Magnetic Particle Testing
Magnetic Particle Testing
PERSONNEL TRAINING
PUBLICATIONS
Duane Badger
NDT Assistance
Copyright © 2006 by The American Society for Nondestructive Testing, Inc. All rights
reserved. ASNT is not responsible for the authenticity or accuracy of information herein, and
published opinions or statements do not necessarily reflect the opinion of ASNT. Products or
services that are advertised or mentioned do not carry the endorsement or recommendation of
ASNT.
ASNT exists to create a safer world by promoting the profession and technologies of
nondestructive testing.
ISBN-13: 978-1-57117-153-5
ISBN-1O: 1-57117-153-3
Student Guide
The materials in this training package are designed to provide basic knowledge of the
fundamentals of magnetic particle testing. The training program that you are participating in
will contain the following classroom hours of instruction to present the information
approximately one hour per lesson. Level II training will include lectures on all 15 chapters
with an average of approximately one hour per lesson, with emphasis on chapters 7
through 15.
The student shall assume the responsibility for reading all assignments to include the
associated with the Student Guide, attend all lectures, and participate in classroom
discussions. Short exams will be administered after each lesson to provide the student with
Your training package will contain the following materials with specific instructions and
A. Student Guide Introduction that will outline the purpose, content, and use of
training material.
B. Magnetic Particle Testing Classroom Training Book that will serve as the
The copy of the electronic lecture is identical to the presentation used by the
instructor during the lectures on each chapter. During the lecture the student
should use the Student Guide to make additional notes, and it will also be
D. Quizzes. The instructor may elect to remove the quizzes from your packet
prior to starting the course and administer them as each lesson is completed. A
quiz will be furnished for each of the lessons in the training course.
The reading assignments will be made by the instructor and will correlate with the lectures.
The Magnetic Particle Testing Classroom Training Book published by ASNT follows the
Lesson 3 - Magnetization
Lesson 5 - Equipment
Lesson 6 - Discontinuities
Lesson 7 - Principles
Lesson 11 - Demagnetization
Lesson 12 - Equipment
The following materials are available from ASNT and is suggested where the student wants
Testing.
2. ASNT Level II Study Guide: Magnetic Particle Testing Method, second edition.
3. ASNT Level II! Study Guide: Magnetic Particle Testing Method, second edition.
Lesson 3 - Magnetization 11
Circular Fields 11
Contact Plates 11
Prods 11
Discontinuity Detection 12
Longitudinal Field 12
Field Direction 12
Magnetization Methods 13
Test Materials 13
Wet Particles 13
Dry Particles 13
Longitudinal Magnetization 14
Field Strength 14
Lesson 3 - Quiz 17
Lesson 6 - Discontinuities 31
Inherent Discontinuities 31
Processing Discontinuities 31
Inservice Discontinuities 32
Summary 32
Lesson 6 - Quiz 35
Lesson 7 - Principles 37
Theories 37
Techniques 38
Magnetic Field Theory 39
Magnetic Domains 39
Magnetic Poles .39
Magnetic Fields .40
Magnetized Ring .40
Bar Magnet 40
Effect of Materials on Electromagnetic Fields 040
Units of Measure for Magnetic Flux 040
Magnetic Hysteresis Al
Magnetic Permeability Al
Lesson 7 - Quiz 43
Lesson 11 - Demagnetization 71
Purpose 71
Principles of Operation for Demagnetization 71
Reasons to Demagnetize 71
Justification for Demagnetizing 71
Reasons Not to Demagnetize 72
Demagnetization Methods 72
Heating Above Curie Point 72
Electromagnetic Demagnetization 72
Alternating Current Demagnetization 72
Direct Current Magnetization 72
Yoke Demagnetization 73
Demagnetization Practices 73
Lesson 11 - Quiz 75
Lesson 12 - Equipment 77
Stationary EqUipment 77
Portable Equipment 77
Dry Powder Test Equipment 77
Powder Selection 77
Powder Applications 77
Applicators 78
Wet Visible Test Equipment 78
Student Guide: Magnetic Particle Testing vii
Particle Selection 78
Wet Fluorescent Method 78
Test Variables 78
Surface Preparation 79
Suspension Application 79
Water Problems 79
Suspension Preparation 79
Dry Powder Liquid Concentrate 79
Paste Concentrate 80
Water Baths 80
Fluorescent Background Check 80
Oil Based Media Contamination Check 80
Settling Test 80
Fluorescent Test Materials 80
Maintenance 80
Ultraviolet Radiation 81
Lesson 12 - Quiz 83
INTRODUCTION
1. Cost effective.
2. Simple.
3. Fast.
4. The most reliable nondestructive testing method for detecting surface and near
surface discontinuities.
5. Able to detect discontinuities in ferromagnetic materials that are not open to the
surface.
