4.5.

5 Selecting equipment to be updated


Select in the “Equipment” pull-down menu and check
that the IP address (initial value: 172.16.60.124) of the equipment is displayed in “Equipment IP”.

When the IP address of JLN-740 has been changed due to the


installation condition, enter the IP address after the change.

When is not
displayed in the “Equipment” pull-down menu, the tool may not be
the JLN-740 update tool. Restart the update tool for JLN-740 as
indicated in Figure 4.1-1.

4-23 4 Program Update


4.5.6 Acquiring a software list from the equipment
After inputting the IP address of the equipment, acquire a list of software to be updated in the
equipment by clicking on . When the software list is acquired correctly, the dialogue that is
shown below is displayed.

Figure 4.5-2 Dialogue displayed after the correct acquisition of a list

4 Program Update 4-24


After correct acquisition of a software list, the target software names and the versions are displayed in
a list. If the software to be updated does not exist in the equipment in such a case where optional
equipment is not connected due to the installation condition, “(Ver. Unknown)” is displayed.

The version number of the screen capture is the one currently


under development. Since the version number may be different
from the one that is currently operating in the product, replace the
number as required.

Figure 4.5-3 List of software targeted for update (example)

4-25 4 Program Update


4.5.7 Selecting an update file
After selecting in the “Software” pull-down menu, click
on and select an update file.
When a correct update file is selected, the information of the selected file is displayed in “Update file”,
“Version”, and “Information”, and is enabled.

4 Program Update 4-26


4.5.8 Updating
Click on to start updating.
Update processing progresses in the order of Transmission→Update→Reboot and the progress
status is displayed on the status bar at the bottom of the window.

Transmit the update file that was specified in the update tool to the equipment. It takes about
1 minute.

4-27 4 Program Update


In this case, the “Updating software” dialogue is displayed on the main display.

After completion of the transmission, the flash ROM is rewritten. It takes about 1 minute.
In this case, the “Updating software” dialogue remains displayed on the main display.

4 Program Update 4-28


After the flash ROM has been rewritten, the JLN-740 is rebooted. It takes about 1 minute.

When update is completed successfully, the “Update completed.” dialogue is displayed.


The update of CQD-7040 has been completed.

4-29 4 Program Update


4.6 Updating an application of CDF-7400 (JLN-740)

This section describes the procedure for updating an application of a signal processor IF circuit
(CDF-7400) by using the remote update function.

The procedure for updating the application by directly connecting a personal computer with the signal
processor without using the remote update function is described in Section 4.12.

4.6.1 Connecting a maintenance PC to the equipment


See below for the conceptual diagram for connecting a maintenance PC tool and the equipment. For
the details of the connection procedure, refer to the Installation Manual.

Main Display Main Display

LAN Cable Maintenance PC


Distribution Distribution
Processor Processor

Power Cable Power Cable

Signal Processor Signal Processor

Coaxial Cable Coaxial Cable

Transducer Transducer

4.6.2 Setting a network


For the details, refer to Section 4.2.

In this manual, the explanation is provided based on the


assumption that the IP address of the maintenance PC is set to
“172.16.60.200”. Replace the IP address according to the personal
computer that is used.

4 Program Update 4-30


4.6.3 Starting the CYC-762 update tool

4.6.4 Inputting the interface setting


To use the remote update function in JLN-740, select in the “Interface” combo box and select
in the pull-down menu.
In this case, check that “PC IP” and “Mask” match the network setting (Section 4.2) of the
maintenance PC.

When the maintenance PC has multiple LAN adapters, select the adapter of the maintenance PC that
is connected to the equipment from the “Adapter” pull-down menu.

After completing the setting up to this stage, click on . If the setting is correct, the dialogue
that is shown in Figure 4.6-1 is displayed and changes to .

Figure 4.6-1 Dialogue displayed after the correct setting

4-31 4 Program Update


4.6.5 Selecting equipment to be updated
Select in the “Equipment” pull-down menu and check
that the IP address (initial value: 172.16.60.124) of the equipment is displayed in “Equipment IP”.

When the IP address of JLN-740 has been changed due to the


installation condition, enter the IP address after the change.

When is not
displayed in the “Equipment” pull-down menu, the tool may not be
the JLN-740 update tool. Restart the update tool for JLN-740 as
indicated in Figure 4.1-1.

4 Program Update 4-32


4.6.6 Acquiring a software list from the equipment
After inputting the IP address of the equipment, acquire a list of the software to be updated in the
equipment by clicking on . When the software list is acquired correctly, the dialogue that is
shown in Figure 4.6-2 is displayed.

