*
MURPHY
ENGINEERING, PROCUREMENT, CONSTRUCTION
AND COMMISSIONING (EPCC) OF PATRICIA
SATELLITE (PT-SA) TOPSIDE FOR SK309/311
SARAWAK DEVELOPMENT PROJECT
CONTRACT NUMBER : C-11-027-ME-039-12MFN
MAGNETIC PARTICLES EXAMINATION PROCEDURE
(MSET DOC. NO : MSETE/M3-297/WP/314)
KENCANA
PETROLEUM
panes Reviewed
wecked | Reviewed i
Deseription Author ‘APBFENEE | Approved
impo Ket KHL
7906.10] _Tesved For Review AH [LARS
County | Ficla/Asset | Facility | Discipline “ales
Code | Code Code ‘Sheet Ne
‘Sequence
Code, No.
Doe. Type
Mom | sk3os MEI7 PR 004 | 25:te ee
Doe No
MOw-SK90071-MEI7-PR-
0005-00,
Page 20125,
REVISION STATUS / SUMMARY OF CHANGES.
Revision | Revised Paragraphs
Revision Description
Reason For Revision‘ee Na:
* ow. k200-71 817.
MURPHY | ™°*=7¢Panncte tesnne proceoure oaosac
Page 3ol 25
TABLE OF CONTENTS,
40 SCOPE PS.
20 DEFINITION P5
3.0 PERSONNEL QUALIFICATION P5
40 ESSENTIAL VARIABLES PS
5.0 EQUIPMENTS PS
60 EQUIPMENT CHECK AND CALIBRATION Po
7.0 TESTING PROCEDURE P10
80 VIEWING P13
9.0 INTERPRETATION OF INDICATION P13
10.0 EVALUATION OF INDICATION P14
11.0 ACCEPTANCE AND REJECTION CRITERIA Pia
12.0 REPORTS. Pts
13.0 DE-MAGNETIZATION P16
14.0 POST CLEANING P16Dee No
Mow.sk208.71-MES7-PR.
saGnpeiy | eermemereen | EE
15.0 RE-TESTING P16
APPENDIX | SAMPLE REPORT P18
APPENDIX II TECHNIQUE SHEET #1 P19
APPENDIX II TECHNIQUE SHEET #2 P21
APPENDIX IV TECHNIQUE SHEET #3 P23Doe No
& Mow.skage-71-ME17-PR.
MAGNETIC PARTIC {OCEDURE 0005-00
MURPHY T1¢ PARTICLE TESTING PR sone
Page Sof 25
1.0 SCOPE
1.1. This procedure describes the method for Magnetic Particle Examination to
be employed by MSET Engineering Corporation Sdn. Bhd
1.2. This procedure is written in accordance with the requirements of ;
1) Section V of ASME Code, Edition 2010,
2) Section Vill Division 1 or 2, Edition 2010,
3) MOM-SK309-71-ME-SP-0013-00 REVO, Technical Spec. for
Pressure Vessel
2.0 DEFINITIONS
a. CLIENT Company or organization for whom MSET Engineering
Corporation Sdn. Bhd. performing the work.
b. NDT Non Destructive Testing
c. ASNT The American Society for Non Destructive Testing
3.0 PERSONNEL QUALIFICATION
3.1 All personnel carrying out Magnetic Particle Testing in accordance with this
procedure shall be qualified in Magnetic Particle Testing Level Il or Level
Il to. MSET Engineering Corporation Sdn. Bhd. written practice
(MSETeMPI355 ISSUE 1 Rev. 1) using SNT-TC-1A as a guide.
4.0 ESSENTIAL VARIABLES
44 All required Essential Variable as per Table 1 shall be specified in
Technique Sheeti ae
Doe No
MOw.SK30071-MEI7-PR
Rew A
Page of 25
42
43
‘A change of requirement in Essential Variable from the specified value, or
range of value, shall require separate Technique Sheet to be added and
do not require revision of this procedure.
A change in Essential Variable listed in Table 1 shall require
requalification.