3. Discontinuities disrupt the magnetic flux field and small magnetic particles will
Personnel Qualifications
2. ANSl/ASNTCP-189.
3. NAS 410.
1
4. Each company performing nondestructive testing must have their own written
1. Permanent magnets.
a. Head shot.
b. Prods.
a. Central conductor.
b. Coil.
c. Yoke.
d. Wrapped Cables.
Magnetic Fields
1. Permanent magnets and longitudinally magnetized test objects have a north and
south pole.
contact.
7. A discontinuity will interfere with the magnetic field and form north and south
2. Magnetic particle testing only works on ferromagnetic materials that are strongly
attracted to a magnet.
testing.)
2. Paramagnetic are weakly attracted materials. (Not tested with magnetic particle
testing.)
3. Diamagnetic are weakly repelled. (Not tested with magnetic particle testing.)
4. Nonferromagnetic are not attracted or repelled. (Not tested with magnetic particle
testing.)
1. Magnetic lines of force are strongest and most dense at the poles.
2. Magnetic lines of force always seek the shortest path of least resistance which is
through metal.
•
• Lesson 1
Quiz
1. Magnetic particle testing will detect 5. Magnetic particle testing can be used on
c. Paramagnetic materials.
6. Ferromagnetic material is:
d. Diamagnetic materials.
a. Strongly attracted by a magnet.
•
a. True d. Not capable of being magnetized.
b. False
7. Nonferromagnetic materials are strongly
a. True b. False
b. False
8. A material is easily magnetized if it is:
a. True
9. Diamagnetic materials are weakly
b. False
repelled by a magnetic field .
• a.
b.
True
False
5
10. Magnetic lines of force (magnetic flux):
resistance.
on paramagnetic materials.
a. True
b. False
south poles.
a. True
b. False
a. True
b. False
a. True
b. False
SURFACE DISCONTINUITIES
1. If linear and 90° from the magnetic field direction, discontinuities will form poles
and a leakage field that will attract magnetic particles and form a visible
indication.
SCRATCHES
Scratches sometimes attract particles in a very strong field, and the field strength may have
SUBSURFACE DISCONTINUITIES
I. Subsurface discontinuities are broad and fuzzy and more easily blown or rinsed
2. Subsurface discontinuities are more difficult to detect unless larger than 0.15 cm
(0.06 in.) and especially if over 0.15 cm (0.06 in.) deep when direct current must
be used.
7
Notes
Lesson 2
Quiz
1. Surface discontinuities that disrupt the 4. Wet fluorescent techniques will produce
flow of magnetic flux lines will attract indications that are easiest to see.
a. True
5. The size, orientation and composition of
b. False
subsurface discontinuities does not
a. True
b. False
objects.
a. True
b. False
9
Lesson 3
Magnetization
CIRCULAR FIELDS
1. Through direct magnetization, circular fields are created when current is passed
through a test object (head shot or prods), but these techniques can cause arc
burns.
conductor or other indirect induction is used (no damage possible from arc burns).
3. The magnetic field direction is always 90 0 from the current flow unless distorted
due to the test object shape or the interaction from the two fields created.
Contact Plates
Stationary unit direct contact copper pads for contact head shot must be clean, with no
Prods
3. Prods are not used much because of the danger of arc burns.
4. Heavy bulky equipment with few advantages over the yoke method for surface
discontinuities.
11
Discontinuity Detection
A discontinuity that is 45 to 90° from the magnetic field direction will disrupt the magnetic
field and will create new poles. These poles will attract small particles and form an
Longitudinal Field
3. Yoke.
and verified.
6. The magnetic field strength increases with the number of turns in the coil or
Field Direction
1. The field direction is 90° from current flow for direct current flow through the test
object.
2. The field direction is 90° from the current flow through wrapped cables of a
solenoid.
3. Odd shaped test objects will have distorted field direction that require verification
The following determines the test parameters to be followed by the technique supplied to the
1. The service intended for the test object, and what discontinuities are acceptable for
that service.
2. The sequence of operations including field strength and direction verification and
demagnetization requirements.
3. The method to be used, including current type, direction, required field strength
and direction.
TEST MATERIALS
Wet Particles
Suspended in an oil or water carrier controlled for wettability, corrosion and anti-foaming.
Dry Particles
1. Come in various colors for contrast and different sizes and shapes to form
indications better.
2. The method and type of particles, application will be in the work instruction.
2.
LONGITUDINAL MAGNETIZATION
Field direction will always be 90° to current flow, except for distorted fields in odd
•
shaped test objects.
Field Strength
1. Field strength will always be maximum at the surface of test objects, and also
maximum at the surface closest to the conductor that induced the field.
length.
•
Notes
Lesson 3
Quiz
]. A circular magnetic field is created in a 5. Where current flows through the test
a. True a. True
b. False b. False
2. Prod magnetization is most effective 6. Magnetic field density remains the same
when the prods are spaced 2.5 to 5 cm regardless of the size or shape of the
a. True a. True
b. False b. False
3. The skin effect is when the 7. The right hand rule demonstrates that
Increases. a. True
a. True b. False
b. False
8. The magnetic field created by a head
magnetization and creates a circular the entire length of the test object.
a. True b. False
b. False
17
9. Discontinuities such as surface breaking 15. Wrapping several turns of cable around
b. False a. True
b. False
10. Prods are often used to test welds.
b. False a. True
b. False
11. Coils can accommodate test objects of
multiple shapes.
a. True
b. False
coil.
a. True
b. False
magnet.
a. True
b. False
a. True
b. False
18 Personnel Training Publications
Lesson 4
Principles of Demagnetization
INTRODUCTION
1. Ferrous materials usually retain some residual magnetic field after the magnetizing
current is removed.
2. Hard, alloy or high carbon steels usually have a higher residual field.
A simple and fast process uses an alternating coil to reduce residual magnetism to ideally
zero but less than 2 or 3 G by remagnetizing a test object. Then the technician reduces the
field strength to zero by increasing the distance from the energized alternating current in the
1. Some stationary bench units have a reversing step down demagnetizing coil
which, theoretically, will demagnetize deeper fields induced with direct current
2. With either alternating current or direct current, the initial magnetic field strength
in the coil must first initially be stronger than the residual field present.
3. The demagnetizing force must reverse which is normal for alternating current coils
4. The demagnetizing force must be reduced, usually by distance from the alternating
current coil and an amperage step down circuit for direct current.
19
5. Demagnetization may be aided by vibrations or by striking with a hammer during
6. Residual field strength is measured with a pocket size field indicator that measures
Efficiency of Demagnetization
1. Demagnetization can only be measured accurately if the last field induced was a
longitudinal field.
2. It is very difficult to measure or detect if a round test object has a residual circular
field.
Reasons to Demagnetize
1. Difficulty in postcleaning.
4. Inservice is in a moving test object. Particles attracted to the field can cause wear.
7. Demagnetization is usually not required for assemblies of low carbon steel with
low retentivity, if the material is heat treated above its curie temperature or
Quiz
after the magnetizing current is turned heat treated above its curie point after
a. True a. True
b. False b. False
a. True a. True
b. False b. False
is shut off.
a. True
b. False
a. True
b. False
23
Lesson 5
Equipment
EQUIPMENT SELECTION
2. Automation or manual.
3. Demagnetization requirements.
6. Accessories required.
Purpose of Test
2. The equipment used will be specified in the work instruction for a Level I or II.
technique, approval by a Level III may be required for the technician to follow.
25
Area Tested
1. The area tested will be in accordance with customer and contract requirements, and
a Level III approved written instruction wi11 normally be provided for the Level I
or Level II technician.
MOBILE EQUIPMENT
1. Mobile equipment can be moved around a shop and usual1y moved in a truck.
2. Provides about 3000 to 4000 A, and can be a power supply for prods or
wrapped cables.
3. The power supply is usually 220 or 440 V and may provide either alternating or
STATIONARY EQUIPMENT
I. Stationary horizontal wet units have two adjustable contact heads that can be used
2. The unit has a tank, pump and agitation and circulation system for the wet solution
used for testing, and a nozzle for wetting the test object.
3. The unit has amperage controls, current timers, an air or hydraulic cylinder for
clamping test objects, and usually supplied with three phase 220 or 440 V. It may
provide either low voltage alternating or direct current at high current for
magnetization.
leakage fields, have a low residual field and must be tested for standard
3. Visible or fluorescent.
•
•
Lesson 5
Quiz
b. False a. True
b. False
2. Level I personnel may accept or reject
•
a. True
3. Under the direction of Level II or Level
b. False
III personnel, Level I personnel may
a. True a. True
b. False b. False
supply.
a. True
b. False
• 29
Lesson 6
Discontinuities
INHERENT DISCONTINUITIES
Inherent discontinuities are formed during the initial melting and solidification process.
4. Blowholes are caused by gas pockets escaping the molten metal. Blowholes in
casting may be caused when moisture in the mold flashing turns into steam and
PROCESSING DISCONTINUITIES
discontinuities.
8. Forging flash line tears are caused by improper trimming or stress build-up.
9. Casting cold shuts are caused by molten metal flowing together but not fusing
together.
10. Casting hot tears and shrinkage are caused by unequal cooling and stress build-up.
31
11. Weld lack of fusion and lack of penetration are caused by improper weld joint
12. Weld heat affected zone cracks can be caused by improper preheat, heat input
13. Weld crater cracks are caused by improper technique for stopping the weld.
15. Machine tears are caused by dull tools or excessive depth of cut.
16. Cracks can be caused by a multitude of reasons at any stage during manufacturing
control.
INSERVICE DISCONTINUITIES
1. The most common inservice discontinuities are fatigue cracks and corrosion.
2. Cracks are usually at or near the surface and propagate longer and deeper. They
may also start as internal cracks and propagate longer and toward the surface.
