KENWOOD OSCILLOSCOPE

CS-5200 S E R I E S
CS-5270 CS-5275
CS-5260 CS-5265
CS-5230 CS-5235

INSTRUCTION MANUAL
KENWOOD CORPORATION

© P R I N T E D IN J A P A N B63-0219-00(N)
36511
SAFETY

Symbol i n T h i s l a n u a l

/ f \ T h i s symbol i n d i c a t e s where a p p l i c a b l e c a u t i o n a r y or o t h e r i information i s to be found.

Power Source

This equipment o p e r a t e s from a power s o u r c e that does n o t a p p l y more t h a n 250 V rms


between the supply conductors or between e i t h e r supply conductor and ground. A protective
ground c o n n e c t i o n by way of t h e g r o u n d i n g c o n d u c t o r i n t h e power c o r d i s essential for
safe operation.

Grounding the Product

T h i s equipment i s grounded through t h e g r o u n d i n g c o n d u c t o r of t h e power c o r d . To a v o i d


e l e c t r i c a l shock, plug the power cord i n t o a p r o p e r l y w i r e d r e c e p t a c l e b e f o r e connecting
to the equipment input or output t e r m i n a l s .

Use the Proper Power Cord

Use o n l y t h e power cord and connector s p e c i f i e d f o r your product.

Use the Proper Fuse

To a v o i d f i r e h a z a r d , use a f u s e of the c o r r e c t type.

Do not Operate in Exp Ios i ve Atmospheres

To a v o i d e x p l o s i o n , do not operate t h i s product i n an e x p l o s i v e atmosphere.

Do not Remove Cover or Panel

To avoid personal injury, do n o t remove the cover or p a n e l . Refer s e r v i c i n g to


q u a l i f i e d personnel.
Voltage Conversion
If t h e power s o u r c e i s not a p p l i e d to your p r o d u c t , c o n t a c t your d e a l e r . To a v o i d
e l e c t r i c a l shock, do not perform the v o l t a g e c o n v e r s i o n .
CONTENTS

1 . GENERAL 1

2 . SAFETY AND EQUIPMENT PROTECTION PRECAUTIONS 2

3 . FEATURES 4

4 . SPECIFICATIONS 6

5 . PANEL EXPLANATION • 1 1

5 - 1 Front Panel 1 1

5-2 Rear Panel 20

5-3 Bottom P l a t e • 2 1

5-4 Readout D i s p l a y 2 2

1) Display positions 2 2

2) Display • • ••• 2 2

6 . CHECKING AND ADJUSTMENT PRIOR TO MEASUREMENT 24

7 . OPERATING PROCEDURES 27

7 - 1 S i n g l e T r a c e Operation 27

1) A l t e r n a t i n g current display • 27

2) Trigger • • 27

3) Composite video s i g n a l d i s p l a y 28

7-2 Multi-Trace Operation 29

1) Switching v e r t i c a l o p e r a t i o n modes 29

2) Switching t r i g g e r sources • 29

3) Line trigger 29

7-3 Sweep M a g n i f i e d Operation 30

7-4 Delayed Sweep Operat ion 30

7-5 X-Y Operation 3 1

7 - 6 S i n g l e Sweep Operation 3 1

7-7 Readout Operation • 32

1) Set value (scale factor) display 32

2) Cursor measurement 32
8 . APPLICATIONS . . . . . . . . . . . . . . . . o - . . e S o , c S e , . . . . , . o 3 4

8 — 1 Measuring V o l t a g e Between Two P o i n t s on a Waveform ..... 3 4


8 — 2 E l i m i n a t i o n of u n d e s i r e d s i g n a l components • •••• ............ 3 g

8-3 Measuring D i r e c t Current (DC) Voltage * • 3 7

8-4 Measuring V o l t a g e R a t i o w i t h C u r s o r s 3 9

8 — 5 Measuring S i g n a l s w i t h Low Frequency Components • •••• .......... 3 9

8-6 Measuring S i g n a l s w i t h High Frequency Components 4 0

8-7 Measuring Time Between Two P o i n t s 4 0


8-8 Measuring Time R a t i o w i t h C u r s o r s 4 2
8 — 9 Measuring F r e q u e n c i e s ••••••••••••••••••••••••••••••••••••••••••••••••••• 4 2

8—10 Measuring P u l s e R i s e and F a l l Times 4 4

8-11 Measuring P u l s e J i t t e r s 4 5

8-12 Measuring Phase D i f f e r e n c e s 4 6

8-13 A p p l i c a t i o n s of X-Y O p e r a t i o n 4 8

9 . F U S E REPLACEMENT, LINE VOLTAGE S E L E C T I O N 50

10. OPTIONAL EQUIPMENT 5 1


GENERAL

P l e a s e note t h e f o l l o w i n g when u s i n g t h i s manual.

• T h i s manual c o v e r s 6 o s c i l l o s c o p e models; CS-5270, CS-5260, CS-5230, CS-5275, CS-5265,

and CS-5235. However, e x p l a n a t i o n about t h e p a n e l s , o p e r a t i n g p r o c e d u r e s and a p p l i c a -

t i o n s i s made u s i n g t h e CS-5270. A l t h o u g h most of the c o n t e n t s i s common to 6 models,

p l e a s e note t h a t any p o r t i o n which i s not i s expressed w i t h i n b r a c k e t s [ ] .

• T h i s manual u s e s , i n some d e s c r i p t i o n s , the terms and a b b r e v i a t i o n s shown below:

Channel 1 or Y input CHI input

Channel 2 or X input CH2 input

V e r t i c a l mode (VERT MODE) V. mode

Vertical variable V. v a r i a b l e
Channel 2 i n v e r t .. Inv.

Vertical position yposition


A l t e r n a t e v e r t i c a l a x i s s w i t c h i n g mode A l t . mode

A l t e r n a t e main sweep and delayed sweep o p e r a t i o n A l t . sweep


Main sweep o p e r a t i o n A sweep
Delayed sweep o p e r a t i o n - • • B sweep
Horizontal variable h. v a r i a b l e

Horizontal position • • ^•position


H o r i z o n t a l mode m
ode
T r i g g e r i n g mode ^ mode

T r i g g e r i n g source v e r t , mode — • Vert.

1
2. SAFETY AND EQUIPMENT PROTECTION PRECAUTIONS

1) Check your l i n e v o l t a g e b e f o r e use. The o s c i l l o s c o p e has a f u s e h o l d e r and voltage

s e l e c t o r on the l e f t of the AC i n l e t on the r e a r p a n e l . The v o l t a g e below the W lark

shows t h e 1ine voltage setting of y o u r o s c i 1 l o s c o p e . I f your line voltage is

different f r o m the o s c i l l o s c o p e s e t t i n g , there i s a danger of m a l f u n c t i o n i n g . Do

not connect to the supply u n t i l the v o l t a g e s e l e c t o r matches the supply voltage.

Set to 100 V.

2) The osci1loscope i s equipped w i t h internal components that are highly charged

electrically. For your own p r o t e c t i o n , do not remove the c a s i n g f o r any reason.

3) Do not use the o s c i l l o s c o p e i n the f o l l o w i n g l o c a t i o n s .

• In direct sunlight

• I n e x t r e m e l y hot and/or humid a r e a s


• I n a r e a s a f f e c t e d by high l e v e l s of mechanical v i b r a t i o n
• Around a r e a s w i t h s t r o n g l i n e s of magnetic f o r c e s or impulse v o l t a g e

4) Make sure t h a t the v o l t a g e a p p l i e d to each input t e r m i n a l does not exceed the maximum

specified.

a CH 1 and CH2 input t e r m i n a l s : 500 Vp-p or 250 V (DC + AC peak, 1kHz or less)

^CH3 and Z AXIS input t e r m i n a l s : 100 Vp-p or 50 V (DC + AC peak, 1kHz or less)

Do not under any circumstances apply v o l t a g e to the o u t p u t t e r m i n a l s from e x t e r n a l

power s o u r c e s .

5) I n o r d e r to p r e v e n t the CRT' s f l u o r e s c e n t s c r e e n from s c o r c h i n g , do not adjust the

brightness higher than n e c e s s a r y . Do not l e a v e the s p o t t i n g f u n c t i o n on for long

p e r i o d s of time.
6) The Instrument I s equipped w i t h a u x i l i a r y support. You can use i t either horizontally
or v e r t i c a l l y . Do not p l a c e any o b j e c t on the top of the instrument or p o s i t i o n i t i n
a r e a s where t h e v e n t i l a t i o n h o l e s i n the c a s i n g a r e b l o c k e d . Any blockage wi11
cause the t e m p e r a t u r e of the i n t e r n a l components to i n c r e a s e , r e s u l t i n g in possible
damage.

7) Connect a ground w i r e to the GND terminal for safety.

8) I f the power i s t u r n e d on and o f f r e p e a t e d l y , the o s c i l l s c o p e may malfunction. Once

the power i s turned o f f , l e a v e f i v e seconds before t u r n i n g the power on again.

9) The probes s u p p l i e d w i t h the o s c i l l o s c o p e a r e p r e c i s i o n p a r t s . Handle them w i t h g r e a t


care. [The probes s u p p l i e d w i t h the CS-5230, CS-5260, or CS-5270 have t e r m i n a l s f o r
detect ing readout signals. Do not use them w i t h an o s c i 1 l o s c o p e w i t h no probe
detection function. ]

3
3. FEATURES

High S e n s i t i v i t y : S e n s i t i v i t y up to 1 mV/div.

f i d e Bands : The f r e q u e n c y range i s DC t o 20 MHz ( - 3 dB) a t

1 fflV a n d , 2 mV/div r a n g e and DC t o 100 MHz [DC t o

40 MHz f o r t h e CS-5230 or CS-5235, DC t o 60 MHz f o r

the CS-5260 or CS-5265] ( - 3 dB) f o r 5 mV/div.

HDTV : A s p e c i a l l y d e s i g n e d HDTV s i g n a l t r i g g e r i n g circuit

o f f e r s s t a b l e and a d j u s t m e n t - f r e e synchronization with

the FRAME a n d L I N E signals from smal1 to l a r g e

amplitudes.

Ease of o p e r a t i o n : The o s c i 1 l o s c o p e does not u s e m u l t i - f u n c t i o n knobs.

Thus, an i n e x p e r i e n c e d person may use i t e a s i l y .


S t r a i g h t f o w a r d s w i t c h l a y o u t on t h e p a n e l e l i m i n a t e s

e r r o r s even f o r c o m p l i c a t e d o p e r a t i o n s .

Three-channel v e r t i c a l a x i s : T r i g g e r s i g n a l s may be measured on CH3 w i t h o u t using

the t r i g g e r view f u n c t i o n .

Stepless s e n s i t i v i t y control : The v e r t i c a l a x i s s e n s i t i v i t y may be changed from 1

mV/div to 5 V/div c o n t i n u o u s l y w i t h a r o t a r y s w i t c h .

High Speed Sweep : H i g h speed sweeping i s p o s s i b l e a t a time base o f 5

n s / d i v ( d u r i n g x 1 0 MAG operation)

[10 n s / d i v f o r the CS-5230 or CS-5235],

High Accuracy :Accuracy i s within 3% f o r b o t h vertical axi s

s e n s i t i v i t y and sweep time.

Fixed Synchronization : The f i x e d s y n c h r o n i z a t i o n f u n c t i o n e l i m i n a t e s t r o u b l e -

some s y n c h r o n i z a t i o n .

One Touch X-Y : S w i t c h i n g to X-Y o p e r a t i o n i s v i a a single switch.

Automatic T r i g g e r S i g n a l S e l e c t i o n : T r i g g e r s i g n a l can be s e l e c t e d a u t o m a t i c a l l y according

to t h e VERT MODE c o n t r o l s e t t i n g by s w i t c h i n g t h e

SOURCE c o n t r o l to VERT.

CHI OUTPUT : A channel 1 output t e r m i n a l * i s added f o r CHI i n p u t

signal monitoring.

High S t a b i l i t y , High R e l i a b i l i t y : Hybrid integrated circuits a r e employed in al1

circuits, r e a l i z i n g high standards i n s t a b i 1 i t y and

reliability.

Scale I l l u m i n a t i o n : The s c a l e i l l u m i n a t i o n f u n c t i o n makes i t p o s s i b l e t o

take photographs of waveform d a t a d i s p l a y e d on the CRT

i n a dark room.

4
Delay Time P o s i t i o n : A d e l a y t i m e p o s i t i o n may be s e t q u i c k l y and accu-

r a t e l y w i t h coarse and f i n e controls.

[The f o l l o w i n g f e a t u r e s apply to the CS-5230,. CS-5280 and CS-5270. ]

Readout : The v e r t i c a l axis input sensitivity, sweep t i m e and

other s c a l e f a c t o r s a r e d i s p l a y e d i n c h a r a c t e r s on the

CRT.

