0% found this document useful (0 votes)
47 views59 pages

CELLGARD ES Energy Saver Class II Laminar Flow Biological Safety Cabinet

Uploaded by

arrows25
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
47 views59 pages

CELLGARD ES Energy Saver Class II Laminar Flow Biological Safety Cabinet

Uploaded by

arrows25
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 59

CELLGARD ES Energy Saver

Class II Laminar Flow


Biological Safety Cabinet

Models
NU-477-300E/400E/500E/600E
Bench/Console

Operation & Maintenance Manual

November 2015
(Series 5 and Higher)
Revision 6

NU-477-400E/500E/600E Only

Manufactured By:
NuAire, Inc.
2100 Fernbrook Lane
Plymouth, MN 55447
Toll-Free: 1-800-328-3352
In Minnesota: (763)-553-1270
Fax: (763)-553-0459

OM0215 Page 1 of 59
Congratulations!
You have just purchased one of the finest Laminar Flow Biological Safety Cabinets available. With proper care, maintenance
(testing), and laboratory procedure, this cabinet will give you years of product and personnel protection from particulate
contaminants as prescribed in NSF/ANSI 49 and EN12469. Please read this manual carefully to familiarize yourself with proper
installation, maintenance, and operation of the cabinet.

www.hc-sc.gc.ca
www.cdc.gov/od/ohs/
www.absa.org
www.absa-canada.org
www.ebsa.be
www.inspection.gc.ca
www.who.int
www.biosafety.be
www.hse.gov.uk
www.nsf.org
www.cetainternational.org
www.osha.gov/dts/osta/
www.nuaire.com

OM0215 Page 2 of 59
November/2015
ABOUT THIS OPERATION & MAINTENANCE MANUAL

The information contained in this manual is intended to reflect our current production standard configuration model
along with the more frequently purchased options. Any unique additions/modifications/shop drawings are appended in the back
flap of this manual, along with any modifications and/or additions to procedures as outlined in this manual. A copy of the original
factory test report is also appended to this manual. In case this manual and/or test report is lost or misplaced, NuAire retains a copy
in our files. A replacement copy can be obtained by calling or writing NuAire, Inc. stating the model number and serial number and a
brief description of the information desired.

OM0215 Page 3 of 59
November/2015
CELLGARD ES Energy Saver Class II Laminar Flow Biological Safety Cabinet
Models NU-477-300E/400E/500E/600E Operation & Maintenance Manual

TABLE OF CONTENTS
Section No. 1 ..............................................................General Description
Section No. 2 ..............................................................Models & Features
Section No. 3 ..............................................................Warranty
Section No. 4 ..............................................................Shipments
Section No. 5 ..............................................................Installation Instructions
5.1 ................................................................................ Location
5.2 ................................................................................ Set-up Instructions
5.3 ................................................................................ Testing Methods and Equipment
Section No. 6 ..............................................................Operating the NU-477E
6.1 ................................................................................ Operator Controls & Indicators
6.2 ................................................................................ Operating Guidelines
6.3 ................................................................................ Operating Sequence
6.4 ................................................................................ Ergonomics
6.5 ................................................................................ Cleaning Procedures
6.6 ................................................................................ Hazardous Decontamination Procedure
Section No. 7 ..............................................................General Maintenance
7.1 ................................................................................ Decontamination
7.2 ................................................................................ Fluorescent Lamp, Bulb Replacement
7.3 ................................................................................ HEPA Filter/Motor Replacement
7.4 ................................................................................ Sliding Window Replacement & Adjustment
7.5 ................................................................................ Airflow Calibration
7.6.................................................................................. HEPA Filter Leak Test
7.7 ................................................................................ Airflow Smoke Pattern Test
7.8 ................................................................................ Cleanliness Classification Test for Pharmacy Application
7.9.................................................................................. Main Control Board Description & Replacement
Section No. 8 ..............................................................Error Indicators & Troubleshooting
Section No. 9 ..............................................................Remote Contacts
Section No. 10 ............................................................Optional Equipment
10.1 .................................................................................. Ultraviolet Lamp
Section No. 11............................................................Electrical/Environmental Requirements
Section No. 12 ...........................................................Disposal and Recycle
Insert.............................................................. ............Replacement Parts

MANUAL DRAWINGS
ACD-16006 ........................................................................ NU-477E Airflow Schematic
BCD-16007 ........................................................................ NU-477-300E Specification Drawing
BCD-16008 ........................................................................ NU-477-400E Specification Drawing
BCD-16009 ........................................................................ NU-477-500E Specification Drawing
BCD-16010 ........................................................................ NU-477-600E Specification Drawing
BCD-12634 ........................................................................ NU-477E SST Cap/Security Drain Valve Installation
BCD-12056 ........................................................................ Front Panel

ASSEMBLY DRAWINGS
BCD-05147 ........................................................................ Base Stand Assembly
BCD-05146 ........................................................................ Base Stand Storage Cabinet Assembly
BCD-11817 ........................................................................ Control Center & Front Decorative Panel Assembly
BCD-15926 ........................................................................ Sliding Window Assembly & Adjustment
BCD-11819 ........................................................................ Base Cabinet Assembly

ELECTRICAL SCHEMATICS
BCD-13276 ........................................................................ Main Control Module Schematic
BCD-14300 ........................................................................ NU-477-300E/400E/500E/600E Electrical Schematic

OM0215 Page 4 of 59
November/2015
CELLGARD ES Energy Saver Class II Laminar Flow
Biological Safety Cabinet
Models
NU-477-300E/400E/500E/600E
MANUFACTURED BY:
NuAire, Inc. - Plymouth, Minnesota, U.S.A.

1.0 General Description

1.1 Description
The CELLGARD ES Model NU-477E Laminar Flow Biological Safety Cabinet (LFBSC) is a bench/table top model, optionally
available with a base support stand, for operation as a console model. The CELLGARD ES Model NU-477E utilizes an Energy
Saver DC ECM Motor optimally determined forward curved fan for each model size/width to maximize both energy
efficiency and filter loading capacity. The Energy Saver ECM Motor is controlled to airflow setpoints via a solid-state DC
motor controller with digital dual thermister airflow sensors that provide an automatic compensation (constant volume
control) for both filter loading and line voltage variances.

The Laminar Flow Biological Safety Cabinet, (LFBSC) is a product resulting from the development of the "laminar flow"
principle (see airflow schematic) and the application of environmental controls as required in the field of biological research
or chemical containment. The LFBSC, when used with proper technique, is an effective laboratory aid in obtaining the
optimum control over product quality while reducing the potential for exposure of both product and personnel to airborne
biological or particulate chemical agents in low to moderate risk-hazard research and drug preparation or product
operations, as prescribed by the Center for Disease Control (CDC) Atlanta, Georgia.

The NU-477E bench LFBSC meets the requirements of a Class II, since the cabinet conforms to the following requirements:

1. Maintains an average inflow velocity of 100 LFPM (.51 M/S) through the work access opening.
2. Has HEPA filtered downflow air that is mixed with the inflow air from a common exhaust plenum.
3. Discharges a percentage of air to the outside atmosphere after HEPA filtration.
4. Has all biologically contaminated ducts and plenums under negative pressure or surrounded by negative pressure.

Cabinets used for work with minute quantities of volatile toxic chemicals and trace amounts of radionuclides required as an
adjunct to microbiological studies must be exhausted through properly functioning exhaust canopies.

1.2 Safety Instructions


These safety instructions describe the safety features of the CELLGARD ES Model NU-477E LFBSC.
The safety cabinet has been manufactured using the latest technological developments and has been thoroughly tested
before delivery. However, the cabinet may present potential hazards if it is not used according to the intended purpose
or outside of operating parameters. Therefore, the following procedures must always be observed:
 The safety cabinet must be operated only by trained and authorized personnel.
 For any operation of this unit, the operator must prepare clear and concise written instructions for operating and
cleaning, utilizing applicable safety data sheets, plant hygiene guidelines, and technical regulations, in particular.
o which decontamination measures are to be applied for the cabinet and accessories,
o which protective measures apply while specific agents are used,
o which measures are to be taken in the case of an accident.
 Repairs to the device must be carried out only by trained and authorized expert personnel.
 Keep these operating instructions close to the unit so that safety instructions and important information are
always accessible.
 Should you encounter problems that are not detailed adequately in the operating instructions, please contact your
NuAire Representative of NuAire technical Services.

OM0215 Page 5 of 59
November/2015
1.3 Explanation of Symbols

Safety alert symbol indicates a potentially


! WARNING hazardous situation which, if not avoided, could
result in death of serious injury.

Safety alert symbol indicates a potentially hazardous


situation which, if not avoided, may result in minor
! CAUTION or moderate injury.

CAUTION used without the safety alert symbol


CAUTION indicates a potentially hazardous situation which, if
not avoided, may result in property damage.

Potential electrical hazard, only qualified person


to access.

 NOTE: Used for important information. Flammable Hazard

Biohazard
Hazardous Gases! Personal
Protection Equipment Required.

Ground, Earth

Chemical Hazard
Lead Free

OM0215 Page 6 of 59
November/2015
OM0215 Page 7 of 59
November/2015
2.0 Models & Features

The model NU-477E, CELLGARD ES Class II Laminar Flow Biological Safety Cabinet is manufactured in
four sizes: 3 ft.(.9m), 4 ft.(1.2m), 5 ft.(1.5m), and 6 ft.(1.8m).

OM0215 Page 8 of 59
November/2015
OM0215 Page 9 of 59
November/2015
OM0215 Page 10 of 59
November/2015
OM0215 Page 11 of 59
November/2015
OM0215 Page 12 of 59
November/2015
OM0215 Page 13 of 59
November/2015
3.0 Warranty

NuAire, Inc. warrants Class II w/ECM Motor (CELLGARD ES) that it will repair F.O.B. its factory or furnish without charge
F.O.B. its factory a similar part to replace any material including HEPA filters in its equipment within 60 months after the
date of sale if proved to the satisfaction of the company to have been defective at the time it was sold provided that all
parts claimed defective shall be returned, properly identified to the company at its factory, charges prepaid. Factory
installed equipment or accessories are warranted only to the extent guaranteed by the original manufacturer, and this
warranty shall not apply to any portion of the equipment modified by the user. Claims under this warranty should be
directed to NuAire, Inc. setting forth in detail the nature of the defect, the date of the initial installation and the serial and
model number of the equipment.

This warranty shall not apply to any NuAire product or part thereof which has been subject to misuse, abuse, accident,
shipping damage, improper installation or service, or damage by fire, flood or acts of God. If the serial number of this
product is altered, removed or defaced as to be illegible, the Warranty shall be null and void in its entirety.

The warranty is for the sole benefit of the original purchaser and is not assignable or transferable. Prior to returning any
item, for any reason, contact NuAire for a Return Authorization Number. This number must accompany all returns. Any
product shipped to NuAire without this number will be returned refused shipment or collect freight.

4.0 Shipments

NuAire takes every reasonable precaution to assure that your CELLGARD ES cabinet arrives without damage. Motor carriers
are carefully selected and shipping cartons have been specially designed to insure your purchase. However, damage can
occur in any shipment and the following outlines the steps you should take on receipt of a NuAire CELLGARD ES cabinet to
be sure that if damage has occurred, the proper claims and actions are taken immediately.

4.1 Damaged Shipments

4.1.1 Terms are factory, unless stated otherwise. Therefore, it is important to check each shipment before acceptance.

4.1.2 If there is visible damage, the material can be accepted after the driver makes a notation on the consignee's copy
of the freight bill. Then an inspection must be made to verify the claim against the carrier. This inspection is the
basis of your filing the claim against the carrier.

4.1.3 If concealed damage is found, it is absolutely necessary to NOTIFY THE FREIGHT AGENT AT ONCE, and request an
inspection. Without this inspection, the transportation company may not accept a claim for loss or damage. If the
carrier will not perform the inspection, an affidavit must be prepared stating that he was contacted on a certain
date and that he failed to comply with the request. This along with other papers in the customer's possession will
support the claim.

OM0215 Page 14 of 59
November/2015
5.0 Installation Instructions

5.1 Location
Within the laboratory, pharmacy, etc., the ideal location of the biological safety cabinet is away from personnel traffic
lanes, air vents (in or out), doors and/or any other source of disruptive air currents.

SUGGESTED CABINET LOCATION IN LABORATORY

THE EXHAUST FILTER AREA IS ESPECIALLY SUSCEPTIBLE TO DISRUPTIVE AIR CURRENTS FROM AIR VENTS. The Electronic
Airflow Control System's exhaust probe is located just above the exhaust HEPA filter and if disruptive air currents are
present, the exhaust probe could be influenced by them, and indicate disruptive readings on the front panel display.
If drafts or other disruptive air currents exceed the inflow velocity of the cabinet through the access opening, the potential
exists for contaminated air to exit or enter the work zone area of the cabinet. It depends on the severity of the air current.
REMEMBER: A BIOLOGICAL SAFETY CABINET IS NO SUBSTITUTE FOR GOOD LABORATORY TECHNIQUE.

