0% found this document useful (0 votes)
369 views27 pages

Service Manual For Pm-200a

Uploaded by

Mehdi Mokrani
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
369 views27 pages

Service Manual For Pm-200a

Uploaded by

Mehdi Mokrani
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 27

Patient Monitor

Service Manual
Content

Content
Chapter 1 Introduction ............................................................................................................................. 1
1.1 Overview ............................................................................................................................. 1
Chapter 2 Interface ................................................................................................................................... 2
2.1 System Structure Diagram .................................................................................................. 2
2.2 Socket Interface................................................................................................................... 2
2.3 MAIN Board Interface ........................................................................................................ 3
2.4 MAIN Board Interface ........................................................................................................ 4
2.5 Keyboard Interface .............................................................................................................. 4
2.6 Power Board Interface ........................................................................................................ 5
Chapter 3 Disassembly Guide .................................................................................................................. 6
3.1 System Overview ................................................................................................................ 6
3.2 Front Case ........................................................................................................................... 6
3.3 Rear Case ............................................................................................................................ 7
3.4 Disassembly Procedure ....................................................................................................... 9
3.4.1 Disassembly Closed Case ........................................................................................... 9
3.4.2 Disassembly Front Case .............................................................................................. 9
3.4.3 Disassembly Rear Case ............................................................................................... 9
3.5 Cable Specification ............................................................................................................. 9
3.5.1 AC Power Cable.......................................................................................................... 9
3.5.2 AC Ground Cable........................................................................................................ 9
3.5.3 DC Output Cable ....................................................................................................... 10
3.5.4 Fan Cable ......................................................................................... ……………… 10
3.5.5 Power Switch ............................................................................................................ 10
3.5.6 Battery Cable............................................................................................................. 10
3.5.7 LED Cable(LED1/LED2) ......................................................................................... 10
3.5.8 TFT Inverter Cable .................................................................................................... 11
3.5.9 TFP Cable.................................................................................................................. 11
3.5.10 Keyboard Cable......................................................................................................... 10
3.5.11 Net Cable................................................................................................................... 10
3.5.12 MAIN-FY801 Cable ................................................................................................. 11
3.5.13 Alarm Led Cable ....................................................................................................... 11
3.5.14 Encoder Cable ........................................................................................................... 11
3.5.15 Pump Valve Cable ..................................................................................................... 12
3.5.16 ECG Cable ................................................................................................................ 12
3.5.17 SpO2 Cable ............................................................................................................... 12
3.5.18 Temperature Cable .................................................................................................... 13
3.6 BOM List .......................................................................................................................... 13
Chapter 4 Troubleshooting ..................................................................................................................... 15
4.1 Troubleshooting ................................................................................................................ 15
4.2 Alarm Information ............................................................................................................ 15
Chapter 5 Test.......................................................................................................................................... 20
5.1 Test Equipment.................................................................................................................. 20
5.2 Connection and Inspiration ............................................................................................... 20
5.3 Function Test ..................................................................................................................... 20

Patient Monitor PM-200A Service Manual i


Content

5.3.1 Display Test ............................................................................................................... 20


5.3.2 Key Test ..................................................................................................................... 20
5.3.3 Sound Test ................................................................................................................. 20
5.4 ECG Test ........................................................................................................................... 20
5.5 Respiration Test ................................................................................................................. 21
5.6 Temperature Test ............................................................................................................... 21
5.7 NIBP Test .......................................................................................................................... 21
5.7.1 NIBP Leakage Test .................................................................................................... 21
5.7.2 NIBP calibrations ...................................................................................................... 21
5.8 SpO2 Test .......................................................................................................................... 21
5.9 Recorder Test..................................................................................................................... 21
5.10 Net Test.............................................................................................................................. 22
Chapter 6 Maintenance and Cleaning ................................................................................................... 23
6.1 System Checks .................................................................................................................. 23
6.1.1 Checks before using the monitor ............................................................................... 23
6.1.2 Routine check ............................................................................................................ 23
6.2 General Cleaning ............................................................................................................... 23
6.3 Cleaning Agents ................................................................................................................ 23
6.4 Sterilization ....................................................................................................................... 23

