Mahr-software-usermanual

The document is an operating manual for MarCom Standard/Professional Software version 5.0 by Mahr GmbH, detailing installation, configuration, and usage instructions for connecting measuring instruments. It covers functionalities such as data transfer, settings management, and connecting RF devices. The manual provides a comprehensive guide to ensure users can effectively utilize the software and its features.

Uploaded by

luctiendat21
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
63 views38 pages

Mahr-software-usermanual

The document is an operating manual for MarCom Standard/Professional Software version 5.0 by Mahr GmbH, detailing installation, configuration, and usage instructions for connecting measuring instruments. It covers functionalities such as data transfer, settings management, and connecting RF devices. The manual provides a comprehensive guide to ensure users can effectively utilize the software and its features.

Uploaded by

luctiendat21
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 38

 Bedienungsanleitung

 Operating instructions

MarCom Standard / Professional©


Software

3722558

Version 5.0

Mahr GmbH
Standort Esslingen
Reutlinger Str. 48, 73728 Esslingen
Tel.: +49 711 9312 600, Fax: +49 711 9312 756
0515 [email protected], www.mahr.com


Contents

Introduction .......................................................................................................................3

Installation .......................................................................................................................3

Start program .......................................................................................................................4

Menu bar and configuration........................................................................................................5

1. Program .......................................................................................................................5
1.1 Close window..................................................................................................................5
1.2 Exit .......................................................................................................................5

2. Measuring Instruments................................................................................................6
2.1 Reset Measurement Cycle..............................................................................................6
2.2 Previous .......................................................................................................................6
2.3 Next .......................................................................................................................6
2.4 Edit .......................................................................................................................6
2.5 Request Value.................................................................................................................6
2.6 Add a RF Device.............................................................................................................7
2.7 Delete a RF Device.........................................................................................................7
2.8 Delete All RF Devices.....................................................................................................7
2.9 Refresh .......................................................................................................................7

3. Settings .......................................................................................................................8
3.1 Password protection.......................................................................................................8
3.2 Load .......................................................................................................................8
3.3 Save .......................................................................................................................9
3.4 Save as .......................................................................................................................9
3.5 Measuring Station No. ....................................................................................................9
3.6 RF .......................................................................................................................9
3.7 COM .....................................................................................................................10
3.8 File .....................................................................................................................10
3.9 Virtual Interface Box .....................................................................................................10
3.10 Autostart .....................................................................................................................11
3.11 Languages....................................................................................................................11
3.12 Beep .....................................................................................................................11

4.0 Information (?).............................................................................................................12


4.1 Info .....................................................................................................................12
4.2 Mahr Homepage...........................................................................................................12

5. Parameterize Measuring Instruments.......................................................................13


5.1 MC-I Switch...................................................................................................................13
5.2 Measuring Instrument...................................................................................................14

6. Connecting RF Devices..............................................................................................21
6.1 i-Stick - Integrated Wireless..........................................................................................21
6.2 e-Stick...........................................................................................................................25
6.3 FM2...............................................................................................................................25

7. Virtual Interface Box...................................................................................................28

8. Overview of the Status and Configuration Windows...............................................33

9. Implementing a Remote Control................................................................................34

10. RS232 Settings on a Measuring Instruments...........................................................35

2 Mahr GmbH • MarCom Standard / Professional 5.0© Software




MarCom Professional 5.0 ©

Introduction

With the MarCom Standard Software you can transfer your measured values into a text file, into Microsoft Office
Excel® or via a keyboard code into any application, all you need is a Mahr measuring instrument connected via a
USB port and a Mahr USB data cable or with the integrated wireless receiver (i-Stick). In addition to a USB port it is
also possible to use a serial COM interface with the suitable data cable for data transmission.
If you have purchased the MarCom Professional Software (optional), you can operate up to 127 measuring
instruments (subtract the amount of hubs being used) via the USB hubs and one additional serial COM interface;
as well as transferring the measuring values via a virtual Interface Box to an further program (e.g. CAQ software).
Therefore you can easily centralize the multiple measuring instruments to just one measurement station. With the
aid of a USB foot switch (optional) you can also transfer your measured values from each individual measuring
instrument to a text file, blank sheets, in separate Microsoft Office Excel® columns or to a CAQ software.

Note: When several USB hubs (e.g. for cascading) are connected together these must be operated in
conjunction with an external power supply.

Please read the following operating instructions carefully so that you can become quickly accustomed to the
MarCom Software.

Installation

In order to start the installation of the MarCom Software, please execute the Setup-MarCom.exe file in the root
directory of your data medium and follow the instructions. Once successfully installed you may plug in the USB
cable or the receiver (i-Stick/e-Stick/FM2) and continue with the installation of the driver. If you wish to connect
multiple USB data cables we recommend that you install the cable and driver successively.