3. Cracks can happen as a single cycle fatigue crack or from multiple cycles.
SUMMARY
1. Magnetic particle testing is used to detect surface and some slightly subsurface
2. Some specifications require half wave or full wave rectified current for
Quiz
a. True a. True
b. False b. False
a. True
b. False
a. True
b. False
discontinuities.
a. True
b. False
a. True
b. False
35
Lesson 7
Principles
THEORIES
1. Magnetic particle testing will reveal surface and slightly subsurface discontinuities
in ferromagnetic materials.
3. Magnetic particle testing may be used for raw material, in-process during
weld.
4. Final product testing is used on the final test objects to prevent test objects with
5. Maintenance tests are used for inservice discontinuities that may have been
6. Magnetic particle testing is based on the principle that magnetic flux lines will be
disrupted by a discontinuity, and this distortion will cause some of the magnetic
flux lines to exit and reenter the test object forming poles at the discontinuity. This
is called magnetic flux leakage and will attract fine magnetic particles that form an
2. Establish a suitable magnetic flux in two directions at 90° in the test object axis.
37
4. Test the object for discontinuities under suitable lighting conditions.
6. Demagnetize, if necessary.
TECHNIQUES
Level II technicians will be provided with a standard written procedure or technique for each
particular magnetic particle test. The technique will depend on the following factors.
2. Type of discontinuity.
5. The type and amount of electric current to be used for magnetization. Alternating
current has several advantages for surface and very near (less than 0.15 cm
[0.06 in.]) surface discontinuities. The rapid current reversal and skin effect
provides better particle mobility to the discontinuity flux leakage site. This is less
pronounced in the wet method because particles may be rinsed away by the liquid
flow.
6. The direction of the current flow and the resulting magnetic field direction
Magnetic Domains
domains that will have positive and negative polarity at opposite ends if
magnetized.
field or if current is passed directly through the test object the magnetic domains
align themselves and form a north and south pole in the test object.
Magnetic Poles
Magnets will have two or more opposite polarity poles, calJed north and south poles. These
poles are where the magnetic lines of force leave or enter the magnet. These lines of force
4. They seek the path of least magnetic resistance or reluctance in completing their
5. If a bar magnet is broken or cracked new magnetic poles are formed. Unlike poles
Magnetized Ring
2. A crack 90° from the magnetic field direction will create poles and an external flux
leakage field that will attract magnetic particles and form an indication.
Bar Magnet
1.
2.
If a horseshoe magnet is straightened to a bar magnet, it illustrates a longitudinally
magnetized bar.
Materials with high permeability (soft or low carbon steel) will have a stronger magnetic
field strength at a given magnetizing force than material with lower permeability (high
Magnetic flux density is the number of lines passing transversely through a unit area (gauss) .
• A magnetic hysteresis loop or curve is a graph that plots magnetizing force and magnetic
field strength.
Magnetic Permeability
Quiz
1. A magnetic flux leakage field can be 4. A test object must be strongly attracted
caused by magnetic lines of force leaving to a magnet for magnetic particle testing
discontinuity. b. False
a. True
5. Lines of force form continuous loops
b. False
that are never broken.
test: b. False
b. False
43
8. A magnetic field is made up of flux
object.
a. True
b. False
south poles.
ferromagnetic materials.
a. True
b. False
a. True
b. False
strength.
a. True
b. False
Flux Fields
ELECTRIC CURRENT
1. Alternating current.
Alternating Current
1. Is best for surface and very near surface discontinuities, and full or half wave
voltage, high amperage magnetizing current and rectify to half wave and full wave
direct current.
3. Standard United States alternating current alternates at cycles 60 times per second
discontinuities greater than 0.3 cm (0.125 in.) deep are difficult or impossible to
45
Direct Current
1. Pure direct current is only obtained from batteries or generators and is seldom
used.
•
2. Half wave and full wave direct current will penetrate deeper in the test object than
alternating current.
1. Half wave rectifiers block the negative current flow and creates a pulsing current.
2. Provides the greatest sensitivity for detecting discontinuities that are below the
surface.
•
Full wave rectifiers block the negative current flow and inverts it so the number of positive
pulses is doubled.
Commercial electric power can be provided as three phase alternating current, with each
phase providing part of the total current. All three phases are inverted and rectified to
produce the desired current, but the current draw and electric bill may be reduced by nearly
half.
•
Notes
Lesson 8
Quiz
1. The standard available current in the 5. The three most common types of
United States is 60 cycles per second, electric current used for magnetic
b. 60 ohms. current.
c. 60 Hz. a. True
d. 60 cy. b. False
a. True discontinuities.
b. False a. True
b. False
3. Rectified alternating current is
sometimes referred to as direct current 7. Half wave direct current has the greatest
a. True discontinuities.
b. False a. True
b. False
4. A half wave rectifier:
a. Reverses the current flow. 8. Three-phase full wave direct current has
b. Reverses half the current flow. all the advantages of single-phase full
49
Lesson 9
INTRODUCTION
1. Particles used are a combination of iron and iron oxide with high permeability and
2. Particles are small (0.0005 to 0.015 cm or 0.0002 to 0.006 in.), and some are round
3. Particles may be applied dry or in a wet media and with various high contrast with
the test object surface. The wet media can be spray can mixes, or oil or water
suspensions. Water requires wetting and anti-corrosive additives, and both must be
4. Particles may be used with visible light and various colors or fluorescent that give
Indications are more readily seen if the application is very light to eliminate
confusing background.
