3、AX-900,AX-900A User Manual
3、AX-900,AX-900A User Manual
Edition: B00
No.:046-001290-03
Date:03/2021
Shenzhen Comen Medical Instruments Co., Ltd.
Product Information
Product Name: Anesthesia Machine
Models: AX-900, AX-900A
Software Version: V5
CE information:
Address: Floor 10, Floor 11 and Section C of Floor 12 of Building 1A & Floor 1 to Floor 5 of Building
2, FIYTA Timepiece Building, Nanhuan Avenue, Matian Sub-district, Guangming District, Shenzhen,
Guangdong, 518106, P.R. China
Statement
Shenzhen Comen Medical Instruments Co., Ltd. (hereinafter referred to as Comen or Comen Company)
owns the copyright of this User’s Manual (non-public publication), and has a right to process it as
restricted materials. This User’s Manual may serve as references for operation, maintenance and repair
of Comen products. Anybody else has no right to show to people the contents of the User’s Manual.
The User’s Manual contains exclusive data under the protection of law of copyright. All rights reserved.
Any individual or organization must not reproduce, amend or translate any part of the User’s Manual
without prior written approval from Comen Company.
Edition number of the User’s Manual is subject to upgrading without prior notice due to any changes in
software, technical specification or other causes.
The User’s Manual is only applicable to the anesthesia machines AX-900 and AX-900A manufactured
by the Comen Company.
Warranty
Comen Company will be responsible for the safety, reliability and performance of its products on the
conditions that the following restrictions are complied with:
The products are operated as per the User’s Manual;
The products are installed, maintained and upgraded by personnel approved or authorized by
Comen company.
Storage environment, working environment and electrical environment of the products conform to
the product specifications;
The product serial number label or indication of manufacture is clear and legible so that Comen
I
may identify that the products are authentically manufactured by Comen Company;
The damages are not caused by human factors (such as dropout by accident, deliberate damage
etc.);
Comen Company will provide free services for all products satisfying the Comen's warranty conditions.
Comen may charge service fees for any services not within the product's warranty. User has to bear all
transportation cost (including custom's duties) for sending the products to Comen.
Caution
This equipment must not be operated at home.
Warning
It is not a medical treatment device.
After-sales Service
Provider: After-sales Service Department, Shenzhen Comen Medical Instruments Co., Ltd.
Address:Floor 10 of Building 1A , FIYTA Timepiece Building, Nanhuan Avenue, Matian Sub-district,
Guangming District, Shenzhen, 518106, Guangdong, China
Tel: +86 755 26431236
Fax: +86 755 26431232
Service hotline: 400-700-9488
ZIP: 518106
II
Preface
This User‟s Manual describes the performance, operating procedures and other safety messages of
anesthesia machines AX-900 and AX-900A manufactured by the Company. The manual offers the best
starting point for a new user to begin operating the anesthesia machines.
Scope of Application
The product is applicable to the inhaling anesthesia and respiration management in adult and pediatric
patients during the surgical operation.
Contraindication
Pneumothorax and severe pulmonary insufficiency are prohibited.
Figure
All figures in this User‟s Manual are given for reference only. The menus, setup and parameters given
in the figures may be not completely consistent with those that you see on the anesthesia machine.
III
Blank Page
IV
Contents
Chapter 1 User Responsibility ............................................................................................. 1-1
V
2.5.15 Breathing system hook........................................................................................... 2-37
2.5.16 Battery.................................................................................................................... 2-37
2.5.17 Serial Port .............................................................................................................. 2-37
2.5.18 USB Port ................................................................................................................ 2-37
2.5.19 Network Port .......................................................................................................... 2-38
2.5.20 VGA interface ........................................................................................................ 2-38
2.5.21 Equal-potential Grounding..................................................................................... 2-38
IX
5.10 AGSS Transfer and Receiving System ................................................................................. 5-40
5.10.1 Structur Composition of AGSS .............................................................................. 5-41
5.10.2 Assemble the AGSS ............................................................................................... 5-41
5.10.3 Waste Gas Disposal System ................................................................................... 5-43
XII
10.1 General Introduction ............................................................................................................. 10-1
10.2 BIS module ........................................................................................................................... 10-1
10.3 BIS display ........................................................................................................................... 10-2
10.4 Safety Information ................................................................................................................ 10-4
10.5 BIS Connection..................................................................................................................... 10-4
10.6 Electrode impedance test results window ............................................................................. 10-5
10.7 Setting BIS ............................................................................................................................ 10-6
10.7.1 Set BIS smoothness................................................................................................ 10-6
10.7.2 Sensor replacement confirmation .......................................................................... 10-6
10.7.3 Set filter switch ...................................................................................................... 10-7
10.7.4 Set wave shift ......................................................................................................... 10-7
10.7.5 Set wave speed ....................................................................................................... 10-7
XV
15.14.1 External vacuum suction system specifications ................................................... 15-40
15.14.2 Internal Vacuum Suction System Specifications.................................................. 15-40
15.15 Alarm Specifications .......................................................................................................... 15-41
15.15.1 The Sound Pressure Alarm................................................................................... 15-41
15.15.2 Air Source Pressure Alarm................................................................................... 15-41
15.16 AnestheticVaporizer Specifications .................................................................................... 15-41
17.1 Instructions for Minimizing Environmental Impact during Normal Use ............................. 17-1
17.2 Information for End of Life Management ............................................................................ 17-2
XVI
Chapter 1 User Responsibility
1.1 Declaration
The product shall be assembled, operated, maintained and repaired as per the User‟s Manual. The
product must be regularly checked. In case that the product requires replacement and maintenance due to
fault, unserviceability, damage/loss/wearing/deformation or contamination of components, contact
immediately local customer service center or agent of Comen for help. The product and any components
of the product must be repaired by trained personnel as per the written instruction provided by the
Company. The product must not be modified without written approval from Shenzhen Comen Medical
Instruments Co., Ltd. User of the product shall undertake full responsibility of any faults caused by
misapplication, improper maintenance/repair, damage, or replacement conducted by anybody who is not
authorized by Shenzhen Comen Medical Instruments Co., Ltd
Warning
The product and its correlative standalone assemblies must not be applied in nuclear
magnetic resonance (MRI) environment.
No hazardous substances are generated by the part of this product contacting with anesthetic
agent.
No hazardous substances are generated by the mixture of the air source of this product and
anesthetic agent.
To avoid explosion hazards, inflammable Anesthetics, such as ether and cyclopropane, must
not be applied to the equipment. Only the non-inflammable Anesthetics conforming to IEC
60601-2-13(GB 9706.29) requirements can be adopted. The equipment allows the adoption of
1-1
non-flammable Anesthetics like enflurane, isoflurane, or sevoflurane, and only one type of
Anesthetics can be applied at one time.
Discarded anesthesia machines and packaging materials must be disposed in accordance with
local relevant laws and regulations, or wastes treatment schedule specified by the hospital. In
addition, they must be put in places out of the reach of children, and corresponding measures
shall be taken to prevent their harm to the ambient environment.
The gas flow rate, volume and leakage specifications are expressed at STPD except for those
associated with Anesthetic breathing system are expressed at BTPS.
1.4.1 Symbols
This manual provides several symbols to indicate various important matters and instructions.In order to
operate the machine correctly,please note the following symbols:
Warning
To warn you of the conditions where serious consequence, disadvantageous matters or danger
may occur. Failure to comply with the warning will result in severe personal injury or death
of the user or the patient.
1-2
Caution
To indicate potential danger or unsafe operation. If not avoided, it may lead to mild personal
injury, product malfunction, damages or property loss. It may also give rise to more severe
harm.
Attention
To emphasize the critical announcements and provide explanations so as to better use this
product.
Note
Warning
The device can only be operated by trained and skilled medical personnel.
Before operating the device, the operator must ensure that the device, connecting cables, and
accessories are intact and function properly.
The device can only be connected to a correctly installed power socket with protective
grounding. If the power socket is not connected to a protective grounding wire, disconnect
from the power cord, or use the device’s internal battery for power supply operation.
All analog and digital devices connected to the device must be certified by the specified IEC
standards (such as IEC 60950 Data processing equipment and IEC 60601-1(GB 9706.1)
1-3
Medical electrical equipment standards). All configurations shall comply with the valid
version of IEC 60601-1(GB 9706.1). The personnel responsible for connecting the optional
equipment to the input / output signal port shall be responsible for configuring the medical
system and ensuring that the system complies with the provisions of IEC 60601-1(GB 9706.1).
Multiple auxiliary output power sockets are provided on the rear of the device. These sockets
are used to provide power to additional equipment (i.e. anesthetic vaporizer, gas analyzer,
etc.) of the device. Do not connect other equipment to these sockets as it may affect patient’s
leakage current.
The alarm shall be set according to different patient conditions. Continuous and close
monitoring of patients is the surest way to ensure patient safety.
The physiological parameters and alarm information displayed on the screen of this device
are only for the reference of medical personnel, and cannot be directly used as a basis for
clinical treatment.
Connect the device to AC power supply before the internal battery runs out.
Do not open the device housing. All repairs or upgrades to the device can only be carried out
by personnel trained and authorized by the Company.
You cannot rely on the audible alarm system only to monitor the patient.
Adjusting the volume of the alarm to a small volume may cause the patient to be dangerous.
In order to avoid explosion hazard, flammable anesthetics such as ethyl ether and
cyclopropane cannot be used in this device. Use only non-flammable anesthetics that meet the
requirements of ISO 80601-2-13, IEC 60601-2-13(GB 9706.29), or ISO 8835-2(YY 0635)
standard. This device can use non-flammable anesthetics such as enflurane, isoflurane, and
sevoflurane. Only one anesthetic can be used at a time.
Do not turn off fresh gas until the anesthetic vaporizer is turned off. The anesthetic vaporizer
cannot be turned on without fresh gas. Otherwise, a high concentration of anesthetic vapor
can enter the equipment pipe and the surrounding air, causing harm to people and objects.
Qualified personnel shall check the condition of the patient. In some cases, some
life-threatening situations may occur, but they do not necessarily trigger an alarm.
1-4
Always set alarm limits to trigger an alarm before a dangerous situation occurs. Incorrect
alarm limit settings may cause the operator not to know that the patient condition has
changed dramatically.
Connecting medical device and non-medical device to an auxiliary power socket at the same
time may increase the leakage current, thus exceeding the permissible value.
Electric shock and fire hazard. Do not clean the device when it is turned on and / or powered
on.
In order to prevent electric shock, this device can only be connected to the power wiring with
protective grounding.
There may be a risk of electric shock. This device can only be opened by authorized service
personnel.
Disconnect the network power supply before removing the rear panel of this device or before
servicing this device.
Failure of the central air supply system may cause multiple or even all devices connected to it
to stop working at the same time.
Use a cleaning and disinfection program that meets your disinfection and risk management
regulations.
Refer to the operation and maintenance manual for all disinfecting equipment.
Do not inhale smoke from any disinfection process.
Care must be taken when lifting and operating the anesthetic vaporizer, as its weight may be
greater than expected, depending on the size and shape of the anesthetic vaporizer.
Do not use talc, calcium stearate, corn starch or similar materials to prevent the bellows from
sticking. These materials may enter the patient’s lungs or airways, causing irritation or
damage.
Disposable items may be considered as potentially biohazardous and shall not be reused.
1-5
When disposing of these items, the relevant regulations for the hospital as well as local
pollutants and biohazards shall be observed.
In order to avoid injury to the patient, do not test or maintain the device during use.
Check the performance specifications of the processing system with which the receiving and
transmitting systems are to be used to ensure compatibility.
It is not possible to use this device close to or stacked with other devices. If necessary, the
device shall be closely observed to ensure that it operates properly in the configurations used.
Make sure that the current alarm preset for the device is appropriate for each patient.
In any area, it is dangerous to use different alarm presets for the same or similar devices.
Subject to the size and weight of this device, the device shall be removed by qualified
personnel.
Excessive machine load may result in dumping. Equipment connected to the side of the
machine shall be within the rated weight range to prevent the machine from tipping over.
When moving the device, excessive load on the device may result in danger of dumping.
Before moving the device, remove all devices on the top panel of the device and all monitoring
devices installed on the side of the device. Be careful when moving the device up and down
ramps, turning corners, and crossing the threshold. Do not attempt to run over hoses, cables,
or other obstacles when moving the device.
Leakage may affect accuracy. Test before proper operation to ensure proper operation of the
device. Do not use leaked circuits.
It is recommended to connect the vent of the device to the hospital’s exhaust emission system
to prevent hospital personnel from being exposed to the gases emitted by the device.
Operating the device below the minimum flow rate may result in inaccurate results.
Improperly cleaned materials can cause biological contamination. Use a cleaning and
disinfection program that meets your disinfection and risk management regulations
Users shall follow recommended daily disinfection procedures for this device and any
reusable accessories.
If this device is damaged in any way that endangers the safety of the patient or user, stop
1-6
using the device and attach a visible mark to indicate that the device is not available. Contact
the Company for technical support.
High concentrations of O2 greatly increase the likelihood of fire or explosion. Oil and grease
may burn at the same time. Therefore, oil or grease cannot be used in oxygen-rich
environments where possible.
Use this device only by professional medical personnel. This device may cause radio
interference or interrupt the operation of nearby devices. It is necessary to take
circumvention measures, such as repositioning the orientation or position of the device, or
shielding the place where the device is placed.
Make sure that there is an independent ventilation mode available at any time during the use
of this device.
Before using this anesthesia machine after cleaning or disinfection, power on the system and
follow the on-screen instructions to perform a leak test.
Use of lubricants not recommended by the Company will increase the risk of fire or
explosion. Use the lubricants approved by the Company.
Low pressure regulators and flowmeters are susceptible to high pressure and may explode
under pressure if improperly maintained or disassembled. Only qualified personnel should
replace or disassemble the connector.
Do not disassemble low pressure regulators, flowmeter devices, or connectors under pressure.
Sudden release of pressure may cause injury.
Check the specifications of the AGSS transmitting and receiving systems and the
specifications of this device to ensure compatibility and prevent the receiving system from
being mismatched.
Before each use, carefully inspect all parts of the breathing system. Make sure that all parts
are free of any obstacles or debris that pose a potential hazard to the patient.
The power plug is used to separate the anesthesia system circuit from the power supply. Do
1-7
not place the device where it is difficult to operate the plug.
When connecting an external device through the input / output signal port or replacing the
O2 battery, do not touch the patient to prevent the patient from leaking current exceeding the
standard requirements.
Avoid connecting two or more hose assemblies in series, which can cause pressure and flow
losses.
When the pipe between the exhaust gas treatment system and the AGSS is clogged, the
extraction flow of the exhaust gas treatment system is insufficient, or the exhaust gas
treatment system cannot work, the exhaust gas in the AGSS may overflow the atmosphere at
a speed exceeding 100ml/min. At this time, AGSS is not recommended.
Use of incorrect connectors can be dangerous. Ensure that all components use the correct
connectors.
Avoid using lower nominal pressure flexible connectors to replace high-pressure flexible
connectors.
After changing the CO2 absorbent or installing a CO2 absorption canister, make sure that
CO2 is fully absorbed by the absorbent.
Before moving the device, remove the spare cylinder and the objects on the top plate and
bracket to prevent the device from tipping.
Caution
In order to ensure patient safety, use the parts and accessories specified in this Manual.
This device can operate normally at the level of interference immunity identified in this
Manual. If the interference level is higher than this level, it may trigger an alarm and may
cause mechanical ventilation to stop. Pay attention to false alarms caused by high-intensity
electric fields.
The device may lose its balance if it is tilted more than 10°. Be careful when moving or
placing this device on a slope that exceeds 10°. Do not hang objects on both sides of the device
to avoid excessive imbalance.
Follow the checklist for daily inspections. In the event of a system failure, do not operate this
1-8
device until the fault is cleared.
Before starting the device, the user must be familiar with the information contained in this
Manual. The device must be inspected and repaired by qualified service personnel as
required.
If the device cannot be operated as described in the Manual, it must be inspected and
repaired by qualified service personnel as required before being put back into use.
When the device and its accessories will exceed their service life, they must be disposed of in
accordance with the guidelines for the management of such products and local regulations for
the management of contaminated and biohazardous goods.
Electromagnetic fields will affect the performance of this device. Ensure that all external
equipment used near this device must comply with the appropriate EMC requirements.
Mobile phones, X-rays, and MRI equipment are all possible sources of interference, as they
emit high-intensity electromagnetic radiation.
Make sure that the gas sources of the device always meet the relevant technical specifications.
Prior to clinical use, the device must be properly calibrated and / or tested as described in this
Manual.
If a system malfunction occurs during the initial calibration or test, the operation of this
device shall be stopped until qualified service personnel have eliminated the malfunction.
Do not use abrasive cleaners (such as steel wool or silver polish) to clean the parts.
1-9
Do not use peracetic acid or formaldehyde fumigation.
After maintenance, functional tests, sensor tests, and system tests must be performed before
clinical use.
Only anesthetic vaporizers with a Selectatec interlock system can be applied to this device.
After each replacement of the anesthetic vaporizer, perform a leak test on the breathing
system circuit.
Use cleaners with care. Excessive liquid may enter the device, thus causing damage.
Do not subject any part of this device to high temperature and high pressure unless it is
specifically stated in this Manual that it is a high-temperature and high-pressure part. Clean
the device as specified in this Manual.
The valve of each inhalation and respiration valve assembly on the breathing system is fragile
and must be handled with care when removing the valve seat from the valve assembly.
If the bellows are moistened with water after cleaning, the bellows surface may have creases
that can cause the bellows to fail to unfold. Make sure to wipe out all moisture from the
bellows after cleaning.
Do not connect any non-isolated device to the DB9 connector of this device.
Do not connect any device to the USB port of the device except for the USB storage devices
approved by the Company.
Do not subject the following parts to high temperature and high pressure: airway pressure
gauge, O2 battery, and flow sensor. These parts cannot withstand soaking or the heat and
pressure of high-temperature and high-pressure processing. If the breathing system is
configured with a disinfectant flow sensor, the flow sensor can be subjected to
high-temperature and high-pressure vapor disinfection.
This device is not suitable for use in magnetic resonance (MRI) environments.
In order to ensure accurate measurement and avoid damage to the device, use only cables
and accessories approved by the Company.
Use the accompanying power cord. For replacement, use only power cords that meet the
specifications.
Do not press down on the manual arm or hang a heavy weight on it. Excessive weight may
1-10
bend or damage the manual arm.
Since sudden release of pressure may cause injury, be careful when disconnecting the “quick
connector”.
Avoid factors that may damage the hose assembly, including excessive bending, rolling, wear,
system pressure and temperature exceeding the hose rating, and incorrect installation.
When removing the breathing system, care must be taken to raise and manipulate the
breathing system. Given the importance and shape of the breathing system, these operations
can be tricky.
When the electronic flow control system fails, the backup flow control system will be enabled.
The O2 basic flow of the backup flow control system is 0.2 L/min. The backup flow control
system displays only one total flowmeter. The total flowmeter can display a maximum flow of
15L/min.
The flow control knob of the backup flow control system shall be slowly rotated to avoid
damaging the control valve. Do not rotate the flow control knob when the flowmeter reading
exceeds the range. When the flow control knob is rotated clockwise to reduce the flow, the
flowmeter reading shall reach 0L/min before reaching the clockwise mechanical stop position.
When the knob has reached the stop position, do not rotate it again. Rotate the flow control
knob counterclockwise to increase the flow.
Do not use damaged device or accessories. During normal use, check all cables (such as AC
power cord and patient connection cable) regularly for damage. If damaged, replace it.
The O2 ratio (FiO2) shall be monitored when using an auxiliary O2 / air supply flowmeter. If
O2 monitoring is not performed, the concentration of O2 delivered to the patient will be
unknown.
Unlocking the casters may cause unexpected movement. The operator shall lock the casters
while using this device.
Unmounted devices may slide off the top plate. The device shall be securely mounted on the
top plate.
The voltage on the auxiliary socket shall be the same as the voltage of the socket into which
the device is plugged. Make sure that the voltage rating of the device plugged into the
auxiliary socket is the same as the supply voltage of this device.
During transportation and storage of the anesthetic vaporizer, a plug shall be used to block
1-11
the inlet and outlet of the anesthetic vaporizer, thus preventing impurities from entering the
vaporizer.
Do not use any flow outlet as a handle when moving this device. Flow outlets may be
damaged. Use the metal side bar on the unit to move the device.
Attention
Install the device in a place convenient for observation, operation and maintenance.
The quick opening of the cylinder valve may cause an unexpected pressure difference. Due to
potential fire or explosion hazards from O2 pressure shocks, the cylinder valve shall be
opened and closed slowly.
If the anesthesia machine is configured with a total flowmeter, the total flowmeter will be
calibrated at 100% O2. For other gases or gas mixtures, the accuracy of the flowmeter may be
reduced.
Changes in inlet pressure, outlet resistance, or ambient temperature may affect the accuracy
of the flow rate.
There are regional or national regulations that apply to medical device manufacturers.
The operator shall be located directly in front of the device and within 4 meters of the screen
to observe the display information of the device.
The figures in this Manual are for reference only. The interface may differ depending on the
system configurations and the selected parameters.
Place this Manual near the device so that it can be easily accessed when needed.
The software of this device is developed in accordance with the requirements of YY/T 0708,
thus minimizing the possibility of danger caused by software errors.
This Manual describes the product according to the most complete configurations. The device
you purchased may not have certain configurations or functions.
The device uses its own respiratory pressure monitoring during operation.
The device uses its own respiratory pressure limiting device during operation.
1-12
The device uses its own respiratory capacity monitoring during operation.
The device uses its own integrity alarm system for the breathing system during operation.
The device uses its own continuous pressure alarm during operation.
The anesthesia gas conveying equipment is used in conjunction with the anesthesia gas
monitoring module complying with YY 0601. The patient circuit and the anesthesia gas
monitoring module shall be connected through an adsorption tube.
When using this device, the concentration of anesthetic shall be continuously monitored to
ensure that the output of the anesthetic is accurate.
Before performing all operations and during operation, the level of the anesthetic fluid shall
be checked. Liquid shall be added when the liquid level is below the warning line. For the
addition of anesthetic to the anesthetic vaporizervaporizer and other information, refer to the
instructions for use of the anesthetic vaporizer.
The device system is designed with the anesthesia gas conveying equipment that complies with
YY 0635.3.
The battery of this device is not a user-serviceable part. Only authorized service
representatives can replace the battery. If the system is not used for a long time, contact a
service representative to disconnect the battery. Dispose of the battery in accordance with
relevant local regulations. When the battery reaches its service life, dispose of it in accordance
with relevant local regulations.
The place designated for servicing O2 equipment shall be clean, grease-free, and not used for
repairing other equipment.
1-13
Chapter 2 Product Overview
This chapter gives an overview of the anesthesia machine and its functions.
2-1
Attention
This Manual describes the device according to the most complete configurations. The device
you purchased may not have certain configurations or functions.
2-2
2.1.2 Anesthesia machine AX-900A
Notes Symbol
2-3
Notes Symbol
Flow regulation
Alternating current
Battery
AUDIO PAUSED
Alarm
Network port
Standby
Cylinder O2-inlet
Cylinder N2O-inlet
Cylinder AIR-inlet
2-4
Notes Symbol
O2、N2O、AIR
280~600kPa
Equipotentiality
2-5
Notes Symbol
Isolation transformer
Pipeline Pipeline
Insp
Inspiratory/expiratory flag
Exp
Autoclavable
Not Autoclavable
BY-Pass flag
APL valve
Mechanical ventilation
Material explanation
2-6
Notes Symbol
USB port
Video output
Hot Caution!
2-7
Notes Symbol
Date of manufacture
Serial Number
Address of manufacture
MR Unsafe
This way up
2-8
Notes Symbol
Fragile
Do not stack
Do not roll
center of gravity
recyclable
Environmental protection
Temperature limitation
Humidity limitation
2-9
Notes Symbol
Abbrev Definition
AA Anesthetic
C
CO2 Carbon dioxide
Compl Compliance
CPB Cardiopulmonary bypass
CPAP/PSV Continuous Positive Airway Pressure/Pressure-su
pport ventilation
E
Et Exhale gas concentration
EtCO2 Expiratory-end tidal CO2 concentration
EtO2 Expiratory-end tidal O2 concentration
Exp Expiratory
F
Fi Inspired gas concentration
FiO2 Fraction of inspired oxygen
FiCO2 Fraction of inspired CO2
I
I:E Inspiratory:Expiratory time ratio
Insp Inspiratory
M
MAC Minimum alveolar concentration
2-10
Abbrev Definition
MV Per minute ventilation
MinRate Minimum Rate
N
N2O Nitrous oxide
O
O2 Oxygen
P
P-F Pressure-Flow loop
P-V Pressure-Volume loop
Paw Airway pressure
PCV Pressure-control ventilation
PEEP Positive end expiratory pressure
Pinsp Inspiratory pressure
Plimit Limiting pressure
Pmean Mean pressure
Ppeak Peak pressure
Pplat Plateau Pressure
ΔPps Support pressure
PRVC Pressure regulated volume control ventilation
PSVPro Pressure-support ventilation protection
R
Raw Resistance
S
SIMV-PC Synchronized intermittent mandatory ventilation-
pressure control (pressure controlled)
SIMV-VC Synchronized intermittent mandatory ventilation–
volume control (volume controlled)
SIMV-PRVC Synchronized intermittent mandatory ventilation-
Pressure regulated volume control synchronous
ventilation
T
VT Tidal volume
VTexp Expiratory tidal volume
VTinsp Inspiratory tidal volume
2-11
Abbrev Definition
Tinsp Inspiration time
Trigger Inspiratory trigger
Tslope Pressure slope
Tpause Inspiratory pause
V
VCV Volume control ventilation
Vol Volume
V-F Volume-Flow loop
2-12
2.4 System Construction
2.4.1 Front
1
2 19
18
3
17
4 16
5 15
14
13
6
7 8 9 10 11 12
2-13
11
1 Ventilator/monitor display Backup flow control system button
10 Total flowmeter
Attention
External vacuum suction and internal vacuum suction choose one of them for
configuration,AX-900A can be optionally equipped with external vacuum suction,AX-900 can
be optionally equipped with internal vacuum suction.
The Control Functions to be implemented through the Front of Anesthesia machine AX-900/900A
Item Name Notes
1 Ventilator/monitor For details, refer to Section 2.5.6.3 CONTROL PANEL.
display
2/3 Auxiliary air / oxy It is used for auxiliary air / oxygen flow output display. For details, refer to
gen flow meter Section 2.5.8 AUXILIARY OXYGEN AND AIR FLOWMETER.
2-14
4 module area The CO2 module, AG module and BIS module mentioned in this Manual can
be inserted and identified.
Press the bottom end of the central brake pedal with your foot to , to
apply brake to the anesthesia machine; press the upper end of the central
brake pedal to to release the brake (see the figure below). After the
7 brake is released, the anesthesia machine can be moved.
Central brake
Push the „O2 +‟, oxygen flush button to supply high-flow oxygen to the
breathing system.
10 Total flow meter It is used for displaying the combined flow of oxygen, air and nitrous oxide.
11 Backup flow control Press the button to expand the flow regulation knob of the backup flow
system button control system.
12 Flow regulation It is used for regulating the nitrous oxide, air and oxygen flux. Rotate the knob
knob of the backup counterclockwise to increase the flow. Rotate the knob clockwise to decrease
flow control system the flow.
2-15
13 Storage drawer Provide two storage drawers (lockable).
Set the switch to its position “ ” to enable gas input and turn on the system;
set the switch to its position “ ” to turn off gas input and turn off the system,
as shown in the figure below:
16
System switch
2-16
14 External vacuum suction(AX-900A):
The function button is the same as the internal vacuum suction.
19 Flow control Touch the touch screen to control the input. Set the [Balance gas], [Oxygen]
(full electronic flow and [Air].
meter)
Note
When using a backup flow control system, make sure that the oxygen,air and nitrous oxide
flow controllers are fully closed at the start and end of each ventilation.
The safety system in this device prevents the delivery of hypoxic mixture to the patient. In the
absence of oxygen, it will not deliver nitrous oxide. The gas circuit safety system ensures
that oxygen is set at a minimum of 25% in the mixture of oxygen and nitrous oxide.
2-17
2.4.2 Back
13 12 11
1
2
4
5
10
9
8
1 Main power socket and its system 8 Auxiliary power output socket‟s
circuit breaker circuit breaker
2-18
rface
Warning
Port connectors used for other specific purposes are not compatible with ISO 5356-1, ISO
5356-2 or ISO 594-2. (luer lock).
2-19
2.4.3 Left
2 6
2-20
bottle
6 Auxiliary O2 / air Nozzle for auxiliary O2 / air output. The outlet merges the auxiliary O2
outlet / air supply flowmeter into a single output.
2.4.4 Right
2-21
Warning
Never use antistatic breathing tubes and face masks, which may cause a fire if they are used
nearby surgical equipments for high frequency surgery.
11
10 12
9 13
8 14
7
6 15
4
3
2 1
Fig. 2-7 Breathing System of Anesthesia machine Series
2-22
8 APL (adjustable pressure limiting) valve
8 APL valve
2-23
Artificial Manual bag suspension arm.
Bellows assembly Used to separate the breathing system‟s gas from the driving gas O2 /
11
(including bellows cover) air.
Select manual/spontaneous (Manual bag) or volume control
(ventilator). When manual/spontaneous is selected, set the manual
switch to its position “Manual” ( ). If volume control is selected, set
the mechanical control switch to its position “Mechanical Control”
Manual/mechanical ( ).
13 ventilation switch
14 Inspiratory check valve Allow the inhaled gas to flow to the patient and prevent backflow.
15 Inspiration port Breathing circuit‟s inspiration port.
Attention
The bellows cover is a transparent cover with scale marks from 300 to 1500ml. These scale marks
are for reference only. VT shall be read from the user interface. The delivered VT is the sum of the
bellows displacement and the fresh gas flow.
The values on the APL valve and the airway pressure gauge are for reference only. Calibrated
patient airway pressure is displayed on the user interface.
2-24
2.5.2 Structural Composition of AGSS
7 2
6
3
5
4
2-25
2.5.3 Vacuum suction device
Vacuum suction refers to the suction device with vacuum generated by the venturi device. The intended
use is negative pressure for sputum discharge in clinical respiratory tract and esophagus and surgery. It
mainly consists of negative pressure generator, negative pressure regulating valve, negative pressure indicator,
overflow cup, filter, liquid collecting bottle and suction tube, and is mainly used in the collection of medical
waste liquid. It is designed with overflow protection function to prevent backflow of fully collected waste so
as to ensure the safety of the gas pipeline. The schematic drawing is as follows.