Cursor F u n c t i o n : The v o l t a g e d i f f e r e n c e , c u r r e n t ratio, time d i f f e r e n c e ,

time r a t i o , f r e q u e n c y and p h a s e d i f f e r e n c e may be

measured w i t h two c u r s o r s . They a r e d i s p l a y e d on the

CRT f o r quick and a c c u r a t e measurement.

5
4. SPECIFICATION

Items CS-5230, 5235 CS-5260, 5265 CS-5270, 5275

CRT

Type 150 mm r e c t a n g u l a r tube w i t h an i n t e g r a l g r a t i c u l e

Acceleration voltage Approx. 12 kV

E f f e c t i v e area 8 x 10 d i v i s i o n s ( 1 d i v i s i o n = 10 mm)

VERTICAL AXIS (COMMON TO CHANNELS 1 AND 2 )

S e n s i t i v i t y ( + 1 0 to +35°C) 1 mV, 2 mV/div : ±5X, 5 mV to 5 V / d i v : ±3%

A t t e n u a t ion 1-2-5 s t e p s , 12 ranges, f i n e c o n t r o l between ranges

Input impedance 1 Mfi ± 2%, approx. 25 pF

Frequency 5mV to 5V/div DC: DC to 40MHz DC: DC to 60MHz DC: DC to lOOMHz


response AC: 5Hz to 40MHz AC: 5 Hz to 60MHz AC: 5Hz to lOOMHz
(-3 dB)
(+10 to +35°C) ImV to 2mV/div DC: DC to 20MHz, AC: 5Hz to 20MHz

R i s e time 5mV to 5V/div Approx. 8. 8 ns Approx. 5. 8 ns Approx. 3. 5 ns

(+10 to +35°C) ImV to 2mV/div Approx. 17. 5 ns

S i g n a l d e l a y time The l e a d i n g edge may be checked u s i n g a square wave of t h e


r i s e time l e s s than t h a t of t h i s model.

Crosstalk -40 dB or l e s s ( a t 1 kHz)

AMaximum input v o l t a g e 500 Vp-p or 250 V (DC + AC peak, 1 kHz or l e s s )

VERTICAL AXIS (CH3)

S e n s i t i v i t y ( + 1 0 to +35°C) 0. 1 V / d i v : ±3%

Input impedance 1MQ ± 2%, approx. 25 pF


Frequency response (-3 dB) DC to 60 MHz DC to 100 MHz
( + 10 t o + 3 5 X O DC to 40 MHz

R i s e time ( + 10 to +35°C) Approx. 8. 8 ns Approx. 5. 8 ns Approx. 3. 5 ns

S i g n a l d e l a y time The l e a d i n g edge may be checked u s i n g a square wave of the


r i s e time l e s s than t h a t of t h i s model.

AMaximum input v o l t a g e 100 Vp-p or 50 V (DC + AC peak, 1 kHz or l e s s )

Operat ion S i n g l e t r a c e : CHI, CH2, CH3 or ADD s i n g l e t r a c e o p e r a t i o n


M u l t i - t r a c e : 2 to 4 t r a c e s of CHI, CH2, CH3 and ADD

ALT/CHOP: D i s p l a y by s e l e c t i n g ALT and CHOP


ADD : Composite waveform o f CHI and CH2 s i g n a l s a r e
displayed.

6
Items CS-5230, 5235 CS-5260, 5265 CS-5270, 5275

Chop frequency Approx. 250 kHz ( i n m u l t i - t r a c e o p e r a t i o n )

Channel p o l a r i t y Normal or i n v e r t e d , CH2 only i n v e r t e d

H o r i z o n t a l a x i s (CH2, e x c e p t f o r X 1 0 MAG o p e r a t i o n )

S e n s i t i v i t y ( + 1 0 to +35°C) Same a s v e r t i c a l a x i s (CH2)

Input impedance Same a s v e r t i c a l a x i s (CH2)


Frequency response
r (-3 dB)
( + 1 0 to + 3 5 C ) DC: DC to 1 MHz, AC: 5 Hz to 1 MHz

X-Y phase d i f f e r e n c e 3 ° or l e s s a t 100 kHz

Operation mode X-Y mode i s s e l e c t e d w i t h H. MODE


CHI: Y - a x i s , CH2: X - a x i s

AMaximum input v o l t a g e Same a s v e r t i c a l a x i s (CH2)

SWEEP

Sweep types A : A sweep


ALT : A l t e r n a t e A sweep and B sweep
B : B sweep
X-Y : X-Y o s c i l l o s c o p e o p e r a t i o n

Sweep time A sweep 0. 5s to 0. l//s/div 0.5s to 5 0 n s / d i v


(+10 to +35°C) ±3X ±3%
1-2-5 s t e p s , 1-2-5 s t e p s ,
21 ranges, 22 ranges,
f i n e adjustment f i n e adjustment
between ranges between ranges

B sweep 50ms to 0.1/is/div 50ms to 5 0 n s / d i v


±3% ±3%
1-2-5 s t e p s , 1-2-5 s t e p s ,
18 ranges, 19 ranges,
f i n e adjustment f i n e adjustment
betwen ranges betwen ranges - •
Sweep m a g n i f i e d o p e r a t i o n
( + 10 t o + 3 5 ° C ) X10 ± 5% (±8X f o r over 0.5 n s / d i v )

Linearity ( + 10 to + 35°C) ±3% (±5% i n x l O MAG o p e r a t i o n )

H0LD0FF C o n t i n u o u s l y v a r i a b l e from A sweep NORM p o s i t i o n

Trace separation B sweep i s c o n t i n u o u s l y v a r i a b l e by ± 4 d i v i s i o n s or more


w i t h r e s p e c t to A sweep.

Delayed sweep o p e r a t i o n Continuous delay o p e r a t i o n (AFTER DELAY)


?
Synchronous d e l a y o p e r a t i o n ( B TRG D) : S y n c h r o n o u s w i t h
the t r i g g e r s i g n a l

Delay time Continuous c o n t r o l by 0. 2 t o 10 d i v i s i o n s f o r 0. 5 d i v to


maximum speed sweep

Delay time e r r o r [CS-5235 5265 5 2 7 5 ] :


( + 10 t o + 3 5 ° C ) ±(3X of s e t v a l u e + l X o f f u l l s c a l e ) + ( 0 to 300 n s )
[CS-5235, 5265, 5 2 7 5 ] :
Reading on CRT ± 4 1 ( 0 t o 300 n s )

Delay jitter 10000 : 1 of a v a l u e 10 times a s h i g h a s A sweep s e t t i n g

7
Items CS-5230, 5235 CS-5260, 5285 CS-5270, 5275

TRIGGERING

T r i g g e r modes AUTO : Automatic f r e e running w i t h no s i g n a l


NORM : T r i g g e r e d sweep
FIX : Sweep a t t r i g g e r i n g p o i n t s e t to c e n t e r of s i g n a l
amplitude
SINGLE: S i n g l e sweep mode
RESET : R e s t a r t i n g s i n g l e sweep o p e r a t i o n

T r i g g e r s i g n a l sources VERT: Input s i g n a l s e l e c t e d f o r V. mode


CHI : Channel 1 input s i g n a l
CH2 : Channel 2 input s i g n a l
CH3 : Channel 3 input s i g n a l
LINE: Commercial power l i n e

Trigger coupling AC : AC c o u p l i n g from 10 Hz


(For t r i g g e r s e n s i t i v i t y , HFrej : Low-pass f i l t e r c o u p l i n g up to 30 kHz
see the t a b l e below.) DC : DC c o u p l i n g
TV FRAME: Composite video s i g n a l , v e r t i c a l s y n c h r o n i z a t i o n
separation
TV LINE : Composite v i d e o s i g n a l , h o r i z o n t a l s y n c h r o n i -
zation separation

Trigger s e n s i t i v i t y ( + 10 to + 35°C)

Signal frequency Sensitivity


(Ampli tude)
MODE COUPLING
CS-5230, 5235 CS-5260, 5265 CS-5270, 5275 NORM FIX *

10Hz to 20MHz 10Hz to 40MHz 10Hz t o 50MHz ldiv 1.5div


AC
20M to 40MHz 40M to 60MHz 50M t o 100MHz 1.5div 2div

10Hz to 30kHz 10Hz to 30kHz 10Hz t o 30kHz ldiv 1..5div


HFrej
over 30kHz over 30kHz over 30kHz >min. >min.
NORM
DC to 20MHz DC to 40MHz DC t o 50MHz ldiv 1.5div
DC
20M to 40MHz 40M to 60MHz 50M t o 100MHz 1. 5 d i v 2div

TV-F, -L Composite video s i g n a l 1. 5div

HDTV HDTV video s i g n a l 1. 5div

AUTO : Same a s above s p e c i f i c a t i o n s f o r above 40Hz.

(The t a b l e shows the s e n s i t i v i t i e s i n terms of the amplitude d i s p l a y e d on the CRT. )


(>min. f o r the H F r e j s e n s i t i v i t y shows t h a t the amplitude n e c e s s a r y f o r s y n c h r o n i z a -
tion increases.)

FIX * : Same a s above s p e c i f i c a t i o n s f o r above 50Hz.

8
Items CS-5230, 5235 CS-5260, 5265 CS-5270, 5275

CALIBRATION SIGNAL : POSITIVE SQUARE WAVE, 1 Vp-p ±3%, APPROX. 1 kHz

INTENSITY MODULATION

Input v o l t a g e Darkens a t TTL l e v e l (+5 V ) .

Input impedance Approx. 10 k Q


Frequency response
DC to 3.5 MHz DC t o 5 MHz
( + 10 t o +35°C)
AMaximum input v o l t a g e 100 Vp-p or 50V (DC + AC peak)

CHI OUTPUT SIGNAL (WITH 50 Q LOAD)

Output v o l t a g e Approx. 50 mVp-p/div

Output impedance Approx. 50 Q

Frequency 1, 2 mV 100 Hz to 20 MHz


response
(-3dB) 5 mV to 100 Hz to 40 MHz 100 Hz t o 60 MHz 100 Hz to 100 MHz

TRACE ROTATION : TRACE ANGLE IS ADJUSTABLE WITH A SEMI-FIXED CONTROL ON PANEL.

POWER SUPPLY

Voltage 100/120/220 VAC ± 10%, 207 t o 250 V

Frequency 50 Hz or 60 Hz

Power consumption Approx. 43 W, 39 W Approx. 48 W, 43 W Approx. 48 W, 43 W


Approx. 52VA, 49VA Approx. 56VA, 51VA Approx. 56VA, 51VA

DIMENSIONS AND WEIGHT (VALUES ENCLOSED IN PARENTHESES INCLUDE PROJECTIONS. )

Width 300 mm (300 mm)

Height 150 mm (172 mm)

Depth 400 mm (469 mm)

Weight Aprrox. 8. 7 kg

OPERATING TEMPERATURE AND HUMIDITY


O p e r a t i n g temperature and
0 t o 40°C, 85% RH or l e s s
humidity
S t o r a g e temperature and
- 2 0 to 70°C, 85% RH or l e s s
humidity
ACCESSORIES

Probe PC-33 10 m ±1%, 22 pF±10%, 10:1 [CS-5230]


PC-31 10 MQ ±1%, 12.5 p F ± 1 0 % , 10:1 [CS-5260, CS-5270]
PC-35 10 MQ ±1%, 19.5 p F ± 1 0 % , 10:1 [CS-5235]
PC-39 10 MQ ±1%, 12.5 p F ± 1 0 % , 10:1 [CS-5265, CS-5275]

2 each

I n s t r u c t i o n manual 1 copy

Adjusting screwdriver 1

Power cord 1

Replacemen f u s e 2
[The s p e c i f i c a t i o n s shown below do not apply to the CS-5235, CS-5265 or CS-5275.]

Items CS-5230, 5235 CS-5260, 5285 CS-5270, 5275

Replacement f u s e 1 A x 2 ( f o r 100 V a r e a )
830 m A x 2 ( f o r 200 V a r e a )

READOUT

Set v a l u e s CHI and CH2 s c a l e f a c t o r s ( w i t h probe d e t e c t i o n ) , CH3


s c a l e f a c t o r ( 0 . 1 V / d i v f i x e d , w i t h no probe d e t e c t i o n ) ,
V-UNCAL, ADD, INV, A/B sweep s c a l e f a c t o r s (MAG-converted),
sweep - UNCAL, DELAY TIME, TRIG' D, X-Y

Cursor modes A V I d i s p l a y e d i n v o l t a g e w i t h c o n v e r s i o n a c c o r d i n g to
(Between A R E F and A CHI s c a l e f a c t o r
cursors) A V 2 d i s p l a y e d i n v o l t a g e w i t h c o n v e r s i o n a c c o r d i n g to
In X-Y mode, only AVI CH2 s c a l e f a c t o r
may be s e t . A V 3 d i s p l a y e d i n v o l t a g e w i t h c o n v e r s i o n a c c o r d i n g to
CH3 0.1 V/div
AT : D i s p l a y e d i n t i m e w i t h c o n v e r s i o n a c c o r d i n g to A
sweep s c a l e f a c t o r
l/AT:Displayed i n frequency w i t h conversion according
to A sweep s c a l e f a c t o r

In V, H-VARI or RATI0:Vo 1 1 a g e r a t i o a n d t i m e r a t i o a r e d i s p 1 a y e d ,
UNCAL mode w i t h 5 d i v i s i o n s on the CRT as 1 0 0 1
PHASE:Phase d i f f e r e n c e i s d i s p l a y e d , w i t h 5 d i v i s i o n s on
the CRT as 3 6 0 ° .