1. There are two methods of processing: continuous and residual. Normally the
interpretation of indications.
2. The dry powder application should be very light while the current is applied so
51
3. The wet solution should be applied and the flow diverted at the instant the current
4. The test object should be positioned so gravity flow of the liquid during the current
5. The current is normally applied in 0.5 to I s increments with at least two current
shots.
6. Some indications are slower to form, but usually 5 to 30 s is adequate for special
techniques.
CONTINUOUS METHOD
1. The continuous method is usually required because it is the most sensitive. Usually
2. The timing of the diversion of flow and current application is critical for the wet
continuous method. The test object is wetted so the liquid is draining and the flow
is diverted at the instant the current is applied, and then another shot of current is
The dry continuous method requires the current to remain on while the dry powder is lightly
applied to the test area. Excess powder is gently blown off when the current is applied.
A common, standard wet continuous method requires that the test object is wetted with
solution, the flow diverted, two shots of proper magnetizing current (0.5 s each) are applied
in rapid succession, and indications are evaluated by looking at the residually held indication
1. The residual method may sometimes be allowed for test objects with high
retentivity.
2. The residual method is also always used during testing of hard to see areas, such
3. Test objects should not be allowed to rub against each other when magnetized
4. The residual method (wet or dry) will not have the sensitivity for small
Field Direction
1. Discontinuities located in the direction of current flow have the highest probability
of detection.
3. It is critical to establish a magnetic field in two 90° directions in all areas of the
for this reason a technique is normally not approved until an actual test object or a
the pie gage, system verification with the tool steel ring, and certain other
Circular magnetic fields are used to detect suspected discontinuities that are perpendicular to
Direct Contact
A head shot, prods or other means of passing current through a test object will create a
circular field, but there is a danger of arc strikes or burns where the current contact is made.
Central Conductor
A central conductor will produce a circular field with no danger of burning or ruining the test
object.
Amperage Requirements
sheet, verified with a pie gage, shims or artificial block, or measured with a Hall
2. A rule of thumb for an amperage starting point is 300 to 800 A/in. of test object
SENSITIVITY LEVEL
Amperage
Higher amperage gives better sensitivity, so use as high an amperage as possible without
The proper preparation, maintenance and application technique of the media is critical to
1. A rule of thumb for an amperage starting point is 300 to 800 A/in. of test object
2. Test objects will require rotation to obtain 100% testing and magnetization in
2 to 90° directions, usually using both circular and longitudinal magnetic fields or
Coil Shot
Placing a test object inside a coil is the usual way to longitudinally magnetize the test object.
The effective field will only allow testing for a distance of about 15 to 23 cm (6 to 9 in.) on
Cable Wrap
Flexible insulated copper cable can be wrapped around a test object using three to seven
turns, and connected to a mobile or portable magnetic particle unit to induce a longitudinal
field.
Quiz
retentivity. b. False
sensitivity.
6. Lower amperage gives better sensitivity.
a. True
a. True
b. False
b. False
57
8. After the solution is applied to the test 12. Magnetic flux or field strength may be
object during the wet continuous verified with a Hall effect gauss meter,
a. 0.5 s. encountered.
b. 2 s. a. True
9. The residual method is useful during through the test object or current is
a. True a. True
b. False b. False
b. False a. True
b. False
11. The magnetic field direction for circular
flow. a. True
b. False
a. True
b. False
a. True
b. False
a. True
b. False
INTRODUCTION
Electric current is used to directly or indirectly create a magnetic field in two 90° directions
CIRCULAR MAGNETIZATION
1. A current passing through a wire or bar creates a circular magnetic field in and
2. Discontinuities are detected best when they are 90° to the magnetic field direction.
May be direct magnetization by use of prod or contact plate head shot, or indirect
2. Indirect techniques are used when possible on finished machined test objects.
3. Test objects often require two or more contact (head) shots to get 100% circular
testing.
61
Circular Magnetization with Induced Current
Special techniques and equipment can be used to induce current that creates circular
magnetic fields.
1. A central conductor through a hollow test object or a hole in the test object will
2. Special techniques may be written and approved by a Level III to use cables
LONGITUDINAL MAGNETIZATION
1. A current passing through a coil or cables wrapped around a test object will induce
COIL MAGNETIZATION
induced and will detect transverse discontinuities (90 0 to the magnetic field
direction).
3. To induce a uniform longitudinal field the length of the test object must be at least
4. For round or odd shaped test objects that don't have this length-to-diameter ratio, a
1. Special equipment and qualified techniques may use the magnetic flow technique
the center and ends of the test object to verify adequate field strength along its
entire length.