Overflow valve
Change-over valve
Regulating Valve
Air resistor
Overflow bottle Liquid collection bottle Liquid collection
2-26
4
3 5 6
Note
Refer to the user’s manual supplied with the external vacuum suction system for specific
information.
Warning
FULL mode is strictly prohibited to be used for patient attraction.Excessive negative pressure will
hurt the patient’s tissues and cause injury to the patient.
The device is a high negative pressure suction device.Thoracic suction and abortion are strictly
prohibited.The excessive negative pressure value will hurt the patient’s fragile body tissue and
cause damage to the patient’s life.
1. ACGO Mode:
When ACGO cover is opened and the anesthesia machine is set to its position ACGO, fresh gas
flow passes through the ACGO outlet in the front of the machine, and the interface is changed into
ACGO mode.
Do not use external ventilator on the ACGO. Do not use the ACGO to drive an external ventilator.
2. Non-ACGO mode:
Mechanical ventilation or manual/spontaneous ventilation for patient may be achieved by the
breathing system.
2-28
3. When ACGO offers fresh gas to respiration apparatus:
Mechanical ventilation is disabled.
Pressure gauge, Manual/mechanical ventilation switch, APL valve and Manual bag support
column are spontaneous of external circuit devices.
AX-900 and AX-900A designed to work with Draeger Series Anesthesia Vaporizers for
non-flammable anesthetic gases fed from Selectatec® (registered trademark of Ohmeda)
fixed/interlocking devices.
Caution
Please refer to the user’s manual delivered along with the Anesthesia Vaporizer for specific
information about relevant Anesthesia Vaporizers.
Note
VCV
SIMV-VC
PCV
2-29
SIMV-PC
CPAP/PSV
PRVC
SIMV-PRVC
PSVPro
Manual/Spont
Control knobs
2-30
1 2 3 4 5 6 7
19
18 8
10
11
13
17 16 15 14
12
2-31
concentration
10 System prompt message area
Caution
When anesthesia machine is restarted, patient type before the last shutdown will be reserved.
This flow control system is able to display the real-time flow levels of oxygen and balance gas.
Balance gas can be set to [AIR] or [N2O] or [NULL].
2-32
2
It is used for auxiliary air / O2 flow output. Rotate the auxiliary O2 / air supply flowmeter control
knob counterclockwise to increase O2 / air flow. Rotate the auxiliary O2 / air supply flowmeter
control knob clockwise to reduce the flow.
2-33
2
Caution
Turn anticlockwise the flow control knob to increase the oxygen flow, or turn it clockwise to
reduce the flow.
2-34
2.5.9 High pressure oxygen outlet
Auxiliary output power supply is a type of safe and effective power supply that is alternating current
output by isolation transformer, and mainly serves as auxiliary power supply for patient monitoring
systems.
Warning
To configure the auxiliary output power supply, the equipment connected to the auxiliary
output power supply shall fall within the voltage/current specifications of the auxiliary
output power supply. The equipment connected to the auxiliary output power supply shall
2-35
be equipment specified by the manufacturer; otherwise, the leakage current may exceed
relevant limits, endanger patient or operator, or even damage the anesthesia machine or
external equipment.
If your anesthesia machine is not equipped with an isolation transformer, the equipment
connected to the auxiliary output power supply may increase the leakage current. The
leakage current shall be determined regularly. In order to reduce the total leakage current,
we suggest that you select anesthesia machines equipped with isolation transformer.
Warning
With standard configuration, anesthesia machine may remain stable when it is inclined by
10°. Do not hang any articles from both sides of an anesthesia machine for fear of
off-balance hazard.
Located below the top plate, it illuminates the workbench, allowing users to read the settings on the
dial of the anesthetic vaporizer in a dark room. The brightness of the light can be adjusted by
adjusting the operation ceiling lamp regulator below the top plate.
The lighting switch is located at the bottom left of the top plate, allowing users to read the scale
display of the auxiliary O2 supply flowmeter, auxiliary air flowmeter, and total flowmeter in a dark
2-36
room.
This device is a full-featured anesthesia delivery workstation. The protruding edge of its stainless
steel workbench can prevent items placed on it from rolling out or sliding off the edge of the
workbench. The wraparound armrests on the workbench can better position the device. Two large
drawers are provided for storing items. Dovetail rails on both sides of the device are used to mount
standard accessory arms for monitors and other equipment. In addition, non-slip foot pedals and
central brake are available.
The hook on the side of the breathing system is used to suspend the breathing circuit.
2.5.16 Battery
The anesthesia machine is equipped with an internal rechargeable battery to ensure that the system can
operate normally in the event of a power outage. When the device is connected to an AC power supply,
the battery can be charged regardless of whether the device is turned on or not. In the event of a sudden
power outage, the system will automatically switch to internal battery power without interrupting the
operation. When the AC power supply recovers within a specified time, the battery starts to charge, and
the system stops using the battery while automatically switching to the AC power supply to ensure the
continuous operation of the system.
If the battery power is too low, it will cause the power supply to malfunction. The device will trigger an
advanced alarm and display the alarm information in the technical alarm area. In this case, use an AC
power supply to supply power to the anesthesia machine to get it back to work and charge the battery.
2-37
2.5.19 Network Port
The symbol “ ” on the back cover of the equipment indicates the network port. This port is
used for data transmission between the Medical Systems.
Warning
All the simulation and digital equipment connected with this system must be the products
certified by the designated IEC standards (e.g. IEC 60950 Data Processing Equipment
Standard and IEC 60601-1 Medical Equipment Standard)
on the rear shell of the instrument is used for an external display. The external display
Equal-potential grounding means connecting the shell or metal part of equipment to the ground
lead so as to avoid indirect-contact shock, explosion, fire hazard and transient overvoltage and
interference caused by ground fault, and to protect the personal/equipment safety.
One end equal-potential ground lead shall be connected to the equal-potential ground pole fitted
on the back cover of the equipment, and the other end shall be connected to one of the terminals
of the equal-potential system. If the protective ground system is damaged, the equal-potential
earthed system may undertake the safety function of the protective grounding lead.
Check whether or not the instrument is kept in good working order prior to each operation.
Warning
If the protective ground system works unstably, the equipment shall be powered by internal
power supply.
2-38
Chapter 3 Basic Operations and Guidance
Warning
Alarms given by an anesthesia ventilator indicate that the patient is being subjected to
potential risks. The causes of all alarms shall be made certain so as to make sure the safety of
patients.
Do not place the power plug used to disconnect the monitor from grid power supply in a
position not easily accessible by the operator.
Before using the device on the patient, make sure that the device is installed correctly and
intact.
The operator shall not touch the patient and the charged equipment outside the device at the
same time.
The input / output signal port can only be connected with the specified external device.
2. Plug the power cord into a power socket. Turn on the main power switch on the back of the
device. The AC power indicator lights up when connected to AC power supply. If the battery power is
low, the battery is charging.
a) The alarm indicator flashes once in red-yellow order, and then the buzzer sounds a
“beep”.
b) The system enters the ON state. The control panel screen displays the Comen LOGO.
Then enter the [Self Test] interface.
c) In the [Self Test] interface, the system automatically performs a series of self-tests and
3-1
displays the startup self-test progress bar. After the self-tests are completed, the self-test results are
displayed.
d) After the completion of self-tests, click the [Continue] button to enter the [Check Before
Use] interface.
e) After confirming that everything is normal, click [Continue] to enter the next step “Leak
in vent mode”.
f) According to the prompt on the interface, select “Start” for “Leak in vent mode”. Enter
“Leak in bag mode” in case of passing the test.
If the test fails, refer to “4.5.2 Breathing System Leak Test in Mechanical Ventilation Mode”
for retesting.
g) According to the prompt on the interface, select “Start” for “Leak in bag mode”. Enter the
“Standby” interface in case of passing the test. At this time, after setting the patient information and
selecting the patient type “Adult” or “Child”, select “Ventilation Start” to enter the user interface.
If the test fails, refer to “4.5.3 Breathing System Leak Test in Manual Ventilation Mode” for
retesting.
Warning
To run the equipment for patients, make sure that the system connection is errorless and is
kept in good order and condition, and finish all tests specified in Chapter 4 Tests Before Use.
If the equipment fails to pass the tests, never use the equipment. Please contact immediately
an authorized service representative to repair the equipment.
The anesthesia machine is equipped with built-in charge batteries to make sure that the
anesthesia machine may still be used normally in case of power failure. Once the anesthesia
machine is connected to AC power supply, it may charge the batteries, disregarding whether
or not it is turned on. In case of sudden power cut, the system can power the anesthesia
machine by its batteries so as to avoid interruption of anesthetization. When the AC power
supply resumes at target time, the system can stop battery feed and enable AC power supply
automatically in order to make sure normal work of the anesthesia machine.
Battery feed can be maintained only for a period of time. Once the battery levelis too low,
anesthesia machine may give out a high-level alarm, and display the alarm message [Low
Battery Voltage] in the technical alarm zone.
3-2
Caution
Avoid short circuit of battery.
Caution
The patient type is not optional in ventilation mode.
3-3
3.2 Configure Volume for Alarms, Prompts and Key Operations
1. Set the system switch to its status ON “ ”.
2. On the user screen, select [Alarm] menu → Access the [Sound] menu.
3. Set volume in grade “1-8” respectively for [Alarm Sound Volume] 1, set volume in grade
“0-7” respectively for [Indicate sound Volume] 2 and [Key sound Volume] 3. The length of
cyan bar indicates the current volume grade. If no grid is cyan, the volume is muted, as shown
in the figure below.
3-4
Note
1: The alarm volume is set to adjust the volume of all high, medium, and low priority audible
alarms.
2: The prompt volume is the volume of the prompt message that appears in the alarm prompt
area.
3: The button volume is the volume of the sound produced by touching the soft buttons on the
operation interface via a touch screen.
2. Set the manual/mechanical ventilation switch to manual position, the screen displays
[ Manual/Spont].
3. Select [Alarm] ,push and turn the “control knob” key to switch over between [ON] or [OFF].
4. Select [ON] or [OFF] , push the “control Knob” to confirm the current option.
Warning
During [CPB] is set to [ON] , part of the physiologic alarm messages may not be triggered;
therefore, the setting shall be applied cautiously. The physiologic alarms include: Apnea,
Apnea>2min, Low Paw, High VTexp, Low VTexp,High MV, Low MV.
2. On the user interface, select [Alarm] menu→ Access the [ventilator] 、 [AG] 、
[CO2]or[BIS]1menu, and set the High/low alarm limits for the parameters.
Note
Note 1: See [Alarm setting] given in relevant chapter for details of High/low alarm limits for
parameters of [ventilator] 、 [AG]、[CO2]and[BIS].
3-6
Note
Please refer to the user manual supplied with the external vacuum suction system for
operation information on the external vacuum suction system.
1. According to 5.9.1 Installation of internal vacuum suction, assemble an external pipe collection
system of a internalvacuum suction device to an anesthesia machine;
2. Confirm that the negative-pressure gear switch is in the OFF position, and the negative-pressure
adjustment knob is in the minimum position adjusted counterclockwise;
3. Open the anesthesia machine‟s O2 source, and confirm whether the pressure is within the applicable
range of the anesthesia machine (280-600kPa);
4. Block the patient end of the negative-pressure suction tube, and turn the negative-pressure gear
switch to the FULL position; observe whether the reading on the negative pressure gauge can reach
60kPa or above;
5. Turn the negative-pressure gear switch to the REG position; slowly rotate the negative-pressure
adjustment knob clockwise; observe whether the reading on the pressure gauge changes with the
adjustment; confirm that the negative pressure can be adjusted to 60kPa or above, and that the
pressure can remain steady when not adjusting;
6. Confirm that there is no fault. Adjust the negative pressure to the desired pressure for use.
1. After use, rotate the negative-pressure adjustment knob counterclockwise until the negative
pressure value returns to zero;
2. Turn the negative-pressure gear switch to the OFF position, and the vacuum suction device is
completely turned off.
Warning
Keep the vacuum suction switch in the OFF state when you are not using the vacuum suction
device.
1. According to 5.9.2 Installation of external vacuum suction, assemble an external vacuum suction
device to an anesthesia machine;
3-7
2. Press the driving gas switch to the OFF position; turn the negative-pressure gear switch to the OFF
position as well;
3. Rotate the negative-pressure adjustment knob counterclockwise to the minimum position until it
can no longer be rotated;
4. Open the O2 source, and confirm that the gas source pressure is within the applicable range of the
external vacuum suction (280-550kPa);
5. When the gas source pressure is appropriate, pull the driving gas switch to the ON position;
6. Block the patient end of the suction tube, and turn the negative-pressure gear switch to the FULL
position; observe whether the reading on the negative pressure gauge can reach 60kPa;
7. Turn the negative-pressure gear switch to the REG position; slowly rotate the negative-pressure
adjustment knob clockwise; observe whether the reading on the pressure gauge changes with the
adjustment; confirm that the negative pressure can be adjusted to 60kPa or above, and that the
pressure can remain steady when not adjusting;
8. Confirm that there is no fault. Adjust the negative pressure to the desired pressure for use.
1. After use, rotate the negative-pressure adjustment knob counterclockwise until the negative
pressure value returns to zero;
2. Turn the negative-pressure gear switch to the OFF position, and press the driving gas switch to the
OFF position. The vacuum suction is then completely shut down.
Warning
Please keep vacuum suction switch at OFF when the vacuum suction device is not in use.
3-8
Fig. 3-4
Warning
Click on the system default, vent mode setting restored to system defaults.
Note
3-9
3.7.4 Set Inspiratory:Expiratory Time Ratio
3-12
3. Confirm the parameter setting via the control knob.
The display accuracy of the numerical values on the flow meter is two decimal places in case of flow
rate less than 1 L/min and is one decimal place in case of flow rate greater than 1 L/min.
For this device, the electronic flow control system includes two modes: total flow and single-pipe flow.
The method of setting flow control mode is as follows:
1. In the user interface, select the [Config] menu → Enter the [Flow Meter Config] label.
The total flow control mode of the electronic flow control system is shown in the figure below:
You can make the following settings in the total flow control menu:
The single-pipe flow control mode of the electronic flow control system is shown in the figure below:
You can make the following settings in the single-pipe flow control menu:
1. Set the [Balance gas] to [AIR], [N2O] or [NULL].
2. Set the flow value of balance gas through the soft keypad.
3. Set the oxygen flow value through the soft keypad.
3-13
3.8.3 Optimal Flow
Note
The optimal flow is valid only when the anesthesia system is equipped with the AG module
with paramagnetic oxygen and is in mechanical ventilation mode.
When the data used for optimal flow calculation is invalid, the optimal flow function will be
invalid.
The function of optimal flow indicator is to save anesthetics.This function only serves as a
reminder,and it does not involve the control function of the anesthesia machine control
module.
The principle of optimal flow rate is to monitor oxygen concentration and various anesthetic gas parameters
through the AG module with paramagnetic oxygen,analyze and calculate the parameters to obtain the gas
consumption,identify the relationship between fresh gas supply and gas consumption,and display the prompt
information on the screen according to the relationship identified.
The optimal flow indicator is used for indicating the relationship between the total flow of fresh gas and the
total consumption.The following figure is the optimal flow indicator function.The bar histogram shows the
supply of fresh gas,and the value below the bar graph represents the total flow value.
3-14
Scope Colour Meaning
Too high yellow The triangle indicates in this region means the difference
between total fresh gas flow and total consumption is greater
than 1 L/min,which will cause the waste of anesthetics.
valid green The triangle indicates in this region means the difference
between total fresh gas flow and total consumption is within
1 L/min,which will save the anesthetics.
Too low red The triangle indicates in this region means the total fresh gas
flow is less than the total consumption,which the fresh gas is
in short supply.
The anesthesia system electronically monitors the pipeline gas supply and the pressure of spare gas cylinder.
When the alarm is triggered in case of abnormal gas supply pressure, the pressure value of the gas supply
pressure is displayed in the corresponding color.
3-15
3.9 Backup Flow Control System
The anesthesia machine has a manual mechanical switch for turning on the backup flow. The flow meter
control system of the electronic flow control system fails (such as communication exception and touch screen
failure) or when the system has low battery power, if you turn on the switch above, the device will enter the
backup flow meter control system.
The interface diagram of the backup flow control system is shown below:
After the backupflow control system is expanded, when using the backup flow control system, you can
rotate the needle valve to regulate the flow. The total flowmeter is used to display the total flow. With the O2
concentration value displayed on the interface, you can calculate the O2 flow and balance gas flow in the gas.
The alarm sound off function here can only disable the technical alarm of the backup flow control system and
the failure alarm of the electronic flow control system.
When the electronic flow control system has not failed, you can press the backup flow control system
button to expand the backup flow control system. When you want to turn off the backup flow control system,
close all needle valves and press the [Close physical flowermeter] button on the interface. Select [Yes] in the
pop-up dialog box to shut down the backup flow control system.
Warning
3-16
When using a backup flow control system, make sure that the oxygen ,air and nitrous
oxide flow controllers are fully closed at the start and end of each ventilation.
The backup oxygen flow meter shall be marked with pure oxygen for application. In case of
using gas mixture, the accuracy will be affected.
Note
The total flow meter is calibrated with 100% oxygen. For other gases or gas mixtures, the
accuracy of flow meter may be reduced.
When viewing the readings on a total flow meter, the line of sight shall be at the same level as
the float. In case of different perspectives, the same scale readings may be different.
If there is a difference between the readings displayed on the electronic flow meter and the
total flow meter, take the reading on the electronic flow meter as a reference and the reading
on the total flow meter as the approximate value.
Note
Anesthesia ventilator shall be configured to comply with breathing system in ISO 80601-2-13
and YY 0635.1, together with AGSS in ISO 80601-2-13 and YY 0635.2。
1. Rotate the control knob of the APL valve to adjust the pressure inside the breathing system to
appropriate range.
2. Set the Manual/mechanical ventilation switch to “Manual” position, and
[ Manual/Spont] will be displayed on the screen.
3. If necessary, press the “oxygen flush” button to inflate the manual bag.
4. In [Manual/Spont] mode, the APL valve is used to regulate the peak pressure in the breathing
system and the gas volume in the manual bag. When the pressure in the breathing system
reaches the threshold, the APL valve is opened to discharge the excessive gas in the breathing
system.
3-17
The pressure waveform and flow rate waveform are shown in the figure below:
Fig. 3-5
Note
Make sure Manual/Spontaneous mode is always available when this device is used on
patients.
3-18
8. If necessary, press the “Oxygen flush” button to inflate bellows.
9. Start the mechanical ventilation by pressing the [Ventilation Start] button.
Note
Make sure all parameters are set properly before starting up the new mechanical
ventilation mode.
Ventilator mode settings will restore to system defaults after anesthesia machine shutdown.
3-20
Paw Tresp
(Rate) Ppeak
Pplat
Peep
Flow Ti Te
Tpause
Under the volume controlled VCV mode, a preset tidal volume will be provided. According to the
preset [VT] , [Rate], [Tpause]and [I:E] , the airflow delivered in the inspiration phase will be
calculated by the ventilator. The inspiratory flow sensor detects the inhaled tidal volume in
real-time fashion, and the exhaled tidal volume reaches the preset value through the feedback of
inhaled tidal volume. As the output of the ventilator is adjustable, so the compliance of the
breathing circuit system and the influence of the fresh airflow can be compensated.
Generally, under this mode, a constant flow is delivered during the inspiration phase, with the
pressure increasing in the lungs; the flow is exhaled fastly in the expiration phase, and the
pressure inside the lungs drops rapidly. By setting [Tpause] , the gas distribution in the patient‟s
lungs can be optimized.
When the airway pressure exceeds the limit pressure, the ventilator will provide alarms and switch
to expiration.
Setting [PEEP] (Positive End-Expiratory Pressure) to improve the expiratory-end CO2 discharge
and increase oxygenation during the respiratory process.
3-21
SIMV Circle
Paw
Trigger Window Out of Trigger Window
Ppeak Pplat Psupp
Peep
Trigger
Ti (Pressure Trigger)
Flow Exp%
Tpause
Trigger
No Ventilation(Flow Trigger)
In Trigger Window
Under the SIMV-VC mode, a preset tidal volume will be delivered to a patient at a preset [Rate]. In
the respiratory interval (trigger window), the patient conducts spontaneous respiration at patient‟s
respiratory rate and tidal volume.
The ventilator waits for the patient‟s spontaneous respiration as per the preset interval. [Trigger]
includes flow triggering or pressure triggering. If the spontaneous respiration reaches the threshold of
[Trigger] in the [Trig Window], the ventilator uses the preset tidal volume and inspiratory time to
deliver fresh gas synchronously, or it will conduct mechanical ventilation as per the preset clinical
parameter of [Rate].
Under this mode, spontaneous respiration can obtain the ventilator pressure to support ventilation.
Thus the patient can overcome the resistance in the patient‟s circuit system and the artificial airway, so
as to conduct ventilation with the preset support pressure.
3-22
3.10.2.3.3PCV Mode
Tresp
(Rate)
Paw Pinsp
Peep
Ti Te
Flow
Under the PCV mode, a constant inspiratory pressure will be provided. According to the preset [Rate] and
[I:E] , the inspiratory time can be calculated by the ventilator. The ventilator increases pressure for patient
side of the breathing circuit through a higher initial airflow, and reduce the airflow after the pressure reaches
the preset value in order to maintain the preset inspiratory pressure, until the respiratory time turns to the
expiratory time.
The ventilator pressure sensor monitors the airway pressure of the patient side of the breathing circuit in
real-time fashion. The ventilator maintains the preset pressure through the feedback of flow corresponding to
the pressure.
Under the PCV mode, setting [PEEP] also can improve the expiratory-end CO2 discharge and increase
oxygenation during the respiratory process.
Settings of PCV Mode:
Pinsp
Rate
I:E
Tslope
Plimit
PEEP
3-23
3.10.2.3.4SIMV-PC Mode
SIMV Circle
Paw
Trigger Window Out of Trigger Window
Pinsp Psupp
Peep
Trigger
Ti
(Pressure Trigger)
Flow Exp%
Trigger
No Ventilation(Flow Trigger)
In Trigger Window
Under the SIMV-PC mode, the ventilator use a preset [Pinsp] to conducts ventilation for a patient at a preset
[Rate]. In the respiratory interval (trigger window), the patient conducts spontaneous respiration at patient‟s
respiratory rate and tidal volume.
The ventilator waits for the patient‟s spontaneous respiration as per the preset interval. [Trigger] includes
flow triggering or pressure triggering. If the spontaneous respiration reaches the threshold of [Trigger] in the
[Trig Window] , the ventilator will use the preset inspiratory pressure and inspiratory time to deliver fresh air
synchronously, or it will conduct mechanica lventilation as per the preset clinical parameter of [Rate] .
Under this mode, spontaneous respiration can obtain the ventilator pressure to support ventilation. Thus the
patient can overcome the resistance in the patient‟s circuit system and the artificial airway, so as to conduct
ventilation with the preset support pressure.
Settings of SIMV-PC Mode:
Pinsp
Rate
Tinsp
Tslope
ΔPps
Plimit
PEEP
Trig Window
Trigger
Exp%
3-24
3.10.2.3.5PRVC Ventilation Mode
Paw VC PC
Pplat
Peep
Ti Te
Flow
25% Tpause
The PRVC mode is a controlled ventilation mode in which pressure regulates volume control.
In the PRVC mode, the first respiratory cycle is a trial ventilation for volume control, and trial
ventilation is performed according to preset [VT], [Rate], [I:E] and the default 25% inspiratory pause. The
measured plateau pressure during the inspiratory pause is used as the inspiratory pressure value for the next
respiratory ventilation cycle. Starting from the second respiratory cycle, the inspiratory pressure value,
[Tslope], [Rate], and [I:E] are used for pressure control ventilation. During ventilation, inspiratory pressure
will be automatically adjusted according to the characteristics of the patient's lungs, in order to achieve
transfer of the target tidal volume.
When the adjusted suction pressure reaches the preset [limit pressure], the system will provide alarm and
change to expiration.
The settings of pressure regulating volume control ventilation mode inlude:
VT
Rate
I:E
Plimit
PEEP
Tslope
3-25
3.10.2.3.6SIMV-PRVC Ventilation Mode
Peep
Trigger
Ti Te (Pressure Trigger)
Flow Exp%
25%Tpause
Trigger
(Flow Trigger)
No Ventilation
In Trigger Window
3-26
Trigger
Exp%
3.10.2.3.7CPAP/PSV Mode
Peep
Trigger
(Pressure Trigger)
Flow Exp%
Trigger
(Flow Trigger) No Ventilation
3-27
3.10.2.3.8PSVPro ventilation mode
Trigger
Ti
(Pressure Trigger)
Flow
Exp%
Trigger
No Ventilation(Flow Trigger)
In Trigger Window
3.13 Timer
The timer is seated in the right bottom corner of the ventilator interface. It is helpful for an anesthetist to
record duration of an operation or count time for some special operations during the operation. It is
operated by the control knob and touch control panel. The default state is OFF, as shown in the figure
below.
3-29
3.13.1 Start Timer
1. Press the control knob or tap [Timer] → [Start] to start the timer.
2. The timer starts counting, [Start] changes to [Stop] .
1. Press the control knob or tap [Timer] → [Stop] to stop the timer.
2. After the timer stops counting, the [Stop] changes to [Reset] .
1. Press the control knob or tap [Timer] → [Reset] to reset the timer.
2. After the limit value turns to “00: 00: 00”, the [Reset] changes to [Start] .
3-30
Fig. 3-14
The system can display the monitoring parameter in 2 ways: Large Font Interface and
Non-Large Font Interface. In the large font interface, only the parameter values are displayed;
but in the non-large font interface, both the parameter waveform and parameter values are
displayed.
The default interface of the system is the non-large font interfacet. Select [Config] →
[Setting] → [Big Font] → [On] / [OFF] to display the large font interface or the non-large
font interface.
The large font interface is shown as below.
3-31
Fig. 3-15
At the end of six consecutive respiratory cycles, if the pressure, flow, capacity, and VT
measurements are greater than the boundary, the system will automatically adjust the waveform at the
beginning of the next respiratory cycle.
At the end of six consecutive respiratory cycles, if the pressure, flow, capacity, and VT measurements
are less than the boundary minus the amplitude, the system will automatically adjust the waveform at the
beginning of the next respiratory cycle.
Select [Config] →[Screen] →[Waveform] menu,you can set the waveform style to [line] or
[fill].
Under the non-standby interface, you can monitor the airway pressure waveform and the
parameter values of airway peak pressure, platform pressure, positive end-expiratory pressure,
mean pressure.
The unit of the pressure parameter can be configured. Currently, the system provides three
units: [cmH2O], [kPa] and [mbar] , of which the [cmH2O] is the default unit. You can select
one unit by choosing [Config] → [Setting] → [Pressure Unit] .
3-32
3.14.5 Tidal Volume Monitoring
Under the non-standby interface, you can monitor the real-time flow waveform and expired
tidal volume, inspired tidal volume, per-minute volume.
The display of the flow waveform is optional. To enable/disable this display, you can select
[Config] → [Screen] → [Flow Wave] → [On] / [OFF] .
The display of the [VTinsp] is optional. To enable/disable this display, you can select [Config]
→ [Setting] → [VTi Display] → [On] / [OFF] .
Under the non-standby interface, you can monitor the real-time volume waveform.
The display of the volume waveform is optional. To enable/disable this display, you can select
[Config] → [Screen] → [Volume Wave] → [On] / [OFF] .
When the BIS module communicates with the anesthesia machine normally,and the BIS
display waveform switch is turned on.Under the non-standby interface, you can monitor the
real-time bis waveform.
The display of the volume waveform is optional. To enable/disable this display, you can select
[Config] → [Screen] → [BIS Wave] → [On] / [OFF] .
2. Set Oxygen Sensor to [On] / [OFF] under [O2 Sensor Monitor] menu as required.
3-33
Caution
When you use the oxygen sensor for the first time or replace the oxygen sensor, please check
whether the oxygen concentration monitoring is accurate. If the monitoring error is obvious,
please calibrate the sensor.
When the [O2 Sensor Monitor] is set to [OFF] , FiO2 will display a void value, and the sensor
cannot be calibrated; meanwhile, the oxygen concentration monitoring and the related
alarms of the sensor will be disabled.
When the [O2 Sensor Monitor] is set to [ON], and the Oxygen Monitoring Source is set to
[OFF] , FiO2 will display a void value, and the sensor cannot be calibrated; meanwhile, the
oxygen concentration monitoring and the related alarms of the sensor will be disabled.
3-34
3.16 Spirometry Loop
The spirometry loop reflects mechanical ventilation conditions and patient‟s lung function, such as
patient‟s compliance, circuit leakage condition, airway blocking, etc., which plays an important
role in the clinical test. 2 loops and related respiratory mechanics parameters of the reference loop
are displayed in the interface, which is shown as below:
E F
Fig. 3-16
2 loops can be displayed in the system loop interface, including 3 types of combinations: [P-V]
3-35
loop and [P-F] loop; [P-V] loop and [V-F] loop; [V-F] loop and [P-F] loop. You can switch to
display among the 3 pairs of loops by selecting [Loops] → [Select Loop] . Take [P-V] loop and
[P-F] loop as example, as shown in the figure below:
Fig. 3-17
Select [Loops] → [Save Loop] to save the loop. The saved reference loop will be displayed in
another color, and the related respiratory mechanics parameters of the reference loop will be
displayed on the right of the loop area. Take [P-V] loop and [P-F] loop as example, as shown
in the figure below:
3-36
3.17 Turn Off the System
Please follow the steps below to shut down when you do not intend to use the machine:
2. If the system switch is set to OFF “ ” in bag mode or in any mechanical ventilation mode,
the system will wait for 12s until it is fully turned off. During this 12-second shutdown delay,
the screen will display a 10-second countdown. Click [Shut down now] will immediately shut
down the ventilator. If the device is in mechanical ventilation, the ventilator will continue to
ventilate the patient in the current ventilation mode.
3. If the system switch is set to OFF “ ” in standby mode, the system will wait for 5s until it is
fully turned off. During this 5-second shutdown delay, the screen will display a 3-second
countdown. Click [Shut down now] will immediately shut down the ventilator.
4. When the user turns on the device during a shutdown delay, the countdown will disappear, and
the ventilator will return to its previous state.
5. Turn off the AC power switch. Unplug the power cord to disconnect the power.
6. Remove the gas source hose to disconnect the gas source.
3-37
Chapter 4 Tests Before Use
Warning
Prior to using the equipment, make sure to read the User’s Manual and understand the
operation and maintenance of all components.
If the equipment fails to pass the pre-use tests, do not use it,and contact the Company.
