Error Messages 3500 3flex v4 X
Error Messages 3500 3flex v4 X
ERROR MESSAGES
June 2016
(Ver 4.x)
Error Messages for the 3500 3Flex
New entries have been found in this directory. Refresh the directory inform-
2433
ation?
Cause: Several analyzer files (sample information, analysis conditions, adsorptive properties,
or report options) have been added to this directory by some function other than the
analyzer program.
Action: Click Yes to update the directory information with data from each new file. This oper-
ation may take a minute. Click No to locate the file manually. This option may be feas-
ible if a large number of files have been copied into the directory and the file name is
known.
2434 File [n] — Subset [n] wrote wrong [n] of data, expected [n] bytes.
Cause: An internal processing and/or hardware error has occurred.
Action: Contact your Micromeritics service representative.
The library directory [n] does not exist. Please re-install to make use of Windows
2448
libraries for file management.
The library directory [n] cannot be used. Please re-install to make use of Win-
2448
dows libraries for file management.
Cause: The .INI file contains the directory specified in the error message, however, the
directory has been moved or deleted.
Action: The installation directory will be substituted. The next time a file is opened, use the dir-
ectories list to move to the correct directory.
The export data file directory [n] cannot be used. Resetting to the installation dir-
2448
ectory.
The z-table file directory [n] cannot be used. Resetting to the installation dir-
2448
ectory.
Problem diagnostic directory [n] is invalid. Resetting to the installation dir-
2448
ectory.
2448 User python script directory [n] is invalid. Resetting to the installation directory.
Default script test file directory [n] is invalid. Resetting to the installation dir-
2448
ectory.
Default mass spectrometer file directory [n] is invalid. Resetting to the install-
2448
ation directory.
Cause: A working directory specified in the .INI file is invalid, has been moved, or has been
deleted.
Action: The installation directory will be substituted. The next time a file is opened, use the dir-
ectories list to move to the correct directory.
Sample Defaults may not be edited while this operation is in progress. Do you
2450
wish to save and close the Sample Defaults edit session?
Cause: An automatic analysis (an analysis in which sample files are created using the
defaults) was processing while editing the defaults.
Action: Finish the edit session of the defaults, close the window, then restart the automatic
analysis.
Cause: The instrument is performing an operation which is not recognized by the application.
Action: Click Yes to have the application cancel the operation. Click No to allow the operation
to continue.
Action A: Usually this message is preceded by another message giving the reason for the instru-
ment’s failure to initialize. Refer to the instructions for that message.
Action B: Ensure that the instrument is attached to the computer on the communications port
configured with the Setup program. Verify that the instrument's power switch is in the
ON position and that the light on the front panel is illuminated. Contact a Micromeritics
service representative if this error message continues.
[n] communications port [n] specified in the program control files is already in
2474
use. Unit cannot initialize.
Cause: The communications port assigned to the indicated unit is in use by another program.
Action: End the program using the busy port or use the Setup program to change the com-
munications port assigned to this program.
[n] communications port [n] specified in the program control files cannot be
2475
accessed. Unit cannot initialize.
Cause: The operating system has prevented the application from accessing the com-
munications port.
Action: Review the hardware configuration of the computer. Ensure no other application is
using the port. Contact a Micromeritics service representative if this error message
continues.
2484 The edit session for [n] must be saved before the analysis. Save changes and
2485 The sample file [n] has an invalid status and cannot be used for this operation.
2485 The service test file has an invalid status and cannot be used for this analysis.
Cause: The selected file has a status other than No Analysis.
Action: Select a different sample file, or create a new one and click Replace All to copy para-
meters from the file originally selected.
2486 Could not construct [n] report type. Program will terminate.
2487 Could not start report generator. Error code [n]. Program will terminate.
Cause A: Full rights to the application’s folders and files is required.
Action A: Contact a system administrator to have full rights granted.
Cause B: An internal processing and/or hardware error has occurred.
Action B: Contact your Micromeritics service representative.
2491 Highlighted fields contain errors. Please correct the errors before closing.
Cause: The fields highlighted in red contain invalid entries. The window cannot be closed until
all errors are corrected.
Action: Check the entries, correct the errors, then close the window.
2498 The requested change to the Sample’s status is invalid at this time.
Cause: A request to change the file’s status, for example, from automatically collected to
manually entered could not be done.
Action: Contact your Micromeritics service representative. Record the name of the sample
file in which the problem occurred.
All sample file names that can be created using the sequence number pattern
2500
already exist. You may want to modify the next sequence number.
Cause: No more sample files can be created using the currently entered file name sequence
number.
Action: Go to Options > Default Method, then enter another sequence number.
System resources have reached a dangerously low level. Please close some win-
2501
dows to avoid the loss of data.
Cause: A large number of windows are open and consuming the system resources available
to all applications.
Action: Close one or more windows. Contact your Micromeritics service representative.
2505 Error logger cannot be initialized. Error code [n]. Program will exit.
Cause: An internal processing error has occurred.
Action: Contact your Micromeritics service representative.
2506 Sample file [n] has a No Analysis status and cannot be used for this operation.
Cause: The selected sample file does not have collected data and cannot be used for oper-
ations, for example, reporting.
Action: Enter the name of a file with a status of Complete, Analyzing, or Entered. Altern-
atively, select a sample file from the list box.
2507 The sample has an invalid status and cannot be used for degassing.
Cause: A sample file has been selected which does not have a No Analysis or Prepared
status.
Action: Select a different file with a status of No Analysis or Prepared.
2508 The selected file [n] does not contain a valid script.
Cause: The selected service test file does not contain any steps or the file has become cor-
rupted.
Action: If the service test file has no steps, open the service test file and add steps to the file
using the Service Test Script tab.
Warning: Changing the calibration information will affect the performance of the
2515 instrument. Only qualified service personnel should do this. Do you wish to pro-
ceed?
Cause: The process of performing a calibration operation was started.
Action: Calibration operations should only be done by or under the direction of qualified ser-
vice personnel.
Canceling this dialog will reset the calibration state to what it was when this dia-
2517
log was first opened. Are you sure you want to cancel?
Cause: The calibration has not been accepted.
Action: If the calibration operation was successful, click Accept.
2532 The instrument contains a different software version. Do you want to reset it?
Cause: The application has discovered a different version of software operating in the ana-
lyzer.
Action: If there are no analyzers other than the one connected to the computer, click Yes,
then allow the updated software to load.
The file [n] could not be written. Check that the directory permissions are cor-
2542
rect.
Cause: The person logged into the computer does not have write access to the program
and/or data directory.
Action: Contact your system administrator set up proper privileges for all users
2577 The python directory is missing or some of its contents have been removed.
Cause: When using Advanced reports, a necessary component is missing.
Action: Re-install the software.
The selected file has a status that is not compatible with the operation reques-
2579
ted.