ii Patient Monitor PM-200A Service Manual


Introduction

Chapter 1 Introduction
1.1 Overview
This monitor is suitable for the adult, the pediatric and the neonate, may monitor the
life parameters, such as the electrocardiograph (ECG), the noninvasive blood pressure
(NBP), the oxygen saturation (SpO₂), the respiration rate (Resp), the body temperature
(Temp) , the invasive blood pressure (IBP), carbon dioxide (CO₂) and so on,, can in the
same screen demonstrate maximum 11 waveforms and all information of the parameters
monitored.
The purpose and function of the patient monitor is to monitor: ECG; heartrate;
noninvasive blood pressure (systolic, diastolic, and mean arterialpressures); functional
arterial oxygen saturation; respiration rate and temperature for adult and pediatric patients
in all hospital areas and hospital-type facilities. It can be used during hospital transport and
in mobile, land-based environments, such as ambulances.
Below shows the functions of this monitor:
1) The electrocardiograph (ECG), including: the heart rate, 6 channels ECG
waveforms, the ST segment analysis (can be partially optional and purchased), the
arrhythmia analysis (can be partially optional and purchased);
2) The oxygen saturation (SpO₂), including: the oxygen saturation , pulse rate, pulse
wave;
3) The noninvasive blood pressure (NBP), including: Systolic pressure, diastolic
pressure, mean pressure;
4) The body temperature (Temp): 2 channels body temperature data;
5) The respiration (Resp): Breathing rate, breath waveform.
6) invasive blood pressure (IBP): 2 channels IBP waves and Systolic pressure, diastolic
pressure, mean pressure;
7) carbon dioxide (CO₂): 1 channel CO₂wave and Breath rate, EtCO₂and FiCO₂.

Patient Monitor PM-200A Service Manual 1


Interface

Chapter 2 Interface
2.1 System Structure Diagram

2.2 Socket Interface

Figure 2-1 sensor socket


The following shows sensor socket:
T1: temperature channel 1 sensor socket;
T2: temperature channel 2 sensor socket;
SpO₂: SpO2 socket;
NIBP: NIBP cuff socket;
ECG: ECG cable socket;

2 Patient Monitor PM-200A Service Manual


Interface

IBP-1: IBP channel 2socket;


IBP-2: IBP channel 2 socket;
2.3 MAIN Board Interface

socket Function
J1 Power
J5 Net
J6 TFT inverter
J8 TFT LCD
J9 VGA
J11 Recorder
J13 To PARA board
J17 Keyboard
J20 Speaker

Patient Monitor PM-200A Service Manual 3


Interface

2.4 PARA Board Interface

socket Function
TTL To MAIN board
PUMP PUMP & Valve
VALVE
ECG ECG
SPO2 SPO2
TEMP Temperature
2.5 Keyboard

