0% found this document useful (0 votes)
516 views26 pages

1 - Acclarix LX3 Series Datasheet

The document describes the features and specifications of the Acclarix LX3 Series diagnostic ultrasound system. It provides details on the system architecture, physical and digital channels, transducer ports, LCD monitor, language support, I/O ports, options, and ergonomic design.

Uploaded by

Bijay Gupta
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
516 views26 pages

1 - Acclarix LX3 Series Datasheet

The document describes the features and specifications of the Acclarix LX3 Series diagnostic ultrasound system. It provides details on the system architecture, physical and digital channels, transducer ports, LCD monitor, language support, I/O ports, options, and ergonomic design.

Uploaded by

Bijay Gupta
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

Acclarix LX3 Series

Diagnostic Ultrasound
Version 1.5

Advanced Technique and Features


TAI-Tissue Adaptive Imaging
eSRI-Speckle Reduction Imaging
Frequency Compounding Imaging
Spatial Compounding Imaging
Harmonic Imaging
Digital Multi-Beam forming
Trapezoid Imaging
Adaptive Doppler imaging
Spectrum Enhancement
Dedicated Steer control
Digital Zoom
Full Screen Zoom
Auto Doppler trace
Needle Visualization
Anatomic M mode
TDI mode-Tissue Doppler Imaging
3D/4D Imaging
ECG synchronization
Color M mode
Elastography mode
Auto IMT
eLearn Instruction software
Panorama
B mode one-key Optimization
PW mode one-key Optimization
Color mode one-key Optimization
Auto IMT
Auto OB
Auto NT
Auto vessel diam*

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
System Overview
System Architecture
- Electronic transducer selection

Physical Channels 64
- Ergonomic access to all transducer ports

Digital Channels ≤1382400


- Dedicated cable hook

Gray scale 256


- Removable gel warmer

Beam Forming Eight beam


- Totally 5 transducer holders with removable
transducer cups, including one dedicated
Processor ARM
transducer holder for endo-cavity transducer.
Memory 2 GB
Hard Drive 120 GB/512GB/1TB SSD
Battery
Operating System Android
System Boot-up About 30s
- Rechargeable Li-ion Battery

Boot-up from sleep 4s


- Two batteries work simultaneously, each
battery capacity of 6800 mAh, total
Shutdown 2s
13600mAh capacity.

Dimensions and Weight


- Removable

Dimension Height: 1440mm


- Approximately 1.5 hours of typical ultrasound
exam use for two fully charged batteries.
(+200mm, adjustable)
Width: 544mm
- Two batteries fully charged in about 3 hours.

Depth: 665mm
- Battery level icon displayed on the main
screen.
Net Weight About 55kg
(no transducer and accessories)
- Battery charging indicator on the control
panel.

LCD Monitor
AC Power Requirements
- 21.5’’ high resolution LCD monitor
Voltage 100 -240 V~
- Resolution: 1920 x 1080
Frequency 50 Hz/60 Hz
- Image Size: 1040*780
Environment Requirements
- Tiltable angle: up 30°; down 90°
Operating Environment
- Left/right rotation: ±90°
Ambient temperature 0° to 40°C
- View angle: right 178°, left 178°, up 178°,
Relative Humidity 15%~95% (no
down 178°
condensing)
- Brightness and Contrast adjustable (1-100)
Atmospheric pressure 86kPa-106kPa
- Folds down to horizontal level for transport.
Storage Environment
- Color temperature adjustable
Ambient temperature -20° to 55°C
Relative Humidity 15%~95% (no
Transducer Ports& Holders
condensing)
- Max. five active transducer ports
Atmospheric pressure 70kPa-106kPa
- (Some models equipped with only 3
transducer ports)

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Language Supported
- English System Ergonomic Design
- Chinese  Max. 5 active transducer ports
- German  20°Tiltable touch screen
- French  Control panel lift via dedicated pedal: 20
- Italian cm(model depended)
- Spanish  Display monitor left/right swivel: ±90°
- Russian  Tiltable LCD display
- Portuguese  Standard PC keyboard with dedicated swivel
- Polish for easy folding
 8 segment physical TGC sliders
I/O Ports  Rear storage tray.
- S-Video  Detachable metal frame for holding video
- USB 3.0(one) printer and DVD Drive
- USB 2.0(six)
- HDMI User Interface
- Ethernet Control Panel
 Interactive back-lighting
Wheel  Plastic and Rubber Hard Keys provides tactile
- Diameter: 5 inch feedback

- 4 wheels with brakes  User-defined hardkeys


 Physical trackball
Options  Up lift: 20 cm (Some models do no support
 Transducers lifting)

 Needle Guide Bracket Kits  8 segment TGC sliders


 Printers  Standard PC keyboard with dedicated swivel
 Printer holder for easy folding

 Batteries  High-performance audio speaker integrated


 1TB SSD with the control panel.

 WIFI module  5 transducer holders integrated with the


 ECG module control panel.

 Footswitch  Front and rear handles.


- Single button/Double buttons
- User-defined Functions(Freeze, Save, Touch Screen
Print)  14’’ Touch screen(resolution: 1366 x 768)
 Ultrasound gel warmer  20°tiltable
 DVD drive  Gesture-control
 Standard PC keyboard  6 user-defined touch screen keys
 Support visual Chinese, English QWERTY

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
keyboard, German QWERTZ keyboard and - Shortcut keys for selecting, viewing, deleting,
French AZERTY keyboard for text input exporting images
 Brightness and color temp. adjustable - Display of catheter size reference

