User Manual PM60
User Manual PM60
Pulse Oximeter
Operator’s Manual
© 2007-2018 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights
I
Intellectual Property Statement
otherwise, of Mindray in China and other countries. All other trademarks that appear
in this manual are used only for informational or editorial purposes. They are the
property of their respective owners.
II
Responsibility on the Manufacturer Party
All information contained in this manual is believed to be correct. Mindray shall not be
liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this
product, only if:
all installation operations, expansions, changes, modifications and repairs of
this product are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable
national and local requirements;and
the product is used in accordance with the instructions for use.
WARNING
NOTE
III
Warranty
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation
or other charges or liability for direct, indirect or consequential damages or delay
resulting from the improper use or application of the product or the use of parts or
accessories not approved by Mindray or repairs by people other than Mindray
authorized personnel.
IV
Company Contact
Tel: 0049-40-2513175
Fax: 0049-40-255726
V
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and in
accordance with its function and intended use. Observance of this manual is a
prerequisite for proper product performance and correct operation and ensures patient
and operator safety.
This manual is based on the maximum configuration and therefore some contents may
not apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the
equipment so that it can be obtained conveniently when needed.
Intended Audience
This manual is geared for clinical professionals who are expected to have a working
knowledge of medical procedures, practices and terminology as required for monitoring
of critically ill patients.
Illustrations
All illustrations in this manual serve as examples only. They may not necessarily reflect
the setup or data displayed on your pulse oximeter.
Password
Password is required to access maintenance. The password is 321.
Conventions
Italic text is used in this manual to quote the referenced chapters or sections.
[ ] is used to enclose screen texts.
→ is used to indicate operational procedures.
VI
Contents
1 Safety ......................................................................................................................... 1-1
1.1 Safety Information .......................................................................................... 1-1
1.1.1 Dangers .............................................................................................. 1-2
1.1.2 Warnings ............................................................................................ 1-2
1.1.3 Cautions ............................................................................................. 1-3
1.1.4 Notes .................................................................................................. 1-4
1.2 Equipment Symbols ........................................................................................ 1-5
2
6.4.1 Switching On/Off SpO2 and PR Alarms ........................................... 6-3
6.4.2 Setting Alarm Level ........................................................................... 6-3
6.4.3 Adjusting the Alarm Limits ............................................................... 6-4
6.4.4 Switching On/Off the Alarm Limit Display ...................................... 6-4
6.4.5 Setting SpO2 Sensitivity..................................................................... 6-4
6.5 Measurement Limitations ............................................................................... 6-5
3
B EMC ......................................................................................................................... B-1
4
1 Safety
DANGER
WARNING
CAUTION
NOTE
1-1
1.1.1 Dangers
There are no dangers that refer to the product in general. Specific “Danger” statements
may be given in the respective sections of this manual.
1.1.2 Warnings
WARNINGS
1-2
in a hazard to the patient. Remember that alarm settings should be
customized according to different patient situations and always keeping
the patient under close surveillance is the most reliable way for safe
patient monitoring.
The physiological data and alarm messages displayed on the equipment
are for reference only and cannot be directly used for diagnostic
interpretation.
To avoid inadvertent disconnection, route all cables in a way to prevent a
stumbling hazard. Wrap and secure excess cabling to avoid risk of
entanglement or strangulation by patients or personnel.
Do not open the equipment housings. All servicing and future upgrades
must be carried out by the personnel trained and authorized by our
company only.
1.1.3 Cautions
CAUTIONS
To ensure patient safety, use only parts and accessories specified in this
manual.
At the end of its service life, the equipment, as well as its accessories,
must be disposed of in compliance with the guidelines regulating the
disposal of such products. If you have any questions concerning disposal
of the equipment, please contact us.
Magnetic and electrical fields are capable of interfering with the proper
performance of the equipment. For this reason make sure that all
external devices operated in the vicinity of the equipment comply with
the relevant EMC requirements. Mobile phone, X-ray equipment or MRI
devices are a possible source of interference as they may emit higher
levels of electromagnetic radiation.
Before connecting the equipment to the power line, check that the voltage
and frequency ratings of the power line are the same as those indicated
1-3
on the equipment’s label or in this manual.
Always install or carry the equipment properly to avoid damage caused
by drop, impact, strong vibration or other mechanical force.
1.1.4 Notes
NOTES
Put the equipment in a location where you can easily see the screen and
access the operating controls.
Keep this manual in the vicinity of the equipment so that it can be
obtained conveniently when needed.
The software was developed in compliance with IEC60601-1-4. The
possibility of hazards arising from software errors is minimized.
This manual describes all features and options. Your equipment may not
have all of them.
During normal use, the operator is expected to face the front of the
equipment.