Cursor measurement Resolution : 10 b i t s


Measuring e r r o r : ±4%
Measuring range: ± 3 . 6 d i v i s i o n s or more v e r t i c a l l y from i
CRT c e n t e r .
±4. 6 d i v i s i o n s or more h o r i z o n t a l l y from
CRT c e n t e r .

• T h e s p e c i f i c a t i o n s shown above a r e s u b j e c t to change w i t h o u t notice.


5. PANEL EXPLANATION

5 — 1 Front Panel

Figure 1 Front Panel (CS-5270)

® Cathode ray tube ( C R T )

The e f f e c t i v e d i s p l a y s c r e e n s u r f a c e runs over an a r e a of e i g h t 10 mm d i v i s i o n s ( 8 0 mm)

a l o n g t h e v e r t i c a l a x i s and t e n 10 mm d i v i s i o n s ( 1 0 0 mm) along the h o r i z o n t a l axis.

With an i n t e g r a l g r a t i c u l e e t c h e d on t h e t u b e s u r f a c e , the chance of measurement

e r r o r s due to p a r a l l a x o c c u r r i n g between t h e t r a c e and t h e g r a t i c u l e i s significantly

reduced. T h e r e i s a l s o a % d i s p l a y on the l e f t of the CRT f o r measuring the r i s e time.

© POWER switch (j^ON/JLOFF)

A pushbutton type s w i t c h t h a t t u r n s the power source on and o f f .

P r e s s i n g the s w i t c h t u r n s the power on. P r e s s i n g i t a g a i n t u r n s the power o f f .

(D Pi l o t lamp

L i g h t s up when the power i s turned on.

® SCALE ILLUM control

Controls the i 1 l u m i n a t i o n of the g r a t i c u l e on the CRT f a c e .

11
® R/0 INTEN control [CS-5230, 5260, 5270]
Controls t h e b r i g h t n e s s o f r e a d o u t v a l u e s s u c h a s s e t v a l u e s , measured v a l u e s , e t c ,

d i s p l a y e d on the CRT.

© TRACE ROTA control


C o n t r o l s the s l o p e of the t r a c e l i n e . The slope of the l i n e w i l l change due to e x t e r n a l

i n f l u e n c e s such as the magnetic f i e l d of t h e e a r t h . Use the s c r e w d r i v e r i n c l u d e d with

the a c c e s s o r i e s to a d j u s t the t r a c e l i n e p a r a l l e l w i t h the h o r i z o n t a l a x i s graticule.

® FOCUS control
A d j u s t s the focus to o b t a i n the c l e a r e s t d i s p l a y .

® INTENSITY control

A d j u s t s the b r i g h t n e s s of the t r a c e l i n e .

® CAL terminal
A voltage terminal f o rcalibration. Used f o r ad j u s t i n g t h e probe. O u t p u t s 1 Vp-p,

p o s i t i v e p o l a r i t y , approx. 1 kHz square waves.

® GND terminal
Used to s e t up a common ground w i t h other equipment.

© V. MODE s e l e c t o r
S e l e c t s the v e r t i c a l a x i s o p e r a t i o n modes.

CHI : CHI input s i g n a l i s d i s p l a y e d on the CRT.

ADD : A l g e b r a i c sum of CHI and CH2 s i g n a l s i s d i s p l a y e d on t h e CRT. I f CH2 i s

s e t to INV, d i f f e r e n c e between CHI and CH2 s i g n a l s i s d i s p l a y e d .

CH2 : CH2 input s i g n a l i s d i s p l a y e d on the CRT.

CH3 : CH3 Input s i g n a l i s d i s p l a y e d on the CRT.

ALT/CHOP: S e l e c t s t h e ALT and CHOP modes. I n t h e ALT mode, i n p u t signals are

d i s p l a y e d a l t e r n a t e l y e v e r y sweep i n t h e m u l t i - t r a c e mode. I n t h e CHOP

mode, i n p u t signals are displayed alternately at a r e p e t i t i o n r a t e of

approx. 250 kHz ( i n d u a l - t r a c e o p e r a t i o n ) , r e g a r d l e s s of sweep time.

12
REFERENCE =

A l t e r n a t e ( A L T ) and Chop (CHOP) Modes

When u s i n g t h e s e modes d u r i n g m u l t i - t r a c e o p e r a t i o n , t h e d i s p l a y I s d i v i d e d up
a c c o r d i n g to time.

I n the chop mode, each channel i s s u b d i v i d e d a c c o r d i n g to time w i t h i n each sweep.

N o r m a l l y , t h i s k i n d of measurement i s c a r r i e d out w i t h s i g n a l s of e i t h e r s l o w e r

sweep r a t e from 1 ms/div or low r e p e t i t i o n r a t e s where f l i c k e r i s q u i t e n o t i c e a b l e ,

I n t h e a l t e r n a t e mode, each c h a n n e l i s d i s p l a y e d one a f t e r a n o t h e r . Therefore,

each channel d i s p l a y appears much c l e a r e r . Normally, a f a s t e r sweep i s employed.

© VOLTS/DIV control

The vertical axis sensitivity i s c o n t r o l l e d w i t h t h e CHI v e r t i c a l a x i s attenuator.


The VOLTS/DIV control switches i t i n 1-2-5 s t e p s . Calibrated vert ical axis
s e n s i t i v i t y i s obtained by s e t t i n g V. VARI to the CAL p o s i t i o n .
I n X-Y o p e r a t i o n , the VOLTS/DIV c o n t r o l s e r v e s as a Y - a x i s a t t e n u a t o r .

Figure 1-2

13
® VARIABLE (V. VARI) c o n t r o l

Provides f i n e - c o n t r o l of t h e CHI v e r t i c a l a x i s sensitivity. S e t w i t h t h e VOLTS/DIV

control. The a t t e n u a t o r i s calibrated I f I t is fully turned clockwise t o t h e CAL

position.
In X-Y o p e r a t i o n , the VARIABLE c o n t r o l s e r v e s as a Y - a x i s f i n e a t t e n u a t o r control.

® ^POSITION control
A d j u s t s the v e r t i c a l p o s i t i o n of CHI waveform d i s p l a y e d on the CRT.
S e r v e s as a Y - a x i s p o s i t i o n c o n t r o l d u r i n g X-Y o p e r a t i o n .

® AC-GND-DC switch
S e l e c t s the CHI v e r t i c a l a x i s input s i g n a l c o u p l i n g mode.

AC : Input s i g n a l s a r e AC c o u p l e d , and DC components a r e removed. Low frequency

signal level falls by 3 dB o r more b e l o w 5 Hz i f the 1 : 1 probe or a

coaxial cable i s used. I t falIs below 0. 5 Hz i f a c a l i b r a t e d 10 : 1 probe

i s used.

GND : I n p u t of the v e r t i c a l a m p l i f i e r i s grounded, a l l o w i n g t h e ground p o t e n t i a l

to be c h e c k e d . The input r e s i s t a n c e i s 1 UQ with respect to t h e ground.

Input s i g n a l s w i l l not be grounded.

DC : I n p u t s i g n a l s a r e DC c o u p l e d , a l l o w i n g s i g n a l s c o n t a i n i n g DC components to

be observed.
The s w i t c h s e r v e s as a Y - a x i s input s e l e c t o r d u r i n g X-Y o p e r a t i o n .

® CH1 INPUT terminal


CHI v e r t i c a l a x i s input t e r m i n a l .
S e r v e s as a Y - a x i s input t e r m i n a l d u r i n g X-Y o p e r a t i o n .

® VOLTS/DIV control
CH2 v e r t i c a l a x i s a t t e n u a t o r . Has the same f u n c t i o n as the CHI VOLTS/DIV c o n t r o l .
S e r v e s as an X - a x i s a t t e n u a t o r d u r i n g X-Y o p e r a t i o n .

® VARIABLE (V. VARI)


CH2 v e r t i c a l a x i s attenuator f o r f i n e adjustment. Has t h e same f u n c t i o n a s t h e CHI

V. VARI c o n t r o l .
S e r v e s as an X - a x i s f i n e adjustment a t t e n u a t o r d u r i n g X-Y operation.

® ^ POSITION control
A d j u s t s the v e r t i c a l p o s i t i o n of CH2 waveform d i s p l a y e d on the CRT.

14
® AC-GND-DC switch

S e l e c t s the CH2 v e r t i c a l a x i s input s i g n a l c o u p l i n g mode i n the same way as the CHI


AC-GND-DC c o n t r o l .

Serves as an X - a x i s input s e l e c t o r during X-Y operation.

© CH2 INPUT terminal

CH2 v e r t i c a l a x i s input terminal.

Serves as an X - a x i s input terminal during X-Y operation.

© CH2 INV switch


When the switch i s pressed, the p o l a r i t y of CH2 input s i g n a l i s inverted.

© $ POSITION control

Adjusts the v e r t i c a l p o s i t i o n of CH3 waveform displayed on the CRT.

® CH3 INPUT terminal

CH3 v e r t i c a l a x i s input terminal.

® DELAY POSITION (COARSE), A R E F CURSOR


Coarsely a d j u s t s the delay time between A sweep and B sweep when H. MODE i s s e t to ALT
or B. (COARSE)

A d j u s t s the p o s i t i o n of the A R E F c u r s o r . [The CS-5230, CS-5260 and CS-5270 have the


cursor functions. ]

Figure 1-3 F r o n t Panel (CS-5270)

15
® DELAY POSITION ( F I N E ) , A CURSOR

Fine-adjustment the delay time.

A d j u s t s the p o s i t i o n of the A c u r s o r . [The CS-5230, CS-5260 and CS-5270 have t h e c u r s o r

function.]

@ CURSOR MODE (R/O OFF) [CS-5230, CS-5260 & CS-5270]

S e l e c t s t h e c u r s o r d i s p l a y mode and t u r n s t h e readout d i s p l a y on and o f f , Every time

the s w i t c h i s p r e s s e d , t h e c u r s o r d i s p l a y mode changes i n t h e o r d e r of O f f , ¥ c u r s o r ,

1 c u r s o r , and O f f . When t h e s w i t c h i s p r e s s e d f o r a second o r more d u r i n g readout

display, readout d i s p l a y i n c l u d i n g s c a l e f a c t o r d i s p l a y i s turned off. To t u r n t h e

readout d i s p l a y f u n c t i o n on, p r e s s the s w i t c h a g a i n .

® TRACE SEP control


A d j u s t s t h e v e r t i c a l p o s i t i o n of A sweep w i t h r e s p e c t to A sweep when the H. MODE i s s e t

to ALT and A sweep and B sweep a r e d i s p l a y e d a l t e r n a t e l y .

® HOLD OFF control


A d j u s t s t h e time between t h e end p o i n t o f an A sweep and t h e s t a r t point of another

A sweep. Use t h i s c o n t r o l t o a d j u s t t h e h o l d o f f time and a d j u s t the t r i g g e r l e v e l i f

triggering i s r e q u i r e d f o r complex s i g n a l s . Turning the control f u l l y counterclock-

wise to the NORM p o s i t i o n m i n i m i z e s the hold o f f time.

© H. MODE switch
S e l e c t s the h o r i z o n t a l a x i s d i s p l a y mode a s shown below:
A : Runs i n the A sweep mode.
ALT : D i s p l a y s A sweep and B sweep a l t e r n a t e l y .
B : Runs i n the B sweep mode.
X-Y : S e r v e s a s an X-Y o s c i l l o s c o p e w i t h c h a n n e l 1 a s t h e Y - a x i s and c h a n n e l 2 a s
the X - a x i s , r e g a r d l e s s of the V. MODE s e t t i n g .

© COUPLING switch

S e l e c t s the t r i g g e r s i g n i l c & u p i i n g .
AC : AC c t i p l e i t r i g g e r s i g n a l w i t h o u t DC components i s input t o t h e t r i g g e r

circuit. The AC p o s i t i o n i s used f o r normal waveform o b s e r v a t i o n .

