0% found this document useful (0 votes)
108 views2 pages

ML-0187 Rev K Evolution Data Sheet

Uploaded by

b117094011
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
108 views2 pages

ML-0187 Rev K Evolution Data Sheet

Uploaded by

b117094011
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

eVolution 3e ventilators

Features

Comprehensive Graphics and Trending


High Performance PSOL and Active Exhalation
Servo Controlled for Precise Breath Delivery
Neonatal through Adult
Invasive and Noninvasive
HL7 Compatibility and Virtual Report
Capnography ready
High Flow Oxygen
2ml Tidal Volume
Weaning Target Tool
Lung Model
HEPA Filter (standard on Turbine Model)

High Pressure System Ventilator

Internal Turbine System Ventilator

ML0187 Rev. K, Int'l Version


Page 1
Specifications

eVolution® 3e Essential
This document is intended to provide information to an international audience
outside of the US. Page 2

Modes of Ventilation User Configurable Graphic User Interface (GUI) and Functions
• Assisted Control Mandatory Ventilation (CMV) • Multiple Easy View Tabs (Main, Settings, Monitoring, Alarms, Configuration)
• Synchronized Intermittent Mandatory Ventilation (SIMV) • Graphics Display (2-Loops and 3-Waveforms)
• Spontaneous Ventilation (CPAP + PS, CPAP + VS, SPAP/ • Apnea Back-up
APRV) - User-selected breath type and settings
• High Flow O2 • Smart Nebulizer™
Additional Features • Smart Sigh™
• Auto Control • Humidification Selection (HME, Humidifier, and None)
• Apnea Backup Ventilation • EtCO2 Graphics Display
• SPAP Philosophy Options: Time, Cycle + Time, Cycle + I:E • PaCO2 Time Stamp Entry
• IBW Calculator
Monitored / Displayed Data
• NIV On/Off in all Modes
• Active Exhalation Valve • Delivered oxygen concentrations: 21 - 100%
• Automatic Leak Compensation (up to 60 lpm) • Trend Data (1-72 hrs.) (45 trend parameters)
• Work of Breathing (Imposed and Total), Power, Stress Index • Monitoring Data (57 monitored parameters):
• Maneuvers: PO.1 and PiMax Basic Mechanics Weaning
- Respiratory Rate - Auto - PEEP - Ve Spont
Breath Types
- PEEP - WOB - Vte Spont
• Volume-controlled (V-CMV; V-SIMV) - Ppeak -E - Leak
• Pressure-controlled (P-CMV; P-SIMV, PS/CPAP) - Pmean - SI - P0.1 & PiMax
• Volume Targeted Pressure-controlled (PRVC-CMV; PRVC-SIMV, VS/ PRVC- - Driving pressure - POWER or C20/C - RSBI
• Dual Level PEEP PS) - Vte & Vti - Ptrach & Pmin - Rate Spont
- Equivalent to APRV & BiLevel (SPAP/APRV) - Ve & Vi - Pplateau - Rate Man
- O2 - Cdyn - Ti / Ttot
Breath Triggering - Te & Ti - Cstat & Cstat / kg - Spont % 1h
-0.1 to (- 20) cmH2O - PFi - Ri & Re - Spont % 8h
• Pressure triggering
0.1 - 25 lpm - PFe - Rce - PO.1 / PiMax
• Flow triggering
- I:E or H:L - VT/kg - WOBimp

Additional Settings EtCO2 EtCO2 Graphics Real Time Graphics


- PetCO2 - PCO2 Volume Loop -Flow Volume Loop
• Respiratory rate 1 - 120 bpm - FetCO2 - Pressure Volume Loop
- FCO2 Volume Loop - Flow + Time
• Tidal volume 20 - 3000 ml - Valv
- VCO2/min - PCO2 + Time - Presure + Time
• PEEP / CPAP 0 - 50 cmH2O - Volume + Time
- Vd ana - FCO2 + Time
• Pcontrol 1 - 100 cmH2O - Valv/min - VteCO2 + Time
• Psupport 0 - 100 cmH2O - VtiCO2
- Vd alv
• Peak flow (mandatory) 1 - 150 lpm
- PeCO2
• Peak flow (spontaneous) Up to 200 lpm - VteCO2
• Peak flow (Turbine Output) Up to 300 l/min - Vd/Vt phy
• I-Times 0.2 - 10.0 sec - FeCO2
• Oxygen concentration 21 - 100% Prioritized Alarms (High, Medium and Low)
• Rise time settings 1 - 20
• Exhalation sensitivity % 1 - 80% Auto set alarm feature & 1000 Event log
(of spontaneous peak flow)
Communication Interfaces
• Plateau (insp pause) 0 - 2.0 sec
Ethernet and Nurse call
Physical Data
Neonatal Software Option Part Number: EVM400156
• AC power: 100 - 240 VAC (47/63Hz)
• NCPAP+ • Standard internal battery (Lithium-Ion* & Lead Acid**)
• Respiratory Rate 1 - 150 bpm
• Battery backup: 2 Hours (HP Model), 4 Hours (Turbine Model)
• Tidal volume 2 - 3000 ml • Gas supply pressure, 2 to 6 Bar
• I-Times 0.1 - 10 sec • Screen: LCD 12.1” / Touch Screen
• PS Tmax 0.15 - 5.0 sec HL7 Protocol Compatibility and Optional Remote Viewing System
Additional Software Options
• CliniNet® HL7 Gateway™
• Suction support Part Number: EVM400158
• Automatic tube compensation Part Number: EVM400157 High Performance Internal Technology
• EtCO2 Part Number: EVM400182 • Integrated high performance turbine technology with included HEPA filter*
• Hi FlowO2 Part Number: EVM400184
• Low Flow O2 Capability*

User Initiated Functions Ordering Information

• Manual inspiration • Inspiratory and expiratory • Model EVL100500-HP-NF, High Pressure System without Exhalation Filter
• Alarm silence breath holds • Model EVL100500-T-NF, Internal Turbine System without Exhalation Filter
• 100% oxygen • Standby *Only available on turbine models. **Only available on high pressure models.
Specifications subject to change. eVent Medical recommends using exhalation filters on all systems.

+1 949.900.1917 [email protected] www.event-medical.com