6 ATEX
Zone 1 8
Zone 0
P
2
T
II 1/2G - Ex op is /[op is T6 Ga] db IIC T6 Ga/Gb
II 1/2D - Ex op is /[op is T86° C Da] tb IIIC T86° C Da/Db
3
5 4 DN50 / ANSI 2"
1
Zone 2
Figure 1 Ex setup
1 GPro 500
2 2 3 4…20 mA (pressure and temperature)
3 Junction Box (Ex-e)
4 Ethernet
5 External power supply
6 Purge box for Zone 1 (optional)
7 M400
8 For detailed cross section view – see “The GPro 500 Interface between Zone 0 and Zone 1”
© 11/2022 METTLER TOLEDO 30 688 448 C 9
1
3
2
TDLS GPro 500
4
5
Figure 2 The GPro 500 Interface between Zone 0 and Zone 1
1 Zone 1 region
2 Process window
3 Check valve
4 Zone 0 region
5 Interface sensor head – probe
The process window and the check valve make sure that Zone 0 and Zone 1 are physically separated.
The sensor head is always in Zone 1 and the probe is in Zone 0.
Non-metallic boundary wall Sensor Head
– Material of non-metallic boundary wall: Fused silica glass C 7980
– Temperature range of the non-metallic boundary wall: – 20 – 55 °C
– Maximum pressure of non-metallic boundary wall: 0.5 barg
CAUTION
For intended installation in an Ex classified area,
please observe the following guidelines (ATEX 2014 / 34 / EU1)).
(1)For UK statutory Requirements SI 2016 No. 1107)
Ex classification: Ex II 1/2G - Ex op is/[op is T6 Ga] db IIC T6 Ga/Gb
and
Ex II 1/2D - Ex op is/[op is T86° C Da] tb IIIC T80° C Da/Db
Designation and number of the declaration: SEV 15 ATEX 0131X
IECEx SEV 15.0013X
CML 22 UKEX 2213X
10 30 688 448 C © 11 / 2022 METTLER TOLEDO
WARNING
Repairs of the flameproof joints must be made in compliance with the
constructive specifications provided by the manufacturer.
Repairs must not be made on the basis of values specified in
tables 1 and 2 of IEC 60079-1 Ed. 7.0.
WARNING
In the normal configuration, the temperature at the interface 5 between the
sensor head and the probe may not exceed 55 °C. If the temperature ex-
ceeds 55 °C at the interface to the sensor head temperature class T6 (85 °C)
is no longer valid and the ATEX classification is violated.
WARNING
If the temperature at the interface 5 between the sensor head and the probe
exceeds 55 °C, the Thermal Barrier – see Appendix 2, chapter 2.3
“Accessories” – has to be used in a way that the temperature at the interface
to the sensor head never exceeds 55 °C. If the temperature exceeds 55 °C at
the interface to the sensor head temperature class T6 (85 °C) is no longer
valid and the ATEX classification is violated.
WARNING
The metallic enclosure of the TDL sensor has to be connected by con-
ductive wiring with the grounding system of the plant.
© 11 / 2022 METTLER TOLEDO 30 688 448 C 11
TDLS GPro 500
Figure 3 Label.
1 Product name
2 Manufacturer
3 Country of origin
4 Gas to be measured
5 Product key
6 Part no.
7 Serial no.
8 Ambient temperature limits
9 ATEX markings
10 Power rating
11 Enclosure ratings
12 SIL Mark
13 FW/HW version
1/2" NPT
WARNING -
DO NOT
OPEN
THE DEVICE
Figure 4 Note label.
M6x12
Figure 5 Grounding label.
12 30 688 448 C © 11 / 2022 METTLER TOLEDO