HFrej : T r i g g e r s i g n a l i s passed though a low-pass f i l t e r and input to the t r i g g e r

circuit. The c u t - o f f frequency of the low-pass f i l t e r i s approx. 30 kHz.

DC : DC coupled t r i g g e r s i g n a l i s input to the t r i g g e r c u r c u i t .

TV FRAME: V e r t i c a l s y n c h r o n i z a t i o n p u l s e s a r e e x t r a c t e d from t h e c o m p o s i t e video

s i g n a l s and input t o the t r i g g e r c i r c u i t .

16
TV LIME : Higriabataiid si^nnfBLHrotroizlditeictnriEgfesresiaceifextracted from t h e composite v i d e o

@ SOURCE switch

S e l e c t s the t r i g g e r s i g n a l source.

VERT: T r i g g e r s i g n a l source I s s e l e c t e d w i t h V. MODE.

If thesingle trace (CHI, CH2, CHS o r ADD) I s s e l e c t e d w i t h V. MODE, the

selected Input signal I sthe t r i g g e r signal source. I f t h e A L T mode I s

selected for 2-trace, 3-trace or 4 - t r a c e operation, each Input s i g n a l i s the

t r i g g e r s i g n a l source i n each sweep operation.

NOTE

I f the chop mode i s s e l e c t e d , d i s p l a y i s changed a c c o r d i n g to t h e chop s w i t c h i n g

s i g n a l , which i s not synchronous w i t h the input s i g n a l . Thus, the input signal

cannot be used a s the t r i g g e r s i g n a l s o u r c e . The t r i g g e r s i g n a l source i s f i x e d

to channel 1 i f the chop mode i s s e l e c t e d i n Vert, operation.

CHI : Channel 1 input s i g n a l i s the t r i g g e r s i g n a l source.

CH2 : Channel 2 input s i g n a l i s the t r i g g e r s i g n a l s o u r c e .

CH3 : Channel 3 input s i g n a l i s the t r i g g e r s i g n a l s o u r c e .

L I N E : Waveform of the l i n e v o l t a g e i s the t r i g g e r s i g n a l source.

® AFTER D/B TRIG'D switch

S e l e c t s t h e continuous d e l a y or synchronous delay.

I n t h e c o n t i n u o u s d e l a y mode, B sweep b e g i n s a s soon a s t h e d e l a y time s e t w i t h t h e

DELAY POSITION c o n t r o l has passed a f t e r t h e A sweep s t a r t e d .

I n t h e synchronous d e l a y mode, B sweep begins a t the f i r s t t r i g g e r point a f t e r the delay

time s e t w i t h the DELAY POSITION c o n t r o l has passed a f t e r the A sweep s t a r t e d .

@ HDTV switch

I f TV FRAME c o u p l i n g or TV LINE c o u p l i n g i s s e l e c t e d w i t h t h e HDTV s w i t c h s e l e c t e d (ON),

o p e r a t i o n i s s y n c h r o n i z e d w i t h the HDTV composite video signal.

17
Figure 1-4

® T. MODE switch
S e l e c t s the t r i g g e r i n g mode.
AUTO : Sweep o p e r a t i o n i s executed a c c o r d i n g to the t r i g g e r s i g n a l . O p e r a t i o n runs

f r e e and the t r a c e l i n e appears i f there i s no t r i g g e r s i g n a l .


NORM : Sweep o p e r a t i o n i s executed a c c o r d i n g to the t r i g g e r s i g n a l . D i f f e r e n t from
AUTO, the t r a c e l i n e does not appear i f t h e r e i s no t r i g g e r s i g n a l .
FIX : Sweep o p e r a t i o n i s e x e c u t e d a c c o r d i n g t o t h e c e n t e r o f t h e a m p l i t u d e o f a
s i g n a l s e l e c t e d w i t h the SOURCE s w i t c h , which i s used a s the t r i g g e r point.

I n t h i s mode, the t r i g g e r l e v e l need not be a d j u s t e d .

SINGLE : S e l e c t s s i n g l e sweep mode.

RESET : S e t s the o s c i l l o s c o p e f o r t r i g g e r i n g i n the s i n g l e sweep mode.

® READY indicator

Shows t h a t t h e o s c i l l o s c o p e i s ready f o r t r i g g e r i n g i n the s i n g l e sweep mode. I t lights

by r e s e t t i n g T. MODE and e x t i n g u i s h e s when sweep i s completed.


® LEVEL c o n t r o l
Controls the t r i g g e r level. Specify a t which point of the t r i g g e r s i g n a l slope to

t r i g g e r and t o s t a r t sweep.

18
® SLOPE +/-

S e l e c t s the p o l a r i t y o f t h e s l o p e of the t r i g g e r s i g n a l . W i t h t h e b u t t o n not p r e s s e d


(+), t r i g g e r i n g i s e x e c u t e d a t the l e a d i n g edge of the Input t r i g g e r s i g n a l . W i t h the
b u t t o n p r e s s e d (-)» triggering I s e x e c u t e d a t the t r a i l i n g edge of the I n p u t t r i g g e r
signal.

® A SWEEP TIME/DIV switch

S w i t c h e s 22 ranges of A sweep time i n 1-2-5 s t e p s between 0. 5 second/div and 50 n s / d i v .

T u r n i n g the H. VARI f u l l y c l o c k w i s e to the CAL p o s i t i o n , the v a l u e i s c a l i b r a t e d .

[21 ranges between 0. 5 second/div and 0.1 / i s / d i v f o r the CS-5230 and CS-5235. ]
© B SWEEP TIME/DIV switch

S w i t c h e s 19 ranges of B sweep time i n 1-2-5 s t e p s between 50 ms/div and 50 n s / d i v .

[18 ranges between 50 ms/div and 0.1 jus/ih f o r the CS-5230 and CS-5235. ]

For p r a c t i c a l i t y , we recommend use of a higher-speed range than the A sweep s e t range.

© VARIABLE control (H. VARI)

F i n e - c o n t r o l s the A sweep time c o n t i n u o u s l y i n the range s e l e c t e d w i t h the A-SWEEP TIME/


DIV s w i t c h . T u r n i n g t h e c o n t r o l f u l l y c l o c k w i s e to the CAL p o s i t i o n , the A sweep time
is calibrated.

® POSITION control

A d j u s t s the h o r i z o n t a l p o s i t i o n of the waveform d i s p l a y e d on the CRT.

S e r v e s as an X - a x i s p o s i t i o n c o n t r o l d u r i n g X-Y o p e r a t i o n .

® X 1 0 MAG switch

P r e s s t h i s s w i t c h to magnify the d i s p l a y 1 0 x i n the c e n t e r of the CRT.


I NOTE 1

Do not use x 1 0 MAG i n X-Y o p e r a t i o n . O t h e r w i s e , the waveform can become n o i s y .

19
5 — 2 Rear Panel

Figure 2 Rear Panel

© CH1 OUTPUT terminal

Outputs AC couple CHI v e r t i c a l signal.

Used f o r c o n n e c t i n g a counter when measuring f r e q u e n c i e s .


:
I NOTE : — — — — — • • ; -

• When u s i n g a counter to measure f r e q u e n c i e s , t h e r e i s the p o s s i b i 1 i t y that

a c c u r a t e measurements a r e not d i s p l a y e d due to n o i s e i n t e r f e r e n c e . I n such a case,


a d j u s t the amplitude p r o p e r l y w i t h CHI VOLTS/DIV or V. VARI.

• Cascade c o n n e c t i o n between CHI, CHI OUTPUT, and CH2 i s not p o s s i b l e .

® Z. AXIS INPUT terminal

Input j a c k f o r e x t e r n a l i n t e n s i t y modulation s i g n a l s . I n t e n s i t y modulation i s performed

at TTL l e v e l . ( P o s i t i v e v o l t a g e reduces the i n t e n s i t y . )

I NOTE ^ 1

I f a v o l t a g e over 5 V ( T T L l e v e l ) i s i n p u t , t h e READ OUT d i s p l a y may be d i s t u r b e d

according to frequencies. I n such a case, lower t h e i n p u t to Z. AXIS down to t h e

TTL l e v e l or so.

20
® Fuse h o l d e r , line voltage selector

Use a 1 A f u s e I n 100/120 V a r e a s . Use a 630 mA fuse i n 220/230/240 V a r e a s .

Changing t h e v o l t a g e r a t i n g s h o u l d be done s t r i c t l y according to t h e d i r e c t i o n s I n

s e c t i o n 9. FUSE REPLACEMENT, LINE VOLTAGE SELECTION o n l y a f t e r d i s c o n n e c t i n g t h e power

cord from the power source Inlet,

® Power cord receptacle

A commercialy-used power source input connector.

5 - 3 Bottom Plate

Figure 3 Bottom P l a t e

© AST1G c o n t r o l

C o n t r o l s the t r a c e or spot a s t i g m a t i s m . A s t i g m a t i s m has been a d j u s t e d b e f o r e shipment.

I f necessary, a d j u s t t h e waveform to t h e optimum c o n d i t i o n w i t h t h e A S T I G and FOCUS

controls.

NOTE

Do not use a m e t a l l i c d e v i c e f o r a s t i g m a t i s m adjustment to a v o i d a c c i d e n t s .

Use the i n s u l a t e d s c r e w d r i v e r i n c l u d e d i n the a c c e s s o r i e s .

21
5 - 4 Readout Display [CS-5230, 5260, 5270]

1) Display positions

Figure 4 b e l o w shows t h e d i s p l a y p o s i t i o n s o f t h e s c a l e f a c t o r s and t h e v a l u e s

measured w i t h the c u r s o r .

©Cursor mode

©Cursor ©Cursor measurement


© C u r s o r mode
measurement (©Delay time)

© C H 2 (CH3)
©ADD scale factor (DB sweep scale factor

©CHI scale factor © C H 2 INV ®A sweep scale factor

Figure 4 Readout Display

2) Display
® Cursor mode

The f o l 1 lowing c u r s o r s a r e d i s p l a y e d a c c o r d i n g to the s e t t i n g of the c o n t r o l s .

AVI, A V 2 , A V 3 , A T , 1 / A T , RATIO, PHASE

® Cursor measurement

A v a l u e measured w i t h two c u r s o r s i s d i s p l a y e d .

• NOTE

I f two c u r s o r s a r e too near beyond the measuring l i m i t s i n the 1/AT mode, "?"

i s d i s p l a y e d f o l l o w i n g " 1 / A T " , i n d i c a t i n g t h a t the measured v a l u e has an e r r o r

out of s t a n d a r d .

(D Delay time
If t h e H. MODE i s s e t t o A L T o r B, "DELAY" i s displayed at the cursor mode
p o s i t i o n and t h e d e l a y time i s d i s p l a y e d a t the c u r s o r measurement p o s i t i o n . If
TRIG' D i s s e l e c t e d or t h e c u r s o r measurement i s l o c a t e d w i t h i n 0.2 d i v i s i o n from the
left s i d e of t h e CRT, DELAY i s f o l l o w e d by " ? " , i n d i c a t i n g t h a t t h e measured v a l u e
has an e r r o r .

V o l t a g e o r time cannot be measured w i t h the c u r s o r s d u r i n g d e l a y time d i s p l a y .


22
© CHI s c a l e f a c t o r

The s e n s i t i v i t y p e r 1 d i v i s i o n o f t h e CHI v e r t i c a l a x i s i s d i s p l a y e d when CHI i s


s e l e c t e d w i t h V. MODE. I f UNCAL i s s e l e c t e d , "CHI" i s f o l l o w e d by ">".
(D CH2 s c a l e f a c t o r (CH3 s c a l e f a c t o r )
Displayed i n the same c o n d i t i o n as CHI above.
NOTE

The CH3 s c a l e f a c t o r I s d i s p l a y e d when o n l y CHS I s s e l e c t e d w i t h V. MODE. The


sensitivity i s f i x e d to 0.1 V/div.

® ADD

I f ADD i s s e l e c t e d w i t h V. MODE, V I s displayed.

© CH2 INVERT

" 1 " i s d i s p l a y e d i f INV i s turned on.

® A sweep s c a l e f a c t o r

The time per ( 1 ) d i v i s i o n i s d i s p l a y e d when A, ALT or B i s s e l e c t e d w i t h H. MODE.

If UNCAL i s s e l e c t e d , "A" i s f o l l o w e d by ">". I f x 10 MAG i s t u r n e d on, t i m e i s


also converted ( i . e . , d i v i d e d by 1 0 ) .

I f X-Y i s s e l e c t e d w i t h H. MODE, "X-Y" i s d i s p l a y e d i n t h e A and B s c a l e f a c t o r


positions.
© B sweep s c a l e f a c t o r

The time p e r ( 1 ) d i v i s i o n i s d i s p l a y e d when ALT o r B i s s e l e c t e d w i t h H. MODE. I f


x10 MAG i s turned on, time i s a l s o c o n v e r t e d ( i . e . , d i v i d e d by 1 0 ) .