Where space permits, a clear 6" (152mm) area should be permitted on each side of the cabinet for maintenance purposes.
The electrical outlet into which the cabinet is connected should be readily accessible for maintenance purposes. Do not
position the cabinet to prevent access to the power cord. The power cord plug serves as the disconnect and should
remain readily accessible. If the outlet is inaccessible, such as a conduit (hardwired) connection, then an appropriate
warning label should be applied near the cabinets on/off switch to indicate the circuit breaker on the power distribution
panel should be used. A MINIMUM CLEARANCE OF 6" (152MM) IS REQUIRED FROM THE TOP OF THE CABINET TO THE
CEILING FOR PROPER VENTILATION OF THE EXHAUST EFFLUX.

If this cabinet is used in a pharmacy application, Per OSHA, NIOSH, and ASHP, it is strongly recommended that the cabinet
be exhausted to the outside. In addition, if this cabinet is used in microbiological application with minute quantities of
volatile toxic chemicals and tracer amounts of radionuclides, Per CDC/NIH and NSF it is strongly recommended that the
cabinet be exhausted to the outside. NuAire offers two general categories of exhaust transitions, which will capture the
exhaust efflux from the cabinet. These are:
Canopy, Thimble or Air Gap Exhaust Transitions (with and without integral fan)

 NOTE: THE EXHAUST SYSTEM SHOULD BE FITTED WITH A BACKDRAFT


DAMPER TO PREVENT REVERSING OF AIRFLOW IN THE SYSTEM.

NuAire strongly recommends a canopy or thimble exhaust for most applications. See separate instruction sheets for a
discussion of exhaust transitions and installation requirements.

 NOTE, some countries (i.e. Germany) only allow canopy or thimble type transitions. Verify requirements per Local,
State and Federal code laws.

OM0215 Page 15 of 59
November/2015
5.2 Set-Up Instructions
Remove outer shipping protection (carton or crating). The cabinet is fastened to the base skid and it is usually the best
procedure to leave the skid in place until the cabinet is located in its approximate position to facilitate ease in handling. It
can then be removed from the skid by removing the banding, bolts and screws holding the cabinet to the skid. It may be
necessary to remove the Control Center in order to gain passage through a doorway. It may easily be removed by following
the instructions on drawing
BCD- 11817.

5.2.1 Base Stand Assembly


The base stand is shipped knocked down in a separate carton and is assembled per drawing
BCD-05147 if accompanied with the unit. Remove the banding holding the cabinet to the base skid. Lift the
cabinet from the base skid and place on the floor. Now lift the cabinet on top of the base and bolt the base stand
to the cabinet using two 3/8" - 16 x 3/4" bolts and washers provided for the front base stand tabs and two 1/4"
acorn nuts for the rear weld studs. Place the cabinet in its desired location.

The base stand storage cabinets will usually be shipped according to customer requirements. If it is shipped
unassembled, it can be assembled per drawing BCD-05146. It is recommended that the upper and lower base
stand braces be installed first, then the rear and bottom panels (the end panels are always prefastened). Once
assembled, fasten the cabinet per the above instructions.

5.2.2 Leveling
Using a level placed on the work tray, adjust the leg levelers, first, end to end, then, front to back. The NSF
approved leg levelers provide a  3/4" (20mm) adjustment.

5.2.3 Bench/Security Drain Valve Installation (BCD-12634)


Place the cabinet on the bench with approximately a 2" (51mm) overhang clearance for the drain SST cap or
security drain valve. If the security drain valve is not desired, leave the SST cap in place and place the cabinet in its
desired location and using RTV caulk, seal all around the base of the cabinet and the bench. This provides a tight
seal to prevent bench spills from migrating under the cabinet.

If a security drain valve is desired, (NOTE, CHECK WITH YOUR SAFETY PERSONNEL FOR REGULATORY
REQUIREMENTS (i.e. LOCKING TYPE) OF DRAIN VALVE INSTALLATION) remove the handle from the valve stem
with security tool provided to gain clearance for valve body rotation. Add Loctite 242 (furnished) to the threads
and rotate valve body until secure, with the valve stem (for handle) on the left side. Re-install handle to valve
stem. Adjust the cabinet on bench to provide a 2" (51mm) overhang and seal the interface of the bench and
cabinet, using RTV caulk as above.

OM0215 Page 16 of 59
November/2015
5.2.4 Gas Service
NuAire doesn't recommend the use of natural gas within the BSC, but if gas service is determined to be necessary
for the application, appropriate safety measures must take place. All NuAire BSC's have precautionary warning
labels that say the following:

Use of explosive or flammable substances in this cabinet


! CAUTION should be evaluated by your appropriate safety personnel.

Once the determination has been made by the appropriate safety personnel, the application of natural gas must
be performed in accordance to national, state and local codes. IT IS ALSO STRONGLY RECOMMENDED THAT AN
EMERGENCY GAS SHUTOFF VALVE BE PLACED JUST OUTSIDE THE BSC ON THE GAS SUPPLY LINE.

The gas valve, when this option is installed, will only operate or flow gas when the cabinet blower is on and no
alarm is present. A solenoid valve is installed on the gas supply line for this purpose.

 NOTE, some countries (i.e. Germany) only allow certain types of certified valves to be used for natural gas (i.e.
Germany DVGW Certified). Verify requirements per Local, State and Federal codes/laws.

As previously stated NuAire doesn't recommend the use of natural gas within the BSC and ASSUMES NO
RESPONSIBILITY FOR ITS USE. USE AT YOUR OWN RISK. The Bunsen burner flame within the BSC not only
contributes to heat build-up; is also disrupts the laminar air stream, which must be maintained for maximum
efficiency. IF THE PROCEDURE DEMANDS USE OF A FLAME, A BUNSEN BURNER WITH ON DEMAND IGNITION IS
STRONGLY RECOMMENDED. DO NOT USE CONSTANT FLAME GAS BURNERS. During use, the Bunsen burner
should be placed to the rear of the workspace where resulting air turbulence will have a minimal effect.

5.2.5 Plumbing Services


Service ball valves with the type of service specified by the removable button on the handle, are located in the
work zone. The service ball valves are not recommended for pressure over 75 p.s.i. (5.2 BAR). Reducing valves
should be installed external to the cabinet if necessary. Service ball valves should never be used for oxygen
service. A special needle valve for oxygen service is required and available upon request.

External connection is to 3/8 inch NPT coupling in the inner sidewalls. Connection to plant utilities should be made
with proper materials for the individual service and according to national and/or local codes. Observe all labels
pertaining to the type of service and operating pressure.

5.2.6 Electrical Services


The NU-477E series Biological Safety Cabinets may be "hardwired" (optional) or plugged into an outlet with
protective earthing connection with the standard power cord. The unit requires
230VAC, 50/60 Hz, single phase (current rating varies per cabinet size, reference Electrical/Environmental
Requirements). It is recommended that power to the cabinet, whether hardwired or plug connected, be on its
own branch circuit, protected with a circuit breaker at the distribution panel. A surge protector is strongly
recommended if you are experiencing power related faults.

PLEASE NOTE, THIS UNIT CONTAINS ELECTRONIC BALLASTS FOR THE FLUORESCENT LIGHTING. ELECTRONIC
BALLASTS OPERATE WITH HIGH INRUSH CURRENT. IT IS NOT RECOMMENDED TO USE THIS PRODUCT WITH
GROUND FAULT CIRCUIT INTERRUPTERS (GFCI'S) BECAUSE THE BALLASTS MAY CAUSE THE GFCI TO TRIP.

OM0215 Page 17 of 59
November/2015
5.2.7 Final Assembly
Remove the protective cardboard cover over the exhaust HEPA filter, located under the protective screen if in
place. Attach the exhaust sensor shroud over the exhaust sensor. The shroud should be placed as close as possible
to the exhaust HEPA filter without coming in contact. The probe gasket should be tightly against the sensor shroud
to prevent sneak airflow paths. The exterior surface and viewing glass are easily cleaned with any mild household
detergent cleaner using a soft cloth. Harsh chemicals, solvent-type cleaners and abrasive cleaners should not be
used.

Do not attempt to clean the HEPA filter media. Cabinet interior walls or work surface are easily cleaned with any
mild household detergent cleaner using a soft cloth. Turn the cabinet on and let it operate for 60 minutes before
using it as a LFBSC.

5.3 Testing Methods and Equipment


After installation and prior to use, NuAire recommends that the cabinet be tested or commissioned to factory standards.
As part of testing, the certifier should go through the following initial checklist to assure all aspects of the BSC installation
are complete and ready for testing.
 Review product installation
- Exhaust connection, if present
- Damper valve installed correctly with label toward front, if present
- BSC basestand level
 Verify airflow sensor shroud is in place
- Exhaust flow
 Perform BSC installation tests
- At a minimum, the following tests should be performed:
 HEPA filter leak test
 Downflow velocity test with high/low alarm limits
 Inflow velocity test with high/low alarm limits
 Airflow smoke patterns
The testing methods and equipment required are specified on the factory inspection report included with this manual
(see insert in back cover).

IT IS RECOMMENDED THAT THESE TESTS BE PERFORMED BY A QUALIFIED TECHNICIAN WHO IS FAMILIAR WITH THE
METHODS AND PROCEDURES FOR TESTING BIOLOGICAL SAFETY CABINETS (SEE INSERT).

AFTER THE INITIAL INSTALLATION TEST, NUAIRE RECOMMENDS THAT THE CABINET BE RETESTED AT A MINIMUM ON AN
ANNUAL BASIS AND AFTER EVERY FILTER CHANGE OR MAINTENANCE ACTION OR ANY TIME THE OPERATOR FEELS IT IS
NECESSARY.

Note that the CELLGARD ES cabinets, filters and seals provide premium performance; Quality Control in both design and
manufacturing assure superior reliability. However, protection to both product and operator is so vital that testing to the
performance requirements should be accomplished as stated to ensure biological safety established by the factory
standards.

OM0215 Page 18 of 59
November/2015
Cellgard ES Energy Saver Class II
Laminar Flow Biological Safety Cabinet
Models NU-477-300E/400E/500E/600E

