0% found this document useful (0 votes)
32 views

Incubator

The document contains verification forms for two incubator models, detailing calibration procedures, dates, and frequency of checks. It includes temperature readings before and after calibration, as well as actual readings at various positions within the incubators, with tolerances specified. Additionally, it lists reference equipment used for calibration and their traceability information.

Uploaded by

rizmin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
32 views

Incubator

The document contains verification forms for two incubator models, detailing calibration procedures, dates, and frequency of checks. It includes temperature readings before and after calibration, as well as actual readings at various positions within the incubators, with tolerances specified. Additionally, it lists reference equipment used for calibration and their traceability information.

Uploaded by

rizmin
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as XLSX, PDF, TXT or read online on Scribd
You are on page 1/ 3

Doc.

Ref:ISO/Calbr/Incubator
Edition: 01 Page 1/1

VERIFICATION FORM
Incubator
Equipment Name / Model: Equipment Reference:
Incubator/ET 618-4 C 11553
Calibration Procedure: Date: Frequency:
Comparison Method 10-Oct-13 Yearly
Temperature Pressure Relative Humidity Ref. Equipment Used
Before: 22 °C atm 55 % RH Ref. Thermometer R 0615
After: 22.5 °C atm 56 % RH Ambient R 0965

I. Calibration Data of Testing Equipment

Actual Reading (oC)


Position Set Temperature Tolerance
1 2 3 Average
Top Midle 20.40 20.40 20.40 20.40 21.0 ± 0.5 oC

Middle 21.0 oC 20.20 20.60 20.60 20.47 21.0 ± 0.5 oC

Bottom Middle 20.60 20.60 20.50 20.57 21.0 ± 0.5 oC

………………………………………. ……………………………………
Performed By: Checked By:
Doc. Ref:ISO/Calbr/Incubator
Edition: 01 Page 1/1

VERIFICATION FORM
Incubator
Equipment Name / Model: Equipment Reference:
Incubator/SCT-FEDEL-X C 1155
Calibration Procedure: Cal Date Cal Due On:
Comparison Method 18-Mar-16 18-Mar-17
Temperature Pressure Relative Humidity Ref. Equipment Used
Before: 22 °C SATP atm 55 % RH Ref. Thermometer R 0615
After: 22.5 °C SATP atm 56 % RH Hyg.Thermometr C 1456

I. Calibration Data of Testing Equipment

Actual Reading
Position Set Temperature Tolerance
1 2 3 Average
Top 35.30 35.20 35.10 35.2 35.0 ± 0.5 oC

Middle 35.0 oC 34.50 34.60 34.50 34.5 35.0 ± 0.5 oC

Bottom 34.60 34.50 34.60 34.6 35.0 ± 0.5 oC

ID. Equipment Model Serial Cal. Due Date Traceability No.


R 0615 Readout ThermometerETI2001 D08480487 17-Mar-17 220081222055701
C 1456 Hygro Thermometer ETI 19-Sep-16 222008122235543

………………………………………. ……………………………………
Performed By: Checked By: