BS-100 Dust Monitor User Manual
BS-100 Dust Monitor User Manual
User Manual
V1.1
1.1 Overview
Welcome to choose Dust Monitoring Device. This user manual describes the
installation, operation, inspection and maintenance of device. It is deigned based on
the backscattering principle of the dust particles for the on-line continuous
measurement of particulate pollutants from stationary pollution source.
Caution:
Uses a 5mw, 650nm semiconductor laser, whose laser beam and the reflected
light directly into the eye will cause serious damage.
Do not carry out power-on operation before installing on the stack/ flue.
Do not enter the stack/flue while the device is running. Staring at the laser beam
and its reflected light shall be prohibited.
Device can be used in the real time continuous measurement of the particulate
pollutant concentrations for various emission sources. It can be facilitated with flue
gas monitoring system, used separately or connected into a dust monitoring network
with several such device sharing a front platform. The device is applicable for dust
monitoring of power plants, steel mills, cement plants, etc., and for the process control
of dust collecting equipment and other powder engineering.
1
2 Principle, Characteristics and Technical Specifications
L
D
RT
PD1
PD0
SHEILD
Preceding
Dirver Pretreatment
Modulation Demodulation
Opto-coupler
CPU
isolation
4-20mA
HMI display module
output
Laser backward scattering principle can prevent the laser beam from swinging
when the refractive index is uneven during the mechanical vibration at uneven
temperature.
2
User Manual of Dust Monitoring Device
Size\Weight 158mm×158mm×273mm\3kg
Temperature: -40℃~65℃
Environment requirements
Relative humidity: 0-100% RH
Measuring error ±10%
3
User Manual of Dust Monitoring Device
3.1 Preparations
5
User Manual of Dust Monitoring Device
6
User Manual of Dust Monitoring Device
5 2
4
3
Concentration
Concentration
Set optical path
Concentration
The concept of optical path: optical path = stack inner diameter+ wall thickness+
0.2m, the largest optical path is 2.5 m. Check if device optical path and site flue size
conform to requirements before installation, if not, please contact technical personnel
to solve.
The system is configured with a calibrator test performance of device, not to calibrate
zero point and full span calibration. This device enters the "Z value" and "Full value"
in the calibration interface in cooperation with the remote control in automatic
confirm the enter key for automatic calibration. Don't need to disassemble the rain
cover to accurate calibration fast. There is no effect on online measurement of the
8
User Manual of Dust Monitoring Device
device.
Use the remote control to carry out the system calibration by completing the
following steps:
Under the main interface, press “ ” key on the remote control to enter the
maintenance interface;
Under the maintenance interface, use the “ ”and“ ” keys to select the
“Calibration” menu;
Press “ ” key to enter the password. Then press “ ”and“ ” each for twice in
turn;
Enter the calibration interface, you may select "Z value" and "Full value" and set
zeroing and span calibration. Fine-tune zero point and full scale point with the
guidance of professional and technical personnel .
Press “ ” key to enter the main interface and wait for the reading to be stable.
Fig. 3.5 maintenance interface Fig. 3.6 Full value Fig. 3.7 Adjust comparison coef.
Before shipment, dust monitor has been adjusted in the dynamic range basing on
the users’ feedback. However, owing to the relationship between the light scattering
and the optical characteristics of the dust, its concentration corresponds to the same
signal output for a specific emission source, but different emission sources with the
same emission concentration may have different output signal. This may result in the
factory setup differing from the site conditions of the users. Too low output signals
lose the sensitivity of device, too high output signals exceed the dynamic range after
installation.
9
User Manual of Dust Monitoring Device
If installed at the back of the emission source, with the reading has been 5% lower
than the measuring range, please increase the signal gains before reference testing; if
installed at the back of the emission source with the reading exceeding the full span,
please decrease the signal gains before reference testing.
Use the remote control to adjust the signal gains by completing the following steps:
From the maintenance interface, select “Set ”menu; press “ ” to enter the
After entering the “Set” interface, please select “Gain” and press “ ”to
confirm;
Note:
Unless the dust concentration is especially low, do not select “X8” as the gain level.
The factory default gains of the dust monitor is “X2”, which does not need to be
adjusted if not necessary.