23
6. CHECKING AND ADJUSTMENT
PRIOR TO MEASUREMENT

In order to operate the o s c i l l o s c o p e a t i t s optimum p e r f o r m a n c e level, c a r r y out t h e

f o l l o w i n g c h e c k s and a d j u s t m e n t s b e f o r e making any measurements. The i n s t r u c t i o n s which

f o l l o w c o n c e r n i n g b a s i c o p e r a t i o n t e c h n i q u e s and a p p l i c a t i o n s assume t h a t t h e c h e c k s and

adjustments d e s c r i b e d here have been completed.

1) Adjust the c o n t r o l panel to the f o l l o w i n g s e t t i n g s .

CHI ( o r Y ) and CH2 ( o r X )

V. MODE CHI

^POSITION Center

V. VARI • CAL

VOLTS/DIV 5 V/DIV

AC-GND-DC GND

CH2 INVERT Off

HORIZONTAL

^•POSITION Center

H. VARI CAL

A SWEEP TIME/DIV 2 ms/DIV

x 10 MAG Off

TRIGGERING

H. MODE A

T. MODE AUTO

COUPLING • • AC

SOURCE • VERT

SLOPE - • +

TRIGGER LEVEL Center

A f t e r c h e c k i n g t h e power s o u r c e and v o l t a g e s e l e c t ion, press t h e POWER s w i t c h . The

pi l o t lamp 1 i g h t s up, and a t r a c e l i n e appears i n 10 t o 15 seconds. Check t o see that

r o t a t i n g t h e INTENSITY c o n t r o l clockwise increases the brightness, and r o t a t i n g i t

counterclockwise d e c r e a s e s the b r i g h t n e s s . Then, r o t a t e t h e INTENSITY c o n t r o l fully

counterclockwise and e x t i n g u i s h the t r a c e line to begin preheat ing. F o r t h e most

a c c u r a t e measurement r e s u l t s , i t i s n e c e s s a r y to warm-up the o s c i 1 l o s c o p e f o r about 30

minutes. However, i f you intend only to d i s p l a y waveforms, warm-up i s not n e c e s s a r y .

24
2) A f t e r warm-up a d j u s t t h e I N T E N S I T Y c o n t r o l so t h a t t h e t r a c e l i n e I s easy to s e e ,

and a d j u s t t h e FOCUS c o n t r o l t o o b t a i n t h e c l e a r e s t d i s p l a y Image. Then, u s e t h e

TRACE ROTA c o n t r o l t o b r i n g t h e t r a c e , l i n e p a r a l l e l w i t h t h e h o r i z o n t a l graduation

line.

3) P l u g t h e probe i n t o t h e INPUT t e r m i n a l o f each c h a n n e l . S e t t h e AC-GND-DC c o n t r o l to

DC. A t t a c h t h e CHI probe to the CAL t e r m i n a l . Set the VOLTS/DIV c o n t r o l to 20 mV/div.

[For the CS-5230, CS-5260 or CS-5270, s e t t h e VOLTS/DIV c o n t r o l t o 0.2 V/div. ]

A d j u s t the CHI y POSITION c o n t r o l so t h a t t h e whole waveform c a n be seen. With t h e

waveform i n t h i s p o s i t i o n , c a r r y out probe compensation adjustment a c c o r d i n g to F i g u r e

5 and the probe i n s t r u c t i o n manual.

CD L e f t waveform : E x c e s s i v e compensation
(D Middle waveform: I n s u f f i c i e n t compensation

(3) R i g h t waveform : Correct compensation

Adjust t h e probe trimmer so t h a t a. c o r r e c t l y


compensated waveform i s displayed.

Figure 5 Probe Compensation

4) S e t t h e VERT MODE c o n t r o l t o CH2 and c a r r y o u t c o m p e n s a t i o n a d j u s t m e n t o f t h e CH2

probe. A f t e r c o m p e n s a t i n g b o t h c h a n n e l p r o b e s , u s e t h e CHI probe e x c l u s i v e l y i n

channel 1 and the CH2 probe e x c l u s i v e l y i n channel 2. T h i s i s n e c e s s a r y because there

i s a s i i g h t c a p a c i t a n c e v a r i a t i o n between t h e two c h a n n e l s , and c o n f u s i n g t h e probes

w i l l cause changes i n compensation adjustments.

5) R e t u r n t h e VERT MODE c o n t r o l t o C H I , e a c h c h a n n e l ' s AC-GND-DC c o n t r o l t o t h e GND

setting, each channel's VOLTS/DIV c o n t r o l to 5 V/div, and t h e ^ P O S I T I O N and

POSITION c o n t r o l s to the c e n t e r .

T h i s i s what we r e f e r to a s t h e " i n i t i a l s e t t i n g " condition.

25
Figure 6 I n i t i a l S e t t i n g (CS-5270)
7. OPERATING PROCEDURES

7 — 1 Single Trace Operation

1) Alternating current d i spI ay

With the osci1loscope i n the i n i t i a l setting condition ( r e f e r t o 6. CHECKING AND

ADJUSTMENT PRIOR TO MEASUREMENT), t h e s i g n a l a p p l i e d t o t h e CHI INPUT t e r m i n a l i s

d i s p l a y e d on t h e CRT. Adjust t h e s i g n a l a m p l i t u d e t o an e a s y - t o - m e a s u r e s i z e by

c h a n g i n g t h e VOLTS/DIV c o n t r o l . Use t h e V. VARI c o n t r o l t o change t h e a m p l i t u d e

continuously. However, i f t h i s i s not n e c e s s a r y , l e a v e the s e t t i n g a t CA1.

Then, move t h e waveform to an e a s y - t o - o b s e r v e p o s i t i o n w i t h the SWEEP TIME/DIV c o n t r o l .

The H. VARI c o n t r o l should be s e t to CAL i f p o s s i b l e .

2) Trigger

T r i g g e r o p e r a t i o n i s r e q u i r e d i f the d i s p l a y e d waveform wavers or i s u n s t a b l e .


• Trigger level

T u r n t h e TRIGGER LEVEL c o n t r o l c l o c k w i s e or c o u n t e r c l o c k w i s e t o l o c k t h e waveform.


Some s i g n a l s a r e e a s i e r to see by changing the s l o p e . T h i s o p e r a t i o n w i t h the TRIGGER
LEVEL and SLOPE c o n t r o l s i s c a l l e d t r i g g e r p o i n t s e t t i n g . The o s c i l l o s c o p e s t a r t s
the sweep a t a s e t t r i g g e r p o i n t .

S L O P E " - " RANGE


+ Direction (Rotate Right)
S L O P E " + " RANGE

TRIGGER L E V E L

S L O P E " - " TRIGGER POINT S L O P E " + " TRIGGER POINT - Direction (Rotate Left)

Figure 7 Relationship between Trigger Level and Slope

• Trigger coupling

S e l e c t an intended trigger coupling.

AC : AC c o u p l e d t r i g g e r s i g n a l w i t h o u t DC components i s i n p u t t o t h e t r i g g e r

circuit. The AC p o s i t i o n i s used f o r normal waveform o b s e r v a t i o n higher


than 10 Hz.

HFrej : T r i g g e r s i g n a l p a s s e d though a low-pass f i I t e r i s input to t h e t r i g g e r


circuit. The c u t - o f f f r e q u e n c y of t h e low-pass f i 1 t e r i s approx. 30 kHz.
T h i s p o s i t i o n i s used f o r s i g n a l s w i t h high-frequency noise.

27
DC : DC coupled t r i g g e r s i g n a l i s input to the t r i g g e r c i r c u i t . This position i s

used f o r low-frequency s i g n a l s below 10 Hz.

TV FRAME: V e r t i c a l synchronization pulses a r e e x t r a c t e d from the composite video

s i g n a l s and input to the t r i g g e r c i r c u i t .

TV LINE : H o r i z o n t a l s y n c h r o n i z a t i o n p u l s e s a r e e x t r a c t e d from t h e c o m p o s i t e video

s i g n a l s and input to the t r i g g e r c i r c u i t .

• T r i g g e r mode

S e l e c t an intended t r i g g e r mode.

AUTO : Operat ion r u n s f r e e and t h e t r a c e 1ine a p p e a r s even i f t h e r e i s no t r i g g e r

signal. Use NORM to observe low-frequency waveform below 40 Hz.

NORM : Sweep o p e r a t i o n i s executed according to t h e t r i g g e r s i g n a l . Sweep operation

i s not executed i f t h e r e i s no t r i g g e r s i g n a l .

FIX : Sweep o p e r a t i o n i s executed according to t h e c e n t e r o f t h e a m p l i t u d e o f t h e


t r i g g e r s i g n a l a s the t r i g g e r p o i n t . I n t h i s mode, the t r i g g e r l e v e l need not

be adjusted.

SINGLE: S e l e c t s the s i n g l e sweep mode. (See 7-6)

RESET : S e t s the o s c i l l o s c o p e f o r t r i g g e r i n g i n the s i n g l e sweep mode. (See 7-6. )

3) Composite video signal display


When i n p u t t i n g composite video s i g n a l s , s e t the COUPLING c o n t r o l to e i t h e r TV FRAME or
TV LINE. A l s o s w i t c h the SLOPE c o n t r o l i n accordance w i t h the s i g n a l p o l a r i t y .

[SLOPE TV-F TV-L

Figure 8 Relationship between COUPLING and SLOPE


7 — 2 Mu 11 i-Trace Operation

1) Switching v e r t i c a l operation modes

When two o r more s w i t c h e s . o t h e r t h a i t h e ALT/CHOP a r e a c t i v a t e d w i t h t h e V. MODE,

the CRT s c r e e n d i s p l a y s t h e s i g n a l s a p p l i e d t o the INPUT t e r m i n a l s o f t h e a c t i v a t e d

channels. Changing the sweep time.and s e t t i n g the t r i g g e r p o i n t may be completed I n

the same way a s C l l s i n g l e t r a c e o p e r a t i o n . When the V. MODE c o n t r o l i s s e t t o ALT,

s i g n a l s of t h e s e l e c t e d channels a r e d i s p l a y e d one a f t e r another f o r each sweep. When

it i s s e t t o CHOP, t h e s i g n a l s a r e s u b - d i v i d e d according t o time and d i s p l a y e d on

the CRT. When t h e V. MODE c o n t r o l i s s e t t o ADD, t h e s i g n a l s ( C H I and CH2) a r e

d i s p l a y e d i n the same manner a s i n a s i n g l e channel o p e r a t i o n . I f t h e INV c o n t r o l i s

selected i n t h e ADD mode, t h e d i f f e r e n t i a l ( C H I -CH2) i s d i s p l a y e d . I n o r d e r t o

measure d i s p l a y e d waveforms i n t h e ADD mode, i t i s n e c e s s a r y t h a t t h e VOLTS/DIV

c o n t r o l s e t t i n g s - b e . t h e same f o r both c h a n n e l s .

2) Switching t r i g g e r sources
S e l e c t a t r i g g e r s i g n a l channel w i t h t h e SOURCE s w i t c h . I f VERT i s s e l e c t e d w i t h the
SOURCE s w i t c h and t h e ALT/CHOP s w i t c h i s s e t t o ALT, t h e s i g n a l o f each c h a n n e l i s
used a s t h e t r i g g e r s i g n a l s o u r c e i n each sweep. I f t h e ALT/CHOP s w i t c h i s s e t to

CHOP, only CHI input s i g n a l i s used a s the t r i g g e r s i g n a l source.


. N 0 T E
— _ __ :
_ _ — _ _____ , ; — _
I f the SOURCE s w i t c h i s s e t to VERT, input s i g n a l s to a l l c h a n n e l s . Sweep may be
unstable i f a c h a n n e l c a r r i e s no s i g n a l o r a s i g n a l o f an a m p l i t u d e below t h e
t r i g g e r s e n s i t i v i t y of the o s c i l l o s c o p e .

I f the t r i g g e r c o u p l i n g i s used to s e t TV FRAME or TV LINE, t r i g g e r i n g w i t h p l u r a l


: channels i s disabled. S e t SOURCE t o an intended t r i g g e r i n g channel, i n s t e a d of
VERT.

3) Line trigger

If the s i g n a l of the d i s p l a y e d channel i s synchronous w i t h the l i n e frequency, the

t r i g g e r p o i n t becomes s t a b l e by s e t t i n g the SOURCE s w i t c h to LINE.