Figure 4.6-2 Dialogue displayed after correct acquisition of a list

4-33 4 Program Update


After correct acquisition of a software list, the target software names and the versions are displayed in
a list. If the software to be updated does not exist in the equipment in such a case where optional
equipment is not connected due to the installation condition, “(Ver. Unknown)” is displayed.

The version number of the screen capture is the one currently


under development. Since the version number may be different
from the one that is currently operating in the product, replace the
number as required.

Figure 4.6-3 List of software targeted for update (example)

4 Program Update 4-34


4.6.7 Selecting an update file
After selecting in the “Software” pull-down menu, click
on and select an update file.
When a correct update file is selected, the information of the selected file is displayed in “Update file”,
“Version”, and “Information”, and is enabled.

4-35 4 Program Update


4.6.8 Updating
Click on to start updating.
Update processing progresses in the order of Reboot→Transmission→Update→Reboot and the
progress status is displayed on the status bar at the bottom of the window.

Since internal preparation is required prior to the updating of the software in the signal processor, a
wait time of about 10 seconds is necessary. In the update tool, the display indicates as if the signal
processor is rebooting during the wait time. However, the processor is not rebooting.

4 Program Update 4-36


Initially, the update file that is specified in the update tool is transmitted to the equipment. It takes
about 3 minutes.

4-37 4 Program Update


In this case, the “Updating software” dialogue is displayed on the main display.

After completion of the transmission, the flash ROM is rewritten. It takes about 1 minute.
In this case, the “Updating software” dialogue remains displayed on the main display.

4 Program Update 4-38


After the flash ROM has been rewritten, the JLN-740 is rebooted. It takes about 1 minute.

When update is completed successfully, the “Update completed.” dialogue is displayed.


The update of CDF-7400 has been completed.

4-39 4 Program Update


4.7 Updating an application of CDC-7400 (JLN-740)

This section describes the procedure for updating an application of a signal processor processing
circuit (CDF-7400) by using the remote update function.

The procedure for updating the application by directly connecting a personal computer with the signal
processor without using the remote update function is described in Section 4.13.

4.7.1 Connecting a maintenance PC to the equipment


See below for the conceptual diagram for connecting a maintenance PC tool and the equipment. For
the details of the connection procedure, refer to the Installation Manual.

Main Display Main Display

LAN Cable Maintenance PC


Distribution Distribution
Processor Processor

Power Cable Power Cable

Signal Processor Signal Processor

Coaxial Cable Coaxial Cable

Transducer Transducer

4.7.2 Setting a network


For the details, refer to Section 4.2.

In this manual, the explanation is provided based on the


assumption that the IP address of the maintenance PC is set to
“172.16.60.200”. Replace the IP address according to the personal
computer that is used.

4 Program Update 4-40


4.7.3 Starting the CYC-762 update tool

4.7.4 Inputting the interface setting


To use the remote update function in JLN-740, select in the “Interface” combo box and select
in the pull-down menu.
In this case, check that “PC IP” and “Mask” match the network setting (Section 4.2) of the
maintenance PC.

When the maintenance PC has multiple LAN adapters, select the adapter of the maintenance PC that
is connected to the equipment from the “Adapter” pull-down menu.

After completing the setting up to this stage, click on . If the setting is correct, the dialogue
that is shown in Figure 4.7-1 is displayed and changes to .

Figure 4.7-1 Dialogue displayed after the correct setting

4-41 4 Program Update


4.7.5 Select the equipment to be updated
Select in the “Equipment” pull-down menu and check
that the IP address (initial value: 172.16.60.124) of the equipment is displayed in “Equipment IP”.

When the IP address of JLN-740 has been changed due to the


installation condition, enter the IP address after the change.

When is not
displayed in the “Equipment” pull-down menu, the tool may not be
the JLN-740 update tool. Restart the dedicated update tool for
JLN-740 as indicated in Figure 4.1-1.

4 Program Update 4-42


4.7.6 Acquiring a software list from the equipment
After inputting the IP address of the equipment, acquire a list of software to be updated in the
equipment by clicking on . When the software list is acquired correctly, the dialogue that is
shown in Figure 4.7-2 is displayed.

Figure 4.7-2 Dialogue displayed after the correct acquisition of a list

4-43 4 Program Update


After correct acquisition of a software list, the target software names and the versions are displayed in
a list. If the software to be updated does not exist in the equipment in such a case where optional
equipment is not connected due to the installation condition, “(Ver. Unknown)” is displayed.

The version number of the screen capture is the one currently


under development. Since the version number may be different
from the one that is currently operating in the product, replace the
number as required.

Figure 4.7-3 List of software targeted for update (example)

4 Program Update 4-44


4.7.7 Selecting an update file
After selecting in the “Software” pull-down menu, click
on and select an update file.
When a correct update file is selected, the information of the selected file is displayed in “Update file”,
“Version”, and “Information”, and is enabled.

4-45 4 Program Update


4.7.8 Updating
Click on to start updating.
Update processing progresses in the order of Reboot→Transmission→Update→Reboot and the
progress status is displayed on the status bar at the bottom of the window.

Since internal preparation is required prior to the updating of the software in the signal processor, a
wait time of about 10 seconds is necessary. In the update tool, this wait time is displayed in the update
tool, the display indicates as if the signal processor is rebooting during the wait time. However, the
processor is not rebooting.

4 Program Update 4-46


Initially, the update file that is specified in the update tool is transmitted to the equipment. It takes
about 1 minute.

4-47 4 Program Update


In this case, the “Updating software” dialogue is displayed on the main display.

After completion of the transmission, the flash ROM is rewritten. It takes about 1 minute.
In this case, the “Updating software” dialogue remains displayed on the main display.

4 Program Update 4-48


After the flash ROM has been rewritten, the JLN-740 is rebooted. It takes about 1 minute.

When update is completed successfully, the “Update completed.” dialogue is displayed.


The update of CDC-7400 has been completed.

4-49 4 Program Update


4.8 Updating FPGA data of CDC-7400 (JLN-740)

This section describes the procedure for updating FPGA data of a signal processor processing circuit
(CDC-7400) by using the remote update function.

The procedure for updating FPGA data by directly connecting a personal computer with the signal
processor without using the remote update function is described in Section 4.14.

4.8.1 Connecting a maintenance PC to the equipment


See below for the conceptual diagram for connecting a maintenance PC tool and the equipment. For
the details of the connection procedure, refer to the Installation Manual.

Main Display Main Display

LAN Cable Maintenance PC


Distribution Distribution
Processor Processor

Power Cable Power Cable

Signal Processor Signal Processor

Coaxial Cable Coaxial Cable

Transducer Transducer

4.8.2 Setting a network


For the details, refer to Section 4.2.

In this manual, the explanation is provided based on the


assumption that the IP address of the maintenance PC is set to
“172.16.60.200”. Replace the IP address according to the personal
computer that is used.

4 Program Update 4-50


4.8.3 Starting the CYC-762 update tool

4.8.4 Inputting the interface setting


To use the remote update function in JLN-740, select in the “Interface” combo box and select
in the pull-down menu.
In this case, check that “PC IP” and “Mask” match the network setting (Section 4.2) of the
maintenance PC.

When the maintenance PC has multiple LAN adapters, select the adapter of the maintenance PC that
is connected to the equipment from the “Adapter” pull-down menu.

After completing the setting up to this stage, click on . If the setting is correct, the dialogue
that is shown in Figure 4.8-1 is displayed and changes to .

Figure 4.8-1 Dialogue displayed after the correct setting

4-51 4 Program Update


4.8.5 Selecting the equipment to be updated
Select in the “Equipment” pull-down menu and check
that the IP address (initial value: 172.16.60.124) of the equipment is displayed in “Equipment IP”.

When the IP address of JLN-740 has been changed due to the


installation condition, enter the IP address after the change.

When is not
displayed in the “Equipment” pull-down menu, the tool may not be
the JLN-740 update tool. Restart the dedicated update tool for
JLN-740 as indicated in Figure 4.1-1.

4 Program Update 4-52


4.8.6 Acquiring a software list from the equipment
After inputting the IP address of the equipment, acquire a list of software to be updated in the
equipment by clicking on . When the software list is acquired correctly, the dialogue that is
shown in Figure 4.8-2 is displayed.

Figure 4.8-2 Dialogue displayed after the correct acquisition of a list

4-53 4 Program Update


After correct acquisition of a software list, the target software names and the versions are displayed in
a list. If the software to be updated does not exist in the equipment in such a case where optional
equipment is not connected due to the installation condition, “(Ver. Unknown)” is displayed.

The version number of the screen capture is the one currently


under development. Since the version number may be different
from the one that is currently operating in the product, replace the
number as required.

