BC5130 Manual Book
BC5130 Manual Book
Operator’s Manual
© 2015-2016 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.
For this Operator’s Manual, the issue date is 2016-10.
derivative work of this manual in any manner whatsoever without the written permission of
informational or editorial purposes. They are the property of their respective owners.
All information contained in this manual is believed to be correct. Mindray shall not be liable
for errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this manual.
Mindray is responsible for the effects on safety, reliability and performance of this product,
only if:
all installation operations, expansions, changes, modifications and repairs of this product
are conducted by Mindray authorized personnel;
the electrical installation of the relevant room complies with the applicable national and
local requirements; and the product is used in accordance with the instructions for use.
I
WARNING
It is important for the hospital or organization that employs this equipment
to carry out a reasonable service/maintenance plan. Neglect of this may
result in machine breakdown or personal injury.
Be sure to operate the analyzer under the situation specified in this manual;
otherwise, the analyzer will not work normally and the analysis results will
be unreliable, which would damage the analyzer components and cause
personal injury.
NOTE
This equipment must be operated by skilled/trained clinical professionals.
II
Warranty
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or
other charges or liability for direct, indirect or consequential damages or delay resulting from
the improper use or application of the product or the use of parts or accessories not approved
by Mindray or repairs by people other than Mindray authorized personnel.
Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Address: Mindray Building,Keji 12th Road South,High-tech industrial
park,Nanshan,Shenzhen 518057,P.R.China
Website: www.mindray.com
E-mail Address: [email protected]
Tel: +86 755 81888998
Tel: 0049-40-2513175
Fax: 0049-40-255726
III
Table of Contents
Table of Contents ....................................................................................................................... 1
1
Table of Contents
2
Table of Contents
3
1 Using This Manual
1.1. Introduction
This chapter explains how to use your BC-5130 Operator’s Manual, which is shipped with
your BC-5130 AUTO HEMATOLOGY ANALYZER (hereafter referred as “the analyzer”) and
contains reference information about the analyzer and procedures for operating,
troubleshooting and maintaining the analyzer. Read this manual carefully before operating
your analyzer and operate your analyzer strictly as instructed in this manual.
1-1
Using This Manual
1-2
Using This Manual
1-3
Using This Manual
Format Indication
[××] all capital letters enclosed in [ ] indicate a key name on the
analyzer or external keyboard, such as [ENTER]
"××" bold letters included in " " indicate text you may find on the
screen of BC-5130, such as “Clear”
×× bold letters indicate chapter titles, such as Chapter 1 Using
This Manual.
All illustrations in this manual are provided as examples only. They may not necessarily
reflect setup of the BC-5130 or data displayed.
1-4
Using This Manual
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
WARNING
Please check the firmness of all the doors and covers before running the
analyzer.
Make sure all the safety measurements are adopted. Disable any safety
device or sensor is prohibited.
1-5
Using This Manual
The reagents are irritating to eyes, skin and airway. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them and the contacted areas in the laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
Keep your clothes, hairs and hands away from the moving parts to avoid
injury.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
CAUTION
Please use the analyzer strictly according to this manual.
NOTE
Use the reagents specified by the manufacturer only. Store and use the
reagents as instructed by instructions for use of the reagents.
Check if the reagent tubes are properly connected before using the analyzer.
1-6
Using This Manual
1.6. Symbols
You will find the following symbols in this manual:
CAUTION
Make sure the symbols are in good condition during daily use and maintenance.
CAUTION,
CONSULT ACCOMPANYING
DOCUMENTS.
BIOLOGICAL RISK
USB PORT
NETWORK PORT
1-7
Using This Manual
ALTERNATING CURRENT
SERIAL NUMBER
DATE OF MANUFACTURE
MANUFACTURER
TEMPERATURE LIMITATION
1-8
Using This Manual
1-9
Using This Manual
(1)
To avoid electric shock, disconnect power cord prior to removing or replacing fuse.
(2)
Warning, potential biological risk.
1-10
Using This Manual
(1)
Warning, potential biohazardous risk.
(2)
The sample probe is sharp and potentially biohazardous. Exercise caution to avoid contact
with the probe when working around it.
1-11
Using This Manual
(1)
Do not put hands under the syringe or in the guide slot when the analyzer is running.
1-12
Using This Manual
(1)
Do not put hands under the syringe or in the guide slot when the analyzer is running.
1-13
Using This Manual
(1)
Caution: Class 3B Laser radiation when open and interlocks defeated avoid exposure to the
beam
1-14
2 Understanding the Analyzer
2.1. Introduction
This chapter introduces the parameters, major components, interfaces, buttons, menus,
software help system, operation information and reagent system of the BC-5130 AUTO
HEMATOLOGY ANALYZER.
2-1
Understanding the Analyzer
2.2. Parameters
NOTE
The purpose of this analyzer is to identify the normal patient, with all normal
system-generated parameters, and to flag or identify patient results that
require additional studies.
Parameter
Name Abbreviation CBC CBC + DIFF
Group
White Blood Cell count WBC √ √
Basophils number Bas# / √
Basophils percentage Bas% / √
Neutrophils number Neu# / √
Neutrophils percentage Neu% / √
WBC group (11)
/
number
Abnormal Lymphocytes ALY% / √
percentage
Large Immature Cells LIC# / √
number
Large Immature Cells LIC% / √
percentage
Platelet clump number PLT Clumps# / √
Platelet clump PLT Clumps% / √
percentage
Lipid granule number Lip# / √
Lipid granule percentage Lip% / √
2-2
Understanding the Analyzer
Width
Plateletcrit PCT √ √
Platelet larger cell ratio P-LCR √ √
Platelet larger cell count P-LCC √ √
Histograms
Scattergram
2-3
Understanding the Analyzer
NOTE
"√"means "available under the mode","/"means "not available under the
mode".
ALY%, LIC%, PLT Clumps%, Lip%, NRBC% ,ALY#,LIC#, PLT Clumps#, Lip#,
and NRBC# are for research use only.
2-4
Understanding the Analyzer
WARNING
Please check the firmness of all the doors and covers before running the
analyzer, and make sure they do not get loose when the analyzer is running.
2-5
Understanding the Analyzer
2-6
Understanding the Analyzer
2-7
Understanding the Analyzer
Power switch
Power switch is located on the back of the analyzer.
CAUTION
Do not turn on/off the switch repeatedly in a short time to avoid damaging
the analyzer.
[Aspirate] key
The [Aspirate] key is on the front of the analyzer. Press the [Aspirate] key to start analysis,
dispense diluent or exit from standby mode.
USB/network ports
The USB ports and network ports are on the left side of the analyzer. The USB ports connect
printer, barcode scanner, etc.. The network ports are used to transmit data.
2-8
Understanding the Analyzer
2-9
Understanding the Analyzer
2.5. Buzzer
The buzzer indicates errors of the analyzer. Tap the touch screen orremove the error to clear
the alarming sound of the buzzer.
Press the [Aspirate] key at the 1 long beep When dialog box message is
analysis screens (including given, the buzzer may not
sample analysis, QC, calibration, beep.
Reproducibility, carryover,
background, aging, optical gain
calibration screens) when
analysis cannot be started.
When the analyzer screen turns Turn off the buzzer. If error occurs during the
black and the message "Please shutdown process, please
power off the analyzer" appears turn off the buzzer when the
screen turns black.
2-10
Understanding the Analyzer
2-11
Understanding the Analyzer
NOTE
Store and use the reagents as instructed by instructions for use of the
reagents.
When you have changed the diluents or lyse, implement a background test
to see if the results meet the requirement.
Pay attention to the expiration dates and open-container stability days of all
the reagents. Be sure not to use expired reagents.
2.7.1.Reagents
M-52D Diluent
M-52D diluent provides stable solution environment for blood cells, dilutes blood sample,
generates sheath fluid and provides conductance in blood cell analysis process.
M-52DIFF Lyse
M-52DIFF lyse dissolves RBCs in the blood cell analysis process. It processes WBCs to
intensify the difference of the WBC sub-populations, and incorporates laser scatter and flow
cytometry to realize DIFF channel WBC analysis.
M-52LH Lyse
M-52LH lyse further processes the blood cells that have been processed by M-52DIFF lyse. It
turns hemoglobin into methemoglobin, which then measured by the colorimetric method. It
also processes WBCs to intensify the difference between basophils and other WBC
sub-populations, and incorporates laser scatter and flow cytometry to realize basophil
analysis.
2-12
Understanding the Analyzer
Probe Cleanser
2-13
3 Understanding the System Principles
3.1. Introduction
The measurement methods used in this analyzer are: the Electrical Impedance method for
determining the RBC and PLT data; the colorimetric method for determining the HGB; flow
cytometry by laser for determining the WBC data. Other parameter results are obtained via
calculation.
3-1
Understanding the System Principles
3.2. Aspiration
If you are to analyze a whole blood sample in the open vial sampling mode, the analyzer will
aspirate 15μL (CBC+DIFF mode) or 11.7μL (CBC mode) of the sample.
If you are to analyze a capillary blood sample in the open vial sampling mode, you should first
manually dilute the sample (20μL of capillary sample needs to be diluted by 480μL of diluent,
dilution ratio: 1:25) and then present the diluted sample to the analyzer, which will aspirate
200μL of the sample.