Catalog Number
Catalog Number NU-477-300E NU-477-400E NU-477-500E NU-477-600E
Nominal 3 foot (0.9m) Nominal 4 foot (1.2m) Nominal 5 foot (1.5m) Nominal 6 foot (1.8m)
Performance Specifications
NSF/ANSI 49 NSF/ANSI 49 NSF/ANSI 49 NSF/ANSI 49
1. Personal Protection
EN 12469 EN 12469 EN 12469 EN 12469
2. Product Protection
EN 12469 Class Class II Class II Class II Class II
Bench top/console w/base Bench top/console w/base Bench top/console w/base Bench top/console w/base
Style of Cabinet
stand/storage cabinet stand/storage cabinet stand/storage cabinet stand/storage cabinet
All welded stainless steel All welded stainless steel All welded stainless steel All welded stainless steel
Cabinet Construction 16GA, Type 304 16GA, Type 304 16GA, Type 304 16GA, Type 304
pressure tight design pressure tight design pressure tight design pressure tight design
Diffuser for Air Supply (Metal) Non-flammable Non-flammable Non-flammable Non-flammable
HEPA Filter Seal Type:
Supply Filter-99.99% Eff. on 0.3 microns HEPEX Seal HEPEX Seal HEPEX Seal HEPEX Seal
Exhaust Filter-99.99% Eff. on 0.3 microns Neoprene, Spring-loaded Neoprene, Spring-loaded Neoprene, Spring-loaded Neoprene, Spring-loaded
Fumigation : per EN 12469, Annex J
Yes Yes Yes Yes
Procedure
Standard Services:
Service Coupling (3/8 inch NPT) One, Right Sidewall One, Right Sidewall One, Right Sidewall One, Right Sidewall
Gas Valve/Service Coupling (3/8inch NPT) One, Right Sidewall One, Right Sidewall One, Right Sidewall One, Right Sidewall
Outlet Two, Backwall Two, Backwall Two, Backwall Two, Backwall
Optional Services:
Gas Cocks 3/8" NPT Up to 3 ea. Sidewall Up to 3 ea. Sidewall Up to 3 ea. Sidewall Up to 3 ea. Sidewall
Ultraviolet Light One, Backwall One, Backwall One, Backwall One, Backwall
Standard/Cup Sinks Lt or Rt. Work Surface Lt or Rt. Work Surface Lt or Rt. Work Surface Lt or Rt. Work Surface
Cabinet Size Inches (mm):
Height (Fully Assembled) 61 7/8 (1572) 61 7/8 (1572) 61 7/8 (1572) 61 7/8 (1572)
Height (Minimum for Transport) Width 59 (1499) 59 (1499) 59 (1499) 59 (1499)
Depth (with Control Center) 41 5/8 (1057) 53 5/8 (1362) 65 5/8 (1669) 77 5/8 (1972)
And Armrest Removed 33 (838) 33 (838) 33 (838) 33 (838)
Work Access Opening Inches (mm):
Standard Opening Height/Optional 8 (203) 8 (203) 8 (203) 8 (203)
Standard Inflow Velocity 105 FPM (.53 m/s) 105 FPM (.53 m/s) 105 FPM (.53 m/s) 105 FPM (.53 m/s)
Work Zone Inches (mm):
Height 28 1/2 (724) 28 1/2 (724) 28 1/2 (724) 28 1/2 (724)
Width 34 3/8 (873) 46 3/8 (1178) 46 3/8 (1178) 70 3/8 (1788)
Depth measured at 8 inches (102mm) 26 1/8 (664) 26 1/8 (664) 26 1/8 (664) 26 1/8 (664)
window height
Viewing Window Inches (mm): Fully closed to Fully closed to Fully closed to Fully closed to
Standard is tempered sliding glass 18 (457) open 18 (457) open 18 (457) open 18 (457) open
Required Exhaust CFM (CMH) CFM (CMH) CFM (CMH) CFM (CMH)
Canopy Variable Range (NU-911) 225-450 (383-765) 295-520 (502-884) 365-590 (621-1003) 436-661 (741-1124)
Canopy Fixed Range (NU-907) 269 (457) 359 (610) 445 (756) 532 (904)
Plant Duct Static Pressure Eng./Metric 0.05-0.1"/1.27-2.54mm 0.05-0.1"/1.27-2.54mm 0.05-0.1"/1.27-2.54mm 0.05-0.1"/1.27-2.54mm
H2O H2O H2O H2O
Heat Rejected, BTU, Per Hour (non- 826 1020 1611 1768
vented) 120 157 198 198
(vented)
Electrical: (CE marked):
Volts, AC 50/60 Hz 230 230 230 230
+Amps: Blower/Lights 1.0 1.3 2.0 2.2
Amps: Each Outlet 3 3 3 3
Amps: Total 6 10 11 11
12 ft. Power Cord (one) 14 GA - 3 Wire, 15A 14 GA - 3 Wire, 15A 14 GA-3 Wire, 15 A 14 GA-3 Wire, 15 A
Crated Shipping Weight:*** 475 lbs. /215 kg. 550 lbs. /249 kg. 650 lbs. /295 kg. 730 lbs. /331 kg.
Net Weight 425 lbs. /193 kg. 500 lbs. /227 kg. 600 lbs. /273 kg. 680 lbs. /308 kg.
Sound Pressure Level per ISO 4871**** Not to Exceed 55 dbA ■ Not to Exceed 56 dbA ■ Not to Exceed 58 dbA ■ Not to Exceed 60 dbA ■

■Reference the customer test report for procedure and results.


****Uncertainty is K = 2 dbA, measurement performed per ISO 11201 in normal running mode.
*** Crated shipping weight does not include weight for accessories or options.
+ Based on cabinet with new filters running at 230VAC.

OM0215 Page 19 of 59
November/2015
6.0 Operating the NU-477E

6.1 Electronic Control System

6.1.1 Overview
The electronic control system is designed to service the control requirements of the NU-477E Biological Safety
Cabinet. The control system consists of two electronic modules that will perform the following functions:
 Control blower via solid state switch.
 Control lights via solid state switch.
 Control outlets via solid state switch.
 Control blower DC ECM Motor with solid-state DC Motor Controller that provides automatic compensation
(constant volume control) for both filter loading and line voltage variances.
 Monitor and display airflow system performance via Flow Gard monitor.
 Airflow system alarm setpoints high/low via Flow Gard monitor.

The NU-477E incorporates the use of two electronic modules that improves the cabinet's performance. The Flow
Gard monitor uses a dual thermistor airflow probe located in the exhaust airflow to monitor the cabinet system
function. The Flow Gard monitor indicates through LED's normal operation, as well as high alarm status (HEPA
filter loading) and low alarm status (low airflow). The main control module, through the use of the front panel,
controls the on/off functions of the blower, fluorescent and ultraviolet (optional) lights, and outlets. The main
control module also monitors the sliding window position with a microswitch for both window high and closed
(interlocks optional UV light) positions. Lastly, the main control module includes fan relay contact closures for
interaction with HVAC systems to optimize environmental performance. All the above functions are shown in a
system block diagram (see Figure 1).

Figure 1: Block Diagram

Figure 1

OM0215 Page 20 of 59
November/2015
6.1.2 Front Panel
The control system front panel contains the following functions described in detail (See Drawing BCD-12056).

6.1.2.1 Blower Keys


The blower keys indicate and control ON/OFF power to the blower.

The blower LED will also blink when both the sliding window is fully closed and the blower "on"
switch has been depressed.

6.1.2.2 Light Keys


The light keys indicate and control ON/OFF power to the fluorescent and optional ultraviolet lights

6.1.2.3 Outlet Keys


The outlet keys indicate and control ON/OFF power to the outlets.

6.1.2.4 Window Alarm LED


The window alarm LED indicates when the sliding window is raised above or below
its proper operating height.

6.1.2.5 Cleaning Key


The cleaning key may be pressed to silence the audible alarm for cleaning/loading purposes only.

6.1.2.6 Flow Gard Arrow Adjustment Keys


The arrow adjustment keys allow user interaction for various functions.

6.1.2.7 Flow Gard LED Display


The Flow Gard LED display indicates the system running condition. Green is normal; yellow is caution and
red is alarm.

6.1.2.8 Flow Gard Reset Key


The Flow Gard reset key allows various user interactions for various functions.

6.1.2.9 Audible Alarm


The audible alarm is provided per EN 457, tested at one frequency to a loudness of
13 db over the cabinet’s normal running db level. The audible alarm will activate on all alarm conditions
as long as they are present. The audible alarm will activate for 30 seconds full on, then ring back once
every 5 seconds there after until the alarm condition is cleared.

OM0215 Page 21 of 59
November/2015
OM0215 Page 22 of 59
November/2015
6.1.3 Run Mode Operation
Operation of the cabinet is initiated by plugging the power cord into the appropriate line power. In the power off
condition (cabinet is unplugged); all calibration and running parameters will be stored in the microprocessor's
EEPROM memory. During the power on condition (cabinet is plugged in), the cabinet's blower, lights, and outlet
may be turned on. The Flow Gard monitor will automatically turn on when the blower is on.

6.1.3.1 Airflow Control Mode


The operating airflows within the cabinet (i.e. 60 LFPM (.30 m/s) downflow and 105 LFPM
(.53 m/s) air inflow barrier) are controlled by the DC motor speed control and an exhaust damper. The DC
motor speed control controls the ECM motor by providing a Pulsed Width Modulation (PWM) signal of 2.5 to
8.0 Vdc that is sent to the ECM motor. The ECM motor also sends back a Revolution Per Minute (RPM) signal
to the DC motor speed control. The RPM signal is monitored for changes in comparison to a reference stored
motor loading curve of RPM versus PWM and adjusts as needed to maintain a constant volume of air.

The airflow calibration mode provides a means of adjusting the curve starting point, if necessary, at any point
of the filter loading process. The airflow calibration mode accessed and exited through a front panel key
sequence allows adjustment to achieve the desired airflow. At that point the current RPM and PWM values
will be stored as the starting point for further airflow operation.

6.1.3.2 Sliding Window Operation


The cabinet has a full counter-balanced and removable sliding glass window with two operational features. As
the window is raised above its specified operating height, an audible and visual alarm alerts the operator of
possible compromised personnel protection. NOTE, for "E" Series, the cabinet also contains a low window
alarm, which will activate both audible and visual alarms if the window is below its specified operating height.
If the window is closed, it will perform an interlock function that will automatically turn off the blower or
prevent it from being turned on. In addition; will allow the Ultraviolet light, if installed, to be turned on.

OM0215 Page 23 of 59
November/2015
6.1.3.3 Blower Password Protection/Night Setback Mode
The blower is operated by using a password (a sequence of 3 key strokes) in combination with or without the
night setback mode. The night setback mode allows the DC ECM motor to continue to run at a lower rate with
the sliding window closed allowing the workzone interior to be continually HEPA filtered. If the night setback
mode is selected (JP3), the blower switch must be on and the window closed for it to be activated. In other
words, if JP3 is on, the night setback mode will indicate when the window is closed, (or with about 1/4” gap
between the bottom of the window and the armrest). If JP3 is off the motor will shut off when the window is
closed.

When the window is closed in night setback mode:


 Blower speed is reduced
 The fluorescent Lights shut off
 Blower On/Fluorescent on LED’s will blink
 24 VAC Power to the alarm monitor is disabled
 Fan relay changes state
 Accessary relay changes state
 Outlet power is disabled

A combination of 3 jumpers on the control board will activate either the password and/or the night setback
mode operation. The jumpers are located and labeled J1, J2, and J3 on the control board. The key stroke
password sequence is a selected combination of the hidden key (blower symbol), blower on and blower off.
The table below indicated the jumper combinations and their associated password with or without the night
setback mode.

Password Password Night Setback Password


JP1 JP2 Mode JP3 Sequence
Off Off Off NO PASSWORD
On Off Off on off hidden
Off On Off on hidden off
On On Off off on hidden
Off Off On off hidden on
On Off On hidden on off
Off On On hidden off On
On On On NO PASSWORD

(Default in bold)

OM0215 Page 24 of 59
November/2015
6.1.4 Flow Gard Operation

6.1.4.1 Overview
The Flow Gard monitor uses a dual thermister airflow probe located in the exhaust airflow to monitor the
cabinet system function. The Flow Gard monitor indicates through LED's normal operation (green), as well as
high alarm status (red) (Hepa filter loading) and low alarm status (red) (low airflow). All user interaction is
accomplished through the arrow and reset keys. IT IS RECOMMENDED THAT THE FLOW GARD BE
CALIBRATED ANNUALLY DURING THE TESTING PROCESS.

6.1.4.2 Power-Up Sequence


On power up, the digital display is initialized and every segment of the display turned on for two seconds. The
three LEDs and the audible alarm are also activated. The firmware version number is then displayed for two
seconds.

6.1.4.3 Audible Alarm


The audible alarm will be activated whenever the Low or High Alarm zone is reached. Once the audible alarm
is activated, it will stay on until the alarm condition is cleared.

6.2 Operating Guidelines


The intent herein is to present general operational guidelines that will aid in the use of the Laminar Flow Biological Safety
Cabinet (LFBSC) to control airborne contaminants of low to moderate risk as stated in Technical Report No. FPS
56500000001 prepared by Dow Chemical U.S.A. for the National Cancer Institute, May 1, 1972.

Procedure protocols defined in terms of the barrier or control concepts unique to LFBSC must be developed in order to
obtain a maximum potential for safety and protection. The pre-planning necessary to develop these protocols is based on
several fundamental considerations, each of which will contribute to optimum benefits from the equipment:
a. Know your "safe working area"
b. Minimize disruption of "air curtain"
c. Minimize room activity
d. Utilize unidirectional air flow
e. Employ aseptic techniques

6.2.1 Know Your "Safe Working Area"


The LFBSC safe working area is basically the worktray or depressed area. All work should be performed on or
above the worktray. The area on or above the
front grill is a non-safe working area.

 NOTE: It is important to maintain an air gap on both sides of the worktray before fastening in place.
The work tray as being part of the cabinet system has been designed to load up to 100 lbs. (45.4 kg)
of work materials. Any additional loading should be evaluated by appropriate safety personnel.

OM0215 Page 25 of 59
November/2015
6.2.2 Minimize Penetration of "Air Curtain"
The minimum number of items necessary should be placed into the cabinet to prevent overloading, but the work
should also be planned to minimize the number of times an operator's hands and arms must enter and leave the
air curtain at the open face. The ideal situation is to have everything needed for the complete procedure placed in
the hood before starting, so that nothing need pass in or out through the air barrier at the face until the procedure
is completed. This is especially important in working with moderate risk agents.

Unnecessary rising of the hands inside the cabinet above the level of the work opening should be avoided. This
presents an inclined plane from hands to elbows along which the downflow of air may run to, and possibly out, the
open face.

