AP8F10 Oxygen Concentrator
AP8F10 Oxygen Concentrator
AP8F10
User’s Manual
SAFETY NOTES…………………………………...2-5
FEATURES…………………………………………6-7
HANDLING…………………………………………8
OPERATING & INSTALLATION…….…………9-14
MAINTENANCE………………………………….15-21
DISASSEMBLY………………………………..….22-24
1
SAFETY NOTES
Ⅰ.BEFORE INSTALLATION
The concentrator should always be kept upright to prevent cabinet
damage while being transported.
If the electrical source power voltage is instable that beyond
normal power rang, please add the voltage stabilizer.
Please use eligible, safe power set and junction box.
Non-professional people do not open oxygen concentrator’s
cabinet.
Ⅱ.PUT POSITION
You may select a room in your house where using your oxygen
concentrator would be most convenient. Your concentrator can be
easily rolled from room to room on its casters.
Be certain to place the unit so all sides are at least 10cm (4inches)
away from walls, draperies, furniture, or other obstruction. Do not
place the unit in a confined area.
Oxygen concentrator must avoid heat source, fire source, wetness,
exorbitant or over-low extra ventricular condition.
Sundries and vessel do not be placed on top of the concentrator.
NEVER block the air openings of the unit or place it on a soft
surface, such as a bed or couch, where the air opening may be
2
SAFETY NOTES
blocked. Keep the openings free from lint, hair and the like.
Ⅲ.USING
The use of oxygen therapy requires that special care be taken to
reduce the risk of fire. Users MUST NOT SMOKE while using
this device. Keep all matches, lighted cigarettes or other sources of
ignition out of the room in which this product is located. NO
SMOKING signs should be prominently displayed. Textiles and
other materials that normally would not burn are easily ignited and
burn with great intensity in oxygen enriched air. Failure to observe
this warning can result in severe fire, property damage and cause
physical injury or DEATH.
For optimum performance, do not open or turn off the concentrator
frequently, reset after 3~5 minutes. Shorter periods of operation
may reduce maximum product life.
A spontaneous and violent ignition may occur if oil, grease or
greasy substances come in contact with oxygen under pressure.
These substances MUST be kept away from the oxygen
concentrator, tubing and connections, and all other oxygen
equipment. DO NOT use any lubricants unless recommended by
manufacturer.
Ⅳ. MAINTENANCE
The oxygen concentrator was specifically designed to minimize
routine preventive maintenance at intervals of once per year. Only
professionals of the healthcare field or persons fully conversant
with this process such as authorized or factory trained personnel
should perform preventive maintenance or performance
3
SAFETY NOTES
adjustments on the oxygen concentrator.
Ⅴ. RADIO FREQUENCY INTERFERENCE
Most electronic equipment is influenced by Radio Frequency
Interference (RFI). CAUTION should be exercised with regard to
the use of portable communications equipment in the area around
such equipment.
Ⅵ. TO REDUCE THE RISK OF BURNS, ELECTROCUTION,
FIRE OR INJURY TO PERSONS.
Avoid using while bathing. If continuous usage is required by the
physician’s prescription:
The concentrator must be located in another room at least
2.5m from the bath.
DO NOT come in contact with the concentrator while wet.
DO NOT place or store product where it can drop into water or
other liquid.
DO NOT reach for product that has fallen into water. UNPLUG
IMMEDIATELY.
A product should NEVER be left unattended when plugged in.
This device is to be used only in accordance with the prescription of
a physician and this User’s Manual. If at any time the patient or
attendant conclude that the patient is receiving an insufficient
amount of oxygen, the supplier and/or physician should be contacted
immediately. No adjustments should be made to the flowrate unless
prescribed by a physician.
Close supervision is necessary when this product is used near
children or physically-challenged individuals.
4
SAFETY NOTES
Use this product for only intended use as described in this manual.
DO NOT use parts, accessories or adapters other than those
authorized by manufacturer.
DO NOT connect the concentrator in parallel or series with other
oxygen concentrators or oxygen therapy devices.
Use of some administration accessories or certain humidifiers not
specified for use with oxygen concentrator may impair the
performance.
In certain circumstances oxygen therapy can be hazardous.
Manufacturer recommends that you seek medical advice before
using this product.
Avoid creation of any spark near medical oxygen equipment. This
includes sparks from static electricity created by any type of
friction.
5
FEATURES
Ⅰ. SUMMARY
Oxygen concentrator is intended for individual use in the home. It is
an electronically operated device that separates oxygen from room
air. It provides high concentration of oxygen directly to you through
a nasal cannula. Clinical studies have documented that oxygen
concentrator are therapeutically equivalent to other types of oxygen
delivery systems.
This user’s manual will tell you about your concentrator and will
serve as a reference as you use your concentrator.
Ⅱ. CHRARCTERS
Complete plastic outer shell, safe and reliable.
Timing accumulates function, show total working hours through
the display screen.
Exceed press safety valve helping safe ensure.
Install power loss alarming function, pressure and cycle failure
alarming function, compressor failure alarming function.