3-2
Understanding the System Principles
3.3. Dilution
The aspirated sample will be quickly and precisely diluted in RBC bath and then segmented
into two portions. One of these two portions will then be diluted again and processed by
different reagents. After this, they are ready for analysis.
This analyzer can process two types of blood samples – whole blood samples and pre-diluted
sample.
Aspirate
sample 15ul
Diluent
550ul
3-3
Understanding the System Principles
Pre-diluted Mode
Capillary blood
sample 20ul
Diluent 480ul
Dilution ratio
1:25
Aspirate
200ul of
Diluent sample
358ul
Aspirate 21.6ul of
Dilution ratio sample
1:69.75
3-4
Understanding the System Principles
After a predetermined volume of blood is aspirated and diluted by a certain amount of reagent,
it is injected into the flow cell. Surrounded with sheath fluid (diluent), the blood cells pass
through the center of the flow cell in a single column at a faster speed. When the blood cells
suspended in the diluent pass through the flow cell, they are exposed to a laser beam. The
intensity of scatter light reflects the blood cell size and intracellular density. The low-angle
scattered light reflects cell size, and the high-angle scattered light reflects intracellular density
(nucleus size and density). The optical detector receives this scatter light and converts it into
electrical pulses. Pulse data collected can be used to draw a 3-dimensional distribution
(scattergram).
3-5
Understanding the System Principles
Colorimetric Method
The WBC/HGB dilution is delivered to the HGB bath where it is bubble mixed with a certain
amount of lyse, which converts hemoglobin to a hemoglobin complex that is measurable at
530 nm. An LED is mounted on one side of the bath and emits a beam of monochromatic light,
whose central wavelength is 530nm. The light passes through the sample and is then
measured by an optical sensor that is mounted on the opposite side. The signal is then
amplified and the voltage is measured and compared to the blank reference reading
(readings taken when there is only diluent in the bath), and the HGB is measured and
calculated in the analyzer automatically.
HGB
The HGB is calculated per the following equation and expressed in g/L.
HGB(g/L) = Constant × Log 10 (Blank Photocurrent/Sample Photocurrent)
3-6
Understanding the System Principles
Each pulse is amplified and compared to the internal reference voltage channel, which only
accepts the pulses of certain amplitude. If the pulse generated is above the RBC/PLT lower
threshold, it is counted as a RBC/PLT. The analyzer presents the RBC/PLT histogram, whose
x-coordinate represents the cell volume (fL) and y-coordinate represents the number of the
cells.
3-7
Understanding the System Principles
MCV
Based on the RBC histogram, this analyzer calculates the mean cell volume (MCV) and
expresses the result in fL.
This analyzer calculates the HCT(%), MCH (pg) and MCHC (g/L) as follows:
RBC MCV
HCT
10
HGB
MCH
RBC
HGB
MCHC 100
HCT
12
Where the RBC is expressed in 10 /L, MCV in fL and HGB in g/L.
RDW-CV
Based on the RBC histogram, this analyzer calculates the CV (Coefficient of Variation) of the
erythrocyte distribution width, it is expressed in %.
RDW-SD
Based on the standard deviation of erythrocyte size distribution, this analyzer calculates the
RDW-SD, its unit is fL.
MPV
Based on the PLT histogram, this analyzer calculates the mean platelet volume (MPV, fL).
PDW
Platelet distribution width (PDW) is the geometric standard deviation (GSD) of the platelet
size distribution. Each PDW result is derived from the platelet histogram data and is reported
as 10(GSD).
PCT
3-8
Understanding the System Principles
PLT MPV
PCT
10000
P-LCR
Platelet larger cell ratio (P-LCR) is the ratio of the larger platelet (volume larger than 12fL)
count to the total PLT count. This analyzer calculates the P-LCR based on the PLT histogram
and expresses the result in %. In the following figure, S2 represents the number of larger
platelet cells, and S1+S2 represents the total PLT count.
P-LCC
This analyzer calculates the platelet large cell count (P-LCC) and expresses the result in
9
10 /L.
P- 3-9
4 Installing Your Analyzer
4.1. Introduction
WARNING
Installation by personnel not authorized or trained by Mindray may cause
personal injury or damage your analyzer. Do not install your analyzer
without the presence of Mindray-authorized personnel.
Your analyzer is tested before it is shipped from the factory. International symbols and special
handling instructions tell the carrier how to treat this electronic instrument. When you receive
your analyzer, carefully inspect the carton. If you see any signs of mishandling or damage,
contact Mindray customer service department or your local distributor immediately.
4-1
Installing Your Analyzer
At least 50 cm on each side, which is the preferred access to perform service procedures;
Enough room on and below the countertop to accommodate the reagents and waste
containers;
The diluent container shall be put within 1.0m under the analyzer, lyse containers are
placed inside the analyzer.
The countertop (or the floor) where the analyzer is placed shall be able to withstand at
least 40kg of weight.
4.2.2.Power Requirements
WARNING
Make sure the analyzer is properly grounded.
Before turning on the analyzer, make sure the input voltage meets the
requirements.
CAUTION
Using pinboard may bring the electrical interference and the analysis
results may be unreliable. Please place the analyzer near the electrical outlet
to avoid using the pinboard.
Please use the original power cable shipped with the analyzer. Using other
power cable may damage the analyzer or cause unreliable analysis results.
4-2
Installing Your Analyzer
4.2.3.General Environment
Optimal operating temperature: 10 ℃~30 ℃
The environment shall be free from dust, mechanical vibrations, loud noises, and
electrical interference.
Do not use this analyzer in close proximity to sources of strong electromagnetic radiation
(e.g. unshielded intentional RF sources), as these may interfere with the proper operation.
Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close.
Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
WARNING
Installation by personnel not authorized or trained by Mindray may cause
personal injury or damage your analyzer. Do not install your analyzer
without the presence of Mindray-authorized personnel.
NOTE
Before the analyzer is shipped out, the sample probe is fixed by a plastic
cable tie to avoid damaging the sample probe during transportation.
Remove the cable tie before using the analyzer.
4-3
Installing Your Analyzer
4-4
Installing Your Analyzer
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc.
according to government regulations.
The reagents are irritating to eyes, skin and airway. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them and the contacted areas in the laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
4-5
Installing Your Analyzer
4-6
Installing Your Analyzer
CAUTION
Make sure the diluent pipe and waste pipe are no longer than 1500mm.
The waste and diluent containers must be placed lower than the countertop
that accommodates the analyzer.
4-7
Installing Your Analyzer
4.4. Notes
The analyzer performance may be undermined if it has been placed in environment of
high dustiness.
The surface of the analyzer shall be cleaned and sterilized regularly with alcohol (75%).
The probe wipe block of the analyzer (see Figure 4-5 Front of the analyzer) shall be wiped
with alcohol (75%) regularly.
If any of the pipes or fluidic components is worn out, stop using the analyzer and contact
Mindray customer service department immediately for inspection or replacement.
Check and make sure the reagents, lyse and waste pipes are not pressed or bent.
You must only use the Mindray-specified reagents, otherwise the analyzer may be
damaged or provide unreliable results.
Pay attention to the expiration dates and open-container stability days of all the reagents.
Be sure not to use expired reagents.
4-8
5 Operating Your Analyzer
5.1. Introduction
This chapter provides step-by-step procedures for operating your analyzer on a daily basis.
A flow chart indicating the common daily operating process is presented below.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
WARNING
Do not contact the patients' sample blood directly.
5-1
Operating Your Analyzer
The reagents are irritating to eyes, skin and airway. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them and the contacted areas in the laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
Keep your clothes, hairs and hands away from the moving parts to avoid
injury.
The sample probe tip is sharp and may contain biohazardous materials.
Exercise caution to avoid contact with the probe when working around it.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes,
capillary tubes and so on.
NOTE
Use the reagents specified by the manufacturer only. Store and use the
reagents as instructed by instructions for use of the reagents.
Check if the reagent tubes are properly connected before using the analyzer.
5-2
Operating Your Analyzer
Checking reagents
Check to see if the reagents are expired or frozen. Reagents must be equilibrated for 24 hours
before use.
Check and make sure the reagents, waste and pneumatic unit tubes are properly connected
and not bent.
Check and make sure the power cord of the analyzer is properly plugged into the power
outlet.
Check and make sure enough printer paper is installed. Check and make sure the power cord
of the printer is properly plugged into power outlet, and the printer is properly connected to
the analyzer.
5-3
Operating Your Analyzer
1. Change the power switch at the backside to ON position ('I') will power on the instrument.
2. The indicator light turns on.
3. The analyzer will perform self-test and initialization.
NOTE
Time needed for initializing the fluidic systems depends on how the analyzer
was previously shut down.
Time needed for starting up the analyzer depends on the startup sequences
run. The startup time should be within 8 minutes.
If the results of the first background check do not meet requirement, the
analyzer will perform background check again.
5-4
Operating Your Analyzer
4. Enter the current user name and the password respectively into the "User ID" box and the
"Password" boxes.
5-5
Operating Your Analyzer
NOTE
If the software cannot be started successfully after being launched for
several times, contact Mindray Customer Service Department or the
authorized distributors.
The default user name and password for administrator are both "Admin".
The user name and password may consist of 1-12 letters, and the password
cannot be null.
NOTE
If error occurs during the initialization process (e.g., background check
fails), the analyzer will report the error. See Chapter 11 Troubleshooting Your
Analyzer for the solution.
5-6
Operating Your Analyzer
The system opens different function for the user according to the user level.
The user level depends on the user name and the password when the user
logs in.
If user switching is necessary, tap the "Logout" icon on the system menu.
Enter the desired user name and the password into the pop-up dialog box
and tap the "OK" button to log in.
Running sample with the background abnormal error present will lead to
unreliable results.
5-7
Operating Your Analyzer
5-8
Operating Your Analyzer
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
WARNING
The sample probe is sharp and potentially biohazardous. Do not contact the
sample probe during operations.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes,
capillary tubes and so on.
NOTE
Make sure the probe tip does not contact the sample tube to avoid potential
spillage.
5-9
Operating Your Analyzer
1)Whole blood
sample
3)Predilute sample
1)Enter sample
info.
2)Select “Whole
Blood” or
“Predilute” mode
3)Press Aspirate
5.5.2 Sample analysis key
4)Remove the
sample when hear
the beep sound
5)Run sample
analysis and report
results
2)Print or transmit to
LIS
5.5.3 Processing
analysis results 3)Parameter flags
4)Flag info.
5-10
Operating Your Analyzer
5.5.1.Sample Preparation
The analyzer can run 3 types of samples: whole blood samples, capillary whole blood
samples and pre-diluted samples.
CAUTION
Prepare samples following the recommend procedure of the manufacturer.
CAUTION
Be sure to collect at least 0.5 mL of blood to ensure the accuracy of the
results.
5-11
Operating Your Analyzer
CAUTION
Be sure to collect at least 120uL of capillary whole blood to ensure the
accuracy of the results.
NOTE
Be sure to run the capillary whole blood samples within 3 minutes to 2
hours after being collected.
3) Pre-diluted samples
1. Tap the diluent dispensing icon, the following dialog box pops up.
2. Present a clean tube to the sample probe, press the aspirate key to dispense diluents
(480μL). The dispensing progress bar will be displayed on the screen.
3. To continue with diluent dispensing, repeat the step 1-2.
4. Add 20μL of venous blood or capillary blood to the diluent, close the tube cap and mix it
properly according to your laboratory's protocol.
5. Tap “Cancel” after preparing all the samples, the analyzer will clean the sample probe
automatically.
5-12
Operating Your Analyzer
NOTE
You can also use pipette to aspirate 480μL of diluent.
After mixing the capillary sample with the diluent, be sure to wait 3 minutes
and then remix before running the sample.
Be sure to run the pre-diluted samples within 30 minutes after the mixing.
Be sure to mix any sample that has been prepared for a while before running
it. Do not mix the samples with massive force using swirl mixer.
5.5.2.Sample Analysis
Tap "Sample Analysis" to enter the sample analysis screen. Tap "Mode" button to select
"Whole blood-CBC+DIFF", "Whole blood-CBC", "Capillary whole blood-CBC+DIFF",
"Capillary whole blood-CBC", "Pre-diluted-CBC+DIFF" or "Pre-diluted-CBC" mode.
5-13
Operating Your Analyzer
NOTE
Letters, numerics and all characters (including special characters)
supported by the keyboard are allowed for sample ID entering.
The allowed length of sample ID is [1, 20], and the ID cannot be null.
Select patient gender from the "Gender" pull-down list. There are two options: "Male" and
"Female".
5-14
Operating Your Analyzer
Enter the patient's date of birth into the "Date of Birth" box. Its format must be the same with
the system date format.
The analyzer provides four ways for you to enter the patient’s age – in years, in months, in
days and in hours. The first way is designed for the adult or pediatric patients no younger
than one year; the second for the infant patients one month to two years; the third for the
neonatal no older than one month, and the fourth for the neonatal no older than 48 hours. You
may choose one of the four ways to enter the patient age.
NOTE
If the patient's date of birth is entered, his/her age will be calculated
automatically, and the age field will gray out and cannot be edited.
If the entered date of birth is later than the current system, then it is
considered invalid.
Enter the name of the department into the “Department” box or select it from the
"Department" pull-down list (when there are previously saved records in the list). The saved
contents will be added in the pull-down list automatically.
Enter the number of the patient’s bed into the “Bed No.” box.
Enter the time when the sample is collected into the “Draw Time” box.
Enter the delivery time of analysis into the "Delivery Time" box.
To enter the name of the person who sent the sample for analysis, enter the name into the
5-15
Operating Your Analyzer
“Clinician” box or SELECT the desired name from the “Clinician” pull-down list (if there are
previously saved names in the list). The saved contents will be added in the pull-down list
automatically.
Entering comments
OK
When you have finished entering the work list information, tap the "OK" button to save the
changes and return to the "Sample Analysis" screen.
Cancel
If you do not want to save the entered work list information, tap the "Cancel" button to return
to the "Analysis" screen without saving the changes.
Enter the sample ID in the "Sample ID" box. Tap "OK" to save the ID and close the dialog box,
the ID will be displayed on the screen as the next sample ID.
2) Selecting mode
Make sure the analyzer indicator is green. Select whole blood (CBC+DIFF or CBC), capillary
whole blood (CBC+DIFF or CBC), or pre-diluted (CBC+DIFF or CBC) mode based on your
needs on the mode selection screen. The selected mode will be displayed at the bottom of
the screen.
5-16
Operating Your Analyzer
3) Aspirate sample
Present the sample to the sample probe. Press the [Aspirate] key to start the analysis.
5-17
Operating Your Analyzer
Tap “RUO Screen” button, and the RUO Screen pops up as the following figure shows:
5-18
Operating Your Analyzer
NOTE
During the analysis process, if errors like clog or bubble occur, the analyzer
will automatically display the results of related parameters as invalid, and
alarm information will show on the error information area. See Chapter 11
Troubleshooting Your Analyzer for the way to remove errors.
3) Parameter flags
See the following section for details about parameter flags.
If the parameter is followed by an “H” or “L”, it means the analysis result has exceeded
the upper or lower limit of the reference range (See Section 9.2.4 Ref. range).
If you see “*****”, as opposed to the result, it means the result is invalid; if you see
“+++++” as opposed to the result, it means the result is out of the display range (See
Table 5-1 Display Range for details).
5-19
Operating Your Analyzer
Flag
Flag Meaning Judgment criterion
Type
Interference of PLT clump
The DIFF and BASO channels
WBC Abnormal or NRBC to WBC count and
are unproportionate.
differential may exist:
Many scatter-points in the
Immature cells or blasts
Immature Cell? immature cell area of the
may exist
scattergram
Abnormal lymphocytes or Many scatter-points in the
Abn./Atypical
atypical lymphocytes may abnormal/ atypical lymphocytes
Lym?
exist. area of the scattergram
WBC
Possible presence of PLT Many scatter-points in the PLT
Flag PLT Clump?
clump clump area of the scattergram
Possible presence of lipid Many scatter-points in the lipid
Lipid Particles?
particles particles area of the scattergram
Many scatter-points in the NRBC
NRBC? Possible presence of NRBC
area of the scattergram
9
Leucopenia Low WBC analysis results WBC < 2.50×10 /L
9
Leucocytosis High WBC analysis results WBC > 18.00×10 /L
Low neutrophils analysis 9
Neutropenia NEUT# < 1.00×10 /L
results
5-20
Operating Your Analyzer
5-21
Operating Your Analyzer
5.6. Auto-Standby
When the time for which the analyzer is free from fluidic operations reaches whatyou have set
at the "Setup" screen of the analyzer (default setting is 15 minutes), a dialog box will pop up,
prompting "Entering standby status…".
After entering standby status, the message " Standby. Press the aspirate key to exit. " will be
displayed at the bottom left of the screen.
NOTE
The analyzer will not enter standby status from the Status screen.
If it is time for auto-Standby and the analyzer is reporting error, then the
error must be resolved first.
During this condition, you can still perform any other operations (e.g.,
printing and transmission) other than fluidic operations.
5-22
Operating Your Analyzer
Refer to Section 9.2.5 Maintenance Setup for how to edit waiting time before
entering standby mode.
Aspirate key
After exiting from standby status, the dialog box above will close automatically.
NOTE
When exiting from the standby status, the analyzer will perform different
maintenance operations based on the time consumed entering standby
status.
If error occurs when the analyzer is exiting from the standby status, see
Chapter 11 Troubleshooting Your Analyzer for solutions.
After exiting the standby status, the analyzer will resume its original status.
The Analysis icon will turn into solid green. And the analyzer indicator will
also turn into solid green.
5-23
Operating Your Analyzer
5.7. Shutdown
CAUTION
Do not start up the analyzer immediately after it is shut down. Wait for at
least 10 seconds.
NOTE
To ensure stable analyzer performance and accurate analysis results, be
sure to perform the "Shutdown" procedure to shut down the analyzer after it
has been running continuously for 24 hours.
2. Tap "OK".
3. When dialog box prompting probe cleanser maintenance displays, present probe cleanser
to the sample probe, and press [Aspirate] key. The probe will aspirate probe cleanser
automatically.
4. After shutting down finishes, the message "Please turn off the power of the analyzer!" will
be displayed. Press the Power switch on back of the instrument to power off.