 Note: When working with agents of lower risk, it is not as important for all materials to be placed in the cabinet before
starting, or for the procedure to be completely finished before materials are removed. Also, the time period for a
unit may be continued over a more extended period during which entries and withdrawals from the cabinet may
be made.

6.2.3 Minimize Room Activity


Activity in the room itself should be held to a minimum. Unnecessary activity may create disruptive air currents as
well as interfere with the work of the operator. A person walking past the front of a cabinet can cause draft
velocities up to 175 fpm (.89 m/s), which are sufficient to disrupt the air balance of the laminar flow unit.

6.2.4 Utilize Unidirectional Air Flow


The operator must keep two important facts in mind: (1) The air, as supplied to the work area through filters from
the top, is contaminant free and (2) Airborne contamination generated in the work area is controlled by the
unidirectional flow of parallel air streams in a top-to-bottom direction.

A solid object placed in a laminar air stream will disrupt the parallel flow and consequently, the capability of
controlling lateral movement of airborne particulates. A cone of turbulence extends below the object and
laminarity of the air stream is not regained until a point is reached downstream, approximately equal to three to
six times the diameter of the object. Within the parameters of this cone, particles may be carried laterally by
multidirectional eddy currents.

Transfer of viable materials and manipulations which may generate aerosols should not be performed above
sterile or uninoculated materials. Items should be localized on the work surface in "clean" and "dirty" groups.

OM0215 Page 26 of 59
November/2015
6.2.5 Employ Aseptic Technique
The operator must not assume an attitude of "let the cabinet do it" when performing procedures within a LFBSC.
Properly balanced and properly used cabinets will do an excellent job of controlling airborne contamination and
containing viable agents, but the cabinet will not eliminate contact transmission of contamination. Normal
laboratory contamination control procedures and basic aseptic techniques are necessary to obtain maximum
benefit from the cabinet. For example, open bottle, tube or flask mounts should be kept as parallel as possible to
the downflow to minimize capture of chance particulates. This precaution is merely an extension of good aseptic
technique as practiced on open bench tops. The good laboratory practices designed to minimize creation and/or
release of aerosols to the environment should not be discontinued.

Items of equipment in direct contact with the etiologic agent must remain in the cabinet until enclosed or until
surface-decontaminated. Trays of discard pipettes must be covered before removal from the cabinet (aluminum
foil may substitute for fabricated covers).

If an accident occurs which spills or splatters suspensions of etiologic agent around the work area, all surfaces and
items in the cabinet must be surface-decontaminated before being removed.

Applying a burner flame to flask and tube necks when mating surfaces of sterile assemblies is a conventional
method of minimizing chance contamination. However, the efficiency of this operation is usually related to the
removal of airborne contamination occurring while the item is uncovered. If the manipulation is carried out in an
environment free of airborne particulates, then the need for the flaming operation is essentially removed. This is
one of the additional advantages of the LFBSC - use of the gas burner is seldom necessary. The gas burner flame in
one of these units not only contributes significantly to the heat build-up, it also disrupts the laminar air streams
which must be
maintained for maximum efficiency. If the procedure demands use of a flame, A BUNSEN BURNER WITH ON
DEMAND IGNITION IS RECOMMENDED. DO NOT USE CONSTANT FLAME GAS BURNERS. It should also be placed
to the rear of the workspace where resulting air turbulence will have a minimal effect. If cabinet air is inadvertently
turned off, the flame could damage the HEPA filters.

6.3 Operating Sequence

6.3.1 Start Up
Turn on cabinet blower and lights, check air intake and exhaust portals of the cabinet to make sure they are
unobstructed.

 Note: Some cabinets are equipped with ultraviolet (UV) lights. Good procedure includes the decontamination or
wipedown of cabinet surfaces with chemical disinfectant before work commences. This practice eliminates the
need for UV lights, whose primary utility in this application is inactivation of surface contamination since the filters
effectively remove all airborne contaminants. UV lights, therefore, are not recommended in the LFBSC.

Allow blowers to operate for a minimum of 15 minutes before aseptic manipulations are begun in the cabinet. If
the filtered air exhausted from the unit is discharged into the room, as in some installations, an additional
advantage is obtained from purification (filtration) of the room air circulated through the equipment. Because of
this characteristic contributing to the quality of the laboratory environment, some owners of LFBSC leave them in
operation beyond the time of actual use.

6.3.2 Wipedown
The interior surfaces of the work space should next be disinfected (see Cleaning Procedures section) by wiping
them thoroughly with 70% alcohol or similar non-corrosive anti microbial agents. USE OF CHLORINATED OR
HALOGEN MATERIALS IN THE CABINET MAY DAMAGE STAINLESS STEEL.

6.3.3 Materials & Equipment


The apparatus and materials should next be placed into the cabinet. Care must be exercised that no items be
placed over the front intake grills. Materials should be arranged so that clean, dirty (used), and virus materials are
well separated. Passage of contaminated materials over uninoculated cultures or clean glassware should be
avoided and transfer of viable materials should be performed as deeply into the cabinet (away from open face) as
possible.

OM0215 Page 27 of 59
November/2015
6.3.4 Air Purge
Additional purging of the workspace without user activity should be allowed for 2-3 minutes after materials and
apparatus have been placed in it. This will rid the area of all "loose" contamination that may have been introduced
with the items.

6.3.5 Perform Work


The work can now be performed. The technician performing the work is encouraged to wear a long-sleeved gown
with knit cuffs and rubber gloves. This will minimize the shedding of skin flora into the work area and concurrently
protect the hands and arms from viable agent contamination. At a minimum, the hands and arms should be
washed well with germicidal soap before and after work in the cabinet. For the preparation of antineoplastic
drugs, the following procedures summarize those contained in OSHA Technical Manual TED 1-0.15A, Section VI,
Chapter 2 “Controlling Occupational Exposure to Hazardous Drugs”. The above document should be thoroughly
studied / reviewed prior to drug preparation in the cabinet. It may be found at this website,
www.osha.gov/dts/osta/.

a. A sterile plastic-backed absorbent drape should be placed on the work surface during mixing procedures.
The drape should be exchanged whenever significant spillage occurs, or at the end of each production
sequence.

b. Vials should be vented with a filter needle to eliminate internal pressure or vacuum.

c. Before opening ampoules, care should be taken to insure that no liquid remains in the tip of the ampoule.
A sterile gauze sponge should be wrapped around the neck of the ampoule while opening.

d. Final drug measurement should be performed prior to removing the needle from the stopper of the vial.

e. A non-splash collection vessel should be available in the biological safety cabinet to discard excess drug
solutions.

6.3.6 Terminal Purging & Wipedown


Following completion of work, allow the cabinet to run for 2-3 minute period without personnel activity to purge
the unit. A surface disinfection of the interior surfaces (see Cleaning Procedures section) should be repeated after
removal of all materials, cultures, apparatus, etc. A careful check of grills and diffuser grids should be made for
spilled or splashed nutrients which may support fungus growth and resulting spore liberation that contaminates
the protected work environment.

6.3.7 Paper Catch/Prefilter


A permanent paper catch is installed behind the rear divider panel of the work zone. This area forms the return air
path to the motor/blower; and if the airflow is blocked, it could seriously affect the performance of the cabinet.
Therefore THE PAPER CATCH SHOULD BE CHECKED AND CLEANED NO LESS THAN ON A WEEKLY BASIS; and a daily
basis if procedures dictate the use of paper products. Any paper removed must be properly disposed of as
Contaminated Hazardous Waste. The above procedures also apply to all units configured with a prefilter.

6.3.8 Shut Down


Turn off blowers and lights. Do not use cabinet as a depository for excess lab equipment during periods of non-
operation. If antineoplastic agents are being prepared in the cabinet, it is recommended to let the cabinet run 24
hours per day. This lessens the possibility that contaminants may escape.

OM0215 Page 28 of 59
November/2015
6.4 Ergonomics
Ergonomics, the study or accommodation of work practices is extremely important for proper cabinet usage and user
health and safety. An evaluation of normal work practices should be performed with each user when working in a cabinet.
Evaluation criteria should be at a minimum:
a. Proper user posture
b. Effective workzone layout for work practice
c. Vision or sightlines

For each of the above evaluation criterion, several work aids may be supplied to accommodate the user.
 Ergonomic chair - A six-way articulating seat and back control for personalized adjustment to assure proper user
posture. Be sure feet are resting on the floor, chair foot support or foot rest. Also be sure back is fully supported
with proper chair adjustments.
 Forearm/elbow support - The cabinet is provided with a non-metallic forearm support on the work access opening.
Periodic mini-breaks during work practice should be taken resting forearm to avoid stress and fatigue. Elbow rests
are optional that can provide support for particular work practices, such as pipetting. Also available as an option,
closed cell foam disposable forearm pads to reduce pressure points and add comfort.
 Effective workzone layout - Always prepare your work procedure to minimize reach to avoid neck and shoulder
stress and fatigue. Rotating tables are optional to maximum workzone and minimize reach.
 Vision and sightline - Always prepare your work procedure to eliminate glare and bright reflections on the window.
Keep your window clean and sightlines clear to your effective workzone.

6.5 Cleaning Procedures


Cleaning the cabinet is an important function in terms of both containment and sterility.
Use the following procedure to effectively clean or surface disinfect the cabinet workzone surfaces.
a. Raise the sliding window to a full-open position, if desired.
b. Press the cleaning key on the front control panel to silence the audible alarm during the cleaning process.
c. Apply appropriate disinfecting solution (i.e. coverage plus ™ (Calgon Corp.)) or similar disinfectant to cabinet
surfaces. Most surface disinfectants require a specific contact time depending upon the microbiological agents
used within the cabinet. CONSULT APPROPRIATE DISINFECTANT DOCUMENTATION FOR PROPER APPLICATION
AND SAFETY PRECAUTIONS.

 NOTE: DISINFECTANTS THAT USE CHLORIDES AND HALOGENS WILL CAUSE DAMAGE TO THE STAINLESS STEEL
SURFACES IF LEFT ON FOR LONG PERIODS OF TIME.

d. After the specified contact time, wipe up excess disinfectant. IF THE DISINFECTANT USED CONTAINS CHLORIDES
OR HALOGENS, RE-WIPE ALL SURFACES WITH 70% ALCOHOL OR SIMILAR NON-CORROSIVE ANTI-MICROBIAL
AGENT TO PREVENT DAMAGE TO STAINLESS STEEL SURFACES.

OM0215 Page 29 of 59
November/2015
6.6 Hazardous Drug Decontamination Procedures
This procedure is intended to provide guidance following a spillage and/or periodic maintenance, testing or relocation of
the cabinet. Additional guidance can be provided by the CETA document CAG-005-2007 found at the CETA website:
www.CETAinternational.org.

6.6.1 Preparation
Prior to beginning decontamination activity, personnel should wear proper personnel protection equipment (PPE)
1
i.e. Tyvek isolation gown, 2 pair of Nitrile gloves and a full-faced HEPA filtered respirator. All protective garments
should be contained in 4 mil plastic bags and labeled for disposal as chemotherapy waste after completion of the
procedure. For the purpose of this procedure, detailed procedures for cleaning a Class II BSC can be found in the
2
1990 ASHP Technical Assistance Bulletin Guidelines on Handling Cytotoxic and Hazardous Drugs .

6.6.2 Procedure
a. Make sure that the cabinet remains in operational mode with internal blower on.
b. Open the hinged or sliding view screen and secure in the full open position.
With the view screen in the full open position,
personnel protection is compromised and a
! CAUTION full faced HEPA filtered respirator must be worn.

c. Clean all readily accessible surfaces of the cabinet.


d. Remove perforated metal diffuser screen from the underside of the supply HEPA filter and place on the
cabinet work tray.

 Note: Depending on the model, the diffuser screen is secured to the cabinet by #8-32 screws or 1/4" - 20 acorn
nuts, 3 places. It is purposely a tight fit and is secured to the back wall with projecting threadless studs.

e. Clean both sides of the perforated metal diffuser screen and remove it from the cabinet.
f. Lift the cabinet work tray, clean both sides and remove it from the cabinet.
g. Remove the front perforated grill, place on the cabinet floor and clean both sides. Remove from cabinet.
h. Clean work tray supports.
i. Working from top to bottom, clean all inside surfaces of the cabinet. Take care not to wet the HEPA filter.
If liquid has collected in the plenum drain, aspirate it using IV tubing into an evacuated container. Label
the evacuated container for disposal as chemotherapy waste.
j. Clean the plenum drain area and wipe dry.
k. If the cabinet requires maintenance and/or replacement of the HEPA filters, the operation should be
halted at this point to allow trained personnel to complete replacement of the HEPA and/or maintenance
action required.