29
7 — 3 Sweep m a g n i f i e d Operation

then m a g n i f y i n g a p o r t i o n of the d i s p l a y e d waveform i n terms of time, i n c r e a s i n g t h e sweep

speed may cause the waveform p o r t i o n to be measured to d i s a p p e a r from t h e s c r e e n . Use the

POSITION c o n t r o l t o move t h e waveform p o r t i o n to be m a g n i f i e d t o t h e m i d d l e o f t h e

CRT s c r e e n . Then, p r e s s t h e x 1 0 MAG s w i t c h t o m a g n i f y t h e w a v e f o r m 10 t i m e s i n the

horizontal direction.

7 — 4 DeIayed Sweep O p e r a t i o n

Delayed sweep ( B sweep) i s another way t o observe a waveform p o r t i o n by m a g n i f y i n g i t i n

terms of time. I t a l l o w s t h e m a g n i f i c a t i o n to be s e t f r e e l y and t h e m a g n i f i e d and non-

m a g n i f i e d p o r t i o n s t o be d i s p l a y e d a l t e r n a t e l y by u s i n g the ALT sweep f u n c t i o n .

1) S e t t h e H. MODE c o n t r o l to A. Adjust t h e c o n t r o l s so t h a t t h e p o r t i o n of t h e input

waveform to be observed i s d i s p l a y e d on the CRT.

2) S e t t h e AFTER D/B TRIG' D c o n t r o l t o AFTER. D. S e t t h e H. MODE c o n t r o l to ALT. The


waveform p o r t ion t o be m a g n i f i e d i s intensity-modulated and d i s p l a y e d b r i g h t . I n
F i g u r e 9, t h e upper t r a c e 1 i n e shows t h e n o n - m a g n i f i e d w a v e f o r m , and t h e l o w e r
t r a c e l i n e shows the m a g n i f i e d waveform.

3) P r o p e r l y a d j u s t the time of d i s p l a y i n g the m a g n i f i e d p o r t i o n w i t h the B SWEEP TIME/DIV,

the time p o s i t i o n o f t h e m a g n i f i e d p o r t i o n w i t h t h e DELAY POSITION COARSE and FINE

c o n t r o l s , and the r e l a t i o n s h i p between A sweep and B sweep v e r t i c a l p o s i t i o n s w i t h the

TRACE SEP.

4) Set t h e H. MODE c o n t r o l to B. Only the m a g n i f i e d p o r t i o n i s d i s p l a y e d on the CRT.

• NOTE 1

If t h e m a g n i f i c a t i o n o f A sweep and B sweep e x c e e d s s e v e r a l hundred t i m e s with

AFTER D s e t t i n g , d e l a y j i t t e r can appear. For j i t t e r - f r e e observation, set the

c o n t r o l t o B T R I G ' D. The t r i g g e r s i g n a l s e l e c t e d w i t h t h e SOURCE s w i t c h i s t h e

t r i g g e r s i g n a l f o r B sweep. B sweep s t a r t point i s s e t w i t h the DELAY P0SIT0N and

TRIGGER LEVEL. Thus, the delay time i s f o r r e f e r e n c e o n l y .

30
A Sweep intensified zone to be magnified.

B Sweep.

Figure 9 ALT S l e e p O p e r a t i o n

7 - 5 X-Y Operation

The o s c i l l o s c o p e not o n l y has a l l the f u n c t i o n s o f a c o n v e n t i o n a l osci1loscope but may


a l s o be o p e r a t e d a s an X-Y o s c i 1 l o s c o p e . I n X-Y o p e r a t i o n , signals applied t o t h e CHI
INPUT t e r m i n a l a r e d e f l e c t e d on the Y - a x i s ( v e r t i c a l a x i s ) , signals applied t o t h e CH2
INPUT t e r m i n a l s a r e d e f l e c t e d on t h e X - a x i s ( h o r i z o n t a l a x i s ) , and a L i s s a j o u s p a t t e r n
is displayed. A L i s s a j o u s p a t t e r n makes i t possible t o measure phase d i f f e r e n c e s
between two s i g n a l s and t h e i r r e l a t i v e frequency proportion.
| NOTE 1

Waveforms may be n o i s y by p r e s s i n g the x 1 0 MAG s w i t c h . Do not use the x 1 0 MAG s w i t c h


for X-Y o e p r a t i o n .

7 — 6 Single Sweep O p e r a t i on

S i n g l e sweep o p e r a t i o n i s used to observe a s i n g l e event or occurence.

If the t r i g g e r point o f a s i g n a l t o be o b s e r v e d i s unknown, s e t the t r i g g e r p o i n t using

the s i g n a l or a s i g n a l approximate to i t .

1) S e t t h e T. MODE s w i t c h t o AUTO or NORM. Adjust t h e t r i g g e r p o i n t w i t h t h e TRIGGER

LEVEL c o n t r o l . I f the COUPLING s w i t c h i s s e t t o DC, the t r i g g e r p o i n t may be a d j u s t e d

a c c u r a t e l y f o r a s i g n a l which changes s l o w l y .

2) Set the T. MODE s w i t c h to RESET, and check the READY LED i s l i t .

31
3) Observe t h e s i g n a l . When sweep I s e x e c u t e d once, t h e READY LED goes out. I f obser-

v a t i o n I s continued, repeat step 2 ) .

I NOTE • • • • 1

•A sweep and B sweep a r e a l t e r n a t e d i n t h e s i n g l e sweep mode w i t h t h e H. MODE

s w i t c h s e t to A L L
9
I n s i n g l e sweep o p e r a t i o n I n the m u l t i - t r a c e ALT mode, s e l e c t e d s i g n a l change i n

e a c h sweep c a n n o t be o b s e r v e d s i m u l t a n e o u s l y . S e l e c t chop mode t o o b s e r v e

them s i m u l t a n e o u s l y .

7 - 7 Readout Operation [CS-5230, 5260, 5270]

1) Set value ( s c a l e f a c t o r ) d i s p l a y

The o s c i l l o s c o p e d i s p l a y s t h e s c a l e f a c t o r s when t h e power i s t u r n e d on. A d j u s t the


b r i g h t n e s s w i t h the R/0 INTEN c o n t r o l as n e c e s s a r y . To t u r n o f f t h e readout d i s p l a y ,

p r e s s t h e CURSOR MODE s w i t c h f o r a p p r o x i m a t e l y two s e c o n d s . To t u r n i t on a g a i n ,

press the switch again.


NOTE — — — — — _ — :

• The o s c i l l o s c o p e d i s p l a y s readout d a t a i n terms o f waveforms and time s h a r i n g .


Waveforms may o c c a s i o n a l l y look like i n t e n s i t y - m o d u l a t e d waveforms, depending
on c o m b i n a t i o n s of waveform types and sweep t i m e s . I f the d i s p l a y i s not
suitable f o r o b s e r v a t i o n o r measurement, t u r n t h e r e a d o u t d i s p l a y o f f by t h e
above procedure.

• When a s i g n a l out of t h e e f f e c t i v e a r e a i s d i s p l a y e d , t h e readout d i s p l a y may


waver o r be t h i c k . We recommend t h a t waveforms be d i s p l a y e d i n t h e e f f e c t i v e

area.

2) Cursor measurement

During readout d i s p l a y , every p r e s s of t h e CURSOR MODE s w i t c h changes the c u r s o r mode

i n the o r d e r of O f f , A V , A T , 1/AT, and O f f . Data i s d i s p l a y e d on the upper p a r t of

the CRT.

A V (Voltage measurement mode)

Two horizontal cursors are displayed. D i s t a n c e between t h e s e c u r s o r s i s c o n v e r t e d

into the v e r t i c a l a x i s s e n s i t i v i t y ( V O L T S / D I V ) , and t h e v o l t a g e i sdisplayed. If a

probe i s a t t a c h e d , i t s a t t e n u a t i o n r a t i o i s converted and d i s p l a y e d . CHI and CH2 data

are d i s p l a y e d simultaneously i f s e l e c t e d w i t h V. MODE. I f V. VARI i s s e t t o UNCAL,

v o l t a g e r a t i o (RATIO) i s d i s p l a y e d , u s i n g 5 d i v i s i o n s on the CRT a s 1 0 0 1

32
A T . t / A T (Time and frequency- measurement mode)

Two v e r t i c a l c u r s o r s a r e d i s p l a y e d . The d i s t a n c e between t h e s e c u r s o r s i s c o n v e r t e d

i n t o sweep t i m e , and time and f r e q e n c y , the r e c i p r o c a l of the time, i s displayed

simultaneously. I f H. VARI i s s e t t o UNCAL, AT data I s voltage ratio, using

5 divisions on t h e CRT a s 1 0 0 % , a n d 1 / A T data i s phase difference, using

5 d i v i s i o n s as 360° .

33
8. APPLICATIONS

Since b o t h the v e r t i c a l and h o r i z o n t a l a x e s of the o s c i l l o s c o p e have been c a l i b r a t e d ,

the o s c i l l o s c o p e i s capable of not only d i s p l a y i n g waveforms but a l s o measuring v o l t a g e or

time q u a n t i t a t i v e l y . For t h e s e measurements, be s u r e to r o t a t e the V. VARI and H. VARI

controls f u l l y clockwise to the CAL p o s i t i o n ( u n t i l a c l i c k sound i s h e a r d ) . We recommend

that t h e probe ( i n c l u d e d i n t h e a c c e s s o r i e s ) be u s e d t o m i n i m i z e i n t e r f e r e n c e to the

s i g n a l s you want to measure.

8 — 1 measuring W o l t a g e B e t w e e n Two Po i n t s on a Waveform

Use t h e f o l l o w i n g procedures f o r measuring voltage between two p o i n t s , peak t o peak

voltage, e t c .

1) A p p l y a s i g n a l to the INPUT t e r m i n a l , and a d j u s t the VOLTS/DIV and SWEEP TIME/DIV


controls. Reset the t r i g g e r point i f necessary.

Set the AC-GND-DC c o n t r o l to AC.

2) Use the y POSITION c o n t r o l so t h a t one of the p o i n t s to be measured f a l l s on one of


the h o r i z o n t a l g r a d u a t i o n l i n e s , w h i l e the other p o i n t can s t i l l be observed on the CRT
screen.

3) Use the POSITION c o n t r o l so t h a t the p o i n t to be measured f a l l s on the v e r t i c a l

s c a l e a t the c e n t e r of the CRT.

4) Measure t h e v e r t i c a l distance between the two p o i n t s . Multiply that value by t h e

VOLTS/DIV s e t t i n g .

When t h e 10:1 probe i s used, a l s o m u l t i p l y the v a l u e by t h e a t t e n u a t i o n r a t e of the

probe.

34
Adjust to the center
vertical scale with ,
Points to be measured
^•POSITION.

between two points


Vertical distance

Adjust to the horizontal


scale with £ POSITION.

Figure 10-1 Measuring Voltage between Two Points

Example : I n F i g u r e 10, t h e v e r t i c a l d i s t a n c e between t h e two p o i n t s i s 4.4 d i v . The


voltage i s found as shown below, p r o v i d e d t h e VOLTS/DIV c o n t r o l i s s e t to 0. 2
V/div and the 10:1 probe i s used:

Voltage between 2 p o i n t s = 4.4 ( d i v ) x 0 . 2 ( V / d i v ) x 10 = 8. 8 V

• Cursor measurement [CS-5230, 5260, 5270]

(1) Move the waveform to be measured to an e a s y - t o - s e e p o s i t i o n .

(2) Press t h e CURSOR MODE s w i t c h t o s e l e c t t h e v o l t a g e measurement mode. B r i n g the


A REF and A cursors i n contact with the points t o be m e a s u r e d . Read t h e
measurement.

cursor Measured value

REF cursor

F i g u r e 10-2 Measuring Voltage between Two Points (Cursor measurement)


8 - 2 E i iminat ion of undesi red s i g n a l components

1
By u s i n g the V. MODE c o n t r o l s ADD s e t t i n g , u n d e s i r e d s i g n a l components can be e l i m i n a t e d ,

a l l o w i n g only d e s i r e d s i g n a l components to be d i s p l a y e d .

1) Apply s i g n a l s i n c l u d i n g u n d e s i r e d components to t h e CHI INPUT t e r m i n a l .

Now apply t h e components you want to e l i m i n a t e t o t h e CH2 INPUT t e r m i n a l .

2) S e t t h e V. MODE s w i t c h t o A L T o r CHOP. S e t t h e SOURCE c o n t r o l t o CH2. Set the

t r i g g e r point t o t h e CH2 s i g n a l , and make s u r e that CH2 c o n t a i n s the undesired

components o f C H I .

3) Press t h e INV s w i t c h , and make s u r e that t h e CH2 s i g n a l r e p r e s e n t s the undesired

components i n the r e v e r s e polarity.

When t h e V. MODE s w i t c h i s s e t t o ADD i n t h e s e c o n d i t i o n s , only the desired signal

components a r e d i s p l a y e d on the CRT.