4 Patient Monitor PM-200A Service Manual


Interface

socket Function
J1 To MAIN board
J2 Alarm led
J3 Encoder

2.6 Power Board Interface

Patient Monitor PM-200A Service Manual 5


Disassembly Guide

Chapter 3 Assembly/Disassembly Guide


3.1 System Overview

3.2 Front Case Disassembly

3.2.1 TFT

6 Patient Monitor PM-200A Service Manual


Disassembly Guide

3.3 Rear Case Disassembly

3.3.1 battery

Patient Monitor PM-200A Service Manual 7


Disassembly Guide

3.3.2 battery case

3.3.3 Power board

3.3.4 Rear case

8 Patient Monitor PM-200A Service Manual


Disassembly Guide

3.4 Disassembly Procedure


3.4.1 Disassembly Closed Case
1) Use a screwdriver to remove the two screws fastening the rear case to the front case.
2) Open the front case.
3) Disconnect the keyboard cable, power switch cable, power led cable, TFT cable and
inverter cable from MAIN board and keyboard, remove front case.
3.4.2 Disassembly Front Case
1) Disconnect the keyboard cable, power switch cable and power led cable from
keyboard. Use a screwdriver remove keyboard.
2) Remove the TFT plate.
3) Remove the inverter board.
4) Remove the alarm led board.
3.4.3 Disassembly Rear Case
1) Use a screwdriver to remove the battery.
2) Disconnect the keyboard cable, inverter cable, speaker cable, net cable, power cable,
TFT cable and MAIN-PARA cable from MAIN board, use a screwdriver to remove
MAIN board & PARA board.
3) Remove the cover plate.
4) Remove pump and valves tube.
5) Remove rear case plate;
6) Remove power board;
7) Remove socket plate;
8) Remove speaker and net board;
3.5 Cable Specification
3.5.1 AC Power Cable

3.5.2 AC Ground Cable

Patient Monitor PM-200A Service Manual 9


Disassembly Guide

3.5.3 DC Output Cable

3.5.4 Power Switch

3.5.5 LED Cable(LED1/LED2)

3.5.6 Keyboard Cable

3.5.7 Net Cable

10 Patient Monitor PM-200A Service Manual


Disassembly Guide

3.5.8 MAIN-PARA Cable

3.5.9 Alarm Led Cable

3.5.10 Encoder Cable

Patient Monitor PM-200A Service Manual 11


Disassembly Guide

3.5.11 Pump Valve Cable

3.5.12 ECG Cable

3.5.13 SpO2 Cable

12 Patient Monitor PM-200A Service Manual


Disassembly Guide

3.5.14 Temperature Cable

3.6 BOM List


Material Code Material Name Specs. Qty.
ZJ00139 Front Case 1
ZJ00071 Rear Case 1
ZJ00032 TFT LCD 800*600 12.1’ LCD 1
ZJ00027 Encoder EC11 1
ZJ00167 Recorder UC50-IC 1
P060903 PARA Board PARAVA 1
M060903 MAIN Board SF16 1
ZJ00140 Keyboard KEY-12A 1
ZJ00198 Net board NET 1
ZJ00248 Alarm LED board ALARM12 1
ZJ00107 Pump RPM27C-12E 1
ZJ00106 Valve KSV05B6-12V 2
ZJ00036 Gas tube Ф2.5-Ф6 1
ZJ00111 6-way gas outlet 6-CHA 1
ZJ00030 Battery 2.2AH Battety/14.8V 1
ZJ00144 Speaker Q5 1
ZJ00058 Power Board AP156 1
ZJ00003 Power In Socket AS-01 1
ZJ00150 Ground Pole 1
ZJ00119 TFT plate SF-01 1
ZJ00159 Rear Case Plate SF-02 1
ZJ00211 Power In Plate SF-03 1
ZJ00077 Battery Plate SF-04 1
ZJ00216 Battery Cover Plate SF-05 1
ZJ00118 Temperature Plate SF-06 1
ZJ00078 Sensor Plate SF-07 1
ZJ00157 Recorder Plate SF-08 1
ZJ00103 MAIN-PARA Cable 12CM(4P-6P-2.0MM) 1
ZJ00128 Keyboard Cable 20PIN-20PIN-2.0MM 1
ZJ00059 TFT Cable 30PIN-30PIN 1
ZJ00098 ECG Cable 34CM(6P-2.5MM) 1
ZJ00099 SpO2 Cable 16CM(6P-2.0MM) 1

Patient Monitor PM-200A Service Manual 13


Disassembly Guide

ZJ00246 Temp Cable 4P-2.0 400MM 1


ZJ00101 Alarm LED Cable 48CM(3P-3P-2.0MM) 1
ZJ00163 Speaker Cable 2P-2.54 1
ZJ00163 Pump Cable 2P--2.0 1
ZJ00123 Valve Cable 4P--2.0 1
ZJ00171 Encoder Cable 6P-2.54 1
ZJ00165 Recorder Cable 6P-2.54 1
ZJ00053 ECG socket 6PIN 1
ZJ00007 Temp Socket 6.35-3 2
ZJ00074 SpO2 Socket 7 PIN 1
ZJ00168 NIBP Socket 1