WARNING
Be sure to dispose of reagents, waste, samples, consumables, etc.
according to government regulations.
5-24
Operating Your Analyzer
NOTE
Do not disconnect power during the shutdown process.
Time needed for shutting down the analyzer depends on the shutdown
sequences run. The shutdown time should be within 10 minutes.
If error that will affect shutdown occurs during the showdown process, the
analyzer will resume to its original status and report the error. See Chapter
11 Troubleshooting Your Analyzer for solutions.
5-25
6 Reviewing Sample Results
6.1. Introduction
The analyzer automatically saves analysis results. The BC-5130 can store up to 150,000
analysis results.
You can review all the analysis results, scattergrams and histograms either in table or graph
mode.
6-1
Reviewing Sample Results
6.2.1.Table
The table lists all analyzed samples, including basic sample information like sample ID,
sample state and so on.
6-2
Reviewing Sample Results
NOTE
The table displays the latest sample results at the top.
6.2.2.Graph Review
Tap "Graph Review" button at the table review screen, or tap the "Previous" button at the
graph review screen to view the analysis results of samples.
6-3
Reviewing Sample Results
Other Parameters
Tap “Other Para.” button, and the “Other Para.” screen pops up. In the “Microscopic Para.”
tab, you can see the following parameters.
In the “RUO Screen” tab, you can see the following parameters.
6-4
Reviewing Sample Results
Select one or more sample records, tap "validate", the sample state of the record will be
marked with "validated".
6-5
Reviewing Sample Results
Cancel validate
Select one or more validated sample records, tap "Cancel validate", the "validated" marks will
disappear.
3. Tap "OK" to delete the record, and the dialog box will be closed.
6-6
Reviewing Sample Results
6.2.5.Edit info.
Tap the desired sample result and it will be highlighted. Tap the "Edit Info." button and the
following dialog box will display.
You may edit the sample and patient information, and tap "OK" to save the change. The
information on the table review screen will be refreshed.
6.2.6.Edit results
Tap the desired sample result and it will be highlighted. Tap the "Edit Result" button and the
following dialog box will display.
6-7
Reviewing Sample Results
Modify the results and tap "OK" to save the changes. The information on the graph review
screen will be refreshed.
6.2.7.Search
1. Tap "Search", the following dialog box will display.
2. Enter search conditions into the edit boxes or select them from the pull-down lists.
3. Tap "OK" to start search, the results will displayed in the table.
6-8
Reviewing Sample Results
6.2.8.Print
Print reports as per the default report template
Select sample records to be printed, and then tap "Print" to print them. In the table review
interface, a “printed” mark will be applied to each printed sample in the sample state sector.
NOTE
In sample state sector, “Validated” mark is prior to „Printed‟ mark.
6.2.9.Transmission
Transmit selected data
6-9
Reviewing Sample Results
6.2.10. Export
1 Tap "Export", the following dialog box will display.
6-10
Reviewing Sample Results
6-11
7 Using the QC Programs
7.1. Introduction
The Quality Control (QC) programs consist of strategies and procedures that measure the
precision and stability of the analyzer. The results imply the reliability of the sample results.
QC involves measuring materials with known, stable characteristics at frequent intervals.
Analysis of the results with statistical methods allows the inference that sample results are
reliable. Mindray recommends you run the QC program daily with normal level controls.
A new lot of controls should be analyzed in parallel with the current lot prior to their expiration
dates.
This may be accomplished by running the new lot of controls twice a day for five days using
any empty QC files. The QC files calculate the mean, standard deviation and coefficient of
variation for each selected parameter. The instrument-calculated means of these ten runs
should be within the expected ranges published by the manufacturer.,
This analyzer provides 2 QC programs: L-J QC and X-B QC.
All the samples, controls, calibrators, reagents, wastes and areas contacted
them are potentially biohazardous. Wear proper personal protective
equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory procedures
when handling them and the contacted areas in the laboratory.
WARNING
Keep your clothes, hairs and hands away from the moving parts to avoid
injury.
The sample may spill from the uncapped collection tubes and cause
biohazard. Exercise caution to the uncapped collection tubes.
The reagents are irritating to eyes, skin and airway. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them and the contacted areas in the laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
7-1
Using the QC Programs
CAUTION
Running QC sample with analyzer error will lead to unreliable results. If
errors are reported during QC analysis, remove the errors first and then
continue with the analysis.
NOTE
Use the controls and reagents specified by the manufacturer only. Store and
use the controls and reagents as instructed by their instructions for use.
Store and use the controls and reagents as instructed by instructions for
use of the controls and reagents.
Refer to the instructions for use of the control for its use and storage.
Be sure to mix any control sample that has been prepared for a while before
running it.
7-2
Using the QC Programs
7.2. L-J QC
7.2.1.Editing L-J Settings (for administrators only)
Before running a new lot of controls, you must set up a QC file for each lot of controls.
1 Tap the menu option "QC" > "L-J QC" > "Setup".
Manual Entry
2. Tap "New", or select a QC file without QC results, and then tap "Edit".
3. Tap "Import File".
7-3
Using the QC Programs
5. Tap "OK" to close the dialog box and return to the L-J QC setup screen.
NOTE
The "Import target/limits" check box is selected by default. If it is
deselected, the operator must enter the target and limits of QC
parameters manually.
10 Set QC sample ID: if you are used to analyzing control together with blood samples,
you can set a unique ID for the control. The analyzer will recognize the sample as
control when it reads the unique ID. After the analysis completes, the results will be
saved into the QC file of the QC sample ID.
Manual Entry
7-4
Using the QC Programs
3. Manually enter the lot No. of the controls in the edit box.
NOTE
The lot No. shall not be empty and up to 16 digits can be entered.
You can enter characters, numbers, letters and special characters.
8. Set QC sample ID: if you are used to analyzing control together with blood samples,
you can set a unique ID for the control. The analyzer will recognize the sample as
control when it reads the unique ID. After the analysis completes, the results will be
saved into the QC file of the QC sample ID.
9. Enter the target and limits in the edit boxes according to the package insert of the
same lot of controls.
7-5
Using the QC Programs
Setting limits
You can adjust the format of limits as per the following procedure:
1. Tap "Set Limits".
2. Tap “By SD” to display the limits in the form of absolute value;
Or tap “By CV” to display the limits in the form of percentage.
7.2.2.L-J QC Run
You can select one of the two ways below to run controls:
Put controls together with normal samples, and run the controls under the sample analysis
screen.
After editing the QC information, you can start QC analysis by one of the following ways
according to the selected QC mode.
Whole blood
Pre-diluted
7-6
Using the QC Programs
CAUTION
Running QC sample with error present will lead to unreliable results. If
errors are reported during QC analysis, remove the errors first and then
continue with the analysis.
NOTE
When switching mode from "Pre-diluted" to "Whole Blood", a progress bar
will be displayed while the analyzer runs mode switching sequence.
1. Tap "QC" > "L-J QC" > "Count" to enter the QC count screen.
NOTE
Be sure that the level of the control to be run is the same with the
current QC file, and the control is not expired.
7-7
Using the QC Programs
3. Run QC analysis:
1) Make sure the analysis mode is "Whole Blood" or "Pre-diluted" and the indicator of
the analyzer is green.
2) Shake the vial of sample as instructed by instructions for use of the control to mix
the sample thoroughly.
3) Present the control sample to the sample probe. Press the [Aspirate] key to start QC
run.
4) When you hear the beep, remove the control.
4. When analysis finishes, the QC results will be displayed in the current screen and be
saved in the QC file automatically.
NOTE
Up to 100 QC results can be saved in each QC file.
7-8
Using the QC Programs
After setting special "QC Sample ID" for a control under the QC setup screen, you can put the
control together with normal samples, and run it under the “Count” screen.
When editing worklist or entering next sample information in the "Next Sample" dialog box
before daily analysis, enter the special "QC Sample ID" as "Sample ID".
Based on the QC mode selected, you can choose to run QC analysis from one of the
following ways:
Whole blood
Pre-diluted
Refer to section 5.5.1 Sample Collection and Handling for sample preparation under
whole blood and pre-diluted modes.
2. When it is ready to run a sample (i.e. the status icon and the analyzer indicator is
green), present the sample to the sample probe.
3. When you hear the beep, remove the control.
4. When analysis finishes, the QC results will be displayed in the current screen and be
saved in the QC file automatically.
NOTE
Up to 100 QC results can be saved in each QC file.
NOTE
When switching mode from "Pre-diluted" to "Whole Blood", a progress bar
will be displayed while the analyzer runs mode switching sequence.
Tap "Edit Result" on the QC screen to edit results and tap "OK" to save the edited results. The
edited results will be marked with an "E".
7-9
Using the QC Programs
Operators of administrator access level can restore the edited results to the original
measurement results.
1. Tap "Restore" on the edit result screen.
QC Graph
QC Table
1. Tap "Graph" button on the "L-J QC" screen to enter the L-J QC graph screen.
2. You can tap the arrow buttons on the right of the graph to browse graphs of the
parameters. You can tap the arrow buttons under the graph horizontally to browse all
the QC results.
7-10
Using the QC Programs
NOTE
With the QC files having QC results, when you have modified and saved a
parameter target/limit, other parameter targets/limits may change
accordingly; and all those changed data will be highlighted in yellow.