As to anesthesia gas delivery system, monitoring devices, alarm system and protective
devices intended to serve an anesthesia systems, they are no matter singly used or assembled
for an anesthesia machine, a checklist of the anesthesia system shall be submitted.
Attention
This guideline can be changed according to different situations of local clinical practice.
Such changes shall be subject to appropriate peer review.
It is recommended that you check whether the N2O blocking function and the O2-N2O ratio
are normal before using the device. Use an O2 concentration tester to monitor the
concentration of O2 in the output gas.
Before the anesthesia machine is used on the first patient every day
4-1
Systematic inspection √ √
Alarm test √ √ √
Gas supply pipeline and
√ √
gas cylinder test
Flow control system test √ √
Anesthesia vaporizer
√ √
installation test
Anesthesia vaporizer
√ √
back pressure test
Breathing system leak
√ √ √
test
Oxygen flush test √ √ √
AGSS transfer and
√ √
receiving system test
Vacuum suction device √ √
test
4.1.2 Before the Anesthesia Machine Used on the First Patient Every
Day
1. Verify that the required emergency equipment is kept ready and in good condition.
2. Conduct test and verify that the equipment is kept in good order and condition, and the
components are connected correctly.
3. Verify the connection of gas supply system, and verify that the gas cylinders are already
assembled, and the displayed pressure values are correct.
4. Verify the liquid anesthetic level inside the anesthesia vaporizer is appropriate, and check
that the anesthesia vaporizer fits its mounting perfectly.
5. Verify that the respiration circuit is connected correctly and intact.
6. Verify that there is enough renewed CO2 absorbent inside the CO2 canister.
7. Connect the scavenging system, and verify that the system works well.
8. Turn the system switch to “ON”, the system automatically performs a series of self-tests.
Verify that the system passes all the self-tests.
9. If the system passes the self-tests, perform “Automatic circuit leak testing” and “Manual
circuit leak testing”.
10. Make sure that the oxygen supply is adequate.
11. Select correct patient types like [adult] or [Child].
4-2
12. Start ventilation.
13. Set proper controlling values and alarm limits for the patient.
If you have finished tests as described in section 4.1.2 Before the Anesthesia machine is Used on
the First Patient Every Day, it is unnecessary to conduct this test, that is the test before the
machine is used on each patient, for the first patient..
1. Verify that the required emergency equipment is kept ready and in good condition.
2. Verify that the level of liquid anesthetic inside the anesthesia vaporizer is appropriate.
3. Verify that the respiration circuit is connected correctly and intact.
4. Verify that the breathing system is filled with enough absorbent.
5. Connect the vacuum suction device and verify that the system is working properly.
6. Conduct gas leak testing of the breathing system.
7. Turn off the APL valve (set it to 30cmH2O).
8. Start ventilation.
9. Set proper controlling values and alarm limits for the patient.
preventive maintenance
Please refer to 4.1.2 Before the Anesthesia machine used on the First Patient Every Day.
Warning
Make sure that the breathing system is connected properly and intact.
When installing the absorption canister, check whether the seal ring is installed correctly. If
the seal ring is not properly installed, leakage of the breathing system may occur.
Check the system and make sure that the following requirements are satisfied:
4-3
2. All components are connected correctly.
3. Check the flowmeter, anesthetic vaporizer, barometer, and gas supply pipes for damage.
4. Breathing system is connected properly, and the respiratory pipelines are intact, equip with
sufficient CO2 Pre-Pak absorbent or CO2 bulk absorbent. Manual ventilation device is
available and functional.
Caution
During the pipeline ventilation, do not set the backup cylinder valve to “ON”. Otherwise,
the gas cylinder may be exhausted and result in short supply in case the pipeline ventilation
gets faulty.
4-4
4.2.1.2 O2 Pipeline Test
1. If anesthesia machine is equipped with a backup cylinder, turn off the backup cylinder
valve. Connect the pipeline of O2 supply.
2. Set the system switch to the position “ON” ( ).
3. Rotate the electronic flow control panel to the medium level of measuring range.
4. Make sure that the pressure value indicated by O2 pipeline pressure gauges is in the range
of 280 to 600kPa.
5. Cut off O2 pipeline supply.
6. As O2 pressure drops, alarms” No O2 Pressure”.
7. Make sure that O2 pressure gauge returns to its zero position.
4.2.1.3 N2O Pipeline Test
To conduct N2O pipeline test, turn on O2 first.
Caution
To conduct N2O gas supply pipeline test, turn on O2 first, and make sure that the O2 gas
supply pressure ranges from 280 to 600kPa; Otherwise, N2O flow cannot be regulated.
When N2O pipeline supply is cut off, the system will not given alarms related to the N2O
pressure as N2O pressure drops.
1. If the anesthesia machine is equipped with a spare cylinder, close all spare cylinder valves. Access
to O2 pipe source and N2O pipe source.
2. Set the system switch to the ON “ ” position.
3. In the user interface, select the [Config] menu → Enter the [Flowmeter Config] tab. Set the flow
control to [Single Bar].
4. Set “balance gas” to “N2O” in the electronic flow control system.
5. Adjust the electronic flow control screen to adjust the flow control to the medium level of the
measurement range.
6. Check whether the pressure indication on the N2O pipe pressure gauge is within the range of
280-600kPa.
7. Make sure that the N2O pressure gauge returns to zero.
8. Disconnect the N2O pipe source.
Caution
When air pipeline supply is cut off, the system will not given alarms related to the air
pressure as air pressure drops.
If the values indicated by the backup cylinder pressure gauges decrease greater than 5000kPa
(725psi), it means that gas leak exists. Replace with a new gas cylinder washer as
described in 5.6 Assemble the Anesthesia Vaporizer. Repeat the steps 1~ 6 in section
4.2.2.2 High-pressure Gas Leak Test of O2 Cylinder. If gas leak still exists, do not use the
cylinder.
4-6
4.2.2.3 High-pressure Gas Leak Test of N2O Cylinders
Please conduct high-pressure gas leak test of N2O cylinders as per the steps described in4.2.2.2
high-pressure gas leak test of O2 gas cylinders. If the value indicated in the N2O high-pressure
pressure gauge decreases greater than 700kPa (100psi) in 1 minute, it indicates that gas leak
exists.
high-pressure gas leak test of O2 gas cylinders. If the value indicated in the air high-pressure
pressure gauge decreases greater than 700kPa (100psi) in 1 minute, it indicates that gas leak
exists.
Warning
If nitrous oxide exists and flows through the system during the test, it shall be collected and
removed using a safe and acceptable method.
Improper gas mixture can cause injury to the patient. Do not use this system if
oxygen-nitrous oxide proportioning system is unable to provide appropriate proportion of
oxygen and nitrous oxide.
Caution
When the electronic flow control system fails, the backup flow control system will be
enabled. The basic oxygen flow of the backup flow control system is 0 L/min. The backup
flow control system displays only one total flow meter. The total flow meter can display the
maximum flow of 15 L/min.
Slowly rotate the flow control knob of the backup flow control system. In order to avoid
damaging the control valve, do not rotate the flow control knob again when the reading on
the flow meter is out of range. When rotating the flow control knob clockwise to lower the
flow, the reading on the flow meter shall be up to 0 L/min before the knob reaches the
clockwise mechanical off position. Do not rotate again when the knob has reached the off
4-7
position. Rotate the flow control knob counterclockwise to increase the flow.
Slowly turn on the gas cylinder valve to prevent damage. Do not force to regulate the flow
control knob. After the spare gas cylinder testing, if spare gas cylinder does not used for gas
supply, turn off each gas cylinder valve.
1. Connect the pipe for gas supply or connect the gas cylinder. Turn on the gas cylinder valve to ensure
normal gas supply.
2. Set the system switch to on “ ” position, and the machine enters the standby host interface.
3. Select the [Config] menu → Enter the [Flowmeter config] label to set [Flow Control] to [Single
tube].
4. Set “balance gas” to “AIR” on the display of the full electronic flow meter.
5. Regulate the air flow. Ensure that the reading on the electronic flow meter is consistent with the set
value.
6. Set “balance gas” to “N2O”.
7. Gradually regulate the nitrous oxide flow, so as to ensure that the oxygen flow increases with the nitrous
oxide flow, and that the flows of oxygen and nitrous oxide meet the ratio of 1: 3.
8. Set the oxygen flow and nitrous oxide flow to 5 L/min.
9. Turn off the gas supply of oxygen pipeline and gas cylinder.
10. Press the rapid oxygenation button to release the internal pressure of the machine.
11. Confirm that the technical alarm of “insufficient oxygen supply pressure” shows up, and that the
display value of nitrous oxide flow and oxygen flow turns to zero.
12. Maintain the oxygen flow at 5 L/min. After confirming that the gas supply of the oxygen pipeline or
gas cylinder is turned on, the nitrous oxide flow appears and stabilizes at 5 L/min.
Note
After the use of backup flow control system, turn off all needle valves, then shut down the
backup flow control system.
When viewing the readings on a total flow meter, the line of sight shall be at the same level
as the float. In case of different perspectives, the same scale readings may be different.
1. Connect the pipe for gas supply or connect the gas cylinder. Turn on the gas cylinder valve to ensure
normal gas supply.
2. Set the system switch to on “ ” position.
4-8
3. Press the backup flow system‟s control button to ensure that the backup flow system can pop up
normally. Ensure that the backup flow system pops up in place, and that there is a prompt of “backup
flow meter is on” during the display interface switch of the full electronic flow meter.
4. After confirming that the backup flow control system pops up, visually check whether the flow display
of the total flow meter is 0 L/min.
5. Regulate the nitrous oxide needle valve. Gradually increase the nitrous oxide flow to confirm the
increase in total flow. Turn off the nitrous oxide needle valve to confirm that the total flow recovers to
0 L/min.
6. Regulate the air needle valve. Gradually increase the air flow to confirm that total flow can rise up to
greater than 10 L/min. Turn off the air needle valve.
7. Regulate the oxygen needle valve so that total flow is 2 L/min.
8. Regulate the nitrous oxide needle valve so that total flow is 8 L/min.
9. Disconnect the gas supply of oxygen pipeline and gas cylinder.
10. Press the rapid oxygenation button to release the internal pressure of the machine.
11. Confirm that the technical alarm of “insufficient oxygen supply pressure” shows up after the reading
on the total flow meter is gradually reduced to zero.
12. Confirm that the reading on the total flow meter restores to 8 L/min and the technical alarm of
“insufficient oxygen supply pressure” disappears after turning on the gas supply of oxygen pipeline or
gas cylinder.
13. Turn off the oxygen and nitrous oxide needle valves.
14. After confirming that all needle valves are turned off, select the backup flow meter shutdown button
on the display of the full electronic flow meter to confirm the normal retraction of the backup flow
meter.
Warning
Even if fresh gas contains enough oxygen, it may mix the low oxygen gas in breathing
system.
If N2O exists and flows through the system during the testing, the N2O gas shall be collected
and eliminated as per safe and acceptable methods.
Improper mixed gas may injure the patients. If the oxygen-N2O linked system cannot
4-9
provide well-proportioned O2 and N2O, the system shall not be used.
Caution
When backup gas cylinder testing is over, turn off all the gas cylinder valves if the backup
cylinders are not intended for gas supply.
Turn the gas flow switches slowly, and do not turn them forcibly when the maximum or
minimum flow range is exceeded to protect, the control valve from damage and to avoid
control failure. When flowmeter is adjusted to the minimum value, the reading shall be
zero.
Conduct the flow control system testing as per the following steps when O2 sensor is not
equiped with:
1 0.6 ≥0.2
2 1.5 ≥0.5
3 3.0 ≥1.0
4 7.5 ≥2.5
1 7.5 ≥2.5
2 3.0 ≥1.0
3 1.5 ≥0.5
4 0.6 ≥0.2
8. Cut off oxygen pipeline supply or turn off oxygen gas cylinder.
Caution
When O2 supply is cut off, alarms “No O2 Pressure” is given as O2 pressure drops.
Warning
During steps 6 and 7, the utilized oxygen sensor must be calibrated correctly, and the linked
system must be kept in its functional mode.
Adjust the testing control only (N2O described in step 6 and O2 described in step 7).
Adjust N2O before O2, and regulate the flows according to priority.
Rotate clockwise O2 and N2O flow control knobs respectively to the end (minimum flow).
4-11
Make sure that the O2 flow is increasing, and the measured O2 concentration must be equal to or
greater than 25% in the whole process.
Make sure that the N2O flow is increasing, and the measured O2 concentration must be equal to or
greater than 25% in the whole process.
8. Cut off the O2 pipeline supply or turn off the O2 cylinder valve.
9. Make sure:
10. Rotate clockwise all flow control knobs to the end (minimum flow).
11. Connect the O2 pipeline supply or turn on the O2 gas cylinder valve again.
12. Set the system to standby mode.
Warning
During testing, the anesthetic shall come from the fresh gas outlet. These agents shall be
discharged and collected as per safe and acceptable methods.
To avoid any damage, rotate clockwise the flow control knob to the end (minimum flow or
turn it off) prior to use.
1. Set the system switch to the position “ON”. An alarm might be given.
2. Set the O2 flow to 6L/min.
3. Make sure that the O2 flow is constant, and that the float of oxygen flow meter (for
AX-600) or main flowmeter (for AX-700/AX-700A/AX-800) can move freely..
4. Adjust the concentration of anesthesia vaporizer between 0 ~ 1%. The O2 flow must not
decrease greater than 1L/min in the whole process. If O2 flow decreases greater than
1L/min:
Replace the anesthesia vaporizer with a new one;
4-12
If O2 flow decreases less than 1L/min after the replacement, the old anesthesia
vaporizer is faulty.
If O2 flow decreases still greater than 1 L/min after the replacement, the anesthesia
machine system is faulty.
Caution
During alarm testing, stay in a place where you can observe the alarm lamps and alarm
prompts and hear the alarm sound.
Warning
In accordance with international laws and regulations, oxygen concentration shall be
monitored during the equipment is applied on a patient. If your equipment is not provided
with the said function, please use a monitoring instrument conforming to corresponding
international standards to monitor the oxygen concentration.
Caution
2. Take the O2 sensor out of breathing system and wait for 2~3 minutes; measure the indoor
air, and verify that the measured O2 concentration [FiO2] is approximately at 21%
3. Set the [Low Limit] of [FiO2] : On the screen, select [Alarm] menu → Access [ventilator]
→ Select [FiO2] [Low Limit] menu, and set the low alarm limit of the parameter to 50%.
4. Observe the alarm message area on the screen, make sure that [Low FiO2] is displayed.
4-13
5. Set the [Low Limit] of [FiO2] to a value lower than the current monitored value of [FiO2] ,
and make that the alarm of [Low FiO2] is cleared.
6. Re-install the O2 sensor into the breathing system.
7. Set the [High Limit] of O2 alarm: Select [Alarm] menu → Access [ventilator] → Select
[FiO2] [High Limit] menu, and set the high alarm limit of the parameter to 50%.
8. Connect the manual respiration manual bag to corresponding connector of breathing
system. Push the oxygen flush button to fill the manual/spontaneous manual bag, an
d make sure that the O2 concentration [FiO2] measured by sensor is of approximate
ly 100%.
9. Observe the physiological alarm message on the screen, and make sure that [High FiO2] is
displayed.
10. Set the [High Limit] of [FiO2] alarm to 100%, and make sure that [High FiO2] is cleared.
3. When the MV is lower than the low alarm limit, observe the alarm message area on the
screen, and make sure that [Low MV] is displayed.
4. Set the [High Limit] alarm of [MV] : On the screen, select [Alarm] menu → Access
[ventilator] → Select [MV] [High Limit] menu, and set the high alarm limit of the
parameter to 9.0L/min.
5. When the MV is higher than the high alarm limit, observe the alarm message area on the
screen, and make sure that [High MV] is displayed.
1. Connect the manual respiration manual bag to the corresponding connector of the breathing
system.
3. Rotate the APL valve control knob to the position with the minimum opening pressure.
4. Pinch the manual respiratory manual bag, and make sure that one complete respiratory
cycle takes place.
5. Stop pinching the manual respiratory manual bag,and wait for at least 20± 3 seconds. Make
sure that [Apnea] alarm is displayed on the screen.
6. Pinch the manual respiratory manual bag for several times, and make sure that the [Apnea]
4-14
alarm displayed on the screen disappears.
1. Connect the manual respiratory manual bag to the corresponding connector of the breathing
system.
2. Rotate the O2 flow control knob to the low limit.
3. Rotate the APL valve control knob to the position of 30cmH2O.
4. Set the Manual / Mechanical Control switch to its position “Manual” ( )
5. Push and hold the oxygen flush button for approximately 15 seconds, and make sure that
[Continuous Pressure] alarm is displayed on the screen.
6. Turn on the patient-end outlet, and make sure that [Continuous Pressure] alarm displayed
on the screen disappears.
1. Set the Manual / Mechanical Control switch to its position “Mechanical Control “( )”.
2. On the screen, select [Alarm] menu → access [Ventilator] →Select [Ppeak] [High limit]
menu, and set the [Low Limit] alarm limit of the parameter to 2cmH2O.
3. Remove the manual respiratory leather-bag from the Y-shaped patient-end port.
4. Wait for 20 seconds, observe the alarm message area on the screen, and make sure that
[Low Paw] alarm is displayedon the screen.
5. Connect the manual respiration manual bag to manual respiration leather-bag port on the
breathing system.
6. Make sure that [Low Paw] displayed on the screen disappears.
4. Make sure that the alarm [High FiCO2] / [High EtCO2] are displayed on the screen when the
concentration of inspired CO2 / expiratory-end CO2 are higher than the alarm limit respectively.
5. Set the [Low Limit] menu of [FiCO2] [ETCO2] alarms of [CO2] to 10 mmHg.
6. Set the [Low Limit] of CO2 to a value higher than the standard gas concentration.
7. Make sure that the alarms [Low FiCO2] / [Low EtCO2] are displayed on the screen when the
concentration of FiCO2/EtCO2 are lower than the alarm limit respectively.
Warning
Foreign objects inside the breathing system may block up the gas flow to the patient, and
may result in a casualty accident. Make sure that no testing plugs or other foreign objects
exist inside the breathing system.
The resistance at 2,5, 15 and 30 l/min, and compliance of those breathing accessories,please
refer to the attached specification for details.
The range of internal volume of any Anesthetic breathing system less than 3.5L.
Breathing system shall be equipped with a ventilator conforming to ISO 80601-2-13 and YY
0635.4.
1. Make sure that the breathing system is connected properly and is kept in good condition.
Once the breathing system is disconnected, the anesthesia machine can give an alarm “No
Breathing System”.
2. Make sure that the check valves in the breathing system work fine.
If the inspiratory check valve turns on during inspiration, and immediately turns off when
expiration begins, it indicates that the inspiratory check valve (unidirectional valve) works
fine.
If the expiratory check valve turns on during expiration, and immediately turns off when
inspiration begins, it indicates that the expiratory check valve (unidirectional valve) works
fine.
4-16
4.5.1 Bellows Tightness Test
1. Set the system to standby mode.
2. Set the Manual/mechanical ventilation switch to the position “Mechanical” ( ).
3. Rotate all flow control knobs to the minimum gas flow.
4. Block up the patient-end outlet, and close the breathing system.
5. Push the oxygen flush button to have the folded sack of bellows risen to its top end.
6. Make sure that the pressure indicated in the airway pressure gauge must not exceed
15cmH2O.
7. The folded sack of bellows shall not fall down. If it falls down,it indicates that gas leak
exists in the bellows. Re-install the bellows.
Caution
System leak test includes the leak test of anesthesia breathing system and anesthesia
ventilator.
To conduct system gas leak test, make sure that the breathing system is connected correctly,
and the respiratory pipelines are kept in good condition.
Caution
The progressive gas leak testing can be terminated if you push [Stop] button. That does not
mean the system gas leak testing fails, only means that the testing gets invalid.
If gas leak testing fails, check all possible causes of gas leak, such as leakage from bellows,
breathing system pipeline, CO2 canister and other connecting devices. During the check of
CO2 canister, pay attention to the seal components of canister to find if any CO2 absorbent
particles are attached on the canister, and remove them if any.
If leaks exist in the breathing system, do not use the equipment. Contact in time the
equipment service personnel or after-service department of the Company.
Loose connection between the bellows and the intubation tube will result in leakage of the
breathing circuit, and will affect the VT supply anomaly of the anesthesia machine.
3. Connect the manual respiration manual bag to corresponding connector on the breathing
system.
4. Rotate the APL valve control knob to the position of its maximum value (75cmH2O).
5. Rotate the flow control knob to turn off the O2, N2O and air flow completely.
6. Insert the Y-piece of corrugated pipe into the leak testing plug of Manual/spontaneous
leather-bag port to block up the gas outlet of Y-piece.
7. Push the oxygen flush button to allow the value indicated by the airway gauge rise to
approximately 30cmH2O.
8. Release the oxygen flush button, and select [Leak Test] menu → [Leak in bag mode] .
9. Push the [Start] button. The system begins manual circuit leak test and displays
simultaneously the prompting message: [Testing is Performing] .
10. If the equipment passes the test, the system displays a prompting message [Leak Test
PASS] .Otherwise, it displays a prompting message: [Leak Test FAIL ] . In such a case,
check the connection of breathing system and condition of pipelines. Conduct leak test
again when problems are solved. If gas leak still exists, contact the equipment maintenance
personnel of After-service Department of the Company.
4-18
11. Leaks may also be verified by observing the readings indicated by the airway pressure
gauge during testing. If the readings drop,it indicates that gas leak exists.
3. Connect the manual respiration manual bag to the corresponding connector on the
respiratory circuit.
4. Insert the Y-piece of corrugated pipe into the leak testing plug of Manual/spontaneous
leather-bag port to block up the gas outlet of Y-piece.
6. Push the oxygen flush button to fully fill the manual/spontaneous manual bag.
7. Make sure that the readings indicated by airway gauge are in the range from 20 to
40cmH2O.
8. Rotate the APL valve control knob to the position of the minimum value for the opening
pressure of APL valve (position MIN).
10. Make sure that the reading indicated by the airway pressure gauge is less than 5cmH2O.
11. Push the oxygen flush button, and make sure that the reading indicated by airway pressure
gauge does not exceed 10cmH2O.
12. Rotate the O2 flow control knob to the minimum value, and verify that the reading indicated
by theairway pressure gauge does not drop to below 0cmH2O.
6. Start ventilation.
7. Check whether or not the respiration check valve moves in the open-close cycle. If not, the
check valve is faulty.
4-19
4.6 Ventilator Test
Caution
Ventilator shall be equipped with an anesthesia system conforming to IEC 80601-2-13 and
IEC 60601-2-13(GB 9706.29).
2. Make sure that the relevant parameters and alarm limits of ventilator are properly set. For
specific settings, refer to the chapter “15.11 Principle and Parameter Specifications of the
Ventilator”.
5. Set the parameters like different tidal volumes, respiratory rates and inspiratory/expiratory
ratios of anesthesia machine. Observe the monitored value and set values of the anesthesia
machine, and check whether or not the actual tidal volumes of bellows hood of the
breathing system can meet the clinical requirements.
Caution
Do not block up the pressure compensation port of AGSS during the test.
If the floater cannot rises up, possible causes include the following:
Floater adhesion. Turn the AGSS upside down, and check whether or not the floater may
move up and down freely.
Floater rises slowly. Filtering net is possibly blocked Contact the manufacturer for checks
and maintenance.
The high-flow AGSS transfer and receiving system is not working or the pumping flow rate
is less than 50L / min (normal working rate). Contact the manufacturer for inspection and
repair.
The low-flow AGSS transfer and receiving system is not working or the pumping flow rate
is less than 25L / min (normal working rate) . Contact the manufacturer for inspection and
repair.
4-20
4.7.1 Connection Leakage Test for AGSS and the Exhaust Gas Outlet
of Anesthesia Machine
1. Remove the rear cover of the host,and remove internal corrugated tube that is connectod to the
air-capacitor.
2. Connect the tubing sets to be tested onto the inlet of air-capacitor. Connect the pressure gauge.
3. Remove the corrugated tube connected on the AGSS, and then block this port so that the corrugated
tube can be connected onto the exhaust gas outlet.
4. Ventilate 10 ±0.5ml of air per minute into the tubing sets to be tested. The leakage amount shall be
no more than 100 ml/min in the transfer and receiving system.
5. If the leakage amount exceeds the above value, re-connect the tubing sets of the exhaust gas outlet,
and then retest according to the above steps.
Caution
Before use must check ensure that the vacuum suction system is qualified.
For the safety and health of patients and others,the negative pressure generator switch
should be at the OFF position before opening the power gas source(or inserting into the
socket hole of the terminal),the negative pressure regulator mode selection switch should be
set to the middle position(OFF position) and the adjustment knob should turn
counterclockwise to the zero position.
1. Assemble the external pipe collection system with internal negative pressure to the anesthesia
machine according to the installation instructions;
2. Turn the negative-pressure gear switch to the OFF position;
3. Rotate the negative-pressure adjustment knob counterclockwise until it can no longer be rotated;
4. Open the O2 source, and confirm whether the gas source pressure is within the applicable range of
the anesthesia machine (280-600kPa);
5. Block the patient end of the suction tube, and turn the negative-pressure gear switch to the FULL
4-21
position; observe whether the reading on the negative pressure gauge can reach 60kPa or above;
6. If there is no reading on the negative pressure gauge, check whether the collecting liquidbottle,
suction tube, overflow bottle, and filter are not installed well, or check for damage and gas leakage;
7. After confirming that the gas circuit connection is intact, turn the negative-pressure gear switch to
the OFF position, and observe whether the pressure on the negative pressure gauge has returned to
zero;
8. Turn the negative-pressure gear switch to the REG position; slowly rotate the negative-pressure
knob clockwise; observe the pressure gauge; confirm whether the pressure can be continuously
adjusted and stabilized at a certain pressure;
9. After completing the inspection, turn the negative-pressure adjustment knob counterclockwise to
adjust the negative pressure to the minimum;
10. Turn the negative-pressure gear switch to the OFF position to avoid gas waste.
1. Assemble external negative pressure to the anesthesia machine according to the installation
instructions;
2. Press the driving gas switch to the OFF position; turn the negative-pressure gear switch to the OFF
position as well;
3. Rotate the negative-pressure adjustment knob counterclockwise until it can no longer be rotated;
4. Open the O2 source, and confirm whether the gas source pressure is within the applicable range of
the external negative-pressure system (280-550kPa);
5. Pull the driving gas switch to the ON position;
6. Block the patient end of the suction tube, and turn the negative-pressure gear switch to the FULL
position; observe whether the reading on the negative pressure gauge can reach 60kPa or above;
7. If there is no reading on the negative pressure gauge, check whether the collecting liquid bottle,
suction tube, overflow bottle, and filter are not installed well, or check for damage and gas leakage;
8. After confirming that the gas circuit connection is intact, turn the negative-pressure gear switch to
the OFF position, and observe whether the pressure on the negative pressure gauge has returned to
zero;
9. Turn the negative-pressure gear switch to the REG position; slowly rotate the negative-pressure
knob clockwise; observe the pressure gauge; confirm whether the pressure can be continuously
adjusted and stabilized at a certain pressure;
10. After completing the inspection, turn the negative-pressure adjustment knob counterclockwise to
adjust the negative pressure to the minimum;
4-22
11. Turn the negative-pressure gear switch to the OFF position, and press the negative-pressure driving
gas switch to the OFF position, so as to avoid gas waste.
Note
Refer to the user manual supplied with the external negative suction system for testing
information.
The vacuum suction system must be inspected before use, accordint to the requirement
specified in the user manual supplied with the vacuum suction system.
Caution
During the test, observe the drive gas discharge vent behind the anesthesia machine, so as to
ensure that the discharge vent is unobstructed.
4-23
Chapter 5 Installation and Connection
Warning
If electrosurgical equipment is used, keep their leads away from the breathing system,
oxygen sensor and other components of the anesthesia machine, make sure that the standby
manual/spontaneous equipment of anesthesia machine are ready for use, and ensure that
masked simple respirator are available in case the electrosurgical equipment prevents safe
use of the ventilator. In addition, ensure that all life supporting and monitoring equipment
may be correctly operated.
The equipment is provided with a waste gas exhaust port. The users shall pay attention to
the disposal of residual breathing gase scavenged.
After the absorbent gets dry, it may pose a danger to the patient if it continues to be used.
Appropriate precautions shall be taken to ensure that the soda lime in the CO2 absorption
canister is not dry. After each use of the system, all gas sources shall be turned off in time.
The anesthesia system has an exhaust outlet. During use, pay attention to the disposal of
discharged respiratory residual gas.
Caution
After the use of the equipment, pay attention to the disposal of the breathing system, the test
of the CO2 absorbent inside Canister (carbon dioxide absorbent) and anesthetics inside the
Anesthesia Vaporizer so as to guarantee normal running of the equipment.
Please do not weigh down the manual support column by hands or by hanging other heavy
objects onto it.
5-1
If the difference between airway pressure gauge reading and the parameter value displayed
on the screen is large, contact the Company.
11
10 12
9 13
8 14
7
6 15
4
3
2 1
17 18 19 20 21
16
5-2
3 Canister release device 14 Inspiratory check valve
4 Manual drain valve 15 Inspiratory port
5 Leak test plug 16 Circuit Lock Hook
6 Breathing tube hook 17 Pressure Sampling Port
7 Expiratory check valve 18 Waste Gas Outlet
8 APL (Adjustable Pressure Limiting) Valve 19 Fresh Gas Inlet
9 Airwany pressure gauge 20 Driving Gas Inlet
10 Manual bag port 21 Guide Post Hole
11 Bellows assembly
7 6
1 2 3 4 5
5-3
2. Push the breathing circuit system into the circuit adapter forcibly such
that the breathing circuit system is connected to the circuit adapter
without gap. Verify that the breathing circuit system has been locked.
Warning
When breathing circuit system is assembled onto the circuit adapter, you must verify that the
breathing circuit system is firmly locked. If not, it may be separated from the circuit adapter
during operation, resulting in severe leak of fresh gas and mismeasurement of tidal volumes.
Caution
If it is very hard to push the breathing circuit system in, check whether or not the nuts on the
downside of the breathing circuit system are tightened. The nuts may get stuck on the top of
AGSS if they are not tightened.
If it is very hard to push in or take out the breathing circuit system, it is necessary to apply
small quantity of lubricating oil (du pont Krytox high-performance fluorine grease) onto the
seal rings of air ports of the circuit adapter.
5-4
2 Assemble the manual support column connector to the breathing
system connector, as shown in the right figure.
1. Fit on the manual bag upward, and screw it onto the manual support column.
2 Align the bayonet of bellows cover with the slots on the breathing system. Press downward the
bellows cover to the end. Hold the outer side of bellows cover by both hands, and screw down
it clockwise. as shown in the right figure:
Warning
Before assembling the bellows cover, check whether or not the seal components of breathing
system are normal. If any backing-off or warpage is found, assemble the seal components
properly before bellows cover is assembled.
5-6
5.1.5 Assemble the Flow Sensor
1. Ensure that the direction of arrow marked on the flow sensor is same to that on the breathing
system, and the side with arrows is facing upward, as shown in the figure below:
3. The breathing joint slot is aligned with the upper and lower slot of the flow sensor,as show in the
figure.
4. Align the breathing joint and the lock nut to the flow sensor interface and tighten the lock nut
clockwise.
5-7
4
Warning
When flow sensor is assembled, tighten the respiration connector rotary-cap locknut;
otherwise flow sensor measurement may be disabled.
Caution
To assemble the breathing tube, hold the connectors at both ends of the breathing tube so as
not damage the breathing tube.
2 Assemble the expiratory tube and inspiratory tube onto the expiratory port and inspiratory port
on the breathing system respectively.
Warning
5-8
Before assembling the oxygen sensor, check if the seal rings of oxygen sensor is in good
condition. Replace the oxygen sensor with a new one if no seal rings is installed or the seal is
damaged.
The O2 sensor unpacking combination must be correct,screwed in place and not skew.
Oxygen sensor must be assembled properly; otherwise gas leakage may occur in the
breathing system.
1. Align the oxygen sensor with the oxygen sensor port “ ” on the breathing system, and insert
it into the port and assemble it securely.
2. Insert one end of oxygen sensor cable into the jack of oxygen sensor.
3. Insert the other end of oxygen sensor cable into corresponding oxygen sensor port “ ” on
the main machine, as shown in the right figure:
1. Before assembling the airway pressure gauge, please check whether the slot is unlocked. If locked, press
the buckle to unlock the slot before proceeding to the next step.
5-9
2. Directly insert the airway pressure gauge into the buckle of CPC connector. The airway pressure gauge
is securely assembled if a sound “De” is heard.
Warning
Please observe the following applied provisions for safety protection:
Use of desiccated CO2 absorbent may endanger the patients. Proper preventive measures
shall be taken to guarantee that the CO2 absorbent inside the Canister (carbon dioxide
absorbent) may not get dry. All gas supplies shall be turned off every time when finished
using the system.
Disposable Canister (carbon dioxide absorbent) is a sealed devices, and may not be opened
or refilled with CO2 absorbent.
Do not allow your skin or eyes to be exposed to substance contained inside the CO2
absorbent canister. In case skin or eyes are exposed to the substances, rinse the affected
parts with fresh water immediately, and take medical treatment.
Be sure to assemble and lock up the Canister (carbon dioxide absorbent) properly;
Otherwise, the patient may inhale repeatedly the carbon dioxide he gives off.
5-10
the operation guide and precautions, see the attached specification sheets for attachments.
Before assembling a Canister (carbon dioxide absorbent), check the color of CO2 absorbent
inside the Canister (carbon dioxide absorbent) so as to determine whether or not to change
the CO2 absorbent first.
Every time a case is finished or during operation, check the color of CO2 absorbent, and
take corresponding treatment measures. For details of changes in color of CO2 absorbent,
refer to the label attached on the package of CO2 absorbent. Color of the CO2 absorbent
may possibly restore to its original color during the period of time when it is not in use.
Please take appropriate preventive measures to ensure that the CO2 absorbent inside
canister may not get dry. All gas supplies shall be turned off in time when finished using the
system. If thoroughly dry CO2 absorbent is exposed to anesthetics, it may release carbon
monoxide (CO), and its continuing in service may do harm to the patients. Replace CO2
absorbent in time for the safety of patients.
Please clean CO2 absorbent and replace the sponge of Canister (carbon dioxide absorbent)
regularly; Otherwise CO2 absorbent powder settled inside the Canister (carbon dioxide
absorbent) may go into the breathing system.
Please clean the Canister (carbon dioxide absorbent) rim regularly. Otherwise the CO2
absorbent particles sticking on the rim may result in leakage in the breathing system.
To assemble the CO2 absorbent, check Canister (carbon dioxide absorbent) rim, strutting
piece and seal for attached CO2 absorbent particles. If any, remove the particles; otherwise
they might result in leakage in the breathing system.
Caution
Gradual color change of absorbent inside the canister indicates that absorption of carbon
dioxide. The color change of absorbent is only a rough indication. It is advisable to
determine when to replace the absorbent by carbon dioxide concentration monitoring.
Discolored absorbent shall be discarded. If left standing for several hours, it may regain its
original color giving a misleading indication.
The breathing system of anesthesia machine includes the self closed- circuit system and
non-closed- circuit system. The difference between them is that the former is equipped with
the Bypass function.
5-11
1 Check whether the support and rim of the CO2 absorbent
canister,are attached with absorbent particles or powder. If any,
please remove them. The filled CO2 absorbent cannot be higher
than the “-MAX-” mark on the CO2 absorbent canister.
2 Pinch the CO2 absorbent canister lock catch with the left hand
3 Align the CO2 absorbent canister with the mounting slot of the
canister bracket, as shown in the right figure.
4 Push the CO2 absorbent canister to the end of the mounting slot
of the canister bracket until it is fixed, as shown in the right
5-12
figure:
5-13
5.3 Replace the Canister (carbon dioxide absorbent)
As the breathing system includes the closed-circuit system with the Bypass structure, so the gas
won‟t leak out to the atmospheric air during the process of replacing CO2 absorbent canister. But
be sure to timely replace the CO2 absorbent and install the CO2 canister to prevent CO2 retention.
1. Push the release device of CO2 absorbent canister to remove the CO2 absorbent canister, as shown
in the figure below.
2. The filled CO2 absorbent cannot be higher than the “-max-” mark on the CO2 absorbent canister.
Check whether the CO2 canister strutting piece, seal ring and the rim are attached with CO2
absorbent particles or powder. If any, please remove them. Push the CO2 absorbene canister to the
end of the mounting slot until it is fixed , as shown in the right figure:
3. Lift up the handle of the bracket until it is locked, as shown in the right figure:
5-14
5.4 Replace CO2 Absorbent
Caution
Gradual color change of absorbent inside the canister indicates that absorption of carbon
dioxide. The color change of absorbent is only a rough indication. It is advisable to
determine when to replace the absorbent by carbon dioxide concentration monitoring.
Discolored absorbent shall be discarded as per local correlative laws and Regulations or
waste disposal system of the hospital. If left standing for several hours, it may regain its
original color giving a misleading indication. To avoid the misleading effect, we suggest that
CO2 absorbent shall be replaced by new one prior to each operation, or carbon dioxide
monitor may be used.
1. Disassemble the CO2 absorbent canister, referring to the reverse process of CO2 absorbent
canister assembling in the section 5.2 of this chapter.
2. Pinch the lock catch of the CO2 absorbent canister with your left hand and rotate it rightward,
as shown in the flag, open the canister bracket and take it out with your right hand .
3. Take out the discolored CO2 absorbent.
4. Fill the new CO2 absorbent into the CO2 absorbent canister along the inside periphery to
prevent absorbent entering the ventilation hole of the strutting piece, otherwise, it may
increase airway resistance.
5. Check whether the CO2 canister strutting piece, seal ring and the rim are attached with CO2
absorbent particles or powder. If any, please remove it. The placed CO2 absorbent cannot be
more than the “-max-” mark on the CO2 absorbent canister.
6. Please refer to the procedure in the section 5.2 of this chapter to reassemble the CO2 absorbent
canister.
Warning
To remount Canister (carbon dioxide absorbent) after replacement of the absorbent
(carbon dioxide absorbent), be sure to check Canister (carbon dioxide absorbent) for
locking so as to ensure proper assembling.
5-15
Caution
The CO2 absorbent must not exceed the “-MAX-” level marked on the Canister (carbon
dioxide absorbent)
O2
O2 and N2O
O2 and AIR
Warning
Only medical-use gas supplies are allowed to used. Other types of gas supplies might
contain water, oil or other contaminants.
If the central gas supply system develops a fault, one or more equipments connected to it
may quit work. In such a case,turn on the standby cylinders to make sure normal work of
the anesthesia machine.
When gas supplies are turned off, pressure still exists inside the pipelines. Therefore, release
the gases from the pipelines before you unplug the gas pipes.
1 To connect the gas supply pipeline, check whether or not the seal rings of connectors for
good condition.If the seal rings are damaged, the pipeline cannot be used, and seal rings
must be replaced; otherwise gas leak may take place.
5-16
2 Align the gas supply hose and connectors with the gas-supply inlet on the back of
anesthesia machine and insert it.
3 Ensure that the gas supply hoses are securely connected to the gas supply inlets, and
tighten the hose nuts by hand.
Caution
The gas supply hoses shall meet the standards of ISO 5359 and YY/T 0799.
The hose connectors shall meet the standards of ISO 9170-1 and YY 0801.1.
There are two exhaust components, located on the left and the rear of the workbench, respectively.
There are two exhaust ports, which are the exhaust port of the AGSS and PEEP generating device
and the exhaust port with internalnegative-pressure driving gas.
For waste gas exhaust, the following shall be adopted:
1 PEEP exhaust port and internal vacuum suction vent, which may directly discharge
oxygen gases indoors.
2 Outside diameter of AGSS connector is 30mm, with a taper of 1:20. Please connect
Anesthetic gas scavenging system or passive waste gas disposal system.
Warning
The PEEP exhaust port may continuously discharge small quantity of oxygen. Never block
the outlet; otherwise the anesthesia ventilator cannot work.
If your anesthesia machine is not equipped with active AGSS, please do not connect the
waste gas exhaust port of anesthesia machine to the active waste gas disposal system of the
hospital.
5-17
Warning
If the vaporizer is incompatible with the anesthesia machine, their performance may be
degraded. Please use the vaporizer matching the equipment.
Please use an anesthesia vaporizer that complies with the ISO 80601-2-13 standard. For
installation, addition, discharge, and other information about the anesthesia vaporizer, see
the instructions of the manufacturer of the anesthesia vaporizer.
The position of the anesthesia vaporizer between “0” and the minimum scale above “0” is
not available and may cause accidental injury to the patient.
Care should be taken to lift and operate the anesthesia vaporizer during installation,as the
weight of the anesthesia vaporizer may be greater than expected,depending on the size of the
anesthesia vaporizer.
Only Selectatec series vaporizers can be used. To conduct testing, ensure that the anesthesia
vaporizer is already locked up.
Do not remove the locked anesthesia vaporizer from the anesthesia machine.
The anesthetic vaporizer cannot be used if it is set between “0” and “ON”.
Caution
Atmospheric pressure may differ from the calibration pressure of the anesthesia
vaporizer,which may lead to inaccurate anesthetic output.During the use of the anesthetic
system, the operator should continuously monitor the anesthetic concentration to confirm
the accurancy of the output concentration.
If the top of anesthesia vaporizer is not horizontal, remove the anesthesia vaporizers and
reassemble it. If the anesthesia vaporizer cannot be set horizontally onto the vaporizer base,
5-18
do not use the system.
Lift each anesthesia vaporizer upward in so far as possible such that it may be separated
from the vaporizer base. However, do not pull it forward. Be careful! Do not allow the
anesthesia vaporizer to rotate on the vaporizer base.
1. Hang the Anesthesia Vaporizer onto the vaporizer base of the anesthesia machine, and
ensure that the Anesthesia Vaporizer completely fits the vaporizer base without gap, as
shown in the figure below:
2. Turn clockwise the locking bar to fix the Anesthesia Vaporizer onto the vaporizer base,
as shown in the figure below:
3. Ensure that top of the Anesthesia Vaporizer is horizontal. If not, remove the vaporizer
and reinstall it.
4. To reinstall it, lift up each Anesthesia Vaporizer vertically (90 degrees) such that it is
separated from the vaporizer base, but do not pull it forward. Be careful not to allow
the Anesthesia Vaporizer to rotate on the vaporizer base.
5. When the Anesthesia Vaporizer is separated from the vaporizer base, reinstall the
5-19
vaporizer and perform steps 1~3. If the Anesthesia Vaporizer cannot be positioned
horizontally on the vaporizer base, do not use the system.
7. Test each possible combination. If one or more anesthesia vaporizers can be turned on
simultaneously, remove and reinstall the anesthesia vaporizers , and perform steps 1 to
6.
Warning
Ensure that anesthetics are filled correctly. Anesthetic names are already indicated on the
vaporizers, and they are also marked with different colors. If anesthetics are incorrectly
filled, the actual output concentration of anesthetics may be changed.
Warning
Anesthetics drained from the vaporizer must not be reused, and it shall be disposed as
hazardous chemicals.
Please mark the bottles containing the drained anesthetics as follows: used anesthetics.
Warning
Using no gasket or using more than one gasket may cause leakage.
Do not leave the spare cylinder valve open while using the pipe for gas supply.
Otherwise, in the event of a pipe gas supply failure, gas supply to the gas cylinder may
be exhausted, resulting in insufficient reserve supply.
5-20
5.7.1 Gas Cylinder (1)
T-type handle
5-21
Washer
5. Ensure gas cylinder outlet is kept away from all articles that may be damaged by released
high pressure gas.
6. Turn on and off the gas cylinder valve quickly to remove the dust on the gas cylinder outlet.
10. Test the gas cylinders. For specific steps, refer to “4.2.1 Gas Supply Pipeline Test”.
5-22
4. Open the buckle.
6. Remove the screw cap on the valve of the new gas cylinder.
7. Ensure that the gas cylinder outlet is kept away from any articles that may be damaged by
released high pressure gas.
8. Turn on and off the gas cylinder valve quickly to remove the dust on the gas cylinder outlet.
12. Assemble gas cylinder plugs and composite sealing washers to all gas cylinder buckles.
13. Perform high pressure leak testing. For specific steps, refer to “4.2.1 Gas Supply Pipeline
Test”.
5-23
5.8 Assemble Module
2. Push the module into the place until the lever at its
bottom gives a “Click” sound, indicating that the
module is fixed properly.
5-24
4
5. Insert the other end of the exhaust tube into the exhaust
inlet on the anesthesia machine. A “Click” sound
indicates the tube is installed in place.
5-25
5.8.2 Assemble Sidestream CO2 Module
For details,,please refer to the section of " 5.8.1 Assemble Sidestream AG Module ".
For details, refer to the section of " 5.8.1 Assemble Sidestream AG Module ".
2. Push the module into the place until the lever at its bottom gives a “Click” sound, indicating
that the module is fixed properly. The indicator light on the module turns on, indicating the
module is installed properly.
5-26
3. According to the triangular mark of the connector, insert one end of the connector into the
sample port.
5-27
2. Install the module into the module slot.
3. Push the module into the place until the lever at its bottom gives a “Click” sound, indicating
that the module is fixed properly. The indicator light on the module turns on, indicating the
module is installed properly.
5-28
4. Insert one end of the exhaust tube into the exhaust outlet on the module, and rotate it clockwise
to tighten it.
5. Insert the other end of the exhaust tube into the exhaust inlet on the anesthesia machine. A
“Click” sound indicates the tube is installed in place.
6. Insert one end of the sampling tube into the sampling tube port, then rotate it clockwise to tighten
it.
5-29
5.8.6 Assemble Artema Sidestream AG+O2 Module
For the detailed installation procedure, please refer to 5.8.5 Assemble ARTEMA Sidestream AG
Module.
5-30
5.8.7 Assemble BIS module
1. Firstly paste the BIS sensor (as instructed in the attached document of the BIS sensor) onto the
patient's head.
2. Plug the sensor connector into the BISx device cable connector.
3. Plug the BISx device cable connector into the BIS module connector,
as shown in the figure:
4. Push the module into the place until the lever at its bottom gives a “Click” sound, indicating that
the module is fixed properly.
5. The indicator light on the module turns on, indicating the module is installed properly.
5-31
5.8.8 Disassemble Sidestream AG Module
5-32
3
4. Push the tab at the bottom of the module upwards, and pull
out the module.
For details, please refer to the section of " 5.8.8 Disassemble Sidestream AG Module " .
For details, please refer to the section of 5.8.8 Disassemble Sidestream AG Module.
5-33
5.8.11 Disassemble Mainstream CO2 Module
1. Pull the connector cable out of the sampling port.
2. Push the tab at the bottom of the module upwards, and pull out the module.
2. Press down the clip at the exhaust outlet of the anesthesia machine to pop out the exhaust tube,
and then remove the tube, as shown in the tight figure:
5-34
2
3. Rotate the tightening knob of the exhaust tube counterclockwise , and then pull out the tube.
4. Push the tab at the bottom of the module upwards, and pull out the module.
5-35
5. Push the buckle upwards as shown in the figure to remove the watertrap.
Please refer to 5.8.12 Disassemble Artema Sidestream AG Module for detailed disassembling steps.
5-36
2. Pull the BIS module cable connector out of the BIS module connector,
as shown in the figure:
3. Press and hold the circular button on the patient interface cable with right hand.
Pull out the sensor connector with left hand, as shown in the figure:
1. Place the collecting bottle in the bracket.Connect the suction and the collecting bottle according to the
mark on the collecting bottle.
5-37
2. As shown in the figure,rotate the nut,install the overflow cup,and insert the suction tube into the
overflow protection interface.
overflow protect-i
on interface
3. Connect the suction tube to both sides of the filter.After connecting,connect one end of the tube to the
overflow cup and the other end to the collecting bottle,as shown below:
Caution
When installing the filter to the suction tube,please note that the side with an IN is facing the
collecting bottle.
When using the collecting bottle,always check the filling of the collecting bottle,and replace
or clean the waste liquid in the bottle before it is full,so as to avoid interruption of suction
and delay of treatment.
1. As shown in the figure, connect the back-up oxygen supply hose to the O2 supply port (in case of
5-38
air-driven, use backup air supply hose).
3. Fix external vacuum suction by tightening the four screws on the bracket.
5-39
4. Connect the other end of the backup oxygen supply hose to the external negative pressure suction port
and tighten the nut to fasten it.
Warning
AGSS transfer and receiving system shall work with a breathing system conforming to YY
0635.1 and ISO 80601-2-13.
5-40
5.10.1 Structur Composition of AGSS
2 7
6
3
5
4
No. Notes
1 Waste gas exhaust nozzle connector
2 AGSS exhaust gase outlet
3 Transfer system hose conical connector
4 Pressure Compensation Port
5 Main body of AGSS system
6 Float (red)
7 Flow regulation knob
1 Assemble the waste gas exhaust nozzle connector to the gas outlet of breathing system:
5-41
2
3 Connect the transfer system hose to the waste gas exhaust nozzle connector.
4 Connect the 30mm conical connector of transfer system hose to the inlet of AGSS system.
5 Connect the AGSS exhaust gas outlet to the exhaust gas disposal system of the hospital
through AGSS active scavenging hose.
5-42
5
Caution
Do not block the pressure compensation port during the process of assembling and using
AGSS.
Prior to transport or movement, remove the AGSS system from the anesthesia machine.
2. The AGSS transfer and receiving is a 1L-type low flow low vacuum type, and conforms to ISO
80601-2-13 and YY 0635.2 standards. The adjustable range of pump rate of the AGSS transfer and
receiving system is 25 ~50 L/min.
3. Prior to use, verify that the waste gas disposal system is a high-flow disposal system, and is
able to reach the flow range.
4. Prior to use, verify that connector of the waste gas disposal system is BS 6834 standard
connector.
Caution
During testing, do not block the pressure compensation port of the AGSS transfer and
receiving system.
Warning
The AGSS transfer and receiving system cannot be used with flammable Anesthesia gases.
5-43
If the hose between waste gas disposal system and AGSS is blocked, or the extract flow of
the waste gas disposal system is insufficient, exhaust gas from the expiratory system exceeds
the tidal volume 1L specified in ISO 80601-2-13 and YY 0635.2, or exceeds the required
semi-sine gas flow of 20 times/minute, or the waste gas disposal system fails to work, gas
inside the AGSS may exceed 100 mL/min and overflow into the atmosphere. In such a case,
it is inadvisable to use the AGSS.
5-44
Chapter 6 Alarm
6.1 Overview
Alarms given by the anesthesia machine have the audible and visual indications to the
medical care personnel when the patients using the anesthesia machine show abnormal changes in
vital signs or the anesthesia machine can not work normally due to its fault.
Caution
When the equipment is turned on, the system tests the alarm tones and alarm lamp
functions if they can work normally. If yes, the equipment gives a sound “Beep”, and the
alarm lamp blinks once in red and yellow each. If the tones and alarm lamp functions are
abnormal, do not use the equipment. Please contact the Company immediately.
In case multiple different alarms occur simultaneously, the equipment will give audible and
visual alarms as per the alarm of the highest level among them.
User shall set the alarm volume and alarm limits as per actual conditions of the patients. Do
not only rely on the audible alarm system for patient monitoring. If the alarm tone is
adjusted to a lower volume, patients’ safety may be endangered. User shall pay close
attention to actual clinical status of the patients.
Information such as physiological parameters and alarms displayed in the screen of the
equipment are for clinicians’ reference only, and must not be directly used as basis for
clinical treatment.
A potential hazard can exist if different alarm pre-sets are used for the same or similar
equipment in any single area.
Alarm system could restore the previous alarm setting no matter the duration of power
interruption.
1 Physiological Alarms
Physiological alarms are usually given when the physiological parameter measurement of
the patient exceeds the preset High/Low alarm limits or some physiological abnormality
6-1
of the patient takes place. The alarm messages of physiological alarms are displayed in
the physiological alarm area at the upper part of the screen.
2 Technical Alarms
Technical alarms, which are also known as system error messages, indicate the alarms
triggered when some system function cannot work normally resulting from misoperation
or system malfunction, or the monitoring results are distorted. Alarm messages of
technical alarms are displayed in the technical alarm area at the upper part of the screen.
3 Prompting Messages
Strictly speaking, prompting messages do not fall into the scope of alarms. Apart from the
physiological alarm messages and technical alarm messages, the Anesthesia machine can
display prompting messages which are related to the system state. The prompting
messages usually do not relate to the vital signs of patients. The prompting messages are
displayed in the system prompting message area.
Physiological sign of patient appears abnormal, and corresponding measures may possibly
be taken or treatment may possibly be conducted.
Levels of all technical alarms and some physiological alarms are already preset before the
anesthesia machines are shipped, and cannot be modified by the users. Levels of some
physiological alarms may be modified.
6-2
Visual alarms
Audible alarms
Alarm messages
Parameter flashing
Of which, the alarm levels of visual alarms, audible alarms and alarm messages are presented in
different ways respectively.
Information: Do
6-3
The following marks are added before alarm messages to differentiate level alarm messages:
Information: no “!”mark
: The icon indicates that alarm sound is temporary muted,temporary mute 120S.
Warning
When using the equipment, you cannot just depend upon audible alarms. If the alarm tone
is adjusted to a lower volume, the patient’s safety may be endangered. Keeping the patient
under close surveillance is the most reliable way.
Click on the system default, when the set minimum alarm volume is less than or equal to
level 5, the alarm volume is restored to level 5 when the default value is restored or when
the minimum alarm volume set is greater than level 5, the alarm volume is restored to the
minimum alarm volume when the default value is restored.
6-4
6.4 Set the Alarm Limits
Caution
When the parameter value is higher than high alarm limit or lower than low alarm limit ,
an alarm will be triggered.
Click on the system default, the ventilator parameter alarm high and low limit is restored to
the default alarm high and low limit corresponding to the patient type.
When using the equipment, always ensure that the alarm limits are set to suitable values.
Set High alarm limit and low alarm limit as per clinical requirements. If the settings are
beyond the valid range, the alarm system may be ineffective.
2. On the alarm screen of ventilator, set one by one the alarm [High Limit] and [Low limit] of
parameters like [MV] , [Ppeak] , [FiO2] and [VT] .
3. Or select [Load Default Alarm Limit] to use the default high/low alarm limits.
6-5
Range of High/Low alarm limits:
1 ( low
1(low limit)
limit)
0~ (High 10(high limit)
MV 2~100 L/min 1
limit-2) 5(high
limit)
8 ( low
10(low limit)
limit)
0~ (High 50(high limit)
2~100 cmH2O 1
limit-2) 40(high
limit)
5 ( low
5(low limit)
limit)
0~ (High 1000(high
VT 5~1600 mL 5
limit-5) limit) 200(high
limit)
Warning
If an alarm is cleared in manual mode, alarms such per-minute ventilation amount and
expiratory tidal volume will not be triggered.
6-6
The alarm setting can be saved before and after shutdown for 30S,and the setting before
shutdown can be maintained.
Respironics CO2:
0(low limit)
(low 0~74 0(low limit)
FiCO2 limit+1)~ 4(high limit) mmHg 1 N/A
76 4(high limit)
25(low limit)
(low 0~(high 25(low limit)
EtCO2 limit+2)~ limit-2) 50(high limit) mmHg 1 N/A
150 50(high limit)
6-7
Masimo CO2:
0(low limit)
(low 0~97 0(low limit)
FiCO2 limit+1)~ 4(high limit) mmHg 1 N/A
4(high limit)
99
25(low limit)
(low 0~(high 25(low limit)
EtCO2 limit+2)~ limit-2) 50(high limit) mmHg 1 N/A
50(high limit)
190
2. On the alarm screen of BIS, set the [High Limit] and [Low Limit] for [BIS] alarms, as shown in the
figure below:
6-8
6.4.4 Setting the AG Alarm Limits
2. Set the [High Limit] and [Low Limit] of the parameters of [FiN2O] , [EtN2O] , [FiAA] and
[EtAA] on the [AG] Alarm screen. See the following figure:
6-9
Exhalatio (Low Limit + 0 % ~ (High 0(low limit) 0(low limit) N/A
% 0.1
n of AA 0.2%)~25.0% Limit-0.2%) 5.0(high limit) 5.0(high limit)
2. The [High Limit ] of CO2 alarm levels can be set to, [Medium],default [Medium],not modifiable.
Push the alarm softkey to set the system to alarm pause status, namely all alarm tones of
the system are shielded, and “ ” icon and 120 s counting-down are displayed in the upper
right corner of the screen.
Caution
In the alarm pause status, all alarm modes works normally except for audible alarm.
In alarm pause status, even if there is a newalarm pause technical or physiological alarm,the
alarm will continue to be suspended.
Once 120s counting down expires, the system will automatically exit the current alarm
pause status and reactivate the audible alarm.
6-10
6.6.2 Cancelling the Alarm pause
If the alarm pause soft key is pushed,the system will exit current alarm pause status and
reactivate audible alarm, and the alarm pause icon turn into ,and 120s counting down
displayed in upper right corner of the screen disappear simultaneously.
Caution
When the anesthesia machine is completely powered off or turned off, the stored module
alarm switch may not be deleted.
For specific measures for each alarm, refer to “Chapter 7 Physiological Alarms and Technical
Alarms”.
6-11
6.9 Testing Alarm System
The alarm system includes three types of alarms, i.e. physiological alarm, technical alarm
and prompt message respectively. You can test the alarm system through the condition of visual
alarm, audible, parameter flashing and message. For example:
1. Make preparations according to the “Measure Steps and Examination” in Chapter
CO2 , and enter the [CO2] Alarm screen to set [High Limit] and [Low Limit] of the
parameters of [FiCO2] , [EtCO2] as 15 mmHg and 6 mmHg respectively.
2. On the screen, select [Alarm] →enter the menu of [Sound] and [Indicate Sound
Volume] , and set the volume to “0~7”.
3. When the measured values exceed high limit and low limit of alarm, select [Alarm]
→ [CO2] → [Alarm Level] , and set the [Alarm Level] to [High] and [Medium]
respectively. Observe the changes of sound, light and parameter flashing. See the contents
of “Visual Alarms”, “Audible Alarms”, “Parameter Flashing” in this chapter for details.
Meanwhile, it indicates that CO2 is too high or too low in the physiological alarm area.
4. In normal condition, when the measured values exceed the high limit or low limit of
alarm, the equipment will provide indications in the form of light, sound and message, and
the average delay of alarm does not exceed 5 seconds.
5. Pull out the sampling tube of CO2 from the anesthesia machine, and the following
message prompts out in the technical alarm area: Without Adsorption Tube.
Attention
When several alarms are triggered at the same time, the system will only give the visual and
audible indications for the alarm of highest level.
6-12
Chapter 7 Physiologic Alarms and Technical Alarms
Most of the essential physiological and technical alarm messages are listed in this chapter, however, some
alarm messages are not necessarily listed.
Caution
In this chapter, H indicates high level, M indicates medium level ,L indicates low level.
Corresponding measures are listed for each alarm message. In case the problem still exists after action is
taken, contact the service technician.
Alarm
Alarm messages Causes and Measures
level
Airway peak pressure (Ppeak) is higher than the set value of high alarm
High Paw H limit. Reduce the set value of tidal volume, or increase the set value of
upper Paw alarm limit.
Airway peak pressure Ppeak is lower than the set value of low Paw
Low Paw H alarm limit (lasting for 20 seconds). Increase the set value of tidal
volume, or reduce the set value of high Paw alarm limit.
MV value is higher than the high alarm limit. Reduce the tidal volume,
High MV M
slow the respiratory frequency, or increase the high alarm limit.
MV is lower than the low alarm limit. Increase the tidal volume,
Low MV M
increase respiratory frequency, or reduce the low alarm limit.
High VTexp H Expiratory Tidal Volume value is higher than high alarm limit. Reduce
the preset tidal volume or increase the high alarm limit.
Low VTexp H Expiratory Tidal Volume value is lower than low alarm limit. Increase
the preset tidal volume or reduce the low alarm limit.
FiO2 value is higher than high alarm limit. Reduce fresh-gas oxygen
High FiO2 M
flow or increase the high alarm limit.
FiO2 value is lower than the low alarm limit. Increase fresh-gas oxygen
Low FiO2 H
flow or reduce the low alarm limit.
Two (2) triggering conditions are satisfied simultaneously:
Apnea M 1. Airway pressure is continuously lower than (PEEP +3) cmH2O for
more than 30 seconds.
7-1
Alarm
Alarm messages Causes and Measures
level
2. Expiratory tidal volume is continuously lower than 10ml for more
than 30 seconds.
Increase the set values of tidal volume and respiratory frequency, or set
it to Manual/spontaneous mode.
No respiration takes place within the latest 120 seconds. Check status of
Apnea>2min H the patient. Use Manual/spontaneous mode to aid the patient to breathe.
Check whether or not any pipeline drops out.
Paw value is higher than Plimit. Increase the Plimit or reduce the tidal
Pressure Limiting L
volume or the respiratory rate.
Continuous Pressure H In the respiratory circuit, Paw value is higher than the continuous
airway pressure. If the continuous airway pressure is too high, the
equipment reports an alarm lasting for 15 seconds.
Pressure is 10cmH2O lower than the atmosphere. Check whether or not
the patient is conducting autonomous respiration. Increase the fresh gas
Negative Pressure H flow. Check whether or not there is high air flow through the
scavenging system. If yes, check the negative pressure relief valve on
the receiver.
Apnea Ventilation M Can not detect patient‟s respiration over the set Apnea time, and the
ventilation for Apnea is undergoing. Check patient‟s respiration or
increase the setting of Apnea time.
High EtCO2 M EtCO2 concentration is higher than the alarm high limit. Increase the
high limit of the alarm setting.
Low EtCO2 M EtCO2 concentration is lower than the alarm low limit. Reduce the low
limit of the alarm setting.
High FiCO2 M FiCO2 concentration is higher than the alarm high limit. Increase the
high limit of the alarm setting.
Low FiCO2 M FiCO2 concentration is lower than the alarm low limit. Reduce the low
limit of the alarm setting.
High FiO2 M (When using the module with paramagnetic oxygen sensor) FiO2 value
is higher than the alarm high limit. Decrease the O2 flow in the fresh air
or increase the high limit of the alarm setting.
Low FiO2 M (When using the module with paramagnetic oxygen sensor) FiO2 value
is lower than the alarm low limit. Increase the O2 flow in the fresh air or
reduce the low limit of the alarm setting.
7-2
Alarm
Alarm messages Causes and Measures
level
High EtN2O M EtN2O concentration is higher than the alarm high limit. Decrease N2O
flow or increase the high limit of the alarm setting.
Low EtN2O M EtN2O concentration is lower than the alarm low limit. Increase N2O
flow or reduce the low limit of the alarm setting.
High FiN2O M FiN2O concentration is higher than the alarm high limit. Decrease N2O
flow or increase the high limit of the alarm setting.
Low FiN2O M FiN2O concentration is lower than the alarm low limit. Increase N2O
flow or reduce the low limit of the alarm setting.
High EtHAL M EtHAL concentration is higher than the alarm high limit. Decrease HAL
flow or increase the high limit of the alarm setting.
Low EtHAL M EtHAL concentration is lower than the alarm low limit. Increase HAL
flow or reduce the low limit of the alarm setting.
High FiHAL M FiHAL concentration is higher than the alarm high limit. Decrease HAL
flow or increase the high limit of the alarm setting.
Low FiHAL M FiHAL concentration is lower than the alarm low limit. Increase HAL
flow or reduce the low limit of the alarm setting.
High EtENF M EtENF concentration is higher than the alarm high limit. Decrease ENF
flow or increase the high limit of the alarm setting.
Low EtENF M EtENF concentration is lower than the alarm low limit. Increase ENF
flow or reduce the low limit of the alarm setting.
High FiENF M FiENF concentration is higher than the alarm high limit. Decrease ENF
flow or increase the high limit of the alarm setting.
Low FiENF M FiENF concentration is lower than the alarm low limit. Increase ENF
flow or reduce the low limit of the alarm setting.
High EtISO M EtISO concentration is higher than the alarm high limit. Decrease ISO
flow or increase the high limit of the alarm setting.
Low EtISO M EtISO concentration is lower than the alarm low limit. Increase ISO
flow or reduce the low limit of the alarm setting.
High FiISO M FiISO concentration is higher than the alarm high limit. Decrease ISO
flow or increase the high limit of the alarm setting.
Low FiISO M FiISO concentration is lower than the alarm low limit. Increase ISO
flow or reduce the low limit of the alarm setting.
High EtSEV M EtSEV concentration is higher than the alarm high limit. Decrease SEV
flow or increase the high limit of the alarm setting.
7-3
Alarm
Alarm messages Causes and Measures
level
Low EtSEV M EtSEV concentration is lower than the alarm low limit. Increase SEV
flow or reduce the low limit of the alarm setting.
High FiSEV M FiSEV concentration is higher than the alarm high limit. Decrease SEV
flow or increase the high limit of the alarm setting.
Low FiSEV M FiSEV concentration is lower than the alarm low limit. Increase SEV
flow or reduce the low limit of the alarm setting.
High EtDES M EtDES concentration is higher than the alarm high limit. Decrease DES
flow or increase the high limit of the alarm setting.
Low EtDES M EtDES concentration is lower than the alarm low limit. Increase DES
flow or reduce the low limit of the alarm setting.
High FiDES M FiDES concentration is higher than the alarm high limit. Decrease DES
flow or increase the high limit of the alarm setting.
Low FiDES M FiDES concentration is lower than the alarm low limit. Increase DES
flow or reduce the low limit of the alarm setting.
H BIS value is higher than the alarm high limit. Check the anesthetics
High BIS
supply. Increase the anesthetics flow rate or alarm high limit.
H BIS value is lower than the alarm low limit. Check the anesthetics
Low BIS
supply. Reduce the anesthetics flow rate or alarm low limit.
Alarm
Alarm messages Causes and Measures
level
No calibration data was found in memory or the calibration data did not
Calibrate Flow
L match. Use Manual/spontaneous mode to assist patient to breathe.
Sensor
Calibrate the flow sensor.
No calibration data was found in memory or the calibration data did not
Calibrate Pressure
L match. Use Manual/spontaneous mode to assist patient to breathe.
Sensor
Calibrate the pressure sensor.
No calibration data was found in memory or the calibration data did not
Calibrate O2 Sensor L
match. Calibrate or replace oxygen sensors.
Power Failure H Three way valve 12V,safety valve 5V or 12V,proportional valve
7-4
Alarm
Alarm messages Causes and Measures
level
7.5V,internal AD reference voltage 3.3V and external AD reference
voltage 2.5V,at least one way power supply failure. Use
Manual/spontaneous mode to assist patient to breathe. Contact the
manufacturer for service.
Self Check Error L MCU hardware error. Contact the manufacturer for service.
Watchdog Failure The external watchdog is burnt out and out of order. Please contact the
H
manufacturer for service.
Internal A/D The internal ADC is out of order. Please contact the manufacturer for
H
Converter Failure service.
External A/D The external ADC is out of order. Please contact the manufacturer for
H
Converter Failure service.
Connection or control of zero valve gets faulty. The device can still
Zero Valve Failure L work, but the monitoring is unreliable. Use manual/spontaneous mode
to aid the patient to breathe if necessary.
Connection or control of expiratory valve is faulty. Use
Expiration Valve
M Manual/spontaneous mode to assist patient to breathe. Please contact
Failure
the manufacturer for service.
Flow Valve Failure M There exists connection or control failure in the flow valve. Use
Manual/spontaneous mode to assist patient to breathe. Please contact
the manufacturer for service.
M Connection of safe valve is faulty. Use Manual/spontaneous mode to
Safe Valve Failure
assist patient to breathe. Please contact the manufacturer for service.
Safe Valve Control Connection or control of safe valve is faulty. Use Manual/spontaneous
H
Failure mode to assist patient to breathe.
Pressure of oxygen supply is inadequate. Make sure O2 supply of
No O2 Pressure H adequate pressure is connected;If an air supply is connected, you may
use Manual/spontaneous to aid the patient to breathe.
Breathing system is not assembled, or the connecting wires of breathing
No Breathing
H system mounting are connected incorrectly. Contact the manufacturer
System
for service.
O2 Flush Failure The button of “O2 Flush”cannot be pressed down. Or the oxygen flush
M cannot be started after the button is pressed down. Contact the
manufacturer for service.
7-5
Alarm
Alarm messages Causes and Measures
level
ACGO in Use L The ACGO is being used. Check the cover condition of ACGO.
UCOS Error The software is out of order. Please contact the manufacturer for
L
service.
Monitor Communication failure between monitor board and backup CPU.The
H
Communication Fail serial line is not connected or the chip of the serial port is broken.
7-6
Alarm
Alarm messages Causes and Measures
level
With Bk CPU
Bk CPU
Communication failure between backup CPU and host.
Communication H
The serial line is not connected or the chip of the serial port is broken.
Failure With Host
Host
Communication failure between backup CPU and host.
Communication Fail H
The serial line is not connected or the chip of the serial port is broken.
With Bk CPU
Host H Communication failure between monitor board and host.
Communication The serial line is not connected or the chip of the serial port is broken.
Failure With
Monitor
VPM
Abnormal communication between VPM and EFM main control board.
Communication H
Please contact the manufacturer for service.
Failure With EFM
FS Communication
Abnormal communication between FS and VPM.
H
Failure With VPM Please contact the manufacturer for service.
Key Board Failure Communication failure between key board and host.
H
The serial line is not connected or the chip of the serial port is broken.
7-7
7.2.2 Full electronic flowmeter Alarms
Alarm
Alarm messages Causes and measures
level
No calibration data was found in memory or the calibration data did not
Calibrate the Gas
L match.
Pressure
Please contact the manufacturer for service.
No calibration data was found in memory or the calibration data did not
Calibrate EFM O2
L match.
Flow valve
Please contact the manufacturer for service.
Drive Gas selector Drive gas select valve connection line fault or select valve fault.
H
valve Failure Please contact the manufacturer for service.
EFM O2 select Oxygen select valve connection line fault or select valve fault.
M
valve err Please contact the manufacturer for service.
EFM N2O select Nitrous Oxide select valve connection line fault or select valve fault.
M
valve err Please contact the manufacturer for service.
EFM air gas select Air select valve connection line fault or select valve fault.
M
valve err Please contact the manufacturer for service.
EFM O2 Flow Oxygen flow valve connection line fault or select valve fault.
M
valve err Please contact the manufacturer for service.
EFM balance gas Balance gas flow valve connection line fault or select valve fault.
M
flow valve err Please contact the manufacturer for service.
EFM O2 Flow Oxygen flow sensor connection line fault or select valve fault.
H
Sensor Failure Please contact the manufacturer for service.
EFM balance gas fl Balance gas flow sensor connection line fault or select valve fault.
H
ow sensor err Please contact the manufacturer for service.
Backup flowmeter Backup flowmeter gata valve connection line fault or select valve fault.
H
select valve err Please contact the manufacturer for service.
Backup flowmeter Backup flowmeter drive valve connection line fault or select valve fault.
M
drive valve err Please contact the manufacturer for service.
Microswitch of the back up flowmeter connection line fault or
Back-up Flowmeter microswitch fault,or the back up flowmeter did not jam and did not pop
M
Abnormal Status up completely.
Please contact the manufacturer for service.
7-8
O2 and N2O ratio a The oxygen flow valve or balance gas valve flow valve is abnormal.
H
bnormal Please contact the manufacturer for service.
Fresh gas is less than 100mL/min.Please check if the air pressure is
No Fresh Gas L sufficient.If the air pressure is sufficient,the flow valve may fault,please
contact the manufacturer for maintenance.
7-9
7.2.3 Battery Alarms
Alarm
Alarm messages Causes and measures
level
The battery voltage is below 9.5V. Please connect alternating
Low Battery current immediately. In case of power cut-off, use Manual/spontaneous
H
Voltage to aid the patient to breathe. If the battery cannot be fully charged within
24 hours, contact the manufacturer for service.
The battery voltage is below 9.1V. Please connect AC supply
System Down For immediately. In case of power cut-off, use Manual/spontaneous to aid
H
Battery Depletion the patient to breathe. If the battery cannot be fully charged within 24
hours, contact the manufacturer for service.
Battery is not assembled, or the battery cable is disconnected with
No Battery M
power module. Please contact the manufacturer for service.
On Battery Power L Battery is in use. Please check the AC power connection status.
7-10
7.2.4 AG Module Alarms
Need O2 Span L Oxygen sensor need to be calibrated in pure oxygen. Calibrate oxygen
Calibrate sensor in pure oxygen. If user can‟t meet the requirement, please return
it back to factory.
O2 Sensor Error L Module fails. Reseat oxygen sensor orreboot.
O2 Port Failure L
Hardware H
Error,Replace
Sensor
Sampling tube blocking. Check and replace the
AG Sampling Line
L sampling tube, if the faultpersists, please contact themanufacturer for
Clogged
maintenance.
7-11
Sampling tube is not connected or in badcontact. Check and replace the
AG No Sampling
L sampling tube, if the faultpersists, please contact themanufacturer for
Line
maintenance.
Adapter abnormal. check and replace the adapter,if fault persists, please
Replace AG
M contactthemanufacturer for
Adapter
maintenance.
L Adapter is not connectedor in bad contact. Check and replace the
AG No Adapter
adapter,if fault persists, please contactthemanufacturer for maintenance.
CO2 Outside L
Specified Accuracy
Range
O2 Outside L
Specified Accuracy
The measured valueexceeds the claimed scope of accuracy. Please
Range
follow the precisionrange of measurement claimed
N2O Outside L
by manufacturers.
Specified Accuracy
Range
AX Out Of L
Accuracy Range
Temp Out Of Range L
Atmospheric
M
Pressure Beyond
L
Module fails. Return to factory for
AG Speed Out Of maintenance.
Bounds
AG Factory L
Calibration Lost
Zero reference L AG Analyzer requires zero. enter AG settings for zero Here zero is equal
7-12
calibration required to calibration.
AG L AG Analyzer does not recognize the existence of
Conc.Unreliable other gas. check the gas.
Replace O2 Sensor L Beyond usage time. Replace oxygen sensor.
AG Software Error L Software error occurs. Reboot
There is only one mainnesthetic gas waveformand value shown in
Found Two AG
M thekinds of anesthesia gasfor tips. Check the patient'sphysiological signs
Gases
andanesthesia.
AG Analyzer operatingmode is on standby. Measurement mode of AG
AG Is Sleeping L
Analyzer is selected.
7-13
Accuracy Range
CO2 Temp Out Of L The module has a fault. Return to us for repair.
Range
CO2 Span L
Calibration Error
(Masimo)
CO2 Factory L
Calibration Lost
(Masimo)
CO2 Speed Out L
Of Bounds
(Masimo)
Atmospheric L
Pressure Beyond
(Masimo)
CO2 Span L The CO2 span is being calibrated. Return to us.
Calibrating......(M
asimo)
Replace CO2 L The adapter has a fault. Check and replace the adapter. If the fault still
Adapter (Masimo) exists, contact us for service.
CO2 No Adapter L The adapter is in poor contact or not connected. Check and replace the
(Masimo) adapter. If the fault still exists, contact us for service.
CO2 Need Zero L The CO2 module needs zeroing. Enter [CO2 Setup] to select [Zero].
(Masimo)
CO2 Software L The software has a fault. Restart the Monitor.
Error (Masimo)
CO2 Hardware L The hardware has a fault. Check and replace the sensor. If the fault still
Error (Masimo) exists, contact us for service.
7-14
Monitor to try again. If the error still exists, contact us for service.
BIS COMM H The BIS module fails to communicate with the main system. Restart the
STOP Monitor to try again. If the error still exists, contact us for service.
H The BIS module fails to communicate normally with the main system.
BIS Comm Error Restart the Monitor to try again. If the error still exists, contact us for
service.
L The alarm limit of parameter BIS is accidentally changed. Contact us for
BIS Alm Lmt Err
service.
L The measured value of parameter BIS goes beyond the specified
BIS overrange
measurement range. Contact us for service.
L BIS sensor is connected unreliably. Check if the BIS sensor is connected
BIS No Sensor
properly. If the fault still exists, contact us for service.
L Electrode of the sensor falls off or the skin is too dry. Check patient cable
impedance too
connection and check electrode-to-skin contact. If necessary, clean and
high
dry the skin.
IMPEDENCE L Prompt information for ten-minute ground electrode impedance check.
CHECKING No action necessary
H Data communication between monitor and BIS module stops. Reinitialize
BIS COMM
BIS module. If there is still malfunction, please contact the manufacturer
STOP
to maintain it.
L Sensor is out of its service life. The initialization of sensor can be still
BIS SENSOR conducted and the data waveform data and parameter can be also
EXPIRED captured. This alarm merely reminds the users that this sensor is out of
service life, but it depends on users whether to replace it.
INVALID BIS L Invalid sensor. Change Sensor
SENSOR
L During the process of initialization and monitoring, the sensor over
REPLACE THE
current and incorrect ground impedance have been detected by the
SENSOR!
system. Change Sensor.
POOR SIGNAL L Weak signal. SQI value is too low. Check the status of patient and the
QUALITY connection of sensor.
(SQI﹤50%)
BAD SIGNAL L Poor Signal. SQI value is too low. Check the status of patient and the
QUALITY connection of sensor.
(SQI﹤15%)
7-15
7.3 Prompt Information
7-16
Chapter 8 CO2 Monitoring
8.1 Overview
CO2 monitoring function of the anesthesia machine adopts infrared absorption technology to
determine the CO2 concentration inside the respiratory circuits of the patient.. Its principle is
based on such a fact that CO2 molecules have an attraction for energy contained by infrared light
of specific wavelength, and the quantity of absorbed energy is direct related to CO2 concentration.
When infrared light emitted by an infrared light supply penetrates a gas sample containing CO2,
part of the energy may be absorbed by CO2 entrained in the gas. A photoelectric detector is set up
on the opposite side of the infrared light supply to determine the residuary energy of infrared light
and convert it into electrical signals. The electrical signals are compared with the energy infrared
light supply and adjusted, and the concentration of CO2 contained in the gas sample may be
precisely reflected.
Sidestream
Sample the respiratory gases inside the patient-respiration airway at constant sampling flow,
and analyze the samples by CO2 sensor built in the module.
Mainstream
For this measurement, CO2 sensors are installed on the airway adapter that plugs directly
into the patient's breathing system.
1. CO2 waveform.
Warning
In accordance with international standards, CO2 concentration shall be monitored while the
equipment is connected with a patient. If your equipment does not have the function, please
use a monitor conforming to corresponding international standards for CO2 concentration
monitoring.
Caution
To make sure the safety of patient, perform CO2 monitoring while use this equipment.. If
8-1
your equipment does not have the function of CO2 monitoring, please use an equipment with
CO2 monitoring function which conforms to corresponding international standards.
2 5
3 Gas outlet
8-2
8.2.2 MASIMO CO2 Module (Mainstream)
2 4
2 5
8-3
1 Module working indicator 4 Sampling line port
3 Gas outlet
2 4
samples gas from the respiratory circuit through the Nomoline Family sampling line at a rate of 50
sml/min, making measurements of CO2 possible for adult, pediatric patients.
The Nomoline Family sampling lines incorporate a unique water separation (NO MOisture) section,
which removes condensed water. The NOMO section is also fitted with a bacteria filter that protects
the gas analyzer from water intrusion and cross contamination.
As long as no sampling line is connected, the CO2 module remains in a low-power sleep mode. Once
the sampling line is connected, the CO2 module switches to measuring mode and starts delivering gas
8-4
data.
The Nomoline Family sampling lines are available in a wide variety of versions for both intubated and
spontaneously breathing patients and in both disposable and re-sposable configurations –intubated
patients can for instance be monitored using the disposable Nomoline Airway adapter Set or a
com-bination of the multiple patient use Nomoline Adapter and a disposable Nomoline Extension /
T-adapter. Spontaneously breathing patients could similarly be monitored using a disposable
Nomoline Nasal CO2 Cannula or a combination of the multiple patient use Nomoline Adapter and a
disposable Nomoline Nasal CO2 Cannula with Luer Connector.
Figure 1.The disposable Nomoline Airway Adapter Set is an alternative to using a combination of the
multiple patient use Nomoline Adapter and a disposable Nomoline Extension / T-adapter.
The Nomoline Adapter may be used with other third party sampling lines and cannulas. Please
however note that the Nomoline Family of sampling lines are designed for optimal performance and
measure-ment fidelity when used with the CO2 module. For instance, when connecting to a
respiratory circuit, the Masimo T-adapter provides a central gas sampling point thereby minimizing the
risk of sampling line occlusion (see below)
Figure 2. For optimal water handling, always use T-adapters with the sampling point in the center of
the adapter, as shown to the left above.
Note
Using sample lines or cannulas with larger inner diameter than 1 mm will increase
8-5
module’s total system response time.
Caution
The section is only applicable to Sidestream and Mainstream CO2 module of anesthesia
machines.
Warning
MASIMO Sidestream CO2 module cannot be used in flammable anesthetic gases.
a) Install the module to the corresponding position of the anesthesia machine plug slot.
b) Connect one end of Nomoline sampling line to the input port of CO2 module, and the
other end of the sampling line to the patient.
c) Turn the system switch to the right side to start the anesthesia machine. The indicator
at the top left corner of the CO2 module is in green.
e) Click [Zero Sensor]. While zeroing, the screen shows: In zero calibration, please wait
for 30s, and start to measure CO2 after the prompt is finished.
f) When CO2 monitoring function is turned on, its working mode is “measure” by
default. However, to ensure the set mode, access [CO2 module] and verify that its
[work mode] is set to [Measure] mode.
g) Connect the gas sample outlet to the scavenging system, or allow the gas to flow back
to the patient circuit.
8-6
h) Check the equipment as per “Chapter 4 Tests Before Use”.
Sampling tube
Sampling Line Sampling
SamplingLine
tube
Before connect the Nomoline sampling line to patient respirotary circuit, perform the
following steps:
a) Connect the sampling line to the gas inlet port of CO2 module;
c) Verify that the indicator at the top left corner of the CO2 module is in green..
d) Exhale to the sampling line, and verify that valid CO2 waveforms and values are
displayed on the anesthesia machine.
f) Check whether or not a clogging alarm is triggered and flashing red light is on gas
inlet port
g) If necessary, perform leak check of patient circuit that is connected to the sampling
line.
Caution
The end of airway adapter connecting to the sampling line shall point upwards to prevent
condensed water drops from entering the sampling line and resulting in clogging.
Measuring Procedures
Set the mainstream CO2 module to perform gas analysis:
a) Install the mainstream module to the corresponding position of the anesthesia machine slot .
8-7
b) Plug the end of the 12Pin connector of the patch cord into the 12Pin data interface of the
mainstream CO2 module. Connect the other end to the MASIMO mainstream CO2 probe. Install the probe on
the airway adapter. Finally, connect the airway adapter to the patient‟s breathing circuit.
c) Turn the system switch to the right side to start the anesthesia machine. The indicator at the top
left corner of the CO2 module is in green.
d) In the [Config] menu, select [Gas Module], and then set up options of [Unit], [O2Compensation],
[Balance gas], [Altitude], etc.
e) Click [Zero Sensor]. While zeroing, the screen shows: In zero calibration, please wait for 30s, and
start to measure CO2 after the prompt is finished.
f) When CO2 monitoring function is turned on, its working mode is [measure] by default. However,
to ensure the set mode, access [CO2 module] and verify that its [work mode] is set to [Measure]
mode.
Caution
The end of the airway adapter connecting to the sampling line shall point upwards to
prevent the condensed water droplets from entering the sampling line and resulting in
clogging.
Measuring Procedures
Set the sidestream CO2 module to perform gas analysis:
8-8
a) Install the sidestream module to the corresponding position on the anesthesia machine.
b) Connect one end of Respironics sampling line to the sampling port of Respironics CO2 module, and
the other end to the patient end.
c) Turn the system switch to the right side to start the anesthesia machine. The indicator at the top left
corner of the CO2 module is in green.
d) In the [Config] menu, select [Gas Module], and then set up options of [Unit],
[O2Compensation], [N2O compensation], [Zero Sensor], etc.
e) Click [Zero Sensor]. While zeroing, the screen shows: In zero calibration, please wait for 30s, and
start to measure CO2 after the prompt is finished.
f) When CO2 monitoring function is turned on, its working mode is [measure] by default. However,
to ensure the set mode, access [CO2 module] and verify that its [work mode] is set to [Measure]
mode.
g) Connect the gas sample outlet to the exhaust system or allow the gas to flow back to the patient
circuit.
h) Check the equipment as described in Chapter 4 Tests Before Use.
i) Testing result is normal, start the CO2 monitoring.
Checks before Use
Before connect Respironics sampling line to the patient's respiratory circuit, perform the the following
steps:
a) Connect the Respironics sampling line to the Respironics CO2 module gas inlet.
b) Verify that the working indicator at the top left corner of the mainstream module is constantly in
green (indicating that the plug-in module is powered on normally)
c) Expire to the adapter on the sampling line, and verify that valid CO2 waveforms and values are
displayed on the anesthesia machine.
d) Block the sampling line mouth for 10 second.
e) Check whether the clogging alarm is displayed on anesthesia machine.
f) If necessary, perform a leak test on the patient circuit connected to the sampling line.
Measuring procedures
Set the mainstream CO2 module to perform gas analysis:
a) Install the mainstream module to the corresponding position on the anesthesia machine.
b) Plug the end of the 12Pin connector of the patch cord into the 12Pin data interface of the
mainstream CO2 module. Connect the other end to the mainstream CO2 probe. Install the probe on the airway
adapter. Finally, connect the airway adapter to the patient‟s breathing circuit.Turn the system switch to the
right side to start the anesthesia machine. The indicator at the top left corner of the CO2 module is in green.
8-9
c) In the [Config] menu, select [Gas Module], and then set up options of [Unit],
[O2Compensation], [N2O compensation], [Zero Sensor], etc.
d) Click [Zero Sensor]. While zeroing, the screen shows: In zero calibration, please wait for 30s, and
start to measure CO2 after the prompt is finished.
e) When CO2 monitoring function is turned on, its working mode is [measure] by default. However,
to ensure the set mode, access [CO2 module] and verify that its [work mode] is set to [Measure] mode.
f) Check the equipment as described in Chapter 4 Tests Before Use.
g) Testing result is normal, start the CO2 monitoring.
Checks before Use
Before connect the adapter to the patient's respiratory circuit, perform the the following steps:
a) Verify that the 12-pin adapter cable is connected to the 12-pin connector of the mainstream CO2
module
b) Verify that the working indicator at the top left corner of the mainstream module is constantly in
green .
c) Expire to the adapter and verify that valid CO2 waveforms and values are displayed on the
anesthesia machine.
Note
The end of the gas adapter connecting to the gas sampling line shall point upwards to
prevent the condensed water drops from entering the gas sampling line and resulting in
clogging.
8-10
8.6 Set the CO2
On the screen, select [Config] menu → [Gas Module] or press the CO2 setting button on the module to enter
[Gas Module], and you can set the CO2 parameters.
Standby:
When CO2 module remains in its standby mode, the gas pump stops so as to extend the lifespan of
module; in addition, a prompt “CO2 Module Standby” is displayed in the message area.
Measure:
When CO2 module is working, the indicator of CO2 module is constantly in green, and measured data is
sent to the anesthesia machine simultaneously.
In normal condition, when CO2 module is connected to the anesthesia machine, the module
automatically set the operating mode to “Measure”. However, the user must verify that the CO2 module
is set to measuring mode.
When the anesthesia machine is used for the first time, the [Work Mode] in CO2 monitoring is set to
[Measure] by default . If CO2 monitoring is in standby mode, you can start CO2 monitoring by selecting
[Config] menu → [Gas Module] → [CO2 Module]→ [Work Mode] , and set it to: [Measure], or by
8-11
pressing the Measure/Standby button on the module to switch between the Standby/Measure mode.
Note
During sensor calibration, press the Measure/Standby button cannot switch the operating
mode. At this time, the Measure/Standby button on the screen does not function either.
When anesthesia machine is restarted, all the setting of CO2 module before the last
shutdown will be reserved.
Warning
Please set all sorts of compensation in accordance with actual conditions; otherwise
measuring results may deviate from actual values, resulting in misdiagnosis.
3. Set the gas compensation concentration of [CO2 module] in accordance with actual
conditions:
[O2 Compensation] :
Three options i.e. [High], [Medium] and [Low]. [High] indicates oxygen compensation is
85%;[Medium] indicates oxygen compensation is 50 %; [Low]indicates oxygen
compensation is 21%. When an O2 module is equipped, an O2 sensor automatically performs
O2 compensation according to the O2 concentration. Therefore, this option is in the default
state. When no O2 sensor provides automatic O2 compensation, the compensation level can
be manually set according to the actual O2 concentration (as shown in the table below).
0-30vol% 21
30-70vol% 50
8-12
70-100vol% 85
[N2O Compensation]
N2O compensation can be set to ON or OFF. If the CO2 module can measure N2O,
it can automatically perform N2O compensation based on the N2O concentration;
therefore, the selection is by default. If the CO2 module cannot measure N2O, the
compensation level can be set manually per actual N2O concentration(as shown in the
table below).
0-30vol% 0
30-70vol% 50
4. Plug in CO 2 module,the main screen to view the CO2 waveform. The CO2 waveform
is shown as the figure below.
8-13
Exhaust tube
Warning
Emitted gas shall be re-entered into the patient circuit or discharged into the discharge
system.
If the collected gas sample is to be supplied for breathing, always use a bacterial filter on
the exhaust side.
If the suction pressure is too high during discharge, the sample flow rate may be affected.
If the positive or negative pressure in the patient circuit is too large, the sample flow rate
may be affected.
Check whether the gas sample flow rate is too high for the given patient category.
8.8.1 Zeroing
For sidestream CO2 module, it is necessary to determine the zero reference level of carbon
dioxide measurement. The purpose is to eliminate the impact of baseline drift occurring during
measuring, and make sure the correctness of measurements. The zeroing calibration is herein
referred to as “Zeroing”.
1. Automatic Zeroing
Sidestream CO2 module can automatically perform zeroing by switching the gas
sampling from respiration circuit to ambient air. Sidestream CO2 module spends less
than 3 seconds to perform automatic zeroing once, and the frequency is 1time every
day.
2. Manual Zeroing
8-14
Sidestream CO2 module may automatic conduct zeroing when necessary. User may also
perform manual zeroing when necessary: access [Config] menu, open the [Gas Module]
menu, and select [Zero Sensor] in [CO2Module] . It is unnecessary to disconnect the
patient gas circuit prior to zeroing.
Warning
Since successful zeroing requires ambient air (21% O2 and 0% CO2) in the sidestream CO2
module, be sure to place the CO2 module in a well ventilated location. Avoid breathing
near the sidestream CO2 module before and after performing the zeroing procedure.
8.8.3 Calibration
CO2 module does not require routine calibration. However, it shall be calibrated every other year or
when the deviation of measured values gets too high.
To prevent cleaning liquids and dust from entering the sidestream CO2 module through its connector,
keep the Nomoline Family sampling line connected while cleaning the sidestream CO2 module.
Warning
Nomoline adsorption tube is not a sterile device. In order to avoid damage, do not perform
high-pressure sterilization on any part of the adsorption tube.
8-15
8.8.5 Lighting signals of the CO2 module
Indication Status
Steady green light System OK
Blinking green light Zeroing in progress
Steady red light Sensor error
Blinking red light Check sampling line
1. The following factors are known to have adverse effects on indication performance:
2. Gas detector
Volume percentage is the unit for detecting gas concentration. Definition of gas concentration is as
follows:
The collective pressure of mixed gases is detected by the cup pressure sensor in the CO2 module.
It can be converted to other units using the actual atmospheric pressure sent from the CO2 module.
For example: 5.0 vol% carbon dioxide @ 101.3 kPa 0.05 x 101.3 x 750 / 100 = 38 mmHg
3. Impact of humidity
The partial pressure and volume percentage of carbon dioxide, nitrogen monoxide, oxygen and
anesthetic gases depend on the water vapor content of the gas detected. Oxygen detection will be
calibrated. After which, the displayed value in normal environmental temperature and humidity will
8-16
be 20.8 vol% instead of the actual partial pressure. 20.8 vol% oxygen corresponds to the actual
oxygen concentration (vapor concentration is 0.7 vol%) of the air in the room. (e.g., when the
atmospheric pressure is 1013 hPa, it corresponds to 25°C and 23% RH).It always displays the real
partial pressure under existing humidity level when measuring carbon dioxide, nitrous oxide and
anesthetic gases (such as all the gases detected by the infrared pool).
The water vapor in the respiratory gas can get saturated (BTPS) in the patient‟s pulmonary alveolus
at body temperature.
When the respiratory gas is collected and put into the sampling line, its temperature becomes close
to environmental temperature before entering the sidestream CO2 module. No water can get into
CO2 module when all condensed water is removed by Nomoline. The relative humidity of the gas
collected is about 95%.
Carbon dioxide value under BTPS can be calculated with the following formula:
In which:
3.8 = Typical partial pressure [kPa] of the water vapor condensed between patient‟s respiration
circuit and CO2 module
8.8.7 Warnings
Warning
The sidestream CO2 module is intended for use by authorized healthcare professionals only.
Carefully route the sampling line to reduce the risk of patient entanglement or strangulation.
Do not lift the CO2 module by the sampling line as it could disconnect from the CO2 module,
causing the CO2 module to fall on the patient.
Dispose Nomoline Family sampling lines in accordance with local regulations for biohazardous
waste.
Use only airway T-adapters with the sampling point in the center of the adapter.
8-17
Do not use T-adapter with pediatrics, as this adds 7 ml dead space to the patient circuit.
Do not use the CO2 module with metered-dose inhalers or nebulized medications as this may clog
the bacteria filter.
Since a successful zeroing requires the presence of ambient air (21% O2 and 0% CO2), ensure
that the CO2 module is placed in a well ventilated place. Avoid breathing near the sidestream
CO2 module before or during the zeroing procedure.
Never sterilize or immerse the sidestream CO2 module in liquid.
The sidestream CO2 moduleis intended only as an adjunct in patient assessment. It must be used
in conjunction with other assessments of clinical signs and symptoms.
Measurements can be affected by mobile and portable RF communications equipment. Make
sure that the sidestream CO2 module is used in the electromagnetic environment specified in this
manual.
Replace the sampling line if the sampling line input connector starts flashing red, or the medical
backboard device displays a “Check sampling line” message.
No modification of this equipment is allowed without authorization of the manufacturer. If this
equipment is modified, appropriate inspection and testing must be conducted to ensure continued
safe operation.
The sidestream CO2 module are not designed for MRI environments.
During MRI scanning, CO2 module must be placed outside the MRI suite.
Use of high frequency electrosurgical equipment in the vicinity of the CO2 module/medical
backboard device may produce interference and cause incorrect measurements.
Do not apply negative pressure to remove condensed water from the Nomoline Family sampling
line.
Too strong positive or negative pressure in the patient circuit might affect the sample flow.
Strong scavenging suction pressure might affect the sample flow.
Exhaust gases should be returned to the patient circuit or to a scavenging system.
Due to the risk of patient cross-infection, always use a bacteria filter on the exhaust port side if
sampled gas is intended to be re-breathed.
Do not place the CO2 module gas analyzer in any position that might cause it to fall on the
patient.
Do not re-use disposable single-patient use Nomoline Family sampling lines due to the risk of
cross contamination.
Do not sterilize or immerse Nomoline Family sampling lines in liquid.
Do not operate the sidestream CO2 module if the enclosure is damaged.
Do not use the Nomoline Airway Adapter Set pediatric with pediatric patients.
8-18
8.8.8 Sampling line clogged
If the anesthetic gas passage is clogged, an [Sampling line clogged] message will be displayed
on the screen. In this case, replace the sampling line.
Warning
The CO2 module should be securely mounted in order to avoid the risk of damage to the
CO2 module.
Do not operate the sidestream CO2 module outside the specified operating environment.
For professional use. See instructions for use for full prescribing information, including
indications, contraindications, warnings, precautions and adverse events.
Do not use sidestream CO2 modules with metered-dose sprays or atomization treatments.
Otherwise, the bacteria filters may be clogged.
8.8.9 Consumables
The Nomoline Adapter is a multiple-patient use product.
The Nomoline Adapter should be replaced according to good clinical practice or when an
occlusion message appears.Occlusion occurs when the sample flow is too low. This is
indicated by a flashing red together with a message on the medical backboard device.
8.8.10 Maintenance
The user shall check the gas readings regularly, and contact the manufacturer engineers for
maintenance if any anomaly is detected.
Dip a cloth in 75% medicinal alcohol, water solution (bleach) containing 10% sodium hypochlorite,
disinfecting spray cleaner (like Steris Coverage SprayHB), ammonia water or mild soapy water, wash it
with water, wring it and then use it to clean the sensor. Dry it and make sure its sight glass is clean
8-19
before using the cleaned sensor.
Wash the airway adapter with mild soapy water, dip it in the disinfectant, like 75% medicinal
alcohol, water solution (bleach) containing 10% sodium hypochlorite, 2.4% glutaraldehyde solvent
(Cidex Plus, Steris System 1, etc.) or ammonia water, and then rinse it with sterile water and dry it.
The above disinfection methods allow you to reuse the airway adapter for 100 times.
8.9.5 Zeroing
Please zero before using CO2,The purpose is to eliminate the impact of baseline drift occurring
during measuring, and make sure the correctness of measurements.
Usually, the CO2 sensor will be auto zeroed when necessary. You can zero it manually when you
consider it necessary: access [Config] menu, open the [Gas Module] menu, and select [Zero Sensor] in
[CO2Module] to zero the CO2 sensor. Keep the patient circuit in the ambient air (21% O2 and 0% CO2)
for 30s in the zeroing process,when the zero prompt for 30s of the interface is finished,it means that the
zero is completed.
8-20
Chapter 9 Monitoring AG (Anesthetic Gas)
9.1 Overview
Anesthetic Gas (AG) can be used to measure the Anesthetic Gas and Respiratory Gas of the
patient under anesthetization. The Anesthetic Gas module provides the numerical values at the
end of expiration (Et) and the inspiration (Fi) of the following gases.
AG monitoring provides:
EtCO2 waveform
Monitored Parameters: FiO2, EtO2, EtCO2, FiCO2, EtN2O, FiN2O, EtAA, FiAA and MAC
AA stands for one of the Anesthetic Gases among Des (Desflurane), Iso (Isoflurane), Enf
Warning
concentration monitoring is required. If the device you are using is not equipped with this
function, monitor the AG concentration using a monitor that complies with relevant
international standards.
Attention
In accordance with international standard (ISO 80601-2-55), AG shall be monitored while the
equipment is connected with a patient. If your equipment does not have the function, please use a
The Anesthetic Gas can be analyzed by various measurement principles. The Dispersive
9-1
Infrared (DIR)Method or the Nondispersive Infrared(NDIR)Method are commonly used to
insulate the absorption characteristics of the gas sample. The DIR Method is to use a single
optical light filter and a prism or a diffraction grating to separate the wave length of each kind of
anesthetic. And NDIR Method is to get the infrared light go through several narrow-band light
filters and determine what kind of gas exists in the mixed gas.
The most commonly used gas analytical method is the one based on the medium of NDIR Method.
The measurement principle is based on that many gases absorb the infrared energy of a certain
wave length.
The multiple gas analyzer of by-pass flow, sampling and infrared usually sucks up the gas sample
from the junction of the patient‟s breathing circuit and air passage device (such as mask, tracheal
tube or laryngeal mask tube).The rate at which the modern gas analyzer gets a sample from the
breathing circuit is between 50 ml/min and 250ml/min. The sample gas goes through a little cup
or a sample room, passing by the infrared transmitter, light filter and infrared detector. The signal
sent out by the infrared detector is in proportion with the infrared energy absorbed by the gas. In
order to quantize and identify various gases, such as laughing gas, CO2 and the five type of
inhalational an esthetic gases, several light filters are needed. The detected signal is amplified, and
is converted by the complex calculation of the microprocessor. Note that oxygen cannot be
The gas analyzer analyzes the oxygen by the affiliated technologies, such as the paramagnetic or
MAC (MAC: minimum alveolar concentration ) is a standard for comparing the effect of the
inspiratory Anesthetic Gases. The MAC value stands for the density of Anesthetic Gas in the
pulmonary alveolus (one barometric pressure) which is the density that will not provoke muscular
If the mechanism of determining MAC value is carried out in the host equipment, the algorithm
applied in the calculation process must be recorded in detail. The following formula can be used
to calculate and show the MAC value by the density of the (exhaled) air at the end of the
9-2
exhalation.
For example, when using one or several anesthetics, the anesthesia module measures and get that
a 60-year-old patient‟s air at the end of the inhalation contains 4% DES, 0.5% HAL and 50% N2O.
4% 0.5% 50%
MAC -0.00269*20
-0.00269*20
2.08
6% *10 0.75% *10 100% *10-0.00269*20
Attention
The above formula is only applicable to patients over 1 year old, 1 year old or below is
The above formula does not take the altitude, other personal factors into consideration.
Paramagnetic Oxygen Analysis is to measure the suction produced by the high magnetic fields
against the oxygen molecules in the mixed gas. The paramagnetic analyzer identify oxygen
and other gases according to the magnetization susceptibility of the magnetic field.
As oxygen is paramagnetic, the oxygen will be attracted by the magnetic field, but most of
other gases will not be attracted. In terms of proportion, if the magnetization susceptibility of
the magnetic field of oxygen is designated as 100,that of most other gases is close to zero.
9-3
• Generally dispense with maintenance
9-4
9.6 Identifying the AG Module
2 5
3 Gas outlet
Note
The AG Module has been equipped with automatic atmospheric pressure compensation.
9-5
9.6.2 MASIMO AG+O2 Module (Sidestream)
2 5
3 Gas outlet
9-6
9.6.3 ARTEMA AG Module (Sidestream)
2 5
3
6
9-7
9.6.4 ARTEMA AG+O2module (sidestream)
1
2 5
3
6
9-8
Attention
The watertrap is used to collect condensing water droplets from the sampling gas to prevent
them from entering the module.When the water collected in the watertrap reaches a certain
amount,the water must be poured out before it can be used again,so as to avoid blocking the
airway.
The watertrap contains filtering material to prevent bacteria/water vapor and patient
secretions from entering the module.When used for a long time,dust or other foreign matter
will reduce the air permeability of the filter material in the watertrap,and in serious cases,the
air passage will be blocked,At this time,the watertrap must be replaced.It is recommended to
replace the watertrap once a month or when a leak/damage or serious contamination is
found.
9-9
9.7 AG module Measurement preparation
Measuring Procedures
Set the anesthesia machine to start gas analysis:
a) Install the sidestream AG module to the corresponding position on the anesthesia machine.
b) Connect the Nomoline sampling line to the gas inlet of the AG module.
c) Turn the system switch to the right side to start the anesthesia machine. The indicator at the top left
corner of the AG module is in green.
d) In the [Config] Menu→ select [Gas Module] → [AG Module]→set the items, such as [Unit] , [O2
Compensation] , [Zero Sensor].
e) When the AG module is turned on, the working mode of the module is [Measure]. But in order to ensure
that it is in the proper working state, please by all means enter the [AG Module] to make sure whether
the [Work Mode] is in the [Measure] state.
f) To connect the outlet of the sample gas to the scavenging system, or to make the gas to flow back to the
patient‟s circuit.
g) If it is green LED indication, Analyzer is available.
h) To carry out inspection before use according to the statement in the “Chapter 4 Tests Before Use(Section
4.5)”.
i) If the inspection is normal, start to monitor the Anesthetic Gas.
Sampling Sampling
line line
9-10
d) Breathe out air to the sampling pipe, and check whether the anesthesia machine shows the effective
wave pattern and values of CO2.
e) Use the finger tip to block up the sampling line, and hold on for 10s.
f) Examine whether there is obstruction warning and if the luminescent sample tube shows a red flashing
light.
g) Under proper circumstances:Carry out enclosure check on the patient‟s circuit that is linked with the
sampling line.
Measuring Procedures
a) Install the sidestream AG module to the corresponding position on the anesthesia machine.
b) Connect the Nomoline sampling line to the gas inlet of the AG module.
c) Turn the system switch to the right side to start the anesthesia machine. The indicator at the top left
corner of the AG module is in green.
d) If there is oxygen sensor, enter the maintenance mode, and choose [Config]→[System Config]→[O2
OPTION]→ [ON] .
e) Click [Zero Sensor]. While zeroing, the screen shows: In zero calibration, please wait for 30s, and start
to measure CO2 after the prompt is finished.
f) In the [Config] Menu→ select [Gas Module] → [AG Module]→set the items, such as [Unit] , [O2
Compensation] , [Zero Sensor].
g) When the AG module is turned on, the working mode of the module is “Measure”. But in order to
ensure that it is in the proper working state, please by all means enter the [AG Module] to make sure
whether the [Work Mode] is in the [Measure] state.
h) To connect the outlet of the sample gas to the scavenging system, or to make the gas to flow back to the
patient‟s circuit.
j) To carry out inspection before use according to the statement in the “Chapter 4 Tests Before Use
(Section 4.5)”.
9-11
Sampling Sampling
line line
Before connecting the Nomoline sampling line to the breathing circuit, carry out the following
steps:
a) Connect the sampling line to the interface of gas inlet of the Anesthesia Module.
b) Check whether the green light of AG Module is steadily on or not(The indication Module is
normal.).
c) Check whether the connection port of the luminescent sample tube is steadily on or not(The
indication system is normal.).
d) As for the Anesthesia Module furnished with oxygen option:Check whether the oxygen reading on
the host equipment is correct or not(21%).
e) Breathe out air to the sampling pipe, and check whether the anesthesia machine shows the effective
wave pattern and values of CO2.
f) Use the finger tip to block up the sampling line, and hold on for 10s.
g) Examine whether there is obstruction warning and if the luminescent sample tube shows a red
flashing light.
h) Under proper circumstances:Carry out enclosure check on the patient‟s circuit that is linked with the
sampling line.
Measuring Procedures
9-12
a) Install the sidestream AG module to the corresponding position on the anesthesia machine.
b) Select the appropriate Watertrap according to the patient type and push the watertrap firmly into the
watertrap socket.
c) Turn the system switch to the right side to start the anesthesia machine. The indicator at the top left
corner of the AG module is in green.
d) In the [Config] Menu→ select [Gas Module] → [AG Module]→set the items, such as [Unit] ,
[O2 Compensation] , [Zero Sensor].
e) When the AG module is turned on, the working mode of the module is “Measure”. But in order to
ensure that it is in the proper working state, please by all means enter the [AG Module] to make
sure whether the [Work Mode] is in the [Measure] state.
f) To connect the outlet of the sample gas to the scavenging system, or to make the gas to flow back to
the patient‟s circuit.
g) Connect one end of the sample tube to the sink and the other end to the patient‟s circuit.
h) To carry out inspection before use according to the statement in the “Chapter 4 Tests Before Use
(Section 4.5)”.
Warning
Children can not use the adult sink,or may cause damage to the patient.
The sink is used to collect condensing water drops in the sample tube to prevent water drops from
entering the module.When the water collected by the sink reaches a certain amount,the water
must be discharged before it can continue to be used,so as to avoid blocking the the gas path.
Make sure all connections are firm and reliable. Any leakage will result in the inclusion of
ambient air in the patient’s respiratory gas, which leads to a wrong reading.
9-13
9.7.4 ARTEMA AG+O2 Module(Sidestream)
9.8 Setting AG
Set the following items on the screen by pressing hotkey [Config] → [Gas Module] → [AG
Module] . or Press the AG setting button on the module to enter [Gas Module], in which you can
set the AG parameters.
Standby: When the anesthesia module is in standby mode, the air pump stops working for the
purpose to extend the life of the module. Meanwhile, it is displayed in the information area that
“The anesthesia module is standing by”.
Detection: When the anesthesia module is working, the green indicator light is steadily on, and
detection data is sent to the monitor.
When the anesthesia module is connected to the monitor on the normal anesthesia interface, the
anesthesia module can automatically change its work mode into “detection mode”, but the user
9-14
must confirm that it is under detection mode.
When the anesthesia apparatus is turned on for the first time, the default [Work Mode] is
[Measure] . If the current anesthesia module is in standby mode, user can start the anesthesia
module on the screen by pressing hotkey [Config] → [Gas Module] → [AG Module] →
[Work Mode] → [Measure] . Or press the Measure/Standby button on the module to switch
between the Standby/Measure mode.
Note
During calibration of the sensor, press the Measure/Standby button cannot switch the
operating mode. At this time, the Measure/Standby button on the screen does not function
either.
When anesthesia machine is restarted, all the setting of AG module before the last
shutdown will be reserved.
On the screen, select [Config] menu → [Gas Module] → [AG Module] → [CO2 Unit] , and
set it to: [mmHg] , [%] or [kPa] .
Warning
Please set the oxygen compensation based on the actual conditions, otherwise the detection result
may severely deviate from the actual value, which may lead to misdiagnosis.
There are 3 options for O2 Compensation, namely “High”, “Medium” and “Low”. “High” refers to
85% of O2 Compensation; “Medium” 50%; and “Low” 21%. When an O2 module is equipped, an
O2 sensor automatically performs O2 compensation according to the O2 concentration. Therefore,
this option is in the default state. When no O2 sensor provides automatic O2 compensation, the
compensation level can be manually set according to the actual O2 concentration (as shown in the
table below).
9-15
Oxygen concentration(%) Compensation level
0~30 High(21%)
30~70 Middle(50%)
70~100 Low(85%)
9.8.4 Zeroing
The infrared gas analyzer needs to establish a zero reference level for the CO2, N2O and anesthetic
agent gas measurement. This zero calibration is here referred to as "zeroing".
1. Automatic zeroing
AG module performs automatic zeroing by switching the gas samples from the breathing
circuit to the environmental atmosphere. The automatic zeroing is performed every 24
hoursand takes less than 10 seconds. If the AG module is equipped with an oxygen sensor,
automatic zeroing shall include calibration of air within the oxygen sensor
2. Manual zeroing
The AG module can perform automatic zeroing when necessary. User can also perform
manual zeroing by entering [Config] hotkey, opening the [Gas Module] manual, and
choosing [Zeroing] in [AG Module] . It is unnecessary to disconnect the patient‟s air way
during zeroing.
Warning
Since a successful zeroing requires the presence of ambient air (21% O2 and 0% CO2), ensure
that the module is placed in a well ventilated place. Avoid breathing near the AG module before
or during the zeroing procedure.
ISA analyzer can recognize the anesthetics automatically. When the content of one anesthetic is
under the limit, while another anesthetic accounts for a major part, the anesthesia apparatus can
detect this change, and indicate the name and data of an anesthetic.
9.12 Warnings
Please refer to content with corresponding of this section of " 8.8.7Warnings " part.
9-17
9.13 Sampling line clogged
Please refer to content with corresponding of this section of " 8.8.8Sampling line clogged " part.
9.15 Consumables
Please refer to content with corresponding of this section of " 8.8.9Consumables " part.
9.16 Maintenance
Please refer to content with corresponding of this section of " 8.8.10 Maintenance " part.
9-18
Chapter 10 Monitoring BIS
Warning
The conductive parts of the sensor and interface shall not be in contact with other conductive
parts, including the ground contact.
In high-frequency surgery, to avoid the risk of burns, the BIS sensor shall not be placed between
the surgical site and the electrosurgical device return electrode.
When using a defibrillator on a patient, BIS sensors shall not be placed between the defibrillator
pads.
Clinical efficacy, risk / benefit and application of BIS function have not been fully validated in
pediatric patients.
To reduce the risk of burns during use of the brain stimulation device (e.g. cranial stimulation of
motor-induced potentials), place the stimulation electrodes as far away from the BIS sensor as
possible and ensure that the sensor is placed as indicated in the instructions on the package.
10-1
1
2 4
10-2
Reflecting the electric
activity of muscle activity
EMG Electromyogram dB 0-100
and high frequency
artifacts.
SR is the percentage of
time the EEG is considered
SR EEG suppression ratio % 0-100
inhibited in the last 60
seconds of the session.
The SQI value reflects the
quality of the signal and
SQI Signal quality index provides the reliability of % 0-100
the BIS, SR values during
the last minute.
BIS
Once
Numeric BIS Numeric Update
frequency range / per
Update Frequency
second
Frequency
0.25 Hz
to 100
EEG
bandwidth EEG Bandwidth Hz Hz (-3
Bandwidth
dB) +/-
10%
When "impedance check in progress”, “the electrode impedance too high”, “poor signal quality”, or
“replacing the sensor” occurs, the measured values of the above four parameters will be affected.
EMG:
10-3
EMG﹤55dB; acceptable EMG.
EMG≤30dB: Optimal EMG.
SQI:
Warning
The BIS shall not be taken as the only reference to adjust the anesthesia dose.
The conductive parts of the sensor shall not be in contact with other conductive parts,
In high-frequency surgery, to reduce the risk of burns, the BIS module shall not be placed
between the surgical site and the electrosurgical device return electrode.
When using a defibrillator on a patient, BIS sensors shall not be placed between the
2) Fix the module in place on the patient through the clip on the back of the BIS module, which shall
4) Attach the BIS sensor to the designated location on the patient as instructed by the sensor.
5) Connect the BIS sensor to the patient cable. Once the device detects a valid sensor, the impedance
of all electrodes will be automatically measured and the result will be displayed in the electrode
10-4
BIS plug-in module
Electrode placement
No. of electrode
Position of Electrode Placement in English
pad
3 On temple
Warning
It is of great importance that you place the electrode in the correct position to maintain proper
The electrode can be placed on the left or right side of the scalp.
Do not place the BIS module over the patient's head so as o avoid any danger to the patient.
Pleasure make sure the patient's skin is dry. Wet sensors or salt bridges may cause false BIS
10-5
Impedance
value (KΩ)
Click to [Impedance Test Results] to update the electrode impedance test results.The measured status
and impedance values are displayed in the electrode impedance test results window:
Symbol Status Measures to Be Taken
Smoothness determines how anesthesia machines averages the BIS data. The smaller the value is, the
anesthesia machine responds more sensitively to changes in patient state. The larger the value is, the more
smooth the BIS trend is, with less variation and less artifact interference.
1) Select [Config] →[BIS Module]→ [BIS smooth rate].
2) Select [10s], [15s] or [30s].
1) When the instrument is in the process of testing, a technical alarm "Replacing the sensor" appears,
10-6
replace the sensor.
2) Select [Config] →[BIS Module]→ [Confirm sensor replacement], suggesting that "Already
replaced?," select [Yes].
Note
When anesthesia machine is restarted, all the setting of BIS module before the last
shutdown will be reserved.
10-7
Chapter 11 Trend and Logs
On the screen, select [Trend] → [Table] to open the window shown in the figure below:
1. Select [Res.], and select [1min], [2min], [4min], [5s] or [30s] as required.
2. You can view the Trend Table by one of the follow ways:
Select [ ] and [ ] buttons of parameter to move the Trend Table leftwards or rightwards and
view the data of other parameters.
Select [ ] or [ ] buttons of time to turn the pages of Trend Table leftwards or rightwards and
view more data.
11-1
11.2 Trend Graph
Graph is used to review the trend of parameter values over the time. The measured
physiological values corresponding to the time point are drawn into a curve which illustrates
the patient parameter‟s trend. Trend graph provides data review for VTe, VTi, MV, Rate,
Ppeak, FiO2, EtCO2, Plat, PEEP, Pmean,BIS, etc. for continuous 60 operating hours at the
resolution of 5s. The trend graph refreshes when the machine is restarted.
Select [Trend] → [Graph] to open the window shown in the figure below:
Attention
After the anesthesia system is powered down or restarted after shutdown, the trend graph
will be cleared.
D
A
G H K
Select Parameter for review: Select [Trend] → [Graph] → [Param], and click the [Param] button
11-2
to the desired parameter. The parameters are VTe, MV, Ppeak, FiO2 and EtCO2 etc.
View Trend Curve: Select [Trend] → [Graph] → [Time], and click the [Time] button to move the
trend curve at an interval of a whole page.
Move Cursor: Select [Trend] → [Graph] → [Cursor], and click [Cursor] button to move the cursor
at an interval of 5s. The time and parameter values corresponding to the cursor place are displayed at
right side of the trend graph,. The time and parameter values change as the cursor moves. You can
change the time interval of the cursor by selecting [Trend] → [Graph] → [Res.]. The range is 5s- 4min.
On the screen, select [Log] to open a window as shown in the figure below:
Alarm log can store all physiologic alarms, technical alarms, and setting information. The logging
is timer-sequenced, and the latest event appears foremost.
1. Click [+] behind the log information to expand the physiological parameter
information during logging,click [-] tucked the physiological parameter information.
2. Select [Pre page] or [Next page] to view the alarm log one by one.
11-3
3. Select [Clear] to delete all logs.
4. Select menu “ ”, and select the arrow button to choose the desired log
filter.You can select [All](Display All), [Tech](Technical alarm), [Phys]( Physiologic
Alarm) , [Indicate], [Alarm record] or [Setting] .
Caution
When the anesthesia machine is completely powered off or turned off, the stored alarm log
are not deleted, and the log contents remain,but the shutdown time will not be saved in log.
If auditory alarms do not generate any longer, you can access the alarm log to view the
events that trigger alarms.
11-4
Chapter 12 Maintenance, Cleaning and Disinfection
Warning
Observe the applicable regulations for safety protection.
Carefully read the safety instructions of each cleaning agent to understand the applicable
materials.
Read carefully the instructions for operation and maintenance of all sterilization equipment.
Wear safety gloves and spectacles. Damaged oxygen sensor may cause leakage and result in
inflammation (including potassium hydroxide).
Reusing non-sterilized breathing system and its reusable attachments may cause cross
infection; therefore, they shall be sterilized prior to each surgical operation.
To prevent the breathing system from leaking, all components must not damaged during
disassembling and reassembling, and correct mounting shall be guaranteed, especially the
assembling of the seal rings. To conduct cleaning and sterilization, guarantee the
applicability of cleaning and sterilization methods to the components, and guarantee the
correctness of the cleaning and sterilization methods.
Please perform removal and mounting as described in this chapter. For details of further
removal and assembling, contact After-service Department of the Company. Incorrect
removal and assembling may cause leak in the breathing system, and impact normal of the
equipment.
For the maintenance, cleaning, disinfection of the external vacuum suction system, please
refer to the User Manual of vacuum suction system.
12-1
Caution
Prior to initial use, the equipment shall be cleaned and disinfected as required. The cleaning
and sterilizing methods are described in this chapter.
To avoid damages to the equipment, refer to data provided by manufacturer if you have any
questions about the cleaning agents.
Never use organic, halogenated or petroleum-base solvents, glass cleaners, acetone or other
irritative cleaning agents.
Please don’t use any abrasive cleaning agents (for example steel wool, silver polishing
materials or cleaning agents).
Do not allow any liquid to infiltrate into shell body of the equipment.
For parts made of synthetic rubber, the soak time must not exceed 15 minutes so as not to
lead to expansion or accelerated aging.
Only the parts marked with 134ºC allow sterilization by high-temperature vapor.
The maximum using times of high temperature and high pressure sterilization of respirator
y circuit is 2000 times.
The breathing circuit must be cleaned and sterilized before use by each patient.
2. Use distilled water to wet the lint-free cloth and properly clean the display, work table, and
the outer surface of the main unit.
3. Please disinfect with sodium hypochlorite solution, then use medical alcohol (75%) deterge
nt, PH value between 7.0 and 10.5, use a lint-free cloth to wipe the display, workbench, the
outer surface of the main machine, etc.
4. The casing is free from signs of deterioration that can be detected by the naked eye, and the
marking of the silk screen does not change.
Warning
Liquid infiltrating into the control assemblies may damage the equipment or cause personal
injury. During cleaning the housing, ensure that no liquid enters into the control
assemblies, and the equipment shall be disconnected from AC supply. Ensure that AC
12-2
supply is reconnected only when the cleaned components get dry thoroughly.
Caution
Use only dry soft lint-free cloth to clean the display screen, and do not use liquid to clean it.
12-3
system circuit
2 Expiratory/ Inspiratory check valve 10 Sealing Washer
cap
3 Expiratory/ Inspiratory check valve 11 Collar Clamp
pedestal
4 Expiratory/ Inspiratory check valve 12 POP-OFF Valve
5 Flow sensor 13 Folded Sack Holder
8 CO2 canister
Components marked with 134 °C are high temperature and pressure resistant, They can be
hand-washed or machine-washed (using soft cleaning agent with pH <10.5), and then be
rinsed thoroughly and air-dried. All components, exception for the oxygen sensors, airway
gauges and disposable flow sensors, can be washed.
If the flow sensors are made of plastics, please refer to the directive rules given in “ 12.3.8
Flow Sensor”.
Warning
CO2 absorbent is a type of high-causticity substance, and is very harmful to eyes, skin and
breathing system of human beings. In case any body parts are stained with CO2 absorbent
by accident, wash parts with water. If irritation is not eliminated after washing, see a
12-4
doctor for help immediately.
2 Move outward the oxygen sensor along the channel, and take it out.
Caution
Never immerse an O2 sensor or its connector in any type of liquid. Dispose of the O2 sensor
according to the manufacturer’s specifications.
Caution
To disassemble the Breathing Tube, hold the connectors at both ends of the Breathing Tube
so as not to damage the Breathing Tube.
Do not reuse filters. Discarded filters shall be disposed as per local laws and regulations, or
hospital disposal regulations. Do not directly throw them away.
2 Remove the filter from Y-piece, and then remove the Y-piece.
3 Remove the expiratory hose from expiratory connector,and inspiratory hose from
inspiratory connector of the breathing system.