6.6.3 Assembly
a. Replace front (if removed) grill.
b. Replace the work tray and carefully tighten the thumbscrews.
c. Replace perforated metal diffuser screen over the underside of the supply HEPA filter.
d. Wipe down all exposed surfaces of the work area with 70% isopropyl alcohol.
e. Prepare for aseptic operation.

¹ Available from Lab Safety Supply, Janesville, WI 53547-1368, or other laboratory, industrial, or hospital supply distributors.

²American Society of Hospital Pharmacists. 1990. ASHP Guidelines on Handling Cytotoxic and Hazardous Drugs Am. J. Hosp. Pharm. 47:1033-49

OM0215 Page 30 of 59
November/2015
7.0 General Maintenance

All maintenance actions on this equipment must be performed by a qualified


technician who is familiar with the proper maintenance procedures required
! CAUTION for this equipment. This includes both testing as well as repair.

7.1 Decontamination
No maintenance should be performed on the interior of the CELLGARD ES cabinet (area behind access panels) unless the
cabinet has been microbiologically decontaminated, is known to biologically clean, or known to be chemically inert. Surface
disinfection is performed as specified in the Cleaning Procedures section.

Hazardous Gases! Personal Protection Equipment Required.


! CAUTION
A disinfection using formaldehyde must be performed in accordance with the specifications of
NSF 49/1992, Annex G or EN 12469, Annex J.

This procedure presents considerable risks and must be performed only by specially trained
and authorized service personnel in accordance with applicable national safety regulations
(e.g. Germany: TRGS 522).

The formaldehyde is vaporized within the tightly sealed sample chamber. The quantity of the
applied formaldehyde depends on the volume of the sample chamber in the safety cabinet that
is to be disinfected. The formaldehyde evaporates immediately after reaching its boiling point;
the minimum reaction time is 6 hours. Therefore, the formaldehyde should be neutralized after
the specified reaction time by vaporizing ammonia.

Flammable Hazard!
! CAUTION

Formalin is flammable. The auto-ignition temperature of formalin is 430 C (820° F).


With a volume percentage of 7.75% in dry air, formaldehyde vapor may explode.
For vaporization, do not use heating devices reaching temperatures above 250° C (477° F).

! CAUTION Chemical Hazard!

Formalin in reaction with hydrogen chloride will form BCME which is a hazardous chemical.
When using formalin, all residues of hydrogen chloride in the work chamber of the cabinet must be
removed.

If microbiological decontamination is necessary, use the following procedure:


1. Disconnect power to the cabinet.
Remove screws at each upper side of the control center and allow the control center to rotate down, resting on
the safety straps.
Disconnect electrical connectors on left side.
Disconnect electrical from right side. (Be sure to note the location of the supply and exhaust sensor wires before
disconnecting them from the main board).
Loosen safety plate next to left hinge.
Remove control center by disconnecting safety straps and moving control center to the left off the slip hinges.
Fold and pinch tubing to seal.

OM0215 Page 31 of 59
November/2015
2. Remove the front decorative panel via top/front fasteners.
3. Remove left and right window farings via upper, middle, and top fasteners.
At this point, the sliding window can be removed or allowed to hang from its top position.
If the window is allowed to hang, use duct tap to secure to the cabinet during the remaining process.
4. Remove armrest via fasteners.
5. Remove downflow and exhaust probe shrouds. Remove only the downflow probe by turning the locking ring
counter clockwise.
6. Remove the diffuser screen and window wiper seal.
7. Place decontamination equipment inside the work area. Reference decontamination procedure,
per EN 12469, Annex J, using the following chart to calculate chemical requirements.

Cabinet Size 300E 400E 500E 600E


Cabinet 58-7/8 x 28 x 34-3/8 in. 58-7/8 x 28 x 46-3/8 58-7/8 x 28 x 58-3/8 58-7/8 x 28 x 70-3/8
Dimensions (1.50 x .711 x .873 m) (1.50 x .711 x 1.18 m) (1.50 x .711 x 1.48m) (1.50 x .739 x 1.88 m)
Cabinet 32.79 cu. ft. 44.24 cu. ft. 55.69 cu. ft. 67.14 cu. ft.
Volume (.929 cu. m) (1.25 cu. m) (1.58 cu. m) (1.90 cu. m)

 Note: The outlets in the work area are energized as long as the cabinet is plugged in and switched on the front panel.
Unplug the cabinet before decontamination equipment is plugged into these outlets.
The control centers electrical connectors may be re-attached to utilize the cabinet's interior outlets and fan
during the decontamination. (See step 8)

8. Set control center on its side (with the ballast end down and the blower capacitor end up) on the left side of
cabinet. Then plug in the 16-pin connector cable from the cabinet to the control center.
In order to activate the blower, a jumper wire must be used to bridge the B/UVT and B/UV - contacts on the main
board. Reconnect power to the cabinet. The outlets and blower can then be activated.
9. Seal front and top openings using pressure plates.
a. The front plate is attached using the following steps:
1) Remove screws in SST header just above workzone opening.
2) Place front seal plate over bottom row of studs with the plate gasket next to the work access
opening and the remaining holes should line up.*
3) Fasten the plate using the fastening screws and nuts provided.
b. The top seal plate is attached using the following steps:
1) Remove filter guard above exhaust HEPA filter.
2) Remove exhaust sensor shroud via fasteners.
3) Place top seal plate over the studs and attach fastening nuts provided.

 Note: *Electrical
Front and top seal plates are obtained by contacting NuAire Representative or Distributor
service can be provided to the inside of the cabinet via the electrical bulkheads
on the front pressure plate. Access to the neutralizing plate can be obtained through the
liquid tight fitting on the front pressure plate.

BE SURE CABINET IS TOTALLY SEALED TO PREVENT ANY


! CAUTION LABORATORY EXPOSURE TO DECONTAMINATION GAS.

10. Perform decontamination procedure per EN 12469, Annex J.

Normally, no preventive maintenance is required on the interior of the cabinet (i.e., the area behind the access panel
containing the HEPA filters and motor (blower assembly). All required adjustments in order to maintain proper cabinet
airflows are external to the cabinet interior. The motor is lubricated for life and is thermally protected with automatic
reset.

¹ Available from Lab Safety Supply, Janesville, WI 53547-1368, or other laboratory, industrial, or hospital supply distributor

OM0215 Page 32 of 59
November/2015
7.2 Fluorescent Lamp, Bulb Replacement
The two (T8) fluorescent bulbs are cool white, rapid start and placed external to the cabinet to aid maintenance and
minimize heat build-up within the cabinet. The life rating of the bulb is 9000 hours based on three hour burning cycles.

To replace a bulb, it is necessary to remove the lamp assembly.

1. Switch Cabinet Light Switch off.


2. Remove the screws at each upper side of the Control Center and allow the Control Center to rotate down, resting
on the safety straps.
3. The bulb is now directly exposed for replacement.
4. The bulb is removed by displacing the bulb to one side against the compressible lamp holder and lifting out the
bulb.
5. Reverse the procedure to reinstall the lamp assembly being careful not to pinch the safety straps, cable or tubing
during closure of the control center.

7.3 HEPA Filter/Motor Replacement (Drawing BCD-11819)


The HEPA Filters under normal usage and barring an accident (a puncture), do not need replacement until the efflux
velocity cannot be maintained or the access inflow velocity cannot be maintained at 100 LFPM (.51 m/s) (min.). This may
permit the average downflow velocity to be as low as 55 LFPM (.28 m/s).

The HEPA Filters should not be replaced until the entire cabinet has been decontaminated or known to be biologically
"clean".

7.3.1 Procedure (see Drawing BCD-11819)

Disconnect electrical power from the unit before


! CAUTION attempting any maintenance action.

Step 1: Remove screws at each upper side of the control center and allow the control center to rotate down,
resting on the safety straps. Second, remove the front decorative panel, which is held into position
by (3) knurled nuts on the top edge and snap fit bullet catches on the bottom.
Step 2: Place sliding window into lowest position and remove front filter panel, which is held into position by
Phillip pan head screws. Once the screws are removed, the panel is held into position by smooth
weldstuds located on the top corner of the front filter panel. Use the window stop brackets as handles to
remove the panel.

CAUTION: Screws are used in lieu of acorn nuts, and lockwashers.


The screws have O-rings and should be replaced if damaged or badly deformed.

The interior of the cabinet is now fully exposed for replacement of the filters and/or motor/blower.

OM0215 Page 33 of 59
November/2015
Step 3: Filter Removal
It is not always necessary to replace both the supply and exhaust filters at the same time. If during the
course of certifications, the downflow falls off while the exhaust increases the supply filter is "loading"
faster than the exhaust filter and only the supply filter may need replacement. The opposite might also
happen depending upon many factors.
a. To remove the supply filter:
1. Unscrew the 3 hand knobs (about 3 turns away from filter) located behind the supply
HEPA filter. This will relieve the pressure to the rear of the plenum.
2. Relax the 4 plenum screws in front of the supply HEPA filter.
3. Lift the permanent plenum and hold up with wire strap.
4. Carefully remove the supply filter. Direct exposure should be avoided.
CAUTION: Dispose of spent HEPA filters properly.
Avoid direct contact to "dirty side" of the filters. Label toxic waste.
b. To remove the Exhaust HEPA:
1. Relax the exhaust filter seal loading mechanism by turning the four threaded bolts
counterclockwise until one can see a definite release of the loading springs.
2. Pull the exhaust choke tray free and remove the filter. It is not necessary to remove the
tray, although it is free to move forward several inches, if necessary, to free the HEPA filter.
Step 4: Filter Installation
When installing new filters, USE ONLY NUAIRE SPECFIED FILTERS FOR REPLACEMENT.
(Filters shall conform to EN13091 and EN 1822-1 Class H14)

Description: Supply HEPA Filter Exhaust HEPA Filter


Efficiency: 99.995% @ 0.3 Micron 99.995% @ 0.3 Micron
Airflow Rating: 100 fpm @ .60  .05" w.g. per sq. ft. 250 fpm @ .70  .05" w.g. per sq. ft.
Frame Type: Metal Metal

NU-477-300E
NuAire Part Number: A-980962-01 A-980962-05
Filter Size: 21" (533mm) x 32" (813mm) x 3" (76mm) 20" (508mm) x 14" (366mm) x 11 1/2" (292mm)
Filter Manufacturer: Camfil Farr Camfil Farr

NU-477-400E
NuAire Part Number: A-980962-02 A-980962-06
Filter Size: 21" (533mm) x 44" (1118mm) x 3" (76mm) 20" (508mm) x 20” (508mm) x 11 1/2" (292mm)
Filter Manufacturer: Camfil Farr Camfil Farr

NU-477-500E
NuAire Part Number: A-980962-03 A-980962-07
Filter Size: 21" (533mm) x 56" (1422mm) x 3" (76mm) 20" (508mm) x 26" (660mm) x 11 1/2" (292mm)
Filter Manufacturer: Camfil Farr Camfil Farr

NU-477-600E
NuAire Part Number: A-980962-04 A-980962-08
Filter Size: 21" (533mm) x 68" (1727mm) x 3" (76mm) 20" (508mm) x 32" (813mm) x 11 1/2" (292mm)
Filter Manufacturer: Camfil Farr Camfil Farr

a. To install the supply filter, simply reverse the procedure outlines in Step 3a, above.

Note: Be sure to open the choke plate fully before inserting the filter into the tray. This will assist in adjusting the airflow.

b. To install the exhaust filter, apply a thin layer of silicone grease to the top and bottom gaskets of
the filter and carefully insert into the exhaust choke tray.
Position the filter frame within the outside walls of the exhaust opening On the top of the hood.
Tighten the spring loaded bolts, 4 places, depressing the gasket material by 1/8 inch (3mm).

OM0215 Page 34 of 59
November/2015
Step 5: Motor/Blower Assembly Removal
a. It is recommended that the motor/blower to be removed as a single unit. To remove, disconnect
electrical connections to the motor, remove the HEPEX pressure plenum and unbolt the
motor/blower assembly from the roof of the cabinet (4 places). Always inspect the rubber
isolation motor mounts and replace those that are cracked or visibly show stress.
b. Replace the motor exactly as originally installed in the blower housing, paying particular
attention to the correct electrical connections (see Electrical Schematic).
c. Re-install the new motor/blower assembly.

7.4 Sliding Window Replacement & Adjustment


The sliding window replacement is accomplished by removing the front decorative panel, control center, and window glide
assemblies. The sliding window adjustment may be required due to everyday use over the life of the cabinet. The left
window glide is stationary since it contains the microswitches that monitor window height. The right window glide is
adjustable by a set screw and tension screw method (see Drawing BCD-11686). When adjusting the sliding window, be sure
to verify proper microswitch operation. If the sliding window is too loose, the sliding window will not properly activate the
microswitches, thus causing potential operational malfunctions to occur. In Addition, the sliding window retention, or
ability to slow the rate of fall, if a window counter balance experiences a fault. It is also required to assure proper window
function (see Sliding Window Retention Verification in the Inspection Report).