Tap the "Print" icon in the status bar to print information of the current QC file and the QC
graph of all parameters.
NOTE
The green vertical line and values of the corresponding QC points will not
be printed.
1. Tap "Table" button on the "L-J QC" screen to enter the L-J QC table screen.
2. You can tap the arrow buttons on the right of the graph to browse all QC records. You
can tap the arrow buttons under the graph horizontally to browse all the parameter
results.
7-11
Using the QC Programs
NOTE
With the QC files having QC results, when you have modified and saved a
parameter target/limit, other parameter targets/limits may change
accordingly; and all those changed data will be highlighted in yellow.
NOTE
The operation will be recorded in the system log.
You can tap the "Print" icon in the status bar to print the QC table.
Transmission
7-12
Using the QC Programs
3. Tap "OK" to start transmitting specified results to the data management software.
NOTE
If “Auto Communicate” is enabled for sample results (see 9.2.1 System
Setup), and a sample is run during the transmission of the QC data, then
only when the QC data transmission finishes will the auto-communication of
the sample result start.
The QC data saved after the QC data transmission starts will not be
transmitted in this cycle.
Export
2 The system will detect the USB and export data automatically.
7-13
Using the QC Programs
7-14
Using the QC Programs
At the X-B QC setting screen, you may activate/deactivate X-B QC, set target/limits, and
configure the sample validity setup.
7-15
Using the QC Programs
1. In the “Samples /Batch” edit box, you may enter the amount of samples [within the
range 20 (default) to 200] to be included in calculating for an X-B QC point.
2. Activate/deactivate X-B QC. If X-B QC is activated, the samples meeting validity
requirements will be included in X-B QC.
Set target/limits
Before the X-B QC analysis, you shall set up the target and limit for each parameter at the
X-B QC setup screen.
NOTE
The units of target/limits of all parameters are the same as those in the
parameter unit setup screen.
1. In the “Target/Limit” area of the X-B QC setup screen, enter the targets and limits in
the “Target/Limit” table.
NOTE
Do not leave any of the targets and limits for the QC parameters
blank.
When first use, the default setting will provide the initial values for
the targets and limits of all QC parameters.
In X-B QC, following sample results will be considered as invalid and cannot be used in the
QC calculation.
Background results;
Calibration data;
7-16
Using the QC Programs
Results generated while there are errors which could affect the accuracy of the
results (insufficient aspiration volume or clogging for example).
"Sample Validity Setup" is to set up the valid ranges of RBC, MCV, MCH and MCHC results.
Only when the results of all these four parameters are within the specified ranges, the sample
results can be used for X-B QC calculation. Do as follows to set the sample validity:
The default validity range of each parameter is shown in the following figure.
7-17
Using the QC Programs
NOTE
In the sample validity setup, the upper limit shall be no smaller than the
lower limit.
The entered valid ranges of the RBC parameters should not exceed their
linearity ranges; the entered valid ranges of other parameters should not
exceed their display ranges.
All the entries shall be numbers with only one decimal point. The length of
the number entered cannot be longer than the length of the text box.
Once the validity range is changed, the previous results before the change
will be discarded. For example, assuming 20 valid samples are needed for
the X-B QC calculation, and you have already acquired 10 groups of valid
sample results; then you change the validity range settings. Then only valid
sample results generated afterwards will be used in the QC calculation.
The units of lower and upper limits of all parameters are the same as those
in the parameter unit setup screen. See section 9.2.4 Setup - Parameter Unit
Setup.
Setting limits
You can adjust the format of limits as per the following procedure:
1. Tap "Set Limits".
7-18
Using the QC Programs
2. Tap “By SD” to display the limits in the form of absolute value;
Or tap “By CV” to display the limits in the form of percentage.
Restore defaults
If you want to restore the default targets and limits of the parameter, tap "Restore Defaults".
The default values of the target and limits of each parameter are as follows:
7.3.3.X-B QC Run
After editing X-B setup, the system will start X-B run automatically.
After every 20~200 results (determined by the setting) are obtained, the system will perform
the X-B calculation at once. You can review the result in X-B QC graph or X-B QC table.
QC Graph
QC Table
7-19
Using the QC Programs
1. Tap the menu option "QC" > "X-B QC" > "Graph", the following screen will display.
2. Select QC file No., the information of the file and the QC graph will be displayed on the
screen.
3. You can tap the arrow buttons under the graph horizontally to browse all the QC results.
7-20
Using the QC Programs
3. You can tap the arrow buttons on the right of the graph to browse all QC records.
The delete, print and export operations can all be performed in the same way as stated in the
L-J QC table review section.
7-21
8 Calibrating Your Analyzer
8.1. Introduction
Calibration is a procedure to standardize the analyzer by determining its deviation under
certain specified conditions. In order to get accurate sample analysis results, you should
calibrate the analyzer per the procedure below when necessary.
There are three calibration programs available on this analyzer: manual calibration, auto
calibration using calibrators and auto calibration using fresh blood samples.
All the parameters or part of the parameters of WBC, RBC, HGB, MCV and PLT can be
calibrated by the calibration programs.
All the samples, controls, calibrators, reagents, wastes and areas contacted them
are potentially biohazardous. Wear proper personal protective equipment (e.g.
gloves, lab coat, etc.) and follow safe laboratory procedures when handling them
and they contacted areas in the laboratory.
WARNING
The reagents are irritating to eyes, skin and airway. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them and they contacted areas in the laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
Keep your clothes, hairs and hands away from the moving parts to avoid
injury.
CAUTION
Do not reuse disposable products such as collection tubes, test tubes,
capillary tubes and so on.
8-1
Calibrating Your Analyzer
NOTE
Be sure to use the Mindray-specified disposable products including
evacuated blood collection tube, anticoagulant collection tubes and
capillary tubes etc.
Use the calibrators and reagents specified by the manufacturer only. Store
and use the calibrators and reagents as instructed by their instructions for
use.
8-2
Calibrating Your Analyzer
You are going to use this analyzer for the first time (usually done by a Mindray-authorized
representative when installing the analyzer).
NOTE
All of the measured parameters must be calibrated before readings of this
analyzer can be used as valid analysis results.
8-3
Calibrating Your Analyzer
4. It is recommended that you create a log table for your analyzer. This log table should
contain all necessary information that is pertinent to your analyzer. Suggested items that you
may want to include in the log table are: calibration date, supplier of calibrator, lot number,
expected results and limits, and result of background check.
NOTE
Be sure to use the evacuated collection tubes recommended in the
Appendix.
If fresh blood samples are used for reproducibility test, make sure the
sample volume is enough to support the test.
8-4
Calibrating Your Analyzer
NOTE
If you log in at the operator access level, you can only view the calibration
factors. To perform calibration, please log out and then log in at the
administrator access level.
1. At the "Manual” calibration screen, check the calibration factors and calculate the new
factors per the following equation:
98.90% 8.4
New factor= =101.81%
8.16
The calculated calibration factors shall be between 75.00%~125.00%. In case of an invalid
8-5
Calibrating Your Analyzer
calibration factor, try to find out the reason (possible reasons include, calibration material not
thoroughly mixed, mis-operation, etc.). Then re-calibrate the analyzer and re-calculate the
calibration factors.
2. Enter the new calibration factors into the factor cell of the parameter that requires
calibration.
3. When you switch screen after entering the new calibration factor, a prompt will display.
If the entered calibration factors are valid, a dialog box will display asking you to save the new
factor when you are exiting the screen. And the calibration date of the corresponding
parameter changes to current system date.
If the entered calibration factors are invalid, a dialog box will pop up prompting "Invalid entry"
when you are switching to another screen. The new calibration factor will not be saved, and
the calibration date will not be refreshed.
Other Operations
If the calibration factors are valid but not saved, a dialog box will display asking you to save the
factors.
Tap "Yes" to save and print the factors. Or tap "No" to cancel the operation without saving or
printing them.
8-6
Calibrating Your Analyzer
NOTE
The calibrator calibration can be performed under Whole Blood -CBC+DIFF,
Whole Blood -CBC, pre-diluted -CBC+DIFF and pre-diluted -CBC mode.
See the instruction for use of the calibrators for the lot No., expiration date
and the targets.
The out-of-range CV% does not influence the display of calibration factors.
8-7
Calibrating Your Analyzer
When the current running is done, if there is a parameter whose calibration data is out of
its linearity range but still within the display range, then the calibration data will be
displayed in the list and a message box will also pop up.
Tap “OK” to close the message box, and the data will be deleted from the table without saving
automatically.
When the running is done, if there is a parameter whose calibration data is out of the
display range, then the non-numeric parameter values "***" will be displayed in the list
and a message box will pop up.
Tap “OK” to close the message box, and the data will be deleted from the table without saving.
The valid results within the linearity range will be displayed directly. Valid calibration results will
be marked with "√” per the default setting, and will be taken to calculate calibration factors.
8. If the calibration factors have not been calculated but you switch to another screen, then a
message box will pop up.
8-8
Calibrating Your Analyzer
Tap "Yes" to switch to another screen while discarding the calibration data and closing the
message box. The original calibration factors remain.
9. When calibration count has been performed to a sample for n times (n≥5), the analyzer will
calculate the Mean, CV% and calibration factors of all the calibration data marked with "√"
(calibration data of the first run is not marked with "√", so it is not included in the calculation).
You can select several data to calculate the calibration factors, but only after at least 5 groups
of the data are marked with "√" can you get the calibration factors. The calibration factors will
be refreshed whenever you select “√” or deselect “√”.
When the amount of valid calibration data in the list reaches 10, a message box "Calibration is
completed!" will pop up. Then, if you press the aspirate key again, the analyzer will beep
without starting analysis.
10. There may be two cases when you are switching to another screen:
If the calibration factors of any parameter is out of the range [75%-125%] or the CV% of
any parameter exceeds the reproducibility range, then the calculated calibration factors of
all parameters will not be saved and a message box will also pop up.
8-9
Calibrating Your Analyzer
Tap "Yes" to close the dialog box and switch to another screen. The calibration factors and
dates of all parameters will not be changed.
If the calculated calibration factors of all parameter are within the range [75%-125%] and
the CV% of all parameter are also within the reproducibility range, then a message box
will pop up.
Tap "Yes" to save the new calibration factors while closing the message box and switching to
another screen.
Other Operations
If the calibration factors are invalid, tap “Print”, the dialog box "New calibration factor is invalid."
will display.
If the calibration factors are valid but not saved, tap "Print", a dialog box will display asking you
to save the factors.
8-10
Calibrating Your Analyzer
Tap "Yes" to close the dialog box, save and print the calibration results. Or tap "No" to cancel
the operation without saving or printing them.
8-11
Calibrating Your Analyzer
If the results are out of the linearity range but still within the display range, a dialog box will pop
up when the results are displayed in the table.
Tap “Yes” to close the message box, and the data will be deleted from the table without saving.
If the results are out of the display rage, the non-numeric parameter values "***" are obtained
and a dialog box will pop up.
8-12
Calibrating Your Analyzer
Tap “Yes” to close the message box, and the data will be deleted from the table without saving.
The valid results within the linearity range will be displayed directly.
Valid calibration results will be marked with "√” per the default setting, and will be taken to
calculate calibration factors.
9. When calibration count has been performed to a sample for n times (n≥5), the analyzer will
calculate the Mean, CV% and calibration factors of all the calibration data marked with "√"
automatically.
You can select several data to calculate the calibration factors, but only after at least 5 groups
of the data are marked with "√" can you get the calibration factors. The calibration factors will
be refreshed whenever you select “√” or deselect “√”.
When the amount of valid calibration data in the list reaches 10, a message box "Calibration
with the current blood sample is completed." will pop up when you start calibration again.
10. Select other calibration sample ID from the "Current Sample ID" pull-down box, analyze
other samples according to step 7-9 above to obtain the calibration factors of all samples.
11. There may be several cases when switching to another blood sample:
If the calibration factors of the blood sample are invalid or the CV% of any parameter exceeds
the reproducibility range, a dialog box will pop up when switching to another blood sample.
Tap "Yes" to empty the entered target of the current sample, all the calibration data obtained
and each calculated value including calibration factors, then close the dialog box and switch to
another blood sample.
If the calibration factors have not been calculated, a dialog box will pop up.
8-13
Calibrating Your Analyzer
Tap “Yes” to empty the entered target of the current sample and all the calibration data
obtained, then close the dialog box and switch to another blood sample.
If the calibration factors of the sample are valid and the CV% of all the parameters do not
exceed the reproducibility range, you can switch to another blood sample directly.
12. After calibration factors of at least 3 fresh blood samples are obtained, tap the "Calculate"
button to enter the screen of calibration calculation.
Select or deselect the calibration factors of a blood sample for the calculation of the mean
calibration factors by tapping the check boxes before the calibration factors.
When 3 or more groups of calibration factors are checked, CV% will be re-calculated
automatically based on the checked calibration factors.
When 3 or more groups of calibration factors are checked, the mean calibration factor will be
re-calculated automatically based on the checked calibration factors. The mean calibration
factors are regarded as invalid if the deviation of absolute value between the calibration factors
included in calculating the mean and the original calibration factors reaches or exceeds 5%; a
dialog box will pop up when you exit the current fresh blood calibration screen.
8-14
Calibrating Your Analyzer
Tap "Yes" to close the dialog box and exit with the current calibration data emptied, and then
switch to another screen.
Tap "No" to return to the current screen. Invalid mean calibration factors are followed with a "?",
and displayed in red.
13. If the mean calibration factors have not been calculated, when you exit the fresh blood
screen or switch to another calibration mode, a dialog box will pop up.
Tap "Yes" to discard the calibration data, close the dialog box, and switch to another screen or
calibration mode. The original calibration factors and date remain the same.
14. If the calculated mean calibration factors are valid, when exiting the fresh blood calibration
screen or switching to another calibration mode, a dialog box will pop up.
Tap "Yes" to save the current mean calibration factors. Then, you can switch to another screen
or calibration mode. Tap "No" to close the dialog box and switch to another screen or
calibration mode without saving the mean calibration factors and all the calibration data.
8-15
Calibrating Your Analyzer
Other Operations
If the mean calibration factors are invalid, tap print, the dialog box "Calibration factor is invalid."
will display.
If the mean calibration factors are valid, you can tap "Print" to print the calibration factors of a
group (or more) of blood samples in table form, no matter whether they are selected ("√") or
not. The results obtained in the calibration process and the mean calibration factors can also
be printed.
8-16
9 Customizing the Analyzer Software
9.1. Introduction
The BC-5130 is a flexible laboratory instrument that can be tailed to your working
environment. You can use the “Setup” program to customize the software options as
introduced in this chapter.
For the security of the settings and data, two access levels are provided to the operators of
the analyzer. The administrator access level provides the operator with access to more
functions or settings, some of which can be configured to be accessible to operators.
9-1
Customizing the Analyzer Software
Print
Tap the menu option "Setup" > "System Setup" > "Date/Time" to enter the "Print Setup"
screen as shown below. You can set up the following contents:
Print setup
Printing content
Auto print
9-2
Customizing the Analyzer Software
Print setup
Print device
There are 2 types of printing device available: printer and recorder. You can select either of
them from the pull-down list.
Printer driver
Paper
Tap the pull-down list to select the paper type of the reports to be printed.
Parameter language
9-3
Customizing the Analyzer Software
Tap the pull-down list to select the parameter language of the reports.
Copies
Enter the number of copies to be printed for each report into the edit box "Copies".
Report title
Report template
Applicable report templates for printer are shown as below:
Printing content
You can choose to select the functions based on your needs by tapping on the check boxes.
9-4
Customizing the Analyzer Software
Auto print
Communication setup
Tap the menu option "Setup" > "System Setup" > "Communication" to enter the
communication setup screen as shown below. You can set up the following contents:
Protocol Setup
Transmission Mode
9-5
Customizing the Analyzer Software
Protocol Setup
Tap the "IP Address", "Subnet Mask" and "Default Gateway" edit boxes to enter the contents.
Communication Protocol
Tap the "Comm. Protocol" pull-down list to select the communication protocol.
Transmission Mode
You can choose to select the functions based on your needs by tapping on the check boxes.
Auto retransmit
Auto Communicate
Not to be transmitted
Bitmap
Data
9-6
Customizing the Analyzer Software
This screen shows the values that may trigger flag alarm. Operators of administrator access
level can modify the reference values of the flags, which indicated the possibility of triggering
flags. The lower the values are, the higher the possibility can be.
9-7
Customizing the Analyzer Software
9-8
Customizing the Analyzer Software
9-9
Customizing the Analyzer Software
NOTE
The analyzer SN cannot be edited.
The date of installation is the date the analyzer is installed by default. It can
be edited, but cannot be later than the current system date.
9.2.2.Access Setup
Tap "Setup" > "Access Setup" in the menu to enter the following screen.
9-10
Customizing the Analyzer Software
Modify password
9-11
Customizing the Analyzer Software
NOTE
The password cannot be null, and 12 characters can be entered at most.
NOTE
The user ID cannot be null, and 12 characters can be entered at most.
9-12
Customizing the Analyzer Software
Delete user
NOTE
The current login user cannot be deleted.
9.2.3.Auxiliary Setup
Tap "Setup" > "Auxiliary Setup" in the menu to enter the following screen. You can set up the
following contents:
Setting of the next sample
Setting of the first sample after startup
Other settings
9-13
Customizing the Analyzer Software
Operators may:
Customize the first sample ID after startup by entering it into the edit box. Or select to
continue with the sample ID before last shutdown.
9-14
Customizing the Analyzer Software
Other settings
Flags
Operators may set up the suspect flag by entering a character into the edit box, or selecting a
letter from the pull-down list (the default character is "R").
Operators may set up the high/low flag by entering two characters into the edit boxes, or
selecting two letters from the pull-down lists (the default character of high flag is "H", and that
of low flag is "L").
9.2.4.Parameter Setup
Parameter unit setup
Tap the menu option "Setup" > "System Setup" > "Reference Unit Setup" to enter the screen
as shown below. You can set up parameter unit at this screen.
9-15
Customizing the Analyzer Software
Under each unit system, you can tap the "Unit" cell to customize the parameter unit.
Tap the "Default" button to restore the default units.
NOTE
The units displayed will be different when different unit system is selected.
9-16
Customizing the Analyzer Software
Select a reference group and tap "New" or "Edit" to enter the reference group setup screen.
