Catlog - Model-SE-LSC
Catlog - Model-SE-LSC
Preface
Main Features:
Applications: Weight/Force Display, Peak Value Detection&Display Holding, Setpoint DO
Output and AO/Digit Transmission.
EMC design with high anti-jamming capability, suitable for industrial environment.
32-bit ARM CPU with 48MHz clock & high arithmetic speed.
6+6 Red LED digital tubes for English character and digit display.
4-key English keypad for Menu&Shortcut mode operation.
24-bit High-precision and high-speed ∑-△A/D conversion module with 1/1,000,000 internal
resolution and max. sampling frequency 3200Hz.
Special anti-vibration digital filtering algorithm for ensuring the weighing stability and accuracy
when there is strong vibration on the load receptor, and the rapid response capability when the
weight signal changes.
Max. Connection Quantity: 8 Loadcells (350Ω).
Zero Upon Power Up, Zero Auto-tracking, Manual/Auto Zero and Zero Calibration available.
Load Calibration, Data Calibration, Segmenting Span Correction and Segmenting Weight
Calculation available.
Auto-locking, Key-locking, Key-unlocking, Digital Setting&Calibration and I/O Testing
functions available.
3 Definable normally open switch inputs [DI] and 3 definable normally open transistor
switch outputs [DO].
1 Optional&definable current signal output [AO: 0~20mA].
1 Optional&definable voltage signal output [VO: 0~10V].
RS232, optional RS485 and CANBUS communication ports for connecting Host IPC/PLC
and LED Remote Display.
With the multitasking mode, the weighing/force measuring process, alarm output and signal
transmission will not be interrupted by parameter setting and the other operations.
1
Contents
1. SAFETY PRECAUTION .................................................................................................................3
2. TECHNICAL SPECIFICATIONS ....................................................................................................... 4
3. USER INTERFACE...................................................................................................................................... 6
3.1 USER INTERFACE DIAGRAM .............................................................................................................................................. 6
3.2 STATE INDICATION.......................................................................................................................................................................... 6
3.3 KEYPAD OPERATION...................................................................................................................................................................... 7
3.4 ALARM SIGNS ........................................................................................................................................................................ 7
4. INSTALLATION&CONNECTION..................................................................................................... 8
4.1 INSTALLATION ..............................................................................................................................................................................................8
4.2 TERMINAL............................................................................................................................................................................... 9
5. OPERATION PROCEDURE......................................................................................................... 11
6. FUNCTIONS&OPERATION ............................................................................................................. 12
6.1 MAIN DISPLAY INTERFACES ........................................................................................................................................... 12
6.1.1 Weight Value and Percent [L] .................................................................................................... 12
6.1.2 Weight Value, Peak Hold Value [P] ........................................................................................... 12
6.1.3 Weight, AO Output Value [Ao].................................................................................................. 12
6.1.4 Weight, VO Output Value [Vo] ................................................................................................................... 13
6.1.5 Weight, Communication State [Co] ........................................................................................... 13
6.2 MAIN MENU......................................................................................................................................................................... 14
6.3 F1-SET PARAMETER SETTING ....................................................................................................................................... 16
6.3.1 Weighing Parameters (SCAL) ................................................................................................... 16
6.3.2 Calibration Parameters (CALP)................................................................................................. 18
6.3.3 Setpoint Parameters (SEtP) ....................................................................................................... 20
6.3.4 Communication Parameters (SErP) ........................................................................................... 22
6.3.5 Display Parameters (dISP)......................................................................................................... 23
6.3.6 A Sample of Parameter Setting .................................................................................................. 24
6.4 F2-CAL SYSTEM CALIBRATION............................................................................................................................................. 25
6.4.1 Zero Calibration (ZEro)............................................................................................................. 25
6.4.2 Data Calibration (dAtA) ............................................................................................................ 26
6.4.3 Load Calibration (LoAd) ........................................................................................................... 28
6.4.4 Segmenting Span Correction (SEgC) ........................................................................................ 29
6.5 F5-LOC KEY-LOCKER................................................................................................................................................................ 30
6.5.1 Key-unlocking (oPEn) ............................................................................................................... 30
6.5.2 Key-locking (Locc) ................................................................................................................... 30
6.5.3 Password Set (PASS).............................................................................................................................. 31
2
1. Safety Precaution
Application environment
Make sure that this product works under the environment where is accord with the technical
specifications.