Set Set
Gain Range Mean Gain
x1 x2 x4 x8
R
Enter
Fig. 3.8 Select and setup the gain Fig. 3.9 Select the gain
10
4 Range Adjustment
The general device has terms such as zero point, range and linearity.
Some concepts of backward scattering method is quite different with general.
First, scattered light created by the laser beam project on the particles is directly in
proportion to the scattering area of the particles (this is the single-scattering case), and
the scattering area of the particle mass under a specific range of particle size is in
directly portion to the concentration of the particle mass.
The zero point in the scattering method indicates there is no scattered light. The
linearity of the scattering refers to the linear relation between the output current and
the scattering intensity, rather than with a linear relationship with the particulate
concentration. Because the later is related with the emitting source and the regression
model, while the former is actually disclosing the actual optical/electrical
characteristics of the device.
On the basis of the above, the factory ranges is not accurate. An accurate one should
be determined through reference. Please refer to the corresponding sections for the
setup of reference and range.
The site calibration (weighing comparison). Install device in the flue, and set the
appropriate range, but cannot ensure the output value (X) is equal to the real value(Y).
While need calibrate through the reference experiment, namely weighing comparison,
to determine B and K values of the equation Y = KX + B to adjust the output value,
make it close to the true output values. General comparison duration is once for a
quarter. Shorten the cycle according to actual situation.
Select “Set” from the maintenance interface, and press “ ” key to enter the
Enter the set screen, then press “ ”and“ ” to select measuring range.
11
User Manual of Dust Monitoring Device
Set Set
Gain Range Mean Range 0-500mg/m3
R Enter
12
5 System fault diagnosis and maintenance
5.1 Maintenance
Essence of the failure: The device might have been damaged. Handling method:
Return it to the factory for repairing.
Instructions for failure diagnosis:
(1) The laser does not illuminate;
13
User Manual of Dust Monitoring Device
(2) The output remains the same when the calibration module is used to calibrate the
zero point and range; the device output remains the same with or without a piece of
white paper placed 500mm in front of the device;
(3) The 4-20ma output remains the same.
Essence of the failure: The device setting is incorrect. Handling method: Consult with
the factory and handle the problem on site.
Instructions for failure diagnosis:
(1) The displayed concentration is 999.9 mg/m3;
(2) The range is exceeded;
(3) The displayed concentration is 0 or very low;
(4) There is no flue gas coming out of the flue but the device output indicates high
dust concentration; but the zero point is accurate when a calibration module is used
for calibration;
(5) The laser does not illuminate.
Instructions for failure handling:
(1) The device optical path does not match the field conditions on site;
(2) The device optical path does not match the field conditions or the actual smoke
dust concentration is very high;
(3) The laser is zeroing calibrated or gain altered (increased);
(4) The device optical path does not match the field conditions on site;
(5) The laser is switched off in the device self-inspecting menu.
Essence of failure: Incorrect installation and wiring. Handling method: Consult with
the factory
Instructions for failure diagnosis:
(1) The device does not work well due to incorrect wiring;
(2) Dust easily accumulates on the lens.
Instructions for failure handling:
14
User Manual of Dust Monitoring Device
15
6 Installation instruction
(1) The standard dust detector accessories are configured for onsite working condition
with negative pressure. If the on-site working condition is positive pressure, then the
dust analyzer needs an air blower to purge. The wind pressure of air pleasure should
be slightly bigger than the pressure in the stack. Otherwise, after running a while, it
will pollute the lens which is inside of the equipment base and influence the measure
accuracy. If sometimes is positive pressure and sometimes is negative pressure, it also
strongly recommends to install an air blower to purge, which could reduce the
maintenance work of optical path.
(2) The location of installation point should select a place with uniform dust
concentration, steady flow rate and not at corner as well. The instrument should install
vertically to the flue pipe and stack wall.
(3) The instrument wiring instruction: there is wiring connection identification
number on the power interface. 1 connects to +24V, 2 connect to 4-20mA signal line,
3 connect to GND, 4 connect to 485A, 5 connect to 485B.
(4) Asbestos gasket, used for buffer and thermal insulation, install between flange and
instrument. If the chimney temperature is above 300℃, it is needed to add more
quantity and more thick asbestos gasket for thermal insulation. The highest
temperature the instrument can burden 65℃.