Cause: The selected sample file cannot be used for the selected operation. For example, a
sample file with a Complete status cannot be used for analysis.
Action: Select a different sample file.
Cause: Files necessary for the application are missing or have been corrupted.
Action: Re-install the software.
Please re-install to make use of Windows libraries for file management. The fol-
2585
lowing libraries are missing: [n]
Cause: This message is triggered on application start up if any of the library files used by an
application, do not exist on disk.
Action: Add the library into the libraries.
2588 Sample file [n] is a [n] file and can not be analyzed on this instrument.
Cause: The specified sample file is from another Micromeritics application which is not usable
with this application.
Action: Select a different sample file.
2592 The selected file has an extension that is not supported by this operation.
Cause: The selected file does not have a supported file extension.
Action: Open the adsorptive properties file. Open the .FPI file selector and select another file
with a supported file extension.
Warning: The selected file uses Unicode (UTF-8 or UTF-16LE) encoding with
2593
non-ASCII characters.
The selected file uses Unicode [UTF-16BE or UTF-32) encoding and could not
2594
be read.
Cause: The Advanced report is not saved in ASCII format, which is required by Python.
Action: Edit the file and save it in ASCII (ANSI) format.
2595 The selected file is too large (maximum allowed size is [n]).
Cause: The Advanced report is too large.
Action: Edit the file and reduce the size.
All rows of [n] do not have the same number of entries. Some rows will be filled
2598
or truncated.
Cause: The gas compressibility table for the analysis gas contains inconsistencies.
Action: Load a correct fluid property information file .
[n] does not have the correct format. Please load a fluid property information
2599
file.
Cause: The selected file was not a fluid property file.
Action: Select a file with fluid property information.
2606 Can not load [n] because the fluid property table is too large.
Cause: The fluid property information file is too large to be stored in a sample file.
Action A: Verify that the selected file is really a fluid property information file.
Action B: Delete unneeded rows and columns from the compressibility factor table.
The fluid property file [n] is obsolete. Contact Micromeritics for the current FPI
2607
file for the adsorptive.
Cause: An old FPI which may be missing important information was loaded.
Action: Contact Micromeritics to get an updated FPI file for that adsorptive.
4005 Pressures were not smoothed. Not enough pressures below 0.10 [n].
Cause: The Smooth pressures below 0.10 P/P0 option was selected on the Report Options
window. There must be at least 10 pressures within this range for smoothing to occur.
Action: Deselect this option to disable this message.
4011 Analysis gas in sample file does not match analysis gas in unit.
Cause: An attempt was made to start an analysis using a sample file in which the analysis gas
specified does not match the analysis gas entered in the unit configuration.
Action A: If the wrong adsorptive was selected in the sample file, change the adsorptive in the
file.
Action B: If necessary, attach the appropriate gas cylinder, then enter the gas in Unit Con-
figuration.
4012 Psat gas in sample file does not match any gas in the unit.
Cause: If using Measure psat of a gas in Po and T options in Analysis Conditions, the selected
gas is not one of the selected gases in Unit Configuration.
Action A: If the incorrect psat was selected, change the psat gas.
Action B: If the gas was recently connected to the instrument, update the Unit Configuration.
4013 The Incremental Dosing option is not available on the selected unit.
Cause: The MicroPore option was selected but the instrument does not have the micropore
option.
Action: Install the MicroPore option before starting the analysis.
Damage to the instrument will result if the sample [n] has not been manually
4017
evacuated. Have you evacuated the sample?
Disabling this option may damage the instrument. Are you sure that the sample
4020
should not be backfilled?
Cause: Backfill sample at start of analysis was not selected on the Sample Backfill Options
The entered [n] value ([n] and Temperature Option of the Analysis Conditions is
4021 outside the range of the pressures listed in the Psat vs Temperature Table
(Adsorptive Properties).
Cause: The entered P0 value is not within the range of pressures selected for analysis.
Action A: Enter a new P0 value.
Action B: Add more pressures and corresponding temperatures to the Analysis Conditions pres-
sure table to include the presently selected P0 value.
The entered bath temperature value ([n] and Temperature Options of the Ana-
4022 lysis Conditions) is outside the range of the temperatures listed in the Psat vs
Temperature Table (Adsorptive Properties).
Cause: The entered bath temperature is outside of the range of temperatures specified in the
Adsorptive Properties.
Action A: Change the entered temperature.
Action B: Change the adsorptive.
Action C: Add more temperatures and corresponding pressures to Adsorptive Properties.
The file [n] cannot be prepared for analysis. It is open for editing and contains
4023
errors.
Cause: An attempt was made to start an analysis using a file that contains errors and is open
for editing.
Action: Finish editing this file, save and close it, then start the analysis.
4024 Backfill gas in sample file does not match any gas in the unit.
Cause: The backfill gas specified in the sample file does not match the analysis gas entered in
the Unit Configuration.
Action A: If the wrong backfill gas was selected in the sample file, change the backfill gas in the
file.
Action B: If necessary, attach the appropriate gas cylinder, then enter the gas in the Unit Con-
figuration.
4027 Fewer than two sample files have data suitable for heat of adsorption reports.
Cause: Less than two of the selected sample files for heat of adsorption reports contain appro-
priate data.
Action: Edit the Quantity Adsorbed table, or select other sample files.
4029 At least two fitted data points are needed for Alpha-S calculations.
Cause: Fewer than two data points fall within the selected Alpha-s range.
Action: Edit either the calculation pressure in the fitted Alpha-s range, or use a different
reference curve.
4030 P0 cannot be measured at intervals when dosing from the Psat tube.
Cause: The dosing option in Adsorptive Properties conflicts with the P0 option.
Action: Either edit Adsorptive Properties and change the dosing option, or edit the P0 option
in Analysis Conditions.
4031 Not enough points with a relative pressure in the range [n].
Cause: Fewer than two data points selected for the Dubinin report fall within the selected rel-
ative pressure range.
Action: Edit the calculation pressure range or the fitted relative pressure range.
Some summary reports could not be produced because they require the Micro-
4032
pore option.
Cause: Some of the summary reports requested were not produced because the micropore
option was not installed.
Action: Edit the summary report and deselect the micropore options.
vath-Kawazoe window.
Computations failed while processing the primary data set. No reports will be
4037
produced.
Cause: The preparation of data for reporting could not be successfully completed. No Hor-
vath-Kawazoe reports will be produced. This message will always be preceded with
another one containing additional information.
Action: Refer to the error message number which preceded this one for an explanation.
4039 The isotherm does not meet the constraints of the Cheng/Yang assumption.
Cause: The Cheng/Yang correction to the Horvath-Kawazoe equation requires the value of
the monolayer volume as calculated from the Langmuir equation. The isotherm does
not correlate to the Langmuir assumption with a coefficient of 0.98 or more. The cor-
rection is not applicable to this isotherm or to the range of the data points selected.