14 Patient Monitor PM-200A Service Manual


Troubleshooting

Chapter 4 Troubleshooting
4.1 Troubleshooting
Problem Recommended Action
No Power up With external power supply connected, push power switch, check
the green POWER LED on the front panel, if it is not lit, replace
power cable and power board.
LCD screen is totally Ensure that the backlight cable is connected to the MAIN board.
black or white. Ensure that the LCD connector is properly connected to the
MAIN board and LCD. If problem persists, replace backlight
board or MAIN board.
LCD screen is Ensure that the LCD connector is properly connected to the
illuminated, but no data MAIN board and LCD. If problem persists, replace MAIN board.
is visible.
“COMM. STOP ERR” Ensure that the MAIN-PARA cable is properly connected to the
MAIN board and FY00801 board. If problem persists, replace
PARA board.
Key or encoder do not Replace keyboard cable or keyboard. If problem persists, replace
work MAIN board.
No sound Replace speaker cable or speaker. If problem persists, replace
MAIN board.
Recorder do not work Ensure that recorder’s power led is on. Ensure the recorder cable
is properly connected to MAIN board. If problem persists,
replace recorder or MAIN board.
ECG no signal Check ECG cable,If problem persists, replace PARA board.
NIBP can’t inflate Ensure there is pump inflation sound when NIBP start, if no
pressure up sound, replace pump cable or pump. If pump is ok, check if there
is leakage in valve, replace valve cable or valve when there is
leakage, If problem persists, replace recorder or PARA board.
SPO2 no signal Ensure SPO2 cable is ok, If problem persists after changing
SpO2 cable, replace PARA board.
Respiration no signal Check ECG cable,If problem persists, replace recorder or PARA
board.
Temperature no signal Check temperature cable and temperature cable type setting,If
problem persists, replace recorder or PARA board.
IBP no signal Ensure IBP cable is ok,If problem persists, replace PARA board.
CO2 no wavefrom Ensure CO2 cable is ok,If problem persists, replace CO2
module.
“CO2 COMM STOP” Ensure that the MAIN-CO2 cable is properly connected to the
MAIN board and CO2 module. If problem persists, replace CO2
module.
4.2 Alarm Information
The alarm Trigger condition Recommended Action
information
** ECG LEAD OFF RL or more than 2 ECG leads check the ECG lead
falls off connection
** ECG LEAD RA RA lead fall off check the ECG lead
OFF connection

Patient Monitor PM-200A Service Manual 15


Test

** ECG LEAD LA LA lead fall off check the ECG lead


OFF connection
** ECG LEAD LL LL lead fall off check the ECG lead
OFF connection
** ECG LEAD V OFF V lead fall off check the ECG lead
connection
** MODULE INIT Module self-checking mistake Restart the machine, if
ERR error still existed,
contact the factory
service
***MODULE COMM The module and the main engine Restart the machine, if
STOP communication have the problem error still existed,
contact the factory
service
** MODULE COMM The module and the main engine Restart the machine, if
ERR communication have the problem error still existed,
contact the factory
service
** PARA ALARM The parameter of the alarm limit contact the factory
LMT ERR is modified by the accident service
** RANGE EXEED The parameter observed value has contact the factory
exceed the measurement scope service
which the system can carry on
** SpO ₂ SENSOR SpO₂ sensor does not connected Check SpO ₂ sensor
OFF connection
** SpO ₂ FINGER The finger fall off from SpO ₂ Check SpO ₂ sensor
OFF sensor connect with the finger
SEARCHING SpO₂ sensor connect bad or the Check SpO ₂ sensor
PULSE... patient move the arm connection situation and
patient's current
condition
** Temp1 SENSOR The body temperature channel 1 Check temperature
OFF sensor do not connect sensor connection
** Temp2 SENSOR The body temperature channel 2 Check temperature
OFF sensor do not connect sensor connection
** WATCHDOG ERR Main engine watch-dog Restart the machine, if
self-checking defeat wrong still existed,
contact the factory
service
** SYSTEM TIME The system clock has not set Change the system time
LOST as the current time, if
error still existed, related
the factory to carry on
the service
** 12V HIGH The 12V voltage examination Restart the machine, if
exceeds the normal voltage scope error still existed,

16 Patient Monitor PM-200A Service Manual


Troubleshooting

contact the factory


service
** 12V LOW The 12V voltage examination is Restart the machine, if
lower than the normal voltage error still existed,
scope contact the factory
service
** 3.3V HIGH The 3.3V voltage examination Restart the machine, if
exceeds the normal voltage scope error still existed,
contact the factory
service
** 3.3V LOW The 3.3V voltage examination is Restart the machine, if
lower than the normal voltage error still existed,
scope contact the factory
service
**BAT HIGH The battery voltage examination Restart the machine, if
exceeds the normal voltage scope error still existed,
contact the factory
service
**BAT LOW The battery capacity is Meets the alternating
insufficient current to carry on the
charge immediately to
the battery
* NBP LOOSE CUFF Reconnects the blood
The cuff has not connected
pressure cuff
* NBP AIR LEAK check the pipe
connection situation or
The cuff has not connected good replace cuff, if the
or the air course leaks air breakdown still existed,
please contact the
factory service
* NBP DEFLATE ERR check the tube
When blood pressure connection or replace
measurement deflates has the cuff, if the error still
problem existed, please contact
the factory service
* NBP WEAK Examined the patient
SIGNAL type set whether
When blood pressure
correctly, check the tube
measurement the pulse signal too
connection or replace
weak, is unable to calculate the
cuff, if the error still
blood pressure
existed, please contact
the factory service
* NBP OUTOF When blood pressure check he tube connection
RANGE measurement the blood pressure or replace cuff, if the
or the pulse signal exceeds the error still existed, please
normal range, is unable to carry contact the factory
on the measurement service
* NBP MOVEMENT Patient arm move Check the patient

Patient Monitor PM-200A Service Manual 17


Test

situation or replace cuff,


if the error still existed,
please contact the
factory service
** NBP OVER The pressure value exceeds the check the pipe
PRESSURE measurement scope connection situation or
replace cuff, if the error
still existed, please
contact the factory
service
* NBP SATURATE When blood pressure Check the patient
measurement the pulse signal situation or replace cuff,
exceeds the normal range, is if the error still existed,
unable to carry on the please contact the
measurement factory service
* NBP PNEUMATIC The cuff has not connected good check the pipe
FAIL or the air course leaks air connection situation or
replace cuff, if the error
still existed, please
contact the factory
service
** NBP SYSTEM Blood pressure system self-check Restart the machine, if
ERR defeat the error still existed,
please contact the
factory service
** NBP TIME OUT Blood pressure measurement Restart the machine, if
overtime the error still existed,
please contact the
factory service
** NBP CUFF TYPE Patient type for adult when has Check the patient type or
WRONG used the neonate cuff replace cuff, if the error
still existed, please
contact the factory
service
** NBP MEASURE This blood pressure measurement Check the patient
FAIL has not been able to calculate the situation or replace cuff,
blood pressure if the error still existed,
please contact the
factory service
** NBP RESET ERR When blood pressure Restart the machine, if
measurement exceptionally reset the error still existed,
please contact the
factory service
**RECORDER OUT There is no paper in recorder or Put paper in recorder and
OF PAPER the recorder door is open close the recorder door
**IBP1 SENSOR OFF IBP sensor does not connected Check IBP sensor
connection
**IBP2 SENSOR OFF IBP sensor does not connected Check IBP sensor

18 Patient Monitor PM-200A Service Manual


Troubleshooting

connection
**IBP1 NEED ZERO IBP1 has not been zeroed Zero IBP channel 1
sensor
** IBP1 NEED ZERO IBP2 has not been zeroed Zero IBP channel 2
sensor
**CO₂ STANDBY CO₂ is on standby mode Set CO₂ to run mode
***CO₂ COMM STOP CO₂module and the main engine Restart the machine, if
communication have the problem error still existed,
contact the factory
service

Patient Monitor PM-200A Service Manual 19


Test

Chapter 5 Test
5.1 Test Equipment
FLUKE MPS450 or MEDISIM 300B ECG simulator
FLUKE BP Pump 2L or CuffLink NIBP simulator
FLUKE Index2 or SmartSat SpO2 simulator
Pressure calibrator or mercury manometer
Resistor box
Ruler
Network cable
Central monitor station
5.2 Connection and Inspiration
Connect the simulator, power cable and sensor cable to the patient monitor; turn on the power.
Patient monitor shall display startup picture and then enters monitor screen.
5.3 Function Test
5.3.1 Display Test
After patient monitor enter monitor screen, the characters and waves should display clearly without
dithers and blurs.
5.3.2 Key Test
Press every key on the keypad, the patient monitor shall execute the corresponding function as
instructed in the operation manual. The encoder shall execute the corresponding as well.
5.3.3 Sound Test
There should be key sound when rotate encoder; there should be alarm sound when triggers alarm.
5.4 ECG Test
5.4.1.1 ECG Calibrate
Set ECG channel 1 and channel 2’s gain to: ×1 and open ECG calibrate switch, ECG wave should
be a 1mV, 1Hz square wave, check the wave amplitude should be 10mm±0.5mm.
5.4.1.2 ECG Gain
Set ECG channel 1’s gain to: ×1, ECG channel 1 lead to II and set ECG simulator to input a 1mV
1Hz square wave, ECG1 wave should be a 1mV, 1Hz square wave, check the wave amplitude should be
10mm±0.5mm.
Set ECG channel 2’s gain to: ×1, ECG channel 2 lead to II and set ECG simulator to input a 1mV
1Hz square wave, ECG2 wave should be a 1mV, 1Hz square wave, check the wave amplitude should be
10mm±0.5mm.
5.4.1.3 ECG Lead
Set ECG simulator to input normal ECG wave, Select all ECG lead and gain, ECG wave should be
correct according to simulator.
5.4.1.4 Lead Off
Disconnect ECG lead from ECG simulator, there should be correct ECG lead off alarm.
5.4.1.5 Heart Rate
Set ECG simulator heart rate to be 30, 80, 300, the patient monitor’s heart rate should be 30, 80 and
300 respectively and error is below 1%.
5.4.1.6 Pace
Set “PACE” switch on and set simulator to input PACE signal, there should be PACE mark on ECG
wave.

20 Patient Monitor PM-200A Service Manual


Test

5.5 Respiration Test


Set the base impedance to 1K and resistance respectively to 0.5Ω and 3Ω, RESP Rate respectively
to 30 and 120, the system shall measure the RESP Rate to be 29~31 and 118~122.
5.6 Temperature Test
Set temperature sensor type to “10K” and resistor box to 30832Ω、19872Ω、8054Ω、6532Ω、
5330Ω,the temperature should be 1℃、10℃、30℃、35℃、40℃ respectively,error below ±0.1℃。
5.7 NIBP Test
1) Set NIBP simulator to adult mode and pressure to 255/195/215 mmHg, set patient monitor
to adult mode and start NIBP test, when measurement finish, systolic, mean, diastolic
pressure should be 255±8mmHg,215±8mmHg 、195±8mmHg.
2) Set NIBP simulator to neonate mode and pressure to 120/80/90 mmHg, set patient monitor
to pediatric mode and start NIBP test, when measurement finish, systolic, mean, diastolic
pressure should be 120±8mmHg,90±8mmHg 、80±8mmHg.
3) Set NIBP simulator to neonate mode and pressure to 120/80/90 mmHg, set patient monitor
to neonate mode and start NIBP test, when measurement finish, systolic, mean, diastolic
pressure should be 40±8mmHg,20±8mmHg 、10±8mmHg.
5.7.1 NIBP Leakage Test
When the cuff is connected may use “Leakage Test” function to start air course charge
process, thus to discover whether the air way’s airtight condition is good or not. If the test
passes, the system will not make any prompt; If do not passed, then in the noninvasive
blood pressure parameter area will have the corresponding wrong prompt.
5.7.2 NIBP calibrations
Using a pressure gauge (or mercury sphygmomanometer) to carry on the calibration, choose
“NIBP Calibration” in the “Service” menu to start calibration.
Connect the pressure gauge, the cuff through a 3-way tube to the blood pressure trachea jack on the
monitor, setting the monitor as “calibration” mode, then charge the cuff using a air pump,
first make the pressure to 250 mmHg, then slowly deflates, when the monitor display 200,
150 and 50 mmHg, the difference between the standard pressure gauge value and the
monitor pressure value should in 3 mmHg. If the value exceeds 3 mmHg, please contact
our company’s attendant.
5.8 SpO2 Test
Put finger to SpO2 sensor and connect SpO2 sensor to the patient monitor, SpO2 value should be
above 95%.
Connect SpO2 sensor to SpO2 simulator and set simulator to 70%、
75%、80%、85%、90%、
95%、
100%,SpO2 value on patient monitor should be 70%、75%、80%、85%、90%、95%、100%
respectively,error below 2%.
5.9 IBP Test
Connect the simulator with a IBP cable to the monitor and zero IBP, set simulator to input a 120/80
ART wave, the monitor should display IBP wave and value, error should below 2%.
5.10 CO2 Test
Connect the water trap to the socket, set CO2 work mode to “RUN” and wait for the monitor to
power up over 30 minites, input a standard 5~10% CO2 gas to the water trap, the monitor should display
CO2 wave and value, error should below 3mmHg.
5.11 Recorder Test
Press “RECORD” key, the recorder should output ECG waveform accordant to display. Take out
recorder paper, there should be “RECORDER OUT OF PAPER” alarm on patient monitor.

Patient Monitor PM-200A Service Manual 21


Test

5.12 Net Test


Connect patient monitor and central monitor station with network cable and start up patient monitor
and central monitor station, the net icon on patient monitor should be “ ”, there should be
waveform and parameter displayed on the central monitor station.

22 Patient Monitor PM-200A Service Manual


Maintenance and Clearing

Chapter 6 Maintenance and Cleaning


6.1 System Checks
6.1.1 Checks before using the monitor
1) Check if there is any mechanical damage;
2) Check if all the outer cables, inserted modules and accessories are in good condition;
3) Check if all the monitoring functions of the monitor can work normally so as to make sure
that the monitor is in good condition.
If you find any damage on the monitor, stop using the monitor on patient, and contact the
biomedical engineer of the hospital or Customer Service Department immediately.
6.1.2 Routine check
The overall check of the monitor including the functional safety check, must be performed by
qualified personnel once every 6 to 12 month or each time after fix up. All checks that need to open the
monitor enclosure must be performed by qualified service personnel.
6.2 General Cleaning
The Patient Monitor must be kept dust-free.
It is recommended that you should clean the outside surface of the monitor enclosure and the
display screen regularly. Only use non-caustic detergents such as soap and water to clean the monitor
enclosure.
1) Avoid using ammonia-based or acetone-based cleaners such as acetone.
2) Most cleaning agents must be diluted before use. Dilute the cleaning agent as per the
manufacturer's direction.
3) Do not use the grinding material, such as steel wool etc.
4) Do not let the cleaning agent enter the monitor. Do not immerse any part of the system into
liquid.
5) Do not leave the cleaning agents at any part of the equipment.
6.3 Cleaning Agents
The following solutions can be used as detergent:
1) Diluted Ammonia Water
2) Diluted Sodium Hyoichlo (Bleaching agent)
3) Diluted Formaldehyde 35% -- 37%
4) Hydrogen Peroxide 3%
5) Alcohol
6) Isopropanol
6.4 Sterilization
To avoid extended damage to the equipment, sterilization is only recommended when stipulated as
necessary in the Hospital Maintenance Schedule. Sterilization facilities must be cleaned first.
Recommended sterilization materials: Ethylate, and Acetaldehyde.

Patient Monitor PM-200A Service Manual 23

You might also like