Main Screen Display User Feedback Field


Information Field - Illustration of trackball and trackball keys
- EDAN logo - Cine bar
- Hospital name - Exit icon for exiting RawData review status.
- Date - The active function of user-defined key F1/F2.
- Time - PIP display
- Patient ID
- Patient Name Status Bar
- Patient Gender - Utility Icon(access to Utilities function)
- Patient Age - Image Store Icon
- Transducer model - USB Icon
- Exam Preset - Printer Icon
- LMP/BBT, GA, EDD - WIFI Icon
- User name - Network Transfer Status Icon
- Hard Drive Icon
Image Field - Battery Icon
- Mechanical Index (MI) - DVD Icon
- Thermal Index (TI)
- Imaging parameters User Login Management
- Gray Scale bar - Supports User Login at boot up and at exiting
- Depth Scale the sleep mode.
- Center Mark - Supports user type of Administrator and
- Measured result window Operator
- TGC curve - Supports switching users without powering
off the system.
Measurements Menu Field - Support Emergency login for emergency use.
- Available generic and application
measurements for current exam preset. Other Features
- Pre-categorized measurement groups.  eLearn instruction tool for basic scanning and
- Consistent with the display on Measurement nerve blocks.
Touch Screen(14-inch). - Support instructions for OB&GYN, Nerve
block, and GI(ABD, Cardiac, etc.)
Thumbnail Field(Clipboard) scanning.
- All captured static images and cine clips - Provides descriptions of Transducer

- Quick viewing the thumbnail in the image area position, Scan technique, Standard

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
ultrasound image, Anatomy, Needle - EM MSK
guide, tips, etc. - Nerve
- The illustration pictures can be enlarged - Sup Nerve (Superficial Nerve)
to full touch screen display by pressing it. - Shoulder
 One-key full screen zoom(3 levels) by - HIP
user-defined key F1/F2 or user-defined touch - Adult Cardiac
screen keys. - Pediatric Cardiac
- EM Adult Card
- Pediatric Abdomen
Exam Presets
- Neonatal Abdomen
 System pre-defined exam presets include
- Neonatal Head
(Transducer specific) :
- TCD
- ABD
- PICC
- Abd Difficult
- Needle-SMP
- EM Abd
- Aorta  User customizable presets: Copy, Delete,
Save as and rename
- Lung
- FAST  Exam presets are configurable in Set-up.

- Early OB  Supports a second page, up to 30 presets per


transducer.
- OB
- EM OB  Each preset can share comment, body mark,
and measure presets.
- Fetal Echo
- GYN  The display order of the exam presets of each
transducer can be adjusted per user's needs.
- IVF
- Urology
Annotations
- Prostate
Comments
- Thyroid
- Breast
- User-programmable home position

- Testis
- Arrow with user controlled orientation

- Carotid
- Five language soft keyboard: Chinese, English,
French, German, Russian.
- Up Ext A (Upper Extremity Artery)
- Up Ext V (Upper Extremity Vein)
- Block move and delete for separate blocks of
text
- Low Ext A (Lower Extremity Artery)
- Low Ext V (Lower Extremity Vein)
- Smart text replacement for predefined text
(e.g., Long replaces Trans with one keystroke)
- Vascular
- Vascular Access
- 545 pre-defined comments

- EM Vas
- User customizable

- Spine
- English Comments Library is supported when
the system language is not in English.
- MSK
- Sup MSK (Superficial MSK)

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Body Mark
- Up to 143 Body Mark graphics in library
- Support separate body mark in Dual/Quad
mode.

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Imaging
Imaging Mode Combinations
Imaging Modes
B-mode
- B/C(PDI or DPDI), Single

M-mode
- B/C(PDI or DPDI), Dual

- M-mode
- B+B/C(PDI or DPDI),Dual live

- Linear Anatomic M mode


- B+Color (PDI or DPDI)+M

- Curved Anatomic M mode


- B+PW (Duplex)

- Color M mode
- B+PW (Update)

Color Doppler
- HPRF

- Velocity-based color Doppler


- B/C(PDI or DPDI)+PW (Triplex)

- PDI
- B/C(PDI or DPDI)+PW (Update)

- DPDI
- B+CW (Update)

PW Doppler
- B/C(PDI or DPDI)+CW (Update)

CW Doppler
- B+TVI

TDI mode
- B+TVI+TEI

- TVI
- B+TVI+TVD(Update)

- TEI
- B+TVI+TVD(Triplex)

- TVD
- B+TVM (Update)

- TVM
- B+E
Imaging Parameters
3D/4D mode
B- mode(Live imaging)
Elastography Mode
Image Type Detail/General/Penetration
Display Modes
Auto(one-key TGC, Gain
Dual Imaging
optimization)
- Available for B and Color(PDI/DPDI) mode.
Digital Zoom x0.8-x2.0
- Displays two image side-by-side, two frozen
or one active/one frozen. Display Depth 1-45cm
Frequency 1-17MHz
- Allows to switch between two images.
Max. 5 fundamental + 5
- Measurements and calculations are
supported on each image and across the dual harmonic

images. Frequency display Bandwith, Frequency points

- Annotations are supported on each image. eSRI Off, Low, Med, High

Quad Imaging FOV Small, Med, Large, Full

- Available for B and Color(PDI/DPDI) mode. Steer 0°, ±10°

- Displays images in four quadrants, four frozen Gain 0-100dB, 1dB/step


or one active/three frozen. TGC 8 segments
- Allows to switch between four images. LGC 8 segments
- Measurements and calculations are Dynamic Range 40-96dB, 2dB/step
supported on each image. Line Density Low, Med, High
- Annotations are supported on each image.

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
≥512 lines - Map
Frame Rate Linear Transducer: Max. - Up/Down Flip
532f/s; - Left/Right Flip
Convex Transducer: Max. - Quick Rotation
48f/s at 18cm depth and full
FOV; M- mode(Live imaging)
Phased transducer: Max. Sweep Speed Fast/High/Med/Low/ Slow
77f/s at 18cm depth and full Corresponds to sweep time of
FOV. 1s, 2s, 4s, 8s and 12s per
Map 11 Types screen respectively.
Persistence Off, Low, Med, High Line Persist Off, Low, Med, High
Focus Position Max. 16 positions, Map 11 Types
adjustable Colorize On, off
Focus Number 1-3, adjustable Tint 5 Types
Colorize On, off Gain 0-100dB, 1dB/step
Tint 5 Types Frequency 1-17MHz
Up/Down Flip Max. 5 fundamental + 5
Left/Right Flip harmonic

Spatial On, off (max 3angles) Dynamic Range 40-96 dB, 2dB/Step
Compounding Strip size Full, large, Med., small
Trapezoid 8,12,20, off (Linear Side-by-side On(Left/Right)
transducer) Off(Up/Down)
Max. 10 ° left/right Acoustic Power 10%-100%, 10%/step
extended angle Linear Anatomic On, off
Acoustic Power 10%-100%, 10%/step M Mode Up to 3 linear sample lines
0°,90°,180°,270° Adjustable angle of each
On, Off sample line
Quick Rotation Real-time speed indicator Curved On, off
Panorama 360° rotation of Panoramic Anatomic M Free-hand drawing of sample
image Mode line; Sample line supports
B- mode(Post-processing & retrospective) edition and deletion.

- Gain M-mode(Post-processing & retrospective)


- TGC - Gain
- LGC - TGC
- Zoom - Dynamic range
- Dynamic range - Colorize
- eSRI - Map
- Colorize - Stripe Size
- Side-by-side

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Color/PDI/DPDI Mode(Live imaging) Color Hide On, off
Image Type High Flow/Mid Flow/Low Flow Panorama On, Off
Dual Live Real-time speed indicator
ROI size/position 360 ° rotation of Panoramic
Frequency Max. 5 levels image
Gain 0-100dB, 1dB/step Auto(one-key Gain
Line Density Low, Med, High optimization)
Dynamic Range 10-70 dB, 5dB/step Color/PDI/DPDI Mode
Not available for Color mode (Post-Processing & Retrospective)
Frame Rate Linear Transducer: Max - Zoom
350f/s; - Color map
Convex Transducer: Max. 8f/s - Invert (Not available for PDI mode)
at 18cm depth, full FOV and - Baseline
biggest ROI; - Color Hide
Phased transducer: Max. - VelDistr
14f/s at 18cm depth, full FOV
and biggest ROI. PW mode(Live imaging)
Persistence Off, Low, Med, High Image Type High Flow/Mid Flow/Low
Smooth Off, Low, Med, High Flow
Wall Filter Low, Med, High HPRF Automatic invocation to
Color Map 8 Types maintain gate location/scale

Steer Angle 0°,±15°, ±30° Auto Trace


(L12-5Q,Thyroid) Trace Side Up, down, both
0°,±10°,±20° Duplex Max. FR: 43f/s, depends on
(L12-5Q,Low Ext V) transducer
0°,±10°, ±15° Triplex Max. FR: 11f/s, depends on
(L12-5Q,Up Ext A) transducer
0°,±5°,±10° Frequency 2 levels
(L17-7SQ,MSK) PRF 0.9- 14.7kHz
PRF 0.6- 11.4kHz Max.PRF 0.34–27.9kHz
Max.PRF 0.1-21.7 kHz (L12-5HQ, Vasc Acc)
(L12-5HQ, Vasc Acc) Gain 0-100dB, 1dB/step
Baseline 25 levels Dynamic Range 10-70 dB, 5dB/step
(Not available for PDI mode) Wall Filter Low, Med, High
Threshold 0-100 Sweep Speed Fast/High/Med/Low/ Slow
Invert On, off Corresponds to sweep time of
(Not available for PDI mode) 2s, 3s, 4s, 6s and 8s per
Acoustic Power 10%-100%, 10%/step screen respectively.

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Baseline 9 levels - Invert
Angle Correction -80° to 80° - Strip size
Quick Angle -60°/-45°/-30°/-15°/0° - Auto trace
/15°/30°/45°/60 ° - Trace side
Steer 0°,±15°, ±30° - Quick Angle
(L12-5Q,Thyroid) - Side-by-side
0°,±10°,±20°
(L12-5Q,Low Ext V) CW mode(Live imaging)
0°,±10°, ±15° Image Type High Flow/Mid Flow/Low Flow
(L12-5Q,Up Ext A) PRF 1- 100kHz
0°,±5°,±10° Gain 0-100dB, 1dB/step
(L17-7SQ,MSK) Dynamic Range 10-70 dB, 5dB/step
Invert Wall Filter Low, Med, High
Volume 0-99 Sweep Speed Fast/High/Med/Low/ Slow
Map 11 Types Corresponds to sweep time of
Colorize On, off 2s, 3s, 4s, 6s and 8s per
Tint 5 Types screen respectively.

Gate Size 0.5-40 mm Baseline 9 levels

Strip size Full, large, Med., small Angle Correction -80° to 80°

Auto Gain, DR or Scale/Baseline, Quick Angle -60°/-45°/-30°/-15°/0°


(One-key user configurable /15°/30°/45°/60 °
Optimization) Invert
Acoustic Power 10%-100%, 10%/step Volume 0-99
PW velocity Max. 4.5m/s (correct angle Map 11 Types
60°); Colorize On, off
Max. 13m/s (correct angle Tint 5 Types
80°); Strip size Full, large, Med., small
Min. 2mm/s (Non-noise Acoustic Power 10%-100%, 10%/step
signal) CW velocity Max. 72m/s
Side-by-side On(Left/Right) Min. 1cm/s
Off(Up/Down) Auto Trace
PW Mode (Post-Processing & Retrospective) Trace Side Up, down, both
- Gain Side-by-side On(Left/Right)
- Dynamic Range Off(Up/Down)
- Colorize CW Mode (Post-Processing & Retrospective)
- Map - Gain
- Baseline - Dynamic Range
- Angle Correct - Colorize

