amorphous (no clearly defined
shape)
Malignant calcifications
- pleomorphic
- branching
26
Malignant Calcifications
Pleomorphic
Linear,
Branching
27
Benign Calcifications
Bilateral CC and MLO views
demonstrate linear and
parallel calcifications
Vascular Calcifications
-Calcifications within the
arterial wall
- Seen in 8–9% of all
screening mammograms
- Increased frequency with
advanced age, diabetes and
renal dialysis
28
Benign Calcifications
Punctate, diffuse
calcifications, vascular
calcifications
29
Malignant Calcifications
Spot-magnification
demonstrate a segmentary
area of fine linear
branching calcifications
30
31
Mammography
Equipment
32
Mammography Unit
Dedicated machine to
detect breast cancer
Low kVp technique is
used to enhance low
subject contrast of
breast tissue
33
Mammography Unit
Target
Filter
Target angle
Focal spot size
Grid
Compression
34
Mammography Unit
Breast tissue compose of adipose, fibrous and glandular
tissue, all with low atomic number
Mammography needs low energy x-ray
17-24 keV most useful in maximizing contrast by
enhancing differential absorption of breast tissue
just penetrate the tissue without overpenetration
35
Target & Filter
Target material
- Molybdenum (Z=42) , rhodium (Z=45), tungsten (Z=74)
• Filter material
- Molybdenum, rhodium, silver
- eliminate very low-energy x-ray which only contribute
to patient dose and high-energy photon result in
decrease in radiographic contrast
36
Target & Filter Combinations
Target and filter combinations
- Molybdenum + Molybdenum filter
- Rhodium + Rhodium filter
- Tungsten + Rhodium / Silver filter
Among three target and filter combinations, which
one is better for generating x-ray at 17-24 keV ?
37
Target & Filter Combinations
X-ray emission spectra at 26
kVp of Molybdenum (Z=42) +
30μm Molybdenum filter
Characteristic x-ray
- ~ 17keV
=> effective for fatty breast
38
Target & Filter Combinations
X-ray emission spectra at 28
kVp of Rhodium (Z=45) +
50μm Rhodium filter
Characteristic x-ray
- ~ 22keV
=> good for dense breast
=> better penetration but
loss in radiographic contrast
39
Target & Filter Combinations
Digital mammography system
Tungsten (Z=74) + Rodium /
Silver filter
Higher energy produced
~30keV
Wider dynamic range of
detector
Superior image resolution
compared with Mo tube
Reduction in radiation dose
40
Target Angle & Tube Tilt
Target angle
- General x-ray 5-15 degree
- Mammography 22 -24
degree
Tube tilt
- 6 degree
41
Target Angle & Tube Tilt
Target angle & tube tilt
Ensure coverage of
24x30cm with ~65 cm SID
Allow central ray run
parallel to the chest wall,
no tissue is missed
Greater tissue thickness
is positioned at cathode,
while nipple is positioned
at anode
42
Target Angle & Tube Tilt
43
Target Angle & Tube Tilt
44
Focal Spot
Small focal spot size is used to improve spatial
resolution
Large focal spot size ~ 0.3mm for routine and
compression use
Small focal spot size ~ 0.1mm for magnification
use for detection of microcalcifications
45
Mammography grid
Linear High Transmission
Cellular (HTC)
Structure Linear Crosshair pattern,
copper
Interspace material Carbon fiber Air
Transmission 50% 75-80%
Grid ratio 5:1 8:1
Scatter control One direction Two directions
46
High transmission cellular
(HTC) grid
- Similar radiation dose
compared to linear
grid
- => Better scatter
radiation control
- => Higher contrast
47
Compression Device
Reduce tissue thickness
=> Reduce scatter radiation
=> Increase radiographic contrast
Improve visualization of breast
tissue with less superimposition
=> improve uniformity
Reduce movement unsharpness as
the breast is immobilized
Reduce magnification
=> Reduce geometric unsharpness
=> increase resolution
48
Digital Mammography
• CR
- Photostimulable phosphor imaging plate
• Full field digital Mammography (FFDM)
- Direct image capture detector (Amorphous Selenium)
- Indirect image capture detector (Cesium Iodide)
X-ray generate mechanism, positioning technique and patient
management skills are same!