TABLE L; Essential variables
Jie [cent arises
1 [Magnetising technique
2 [Magnetzing current type or amperage oulide range speciied
3 [Surface preparation
«4 [Magnetic particles (Huorescarl Nile, col, panicles size,
wet/dry)
[Method of parides application
Method of excess patical remonl
Vinimurn fight rtensity
[Coatingthickness greater than spected
Perfomance demonstration
[Surface temperature culside recommended
TABLE; Nonfscential variables
fie. [essential Variatter
{_|Shupe orsite ofthe exanimtion obeat
2. [Equipment cf the s3m2 type
3 [Temperature (within those speclied By m snufselure or
evcurly qualfied
7 Oemagnetisng lehrigae
5
é
ost examination cleaning technique
ersonrel qualification requirement
5.0 EQUIPMENT
84
Magnetizing EquipmentDoe No
MOMSK309-71-MEI7-PR-
+
MURPHY | ™°xetePaencte testo eroceoure sons
age 7 0125,
5.1.4 Permanent magnetic yoke or AC yoke shall be used for the
magnetizing equipment.
5.1.2 Permanent magnetic and AC yokes shall be capable of lifting
calibration block in Para 5.4, made of the same material to be
inspected with the poles set at the pole spacing to be used.
5.1.3 Yoke used shall be specified in Technique Sheet.
5.2. Magnetic Particle
5.2.1 General
5.2.1.1 Either dry magnetic particle or wet magnetic particle, and
either color contrast or fluorescent may be used in
accordance to this procedure.
6.2.1.2 The shape and size of all magnetic particles, either wet or
dry, suspension vehicle and concentration shall be in
accordance with SE-709.
5.2.2 Dry Particle
5.2.2.1 If diy particle is used, the color of the particle (dry
powder) shall provide adequate contrast with the surface
being examined
5.2.2.2 Dry magnetic particle examination shall not be performed
if the surface temperature of the part exceeds 315°C.
5.2.2.3 Dry particle used shall be specified in Technique Sheet.‘Doe Wo
MoM.SK300-71-ME7 PR.
*
MURPHY |. ™2v#"ePsnnour rest proceoure es
Pages ct 25
5.2.3. Wet Particle
5.2.3.1 If wet particle is used, the color of the particle shall provide
adequate contrast with the surface being examined.
5.2.3.2 Wet magnetic particle examination shall not be performed
if the surface temperature of the part exceeds 55°C.
5.2.3.3 Wet particle used shall be specified in Technique Sheet.
5.3. Contrast Aids
5.3.1. White background paint shall be used when conducting magnetic
particle testing using wet color contrast technique.
5.3.2 White background paint shall be of an approved type, to present a
high contrast background to any color magnetic ink indicating the
presence of the defects
5.3.3 The contrast paint shall be applied by aerosol can spray to give a
thin, even coating and is quick drying,
5.3.4 White background paint used shall be specified in Technique Sheet.
5.4 Calibration Block
5.4.1 A minimum 18 kg calibration biock shall be used to test the strength
‘of magnetizing force of permanent magnetic yoke and a minimum
45. kg calibration block shall be used to test the strength of
magnetizing force of the AC yoke,Deena:
* (MONE-SK309-71-41E17-PR-
MURPHY | ™ox7e°asncieresmna proceoure cots
Page 90126
5.4.2 Each calibration block shall be weighed with a scale from a
reputable manufacturer and stenciled with the applicable nominal
weight prior to first use. A calibration block need only be verified
again if damaged in a manner that could have caused potential loss
of material
5.5 Flux Indicator
554
55.2
The strength and direction of induced magnetic field in the material
shalll be checked using flux indicator.
Only ASME Magnetic Field Indicator/Pie shaped magnetic particle
field indicator shall be used
6.0 EQUIPMENT CHECKS AND CALIBRATION
6.1 Yoke Magnetizing Force
644
613
614
‘The magnetizing force of yokes shall be calibrated annually or after
any repair or overhaul. This test shall be made by measuring the
yoke's lifting power. All calibration results shall be recorded and a
calibration certificate issued.
Daily verification shall be carried out to ensure that the calibration
criteria have been maintained. This shall be conducted at the
beginning and at the end of each shift
Permanent magnetic yoke must be capable of lifting and holding an
18 kg calibration block and AC yoke must be capable of lifting and
holding a 4.5 kg calibration block during magnetization at the
maximum poles spacing to be used for inspection.
Yoke that fails to lift the calibration block shall not be used for
inspection.2°
MURPHY
‘Doe Na:
MoM-SKA00-71-MEI7-PR-
MAGNETIC PARTICLE TESTING PROCEDURE 0005-00
Rew A
Page 10025
6.2 Magnetic Field Strength and Direction on Test Object
6.24
622
623
It may be necessary to check and demonstrate the presence and
direction of magnetic flux in test object during Magnetic Particle
Testing
This test shall be conducted at the beginning and on the test object
to be tested on each Magnetic Particle Testing.