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
- Map PRF 0.9- 5.9kHz
- Baseline Frequency 2 levels
- Angle Correction Gain 0-100dB, 1dB/step
- Invert Dynamic Range 10-70 dB, 5dB/step
- Strip size Wall Filter Low, Med, High
- Auto trace Sweep Speed Fast/High/Med/Low/ Slow
- Trace side (Corresponds to sweep time
- Quick Angle of 2s, 3s, 4s, 6s and 8s per
- Side-by-side screen respectively.)
Baseline 9 levels
TVI Mode (Live imaging) Angle Correction -80° to 80°
Image Type High Flow/Mid Flow/Low Flow Quick Angle -60°/-45°/-30°/-15°/0°
Dual Live B+ TVI /15°/30°/45°/60 °
ROI size/position Adjustable Invert On, Off
Frequency 2 levels Volume 0-99
Gain 0-100dB, 1dB/step Map 11 types
Line density Low, Med, High Colorize On, off
Persistence Off, Low, Med, High Tint 5 Types
Smooth Off, Low, Med, High Gate Size 0.5-40 mm
Wall Filter Low, Med, High Strip size Full, large, Med., small
Color Map 10 types Acoustic Power 10%-100%, 10%/step
PRF 0.6- 3.5kHz Auto Trace
Baseline 25 levels Trace Side Up, down, both
Threshold 0-100 Side-by-side On(Left/Right)
Invert On, off Off(Up/Down)
Acoustic Power 10%-100%,10%/step TVD(PW-TDI)(Post-Processing & Retrospective)
Color Hide On, off - Gain
TVI Mode (Post-Processing & Retrospective) - Dynamic Range
- Zoom - Colorize
- Baseline - Map
- Color map - Baseline
- Invert - Angle
- Color Hide - Quick Angle
- Invert
TVD mode(Live imaging) - Auto trace
Image Type High Flow /Mid Flow/Low
- Trace side
Flow
- Gate Size
Triplex B+TVI+TVD
- Side-by-side

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
3D/4Dmode(Live imaging) Review and Post-Processing functions
Acquisition 3D, 4D Cine Review
modes - Frame by frame manual review
Visualization Volume rendering, Multi-Slice - Auto playback with 6-level speed adjustable
modes - Start frame and end frame are selectable for
Multi-Slice Max. 21 slices can be cine loop review
displayed on the same screen; - Independent cine review in Dual/Quad mode.
Distance between each slice - Maximum cine memory depends on
is 0.5-10.0mm transducers and image parameters:
VOI size/Position Adjustable - 100000 frames for B mode

Render modes Surface, Max. - 30000 frames for Color mode

3D clip - 180s for M mode

Cut tools Trace, Box, Eraser - 240s for PW/CW Doppler mode

Cut functions Undo, Undo all, Redo


Display formats Single 3D, Dual(A-plane + 3D), Post-Processing Features

Quad(A/B/C Planes + 3D) All the image/cine is stored in Raw Data format in
local disk. The following Post-Processing features
3D parameters Threshold, Smooth,
are available when in image/cine review of current
Brightness, Contrast, Tint
exam or the stored exam.
eFace EDAN auto show face
- Adjusting imaging parameters
4D frame rate Max. 6vps
- Measurements
- Annotations
Elastography mode(Live imaging)
- Storing static image/ cine loop
Opacity 1, 2, 3, 4 levels
Smooth Off, Low, Med., High
Persistence Off, Low, Med, High
Map 0-6
DR 0-5
Invert On, Off

Elastography Mode
(Post-Processing & Retrospective)
- Opacity
- Map
- DR
- Invert

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Transducers and Biopsy Guide
Transducer Applications

Transducer Applications Transducer Applications


C5-1Q Abdomen C5-2Q Abdomen
Fetal / Obstetrics Fetal / Obstetrics
Urology Urology
Gynecology Gynecology
Musculoskeletal Musculoskeletal

L17-7HQ Small Parts L12-5Q Small parts


Peripheral Vascular Peripheral Vascular
Musculoskeletal Abdomen
Musculoskeletal
MC8-4Q Pediatric L17-7SQ Intra-operative
Abdomen Musculoskeletal
Neonatal Peripheral Vascular
Musculoskeletal
Peripheral Vascular

P7-3Q Adult Cardiac MC9-3TQ Pediatric


Pediatric Abdomen
Abdomen Neonatal
Pediatric Cardiac Musculoskeletal
Neonatal cephalic Peripheral Vascular
C6-2MQ Fetal / Obstetrics P5-1Q Adult Cardiac
Abdomen Abdomen
Gynecology Pediatric Cardiac
Urology Adult Cephalic

E10-3BQ Fetal / Obstetrics E8-4Q Fetal / Obstetrics


Gyncecology Gyncecology
Trans-vaginal Trans-vaginal
Trans-rectal Trans-rectal
Urology Urology
E10-3HQ Fetal / Obstetrics L12-5HQ Small Parts(Breast,
Gyncecology Testes, Thyroid)
Trans-vaginal Peripheral Vascular
Trans-rectal
Musculoskeletal
Urology
Abdominal

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Transducer Specifications

Transducer C5-1Q C5-2Q L12-5Q

Transducer Type Convex, Crystal Convex Linear

Bandwidth@ -6dB 2-5MHz 2-5MHz 5-11MHz

Centeral Frequency 3.25MHz 3.5MHz 8.0MHz


H2~4/H3~5/H2~
B Harmonic H2~4/H3~5
5/ H6~10/H7~12
Frequencies(MHz) / H2~5/ H3~4*
H3~6
B Fundamental
2~4/3~5/2~5 2~4/3~5/2~5 5~8/6~10/7~11
Frequencies(MHz)

Spectrum Doppler
2.3/3.0 2.3/3.0 4.7/5.7
Frequencies(MHz)