Note: Once the wireless receiver (i-Stick/e-Stick) has been installed, up to 8 wireless measuring
instruments can be used. Should you require a further 8 wireless measuring instruments a
second i-Stick/e-Stick should be plugged into a free USB interface. A total of maximum
4 i-Stick/e-Stick wireless receivers with 32 measuring instruments can be connected
simultaneously.

Minimum system requirements:

• MS Windows 8, 7
• One free USB-interface 2.0 or higher
• Up to 500 MB of available hard disk space
• Recommended: MS-Excel 97 or higher

Mahr GmbH • MarCom Standard / Professional 5.0© Software 3




Start program

After starting the program a small status window will appear in the screen in which you are to determine where the
program files and templates are to be stored (directory) in the future, in either documents or public documents.

Once this has been defined, as soon as the program is started the following status window will appear on the
screen. In this status window you can instantly see whether a measuring instrument has been connected and
recognized (green signal lamp), plus the last transferred measured value will appear.
This window must always be open so that the measured values can be transferred!

Open the configuration window with a double click on the status window.

In this configuration window the connected measuring instruments will be displayed and configured. Normally, after
the first installation only the free serial COM interface will be shown.

Now, you can plug in the USB data cable, please follow the instructions regarding the driver installation contained
in the brief guide. Once the driver is successfully installed, the new connected measuring instrument will appear in
the configuration window. Please plug in the second USB data cable and also install the driver; with multiple USB
data cables repeat these steps accordingly (only with MarCom Professional). When you wish to connect a wireless
measuring instrument please refer to chapter 6 “Connecting RF Devices”.

When the driver is successfully installed, the new measuring instrument will appear in the configuration window,
along with the description of the connected USB data cable and a description of the measuring instrument. In order
to activate the new data cable, click on “Measuring Instruments – Refresh”.
The connected measuring instrument can be edited (see chapter “2.4 Edit”).

4 Mahr GmbH • MarCom Standard / Professional 5.0© Software




Menu bar and configuration

1. Program

1.1 Close window

With this menu you can close the configuration window; alternatively, you can click on the cross with the red
background in the top right corner of the configuration window. Even when the configuration window is closed
the status window is still open and it is still possible to transfer measured values. With a double click upon the
status window the configuration window will reopen.

1.2 Exit

If you click on Exit the complete MarCom Program including the status window will be closed, no further
measured values will be transferred. Alternatively you can click on the cross with the red background in the
top right corner of the status window.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 5




2. Measuring Instruments

2.1 Reset Measurement Cycle

In the parameter window the MC-I data cable (e.g. foot switch), or a measuring instrument; the defined
measuring cycle will be reset to zero. See sections “5.1 MC-I Switch” and “5.2 Measuring Instrument”.
Alternatively, the measuring cycle can also be reset clicking the Reset button in the status window.

2.2 Previous

The red field highlighted in the configuration window will move to the preceding measuring instrument e.g.
from USB2 to USB1. The highlighted measuring instrument can then be edited.
Tip: You can also select a measuring instrument to be edited by clicking upon it.

2.3 Next

The field highlighted in red in the configuration window will return to the following measuring instrument e.g.
from USB 2 to USB 3. The highlighted measuring instrument can then be edited.
Tip: You can also click upon a measuring instrument to edit it.

2.4 Edit

With Edit the parameter window of the aforementioned red highlighted measuring instrument will open.
Here the settings for the connected measuring instrument can be made. See ”5. Parameterize Measuring
Instruments”
Tip: The parameter window can also be opened with a double click on the image of the measuring
instrument in the configuration window.

2.5 Request Value

For test purposes a measuring value from a measuring instrument can be requested.

6 Mahr GmbH • MarCom Standard / Professional 5.0© Software




2.6 Add a RF Device

Add a new RF device to the MarCom Software. In the submenu there are 3 RF systems available.

i-Stick
Integrated Wireless system based on ANT+ technology.

e-Stick
A new series of external RF module units which are similar to FM2; however, with a new and compact USB
receiver which uses low Energy RF (wireless) technology.

FM2
An older range of external RF module unit, these can be connected via the cable interface.

For further detailed information regarding how to add an RF device, please refer to chapter “6 Connecting
integrated wireless measuring instruments”.

2.7 Delete a RF Device

Connected RF devices can be deleted in the MarCom configuration. To reinstall an instrument in the MarCom
Software, please refer to chapter “2.6 Add a RF device”.

2.8 Delete all RF Devices

All connected RF devices can be deleted in the MarCom configuration. To reinstall these instruments in the
MarCom Software, please refer to chapter “2.6 Add a RF device”.