49
Image Formation
50
Computer Aided Diagnosis (CAD)
• Designed to increase productivity and accuracy of the
radiologist interpretation
• Work by first mapping normal healthy breast tissue
• Compare the images with normal reference maps
=> highlight suspicious areas
=> increase detection rate of breast cancer by ~10 %
51
Computer Aided Diagnosis (CAD)
52
Computer Aided Diagnosis (CAD)
53
Digital Mammography
• Limitations
- High resolution images
=> file size ~ 9M/ image
=> storage capacity
- High resolution workstation (5M) is required
- Computer aided diagnosis (CAD)
=> False positive cases
=> Unnecessary further examination and biopsy done
54
QA / QC for Mammography
55
Quality Assurance for Mammography
• the overall management program that consistently
provide high image quality and enhance patient
care
• Includes collection and evaluation of data
=> such as repeat radiographs, image quality,
patient types, technical effectiveness, efficiency
and in-service education
56
Quality Control for Mammography
• Component of a quality assurance program
responsible for assessing image quality and
equipment integrity
• Actual test to ensure the standards of the quality
assurance program are met
57
Quality Control for Mammography
• Personnel includes
- Radiologists
=> Responsible for medical auditing, detect any
discrepancies that may occur
- Medical physicists and service engineer
=> perform operation of equipment monitoring
- Radiographers
=> carry out routine testing procedures
58
QC test conducted by radiologists
Weekly
- Film rating
59
QC test conducted by Medical
physicist or service engineer
Annual
- Mammography unit assembly evaluation
- Collimation assessment
- kVp accuracy and reproducibility
- Beam quality (HVL) assessment
- Evaluation of system resolution
- AEC performance
- Breast entrance exposure and radiation output
60
QC test of digital system conducted
by radiographer
Weekly Biweekly
- Detector Flat Field Calibration • Compression Thickness
- Detector Evaluation Indicator
- Repeat Analysis
- Phantom Imaging
- Visual checklist
Semi-annually
• Compression force
- DICOM Printer Evaluation
61
Detector Flat Field Calibration
- To ensure system is calibrated properly and field uniformity
- Use 4 cm gain calibration phantom
- Follow screen instructions
Before the calibration, clean the phantom and detector surface
- Artefacts would affect flat field calibration
62
Detector Evaluation
- Detector Evaluation
- Artefact Evaluation
- Signal-to-Noise Ratio Calculation (SNR > 40)
- Contrast-to-Noise Ratio Calculation (within 15% of
the baseline CNR)
63
Repeat Analysis
- To determine the number and
cause of repeated/rejected
mammograms.
- => evaluate mammography
tube error and human error
(positioning technique and
patient movement)
- Repeat- reject analysis ( < 5%)
=> ? Need for training
=> ? Call for service
64
Phantom Imaging
- To assure the image quality due to
the imaging system are maintained
at optimum levels.
- Phantom image can be evaluated on
diagnostic review workstation (5M)
and printed film
- Minimum: 5 fibres, 4 specs, 4
masses
- Discrepancy in number gradually
Indicate deterioration of the
imaging system
Identify from tube or review
Phantom image
workstation or DICOM printer
65
DICOM Printer Evaluation
- To assure consistency of
DICOM printer performance.
- Measure the optical density
of the 10%, 40% and 90%
patches on the SMPTE film
=> value of 10-40%, 40% and
90% must fall +/- 0.15 of the
control limits
66
Compression Thickness Indicator
- To assure the indicated compression thickness is within
tolerance.
- Using the Auto-Filter mode, filter and kVp are
determined by compression thickness.
- Compression indicator should be accurate, so that
appropriate exposure factor and setting are used.
67
Compression Thickness Indicator
- ACR phantom thickness (~4.2
cm)
- Use 7.5 spot compression
paddle
- Apply full compression ~ 30
lbs
- Displayed thickness +/-
0.5cm of the measured
thickness
68
Compression Force
Compression force range from 25-40 lbs
Ensure provide adequate compression
but not too much and that compression
is controlled.
=> adequate separation of breast tissue
but not causing injury to patient
• Apply full compression and use bathroom
scale to measure compression force
69
Breast Tomosynthesis
Tradition mammography is a single,
two dimension (2D) image of the
breast. Breast tissues overlapping
may mimic or hide cancer.