YOKE MAGNETIZATION
I. Hand-held portable yokes can be used to induce a longitudinal field in a test object
or weld.
2. Yokes are required to have specified lifting force and tested with dead weight
checks.
3. When used for odd shaped test objects, a pie gage or other magnetic penetrameter
It is essential and critical that test objects are magnetized in two 90° directions in order to
Each magnetic particle testing procedure and technique must specify which units of
1. Diamagnetic materials have permeability slightly less than that of a vacuum and
2. Paramagnetic materials have permeability slightly greater than air and may be
3. Ferromagnetic materials have permeability much greater than air and are strongly
attracted to a magnet and may retain a residual field after the magnetic force is
removed.
Prods
1. A good rule of thumb for prod testing is 90 to 125 A/in. of prod spacing, use a prod
2. The magnetic field strength and direction must be verified with a pie gage, shim or
other artificial discontinuity at the edges and center of the area being tested.
Yokes
The yoke field strength is usually verified periodically by dead weight lifting power. The
field strength and direction induced in a test object should be verified with a pie gage, shim
or other artificial discontinuity at the edges and center of the area being tested.
1. When test objects are placed in a coil, the induced magnetizing force is calculated
using ampere turns (number of turns or wraps of cable times the amperage passed
through the coil). A Hall effect meter measures the resulting longitudinal magnetic
2. Do not relate ampere turns to magnetic field strength unless this has been
3. Formulas have been written and rules governing what the conditions must be so
amperage.
4. The actual resulting magnetic field must be verified by test objects with artificial
discontinuities, a Hall effect meter, pie gage, shims or other approved devices.
1. When a long test object is placed in a multi-turn coil (usually five turns) and a
current is passed through the coil, a longitudinal field is produced parallel to the
2. If the proper amperage is selected the magnetic field will be strong enough to
test a section about equal to the radius of the coil on each side of the coil.
1. Coils have been designated as low, intermediate and high fill factor coils. This and
other rules were developed so formulas could be written to obtain a starting point
2.· A low fill factor coil is when the cross-sectional area is ten or more times the
3. An intermediate fill factor coil is when the cross-sectional area is more than two
and less than ten times the cross-sectional area of the test object.
Student Guide: Magnetic Particle Testing 65
~'~l~!!pe
4. High fill factor coil is when cable wraps are used or the cross-sectional area is less
then two times the cross-sectional area of the test object (including hollow
portions).
INTERPRETATION OF INDICATIONS
All indications must be interpreted to verify that a false or nonrelevant indication is not
hiding or covering a relevant indication. This may require lowering amperage and
Relevant Indications
Nonrelevant Indications
Are caused by magnetic leakage fields that are not discontinuities and may be excessive
False Indications
Are not the result of magnetic leakage fields and may be particles held by gravity or foreign
Quiz
flow. b. Yoke.
of the material.
5. An indirect method of inducing a
d. At 90° to the current direction.
longitudinal magnetic field is with a:
a. True
6. A long, uniform, hollow test object
b. False
magnetized with a central conductor:
3. Prods may be used to induce a a. Will have a north and south pole at
69
7. In order to induce a uniform 10. The mathematically calculated starting
longitudinal field in a coil, the test magnetizing force to use for a test
b. 4. a. 9000 A turns.
c. 6. b. 1800 A.
d. 15. c. 4500 A.
d. both a and b.
8. When a ring or wheel is placed inside a
e. none of the above.
coil, radial or transverse discontinuities
can be detected if the test object is 11. An intermediate fill factor coil/test
rotated 90 0 and repositioned in the coil object combination is when the cross-
a. True
9. The only type of material that can be
b. False
tested by magnetic particle testing is:
c. Ferromagnetic. a. 2.
c. 10.
d. 16.
Demagnetization
PURPOSE
The purpose of demagnetization is to demagnetize to less than +/-3 G after required tests,
which is required by most specifications and is normally easily accomplished. Some highly
retentive, odd shaped test objects or distorted fields may require special techniques.
The simplest method is to pass the test object through an alternating coil that will magnetize
the test object to a higher level than the residual field. Then, by reducing the magnetizing
force by increasing the distance from the coil or by reducing the amperage, the residual field
Reasons to Demagnetize
The reason to demagnetize is to ensure the residual field will not interfere with further
field does not have any detrimental physical effects on the test object.
1. The residual field may interfere with nearby instruments that work on magnetic
principles.
2. The residual field may cause difficulty in post cleaning. Magnetic media from the
71
3. The residual field can attract chips during machining or interfere with welding
4. The residual field can attract particles in a moving test object and cause wear.
2. Welded assemblies that have low retentivity usually don't have a residual field
DEMAGNETIZATION METHODS
Heating above the curie point (650 to 870°C [1200 to 1600 OF] for ferrous alloys) causes the
material to lose all magnetism characteristics, and upon cooling they are ferromagnetic again
Electromagnetic Demagnetization
Accomplished by creating a slightly higher magnetic field than the residual and gradually
Is normally the easiest, fastest and most effective, and normally a 2 to 5 s operation.
