User Manual VC-2000
※ Special Features
- Provides 3 different screen mode : Main, Trend, Event
- Automatically save the Trend data of the day when data of the day are generated
- Save Trend Data of 30 days.
- Trend data viewer for the data collected from VP-1000 and VP-1200, VP-700.
2
User Manual VC-2000
3
User Manual VC-2000
Contents
Preface
Preface
CAUTION
Important! Before use, please carefully read this
manual, accessory directions for use, all other
precautionary information and specifications.
Please keep this manual in a place where you can reach
at any time for your reference.
This device shall be used under direction of licensed
medical person.
This device is installed on the PC, so patient data must
be secured from PC hacking and virus.
4
User Manual VC-2000
WARNING
Free software available only for VOTEM patient monitors
We do not take any responsibility.
You agree that your use of this software is not the responsibility of
VOTEM in the event of a problem.
Never use it for any other purpose.
Get your computer safe from computer viruses and hacking.
5
User Manual VC-2000
How to reach us
You may contact us at our telephone number of sales, services, and supplies.
VOTEM <www.votem.kr>
* Please give us the Serial Number, Purchasing date and when trouble occurs and service
is needed.
6
User Manual VC-2000
Some symbols are used in this manual are to help users or patients to use
this product safely and appropriately, and to prevent any risk to patients or
damage to materials. Please read and understand all warnings and
precautions.
WARNING
“WARNING” is used to indicate the presence of a hazard which can
cause severe personal injury, death or substantial property damage if
the warning is ignored.
CAUTION
“CAUTION” is used to indicate the presence of a hazard which will or
can cause minor personal injury damage if the caution is ignored.
NOTE
“Note” is used to notify the user of installation, operation, or
maintenance information which is important but not hazard related.
7
User Manual VC-2000
Screens
Main
Screen
Trend
Screen
8
User Manual VC-2000
Screens
Event
Screen
9
User Manual VC-2000
… Patient Monitors …
HUB
10
User Manual VC-2000
2. AP Wireless connect
Patient Monitors …
AP Wireless
11
User Manual VC-2000
1. Click My Network Icon and then click again the right button of mouse and click
‘Properties’ in the menu.
2. Click In Local Area Connection, click the right button of mouse and click ‘Properties’
in the menu again.
12
User Manual VC-2000
3. In Local Area Connection Properties, check ‘Internet Protocol (TCP/IP)’ and then
click ‘Properties’ in the menu.
IP Address 192.168.10.15
Subnet Mask(U) 255.255.255.0
Default Gateway(D) 192.168.10.1
13
User Manual VC-2000
14
User Manual VC-2000
(2) Please select “I accept the license agreement”. And Click [Next>] button.
15
User Manual VC-2000
16
User Manual VC-2000
(1) To set up Central monitoring system Data base Driver, go to “1bed setup →
Volume → setup.exe”.
(3) Please select “I accept the license agreement”. And Click [Next>] button.
17
User Manual VC-2000
18
User Manual VC-2000
If the program does not work properly, install the Microsoft .NET Framework.
19
User Manual VC-2000
Run as administrator
20
User Manual VC-2000
“CAUTION”
All IP Address of Central PC and connected patient monitors shall be different each
other..
Bed No. of each patient monitor shall be different each other.
◆ IP Address setting
Only 4th digit of IP Address can be changed. (192.168.10.XXX)
2 Setup for each monitor which will be connected with VC-2000 same as no.1.
21
User Manual VC-2000
◆ BED Setup
On Main screen, click Menu button and select “SETUP BED NUMBER
SET” and Setup BED Number. (Range: 1 ~ 99)
2 Setup for each monitor which will be connected with VC-2000 same as no.1
IP Addres BED
192.168.10.31 1
22
User Manual VC-2000
23
User Manual VC-2000
24
User Manual VC-2000
Successfully connect.
(2) If you double click central number, you can connect with wanted bed number.
a. Manual
In case of Manual function, you can connect as follows.
(1) If you double click central number, you can connect with wanted bed number.
Successfully connect
25
User Manual VC-2000
1. Menu Bar.
26
User Manual VC-2000
1
SD CARD port on the back of the
monitor. Then, click the “Menu
button” on Trend screen and select
TREND SAVE.
Changing to "SAVE 15SEC" or "SAVE 1MIN" will initialize the data and save it from the
Note
beginning.
27
User Manual VC-2000
You can start/stop NIBP measurement of patient montior which is connected to VC-
2000 manually
28
User Manual VC-2000
29
User Manual VC-2000
30
User Manual VC-2000
31
User Manual VC-2000
b. Event Viewer
Event Viewer will be shown on Event screen for selected bed number.
32
User Manual VC-2000
① Bed select
You can select wanted bed number. However, if there is no event data, bed will
not appear on the bed selection menu.
② Event List
Event date is displayed. The list include of bed number and the time which the
event occurred.
③ Delete File
You can delete a selected event data from the list on the left side of the screen.
④ Delete All
You can delete all event data from selected bed.