7.5 Airflow Calibration


The NU-477 airflow calibration consists of adjustments to balance the airflow within the cabinet. THIS WORK SHOULD BE
DONE ONLY BY A QUALIFIED TECHNICIAN WHO CAN MEASURE THE AIRFLOW FROM THE FILTERS WITH A SUITABLE
VELOMETER. NuAire provides two adjustments to balance the airflow within the cabinet.
These are:
a. PWM signal adjust via DC motor speed control.
b. Exhaust filter choke

The PWM signal adjustment establishes the motor speed controls curve starting point of the programmed internal
reference performance curve while the choke adjusts or balances the exhaust airflow as well as makes up for filter
resistance tolerances. Since it has been NuAire's experience that the filters may not "load" evenly, choke adjustments may
be necessary for proper cabinet performance.

OM0215 Page 35 of 59
November/2015
Stored Curve versus Fixed Duty Cycle Running Mode
Normally for the NU-477 application, a stored motor loading curve is used for each size of cabinet established by setting
manual jumpers on the main control board (as referred to on the Electrical Schematic). A fixed duty cycle is also available
for special application cabinets where the stored motor loading curve performance doesn’t meet the special application
cabinet requirements,

The stored curve versus fixed duty cycle running mode can be accessed through the following designated key stroke
sequence.
 Press and hold the ALARM SILENCE key
 Press OUTLET ON, OUTLET OFF and release the ALARM SILENCE key. (If the sequence is properly entered, either
the fluorescent light LED would blink for stored curve running or UV LED would blink for fixed duty cycle running.
Pressing either key will change modes.)
 Press the ALARM SILENCE key to enter the desired running mode. The calibration process and key strokes are the
same for both modes.

DC ECM motor PWM signal DC voltage should also be monitored and recorded upon final calibration. The DC voltage may
be measured using a digital voltmeter. The two test points to measure DC ECM motor voltage are located on the DC motor
connector on the main control board.

The cabinet is considered to be certifiable if the following airflow measurements are present:
a. Downflow average: 60 LFPM  5 LFPM (.30 m/s  .025 m/s).
b. Inflow average: 105 LFPM  5 LFPM (.53 m/s  .025 m/s) using the direct inflow measurement method
or constricted 3 inch (76mm) high access opening measurement method.

The airflow calibration mode is accessed through the following designated key stroke sequence.
 Press and hold the alarm silence key
 Press light on, light off and release the alarm silence key
(If the sequence is properly entered, the outlet LED will blink at 1 second intervals and the blower will turn on.)

Once in airflow calibration mode the PWM signal/DC voltage can be increased (increase airflow) by pressing the UV light
key. The PWM signal can also be lowered (decrease airflow) by pressing the FL light on key. During the process of pressing
these keys the associated LED will indicate a change is being made. When a minimum or maximum PWM signal/DC voltage
is reached the associated LED will remain lit indicating that no further adjustment in that direction is possible.

 Note: Once at the desired airflow setting (after the last increase or decrease key change), the cabinet must run an
additional 60 seconds at this airflow setting for the system to gather averaging data before exiting the calibration
mode that enters the calibration setpoint into the system.

 Note: It is possible to return the PWM signal to its original default setting. While in airflow calibration mode press
and hold both FL light on and FL light off keys simultaneously for 3 seconds, an audible alarm will beep on each of
the 3 seconds. Upon completion of the process, both the FL light on and UV light LED’s will blink 3 times indicating a
successful reset has occurred.

Once the airflow calibration is complete; exit the airflow calibration mode by pressing the alarm silence key.

OM0215 Page 36 of 59
November/2015
7.5.1 Downflow Calibration
Step 1:  Place a velometer in the cabinet workzone on the horizontal plane 4 inches (102mm) above the
bottom edge of the viewing window. Spot check several points on the recommended downflow
velocity test grid found in table 7.0.
Step 2:  If necessary, enter the airflow calibration mode, as stated above and adjust the PWM signal/DC
voltage to achieve desired airflow.
Step 3:  Proceed to inflow calibration.

7.5.2 Inflow Calibration


Step 1:  Measure the inflow velocity using the recommended procedure found in Table 7.0. If necessary,
adjust the exhaust filter choke, located under the front decorative panel, to achieve the correct
average inflow velocity within the stated range of 105  5 LFPM (.53  .025 m/s).
 Less than 100 LFPM (.51 m/s);
First, open the choke plate or make sure it is open. If this is insufficient, then increase the PWM
signal/DC voltage..
 Greater than 110 LFPM (.56 m/s);
First, adjust the PWM signal/DC voltage to achieve 1/2 the exhaust excess, then close the choke
plate to achieve the balance. In this fashion, the downflow should remain nearly constant (i.e.
what the reduced speed took away, the choke plate restores).

Note: The choke plate adjustment requires a standard blade screwdriver. To adjust, loosen the liquid-
tight fitting around the choke adjustment shaft. While monitoring the exhaust flow to check
position, turning the choke adjustment shaft clockwise will open the choke while turning
counter clockwise closes the choke.

Step 2:  Once exhaust adjustment is complete, return the downflow calibration and then check average
downflow velocity. If the downflow average remains within the correct range, the calibration is
complete. If not, readjust as necessary to obtain the correct calibration range. Once entire
cabinet has been balanced, tighten liquid-tight fastener around choke adjustment shaft.

OM0215 Page 37 of 59
November/2015
7.5.3 Flow Gard Calibration
Password
To enter into calibration mode, a 5 keystroke code must be entered.
 Press the [] [] [RESET] [] [] in sequence.
The keypad will stay unlocked until the same 5 keystroke code is entered again.
A single audible beep will confirm the 5 keystroke code sequence was entered correctly.

To calibrate the Flow Gard monitor, the cabinet must first be certified or set to nominal airflow values.
Once the cabinet nominal airflows are set, perform the following procedure:
 Press and hold the [RESET] key for 10 seconds until the display indicates "CAL".
 Press the [RESET] key again. Display should indicate ".53" and "PGM".
 Press the [] or [] key to match the actual calculated inflow velocity.
 Press the [RESET] key to enter the calibration sequence. The monitor will perform the following sequence:
1) 2 short beeps
2) 20 second averaging countdown
3) 2 short beeps, end of sequence
4) Display will indicate "CAL" if successful or "Err" if not successful.
 If the calibration is successful, press and hold the [RESET] key for 2 seconds, the display should indicate normal
readings.
 If the calibration is not successful, press the [RESET] key to acknowledge the error and re-enter the "CAL"
mode. Below, is the calibration error code, along with the cause and correction. Once reviewed, try to
recalibrate using the above sequence.
 Alarm limits are set at the factory to the following specifications.
1. Supply air high velocity limit shall alarm before any airflow reading is 20% above the average
velocity.
2. Supply air low velocity limit shall alarm before any airflow reading is 20% below the average
velocity.
3. Alarm shall also activate before the corrected inflow velocity is 90 fpm (.46m/s) or lower

Error Code Cause and Correction


Airflow below the instrument's calibration range. Check exhaust probe shroud position, verify airflow.
ErL
Airflow above the instrument's calibration range. Check exhaust probe shroud position, verify airflow.
ErH
Too much variation in airflow. Check exhaust probe shroud position, probe gasket
Err
or room air currents in the exhaust airflow area.
Move probe shroud close to the top of HEPA filter.
The calibration point us too close to the low Reduce the low alarm point.
Erc
alarm setpoint Remove the piece of foam that is between the two
diffuser screens inside of the probe shroud.
The calibration point us too close to the high Move probe shroud further from HEPA filter.
ErF
alarm setpoint. Increase the high alarm setpoint.

The incorrect values will not be stored in memory. The monitor will continue to use the previous calibration values
until a correct calibration is successfully completed.

7.5.3.1 Alarm Setpoint Display


The alarm setpoints may be viewed during calibration mode. Press the [] key, the display will toggle
between the current reading and the High alarm setpoint. Press the [] key, the display will toggle between
the current reading and Low Alarm setpoint. Note: Only hold the keys down for 2 seconds at a time.

OM0215 Page 38 of 59
November/2015
7.5.3.2 Alarm Setpoint Calibration
If desired, the Flow Gard high or low alarm limit may be adjusted. During the calibration mode, perform the
following sequence:
 Press and hold either [] or [] key for 5 seconds, display will indicate "PGM" and the current high
or low alarm setpoint.
 Using the [] or [] keys, adjust to the desired alarm limit value.
 Press the [RESET] key to enter the new alarm setpoint value. Monitor will give 2 quick beeps to
acknowledge the saved value and return to run mode.

7.5.3.3 Parameter Configuration Mode


The parameter configuration mode allows limited user interaction for the following table items.

Configuration Parameters
The following table shows the factory default settings for the monitor's various programmable parameters.
A default reset restores these settings.

Configuration Parameter Factory Default


CAL 105 fpm (.53 m/s)
P01 –Digits enabled/disabled Enabled
P02 -Units of measure Metric (m/s)
P03 -Temporary horn disable timer 0 (infinite)
P04 - Warning-to-alarm transition delay timer 1 second
P05 -Alarm-to-warning transition delay timer 5 seconds
P06 -Low alarm warning offset (display will read 2 fpm (0.01m/s))
dEF -Default reset Resets P01 - P11 parameters to factory defaults

To access the parameter configuration mode, perform the following:


 Press and hold [RESET] key for 10 seconds until the display indicates "CAL".
 Press the [] or [] keys to scroll through the menu selections.
 Once the menu selection is found, press [RESET] key to open parameter for change.
 Press the [] or [] keys to alter the parameter value.
 Press the [RESET] key to enter the parameter value desired. The display will flash "PGM" once, and
monitor will give 2 quick beeps to acknowledge the parameter value. The monitor will then return to
the menu.

To exit the Configuration menu, press and hold the Test/Reset button for two seconds. The monitor will also
time out and exit the Parameter Configuration menu after one minute without keypad activity.

 Note: Remember to enter the Password again to lock the keypad.

OM0215 Page 39 of 59
November/2015
P01 Digits Enabled/Disabled
Digits can be enabled or disabled by entering this configuration. The PGM descriptor will turn on and by pressing either
the Up or Down button toggles between the two settings. The status indicators and icons will not be turned off. After
selection, press Test/Reset button to save. Monitor will give 2 quick beeps to acknowledge save and return to the P01
Configuration menu. Advance to another configuration or exit by holding Test/Reset button for 2 seconds.

P02 Units of Measure


Velocity can be displayed in feet per minute (fpm) or meters per second (m/s). After entering this Configuration menu
selection, the monitor will turn on the PGM descriptor and display current units. Press the Up or Down Button to change
setting. Press Test/Reset button to save. The monitor will give 2 quick beeps to acknowledge save and return to the P02
Configuration menu. Advance to another selection or exit by holding Test/Reset button for 2 seconds.

P03 Temporary Horn Disable Timer


During an alarm condition, the alarm can be temporarily silenced by pressing the Test/Reset button. Normally, the alarm
will be silenced for the duration of the current alarm condition. Using the Horn Disable Timer, the monitor can be
configured to enable the horn to come back on after a specified number of minutes. It can also be configured so that the
horn cannot be silenced at all (0), which is the default condition for E models.

After entering this Configuration Parameter, the time can be set to any value from 0-255. If set at 255 and alarm
temporarily silenced by pressing the Test/Reset Button, the alarm will not come on again until this alarm condition clears
and another alarm event occurs. If it is desired to not be able to silence the alarm, set to 0. Any value between 0-255
indicates in minutes the time in which the horn will alarm again if the alarm condition is not corrected. Make any changes
and press Test/Reset Button. Monitor will give two quick beeps to acknowledge save and go back to Configuration Menu.

P04 Warning-to-Alarm Transition Delay Timer


The yellow warning to red alarm transition time is the delay period, in seconds, that a given airflow condition must remain
present before the monitor will go into the appropriate alarm zone. This feature prevents the monitor from toggling back
and forth between zones when a condition is on the border.

This Configuration Parameter sets the warning-to-alarm transition timer. After this configuration menu selection is
entered, the monitor will turn on the program mode PGM descriptor and display the current value for the warning-to-alarm
transition timer. This timer can be set from 0 to 255 seconds.

When desired setting is displayed, press Test/Reset Button. The horn will give two quick beeps to acknowledge save and
return to the Configuration Menu.