Some stationary units have a reversing demagnetization circuit that will reverse and lower
the direct current and lower the residual. These units are better than alternating current for
An alternating current yoke can be used to demagnetize by simply magnetizing the test
object by close proximity with the yoke energized and moving the test object away from the
yoke.
Demagnetization Practices
1. The Earth's magnetic field usually does not interfere with shop demagnetizations.
2. If long test objects are stored with their axis north and south, they will pick up a
3. Striking a test object or vibrating it during demagnetization will aid the process for
•
• Lesson 11
Quiz
•
treatment.
axis north to south .
b. Require demagnetization before
b. False treatment.
a. True b. False
b. False
• 75
7. The magnetic field strength at the start
strength.
a. True
b. False
demagnetized by:
field.
current to zero.
Equipment
STATIONARY EQUIPMENT
1. Mechanized equipment.
2. Semiautomatic equipment.
PORTABLE EQUIPMENT
2. Powder color choice is dependent upon which powder will provide the best
contrast.
Powder Selection
1. Gray powder.
2. Black powder.
3. Red powder.
4. Yellow powder.
Powder Applications
77
Applicators
1.
2.
Squeeze bottle.
Particle Selection
3.
4.
5.
Suspension characteristics.
Particles.
Media.
•
WET FLUORESCENT METHOD
When exposed to ultraviolet radiation, fluorescent dye coated magnetic particles glow.
Test Variables
1. Bath strength.
3. Current/suspension application.
78
4. Lighting conditions.
• 1.
2.
Dust, rust, loose scale and oil or grease should be removed.
Paint and plated coatings over 0.01 cm (0.005 in.) thick should be stripped.
Suspension Application
necessary.
3. Bath strength and immersion time effect the size of the indication.
Water Problems
• 2.
3.
Wetting agent, rust inhibitors and other water conditioners must be used with water
baths.
Suspension Preparation
1. Pour the required amount of powdered concentrate into the bath liquid in the tank.
Water Baths
Settling Test
Maintenance
1. Ultraviolet lamps should be energized for at least 15 min before the test is
performed.
Quiz
a. True
5. Maintenance magnetic particle tests
b. False
may be performed through thin coatings
2. Small, hand-carried portable units will (less than 0.003 in.) using a yoke.
3. Mobile or stationary units will usually suspension and particles and may select
supply up to about 4000 A low voltage a good allowable color contrast with the
a. True b. False
b. False
83
7. If wet suspensions are used from
a. True
b. False
fluorescent.
a. True
b. False
Types of Discontinuities
INHERENT DISCONTINUITIES
2. Cold shuts are caused in a casting when metal splashes or flows together but
3. Ingot pipe is caused by shrinkage at the top of the ingot, or can form in the riser in
a casting. This top area of the ingot (called the hot top or casting riser) is nonnally
4. Unequal cooling and stress build-up cause casting hot tears and shrinkage
especially between thick and thin section changes that are near 90°. These are
linear indications that may appear to look like a wide crack or void.
5. Blowholes in the ingot are caused by gas pockets escaping, and in a casting may
6. Porosity is trapped gas pockets in the cooling metal and may occur in the ingot,
casting or in a weld.
After rolling or fonning, they are often linear with the grain direction and detected
9.· Alloy segregation occurs because alloy elements have cohesion and sometimes
group together instead of alloying uniformly in the ingot or casting and may show
8S
l.
2.
Seams.
Laminations.
PRIMARY PROCESSING DISCONTINUITIES
•
3. Stringers.
4. Cupping.
5. Cooling cracks.
8. Hydrogen flakes.
9. Welding discontinuities.
a. Cold cracking.
b. Hot cracking.
•
c. Lamellar tearing.
d. Lack of fusion.
e. Lack of penetration.
f. Porosity.
g. Inclusions.
h. Undercut.
1. Overlap.
l. Cracks.
3. Pickling cracks.
4. Machining tears.
86
5. Plating cracks.
2.
Fatigue cracking.
Creep cracking.
INSERVICE DISCONTINUITIES
4. Hydrogen cracking .
Quiz
solidifies. b. Burst.
rejected. a. Seam.
object to fail.
6. A burst may be found in a:
Ii hear.
a. True
b. False
89
7. A common discontinuity in a weld or 12. A slag inclusion in a weld is normally
a. A crack. b. False
b. Porosity.
c. Shrinkage.
d. Lamination.
manufacturing.
a. True
b. False
a. Lack of fusion.
b. Slag inclusions.
c. Lack of penetration.
d. Undercut.
single overloading.
a. True
b. False
a. True
b. False
Evaluation Techniques
REFERENCE STANDARDS
Rules of thumb are okay for a starting point, but monitoring must be used to verify 100%
SYSTEM EVALUATION
Standard test objects with known discontinuities can be used to check system performance.