Do not open the shell before power-off.
Cable Laying
Weighing signal, analog signal and communication signal cables should be laid in pipes, and
do not lay them together with power cables.
Power Supply
The power supply of the controller should be separated from the power supply of the driving
devices.
Please make sure that the inputted voltage is correct before power-on.
Environmental Protection
While being discarded as worthless, the product should be processed lawfully as leady
industrial waste for environment protection.
Other Notes
The installation, wiring and maintenance should be operated by the engineers with the
relevant professional knowledge and safety operation ability.
Although being not described in this manual, the relevant safety operating procedures and
standards should be followed.
3
2. Technical Specifications
Executing Standards
PRC GB/T 7724-2008《National Standard for Electronic Weighing Meter》.
PRC JJG 649-2016《National Verification Regulation for Digital Weighing Indicators
(WeighingIndicators)》.
OMIL R76-1: 2006《Non-automatic Weighing Instruments》International
Recommendation.
Accuracy Grade:○
Ⅲ.
Number of Verification Scale Interval: n=3000.
Display
6+6 Red LED digital tubes for English character and digit display.
Weight Display Range: -9,999~+99,999.
Scale Capacity: Setting Range 1~99,999.
Scale Division: Optional 1, 2, 5, 10, 20, 50, 100, 200, 500.
Display Resolution: 1/50,000.
Decimal Point Position: Optional 0, 0.0, 0.00, 0.000, 0.0000.
Display Refreshing Time: Setting Range 0.01~1.00s.
Keypad
4-key English keypad for Menu&Shortcut mode operation.
4
COM3: Optional CANBUS.
Connectable: Host IPC/PLC and LED Remote Display.
Operating Specification
Operating Voltage: DC12-24V.
Max. Power Consumption: 3W.
Outline Size: 71.5×113×57.9mm [W×H×D ].
Guide rail type installation.
Guide Rail Groove Size: 35×1.6 mm [W×H×D ].
Operating Temperature: -25℃ to +45℃.
Storage Temperature: -30℃ to +60℃.
Relative Humidity: Max. 85%RH.
Weight: Approx. 167g.
5
3. User Interface
12 11 10 9 8 7 6 5 4 3 2 1
GNDmA+ Vo+ GNDRXD TXD A+ B- H L DC+ DC-
AH
AL
RUN
■ ► ▲
LOADCELL DI/DO
S- S+ E- E+ SHD DI1 DI2 DI3 DO1 DO2 DO3COM
1 2 3 4 5 6 7 8 9 10 11 12
6
3.3 Keypad Operation
If there is not any keypad operation in one minute and it’s not in the processes of ‘F2 System Calibration’ & ‘F6
Factory Adjustment’, the controller will return to ‘Main Display Interface’ automatically.
7
Preheating Time Countdown Refer to parameter [128] ‘Zero Upon Power Up’, [129]
tXX.XX
[min.sec]. ‘Power Up Zero Time’ and [130] ‘Power Up Zero Limit’.
oV-Zr Over ‘Power Up Zero Limit’. Wait for the preheating time over or press any key to exit.
8
4. Installation&Connection
4.1 Installation
71.5 57.9
113
35
Plastic Fastener
1.6
9
4.2 Terminal
LOADCELL DI/DO
S- S+ E- E+ SHD DI1 DI2 DI3 DO1 DO2 DO3COM
1 2 3 4 5 6 7 8 9 10 11 12
3 E- Excitation Voltage -.
10
12 11 10 9 8 7 6 5 4 3 2 1
GNDmA+ Vo+ GNDRXD TXD A+ B- H L DC+ DC-
1 DC- DC Input -.
2 DC+ DC Input +.
3 L Data -.
4 H Data +.
5 B- Data -.
6 A+ Data +.
10
5. Operation Procedure
Scale Setting
Zero Calibration
[Data Calibration]
Load Calibration
[Segmenting Correction]
Setpoint Setting
Other Settings
Key-locking
11
6. Functions&Operation
AH
AL
RUN
AH
AL
RUN
AH
AL
RUN
AO: 0.00~20.00mA.
