The document is a calibration certificate issued by SONEL S.A. for an electrical installation meter, detailing the calibration methods, environmental conditions, and results of various electrical measurements. It confirms the traceability of measurements to the International System of Units and includes specific measurement results with associated uncertainties and permissible errors. The certificate is accredited by the Polish Centre for Accreditation, ensuring compliance with international standards.

Uploaded by

Max Eddisson
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
23 views7 pages

2023-20-09 c50879 Swiadectwo en Akr

The document is a calibration certificate issued by SONEL S.A. for an electrical installation meter, detailing the calibration methods, environmental conditions, and results of various electrical measurements. It confirms the traceability of measurements to the International System of Units and includes specific measurement results with associated uncertainties and permissible errors. The certificate is accredited by the Polish Centre for Accreditation, ensuring compliance with international standards.

Uploaded by

Max Eddisson
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 7

SONEL S.A.

Testing and Calibration Laboratory


Wokulskiego 11
58-100 ŚWIDNICA
phone: +48 74 85 838 00, e-mail: [email protected]

Calibration laboratory accredited by Polish Centre for Accreditation, a signatory to EA MLA


and ILAC MRA that include recognition of calibration certificates.
Accreditation No AP 173

CALIBRATION CERTIFICATE
Date of issue: 20 September 2023 Certificate No: Page 1/7

OBJECT OF Meter for electrical installation parameters


CALIBRATION type: MPI-530-IT, serial No: , manufacturer: SONEL S.A.

APPLICANT

CALIBRATION According to IW01 ''Calibration of digital meters'', issue 2.3, date 29.08.2022, IW07
METHOD ''Calibration of Short-ciruit meter'', issue 1.3, date 29.08.2022, IW09 ''Calibration of
RCD meters'', issue 2.1, date 29.08.2022.

ENVIRONMENTAL Ambient temperature: (23,3 ÷ 24,3) °C


CONDITIONS Relative humidity: (43,7 ÷ 45,5) %

DATE AND PLACE 20 September 2023


OF CALIBRATION Testing and Calibration Laboratory. Wokulskiego 11, 58 - 100 ŚWIDNICA

TRACEABILITY This certificate is issued under the agreement EA MLA in the field of calibration and
provides traceability of measurement results to the International System of Units (SI).

CALIBRATION The results of calibration have been presented on page 2/7 to 7/7 including
RESULTS uncertainty of measurement. Points outside the range of accreditation are marked #.
Presented results relate only to the items calibrated.
UNCERTAINTY OF Uncertainty of measurement has been evaluated in compliance with EA-4/02
MEASUREMENTS M:2022.The expanded uncertainty as signed corresponds to a coverage probability
of 95 % and the coverage factor k = 2.
CONFORMITY
WITH
REQUIREMENTS

SONEL S.A
Testing and Calibration Laboratory
Metrology Specialist

This certificate without approval Laboratory shall not be reproduced except in full.
CALIBRATION CERTIFICATE issued by ACCREDITED LABORATORY No AP 173

Date of issue 20 September 2023 Certificate No: Page 2/7

CALIBRATION Calibration results are the following:


RESULTS

1. AC 50 Hz voltage.
Max
Reference value of Measured value of Measurement
Range Measurement error permiss.
voltage voltage uncertainty
error
V V V V V V
299,9 230,00 229,90 -0,10 0,18 5,00
500 400,0 400,0 0,0 0,7 10,0

2. AC 50 Hz current - C3 clamps.
Max.
Reference value of Measured value of Measurement
Range Measurement error permiss.
current current uncertainty
error
mA mA mA mA mA mA
99,9 90,00 89,90 -0,10 0,14 18,80
999 900,0 901,0 1,0 1,4 78,0
A A A A A A
9,99 9,000 9,000 0,000 0,014 0,590
99,9 90,00 91,00 1,00 0,16 5,45
999 900,0 907,0 7,0 1,4 52,7

3. AC 50 Hz current - F3 clamps.
Max.
Reference value of Measured value of Measurement
Range Measurement error permiss.
current current uncertainty
error
A A A A A A
9,99 9,00 8,96 -0,04 0,11 3,11
99,9 90,00 89,90 -0,10 0,16 4,10
999 900,0 897,0 -3,0 1,4 14,0
kA kA kA kA kA kA
3,00 1,900 1,890 -0,010 0,006 0,042

4. Fault loop parameters (Z L-PE).


Max.
Measurement
Parameter Reference value Measured value Measurement error permiss.
uncertainty
error
- Ω Ω Ω Ω Ω
Z 0,124 0,148 0,024 0,006 0,036
R 0,109 0,140 0,031 0,006 0,056
# X 0,058 0,049 -0,009 0,007 0,056
Z 2,110 2,131 0,021 0,006 0,136
R 2,109 2,130 0,021 0,006 0,156
# X 0,058 0,061 0,003 0,011 0,156

