HP Sonos 4500 5500 Transducer Guide
HP Sonos 4500 5500 Transducer Guide
T R A N SD U C E R R E F E R E N C E
SONOS 5500/4500
ImagePoint Hx
PH I L I PS ME D I C AL SY ST E MS
English
M2424-99300-06
User’s Guide
Transducer Reference
Preface
1 Transducer Setup
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Contents-1
Disinfectants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Glutaraldehyde-based Disinfectants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Non-Glutaraldehyde-Based Disinfectant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Chemicals that Damage Transducers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Contents-2
Contents
3 Endocavity Transducers
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Contents-3
4 TEE Transducers
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Contents-4
Contents
5 Intraoperative Transducers
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Contents-5
Defibrillators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Accident Prevention . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Contents-6
Preface
Chapter 2 describes the care required for your Philips SONOS and ImagePoint
HX transducers. The handling, storing, general cleaning, and disinfecting of all
supported transducers is discussed, as well as the ordering of biopsy starter kits.
NOTE The most current information on cleaning transducers can be found on the fol-
lowing Philips web site:
www.medical.philips.com/transducercare
Preface-1
Safety Questions
Additional Information
You must be thoroughly familiar with the operation of the Philips SONOS 5500/
4500 or ImagePoint ultrasound system to use the Philips specialty transducers
described in this manual.
Use this guide in conjunction with the following books contained in the
Reference Manual binder:
Additionally, there are several specialty guides that describe applications and
optional packages, such as Stress Echocardiography, Contrast, 3-D, Acoustic,
and Integrated Digital Interface.
Preface-2
1 Transducer Setup
Connecting Transducers
Vertical
6apc0148
Horizontal
6apc0149
1 With the latch in the vertical position, plug the transducer adapter into
the slot.
6apc0160
Transducer adapter
2 Push the connector in firmly so the pins are seated correctly. Turn the
latch to the horizontal position.
6apc0161
LEFT RIGHT
AMP-style transducer
6apc0162
1 Plug one adapter into the right slot and one into the left slot.
Connection Tips
• If you feel any resistance when inserting the connector, don’t try to force it in
place. Check the latch and carefully reinsert.
• If “NO Probe” displays on the screen, either the system is running an identi-
fication check or the transducer is not properly connected. Check that you
have selected the correct slot (Left, Center, Right, or Non-image under
Probe) on the SONOS 4500 or 5500.
• If “?? Probe” displays, the transducer is not latched into position. This mes-
sage could indicate that the transducer connector is not locked into place.
• Before disconnecting the transducer, unlock the transducer connector by
turning the latch to the vertical position.
• When using a TEE transducer, always make sure the TEE tip and shaft are
straight before connecting the transducer.
• Remove any other transducer connected to the system prior to connecting
the Pediatric Biplane.
Calibration:
NOTE When the transducer is properly connected, and depending on which transducer
you are using, either the frequency or label information displays in the upper left
corner or below the grayscale bar.
Activating Transducers
This section shows you how to activate transducers on your ultrasound system.
Touch Probe. Then touch the softkey indicating where the transducer is
connected.
Probe
ImagePoint Hx
Up to three imaging To activate a different transducer, press Probe. Each time you press this
transducers and one control another transducer is activated. A label for the active transducer
dedicated Doppler
transducer may be
appears below the grayscale bar.
connected to your
system at one time. A S D F G H J K L ’
System
Z X C V B N M <, . ?
Shift Mic
Filter Power
Size Position
Body Marker Zoom
Frequency
Probe
Angle Calcs Dual
Baseline Hx M Mode
Dynamic
Left Right Range
Angio Caliper
2D
Scale Tape
CW Trace GAIN
Focus
Print 2
PW Spectral Erase
Gain Freeze
Color Print 1
Depth
Enter
6APC0045
Probe Control
This section tells you how to change the transducer frequency on your ultrasound
system.
2D
Frequency
Fusion
See the Philips SONOS 4500 and SONOS 5500 System Basics User’s Guide for
more information on the Frequency Fusion icon.
ImagePoint Hx
If the transducer is Press Frequency to change the transducer’s operating frequency.
already operating at the
highest frequency,
’
pressing Frequency A S D F G H J K L
System
selects the lowest Z X C V B N M <, . ?
Shift Mic
frequency. If the
transducer has only one Volume Patient Preset Home Clear Text
Filter Power
Size Position
Body Marker Zoom
Frequency
Probe
Angle Calcs Dual
Baseline Hx M Mode
Dynamic
Left Right Range
Angio Caliper
2D
Scale Tape
CW Trace GAIN
Focus
Print 2
PW Spectral Erase
Gain Freeze
Color Print 1
Depth
Enter
6APC0045
Frequency control
The following table lists transducers that can be used with the SONOS 4500 and
SONOS 5500.
OmniPlane II TEE T6210 21367A * Adult and Pediatric Cardiac Requires an adapter
Sector Transducer
* Transducers supported from previous software revisions. Not available for sale
with new system orders.
** The 21228B, 21223B, and 21221B transducers are supported only on SONOS
4500 and 5500 Revision B.2 and higher.
ImagePoint Hx Transducers
The following table lists transducers that can be used with the ImagePoint Hx.
* Transducers supported from previous software revisions. Not available for sale
with new system orders.
Specialty Transducers
Endocavity Transducers
The following table defines the necessary revision levels for using endocavity
transducers on the indicated Philips imaging systems.
Endovaginal/
Philips Ultrasound Endovaginal Endorectal
Endorectal 21336A
System 21370A (E6514) 21371A (E7511)
(E6509)
ImagePoint and Rev. A.1 and higher Rev. A.0 and higher Rev. A.0 and higher
ImagePoint Hx
TEE Transducers
The following table defines the necessary revision levels for using TEE
transducers on the indicated Philips imaging systems.
Pediatric
Philips OmniPlane OmniPlane II OmniPlane II miniMulti
Biplane
Ultrasound 21364A 21367A 21369A 21381A
21366A
System (T5012) (T6210) (T6210) (T6207)
(T7506B)
SONOS 5500 Rev. A.0 and Rev. A.0 and Rev. A.0 and Rev. A.0 and Rev B.2.2. and
higher (with higher (with higher higher (with higher
transducer transducer transducer
adapter) adapter) adapters)
Pediatric
Philips OmniPlane OmniPlane II OmniPlane II miniMulti
Biplane
Ultrasound 21364A 21367A 21369A 21381A
21366A
System (T5012) (T6210) (T6210) (T6207)
(T7506B)
SONOS 4500 Rev. B.0 and Rev. B.0 and Rev. B.0 and Rev. B.0 and Rev B.2.2. and
higher (with higher (with higher higher (with higher
transducer transducer transducer
adapter) adapter) adapters)
ImagePoint and N/A Rev A.1 and Rev. A.0 and N/A N/A
ImagePoint Hx higher higher
(with transducer
adapter)
Intraoperative Transducers
The following table defines the necessary revision levels for using intraoperative
transducers on the indicated Philips imaging systems.
SONOS 5500 Rev. A.0 and higher Rev. B.0 and higher Rev. A.0 and higher
SONOS 4500 Rev. B.0 and higher Rev. B.0 and higher Rev. B.0 and higher
Introduction
This chapter contains information on the general care, handling, and storing of
Philips transducers for SONOS 5500/4500 and ImagePoint Hx systems. The
general cleaning of external transducers is discussed, as well as the cleaning and
disinfecting of specialty transducers, such as the endocavity, biopsy-capable,
transesophageal, and intraoperative transducers.
Handling Transducers
Although a transducer is designed for durability, use care when handling it.
Dropping or banging the transducer could damage the acoustic cap and
piezoelectric crystals. Cuts in the transducer cable or cracks in the housing could
destroy the electrical safety features of the transducers. This damage is not
covered by the warranty or your service contract.