Figure 4.8-3 List of software targeted for update (example)

4 Program Update 4-54


4.8.7 Selecting an update file
After selecting in the “Software” pull-down menu, click
on and select an update file.
When a correct update file is selected, the information of the selected file is displayed in “Update file”,
“Version”, and “Information”, and is enabled.

4-55 4 Program Update


4.8.8 Updating
Click on to start updating.
Update processing progresses in the order of Reboot→Transmission→Update→Reboot and the
progress status is displayed on the status bar at the bottom of the window.

Since internal preparation is required prior to the updating of the FPGA data in the signal processor, a
wait time of about 10 seconds is necessary. In the update tool, the display indicates as if the signal
processor is rebooting during the wait time. However, the processor is not rebooting.

4 Program Update 4-56


Initially, the update file that is specified in the update tool is transmitted to the equipment. It takes
about 20 minutes.

4-57 4 Program Update


In this case, the “Updating software” dialogue is displayed on the main display.

After completion of the transmission, the flash ROM is rewritten. It takes about 15 minutes.
In this case, the “Updating software” dialogue remains displayed on the main display.

4 Program Update 4-58


After the flash ROM has been rewritten, the JLN-740 is rebooted. It takes about 1 minute.

When update is completed successfully, the “Update completed.” dialogue is displayed.


The update of FPGA data of CDC-7400 has been completed.

4-59 4 Program Update


4.9 Updating an application of NWZ-650SDR (JLN-740)

This section describes the procedure for updating an application of a remote display (NWZ-650SDR) by
using the remote update function.

The procedure for updating the application by using a micro SD card without using the remote update
function is described in Section 4.15.

4.9.1 Connecting a maintenance PC to the equipment


See below for the conceptual diagram for connecting a maintenance PC tool and the equipment. For
the details of the connection procedure, refer to the Installation Manual.

Main Display Main Display

Remote Display Remote Display

LAN Cable Maintenance PC


Distribution Distribution
Processor Processor

Power Cable Power Cable

Signal Processor Signal Processor

Coaxial Cable Coaxial Cable

Transducer Transducer

4.9.2 Setting a network


For the details, refer to Section 4.2.

In this manual, the explanation is provided based on the


assumption that the IP address of the maintenance PC is set to
“172.16.60.200”. Replace the IP address according to the personal
computer that is used.

4 Program Update 4-60


4.9.3 Starting the CYC-762 update tool

4.9.4 Inputting the interface setting


To use the remote update function in JLN-740, select in the “Interface” combo box and select
in the pull-down menu.
In this case, check that “PC IP” and “Mask” match the network setting (Section 4.2) of the
maintenance PC.

When the maintenance PC has multiple LAN adapters, select the adapter of the maintenance PC that
is connected to the equipment from the “Adapter” pull-down menu.

After completing the setting up to this stage, click on . If the setting is correct, the dialogue
that is shown in Figure 4.9-1 is displayed and changes to .

Figure 4.9-1 Dialogue displayed after the correct setting

4-61 4 Program Update


4.9.5 Selecting the equipment to be updated
Select in the “Equipment” pull-down menu and check
that the IP address (initial value: 172.16.60.124) of the equipment is displayed in “Equipment IP”.

When the IP address of JLN-740 has been changed due to the


installation condition, enter the IP address after the change.

When is not
displayed in the “Equipment” pull-down menu, the tool may not be
the JLN-740 update tool. Restart the dedicated update tool for
JLN-740 as indicated in Figure 4.1-1.

4 Program Update 4-62


4.9.6 Acquiring a software list from the equipment
After inputting the IP address of the equipment, acquire a list of software to be updated in the
equipment by clicking on . When the software list is acquired correctly, the dialogue that is
shown in Figure 4.9-2 is displayed.

Figure 4.9-2 Dialogue displayed after the correction acquisition of a list

4-63 4 Program Update


After correct acquisition of a software list, the target software names and the versions are displayed in
a list. If the software to be updated does not exist in the equipment in such a case where optional
equipment is not connected due to the installation condition, “(Ver. Unknown)” is displayed.

The version number of the screen capture is the one currently


under development. Since the version number may be different
from the one that is currently operating in the product, replace the
number as required.

Figure 4.9-3 List of software targeted for update (example)

4 Program Update 4-64


4.9.7 Selecting an update file
After selecting in the “Software” pull-down menu, click
on and select an update file.
When a correct update file is selected, the information of the selected file is displayed in “Update file”,
“Version”, and “Information”, and is enabled.

4-65 4 Program Update


4.9.8 Updating
Click on to start updating.
Update processing progresses in the order of Transmission→Update→Reboot and the progress
status is displayed on the status bar at the bottom of the window.

Initially, the update file that is specified in the update tool is transmitted to the equipment. It takes
about 20 minutes for this processing.

4 Program Update 4-66


In this case, the “Updating software” dialogue is displayed on the main display and the remote display
(NWZ-650SDR) that is in the process of update.

4-67 4 Program Update


After completion of the transmission, the flash ROM is rewritten. It takes about 1 minute.
In this case, the “Updating software” dialogue remains displayed on the main display and remote
display (NWZ-650SDR) that is in the process of update.

4 Program Update 4-68


After the flash ROM has been rewritten, NWZ-650SDR to be updated reboots and installation
processing starts in NWZ-650SDR. Subsequently, JLN-740 reboots. It takes about 10 minutes for this
processing.

When update is completed successfully, the “Update completed.” dialogue is displayed.


The update of NWZ-650SDR has been completed.

4-69 4 Program Update


4.9.9 Update file of a remote display (NWZ-650SDR)

Many types of update files (extension: *.jrc) are available for a remote display (NWZ-650SDR) and an
appropriate file must be selected according to the installation condition.

Display the menu by selecting “MENU”→ “SDME ADV”→ “MANAG NUMB” of a remote display
(NWZ-650SDR) at installation and select a file from the five types of software control number avoiding
duplication within the equipment.

Software
No. Remote control Update file name Remarks
display
number
1 10010701 NWZ650SDR_RD-1_*.jrc
2 10010901 NWZ650SDR_RD-2_*.jrc
3 NWZ-650SDR 10011101 NWZ650SDR_RD-3_*.jrc
Used in JLN-720 only.
4 10011301 NWZ650SDR_RD-4_*.jrc
Not used in JLN-740.
5 10011501 NWZ650SDR_RD-5_*.jrc
6 10011701 NWZ840SDR_RD-1_*.jrc
7 10011901 NWZ840SDR_RD-2_*.jrc
8 NWZ-840SDR 10012101 NWZ840SDR_RD-3_*.jrc
Used in JLN-720 only.
9 10012301 NWZ840SDR_RD-4_*.jrc
Not used in JLN-740.
10 10012501 NWZ840SDR_RD-5_*.jrc

* In wild card “*”, an alphanumeric character string is inserted for each version.

Software control number:

Software control number:

Software control number:

Software control number:

Figure 4.9-4 Relationship between a software list and software control numbers

4 Program Update 4-70


4.10 Updating an application of NWZ-840SDR application (JLN-740)

This section describes the procedure for updating an application of a remote display (NWZ-840SDR) by
using the remote update function.

The procedure for updating the application by using a micro SD card without using the remote update
function is described in Section 4.16.

4.10.1 Connecting a maintenance PC to the device


See below for the conceptual diagram for connecting a maintenance PC tool and the equipment. For
the details of the connection procedure, refer to the Installation Manual.

Main Display Main Display

Remote Display Remote Display

LAN Cable Maintenance PC


Distribution Distribution
Processor Processor

Power Cable Power Cable

Signal Processor Signal Processor

Coaxial Cable Coaxial Cable

Transducer Transducer

4.10.2 Setting a network


For the details, refer to Section 4.2.

In this manual, the explanation is provided based on the


assumption that the IP address of the maintenance PC is set to
“172.16.60.200”. Replace the IP address according to the personal
computer that is used.

4-71 4 Program Update


4.10.3 Starting the CYC-762 update tool

4.10.4 Inputting the interface setting


To use the remote update function in JLN-740, select in the “Interface” combo box and select
in the pull-down menu.
In this case, check that “PC IP” and “Mask” match the network setting (Section 4.2) of the
maintenance PC.

When the maintenance PC has multiple LAN adapters, select the adapter of the maintenance PC that
is connected to the equipment from the “Adapter” pull-down menu.

After completing the setting up to this stage, click on . If the setting is correct, the dialogue
that is shown in Figure 4.10-1 is displayed and changes to .

Figure 4.10-1 Dialogue displayed after the correct setting

4 Program Update 4-72


4.10.5 Selecting the equipment to be updated
Select in the “Equipment” pull-down menu and check
that the IP address (initial value: 172.16.60.124) of the equipment is displayed in “Equipment IP”.

When the IP address of JLN-740 has been changed due to the


installation condition, enter the IP address after the change.

When is not
displayed in the “Equipment” pull-down menu, the tool may not be
the JLN-740 update tool. Restart the dedicated update tool for
JLN-740 as indicated in Figure 4.1-1.

4-73 4 Program Update


4.10.6 Acquiring a software list from the equipment
After inputting the IP address of the equipment, acquire a list of software to be updated in the
equipment by clicking on . When the software list is acquired correctly, the dialogue that is
shown in Figure 4.10-2 is displayed.

Figure 4.10-2 Dialogue displayed at the correct acquisition of a list

4 Program Update 4-74


After correct acquisition of a software list, the target software names and the versions are displayed in
a list. If the software to be updated does not exist in the equipment in such a case where optional
equipment is not connected due to the installation condition, “(Ver. Unknown)” is displayed.

The version number of the screen capture is the one currently


under development. Since the version number may be different
from the one that is currently operating in the product, replace the
number as required.

Figure 4.10-3 List of software targeted for update (example)

4-75 4 Program Update


4.10.7 Selecting an update file
After selecting in the “Software” pull-down menu, click
on and select an update file.
When a correct update file is selected, the information of the selected file is displayed in “Update file”,
“Version”, and “Information”, and is enabled.

4 Program Update 4-76


4.10.8 Updating
Click on to start updating.
Update processing progresses in the order of Transmission→Update→Reboot and the progress
status is displayed on the status bar at the bottom of the window.

Initially, the update file that is specified in the update tool is transmitted to the equipment. It takes
about 20 minutes for this processing.

4-77 4 Program Update


In this case, the “Updating software” dialogue is displayed on the main display and the remote display
(NWZ-840SDR) that is in the process of update.

4 Program Update 4-78


After completion of the transmission, the flash ROM is rewritten. It takes about 1 minute.
In this case, the “Updating software” dialogue remains displayed on the main display and remote
display (NWZ-840SDR) that is in the process of update.

4-79 4 Program Update


After the flash ROM has been rewritten, NWZ-840SDR to be updated reboots and installation
processing starts in NWZ-650SDR. Subsequently, JLN-740 reboots. It takes about 10 minutes for this
processing.

When update is completed successfully, the “Update completed.” dialogue is displayed.


The update of NWZ-840SDR has been completed.

4 Program Update 4-80


4.10.9 Update file of a remote display (NWZ-840SDR)

Many types of update files (extension: *.jrc) are available for a remote display
(NWZ-650SDR/840SDR) and an appropriate file must be selected according to the installation
condition.

Display the menu by selecting “MENU”→ “SDME ADV”→ “MANAG NUMB” of a remote display
(NWZ-650SDR/840SDR) at installation and select a file from the five types of software control number
avoiding duplication within the equipment.

Software
Remote
No. control Update file name Remarks
display
number
1 10010701 NWZ650SDR_RD-1_*.jrc
2 10010901 NWZ650SDR_RD-2_*.jrc
3 NWZ-650SDR 10011101 NWZ650SDR_RD-3_*.jrc
Used in JLN-720 only.
4 10011301 NWZ650SDR_RD-4_*.jrc
Not used in JLN-740.
5 10011501 NWZ650SDR_RD-5_*.jrc
6 10011701 NWZ840SDR_RD-1_*.jrc
7 10011901 NWZ840SDR_RD-2_*.jrc
8 NWZ-840SDR 10012101 NWZ840SDR_RD-3_*.jrc
Used in JLN-720 only.
9 10012301 NWZ840SDR_RD-4_*.jrc
Not used in JLN-740.
10 10012501 NWZ840SDR_RD-5_*.jrc

* In wild card “*”, an alphanumeric character string is inserted for each version.

Software control number:

Software control number:

Software control number:

Software control number:

Figure 4.10-4 Relationship between a software list and software control numbers

4-81 4 Program Update


4.11 Updating an application of NWZ-510SDW (JLN-741)

This section describes the procedure for updating an application of a main display (NWZ-510SDW) by
using micro SD cards.

JLN-741 does not have a remote update function.

4.11.1 Checking the equipment configuration


When updating an application of a main display (NWZ-510SDW) by using micro SD cards, it is not
necessary to change the equipment configuration.

Prepare two micro SD cards. Refer to Sections 4.11.2 and 4.11.3 for rewriting the internal data of the
two micro SD cards.

Main Display

Micro SD card × 2

Signal Processor

Transducer

It is not necessary to change the equipment configuration even if


an application of a JLN-740 main display (NWZ-510SDW) is to be
updated by using micro SD cards.

4 Program Update 4-82

You might also like