Color Doppler
2.9/3.2/3.5 2.2/2.7/3.2 5.2/5.9
Frequencies(MHz)

Elements 160 128 128

Footprint NA NA 38mm

Convex Radius 50mm 60mm NA

FOV 64° 60° NA

Display Depth 45cm 45cm 11cm

Biopsy Guide No Yes Yes

Cable Length 2.0m 2.0m 2.0m

Transducer L17-7HQ L17-7SQ MC8-4Q MC9-3TQ

Transducer Type Linear Linear Micro Convex Micro Convex

Bandwidth@ -6dB 7-15MHz 7-15MHz 4-9MHz 3-9MHz

Central Frequency 12.0MHz 12.0MHz 6.2MHz 6.4MHz


B Harmonic H4~7/H5~7/H5~
H9~13/H10-17 H9~13/H10-17 H5~8/H6~9
Frequencies(MHz) 8/H5~10/H6~10
B Fundamental 7~11/8~13/9~1 7~11/8~13/9~1 4~5/4~6/4~7/
3~6/4~7/5~9
Frequencies(MHz) 5 5 5~8/6~9
Spectrum Doppler
6.7/8.0 6.7/8.0 4.2/5.0 3.6/4.5
Frequencies(MHz)

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Color Doppler
6.7/8.0 6.7/8.0 4.2/5.0 3.6/4.5
Frequencies(MHz)

Elements 192 128 128 128

Footprint 38mm 26mm NA NA

Convex Radius NA NA 15mm 10mm

FOV NA NA 100° 150°

Display Depth 11cm 11cm 15cm 15cm

Biopsy Guide Yes No Yes Yes

Cable Length 2.0m 2.0m 2.0m 2.0m

Transducer P5-1Q P7-3Q C6-2MQ

Transducer Type Phased Phased Wobbler

Bandwidth@ -6dB 1-5MHz 3-7MHz 2-5MHz

Central Frequency 2.7MHz 5.0MHz 3.9MHz


B Harmonic H2~4/H3~5/H2~
H2~4/H3~5 H5~7/H6~8
Frequencies(MHz) 5
B Fundamental
1~3/2~4/2~5 3~5/4~6/5~7 2~4/3~5/2~5
Frequencies(MHz)
Spectrum Doppler
2.0/2.2 2.7/3.8 2.6/3.0
Frequencies(MHz)
Color Doppler
2.0/2.5 2.5/3.0 2.6/3.0/3.3
Frequencies(MHz)

Elements 64 96 128

Footprint 16 mm 15 mm NA

Convex Radius NA NA 40mm

FOV 90° 90° 64°

Display Depth 30cm 18cm 30cm

Biopsy Guide Yes No No

Cable Length 2.0m 2.0m 2.0m

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Transducer E8-4Q E10-3BQ E10-3HQ L12-5HQ

Transducer Type Intra-cavity Intra-cavity Intra-cavity Linear

Bandwidth@ -6dB 4-8MHz 4-8MHz 4-8MHz 5-11MHz

Centeral Frequency 6.2MHz 6.5MHz 6.5MHz 8.0MHz

H5~6/H5~8/ H5~6/H5~8/H5~ H6~10/H7~12/H


B Harmonic H5~6/H5~8/
H5~10/H6~11/ 10/H6~11/ 8~10/ H8~11/
Frequencies(MHz) H5~10/H6~11
H6~13 H6~13* H8~12
B Fundamental 5~8/6~10/7~11/
4~6/4~7/5~8 4~6/4~7/5~8 4~6/4~7/5~8
Frequencies(MHz) 8~10
Spectrum Doppler
3.6/5.0 3.6/4.8 3.6/4.8 4.7/5.2
Frequencies(MHz)
Color Doppler 5.2/5.9/7.2/7.5/
3.6/4.7 4.0/4.5/5.0 4.0/4.5/5.0
Frequencies(MHz) 7.9

Elements 128 192 192 192

Footprint NA NA NA 38mm

10mm 10mm
Convex Radius 10mm NA

FOV 150° 190° 190° NA

Display Depth 14cm 14cm 14cm 11cm

Biopsy Guide Yes Yes Yes Yes

Cable Length 2.0m 2.0m 2.0m 2.0m

NOTE: The asterisk "*" indicates that the frequency type is related to the transducer version. Among them:
optimized C5-2Q probe (02.01.212622015) add harmonic frequency type H3~4 and optimized E10-3HQ
probe (02.01.214791011) add harmonic frequency type H6~13.

Biopsy Guide

 Needle Guide
- Supports guide lines of multiple angles.
- Supports single and parallel guide line
- Supports depth and length mark on guide line.
- Support guide line calibration.

 Need Visualization
- Supports three needle inserted angles for linear transducers

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
 Center Line
- Center Line is a vertical dotted line displayed at the middle of the image field, representing the
middle of ultrasound beam.It helps to locate the position and depth of a target disease focus for
out-of-plane biopsy, lithotripsy and etc

 Supported Needle Guided Brackets

Model Type Angle/Depth Description

For use with the E8-4Q,


BGK-CR10UA In-plane 2°
Supports: 16G-18G

For use with the


BGK-002 In-plane 38°, 46°, 58° L12-5Q/L17-7HQ/L12-5HQ,
Supports: 14G-23G
For use with the
0.5cm, 1.0cm, 1.5 cm,
BGK-003 Out-of-plane L12-5Q/L17-7HQ/L12-5HQ,
2.5 cm, 3.5 cm
Supports: 21G-22G
For use with the MC9-3TQ,
BGK-004 In-plane 12°, 20°
Supports: 14G-23G