When you are not using the transducer, place it in the transducer holder located
on the top right side of the system to assure safe, convenient storage. For more
infomation on storing transducers, see "Storing Transducers" on page 2-3.
When you are shaking an ultrasound gel bottle, be careful not to hit the
CAUTION
transducer face with the tip of the bottle. Striking the transducer face with a hard
object can damage the transducer. This damage is not covered by the warranty or
your service contract.
All transducers are exposed to many potentially damaging situations during use
and cleaning. Periodically check the transducer for the following signs of
damage:
Ergonomic Grip
Philips has developed a slip-on reusable grip to provide alternative grip options
for the s3, s4, and s8 transducers. To order grips, contact your local Philips
representative.
1 Wipe with a dry or water-moistened soft cloth. You can also use isopropyl
alcohol solution (rubbing alcohol), a 10% bleach solution or a 2.4% glutaral-
dehyde-based disinfectant. For a list of the glutaraldehyde-based disinfectants
that Philips recommends for specific transducers, see page 2-11.
2 Remove any cleaning solution residue with a soft cloth moistened in water.
Do not allow cleaning solutions to air-dry on the grip. For further instructions
on how to minimize the effects on transducers from disenfectants, see page 2-
12.
Storing Transducers
If a carrying case is provided with your transducer, always use the carrying case
to transport the transducer from one site to another. To prevent kinking, carefully
place the transducer in the case. Before closing the lid, make sure no part of the
transducer is protruding from the case. Wrap the case in plastic material
containing air pockets (bubble wrap), and pack the wrapped case in a cardboard
carton.
External-use Transducers
Be sure to clean and disinfect the external transducers before storing them.
• Store transducers in the transducer holders on the side of the ultrasound sys-
tem.
Endocavity Transducers
Be sure to clean and disinfect the endocavity transducers before storing them.
• Store in one of the transducer holders on the side of the ultrasound system.
• Store the transducer separately from other instruments to avoid inadvertent
transducer damage.
• If storing the transducer in the original case or in a drawer, make sure the
transducer is thoroughly dry.
• Avoid storing the transducer in areas of temperature extremes or in direct
sunlight.
TEE Transducers
Be sure to clean and disinfect the TEE transducers before storing them.
• Never store a TEE transducer in the carrying case, except to transport it.
• Be sure the distal tip is straight and protected before storing the transducer.
• To avoid damaging the shaft or steering mechanism, do not bend or coil the
flexible shaft of the transducer in less than a one-foot diameter (30 cm)
circle.
• Store the transducer on a securely mounted wall rack, with slots properly
spaced to prevent the transducer from falling.
• Store the transducer separately from other instruments to avoid accidental
transducer damage.
• Avoid storing the transducer in areas of temperature extremes or in direct
sunlight.
Intraoperative Transducers
Be sure to clean and disinfect the interoperative transducers before storing them.
When storing the Model 21390A intraoperative transducer on the system, use the
NOTE
cable management clips with the transducer cable.
• If storing the transducer in the original case, make sure the transducer is
thoroughly dry.
• Avoid storing the transducer in areas of extreme temperatures or in direct
sunlight.
For proper transmission of the acoustic beam, use the ultrasound transmission gel
supplied by Philips, or another glycol, glycerol, or water-based couplant. Do not
use mineral oil, oil-based couplants, or other non-approved material because it
might damage the transducer.
WARNING Do not use gels that contain any of the following agents:
-- acetone
-- methanol
-- denatured ethyl alcohol
-- mineral oil
-- iodine
-- any lotions or gels containing perfume
Latex Materials
The Philips’ diagnostic ultrasound systems and transducers in this manual do not
NOTE
contain natural rubber latex that contacts humans. Natural rubber latex is not
used on any ultrasound transducer, including transthoracic, intraoperative, and
transesophageal transducers. It also is not used on Philips ECG cables for the
products in this manual.
Biopsies can be performed with a variety of probes. The following table lists
transducers that support the use of biopsy needle guides and the biopsy kit used
by each.
For a list of the CIVCO kit numbers to order the appropriate replacement biopsy
kits for your endocavity transducer, see “Accessory Equipment” on page 3-12
Biopsy Precautions
Always follow these basic precautions when using a biopsy needle guide and
bracket:
• All biopsy procedures must be performed using highly controlled sterile pro-
cedures. Be sure that the biopsy needle package is not opened or damaged
and that the date has not expired.
• Before using the biopsy setup for the first time, it is strongly recommended
that you familiarize yourself with each possible needle position, procedure,
probe, and screen display. See Chapter 2 of the ImagePoint Hx User’s Guide
and Chapter 5 of the SONOS System Basics guide for information on using
biopsy transducers.
• Most needle guides are shipped sterile. Verify on the package that the guide
is sterile.
• Use a sheath with the biopsy needle.
• Most needle guides are disposable. Do not re-use disposable needle guides.
• After use, remove the needle guide from the bracket and discard the needle
guide.
• Most needle guide brackets are not shipped sterile. Follow the manufac-
turer’s instructions to disinfect the needle bracket.
• For a description of system controls for biopsy use, refer to Chapter 2 of the
ImagePoint Hx User’s Guide and Chapter 5 of the SONOS System Basics
guide.
• For biopsy needle guide instructions in this book see “Biopsy Precautions”
on page 2-7 and “Biopsy Guide Instructions” on page 2-8.
The following categories identify uses and required cleaning and disinfection
standards:
Philips recommends that you use a protective sheath with transducers used in
endocavity and biopsy procedures and a sterile protective sheath with TEE
transducers.
WARNING Never sterilize the transducer with sterilization techniques such as auto-
clave, ultraviolet, gamma radiation, gas, steam, or heat sterilization tech-
niques. Severe damage will result.
Establish and clearly post a disinfecting procedure that includes all required
NOTE
disinfection steps.
Disinfectants
The following high-level disinfectants have been tested with your Philips
transducers, as specified below, and are compatible with the material used in the
transducers. Use of any other disinfectants can void your system warranty and
service contract.
Glutaraldehyde-based Disinfectants
Philips recommends the following 2.4% glutaraldehyde-based disinfectants for
all transducers:
• aidal plus *
• abcoCIDE 14 *
• abcoCIDE 28 *
• Cidex™ Activated Dialdehyde Solution
• Cidex Formula 7™ Long-Life Activated Dialdehyde Solution
• Cidex Plus™ 28 Day Solution
• Cetylcide-G® Concentrate and Diluent Concentrate
• MaxiCide Plus *
• MedSci 3% Glutaneldehyde
• Metricide® 28 Day Long-Life Activated Dialdehyde Solution
• Metricide Plus 30® Long-Life Activated Dialdehyde Solution
• Omnicide™ 14NS *
• Omnicide™ 28 *
* Those disinfectants noted with asterisks are not FDA cleared for use in the
United States. The FDA maintains a list of 510k-cleared high-level disinfectants
at www.fda.gov/cdrh/ode/germlab.html
Non-Glutaraldehyde-Based Disinfectant
CIDEX® OPA Solution (An ortho-phthalaldehyde solution manufactured by
Advanced Sterilization Products) is a non-glutaraldehyde-based product that can
be used to high-level disinfect the transesophageal (TEE) and intraoperative
transducers listed below:
WARNING If you use CIDEX® OPA, a residual solution may remain on your transducers
if you do not carefully follow the manufacturer's instructions. Residual
CIDEX® OPA on TEE transducers may cause the following:
• Temporary staining of the mouth and lip area
• Irritation or chemical burns of the mouth, throat, esophagus and stom-
ach
CAUTION The use of any non-glutaraldehyde-based disinfectant other than CIDEX® OPA
on the 21364A, 21367A, 21369A, 21390A, and 21336A transducers may cause
damage to your transducer. Damage caused by such use is not covered by the
warranty or your service contract.
