LAB 2 THERMO Temp Measurement
LAB 2 THERMO Temp Measurement
LAB MODULE
THERMODYNAMICS I
LEVEL OF COMPLEXITY
1 2 3 4 5 6
KNOWLEDGE REPETITION APPLICATION ANALYSIS EVALUATION SYNTHESIS
√ √ √
1
THERMODYNAMICS LABORATORY MODULE
EXPERIMENT 2
1. OBJECTIVE
2. INTRODUCTION
3. EQUIPMENT
4. PROCEDURE
5. RESULTS
7. CONCLUSION
2
THERMODYNAMICS LABORATORY MODULE
EXPERIMENT 1
Temperature Measurement Method
1. OBJECTIVE
2. INTRODUCTION
2.1 Background
The temperature of a substance does not measure its heat content but rather the average
kinetic energy of its molecules resulting from their motions. A one-pound block of iron and a two-
3
THERMODYNAMICS LABORATORY MODULE
pound block of iron at the same temperature do not have the same heat content. Because they are at
the same temperature the average kinetic energy of the molecules is the same; however, the two-
pound block has more molecules than the one-pound block and thus has greater heat energy.
The scale we use to measure temperature is "degrees" (°). There are three temperature
scales that are used today.
i. The Kelvin (K) scale is used by scientists and for astronomical temperatures.
ii. The Celsius scale (°C) is used in most of the world to measure air temperatures.
iii. The Fahrenheit scale(°F) is used to measure temperatures at or near the surface.
All three temperature scales are related to each other through the "triple point of water". The
triple point of water is the temperature at which water vapor, liquid water, and ice can coexist
simultaneously. The triple point occurs at 0.01 °C (273.16 K or 32.02 °F). To convert from one
temperature scale to another, we need to use the equations as below:
a) Partial Immersion
4
THERMODYNAMICS LABORATORY MODULE
b) Total Immersion
Total immersion thermometer need to be immersed up to the liquid temperature mark on the
thermometer. Since the thermometer column is fully immersed, this thermometer is the most
accurate.
2.3 Thermocouple
5
THERMODYNAMICS LABORATORY MODULE
Exposed junction thermocouple is fast responding but the thermocouple itself is unprotected
and subject to corrosion from the environment. Also, the smaller the probe sheath diameter, the faster
the response. Often the thermocouple is located inside a metal or ceramic shield that protects it from
a variety of environments. Metal-sheathed thermocouple is also available with many types of outer
coating, such as polytetrafluoro ethylene for trouble-free use in corrosive mediums.
2.4 Thermistor
Thermistor as shows in figure 3 is a thermally sensitive resistor and has, according to type, a
negative (NIC) or positive (PTC) resistance/temperature coefficient. Thermistor thermometry is based
on the principle that metal oxides change resistance with a change in temperature. Resistance
decreases as the temperature decreases. The meter where it is converted and displayed as a
temperature reading detects this resistance change. Thermistor has excellent accuracy over
biological or ambient temperature ranges when compared to RTDs or Thermocouples. Response
time is generally faster than RTDs.
6
THERMODYNAMICS LABORATORY MODULE
Thermistor has a semiconductor material which changes its electrical resistance as a function
of temperature. Extension wires used with thermistor can be plain copper wire. Thermistor offers
accuracy similar to RTD within narrow temperature ranges near to ambient temperature. It is
generally responses faster comparatively. Since thermistor standards vary, care must be taken to
match the instrumentation to the sensor. The resistance-temperature relationship of a thermistor is
negative and highly nonlinear. Thermistor is usually designated in accordance with it's resistance at
25°C.
A typical RTD as shows in figure 4 consists of a fine platinum wire wrapped around a mandrel
and covered with a protective coating. Usually, the mandrel and coating is glass or ceramic.
Depositing can also make the platinum as a film on a substitute and then encapsulating it. RTD is
wire wound and thin film device that work on the physical principle of the temperature coefficient or
electrical resistance of metals.
The electrical resistance of the RTD changes as a function of temperature. Circuitry similar to
Wheat stone bridge is built into control designed for use with RTD. Constant current into the bridge
produces an output voltage that varies with temperature. Lead wire resistance can significantly affect
the RTD measurement. This is typically corrected using a third (compensating) lead wire.
7
THERMODYNAMICS LABORATORY MODULE
RTD is nearly linear over a wide range of temperatures and can be made small enough to
have response times of a fraction of a second. The classical resistance temperature detector (RTD)
construction using platinum was proposed by C.H.Meyers in 1932. This requires an electrical current
to produce a voltage drop across the sensor that can be then measured by a calibrated read-out
device.
The bimetallic thermometer as shows in figure 5 uses a bimetal, which is composed of two
types of metals with different thermal coefficients of expansion and they are wound into a helical form,
change according to temperature is transmitted to the indicator. This thermometer is simple in
construction and reasonably priced.
8
THERMODYNAMICS LABORATORY MODULE
Equipment Note
This Temperature Measurement Bench (Model: HE 151) has been designed to demonstrate the
fundamental temperature measuring techniques using thermocouples, mercury in glass thermometer,
resistance temperature detector (RTD), thermistor, bimetallic temperature indicator and etc.
Temperature measurement is used to measure air temperature, boiling water temperature, ice-point
temperature and wet or dry bulb temperature. Temperature can be measured via a various ranges of
sensors. All of these sensors infer temperature by sensing changes in physical characteristics.
9
THERMODYNAMICS LABORATORY MODULE
4. PROCEDURE
10
THERMODYNAMICS LABORATORY MODULE
11
THERMODYNAMICS LABORATORY MODULE
12
THERMODYNAMICS LABORATORY MODULE
5. RESULTS
13
THERMODYNAMICS LABORATORY MODULE
6.2 Why there are differences in temperature readings among measurement devices?.
6.3 Which device gives the fast response during taken measurement?
6.4 To continuously measure temperature in 24 hours within 200 meter distance which device
could you propose and why?
6.5 From you own reading, give the temperature measurement device that could be used to
measure melted aluminium which having temperature > 400°C.
14
THERMODYNAMICS LABORATORY MODULE
7. CONCLUSION
Based on data and discussion, make your overall conclusion by referring to experiment objective.
15