12-5
2
12.2.4 Disassemble the Manual bag
Just remove the manual bag from the breathing system, as shown in the figure below:
2. When the CPC connector latch is released, pull out the airway gauge.
12-6
12.2.6 Disassemble the Manual Support Column
1 Rotate the locking nut counterclockwise.
2 Hold the manual support column with your right hand, and remove the manual support
column from breathing system with your left hand.
12-7
2 Remove the folded sack from the folded sack holder.
3 Press the collar clamp towards the center, and remove the folded sack holder.
12-8
Warning
Never disassemble the pressure relief valve. Otherwise, the pedestal and diaphragm may be
damaged and further endanger patient safety.
1
2 Remove the breathing tube connector from the breathing interface horizontally.
12-9
3 Pull the flow sensor out of the respiration port horizontally.
12-10
12.2.11 Disassemble the Breathing Circuit System
1 Ensure that all the above-mentioned assemblies
are disassembled. Then hold up the breathing
circuit system by one hand, and push and open
the fixation latch of the circuit adapter by the
other hand.
Caution
If it is very hard to push in or take out the breathing circuit system, apply lubricating oil onto
seal ring of airway connector of circuit adapter to reduce the frictional force.
12-11
2 Remove the 30mm outer cone connector of hose of the transfer system.
3 Remove the 30mm internal cone connector of hose of the transfer system.
4 Lift up the AGSS system from AGSS bracket and take it out.
12-12
components from anesthesia machine;
2. Hold the AGSS cover with left hand, and rotate it to the right, as shown in the following figure
4. Take out the filter of AGSS transfer system, as shown in the following figure
12-13
5. Clean the filter.
12-14
2. Unscrew the four screws between the external vacuum suction system and its support.
4. Disconnect the backup oxygen supply hose / backup air supply hose from the gas supply port.
12-15
12.2.13.2 Disassemble Internal Vacuum Suction System
Pull out the suction tube,take out the collecting bottle,and discard the filter.
Filter
Suction
tube
collecting
bottle
Caution
When replacling the filter,please follow the local treatment rules to dispose the old filter.
The flow sensors are plastic products, and the specific cleaning procedures are described in
12.3.8 Flow Sensor.
Warning
Never use talcum, zinc stearate, calcium carbonate, cornstarch or similar materials to avoid
12-16
adhesion. These materials might access lung of the patient or the gas duct, resulting in irritation
or damage.
Never soak breathing system and oxygen sensor into the liquid together or treat them in
high-pressure and high-temperature environment.
Check the components for damages, and replace them when necessary.
The breathing circuit must be cleaned and sterilized before use by each patient.
The maximum using times of high temperature and high pressure sterilization of respiratory
All components of breathing system of the anesthesia machine can be cleaned and sterilized.
Different components have their own cleaning and sterilizing requirements.
Components of breathing system of the anesthesia machine shall be cleaned and sterilized in time
as per actual conditions in order to avoid cross infection of patients served by the anesthesia machine.
The following cleaning ,disinfecting and sterilizing methods for components are recommended by
the Company.
Table 12.3-1 Recommended method for cleaning and disinfecting various components
Intermediate
Component Cleaning Sterilization
disinfection
CO2 Canister ★
Oxygen sensor ★
Breathing Tube,Y-piece and mask ★
Manual bag ★
Airway pressure gauge ★
Manual support column ★
Bellows Assembly ★
Flow sensor ★
Expiratory Check Valve Assembly ★
Inspiratory Check Valve Assembly ★
★: It indicates that this recommended cleaning and disinfection method can be utilized.
Cleaning: Gently wipe the surface of the cleaning parts with a damp cloth or cleaning cotton
that has been soaked in a soft, recommended cleaning agent (see Table 12.3-2 recommended
12-17
cleaning disinfectant table) to ensure that all surfaces of the parts are cleaned while This
cleaning requirement complies with the disinfection regulations and procedures of the medical
institution. After cleaning, gently wipe the surface of the cleaning part with a lint-free cloth or
medical cleaning cotton that has been soaked in medical clean water (recommended water
temperature is 40 °C). Finally, dry it with a dry, lint-free cloth.
Sterilization: First use the recommended cleaning agent (see Table 12.3-2 recommended
cleaning disinfectant table) to rinse clean parts, then use high temperature and high pressure
steam sterilization, and require this sterilization to meet the disinfection regulations and
procedures of reference medical institutions, recommended method 1: Maintain effective
sterilization time at temperature 121 °C for at least 30 min; recommended method 2: maintain
effective sterilization time at temperature 134 °C for about 4 min-6 min; After disinfection and
sterilization, remove the parts and drain the remaining distilled water, then rinse with medical
clean water (recommended water temperature of 40 °C), and dry it naturally or in a ventilated
place with a temperature of less than 70 °C.
The following table is the recommended cleaning and disinfecting agent:
The cleaning disinfectants listed below have been tested that didnot damage the breathing
components. The listed cleaning disinfectants may not be available in all countries or regions.
Please comply hospital‟s instructions for the cleaning disinfectants.
Caution
For the cleaners that are not in the cleaner list, their PH value should be between 7.0 and
10.5.
12-18
disinfection method
Cleaning Medical clean water /
Soap water (weakly alkaline) /
Pure distilled water /
Intermediate Medical alcohol 75%
disinfection
Sodium hypochlorite solution 10% effective chlorine
Caution
Check the seals before installing the CO2 canister assembly. If not, replace the seals.
1. Please refer to12.2.1 Disassemble the CO2 Canisterfor Removal of the CO2 canister.
2. Clean and sterilize the CO2 canister as per the methods recommended in the table 12.3-1
recommended methods for cleaning and disinfecting various components given in 12.3
Clean,Disinfect,andAssemblethe Breathing System.
3. After cleaning and sterilization is completed,the CO2 canister is thoroughly dried, pour CO2
absorbent into the CO2canister.
4. Please refer to 5.2 Install the CO2 Absorbent Canister,assemble the CO2 Absorbent Canister onto
the breathing system.
5. Check and test the system before use. For details, see 4.5.2 Breathing System Leak Test in
Mechanical Ventilation Mode.
Warning
Never soak breathing system and oxygen sensor into the liquid together or treat them in
high-pressure and high-temperature environment.
12-19
condensed water from the measuring surface, and reassemble the oxygen sensor into the
breathing system.
1. Please refer to 12.2.2 Disassemble the oxygen sensor for disassembling the oxygen sensor;
2. Please clean the oxygen sensor as per the methods recommended in the table 12.3-1
recommended methods for cleaning and disinfecting various components given in 12.3
Clean, Disinfect,Sterilize and Assemble the Breathing System.
4. Check and test the system before use. For details, see 4.5.2 Breathing System Leak Test in
Mechanical Ventilation Mode.
Caution
When you assemble, clean or sterilize the breathing tube, hold the connector fitted at both
ends of the breathing tube, so as not to damage the breathing tube.
1. See 12.2.3 Disassemble the breathing tube, Y-piece and mask for removal of the breathing
tube, Y-piece and mask.
2. Clean and sterilize the Breathing Tube,Y-piece and mask as per the methods recommended
in the table 12.3-1 recommended methods for cleaning and disinfecting various components
given in 12.3 Clean,Disinfect,Sterilize and Assemble the Breathing System.
3. After cleaning and sterilization is completed,the Breathing Tube,Y-piece and mask are
thoroughly dried, please refer to5.1.6Assemble the Breathing Tube,Y-piece and mask, assemble
the Breathing Tube ,Y-piece and mask onto the breathing system.
4. Check and test the system before use. For details, see 4.5.2 Breathing System Leak Test in
Mechanical Ventilation Mode.
2. Clean and sterilize the manual bag as per the methods recommended in the table 12.3-1
recommended methods for cleaning and disinfecting various components given in12.3
Clean, Disinfect, Sterilize and Assemble the Breathing System.
3. After cleaning and sterilization is completed,verify that the manual bag is thoroughly dry,
and then assemble the manual bag as per5.1.3Assemble the Manual bag.
4. Check and test the system before use. For details, see 4.5.3 Breathing System Leak Test in
12-20
Manual Ventilation Mode.
2. Clean the airway pressure gauge as per the methods recommended in the table 12.3-1
recommended methods for cleaning and disinfecting various components given in12.3
Clean, Disinfect,Sterilize and Assemble the Breathing System.
3. After cleaning is completed,verify that theairway pressure gauge is thoroughly dry, and
then assemble the airway pressure gauge as per5.1.8 Assemble the Airway Pressure Gauge.
4. Check and test the system before use. For details, see 4.5.2 Breathing System Leak Test in
Mechanical Ventilation Mode.
2. Clean and sterilize the Manual support column as per the methods recommended in the
table 12.3-1 recommended methods for cleaning and disinfecting various components given
in12.3 Clean, Disinfect,Sterilize and Assemble the Breathing System.
4. Check and test the system before use. For details, see 4.5.3 Breathing System Leak Test in
Manual Ventilation Mode.
Caution
The folded sack assembly shall not be soaked in warm water and cleaning solution for more
than 15 minutesto prevent expansion or aging.
When you air-dry the folded sack, hang and expand it adequately. Otherwise, adhesion
might occur in the folded sack.
Disassemble the bellows assembly for cleaning; otherwise, drying may take a long time.
1. See 12.2.7 Disassemble the Bellows Assembly for Removing the Bellows Assembly.
12-21
2. Clean and sterilize the bellows assembly as per the methods recommended in the table
12.3-1 recommended methods for cleaning and disinfecting various components given
in12.3 Clean, Disinfect,Sterilize and Assemble the Breathing System.
3. After cleaning and sterilization, spread out the components of the bellows assembly and
hang them in a ventilated place with a temperature less than 70 °C to dry;.
4. When the bellows assembly is thoroughly air-dried, check the components for damages,
and then conduct assembling as per the procedures described in 5.1.4 Assemble the
BellowsComponents. Connect the bellows assembly, ventilator and breathing system.
5. Check and test the system before use. For details, see 4.5.1 Bellows Tightness Test.
2. Clean the flow sensor as per the regulations issued by the hospital or as per the methods
recommended in the table 12.3-1 recommended methods for cleaning and disinfecting
various components given in 12.3 Clean, Disinfect, Sterilize and Assemble the Breathing
System.
3. After the cleaning and disinfection is completed, after the flow sensor is completely dry,
please refer to 5.1.5 Assemble the Flow Sensor to install the flow sensor;
4. Check and test the system before use. For details, see 4.5.2 Breathing System Leak Test in
Mechanical Ventilation Mode.
Caution
Never have the flow sensors in high-pressure and high-temperature environment for
disinfection.
Never use high pressure gas or brush to clean the flow sensors.
Never use unapproved cleaning agent containing polycarbonate.
Do not clean the inner surface of flow sensor, only use a damping cloth to wipe its outer
surface.
Warning
To assemble a flow sensor, tighten the locknut of respiration port; otherwise the measuring
function of flow sensors may be disabled.
The respiration port that is connected to the respiration hose shall face downward;
12-22
Otherwise condensed moisture may flow into the breathing system and impact the flow
sensor measurement.
2. Clean and sterilize the Expiratory check valve assembly as per the methods recommended
in the table 12.3-1 recommended methods for cleaning and disinfecting various
components given in12.3 Clean, Disinfect, Sterilize and Assemble the Breathing System.
3. After the cleaning and sterilization is completed, after the Expiratory check valve assembly
is completely dry, assemble the Expiratory check valve assembly according to the reverse
process of dismounting the Expiratory check valve assembly, according to 12.2.9
Disassemble Expiratory Check Valve Assembly.
4. Check and test the system before use. For details, see 4.5.5 Check Valve Inspection and Test.
Warning
Never disassemble the check valve diaphragm and check valve cover.
2. Clean and sterilize the Inspiratory check valve assembly as per the methods recommended
in the table 12.3-1 recommended methods for cleaning and disinfecting various
components given in12.3 Clean, Disinfect, Sterilize and Assemble the Breathing System.
3. After the cleaning and sterilization is completed, after the Inspiratory check valve assembly
is completely dry, assemble the Inspiratory check valve assembly according to the reverse
process of dismounting the Inspiratory check valve assembly according to 12.2.10
Disassemble Inspiratory Check Valve Assembly.
4. Check and test the system before use. For details, see 4.5.5 Check Valve Inspection and Test.
Warning
Never disassemble the check valve diaphragm and check valve cover.
12-23
12.3.11 Breathing Circuit system
1. After ensuring that the components described above have been removed, see 12.2.11
Disassemble the Breathing Circuit System for Disassembly of the Breathing Circuit
System.
2. Clean and sterilize the breathing circuit system as per the methods recommended in the
table 12.3-1 recommended methods for cleaning and disinfecting various components
given in12.3 Clean,Disinfect,Sterilize and Assemble the Breathing System.
3. After sterilization is completed, the breathing circuit system is hung upside down in a
ventilated place with a temperature less than 70 °C to dry;
4. After cleaning and sterilization, after the Breathing System has completely dried, please
refer to 5.1.1 Assemble the Breathing Circuit Systemto install the breathing circuit system,
and then install the components as described above to ensure the formation of a complete
breathing system.
5. Check and test the system before use. For details, see 4.5.2 Breathing System Leak Test in
Mechanical Ventilation Mode.
2. Shake the removed filter and shake the dust and impurities from the filter net until it
reaches a satisfactory cleaning effect;
3. Clean the AGSS transfer and receiving system as per the methods recommended in the
table 12.3-1 recommended methods for cleaning and disinfecting various components
given in12.3 Clean, Disinfect, Sterilize and Assemble the Breathing System.
4. After the cleaning is completed, after the AGSS Transfer and Receiving System is
completely dry, please refer to 5.10.2 Assemble the AGSS to install the AGSS system.
5. Check and test the system before use. For details, see 4.7 AGSS Transfer and Receiving
System Test.
Caution
Before use must check ensure that air pressure in the rang of 280-550kPa for external
vacuum suction system.
12-24
The maximum negative pressure of vacuum suction system is not less than 75kPa.
Warning
The negative pressure regulator is prohibited from cooking in the sterilizer, and the use of
liquid disinfectant which is harmful to plastics is prohibited to disinfect the negative
pressure regulator, which will cause serious damage to the plastic parts of the negative
pressure regulator. Influence the function of negative pressure regulator.
5. When not in use, the overflow cup and the collecting bottle can be separately stored and
disinfected;
6. After the negative pressure suction system is disinfected, vacuum suction should be placed in a
well-ventilated environment,so that residual moisture or disinfectant can be volatile.Then please
refer to 5.9 Installation of the Vacuum Suction System to install the vacuum suction system.
12-25
Warning
The life of collection bottle and overflow cup is ten years.
Note
Please refer to the user manual supplied with the external vacuum suction system for
cleaning, disinfection and installation information.
12.3.14 Battery
Caution
To extend the service life of batteries, use the battery at least once a month, and charge them
when the battery charge is used up.
Please check and replace batteries regularly. The service life of batteries depends on
frequency of use and service time. Based on proper maintenance and storage, the service life
of battery is approximately 3 years. In case they are improperly used, their service life may
be reduced. It is recommended that batteries be replaced every 3 years.
In case of battery fault, contact the personnel of the manufacturer for replacement. The
user must not replace it by themselves.
The time the battery is powered depends on the configurations and operation of the device.
After the main power supply is interrupted, when the “ON-OFF” button is kept on, after the
interruption time exceeds 30s, there is internal power supply that can support normal
operation.
12-26
Chapter 13 Maintenance and Failure Recovery
Warning
Do not use the faulty anesthesia machine.All repairs and maintenance shall be carried out
by authorized service representatives.
Adopt a cleaning and disinfection plan that meets you disinfection and risk management
requirements.
Please don’t use lubricant containing oil or grease, which may have a fire or explosion
hazard when O2 reaches a certain concentration.
Used equipment might be contaminated by blood or body fluid. Please observe relevant
disinfection control and safety regulations.
Moving parts and detachable components may cause hazard of hand nipping/crushing;
therefore, move or replace system components with more care.
Do not use faulty equipment. Contact the service representative authorized by the Company
to fulfill all required maintenance, or the qualified professionals to fulfill the replacement and
maintenance of parts listed in the User‟s Manual.
When maintenance is over, test the equipment and ensure that the equipment works normally
and meet the requirements of specifications.
Caution
Personnel without experiences in maintenance of such equipment must not service the
equipment.
Replace damaged parts with the ones produced or sold by the Company. When replacement
is over, perform testings to ensure that the equipment conform to the specification
requirements of manufacturer.
If service and support are required, contact the after-service department of the Company.
To find out further product information and relevant technical data, contact the
after-service department of the company, and we may provide documentary data about
13-1
some components.
Do not immerse the O2 sensor or its connecter in any type of liquid.
Do not clean the inner surface of the O2 sensor.
Caution
In this schedule, the minimum maintenance frequency is based on a typical use of 2000
operating hours per year. If the actual use time is longer than 2000 hours, the frequency of
equipment maintenance shall be higher in a year.
When cleaning and installing,please check whether the parts and sealing rings are damaged
and replace or repair if necessary.
Minimum maintenance
Maintenance
frequency
Replace the vaporizer base and seal rings of the breathing system port. For
Yearly details, contact the after-service department of the Company.
CO2 module calibration.
Replace the built-in battery. For details, please contact the after-service
Every 3 years
department of the Company.
13-2
Minimum maintenance
Maintenance
frequency
Use a new washer on the backup cylinder port every time a new backup cylinder
is to be assembled.
If the colour of CO2 absorbent varies, replace CO2 absorbent of the canister.
If the measured deviation of oxygen sensors is too high and cannot be corrected
When necessary after being calibrated many a time, replace the oxygen sensors.
If the seal ring of flow sensor is damaged, the diaphragm is cracked or
deformed, or the sensors show deformation or cracking, replace the flow
sensors.
If the hoses and silicone hose of the transfer system are broken, replace them.
13.4 O2 Calibration
Warning
Please do not implement the calibration procedure when the system is connected to a
patient.
When calibrating an oxygen sensor, the environmental pressure must be identical with the
environmental pressure where oxygen transfers in the breathing system. If not, the
monitoring values may exceed the limits.
Before calibrating an oxygen sensor, disassemble the oxygen sensor. Verify that oxygen
sensor and the assembling place do not have accumulated water, and then re-assemble the
oxygen sensor.
If oxygen sensor is not equipped with or not intended to use, it is unnecessary to perform O2
calibration.
13-3
Caution
If calibration fails, check whether or not some technical alarm is given. Take measures to
remove the alarm, and then calibrate the O2 sensor again.
If calibration fails many a time, replace the oxygen sensor, and perform calibration over
again. If calibration still fails, contact the service personnel or the Company in time.
The discarded oxygen sensors shall be treated as per relevant regulations for biological
hazards, and please don’t have them burnt.
1. Verify that the system is in its standby mode; otherwise push the standby soft key to
access the screen of [Standby].
2. Select [O2 Calibration] menu → [21% O2 Calibration] , and open the [21% O2
Calibration] menu, and access the 21% O2 Calibration screen.
3. Remove the oxygen sensor from breathing system, and place it in the air for 2~3 minutes.
The disassembling procedures are described in 10.2.2 Disassemble the Oxygen Sensors
4. In the [21% O2 Calibration] menu, select [Timer] button to time for 3 minutes. When
timing is over, push the [Start] button, the system starts “21 % O2 Calibration”, and
[Calibrating] is displayed on the screen.
5. In process of calibration, push the [Stop] button to stop the calibration in progress.
6. If the equipment passes the testing, a prompting message [Calibration Result PASS] is
displayed on the screen. Otherwise, prompting messages [Calibration Result FAIL] and
[Repeat] will be displayed, and recalibration is required in such a case.
7. Select [Exit] .
Caution
If 100% O2 calibration fails, check whether or not some technical alarm is given. Take
measures to remove the alarm, and then calibrate the O2 sensor again.
If calibration fails many a time, replace the oxygen sensor, and perform 21% O2 calibration
13-4
over again. When 21% O2 calibration is successful, perform 100% O2 calibration. If 100%
O2 calibration still fails, contact the service personnel or the Company in time.
1. Ensure that “21% O2 Calibration” is finished and calibration is successful. Ensure that
[No O2 Pressure] does not occur.
2. Verify that the system is in its standby mode; otherwise push the standby soft key to
access the interface of [Standby] mode.
3. Select [O2 calibration] menu → [100% O2 calibration] , and open the [100 % O2
calibration] menu, and access the 100 % Oxygen calibration screen.
6. Turn on O2 supply, the regulated flow shall be at least 8L/min; turn off other gas supply.
7. Wait for 2~3 minutes, or in the [100% O2 Calibration] menu, select [Timer] button to
time for 3 minutes. When timing is over, push the [Start] button, the system starts “100%
Oxygen Calibration”, and [Calibrating] is displayed on the screen.
8. In process of calibration, if you push the [Stop] button, the calibration in progress will be
stopped.
9. If the equipment passes the testing, a prompting message [Calibration Result PASS] is
displayed in the menu. Otherwise, prompting messages [Calibration Result FALL] and
[Repeat] will be displayed, and recalibration is required in such a case.
2. Use a small flat screwdriver to open the latch of the airway gauge lens,and then remove
it.
3. Use a screwdriver to adjust the zeroing screw, and adjust the pressure gauge pointer to
zero.
13-5
4. Set the Manual/mechanical ventilation switch to its mechanical control mode.
5. Insert the Y-piece into the leak hunting plug to seal the respiratory airway.
6. Push the oxygen flush valve repeatedly to make the pressure gauge pointer swinged.
7. Remove the Y-piece from leak hunting plug and release the oxygen flush button. Then
check whether or not the gauge pointer returns to zero
9. If pointer can return to return to zero normally, directly press the lens of airway gauge,
and assemble the gauge properly. If the pointer still fails to return to zero, please contact
the after-service department of the Company.
Check the hose of transfer system. If there is any damage, please replace it.
Impurities and dusts may clog the filter during using AGSS transfer system. Use the following two ways
When using the filter, if the AGSS float isn‟t floating, the filter may be clogged.
Place the AGSS canister at horizontal level to see whether the filter is clogged.
If the filter is clogged, please refer to section 12.3.12 AGSS Transfer and Receiving System to clean or
replace it.
Warning
Removal and maintenance of equipment is recommended if liquid or solids are drawn into
the vacuum pump.
13-6
13.7.1 Internal Vacuum Suction System Maintenance
Check whether the overflow cup, liquid collection bottle, filter, suction pipe and suction head are damaged. If
there are any damages, make replacement in time. If the vacuum system is unable to function normally due to
internal system failure of the machine, contact the manufacturer for inspection and maintenance in time.
After each use, remove all the suction pipe, filter, suction head and liquid collection bottle.
Note
Please refer to the user manual supplied with the external vacuum suction system for
Maintenance information.
7. Keep the ventilation mode and allow the instrument to run for approximately 1~2 hours,
till the water accumulated inside the breathing system is removed.
13-7
the water accumulated with the manual drain valve before use again.
Please drain with the manual drain valve in the following steps:
1. Hold the manual drain valve and press upwards to open the drain valve. The water
accumulated will be discharged, as shown in the right figure:
2. After the water accumulated has been discharged, release your hand, the manual drain
valve will auto reset.
Attention
After the water accumulated has been discharged, please reset the manual drain valve, and
make sure that the anesthesia machine can work normally.
13-8
Chapter 14 Accessories
Warning
Only the accessories specified in this chapter may be used. Using other accessories may lead
to incorrect measured values or equipment failure.
Disposable accessories can be used only once, and their reuse may cause performance
reduction or cross infection.
All accessories intended to contact human body shall meet the organism compatibility
required by ISO 10993-1 Standard. They shall be compatible with non-inflammable
anesthesia gasses and anesthetics and are expected to cause no adverse reaction when they
are exposed to human body, and they shall not work with inflammable anesthesia gasses.
The discarded accessories shall be treated as per local laws and regulations, or hospital
regulations, Do not throw them directly.
Accessories that are not in direct contact with the human body do not require disinfection.
The detailed method for disinfection of accessories directly in contact with the human body
is described in the attached Manual.
When the device and its accessories reach the end of their service life, they must be disposed
of in accordance with the guidelines for the management of such products as well as local
regulations for contaminated and biohazardous goods.
To use additional bacteria filter if sample gas is to be returned to the breathing system.
14-1
(sidestream)
Access PN Recommended
ory Remarks Model replacement manufacturer
name period
mainstream O.1062
CO2/AG adapter 60
14-3
adapter (with
dehumidification
tube)
PN Recommended
Name Remarks model replacement manufacturer
period
14-4
3 L)
14-5
Pipeline 040-000274-00 1 year
High temperature
assembl 72201
L-type connector
y
14-6
disposable)
14-7
Chapter 15 Installation and Specifications
80
71 75 76
18 78 88
73 74 NO
NC
19
72
79 89
17 24 77 87
污气处理
84 91
83
81 82
16
23 92
9 70 85 86
90
7 P 11 57 58 69
1 59
54
15 52
48 53
9 93 26 29
20 31
7 P 11 25 27 28 30 95
2
13 67
10 49 68
66
8 P
3 94
12
14 33
9 21 50 60
32 34 42 55
7 P 11
4 59
39 65
43
13 41
38 40 57
44 51
36 NO 61
9 NC
22 35 37
7 P 11
5 56 64
45 O2 P
13
10 63
62
8 P
6 58
12
14 46 47
Serial Serial
Description Description
No. No.
Two-way two-port
1 Pipeline oxygen inlet 51 electromagnetic valve
(always on)
Two-way three-port
4 Pipeline air inlet 54
electromagnetic valve
15-1
7 Filter 57 ACGO
Module exhaust
19 Gas nozzle 69
channel
Tree-way two-port
Pressure regulating valve
20 70 electromagnetic valve
(0.2MPA)
(always on)
Two-way two-port
26 electromagnetic valve 76 Bellows
(always off)
15-2
28 Filter 78 Manual machine control
Electromagnetic
29 79 Manual bag
proportional valve
Two-way two-port
31 Flow sensor 81 electromagnetic valve
(always off)
Electromagnetic
32 Filter 82
proportional valve
Two-way two-port
33 electromagnetic valve 83 Peep valve
(always off)
Safety valve
34 Check valve 84
(10cmH2O)
Two-way two-port
36 electromagnetic valve 86 Gas resistance
(always off )
42 Filter 92 Atmosphere
Pressure switch
43 Check valve 93
(0.2MPA)
Two-way five-port
Tubular flow meter
44 94 bistable mechanical
(0-15L/min)
valve
46 Evaporator 96 /
Gasses are delivered from gas supplies to the system through connected pipeline or gas cylinders.
Pipeline gas supplies include 3 types i.e. O2 N2O and AIR, they enter into the system respectively
through pipeline gas supply ports 1, 2,4,5, and their operating pressure measured at the flowmeter
front-end is 200kPa. Backup cylinder gas supplies include 3 types i.e. O2, AIR and N2O(only allow to
select two of the three gases. Oxygen and nitrous oxide are taken as the examples in the figure below),
they enter into the system respectively through backup-cylinder gas supply ports3 and 6. Their ranges of
operating pressure are respectively 6.9 ~15 Mpa and 4.2~6Mpa, and are reduced to 300 ~500kPa by
pressure regulat or 14. Each type of port is provided with a clear flag and is capable of anti-misplug
function to prevent users from connect gas supplies incorrectly. Insides of all ports are fitted with filter
and check valve, and pressure gauges with color codes are used to display the pressure pipeline gas
supplies and backup cylinders. Pressure relief valve 13 is used to avoid too high input pressure of gas
supplies.
All connections are fitted with flagged gas-supply inlet port connectors, filters and check
valves. Pressure gauge displays the pressure of gas cylinders and pipelines. Regulator can
reduce the gas cylinder pressure to a proper systemic pressure. Pressure relief valve can assist
to protect the system from being damaged by high pressure.
Connections of all air cylinders are fitted with collar clamp plugs.
When supply pipelines are connected, keep the gas cylinder valves in their OFF
position.
When the system is not in use, cut off the gas supply lines.
Warning
When pipeline gas supply is in use, do not set the cylinder valve to “ON”. Cylinder gas
supply will exhaust. In such a case, gas supplies may be inadequate when a pipeline fault
occurs.
15.1.3 O2 Flow
O2 is directly transmitted to the O2 channel of gas mixer at line pressure or regulated gas
cylinder pressure. O2 may be also directly transmitted to the respiration machine if O2 is set
as driving gas. If the pressure is too low, alarm may be displayed in the display screen.
Secondary controller may reduce the pressure of quick charging valve and auxiliary O2
supply flowmeter.
15-4
When O2 button is pushed to start O2 charging, quick charging valve may provide high O2
flow (between 25 and 75l/min) to the fresh gas outlet.
Warning
When the pressure of O2 supply is lower than 100 kPa, N2O supply is automatically cut off
by O2-N2O cut-off valve, but the air supply will not be affected.
15-5
19
18 20
21
22
1 2 7 24
25
26
3 27
23 28
8
29
4 30
9
6 5 10 17
31 32
11
34
35
12 13
36 37
33
41
38
14 15 42
39 43
16
40 44 45
46
50
49 51 47
52 48
15-6
No. Component No. Component
Single tube flowmeter backlight AIR、O2 source driven selector valve
13 39
board
14 Workbench lighting switch 40 Sensor board
15 Workbench lighting board 41 Switch signal 2
16 Circuit Heater 42 Pipeline pressure sensor
17 DC/DC power board 43 Cylinder pressure sensor
18 Keyboard 44 Sensor signal adaption board
19 Shuttle 45 Flow sensor
Mechanical/Electronic flowmeter
20 Indicator light 46
selector valve
21 Alarm light 47 O2、N2O、AIR gate valve
22 Buzzer 48 Proportional valve
23 Main control board 1 49 Main control board 2
24 LCD screen 1 50 LCD screen 2
25 Touch screen 1 51 Touch screen 2
26 Alarm speaker 52 USB2
Classified by category of anti-electric shock Class I, Device With Internal Power Supply, Normal
Mobile Device
Classified by degree of anti-electric shock Defibrillation-proof Type BF Apply Part
Defibrillation-proof recovery time BIS: < 30 s; Others: <5 s
Classified by the safety degree under coexistence of Not suitable to apply in the place with flammable
flammable anesthetic gas and air or oxygen or anesthetic gas
nitrous oxide
Classified by the work mode Continuous-working Device
Classified by degree of water proof IPX0 Type
Applicable standard IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC
60601-1-8, IEC 60601-1-9, ISO 80601-2-13, IEC
60601-2-26, ISO 10079-3
The anesthesia machine is integrated with pressure limit device, expiratory gas volume
monitor, breathing system equipped with alarm system, pressure measurement device,
15-7
anesthesia breather system, vacuum suction system,anesthesia gas delivery device, anesthesia
respiration machine, O2 monitor and CO2 monitor,and it can be installed anesthesia-gas
purification/transfer and receiving system. Where:
AX-900 and AX-900A anesthesia machines conform to the standard of ISO 80601-2-13;
Pressure limit device, expiratory gas volume monitor and breathing system equipped with
an alarm system conform to Standard IEC 60601-2-13(GB 9706.29);
Pressure measurement device and anesthesia ventilation system conform to Standard ISO
80601-2-13 and YY 0635.1;
Anesthesia gas purification, transfer and receiving system conforms to Standard ISO
80601-2-13 and YY 0635.2;
Anesthesia gas delivery device conforms to the Standard ISO 80601-2-13 and YY 0635.3;
AC mains
Output voltage 220 to 240 V 220 to 240 V 220 to 240 V 220 to 240 V
AC mains
15-8
Fuse T10 AL/250 V
Output voltage 100 to 127 V 100 to 127 V 100 to 127 V 100 to 127 V
Warning
The system connected with auxiliary power must be certified by the designated IEC
standards (e.g. IEC 60950 Data Processing Equipment Standard and IEC 60601-1 Medical
Equipment Standard).
