SCS-1000

Sysmex Calibrator System Assay Sheet

4135 0525 16-Jun-2024

XQ-Series XP-Series pocH-100i** XS-Series*


Assay Acceptable Assay Acceptable Assay Acceptable Assay Acceptable
Parameter Target Limits Target Limits Target Limits Target Limits
WBC K/uL 7.265 7.030 - 7.500 6.88 6.58 - 7.18 7.27 6.96 - 7.58 7.378 7.087 - 7.669
RBC M/uL 4.445 4.325 - 4.565 4.335 4.248 - 4.422 4.438 4.349 - 4.527 4.472 4.403 - 4.541
HGB g/dL 12.73 12.57 - 12.89 12.22 12.10 - 12.34 12.44 12.32 - 12.56 12.63 12.53 - 12.73
HGB mmol/L 7.90 7.80 - 8.00 7.59 7.51 - 7.67 7.72 7.64 - 7.80 7.84 7.78 - 7.90
HCT % 34.61 33.45 - 35.77 33.83 33.08 - 34.58 36.00 35.20 - 36.80 37.56 36.73 - 38.39
MCV fL - - - - 78.03 77.17 - 78.89 81.12 80.23 - 82.01 83.99 82.92 - 85.06
PLT K/uL 248.3 236.3 - 260.3 254.2 241.5 - 266.9 240.1 228.1 - 252.1 245.1 236.4 - 253.8

SCS-1000 ASSAY TERM DEFINED


Assay Target - This is the assigned value for calibration.
Acceptable Limits - These limits represent the interval around the Assay Target that can be attributed to the expanded uncertainty of the total traceability chain.
A calibrator mean (n=5) that falls within these limits indicates an accurately calibrated instrument.

* XS-1000i/XS-800i - Assay target for WBC only for operation in CBC+Diff mode
**pocH-100i - Assay Target for WBC only for systems operating under software version 00-18 and following

Revised 07.2023