1-4
1.2 Equipment Symbols
Caution
Input/Output
AUDIO PAUSED
Left/Right button
Up button
Down button
Date of manufacture
Manufacturer
Serial number
1-5
DEFIBRILLATION-PROOF TYPE BF APPLIED PART
1-6
2 The Basics
2.1 Introduction
2.1.1 Intended Use
The pulse oximeter is intended for continuously monitoring, spot checking, displaying,
storing and transferring oxygen saturation and pulse rate of single adult, pediatric and
neonatal patients in hospitals, emergency treatment, patient transport and home care.
WARNING
2.1.2 Contraindications
None.
2.1.3 Components
This pulse oximeter consists of a main unit and an SpO2 sensor.
2-1
2.2 Main Unit
2.2.1 Front View
8
6 7
5 2
9
3
10
1. Display screen
2. Power button
Press this button to turn the pulse oximeter on after the batteries are installed.
Press and hold it for 2 seconds to turn the pulse oximeter off.
3. Power indicating lamp
It is a LED that lights green and yellow. The status of the LED is specified as
follows:
2-2
Green: when the pulse oximeter is placed in the Charger stand and the AC
mains is connected, or when the battery is fully charged if a lithium battery
is used.
Yellow: when a lithium ion battery is used and is being charged.
Off: When the AC mains is not connected.
4. Alarm indicating lamp
When an alarm occurs, this lamp will light up as defined below:
High level alarms: the lamp quickly flashes red.
Medium level alarms: the lamp slowly flashes yellow.
Low level alarms: the lamp lights yellow without flashing.
5. Press this button to pause or reactivate the alarm sound.
6. Left button
Press this button to enter the main menu or select the highlighted menu item.
7. Right button
Press this button to return to the previous menu or exit the current menu. In the
case that no menu is opened, you can press and hold this button for 2 seconds to
lock or unlock buttons except the Power button.
8. Up button
Press this button to move the cursor upwards, increase the value of selected menu
item or raise the beat volume.
9. Down button
Press this button to move the cursor downwards, decrease the value of selected
menu item or lower the beat volume.
10. Power supply connector
It is used to connect the Charger stand.
2-3
2.2.2 Rear View and Right View
4
2
1. Multifunctional connector
It is used to connect an SpO2 sensor to measure the oxygen saturation or connect a
personal computer through a PC communication cable to export the trend data.
2. Speaker
3. Battery door
4. Cord hole
2-4
2.3 Display Views
The following figures show the layout of the wave screen and the normal screen.
3 4
2 5 2
1 5
7
8
1. Waveform Area
This area displays Pleth waveform (Pleth). The label of this waveform is shown at
the top left corner.
2. SpO2 area
3. Technical alarm area
This area shows the technical alarm message, prompt message and the pulse
oximeter’s operating mode. When multiple messages come, they will be displayed
circularly.
4. Alarm status area
1
2
3 5
4
6
1. SpO2 label
2. SpO2 unit
3. SpO2 high alarm limit
4. SpO2 low alarm limit
5. Oxygen saturation reading
6. Pleth bar
2-6
2.3.2 PR Area
1
2
3 5
4
1. PR label
2. PR unit
3. PR high alarm limit
4. PR low alarm limit
5. Pulse rate reading
1 2 3 4
1. Patient ID
2. Patient category
Patients are classified into three categories: adults, pediatrics and neonates.
3. Battery symbol
4. Real-time clock
2-7
FOR YOUR NOTES
2-8
3 Getting Started
Before unpacking, examine the packing case carefully for signs of damage. If any
damage is detected, contact the carrier.
If the packing case is intact, open the package and remove the equipment and
accessories carefully. Check all materials as per the packing list and check for any
mechanical damage. Contact us in case of any problem.
NOTE
Save the packing case and packaging material as they can be used if the
equipment must be reshipped.
WARNING
3-1
3.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in
this manual.
When the equipment is moved from one place to another, condensation may occur as a
result of temperature or humidity difference. In this case, never start the system before
the condensation disappears.
WARNING
Make sure that the operating environment of the equipment meets the
specific requirements. Otherwise the equipment may not meet the
specifications claimed in this manual and unexpected consequences, e.g.
damage to the equipment, could result.
3-2
NOTE
If the pulse oximeter has not used for a prolonged time, a technical alarm
[Clock Need Set] may occurs when the pulse oximeter is started. In this
case, set the clock. In the continuous monitoring mode, changing the
clock may clear the trend data of this mode.
WARNING
3-3
FOR YOUR NOTES
3-4
4 Basic Operations
The work mode is shown in the technical alarm area. When other technical alarms or
prompt messages occur, the work mode and the message will be displayed circularly.
NOTE
The trend data of the continuous monitoring mode will be cleared if you
change the continuous monitoring mode to the spot-checking mode.
4-1
4.2 Admitting a Patient
WARNING
Be sure to select correct patient category setting for your patient before
measurement. Wrong patient category may result in patient hazard due
to improper alarm limits.