12
6.1.4 Weight, VO Output Value [Vo]
AH
AL
RUN
VO: 0.00~10.00V.
AH
AL
RUN
13
6.2 Main Menu
14
Main Menu Second Menu
Special for manufacturer.
【mV】Exfactory Span Adjustment: Use standard weighing
test equipment to adjust the weighing signal interface for
-SPAn-
normalizing
Span Coefficient to 1.
-AdtS- Weighing Signal Linearity Test.
Factory AO [0~20mA] Zero/Full Adjustment.
F6.FAC -AVZF-
Adjustment VO [0~10V] Zero/Full Adjustment.
AO [0~20mA] Linearity Test.
-AVtS-
VO [0~10V] Linearity Test.
-dIdo- DI/DO Input&Output Test.
-dEFU- RAM Reset: Reset to factory defaults.
-dStS- Display/DO Reliability Test.
Only for query.
--VEr- Version No.
Product
F7.InF --Sn-- Serial No.
Information
-dAtE- Exfactory Date.
-HArd- Hardware Configuration.
15
6.3 F1-SET Parameter Setting
Scale Division
103 dV 1~500 1
1, 2, 5, 10, 20, 50, 100, 200, 500
-99999~ 0 Zero Value
104 ZE
+99999 [*] Only for query.
Span Coefficient
1.0000
105 SP >0 Max. display value: 9.9999.
[*]
Only for query.
Stability Range [Division]
Set value = 0: No ‘Stability Judging’.
106 Sr 0~5 0 Set value > 0: Weight Variance per [107] ‘Stability
Time’ being in [106] ‘Stability Range’ means
‘Weight is stable’.
107 St 0.1~9.9 0.5 Stability Time [s]
16
No. Sign Range Default Description Set
Sampling Frequency [Hz]
0: 5 Hz
1: 200 Hz
108 Fc 0~5 0 2: 400 Hz
3: 800 Hz
4: 1600Hz
5: 3200Hz
Anti-vibration Digital Filter1
Set Value Cutoff Frequency
0 None
1 11.2Hz
2 8.0Hz
3 5.6Hz
109 Ft 0~9 5 4 4.0Hz
5 2.8Hz
6 2.0Hz
7 1.4Hz
8 1.0Hz
9 0.7Hz
Weight Display Smooth Filter2
The bigger set value of Filter2 will make the weight
110 Sb 1~128 1
display value more stable with slower response
speed.
Weight Display Multiplier
0.01~
111 EL 1 It’s for enlarging or lessening the weight dispay
9999.99
value.
17
6.3.2 Calibration Parameters (CALP)
0: oFF
123 nZ 0~50000 50 changing is within this range, Manual Zero and Auto
127 Un 0 0 Unused
18
No. Sign Range Default Description Set
131 L1 1000 Correction Point’s Loading Weight 1~10
132 L2 2000 Inputted Loading Weight value for Segmenting
133 L3 3000 Correction.
134 L4 4000 Demand: L1≤L2≤…≤L10.
135 L5 1~99999 5000 Only for query.
136 L6 [*] 6000
137 L7 7000
138 L8 8000
139 L9 9000
140 LA[L10] 10000
141 d1 10000 Correction Point’s AD Value 1~10
142 d2 20000 AD Value detected via Segmenting Correction.
143 d3 30000 Demand: d1≤d2≤…≤d10.
Only for query.
144 d4 40000
-99999~
145 d5 50000
+99999
146 d6 60000
[*]
147 d7 70000
148 d8 80000
149 d9 90000
150 dA[d10] 99999
151 C1 Span Correction Ratio 1~10
152 C2 Span Correction Ratio of Linearity Segment got by
153 C3 Segmenting Correction.
154 C4 Linearity Segment 1:[Zero Value]~d1.
0.0100~
155 C5 Linearity Segment 2:d1~d2.
99.9999 1.0000
156 C6 Linearity Segment 3:d2~d3.
[*]
157 C7 …
158 C8 Linearity Segment 10:d9~[Max. AD
Value].Max. display value: 9.9999.