⑤ Wave select
You can select 3 Waves(LeadⅠ, LeadⅡ, LeadⅢ, Lead AVR, Lead AVL, Lead
AVF, Lead V1, Lead V2, Lead V3, Lead V4, Lead V5, Lead V6, RESP, SpO2,
IBP1, IBP2, IBP3, IBP4, EtCO2).
⑥ Wave
Displays wave between -10 to +10 seconds based on time which event
occurred. You can move to the bottom scroll to control the time.
⑦ Print
The information which is displayed in Event viewer like wave, numeric, alarm list
and event occurrence time can be printed.
⑧ Exit
Exits the Event Viewer.
⑨ Numerical Data
Display Numerical data(HR, ST, SpO2, RESP, EtCO2, IBP1, IBP2, IBP3, IBP4,
NIBP, Temp1, Temp2, Temp3, Temp4)
⑩ Alarm List
Displays the alarms which occurred.
33
User Manual VC-2000
Event Viewer can print as Event Wave form and numerical Data form.
34
User Manual VC-2000
c. Wave Reviewer
Wave Reviewer will be shown only on Wave review Screen of selected Bed
number and you can select wanted bed number.
35
User Manual VC-2000
① Bed select
You can select wanted bed number. However, if there is no Wave review data,
the bed select menu will not be displayed.
② Review List
Displays a list of saved Wave review data. The list displays the Bed number and
the time whicht the Wave review saved.
③ Delete All
Delete all Wave review data of selected Bedno.
④ Wave select
You can select 3 Waves(LeadⅠ, LeadⅡ, LeadⅢ, Lead AVR, Lead AVL, Lead
AVF, Lead V1, Lead V2, Lead V3, Lead V4, Lead V5, Lead V6, RESP, SpO2,
IBP1, IBP2, IBP3, IBP4, EtCO2).
④ Wave
Displays wave of +20 seconds based on time which wave review data saved.
You can move bottom scroll to control the time.
⑤ Print
The information which is displayed in Wave reviewer like wave, numeric, alarm
list and event occurrence time, etc can be printed.
⑥ Exit
Exits the Wave Reviewer.
⑦ Numerical data
Display Numerical data(HR, ST, SpO2, RESP, EtCO2, IBP1, IBP2, IBP3, IBP4,
NIBP, Temp1, Temp2, Temp3, Temp4)
⑧ Alarm List
Displays the alarms which occurred.
36
User Manual VC-2000
Wave Reviewer can be printed as Event Wave form and numerical Data form.
37
User Manual VC-2000
This menu will be turn on/off data saving of Event Viewer and Wave Reviewer. When
the data of the Wave reviewer is turned off, the data of Event viewer is also turned off.
38
User Manual VC-2000
This menu can delete all data of Trend viewer, Event viewer and Wave reviewer.
a. Delete all Trend data
This menu will delete Saved trend data of all bed
b,c. Delete all Event data, Delete all Wave Review data
This menu will delete all saved Event & Wave Review
39
User Manual VC-2000
40
User Manual VC-2000
ICON menu
41
User Manual VC-2000
42
User Manual VC-2000
This menu will be shown only on wave menu when you right click on your mouse.
1. Wave Select
(1) Change the selected wave form
43
User Manual VC-2000
2. Freeze
All wave stops.
44
User Manual VC-2000
1. Main Screen
2 3
1
4
11
12
5
13
14 6
15
10
9 8 7
45
User Manual VC-2000
2. Trend Screen
12 13
46
User Manual VC-2000
Alarms
Warning! : Every time the monitor is used, please check alarm limits and
settings to ensure that they are appropriate for the patient being
monitored.
When an alarm occurs, the screen( portion of default screen) will have flashing red bed
number, red alarm message and red parameter.
47
User Manual VC-2000
48
User Manual VC-2000
4. Select desired Alarm type and adjust limits with arrows placed next to each
numbers
5. Select Ok to confirm
Both of patient monitor and VC-2000 will have same alarm setting for selected device.
Dia 90 50 Resp 40 10
Mean 110 70 Temp 1 39.0 36.0
IBP2 Sys 160 90 Temp 2 39.0 36.0
Dia 90 50 Temp 3 39.0 36.0
Mean 110 70 Temp 4 39.0 36.0
EtCO2 40 10
49
User Manual VC-2000
Before you request for after Check below contents first when
VC-2000 is not operating properly.!
service.
If it’s not connected automatically when you turn off/on PC or patient monitor.
Currently, when you re-boot connected patient monitors or re-start the VC-2000, you
have to connect again VC-2000 with patient monitors. (This is not error or problem
with program.)
50
User Manual VC-2000
[Appendix I]
Output 1 2 3 4 5 6 7 8
Color Orange/White Orange Green/White Blue Blue/White Green Brown/White Brown
Outpu
1 2 3 4 5 6 7 8
t
Orange/Whit Blu Blue/Whit Orang Brown/Whit Brow
Color Green/White Green
e e e e e n
51
VOTEM CO., LTD