CH : S i g n a l w i t h u n d e s i r e d components

ADD : S i g n a l w i t h o u t u n d e s i r e d component

-CH2 : Undesired component

Figure 11 Common Mode R e j e c t i o n

4) E l i m i n a t i o n c a p a b i l i t i e s v a r y w i t h t h e s i z e o f t h e u n d e s i r e d component. I n order to

a t t a i n the b e s t r e s u l t s , d i s p l a y the CH2 s i g n a l w i t h s l i g h t l y h i g h e r VOLTS/DIV setting.

A f t e r p r e s s i n g the INV s w i t c h and a c t i v a t i n g the ADD f u n c t i o n , a d j u s t the CH2 V. VARI

c o n t r o l to g e t a good waveform.

36
8 - 3 measuring Di r e c t Current (DC) Voltage

The o s c i l l o s c o p e ' s v e r t i c a l a m p l i f i c a t i o n I s proven by a d i r e c t c u r r e n t a m p l i f i e r c i r c u i t

c h a r a c t e r i z e d by e x c e l l e n t s t a b i 1 i t y . By s w i t c h i n g t h e AC-GND-DC c o n t r o l t o t h e DC

s e t t i n g , d i r e c t c u r r e n t v o l t a g e may be measured.

1) Apply the s i g n a l to the INPUT t e r m i n a l . Use the VOLTS/DIV and SWEEP TIME/DIV c o n t r o l s

to d i s p l a y the waveform a t an e a s y - t o - s e e size. Adjust the TRIGGER LEVEL c o n t r o l a s

necessary.

2) Set the T. MODE c o n t r o l to AUTO. S e t the AC-GND-DC c o n t r o l to GND. The t r a c e l i n e i s

d i s p l a y e d on t h e CRT. T h i s t r a c e 1ine r e p r e s e n t s t h e ground p o t e n t i a l . Use t h e

^POSITION c o n t r o l to b r i n g the t r a c e 1ine i n c o n t a c t w i t h one o f t h e h o r i z o n t a l

graduation lines. I t i s normally l i n e d up on the OX g r a d u a t i o n l i n e i f the signal i s

positive. Or, i t i s 1 i n e d up on t h e 100% 1 i n e i f the s i g n a l i s negative. Once

1 i n e d up, t h e t r a c e p o s i t i o n becomes t h e r e f e r e n c e p o t e n t i a l , so do n o t t o u c h t h e

^POSITION c o n t r o l d u r i n g measurement processes.

3) S e t the AC-GND-DC c o n t r o l to DC. The s i g n a l i s d i s p l a y e d on t h e CRT w i t h the d i r e c t

c u r r e n t components intact.

I f e i t h e r the VOLTS/DIV or r e f e r e n c e p o t e n t i a l s e t t i n g i s i n a p p r o p r i a t e , the waveform

may d i s a p p e a r from the d i s p l a y s c r e e n . Make sure the s e t t i n g s a r e c o r r e c t .

4) Measure the p o t e n t i a l i n the e q u i v a l e n t procedure to measuring the v o l t a g e between two

points. The p o t e n t i a l s i g n i s p l u s i f above t h e r e f e r e n c e o r minus i f below t h e

reference.

Measuring point adjusted to the center


vertical scale by ^ • POSITION

Ground potential adjusted by


I POSITION (reference line)

F i g u r e 12-1 DC Voltage Measurement

37
5) I f there i s only one s i g n a l t o be m e a s u r e d , a p p l y i t t o CHI, a n d u s e t h e CH2

^ POSITION c o n t r o l so t h a t CH2 d i s p l a y s the ground p o t e n t i a l . I f t h e V. MODE c o n t r o l

I s s e t t o ALT o r CHOP, you w i l l be c e r t a i n o f t h e ground p o t e n t i a l t h r o u g h o u t t h e

p r o c e d u r e . However, make c e r t a i n t h a t t h e ground p o t e n t i a l s o f b o t h c h a n n e l s a r e

always the same.

- Cursor measurement [CS-5230, 5260, 5270]

(1) C a r r y out s t e p s 1 and 2 f o r measuring w i t h t h e s c a l e on the CRT.

(2) P r e s s the CURSOR MODE s w i t c h to s e l e c t the v o l t a g e measurement mode. A l i g n the A R E

F c u r s o r t o the ground p o t e n t i a l .

(3) Set the AC-GND-DC c o n t r o l to DC. A d j u s t t h e A c u r s o r to the p o i n t to be measured.

Read t h e measurement.

Measured value

cursor

REF
cursor

F i g u r e 12-2 DC Voltage Measurement (Cursor measurement)


8 - 4 Measuring Voltage Rat io w i t h C u r s o r s [CS-5230, 5260, 5270]

Overshoot of a square wave, e t c . may e a s i l y be measured w i t h the c u r s o r s .

1) Apply t h e s i g n a l to the INPUT t e r m i n a l . Use the VOLTS/DIV and SWEEP TIME/DIV c o n t r o l s


to d i s p l a y the waveform a t an e a s y - t o - s e e p o s i t i o n .

2) A d j u s t the amplitude to 5 d i v i s i o n s w i t h the V. VARI c o n t r o l .

3) P r e s s t h e CURSOR MODE s w i t c h t o s e l e c t t h e v o l t a g e measurement mode. A l i g n theAREF

c u r s o r to the top of the square wave and the A c u r s o r to the peak of the overshoot.

4) When t h e V. V A R I c o n t r o 1 i s s e t t o UNCAL, r a t i o measurement i s s e l e c t e d and


a c c o r d i n g l y the d i s p l a y e d d a t a r e p r e s e n t s the percentage of the overshoot.

-Readout : Measurement
cursor
REF cursor

Figure 13 Measuring Voltage Ratio with Cursors

8 — 5 Measur i ng S i g n a l s with Low Frequency Components

When t h e o s c i l l o s c o p e ' s AC-GND-DC c o n t r o l i s s e t to AC, t h e r e i s a chance t h a t errors

may o c c u r i n t h e v o l t a g e measurement. This inaccuracy i s c a u s e d by low range c u t - o f f

frequencies. At AC, t h e most a c c u r a t e frequency measurements a r e r e a l i z e d above t h e 20 to

30 Hz range. T h e r e f o r e , when measuring f r e q u e n c i e s below t h i s range, s w i t c h the AC-GND-DC

c o n t r o l t o t h e DC s e t t i n g . I f you a r e u s i n g t h e probe s u p p l i e d w i t h t h e o s c i 1 l o s c o p e ,

a c c u r a t e measurement of f r e q u e n c i e s as low a s 2 to 3 Hz can be r e a l i z e d a t AC.

, 39
8 — 6 M e a s u r i n g S i g n a I s wi t h H i g h F r e q u e n c y Components

Always use a probe when measuring p u l s e s or s i g n a l s of a few hundred kHz or above.

This i s because d i s t o r t i o n w i l l o c c u r i n the waveform's h i g h f r e q u e n c y component due to

the use of long l e a d s . T h i s makes i t d i f f i c u l t to conduct a c c u r a t e waveforms. This i s

also t r u e f o r probes w i t h l o n g ground l e a d s , so keep them a s s h o r t as p o s s i b l e . In

a d d i t i o n , e n s u r e the ground l e a d c l i p i s c o n n e c t e d to a ground p o t e n t i a l c l o s e to the

s i g n a l to be measured.

Point of measurement

Ground pattern

F i g u r e 14 Measuring S i g n a l s with High Frequency Components

8 — 7 M e a s u r i ng T i me B e t w e e n Two Po i n t s

When measuring time between two p o i n t s , measurements can be determined from SWEEP TIME/DIV

and h o r i z o n t a l distance.

1) Display the waveform by a d j u s t i n g each c o n t r o l . Set a l l the VARIABLE c o n t r o l s to the

CAL p o s i t i o n .

2) Use the ^ • P O S I T I O N c o n t r o l t o b r i n g one p o i n t t o be m e a s u r e d in line with a

V e r t i c a l graduation l i n e . Then, use the y POSITION c o n t r o l to b r i n g the o t h e r point

to be measured i n l i n e w i t h the h o r i z o n t a l s c a l e i n the middle of the CRT.

3) Measure t h e h o r i z o n t a l d i s t a n c e between t h e two p o i n t s . Multiply this v a l u e by the

SWEEP TIME/DIV s e t t i n g v a l u e . I f the x 10 MAG function i s a c t i v a t e d , m u l t i p l y the

v a l u e by 1/10.

40
A d j u s t t o t h e vertical
scale with • POSITION

Adjust to horizontal
Horizontal distance center line w i t h
between 2 signals t POSITION

F i g u r e 15-1 Measuring Time Between Two P o i n t s

Example : I n t h e c a s e o f F i g u r e 15, t h e h o r i z o n t a l d i s t a n c e between t h e two p o i n t s i s


5.4 d i v . The time between t h e two p o i n t s i s found a s f o l l o w s , p r o v i d e d t h e
SWEEP TIME/DIV s e t t i n g i s 0.2 ms/div.

Time between two p o i n t s = 5. 4 ( d i v ) x O . 2 ( m s / d i v ) = 1. 08 (ms)


I f the x 1 0 MAG f u n c t i o n i s used:

Time between two p o i n t s - 5.4 ( d i v ) x O . 2 ( m s / d i v ) x 1/10

= 0.108 (ms) = 108 0 * s )


• Cursor measurement [CS-5230, 5260, 5270]
(1) D i s p l a y the waveform to be measured i n an e a s y - t o - s e e p o s i t i o n .
(2) P r e s s the CURSOR MODE s w i t c h to s e l e c t the time and frequency mode. Bring the AREF
and A c u r s o r s i n c o n t a c t w i t h the p o i n t s to be measured. Read the measurement.

Measured value

REF cursor cursor

Figure 15-2 Measuring Time between Two Points (Cursor measurement)

41
8 - 8 Measuring Time R a t i o with Cursors [CS-5230, 5260, 5270]

The duty of a square wave can e a s i l y be measured w i t h t h e c u r s o r s .

1) Apply the s i g n a l to the INPUT terminal. Use the VOLTS/DIV and SWEEP TIME/DIV c o n t r o l s

to d i s p l a y the waveform a t an e a s y - t o - s e e p o s i t o n .

2) A d j u s t the w i d t h of a c y c l e to 5 d i v i s i o n s w i t h the H. VARI c o n t r o l .

3) P r e s s the CURSOR MODE s w i t c h to s e l e c t the time and frequency measurement mode. Align

the AREF cursor to the t r a i l i n g edge o f t h e s q u a r e wave and t h e A cursor to the

l e a d i n g edge.

4) When t h e H. VARI c o n t r o l i s s e t t o UNCAL, r a t i o / p h a s e measurement i s s e l e c t e d and

a c c o r d i n g l y the d i s p l a y e d d a t a r e p r e s e n t s the duty of the square wave.

Measured value

I
- I J J J
" -
J U 1 cursor

Figure 16 ' Measuring Time Ratio with Cursors

8 — 9 M e a s u r i ng F r e q u e n c i e s

S i n c e the frequency i s found as a r e c i p r o c a l of a p e r i o d , measure the time ( p e r i o d ) of one

c y c l e and c a l c u l a t e i t s r e c i p r o c a l v a l u e .

1) Measure t h e time of one c y c l e .

2) F i n d the r e c i p r o c a l v a l u e of the p e r i o d found.

42
1 cycle = 4 0 /TS (5 /-IS/divx 8 div)

Figure 17-1 Measuring Frequency

Example : I n the case of F i g u r e 17, the frequency i s found as shown below, provided the
period i s 40 / « .

3
Freqyency = - = 25X 10 = 25 kHz
40X10-"

• Cursor measurement [CS-5230, 5260, 5270]

(1) Display the waveform to be measured in an easy-to-see position.

(2) Press the CURSOR MODE switch to s e l e c t the time and frequency mode. Bring the AREF

and A c u r s o r s in contact with the points to be measured. Read the measurement.

Measured value

F i g u r e 17-2 Measuring Frequency (Cursor measurement)

43
8 —10 measuring Pulse R i s e and F a l l Times

The r i s e ( f a l l ) time i s found by measuring the time between 10% and 90% of the peak v a l u e .

For t h i s purpose, t h e o s c i l l o s c o p e has a d d i t i o n a l g r a d u a t i o n s a t 10% and 9 0 1

1) Apply the s i g n a l . A d j u s t t h e VOLTS/DIV and V. VARI c o n t r o l s so t h a t t h e a m p l i t u d e i s

5 divisions. S e t t h e H. VARI c o n t r o l to CAL. P r e s s the x 1 0 MAG s w i t c h i f n e c e s s a r y .