Action A: The analysis will be performed without the Cheng/Yang correction. Deselect Apply
Cheng/Yang correction on the Horvath-Kawazoe Report Options window to prevent
this message from appearing on future reports.
Action B: Generate the Langmuir report for the same data points selected for the Horvath-
Kawazoe report. If the Langmuir correlation coefficient can be brought above 0.98 by
removing some points at high relative pressure, remove them, then reproduce the
Horvath-Kawazoe reports.
4040 The value of Qm computed from the Langmuir equation is too low.
Cause: The Cheng/Yang correction to the Horvath-Kawazoe equation requires the value of
the monolayer volume as calculated from the Langmuir equation. The computed
value is less than the volume adsorbed at the largest relative pressure included in the
analysis. The correction is not applicable to this isotherm or to the range of the data
points selected.
Action: The analysis will be performed and the Cheng/Yang correction will be applied to all
points with a volume adsorbed less than the value of Vm. The pore size will not be cal-
culated for data points with an invalid volume adsorbed. Deselect Apply Cheng/Yang
correction on the Horvath-Kawazoe Report Options window to clear this message.
4044 Points in the Langmuir report pressure table lie outside the collected data.
4045 Points in the report pressure table lie outside the collected data.
Cause: Calculation pressure range is not being used. More than one of the report pressure
table points is above the range of the collected data and more than one is below.
Action: Change the report pressure table to be more consistent with the collected data.
4048 Warning: An error occurred while restoring the heat of adsorption report editor.
Cause: The state of the heat of adsorption report editor could not be restored. Default settings
will be used.
Action: No action.
The sample [n] does not have enough data. A minimum of two adsorption points
4049
is required.
Cause: A sample file has been included in the Heat of Adsorption report that does not have
enough data.
Action: Remove the file from the selected file list.
None of the requested quantities adsorbed is within the range of the collected
4050
data of more than one sample file.
Cause: The Heat of Adsorption report failed because the specified quantities are not within
the range of the collected data.
Action: Edit the quantities adsorbed so that they are within the range of the collected data, or
select other sample files.
The sample [n] does not have any data in the range of the requested quantities
4051
adsorbed.
Cause: The sample’s data cannot be interpolated to any of the quantities adsorbed.
Action: Edit the quantities adsorbed so that they are within the range of the collected data.
4052 Fewer than two points are selected for this report.
4052 Fewer than three points provided to the interpolation table.
4052 Fewer than three points are selected for this report.
Cause: At least two (or three) points are required for the BET calculations.
Action: Edit the calculation range in the BET report.
4053 At least two data points must be selected for t-Plot calculations.
Cause: At least two points are required for the t-Plot calculations.
Action: Edit the calculation range for the t-Plot report.
4054 Fewer than two data points are inside the fitted thickness range.
Cause: At least two points must be within the fitted thickness range for the t-Plot calculations.
Action A: Edit the calculation range for the t-Plot report.
Action B: Edit the fitted thickness range in the t-Plot report editor.
4055 A positive BET surface area was not calculated. Please check your BET Report.
A positive Langmuir surface area was not calculated. Please check your Lang-
4056
muir report.
Cause: Fewer than two points were assigned to the requested surface area calculation in the
collected data table.
Action A: Assign more points to the surface area calculation.
Action B: Select a different surface area in the t-Plot report editor.
4057 At least two data points are needed for Freundlich calculations.
4058 At least two data points are needed for Temkin calculations.
Cause: Less than two data points have been selected for the specified report; at least two are
required.
Action: Edit the selection of points on the report's interactive editor or on the Pressures win-
dow for the specified report.
4067 No data points are within the range of pressures in the reference isotherm.
Cause: There are no collected data points within the range of pressures in the reference iso-
therm.
Action: Select data points in the range of the reference isotherm, or select a more appropriate
reference isotherm.
4071 The selected pressures points do not form a valid set for deconvolution.
Cause: The data points selected for analysis do not contain enough information to allow a
DFT data reduction.
Action: At least two points with strictly increasing pressures and volumes adsorbed are
required for a DFT Plus data reduction. Edit the selection of data points on the DFT
interactive editor or on the DFT Pressures window.
4072 The range of pressures selected is too small to deconvolute using this model.
Cause: A null result was found using the selected model.
Action: At least two points with strictly increasing pressures and volumes adsorbed are
required for a DFT Plus data reduction. Edit the selection of data points on the DFT
interactive editor or on the DFT Pressures window.
4073 The analysis gas [n] does not match the model gas [n].
Cause: The model assumes a specific gas, and the sample file uses a different one.
Action: Select a model that assumes the same gas.
4074 The analysis temperature [n] does not match the model temperature [n].
Cause: The temperature for the selected model did not match the analysis temperature.
Action: Select a different model.
4075 The models cannot be located in the models folder. Reinstall the software.
Cause: The models could not be located. They may have been inadvertently deleted or
moved.
Action: Reinstall the software.
Action A: Edit the Isotherm report for the named overlay file, then select a surface area option
for Volume Adsorbed.
Action B: Click Overlays on the Report Options window of the primary file, then remove the
named overlay file from the list.
4078 Slope and Y-Intercept cannot be determined from the selected points.
Cause: The Langmuir report cannot be generated from the selected points.
Action: Edit the calculation pressure range in the Langmuir report pressure window.
4080 The compressibility factor table has [n] pressure entries. Only [n] are allowed.
Cause: The fluid property information file has too many rows of compressibility information.
Action: Delete rows for pressures that are not likely to be encountered in the analysis.
In the compressibility factor table for [n], the 4 temperatures below [n] and 7 tem-
4081
peratures above must span at least 10 [n].
Cause: The fluid property information file does not have enough columns of compressibility
information for temperatures near the analysis temperature.
Action: Add rows for temperatures near the analysis temperature.
Compressibility factors must be greater than zero for temperatures near ambi-
4082
ent and pressures up to the maximum manifold pressure.
Cause: Fluid property information indicates that the analysis gas will condense.
Action A: Check that the correct fluid properties are being used.
Action B: Change the pressure or temperature range for the analysis.
Action: Click No and modify the pressure table before restarting the analysis. Click Yes and
perform the analysis omitting the points that would cause condensation.
The compressibility file [n] is obsolete. Please load the fluid property inform-
4095
ation file (.FPI) for the adsorptive.
Cause: The sample file includes obsolete compressibility information from a *.DAT file.
Action: Load the *.FPI file for the adsorptive.
The Analysis Conditions dosing option Low pressure incremental dose mode cannot
4096
be used with the Adsorptive Properties dosing method Selective Adsorption.
The Analysis Conditions dosing option Maximum volume increment cannot be used
4097
with the Adsorptive Properties dosing method Selective Adsorption.
The Analysis Conditions dosing option First pressure fixed dose cannot be used with
4098
the Adsorptive Properties dosing method Selective Adsorption.
The Analysis Conditions dosing option Selective Adsorption requires the Analysis
4099
Conditions choice of Absolute pressure dosing.
The Analysis Conditions choice of Absolute pressure dosing requires that the Adsorpt-
4100
ive Properties dosing method is set to Normal or Selective Adsorption.
Cause: There is a conflict between Analysis Conditions selections and Adsorptive Properties
selections.
Action: Change the Analysis Conditions dosing option or do not use selective adsorption.
4114 The backfill gas [n] in sample file [n] does not match any gas in the unit.
Cause: The selected backfill gas for the Analysis Conditions is not one of the gases con-
nected to the instrument. Go to Unit [n] > Unit Configuration and verify the gas.
Action: Edit the Analysis Conditions backfill gas and change the gas to one which is con-
nected to the instrument.
4115 The P0 [n] in sample file [n] does not match any gas in the unit.
Cause: The selected gas for the P0 and T for the Analysis Conditions is not one of the gases
connected to the instrument. Go to Unit [n] > Unit Configuration and verify the
gas.
Action: Edit the Analysis Conditions P0 and T and change the gas to one which is connected
to the instrument.
4116 The adsorptive [n] in sample [n] is different from port [n] adsorptive [n].
Cause: The adsorptive selected for one analysis port is different from the adsorptive that is in
the sample file selected for another port. All ports must use the same adsorptive.
Action: Edit the Analysis Conditions and change the adsorptive to match the other ports.
4117 Gas [n] in sample file [n] does not match any gas in the unit.
Cause: The adsorptive selected for the analysis is not one of the gases connected to the
instrument.
Action A: Go to Unit [n] > Unit Configuration and verify the gas.
Action B: Edit the Analysis Conditions and change the adsorptive to one which is connected to
the instrument.
4118 The pre-analysis backfill gases differ. Each port must use the same backfill gas.
The post-analysis backfill gases differ. Each port must use the same backfill
4119
gas.
4120 The analysis gases differ. Each port must use the same analysis gas.
The Psat gases differ. Each port must use the same gas for P0 or Psat meas-
4121
urements.
Cause: The type of gas(specified in the error message) selected for one port in the analysis is
different from the adsorptive that is in the sample file selected for another port. All
ports must use the same adsorptive.
Action: Edit the Analysis Conditions and change the gas to match the other port(s).
The dosing source gases differ. When dosing from a sample port or the Psat
4122 tube and the dosing source is charged from a gas inlet, each port must use the
same dosing source gas.
Cause: An incompatibility exists between the dosing method and the analysis gases in the
sample files for the selected ports.
Action: When dosing from a sample port or the Psat tube and the dosing source is charged
from a gas inlet, the sample files selected for all ports must use the same dosing
source gas. Check the analysis adsorptive and analysis adsorptive properties in the
analysis conditions of the selected sample files.
Helium adsorptive properties could not be found in the adsorptive properties lib-
4123
rary.
Cause: There is no Adsorptive Properties file for helium available to the application.
Action: An Adsorptive Properties file for helium was installed with the software. It was likely
deleted. Either re-install the software, or create an Adsorptive Properties file for
helium.
Dosing from port [n] cannot be done because a sample is selected for analysis
4124
on that port.
Cause: The Analysis Conditions of a sample file calls for dosing from a port which has a
sample file assigned to it for analysis.
Action: If the Analysis Conditions are correct for the currently selected sample file, clear the
sample file selection from the port used for dosing. Otherwise, select a different
sample file.
Determination of P0 from the dosing source can only be done if dosing is done
4125
from the Psat tube, sample port, or vapor source.
Cause: If measuring Psat over the source, one of the dosing options listed in the error mes-
sage must be selected.
Action: Select one of the dosing options from the error message.
Warning: Measured ambient temperature is not available for calculating the max-
4126
imum manifold pressure. Analysis will continue using [n].
Cause: Measured ambient temperature is not available. Analysis will continue with a reas-
onable room temperature
Action: Displayed as a warning. Continue the analysis.
The compressibility factor table has [n] temperature entries. Only [n] are
4127
allowed.
Cause: The fluid property information file has too many columns of compressibility inform-
ation.
Action: Delete columns for temperatures that are not likely to be encountered in the analysis.
Action: Edit Analysis Conditions / Adsorptive Properties and choose a different option.
Cause: Calculations failed for the selected sample file in the BJH report.
Action: Select a different sample file.
4150 No valid points within the specified [n]s pore size interval.
Cause: A BJH min/max pore diameter range selection which is too narrow.
Action: Make the pore diameter range wider .
The computer does not have the communications port specified for the
4400
SmartPrep(s). Cannot initialize.
Cause: The communications port associated with this unit was not valid.
Action: Run the setup program and set up the unit on a valid port.
4401 The communications port specified for the SmartPrep(s) is already in use. Can-
not initialize.
Cause: The communications port associated with this instrument is in use by some other pro-
gram in the system.
Action: Close the other program to release the port. Restart the analysis application.
The communications port specified for the SmartPrep cannot be accessed. Can-
4402
not initialize.
4403 Cannot communicate with SmartPrep Unit [n].
Cause: The communications port associated with this unit was not valid.
Action: Run the setup program and set up the unit on a valid port.
4407 Error searching for installed Smart VacPreps. The registry could not be read.
Cause: The application was not installed properly.
Action: Reinstall the application. Contact a Micromeritics service representative if this error
message continues.
The [n] in [n] already has Smart VacPrep S/N [n]. The Smart VacPrep must be
4408
removed from the [n] before it can be added.
Cause: The Smart VacPrep was already installed for another application / unit.
Action: Remove the Smart VacPrep from the installed unit before adding it to the preferred
unit.
4409 A free IP address on the same subnet as [n] could not be found.
Cause: All IP addresses on the network for the Ethernet card specified during installation are
in use by other Micromeritics applications on this computer.
Action A: Uninstall unused Micromeritics applications.
Action B: Configure a different Ethernet card for use by the application using the application
installer.
4410 The .INI file could not be updated with configuration for Smart VacPrep S/N [n].
Cause: The application .INI file is opened by another application and could not be updated.
Action: Close all open applications and add the Smart VacPrep again using the Smart
VacPrep menu.
4413 Error overheating on port [n]. Current = [n], Target = [n], Limit = [n].
Cause: The temperature of the indicated mantle exceeded the maximum allowed value.
Action: Ensure the power and thermocouple connectors for the mantle are properly installed.
Contact a Micromeritics service representative if this error message continues.
Action: Ensure the thermocouple is plugged in. Contact a Micromeritics service rep-
resentative if this error message continues.
Degas transducer zero calibration failed. Current Offset = [n] counts, Current
4415
Pressure = [n], New Offset = [n] counts, Nominal = [n] counts.