You can set up the name, lower and upper limits of age and parameter range.
9-17
Customizing the Analyzer Software
Tap the "Set to Default" button, the reference ranges of the selected factory reference group
can be restored to the default settings.
NOTE
The name of the reference group cannot be null.
The names of the customized reference groups shall not repeat the names
of the 5 default groups, and they shall not repeat each other either.
Select a reference group and then tap "Set to Default" to set it as default reference group.
NOTE
The name, lower and upper limits of age and gender of the factory reference
groups cannot be modified.
To modify the reference range of a reference group, select the group from the reference
group list on the left, and then tap the cells of upper and lower limits in the table and re-enter
the values.
To restore the reference ranges to default, tap the “Default” button on top right of the screen.
When "Match customized ref. group first" is selected, if the age ranges of the customized
reference group and the default reference group contradicts with each other, the customized
reference group will be matched first on the sample analysis and review screens.
9-18
Customizing the Analyzer Software
Tap the “New” button to add a new row in the table, and then you can enter the name of the
parameter in the row.
Delete
Select a row in the table, tap the “Delete” button to delete the parameter.
NOTE
You can add up to 40 microscopic parameters.
The reconfigured setup will not be applied to sample records which already
have microscopic results saved, but only applied to sample records with
unsaved microscopic results and records attained after the new setup is
applied.
9-19
Customizing the Analyzer Software
Standby
Tap the text box "Wait" and enter the waiting time before entering the standby status. The
range allowed is 10 -30 minutes, and the default setting is 15 minutes.
9.2.6.Reagent Setup
Tap "Setup" > "Reagent Setup" in the menu to enter the following screen.
9-20
Customizing the Analyzer Software
It is recommended that you replace the reagents when their residue volume icons turn from
BLUE to RED.
This function may also be used to refill reagent inside the fluidic system when a new
container of reagent is loaded.
NOTE
The reagents must be kept still for at least for a day after long-term
transportation.
When you have changed the diluents or lyse, run a background test to see if
the results meet the requirement.
Diluent
DIFF lyse
9-21
Customizing the Analyzer Software
LH lyse
1. Tap the reagent you want to replace, and then tap "Setup".
NOTE
Please keep the diluent container from severe shock or crashing against
other object. Otherwise, the alarming would be unreliable.
9-22
Customizing the Analyzer Software
RBC gain
Tap the MCV-G "Set Value" cell, and enter the new value of RBC gain.
9-23
Customizing the Analyzer Software
HGB gain
NOTE
The gains of LAS, MAS and WAS cannot be modified.
9-24
Customizing the Analyzer Software
Tap "Yes" to save the settings and switch to the corresponding screen. Tap "No" to switch to
the corresponding screen without saving the settings.
9-25
10 Servicing Your Analyzer
10.1. Introduction
Preventive and corrective maintenance procedures are required to keep the analyzer in a
good operating condition. This analyzer provides multiple maintenance functions for this
purpose.
This chapter introduces how to use the provided functions to maintain and troubleshoot your
analyzer.
All the analyzer components and surfaces are potentially infectious, take
proper protective measures for operation or maintenance.
WARNING
The reagents are irritating to eyes, skin and airway. Wear proper personal
protective equipment (e.g. gloves, lab coat, etc.) and follow safe laboratory
procedures when handling them and the contacted areas in the laboratory.
If reagents accidentally spill on your skin or in your eyes, rinse the area with
ample amount of clean water; seek medical attention immediately.
CAUTION
Improper maintenance may damage the analyzer. Operators must follow the
instruction of this Operator's Manual to perform maintenance operations.
Exercise caution to avoid contact with the sharp sample probe when
performing maintenance.
The following table lists the tools that may be used in maintenance.
No. Tools
1. Cross-headed screwdriver
2. Slotted head screwdriver
3. Medical gloves
4. Alcohol
10-1
Servicing Your Analyzer
10.2.1. Maintenance
Tap "Service" > "Maintenance", and select the "Maintenance" tab to enter the following
screen.
Unclog aperture
Unclogging includes zapping and flushing the aperture. When clog error us reported, you
should unclog the aperture.
The unclogging procedures are:
1. Tap the "Unclog" button to start unclogging.
2. When the progress ends, a message will display indicating "Maintaining finished!".
3. Do the above procedures to continue unclogging aperture if necessary. If the error
persists, perform probe cleanser maintenance of the related channels.
10-2
Servicing Your Analyzer
when analyzer shuts down due to abnormal power break-off; probe cleanser
maintenance must be performed after it is started up again.
4. After aspirating probe cleanser, the analyzer performs probe cleanser soak
automatically, and a progress bar will display indicating the progress.
5. When the progress ends, the following dialog box will display, tap “OK” to close the
dialog box.
10-3
Servicing Your Analyzer
10.2.2. Cleaning
You should clean the following components when:
WBC and (or) HGB background results exceed their limits, perform WBC bath cleaning.
If WBC bath cleaning does not solve the problem, perform WBC probe cleanser
maintenance.
RBC and (or) PLT background results exceed their limits, perform RBC bath cleaning. If
RBC bath cleaning does not solve the problem, perform RBC probe cleanser
maintenance.
there are too many particles in the scattergram of background results, perform WBC
bath cleaning. If WBC bath cleaning does not solve the problem, perform WBC probe
cleanser maintenance.
Tap "Service" > "Maintenance", and select the "Cleaning" tab to enter the following screen.
10-4
Servicing Your Analyzer
Fluidics
Flow cell
Sample probe
WBC bath
RBC bath
10-5
Servicing Your Analyzer
Pack-up
If the analyzer is not to be used for over 2 weeks, you should perform this procedure.
Do as follows to pack up:
1. Tap "Pack-up", the dialog box "Start pack-up?" will pop up.
2. Tap "Yes" to perform the pack-up procedure. The following dialog box will display.
3. Take out the tubes as instructed and then tap "OK" to drain the fluidics.
10-6
Servicing Your Analyzer
4. The following dialog box will display after draining the fluidics.
5. Put the tubes into distilled water as instructed, and tap "OK" to start priming.
6. When the priming progress ends, the following dialog box will be displayed.
7. Take out the tubes as instructed and then tap "OK" to drain the fluidics again.
10-7
Servicing Your Analyzer
8. The following dialog box will be displayed after draining the fluidics.
NOTE
This software can still be used after the pack-up.
Reset
When major components of the analyzer have been replaced, or the fluidic system has been
serviced, you must reset the fluidics.
Do as instructed below:
1. Tap "Reset Fluidics", a dialog box will pop up asking you to confirm the operation.
2. Tap "OK" to start initialization, the message "Resetting fluidics. Please wait..." will be
displayed.
3. When the progress ends, a dialog box will display indicating “Resetting fluidics
finished!”.
4. Do the above procedures to continue resetting fluidics if necessary.
10-8
Servicing Your Analyzer
10-9
Servicing Your Analyzer
You may view the error info., parameter modification info. and records of daily operation in
the log.
The “Log” screen records all activities of the analyzer. It contributes significantly to searching
for operation history and troubleshooting the analyzer.
NOTE
The oldest record will be overwritten automatically when number of log
records reaches the utmost.
Exporting logs
1. Tap "Export", the following dialog box will display.
10-10
Servicing Your Analyzer
10-11
Servicing Your Analyzer
NOTE
If the status is outside normal range, it will be highlighted with red
background.
10.5.1. Counter
The counter counts the running times of the analyzer and the occurrence times of some
major parameters.
Viewing details
You may tap the "Detail..." buttons following "Runs", "QC Runs" or "Calibration Runs" to view
the related details.
10-12
Servicing Your Analyzer
10-13
Servicing Your Analyzer
10-14
Servicing Your Analyzer
10.5.4. Sensor
Tap "Status" > "Sensor" in the menu to enter the following screen.
You may check the sensor status of the analyzer.
10-15
Servicing Your Analyzer
10-16
11 Troubleshooting Your Analyzer
11.1. Introduction
This chapter contains information that is helpful in locating and correcting problems that may
occur during operation of your analyzer.
NOTE
This chapter is not a complete service manual and is limited to problems
that are readily diagnosed and/or corrected by the user of the analyzer.
11-1
Troubleshooting Your Analyzer
Red: fatal error. When this kind of error occurs, the analyzer will stop running
immediately, and any further operation is prohibited.
Orange: error that stops operation. When this kind of error occurs, the analyzer will stop
running immediately.
Blue: error that restricts certain operations. When this kind of error occurs, the analyzer
can still continue with the current operation, but any other operations related to the error
will be restricted.
The name and troubleshooting method of the errors are displayed. Names of the errors are
displayed by the order of their occurrence.
You may tap to select the error, and view its troubleshooting information in the
troubleshooting box. The troubleshooting information of the first error is displayed by default.
Please follow the troubleshooting to resolve the error by sequence.
11-2
Troubleshooting Your Analyzer
Remove error
Tap the "Remove Error" button to clear all the errors that can be removed automatically. For
the errors that cannot be removed automatically, follow the troubleshooting method to solve
them.
Tap "Close" to close the dialog box, but the errors will still be displayed in the error info. area
on the screen. Tap the error info. area again, the dialog box will be displayed.
