PH700 en
PH700 en
Instruction Manual
This instrument is an outstanding combination of advanced electronics design and sensor technology. It’s the
most economical choice of a reliable lab-grade benchtop pH meter. Please read this maual carefully to
properly use and maintain the meter. Apera Instruments reserves the right to update the contents of this
manual without giving prior notices.
2 Technical Specifications
Range 0 to 14.00 pH
Resolution 0.01/0.1 pH
Accuracy ±0.01 pH ±1 digit
pH
Temperature
0 to 100 ˚C (Auto or Manual)
Compensation
Calibration 1 to 3 points automatic calibration
Range ±1999 mV
mV Resolution 1 mV
Accuracy ±0.1% F.S ±1 digit
Range 0 to 100˚C; 32 to 212°F
-1-
3 What’s in the Box?
Content Quantity
3.1 PH700 Benchtop pH Meter 1
3.2 201T-F Plastic 3-in-1 Combination pH Electrode 1
3.3 pH standard buffer solutions (4.00/7.00/10.01 pH) 1 for each 50ml
3.4 9V Power Adapter 1
3.5 Electrode Holder 1 set
3.6 Instruction Manual 1
4 Instrument Description
Figure-1
(6) — The pH value displayed during calibration, numberings for storage and recall, and reminder icons
(7) — Icons for data storage and recall — M+: store measuring data; RM: Recall measuring data
Figure-2
3.2.1. Keypad operations
Short press ----- <1.5 seconds; Long press ----- >1.5 seconds.
-3-
4.3 Connectors
Figure - 3
-4-
Clear stored data
The meter can store 50 groups of data. Once the storage is full, if user
presses again, the FUL icon will be displayed on LCD (see Figure-6).
The storage needs to be cleared in parameter setting P4 so as to keep storing
data. The numberings in M+ indicates the total number of stored data; The
numberings in RM indicates the number of stored data in current mode
(ph/mV). For example, if 20 groups of data is stored in pH mode, and 10 groups Figure - 6
in mV mode, then the numberings for M+ will be “M+30” in both pH and mV mode; the numberings for RM in
pH mode would be “RM20”, in mV mode “RM10”.
5 pH Measurement
5.1 Preparation
Install the electrode holder
The electrode holder is composed of a metal stand base, a supporting pole, and an electrode clip. First, install
the metal stand base to the meter (screw it on). Please note that the metal base can be installed on either side
of the meter. Then install the supporting bar on the metal stand base, and then install the electrode clip on the
supporting bar.
-5-
Technical Specifications of the 201T-F pH Electrode
Measurement Range: 0 - 14 pH, 0 - 80˚C (32 – 176˚F)
Junction: Single Ceramic
Reference Electrode: Ag/AgCl
Connector: BNC/RCA
Size: ø12*160 mm
Temperature unit: 30K Thermistor
Electrode Connectors
The pH electrode has two connectors: the BNC connector connects the pH probe; the RCA connector connects
the temperature sensor. Plug these two connectors into ‘pH/mV” and “TEMP” sockets. Please note not to pull
the cables in case of poor contact. Please keep the connectors clean and dry. Refer to section 4.7 regarding
how to properly maintain the pH electrode.
Table-2
3-Point Calibration
The meter can adopt 1 to 3 points of calibration. The 1st point must be using 7.00 pH (or 6.86 pH) buffer
solution, and then choose other buffers to do 2nd point and 3rd point. Please refer to Table-3 for details. In
the process of calibration, the slope of acidity range and alkalinity range will be displayed, which indicates
the health of your probe. The closer the slope is to 100%, the better the condition of the probe is.
-6-
Table-3 Calibration Mode
Calibration
USA NIS When to adopt
icon
1-Point
7.00 pH 6.86 pH accuracy≤ ±0.1 pH
Calibration
-7-
5.4 pH Meter Calibration
1) Press to enter calibration mode. CAL1 icon will flash in the upper
right corner of the LCD. 7.00 pH will flash in the lower right corner of the
LCD, reminding you to use pH 7.00 buffer to conduct 1st point of
calibration.
2) Use distilled water to rinse off electrode and then dry it (with a tissue).
Dip it into pH 7.00 buffer solution, stir gently and let it stand still and wait
for the reading to become stable. In the lower right corner of LCD, the
process of auto recognizing the buffer solution will be displayed. Pressing
before the buffer is recognized will generate Er2 (please refer to
table 5).
3) When the meter locks 7.00 pH in the lower right corner, stable
icon displays on LCD. Then press key to calibrate the meter.
End icon appears after calibration is done.
The 1st point calibration is finished. In the meanwhile, CAL2 will flash at
the upper right corner, and 4.00 pH & 10.01 pH will flash alternately at the
bottom right, indicating using pH4.00 or pH10.01 buffer solution
to make the 2nd point calibration.
4) Take out the pH electrode, rinse it in distilled water, dry it, and dip
it into pH 4.00 buffer solution. Stir the solution gently and let stand still in
the buffer solution until a stable reading is reached. The meter’s display
will show the recognition process of calibration buffer solution at the lower
right corner. When the meter recognizes 4.00 pH, stable icon
displays on LCD. Press key to calibrate the meter. End icon and
electrode slope of acidity range will be displayed after calibration is done.