1 After every patient exam, wipe the ultrasound transmission gel off the
transducer.
2 Wipe the transducer and cable with a dry or water-moistened soft cloth.
WARNING Isopropyl alcohol may damage parts of your transducer. This damage is not
covered by the warranty or your service contract.
C D
B A - cable
A
B - strain relief
C - strain relief/housing joint
D - housing
E - Do not use alcohol in this area
F - You may use alcohol in this area
1 inch
2.5 cm
F
E
2 Wash the transducer head and cable with soap and water to remove any pro-
tein buildup; however do not rinse or immerse the cable near the connector or
the connector itself.
3 Disinfect the transducer and the cable with one of the disinfectants listed in
“Glutaraldehyde-based Disinfectants” on page 2-11 or “Non-Glutaraldehyde-
Based Disinfectant” on page 2-12.
4 Remove the transducer from the disinfectant and thoroughly rinse with sterile
water.
5 Check the transducer for any residual infectious material. If any is present,
disinfect the transducer again.
WARNING Never sterilize the transducer with sterilization techniques such as auto-
clave, ultraviolet, gamma radiation, gas, steam, or heat sterilization tech-
niques. Severe damage will result.
2 Wash the transducer head and cable with soap and water to remove any pro-
tein buildup; however do not rinse or immerse the cable near the connector or
connector.
3 Disinfect the transducer and cable with one of the glutaraldehyde-based disin-
fectants listed on page 2-11.
• Do not rinse or immerse the connector or the portion of the cable near
the connector.
• Avoid transducer contact with strong solvents such as acetone, freon,
and other industrial cleansers.
• Do not soak the transducer for extended periods of time, such as over-
night.
4 Remove the transducer from the disinfectant and thoroughly rinse with sterile
water.
5 Check the transducer for any residual infectious material. If any is present,
remove it and disinfect the transducer again.
WARNING Never sterilize the transducer with sterilization techniques such as auto-
clave, ultraviolet, gamma radiation, gas, steam, or heat sterilization tech-
niques. Severe damage will result.
Clean and disinfect the transducer immediately after each use to protect patients
and personnel from a variety of pathogens. Establish and clearly post a cleaning
procedure that includes the following steps:
2 Remove all organic matter and other residue, using the following procedure.
a. Soak gauze pads in mild, soapy water. Do not use iodine-based soaps.
Philips recommends using an enzymatic cleaner to assist in removing
protein residuals. A common enzymatic cleaner is one with a pH of 6.0
to 8.0 (e.g., Metrizyme) and contains diluted concentrations of surfa-
cants, alcohols, salts, and acids. These cleaners are further diluted dur-
ing use.
b. Wipe the distal tip and flexible shaft up to the control housing
(steering mechanism) with the gauze pads.
3 Rinse the distal tip and flexible shaft well, using water.
Do not rinse or immerse the control housing, cable, or connector. For the Model
CAUTION 21381A, miniMultiplane, do not immerse past the 65cm marker.
5 Disinfect the distal tip and flexible shaft by placing them in one of the disin-
fectants listed in “Glutaraldehyde-based Disinfectants” on page 2-11 or
“Non-Glutaraldehyde-Based Disinfectant” on page 2-12. This disinfectant
procedure is only allowed with certain models (see the list). Be sure to use a
disinfectant recommended for the transducer you are using.
7 Check the transducer for any residual organic material. If any is present,
remove it and disinfect the transducer again.
8 Dry the distal tip and flexible shaft with a soft cloth or pad.
9 Lightly wipe the steering mechanism with a pad moistened with 70% isopro-
pyl alcohol.
High-level disinfection and the use of a sterile sheath is the accepted method of
infection control for this critical category of device.
When the 21390A probe is used for animal studies, the animal use label
must be attached. See “Animal Use Label” on page 5-3.
2 Wash the transducer head and cable with soap and water to remove any pro-
tein buildup; however, do not rinse or immerse the connector itself or the
length of cable near the connector.
3 Disinfect the transducer and cable with one of the glutaraldehyde-based disin-
fectants listed on page 2-11. For the model 21390A (15-6L) transducer, you
may use either a glutaraldehyde-based disinfectant or Cidex® OPA. See
“Non-Glutaraldehyde-Based Disinfectant” on page 2-12 for this information.
• Do not rinse or immerse the connector or the portion of the cable near
the connector.
• Avoid transducer contact with strong solvents such as acetone, freon,
and other industrial cleansers.
• Do not soak the transducer for extended periods of time, such as over-
night.
5 Check the transducer for any residual infectious material. If any is present,
remove it and disinfect the transducer again.
WARNING Never sterilize the transducer with sterilization techniques such as auto-
clave, ultraviolet, gamma radiation, gas, steam, or heat sterilization tech-
niques. Severe damage will result.
Overview
Patient Safety
You can help ensure patient safety when using an endocavity transducer by:
• Verbally preparing each patient for the procedure prior to the exam.
• Scrutinizing the entire transducer and testing all of the controls before each
use.
• Using mandatory protective equipment, such as a market-approved sheath or
condom, during an exam. See “Ordering Biopsy Kits and Sheaths” on page
3-12.
• Inserting and operating the transducer properly.
• Not allowing water or other liquids to drip into the transducer connector or
the interior of the system or onto the keyboard.
The following table summarizes patient safety problems, how to prevent them,
and the pages in this guide where details are provided.
Mechanical Severe trauma, cuts, Inspect transducer using both sight and “Checking the
damage bleeding, perforations touch before exam. Transducer” on page
2-2
Insufficient Spread of illness or Thoroughly clean and disinfect transducer “Cleaning and
cleaning protocol disease after each use. Cover transducer with Disinfection of
protective sheath. Cover imaging system Transducers Used in
with disposable drape if highly pathogenic Endocavity
organisms known or suspected. Procedures” on
page 2-15
Improper Cuts, bleeding, ligament Never use force when inserting or “Endocavity Exam
insertion or damage, perforations removing transducer. Guidelines” on page
withdrawal 3-10
Any other Potentially serious Do not use transducer. Contact your Philips
irregularity consequences service
representative
NOTE The diagnostic ultrasound systems and transducers discussed in this manual do
not contain natural rubber latex that contacts humans. Natural rubber latex is not
Equipment Operation
Electrical Safety
Defibrillators
The outer insulating layer of the transducer, if intact, is designed to withstand
electrical stress created by defibrillation. A small hole in the outer layer of the
transducer, however, opens a conductive path to grounded metal parts in the
transducer. The secondary arcing that could occur during defibrillation could
cause patient burns. The risk of burns is reduced but not eliminated by using an
ungrounded defibrillator. Therefore, remove the transducer from the patient
before defibrillation. A disposable sheath provides no protective electrical
insulation against defibrillation.
Accident Prevention
Use defibrillators that do not have grounded patient circuits. Refer to the
defibrillator service guide, or consult with a biomedical engineer to determine
whether or not a defibrillator patient circuit is grounded.
Endovaginal/Endorectal
(Model 21336A)
6APCkim1
Endovaginal/Endorectal
(Model 21336A)
Specs: Length (transducer & cable) – approximately 263 cm
Length of handle & tip – 290 mm
Radius of curvature – 8.7 mm
Endovaginal
(Model 21370A)
404e001
Endorectal
(Model 21371A)
404e002
WARNING Do not use gels that contain any of the following agents:
-- acetone
-- methanol
-- denatured ethyl alcohol
-- mineral oil
-- iodine
-- any lotions or gels containing perfume
Check with the gel manufacturer if you have any questions regarding the contents
of the gel.
404e007
CAUTION Use acoustic coupling gel only inside the sheath; do not use coupling gel outside
of the sheath. For proper transmission of the acoustic beam, use the ultrasound
transmission gel supplied by Philips, or another glycol, glycerol, or water-based
couplant inside the sheath. Do not use mineral oil, oil-based couplants, or other
non-approved material because they might damage the transducer.
2 Place the sheath over the transducer or unroll the sheath until it covers the
insertable portion of the transducer.
404e008
3 Use an elastic band to hold the sheath in place at the transducer handle.
4 Gently pat the sheath to remove any air pockets. Check for tears or damage
before proceeding.
404e015
• A thorough inspection procedure is required for the safety of both you and
the patient, and to ensure the continued correct functioning of the transducer.
• Forceful insertion, manipulation, or withdrawal of a transducer can result in
lacerations, bleeding, perforation, tearing of adhesions, and ligament dam-
age.
• Refrain from handling the tip of the transducer.
• Use the handle of the transducer as much as possible.
Accessory Equipment
Use the following CIVCO numbers to order the appropriate replacement biopsy
kit for your transducer:
This section describes how to use the biopsy needle guide with the endocavity
transducers. The needle guide mounts onto the transducer to guide a biopsy
needle during a biopsy or aspiration procedure.
• Most needle guides are shipped sterile and are disposable. DO NOT reuse
disposable needle guides. Please verify on the packaging that the needle
guide is sterile.
• After use, remove the needle guide from the transducer and discard the dis-
posable guide.
WARNING The biopsy needle guide and endocavity transducer must be cov-
ered with a sheath during biopsy procedures.
2 Locate the attachment guides (tabs) found on the underside of the needle
guide. Note: these guides fit into the notches on the transducer shaft. The
illustrations below show the location of the notches on an uncovered trans-
ducer.
404e006
6APCkimt
404e006
6APCkimta
4 Check the sheath on the transducer to make sure it did not tear or get damaged
in the process of installing the needle guide.
5 Place an adequate amount of ultrasound coupling gel at the tip of the inside
surface of a second sheath.
6 Add the second sheath over the transducer and needle guide, making sure that
there are no air bubbles between the two sheaths.
404e009
7 The following picture illustrates how the needle pierces the outer sheath.
WARNING Do not insert the needle into the guide until the transducer is
inserted into the patient.
404e010
Overview
• 21364A, the OmniPlane transducer, enabling 360o views of the heart, steer-
able CW Doppler, dual frequency, and frequency agility. Additional features
on the SONOS 4500 and 5500 include frequency fusion and adaptive color
flow.
• 21367A and 21369A, the OmniPlane II transducers, enabling 360o views of
the heart, steerable CW Doppler, dual frequency, and frequency agility.
Additional features on the SONOS 4500 and 5500 include frequency fusion
and adaptive color flow.
The difference between the two transducers is their connector to the ultra
sound system. The 21367A uses an AMP-style connector with a transducer
adapter and the 21369A connects directly to the ultrasound system using a
Cannon®style connector.
• 21381A, the miniMultiplane transducer, with a small tip array, enabling 180o
rotation for 360o views of the heart. Additional features of this small, light-
weight transducer include steerable CW Doppler and PW Doppler studies.
• 21366A, the Pediatric Biplane transducer, enabling both transverse and lon-
gitudinal views, and steerable CW Doppler for examinations of pediatric and
small adult patients.
Audience
Philips TEE transducers are designed for use by physicians who are properly
trained in esophagogastroscopic techniques according to currently approved
relevant medical practices. It is recommended that physicians operating any
Philips TEE transducer are:
Patient Safety
Philips transducers are built to be safe, sturdy, and reliable. However, on a few
occasions patients have experienced complications due to improper use of a TEE
transducer. This section provides information on patient safety when using a TEE
transducer.
Philips recommends that To operate a TEE transducer, you must be a physician who is properly trained in
you practice using the esophagogastroscopic techniques, according to currently approved relevant
controls before medical practices.
performing every
procedure mentioned
here. You also must be thoroughly familiar with the safe operation, care, and
maintenance of the Philips ultrasound imaging system used with the transducer,
as well as proficient at interpreting the images generated. As always, use
informed judgement when selecting patients for TEE exams.
You can help ensure patient safety when using a TEE transducer by:
• Verbally preparing each patient for the procedure prior to the exam.
• Scrutinizing the entire transducer and testing all of the controls before each
use.
• Inserting, removing, and operating the transducer properly.
• Ensuring that the transducer handle does not rest on or touch the patient.
• Using protective equipment, such as a bite guard and a market-approved
sheath, during a TEE exam. See “Accessory Equipment” on page 4-48.
• Not allowing water or other liquids to come in contact with the transducer
connector, the interior of the system, or onto the keyboard.
NOTE To avoid damaging gastroscope cables, be sure that the distal tip is in the neutral
(straight) position when inserting a transducer into or removing it from a sheath.
The following table summarizes patient safety problems, how to prevent them,
and the pages in this guide where details are provided.
Mechanical Severe trauma, cuts, Inspect transducer using both sight and page 4-32
damage bleeding, perforations touch before exam.
Electrical damage Esophageal burns Check transducer for frayed insulation, kinks, page 4-7
or other abnormalities. Follow electrical
safety checking procedures.
Biting, scraping Tooth damage, esophageal Use a bite guard. page 4-48
transducer burns
Insufficient Spread of illness or disease Thoroughly clean and disinfect transducer page 2-18
cleaning protocol after each use. Cover tip and shaft with
sheath. Cover imaging system with
disposable drape if highly pathogenic
organisms are known or suspected.
Improper insertion Esophageal cuts, bleeding, Never use force when inserting, removing, or page 4-34
or withdrawal ligament damage, manipulating transducer. During insertion,
perforations lock the medial/lateral controls. During
withdrawal, place the medial/lateral controls
in free-wheeling mode.
Pressure necrosis Death of esophageal lining Keep deflection controls in free-wheeling page 4-34
tissue mode and unplug transducer from system
when not imaging. Minimize the pressure
applied to deflection area and distal tip. Do
not let distal tip displace a tissue area for
more than five minutes.
Increased Esophageal burns Use the TEE preset which has been page 4-38
transducer established to minimize the effects of
temperature temperature. For febrile patients, use the
auto cool feature.
Improper patient Transient unilateral vocal Never use the transducer during any page 4-34
position cord paralysis procedure requiring extreme neck flexion,
such as sitting craniotomies.
Non-isolated Electrical burns Only use isolated-output electrosurgical units page 4-7
ESUs (ESUs); the ESU label or service guide, or
your biomedical department, should identify
whether or not the ESU is isolated. Unplug
transducer from the system when you are not
imaging.
NOTE The Philips’ diagnostic ultrasound systems and transducers in this manual do not
contain natural rubber latex that contacts humans. Natural rubber latex is not
used on any ultrasound transducer, including transthoracic, intraoperative, and
transesophageal transducers. It also is not used on Philips ECG cables for the
products in this manual.
Equipment Operation
• Cuts and abrasions on the transducer and insulation from teeth or sharp
instruments, such as scalpels, scissors, and clamps.
Review the following table to familiarize yourself with specific problems, how to
avoid them, and the pages in this guide where details are provided. Philips also
strongly recommends that you clearly post stringent protocols for TEE transducer
care, based on the information in this guide, to minimize the chance of damage.
Current leakage Serious electrical hazards Check transducer for cuts, frayed page 4-7
insulation, kinks, or other abnormalities.
Follow electrical safety checking
procedures.
Biting transducer Mechanical and electrical Cover patient’s teeth with a bite guard. page 4-48
hazards Cover the distal tip and flexible shaft with
a sheath (optional).
Forcing Breaks steering mechanism Operate deflection controls gently. page 4-11
deflection
controls
Incorrect storage Possible damage to highly Suspend transducer from a wall- page 2-3
sensitive elements, cuts in mounted rack and cover distal tip with tip
flexible shaft protector when not in use.
Internal Severe transducer damage that Never sterilize transducer using bleach, page 2-18
exposure to affects image quality, steering steam, heat, or ethylene oxide. Never
liquids mechanism, and electrical immerse steering mechanism in any
safety disinfectant or liquid.
Any other Potential severe damage Do not use transducer. Contact your
irregularity Philips service
representative
Electrical Safety
The Philips ultrasound systems and transducers discussed in this manual comply
with common medical device electrical safety standards.
Leakage Current
If the outer layer of the shaft is punctured or cracked, a patient’s esophagus could
be exposed to chassis leakage current. This leakage current is not hazardous
provided that the ground connector (third wire) in the ultrasound system power
cable is intact and connected to a properly grounded wall outlet. Even if the
ground connector breaks, leakage current will not exceed 100 microamperes
when connected to a 120 volt outlet, or 200 microamperes when connected to a
240 volt outlet.
Leakage hazards are further reduced when the ultrasound system is plugged into
an isolated power outlet, which is standard in most operating rooms.
Electrosurgical Units
Electrosurgical units (ESUs) and other devices intentionally introduce radio
frequency electromagnetic fields or currents into patients. Because imaging
ultrasound frequencies are coincidentally in the radio frequency range,
ultrasound transducer circuits are susceptible to radio frequency interference.
While an ESU is in use, severe noise interferes with the black and white image,
and completely obliterates the color image.
Combinations of failures in an ESU or other device and in the outer layer of the
transducer shaft can cause electrosurgical currents to return along the transducer
conductors. This could burn the patient, and the ultrasound system and transducer
could also be damaged. Be aware that a disposable sheath provides no protective
electrical insulation at ESU frequencies.
Pacemakers
Ultrasound equipment in normal operation, as with other medical electronic
diagnostic equipment, uses high-frequency electrical signals that can interfere
with pacemaker operation. Though the possibility of interference is slight, be
alert to this potential hazard and halt system operation immediately should you
note interference with a pacemaker.
Defibrillators
The outer insulating layer of the transducer, if intact, is designed to withstand
electrical stress created by defibrillation. A small hole in the outer layer of the
transducer, however, opens a conductive path to grounded metal parts in the
transducer. The secondary arcing that could occur during defibrillation could
cause patient burns. The risk of burns is reduced but not eliminated by using an
ungrounded defibrillator. Therefore, disconnect the transducer and remove the
transducer connector from the system before defibrillation. A disposable sheath
provides no protective electrical insulation against defibrillation.
Accident Prevention
Visually and tactually inspect a TEE transducer for bumps, cracks, and cuts prior
to each TEE exam. A small bump on the shaft surface could indicate that a strand
from the ground shield has broken and is beginning to puncture the outer layer. If
you suspect a problem with the flexible shaft, perform the electrical safety
checking procedure provided in the next section.
Use ESUs that have isolated outputs. Return fault/ground fault detection circuits
provide additional protection. (To determine if an ESU has an isolated output,
read the label on the ESU, refer to the ESU service guide, or ask a biomedical
engineer.)
Use defibrillators that do not have grounded patient circuits. Refer to the
defibrillator service guide, or consult with a biomedical engineer to determine
whether or not a defibrillator patient circuit is grounded.
Before defibrillation, always disconnect and remove the transducer from the
system.
Require periodic electrical safety checks to ensure that the grounding system in
your area remains intact.
If the transducer is left in a patient during periods when imaging is not taking
place, unplug the transducer from the system to reduce the possibility of leakage
current or ESU interaction. Also make sure that the deflection control brakes are
off and that the transducer is in free-wheeling mode.
NOTE A leakage current detector (the Dale 800) is listed on the Philips Transducer web
site at www.medical.philips.com/transducercare. For an example how to test cur-
rent leakage, using a Dynatech Nevada 232 Safety/ECG Analyzer, see “Leakage
Current Testing” on page 5-6.
This section discusses the basic parts of a TEE transducer and lists important
features of the following TEE transducers:
Deflection controls
Control housing
Distal tip
Deflection area
Flexible shaft
Transducer connector
008E11C
The deflection controls on the TEE transducers discussed in this manual move
the deflection area, located between the distal tip and flexible shaft. The
deflection area bends when you operate the controls, permitting anterior,
posterior, and lateral positioning for most transducers.
Never apply excessive force when operating the deflection controls during
imaging. Don’t allow the transducer to remain at a maximum deflection for
long periods of time.1.
WARNING To prevent tissue damage Philips recommends that the tip of the transducer
be straightened and both friction brakes released before withdrawing the
transducer from the patient. In the neutral position the tip is straight when
both white circles on the control wheels are aligned and point toward the
center of the array rotation button.
1. Urbanowitz, John, H., et al., “Transesophageal Echocardiography and its Potential for Esophageal
Damage,” Anesthesiology 72:1, 1990.
2. Radwin, Martin, et al., “Transesophageal Echocardiography: Intubation Techniques,” Philips
Application Note 5091-2804E, 1992.
Temperature Controls
The TEE transducers discussed in this manual have temperature sensors in the
distal tip. An automatic warning appears on the ultrasound screen if the distal tip
temperature exceeds a specified limit, and the system discontinues scanning.
Use the TEE auto cool feature to enter above-normal patient temperatures.
Entering a patient’s temperature enables the system to calculate tip temperature
more accurately, and can prevent unnecessary interruptions while scanning. See
page 4-38 for details of the auto cool feature.
OmniPlane
(Model 21364A)
210e003a
Features: Enables high resolution imaging and 360o views of the heart,
unobstructed by lungs and ribs. Incrementally rotates from the
horizontal plane at 0o, to the vertical plane at 90o, to a left/
right reversed horizontal plane at 180o. Enables lateral and
anterior/posterior tip movement, with an adjustable friction
brake. Capable of color flow imaging, steerable CW and PW
Doppler, and frequency agility. Tip surface constantly moni-
tored for patient safety. Convenient hanging ring.
NOTE Philips recommends that you only use the OmniPlane transducer on patients
weighing at least 40 kg (88 lbs), to ensure the esophagus can comfortably accom-
modate the transducer.
OmniPlane II
(Model 21367A and 21369A)
5apc040
Features: Enables high resolution imaging and 360o views of the heart,
unobstructed by lungs and ribs. Incrementally rotates from the
horizontal plane at 0o, to the vertical plane at 90o, to a left/
right reversed horizontal plane at 180o. Enables lateral and
anterior/posterior tip movement, with an adjustable friction
brake. Capable of color flow imaging, steerable CW and PW
Doppler, and frequency agility. Tip surface constantly moni-
tored for patient safety. Convenient hanging ring.
NOTE Philips recommends that you only use the OmniPlane II transducer on patients
weighing at least 25 kg (55 lbs), to ensure the esophagus can comfortably accom-
modate the transducer.
Medial/lateral control
Anterior/posterior control
The larger inner wheel controls anterior/posterior movement, while the smaller
outer wheel controls medial/lateral movement. Aligning the white circles on each
wheel with the center of the array rotation buttons places the tip of the TEE probe
into the neutral position.
Each wheel has a friction brake that holds the tip position without locking it in
place. Use the minimal setting needed to maintain tip position. This allows the tip
to straighten if it meets additional resistance. The brake for the larger wheel
(anterior/posterior movement) is the lever with the arrow on it. Push the lever in
the direction of the arrow to increase the resistance, or holding power, of the
brake. The brake for the smaller wheel (medial/lateral movement) is the knob
that sits on top of the wheel. Turn this knob clockwise to increase resistance.
CAUTION When repositioning or withdrawing a TEE transducer, put the controls in free-
wheeling mode. If the transducer is left in the patient for an extended period,
unplug the transducer from the system to reduce the possibility of leakage current
or ESU interaction. Also, make sure the transducer is in free-wheeling mode.
You can rotate the OmniPlane array to achieve a 360o view of the heart in the
following ways:
a. To increase rotation, press the array rotation button that is distal to the
system.
c. Turn Seek Angle to locate a specific angle. When you touch Seek, the
array rotates to the specified angle.
210e005a
Before connecting the transducer, be sure the tip and shaft are straight. When you
lock the connector into the transducer slot, the OmniPlane transducer recalibrates
itself to the 0o (horizontal plane) position. As with any array movement, the
motor makes a humming sound while recalibrating.
If you see a “Calibration failed” message on the screen, you can continue to
image. However, the transducer may lack the full 180o rotation capacity, and the
rotation displayed on the screen could be inaccurate.
1 Make sure that the transducer’s shaft and tip are straight.
1 Make sure that the transducer’s shaft and tip are straight.
If you still see the message, contact your Philips service representative.
miniMultiplane
(Model 21381A)
WARNING The miniMulti TEE transducer is not intended for fetal imaging.
Philips recommends that you only use the miniMulti TEE transducer on patients
weighing at least 3.5 Kg (7.7lbs.)
CAUTION The ability of a patient to swallow or accommodate the transducer should be con-
sidered.
The smaller wheel on the handle is for controlling the transducer tip deflection.
This wheel has a locked and freely moving mode. In the locked mode the
movement of the deflection wheel is restrained. This is used to hold the tip in a
certain position. A metallic ring around the body of the handle, which clicks on/
off, controls the lock.
WARNING Check if the maximum deflection of the tip is 120° upward and 90° down-
wards. If the up/down deflection shows an unwanted amount of free play or
exceeds the maximal deflection angles given above, do not use the trans-
ducer. Contact the service organization to re-adjust the steering of the
Lock operation
To protect the patient and the transducer, unlock the deflection control when
inserting or withdrawing the transducer.
White
Unlock
Red
Lock
Short Axis
Angle = 0°
Long Axis
Angle = 90°
Short Axis
Angle = 180°
CAUTION Do not use excessive force on the scanplane rotation control wheel at its outer
positions as this may damage the endoscope mechanism.
Pediatric Biplane
(Model 21366A)
0008e9a
Be sure to read Features: Enables high resolution imaging unobstructed by lungs, ribs,
“Biplane Images” on or scar tissue. Capable of showing both transverse and longi-
page 4-27 for more tudinal images as well as color flow imaging, PW Doppler,
information.
and steerable CW Doppler.
NOTE Philips recommends that you only use the Pediatric Biplane transducer on
patients weighing at least 3 kilograms (6.6 lbs).
Medial/lateral control
Anterior/posterior control
The larger inner wheel controls anterior/posterior movement, while the smaller
outer wheel controls medial/lateral movement. Align the white circles on each
control wheel perpendicular to the control handle for a neutral position of the
TEE tip.
Each wheel has a detent that holds the tip position without locking it in place.
Use the minimal setting needed to maintain tip position. This allows the tip to
straighten if it meets additional resistance. The knob detents operate as illustrated
in Figure 4-5.
CAUTION When repositioning or withdrawing a TEE transducer, put the controls in free-
wheeling mode. Also whenever the transducer is not being used during a proce-
dure, make sure it is in free-wheeling mode and unplugged from the system.
Biplane Images
The default orientation for both transverse and longitudinal images is a sector
with the apex at the top. A default longitudinal image is shown below.
5.0 MHZ-V
300e004
You can configure the longitudinal image so that it is always initially inverted,
which makes the distinction between transverse and longitudinal images more
obvious. If you configure inverted longitudinal images, transverse images retain
apex-at-the-top orientations. You also can reverse the longitudinal image and
orientation icon.
2D COLOR PW CW ANGIO
300e005
2D COLOR PW CW ANGIO
Secondary
Controls
300e005
Before each TEE exam, carefully inspect the transducer and try the controls, as
described in the following sections.
Carefully inspect the entire surface of the distal tip and flexible shaft for
protrusions, holes, dents, abrasions, cuts, burrs, or cracks that could be extremely
hazardous to both you and your patient. Carefully feel the tip and shaft for
problems, and visually scrutinize the entire transducer. If you suspect an
electrical problem, follow the electrical safety checking procedure on page 4-7.
Use the deflection controls to position the tip every way possible, both to ensure
that the controls work properly and to get used to the feel of the particular TEE
transducer. Make sure that the controls operate smoothly without binding, and
that you can achieve all possible positions easily before introduction of the TEE
transducer into a patient.
• Consider the patient’s size and ability to accommodate the transducer tip and
shaft.
• Check for any history of gastroesophageal disease or difficulty swallowing.
• Evaluate the potential overall effects of mediastinal radiation, chemotherapy,
anticoagulation, or steroid therapy.
• Remain aware that you may discover unsuspected esophageal pathology
during an exam. Particularly with pediatric patients, be alert for congenital
problems with the esophagus or stomach.
• Use the auto cool feature when examining a patient with an above-normal
temperature. The auto cool feature is described on page 4-38.
Exam Guidelines
During a TEE exam, an assistant can provide oral and pharyngeal suctioning of
the patient, and can monitor the patient’s blood pressure and general responses.
For unexpected occurrences, an emergency cart with basic life support
equipment should be ready. Throughout the exam it is important to carefully
monitor the patient’s reactions and assure that ventilation and vital signs are
stable.
In the operating room, do not use TEE transducers during surgical procedures
requiring extreme neck flexion, such as sitting craniotomies.1, 2
This problem has occurred rarely, but its consequences can be serious. Please see
“Tip Fold-Over Problem” on page 4-36 for details.
Do not let the distal tip displace any one segment of tissue for more than five
consecutive minutes. Also make sure the deflection area and the distal tip are in a
position of least potential pressure, and that the transducer is in a free-wheeling
mode and unplugged whenever you are not imaging.1
Philips recommends that you stop TEE scanning and unplug the transducer from
the system during periods of poor perfusion, circulatory arrest, or the
hypothermic phase of open heart surgery. To discontinue scanning, unlock the
transducer connector according to the instructions in “Connecting Transducers”
on page 1-1.
Before each TEE exam, carefully inspect the transducer, as described earlier
in this chapter.
A thorough inspection procedure is required for the safety of the patient and
yourself, and to ensure the continued correct functioning of the transducer.
Refrain from handling the distal tip, and do not clamp any part of the flexi-
ble shaft.
If you must handle the distal tip, grasp it on the sides. Do not touch the top or
bottom. Support the transducer’s proximal head, either by having an assistant
1. Urbanowitz, John H., et al., “Transesophageal Echocardiography and its Potential for Esophageal
Damage,” Anesthesiology 72:1, 1990.
Problem Prevention
The following factors could cause the tip to fold over. Other factors could also be
involved; the list below is not exhaustive.
Insertion technique
You might find transducer insertion easier if you guide the transducer into the
patient’s mouth with your fingers. You also might want to block medial/lateral tip
movement.
NOTE All patients should wear a bite guard during a TEE exam. A bite guard protects
against dangerous transducer mechanical and electrical malfunction caused by
involuntary biting. Even anesthetized patients require bite guards to prevent dam-
age to both their teeth and the transducer. See “Bite Guard” on page 4-48 for
information on the bite guards available from Philips.
Definitely avoid the following situations when inserting any TEE transducer into
a patient:
Problem Recognition
The TEE transducer tip could be folded over in the patient if you encounter:
Problem Correction
If you suspect the transducer tip is folded over, our physician consultants
recommend that you gently try to manipulate the transducer. If the tip is not
locked or jammed in a doubled-over position and you can move it forward,
advance the transducer into the stomach. Then straighten the tip and remove the
transducer.
If you cannot move the tip in any direction, our consultants recommend that you
x-ray the patient to evaluate the situation. You might also want to involve a
gastroenterologist or anesthesiologist.
NOTE The patient temperature displayed on the ultrasound screen is always either 37oC
or the temperature that you manually enter. The system does not report actual
patient temperature.
If the distal tip temperature reaches 41oC (105.8oF), a warning message appears
and the transducer temperature displays in inverse video. If the temperature
reaches 44oC (111.2oF), a warning appears with the patient and transducer
temperatures, and the system stops imaging until the distal tip cools to 43.5oC or
lower.
WARNING To avoid risk of esophageal burn for adult patients, minimize the time spent
imaging at distal tip temperatures in excess of 43oC. Exposure should be
limited to 10 minutes or less at 43oC or above.
WARNING To protect the patient and ensure the effectiveness of the auto cool feature,
always enter the correct patient temperature.
Each time you turn off or reset the system, the system assumes a patient temperature
of 37oC. To ensure patient safety, always enter a patient’s temperature if it is above
normal.
1 Touch Probe.
Probe
Probe
Pat Temp
37.0oC
3 The current patient temperature displays beside the PAT T: label on the
imaging screen.
o
T:3 7.0 C
PAT
6apc0158
Filter Power
Size Position
Body Marker Zoom
Frequency
Probe
Angle Calcs Dual
Baseline Hx M Mode
Dynamic
Left Right Range
Angio Caliper
2D
Scale Tape
CW Trace GAIN
Focus
Print 2
PW Spectral Erase
Gain Freeze
Color Print 1
Enter Depth
6APC0045
Probe Control
PAT T:37.0oC
TEE T:37.0oC
300e004
A less-than sign (<) after TEE T indicates that the transducer’s distal tip
temperature is below the patient temperature (PAT T) assumed by the system
(either 37oC or the temperature you entered).
A colon (:) after TEE T indicates that the tip temperature is greater than or equal
to the patient temperature.
If the distal tip reaches 41oC (105.8oF) during imaging, a message appears on the
screen (Figure 4-10.).
PAT T:37.0oC
TEE T:41.0oC
300e004
If the distal tip reaches 44oC (111.2oF) during imaging, a message displays on the
screen (Figure 4-11.).
PAT T:37.0oC
TEE T:44.0oC
300e004
If you are using a Biplane transducer, the system displays both the horizontal
(TEE H) and vertical (TEE V) temperatures. If either one of these temperatures
reaches 44oC, imaging stops and does not resume until the temperatures for both
arrays are below 43.5oC.
Resuming Imaging
If the distal tip temperature drops to 43.5oC, the system resumes imaging. If the
auto cool message persists longer than one minute, or if an error message
appears, contact your Philips ultrasound service representative.
WARNING The “Reconnect the transducer.” error message is often caused by a poorly
seated transducer connector, but could be caused by a failure in the auto
cool safety logic. In the case of a logic failure, distal tip temperatures could
reach 46.5oC in hyperthermic patients (40 – 41oC) before the error causes
scanning to stop. At this temperature, esophageal burns may occur.1 , 2
2 Press Reset
Reset .
If the system does not resume imaging immediately, contact your Philips service
representative.
Probe
Even when this control 2 Touch TEE Temp Display to turn the temperature display on or off.
is off, the system
displays the
transducer Probe
temperature if it
approaches the
temperature limit.
TEE Temp
Display
TEE Temp
oC
1 Press Setup .
2 Highlight Preset Parameters from the menu and then press Enter .
3 Using the trackball, highlight the parameters you want to change:
If you choose not to have the temperatures always displayed, the patient
temperature only shows if you enter a patient temperature above 37oC. During an
auto cool warning or activation, both the patient and distal tip temperature
appear, regardless of the setting.
You should follow your institutional guidelines for post TEE exams. In addition,
you might want to include the following recommendations in your guidelines.
These recommendations are neither exhaustive nor restrictive, but suggest steps
you might want to incorporate into your post TEE exam routine.
Accessory Equipment
Each TEE transducer comes with a disposable bite guard and tip protector. This
section describes these accessories, and provides ordering information. It also
mentions circumstances that warrant the use of a disposable drape.
0008e2a
CAUTION The M2203A bite guard strap contains natural rubber latex, which may cause
allergic reactions.
Tip Protector
When not using a carrying case to transport a TEE transducer, you should use a
tip protector on its distal tip. The tip protector helps prevent serious damage to
the transducer’s lens. Philips supplies tip protectors designed for each of its TEE
transducers.
To order the following tip protector in the United States, contact the Philips
Support Materials Center at 1-800-722-0659. In other countries, contact your
local Philips sales representative.
Disposable Drape
During exams where you believe contamination of the imaging system can occur,
it is recommended that you take universal precautions and cover the imaging
system with a disposable drape. Consult your hospital’s rules regarding
equipment use in the presence of infectious disease.
Overview
• 21390A, the 15.0 - 6.0 MHz UltraBand linear transducer, provides high-res-
olution near field imaging to assist surgeons in assessing arterial reconstruc-
tions, vein bypass grafts, or revision for minute defects. It allows trapezoidal
imaging. The 21390A provides a small profile, making it easy to maneuver,
and enabling interrogation of a site with restricted access or within a very
small incision.
• 21380A, the high-frequency UltraBand Sector transducer, enabling excellent
2-D resolution by providing imaging in a range of frequencies, up to 12.0
MHz. Provides capability for intraoperative imaging. Small, lightweight
design with a thin, flexible cable permits easy manipulation for neonatal and
pediatric patients and on the organ surface.
• 21358B, the trapezoidal linear array transducer, used for trapezoidal imaging
in the operating room for greater opportunities to access wide-field imaging
during detailed surgical procedures.
Audience
Philips intraoperative transducers are designed for use under the guidance of
physicians who are properly trained in intraoperative ultrasound imaging
techniques, according to currently approved relevant medical practices. It is
recommended that physicians operating any Philips intraoperative transducer are:
Intended Use
You can also use the transducers discussed in this chapter for adult, pediatric, and
neonatal heart and vessel imaging.
NOTE For intraoperative use on the heart and central circulatory system, only the
21390A 15-6L intraoperative transducer is approved in EU countries.
When the 21390A probe is used for animal studies, the animal use label
must be attached.
1 Wipe the side of the connector, opposite the serial number label, with an iso-
propyl alcohol moistened cloth and dry with a soft cloth.
2 Place the label on the transducer connector housing at the indented area
(opposite the serial number label).
Patient Safety
This section lists conditions that can adversely affect patients when you use an
intraoperative transducer.
You can help ensure patient safety when using an intraoperative transducer by:
NOTE The diagnostic ultrasound systems and transducers discussed in this manual do
not contain natural rubber latex that contacts humans. Natural rubber latex is not
used on any ultrasound transducer, including transthoracic, intraoperative, and
transesophageal transducers. It also is not used on SONOS or ImagePoint ECG
cables.
Equipment Operation
• Cuts and abrasions on transducer insulation and lens from sharp instruments
such as scalpels, scissors, and clamps.
• Improper disinfection techniques, causing fluid to enter the transducer.
• Damage caused by dropping the transducer on a hard surface.
NOTE If you find any signs of damage to the transducer, patient safety may be in
question. Do not use the transducer and contact your Philips service representa-
tive.
Philips strongly recommends that you clearly post stringent protocols for
intraoperative transducer care, based on the information in this guide, to
minimize the chance of damage.
Electrical Safety
All Philips ultrasound systems and transducers comply with common medical
device electrical safety standards.
Leakage Current
SONOS and ImagePoint transducers that are approved for intraoperative use are
labeled on the transducer connector as type CF in accordance with IEC
60601-1. Type CF classification indicates that the degree of protection from
electrical shock afforded by the transducer is suitable for all patient applications
including direct cardiac and intraoperative applications.
NOTE Attach the transducer to the ultrasound system and test with an appropriate safety
analyzer to ensure that IEC 60601-1 type CF leakage limits are not exceeded.
The following procedure tests the current leakage, using a Dynatech Nevada 232
Safety/ECG Analyzer. This procedure shows one example. The procedure for
your safety analyzer may be very different.
NOTE Another leakage current detector (the Dale 800) is listed on the Philips Trans-
ducer web site at www.medical.philips.com/transducercare.
Hot System
Transducer
Under
Neutral immersed in
Test
saline solution
AC Micro
Ground green or Main power switch “ON” ammeter
green/yellow
6 Set the safety analyzer Leads switch to the ALL (all ECG Leads) position.
7 Connect a lead to any ECG terminal on the safety analyzer and immerse the
other end into the saline solution.
10 Hold the safety analyzer Ground Open switch in the Open position (Single
Fault condition) and note the leakage reading on the output display.
11 Repeat Steps 9 and 10 with the Polarity switch in the Reverse position. The
applicable limits are:
10 microamps rms (Normal condition)
50 microamps rms (Single fault condition)
Test # 2 - Transducer Leakage Current Test with Mains Voltage Applied (Sink)
6 Set the safety analyzer Leads switch to the Isolation Test position.
7 Connect a lead to any ECG terminal on the safety analyzer. Leave the other
end of the lead disconnected for now.
8 Hold the Isolation Test button down and note the leakage reading on the out-
put display. This is the correction factor which will be subtracted from the
final reading.
9 Immerse the other end of the lead into the saline solution.
10 Hold the Isolation Test button down and take the leakage reading again. Sub-
tract the correction factor found in step 8 to get the accurate leakage measure-
ment. The leakage must be less than the following limit:
50 microamps rms
Defibrillators
The outer insulating layer of the transducer, if intact, is designed to withstand
electrical stress created by defibrillation. A small hole in the outer layer of the
transducer, however, opens a conductive path to grounded metal parts in the
transducer. The secondary arcing that could occur during defibrillation could
cause patient burns. The risk of burns is reduced but not eliminated by using an
ungrounded defibrillator. Therefore, remove the transducer from the patient
before defibrillation. A disposable sheath provides no protective electrical
insulation against defibrillation.
Accident Prevention
Use defibrillators that do not have grounded patient circuits. Refer to the
defibrillator service guide, or consult with a biomedical engineer to determine
whether or not a defibrillator patient circuit is grounded.
This section lists important features of the 21390A, 21380A, and 21358B models
of intraoperative transducers.
Also discussed in this section is how to prepare the transducers for use in
intraoperative applications.
Features: UltraBand 15.0 - 6.0 MHz, 128 element transducer which pro-
vides high-resolution near field imaging to assist vascular sur-
geons in assessing arterial reconstructions, vein bypass grafts,
or revision for minute defects. The 21930A allows trapezoidal
imaging. It provides a small profile, which is easy to maneu-
ver, enabling interrogation of a site with restricted access or
within a very small incision. Capable of color flow imaging,
Angio, PW Doppler, and Intelligent Doppler.
High-Frequency UltraBand
(Model 21380A)
s12
6apc0165
If you are using the Models 21380A, 21390A, or 21358B transducers, use a
sheath filled with sterile gel or saline according to the directions in Steps 1 and 2.
1 Place 20 cc of sterile gel or saline into the sheath. Visually inspect the sheath
for leaks before proceeding.
2 Insert the transducer into the sheath and unfurl the sheath until it covers the
transducer and its cable. The sheath must be unfurled far enough to maintain
the sterile field.
416e015 6apc0144
3 The use of a sterile elastic band or clip to hold the proximal end of the sheath
in place is recommended.
4 Ensure that wrinkles over the face of the transducer are minimized. Check the
sheath for tears or damage before proceeding.
5 When operating the transducer, make sure that proper orientation is main-
tained to avoid interpretation confusion.
6 Imaging improves with adequate coupling between the patient surface and
sheath surface. Sterile water works as a good acoustic coupling agent during
surgery.
NOTE Make sure that the imaging surface is moist so that good acoustic contact is
achieved. A dry surface results in excessive artifact.
Disposable Drape
During exams where you believe contamination of the imaging system can occur,
it is recommended that you take universal precautions and cover the imaging
system with a disposable drape. Consult your hospital’s rules regarding
equipment use in the presence of infectious disease.
Accessory Equipment
Table 6-1 Supplies and Accessories For Use With SONOS and ImagePoint Transducers
ULTRASOUND GEL
Basin 21110A
Ordering Sheaths
To order transducer sheaths for the Models 21380A, 21390A, and 21358B
transducers, contact:
If you use a transducer sheath other than the one listed above, you must use a
market cleared sterile sheath and ensure that the following minimum dimensions
are met.
Overview
Electrical Safety
Defibrillators
The outer insulating layer of the transducer, if intact, is designed to withstand
electrical stress created by defibrillation. A small hole in the outer layer of the
transducer, however, opens a conductive path to grounded metal parts in the
transducer. The secondary arcing that could occur during defibrillation could
cause patient burns. The risk of burns is reduced but not eliminated by using an
ungrounded defibrillator. Therefore, remove the transducer from the patient
before defibrillation. A disposable sheath provides no protective electrical
insulation against defibrillation.
Accident Prevention
Use defibrillators that do not have grounded patient circuits. Refer to the
defibrillator service guide, or consult with a biomedical engineer to determine
whether or not a defibrillator patient circuit is grounded.
NOTE The Philips’ diagnostic ultrasound systems and transducers in this manual do not
contain natural rubber latex that contacts humans. Natural rubber latex is not
used on any ultrasound transducer, including transthoracic, intraoperative, and
transesophageal transducers. It also is not used on Philips ECG cables for the
products in this manual.
.
Transthoracic OmniPlane
(Model 21349A)
6APC0154
You can rotate the Transthoracic OmniPlane array to achieve a 360o view of the
heart by using the array rotation controls on the ultrasound system’s touch panel.
3 Turn Seek Angle to locate a specific angle. When you touch Seek, the array
rotates to the specified angle.
4 Touch Slow Speed to slow down the speed of the up and down rotation.
You can move the array up to 180o. An icon shows the current degree of rotation.
Depending on image orientation, the icon appears in the upper or lower right side
of the imaging screen. Because the center of the array is the pivot point, you can
achieve a 360o view, as shown in the following figure.
0° 0° 180°
0° 0°
0° 45° 180°
45 °
45 °
0° 180° 90 °
90°
90 °
0° 135 ° 180°
135 ° 135 °
0° 180° 180°
180 ° 180 °
6apc0151
When you lock the connector into the transducer slot, the Transthoracic
OmniPlane transducer calibrates itself to the 0o (horizontal plane) position. As
with any array movement, the motor makes a humming sound while calibrating.
If you see a “Omni Transducer Calibration failed” message on the screen, you
can continue to image. However, the transducer may lack the full 180o rotation
capacity, and the rotation displayed on the screen could be inaccurate.
If you still see the message, contact your Philips service representative.
Index-1
Index
Index-2
Index
Index-3
Index
Index-4
Index
Transthoracic OmniPlane
transducer
array rotation controls
6-4
calibration failed
message 6-6
connection information
1-15
defibrillator safety issues
6-1
model 21349A described
6-1
revision levels for using
1-15
specifications for 6-3
W
warranty
coverage limits indicated
2-13, 2-15, 2-17, 2-18,
2-20, 2-22
web site
for cleaning transducers
Preface-1, 2-9
for FDA-cleared
disinfectants 2-12
leakage current testing
product 5-6
what gels to use 2-5
Index-5
Index
Index-6