For use with the E10-3BQ,


BGK-005 In-plane 0°
Supports: 16G-18G

For use with the E10-3HQ,


BGK-006 In-plane 1°
Supports: 16G-18G

For use with the C5-2Q,


BGK-007 In-plane 18°, 25°, 35°
Supports: 14G-23G

For use with the P5-1Q


BGK-008 In-plane 12° , 22°
Supports: 14G-23G

For use with the C5-1Q


BGK-009 In-plane 14°, 20°, 32°
Supports: 14G-23G

For use with the MC8-4Q,


BGK-012 In-plane 11°, 20°, 37°
Supports: 14G-23G

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Measurements
 Default measurement unit options
- Distance: mm, or cm  HR

- Area: mm2, or cm2  Tei index: COT, ET


Doppler mode
- Volume: mm3, or cm3
 PS
 Caliper Size: switch automatically according
to the distance (3 sizes)  ED

 Dynamic display of measurement results  RI

 Reposition caliper  PI

 Pre-categorized measurement groups based  PS,ED,RI,S/D

on clinical applications; Configurable in  Time


Measure Preset. Measured results of each  HR
measurement are configurable in Measure  Manual Trace
Preset.  Spline Trace
 Measurements displayed on main screen and  Auto Trace(Max. 15 measured results are
touch screen are consistent. configurable)

 Velocity
General Measurements  PGMax
B-mode  PGMean
 Distance(2-point,Ratio D1/D2)  Volume Flow
 Circumference/Area (Ellipse, Trace,  Tei index: COT, ET
Spline,Ratio A1/A2)  dp/dt
 Angle(3-point) Elastography mode
 Volume(3-distance, Ellipse+ 1 distance)  Eratio(Ellipse, Trace)
 Stenosis
- %Dist Stenosis(Distance) Application Measurements/calculations
- % Area Stenosis (Ellipse, Trace, Spline) Abdomen
 Vessel B-mode:
- Vessel Diamemter(2-point, Ellipse)  Liver
- Volume flow area - Length, Width, Height
- IMT - Volume(calculation)
M-mode - Portal Vein Diameter
 Distance(2-point method) - Common Hepatic Duct
 Ratio D1/D2(2-point method)  Gallbladder
 Time - Length, Height
 Slope - Gallbladder Wall Thickness

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
- Common Bile Duct - Endometrium Thickness

 Pancreas - Uterus body


- Head, Body, Tail, Duct - Uterus volume

 Spleen - UT Cavity
- Length, Height  Cervix
 Renal - Length, Width, Height
- Length, Width, Height - UT-L/CX-L(calculation)
- Volume(calculation)  Ovary
- Renal Cortex Thickness - Length, Width, Height

 Aorta Diameter - Ovary volume

 Bladder  Follicle
- Pre-void bladder (Length, Width, Height)  Cyst
- Post-void bladder Length, Width, Height)  Fluid POD
- Micturated Volume  Pelvic Floor
- Pre-void bladder volume - BSD(R)
- Post-void bladder volume - BSD(S)
PW mode: PW mode:

 Abdominal Aorta  Uterine Artery


 Superior Mesenteric Artery  Ovary Artery
 Inferior Mesenteric Artery
 Hepatic Artery Obstetrics

 Splenic Artery B-mode:

 Renal Artery BPD, HC, AC, FL, HUM, CER,


 Fetal
 Portal Vein OFD, NF, TAD, APAD, THD,
Biometry
 Inferior Vena Cava APTD, TTD, FTA

 Main Portal Vein CRL, BPD, FL, HUM, NT, GS,


 Early Gest
 Hepatic Vein YS, AF

 Middle Hepatic Vein  Long HUM, ULNA, RAD, TIB, FIB,


 Splenic Vein Bones Foot

 Superior Mesenteric Vein  Fetal CER, NT, NF, LVW, CM,


 Inferior Mesenteric Vein Cranium NB,PNT, BOD

 AFI Q1, Q2,Q3,Q4


Gynecology
 Ovary Ovary L, Ovary W, Ovary H
B-mode:
LV Diam, LA Diam, RV Diam, RA
 Uterus  Chamber
Diam
- Length, Width, Height

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
LVOT Diam, Ao Asc, Ao Arch,  PV PV Diam
 LVOT/AO
Ao Isthmus, Desc Ao  RVDs RVDs
RVOT Diam, MPA Diam, Ductus RA length, RA Width, LA length,
 RVOT/PA  LA/RA
A LA width
PW mode:  LVM(A-L)
 MCA (Left LVAd Sax Epi, LVAd Sax Endo,
 Umb. A Ventricular LVAd Apical

 Planenta A Mass)

 Ovary A  LVM (T-E)


 Ut. A (Left LVAd Sax Epi, LVAd Sax Endo,

 Fetal Ao Ventricular a, d
Mass)
 Desc Aorta
 SMA  LVM

 IVC (Cube)(Left
LVSTd, LVIDd, LVPWd
Ventricular
 Ductus V
Mass)
 FHR
 MV  MVA

 TV  AVA

 MPV PW mode:

 Ductus A E/A, MV PHT, MV VTI, IVRT, MV


M-mode: A Duration, MV DecT, MR
 Mitral Valve
 FHR Vmax, MR VTI, MV E
Duration,MV HR

Cardiac  Tricuspid
TV VTI, TV Vmax, TV E/A, TV HR
B-mode: Valve

 LV A4C Dias., A4C Sys., A2C Dias., LVOT VTI, LVOT Vmax, AV VTI,
Simpson A2C Sys. AV HR, AV Vmax, AV Accel

LVSTd, LVIDd, LVPWd, IVSTs,  AV Time, AV Decel Time, AR VTI,


 LV Study AR Vmax, AR Accel Time, AR
LVIDs, LVPWs
PHT, AR Decel Time
 LV/RV LVIDd, LVIDs,RVAWd, RVIDd
PV VTI, PV Vmax, PR Vmax, PV
 LA/Ao LA,AoD  PV
Accel Time ,PV HR
 Aorta Ao Asc, AoD
PVein S Vel, PVein D Vel, PV A
 RVOT  PV Vein
RVOT Diam Vel, PV A Dur
Diam
Hep S Vel, Hep D Vel, Hep A
 LVOT Diam LVOT Diam  Hep Veins
Vel, Hep A Dur