System Standardization
Consistent sensitivity can also be verified for the whole system by using known defect
standards.
Parametric Evaluations
Evaluation of magnetic particles and test object variables are used to generate or modify
Technique Development
1. Proof of adequate field strength and 100% coverage of the test object.
3. Instruction sheet for other technicians for the same or similar test objects.
91
1.
STANDARD TOOL STEEL RING
The ring is placed on a central conductor, and three or more different amperages
are passed through the central conductor to create a uniform circular field.
•
2. The side drilled holes that show on the outside surface are recorded and used as a
If the test is conducted the same each time, the circular field in the ring will show subsurface
indications from the side drilled holes. If the results are not the same there could be a
problem with the suspension or particles, the ring or with the magnetizing equipment.
When a new ring is purchased, it should be tested to verify it produces the indications
required by the specification. If new magnetizing equipment or particles are purchased, the
There are several reference blocks that can be used for information on system and
verification of suspension and particle performance for additional information. The system
performance checks specified in the contract referenced specifications would still have to be
1. A Hall effect meter measures the number of magnetic flux lines at 90 0 from the
test object. A probe often requires a plastic fixture to take the readings at the right
orientation.
2. The Hall effect meter may be used when the current is energized and will measure
May be used for special cases to measure magnetic saturation points if qualified techniques
are developed .
Quiz
1. Before magnetic particle testing, a 5. The standard tool steel ring is used to
detected.
2. A technician can rely solely on
c. That the suspension and
empirical rules for establishing
fluorescence is satisfactory.
magnetic particle testing procedures.
d. That the system is working.
a. True
b. False
95
Lesson 15
Quality Control
INTRODUCTION
Properly trained technicians using proper material, techniques and procedures correctly will
New Materials
instructions for storage are followed, materials should not degrade except by contamination
In-Use Materials
1. In-use procedures are controlled by the company. Contamination from poor storage
2. Materials used in open tanks are most often contaminated from improperly
precleaned test objects or from allowing grinding operations near the tanks.
MATERIAL CONTAMINATION
1. Evaporation losses.
2. Heat degradation.
3. Process degradation .
• 97
MATERIAL CONTROL REQUIREMENTS
New materials are checked prior to use and must also pass all the tests used for the in-use
materials.
Ultraviolet radiation used for fluorescent testing is filtered so the wavelength range from 320
to 400 nm (usually 365 nm) is used. This wavelength will emit visible light when it strikes
the fluorescent dye used for testing as a yellow-green color with a wavelength of 510 to
550 nm.
Ultraviolet radiation intensity used for testing must be a minimum of 1000 Jl W/cm 2
20 Ix (2 ftc).
2. The testing booth should have dark walls, and technicians should not wear light
VARIABLES
Intensity is normally measured with a light meter dajly or weekly because the lamps lose
intensity as they age. Turning the light off and on reduces the bulb life.
PROPER PROCEDURE
2. The design engineer is responsible for referencjng the proper procedure and
3. For required maintenance, the owner of the test object or assembly is responsible
Variation in Technique
3. Sequence of tests.
New test objects usually require half or full wave direct current to detect subsurface
discontinuities, and may also use alternating for better surface sensitivity.
1. Determine whether direct induction may ruin the test object. Finished machined
2. The amount of current to start the test may be determined by experience or formula
guidelines.
3. The proper current values will be verified for each area of the test object using a
pie gage, notched shims or a Hall effect meter. The current raised or lowered to
Selection of Media
Selection of Technique
1. The continuous method must be used for required tests unless the residual method
is authorized in writing.
2. The residual method should only be used as and aid in test indications detected
with the continuous method, and for information tests of materials with high
retentivity.
two directions.
discontinuities.
Head Shot
1. Establishing a circular field using a direct contact (head shot) 1000 A/in. of
outside diameter will normally create an adequate field (and may be too high for
odd shaped test objects). Starting current levels are adjusted up or down, and the
2. Caution must be used to avoid arc strikes or burning the test object.
Central Conductor
1. The strongest field will be next to the central conductor, so normally the test object
2. The larger the central conductor, the more circumference of the test object that can
be tested.
Longitudinal Field
1. Ampere turns = 45 000 divided by the length divided by the diameter. Then the
ampere turns divided by the number of turns in the coil will give the starting
amperage.
2. The strongest field will be inside the coil right at the inside surface of the coil.
Optimum sensitivity will be alternating current for fine surface discontinuities, wet
•
Notes
•
Lesson 15
Quiz
detected. b. False
a. True
6. The ultraviolet radiation intensity is
b. False
normally measured with a light meter
a. True b. False
b. False
7. Too much current will produce
a. True a. True
b. False b. False
thereafter.
• a.
b.
True
False
105
8. New test objects usually require half or
sensiti vity.
a. True
b. False
standards.
a. True
b. False
a. True
b. False