CARESCAPE R860 TRM 2065496-001
Installation Checklist
.
Customer Name: Service Record
Number:
System ID or Equipment ID:
Serial Number:
Software version:
Performed by:
Date:
Inspect the system
.
Procedure Expected results Outcome Comments
Pass / Fail
• Check for shipping damage. If • No shipping
any damage is found, take damage was
photos and notify the shipping observed.
carrier. • Ventilator is
• Unpack and assemble the unpacked and
ventilator. Refer to "Setup and properly
Connections" in the User’s assembled.
Reference Manual (URM). • Protective tape has
• Remove protective tape from been removed from
cart casters, if applicable (cart cart casters.
only). • Casters roll and
• Verify that casters are not lock as needed and
loose or damaged. (Refer to are firmly mounted
"Replacing the cart caster" in to machine base
the Technical Reference
Manual TRM) for detailed
information).
1
Power cords
.
Procedure Expected results Outcome Comments
Pass / Fail
• Verify AC power cord • Power cord
configuration. configuration is correct.
• Check that all AC • All power cords are
power cords are fully seated correctly.
seated.
• Power clamp screws
• Check and if necessary are tightened to 1.0
tighten the following: Nm and the retainer
power clamp screws to and outlet screws are
1.0 Nm; tighten the tightened to 0.45 Nm.
inlet power cord
retainer
(1503-3033-000) screw
to 0.45 Nm; tighten the
outlet power cord
retainer
(1505-3041-000) screw
to 0.45 Nm.
Verify annunciators
.
Procedure Expected results Outcome Comments
Pass /Fail
• Power up system and listen • Buzzer sounds for
for the buzzer sounds. 2 seconds when
• Verify main audio tones. switch is turned On.
• Verify the display unit green • Main audio alarm
LED illuminates. sounds briefly just
• Verify the display alarm LED before the system
bar is active. startup screen
displays.
• Verified the green
LED illuminates.
• LEDs blink: red,
yellow, blue.
2 2065496-INC A 10 2015
Configuration/System tests
.
Procedure Expected results Outcome Comments
Pass / Fail
• Verify the factory installed • Factory installed
ventilation options match the options match the
configuration purchased with configuration
the machine. purchased.
• Select Service log and Alarm • Reviewed error and
log and review error and alarm log entries to
alarm log entries that were identify any
entered previous to the facility installation failures.
installation. These are log • Touchscreen is
entries that were generated calibrated and
during the normal tracks touches
manufacturing process. Look accurately.
for any new log entries that • Display indicates
have occurred after that calibration
installation. Use the Reset passes for each of
logs (if you have SP04) after the following:
reviewing alarm entries.
• O2 FCV
• Select the Touchscreen tab
• Air FCV
and perform the Touchscreen
• Exhalation
calibration.
valve
• Access the Service
Calibration menus and • Verified Calibration
perform: Flag is set to Off
(unchecked)
• O2 FCV
• “Calibration
• Air FCV
Required” message
• Exhalation valve
is not displayed.
• Set Calibration flag to Off • All elements of the
(uncheck). System checkout
• Power system Off then On pass.
and verify the “Calibration • The “On Battery”
Required” message is not alarm and the
shown in the lower left portion battery indicator
of the display. icon display.
• Perform System Check
(automated checkout)
• Perform Power Failure test.
2065496-INC A 10 2015 3
Electrical safety test
.
Procedure Expected results Actual results Outcome Comments
Measure the ground The measured Pass / Fail
resistance. ground resistance is
less than 0.2 Ohms.
Measure the leakage The measured Pass / Fail
current. leakage current is
less than 300
μAmps @ 120 / 100
Vac and is less than
500 μAmps @ 220 /
240 Vac.
4 2065496-INC A 10 2015
CARESCAPE R860 TRM 2065496-001
EVair compressor (if installed)
Refer to the EVair Technical Reference Manual 2066031-001 for
detailed instructions.
Installation Checklist
.
Customer Name: Service Record
Number:
System ID or Equipment ID:
Serial Number:
Software version:
Performed by:
Date:
Inspect the compressor
.
Procedure Expected results Outcome Comments
Check the AC power strain relief Power cord is secured Pass / Fail
clamp. by the strain relief
clamp.
Verify annunciator
.
Procedure Expected results Outcome Comments
Display LCD illuminates when Display illuminates when Pass / Fail
compressor is plugged in AC. power is turned on.
5
Pressure 01 functional test
.
Procedure Expected results Actual results Outcome Comments
Perform the Pass / Fail
• The difference
compressor pump-up
between the two
test.
displayed
pressures does
not exceed 60
kPa (0.6 bar, 8.7
psi).
• The compressor
pressure reached
300 kPa (3.0 bar,
43.5 psi) within
15 seconds or
less.
Pressure threshold settings
.
Procedure Expected results Actual results Outcome Comments
Perform the following Pass / Fail
• The lower
steps if the
pressure
compressor is not
threshold at the
going to be used as
AIR IN for the
the primary air source
automatic
(backup air source
backup start is
only).
240 to 320 kPa
• Verify the (34.8 to 46.4 psi,
pressure 2.4 to 3.2 bar).
threshold setting • The higher
for the backup pressure
mode is threshold at the
appropriate for AIR IN for the
the facility. automatic
• Connect the backup stop is 20
Medical Air to 100 kPa (2.9
pipeline hose to to 14.5 psi, 0.2 to
the comoressor 1.0 bar).
AIR IN
connection.
6 2065496-INC A 10 2015
Electrical safety test
.
Procedure Expected results Actual results Outcome Comments
Measure the ground The measured Pass / Fail
resistance. ground resistance is
less than 0.2 Ohms.
Measure the leakage The measured Pass / Fail
current. leakage current is
less than 300
μAmps @ 120 / 100
Vac and is less than
500 μAmps @ 220 /
240 Vac.
2065496-INC A 10 2015 7
8 2065496-INC A 10 2015