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
 RVSP TR Vmax ,RA Pressure - Pre-void Bladder (Length, Width, Height,
volume)
 AVA(VTI) LVOT VTI,AV VTI
- Post-void Bladder (Length, Width, Height,
Sa Medial, Ea Medial, Aa
volume)
 TDI Medial, Sa Lateral, Ea Lateral,
Aa Lateral
- Micturated Volume
 Prostate
MR Trace, AR Trace, TR Trace,
 PISA - Length, Width, Height
PR Trace
 Seminal
 Qp/Qs LVOT VTI,RVOT VTI
- Length, Width, Height
M- mode:
 Testis
LVSTd, LVIDd, LVPWd, IVSTs,
 LV Study - Length, Width, Height
LVIDs, LVPWs
PW mode:
 LV/RV LVIDd, LVIDs,RVAWd, RVIDd
 Renal Artery
 Time LVET, LV PEP, RV PEP
 Arcuate Artery
MV D-E Exc, MV D-E Slope, E-F  Segmental Artery
 Mitral Valve Slope, EPSS, MV E-E Sep, MV  Interlobar Artery
A-C Interval, MAPSE

 TAPSE TAPSE Small Parts


 LA/Ao LA, AoR Diam, RVOT Diam,ACS B-mode:

 HR HR  Thyroid

 LVM - Length, Width, Height

(Cube)(Left - Thyroid Isthmus


LVSTd, LVIDd, LVPWd
Ventricular  Breast
Mass) - Lesion1, Lesion2, Lesion3, Lesion4,

 IVC-CI IVCmax, IVCmin, , IVC-CI(cal) Lesion5

C- mode:  Testis
- Length, Width, Height
MR Rad,MR Als. Vel,AR Rad,AR
PW mode:
 PISA Als. Vel, TR Rad, TR Als. Vel, PR
Rad, PR Als. Vel  Superior Thyroid Artery
 Inferior Thyroid Artery

Urology
Vascular
B-mode:

 Renal B-mode:

- Length, Width, Height Common Carotid Artery


 Carotid
- Renal Cortex Thickness Intima-Media Thickness,
Internal Carotid Artery
 Bladder

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Intima-Media Thickness, Vein, Posterior Tibial Vein,
Carotid Artery Bifurcation Anterior Tibial Vein, Small
Intima-Media Thickness Saphenous Vein ,Volume Flow
PW mode: PW mode:
Common Carotid Artery, Anterior Cerebral Artery, Middle
External Carotid Artery, Internal Cerebral Artery, Posterior
Carotid Artery, Vert Artery, Cerebral Artery, Anterior
 Cephalic
Subclavian Artery, HR, Volume Communicating Artery,
Flow Posterior Communicating
PW mode: Artery, Basilar Artery, Vertebral

 Upper Subclavian Artery, Axillary Artery, Internal Carotid Artery

Extremity Artery, Brachial Artery, Ulnar  Volume B mode:


Artery Artery, Radial Artery, Flow Volume Flow Area
HR ,Volume Flow
B mode:
PW mode: Stenosis% Distance
Subclavian Vein, AxillaryVein,
 Stenosis%
Stenosis% Area (Ellipse, Trace,
 Upper
Brachial Vein, Cephalic Vein, Spline)
Extremity
Basilic Vein, Ulnar Vein, Radial
Vein  Vessel
Vein, Median Cubital B mode: Vessel Diam
Diam
Vein,Volume Flow

PW mode:
Pediatrics
Common Femoral Artery, Deep
B-mode:
Femoral Artery, Superficial
Femoral Artery, Common Iliac
 Left lateral ventricle
 Lower  Right lateral ventricle
Artery, External Iliac Artery,
Extremity
Internal Iliac Artery, Popliteal  left trigone
Artery
Artery, Peroneal Artery,  right trigone
Posterior Tibial Artery, Anterior  Hip joint(with dislocation type)
Tibial Artery, Dorsalis Pedis - HIP Angle
Artery, HR,Volume Flow - HIP d/D
PW mode:
Common Femoral Vein, Deep Emergency
 Lower Femoral Vein, Superficial  EM Abd package
Extremity Femoral Vein, Common Iliac  EM OB package
Vein Vein, External Iliac Vein, Internal  EM Card package
Iliac Vein,Great Saphenous
Vein, Popliteal Vein, Peroneal

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Reports  Cine clips supports prospective and
retrospective storing.
 Editable worksheet
 The length of cine clip is configurable.
 Report type: ABD, GYN, OB, URO, VAS, SMP,
FETAL, CARD, PED, Nerve, MSK  Prospective storing: max. 2 min length of clip
can be stored in real-time scanning.
 Comments/Findings section
 Retrospective storing: all the clip data in the
 Supports fetal growth curve and grow bar
cine buffer can be stored in cine review mode,
display; supports data display of max. 4 fetus
max.2 min.
 Supports Fetus Score *and adding Fetus
 Supports up to 150,000(for 512GB hard
Score to the report
disk)or300,000(for ITB hard disk) lossless
 Supports data of multiple fetus
single frames.
 User-imported Report Header
 Supports cine clips exported to USB disk of:
 User-defined hospital logo
 Multiple number of selected images
Exam Database
 Multiple layouts of image in report.
Support exam storage without patient information
 Support select all images to add into the
Support exam query
report
Support review current exam or prior exam
 Report Layout supports auto adjust.
Support review images of an exam
 Support zoom in preview
Support review report of an exam
 Support Export as PDF format
Support export images as BMP,JPEG, TIFF, Raw
 Support print by report printer.
Data or DICOM format
 Support custom report information
Support export cine clip as AVI, MP4, WMV, Raw
 Support display BMI and BSA
Data or DICOM format
 Some information supports to adjust its
Support export Report as PDF format
display order in the report
Support export exams (including patient
 (Measurements/Growth Curve/Image/Fetus
information, images)
Score Result/Findings/Comment)
Support compare images
 Support display the time of system’s first use
displayed in the report
Exam Archiving
All Clips and Static images stored on the system
Image Storage& Exam Archiving are stored internally in Raw Data format. They can
Image Storage be archived to other storage device for long-term
 Static image/Cine clip is stored in local disk in storage as described below..
Raw Data format. - Archived to DICOM server.
 Two dedicated hard keys on the console for - Archived to FTP server.
capturing static image and cine clips - Archived to USB device.
respectively.
- Note: If a clip length exceeds 3s, when