2.9 Refresh

The USB port will be examined for a detached or a newly connected cable, additionally the configuration
window will return to the last saved version.
Attention: When the cable is either plugged in or out the configuration will automatically be refreshed!
Amended settings and settings that have not been saved will be lost!

Mahr GmbH • MarCom Standard / Professional 5.0© Software 7




3. Settings

3.1 Password Protection

Here you can determine a password for your MarCom configuration, thus protecting against unauthorized
third-party access. Simply enter any desired password, repeat the process a second time, and subsequently
activate the password protection by pressing Activate.

In order to revoke the password protection, enter the password and then press Abort. To reactivate a password
simply enter any desired password, repeat the process a second time, subsequently activate the password
protection with the Activate icon.
Attention: Please store your password carefully otherwise you cannot deactivate the password protection!

3.2 Load (Only available with MarCom Professional)

Here the previously saved configuration settings can be reloaded.

8 Mahr GmbH • MarCom Standard / Professional 5.0© Software




3.3 Save (Only available with MarCom Professional)

The actual configuration settings will be saved under the actual opened file (e.g. DEFAULT.MCC).
Attention: previously saved configuration settings will be overwritten!

3.4 Save As (Only available with MarCom Professional)

The actual configuration settings can be saved under a different file name, therefore it is possible to save
and reuse multiple configuration settings. Note: If the cable does not have a USB plug, which is required
for stored configuration, it will not be shown. If an additional cable is attached which is not for the stored
configuration, then this will be shown in its initial non-configured state.

3.5 Measuring Station No.

When using several integrated wireless receiver to transmit the measured values to different PCs that are in
close proximity we therefore recommend that before connecting each of the integrated wireless measuring
instruments, every PC should be assigned on the with a different Measuring Station No. (No. 1 to 64 are
possible). Therefore, you will avoid inadvertently sending measured values from an integrated wireless
receiver to a PC (measuring station) which has not been appropriately assigned.

3.6 Wireless (Radio Frequency)

The user can specify that when the digital indicator is switched off whether MarCom continues to seek
a wireless connection with the digital indicator or not. When “Disconnect Wi-Fi” has been selected and
subsequently the digital indicator is switched off the integrated wireless connection will also be disconnected.
Once the digital indicator is switched on, the connection has to be once again selected in both the MarCom
software and on the digital indicator. If this “Disconnect Wi-Fi” mode has not selected then once the digital
indicator is switched on again the wireless connection is found automatically.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 9




3.7 COM

This menu offers a clear overview of the serial COM interface that is being used; they can be shown or
hidden in the configuration window. (Only COM1 is available with MarCom Standard).

3.8 File

The global Excel file settings for all measuring instruments are conducted here.

If global is activated, then the following settings under ”Open As” are valid for all measuring instruments.
The measured values of all measuring instruments are then transferred into a combined Excel file, however
the settings under ”Open As” in the parameter window for measuring instruments are marked in a light gray
and thus cannot not be edited.

If global is deactivated, then the following settings under ”Open As” in the parameter window of the
individual measuring instruments can be altered separately, therefore for example the values from each
individual measuring instruments can be send to a different Excel file.

With ”Open As” you can select an Excel file and the path (push button with 3 dots) for all the measuring
instruments. Additionally you can choose whether to open a particular template (.xlt or .xltx) or a particular
file (.xls or .xlsx) for data acquisition.

3.9 Virtual Interface Box (only with MarCom Professional)

One or two virtual interface boxes can be activated, these interface boxes must be assigned to an interface
(e.g. COM 3) and to a box type (e.g. Mux50 Mahr-Box).
Additionally, to each designated Virtual Box a Foot Switch can also be assigned, which can be later used to
trigger the data request.

10 Mahr GmbH • MarCom Standard / Professional 5.0© Software




Note: Once the type of box has been selected, a tick must be positioned in the respective check box. This is
important so that the box will appear subsequently in the configuration window as an option.

In the Parameter window the individual measuring instruments can be subsequently assigned to a virtual
interface box, the individual measuring instrument is effectively connected directly to the determined virtual
interface box (see Chapter 7 “Virtual Interface Box”). Therefore you can connect a USB or an
integrated wireless measuring instrument simply to third-party software such as, for example, CAQ program,
whether you are connected via an interface box (Multiplexer).

3.10 Autostart

Specify whether MarCom starts automatically with a Windows start, this is useful when for example using
MarCom as a virtual Interface Box.

3.11 Language

Select a different language.