Authorized by:
CALIBRATION CERTIFICATE issued by ACCREDITED LABORATORY No AP 173

Date of issue 20 September 2023 Certificate No: Page 3/7

4. Fault loop parameters (Z L-PE) - cont.


Max.
Measurement
Parameter Reference value Measured value Measurement error permiss.
uncertainty
error
- Ω Ω Ω Ω Ω
Z 19,109 19,129 0,020 0,013 0,985
R 19,109 19,129 0,020 0,013 1,005
# X 0,058 0,118 0,060 0,007 1,005
Z 190,11 188,92 -1,19 0,12 9,81
R 190,11 - - 0,12 -
# X 0,06 - - 0,01 -
Z 0,804 0,825 0,021 0,006 0,070
R 0,697 0,737 0,040 0,006 0,090
# X 0,402 0,370 -0,032 0,006 0,090
Z 2,360 2,374 0,014 0,007 0,148
R 2,239 2,262 0,023 0,007 0,168
# X 0,746 0,722 -0,024 0,007 0,168

5. Fault loop parameters (Z L-PE RCD).


Max.
Measurement
Parameter Reference value Measured value Measurement error permiss.
uncertainty
error
- Ω Ω Ω Ω Ω
Z 0,124 0,110 -0,014 0,013 0,107
R 0,109 0,095 -0,014 0,009 0,107
# X 0,058 0,055 -0,003 0,009 0,107
Z 2,110 2,105 -0,005 0,011 0,227
R 2,109 2,095 -0,014 0,009 0,227
# X 0,058 0,060 0,002 0,011 0,227
Z 19,109 19,125 0,016 0,016 1,247
R 19,109 19,125 0,016 0,014 1,247
# X 0,058 0,055 -0,003 0,009 1,247
Z 190,11 190,45 0,34 0,15 11,91
R 190,11 - - 0,13 -
# X 0,06 - - 0,06 -
Z 0,804 0,790 -0,014 0,011 0,148 0
R 0,697 0,695 -0,002 0,009 0,148 0
# X 0,402 0,370 -0,032 0,008 0,148 0
Z 2,360 2,340 -0,020 0,011 0,242
R 2,239 2,225 -0,014 0,009 0,242
# X 0,746 0,720 -0,026 0,011 0,242

Authorized by:
CALIBRATION CERTIFICATE issued by ACCREDITED LABORATORY No AP 173

Date of issue 20 September 2023 Certificate No: Page 4/7

6. Time interval of RCD response.

Reference value of Measured value of Measurement Max. permissible


Measurement error
time interval time interval uncertainty error

ms ms ms ms ms
10,0 10,0 0,0 1,1 2,2
40,0 40,0 0,0 1,1 2,8
490,0 490,0 0,0 8,2 11,8

7. AC current (RCD test current).

Nominal quantity Measured value of Measurement Nominal indication


Shape / multiplier value current uncertainty interval

mA mA mA mA mA
+ SIN / x 0,5 15 14,50 0,09 13,80 15,00
+ SIN / x 1 30 31,65 0,14 30,00 32,40

8. AC 50 Hz voltage (touch voltage).

Reference value of Measured value of Measurement Nominal indication


IΔn
voltage voltage uncertainty interval

mA V V V V V
24,903 27,000 0,063 24,903 28,639
30
42,00 45,40 0,11 42,00 48,30

9. AC 50 Hz resistance (resistance-to-earth for TT).

Reference value of Measured value of Measurement Nominal indication


IΔn
resistance resistance uncertainty interval

mA kΩ kΩ kΩ kΩ kΩ
0,8301 0,8800 0,0058 0,7801 0,9631
30
1,4001 1,4800 0,0059 1,3501 1,5901
0
10. DC resistance (insulation resistance measurement). 0
Test voltage 50 V. 0

Reference value of Measured value of Measurement Max. permissible


Measurement error
resistance resistance uncertainty error

kΩ kΩ kΩ kΩ kΩ
50,0 50,0 0,0 1,1 9,5
MΩ MΩ MΩ MΩ MΩ
4,000 4,010 0,010 0,070 0,200 0
40,00 40,00 0,00 0,70 2,00 0
230,0 231,0 1,0 4,1 14,9 0
Authorized by:
CALIBRATION CERTIFICATE issued by ACCREDITED LABORATORY No AP 173

Date of issue 20 September 2023 Certificate No: Page 5/7

Test voltage 1000 V.