(5) The function of measuring module: the measuring module, as spare parts for
maintenance, is used to calibrate 4-20mA and test the performance of instrument. If
the difference of LV value on two side of measuring module is larger than 100, it
prove the working operation is good.
(6) The function of remote controller: Easy maintenance and commissioning.
(7) The reserved sampling port should be convenient to see the laser beam from the
instrument to avoid the laser beam touch the flange and mounting wall.
16
7 Commissioning
If newly installed dust analyzer shows 0 during furnace shutdown and displays data
when opening furnace, it is considered that this analyzer is in good working and no
need of any commissioning operation.
If the instrument occurs below problems when opening furnace, please kindly check
the below suggestion.
In general, LV value is quite low which is about 10. Below ways is sequenced but not
separate. If the first method is not workable and please try the second way as follow.
If the first method is workable, it is unnecessary to try the second method.
Method one: Zeroing by covered the laser device.
Unscrewed the black shell, the laser device can be seen in the middle. Then use the
barrier to cover the round hole front of the laser device. In this condition, Press “S” on
the remote control and then enter into calibration interface to zeroing. Once press
“OK”, the zeroing operation is completed. (This step does not need of calibration
module and span calibration.)
Method two: Increase the gain of software or electrical bit of hardware.
In the main interface of dust analyzer, press “S” on the remote control to enter into
maintenance interface, selecting into gain setting which can be used to increase the
gain.
There are 2 blue electrical bits under the second circuit board under the instrument
display. The left is gain electrical bit which shows “gain” on the silkiness and shift
again crosswise to increase the hardware gain.
Method three: increase the k value.
In the main interface of dust analyzer, press “S” on the remote control to enter into
maintenance interface, selecting into calibration option to set “K” valve. It can be
realized by adding K value.
Method four: By increasing optical path to expand the measuring range.
Under the LV value interface to adjust angle of the laser device, uses M4 hex wrench
to adjust two screws at backward of laser device. One screw is to adjust the angle of
17
User Manual of Dust Monitoring Device
laser, and the other has fixation ability. The higher angle will increase the optical path
and LV value. Once the LV value reached 30-50, to tighten the screw which has flat
wisher, it is fine to fixed optical path at this point. In this condition, Press “S” on the
remote control and then enter into calibration interface to zeroing. Once press “OK”,
the zeroing operation is completed. (This step does not need of calibration module and
span calibration.)
The long screw (left) is used to adjust the angle of laser device, and short screw (right)
has the fixation ability.
If it is a new analyzer, please confirm whether the flange hole is missed or not, which
leads to the laser shot on the installation wall. The laser does not enter into flue pipe
so that it causes the full range.
If it is already installed for a period of time, the first task it to check the optical lens is
polluted and the flange end is blocked or not. At last please check that optical path
matches the chimney size. The optical length calculating method as follow: inner
diameter of flue pipe and plus thickness of wall. The largest optical path is 2,5 meter.
When there is large difference between theoretical optical path and instrument optical
path, it can be adjusted by following methods. Please noted that the operation should
be under guidance of technical engineer.
Method: Shorten the optical path and increase the optical path.
18
User Manual of Dust Monitoring Device
Under the LV value interface to adjust angle of the laser device, uses M4 hex wrench
to adjust two screws at backward of laser device. One screw is to adjust the angle of
laser, and the other has fixation ability. The higher angle will increase the optical path
and LV value. Once the LV value reached 30-50, to tighten the screw which has flat
wisher, it is fine to fixed optical path at this point. In this condition, Press “S” on the
remote control and then enter into calibration interface to zeroing. Once press “OK”,
the zeroing operation is completed. (This step does not need of calibration module and
span calibration.)
19
8 Warranty Policy
The device has been manufactured and designed based on the latest national standard,
and meets the requirements of ISO9001 quality management system.
The defective products must be returned to us, attached with a detailed problem
description. The company reserves the right to charge the field maintenance fees if no
failure and only improper use is discovered. The company assumes no responsibilities
for any loss or damage caused directly or indirectly by improper use or operation by
the buyer or any other party.
The manual relate only to device and components sold to the buyer, authorized
dealers, distributors and representatives. The warranty period specified in the manual
will not be extended due to any handling work during the service life.
20