The device provides four power sockets for auxiliary equipment of anesthetic systems (eg,
evaporator, gas analyzer), do not connect other devices with the sockets, it may effect the
patient's leakage current. Overload is not allowed.
Internal battery
Length: 5m
15-9
15A for 100 -127V~
1. General
15-10
Respiration rate 0 to 150 ±1 breaths/min
Fi and ET Fi and ET are displayed after one breath and have a
continuously updated breath average.
ET will typically decrease below nominal value
(ETnom ) when respiration rate (RR) exceeds the RR
threshold (RRth) according to the following formulas:
CO2
CO2 ET=ETnom×(125RR⁄) for RRth >125
OR+/AX+
CO2 ET=ETnom×√(70RR⁄) for RRth >70 N2O, O2,
DES, ENF, ISO, SEV ET=ETnom×√(50RR⁄) for
RRth >50 HAL ET=ETnom×√(35RR⁄) for RRth >35
Automatic agent identification OR+/AX+: Primary and secondary agent.
3. Gas Analyzer
15-11
OR+/AX+
< 4 seconds
(with 2 m Nomoline Airway Adapter Set sampling line)
4. Gas
The accuracy of all detected values meet the requirements of ISO 80601-2-55 and EN 864:1996.
Accuracy for standard conditions The following accuracy apply to a dry gas at 22±5°C and 1013±40hPa.
Gas Accuracy
CO2 ±(0.3 kPa + 4% of reading)
N2O ±(2 kPa + 5% of reading)
Agents ±(0.2 kPa + 10% of reading)
O2 ±(2 kPa + 2% of reading)
Breath detection: adaptive threshold value, minimum CO2 concentration change is 1vol%
Threshold value of Anesthetic gases: main Anesthetic gas (OR+/AX+):0.15 vol% When an Anesthetic gas
is marked, its concentration will be reported even if it is below 0.15 vol%.
15-12
Carbon dioxide Anesthetic nitrogen
Gases or water vapor Gas concentration
Carbon dioxide AX+ gas monoxide
nitrogen monoxide 4) 60 vol% _2) _1) _1) _1)
HAL4) 4 vol% _1) _1) _1) _1)
Enflurane, isoflurane, +8% of the _2) _1) _1)
5 vol%
sevoflurane reading 3)
+12% of the _1) _1) _1)
Desflurane 15 vol%
reading 3)
Xe(xenon) 80 vol% -10% of the reading 3) _1) _1)
He(helium) 50 vol% -6% of the reading 3) _1) _1)
Quantitative spray Quantitative spray
C2H50H(ethanol) 0.3 VOI% _1) _1) _1) _1)
C3H70H(isopropanol) 0.5 VOI% _1) _1) _1) _1)
CH3COCH3(acetone) 1 vol% _1) _1) _1) _1)
CH4(methane) 3 vol% _1) _1) _1) _1)
CO(carbon monoxide) 1 vol% _1) _1) _1) _1)
Note 1: The abovementioned “Accuracy – all conditions” includes all negligible interferences and impacts.
Note 2: The abovementioned “Accuracy – all conditions” includes negligible interferences and impacts
when concentrations of nitrogen monoxide and oxygen are correctly set.
Note 3: Interferences at specified gas concentrations. For example, 50 vol% helium can lower the carbon
dioxide reading by 6%. That is to say, when a gas mixture with 5.0 vol% carbon dioxide and 50 vol%
nitrogen is detected, the reading of carbon dioxide concentration is normally (1-0.06)* 5.0 vol%=4.7 vol%
carbon dioxide.
Name Specifications
EtCO2 complies with the requirements of standard YY0601
EtCO2 specification of Masimo (Mainstream)
CO2
measurement 0mmHg~190mmHg,0~25%(at760mmHg)
range
15-13
CO2 resolution 1mmHg or 0.1kPa or 0.1%
All conditions:
CO2 accuracy
±(0.3kPa+4% of the readings)
Total system
<1s
response time
Respiration rate 0~150rpm
Whether there
is automatic
barometric Configuration
pressure
compensation
Warm up time 10s
15-14
differential value between the selected time period. Minimum of 5 mmHg
baseline and the between peak and valley of waveform required.
CO2 value shall be 5 mmHg. Time Period Selections: 10 second, 20 second
CapnoTrak sidestream:ETCO2 and Respiration
Rate accuracy is verified by using a solenoid test
setup to deliver a square wave of known CO2
concentration to the device. 5% and 10% CO2
ETCO2 and
concentrations were used and respiration rate was
Respiration rate
/ varied over the range of the device. Pass/Fail
accuracy
criteria was a comparison of the respiratory rate
method
output from the sensor to the frequency of the
square wave. EtCO2 measurements at those rates
were compared to the CO2 readings under static
flow conditions.
CapnoTrak sidestream:Range: 0, 2 to 100 breaths
Respiration Range: 0 to 150 breaths per minute
per minute (br/m)
Rate (BPM)
Accuracy: ±1 breath per minute
Calculation Accuracy: ±1 breath
Method: 8 breath averaging
LoFlo sidestream: <4seconds - includes transport
time and rise time(Typical rise time:250ms)
CapnoTrak sidestream:Less than 4 seconds.
Total system Includes transport time and rise time with water
<1s
response time filter assembly and airway adapter.(Up to an
additional 3 seconds for sidestream sampling
cannulas with dehumidification and extension
tubing)
CapnoTrak sidestream:Less than 340ms.(Up to
an additional 70ms for sidestream sampling
CO2 Rise Time /
cannulas with dehumidification and extension
tubing)
Warm up time 2min /
Whether there
is automatic
barometric None /
pressure
compensation
15-15
15.5.4 Artema AG gas analysis Specification
Name Specification
BIS:0.0~100.0
ESR:0.0~100.0%
15-16
15.7 EMC and Radio Management Compliance
Anesthesia machine AX-900 and AX-900A conform to EMC Standard IEC 60601-1-2.
Caution
The user needs to install and use according to electromagnetism compatibility information
which is attached with it.
Portable and mobile RF communication devices may influence Anesthesia machines AX-900
and AX-900A performance, so Anesthesia machines AX-900 and AX-900A should be kept
away from them during using.
Warning
Anesthesia machines AX-900 and AX-900A should not be used adjacent to or stacked with
other equipment and that if adjacent or stacked use is necessary, the Anesthesia machines
AX-900 and AX-900Ashould be observed to verify normal operation in the configuration in
which it will be used.
When the plug-in slot is plugged into any one of the three modules, Masimo CO2
(mainstream), Respironics CO2 (mainstream) and BIS, EMC is of Class A. When the plug-in
slot is not plugged into these three modules, EMC is of Class B.
Table 1
Guidance and manufacturer’s declaration –electromagnetic emissions
The Anesthesia machines AX-900 and AX-900A is intended for use in the electromagnetic environment
specified below. The customer or the user of the SECP-II should assure that it is used in such an
environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions The Anesthesia machines AX-900 and AX-900A uses RF
CISPR 11 energy only for its internal function. Therefore, its RF
Group 1
emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions The Anesthesia machines AX-900 and AX-900A) is
Class B
CISPR 11 suitable for use in all establishments other than domestic
15-17
Harmonic emissions and those directly connected to the public low-voltage
Class A
IEC 61000-3-2 power supply network that supplies buildings used for
Voltage fluctuations / domestic purposes.
flicker emissions Complies
IEC 61000-3-3
Table 2
Guidance and manufacturer’s declaration – electromagnetic immunity
The Anesthesia machines AX-900 and AX-900A is intended for use in the electromagnetic environment
specified below.
The customer or the user of the Anesthesia machines AX-900 and AX-900A should assure that it is used in
such an environment.
IEC 60601 test Electromagnetic
Immunity test Compliance level
level environment –guidance
Electrostatic ±8 kV contact ±8 kV contact Floors should be wood,
discharge (ESD) ±2,4,8,15 kV air ±2,4,8,15 kV air concrete or ceramic tile. If
floors are covered with
IEC 61000-4-2 synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast ± 2 kV for power ± 2 kV for power Mains power quality should
transient/burst supply lines supply lines be that of a typical
IEC 61000-4-4 ± 1 kV for ± 1 kV for commercial or hospital
input/output lines input/output lines environment.
100 kHz repetition 100 kHz repetition
frequency frequency
Surge ± 0.5kV, ± 1 kV ± 0.5kV, ± 1 kV Mains power quality should
line(s) to lines line(s) to lines be that of a typical
IEC 61000-4-5 ± 0.5kV, ± 1 kV, ± ± 0.5kV, ± 1 kV, ± 2 commercial or hospital
2 kV line(s) to kV line(s) to earth environment.
earth
15-18
Voltage dips, short 0 % UT; 0.5 cycle 0 % UT; 0.5 cycle Mains power quality should
At 0°, 45°, 90°, At 0°, 45°, 90°,
interruptions and be that of a typical
135°, 180°, 225°, 135°, 180°, 225°,
voltage variations 270°and 315° 270°and 315° commercial or hospital
on power supply environment. If the user of
0 % UT; 1 cycle 0 % UT; 1 cycle
input lines the Anesthesia machines
and and
70 % UT; 25/30 70 % UT; 25/30 AX-900 and AX-900A
IEC 61000-4-11 cycles cycles requires continued operation
Single phase: at 0° Single phase: at 0° during power mains
interruptions, it is
0 % UT; 250/300 0 % UT; 250/300
cycles recommended that the
cycles
Anesthesia machines
AX-900 and AX-900A be
powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
Power frequency 30 A/m(50/60 Hz) 30 A/m
fields
Table 3
15-19
AX-900A, including cables, than the
recommended separation distance
calculated from the equation
Conducted RF 3Vrms (emf), 3Vrms (emf), applicable to the frequency of the
IEC 61000-4-6 6Vrms (emf) in 6Vrms (emf) transmitter.
ISM and amateur in ISM and Recommended separation distance
Radiated RF radio bands amateur radio
d = 1.2 P
IEC 61000-4-3 150kHz to 80MHz bands
d = 1.2 P 80MHz to 800MHz
3V/m 3V/m
d = 2.3 P 800MHz to 2.7GHz
80MHz to 2.7GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, ashould
be less than the compliance level in
each frequency range. b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
15-20
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and VT broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
Anesthesia machines AX-900 and AX-900A is used exceeds the applicable RF compliance level above,
the Anesthesia machines AX-900 and AX-900A)should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating
the Anesthesia machines AX-900 and AX-900A.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
15-21
absorption and reflection from structures, objects and people.
Size 689mm*800mm*1400mm
Weight 128kg (standard configuration) (without anesthesia vaporizer and gas cylinder)
Maximum Bearing Weight The maximum bearing weight of the entire machine is 210kg
15-22
Top Plate
Dimensions with
508mm*313mm*380mm
Additional Accessory
Workbench
Maximum supporting
Maximum supporting weight of workbench is 20kg
weight
Dovetail Rail
Maximum Installation
500mm*750mm*500mm
Dimension
Handrail
Length 750mm
Drawers
Top: 1kg
Maximum Bearing Weight
Bottom: 3kg
Top: 462mm*287mm*141mm
Drawers
Bottom: 437mm*287mm*245mm
Length: 425mm
Size
Height: 240mm
Loop Hook
Caster Wheels
Display Screen
15-23
Type TFT LCD, allowing touch control
Brightness Adjustable
LED Indication
When the system is powered by battery, battery indicator lamp flashes as per a
frequency of 1 Hz.
1 piece (yellow, red. It only flashes in red when high-level and medium-level
Alarm indicator lamp
alarms occurs simultaneously)
Audible Indication
It gives out alarm tone and key-stoke tones; supports multiple-level volume
Speaker
function; the alarm tones conform to IEC 60601-1-8and YY 0709.
Buzzer It may give out alarm tone in case the system cannot work normally.
Connector
1 AC supply connector
Power supply
4 auxiliary output power-supply connectors
1 RJ45 connector
1 VGA connector
15-24
Host
Relative humidity
Item Temperature (ºC) Atmospheric pressure (kPa)
(Non-condensation)
Gas Supply
Pipeline gasses O2,N2O,AIR
Backup gas-cylinder gasses O2,N2O,AIR
Pipeline gas connection NIST
Backup cylinder connection YOKE-CGA
Pressure range at inlet 280~600kPa
Filter 60-80um
15-25
Oxygen / nitrous oxide / air 0 L/min~15L/min
Display range
range:
ACGO
Rapid Oxygenation
When the button of “Rapid Oxygenation” is pressed, the quick inflation
Oxygenation
valve provides the fresh gas outlet with high flow (25-75 L/min) of oxygen.
15-26
15.10.5 Breathing System Specifications
Compliance of breathing
system and its cycle
Adult mode ≤4 mL/100Pa, pediatric mode ≤3 mL/100Pa
absorption assembly
(manual/spontaneous mode)
Leak in CO2 canister The leakage shall not be greater than 50ml/min at 3kPa.
The leakage shall not be greater than 50ml/min at 3kPa (APL valve scale
APL valve leak
mark is 75)
CO2 Absorption Apparatus
Volume of CO2 absorption
Approximately 2000ml
apparatus
The breathing circuit is equipped with heating function,which can effectively remove water accumulation in
the circuit.
Manual bag end Taper coaxial fitting of 22mm (outside) and 15 (inside)
APL Valve
15-27
The expiratory pressure/flow rate characteristics at a fresh gas flow rate of
10±1 L/min, or the maximum fresh-gas inlet flow rate specified in the
Expiratory impedance instructions for use (whichever is greater) of the Anesthetic breathing
system,including the flow at 15L/min, 30L/min, 60L/min,the expiratory
impedance of the breathing system should not exceed 0.6 kPa.
The inspiratory pressure/flow rate characteristics at a fresh gas flow rate of
10±1 L/min, or the maximum fresh-gas inlet flow rate specified in the
Inspiratory impedance instructions for use (whichever is greater) of the Anesthetic breathing
system,including the flow at 15L/min, 30L/min, 60L/min,the Inspiratory
impedance of the breathing system should not exceed 0.6kPa.
Pressure-flow Curve of APL Valve
3 0.17 0.18
10 0.21 0.22
20 0.26 0.27
30 0.33 0.34
40 0.42 0.43
50 0.53 0.54
60 0.71 0.73
70 0.93 0.94
Expiratory impedance of breathing system cycle absorption assembly (CO2 canister is filled up with
“Medisord TM” CO2 absorbent)
Inspiratory impedance of breathing system cycle absorption assembly (CO2 canister is filled up with
“Medisord TM” CO2 absorbent)
15-28
15.11 Principle and Parameter Specifications of the Ventilator
15.11.1 Principle
Volume and pressure measurements are provided by the flow sensors. Each flow sensor are
connected to the monitoring module through 2 pieces of pipes. The monitoring module
measures the change in pressure of gas flow that passes through the flow sensors, while
pressure varies along with the flow.
Ventilator uses the values related to volumes and alarms on the basis of the data provided by
the expiratory gas flow sensors. Ventilator utilizes the other inhaling flow sensor to adjust its
output so as to adapt to the variance in fresh gas flow, minor gas leak, and the gas compliance
of the respiration circuit. Patient circuit allows compliance compensation. To obtain further
higher high accuracy, small quantity of gas conducts infiltration due to gas resistance so as to
aid to maintain the constant pressure of expiratory valve.
15-29
V-PC,CPAP/PSV,PRVC,P
SVPro,SIMV- PRVC
SIMV-PC,SIMV-VC,CPA
ΔPps(Support Pressure) 3~60 cmH2O 1 cmH2O P/PSV,PSVPro,SIMV-
PRVC
VCV,SIMV-VC,PCV,SIM
PEEP (Positive
OFF, 3~30 cmH2O 1 cmH2O V-PC,CPAP/PSV,PRVC,P
end-expiratory pressure)
SVPro,SIMV- PRVC
15 mL ~ 1500
mL(VCV,SIMV-VC,S
IMV-PC,CPAP/PSV,P
15~100 ml: 5 ml
RVC,SIMV-PRVC,PS VCV,SIMV-VC,PRVC,
VT (Tidal volume) VPro); 100~300 ml: 10 ml
SIMV- PRVC
In the PCV mode,tidal 300~1500 ml: 25 ml
volume can be
detected to 5ml.
VCV,SIMV-VC,PCV,SIM
Rate(Respiratory Rate) 4~100 bpm 1 bpm V-PC,PRVC,PSVPro,SIM
V- PRVC
Apnea.IE(Apnea 4:1~1:8
0.5 CPAP/PSV
Respiratory Ratio)
OFF, 5%~60% of
Tpause (Inspiratory Pause) 1% VCV, SIMV-VC
inhaling time
SIMV-VC,SIMV-PC ,SIM
Rate (SIMV Frequency) 4~60 bpm 1 bpm
V- PRVC
level) 1% CPAP/PSV,PSVPro,SIMV-
PRVC
15-30
Minrate 2~60bpm 1bpm CPAP/PSV
SIMV-VC,SIMV-PC,PSVP
Tinsp (Inspiratory time) 0.2~5.0 s 0.1 s
ro,SIMV- PRVC
Performance of Ventilator
Maximum inspiratory flow shall not lower than 120L/min when gas supply
Inspiratory flow
pressure is 280 kPa.
Pressure limit controlling 1. Controlled by the electronic relief valve fitted inside the ventilator;
means for ventilator 2. Controlled by the mechanical relief valve fitted inside the ventilator.
MV(Per-minute
0~100 L/Min
ventilation amount)
VT(Inspiratory and
0~3000 mL
expiratory tidal volume)
FiO2(Oxygen
18~100%
concentration)
I:E(Inspiratory- expiratory
4:1~1:12
ratio)
15-31
Compl(Compliance) 0 ml/cmH2O~300 ml/cmH2O
O2 concentration of
18%~100%;
oxygen sensor
a) BIS:0.0~100.0
d) ESR:0.0~100.0%
Parameters
15 mL~60 mL:±10 mL;
60 mL~210 mL(except 60 mL):±15 mL;
VT
210 mL~1500 mL (except 210 mL): ±7% of set value.
15-32
Inspiratory pressure: ±2.5 cmH2O or ±7% of setting value, whichever is greater;
Limit pressure: ±2.5 cmH2O or ±7% of setting value, whichever is greater;
End-expiratory positive pressure: the error is not defined at the OFF state
PCV 3 cmH2O~30 cmH2O: ±2.0 cmH2O or ±8% of setting value, whichever is greater;
Support pressure: ±2.5 cmH2O or ±7% of setting value, whichever is greater;
Apnea pressure: ±2.5 cmH2O or ±7% of setting value, whichever is greater;
Trigger pressure: ±2 cmH2O
Trigger flow
±1 L/min
rate
Inspiratory Stop
±10%
level
15-33
Full electronic flow meter:
Single-pipe control mode: In the case that the oxygen flow control
accuracy is within 10% to 100% of the full scale, the scale accuracy
shall be within ±10% of the indicated value;
In the case that the air / nitrous oxide flow control accuracy is within
10% to 100% of the full scale, the scale accuracy shall be within
±10% of the indicated value;
Main flow control Total flow control mode: The oxygen concentration (air balance gas)
system control error is not greater than ±3% (V/V); the oxygen
Flow control
concentration (nitrous oxide balance gas) control error is not greater
system
than ±3% (V/V);
Flow control accuracy: In case of full scale within 10% to 100%, the
scale accuracy shall be within ±10% of the indicated value;
Backup flow control mode: The control accuracy is less than ±10%
of the reading within the pure oxygen flow range of 0 L/min to 10
L/min. Other ranges are not defined.
Auxiliary gas When the full scale is between 10% and 100%, the accuracy of the
flow control scale should be within ± 10% of the indicated value, and the other
system ranges are not defined.
Measurement Parameters
0~60ml(excluding 60ml): :±10 ml;
60ml ~ 3000ml: ± 20ml or ± 7% of the actual reading, whichever is greater, the other
VTexp
ranges are not defined.
Inspiratory tidal ± 20ml or ± 7% of the actual reading, whichever is greater, the other ranges are not
volume defined.
Pressure monitoring error: -20 cmH2O~120 cmH2O: ±2.0 cmH2O or ± 4% of setting
value, whichever is greater; the other ranges are not defined.
End-expiratory positive pressure error: 0 cmH2O~70 cmH2O: ±2.0 cmH2O or ± 4% of
setting value, whichever is greater; the other ranges are not defined.
Paw
Platform monitoring error: 0 cmH2O~120 cmH2O: ±2.0 cmH2O or ±4% of setting value,
whichever is greater; the other ranges are not defined.
Average pressure monitoring error: -20 cmH2O~120 cmH2O: ±2.0 cmH2O or ± 4% of
setting value, whichever is greater; the other ranges are not defined.
Rate ±1 bpm or ±5% of the setting value, whichever is greater; the other ranges are not defined.
15-34
2:1~1:4:±10% of actual reading
I:E 4:1~2:1 and 1:4~1:12 ±25% of actual reading
Other ranges are not defined.
0 L/min~30 L/min: ±1 L/min or ±15% of the setting value, whichever is greater; >30
MV
L/min: not defined.
0 ml/cmH2O~250 ml/cmH2O: ±0.5 ml/cmH2O or ±15% of the actual reading, whichever
Compl
is greater, the other ranges are not defined.
0 cmH2O/(L/s) ~ 20 cmH2O/(L/s) : ±10 cmH2O/(L/s) ; 20 cmH2O/(L/s) ~ 500
Resistance
cmH2O/(L/s):±50% of the actual reading;the other ranges are not defined.
Full electronic flow meter:
Single-pipe control mode: In the case that the oxygen flow
monitor accuracy is within 10% to 100% of the full scale, the
scale accuracy shall be within ±10% of the indicated value;
In the case that the air / nitrous oxide flow monitor accuracy is
Main flow control
Flow control within 10% to 100% of the full scale, the scale accuracy shall be
system
system within ±10% of the indicated value;
Backup flow control mode: within 10% to 100% of the full scale,
the scale accuracy shall be within ±10% of the indicated value,
and the other ranges are not defined;
Auxiliary gas flow When the full scale is between 10% and 100%, it should be within
control system ±10% of the indicated value, and the other ranges are not defined.
O2 concentration
of oxygen ,the other ranges are not defined.
±3%(V/V)
sensor
Alarm Setting
Parameter Setting range Remarks
15-35
Preset Low alarm limit 0mmHg~High limit -2mmHg low limit
(Respironics
CO2)
FiCO2 Alarm High alarm limit Low limit +1mmHg~76mmHg High limit is greater than
Preset
Low alarm limit 0mmHg~74mmHg low limit
(Respironics
CO2)
EtCO2 Alarm High alarm limit Low limit +2mmHg~190mmHg High limit is greater than
Preset
Low alarm limit 0mmHg~High limit -2mmHg low limit
(Masimo
CO2)
FiCO2 Alarm High alarm limit Low limit +1mmHg~99mmHg High limit is greater than
Preset
Low alarm limit 0mmHg~97mmHg low limit
(Masimo
CO2)
15-36
FiCO2 High (Low limit +2mmHg)~99mmHg Upper limit greater
alarm than lower limit
limit
Low 0mmHg~(High limit -2mmHg)
alarm
limit
EtN2O High (low limit + 2%) ~100 % Upper limit greater
alarm than lower limit
limit
Low 0%~(High limit - 2%)
alarm
limit
FiN2O High (low limit + 2%) ~100% Upper limit greater
alarm than lower limit
limit
Low 0%~(High limit - 2%)
alarm
limit
EtHAL/EtEN High (low limit +0.2%) ~25.0% Upper limit greater
F/EtISO/EtS alarm than lower limit
EV/EtDES limit
Low 0%~(High limit - 0.2%)
alarm
limit
FiHAL/FiEN High (low limit +0.2%) ~25.0% Upper limit greater
F/FiISO/FiSE alarm than lower limit
V/FiDES limit
Low 0%~(High limit - 0.2%)
alarm
limit
Note: * Typical conditions for accuracy measure:
15-37
15.12 Principle and Specifications of Oxygen Sensors
Oxygen sensor is a type of electrochemical equipment. Oxygen penetrates into the battery
through a diaphragm and oxidize the metal electrodes. This oxidation generates a current
that is directly proportional to the oxygen partial pressure formed on the transducing surface
of electric poles. Metal electrodes are progressively eliminated in the oxidation process.
For oxygen monitoring, signal processing and analysis circuit are adopted to translate battery
signaling into corresponding percentage values of oxygen concentration. System displays the
value, and compare it to the stored alarm limits. If the value falls outside the limits,
anesthesia machine gives an appropriate alarm.
Oxygen sensor
Period of Validity Term of validity shall not exceed 13 months after the
15-38
package is unpacked (under the terms defined by the
manufacturer; otherwise the valid time may be somewhat
different)
Weight 2.2Kg
Dimension 535×120×155mm(H×W×T)
Applicable laws and ISO 80601-2-13 and YY 0635-2
regulations
Pressure relief device Atmospheric pressure compensation port
AGSS-H: The float drops below the "MIN" mark on the viewing window
when the system is not operating or when the suction flow rate is less than
50 L / min.
System status indication
AGSS-L: The float drops below the "MIN" mark on the viewing window
when the system is not operating or when the suction flow rate is less than
25 L / min.
15-39
norma impedance
l 75L / min intake flow No more than 3.5cmH2O
condit impedance
ions Induced flow Not more than 50ml / min
Rated maximum suction flow AGSS air outlet no lower than 10cmH2O
pressure drop impedance
Rated minimum suction flow AGSS air outlet no lower than 20cmH2O
pressure drop impedance
Overflow Not more than 100ml / min
Leakage Under inlet air condition of 10 ±0.5L / min , less than 90ml / min
75L / min intake flow
Under No more than 10cmH2O
impedance
single
Induced flow Not more than 100ml / min
failure
Rated maximum suction flow AGSS air outlet pressure drop resistance no greater than
condit
pressure drop impedance 0.5cmH2O
ion
Overflow May exceed 100ml/min
Please refer to the user manual supplied with the external vacuum suction system for the specifications
of the external vacuum suction system.
15-40
Venturi vacuum suction
The performance category high vacuum/high flow
Maximum vacuum In case of gas source pressure within the range of 280kPa to 600kPa, the
maximum vacuum sputum excretion generated is greater than or equal to
75 kPa.
Maximum suction flow In case of gas pressure within the range of 280kPa to 600kPa, the
maximum suction flow generated is greater than or equal to 30 L/min (not
configured with the liquid collection bottle and filter).
Alarm
15-41
Chapter 16 Different types of Anesthesia Machine
Model
AX-900A AX-900
References
16-1
Model
AX-900A AX-900
References
16-2
Model
AX-900A AX-900
References
16-3
Chapter 17 Consideration for Environmentally
Conscious Design
EXPECTED SERVICE LIFE; and the Hospital. Avoid overusing the cleaning
reagents and other substances. For the reusable
accessories, clean it with specified reagent and put
away, and for the disposable one, deal with it in a
collective way and put into the specified location
where complying with the rules and regulations of
the Local and the Hospital. If not specified, please
follow the rules and regulations of the Local and the
Hospital.
Use the specified accessories and cleaning and
2) Instructions on how to use and maintain the
disinfection reagent to avoid harm to the machine
ME EQUIPMENT in order to minimize the
and accessories and reduction of the service life. Use
ENVIRONMENTAL IMPACT during its the medical device strictly following the instruction
EXPECTED SERVICE LIFE; manual. And for maintaining the medical device,
always dilute according to the manufacturer‟s
instructions or use lowest possible concentration.
Never use bleach. Do not mix disinfecting solutions
17-1
(such as bleach and ammonia) as this may result in
hazardous or poisonous gases or liquids. When there
is a need to maintain, please follow the Instruction
for Use or follow the rules and regulations of the
Hospital.
During normal use of this device, it will consume
3) Consumption during NORMAL USE (e.g.
electricity (alternate current and direct
energy, consumable materials/parts,
current-battery). The disposable electrode is also
disposables, water, gasses, consumed and shall be disposed following the rules.
17-2
Table 2 The requirements of Clause 4.5.3 and Instructions provided by manufacturer
The requirements of Clause 4.5.3 Instructions provided by manufacturer
1) The location of components and parts The battery is located on the back of the device.
within the ME equipment that contain stored Capacitors may contain stored energy or may
energy or pose other hazards that can result in pose other hazards, assembled on the PCB
an unacceptable risk to disassemblers or others boards within the device.
and methods for controlling such risks.
2) The identity and location of hazardous The battery is located on the back of the device.
substances requiring special handling and Capacitors may contain stored energy or may
treatment pose other hazards, assembled on the PCB
boards within the device.
3) Disassembly instructions sufficient for the For other hazards that may result in
safe removal of these hazardous substances unacceptable risk, the main concern is the
including radioactive sources and induced handling with battery: Risk of fire, explosion,
radioactive materials within the ME equipment. or burns. Do not crush, puncture, disassemble
or short circuit the battery. Do not dispose of
the battery in fire or water. Do not place the
battery in an environment whose temperature is
above 60℃ (140℉). Store the battery in the
-20℃(-4℉)to 60℃(140℉) environment. Use
the specified charger only. Read instructions for
use. Maximum Recommended Ambient is 45℃
(125℉).
Dispose of used batteries promptly and in an
environmentally-responsible manner. Do not
dispose of the battery in normal waste
containers. Consult your hospital administrator
to find out about local arrangements.
As for disposing of the medical device, to avoid
contaminating or infecting personnel, the
environment or other equipment, make sure you
disinfect and decontaminate the medical device
appropriately before disposing of it in
accordance with your country‟s laws for
equipment containing electrical and electronic
parts. For disposal of parts and accessories such
as thermometers, where not otherwise
17-3
specified, follow local regulations regarding
disposal of hospital waste.
17-4