Single Patient Dialysis System TR-3000 MIRAI User Manual
1. Specifications
1.1 Overview
Single Patient Dialysis System“TR-3000 MIRAI” is the dialysis type artificial
kidney unit to be used to eliminate harmful substance from blood with the dialyzer.
The system is equipped with the capabilities of mixing rich flowing liquid with
clean water to prepare dialysate of predetermined composition, control of its
conductivity, as well as heating the dialysate to temperatures close to a patient’s
temperatures, control of its temperatures, control of dialysate flow flowing into a
dialysis unit, and monitoring internal pressure of the blood circuit, dialysate circuit
internal pressure, etc.
In addition, a user can selectively use the following two treatments, depending on
a patient’s constitution and conditions: acetate dialysis and bicarbonate dialysis.
Through chamber method, this system is capable of automatically controlling UF
amount from a patient in accurate and easy manner, by creating a sealed circuit with no
inlet/outlet other than a membrane on the dialyzer, without lowering dialysis efficiency.
1-2
1. Specifications
1-3
Single Patient Dialysis System TR-3000 MIRAI User Manual
1-4
1. Specifications
VP alarm Venous pressure goes out of upper/lower limits of a set alarm range.
(For setting of alarms, the device is equipped with one-touch set/release
capability.)
DP alarm Dialysate pressure goes out of upper/lower limits of a set alarm range.
(For setting of alarms, the device is equipped with one-touch set/release
capability.)
Dialysate temperature Dialysate temperature exceeds the upper limit alarm value (41°C).
alarm (41°C)
Dialysate temperature A device to prevent overheating is actuated.
alarm (overheating)
Dialysate temperature Dialysate temperature drops below the lower limit alarm value.
alarm (lower limit)
Temp. (Con) TH break Broken temperature thermistor is detected.
Temp. (Ind) TH break Broken temperature thermistor is detected.
Chamber alarm Abnormal condition occurs in changeover time of the chamber.
Dialysate supply alarm Supply water is stopped or has its flow rate considerably reduced.
Blood pump cover open Cover of the blood pump is open.
Blood pump alarm 1 The revolving speed of the blood pump is considerably different from the
setting.
Blood pump alarm 2 The direction of blood pump rotation is reversed.
UF pump alarm The revolving speed of the blood pump is considerably different from the
setting.
Blood leak alarm Blood the amount of which exceeds a defined value has leaked into the
dialysate.
Air bubble alarm Air bubble contained in the blood circuit is detected.
Closed circuit leak Result of on-line self-diagnosis exceeds a judged value.
Closed circuit alarm Mass of air mixes into the closed circuit.
Disinfection preparation “Rinse” or “Disinfect” switch is pressed with no bypass connector set in
alarm the bypass hanger.
Reset Clamp SW ON Clamp has not been released after recovery from abnormal conditions.
Clamp malfunction Clamp mistakenly blocks blood circuit.
Syringe completion Internal pressure inside the syringe pump has abnormally increased.
Syringe pump has reached a stroke end.
S/N blood taking Blood taking operation times out while the syringe needle is still running.
S/N blood returning Blood returning operation times out while the syringe needle is still
running.
Clogged flow The float of flow meter remains down for 2 consecutive seconds or longer.
Dialysate conductivity Dialysate conductivity goes out of the upper/lower limits of a set alarm
alarm range.
Conc. Pump alarm Concentrate pump does not rotate.
Bypass alarm Bypass switching valve behaves abnormally.
1-5
Single Patient Dialysis System TR-3000 MIRAI User Manual
Disinfection Disinfection has not been completed properly due to abnormal conditions
while rinsing or disinfection.
Blood excessive UF rate is larger than a defined value, against blood flow.
concentration
Goal time of timer 1 Set time of timer 1 has been reached.
Goal time of timer 2 Set time of timer 2 has been reached.
1.5.2 Warnings
1-6
1. Specifications
1.6 Options
NaCl infusion function Dialysate conductivity can be changed as necessary, by infusing set
amount of NaCl solution of 10 [%].
Automatic blood Blood pressure can be measured automatically at set intervals.
pressure measurement Blood pressure can also be measured manually, independent of set
function time intervals.
Data communications Use of RS-232C port enables communications with a host
function computer.
Hot water rinse function When rinsing, supply water is heated so that any deposits in the
water system circuit that cannot be removed by water at ordinary
temperatures can be washed away
Automatic priming Addition of a blood clamp unit enables used of automatic priming
function function.
Concentrate This is to be used when concentrate of dialysate is supplied from
centralized piping one container to more than one private dialyzer.
bracket
Nurse call switch This is a nurse call switch.
Compatibility with The device is for checking various settings of a dialyzer. Possible
input/output device to change data, etc.
(Zaurus)
Holder for endotoxin This is a holder in which an endotoxin removal filter can be placed.
filter
Priming funnel This is a funnel to receive drainage at the time of priming.
Pole dedicated to This is a pole dedicated to placement of infusion pump to be used
supplementary liquid for supplementary liquid.
1-7