3.12 Beep

A beep (audible signal tone) as a Wave file can be selected (enabled / disabled) for the transmission of
measured data and for the end of a measuring series.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 11




4. Information (?)

4.1 Info

Information regarding the MarCom Software, for example which version.

4.2 Mahr Homepage

www.mahr.com

12 Mahr GmbH • MarCom Standard / Professional 5.0© Software




5. Parameterize Measuring Instruments

5.1 MC-I Switch

The MC-I Cable is fitted on one end with a jack bush so that it can connected to a relay (e.g. foot switch
4102058) with a jack plug (according to JIS C 6560, 3.5 mm), therefore multiple foot switches can be utilized
via USB hubs to control individual or multiple measuring instruments.

• Connection: Description of the connected data cable (e.g. MC-I).

• Device: Description of the connected device type (e.g. switch)

• Description: A more detailed description can be entered (e.g. foot switch) in order to clearly identify
the switch

• Measurements per Cycle:


Enter the number of measurements per cycle for example: when you enter 3, the
foot switch must be operated 3 times before a cycle can begin, this is important to
the allocation of the measuring instruments.

Example: If by ”Measurements per cycle” 3 is entered then select parameterize a measuring


instrument (see 5.2 Measuring Instruments) as “Data Request” via MC-I Switch.
Select the appropriate switch and the option “with Measurement No. 2”, then every
second time (2 times) the foot switch is operated a measured value will be transferred
in the cycle. The first and third (1 and 3) are not activated, therefore no measured data
will be transferred. In this situation another instrument could be allocated in in this first
or third cycle. When measuring instruments with “Measurement No. 1” and
“Measurement No. 3” have been allocated, those measured values which are obtained
by the first and third activation of the foot switch transferred.

Note: Please make sure that when entering in “Measurements per cycle” and “with
Measurement No.” that in the parameter window the measuring instruments
correspond logically. For example do not enter “with Measurement No” 5, although
by “Measurements per cycle” only 3 has been entered.

• Actual Measurement No:


Here the actual status of the cycle is shown, if necessary one can skip to a certain
point in the cycle.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 13




5.2 Measuring Instrument

In this menu all the most important settings for the individual measuring instruments can be administered.

• Connection cable: Select the connected data cable (e.g. 16EXu or Opto Duplex). When connecting
the USB data cable, the cable will automatically be recognized and cannot be
edited.

• Instrument: Select the connected measuring instrument (e. g. MarCator 1086) the selected
instrument will also be displayed as a picture for easier recognition.

• Description: Enter an appropriate description of the connected measuring instrument (e.g. Dial
Indicator 2) or the measuring task being conducted.

• Data Request: Select a data request, request measured values

With “None” the value must be send with the measuring instrument, for example
with the Data button on the data cable or on the measuring instrument.

Note: Regardless of which data request has been selected the measured value can
always be send directly from the instrument.

If the window is a light gray, the contents cannot be selected and the connected
measuring instrument will not support request measured values. In this instance the
measured value can be sent from the measuring instrument (e.g. by pressing the
DATA button).

14 Mahr GmbH • MarCom Standard / Professional 5.0© Software




Via a “keypad”, request measured values with a predetermined button on the keyboard e.g. F1, F2 up to F12.

Note: The functions buttons F1, F2 etc. are deactivated in MS Excel!

Via a “MC-I Switch” (foot switch). Here you can select additional values with either “per Measurement” (i.e.
each time the switch is operated a value will be sent) or “with Measurement No.” (i.e. the values that are to be
send to the measuring cycle after X-times activation of the switch).
How to define the measuring cycle can be found in section 5.1 MC-Switch.

Via the “Timer”; here you can set the time control of a cyclical data request. The cycle can be entered in
hours, minutes and seconds. The timer cycle will be shown as a Timer symbol in the configuration window,
the symbol can be used to start or stop the timer.

Additionally, a pull-down menu for starting and stopping the timer is available. You can select whether the
timer is to be also started and stopped via a function key (e.g. F1), MC-I switch (e.g. foot switch) or global via
the timer button in the status window. In this way several timers for various measuring instruments can be
simultaneously started and stopped.

Note: The Timer key to start and stop the timer is with selection of a transfer to using a
“keyboard code” is not available and is not practical.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 15




Zero setting:

When using a RF device an additional point appears in the menu.

Here the type of action can be selected in which the RF device gives the command for zero setting. Whilst
“Keypad” and “MC-I Switch” function identically to that of Data Request it is also possible to create an
additional button on the user window which is clicked with the mouse.

• Transfer to: Select the data format in which the values should be sent to (e.g. MS Excel, text
file, keyboard code or a virtual interface box).

With “Text file” the measured values taken from a measuring instrument will be transferred to a text file; you
can enter the path and the description of the text file with “File” and then “Text file-Settings”.