P05 Alarm-to-Warning Transition Delay Timer


The red alarm to yellow warning transition time is the delay period in seconds that an airflow condition must remain
present before the monitor will go into the appropriate warning zone. This feature prevents the monitor from toggling back
and forth between zones when a condition is on the border.

This Configuration Parameter sets the alarm-to-warning transition timer. After this configuration menu selection is
entered, the monitor will turn on the program mode PGM descriptor and display the current value for the alarm-to-warning
transition timer. This timer can be set from 0 to 255 seconds. Once the desired setting is displayed, press the Test/Reset
button. The horn will give two quick beeps to acknowledge save and return to the Configuration Menu.

OM0215 Page 40 of 59
November/2015
P06 Low Alarm Warning Offset
The low warning offset defines the starting point of the low warning zone. It is a value (in the current unit of measure)
that is added to the low alarm setpoint. It determines when the yellow low warning light comes on.

Example: If the low alarm is set at 700 fpm and the low alarm warning offset is set at 50 fpm, the yellow low
warning light will come on at 750 fpm.

This Configuration Parameter sets the low alarm warning offset. After this configuration menu selection is entered, the
monitor will turn on the program mode PGM descriptor and display the current value for the low alarm warning offset.
When the desired setting is displayed, press the Test/Reset button. The horn will give two quick beeps to acknowledge
save and return to the Configuration menu.

P07
P08
P09 None of these are in use
P10 at this time.
P11
P12

dEF Reset Configuration Parameters to Factory Default Settings


All the parameters configurable by the user can be reset to the factory defaults located in the memory of the Biological
Safety Cabinet monitor. After this configuration menu selection is entered, the monitor will turn on the program mode
PGM descriptor and the display will show: rES

Press the Test/Reset button. The PGM descriptor will flash once and the horn will give two quick beeps to acknowledge
that the configuration settings have been set to their factory default settings.

The monitor will return to the dEF Configuration menu selection. Press the Up and Down buttons to advance to another
Configuration Parameter. Press and hold the Test/Reset button for 2 seconds to exit the Configuration menu.

7.6 HEPA Filter Leak Test


In order to check filter and filter seal integrity, the HEPA filter media and seals must be directly accessible, by the measuring
instrument. The challenge material (i.e. PAO) should be supplied in the rear center of the workzone over the intake slots.
The upstream challenge port being common for both filters in located on top of the cabinet.

7.6.1 Supply Filter


The diffuser plate placed below the HEPA to protect the filter during normal usage may be removed as follows:
The diffuser is secured to the cabinet shell by #1/4-20 acorn nuts located immediately behind the front viewing
window. After removing the fasteners, drop the front of the diffuser plate several inches and pull forward gently.
Note that the diffuser is purposely a tight fit - it is held to the back wall of the cabinet interior by a light push - fit
with projecting studs.

OM0215 Page 41 of 59
November/2015
7.6.2 Exhaust Filter
The exhaust filter is typically more difficult to check since protective grills, charcoal filters, or exhaust transitions
could cover the filter. Access panels are usually provided and should be removed. If an air gap exhaust transition
is provided, the air gap must be sealed with duct tape or other suitable means to prevent contaminated air from
migrating into the exhaust efflux. All exhaust blowers/fans should be turned off during the check.

 NOTE: if the upstream challenge port is deemed contaminated and not accessible, use both downflow and exhaust
volume for determining challenge concentrations. Use following area information below with average downflow
velocity and spot-check exhaust velocities as measured to determine volume (CFM) (CMH).

*Supply Area Exhaust Area


Model Size 2 2 2 2
(ft )(m ) (ft )(m )
300E 6.33 (.588) 1.61 (.150)
400E 8.53 (.792) 2.38 (.221)
500E 10.74 (.998) 3.15 (.293)
600E 12.95 (1.203) 3.92 (.364)
* Measured 4 inches above the bottom edge of the window.

Laskin Nozzle Concentration Formula

# Nozzles x 135 CFM x 100 ug/L Challenge


=
Downflow (CFM) + Exhaust (CFM) Concentration (ug/L)

# Nozzles x 229 CMH x 100 ug/L = Challenge


Downflow (CMH) + Exhaust (CMH) Concentration (ug/L)

7.7 Airflow Smoke Pattern Test


The airflow smoke pattern test is performed using a smoke source (i.e. smoke tubes) in and around the cabinet workzone
and access opening to determine a visual representation of the cabinet’s containment performance.
To perform the test, the smoke source should be passed through the following areas:
1. A smoke source shall be passed from one end of the cabinet to the other, along the center line of the work surface,
at a height of 4 inches (102mm) above the top of the access opening.
2. A smoke source shall be passed from one end of the cabinet to the other, 1 inch (25mm) just inside the view
screen, at a height 6 inches (152mm) above the top of the access opening.
3. Pass a smoke source along the edges of the entire perimeter of the work opening approximately 1.5 inches
(38mm) outside the cabinet, with particular attention paid to corners and vertical edges.
4. Pass a smoke source 2 inches (51mm) from the sides up inside of the window at the side channel seals, and along
inside of the cabinet along the top of the work area or immediately below the wiper gasket.

The criteria used to evaluate the smoke patterns is the following:


1. The smoke inside the cabinet shall show smooth downward flow with no dead spots or reflux.
2. No smoke shall escape from inside the cabinet.
3. No smoke refluxes out of the cabinet once drawn in, nor does smoke billow over the worksurface or penetrate
onto it.
4. No smoke shall escape from the cabinet.

OM0215 Page 42 of 59
November/2015
7.8 Cleanliness Classification Test for Pharmacy Application
1
If this cabinet is going to be used within pharmacy, per USP797 , the cabinet must be tested to assure compliance to ISO
2
14644-1:1999, Cleanrooms and Associated Controlled Environments, Part 1: Classification of Air Cleanliness . The
cleanliness classification test is performed using a particle counter to measure particle counts within the cabinet workzone.
Turn on cabinet and let warm up for several minutes. Turn on particle counter and flush out sample tubing line to remove
latent particles. Set the particle counter to measure 0.5 micron or larger particles at the appropriate measuring rate.
3
“Operational Particle Count Test ”
Position the particle counter isokinetic probe at a point 6 inches (152mm) upstream of the aseptic manipulation area (hand
convergence point) and mounted so as not to interfere with the operator’s hand movement. The pharmacy operator will
simulate IV manipulation during the particle count test using non-hazardous materials. A minimum of three (3) 1-minute
particle counts shall be sampled and recorded while the user simulates aseptic compounding manipulations.

“At Rest Particle Count Test”


Take 5 test points in 1-minute intervals on a grid, in a horizontal plane as measured approximately 6-inches (152mm) above
the worksurface. The grid location is designed as the workzone centerpoint and each corner measured 6-inches (152mm)
from the inside perimeter.

Record the 5 particle count values for each of the test points over the 1-minute sample time. All final count particle
concentrations and calculated 95% upper confidence limit shall not exceed 3520 particles per cubic meter (ppcm) or (100
particles per cubic feet (ppcf).

1
USP28-NF23: United Stated Pharmacopeial Convention, Inc., 12601 Twinbrook Parkway, Rockville, MD
20852, USA, www.usp.org.
2
ISO 14644-1:1999 Cleanrooms and Associated Controlled Environments-Classification of Air Cleanliness, International
Organization for Standardization, Case Postale 56, CH-1211 Geneve 20, Switzerland
3
CAG-002-2006: CETA Compounding Isolator Testing Guide, Controlled Environment Testing Association, 1500 Sunday
Drive, Suite 102, Raleigh, NC 27607, USA, www.cetainternational.org

OM0215 Page 43 of 59
November/2015
Table 7.0
Recommended Measurement Methods for Cabinet Downflow & Inflow

A. Downflow Measurement
a. Recommended Instruments: TSI 8355 Thermoanemometer
b. Procedure:
Supply filter efflux is measured on a grid, in a horizontal plane 4 inches (102mm) above the bottom edge of the
window. No reading should be taken closer than 6 inches (152mm) from the inside perimeter.
c. Test Data - Inches (mm):

300E 4.297 12.891 21.485 30.078


(109) (327) (546) (764)
400E 5.797 17.391 28.982 40.578
(147) (442) (736) (1031)
500E 7.297 21.891 36.485 57.078
(185) (556) (927) (1297)
600E 8.797 26.391 43.985 61.578
(223) (670) (1117) (1564)
6.348
(163)
19.312
(491)

Number of Readings: Average Velocity ft./min.(m/s)

d. Acceptance Criteria:
1. Average downflow velocity = 55 to 65 fpm (.28 to .33 m/s)
2. Individual readings must be within +20 percent of the average downflow velocity.

B. Inflow Measurement
a. Recommended Instrument: Shortridge Flowhood ADM-870 or TSI 8355 Thermoanemometer.
b. Primary Procedure:
The primary procedure to determine inflow velocity uses a Direct Inflow Measurement (DIM) Instrument (i.e.
Shortridge Flowhood). The DIM Instrument can be used directly on the cabinet with NO CORRECTION FACTORS
REQUIRED if operated in the local density default mode. NSF has tested the cabinet and established listed air
velocities expressed in local density. The DIM Instrument should be equipped with a flowhood that is as close as
possible to the width of the cabinet (i.e. NU-477-400 should use 1 x 4 foot flowhood). The DIM Instrument should
also be duct taped to the cabinet to prevent any sneak air paths from occurring.

The DIM Instrument will read inflow volume (i.e. CFM). Use the window access opening area to calculate inflow
velocity.

Alternate Procedure:
The alternative procedure to determine inflow velocity uses a thermoanemometer in a constricted window access
opening of 3 inches (76mm) with the armrest removed. Inflow air velocity is measured in the center of the
constricted opening 1-1/2 inches (38mm) above the work access opening on the following specified grid. Use the
correction factor table to calculate the inflow velocity.

c. Test Data - Inches (mm):


1. DIM Measurement

3 3 2 2
Inflow Volume t. /min.(m /s) Access Opening ft. (m ) Inflow Velocity ft./min(m/s)

OM0215 Page 44 of 59
November/2015
2. Constricted 3 inch (76mm) high access opening measurement - Inches (mm):

4 8.3966 12.792 17.188 21.584 25.980 30.375


300E
(102) (213) (325) (437) (548) (660) (771)
4 8.264 12.528 16.792 21.056 25.320 29.584 33.848 38.112 42.375
400E
(102) (210) (318) (426) (535) (643) (751) (860) (968) (1076)
4 8.198 12.396 16.594 20.792 24.990 29.188 33.386 37.584 41.782 45.980 50.178 54.375
500E
(102) (208) (315) (421) (528) (635) (741) (848) (955) (1061) (1168) (1274) (1381)
4 8.158 12.316 16.474 20.632 24.790 28.948 33.106 37.264 41.422 45.580 49.738 53.896 58.054 62.212 66.375
600E
(102) (207) (313) (481) (524) (630) (735) (841) (946) (1052) (1158) (1263) (1369) (1475) (1580) (1686)

Average Velocity of Constricted Area ft./min.(ms)


Number of Readings:

2 2 3
Average Velocity fpm (mps) X Constricted Area ft (m ) = Constricted CFM (m /s)
of Constricted Area Area Volume

CFM (m /s)  8" (203mm)


3 2 2
Constricted Area Volume ft (m ) = Average Velocity fpm (mps)
Access Window Area of 8" (203mm)
Access Window Area

Average Velocity of 8" (203mm) fpm (mps) X Correction Factor =Average Inflow Velocity fpm (mps)
Access Window Area

d. Acceptance Criteria:
1. Access Opening Inflow Velocity = 100 to 110 fpm (.51 to .56 m/s)

Areas/Correction Factors for Calculations


3" (76mm) Constricted 8" (203mm) Correction Factor for
Cab. Size
Window Access Area Window Access Opening 8" (203mm) Window
.72 1.91
300E 1.22
(.067) (.177)
.97 2.58
400E 1.22
(.090) (.240)
1.22 3.24
500E 1.22
(.113) (.301)
1.47 3.91
600E 1.18
(.137) (.363)

OM0215 Page 45 of 59
November/2015
7.10Main Control Board Description & Replacement
The main control board consists of one Printed Circuit Board (PCB) assembly. The assembly consists of a power supply,
relay logic and an independent motor speed control.

Disconnect electrical power from the unit before attempting any maintenance action.

The main control board is fastened to the control center with (6) 6-32 studs, lockwashers and nuts. All electrical
connections are made with removable terminals and/or Faston connectors. All AC circuits are fuse protected and when
replacement is necessary, USE ONLY FUSES OF SAME TYPE AND RATING FOR PROTECTION AGAINST RISK OF FIRE.