Cause: The pressure transducer offset exceeds the recommended limit.
Action: Ensure that the vacuum pump is on. Repeat the pressure offset calibration. Contact a
Micromeritics service representative if this error message continues.
Degas transducer scale calibration failed. Reference = [n], Current = [n], New
4416
Scale = [n] / count, Nominal = [n] / count.
Degas vacuum gauge low point calibration failed. Reference = [n] / count, Cur-
4417
rent = [0] counts.
Cause: The transducer offset calibration was rejected.
Action: Contact your Micromeritics service representative.
Degas vacuum gauge high point calibration failed. Reference = [n], Current = [0]
4418
counts.
Cause: The vacuum gauge calibration was rejected.
Action: Contact your Micromeritics service representative.
4421 Smart VacPrep S/N [n] is busy and could not be removed.
Cause: The Smart VacPrep cannot be removed because it is currently performing an oper-
ation.
Action: Wait until the Smart VacPrep completes the current operation and try again.
The file [n] on port [n] could not be opened. Check if the sample file is already in
4424
use for editing or analysis.
Cause: The selected sample file is already open by this or another application.
Action A: The selected sample file is damaged.
Action B: Select another sample file.
The sample [n] on port [n] has an invalid status and cannot be used for degass-
4425
ing.
Cause: The status of the file is not consistent with the current operation.
Action: Select a sample file that has not been used for an analysis. Only sample files with a
status of No Analysis or Prepared may be selected.
The file does not contain any mass spectrometer data or is missing required
4602
fields. Nothing was imported.
Cause: The selected file does not contain mass spectrometer data.
Action: When exporting the data using the analysis application for the mass spectrometer,
ensure that the signals of interest are selected for export. Export the data to a sup-
ported text format then import the file into the Micromeritics application.
This sample does not contain an experiment which can be used in a [n] cal-
4604
ibration.
Cause: The sample file selected in the calibration editor does not have valid peaks for the type
of calibration.
Action: Select a sample file that contains the appropriate experiment type.
The calibration data for experiment [n] may already be loaded. You may delete
4605
duplicates after loading.
Cause: Before reloading the peak information for a sample file into the calibration editor, all
previously loaded peaks must first be deleted.
Action: Delete all peaks from the calibration editor, then try loading the sample file again.
The file [n] was not created with this instrument or does not contain calibration
4606
data.
Cause: The selected sample file contains data collected on an instrument different from the
instrument associated with the current calibration editor.
Action: Choose a sample file containing calibration data collected on the current instrument.
4607 The calibration data file [n] was not created from an imported mass spec file.
Cause: The selected sample file does not contain mass spectrometer data.
Action: Choose a sample file containing mass spectrometer calibration data collected.
Unable to update the adsorptive property file. Ensure that the selected adsorpt-
4608
ive is not opened for editing.
Cause: The selected adsorptive property file could not be written.
Action: Ensure that the adsorptive property file for the gas selected in the gas flow constant
calibration wizard is not currently open in an adsorptive property editor.
4609 Unexpected I/O error accessing data for an experiment in sample [n].
Cause: A sample file for a running analysis could not be opened with exclusive write access.
Action: Ensure that no applications other than the instrument application has access to the
sample file during an analysis.
4610 Cannot append to an experiment in sample [n]. Maximum size may be exceeded.
Cause: Collected data could not be written to a sample file for a running analysis. This may be
due to the maximum allowed file size for the operating system.
Action: Reduce the amount of data collected during analysis by reducing the data collection
rate or reducing the number of experiments in the analysis conditions task list.
4611 At least two peaks are required to determine first order kinetics.
Cause: An experiment selected in this report contains less than the two peaks required by the
report.
Action: Identify the peaks in the experiment before reporting or deselect the experiment in the
report options before generating the report.
4614 At least 6 peaks are required for loop calibration; the peak table contains [n].
Cause: The number of peaks for the loop calibration is less than six.
Action: Using the experiment peak editor, create baselines for the injection peaks. There
must be at least six injections and corresponding peaks to calculate the loop volume.
New steps cannot be inserted in the synopsis at any point that precedes the
4621
step that is currently being executed.
Cause: The analysis has already progressed past the point at which an attempt was made to
insert a step; therefore the step would not be performed.
Action: Insert the new analysis step after the step or repeat loop that is currently being
executed.
An error occurred while loading the application control information. Data entry
6000
cannot be performed. (Code [n]).
Cause: An error occurred accessing the control information disk file required by this applic-
ation.
Action: The disk drive may have failed or be corrupt. Run diagnostics on the disk drive.
6004 When dosing from the Psat tube, [n] must be entered or calculated.
Cause: The dosing method selected requires the entered or calculated P0 option.
Action: Check the analysis adsorptive properties and the P0 and T options in the analysis con-
ditions of the sample file.
Analysis with krypton requires a 10 mmHg transducer. Port [n] does not have a
6005
10 mmHg transducer.
Cause: Krypton was selected as the adsorptive for a port which does not support it.
Action: Select a different sample file for this port or change the adsorptive in the Analysis
Conditions of the sample selected for the port.
The pressure table has points that may be below port [n] minimum reading of
6006 [n]. Points outside the range of the transducers will be skipped. Do you want to
continue with the analysis?
Cause: Pressures specified in the Analysis Conditions are below the minimum values allowed
for the port.
Action: Click Yes to continue the analysis, omitting the below-minimum points or click No and
edit the pressures in the Analysis Conditions.
The expected saturation pressure [n] exceeds the maximum pressure for this
6009
port [n].
Cause: The Analysis Conditions specify pressures which will produce a saturation pressure
greater than can be measured by the selected port.
Warning: The entered saturation pressure [n] does not agree with the expected
6010
saturation pressure [n] at [n].
Cause: The entered saturation pressure does not agree with the expected saturation pres-
sure at the analysis temperature.
Action: Verify the value of the saturation pressure in the P0 and T options in the analysis con-
ditions of the sample file
6012 The target pressure [n] exceeds the expected saturation pressure [n] at [n].
Cause: Target pressures exceed the saturation pressure determined from the fluid property
information.
Action: Remove targets that are above the saturation pressure.
The fluid property information does not have enough data near [n] to calculate
6014
the saturation pressure.
Cause: When absolute pressure dosing is used, the pressure table is checked to ensure that
the saturation pressure of the adsorptive is not exceeded. This error indicates that the
saturation pressure could not be determined because the sample temperature is not
in the range of the saturation pressures or sublimation pressures in the fluid prop-
erties. This check is skipped if the sample temperature is above the critical tem-
perature of the adsorptive.
Action: Ensure that the entered sample temperature is correct. Ensure that the correct
adsorptive is selected and that it has the displayed name of the fluid properties is cor-
rect. Contact your service representative if there is a problem with the adsorptive prop-
erties.