Note: When using several measuring instruments the values of each individual
measuring instrument should be transferred into separate text file.

16 Mahr GmbH • MarCom Standard / Professional 5.0© Software




With “Excel” the values are transferred into Microsoft Excel. If global (see 3.8 File) is deactivated, you can
select a separate Excel file or template as the destination. If global (see 3.8 File) is activated then the
measured values of all measuring instruments are transferred into a common Excel file or template. The set
up possibilities in “Open As” for the individual measuring instruments are deactivated, the global Excel file
settings can be conducted; see section “3.8 File”. Additionally, you can enter in the respective “Sheet”, the
respective “Column” and the respective “Start Row” in which the values will be prompted.

Note: When a write protected Excel file or template is used, the write protection must be
removed from column 256 (IV)!

Use “Insert in” to select whether the measured values will be transferred into a column or row.

With “Sheet” enter the name of the sheet; with “Column” select a column and with “Start row” select the
start row for the measured values. Once this has been conducted, define last row “End row” in which the
measured values will be transferred to. Finally determine what should happen once the measured values
have been transferred to the last column / row (cell):

With “Conclude measurement” the transmission will stop and a message window will appear on the screen,
informing you that the measurement has been concluded.

With “Continue measurement in next column/row” measurement will continued in the next column or row,
here it is possible to define an additional end column and/or end row. In this way yon can transfer measured
values to columns or rows in a zigzag form from a measuring instrument.
At the end of the defined last column and row, appears a message that the series of measurements has
been concluded.

With “Continue measurement in...” appears another bar in which you can define where (Sheet, column etc.)
the measurements should continue. “Continue measurement in...” can be cascaded up to 99 times.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 17




When “Insert in” row is selected, then the above description applies according to the row.

In the field “Actual”, the cycle and the actual cell (row and column) in which the data is to be transferred is
shown (defined). These values can be also changed, for example after deletion of measured values into a
certain cell to be transferred. In this way data can be transferred to cells in which the previous value was
deleted. Once the empty cells are used, “fill up” the transmission normally, continues without overwriting the
previously transferred value.

The measurement cycle can be reset be either in the menu “Measuring Instruments - Reset Measuring
Cycle “ (see 2.1 Reset Measuring Cycle) or the push button “Reset” in the status window.

With “keyboard code” transfer to, the measured values of the measuring instrument are transferred into the
actual current position of the cursor. Therefore, it is possible to transmit the measured value independently of
the software application. Measured values can therefore be transferred into MS Access, OpenOffice or into a
CAQ program.

Keyboard code settings: you can define 2 end markers (End marker 1 and End marker 2), once a measured
value has been transferred the sequence of the End markers will be conducted, i.e. if, for example ENTER is
selected, that corresponds to pressing the enter key on the keyboard (measured value is transferred and the
position of the measured value depends upon the chosen End markers / sequence). In this way you can jump
to any field in your form or table after a measured value has been transmitted.

18 Mahr GmbH • MarCom Standard / Professional 5.0© Software




In order to select the End marker you must click on the ENTER field and select the required character from
the Pull-Down menu.

With the decimal separator select the decimal separator according the type used in the country you are in or
according to type used in the application.

With “virtual interface box code” the measured values from the measuring instrument will be transferred to a
virtual interface box, which for example is connected to a CAQ software.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 19




Select a virtual interface box from the pull down menu (activate beforehand, see 3.9 Virtual Interface Box).
Subsequently use the mouse to click on a free connection port number in order to virtually connect the
measuring instrument to the interface box; the port connection port number for the measuring instrument
is marked with an X in the green box. If the connection port number is allocated to another measuring
instrument the box will appear in dark gray and cannot be selected as a connection port. Please refer to
Chapter 7 Virtual Interface Box for further information.

20 Mahr GmbH • MarCom Standard / Professional 5.0© Software




6. Connecting RF Devices

6.1 i-Stick - Integrated Wireless

To install a new Integrated Wireless (RF) device, please proceed as follows:

1. Start the MarCom Software

2. Insert the i-Stick receiver in a free USB interface

3. Windows usually displays a “Found new Hardware” window

4. Install the USB2 driver for the i-Stick receiver (the new Hardware indicates the installation path of the
”USB2 driver” on the MarCom data medium and install the driver); once installed the receiver can be
connected up to 8 integrated wireless measuring instruments. Should you require a further 8 integrated
wireless measuring instruments a second i-Stick receiver should be plugged into a free USB interface. A
total of max. 4 i-Stick wireless receivers with 32 integrated wireless measuring instruments can be
simultaneously connected.

