Printopp
Printopp
CHAPTER
LEARNING OUTCOMES
Measurement of colors is the historical basis of quantifying the concentrations of the unknown analytes in the
clinical chemistry laboratory. Portable spectrophotometers are utilized to determine the transmittance which is then
converted to absorbance following Beer's Law.
Currently, fully automated biochemistry analyzers have replaced manual photometric readings to meet the desired
turnaround time of each assay. The principles and distinguishing characteristics of each analytical method are
presented in this chapter.
I, COLORIMETRY
a. Spectrophotometric measurement - is the measurement of light intensity in a narrower wavelength.
b. Photometric measurement - is the measurement of light intensity using a specific wavelength.
40 REVIEW HANDBOOK IN CLINICAL CHEMIS-
The primary analytical utility of spectrophotometry or filter photometry is the isolation of discreet portin
of the spectrum for purposes of measurement.
A. Spectrophotometry
It involves measurement of the light transmitted by a solution to determine the concentration of the
ligh
absorbing substances in the solution.
A spectrometer is a device that measures the wavelengths of light or the intensity of radiation.
Single-beam Spectrophotometer
It is the simplest type of an absorption spectrophotometer.
It is designed to make one measurement at a time at one specified wavelength.
The maximum absorption of the analyte must be known in advance when a single-beam instrument is used.
Double-beam Spectrophotometer
It is an instrument that splits the monochromatic light into two components: one beam passes through
Light
Sample PM tube
(1) Stable source of radiant energy, (2) filter that source Monochromator
isolates a specific region of the electromagnetic spectrum, A/D Display
(3) sample holder, (4) radiation detector, (5) signal
processor, and (6) readout device
It minimizes unwanted or stray light and prevents the entrance of scattered light into the monochromator
system.
Stray light refers to any wavelengths outside the band transmitted by the monochromator; it does not
originate from the polychromatic light source; it causes absorbance error.
maximum absorbance that a spectrophotometer can achieve.
Stray light limits the
Straylight is the most common cause of loss of linearity at high-analyte concentration.0p 2 7291901
3. Monochromator
absorbence
It isolates specific or individual wavelength of light.
Kinds of Monochromators
Prisms
These are simple, least expensive, not precise, but useful monochromators.
These are made by placing semi-transparent silver films on both sides of a dielectric such as magnesium
fluoride.
produce monochromatic light based on the principle of constructive interference of waves-light
Filters
the second surface.
waves enter one side of the filter and are reflected at
have a low transmittance of the selected wavelength.
Filters usually pass : wide band of radiant energy and
4, Exit Slit
of the spectrum to reach
It controls the width of the light beam (bandpass); allows only a narrow fraction
the sample cuvette.
Bandpass is the total range of wavelengths transmitted.
1/5 the natural bandpass of the
Accurate absorbance measurement requires a bandpass less than
spectrophotometer.
42 REVIEW HANDBOOK IN CLINICAL CHEMISTRY
The degree of wavelength isolation is a function of the type of device used and the width of entrance and exit
slits.
narrower the bandpass
the bandpass, that is, the
Spectral purity of the spectrophotometer is reflected by
the greater the resolution.
Kinds of Cuvettes
Alumina silica glass - most commonly used (available in 350 nm to 2000 nm)
Quartz/plastic = used for measurement of solution requiring visible and ultraviolet spectra
Borosilicate glass
Soft glass
Notes to Remember
Cuvettes with scratches on their optical surface scatter light and should be discarded.
Silica cuvettes transmit light effectively at wavelengths 2220nm.
alkali slowly dissolves glass
Alkaline solutions should not be left standing in cuvettes for prolonged periods because
producing etching.
The path length of cuvettes is 1 cm, although much smaller path lengths are used in automated systems.
10 cm, increasing the absorbance
To increase sensitivity, some cuvettes are designed to have path lengths equivalent to
for a given solution by factor of 10.
6. Photodetector
It detects and converts transmitted light into photoelectric energy.
It detects the amount of light that passes through the sample cuvette.
Kinds of Photodetectors
Barrier Layer Cell/Photocell/Photovoltaic Cell
It is the simplest detector; least expensive; temperature-sensitive.