4-2
4.3 Selecting the Screen
To select a screen to be viewed,
1. Select [Menu]→[System].
2. Select [Screen] and toggle between [Normal] and [Wave].
NOTE
If the the pulse oximeter is used outdoors or the ambient light is strong,
set the screen brightness to a higher level.
4-3
4.6 Setting the Clock
To set the clock,
1. Select [Menu]→[System]→[Clock >>].
2. Set [Date] and [Time].
3. Select [Format >>]. In the [Format] menu,
Set [Date Format] to [yyyy-mm-dd], [mm-dd-yyyy] or [dd-mm-yyyy].
Set [Time Format] to [24 h] or [12 h].
4 Press the Right button to return to the previous menu in the spot-checking mode, or
press the Right button and follow the prompt in the continuous monitoring mode.
NOTE
and 10 is the maximum volume. When [Beat Vol] is set to 0, the icon is
shown in the PR area.
You can increase/decrease the beat volume by pressing the Up/Down button in the case
that no menu is opened.
During SpO2 monitoring, the pitch of the pulse tone changes as the patient’s oxygen
saturation level changes. The pitch of the tone rises as the oxygen saturation level
increases and falls as the oxygen saturation level decreases.
4-4
4.7.2 Setting the Key Volume
To set the key volume,
1. Select [Menu]→[General Setup].
2. Set [Key Vol] to a value between 0 and 10. 0 means the key volume is turned off,
and 10 is the maximum volume.
In the demo mode, the message [Demo Mode] is shown in the technical alarm area. To
exit the demo mode, press the Right button to return to the screen displayed before you
enter the demo mode.
WARNING
The Demo mode is for demonstration purpose only. To avoid that the
simulated data are mistaken for the monitored patient’s data, you should
not enter the Demo mode during a patient is being monitored. Otherwise,
improper patient monitoring and delayed treatment could result.
4-5
4.10 Entering/Exiting the Standby Mode
4.10.1 Entering the Standby Mode
In the spot-checking mode, in situations where reviewing or exporting trend data is not
performed, the pulse oximeter will automatically enter the standby mode if no button
operation and SpO2 signal are detected for 1 minute and no "Battery Too Low" alarm
occurs. In the standby mode,
The backlight is turned off.
The standby screen is displayed.
Alarms excetp the “Battery too Low” alarm are disabled.
In the standby mode, if the “Battery too Low” alarm occurs, the system gives an audible
alarm which is defined as follows:
Double beep
Alarm volume: 2
Alarm interval: 30 seconds.
If you exit the standby mode at this time, the “Battery too Low” alarm will be given in
the form of techinical alarm.
4-6
4.11 Setting Auto Poweroff
In the spot-checking mode, you can select to shut off the pulse oximeter automatically:
1. Select [Menu]→[System]→[Maintenance >>]→enter the required password.
2. Select [User Maintenance >>], and then select [Auto Quit] to toggle between:
[Allowed]: The pulse oximeter will shut down automatically if no button
operation and SpO2 signal are detected for 5 minutes in the standby mode.
[Unallowed]: The pulse oximeter will not shut down if no button operation
and SpO2 signal are detected for 5 minutes in the standby mode.
4.12 Configuration
The pulse oximeter provides factory configuration and user configuration. The
configuration is loaded and saved according to the operating mode and patient category.
SpO2 and PR alarm limits are subject to the patient category. The user configuration will
be loaded first when the pulse oximeter is started. If the user configuration is not
available, corresponding factory configuration will be loaded. Refer to C Factory
Defaults.
The factory configuration will be restored according to the current work mode and
patient category.
4-7
4.12.2 Saving the User Configuration
You can change the pulse oximeter’s settings and save the changed settings as user
configuration.
The user configuration will be saved according to the current work mode and patient
category.
The system will select an appropriate user configuration according to the current work
mode and patient category. If the user configuration of this type is not available,
corresponding factory configuration will be restored.
4-8
5 Alarms
Alarms, triggered by a vital sign that appears abnormal or by technical problems of the
pulse oximeter, are presented to the user by visual and audible alarm indications.
1. Physiological alarms
Physiological alarms, also called patient status alarms, are triggered by a monitored
parameter value that violates set alarm limits or an abnormal patient condition.
Physiological alarm messages are displayed in the physiological alarm area.
2. Technical alarms
Technical alarms, also called system status alarms, are triggered by a device
malfunction or a patient data distortion due to improper operation or system
problems. Technical alarm messages are displayed in the technical alarm area.
3. Prompt messages
As a matter of fact, prompt messages are not alarm messages. Apart from the
physiological and technical alarm messages, the pulse oximeter will show some
messages telling the system status. Prompt messages are displayed in the technical
alarm area.
5-1
5.2 Alarm Levels
By severity, the pulse oximeter’s physiological alarms can be classified into three
categories: high level alarms, medium level alarms and low level alarms.
The pulse oximeter’s technical alarms can be classified into three categories: high level
alarms, medium level alarms and low level alarms.
The levels for some technical alarms and physiological alarms are predefined before the
pulse oximeter leaves the factory and cannot be changed.
For different alarm levels, the alarm lamp, alarm tone and alarm messages presented are
different.
5-2
5.3.1 Alarm Lamp
If a technical or a physiological alarm occurs, the alarm lamp will flash. The flashing
color and frequency match the alarm level as follows:
High level alarms: the lamp quickly flashes red.
Medium level alarms the lamp slowly flashes yellow.
Low level alarms: the lamp turns yellow without flashing.
For physiological alarms, the asterisk symbols (*) before the alarm message match the
alarm level as follows:
High level alarms: ***
Medium level alarms: **
Low level alarms: *
Additionally, the system uses different background colors for the alarm message to
match the alarm level:
High level alarms: red
Medium level alarms: yellow
Low level alarms: yellow
5-3
NOTE
5-4
5.4.2 Changing the Alarm Volume
When alarm volume is set to 0, the alarm sound is turned off and a symbol
appears in the alarm status area. In the case that alarm sound is switched off, the alarm
lamp remains lit and alarm message remains presented. The audible alarm is reactivated
automatically when:
The pulse oximeter is shut down and restarted;
The factory configuration is loaded;
The user configuration is loaded.
When a factory or user configuration is selected, the alarm volume of the pulse oximeter
may be lower than the minimum alarm volume. In this case, the alarm volume is
automatically adjusted according to the minimum alarm volume.
WARNING
When the alarm sound is switched off, the pulse oximeter will give no
audible alarm tones even if a new alarm occurs. Therefore the user
should be very careful about whether to switch off the alarm sound or
not.
Do not rely exclusively on the audible alarm system for patient
monitoring. Adjusting alarm volume to a low level may result in a hazard
to the patient. Always keep the patient under close surveillance.
5-5
5.4.3 Pausing the Alarm Tones
The audible alarm is paused, but the alarm lamp remains lit and the alarm
message remains displayed;
The remaining alarm pause time is displayed in the alarm status area;
Audible alarm starts again automatically after the alarm pause period expires. You can
also press the key to restart the audible alarm. The audible alarm is reactivated
automatically when:
The pulse oximeter is shut down and restarted.
The pulse oximeter is switched to a new operating mode
The pulse oximeter enters or exits the standby mode
The pulse oximeter enters or exits the demo mode.
Alarm pause time can be adjusted and the default is 2 minute. To change the alarm
pause time,
1. Select [Menu]→[System]→[Maintenance >>]→enter the required password.
2. Select [Alarm >>] and then set [Audio Paused] to an appropriate time.
5-6
5.5 Setting the Alarm Level
The levels of all technical alarms except [Sensor Off] are predefined before the pulse
oximeter leaves the factory. To set the alarm level of [Sensor Off],
1. Select [Menu]→[Alarm Setup];
2. Select [Sensor Off] and choose a desired alarm level.
If you enter the demo mode after the [Sensor Off] alarm occurs, the alarm will be
presented in the form of prompt message when you exit the demo mode.
In the continuous monitoring mode, the alarm level of [Sensor Off] can be set to [High],
[Med] or [Low]. If theSpO2 sensor falls off before a valid SpO2 signal is obtained, the
pulse oximeter will present [Sensor Off] in the form of prompt message.
In the spot-check mode, the alarm level of [Sensor Off] can be set to [High], [Med],
[Low] or [Off].
If the [Sensor Off] alarm is not switched off, the pulse oximeter will present
the alarm in the form of prompt message in the case that the SpO2 sensor falls
off before an effective SpO2 signal is obtained.
If the pulse oximeter enters the standby mode after the [Sensor Off] alarm
occurs, it will present the alarm in the form of prompt message when the
standby mode is exited.
5-7
5.6 Switching On/Off the Reminder Tone
When the alarm volume is set to 0, the pulse oximeter gives a reminder tone of two
beeps every three minutes if reminder tone is switched on.
The reminder tone is switched off by default. The setting of the reminder tone is saved
even if the pulse oximeter is turned off.
5-8
6 Measuring SpO2
6.1 Introduction
SpO2 measuring is a non-invasive technique used to measure the amount of oxygenated
haemoglobin and pulse rate by measuring the absorption of selected wavelengths of
light. The light emitted by a red and infrared light-emitting diodes passes through the
tissue and is converted into electrical signals by a photodiode. This device is calibrated
to display functional oxygen saturation.
2 3
4
6-1
6.2 Safety
WARNING
Use only SpO2 sensors specified in this manual. Follow the SpO2 sensor’s
instructions for use and adhere to all warnings and cautions.
Check the SpO2 sensor and its package for any sign of damage before use.
Do not use the sensor if any damage is detected.
When a trend toward patient deoxygenation is indicated, blood samples
should be analyzed by a laboratory co-oximeter to completely understand
the patient’s condition.
Do not use the pulse oximeter and the SpO2 sensor during magnetic
resonance imaging (MRI). Induced current could cause burns.
Prolonged continuous monitoring may increase the risk of undesirable
changes in skin characteristics, such as irritation, reddening, blistering or
burns. Inspect the sensor site every two hours and move the sensor if the
skin quality changes. Change the application site every four hours. For
neonates or patients with poor peripheral blood circulation or sensitive
skin, inspect the sensor site more frequently.
As with electrosurgical unit, carefully route patient cabling to avoid
entanglement.
Do not use the SpO2 sensor on a limb with an intravenous infusion or
arterial catheter in place.
Do not use the SpO2 sensor on a limb where the NIBP cuff is applied. This
may result in inaccurate SpO2 reading due to blocked blood flow during
cuff inflation.
Do not apply the equipment to a patient younger than one year old if the
operating temperature is higher than 40℃. This may lead to skin burn.
6-2
6.3 Applying the Sensor
1. Select an appropriate SpO2 sensor according to the patient category and weight.
2. Remove colored nail polish from the application site.
3. Apply the SpO2 sensor to the patient.
4. Connect the SpO2 extension cable to the pulse oximeter.
switched off and the symbol will be shown in the SpO2 or PR area.
6-3
6.4.3 Adjusting the Alarm Limits
6-4
To set the SpO2 sensitivity,
1. Select [Menu]→[General Setup].
2. Set [Sensitivity] to [High], [Med] or [Low], whose averaging time is respectively
7 seconds, 9 seconds and 11 seconds.
6-5
FOR YOUR NOTES
6-6
7 Data Management
NOTE
The stored data will not be cleared in case of power failure or power off.
It's recommended to export the data before the memory is filled up.
7-1
If the patient ID is changed, a new patient is considered to be admitted and the
new measurements will be stored under the new patient ID. The data under
the old patient ID will be cleared.
When the memory is filled up, the oldest data will be overwitten by the new data.
7-2
7.2 Reviewing Trend Data
The patient’s history physiological data can be stored and displayed in the form of a
trend table. Reviewing the trend data helps you to understand changes in the patient
condition.
In the trend window, SpO2 and PR readings beyond the alarm limits are displayed in red
and the blank lines indicate that the pulse oximeter is shut off during that period.
7-3
Select [Start Time >>] to set the [Date] and [Time] from which you
want to review.
Select [Delete All] to delete all the trend data under the current patient
ID.
Press the Right button to exit the trend window.
In the trend window, SpO2 and PR readings beyond the alarm limits are displayed in red.
If you have changed the system time before entering the trend window, the trend data
time before the system time is changed remains unchanged.
7-4
Select [Select ID] to review the history data as well as the maximum,
minimum and average measurements of the selected paitent.
Select [Delete Selected] to delete the trend data under the selected
patient ID.
Select [Delete All] to delete the trend data under all patient IDs.
Press the Right button to exit the trend window.
When the trend exporting window is entered, all operations except exiting the trend
export window and turning off the pulse oximeter are disabled. To exit the trend
exporting window, press the Right button and select [OK].
Refer to the PC data management system software for detailed information on trend
data exporting.
7-5
FOR YOUR NOTES
7-6
8 Battery
8.1 Overview
The pulse oximeter is designed to operate on three 1.5V alkaline AA batteries or a
rechargeable lithium ion battery.
When the alkaline batteries are used, the battery icon indicates the battery status as
follows:
Indicates that the batteries work correctly. The solid portion
represents the current power level of the batteries in proportion to its
maximum power level..
Indicates that the batteries have low power level and need to be
replaced.
When a lithium ion battery is used, the battery icon indicates the battery status as
follows:
Indicates that the battery works correctly. The solid portion
represents the current charge level of the battery in proportion to its maximum
charge level.
Indicates that the battery has low charge level and need to be
charged.
8-1
If the battery capacity is too low, a technical alarm will be triggered and the [Battery
Too Low] message displayed. At this moment, replace the batteries if the alkaline
batteries are used, or charge the battery if a lithium ion battery is used. Otherwise, the
pulse oximeter will shut down automatically when the battery is depleted. If interrupting
patient monitoring is not allowed at this moment, you can place the pulse oximeter in
the Charger stand and connect the AC mains. In this case, the pulse oximeter will give
prompt message [Battery Type Err] if the alkaline batteries are used. You must pay
close attention to the power supply to the pulse oximeter and replace the alkaline
batteries or charge the lithium ion battery as soon as interruption of patient monitoring
is permissible.
You can charge the lithium ion battery using a Charger stand whether the pulse oximeter
is turned on or off. However, monitoring a patient while the battery is being charged is
not recommended.
NOTE
WARNING
8-2
8.2 Installing the Batteries
8.2.1 Opening the Battery Door
1. Remove the pulse oximeter from the Charger stand and disconnect the SpO2
sensor.
2. Use the key to loose the screw that secures the battery door to the pulse oximeter.
3. Press the battery door, push it downwards and remove the battery door.
8-3
8.2.2 Installing the Alkaline Batteries
1. Insert the AA alkaline batteries in the battery compartment, aligning the + on each
battery with the + shown inside the battery compartment.
2. Close the battery door and push it upwards.
3. Tighten the screw that secures the battery door to the pulse oximeter.
Caution
Do not run the pulse oximeter using alkaline batteries of different types
or capacities at the same time.
2. Insert the lithium ion battery in the battery compartment, aligning the + on the
battery with the + shown inside the battery compartment.
3. Close the battery door and push it upwards.
4. Tighten the screw that secures the battery door to the pulse oximeter.
8-4
8.3 Charging the Lithium Ion Battery
Pulse oximeter
WARNING
Do not use the charger stand when the alkaline batteries is depleted or no
battery is installed.
Monitoring a patient while the battery is being charged is not
recommended.
8-5
8.4 Conditioning the Lithium Ion Battery
A lithium ion battery needs at least two conditioning cycles when it is put into use for
the first time. A battery conditioning cycle is one complete, uninterrupted charge of the
battery, followed by a complete, uninterrupted discharge of the battery. A lithium ion
battery should be conditioned regularly to maintain its useful life. Condition a battery
once when it is used or stored for two months, or when its run time becomes noticeably
shorter.
8-6
8.5 Checking the Lithium Ion Battery
The performance of a rechargeable lithium ion battery may deteriorate over time. To
check the performance of a battery, follow this procedure:
1. Disconnect the pulse oximeter from the patient and stop all monitoring and
measuring procedures.
2. Place the pulse oximeter in the Charger stand and connect the AC mains. Allow the
battery to be charged uninterruptedly for above 2 hours.
3. Disconnect AC mains and allow the pulse oximeter to run on the battery until it
shuts off.
The operating time of a battery reflects its performance directly. If the operating time of
a lithium ion battery is noticeably shorter than that stated in the specifications, replace it
or contact your service personnel.
NOTE
8-7
8.6 Disposing of the Batteries
Batteries that are damaged or depleted should be replaced and discarded properly.
Dispose of used batteries according to local regulations.
WARNING
8-8
9 Maintenance and Cleaning
Use only the substances approved by us and methods listed in this chapter to clean or
disinfect your equipment. Warranty does not cover damage caused by unapproved
substances or methods.
We make no claims regarding the efficacy of the listed chemicals or methods as a means
for controlling infection. For the method to control infection, consult your hospital’s
Infection Control Officer or Epidemiologist.
Keep you equipment and accessories free of dust and dirt. To avoid damage to the
equipment, follow these rules:
Always dilute according the manufacturer’s instructions or use lowest
possible concentration.
Do not immerse part of the equipment in the liquid.
Do not pour liquid onto the equipment or accessories.
Do not allow liquid to enter the case.
Never use abrasive materials (such as steel wool or silver polish), or erosive
cleaners (such as acetone or acetone-based cleaners).
9-1
WARNING
Be sure to shut down the system and disconnect all power cables from the
outlets before cleaning the equipment.
No modification of this equipment is allowed.
The service personnel must be properly qualified and thoroughly familiar
with the operation of the equipment.
The safety checks or maintenance involving any disassembly of the
equipment should be performed by professional servicing personnel.
Otherwise, undue equipment failure and possible health hazards could
result.
The responsible hospital or institution shall carry out all cleaning and
disinfection procedure specified in this chapter.
Do not open the equipment housings. All servicing and future upgrades
must be carried out by the service personnel.
CAUTION
NOTE
9-2
9.1 Safety Checks
Before first use, or at least every two years, or whenever your pulse oximeter is repaired
or upgraded, a thorough inspection should be performed by qualified service personnel
to ensure the reliability.
9.2 Cleaning
Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots
of dust and sand in your place, the equipment should be cleaned more frequently. Before
cleaning the equipment, consult your hospital’s regulations for cleaning the equipment.
9-3
To clean your equipment, follow these rules:
1. Shut down the pulse oximeter and disconnect it from the power line.
2. Clean the display screen using a soft, clean cloth dampened with a glass cleaner.
3. Clean the exterior surface of the equipment using a soft cloth dampened with the
cleaner.
4. Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.
5. Dry your equipment in a ventilated, cool place.
9.3 Disinfecting
Disinfection may cause damage to the equipment and is therefore not recommended for
this pulse oximeter unless otherwise indicated in your hospital’s servicing schedule.
Clean the pulse oximeter before disinfecting it.
CAUTION
9.4 Disposal
Dispose of the pulse oximeter in accordance with local environment and waste disposal
laws and regulations. For the disposal of SpO2 sensor, follow local regulations regarding
disposal of hospital waste.
9-4
10 Accessories
WARNING
Use only accessories specified in this manual. Using other accessories may
cause damage to the pulse oximeter.
Disposable accessories are designed for single-patient use only. Reuse of
them may cause a risk of contamination and affect the measurement
accuracy
Check the accessories and their packages for any sign of damage. Do not
use them if any damage is detected.
The SpO2 sensor mentioned in this chapter meet the bio-compatibility
requirements and complies with ISO 10993-1, ISO 10993-5 and ISO
10993-10 standards.
SpO2 Sensor
Type Model Applicable patient Wavelength* PN
520A Adult (finger clip) 520A-30-64101
Pediatric (finger
520P 520P-30-64201
Single clip) 660 nm
patient use 520I Infant (toe sensor) 905 nm 520I-30-64301
Neonate (foot
520N 520N-30-64401
sensor)
512E 512E-30-90390
Adult (finger clip)
512F 660 nm 512F-30-28263
512G Pediatric (finger 905 nm 512G-30-90607
512H clip) 512H-30-79061
DS-100A Adult (finger clip) 9000-10-05161
Pediatric/Infant
OXI-P/I 660 nm 9000-10-07308
(finger clip)
Reusable 890 nm
Adult/Neonate
OXI-A/N 9000-10-07336
(finger clip)
ES-3212-9 Pediatric (ear clip) / 0010-10-12392
Adult (multi-site
sensor) 660 nm
518B 518B-30-72107
Neonate (multi-site 905 nm
sensor)
* The maximum optical output power of SpO2 sensors is less than 18 mw. Information
on wavelength range and the maximum optical output power can be especially useful to
clinicians, for example, when performing photodynamic therapy.
10-2
Other Accessories
Description PN
SpO2 Extension cable 0010-20-43075
Protective cover 0852-21-77412
Carrying case 0852-10-77701
Charger stand 0000-10-11263
PC Communication cable 0850-20-30725
Lithium ion battery(LI11S001A) M05-010004-08
Mounting clamp 0852-30-77537
10-3
FOR YOU NOTES
10-4
A Product Specifications
Physical specifications
Width × Height × Depth 56×124×30 mm
Max. weight < 300g (full configuration, including the batteries)
A-1
Charger stand
Input voltage 100 to 240 VAC, 50/60 Hz
Output voltage 5 VDC
Output current 1.2 A
Output power 6W
Alkaline batteries
Quantity 3
Specification 1.5 V, AA
Capacity 2000 mAh
Run time 36 hours with SpO2 monitored continuously, audio
indicators off and backlight brightness set to minimum
using new, full power batteries at ambient temperature
25℃.
Shutdown delay Min. 5 minutes after the low battery alarm first occurs.
A-2
Hardware specifications
Display Color TFT, 2.4”, 320×240 pixel
Power indicating lamp 1, lighting green and yellow
Loudspeaker 1; Gives audible alarm (45 to 85dB) and button tone;
Supports Pitch Tone and multi-level volume;
Alarm tones meet the requirement of IEC 60601-1-8.
Alarm indicating lamp 1, lighting red and yellow
Multifunctional connector 1, 9-pin type D connector
Power supply connector 1, used to connecting the Charger stand
Data storage
Measurement specifications
SpO2
Measurement validation: The SpO2 accuracy has been validated in human studies
against arterial blood sample reference measured with a CO-oximeter. Pulse oximeter
measurements are statistically distributed, and only about two-thirds of the
measurements can be expected to fall within the specified accuracy compared to
CO-oximeter measurements.
Range 0 to 100%
Resolution 1%
A-3
Accuracy 70 to 100%: ±2% (measured without motion in
adult/pediatric mode)
70 to 100%: ±3% (measured without motion in neonatal
mode)
70 to 100%: ±3% (measured with motion)
0% to 69%: Unspecified
Refreshing rate 1s
Averaging time 7 s (When the sensitivity is set to High)
9 s (When the sensitivity is set to Med)
11 s (When the sensitivity is set to Low)
PR
Range 18 to 300 bpm
Resolution 1 bpm
Accuracy ±3 bpm (measured without motion)
±5 bpm (measured with motion)
Refreshing rate 1s
Averaging time 7 s (When the sensitivity is set to High)
9 s (When the sensitivity is set to Med)
11 s (When the sensitivity is set to Low)
A-4
B EMC
The device meets the requirements of IEC 60601-1-2: 2014.
WARNING
B-1
special environment, such as magnetic resonance imaging environment,
the equipment/system may be disrupted by the operation of nearby
equipment.
B-2
Voltage Complies connected to the public low-voltage
fluctuations power supply network that supplies
NOTE
Other devices may affect this device even though they meet the
requirements of CISPR.
B-3
If the device is operated within the electromagnetic environment listed in Table
Guidance and Declaration —Electromagnetic Immunity, the system will remain safe
and provide the following essential performance:
■ Operating mode
■ Accuracy
■ Function
■ Accessories identification
■ Data stored
■ Alarm
■ Detect for connection
B-4
transient/burst power supply power supply should be that of a
IEC 61000-4-4 lines lines typical commercial
±1 kV for ±1 kV for or hospital
input/output input/output environment.
lines lines
(length (length greater
greater than 3 than 3 m)
m)
B-5
battery.