Compressor has got hot protection that ensures the safety of the
compressor and concentrator.
With double flux function.
Ⅲ. SPECIFICAIONS
1. Flow Range: 1~10L/min
NOTE: total flux 10L/min, for two patients use at same time.
2. Oxygen Concentration: 95.5%~87%
3. Maximum Outlet Pressure: 40kPa~70kPa (5.8psi~10.2psi)
4. Pressure Relief Mechanism Operational at:
250kPa±25kPa (36.25psi±3.63psi)
6
FEATURES
5. Altitude: Up to 1230 meters (4000 ft) above sea level without
degradation of concentration levels. From 1230 meters (4000 ft) to
4000meters (13129 ft) below 90% efficiency.
6. Sound Level: ≤60dB(A)
7. Power Supply:
□AC220V±10%
□50Hz
2.2amps(AC220V~240V)
8. Input Power: 850VA
9. Weight: 34.5kg ( 75.9 lbs )
10. Dimension:
H445mm (17.5″) ×W372mm(14.7″)×D680mm(26.8″)
11. Minimum Operating Time: 30minutes
Electric classification: II class, type B applied part.
Work system: work continuously.
Keep it far from flammable and explosive gas area.
Ⅳ.NORMAL WORKING CONDITION
Temperature range: 10℃~35℃(50℉~95℉)
Relative humidity: 20%~60%
Atmosphere pressure: 86kPa~106kPa (12.47psi~15.37psi)
NOTICE: Please place the device in normal working temperature
circumstance for at least 4 hours if the storage, transportation
temperature beneath 5℃(41℉).
7
HANDLING
Ⅰ. UNPACKING
Check for any obvious damage to the carton or its contents. If
damage is evident, please notify the carrier or local dealer.
Remove all loose packing from the carton.
Carefully remove all the components from the carton.
NOTE: Unless the oxygen concentrator is to be used immediately,
retain containers and packing materials for storage until use of the
concentrator is required.
Ⅱ. INSPECTION
Examine exterior of the oxygen concentrator for nicks, dents,
scratches or other damages. Inspect all components.
Ⅲ. STORAGE
Store the repackaged oxygen concentrator in a dry area.
DO NOT place other objects on top of the repackaged concentrator.
8
OPERATING & INSTALLATION
Ⅰ.FEATURE VIEW
Elapsed time meter
Oxygen outlet 1 Flowmeter
Fault indicator light (red)
Oxygen outlet 2 Power Indicator Light (green)
O2 sensor indicator
light (yellow)
Humidifier Circuit breaker
outlet
Cabinet
9
OPERATING & INSTALLATION
Ⅱ.PREPARE WORK
Unscrew the cover of the humidifier, fill the purify water (or
distilled water) into the humidifier bottle between the maximum
and minimum water level lines, and then screw the humidifier
bottle. (If needed, add other medicine into the water, please
according to the doctor’s suggestion.)
Screw the humidifier bottle absorbing connector into the cover of
the humidifier, then insert the humidifier to the elastic belt on the
left side of the unit, and connect the other end of the cannula to the
oxygen outlet.
Plug in power supply.
Press down the power switch, the concentrator works.
Flowmeter knob
Black floater
Flowmeter
Figure 3
10
OPERATING & INSTALLATION
line on the flowmeter. Next, turn the flow knob until the ball rises
to the line. Now, center the ball on the L/min line prescribed.
CAUTION: If the flowrate on the flowmeter ever falls below
0.5L/min, check tubing or accessories for blocked or kinked
tubing or a defective humidifier bottle.
Meanwhile, humidifier bottle will have air bubble around the sieve
core. Then, oxygen comes from the humidifier outlet.
2. Connect nasal oxygen cannula to humidifier outlet, the other end
match the patient.
Note: The time of absorbing oxygen and the flowrate range according
to the doctor’s advices.
Figure 4
11
OPERATING & INSTALLATION
Ⅴ.ALARM SIGNAL
Oxygen concentration indicator
(If your unit features the O2 sensor feature)
This feature monitors the purity of oxygen generated by the oxygen
concentrator. If O2 concentration falls below factory preset
standards, indicator lights on the control panel will illuminate.
Initial startup of the concentrator
NOTE: Concentrator may be used during the initial start warm-up
time (approximately 30 minutes) while waiting for the O2
concentration to reach maximum.
When the unit is turned on, the green light will come on (O2
concentration greater than 82%). After 5 minutes, the oxygen sensor
will be operating normally and will control the indicator lights
depending on oxygen concentration values. The explanations of the
indicator light functions are as follows.
Explanation of the indicator lights:
Symbol Status Indicator Lights
SYSTEM OKAY
GREEN
(O2 concentration is greater than 82%.)
O2 concentration is greater than 50% and
YELLOW
less than 82%.
SYSTEM FAILURE
(O2 concentration is less than 50%; Or,
RED
High/Low cycle pressure failure alarm; Or,
Compressor circuit failure alarm )
12
OPERATING & INSTALLATION
1. O2 concentration is greater than 82%. -- Green light illuminates,
and LED panel shows elapsed time. Normal Operation.
2. O2 concentration is greater than 50% and less than 82%. --
Green with yellow light illuminate, and LED panel shows
elapsed time. Call Supplier Immediately. You may continue to
use the concentrator unless instructed otherwise by your supplier.
Be certain that BACKUP OXYGEN is nearby.
3. O2 concentration is less than 50%. -- Red light illuminates,
continuous audible alarm sounds, and LED panel shows word
“LO” . Total unit shutdown. Switch immediately to backup
oxygen supply. Call supplier immediately.
4. Low cycle pressure failure alarm -- Red light illuminates,
continuous audible alarm sounds, and LED panel shows word
“E1”. Total unit shutdown. Switch immediately to backup
oxygen supply. Call supplier immediately.
5. High cycle pressure failure alarm -- Red light illuminates,
continuous audible alarm sounds, and LED panel shows word
“E2”. Total unit shutdown. Switch immediately to backup
oxygen supply. Call supplier immediately.
6. Compressor’s open circuit failure alarm -- Red light illuminates,
continuous audible alarm sounds, and LED panel shows word
“E3”. Total unit shutdown. Switch immediately to backup
oxygen supply. Call supplier immediately.
7. Compressor’s short circuit failure alarm -- Red light illuminates,
continuous audible alarm sounds, and LED panel shows word
“E4”. Total unit shutdown. Switch immediately to backup
13
OPERATING & INSTALLATION
oxygen supply. Call supplier immediately.
Ⅵ.TURNING OFF
Take off nasal oxygen cannula from humidifier outlet first, close
power switch, and then cut off power source.
Type B
Class Ⅱ Equipment
applied part
Off On
(power. disconnection (power. connection to
from the mains) the mains)
Breaker No smoking
14
MAINTENANCE
NOTE: Power should be disconnected before beginning preventive
maintenance on the concentrator. The concentrator is specifically
designed to minimize routine preventive maintenance at intervals of
once per year. In places with high dust or soot levels, maintenance
may need to be performed more often. The following must be
performed at a minimum of one year in service to assure years of
additional reliability.
Ⅰ.CLEANING CABINET
Clean the cabinet at least once a month. First cut off the power
supply, and then clean the cabinet with a mild household cleaner
and non-abrasive cloth or sponge. Avoid seeping liquid into the
crack of the cabinet.
Ⅱ.CLEANING FILTER
Please clean and replace the filters in time that is very important to
protect compressor and delay concentrator’s life.
CAUTION: DO NOT operate the concentrator without the filters
installed.
DISASSEMBLY
1. The first stage mesh in filter cleans once half a month.
The first stage filter can be taken out from both sides of concentrat-
or. (Figure 5)
Figure 5
15
MAINTENANCE
2. The second stage mesh in filter cleans once a month.
Plate groove (insert flat-screwdriver)
16
MAINTENANCE
Ⅴ.TROUBLESHOOTING GUIDE
17
MAINTENANCE
Concentrator
doesn’t work, 1) Low oxygen 1) Clean or Replace
red light concentration. filters.
illuminates,
continuous
audible alarm
sounds, and LED 2) If condition persists, discontinue use, contact
panel shows Equipment Provider immediately.
word “LO”.
Concentrator
doesn’t work, 1) Clean or Replace
1) Low pressure alarm.
red light filters.
illuminates,
continuous
audible alarm
sounds, and 2) If condition persists, discontinue use, contact
LED panel Equipment Provider immediately.
shows word
“E1”.
18
MAINTENANCE
19
MAINTENANCE
21
DISASSEMBLY(CUT OFF the power supply first)
22
DISASSEMBLY(CUT OFF the power supply first)
Upward to lift the cabinet. (Figure 10)
Heat
exchanger
Connected
Fan line row
Main controlling
Transformer circuit board
Screws Figure 11
23
DISASSEMBLY(CUT OFF the power supply first)
Inlet sound
buffer box
Compressor
Figure 12
Air valve
Sieve bed
Pressure
regulator
Capacitance
Figure 13
24
OTHER ATTENTION ITEMS
Ⅰ.LIMITATION OF THE TRANSPORT AND STOCK
Temperature Range: -40℃~70℃(-104℉~158℉)
Relative Humidity Range: 10%~100%
Atmosphere Press Range: 86kPa~106kPa(12.47psi~15.37psi)
The oxygen concentrator should be stored in area without erode gas;
avoid shaking and inversion in transportation.
25
OTHER ATTENTION ITEMS
Ⅲ.ELECTRICAL ELEMENTS
Electronic control
~ circuit
Assembled Assembled
Fan Compressor valveⅠ valveⅡ
Ⅳ.ACCESSORIES LIST
1. Nasal oxygen cannula: 1 set
2. Filter core: 1 set
3. Humidifier bottle (including tube): 2 set
4. The base for double flux: 1 set
5. User manual: 1 piece
Manufacturer information:
ALL PRO CORPORATION
No.592 Tonghe Road, Qingdao, China
Service: [email protected]
Tel: +86 186 0532 8290
Made in China
26