Reference value of Measured value of Measurement Max. permissible


Measurement error
resistance resistance uncertainty error

GΩ GΩ GΩ GΩ GΩ
9,50 9,37 -0,13 0,17 0,44

11. DC resistance (continuity measurement with 200 mA current).


Max.
Reference value of Measured value of Measurement
Range Measurement error permiss.
resistance resistance uncertainty
error
Ω Ω Ω Ω Ω Ω
2,000 2,000 0,000 0,006 0,070
19,99
18,000 17,980 -0,020 0,012 0,390
22,00 22,00 0,00 0,06 0,74
199,9
180,00 179,60 -0,40 0,12 3,90
220,0 219,0 -1,0 0,6 7,4
400
360,0 359,0 -1,0 0,6 10,2 0

12. DC resistance (low voltage resistance measurement).


Max.
Reference value of Measured value of Measurement
Range Measurement error permiss.
resistance resistance uncertainty
error
Ω Ω Ω Ω Ω Ω
10,00 9,90 -0,10 0,06 0,60 0
199,9
180,00 179,40 -0,60 0,12 5,70 0
1999,9 1900,0 1894,0 -6,0 1,3 60,0 0

13. AC resistance (resistance-to-earth)


4P 50 V 50 Hz
Nominal value
Max.
resistance of aux. Reference value of Measured value of Meas. Meas.
Range permiss.
electrodes resistance resistance error uncert.
error
RH RS
Ω Ω Ω Ω Ω Ω Ω Ω
98 98 0,110 0,130 0,020 0,006 0,042
9,99 98 98 0,510 0,530 0,020 0,007 0,050
98 98 9,010 9,010 0,000 0,014 0,220
99,9 98 98 90,01 90,10 0,09 0,14 2,10
999 98 98 900,0 903,0 3,0 1,4 21,0
0

Authorized by:
CALIBRATION CERTIFICATE issued by ACCREDITED LABORATORY No AP 173

Date of issue 20 September 2023 Certificate No: Page 6/7

4P 25 V 50 Hz
Nominal value
Max.
resistance of aux. Reference value of Measured value of Meas. Meas.
Range permiss.
electrodes resistance resistance error uncert.
error
RH RS
Ω Ω Ω Ω Ω Ω Ω Ω
9,99 98 98 0,510 0,530 0,020 0,007 0,050

3P 50 V 50 Hz
Nominal value
Max.
resistance of aux. Reference value of Measured value of Meas. Meas.
Range permiss.
electrodes resistance resistance error uncert.
error
RH RS
Ω Ω Ω Ω Ω Ω Ω Ω
98 98 0,516 0,500 -0,016 0,007 0,050
9,99
98 98 9,016 8,990 -0,026 0,014 0,220
99,9 98 98 90,02 90,00 -0,02 0,14 2,10
999 98 98 900,0 902,0 2,0 1,4 21,0

3P 25 V 50 Hz
Nominal value
Max.
resistance of aux. Reference value of Measured value of Meas. Meas.
Range permiss.
electrodes resistance resistance error uncert.
error
RH RS
Ω Ω Ω Ω Ω Ω Ω Ω
98 98 0,516 0,500 -0,016 0,007 0,050
9,99
98 98 5,016 5,020 0,004 0,009 0,140

14. AC resistance (selective measurement of earthing with clamps)


3P+C 50 V 50 Hz
Nominal value
Max.
resistance of aux. Reference value of Measured value of Meas. Meas.
Range permiss.
electrodes resistance resistance error uncert.
error
RH RS
Ω Ω Ω Ω Ω Ω Ω Ω
98 98 0,516 0,500 -0,016 0,007 0,081
9,99
98 98 9,016 8,980 -0,036 0,014 0,761
99,9 98 98 90,02 89,90 -0,12 0,14 7,60
999 98 98 900,0 900,0 0,0 1,4 21,0

Authorized by:
CALIBRATION CERTIFICATE issued by ACCREDITED LABORATORY No AP 173

Date of issue 20 September 2023 Certificate No: Page 7/7

14. AC resistance (selective measurement of earthing with clamps) - cont.


3P+C 25 V 50 Hz
Nominal value
Max.
resistance of aux. Reference value of Measured value of Meas. Meas.
Range permiss.
electrodes resistance resistance error uncert.
error
RH RS
Ω Ω Ω Ω Ω Ω Ω Ω
9,99 98 98 0,516 0,510 -0,006 0,007 0,081

15. AC resistance (selective measurement of earthing with double clamps)


50 Hz
Max.
Reference value of Measured value of Measurement
Range Measurement error permiss.
resistance resistance uncertainty
error
Ω Ω Ω Ω Ω Ω
9,99 9,016 8,940 -0,076 0,014 0,942
19,9 18,02 17,80 -0,22 0,07 2,20
99,9 90,02 88,20 -1,82 0,14 18,40

Measurements were made with clamps C-3 nr 119572 UJS, N-1 nr 106719 XCS delivered with the meter.

Authorized by: