Optix 2.2 Guide
Optix 2.2 Guide
Gencoa Ltd
4 De Havilland Drive
Liverpool
L24 8RN
United Kingdom
Tel: +44 (0)151 486 4466
Fax: +44 (0)151 486 4488
Email: [email protected]
www.gencoa.com
1 www.gencoa.com/optix
Optix Sensor Manual
Table of Contents
1. Product introduction ................................................................ 4
4. Operation ...............................................................................12
5. Maintenance ..........................................................................15
...................................................................................................... 16
6. Troubleshooting .....................................................................18
7. Specifications.........................................................................20
2 www.gencoa.com/optix
Dimensions and physical data ....................................................... 20
3 www.gencoa.com/optix
1. Product introduction
Optix is a multi-purpose instrument for gas sensing in any vacuum environment, functioning
through a wide range of operating pressures to cater for most industrial vacuum production
processes without any requirement for differential pumping.
2. Safety information
Never run the power cables through a pinch point such as a door or panel.
Never use nails, staples or similar to fasten the power cables in place.
Always ensure that the mains power plug is fully inserted into the adapter socket to
ensure a secure electrical connection.
Exercise caution plugging the Optix mains adapter into multiple socket connectors. An
overload on one socket could cause a fire or electric shock possibly resulting in
serious injury.
Never plug the mains adapter into a power source that does not correspond to both
the voltage and frequency specified on the regulatory label on the mains adapter.
4 www.gencoa.com/optix
Never attempt to connect or disconnect the mains power cable or plug with damp or
wet hands.
Never allow any liquids to spill inside your Optix sensor or PSU. Exposure to liquid or
moisture can cause fire or electric shock possibly leading to serious injury. Should
water contamination occur you should immediately carry out the following actions: -
o Turn off the Optix using the on/off switch located on the front fascia of the PSU.
o Disconnect the power cable from both the source and the PSU.
Never locate the Optix PSU in an area where it will be exposed to excess heat (30°C
Maximum) such as in direct sunlight as this may result in a system failure or
malfunction.
Never keep the Optix PSU or sensor in an area where it will be exposed to humid
conditions as this may result in a system failure, malfunction or loss of data.
The High Voltage (HV) cable that is connected between the PSU and the Optix can deliver
voltages of up to 3kV. Care must be taken when connecting and disconnecting this cable.
When connecting the HV cable to the sensor, ensure that the PSU on/off switch is off
and that the PSU DC input power plug is disconnected.
When disconnecting the HV cable to the sensor, ensure that the PSU on/off switch is
off and that the PSU DC input power plug is disconnected.
Never attempt to turn the PSU on when the HV cable is not connected at both ends
5 www.gencoa.com/optix
3. Getting started
6 www.gencoa.com/optix
3.2. System description
The Optix gas sensor consists of a plasma generator, a sensing unit and a power
supply (PSU).
1. On/off switch – Turns the power to the PSU and the Optix sensor on or off
▪ Up position = ON
2. Power LED – Solid light when power is being supplied to the PSU and Optix
sensor. A flashing light indicates that the PSU is ready for operation.
4. HV LED – Lit indicates that the PSU high voltage output is engaged.
5. Plasma LED – Lit indicates that a plasma has been struck inside the plasma
generator.
7 www.gencoa.com/optix
PSU Rear fascia
3. DC in – Connect the DC power adapter here. The Optix is supplied with a 24V
DC power adapter. If you wish to supply your own source of DC power it must
be in the range 18V to 28V and capable of supplying at least 750mA.
4. COMMS – Connect the sensor communication cable here. Ensure that the cable
is firmly locked in the socket.
8 www.gencoa.com/optix
Optix sensor and plasma generator
1
3
2 5
2. High voltage input – Connect the high voltage cable from the PSU here
3. User – The User port has 4 digital outputs that can be assigned to become
active when a species measurement exceeds a user defined threshold. When
the trigger point is reached the specified output will relay Vcc.
The User port also contains a 0V – 10V analogue output that can be assigned to
a user defined gas species (see page 38 for details on how to configure the
output)
9 www.gencoa.com/optix
Pin 5 – Out 4 (Vcc = Device ready)
Pin 6 – Analogue out (0V – 10V)
Pin 7 – Reserved
Pin 8 – Ground Analogue output
Pin 9 – Ground Vcc
5. RS232 – Allows for serial commands to be sent to the Optix sensor and for data
to be received.
Pin 1 – NC
Pin 2 – Tx
Pin 3 – Rx
Pin 4 – NC
Pin 5 – 0V
Pin 6 – NC
Pin 7 – NC
Pin 8 – NC
Pin 9 – NC
3.3. Installation
1. Install the Optix sensor on a KF25 flange using the supplied meshed O-ring. The
Optix sensor can be mounted in any orientation.
2. Connect the High Voltage cable from the PSU High Voltage output to the Optix
sensor High Voltage input
10 www.gencoa.com/optix
3. Connect the COMMS cable between the PSU and the Optix sensor
5. Turn the PSU and sensor on via the ON/OFF switch on the PSU front facia.
Note: The meshed KF25 O-ring is required to eliminate any interaction between
plasmas in the monitored vacuum chamber and the plasma generated by the
Optix sensor.
11 www.gencoa.com/optix
4. Operation
The Optix sensor uses the principal of remote plasma emission monitoring to identify
and monitor gaseous species within a vacuum. The principal of operation is described
in the following schematic:
Note: The sensor is only able to detect gaseous species. Species, such as
metallic vapors, will condense before reaching the plasma generator.
12 www.gencoa.com/optix
B) The light generated by the plasma is collimated and fed into an optical
spectrometer. The spectrometer resolves the intensity of the light at discrete
wavelengths at approximately 0.3nm intervals between 200nm and 900nm.
C) Gaseous species emit light at defined wavelengths – for example H emits light
at 656nm, 486nm and 410nm. Therefore, by monitoring the intensity of light at
specific wavelength the presence and changing concentrations of gases can be
determined.
The Optix has two modes of operation: OES and RGA mode.
In OES mode the readings are relative to each other, or in other words, ratios rather
than absolute readings. Use this mode if ratios of gases are a better measurement or if
using the Optix to analyse an external plasma source.
In RGA mode the Optix calculates partial pressures of the selected gases.
The Optix sensor is designed to operate from a total pressure of 1x10-7 mbar to
0.5 mbar. At pressures below 1x10-7 there will still be a plasma generated, however the
light intensity will be too low to be usable for most applications. Above 0.5 mbar a
plasma can be sustained, however the proportional relationship between partial
pressure and light intensity is no longer maintained. Additionally, contamination of the
plasma generator can occur.
13 www.gencoa.com/optix
The Optix is interlocked via its integrated pirani pressure gauge. By default the Optix is
set to enable the plasma below 0.5 mbar and disable the plasma above 0.5 mbar. The
pressure for the plasma on/off switch can be adjusted via the device settings. See page
29 (Starting Pressure) for details on how this can be adjusted.
14 www.gencoa.com/optix
5. Maintenance
Internal description
1. Anode
2. Cathode
3. Window protector
4. Sapphire window
Periodic cleaning of the plasma generator’s electrodes may be required to maintain stable
operation. The frequency of cleaning will depend on the pressures and composition of gases
in the vacuum. Using the sensor with condensing species will require more frequent cleaning.
1 2
15 www.gencoa.com/optix
2. Remove the plasma generator cover and power feedthrough assembly
3 4
4. Clean the anode (above the striking collar) to a reflective finish by using an abrasive
such as Scotch-brite combined with isopropyl alcohol. Be careful not to bend the
anode or damage the striking collar.
2. Remove the magnetic array then remove the 6 clamp bolts and the 2 half-clamps. The
internal surface of the cathode can then be cleaned with an abrasive such as Scotch-
brite combined with isopropyl alcohol.
16 www.gencoa.com/optix
Cleaning the sapphire window
Prolonged use of the sensor with condensing gases may cause a coating buildup to
occur on the sapphire window. Over time this may result in a degradation of the signal
intensity. The sapphire window may be cleaned or replaced by the following steps:
1 2
1. Unscrew 2 bolts connecting the bracket between the plasma generator and the
sensor case. Note: Do not unscrew 3 bolts on the sensor case as this may lead
to loss of calibration.
Carefully separate the plasma generator from the sensor case. Do not
excessively bend the two components whilst performing this action.
2. Remove the 4 screws from the locator part to expose the sapphire window.
Once removed, any coating on the window can be removed with fine abrasive
such as 1500 grit (or greater) emery paper.
17 www.gencoa.com/optix
6. Troubleshooting
Cause / Action: If the plasma LED on the PSU is lit this would indicate that a short circuit has
developed between the plasma generator anode and cathode. In this case, check that no
debris is present inside the plasma generator and contact Gencoa.
Cause / Action: If the total pressure is below 1x10-6 mbar the plasma may not strike. It may
takes upwards of 1 minute for the plasma to strike at lower pressures. The striking time will
increase if the sensor becomes contaminated. See section xx for instructions on how to clean
the plasma generator.
Cause / Action: This is caused by debris entering the plasma generator and adhering to the
anode. This in turn creates a charge concentration point that can result in a glowing spot that
will interfere with senor readings. The anode can be cleaned by gently wiping with a clean, dry
cloth to remove any debris. Ensuring that debris does not enter the plasma generator during
venting cycles will prevent this from occurring.
Cause / Action: The USB connection is susceptible to interference from high frequency noise,
such as that emitted by high frequency power supplies. If connection problems when high
frequency power supplies are used in the vicinity of the sensor then consider switching to the
Ethernet connection. The Ethernet connect is far less susceptible to high frequency noise.
18 www.gencoa.com/optix
Problem: A dead pixel (a single wavelength peak) is showing on the spectrum
Cause / Action: This is can be removed by taking a reference of the dark spectrum (i.e.
with plasma off) and selecting “use reference spectrum” in the sensor options.
Spectrum overlap
A single species can emit light at multiple wavelengths, therefore there is the possibility
that two species can have overlapping emission peaks. This is a similar effect observed
when using mass spectrometers where two species share the same mass. The
wavelengths used in the Optix default species selection library have been chosen to not
overlap with other species. However, care must be taken if selecting custom
wavelengths to account for this effect.
19 www.gencoa.com/optix
7. Specifications
Electrical
20 www.gencoa.com/optix
Operating data
Communication interfaces
Software
21 www.gencoa.com/optix
Part 2 – Process Monitoring Software
1 2 3 4 5 6 7
22 www.gencoa.com/optix
1.1 Add Optix sensor
An Optix unit can be connected to the Process Monitoring Software using an ethernet cable
or a USB cable.
Troubleshooting
• The computer running the Process Monitoring Software needs to be on the same
subnet as the Optix sensor i.e. the computer must share the first 3 entries of the
IP address with the Optix sensor.
• The Optix sensor cannot share an IP address with another device on the same
network. The Optix IP address can be changed in the Device Settings screen.
23 www.gencoa.com/optix
1.1.2 USB Connection
• Select the COM port from the dropdown list and press OK.
Note: If no COM port is shown in the dropdown list and you are using either Windows
7 or 8 then you may need to install the required USB drivers. To do this, locate the
Optix install directory (the default directory is C:\Program Files (x86)\Gencoa Ltd\OptiX
Software) and run the program serial_install
Toggles between displaying and hiding the list of connected Optix devices.
24 www.gencoa.com/optix
1.3 General settings
25 www.gencoa.com/optix
• Layout settings allows for customisation
of the Process Monitoring Software.
26 www.gencoa.com/optix
1.4 Start data saving
• When data saving is enabled all species contained within the species list will be saved
at the time interval specified in the setting dialog box.
• Ends the recording of data, and generates a .csv file in the data saving
directory.
27 www.gencoa.com/optix
2. Software Overview – Device tab
1 2 3 4 5 6 7 8 9 10 11 12
16 17
13 14
15
18
28 www.gencoa.com/optix
2.2 Disconnect
Disconnects from the Optix sensor. The disconnected device will still show in
the device list.
• Starting Pressure is the pressure at which the Optix plasma will automatically turn
on and off at if “Start Automatically” is selected in the Power Supply Settings window.
0.5 mbar is the default value.
29 www.gencoa.com/optix
2.5 Turn PSU plasma on/off
Turn the power supply high voltage output on or off in order to strike the
sensor’s plasma. A plasma will ignite under the conditions of the sensor’s
operating range.
The icon will turn yellow when the high voltage is on.
If the “Start Automatically” option is enabled (2.6) then the status will be
displayed by the colour however you cannot manually turn the plasma on or off.
Start Automatically
SetPoint
• Sets the current setpoint (in µA) of the plasma generator power supply. Current
setpoints of 50, 100, 200 and 300 are selectable. The default current setpoint is 200
µA.
Degas
• The degas function increases the power supply current setpoint to 1000µA in order to
accelerate ion induced desorption of species within the sensor. This is particularly
effective at removing water vapour from within the sensor
30 www.gencoa.com/optix
2.7 View Spectrum plot
• The spectrum view displays the light spectrum as read by the Optix spectrometer.
There are a number of features that can be enabled to enhance the interpretation of
the spectrum.
• A reference spectrum can acquired at any time by using the Refresh Reference
button. Using the Show Reference button will display the reference spectrum
alongside the current “live” spectrum.
• The reference spectrum can be subtracted from the live spectrum by enabling the
Use Reference Spectrum option from within the sensor settings menu. This can be
particularly useful when a “baseline” condition has been established, where this
feature is used to “zero” the sensor readings.
• Enabling Auto Scaling will automatically resize the spectrum to keep all peaks in
view. Note: Disabling Auto Scaling will allow manipulation of the spectrum view using
the same methods as used for the trend view.
31 www.gencoa.com/optix
Adding and saving spectra
• The spectrum can be saved by opening the view spectra icon and right clicking
on the active spectrum to save to a .csv file.
• Spectra can also be saved to a csv via the icon or a screenshot via
• Saved spectra can be added to the spectrum view by opening the add spectrum icon
Auto ID
The Auto ID feature uses a peak detection algorithm combined with a reference library
of emission lines to identify spectrum peaks.
The emission library has been constructed based on experimental trials of the Optix
sensor, combined with referenced emission data. The library is tailored to avoid
displaying multiple identifications for a single peak, this in order to reduce the
probability of misidentification. Therefore, some peaks may not be identified if they
often overlap with multiple species.
32 www.gencoa.com/optix
Auto ID settings
• Hit interval Sets the distance either side of peak over which the algorithm will
search to find an entry in the ID database. Setting this value too high will result in
multiple identifications for a single peak. Setting too low may result in peaks not being
identified.
33 www.gencoa.com/optix
2.8 Sensor settings
No of Scan Sets the number of scans averaged for each updated data point. Increasing
the number of scans will improve signal to noise ratio at the expense of update speed.
Use Reference Spectrum Select whether to subtract the most recent reference spectrum
from the active spectrum received from the spectrometer. This will “zero” the spectrum
reading.
Use Electrical Dark Selects whether to use correction for thermal effects on the
spectrometer
Use non Linearity Coefficient Selects whether to enable linearity correction for light
intensity
Refresh Reference Captures the current active spectrum as a new reference spectrum.
This can either be recalled and overlaid or subtracted from the active spectrum.
Spectrum Range Selects the spectral range to be acquired from the spectrometer.
Shift Pixel Applies a global offset to the spectrum by the specified number of pixels. This
can be used to adjust the wavelength calibration of the spectrometer.
34 www.gencoa.com/optix
Use Correction Selects whether to apply the irradiance calibration to the spectrum
The species settings screen assigns up to 20 species to be monitored by the Optix sensor.
These settings are then stored and processed on the Optix sensor.
Enable Double click on the “tick” to decide whether the species will appear in the
monitoring charts. Note: this setting affects which species are saved when data saving is
enabled.
Coefficient Applies a coefficient to the species reading. Can be useful for bringing species
with different signals levels to be visible at the same level on the monitoring plots.
Min Sets the minimum trigger point for the species (see Trigger Settings)
Max Sets the maximum trigger point for the species (see Trigger Settings)
35 www.gencoa.com/optix
# Sample Sets the number of samples to be taken for each species readings. Increasing
the number of samples will improve the signal to noise ratio for that species, at the expense
of update speed.
Unlike the Number of Scans setting, which is applied to the whole spectrum, the #
Sample setting is per species. Therefore, individual species can be configured for
optimal signal to noise ratio vs speed of update.
The RGA Mode of the Optix adds some additional options to the species setup screen:
Gas When enabled the species is used as part of the partial pressure calculation.
Make sure that “Gas” is enabled for all significant species present in the vacuum
system for most accurate partial pressure measurements.
36 www.gencoa.com/optix
Description
Selection defines the species by a central wavelength and a specified bandwidth in pixels
(each pixel is approximately 0.3 nm).
Selection Modes:
Sum Sums the value of the selected wavelength plus the values of the
wavelengths inside the bandwidth
Average Averages the value of the selected wavelength and those inside
the specified bandwidth
Tracking Identifies and uses the highest value within the selected
wavelength and bandwidth range.
Large shifts in temperature around the sensor can cause the location of
peaks to move by up to 3 pixels. Using the Tracking mode will automatically
correct for wavelength shifts of peaks within the specified bandwidth.
Combination defines the species by a combination of two other species – e.g. a ratio
Constant defines the species by a constant number – useful for displaying a reference level
on the trend plot.
37 www.gencoa.com/optix
Bandwidth Sets the number of pixels either side of the selected peak to be either
averaged, summed or tracked, depending on the Mode selection.
The trigger settings determine when the relay outputs on the sensor’s User port become
activated. There are 3 triggers that are user configurable, the 4th trigger on the User port is
always assigned to the sensor operational status.
Persistence Sets whether the trigger remains active after the species crosses the trigger
threshold.
38 www.gencoa.com/optix
Time On Sets the time that the trigger remains active after the species crosses the
trigger threshold. This setting is only applicable if persistence is turned off.
Duration Sets the time that the species must remain inside the trigger condition for the
trigger to become active. This can be used to prevent noise from activating the trigger.
Species Selected Selects the species number to be used for the trigger
Select which species reading is sent to the Optix analogue output. The
output will be scaled to 0-10V between the minimum and maximum
readings.
The species table summarises the selected wavelength, the assigned colours, the current
value of the reading (in % of the maximum spectrometer reading) and the minimum and
maximum trigger points.
39 www.gencoa.com/optix
2.14 Species donut
The species doughnut displays the ratio of intensities of the currently enabled species.
Species that are below their minimum trigger point are omitted from the
doughnut.
Displays each species proximity to the selected maximum and minimum values
40 www.gencoa.com/optix
The species bar chart displays enabled species and their proximity to their maximum and
minimum trigger points. The name of the species will turn red when maximum or minimum
trigger points are reached.
The current total pressure reading derived from the plasma generator electrode voltage,
current and integrated pirani gauge.
Each dot represents the status of the 3 user definable triggers (Filled blue is an
activated trigger)
The trend view displays the values of enabled species with time. Historic trends of species
that are not currently enabled will be displayed after they are re-enabled.
Pausing Pausing the update of the plot is performed by moving the scroll bar below the
plots x axis or by clicking the pause icon in the lower right.Un-pausing the plot is performed by
pressing the play button in the centre of the plot.
The trend view will auto-scale so that the species with the greatest reading is always visible.
When the plot is “live” (i.e. not paused) bounds will be placed on the ability to zoom so as to
keep all trends within view. Pausing the plot will enable complete control of the plot
manipulation.