5. Double click on the Mahr Logo

Once the driver of the i-Stick receiver has been successfully installed; “1 USB-Device” will appear in
the top left of the window.

6. Select the Menu “Measuring instruments  Add RF Device  i-Stick”

Mahr GmbH • MarCom Standard / Professional 5.0© Software 21




7. The “Parameterize Measuring Instruments” window will appear.

8. Select a channel (01 to 03). We recommend that when connecting for the first time, that you leave
channel set at 01, only change the channel if at a later date you experience any problems with
transmission, then change the channel in the MarCom Software and the integrated wireless
measuring instrument.

Note: The channel number must match the channel number of the measuring instrument!

9. Select the measuring instrument that is to be connected (e.g. MarCal 16EWRi or MarCator 1086Ri).

10. Enter an appropriate description for the Measuring Instrument (e.g. RF1)

11. Allot the measuring instrument an individual instrument number (from 001 to 999). The prefix e.g. 01 is
the measuring station, this can be set under “Settings  Enter a Measuring Station No.” (see
Chapter 3.5 Measuring Station No.). The Measuring Station No. enables that several measuring
positions in close proximity can be used as each instrument has its own individual number. Therefore,
you will avoid inadvertently sending measured values from an integrated wireless receiver to
a PC (Measuring Station) which has not been appropriately assigned.

Note: Never assign the same instrument number to several integrated wireless measuring
instruments!

22 Mahr GmbH • MarCom Standard / Professional 5.0© Software




12. In the Checkbox “ECO” you can activate the energy saving mode for the wireless data transmission, thus
reducing energy consumption of the integrated wireless wire measuring instrument up to 30 %.

Note: In the “ECO mode”, the transmission speed will be reduced; rapid transmission
intervals of <7 seconds are not recommended or possible. If the “ECO mode” is to be
activated then the “ECO mode” on the integrated wireless measuring instrument must
also be activated!

13. All other settings such Data Request, Transfer, etc. can be configured in MarCom just as if using a cable
(see Chapter “5.2 Parameterize Measuring Instruments”).

14. Subsequently, you must press the “Connect RF-Device” button in order to establish a connection. The
MarCom software will search for active integrated wireless measuring instruments; next select the
connected measuring instrument (e.g. MarCal 16EWRi or MarCator 1086Ri).

Mahr GmbH • MarCom Standard / Professional 5.0© Software 23




15. Switch on the wireless measuring instrument.

16. Press the “Menu button” on the wireless measuring instrument for > 2 seconds.

17. “d OFF” (Digital Indicator) or “UnLoc” (Digital Caliper 16 EWRi) will appear in the display. With
“UnLoc”press the “Menu” button 3 times until “OFF d” appears in the display (the wireless function is
deactivated).

18. Press either the “ON/OFF” button and/or the “OI” button to activate the wireless function.

19. „d - - - -“ appears in the display (at present noinstrument number has been allocated).

Note: If a instrument number has already been allocated (e.g. d 01002) this can be deleted by
pressing the “PRESET” button (on a Digital Indicator) or the “OI” button (on a Digital
Caliper), a new instrument number can now be assigned.

20. Press the “DATA” button on the wireless measuring instrument to establish a connection.

21. The “Wireless symbol” in the digital indicator’s display will blink, once the connection has been
established the wireless symbol will remain in the display; additionally the allocated instrument number
will also appear in the display (digital indicator).

22. With a digital indicator briefly press the “Menu” button to exit the menu. When using a Digital Caliper
16 EWRi, press and hold the “Menu” button to exit the menu.

23. The “Connection symbol” in the top left corner becomes green. To exit the Parameterize Measuring
Instruments window, press OK, this also confirm the setting in MarCom.

24. The first measuring instrument is connected, if further integrated wireless measuring instruments are
required, repeat from step 6 . . .

24 Mahr GmbH • MarCom Standard / Professional 5.0© Software




6.2 e-Stick

The e-Stick wireless module connection is identical to connecting an FM2 System. (see „6.3 FM2“).

6.3 FM2

To install a new FM2 wireless module device, please proceed as follows:

1. Start the MarCom Software

2. Insert the FM2 receiver in a free USB interface

3. Windows usually displays “Found new Hardware” window

4. Install the FTDI driver for the FM2 receiver


If this does not happen automatically, the driver we have to be manually installed (the Device Manager,
use the installation path ”FTDI Driver” on the MarCom data carrier and install the driver).

5. Double click on the Mahr Logo

Once the driver of the FM2 receiver has been successfully installed; “1 USB-Device” will appear in
the top left of the window.

6. Select the Menu “Measuring instruments  Add RF Device  FM2”

Mahr GmbH • MarCom Standard / Professional 5.0© Software 25




7. The “Parameterize Measuring Instruments” window will appear.

8. Select a channel (01 to 05). We recommend that when connecting for the first time, that you leave
channel set at 01, only change the channel if at a later date you experience any problems with
transmission, then change the channel in the MarCom Software and the FM2 sender.

9. Enter an appropriate description for the Measuring Instrument (e.g. FM2 - 1)

10. Once a request type has been selected (triggering the sending of the measured values to a PC) a new
box will appear “Request Timeout”. In this box you are able to set the stand-by duration of the transmitter
module, maximum duration is 30 minutes. Due to technical reasons with this module it is not possible for
periods longer than 30 minutes.

This internal timer will be automatically reset after each measurement.

11. All other settings such as Request of, Transfer to etc. can be configured just as a cable connection in
MarCom (see Chapter “5.2 Parameterize Measuring Instruments”).

26 Mahr GmbH • MarCom Standard / Professional 5.0© Software




12. To establish a connection with the sender, the Connect button in the top left must be activated; the
MarCom Software only scans after active senders in the transmission range.

13. Press the data button on the wireless module so long (ca. 5 seconds) until the red LED on the wireless
sender is extinguished and the green LED briefly lights up.

14. The Serial No. should appear in the configuration window, the connection window should then close.
Press the “OK” button to close configuration window.

15. If you changed the operation time “Request Timeout” (see section 10) you will be asked to press the
connection button again on the sender. By this you can transfer the new time setting to the module. This
will be only possible if you have already connected the transmitter. Afterwards the configuration window
closes automatically.

16. The Sender must be returned to the measuring mode, to do so briefly press the data button on the
sender. The blue LED is now continuously lit.

Note: The sender is now ready for request, as previously set, see section 10. This means
after this is completed and the blue LED is extinguished, it first possible to enable a
remote request by pressing the data button on the receiver.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 27




7. Virtual Interface Box (only with MarCom Professional)

In order to use the new Mahr virtual interface box and conveniently connect you Mahr measuring
instrument to a CAQ software, you must firstly install the actual MarCom Professional 5.0. The driver for the
virtual interface box is automatically installed when installing MarCom.

Please proceed as follows:

1. When starting MarCom, the small status window appears on the screen, to open the configuration window
double click on the Mahr logo

2. Before you connect the measuring instrument to the virtual interface box, you must beforehand select and
activate a virtual interface box. To do this open the “Emulation of Interface Box” menu under “Settings”.

3. In the following dialog window you can activate up to 4 virtual interface boxes. These virtual interface boxes
must be allocated an interface (e.g. COM 3) and a box type (e.g. Mux50 Mahr-Box), and subsequently the
appropriate check box must be activated. Furthermore the allocation of a foot switch for data request to
each Interface Box can also be conducted in this window.

28 Mahr GmbH • MarCom Standard / Professional 5.0© Software




In the configuration window the individual measuring instruments can be selected which will be subsequently
the transfer to in the virtual interface box, the individual measuring instruments are thus virtually connected
to the virtual interface box (see following). Therefore; you can connect a USB or an integrated wireless
measuring instrument simply to a third party software, for example, to a CAQ program, as if you are
connected via an interface box (Multiplexer).

4. Please proceed as explained in Chapters “5 Parameterize Measuring Instruments” and 6 “Connecting


RF Devices” with the connection of your measuring instruments (e.g. integrated wireless measuring
instruments, USB data cable, MC-I foot switch, etc.).

5. To connect individual measuring instruments to the inputs on the virtual interface box, please open the
configuration window of the appropriate measuring instrument.
In this window please select as “Transfer to” the Virtual Interface Box instead of the standard setting;
Excel.

In the pull down menu select Virtual Interface Box, which you have previously activated in “Settings – Virtual
Interface Box.” Subsequently use the mouse to click on a free connection port number in order to virtually
connect the measuring instrument to the interface box. The port connection port number for the measuring
instrument is marked with an X in the green box.

Mahr GmbH • MarCom Standard / Professional 5.0© Software 29




If the “Select port number” is allocated to another measuring instrument the box will appear in dark gray and
cannot be selected.

30 Mahr GmbH • MarCom Standard / Professional 5.0© Software




6. Once the “Select port number” has been selected, close the configuration window with OK and
continue with the next measuring instrument. When all your measuring instruments have been virtually
connected to the interface close both configuration windows.

Note: To transfer measured values, the virtual interface box must be activated and the small
status window is active.

In order that you do not have to manually start the MarCom software after every new start or reboot of the
PC, you can add MarCom to the “Windows Autostart”, to do this activate the Autostart in the Settings menu
(please refer to Chapter 3.10 Autostart).

