Physical Dimensions of Solid Plastics Specimens: Standard Test Methods For
Physical Dimensions of Solid Plastics Specimens: Standard Test Methods For
for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: D5947 − 24
Test Method Apparatus Elastic Modulus RangeA Diameter of Presser Foot Pressure on Specimen,
MPa or Spindle, mm Approximate, kPa
A1 A from >35 to <275 6.4 40 to 180
A2 A from >276 to <700 6.4 40 to 300
A3 A >701 6.4 40 to 900
B B 6.4 unknown
C C 6.4 to 12.7 5 to 900
D D 6.4 to 12.7 5 to 900
H C 6.4 30
A
Determined by Test Method D638 or Test Method D790.
force necessary to just prevent a change in the indicator reading 8.2.1.1 Close the micrometer on the screw-thread-pitch wire
must be more than the minimum permissible force specified for or plug gauge according to the calibration procedure of 8.6.2 or
a specimen. 8.6.3, as appropriate;
8.2.1.2 Observe and record the thickness indicated;
7. Test Specimens 8.2.1.3 Move the screw-thread-pitch wire or plug gauge to a
7.1 The test specimens shall be prepared from plastics different position between the presser foot and anvil, and repeat
materials in sheet, plate, or molded shapes that have been cut 8.2.1.1 and 8.2.1.2; and
to the required dimensions or molded to the desired finished 8.2.1.4 If the difference between any pair of readings is
dimensions for the particular test. greater than 2.5 µm, the surfaces are not parallel.
7.2 Prepare and condition each specimen to equilibrium in 8.3 Lacking a detailed procedure supplied by the instrument
accordance with Practice D618 unless otherwise specified by manufacturer, confirm the requirements for parallelism of
the relevant ASTM material specification. dial-type micrometers given in 6.3.1.2 by placing a hardened
steel ball (such as that used in a ball bearing) of suitable
7.3 For each specimen, take precautions to prevent damage
diameter between the presser foot and anvil. Mount the ball in
or contamination that might affect the measurements adversely.
a fork-shaped holder to allow it to be moved conveniently from
7.4 Unless otherwise specified, make all dimension mea- one location to another between the presser foot and anvil. The
surements at the standard laboratory atmosphere in accordance balls used commercially in ball bearings are almost perfect
with Practice D618. spheres having diameters constant within 0.2 µm.
8. Calibration (General Considerations for Care and Use NOTE 2—Exercise care with this procedure. Calculations using the
equations given in X1.3.2 show that the use of a 680 g mass weight on a
of Each of the Various Pieces of Apparatus for ball between the hardened surfaces of the presser foot and anvil can result
Dimensional Measurements) in dimples in the anvil or presser foot surfaces caused by exceeding the
8.1 Good testing practices require clean anvil and presser yield stress of the surfaces.
foot surfaces for any micrometer instrument. Prior to calibra- 8.3.1 Observe and record the diameter as measured by the
tion or dimensional measurements, clean such surfaces by micrometer at one location.
inserting a piece of smooth, clean bond paper between the anvil 8.3.2 Move the ball to another location and repeat the
and presser foot and slowly moving the bond paper between measurement.
the surfaces. Check the zero setting frequently during measure- 8.3.3 If the difference between any pair of readings is
ments. Failure to repeat the zero setting may be evidence of dirt greater than 2.5 µm, the surfaces are not parallel.
on the surfaces. Avoid pulling any edge of the bond paper 8.4 Lacking a detailed procedure supplied by the instrument
between the surfaces to reduce the probability of depositing manufacturer, confirm the flatness of the anvil and the spindle
any lint particles on the surfaces. Periodic verifications with the surface of a micrometer or dial gauge by the use of an optical
gauge blocks shall be conducted. flat that has clean surfaces. Surfaces shall be flat within 1 µm.
8.2 The parallelism requirements for machinists’ microm- 8.4.1 After cleaning the micrometer surfaces (see 8.1), place
eters demand that observed differences of readings on a pair of the optical flat on the anvil and close the presser foot as
screw-thread-pitch wires or a pair of standard 6.4-mm nominal described in 8.6.2, 8.6.3, 8.6.4, or 8.6.5, as appropriate.
diameter plug gauges be not greater than 2.5 µm. Spring-wire 8.4.2 When illuminated by diffused daylight, interference
stock or music-wire of known diameter are suitable substitutes. bands are formed between the surfaces of the flat and those of
The wire (or the plug gauge) has a diameter dimension that is the micrometer. The shape, location, and number of these
known to be within 61 µm. Diameter dimensions may vary by bands indicate the deviation from flatness in increments of half
an amount approximately equal to the axial movement of the the average wavelengths of white light, which is taken as 0.25
spindle when the wire (or the plug gauge) is rotated through µm.
180°. 8.4.2.1 A flat surface forms straight parallel fringes at equal
8.2.1 Lacking a detailed procedure supplied by the instru- intervals.
ment manufacturer, confirm the parallelism requirements of 8.4.2.2 A grooved surface forms straight parallel fringes at
machinist’s micrometers using the following procedure: unequal intervals.
ANNEX
(Mandatory Information)
APPENDIX
(Nonmandatory Information)
SUMMARY OF CHANGES
Committee D20 has identified the location of selected changes to this standard since the last issue, D5947 - 18,
that may impact the use of this standard. (April 15, 2024)
(1) Added manufacturer’s recommendations to 6.1.4 (4) Changed language in 8.6.1 and A.1.1 for when and how
(2) Changed 6.1.6 to best practice. calibration is required.
(3) Moved Note 2 to 8.1.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or [email protected] (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/