Note: Pausing the plot does not mean losing data. Upon un-pausing the plot any data
taken during the pause period will be updated on the plot.
41 www.gencoa.com/optix
Zooming Zooming in on an area of the plot is performed by dragging a box over the
desired area from left to right. To zoom out to the default scale drag a box anywhere on the
plot from right to left.
Moving through time The plot can moved backwards through time by dragging the
scroll bar below the x axis of the plot
Rescaling the x and y axis The x and y axis can be rescaled by clicking (and
holding) the handles at the top or bottom of the axis – then dragging to re-scale.
Note: When in pause mode, full control is granted over the scaling. When un-paused
the plot will always rescale to fit the largest and smallest species values.
Free plot movement The plot can be freely moved in both the x and y axis by
holding the middle mouse button on any location on the plot and moving the mouse in the
desired direction. Note: This is only available in pause mode.
42 www.gencoa.com/optix
Part 3 – RS232 Communications
The Optix acts as a slave device when using RS232 communications, i.e. it never initiates the
transmission. The master device transmits to the Optix which executes the communication
and replies to the master device.
Get values (V), parameters (P) or information (I) with the G command
43 www.gencoa.com/optix
1 / Device
G1P
G1I
44 www.gencoa.com/optix
G1V
2 / PSU
G2P
Command codes
45 www.gencoa.com/optix
Status codes
G2I
G2V
46 www.gencoa.com/optix
3 / Sensor
G3P
47 www.gencoa.com/optix
G3I
5 / Species
G5P”species index”
48 www.gencoa.com/optix
G5V”species index”
6 / Triggers
G6P”trigger index”
Additional commands:
G6V”trigger index”
49 www.gencoa.com/optix
G6V3
7 / Analogue output
G7P
G7V
50 www.gencoa.com/optix
Part 4 – OPC server
OPC clients connect to the server using Microsoft COM or, if located on a
different host, Distributed COM (DCOM). The Optix OPC Server and the
controller communicate using a protocol which is transparent to OPC clients.
This protocol is proprietary to Gencoa Ltd.
The Optix OPC Server runs as a Windows form. This form can however be
made invisible which means that it will not appear on the task bar.
51 www.gencoa.com/optix
OPC Data Access Interface
The OPC Data Access interface allows multiple Data Access clients to access
any Optix controller tag.
The Optix OPC Server conforms to version 3.00 of the Data Access Custom
Interface Standard published by the OPC Foundation.
Software Installation
This chapter describes the installation process for the OPC Portal Server.
Note: You cannot install the Optix OPC Server as an upgrade from a previously
installed version. You first need to remove the old version before starting a new
installation.
In This Chapter
52 www.gencoa.com/optix
The installation process requires that you complete a number of tasks in the
following order:
1) Make sure you have a Windows administrator account with the privilege.
2) Install the OPC Core Components Redistributable.
3) Install the Optix OPC Server.
4) Set up Distributed COM (DCOM) if required.
Note: Before you begin this task, you must have a Windows administrator
account.
Note: Before you begin this task, you must have a Windows administrator
account and you must have installed the OPC Core Components
Redistributable.
53 www.gencoa.com/optix
3) Click to install the ‘Optix OPC Server Setup.msi’.
► The Optix OPC Server Installer starts.
4) Click Next.
5) At the Select Installation Folder step
Do the following:
a. Accept the default folder for the installation.
b. Select the radio button for Everyone.
c. Click Next.
6) At Confirm Installation, click Next to start the installation.
7) When Installation Complete appears, click Close.
► You have now installed the Optix OPC Server
If you are implementing an HMI interface and developing your client (i.e. using
Cimplicity), it is possible that the user launching the program (i.e. ‘System User’)
might be different than the one starting the server (i.e. User logged on the
computer). If both user do not have the same ‘Rights’ it might limit the number of
functions/parameters that you can access.
To configure the ‘default user account’ that will be running the Optix OPC
Server, do the following:
54 www.gencoa.com/optix
► Expand the My Computer item and select DCOM Config. Locate the
Optix OPC Server in the list.
12) Right-click to open the properties for the server, select the Identity tab.
55 www.gencoa.com/optix
► Set the user account to The launching user.
56 www.gencoa.com/optix
Server Interface
In order to communicate with the Server through an OPC client, you first need to
register the server into the windows registry. After you clicked on this button (and if you
acknowledge the popup window) it will call the ‘regserver’ command and shutdown the
application. You can restart the Server from the Start Menu or connect to it from your
client application (the Server will automatically open when the first client connect to it).
If you wish to uninstall the Server from the computer, it is cleaner to first unregister it.
This will remove the server entry from the Windows registry. After you acknowledge the
popup window the application will shut down. You can completely remove the Optix
Server from the program options in the control panel.
57 www.gencoa.com/optix
Prog ID and CLSID
OPC server names are DCOM ProgIDs, human readable names for components that
have numeric names (CLSID's) in Windows. The CLSID is what is used by the OPC
Client when it actually tries to connect to the OPC Server. In the case of this Optix
Server, resolving the OPC Server name (ProgID) Optix.OPCServer.1 will return a
CLSID of {83D2074F-D746-4138-BF4D-EA988D182011}.
Connected Clients
This will displayed the number of OPC clients currently connected to the Server.
Interface Visibility
Most servers run as services and it is not straight forward to see if they are running
and/or their status. To help with the design/test of the client application the above
window is shown. When you finish your testing or if you do not wish for this window to
be displayed you can make it invisible.
Note: There is also an OPC tag to make this window visible or invisible.
Log Area
Any error code returned by the server that is not related to a COM/DCOM error will be
appended in this text area together with a short description.
The COM/DCOM errors are shown in the result section of the OPC item/tag.
Connection Status
Green: Connected
Red: Not connected
58 www.gencoa.com/optix
Close the Server
You can use this button to properly shutdown the server. This will notify the OPC
Clients connected to the Server that you want to close it, giving each one enough time
to properly disconnect.
59 www.gencoa.com/optix
Tag Description
Note: Tags within the Device group only update when the initial
connection is made to the Optix. Other tags will update at the specified
UpdateTime and only if their value has changed.
IP R/W String IP address of the Optix. Must be set to match the unit to be
connected.
Connect R/W Bool Connection status between the OPC server and the Optix. Set this
to true in order to connect to the Optix for the first time.
AutoConnect R/W Bool Enable or disable auto connection to the Optix. Connect must be first
set to True
UpdateTime R/W Integer Update time in ms for communication between the server and the
Optix.. 250 ms minimum value
60 www.gencoa.com/optix
PressureCoefficient R Real Pressure correction coefficent
Status R/W Bool High voltage status of the PSU. True = on, False = off. Use this to
turn the plasma on or off
Pulse R/W Integer Activates the pulse mode if a pulse power supply is present
AutoStart R/W Bool Activates the plasma autostart (will over-ride status writes)
UseReference R/W Bool If True will subtract the reference spectrum from the active spectrum
61 www.gencoa.com/optix
UseCorrection R/W Bool Use irradiance calibration
ResetReference R/W Integer Writing any value will reset the reference spectrum
NoOfSample R/W Integer Number of samples to be averaged for the species reading
UseforPP R/W Bool Selects if to use this species for the partial pressure calculation
Value R Real The species reading in % of full scale or partial pressure (in RGA
mode)
Status R Integer 1 = Below Min, 2 = Between Min and Max, 3 = Above Max
62 www.gencoa.com/optix
SpeciesIndex R/W Integer Index of the species to be used for the trigger
Condition R/W Integer 0 = Below Min or Above Max, 1 = Below Min, 2 = Between Min and
Max, 3 = Above Max
Status R/W Bool Active status of the trigger. Trigger can be cleared by writing 0
Min R/W Real Min value for the output scaling in % or partial pressure depending
on the sensor mode
Max R/W Real Max value for the output scaling in % or partial pressure depending
on the sensor mode
Return Codes
Note, the last error code will persist until cleared by a “0” write
Value Description
63 www.gencoa.com/optix