Some studies showed that
mammography diagnosed a large
number of women with breast disease
need not to be treated (benign
lesions), resulting in unnecessary
biopsy and operation.
70
Breast Tomosynthesis
Respond to those high false positive
rate and high negative rate due to
dense breast, 3D mammography have
been invented, called tomosynthesis.
71
Breast Tomosynthesis
A new tool to reduce tissue
overlapping
Multiple low dose images are
acquired from various angles
as the x-ray tube moves in
an arc
Enable the acquisition of a
three-dimensional (3D)
volume of thin-section data
72
Breast Tomosynthesis
Basic technologic principles of breast tomosynthesis
©2007 by Radiological Society of North America 73
Breast Tomosynthesis
Image data acquired from
different angles are
reconstructed to provide
separate depiction of two
overlapping structures located
in different planes.
With tomosynthesis,
radiologists can page through
the breast just like the CT
slices.
74
False positive
75
False positive
76
False negative
77
False negative
Conventional mammograms 3D mammogram of the
Breast cancer hidden same breast reveal a
malignant tumour
78
Other Imaging Modalities
79
Ultrasonography
• Radiation-free
• Good for clarification of
abnormalities seen on mammography
• Physicians suggest to do
mammography and ultrasound for
patients with dense breasts to improve
cancer detection rate
80
Ultrasonography
Indications
• Characterize mass as cystic or solid
• Characterize solid mass as probably
benign, intermediate or malignant
=> help distinguishing malignant and
benign breast disease, such as cysts,
fibroadenoma
• Investigate palpable lesion
• Allow image-guided biopsy
81
Contrast MRI
• Functional study, require use of
gadolinium contrast
• Based on the fact that a carcinoma is
usually highly-vascular Without contrast
=> contrast medium will quickly
accumulate in the tumor With contrast
• Sensitivity to lesions below 1cm ~
90%, but specificity 30-70%
82
Contrast MRI
Indications
• Map tumor extent and staging newly diagnosed
breast cancer
• Monitor treatment response
• Screen women who difficult mammogram such
as dense breasts, breasts with breast implant
and surgical scar
• Investigate unknown cause of axillary
adenopathy
• Check silicone rupture
83
Contrast MRI
• Limitations
- high false positive rate because of vascularization of
noncancerous abnormalities that can mimic cancer
=> overdiagnosis including unnecessary additional biopsy for
patients
- Can’t detect microcalcifications
- Difficult to perform biopsy
- Expensive
- Time consuming
84
PET/CT
- Pre-surgical planning, staging of breast
cancer
- Evaluation of chemotherapy response in
both patients with locally advanced
breast carcinoma and those with
metastatic disease
- Monitor of recurrence
85
~ The End ~
86
References
ACR PRACTICE GUIDELINE FOR THE PERFORMANCE OF SCREENING MAMMOGRAPHY, 2004
Astley, S and F Gilbert. Computer-aided detection in mammography. Clinical Radiology 2004;59:390-399.
Bashong, S. Radiologic Science for technologists: physics, biology and protection, 9th ed.,
Mosby/Elservier:2008
Cardenosa, G. The Breast Imaging Companion, 2nd Ed. Philadelphia:2001
Lee, L., Fundamentals of mammography 2nded., Edinburgh:2003.
Peart, O. Mammography & Breast Imaging just the facts, 1st Ed. McGraw-Hill Education (Asia):2005
Papp, J. Quality management in the imaging sciences, 3rd ed., Mosby/Elservier:2006
Pisano, E, Yaffe, M & Kuzmiak, C, Digital Mammography, 1st ed., Philadelphia:2004.
Wagner, J., Procedures and documentation for mammography and quality management 1st ed., McGraw-
Hill: 2000
Hong Kong Cancer Registry, http://www3.ha.org.hk/cancereg/:2013
• Park J M, Franken EA, Garg, M, Fajardo, LL & Nikalason, LT. Breast Tomosynthesis:Present
Considerations and Future Applications. Radiographics 2007;27:S231-S240.
• Rosen E L, Eubank, WB & Mankoff, DA . FDG PET, PET/CT and Breast Cancer Imaging. Radiographics
2007;27:S215-S229.
• Kneeshaw, P et.al. Current applications and future direction of MR mammography. British Journal of
Cancer 2003;88:4-10.
87