US Site Preparation VOLUSON E6 E8 E10

The document provides recommendations for an ultrasound room layout including dimensions of the ultrasound system, required inputs and outputs, and environmental and electrical specifications. An ultrasound room should include a door, film viewer, examination table, ultrasound system, and dedicated power outlet among other items. The system operates between 10 to 30 degrees C, requires 100-240 VAC power, and has a weight of 150 kg.

Uploaded by

jrexscci
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
38 views1 page

US Site Preparation VOLUSON E6 E8 E10

The document provides recommendations for an ultrasound room layout including dimensions of the ultrasound system, required inputs and outputs, and environmental and electrical specifications. An ultrasound room should include a door, film viewer, examination table, ultrasound system, and dedicated power outlet among other items. The system operates between 10 to 30 degrees C, requires 100-240 VAC power, and has a weight of 150 kg.

Uploaded by

jrexscci
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Physical Dimensions and Weight (mm): Recommended Ultrasound Room Layout

Please Note: The picture may vary according to order configuration.

1. door – at least 76 cm (2.5 foot /


30 inches)
2. film viewer
3. counter top, sink with hot and
cold water, supplies storage
4. linen supply
5. probes/supplies
6. examination table
7. footswitch
8. stool
9. ultrasound system
10. external peripherals
11. dedicated power outlet - circuit
breaker protected and easily
accessible
12. network interface
13. distance from wall or objects
14. cabinet for software and
manuals
150 kg
330.0 lbs.

External Inputs and Outputs :


1280x1024 1280x800 Pal-720x576
NTSC-720x480
Environmental Requirements Electrical Specifications
Operating Temperature 10 to 30 C (50 to 86 F)
Voltage 100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Operating Humidity 30 to 80% rH, non-condensing
2730 BTU/hour +300 Btu for each person. Power Consumption 800 VA, including all on board peripherals.
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -10 to 50 C (14 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity < 90% rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise Output max. 60 dB(A)

Please Note: For more information please refer to the product’s service & user manuals.