Note: UT is the A.C. mains voltage prior to application of the test level.
IEC
Immun 60601 Complia Electromagnetic
ity test Test level nce level environment - guidance
Conduct 3 Vrms 3 Vrms Portable and mobile RF
ed 150 kHz to communications equipment
disturba 80 MHz should be used no closer to any
nces part of the device, including
6 Vrms 6 Vrms
induced cables, than the recommended
by RF in ISM separation distance calculated
fields bands and from the equation applicable to
B-6
IEC610 amateur the frequency of the transmitter.
00-4-6 radio Recommended separation
a
bands distance:
between 3 .5
d = P 150k to 80
0,15 MHz V
and 80 MHz
MHz
B-7
0-4-3 9 V/m 9 V/m Interference may occur in the
704–787
vicinity of equipment marked
MHz,
5100–5800 with the following symbol:
MHz
.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
a
The ISM (industrial, scientific, and medical) bands between 150 kHz and 80
MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957
MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The
amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0
MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1
MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0
MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and
50,0 MHz to 54,0 MHz.
b
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the device.
c
Over the frequency ranges 150 kHz to 80 MHz, field strengths should be
B-8
less than 3V/m.
100 12 3.5 7
B-9
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
B-10
C Factory Defaults
This section lists the most important factory default settings. These settings are not
user-adjustable. However, you can restore the factory default settings if necessary.
In the tables below, column “A” indicates whether this item is affected by factory
configuration or user configuration.
“√” means “Yes”: The user configuration will be loaded when the pulse
oximeter is restarted; if the user configuration is not available, the factory
configuration will be restored.
“×” means “No”: The changed settings will be saved when the pulse oximeter
is shut off and will be loaded when the pulse oximeter is restarted.
C-1
C.2 System Setup
C-2
D Alarm Messages
This section lists only the most important physiological and technical alarm messages.
Some messages appearing on your pulse oximeter may not be included.
In the tables below, column “L” indicates the default alarm level: “H” means high, “M”
means medium and “L” means low; “*” means the alarm level is user-adjustable.
The “Cause and actions” column gives recommendations to instruct you to troubleshoot
the problems. If the problem persists, contact your service personnel.
D-1
D.2 Technical Alarm Messages
Alarm messages L Cause and action
SpO2 Self Test Err L The pulse oximeter failed the power-on self-test. Restart
the pulse oximeter.
SpO2 Comm Err L The pulse oximeter encountered a communication error.
Restart the pulse oximeter.
SpO2 Comm Stop L An error occurred to the SpO2 module or there was a
communication problem. Restart the pulse oximeter.
SpO2 Overrange L Measured SpO2 value is beyond the specified
measurement range. Check the patient’s condition.
PR Overrange L Measured PR value is beyond the specified measurement
range. Check the patient’s condition.
Sensor Off L* The SpO2 sensor detached the patient or the pulse
No Sensor L oximeter, or there was a fault with the SpO2 sensor, or an
unspecified SpO2 sensor was used. Check that the sensor
application site and the sensor type are correct, and make
sure that the sensor is undamaged. Reconnect the sensor if
the sensor is disconnected or use a new sensor if the
sensor is damaged.
SpO2 Low Perf L The signal detected is weak or the signal quality is poor.
Check the patient’s condition. Change the sensor
application site. If the problem persists, replace the sensor.
Voltage Too High L The system power supply fails. Restart the pulse oximeter.
Voltage Too Low L
Battery Too Low M The battery power is low. Replace the batteries if alkaline
batteries are used or charge the battery if a lithium ion
battery is used.
Battery Error L A problem occurs when the lithium ion battery is being
charged. Check the battery for damage. If yes, replace the
battery.
Power Comm Err M Communication problem occurred to the power supply
part during the power-on self-test or operation. Restart the
pulse oximeter.
Clock Need Set L The real-time clock is reset. Set the clock.
D-2
E Symbols and Abbreviations
E.1 Units
A ampere
bpm beats per minute
ºC centigrade
g gram
kHz kilohertz
MHz megahertz
GHz Gigahertz
h hour
Hz hertz
k kilo
kg kilogram
kPa kilopascal
m meter, minute
M mega
min minute
mm millimeters
ms millisecond
mW milliwatt
s second
nm nanometer
ppm part per million
V volt
µA microampere
E-1
E.2 Symbols
– minus
– negative
% percent
/ per; divide; or
+ plus
= equal to
< less than
> greater than
≤ less than or equal to
≥ greater than or equal to
± plus or minus
× multiply
© copyright
E-2
E.3 Abbreviations
E-3
FOR YOUR NOTES
E-4
P/N:0852-20-77462(8.0)