2) Use the y POSITION c o n t r o l t o move t h e waveform between 10% and 9 0 1 Then, move t h e

10% p o i n t to a v e r t i c a l graduation l i n e w i t h the POSITION. Measure t h e h o r i z o n t a l

distance between t h e 10% and 9 0 % p o i n t s . The t i m e i s found from t h e h o r i z o n t a l

distance.

Adjust to the vertical scale


with • POSITION

Rise time

Figure 18-1 Measuring R i s e Time

3) When m e a s u r i n g h i g h speed r i s e (fall) time, you must a l l o w f o r t h e o s c i l l o s c o p e ' s

inherent time f a c t o r . The t r u e r i s e time f o r a measurement of 20 n s i s found a s

follows:

2 2
to = V t m — t r tm: A c t u a l l y measured v a l u e

t r : Oscilloscope's inherent r i s e time


2 2
= V20 - 8 . 8 = 18. 0ns [CS-5230, 5235]

However, t h i s c o r r e c t i o n f a c t o r i s not s i g n i f i c a n t when the a c t u a l l y measured v a l u e i s

above 30 n s .

• Cursor measurement [CS-5230, 5260, 5270]


(1) C a r r y out s t e p 1 of o b s e r v a t i o n w i t h the s c a l e on the CRT.

44
(2) P r e s s the CURSOR MODE s w i t c h to s e l e c t the time and frequency mode. A l i g n the A R E F

c u r s o r t o the 101 point and the, A c u r s o r to the 901 p o i n t . Read the measurement.

Measured value

REF cursor cursor

Figure 18-2 Measuring R i s e ( F a l I ) Time

8 —11 Measuring Pulse Jitter

P u l s e j i t t e r w i t h a r e l a t i v e l y long c y c l e may be measured u s i n g the B sweep f u n c t i o n .

1) Apply the s i g n a l t o the INPUT t e r m i n a l . A d j u s t the a m p l i t u d e t o 5 d i v i s i o n s w i t h the


VOLTS/DIV and V. VARI c o n t r o l s . S e t the VOLTS/DIV and H. VARI t o CAL. Display the
waveform i n an e a s y - t o - s e e p o s i t i o n w i t h the SWEEP TIME/DIV c o n t r o l .

P r e s s the x 1 0 MAG s w i t c h i f n e c e s s a r y .

2) S e t t h e H. MODE c o n t r o l t o ALT o r AFTER' D. Adjust t h e B SWEEP TIME/DIV and DELAY

POSITION c o n t r o l s so t h a t the p u l s e edge to be measured i s i n t e n s i t y - m o d u l a t e d .


:
| NOTE 1

If t h e m a g n i f i c a t i o n o f A sweep and B sweep i s h i g h e r than s e v e r a l hundred


m a g n i f i c a t i o n s when the AFTER. D f u n c t i n i s used, d e l a y jitter i s produced. The

delay jitter l e v e l of the o s c i l l o s c o p e i s l e s s than 0.1 V a t 100 m a g i n i f i c a t i o n s .

45
Pulse jitter

Figure 19 Measuring Pulse J i t t e r

3) Set the H. MODE c o n t r o l to B. Measure the j i t t e r w i d t h a t the edge of the j i t t e r .


The j i t t e r i s found by m u l t i p l y i n g t h e j i t t e r w i d t h by t h e r e a d i n g of t h e B SWEEP
TIME/DIV.

Example : The j i t t e r time i s found by m u l t i p l y i n g t h e j i t t e r w i d t h by a r e a d i n g of t h e B


SWEEP TIME/DIV. I f t h e j i t t e r w i d t h i s 1.6 d i v and t h e r e a d i n g of t h e B SWEEP
TIME/DIV i s 0. 2 jus, the j i t t e r time i s found as shown below:
Jitter time = 1.6 ( d i v ) x Q . 2 (jus) = 0. 32 jus

8 — 12 M e a s u r i ng P h a s e Di f f e r e n c e s

With the d u a l - t r a c e f u n c t i o n , phase d i f f e r e n c e can be measured between, f o r example, two


s i n e wave s i g n a l s of i d e n t i c a l frequency.

1) Apply two s i g n a l s to the r e s p e c t i v e INPUT t e r m i n a l s . A d j u s t the VOLTS/DIV and V. VARI

c o n t r o l s so t h a t t h e two s i g n a l s have i d e n t i c a l amplitude.

2) Adjust t h e SWEEP TIME/DIV and H. VARI c o n t r o l s so t h a t one p e r i o d of t h e waveforms


is 8 div.

3) Use the y POSITION c o n t r o l s of both c h a n n e l s t o b r i n g t h e waveforms t o t h e c e n t e r


of the CRT.

4) Measure the h o r i z o n t a l distance between c o r r e s p o n d i n g p o i n t s on both s i g n a l s . There


i s a phase d i f f e r e n c e of 45 degrees f o r e v e r y d i v i s i o n .

46
1 cycle

Reference signal

Comparison
signal Phase difference

Figure 20-1 Measuring Phase Difference

5) I f phase d i f f e r e n c e I s n e g l i g i b l e and the h o r i z o n t a l d i s t a n c e i s s h o r t , p r e s s the x 1 0


MAG s w i t c h i n t h i s c o n d i t i o n .

W i t h t h e x 10 MAG s w i t c h p r e s s e d , t h e phase d i f f e r e n c e i s 4.5 d e g r e e s f o r e v e r y


division.

• Cursor measurement [CS-5230, 5260, 5270]

(1) C a r r y out step 1 of o b s e r v a t i o n w i t h the s c a l e on t h e CRT.

(2) A d j u s t t h e SWEEP TIME/DIV nad H. VARI c o n t r o l s so t h a t one p e r i o d of the waveform i s


5 div.

(3) P r e s s t h e CURSOR MODE s w i t c h t o s e l e c t t h e time and f r e q u e n c y mode. A l i g n the


A R E F c u r s o r and A cursor to the corresponding p o i n t s o f two w a v e f o r m s . Read
the measurement.

Measured value

REF cursor cursor

Figure 2 0 - 2 Measuring Phase Difference (Cursor measurement)

47
8-13 Appl i c a t i o n s o f X-Y O p e r a t i on

When o p e r a t i n g t h e o s c i l l o s c o p e i n t h e X-Y mode, L i s s a j o u s p a t t e r n s can be d i s p l a y e d .

W i t h L i s s a j o u s p a t t e r n s , i t i s p o s s i b l e t o f i n d even t h e s l i g h t e s t phase d i f f e r e n c e or

s i g n a l d i s t o r t i o n and r e l a t i v e frequency proportions.

1) Apply the s i g n a l to be measured to the CHI INPUT t e r m i n a l . Apply the r e f e r e n c e s i g n a l

to the CH2 INPUT t e r m i n a l .

2) Set the H. MODE c o n t r o l to X-Y.

3) Adjust t h e VOLTS/DIV and V. VARI c o n t r o l s of b o t h c h a n n e l s t o o b t a i n an a c c e p t a b l e

display.

4) I n phase d i f f e r e n c e measurement w i t h L i s s a j o u s p a t t e r n s , measured v a l u e s w i 1 1 n o t

change by t u r n i n g the V. VARI c o n t r o l . Use an e a s y - t o - s e e d i s p l a y f o r measurement.

. B
<f> = sm —

or
B
sin</> = —
A

where <>
/ is the phase difference

F i g u r e 21 Measuring Phase D i f f e r e n c e with L i s s a j o u s Patterns

The f o l l o w i n g r e p r e s e n t L i s s a j o u s p a t t e r n s i n d i c a t i n g t h e presence of s i g n a l d i s t o r t i o n or

phase d i f f e r e n c e .

Amplitude distortion, no No amplitude distortion, No amplitude distortion,


phase discrepancy no phase discrepancy 1 8 0 ° phase discrepancy

Amplitude distortion, No amplitude distortion, No amplitude distortion,


phase discrepancy 9 0 ° phase discrepancy phase discrepancy

Figure 22 Typical L i s s a j o u s Patterns

48
The f o l l o w i n g r e p r e s e n t L i s s a j o u s p a t t e r n s when input frequency p r o p o r t i o n s a r e changed

0° 45° 90° 135° 180° Frequency Proportion (CH1[Y]:CH2[X1)

1 :1

1 :2

1 :3

Figure 23 L i s s a j o u s P a t t e r n s when Frequency Proportions Are Changed

49
9. FUSE REPLACEMENT, LINE VOLTAGE SELECTION

Before conducting the f o l l o w i n g operation, t u r n the power o f f and unplug t h e power cord

from the w a l l socket.

Fuse Replacement

If the fuse blows, the osci1loscope w i 1 1 not o p e r a t e . Locate the cause. I f the

oscilloscope i s not d e f e c t i v e , remove t h e cap from t h e f u s e h o l d e r on the r e a r p a n e l

with a standard type s c r e w d r i v e r , then r e p l a c e the f u s e w i t h a new f u s e . (See Figure

24.)

100 or 120 V : 1 A (Time l a g )

220, 230 or 240 V : 630 mA (Time l a g )


| NOTE 1

If any of the f o l l o w i n g cases apply, please contact your dealer, or our

distributor.

• I t i s not c l e a r why the fuse has blown, or the o s c i l l o s c o p e seems to be d e f e c t i v e .

• The o s c i l l o s c o p e uses a time l a g type f u s e . I f a time l a g type f u s e of the above-

shown r a t i n g i s not a v a i l a b l e .

Changing Voltage S e l e c t ion


Remove t h e f u s e h o l d e r and v o l t a g e s e l e c t o r from t h e r e a r p a n e l w i t h a s t a n d a r d type
screwdriver. F i t the 1ine voltage s e l e c t o r to match the s u p p l y voltage rat ing
( T mark). I f 100 or 120 V i s to be changed i n t o 220, 230 or 240 V, the 1 A f u s e should
be r e p l a c e d w i t h a 630 mA f u s e . (See F i g u r e 24. )

Fuse Holder

F i g u r e 24 Changing Voltage Requi rements

50
10. OPTIONAL EQUIPMENT

The o s c i l l o s c o p e has an o p t i o n a l a c c e s s o r y bag. T h i s bag a t t a c h e s to the o s c i l l o s c o p e and


p r o v i d e s a s t o r a g e space f o r probes, e t c .

How to Attach Accessory Bag (MC-78)

1) Separate t h e a c c e s s o r y bag from the mounting p l a t e .

2) L i n e up t h e f o u r h o l e s on the r i g h t s i d e (viewed from t h e f r o n t ) of the o s c i l l o s c o p e

with the corresponding f o u r h o l e s i n t h e mounting p l a t e . F i x t h e mounting p l a t e to

the o s c i l l o s c o p e c a s e w i t h t h e f o u r n y l o n r i v e t s and w a s h e r s p r o v i d e d . Be s u r e t h a t

the mount i n g p l a t e i s i n t h e v e r t i c a l p o s i t i o n a s shown i n F i g u r e 25. I n s e r t the

grommet i n t o the h o l e s f i r s t , then p r e s s the p l u n g e r s ,

(When removing, j u s t p r y the p l u n g e r s out w i t h a s t a n d a r d type s c r e w d r i v e r . )

3) A t t a c h the bag to the mounting p l a t e w i t h the hooks.

Mounting plate

Washer
Grommet
Nylon rivet
Plunger

Accessory bag

Figure 25 Attaching Accessory Bag

51
CORRECTION
T h i s c o r r e c t i o n i s f o r : C S - 5 2 0 0 SERIES OSCILLOSCOPE
T i t l e : CS-5200 SERIES OSCILLOSCOPE INSTRUCTION MANUAL
Manual P a r t No. : B63-0219-00
Dated : 12/93

Page. 1 7 5. PANEL EXPLANATION

© COUPLING switch

[ ERROR] TV LINE : ••••••••••••••••••••••••••••••••••••••••••••••• from the composite video

[CORRECT] TV LINE : Horizontal synchronization pulses are e x t r a c t e d from the composite video
s i g n a l s and input t o the t r i g g e r c i r c u i t .

Page. 5 0 9. FUSE REPLACEMENT,. LINE VOLTAGE SELECTION

Fuse Replacement

[ ERROR] 100 or 120 V : 1 A (Time l a g )


220, 230 or 240 V : 630 mA (Time l a g )

[CORRECT] 100 or 120 V , : 1 A


220, 230 or 240 V : 630 mA

Note
[ERROR] I f any of the •
distributor.
• I t i s not c l e a r why the fuse
* The o s c i l l o s c o p e uses a time l a g type fuse. I f a time l a g type fuse of the above-
shown r a t i n g i s not a v a i l a b l e .

Note
[CORRECT] I f any of the •
distributor.
• I t i s not c l e a r why the fuse
* I f a fuse of the above shown r a t i n g i s not a v a i l a b l e .
A product of

KENWOOD CORPORATION
2-5, 1 -chortle Shibuya, Shibuya-ku, Tokyo 150, Japan