6019 A source for the selected backfill gas could not be found.
Cause: The backfill gas specified in Analysis Conditions/Preparation is not one of the gasses
known to be connected to the instrument. Go to Unit [n] > Unit Configuration and
verify the selected gas.
Action: Go to Unit [n] > Unit Configuration and verify the valve where Nitrogen is con-
nected.
Step [n]. Pulse and loop calibration experiments are not supported by this hard-
6020
ware configuration.
Cause: Pulse and loop calibration experiments require an instrument equipped with an injec-
tion loop. The injection loop was not detected as an installed option for this instrument.
Action: Perform the analysis on an instrument equipped with the injection loop option. If the
instrument is equipped with this option and this error occurs, contact your Micro-
meritics service representative.
6022 Step [n]. Hardware configuration does not support loop valve operations.
Cause: Loop valve operations require an instrument equipped with an injection loop valve.
The injection loop was not detected as an installed option for this instrument.
Action: Perform the analysis on an instrument equipped with the injection loop option or set
the loop valve to the Fill position in the valve actuation step. If the instrument is
equipped with this option and this error occurs, contact your Micromeritics service rep-
resentative.
6025 Step [n]. Carrier gas [n] does not match any gas in the unit.
6026 Step [n]. Loop gas [n] does not match any gas in the unit.
Cause: An unconfigured gas is selected for use by the specified mass flow controller.
Action: If the gas is connected to the instrument, go to Unit [n] > Unit Configuration and
add the gas to the gas selections.
If the gas is not connected to the instrument, either connect the gas and add the gas to
the gas selections in the unit configuration or change the gas selected for use by the
specified mass flow controller for this step.
Step [n]. A carrier gas must be selected and flowing at least 10 sccm before the
6027
detector can be turned on.
Cause: The TCD is set to enabled by this step, but a carrier gas with a flow rate of at least 10
sccm is required to enable the TCD.
Action: Add a gas flow step configured with a carrier gas flow rate of at least 10 sccm before
the step enabling the TCD, or set the TCD to disabled in the specified step and add a
Set Detector step with the TCD enabled to the analysis conditions task list after a step
configuring the carrier gas flow to at least 10 sccm.
6028 Step [n]. A loop gas must be selected and flowing before loop injection.
Cause A: An unconfigured gas is selected for use by the loop mass flow controller.
Action A: If the gas is connected to the instrument, go to Unit [n] > Unit Configuration and
add the gas to the gas selections.
If the gas is not connected to the instrument, either connect the gas and add the gas to
the gas selections in the unit configuration or change the gas selected for use by the
loop mass flow controller for this step.
Cause B: An invalid loop gas is selected for the specified step.
Action B: Correct the selected loop gas in the specified step.
6030 Step [n]. The detector must be turned on before the start of data recording.
Cause: The TCD has not been set to enabled prior to a TCD recording block.
Action: Add a Set Detector step prior to the TCD recording block with the TCD set to
enabled.
Step [n]. Flow of an inert gas for a minimum of 10 minutes is required between
6031
hydrogen and oxygen flows from the same inlet manifold.
Cause: A switch in gas flow from hydrogen to oxygen or oxygen to hydrogen occurs without
an intervening flow of inert gas. Mixing of hydrogen and oxygen in an inlet manifold is
not allowed.
Action: Add a gas flow step with an inert gas selected between gas flow steps of hydrogen
and oxygen.
Step [n]. Combining of hydrogen and oxygen in carrier and loop gas selections
6032
is not allowed.
Cause: Hydrogen and oxygen are selected as carrier and loop gases in the specified step.
Simultaneous gas flows of hydrogen and oxygen is not allowed.
Action: Correct the gas flow step to prevent this condition.
6033 Calculated Po of [n] for [n] at analysis temperature [n] is out of range.
Cause: The P0 calculated from the analysis temperature is outside the allowed range.
Action: Verify the analysis temperature. Verify that the adsorptive properties file for the ana-
lysis gas is available in the param subdirectory of the application directory, and
reselect the analysis gas in the sample file analysis conditions.
6034 Step [n]s. A manual injection gas must be selected in a preceding gas flow step.
Cause: A manual injection was selected in the specified dose step, but a manual injection gas
was not specified in a preceding gas flow step.
Action: Select an injection gas in a gas flow step preceding the dose step configured for a
manual injection.
The gas selected for this task is different from the gases selected for the neigh-
6211 boring tasks. An evacuation task must be added to add this task. Add evac-
uation and proceed?
Cause: A new task was entered and the gas does not match the gas in the previous task. Mix-
ing gases may be dangerous.
Action: Select Yes to automatically insert the evacuation task. Select No to continue editing
the task options.
You cannot delete the only evacuation task between two tasks with different
6211
gases.
Cause: An evacuation task is required between two tasks with different gases in the pre-
paration options in-situ-activation task list of a chemical adsorption sample file.
Action: Enter an evacuation task between the two tasks.
Adding this leak test will require a preceding evacuation. Do you wish to add
6212
both tasks?
Cause: By inserting or deleting a task in the Selected Tasks list on the Analysis Conditions
window, a leak test task was entered without a preceding evacuation task. The leak
test will fail if the sample pressure exceeds 0.5 mmHg.
Action: Click Yes to evacuate the sample between the two tasks. Click No to exclude evac-
uation
The maximum allowed number of tasks is selected. Delete some tasks before
6213
proceeding.
Cause: An attempt was made to insert more than twenty tasks.
Action: No more than twenty tasks may be specified in one sample file. If additional
preparation steps are required, use two sample files. The first sample file should
contain only preparation tasks; the second file should contain any remaining
preparation tasks and an analysis task.
The vapor source heating mantle and the sample furnace cannot both be con-
trolled during analysis. Either Adsorptive Vapor Source Temperature heating
6216
mantle control or Temperature Options Furnace temperature control must be
disabled.
Cause: When the dosing mode for a chemical analysis is set to vapor, simultaneous vapor
and furnace temperature control is not allowed.
Action: Disable either the vapor source temperature heating mantle in the analysis adsorptive
properties or the furnace temperature control in the temperature options.
The Active Metal Table Options dialog box does not specify a stoichiometry factor
6230 for the adsorptive [n] of the element [n]. One (1.0) was substituted for this value
in the calculations.
Cause: A non-zero percent of sample weight for an element was specified in the Active
Metals Table of the sample report options but a stoichiometry factor for the adsorptive
No element in the Active Metals Table Options has a [n] of Sample Weight value
6231
greater than zero. Surface areas will not be calculated.
No element in the Active Metals Table Options has a [n] of Sample Weight value
6231 greater than zero. Metallic Dispersion and Metallic Surface Area will not be cal-
culated.
Cause: The active metals table for the sample does not contain any selected metals.
Action: Edit the active metals table to contain one or more metals with a non-zero percent of
sample weight.