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
selecting to export in DICOM format to a - Displays/hides the ended exams in the
DICOM server or USB stick, only the last 3s of worklist
the clip will be exported for this release. - Query conditions can be configured to
- Burned to DVD disk quickly filter exam information.
- Sent to mobile devices.  MPPS
- The MPPS service automatically sends
the exam status to the MPPS server at
Connectivity
the start and the end of the ultrasound
Network
exam.
 Wired network connection
- Displays Additional Materials item on
 Wi-Fi connection
patient information page.

 Storage Commitment
DICOM 3.0 Service
- Supports using the port information of
 DICOM Storage
the storage commitment server to
- Connectivity to DICOM server for storage
receive storage commitment information
of all static images or cine clips with
- Supports the establishment of a new
patient information.
association for receiving storage
- Manual-Transfer in background on
commitment information
Demand.
 DICOM Query/Retrieve
- Auto-Transfer when store or at exam
- Supports entering key words for query
end.
prior exams from DICOM server.
- Transfer management UI for viewing
- Supports download a queried exam to local
transfer task status, retransferring or
disk for reviewing.
deleting the tasks manually.
 DICOM Print
- Transfer process encrypted.
- Prints the images remotely via a DICOM
- Supports Structured Report transferring:
printer which connects to a DICOM server.
OB, GYN, Cardiac and Vascular.
- Multiple parameters for printing are
 DICOM Modality Worklist
configurable.
- Enables query of the patient worklist
schedule from hospital information
FTP Network Store Service
system to the ultrasound system via
- Supports to transfer exams to FTP
DICOM network connection.
servers for storage in the background.
- Query of worklist on demand or on start
- Transfer management UI for viewing
of exam.
transfer task status, retransferring a task
- Populates the Patient Information screen
or deleting a transfer task.
with patient demographic information
- A PDF report can be sent to FTP server
automatically when one patient is
together with the exam.
selected.

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Cloud Share - HP Color Laser 150nw
- Supports sending image/clips to mobile - HP Laser 103a
devices by scanning the QR code on - EPSONL3118
main screen when CloudShare icon is  Photo printer
clicked. - Cannon CP1300
Time Synchronization
 Network report printer
Sets the system time to synchronize with the
network time
Mini PC
Supported Peripherals
Printers Safety and Regulatory
The system supports the connection of Video The Acclarix LX3 series Diagnostic Ultrasound
printers and report printers. The report can be System have been designed, manufactured and
connected locally via USB connector or remotely tested to comply with the following internationally
via network connection. Printer drivers can be recognized standards:
imported to the system for the support of more - IEC 60601-1: Medical Equipment Safety
report printers. - IEC 60601-1-2: Medical Device
 Video printers Electromagnetic Safety
- SONY UP-X898MD - IEC 60601-2-37: Ultrasonic Medical
- SONY UP-D25MD Equipment Safety

- SONY UP-25MD - IEC 62133: Battery Safety

 Local report printer - IEC 62304: Medical Device Software

- HP LaserJet CP1525n Color Life-cycle Process

- HP Deskjet Ink Advantage 2010 - IEC 62366: Medical Device Usability


Engineering
- HP Deskjet 1010
- EN ISO 14971: Medical Device Risk
- HP Deskjet 1510
Management
- HP DeskJet Ink Advantage Ultra 2029
- ISO 10993-1 Biological evaluation of medical
- HP DeskJet 1112
devices — Part 1:Evaluation and testing within
- Canon iP2780
a risk management process sheet
- HP LaserJet Pro MFP M126nw
- EPSON L310
- HP DeskJet 1050
- HP DeskJet 2050
- HP LaserJet M252n
- EPSON L130
- HP Color LaserJet Pro M254nw
- HP Color Laser 150a

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.
Revision History
Version Revisions Date
1.0 Initial release. 2020-7-27
1.1 Update transducer related information. See the changes highlighted with 2020-11-26
blue color.
1.2 Updated for V1.10 release. See the changes highlighted with blue color. 2021-03-05
1.3 Updated for V2.00 release. See the changes highlighted with blue color. 2021-12-02
1.4 Updated for V2.10 release. See the changes highlighted with blue color. 2022-04-10
1.5 Updated for R2.20 release. See the changes highlighted with blue 2022-08-29
color.

This datasheet applies to AcclarixLX3 series Diagnostic Ultrasound Systems, including AcclarixLX3,
Acclarix LX3 Exp, Acclarix LX3 Super, Acclarix LX25 and Acclarix LX28 models. The configuration
difference between each model is listed in the following table.

Configuration Difference
Models Feature 1 Feature 2 Feature 3
Seminal Vesicle Meas. Testis Meas. Fluid POD
Acclarix LX3 X √ √
Acclarix LX3 Exp √ X √
Acclarix LX3 Super √ √ √
Acclarix LX25 X √ X
Acclarix LX28 √ X X

*Feature is subject to regulatory approval, and may not be available for sale in specific countries.

Features and specifications are subject to change without prior notice. No reproduction, copy or transmission may be
made without written permission. Not all products or features are available in all countries, contact Edan for local availability.

You might also like