Physical Dimensions and Weight (mm): Recommended Ultrasound Room Layout
Please Note: The picture may vary according to order configuration.
1. door – at least 76 cm (2.5 foot /
30 inches)
2. film viewer
3. counter top, sink with hot and
cold water, supplies storage
4. linen supply
5. probes/supplies
6. examination table
7. footswitch
8. stool
9. ultrasound system
10. external peripherals
11. dedicated power outlet - circuit
breaker protected and easily
accessible
12. network interface
13. distance from wall or objects
14. cabinet for software and
manuals
150 kg
330.0 lbs.
External Inputs and Outputs :
1280x1024 1280x800 Pal-720x576
NTSC-720x480
Environmental Requirements Electrical Specifications
Operating Temperature 10 to 30 C (50 to 86 F)
Voltage 100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Operating Humidity 30 to 80% rH, non-condensing
2730 BTU/hour +300 Btu for each person. Power Consumption 800 VA, including all on board peripherals.
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -10 to 50 C (14 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity < 90% rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise Output max. 60 dB(A)
Please Note: For more information please refer to the product’s service & user manuals.