IOLMaster Service Bulletin 532 Update

The document provides information about a new camera module available for the IOLMaster. It identifies the necessary tools, differentiates the old and new camera modules, and provides instructions for adjusting the sensitivity of the new camera module which requires a different voltage level than the old module. Technicians are advised to adjust related components after replacing the camera and potentially recalibrate measurements. The new camera module is only compatible with IOLMaster devices using a 9V digital board.

Uploaded by

jtorres
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
52 views5 pages

IOLMaster Service Bulletin 532 Update

The document provides information about a new camera module available for the IOLMaster. It identifies the necessary tools, differentiates the old and new camera modules, and provides instructions for adjusting the sensitivity of the new camera module which requires a different voltage level than the old module. Technicians are advised to adjust related components after replacing the camera and potentially recalibrate measurements. The new camera module is only compatible with IOLMaster devices using a 9V digital board.

Uploaded by

jtorres
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Service Bulletin

Surgical Products Ophtalmology Service-Bulletin Version


No. 532 No. 2.0

Aktion / Action Status Kostenerstattung / Cost Recovery Gültigkeit / Validity

Sofort Hersteller Bis :


Immediate Manufacturer Until (date):

Nächster Besuch VG oder Vertretung Nach:


First Opportunity Agent / Dealer After (date):

Bei Reparatur Kunde Bis Serien-Nr.:


In the event of repair Customer To (ser. No.):

Kundenwunsch Servicekodierung Ab Serien-Nr.:


At customer request Service Coding From (ser. No.):

Zur Information IC Bis auf Widerruf


For Information Only Until further notice

FC Ersetzt Service Bulletin Nr.: 532 V1.0


This bulletin replaces no.: 532 V1.0

Re: IOLMaster
New Camera module available

Table of contents
1. General Information 2
2. Necessary tools / continuative documentation 2
3. Identification / Differentiation 3
4. Adjusting the sensitivity 4

Version 2.0
Service-Bulletin No. 532 IOLMaster 1/5
1. General Information

Due to a focus level shift of the CCD-chip of some CCD-cameras (mostly on the
1493-442 types), this camera has been modified to a camera exchange kit with
different bases. The kit also contains shim rings to cover all know focus-levels of the
previous generations.
See Bulletin 560 for further information.
The kit is fully backward compatible for all IOLMaster generations, containing Digital-
IOL PCBs with camera operating voltage of 9V. It is not compatible if Digital-IOL PCBs
with a camera operating voltage of 8V are used. In this case the related Digital-IOL
PCB must also be exchanged.
See Bulletin 448 for further information.
Note!
The new camera-module needs a different adjustment
voltage – see chapter 4 at page 4 (Adjusting the sensitivity).

2. Necessary tools / continuative documentation

Tools:
─ standard tools
─ calibrated multimeter
─ IOLMaster calculation-spreadsheet

Continuative documentation:
─ IOLMaster service manual (SM-30-4039)
─ IOLMaster service manual (SM-30-4047)
─ Bulletin 560
─ Bulletin 510
─ Bulletin 448
─ Bulletin 305
─ Bulletin 292

Version 2.0
Service-Bulletin No. 532 IOLMaster 2/5
3. Identification / Differentiation
OLD / 8V replaced
Image with „Blooming“ - line
CCD sensitivity voltage = 3.4 V ± 0.5 V

1022-328

focused image in ALM – mode

OLD / 9V replaced
Image with „Blooming“ - line

focused image in ALM – mode

OLD / 9V replaced
Image with NO „Blooming“ - line
CCD sensitivity voltage = 2.9 V ± 0.3 V

focused image in ALM – mode

NEW / 9V (complete kit)


Image with NO „Blooming“ - line
White bases

1677-370

focused image in ALM – mode

Part-No. Description Status Remark


1022-328 CCD camera, complete (old version / 8V) Replaced Unavailable
1308-774 CCD camera, new complete Replaced No longer available
1493-442 CCD camera, 9V / Version 02, complete Replaced No longer available
1677-370 Kit, CCD camera 9V/V03 (white bases) New Available
1297-413 CCD cable 2 Valid Included with camera

Version 2.0
Service-Bulletin No. 532 IOLMaster 3/5
4. Adjusting the sensitivity

Adjust the R75 potentiometer at the Digital-IOL PCB in such a way, that a voltage of
NEW: 2.9V ± 0.3V (OLD: +3.4V ± 0.5V) is present between the GND and the potentiometer
slider contact (middle contact). Use a calibrated multimeter.

2.9 V

Caution!
The operating voltage (8V / 9V) is not adjustable and depends upon
the hardware-version of the Digital-IOL PCB.
The new type of CCD camera module is only compatible with the
9V version of the Digital-IOL PCB. You cannot use the new camera-
module with an 8V version of the Digital-IOL PCB, even if it the
image looks good. In this case the camera is not within
specification, and this may lead to incorrect readings of the image
analysing software, which could influence the KER, ACD and WTW
results.

Note!
See Bulletin 448 regarding further information of 8V / 9V
determination of Digital-IOL PCB.
See Bulletins 292 and 305 regarding further information about
determination version 1 (1031-843) and version 2 (1078-792)
Digital-IOL PCB.

Caution!
After replacement of the camera module it will be necessary to
perform the adjustment of the surrounding field illumination (R86)
and the Keratometer-illumination (R87) potentiometers on the
Digital I/O PCB.
If the IOLMaster is equipped with Software 5.02 or higher, also the
spot-size calibration (50….55) of the Keratometer-spots must be
readjusted.
See Bulletin 510 for the related procedure.
Also a re-calibration of the KER, ACD and WTW might be
necessary if the measurement results (of the test-eye) are not
within the imprinted tolerances.
See service-manual for the related procedures and further
information.

Version 2.0
Service-Bulletin No. 532 IOLMaster 4/5
78M08 Æ +8 V Å version Æ +9 V Å 78M09G
R28 – Laser adj. power (40µW)
R30 – Laser 100% power (400 µW)
Surr. Field illumination – R86
Keratometer illumination – R87

8V 9V

1031-843 Æ Version 1
1078-792 Æ Version 2

The imprinted number on the board itself is NOT


conclusive to determine between version 1 and 2.
Some boards may have the version 1 number
imprinted but are configured as version 2.
Take the number from the sticker as the reference
for either version 1 or 2 identification.

Note!
Please contact us at czm.service@meditec.zeiss.com for any questions
you may have regarding this Bulletin or the procedures described within.

MAG-SF / Wistuba
2008-10-27

Version 2.0
Service-Bulletin No. 532 IOLMaster 5/5