In the meanwhile, CL3 will flash at the upper right corner of the LCD, and
10.01 pH will flash at the lower right, indicating using pH10.01 buffer
solution to make the 3rd point calibration.
5) Take out pH electrode, rinse it in distilled water, dry it, and dip
it into pH 10.01 buffer solution. Stir the solution gently and let it stand still
in the
Figure-7
buffer solution until a stable reading is reached. The meter’s display will show
recognition process of calibration buffer solution at the bottom right of
LCD. When the meter recognizes 10.01 pH, stable icon displays on LCD. Press
key to calibrate the meter. End icon and electrode slope of alkalinity range
display after calibration is done. The meter returns to the measurement mode, the calibration icons L M
H will be displayed. Please see Figure-7 for the above calibration process.
-8-
6) During the calibration process, press key to exit from the calibration mode. The meter can
perform one-point, two-point and three-point calibration. Calibration guide icons appear on LCD.
5.5 Measuring pH
Screw off the KCL storage bottle, and put it aside (do not dump or spill the KCL solution). Prepare a cup of
distilled water, and rinse the electrode in it for a few seconds. Gently shake the probe to remove excess water,
and dry it with clean tissue paper or Kimwipes (DO NOT rub or wipe the probe, just use paper to dap off excess
water). Gently stir the electrode for a few seconds after it’s dipped into the test solution and then let it stand
still. Wait for the stable measurement (a smiley face icon appears and stays) and then record the
readings. Please refer to Figure-8 for calibration and measurement process of the pH meter.
Notes:
1) This default 201T-F electrode will NOT give accurate and stable pH readings when testing
purified water such as distilled or deionized water. This because those water do not have
enough ions present for the electrode to function properly. To measure distilled or deionized
water’s pH, users need to use a specialized pH electrode with the meter such as LabSenTM
803 Pure Water pH Electrode. Contact us at 1-614-285-3080 for more details. When testing
purified water like spring water or drinking water, it will take longer for the readings to get
stabilized (typically 3-5 minutes) because there is very few ions left to be detected by the
sensor in those purified water.
2) Purified water such as distilled water and deionized water are recommended for only rinsing
the probe for the best result.
3) The 201T-F pH Electrode is NOT suitable for testing high-temperature samples (>176˚F).
Testing high-temperature samples could cause permanent damage to the electrode. A
specialized electrode such as LabSenTM 213 (up to 225˚F) should be used in this situation.
4) The electrode housing is polycarbonate. When using cleaning solutions, take cautions on carbon
tetrachloride, trichlorethylene, tetrahydrofuran, acetone, etc., which will dissolve the housing and
invalidate the electrode.
-9-
5.5.1 According to the pH isothermal measurement principle, the closer the test sample’s temperature is to
the calibration solution’s, the higher the accuracy of the measurement. This principle is recommended to
follow when conducting tests for the best result. Example: If users were to test samples at 150˚F, we
recommend warming up the calibration solutions to the same temperature before performing calibrations in
order to get the most accurate readings.
5.5.2 The instrument has a function to return to factory default setting, which can be set up in P5 (refer to
section 5.6). Returning to factory default setting is to restore the meter to theoretical value (zero potential pH
is 7.00, slope is 100%), and set all the parameters to default settings. When the meter’s calibration or
measurement is performing abnormally, users can use this function to let the meter return to factory default
mode, and then conduct calibration and test again. Please note that this function is irreversible once used.
5.5.3 Self-Diagnosis
In the process of calibration and measurement, the meter has a self-diagnosis function. Please see table 5.
* If you find any air bubble in the glass bulb of the pH sensor, simply shake the probe for a few times
to remove it. The existence of an air bubble in the glass bulb will significantly decrease the stableness
of measurement.
- 10 -
5.6 Parameter Setting
Table-6
- 11 -
5.7 pH Electrode’s Maintenance
5.7.1 Daily maintenance
The soaking solution contained in the supplied protective bottle is used to maintain activation in the glass bulb
and junction. Loosen the cap, take out the electrode and rinse in distilled or water before taking a measurement.
Insert the electrode and tighten the cap after measurements to prevent the solution from leaking. If the soaking
solution is cloudy or moldy, replace the 3M KCL solution.
* We DO NOT recommend using other brand’s soaking solution, since different chemicals can be used
and potential damage can be caused to the electrode)
* The electrode should NOT be soaked and stored in purified or distilled water, protein solution, acid fluoride
solution, or organic lipids. Distilled water or purified water is only recommended for rinsing electrodes.
5.7.5 Clean contaminated glass bulb and junction (please refer to Table-8)
- 12 -
6 mV measurement
- 13 -
7 Recommended pH Electrodes for Specific Applications
8 Warranty
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace
free of charge, at option of APERA INSTRUMENTS(Europe) GmbH, any malfunctioned or damaged product
attributable to responsibility of APERA INSTRUMENTS(Europe) GmbH for a period of TWO YEARS (SIX
MONTHS for the probe) from the delivery.
This limited warranty does not cover any damages due to:
Transportation, storage, improper use, failure to follow the product instructions or to perform any preventive
maintenance, modifications, combination or use with any products, materials, processes, systems or other
matter not provided or authorized in writing by us, unauthorized repair, normal wear and tear, or external
causes such as accidents, abuse, or other actions or events beyond our reasonable control.