The possible error(s) and the corresponding troubleshooting information are listed below:
11-3
Troubleshooting Your Analyzer
the lyse.
3. If the error still exists after replacing the lyse, contact our
customer service department.
1. Tap "Remove Error", and enter the new barcode of the
LH lyse into the reagent setup dialog box.
2. After replacing the lyse container, tap "Apply" to prime
LH lyse expired
the lyse.
3. If the error still exists after replacing the lyse, contact our
customer service department.
1. Tap "Remove Error", and enter the new barcode of the
DIFF lyse into the reagent setup dialog box.
2. After replacing the lyse container, tap "Apply" to prime
DIFF lyse expired
the lyse.
3. If the error still exists after replacing the lyse, contact our
customer service department.
Empty the waste container or use a new waste container.
2. Tap “Remove Error” to see if the error can be removed.
Waste container full
3. If the error still exists, contact our customer service
department.
1. Power off the analyzer directly, and then contact our
Power source error
customer service department.
1. Power off the analyzer directly, and then contact our
Temperature sensor error
customer service department.
1. Power off the analyzer directly, and then contact our
Preheat assembly error
customer service department.
1. Power off the analyzer directly, and then contact our
Laser error
customer service department.
1. Tap “Remove Error” to see if the error can be removed.
Syringe module error 2. If the error still exists, contact our customer service
department.
1. Tap “Remove Error” to see if the error can be removed.
Aspiration module lift
2. If the error still exists, contact our customer service
mechanism error
department.
1. Tap “Remove Error” to see if the error can be removed.
Aspiration module rotary
2. If the error still exists, contact our customer service
mechanism error
department.
1. Tap “Remove Error” to see if the error can be removed.
Background abnormal 2. If the error still exists, contact our customer service
department.
1. Tap “Remove Error” to see if the error can be removed.
Exiting standby mode failed 2. If the error still exists, contact our customer service
department.
1. Tap “Remove Error” to see if the error can be removed.
Replacing diluent failed
2. If the error still exists, contact our customer service
11-4
Troubleshooting Your Analyzer
department.
1. Tap “Remove Error” to see if the error can be removed.
Replacing DIFF lyse failed 2. If the error still exists, contact our customer service
department.
1. Tap “Remove Error” to see if the error can be removed.
Replacing LH lyse failed 2. If the error still exists, contact our customer service
department.
1. Check if there is diluent reagent in the container. If not,
replace the container with a new one.
HGB blank voltage abnormal 2. Tap "Remove Error" to remove the error.
3. If the error persists, contact our customer service
department.
1.Please do "Probe Cleanser Maintenance".
2.Please do "Clear Flow Cell".
3. Power off the analyzer directly, and restart it after a
Liquid pressure overloaded
while.
4. If the error persists, contact our customer service
department.
1. Tap “Remove Error” to see if the error can be removed.
Vacuum pressure abnormal 2. If the error still exists, contact our customer service
department.
1. Tap “Remove Error” to see if the error can be removed.
Preheat bath temperature
2. If the error still exists, contact our customer service
error
department.
1. The ambient temperature shall be within [10℃, 30℃]
Machine temperature too high 2. Diluent temperature shall be within [10℃, 30℃].
3. Tap “Remove Error” to see if the error can be removed.
Diluent temperature is too 1. Diluent temperature shall be within [10℃, 30℃].
high
Diluent temperature is too low 1. Diluent temperature shall be within [10℃, 30℃].
1. Tap “Remove Error” to see if the error can be removed.
Clogging 2. If the error still exists, contact our customer service
department.
1. Tap “Remove Error” to see if the error can be removed.
Aperture voltage abnormal 2. If the error still exists, contact our customer service
department.
Impedance channel signal 1. Please eliminate sources of interference.
abnormal 2. Tap “Remove Error” to see if the error can be removed.
1. Tap "Remove Error" button to remove the error.
2. If the error persists, power off the analyzer and power on
Analysis is abnormal later.
3. If the error persists, contact our customer service
department.
RBC channel is abnormal 1. Please check the analysis mode.
11-5
Troubleshooting Your Analyzer
11-6
12 Appendices
A. Index
Analysis, 5-13 Logs, 10-10
Aspiration, 3-2 maintenance, 10-1
Auto-Sleep, 5-22 Manual Calibration, 8-4
buzzer, 2-11 Parameter flags, 5-19
Calibration, 8-1 Parameters, 2-2
calibrators, 2-14 Prediluted samples, 5-12
Capillary whole blood samples, 5-11 Quality Control, 7-1
Colorimetric Method, 3-6 Reagents, 2-13
controls, 2-14 Setup, 9-1
Dilution, 3-3 Shutdown, 5-24
Electrical Impedance Method, 3-7 software, 1-2
Flow Cytometry by Laser, 3-5 Startup, 5-4
Graph Review, 6-3 Table Review, 6-2
hardware, 1-2 Touch Screen Calibration, 10-9
Initial Checks, 5-3 Transmission, 6-9
Installation, 4-2 troubleshooting, 11-2
Logon, 5-4 Whole blood samples, 5-11
A-1
B. Specifications
B.1. Classification
According to the CE classification, the BC-5130 belongs to In vitro diagnostic medical devices
other than those covered by Annex II and devices for performance evaluation.
B.2. Reagents
Name Model
Diluent M-52D
Lyse M-52DIFF
M-52LH
Probe Cleanser /
B.4. Parameters
Parameter Abbreviation Default Unit
9
White Blood Cell count WBC 10 /L
9
Basophils number Bas# 10 /L
Basophils percentage Bas% %
9
Neutrophils number Neu# 10 /L
Neutrophils percentage Neu% %
9
Eosinophils number Eos# 10 /L
Eosinophils percentage Eos% %
9
Lymphocytes number Lym# 10 /L
Lymphocytes percentage Lym% %
9
Monocytes number Mon# 10 /L
Monocytes percentage Mon# %
9
Abnormal Lymphocyte number ALY# (RUO parameter) 10 /L
Abnormal Lymphocytes
percentage ALY% (RUO parameter) %
B-1
Specifications
9
Large Immature Cell number LIC# (RUO parameter) 10 /L
Large Immature Cell percentage LIC% (RUO parameter) %
9
Platelet clump number PLT Clumps#(RUO parameter) 10 /L
Platelet clump percentage PLT Clumps%(RUO parameter) %
Lipid granule number Lip#(RUO parameter) 109 / L
Lipid granule percentage Lip%(RUO parameter) %
Nucleated red blood cell (NRBC)
number NRBC#(RUO parameter) 109 / L
Nucleated red blood cell (NRBC)
percentage NRBC%(RUO parameter) %
12
Red Blood Cell count RBC 10 /L
Hemoglobin Concentration HGB g/L
Mean Corpuscular Volume MCV fL
Mean Corpuscular Hemoglobin MCH pg
Mean Corpuscular Hemoglobin MCHC g/L
Red BloodConcentration
Cell Distribution Width RDW-CV %
Coefficient
Red Blood of VariationWidth
Cell Distribution RDW-SD fL
Standard Deviation
Hematocrit HCT %
9
Platelet count PLT 10 /L
Mean Platelet Volume MPV fL
Platelet Distribution Width PDW None
Plateletcrit PCT %
Platelet-Large Cell Ratio P-LCR %
9
Platelet Larger Cell Count P-LCC 10 /L
White blood Cell Histogram WBC Histogram None
Red blood Cell Histogram RBC Histogram None
Platelet Histogram PLT Histogram None
Differential Scattergram Diff Scattergram None
B.5.2. Throughput
Open vial-whole blood, open vial-pre-diluted and open vial-capillary whole blood modes: no
less than 60 samples per hour
B-2
Specifications
HGB 0 g/L~300g/L
9 9
PLT 0×10 /L~9999×10 /L
HCT 0%~80%
HCT ≤ 0.5 %
≤ 10 10 / L
9
PLT
B-3
Specifications
RBC ≤±6%
HGB ≤±7%
PLT ≤±15%
B.6.5. Compatibility
Deviation ranges: WBC ≤± 5%, RBC ≤±2%, HGB ≤±2%, PLT ≤±8%, HCT/MCV ≤±3%.
B.6.6. Reproducibility
B.6.7. Carryover
Parameter Carryover
WBC ≤0.5%
RBC ≤0.5%
HGB ≤0.6%
HCT ≤0.5%
PLT ≤1.0%
WARNING
Be sure to use the specified devices only.
B-4
Specifications
NOTE
If the analyzer is to be connected with LIS, the PC must be configured with
dual network cards.
RAM: 1G or above
B.8. Interfaces
4 USB ports
B.10. FUSE
WARNING
Use specified fuse only.
B-5
Specifications
This equipment complies with the emission and immunity requirements of the EN
61326-1:2013 and EN 61326-2-6:2013.
NOTE
It is the manufacturer's responsibility to provide equipment electromagnetic
compatibility information to the customer or user.
B.12. Sound
Maximal sound: 65 dBA
NOTE
Be sure to use and store the analyzer in the specified environment.
B-6
Specifications
Height
Depth
Width
BC-5130 Analyzer
Width(mm) ≤ 325
Dimensions Height (mm)≤ 435 (with foot)
Depth (mm) ≤ 410
Weight ≤25Kg
B.17. Contraindications
None
B-7
P/N: 046-010333-00(1.0)