[n] cannot be produced: No data points were selected for inclusion in the cal-
6232
culations.
Cause: Fewer than two points were selected for the Difference or Sinfelt report calculation.
Action: Select two or more data points for the line fit.
[n] not compatible with the MFC seals. Kalrez seals are recommended for use
6233
with this fluid.
Cause: Kalrez seals are indicated for the fluid specified in a flow task in the preparation
options in-situ-activation task list of a chemical adsorption sample file.
Action: Contact your Micromeritics service representative.
[n] is in the liquid phase at ambient temperature and pressure and cannot be
6234
used with this MFC.
Cause: Fluids selected for flow tasks in the preparation options in-situ-activation task list of a
chemical adsorption sample file must be in the gas phase at ambient temperature and
pressure.
Action: Select another fluid for the flow task.
Cause: Calculations for the primary/repeat/differnce data set failed. Not enough points were
supplied to the line-fit, or the line-fit could not be successfully computed.
Action: Ensure that an appropriate selection of data has been made for the calculation.
None of the requested quantities adsorbed is within the range of the primary
6560
data of more than one sample file.
None of the requested quantities adsorbed is within the range of the repeat data
6561
of more than one sample file.
Cause: The Heat of Adsorption report requires isotherm data to be within the range of the cal-
culations quantity adsorbed values.
Action: Increase the range for the table of quantity adsorbed values.
The sample [n] does not have enough primary data. A minimum of two adsorp-
6563
tion points is required.
The sample [n] does not have enough repeat data. A minimum of two adsorption
6564
points is required.
The sample [n] does not have enough difference data. A minimum of two adsorp-
6565
tion points is required.
Cause: Less than two adsorption points were obtained from the indicated sample for the Heat
of Adsorption report.
Action: Ensure the indicated sample has a valid adsorption isotherm.
The sample [n] does not have any primary data in the range of the requested
6566
quantities adsorbed.
The sample [n] does not have any repeat data in the range of the requested
6567
quantities adsorbed.
Cause: The sample does not have any primary data in the range of the requested quantities
adsorbed.
Action: Extrapolation is not allowed. Ensure that enough isotherm data is inside the range of
the requested quantities adsorbed for interpolation.
[XDCR] Transducer offset calibration rejected (current: [n], new: [n], nominal:
10180
[n], max: [n]).
Cause: There is problem with communication between the instrument and the vacuum
gauge.
Action: Contact your Micromeritics service representative.
Transducer scale calibration rejected (current: [n], new: [n], nominal: [n], min:
10190
[n], max: [n]).
Cause: The transducer scale calibration was rejected.
Action: Contact your Micromeritics service representative.
Turbo pump off during analysis. Turn turbo pump on before resuming ana-
10310
lysis.
Cause: The turbo pump is speed below the threshold for proper operation. The analysis was
suspended.
Action: Ensure the turbo pump is powered on and operational before resuming the analysis.
Time limit exceeded while waiting for the elevator to rise into position (elapsed:
10360
[n] s, max allowed: [n] s, status: [n], alarm code: [n], inputs: [n], position: [n]).
Cause: There is a problem with the elevator.
Action: Contact your Micromeritics service representative.
Time limit exceeded while waiting for the elevator to lower into position
10370
(elapsed: [n] s, max allowed: [n] s, status: [n], alarm code: [n], inputs: [n], pos-
ition: [n]).
Cause A: There is a problem with the elevator.
Action A: Check the dewar. Remove ice if necessary, then restart the analysis. Contact your
Micromeritics service representative.
Cause B: The Psat tube is interfering with elevator movement.
Action B: Ensure the Psat tube is close to the sample tube and the dewar lid is over both the
sample and Psat tubes, then restart the analysis. Contact your Micromeritics service
representative if necessary.
[n] Rotary valve move error detected. Expected position: [n], Actual position:
10414
[n].
10415 [n] Rotary valve initialization error detected. Number of ports: [n] expected [n].
Cause: A rotary valve command did not complete successfully.
Action: Contact your Micromeritics service representative.
Time limit exceeded while waiting for baseline stable (elapsed: [0] s, slope: [3]
10418
%/min, threshold: [3] %/min, remaining duration [SEC] s).
Time limit exceeded while waiting for baseline return (elapsed: [0] s, change: [3]
10419
%/min/min, threshold: [3] %/min/min, remaining duration [SEC] s).
Cause A: The signal did not meet the specified threshold for baseline stability within the spe-
cified time period.
Cause B: The signal did not meet the specified threshold for a return to baseline within the spe-
cified time period.
Action: A number of factors affect the thermal conductivity detector baseline including effect-
iveness of the cold trap, the sample under analysis, flow rate and temperature sta-
bility, gas contamination, as well as the condition of the detector. Verify that the cold
trap is properly prepared. Check for leaks in the carrier path. Verify stable operation of
the carrier mass flow controller and the manifold and furnace temperature sensors
and temperature control. If the problem persists, the thermal conductivity detector fil-
aments may be contaminated or need to be replaced. Contact your Micromeritics ser-
vice representative.
Action: Do not enable the TCD without carrier gas flow at the minimum specified rate. Do not
disable carrier gas flow while the TCD is enabled.
An invalid dynamic chemisorption script step was encountered. (step: [n], offset
10431
in script: [n], type: [n], size: [n])
A dynamic chemisorption script experiment was ended during recording. (step:
10432
[n], offset in script: [n], type: [n], size: [n])
A dynamic chemisorption script experiment was ended during repeat. (step: [n],
10433
offset in script: [n], type: [n], size: [n])
A dynamic chemisorption script step had the wrong experiment or step number.
10434
(step: [n], offset in script: [n], type: [n], script step: [n])
A dynamic chemisorption script record step was encountered during recording.
10435
(step: [n], offset in script: [n], type: [n], size: [n])
Cause: An error was detected in the analysis conditions for the currently running sample.
Action: Contact your Micromeritics service representative.
A dynamic chemisorption script end record step was encountered while not
10437
recording. (step: [n], offset in script: [n], type: [n], size: [n])
A dynamic chemisorption script repeat step was encountered during repeat.
10438
(step: [n], offset in script: [n], type: [n], size: [n])
A dynamic chemisorption script end repeat step was encountered while not
10439
repeating (step: [n], offset in script: [n], type: [n], size: [n])
Cause: An error was detected in the analysis conditions for the currently running sample.
Action: Contact your Micromeritics service representative.
Manifold temperature error detected (manifold: [n], port: [n], heater: [n], heater
10710
target: [n], power: [n]).
Cause: An error was detected during manifold temperature control.
Action: Ensure the cover to the upper cabinet of the instrument is securely fastened. If the
source of the problem has been identified and corrected, close the application pro-
gram, power the instrument off for a few seconds, then turn the power ON. Restart
the application program. Contact a Micromeritics service representative if this error
message continues. Contact the local fire department, if necessary.