7. You have now successfully completed the connection of your measuring instruments to the virtual
interface box Mux-50 and activated the Box (in this example) to the COM 3 interface.
Subsequently you must assign your CAQ software to the new interface box.

In the following screenshots show an example from qs-Stat to Q-Das.

8. In qs-Stat select the Characteristics Type of the Input Type of a Mux-50 compatible interface box (e.g.
Bobe M (Mux), Brecht Mux 8, Mitutoyo Mux50 etc.).

Mahr GmbH • MarCom Standard / Professional 5.0© Software 31




9. Once the Input Type has been selected (in this example a Bobe M (Mux), select the device settings for
this type. Then select the interface for the interface box (in this example COM 3) and a
channel (Port Connection No.) of the connection measuring instrument.

Ill: qs-Stat

Ill: MarCom

10. Finally check the configuration of the interface settings, usually the default settings can be used.

11. The configuration of the virtual interface box and the CAQ software is now completed and you can now
begin your measurements.

32 Mahr GmbH • MarCom Standard / Professional 5.0© Software




8. Overview of the Status and Configuration Windows

Status window

Connect /
disconnect all
RF-Devices
Last sent measured
value is displayed

Global Timer Reset the


(Start/Stop) measuring cycle

Configuration window
Description of the Description of Last sent Destination file and
connected measuring the measuring measured path to where the
instrument task value values are to be sent

Connect/
disconnect
RF-Device

Connected USB plug-in Description of Request of the Data Request via. To set a
RF-Device position the connected measured value (e.g. switch or Indicator to
data cable via the Timer keyboard - button zero
Individual (Start/Stop) F1)
instrument
number

Mahr GmbH • MarCom Standard / Professional 5.0© Software 33




9. Implementing a Remote Control


F5

Page up Page down

No function

The Remote Control (4102221) no external driver is required. Once a USB receiver has be connected
Windows own driver will be installed. After the this driver has be installed you can press a button on the
remote control which simulates a keyboard command, see above illustration.
With MarCom these keyboard commands can be assigned functions.

In the shortlist you can select functions for each button on the remote control. Please be aware the function
buttons “Page up” and “Page down” are blocked for all other applications, once the MarCom is exited these
functions button resume their original function.

34 Mahr GmbH • MarCom Standard / Professional 5.0© Software




10. RS232 Settings on a measuring instrument

With measuring instruments that have a free configurable RS232 interface and cannot be directly connected
to a Serial COM port or a USB port with a USB data cable require a RS232-USB adapter cable. The RS232
parameter must be set up according to the pre-configured MarCom Software, in the majority of cases this is
the standard setting of the measuring instrument, nevertheless these should also be checked.
As to set up these parameters on the appropriate measuring instrument, please refer to the operating
instruments of the measuring instrument.

If an adapter cable is used, the original RS232 data cable of the measuring instrument is required (see Order
no. in brackets).

The settings are as follows:

Adapter cable Millimar USB (for Millimar 1240, C1208, C1210, C1216, C1245, S1840 etc.)
Order no.: 4102331 + (7024634)

Millimar Millitron 1240


Protocol  M1240 R/D  COMPUTER
Format  8 N 1 Baud  9600
Handshake  kein Prot  —
Baud  9600 Interface  RS232C
Char  8n1

Adapter cable 817 USB (for Digimar CX1, CX2, 817 CLM etc.)
Order no.: 4102333 + (7024634)

817 CLM / MarCheck


Menü  13. Data and printer  Interface RS232 Out  User RS232
Baudrate  9600
Data format  1. No parity 8 Bits
Interface mode  1. Handshake ON (CTS)
CX1
Channel  5
Baudrate  9600
Data format  1. No parity 8 Bits
Interface mode  1. Computer or 4. Data manual transmit
Interface mode  1. Computer requires CTS
CX2
FORMAT  8-N-1
HANDSH  NONE
BAUD  9600

Adapter cable MSP 2 USB


Order no.: 4102334 + (4102711)

MSP 2 QuadraChek
Transmit value/ Send meas. list Baud  4800
Baud  4800 FORMAT  7-E-2
Dat  7
Par  E
Stp  2
Close  CR

Adapter cable Millimar 832 USB


Order no.: on Request

Baud  4800

Mahr GmbH • MarCom Standard / Professional 5.0© Software 35




Notes

........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................

36 Mahr GmbH • MarCom Standard / Professional 5.0© Software




Notes

........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................
........................................................................................................................................................

Mahr GmbH • MarCom Standard / Professional 5.0© Software 37


© by Mahr GmbH
We reserve the right to make changes
to our products, especially due to
technical improvements and further
developments.
All illustrations and technical data are
therefore without guarantee.