This shall be done using ASME Magnetic Field Indicator.
6.3 Visible and Fluorescent Light Meters
634
6.3.2
Visible and fluorescent light meters used to measure illumination
evel shall be calibrated at least once a year or whenever the meter
has been repaired,
If the meters have not been used for one year or more, calibration
shall be done before being used
7.0 TESTING PROCEDURE
7.1 Surface Preparation
TAA
Satisfactory results are usually obtained when the surfaces are in;
as-welded, as-rolled, as-cast or as-orged conditions. However,
surface preparation by grinding or machining may be necessary
where surface irregularities could mask indications due to
discontinuities.*
MURPHY
‘Doe Na
Mow.ska0e-71-Me17.9R.
MAGNETIC PARTICLE TESTING PROCEDURE 0008-00
Rew A
Page tt of 25
TAD
TA4
TAT
Prior to magnetic particle examination, the surface to be examined
and all adjacent areas within at least 1 inch (25mm) shall be dry and
free of all dirt, grease, lint, scale, welding flux and spatter, oll or
other extraneous matter that could interfere with the examination.
If coatings are left on the part in the area being examined, it must be
demonstrated that indications can be detected through the
maximum coating thickness; otherwise the coating shall be
removed.
Cleaning may be accomplished using detergents, organic solvents,
de-scaling solutions, paint removers, vapor degreasing, sand or grit
blasting or ultrasonic cleaning methods.
Quick drying contrast paint may be used at areas where insufficient
contrast between background and magnetic particles exist
‘Surface examination shall include checking of any discontinuity
surface openings which may not attract or hold magnetic particles
because of their width.
The shape and size of object to be examined are not limiting
factors, provided the specified area under examination can be
properly examined within the specified time intervals of this
procedure.
7.2 Testing Coverage
7.24
7.22
Testing object requiring Magnetic Particle Testing shall be tested for
their full area (100%) unless partial or spot testing is specified by
Client.
The surface shall be examined visually prior to any examination as
to determine the suitability of the testing technique to be employeddices | — |
‘ee Na:
MOM-SKO08-71-MEI7-PR.
Pago 120128
73
74
75
7.2.3 Two separate examinations shall be performed in each area, with
the lines of magnetic flux during which the second examination to
be approximately 90° to those of the first examination.
Magnetization
7.3.1 The magnetization shall be induced by a continuous magnetization
of Alternating Current Electro-Magnetic Yoke.
7.3.2 The minimum poles spacing shall be 100mm and the maximum
spacing shall be 200mm.
7.3.3. However, the time for each magnetization shall not be too long to
avoid over heating of the yoke
Particles Application
7.4.1 The Ferromagnetic Particles shall be lightly dusted or sprayed on
the work piece whilst the Magnetic Flux is being induced.
7.4.2 Sufficient time shall be allowed for indications to ‘Build up’ before
evaluation.
Excess Particle Removal
7.5.1 Accumulation of excess magnetic particles in examination surfaces
shall be removed by a light air stream or blow.| eer ee
‘Doc No
MOWe-SK300-71-MEI7-PR-
Page 130128,
8.0
9.0
7.5.2 For dry magnetic particles, accumulation of excess magnetic
particles in examination surfaces shall be removed by a light air
stream from a bulb or syringe or other source of low pressure dry
air.
7.5.3 The magnetizing power shall be maintained while removing the
excess magnetic particles.
VIEWING
a4
82
83
84
Each entire surface under examination shall be examined before
proceeding to the next stage in the testing procedure and viewing shall be
carried out after each excitation. Magnification of detail may be obtained
by the use of low-power magnifier. Care shall be taken to ensure that
Magnetic indications have been inspected and recorded,
When viewing using color contrast inks or powder the area under
examination shall be illuminated by daylight or artificial light to @ level of
not less than 100 fe (1000 Ix)
Viewing shall be such that the glare is avoided during inspection
‘When using fluorescent inks or powders, the area under examination shall
be darkened and the inspection of the test surface shall be carried out
under black light conditions, The minimum light level shall be 1000
HWiem? and shall be checked at least once every 8 hours or when the
work station is changed
INTERPRETATION OF INDICATIONS
94
Observation shall be made while applying the magnetic particle during
magnetization process so that the indication may be reviewed before
magnetic field diffusion occurs.Doone:
MoM SK308-7-EI7-PR-
Ea
MURPHY | "2*e0