DESCRIPTION: BLOWER FUSE OUTLET FUSE LIGHT FUSE TRANSFORMER FUSE


FUSE TYPE: TIME-LAG TIME-LAG TIME-LAG TIME-LAG
FUSE SIZE: 1/4 X 1-1/4 INCH 5 X 20MM 5 X 20MM 5 X 20MM
NU-477-300E 5 AMPS 3 AMPS (2) 1 AMP 1/4 AMP (2)
NU-477-400E 6.3 AMPS 3 AMPS (2) 1 AMP 1/4 AMP (2)
NU-477-500E 8 AMPS 3 AMPS (2) 1 AMP 1/4 AMP (2)
NU-477-600E 8 AMPS 3 AMPS (2) 1 AMP 1/4 AMP (2)

The main control board has several configuration jumpers used for different purposes. J1, J2, and J3 are used for the
blower password and night setback mode selection/configuration. J4 and J9 are not used for this application and J6 & J7
are used to configure the ECM motor curve for the appropriate size of cabinet (see chart below).

Cabinet Jumper Configuration


J6 J7 Cabinet Width
Off Off 3 foot
On Off 4 foot
Off On 5 foot
On On 6 foot

The DC ECM motor PWM signal/DC voltage can be measured and monitored. The DC voltage may be measured using a
digital velometer. The two test points to measure the PWM signal/CD voltage are located on the DC motor connector on
the main control board. PWM signal/DC voltage will range between 2.5 and 8.0 Vdc during normal operation and
approximately 2.0 Vdc during night setback mode.

OM0215 Page 46 of 59
November/2015
8.0 Error Indicators & Troubleshooting

Audible alarms and error indicators occur for a variety of reasons. Whenever an alarm condition is present, the audible
alarm and error indicator will be presented and stay on until the error is cleared. When presented with an error indicator,
please perform the following:
Step 1: NOTE ALL ERROR INDICATORS.
When the cabinet is running, any and all red indicators display an error.

Step 2: VERIFY ERROR INDICATORS.


Error indicators can be verified by turning the error function on/off.

Step 3: MONITOR RE-OCCURRENCE OF ERROR INDICATORS.


If re-occurrence of the error indicator is immediate or daily, use guide below to correct the situation.

Error Indicator Troubleshooting Guide


Error Indicator Indicator Correction
Window alarm Sliding window is above or below its Verify standard working height and window micro switch operation.
standard working height or micro switch
is not operating properly.
Cabinet fluorescent lights Check light fuse on main control board in control center.
won't turn on. Check fluorescent lamps.
Check voltage to light ballasts.
Check ballast.
Check light switch.
Cabinet blower won't turn on. Check sliding window for correct operational height.
Check blower fuse on main control board in the control center.
Check voltage to blower on main control board in the control
center. At motor voltage regulator and at bulkhead connector.
Check wiring to blower.
Check blower capacitor.
Check sliding window blower cutoff microswitch.
Check blower motor.
(Note: blower motor has internal thermal protector. Let blower
motor cool off for a minimum of 30 minutes to assure thermal
protector is not open.)
Cabinet outlets won't turn on. Check outlet fuse on main control board in control center.
Check voltage to outlets.
Cabinet ultraviolet light Check light fuse on main control board in control center.
won't turn on. Check ultraviolet lamp.
Check voltage to ultraviolet ballasts.
Check ballast.
Check light switch.
Blower/lights/outlet fuses Check for short on main control board.
continue to blow. Isolate output of circuit by:
disconnecting control center connectors, light circuit, motor
voltage regulator, etc. to isolate the short.
Flow Gard Low Alarm Red LED/LO Check for proper blower operation.
Check paper catch.
Check for proper airflows.
Check calibration.
Check exhaust sensor shroud orientation.
Flow Gard High Alarm Red LED/HI Check cabinet for proper airflows.
Check calibration.
Check exhaust sensor shroud orientation.

OM0215 Page 47 of 59
November/2015
9.0 Remote Contacts

9.1 Fan Relay


The fan relay contacts are single pole normally open contact closure outputs which are activated whenever the blower is
turned on. The contact points are located on the main control module.
Contact ratings are 250 VAC maximum at 2 Amps.

10.0 Optional Equipment

10.1 Ultraviolet Lamp

Ultraviolet light will injure your eyes. Avoid direct viewing at all times.
! CAUTION Personnel should not be present when ultraviolet lamp is on.

10.1.1 Overview
The germicidal ultraviolet is primarily intended for the destruction of bacteria and other microorganisms in the air
or on directly exposed surfaces. Approximately 95% of the ultraviolet radiation's from germicidal tubes are in the
253.7 nanometer region. This is a region in the ultraviolet spectrum which is near the peak of germicidal
effectiveness. The exposure necessary to kill bacteria is the product of time and intensity. High intensities for a
short period of time, or low intensities for a longer period are fundamentally equal in lethal dosage on bacteria
(disregarding the life cycle of bacteria). The intensity of light falling on a given area is governed by the inverse law;
that is the killing intensity decreases as the distance increases from the tube.

The germicidal tube is placed in the cabinet to provide an average intensity of 100 microwatts per centimeter
(for a new tube) falling on a horizontal plane defined by the bottom of the work surface. The minimum
requirement per paragraph 5.12 of NSF Standard 49 is 40 microwatts per square centimeter (ref. NSF Std. #49,
June, 1976).

Since ultraviolet rays will not penetrate ordinary glass, it is recommended that the sliding window be closed while
the ultraviolet light is on within the cabinet; or that personnel leave the cabinet face area.

10.1.2 Operation
The operation of the ultraviolet lamp is accomplished by closing the sliding window and pressing the UV switch
located on the front panel.

10.1.3 Precaution
The rays from germicidal tubes may cause a painful but temporary irritation of the eyes and reddening of the skin,
if of sufficiently high intensity, or if exposure covers a prolonged period of time. For this reason, one should avoid
direct eye and skin exposure to ultraviolet light. If exposure cannot be avoided, it is necessary for personnel to
wear eye goggles or face shields, and long sleeve gowns with rubber gloves.

Since ultraviolet rays will not penetrate ordinary glass, it is recommended that the sliding window be closed while
the ultraviolet light is on within the cabinet; or that personnel leave the cabinet face area.

10.1.4 Maintenance
The output of an ultraviolet lamp deteriorates with burning age. The useful life of the lamp is approximately 7000
hours under specific test conditions.

 Note: Before testing with lamp off, the light may be cleaned with a lint-free cloth dampened with
alcohol or ammonia and water

It is recommended that either a time schedule is established or the tube's output is measured periodically and the
tube replaced when its output falls below 40 microwatts per square centimeter or exceeds 7000 hours of
operation. Lamps should be allowed to operate approximately 5 to 10 minutes (longer when the lamp is in low
temperatures) to warm up sufficiently and wiped clean of dust or dirt before reading the output with a meter.
Even minute amounts of dust will absorb ultraviolet energy.

OM0215 Page 48 of 59
November/2015
Energies Required to Destroy Some Microorganisms by Ultraviolet Radiations(e)

Microwatt Microwatt
Mold Spores seconds Protozoa seconds
per cm/2 per cm/2
Penicillum roqueforti 26,400 Paramecium 200,000(a)
Penicillium expansum 22,000
Penicillium digitiatum 88,000 Nematode Eggs 40,000(b)
Aspergillus glaucus 88,000
Aspergillus flavus 99,000 Algae 22,000(c)
Aspergillus niger 330,000
Rhizopus nigricans 220,000 Virus
Mucor racemosus A 35,200 Baceriophage (E. Coli) 6,600
Mucor racemosus B 35,200 Tobacco Masaic 440,000
Oospora lactis 11,000 Influenze 3,400(d)
Yeasts
Saccharomyces 13,200
ellipsoideus 17,600
Saccharomyces cerevisiae 13,200
Brewers' yeast 6,600
Baker's yeast 8,800
Common yeast cake 13,200
Bacteria
Streptococcus lactis 8,800
Strep. hermolyticus (alpha type) 5,500
Staphylococcus aureus 6,600
Staphylococcus albus 5,720
Micrococcus sphaeroides 15,400
Sarcina lutea 26,400
Pseudomonas fluorescens 7,040
Escherichia coli 7,040
Proteus vulgaris 7,480
Serratia marcescens 6,160
Bacillus subtilis 11,000
Bacillus subtilis spores 22,000
Spirillum rubrum 6,160

References:

(a) Luckiesh, Matthew (1946) Application of Germicidal, Ethyemal and Infrared Energy, D. Van Nostrand o., New York, New York,
pp 253
(b) Hollaender (1942) Aerobiology, A.A.A.S. (for 90% inactivation), pp 162
(c) Ellis, C. and Wells, O.O. (1941) The Chemical Action of Ultraviolet Rays, Reinhold Publishing Corp., pp. 713-714
(d) Hollaender, A., Oliphant, J.W. (1944) The inactivation effect of monochromatic ultraviolet. Radiation on Influenze Virus (for
90% inactivation) Jour. of Bact. 48, pp. 447-454
(e) This table, "Energies Required to Destroy Some Microorganisms by Ultraviolet Radiations" comes from Westinghouse brochure
entitled - "Westinghouse Sterilamp Germicidal Ultraviolet Tubes"

OM0215 Page 49 of 59
November/2015
11.0 Electrical/Environmental Requirements

11.1 Electrical (supply voltage fluctuations not to exceed +/- 10%)


*NU-477-300E 230 VAC, 50/60 Hz, 1 Phase, 6 Amps
*NU-477-400E 230 VAC, 50/60 Hz, 1 Phase, 10 Amps
*NU-477-500E 230 VAV, 50/60 Hz, 1 Phase, 11Amps
*NU-477-600E 230 VAC, 50/60 Hz, 1 Phase, 11 Amps

*CE Certified

11.2 Operational Performance (for indoor use only)


Environment Temperature Range: 60F-85F (15C - 30C)
Environment Humidity: 20% - 60% Relative Humidity
Environment Altitude: 6562 Feet (2000 meters) Maximum

11.3 Light Exposure


Standard Fluorescent Lighting @ 150 ft. candles (1614 LUX) maximum intensity.

11.4 Installation Category: 2.0


Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to
withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example,
in CAT II, which is the category used for instruments in installations supplied from a supply comparable to public mains
such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500 V
for a 230 V supply and 1500 V for a 120 V supply.

11.5 Pollution Degree: 2.0


Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity
caused by condensation.

11.6 Chemical Exposure


Chemical exposure should be limited to antibacterial materials used for cleaning and disinfecting. CHLORINATED AND
HALOGEN MATERIALS ARE NOT RECOMMENDED FOR USE ON STAINLESS STEEL SURFACES. Chamber decontamination
can be accomplished by paraformaldehyde, vapor phased Hydrogen Peroxide or Chlorine Dioxide without degradation of
cabinet materials.

11.7 EMC Performance (classified for light industrial)


Emissions: EN61326
Immunity: EN61326

Class A equipment is intended for use in an industrial environment. In the


documentation for the user, a statement shall be included drawing attention to
! WARNING the fact that there may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted as well as radiated
disturbances.

OM0215 Page 50 of 59
November/2015
12.0 Disposal and Recycle

Cabinets that are no longer in use and are ready for disposal contain reusable materials. ALL components with the
exception of the HEPA filters may be disposed and/or recycled after they are known to be properly disinfected.

 Note: Follow all local, state and federal guidelines for disposal of HEPA filter solid waste.

BIOHAZARD

Prior to any disassembly for disposal the cabinet must be decontaminated


! CAUTION

RECYCLE LEAD FREE

Component Material
Base Cabinet Stainless Steel
Front Grill Stainless Steel
Worksurface Stainless Steel
Window Faring Stainless Steel
Window Glides HDPE
Window Safety Glass
Window Frame Stainless Steel
Front Service Panel Painted Steel
Front Decorative Panel Painted Steel
Control Center Painted Steel
Supply Diffuser Aluminum
Exhaust Filter Aluminum
HEPA Filter Frames Painted Steel
Hepex Bag PVC
Blower Wheel & Housing Steel
Motor Various Steel/Copper
Printed Wiring Assembly Lead Free Electronic
Wire PVC Coated Copper
Ballasts Various Steel, Electronic
Armrest PVC
Connectors Nylon
Hardware Stainless Steel and Steel

 Note: Material type can be verified with use of a magnet with stainless and aluminum being non-magnetic.

OM0215 Page 51 of 59
November/2015
OM0215 Page 52 of 59
November/2015
OM0215 Page 53 of 59
November/2015
OM0215 Page 54 of 59
November/2015
OM0215 Page 55 of 59
November/2015
OM0215 Page 56 of 59
November/2015
OM0215 Page 57 of 59
November/2015
OM0215 Page 58 of 59
November/2015
OM0215 Page 59 of 59
November/2015

You might also like