It is used in filter photometers with a wide bandpass.
It is the basic photodetector for detecting and measuring radiation in the visible region.
It iscomposed of selenium on a plate of iron covered with transparent layer of silver.
It requires external voltage source but utilizes internal electron transfer for current production low
internal resistance.
This cell typically has a maximum sensitivity at about 550 nm and the response falls off to about 10% of the
maximum at 350 and 750 nm.
Phototube
It displays the
output of the detection system.
Example: Galvanometer/ammeter/light-emitting diode (LED) display
Quality Assurance Monitoring of Spectrophotometers
It includes monitoring the
wavelength/photometric accuracy, absorbance check, linearity, and detects straylight.
Wavelength accuracy is the actual wavelength of light that passes through the monochromator.
Didymium or holmium oxide filter is used to check
wavelength accuracy (wavelength quality assurance
calibration).
Didymium glass has an absorption peak around 600 nm.
Holmium oxide has multiple absorption with a sharp peak at 360 nm.
Neutral density filters and dichromate solution verify linearity.
An absorbance check is performed using glass filters or solutions that have known absorbance values for a
specific wavelength - the operator simply measures the absorbance of each solution at a specified wavelength
and compares the results with the stated values.
Absorption spectroscopy is preferred for solutions with absorbance values of less than 2.0.
Absorbance values higher than 2.0 may yield unreliable results, and thereby cause a deviation from Beer's law,
resulting in bending of the linear plot
Beer's Law
It states that the concentration of the unknown substance is directly proportional to the absorbed
light (absorbance or optical density) and inversely proportional to the amount of transmitted light
(% Transmittance).
It mathematically establishes the relationship between concentration and absorbance.
Absorbance (A)
It is the amount of light absorbed.
It is proportional to the inverse log of transmittance.
It is mathematically derived from % T.
A = abc = 2
where:
A= Absorbance
a = Molar absorptivity; absorptivity of the compound under standard conditions
b = Length of light through the solution
c = Concentration of absorbing molecules/solution
Au
Unknown solution: =
AS
where:
Orange Green-blue
601-650
NOTE: If a solution absorbs light of a certain color (2nd column), the observed color of the solution is the complementary color.
Notes to Remember
The amount of light absorbed at a particular wavelength depends on molecular and ion types present and may vary
with concentration, ph, or temperature.
The light path must be kept constant to have absorbance proportional to concentration.
Deviations from Beer's law may be caused by changes in instrument functions or chemical reactions. astile
Instrument deviation is commonly a result of the finite band pass of the filter or monochromator. chordal ddail
Changes in solution pH, ionic strength, and temperature can alter Beer's law.
Turbidity readings on a spectrophotometer are greater in the blue region than in the red region of the spectrum.
Blanking Technique A
It means the blank contains serum but without the reagent to complete the assay.
Reagent blank corrects absorbance caused by the color of the
reagents the absorbance of reagents is
Light
source Readout
Monochromator
Burner
head
Oxidant
Mixing baffles
Aspirating air
Sample
Principle: The unknown sample is made to react with a known solution in the presence of an indicator.
Examples: Schales and Schales method (Chloride test)
EDTA Titration method (Calcium test)
III, TURBIDIMETRY
Principle: It determines the amount of light blocked (reduction of light)
by a particulate matter in a turbid
solution.
spectrophotometers.
to detect bacterial growth in broth cultures, antimicrobial test
Uses: Protein measurements (CSF and urine);
Basic Components: Light source (mercury-arc lamp, a tungsten-filament lamp, light emitting diode, and a laser),
collimator, monochromator, sample cuvette, stray light trap, and photodetector
etector, spectrophotometer
FIGURE 6. Nephelometer
turbidimetry
versus Spectrophotometer.
Light
source Source: Bishop et al., 2010
Detector, nephelometer
forward light scatter
Detector, nephelometer
90° light scatter
V. FLUOROMETRY
Primary filter
Principle: It determines the amount of light emitted by a Source
Sample
molecule after excitation by electromagnetic radiation. holder
It measures the amount of light intensity present over a zero
background.
It uses two monochromators (either filters, prisms, or Secondary
filter
gratings) - wavelength that is best absorbed by the
the
solution to be measured is selected by the primary filter or
excitation monochromator; the incident light is prevented Detector
To avoid potential interference from the excitation signal, FIGURE 7. Filter Fluorometer.
fluorescence measurements detect the emitted light at a Source: Bishop et al., 2010
Fluorescence is the light emission from a single excited state, while phosphorescence is the light emission from
an excited triplet state.
Use: For measurement of porphyrins, magnesium, calcium, and catecholamines
Basic components: Light source (mercury arc or xenon lamp), primary monochromator, secondary
monochromator, cuvette, and photodetector (photomultiplier tube or phototube)
VI. CHEMILUMINESCENCE
Principle: The chemical reaction yields an electronically excited compound that emits light as it returns to its
ground state or transfers its energy to another compound which then produces an emission.
In this method, no excitation radiation is required and no
monochromators are needed because the
chemiluminescence arises from one sample.
It quantifies sample analytes in a wide range of wavelengths not observed in the visible spectrum.
The measurement of light signal is against a dark background since no excitation light is required.
The emission of light is produced from a chemical or electrochemical reaction
and not from absorption of
electromagnetic energy-the light that is emitted is detected at a right angle to the incident light to eliminate
any excitation interference signal.
It involves the oxidation of anorganic compound (dioxetane) by an oxidant (hydrogen peroxide), and the
reactions may occur in the presence of catalysts, such as enzymes, metal ions, and hemin-it
produces
chemiluminescence reaction as it returns to the singlet state.
A typical signal from a chemiluminescent compound rises rapidly with time and reaches a maximum limit when
reagent and analyte are completely mixed.
This method is best when
differentiating two compounds having excitation reaction at the same wavelength but
emit at different wavelengths.
ANALYTICAL METHODS 47
It is widely used nowadays due to its high sensitivity while even more sensitive than fluorometry.
Methods: Chemiluminescence Immunoassay (CLIA)
Uses: Autoantibody testing; measurement of hormones, drugs, vitamins, tumor markers; forensic analysis;
microbial and infectious disease marker studies; and toxicology
Photodetector: Photomultiplier tube (luminometer)
Notes to Remember
In chemiluminescence, the wavelength of the emittedlight is longer than that of the exciting light because some energy
is lost before emission from the excited state due to collisions with the solvent or other molecules.
Thesignal-to-noise ratio in chemiluminescence is very high because light is measured against a nearly dark
background.
The light emission from a single excited state is called fluorescence. INTEM
VII. OSMOMETRY
Principle: Measurement of changes in the colligative properties of solutions that occur owing to variations in
particle concentration.
It is the measurement of the osmolality of an aqueous solution such as serum, plasma, or urine.
When active osmotic particles are added to a solution, osmolality increases and four other properties of the
solution are also affected.
point of the solution remains stable for several minutes once Stirring
wire
equilibrium temperature is reached. Refrigerant
out
sample is rapidly
Procedure: In a refrigeration chamber, the
several degrees below its freezing point
supercooled to Sample Refrigerant in
stirrer to start the freezing process,
then mixed with a
During electrophoresis, proteins are negatively charged (anions) and they move towards the anode.
Electrical power, Support medium, Buffer, Sample, and
Detector
Basic Components:
Buffer: Barbital (pH 8.6)
48 REVIEW HANDBOOK IN CLINICAL CHEMISTRY
The ionic strength of the buffer determines the amount of current and the movement of the proteins for a fixed
voltage. If ionic strength is low, more current is carried by the charge proteins which move a longer distance (fast
mobility). If ionic strength is high, less current is carried by the proteins which move shorter distance (slow
mobility).
A high ionic strength prevents effective separation of electrophoretic bands.
Increasing the strength of the field (increased current) also increases migration.
The more the ph of the buffer differs from the isoelectric point (pl), the greater is the magnitude of the net
charge of that protein and the faster it will move in the electric field.
Electrophoretic Mobility
Electrophoretic mobility is directly proportional to net charge and inversely proportional to molecular size and
viscosity of the supporting medium.
