Smart2Pure Water Purification System :-
1) The Devices Types
A. Smart2Pure Standard
B. Smart2Pure UV
C. Smart2Pure UF
D. Smart2Pure UV/UF
2) Water Types and Applications
A. Type 1 Water → Cell and tissue culture
→ PCR, DNA sequencing
→ Electrophoresis, TOC Measurements, IC
→ HPLC, GC-MS, ICP-MS, AA
B. Type 2 Water → Rinsing lab glassware
→ Supplying general lab equipment
→ Preparing and diluting buffers, reagents, and media
Smart2Pure UV\UF 12L/h
Type 1 Water
» Resistivity at 25°C, MΩ.cm = 18.2
» Conductivity, µS/cm = 0.055
» TOC value , ppb = 1 – 5
» Bacterial content in CFU/ml: <0.01
» Particles, µm/mL: <1
» Endotoxin, EU/ml: <0.001
» floow Rate, L/min = 0.6
Type 2 Water
» Applications: Rinsing lab glassware, supplying autoclaves and ultrapure
water systems, preparing and diluting buffers, reagents, and tissue culture
media.
» Pure water production at 15°C, L/h= 12
» Resistivity at 25°C, MΩ.cm = 15-10
» Conductivity, µS/cm = 0.067-0.1
Storage Tank 30L / 60L
» Polyethylene reservoir is opaque to light and has large opening for easy
and effective cleaning
» Sterile venting filter/reservoir overflow protects against contamination by
microorganisms
» CO2 adsorber prevents airborne contamination from entering the tank
» Recirculation pump protects the high purity water from bacterial growth
during standstills and maintains low conductivity value
» Conical bottom outlet allows for complete draining and efficient cleaning
and disinfection
*Extra Information about Smart2Pure
Product Dimensions HxWxD, mm: 545x305x300
Product Weight kg w/o water: 20
Required feed water quality:
1. Source Tap: water
2. TDS: >500 ppm
3. Turbidity: > 1 N.T.U
4. pH Range: 4 – 11
5. Temperature: °C (2-35)
6. Pressure, psi (bar): 1.4 - 87 (1 - 6)
*Replacement Consumables:
Consumable Estimated Life Time How You Will Know To
Change
Pretreatment Cartridge (RO Up to 12 months Ultrapure cartridge
Membrane) exhausting quickly
Ultrapure Cartridge Up to 12 months •Purity drops below acceptable
limits
•“Limit value Cond.” alarm and
LED on
0.2 μm Filter For System And Up to 12 months Dispensing water flow is
Hand Remote Dispenser reduced
•Ultrapure cartridge is replaced
•Bacteria is present in product
water
Ultrafilter When applicable – up to 2 years •Dispensing water flow is
reduced
•Replace every 2 years
UV Lamp For Type I System When applicable – up to 2 years UV lamp is not lit & Replace
every 2 years
CO2 Vent Filter Up to 12 months Turns from white to blue
UV Light For Storage Reservoir – When applicable – up to 2 years Replace every 2 years
230 V, 50 Hz only
Note: The actual life time of the consumables depend on the quality of the feed
water and the amount of water used by the system.
(When to change the consumables System monitors hours of operation to alert
customer when consumables need to be replaced. Set point limits also monitor
purity).