159 C9
Only for query.
160 CA[C10]
Segmenting Weight Calculation Permission
0 0: oFF
161 SC 0~1
[*] 1: on
Operating Authorization: Administrator.
19
6.3.3 Setpoint Parameters (SEtP)
20
No. Sign Range Default Description Set
212 Un 0 0 Unused
0. ≡ZEro [Manual Zero without Power-down
Protection]
213 dI1 0 1. -ZEro [Zero Calibration with Power-down
0~2
214 dI2 1 Protection]
[*]
215 dI3 2 2. -CLS- [Clear the display value of Peak Hold
Value]
Operating Authorization: Administrator.
DO1/DO2/DO3 Signal
0. ZEro [Weight ≤ Non-load Zero Range]
1. L1 [203]<P1
2. 1≡2 P1≤[203]≤P2
3. b2 [203]>P2
4. LE1 [203]≤P1
5. 1-2 P1<[203]<P2
6. b2 [203]≥P2
7. b1 [203]>P1
216 do1 1
0~10 8. bE1 [203]≥P1
217 do2 2 9. L2 [203]<P2
[*]
218 do3 3 10. LE2 [203]≤P2
Note:
‘L’: Less than; ‘LE’: Less than or Equal to.
‘b’: Bigger than; ‘bE’: Bigger than or Equal to.
‘≡’: Closed interval; ‘-’: Open interval.
Operating Authorization: Administrator.
21
6.3.4 Communication Parameters (SErP)
22
6.3.5 Display Parameters (dISP)
Auto-Locking
0: oFF
902 Lc 0~1 0 minute and it’s not in the processes of ‘F2 System
23
6.3.6 A Sample of Parameter Setting
Modify the parameter ‘[102] Scale Capacity’.
【■】+【►】【▲】: F1-SEt
【 】+【►】【▲】
【 】+【►】【▲】
【 】
24
6.4 F2-CAL System Calibration
After doing ‘System Calibration’, Tare Weight value will return to zero automatically.
【■】+【►】【▲】: F2-CAL
M 【 】+【►】【▲】: -ZEro-
【 】
25
6.4.2 Data Calibration (dAtA)
Input the specification parameter values of loadcell [Total Capacity and Output Sensitivity] according
to the actual configuration of the weighing system to correct Span Coefficient. If there is no access to get the
specification parameter values for Data Calibration, then it’s necessary to do Load Calibration.
【■】+【►】【▲】: F2-CAL
M 【 】+【►】【▲】: -dAtA-
【 】
【 】
【 】
【 】
26
【►】【▲】:
105: New Span Coefficient value.
oLd: Original Span Coefficient value.
Sr: Span Correction Ratio = New Value / Original Value.
Note:
Total Capacity of Loadcells = Loadcell Capacity × Quantity.
Voltage Ratio = Excitation Voltage on the terminal of controller / Excitation Voltage on the terminal of
loadcells.
The rated excitation voltage for loadcells is DC9V. It’s best to measure the actual voltage value.
4-wire connection: The voltage attenuation is big, the voltage on both sides should be measured.
27
6.4.3 Load Calibration (LoAd)
After doing Data Calibration, if there are conditions for Load Calibration, do Load Calibration with
loading standard weight on the weigher to correct Span Coefficient further for higher weighing accuracy.
The loading weight should be bigger than 50% of Scale Capacity value.
【■】+【►】【▲】: F2-CAL
M 【 】+【►】【▲】: -LoAd-
【 】
【 】
【►】【▲】:
105: New Span Coefficient value.
oLd: Original Span Coefficient value.
Sr: Span Correction Ratio = New Value / Original Value.
Ad: AD Value.
If AD Value ≤ Zero Value, display ‘Err’, and it’s not allowed to save
Span Coefficient.