The larger the protein molecule, the more it interacts with the environment, creating a "solvent drag" effect
which slows down the movement of the protein.
A particle without a net charge will not migrate, and it remains in the point of application.
The ions carry the applied electric current and allow the buffer to maintain a constant pH during electrophoresis.
Precautions
Improper alignment of the electrodes may cause the current on one side of the gel to be denser than on the other.
This imbalance drives the migration ofproteins to the side with more current.
The proteins may migrate off the gel and into the buffer if the electrophoresis lasts too long.
If the electrical circuit is interrupted and no current flows, the proteins will not
migrate from the point of
application.
Gels frequently display the "smile artifact,"
in which migration of samples at the center of the gel is farther than
that of the samples at the margins.
Densitometry
It measures the absorbance of stain - concentration of the dye and protein fraction.
It scans and quantitates electrophoretic pattern.
Basic components: Light source, monochromator, scanner, optical system, and photodetector
Notes to Remember
At pH 8.6, the gamma globulins move toward the cathode despite the fact that they are negatively charged a condition
known as endosmosis.
The actual distance traveled by a particular protein migrating in an electrical field is determined by the combined
magnitudes of the electromotive force (a feature of the protein itself and the pH) and the electro-osmotic force
(a function primarily of the support medium).
at the
After electrophoresis, the gel is treated with a mild fixative, such as acetic acid, that precipitates the proteins
positions to which they have migrated. They are then stained, and the gel is dried and cleared of excess stain.
Because electrophoresis separations sometimes necessitate high voltages (50-200 V DC), the power source should
maintain a consistent DC voltage at these depths.
Isoelectric Focusing (IEF)
It separates molecules by migration through a pH gradient.
It is ideal for separating proteins of identical sizes but with different net charges.
by adding acid to the anodic area of the electrolyte cell and adding base to the cathode
ph gradient is created
area.
Proteins move in the electric field until they reach a pH equal to their isoelectric point.
cellulose acetate. t r S STRINI POMSINT9
Supporting media: agarose gel, polyacrylamide gel, and
Advantages of IEF
1. The ability to resolve mixture of proteins
To detect isoenzymes of ACP, CK, and ALP in
serum
2.
as alpha-1-antitrypsin
3. To identify genetic variants of proteins such
4. To detect CSF oligoclonal banding
Capillary Electrophoresis
In this method sample molecules are separated by electro-osmotic flow (EOF),
It utilizes nanoliter quantities of specimens.
in the specimen emerge
early at the capillary outlet because the EOF and the ion
Positively charged ions
movement are in
the same direction.
the specimen
move towards the capillary outlet but at a slower rate.
Negatively charged ions in
ves Servration, quantitation, and determination of molecular weights of proteins and peptides: analysis
of PCR
Terminologies
on pH conditions
1. Amphoteric - has a net charge that can be either positive or negative depending
the fixed support
Electroendosmosis/Endosmosis - is the movement of buffer ions and solvent relative
to
2.
Principle: It determines the components of : mple where the x-axis represents the mass-to-charge (m/z) ratio
and the y-axis is their relative abundance within the sample.
It is based onthe fragmentation and ionization of molecules using a suitable source of energy.
It separates diverse molecules from one another in the gas phase (gas-liquid chromatography) utilizing the so-
called mass spectroscopy-mass spectroscopy (MS-MS) methodology.
It is vital in structural identification and determination of molecular weight.
The relative abundance of each ion produces a distinct mass spectrum of the original molecule.
Uses: Detection and quantification of analytes in human fluids and tissues; utilized in proteomics, metabolomics,
and toxicology; differentiates normal tissues from malignant tissues; quanti-assay of both therapeutic and illicit
drugs; and drug development
Requirement: Before a compound can be detected and quantified by MS, it must be separated by GC.
Advantage: Qualitative and quantitative analysis of complicated samples; in biomedical research
Basic components: Sample inlet, ion source, mass analyzer, and ion detector
Samples: Blood (serum and plasma), urine, tissues, and other body fluids
Sample Preparation
Samples should be volatilized then ionized prior to measurement, to generate charged molecular ions and
fragments.
During Molecular Fragmentation
The original/native molecule ions undergo fragmentation to lower molecular mass forms.
Molecular ions are transported via an inert gas through a tube comprising a section that is in an alternating
electric field.
The higher m/z fragments migrate onto the tube at closer positions to the electron beam, whereas the smaller
cationic fragments migrate farther away from the beam the smaller cationic fragments migrate farther along
the tube than the larger cationic fragments.
A change in voltage happens everytime a fragment "hits" the tube, representing a specific m/z ratio.
The size of thevoltage change indicates the amount of a specific fragment.
Generation of Mass Spectrum
The mass spectrum is achieved whenever the plot of the
voltage changes is compared to the m/z ratio.
Each compound or macromolecule has a unique mass spectrum, which appears in a "fingerprint" pattern.
Given the electron beam intensities, temperature, and other parameters, fragments, as well as their relative
abundances, define the parent molecule, allowing the original molecule present in a patient sample to be
identified.
The ions or analytes are separated by m/z in a mass spectrometer, which is shown in the mass
x-axis of a
spectrum againsttheir relative abundance (y-axis).
The sample is measured by a detector which provides the ion current (charge) for each analyte.
To connect signal intensity to concentration, absolute quantification by mass spectrometry requires an internal
standard, a chemical that is
similar to but not identical to the analyte of interest.
Mass Spectrometry Imaging
Rather than assessing absolute concentrations, this technique focuses on monitoring various biomolecules such
as lipids and peptides.
ANALYTICAL METHODS 51
Sample Vaporizer
Magnet
Sample (Filament)
Injection
Port 1606001 FIGURE 11. Mass Spectrometry.
Magnetic
Field Source: Priyam Study Centre. (n.d.).
Mass Spectrometry. https://www.
priyamstudycentre.com/2022/02/mass-
spectrometry.html
Sample Accelerated
Molecule Ion Beam
Electron
Beam
Detector
method with FDA-approved immunoassays and LC-MS/MS methods. PSI showed a significant correlation and
McPherson and Pincus, 2022).
adequate capabilities (Shiet al,, 2015, as cited in
Desorption Electrospray Ionization (DESI)
It has similarity with MALDI, however, it utilizes a charged solvent stream instead of a
laser.
It is also used in cancer biology specifically in the study of brain cancer markers.
the charged solvent
The charged sample ions that is formed between the sample (blood or urine) and
are
determined by MS.
under ambient environments
It provides a great advantage in biomolecule profiling and imaging capabilities
(McPherson and Pincus, 2022).
X. PROTEOMICS
Principle: It is basically the study of proteins in aid of disease diagnosis.
biomarkers that will assist in the detection of diseases.
It identifies potential
at given time
It involves the characterization of complete set of proteins present in a cell, organ, or organism
(Chandramouli and Qian, 2009).
It determines the actual, real-time condition of the cells.
High-resolution MS are often used to improve the specificity of highly charged peptides and proteins, and can
separate isotopic peptides that differ in their composition (Mann and Kelleher, 2008, as cited in McPherson and
Pincus, 2022).
When coupled with chromatography and mass ratio, proteomics is a reliable method for amino acid sequencing.
Uses: For proteome profiling, comparative expression analysis of two or more protein samples, localization and
XII. CHROMATOGRAPHY
Principle: It is the components in a solution by specific differences in physical-chemical
separation of soluble
characteristics of the different constituents, with a mobile phase and a stationary phase.
Bases of Separation
1. Rate of Diffusion
4. Sample volatility/solubility
5. Distribution between two liquid phases
6. Molecular Size (molecular sieving)
7. Hydrophobicity of the molecule
8. Ionic attractions
ANALYTICAL METHODS 53
9. Differential distribution
between two immiscible liquids
10. Selective separation of substances
11. Differences in adsorption and desorption of solutes
The retention time is utilized to determine analytes versus the standard retention time run in the same
conditions.
Peak is created by the sample analytes and it is proportional to the concentration of the sample.
FORMS OF CHROMATOGRAPHY hon
A. Planar
Paper Chromatography
It is used for fractionation of sugar and amino acid.
Sorbent (stationary phase): Whatman paper
Thin Layer Chromatography (TLC)
It is a semiquantitative drug screening test.
Sample components are identified by comparison with standards on the same plate.
Extraction of the drug is pH-dependent the ph must be adjusted to reduce the solubility of the drug in the
aqueous phase.
When all drug spots including the standards have migrated with the solvent front, it is caused by incorrect
aqueous to nonaqueous solvent mixture.
Biological samples such as blood, urine, and gastric fluid can be used for the test.
Each drug has characteristic Re value
and it must match the R value of the drug standard.
Sorbent: Thin plastic plates impregnated with a layer of silica gel or alumina.
the relative distance of migration from the point application
Retention factor (R) value - is
B. Column
Pressure Regulator
Injection Port
Computer
Controller
and Data
FIGURE 12. Gas Chromatography.
Acquisition Inlet Gas Filters
Oven Source: ScienceDirect Topics. (2017). Gas
Chromatography - An Overview. https://www.
Flow sciencedirect.com/topics/agricultural-and-
Controller biological-sciences/gas-chromatography
Gas Chromatograph
Carrier Gas
Detector
Supply
Transfer Line
Liquid Chromatography
the distribution of solutes between a liquid mobile phase and a stationary phase.
It is based on
The sensitivity of LC-MS/MS may allow lower limits of detection for some analytes, such as steroids, compared
to
immunoassays and other methods while giving clinical laboratories the ability to multiplex, identifying and
quantifying several analytes of interest simultaneously (Jannetto, 2015).
Basic components: Liquid
mobile phase, sample injector (manual or automatic), mechanical pump, column,
detector, and a data recorder
Advantage: Uses lower temperatures for separation of analytes, thereby achieving better separation of
thermolabile compounds
various hemoglobins associated with specific diseases (e.g., thalassemia); rapid HbA1c test (within 5 minutes)
Sensitivity level: Nanomolar to micromolar range .291431
Gel/Gel Permeation/Gel Filtration/ It separates molecules based on differences in their size and shape. As solutes travel through
Size Exclusion/Molecular Sieve the gel, large molecules remain in the mobile phase and are eluted rapidly from the column.
1. Potentiometry
It is the measurement of electrical potential due to the activity of free ions - change in voltage indicates
activity of each analyte.
It measures the differences in voltage (potential) at a constant current.
It follows the Nernst equation - an ideal reference electrode should conform to this equation.
Concentration of ions in a solution can calculated from the measured potential difference between the
be
1.
Saturated calomel electrode - consists of mercury in contact with a solution that is saturated with mercurous
chloride (calomel) and known concentration of potassium chloride
2. Silver-silver chloride - consists of a silver electrode immersed in solution that has
potassium chloride
been saturated
with silver chloride; ideal for temperatures higher than 60° C and react with more sample
components compared to calomel
Protein coating
the ISE membrane would cause a response error--if interference is due
ISE in undiluted specimen can be employed to yield correct electrolyte activity
alternative mode of analysis such
(i.e., equivalent of osmolality).
Causes of Malfunctions of ISE
of counter voltages from liquid junction potentials at the salt bridge, and
Defective ISE membrane, buildup
buildup of proteins at the electrodes
58 REVIEW HANDBOOK IN CLINICAL CHEMIST
Indicator
Reference
Electrode
Electrode
FIGURE 13.
Ion-Selective Electrode.
Source: Lindner and Pendley,
2012
Bridge Electrolyte-1
Inner filling
Solution
Bridge Electrolyte-2
Ion-selective
Liquid Junctions Membrane
Sample Solution
2. Coulometry
It is the measurement of the amount of electricity (in coulombs) at a fixed potential.
It is an electrochemical titration in which the titrant is electrochemically generated and the endpointi
detected by amperometry.
It follows Faraday's law-the number of silver ions generated after ionization can be calculated fron
Faraday's law, and the product of the calculation is exactly equal to the chloride ions in the sample.
Use: Chloride test (CSF, serum, and sweat)
Procedure: A constant current is applied across the two silver electrodes, which liberates silver ions into
specimen at a constant rate. Chloride ions in the sample combine with silver ions to produce insoluble silve
chloride. A pair of indicator and reference electrodes senses the excess silver ions and stops the titration
(McPherson and Pincus, 2022).
Interferences: Bromide, cyanide, and cysteine
3. Amperometry
It is the measurement of the current flow produced by an oxidation reaction.
Several immobilized enzyme electrodes use this amperometry principle.
Use: pO2, glucose, chloride, and peroxidase determinations
Polarography
It is the measurement of differences in irrent at a constant voltage.
It follows the Ilkovic equation.
pO2 measurement
It is determined amperometrically.
In this assay, the oxygen from the blood sample diffuses across a semipermeable membrane and is reduce
at the cathode of a polarographic electrode, thereby producing a measurable current that is directl
proportional to the sample p02 (Trulock 1II, 1990).
It uses Clark electrode.
4. Voltammetry
It is the
measurement of current after which a potential is applied to an electrochemical cell.
It allows the sample to be preconcentrated, thus utilizing minimal analyte concentration.
Advantage: High sensitivity and the capability for multielement measurements (analytes can be detecte
in the parts-per-billion range)-several analytes can be measured simultaneously in a single voltammel"
assay
(POCT)
It is also known as the
decentralized test, bedside test, or near-patient test.
It is an analytical testing performed outside the laboratory usually by nonlaboratorian personnel.
It is used both
in chemical analysis and microbiological tests.
It is
currently routine in all hospitals and has become the standard for patient care in a variety of other health
care settings (Nichols et al., 2020).
A separate certification
is usually not necessary when it is performed under the supervision of the laboratory
director responsible for laboratory services in a health care facility.
Examples: bloodglucose meter, electrolytes, blood gases, enzymes, cardiac markers, and therapeutic drug
monitoring (TDM)
Types of POCT: Wet chemistry, cartridge-based, and strip-based
Disclaimer: The author is not in any way connected with the manufacturer/
distributor of nor promoting this product. The descriptions and photo are
used only for lecture/discussion purposes.
analyte antibody (the antibody that specifically targets the molecule being measured). lodds 19th
This kind of assay is mostly used in TDM, hormone studies, and toxicology.
Chemistry Laboratory b sdal3edW 0f
Examples of Immunoassays/Immunochemical assays used in the Clinical
Enzyme-Linked Immunosorbent Assay (ELISA)
Chemiluminescence Immunoassay (CLIA)
Immunoradiometric Assay (IRMA)
Immunochemiluminometric Assay (ICMA)
Fluorescence Polarization Immunoassay (FPIA)
Enzyme-Amplified Sensitivity Immunoassay (EASIA)
NOTE: For the principles and discussions of the immunoassays, refer to Chapters 19 and 21.
60 ANALYTICAL METHOD~
ASSESSMENT QUIZ
1. Which type of filter is best for measuring stray light?
didymium
b. sharp cut-off d. gratings
2. Which part of the spectrophotometer allows only a narrow light to reach the sample cell?
entrance slit c. exit slit
b. coulometry d. polarography
4. It involves dissociation of analytes by heat energy into atoms and measures minimal concentrations.
a. electrophoresis c. nephelometry
b. fluorometry d. atomic absorption spectrophotometry
5. Ion selective electrode is a "selective" method due to which factor?
C.
tandem mass spectroscopy
b. d. MALDI-TOF
10. What is the
ideal concentration of the buffer in serum protein electrophoresis?
a. pH 5.2 C. pH 7.4
b. pH 6.5 d. pH 8.6
11. In electrophoresis, if
the electrodes are not fixed and properly aligned, the proteins will migrate
on the side with more current.
a. closer
C. very near
b. farther
d. faster
12. Which separation technique is a requirement in mass
spectrometry before the unknown analytes can be
detected and quantified?
a. gas chromatography C. time of flight
b. electrophoresis
13. In electrochemistry, it is a type of pH electrode with
measurement of carbon dioxide.
a sodium bicarbonate buffer utilized in the
a. Clark
C. Calomel
b. Valinomycin d. Severinghaus
ANALYTICAL METHODS 61
14. It
measures proteins with large molecular complexes using scattered light and sensitive enough
quantitate concentrations less than
to even