AMETEK 241CE II Hydrocarbon Dew Point Analyzer2
AMETEK 241CE II Hydrocarbon Dew Point Analyzer2
®
WESTERN RESEARCH
Model 241CE II Hydrocarbon Dew Point
Analyzer
maintenance analyzer for automated additional remote controllers, or costly
Superior Benefits direct measurement of hydrocarbon sample systems. It carries CSA and
zx Accurate and objective direct dew point temperature in natural gas ATEX certification for use in hazardous
measurement of hydrocarbon streams, using a patented chilled areas using an EEx d (flameproof) type
dew point temperature at pipeline mirror design. enclosure. Sample gas flows through
pressure or Cricondentherm a measuring cell mounted on the
zx Proprietary three-stage sample Direct Measurement outside of the flameproof electronics
filter provides superior The Model 241CE II Analyzer uses a enclosure which simplifies sample
protection thermoelectric cooler to control the system maintenance.
zx Designed for unattended temperature of a two surface mirror. A fully integrated sample system
operation with low maintenance During each measuring cycle the package, including proprietary
zx Electricity the only consumable microcontroller cools the mirror at a multiple stage filtration specifically
controlled rate until the optics have designed for natural gas samples,
zx Compact design allows mounting
in optional enclosures detected changes in light intensity that protects the analyzer from common
occur when liquids condense on the natural gas contaminants (aerosols,
two surfaces of the mirror. Differences particulates and liquid slugs) and
in surface tensions between liquid ensures reliable, fast, and accurate
The Need hydrocarbons and other condensables means of detecting the hydrocarbon
Natural gas producers, distributors, enable the analyzer to reject non- dew point temperature in natural gas
and industrial consumers understand hydrocarbon dew points. This applications.
the importance of ensuring that ability along with sensitive optics
natural gas meets various strict and a precisely
quality specifications which typically controlled cooling
include a maximum hydrocarbon dew rate provides
point temperature. Even relatively improved accuracy
small quantities of liquids in natural and repeatability
gas transmission lines can damage compared to a
sensitive metering equipment, plug manual chilled
lines, interfere with odorants, and mirror type device
most importantly, create unsafe which depends
conditions. Liquid drop out is not only on the training
an operational and safety concern, of the operator.
but may also result in errors in other The temperature
critical measurement equipment. of the mirror in
Hydrocarbon liquids in natural gas the 241CE II is
delivered to end-users can cause measured using
damage to gas turbines, upset a Platinum RTD
plants using natural gas as chemical which provides
feedstock, increase emissions from higher accuracy
burners, and compromise safety. and long term
stability compared
In many applications, automated
to thermocouples.
on-line analyzers, such as the Western
Research 241CE II Hydrocarbon Dew The Model
Point Analyzer, are replacing manual 241CE II is
chilled mirror devices or calculation designed for
methods, which have traditionally unattended
been used to determine hydrocarbon operation, in
dew point temperature. For over 15 a complete
years, AMETEK Western Research package, without
has provided an accurate and low- the need for
Model 241CE II Hydrocarbon Dew Point
Analyzer - European version shown
www.ametekpi.com
Model 241CE II Hydrocarbon Dew Point Analyzer
WESTERN RESEARCH
Performance Specifications Applications
Methodology: Chilled-mirror Highest Measurable Dewpoint: zx Pipeline quality and custody
Application dependent, typically transfer
Accuracy: Hydrocarbon dew point
15°C (27°F) below the temperature zx Monitor inlet and outlet of storage
temperature ±1°C (±2°F) facilities
at the analyzer installation
Sensitivity: Temperature ±0.1°C (±0.2°F) zx Control of contractual blending
Maximum Working Pressure: 13.79 operations
Sample Transport: 1/4 inch stainless
MPa (2000 psi) zx Monitor amount of superheat in
steel instrument tubing recommended gas turbine feed gas
(heat tracing may be required) Power: 120 VAC ±10%, 60 Hz or 240
zx Optimizing of switch times for
±10% VAC, 50 Hz
Typical Flow: 1 to 5L/min. (2 to 10 pressure swing adsorption
scfh) depending on sample system Power Consumption: less than 275 W
configuration and pressure Outputs: 4-to-20 mA (isolated, loop- Standard Features
Ambient Temperature: 10°C to 40°C power by customer or optional Outputs
(50° to 104°F) self-powered) zx Hydrocarbon dew point
Physical Dimensions (H x W x D): temperature current output
Cooling Capability: Typically 60°C
zx Digital communications via
(108°F) below the temperature at 838 x 571.5 x 289 mm (33 x 22.5 x Modbus RTU protocol
the analyzer installation. The cooling 11.37 in.)
capability of the analyzer is affected
Sample System
Weight: 45 kg (100 lbs.)
by the ambient temperature, zx Fully integrated sample system,
Approvals and Certifications: including three-stage natural gas
sample gas composition and the Class I, Division 1, Groups C & D filter block and flow control
pressure. (Consult your AMETEK ATEX II 2 G EExdIIBT4 zx Optional regulator for measurement
representative for evaluation of your Russian Ex Proof Certification; at Cricondentherm
application.) 1Exd llB T3 X
CE Compliance: Complies with all User Interface
relevant European Directives
zx On-board microcontroller, display,
and user interface
571.5mm [22.5”]
527mm [20.75”] zx Optional remote start of measuring
cycle available via external input
22mm [0.87”]
267mm [10.5”]
APPROX 574mm [22.6”] DOOR SWING
838mm [33”]
686mm [27”]
HEATSINK
ASSY