28
6.4.4 Segmenting Span Correction (SEgC)
After doing Zero Calibration and Load Calibration [or Data Calibration], it’s allowed to do Segmenting
Span Correction with loading standard weight on the weigher by 10 Correction Points to get Span Correction
Ratio of 10 linearity segments for higher weighing accuracy.
【■】+【►】【▲】: F2-CAL
M 【 】+【►】【▲】: - SEgC -
【 】
30
6.5 F5-LOC Key-locker
【■】+【►】【▲】: F5-Loc
M 【 】+【►】【▲】: -oPEn-
【 】
【 】: If inputted password is
correct, Key-unlocking will be valid.
【■】+【►】【▲】: F5-Loc
M 【 】+【►】【▲】: -Locc-
【 】
【 】: If inputted password is
correct, Key-locking will be valid.
M
M
29
6.5.3 Password Set (PASS)
【■】+【►】【▲】: F5-Loc
M 【 】+【►】【▲】: -PASS-
【 】
【 】
【 】
32
Appendix A. Register Table of Host-Slave MODBUS[ASCII/RTU]
Command
Data Name Type Address Attribute Description
[HEX]
Long 40001 R 03
Weight Long 40003 R 03
40005.0 1: In Peak Value detection process.
40005.1
40005.2
40005.3
40005.4
40005.5
40005.6
40005.7
Running State 1 UnShort 40005.8 R 03
40005.9
40005.10
40005.11
40005.12
40005.13 1: Weight is stable.
40005.14 1: Overload Alarm.
40005.15 1: Controller Fault.
40006.0
40006.1
40006.2
40006.3
40006.4
40006.5
40006.6
40006.7
Running State 2 40006.8
UnShort R 03
[Unused]
40006.9
40006.10
40006.11
40006.12
40006.13
40006.14
40006.15
31
Command
Data Name Type Address Attribute Description
[HEX]
40007.0 1: DO1 ON; 0: DO1 OFF.
40007.1 1: DO2 ON; 0: DO2 OFF.
40007.2
40007.3
40007.4
40007.5
40007.6
40007.7
DO State UnShort 40007.8 R 03
40007.9
40007.10
40007.11
40007.12
40007.13
40007.14
40007.15
0xA500: Unused.
0xA501: Unused.
0xA502: Unused.
0xA503: Unused.
0xA504: Unused.
0xA505: Unused.
0xA506: Unused.
0xA507: Unused.
0xA508: Unused.
0xA509: Unused.
Operation UnShort 40008 W 06
0xA50A: Clear Screen.
0xA50B: Unused.
0xA50C: Clear Alarm.
0xA50D: Manual Zero.
0xA50E: Zero Calibration.
0xA50F: Load Calibration [Do Span
Calibration with the value of
‘Calibrating Weight’ as loading
weight].
0xA510~0xA5FF: Unused.
Calibrating Weight UnLong 40009 R/W 03/10
UnLong 40011 R 03
UnLong 40013 R 03
33
Command
Data Name Type Address Attribute Description
[HEX]
UnLong 40015 R 03
Non-load Zero Range UnShort 40017 R/W 03/06
Weight Setpoint 1 UnLong 40018 R/W 03/10
Weight Setpoint 2 UnLong 40020 R/W 03/10
Weight Peak Value UnLong 40022 R 03
UnLong 40024 R/W 03/10
Data Compared with 0: Real-time Weight.
UnShort 40026 R/W 03/06
Weight Setpoints 1: Weight Peak Value.
UnShort 40027 R/W 03/06
UnShort 40028 R/W 03/06
UnShort 40029 R/W 03/06
UnShort 40030 R/W 03/06
UnShort 40031 R/W 03/06
UnShort 40032 R/W 03/06
UnShort 40033 R/W 03/06
40034-
UnLong W 10
40037
UnShort 40038 W 06
0x02:【■】 (【MENU】
)
Key Code UnShort 40039 R 06
0x04:【►】
0x40:【▲】
0x80:【 】 (【ENT】)
Serial No. UnLong 40040 R 03 6 Digits.
UnShort: Unsigned Short Int; UnLong: Unsigned Long Int; Long: Signed Long Int.
34
Appendix B. Data Frame Format of Continuous Sending [ASCII]
35