Mantle temperature error detected (type: [n], actual: [n], max allowed: [n], target:
10730
[n], power: [n]).
Cause: An error was detected with the mantle temperature.
Action: Contact your Micromeritics service representative.
Mantle breaker open (type: [n], actual: [n], max allowed: [n], target: [n], power:
10740
[n]).
Cause: The circuit breaker to the mantle is open.
Action: Contact your Micromeritics service representative.
Sample temperature control error (actual: [n], scheduled: [n], furnace: [n], tar-
10743
get: [n], rate: [n]/min, power: [n]).
Cause: While controlling the furnace temperature, the difference between the furnace and
sample temperatures exceeded the maximum temperature delta allowed.
Action: Contact your Micromeritics service representative.
Furnace temperature more than [n] different from sample temperature (furnace:
10747
[n], sample: [n], target: [n], rate: [n]/min, power: [n]).
Cause: While ramping the furnace temperature, the difference between the actual and expec-
ted temperature exceeded the maximum temperature delta allowed.
Action: Contact your Micromeritics service representative.
10750 Time limit exceeded during evacuation (target: [n], pressure: [n], elapsed: [n] s).
Cause: The maximum time for evacuating the sample was exceeded. Possible causes are a
leak in the sample tube fitting, a crack in the sample tube, or a poorly degassed
sample.
Action: Check the sample tube and the sample tube fitting. Ensure that the tube is securely
attached to the port. Ensure that the sample is properly degassed, then restart the
analysis.
Mass flow rate is out of tolerance (gas: [n], valve: [n], flow constant: [n], target:
10753
[n]/min, actual: [n]/min, minimum: [n]/min, maximum: [n]/min).
Cause: During in-situ-activation flow task, the difference between the actual and expected
flow rates exceeded the flow rate tolerance.
Action: Check the gas regulator for proper inlet gas pressure as specified in the instrument
operator manual. Contact your Micromeritics service representative.
Time limit exceeded while dosing (gas: [n], valve: [n], target: [n], pressure: [n],
10760
elapsed: [n] s).
Cause A: The specified pressure was not attained. The gas regulator may be set too low or
turned off.
Action A: Set the gas regulator to 10 psig (0.7 bar), then restart the analysis.
Cause B: The gas cylinder is empty.
Action B: Connect a new gas cylinder, then restart the analysis.
Attempts to dose failed on sample port [n]. (qty required:[n, qty dosed: [n],
10770
sample pressure: [n], gas: [n]).
Cause: There was a problem dosing the sample to target pressure. The instrument was
unable to dose the required quantity of gas after several attempts.
Action: Check that the outlet stage of the gas regulator is within specification. Review the ana-
lysis parameters.
Leak test failed (sample port: [n], interval: [n] s, leak rate: [n]/min, max allowed:
10780
[n/min).
Cause: With the sample port valve closed, the sample pressure increased by 0.15 mmHg
before the leak test duration was completed.
Action: Check sample tube fitting and ensure that it is securely attached to the port, then
restart the analysis.
Quantity dosed on sample port [n] has exceeded the maximum of [n] (qty dosed
10790
this point [n], pressure: [n]: [n], gas: [n]).
Cause: An excessive quantity of gas has been dosed into the sample port due to an excessive
quantity of a sample with high pore volume, condensation of gas due to a lower than
expected saturation pressure, or a leak.
Action: Review the analysis conditions, the sample quantity, and the sample tube connection
to identify, then correct the problem before repeating the analysis.
P0-over-sample failed on sample port [n] (pressure: [n], last pressure: [n], [n], rel
10801
pressure: [n], qty ads: [n], doses: [n]).
Cause: Attempts to condense the adsorptive gas in the sample tube have failed due to an
excessive quantity of a sample with high pore volume, adsorptive gas contamination,
a higher than expected saturation pressure, or a leak.
Action: Review the analysis conditions, the sample quantity, the gas supply, and the sample
tube connection to identify, then correct the problem before repeating the analysis.
Warm free-space measurement failed on sample port [n] (qty in free-space: [n],
10830
qty in port: [n], pressure: [n], port vol: [n], port temp: [n]).
Cause: There is a problem with the warm free-space measurement on the sample port.
Action: Ensure that no problem exists with the sample tube or gas connection.
Cold free-space measurement failed on sample port [n] (qty in free-space: [n],
10840
qty in port: [n], pressure: [n], port vol: [n], port temp: [n], warm free-space: [n]).
Cause: There is a problem with the cold free-space measurement on the sample port.
Action: Ensure there is no problem with the sample tube or analysis bath.
Maximum target pressure exceeded in sample port [n] (target pressure: [n], max
10850 instrument manifold pressure: [n], gas: [n], max gas manifold pressure: [n], max
transducer pressure: [n]).
Cause: A target pressure was requested that exceeds the maximum allowed. The maximum
pressure may be based on saturation pressure of the gas at the temperature of the
gas source, the manifold, the sample, or ambient temperature.
Action: Review the maximum allowable pressures in the error message and the analysis con-
ditions to identify and correct the problem before repeating the analysis.
10860 Psat gas [n] is not condensing. (pressure: [n] maximum manifold pressure: [n]).
Adsorptive [n] is not condensing. (pressure: [n] maximum manifold pressure:
10870
[n]).
Cause: The gas specified in the error message is not condensing.
Action: Review the analysis parameters, gas connections, and analysis bath.
Purification of the adsorptive [n] failed at [n] (charge pressure: [n], minimum
10890
allowed: [n]).
10940 The dewar shield was not present when the analysis was started.
Cause: The dewar shield was not detected when the analysis was started.
Action: For best performance, install the dewar shield before starting an analysis.
Power supply voltage error detected (type: [n], voltage: [n] volts, nominal: [n]
10950
volts.
Cause: There is a problem with the power supply voltage.
Action: Contact your Micromeritics service representative.
Line frequency out of tolerance (Frequency: [n] Hz, Minimum: [n] Hz, Maximum:
10980
[n] Hz).
Cause: A problem was detected in the power supplied to the instrument.
Action: Contact your Micromeritics service representative.
10981 Cryostat operation failed or was canceled (errors: [n], target: [n], last: [n]).
Cause: The wait for the CryoStat to reach target temperature was either skipped or failed due
to a communications issue with the CryoStat.
Action: Ensure that the CryoStat, the instrument, and the Windows computer are connected
to the same network switch or hub. Ensure that no other devices are connected on
the same subnet with the same IP address.
Pneumatic pressure out of tolerance (Pressure: [n] psig, Minimum: [n] psig, Max-
10990
imum: [n] psig).
Cause: A problem was detected in the compressed air supply used by the pneumatically actu-
ated, hard seal valves and the furnace.
Action: Consult the Operator Manual for the compressed air supply pressure required for
proper operation of the instrument.