Multiprobe Temperature Verification Kit 4460108A
Multiprobe Temperature Verification Kit 4460108A
Verification Kit
for use with Veriti® Thermal Cycler
User Guide
For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use.
Information in this document is subject to change without notice.
APPLIED BIOSYSTEMS DISCLAIMS ALL WARRANTIES WITH RESPECT TO THIS DOCUMENT, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO
THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TO THE FULLEST EXTENT ALLOWED BY LAW, IN NO EVENT SHALL APPLIED
BIOSYSTEMS BE LIABLE, WHETHER IN CONTRACT, TORT, WARRANTY, OR UNDER ANY STATUTE OR ON ANY OTHER BASIS FOR SPECIAL, INCIDENTAL,
INDIRECT, PUNITIVE, MULTIPLE OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING BUT NOT
LIMITED TO THE USE THEREOF, WHETHER OR NOT FORESEEABLE AND WHETHER OR NOT APPLIED BIOSYSTEMS IS ADVISED OF THE POSSIBILITY OF
SUCH DAMAGES.
TRADEMARKS
The trademarks mentioned herein are the property of Life Technologies Corporation or their respective owners.
Appendix B Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Instrumentation safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Symbols on instruments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
General instrument safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Safety information
Note: For general safety information, see this Preface and
Appendix B, “Safety” on page 51. When a hazard symbol and hazard
type appear by a chemical name or instrument hazard, see the
“Safety” Appendix for the complete alert on the chemical or
instrument.
Safety alert Four safety alert words appear in Applied Biosystems user
words documentation at points in the document where you need to be aware
of relevant hazards. Each alert word—IMPORTANT, CAUTION,
WARNING, DANGER—implies a particular level of observation or
action, as defined below:
Overview
The Multiprobe Temperature Verification Kits (TVKs) are for use
with the Veriti® Thermal Cycler, as described in the Veriti Thermal
Cycler User Guide (PN 4375799).
The Multiprobe TVKs verify the temperature accuracy and non-
uniformity of Applied Biosystems thermal cyclers. The Multiprobe
TVKs must be used only by operators trained in laboratory
techniques and procedures.
The customer is responsible for validation of assays and compliance
with regulatory requirements that pertain to their procedures and uses
of the instrument.
About this Guide This guide describes how to assemble and use the Multiprobe TVKs.
Annual Use the TVK on the sample block(s) to perform the following tests at
Temperature least once a year:
Verification Tests • Temperature Verification Test – Checks the temperature of the
sample blocks against specifications for temperature accuracy.
• Temperature Non-Uniformity Test (TNU) – Checks the
temperature uniformity of the sample blocks.
• Heated Cover Test – Checks the heated cover temperature
accuracy.
Testing Applied Biosystems recommends that you perform each of the three
Frequency tests at least once a year, then as frequently as necessary, depending
on the type of work your laboratory performs.
Based on customer application, the Veriti® Thermal Cycler may
require more frequent temperature testing. The customer is
responsible for determining the needed testing frequency.
TVK
Parameter Specification
Specifications
Accuracy Accuracy of well probes:
• ±0.1 °C (35 °C to 96 °C)
• ±0.2 °C (0 °C to 100 °C)
Accuracy of heated cover probes:
• ±1.0 °C (100 °C to 110 °C)
• ±1.5 °C (35 °C to 125 °C)
Resolution ±0.01 °C
Range
• Well probes, location 1 to 8 • 0.00 °C to 100 °C
• Cover probe location 9 • 35.0 °C to 115.0 °C
,
Environmental conditions
• Temperature • 18 °C to 28 °C
• Relative humidity • 0 to 75%
Storage
• Temperature • 10 to 40 °C
• Relative humidity • 0 to 70%
Multiprobe Modules
IMPORTANT! Use only the multiprobe module that is designed for
the sample block you are testing. See pages 10 and 11 for the TVKs
and multiprobe module part numbers.
Each multiprobe module consists of 8 temperature probes that
measure the temperature of 8 sample wells. Position 9 measures the
heated cover temperature.
IMPORTANT! Applied Biosystems recommends that you use the
multiprobe module and digital thermometer to perform only the
Temperature Verification Test, Temperature Non-Uniformity Test,
and the Heated Cover Test.
Multiprobe
To test the... Use the... Kit PN
Modules for the
Veriti® 96-Well 96-Well Fast Thermal 9-Channel Temperature 4373997
Sample Blocks Cycler, with 0.1-mL sample Verification Kit for VeriFlex™ 0.1-
wells mL 96-Well Sample Blocks
Ribbon cable
Multiprobe
To test the... Use the... Kit PN
Modules for
Veriti® 384-Well 384-Well Thermal Cycler with 9-Channel Temperature 4385338
Sample Blocks 0.02-mL sample wells Verification Kit for 0.02-mL
384-Well Sample Block
Heated cover
temperature test
,
point
Ribbon cable
Digital Thermometer
Front View For additional information, see “TVK Specifications” on page 9.
TNU
AVG
CHAN
Keypad Buttons
Button Function
Top View The top view shows the ports for cable connections:
Operating When you use the digital thermometer to perform the tests in this
Guidelines guide:
• Ensure that the multiprobe module is connected by the ribbon
cable to the input connector at the top of the thermometer.
IMPORTANT! Use only the multiprobe module that is designed
for the sample block(s) you are testing. See pages 10 and 11 for
the TVKs and multiprobe module part numbers.
Connect the 1. Connect the free end of the ribbon cable (from the multiprobe
Digital module) to the input connector port on the digital thermometer.
Thermometer IMPORTANT! Use only the multiprobe module that is designed
for the sample block(s) you are testing. See pages 10 and 11 for
the TVKs and multiprobe module part numbers.
2. Connect the round 8-pin DIN receptacle end of the
communication cable to the thermometer communication port.
Ribbon cable
Communication cable
TVK Serial
Communications
port
Set up the 1. If the heated cover is closed, lift the lever and place the heated
Multiprobe cover in the fully opened position.
Module 2. Place the multiprobe module so that tip #1 sits in well A1.
A1
,
3. Make sure that the ribbon cable leads straight out from the
module, lies flat, and is under the thermal cycler cutout.
Test Pattern of The test wells (gray) are in the six sample blocks. Each sample block
96-Well Sample consists of two columns.
Blocks
A1 A3 A5 A7 A9 A11
C1 C3 C5 C7 C9 C11
E1 E3 E5 E7 E9 E11
H6 H8
,
H2 H4 H10 H12
1 2 3 4 5 6
Run the Test 1. See “Before You Run 96-Well Instrument Tests” on page 16.
2. Power on the thermal cycler, then power on the digital
thermometer.
3. Touch Main MenuTools MenuRun Temperature
Verification.
Testing Zones
Well Position of Probe Tip #1
(Sample Blocks)
1 and 2 A1
3 and 4 A5
5 and 6 A9
Print Test Results You must have a printer connected and configured as described in the
Veriti® Thermal Cycler User Guide.
Press Print to print results.
Save Test Results 1. Connect your USB memory stick to the USB port on the front of
the thermal cycler.
USB port
Run the Test 1. See “Before You Run 96-Well Instrument Tests” on page 16.
2. Power on the thermal cycler, then power on the digital
thermometer.
3. Touch Tools MenuRun TNU Test.
Testing Zones
Well Position of Probe Tip #1
(Sample Blocks)
1 and 2 A1
3 and 4 A5
5 and 6 A9
Print Test Results You must have a printer connected and configured as described in the
Veriti® Thermal Cycler User Guide.
Press Print to print results.
Save Test Results 1. Connect your USB memory stick to the USB port on the front of
the thermal cycler.
USB port
Run the Test 1. See “Before You Run 96-Well Instrument Tests” on page 16.
2. Power on the thermal cycler, then power on the digital
thermometer.
3. Touch Tools MenuHeated Cover Verification.
Print Test Results You must have a printer connected and configured as described in the
Veriti® Thermal Cycler User Guide.
Press Print to print results.
Save Test Results 1. Connect your USB memory stick to the USB port on the front of
the thermal cycler.
USB port
Set Up the 1. Connect the multiprobe module to the digital thermometer. See
Multiprobe page 16 for instructions.
Module 2. If the heated cover is closed, lift the lever and place the heated
cover in the fully opened position.
3. Place the multiprobe module so that tip #7 sits in well P1.
Tip #7
P1 Well
4. Ensure that the ribbon cable leads straight out, lies flat, and is
under the thermal cycler cutout.
5. Slide the heated cover forward, then pull down the lever.
IMPORTANT! Do not crush the multiprobe module ribbon cable
when the heated cover is closed.
,
Run the Test 1. Power on the thermal cycler, then power on the digital
thermometer.
2. Touch Main Menu Tools Menu Run Temperature
Verification.
,
The first screen opens.
Print Test Results You must have a printer connected and configured as described in the
Veriti® Thermal Cycler User Guide.
Press Print to print results.
Save Test Results 1. Connect your USB memory stick to the USB port on the front of
the thermal cycler.
USB port
Run the Test 1. See “Before You Run 384-Well Instrument Tests” on page 32.
2. Power on the thermal cycler, then power on the digital
thermometer.
3. Touch Tools MenuRun TNU Test.
Print Test Results You must have a printer connected and configured as described in the
Veriti® Thermal Cycler User Guide.
Press Print to print results.
Save Test Results 1. Connect your USB memory stick to the USB port on the front of
the thermal cycler.
USB port
Run the Test 1. See “Before You Run 384-Well Instrument Tests” on page 32.
2. Power on the thermal cycler, then power on the digital
thermometer.
3. Touch Tools MenuRun Heated Cover Verification.
Print Test Results You must have a printer connected and configured as described in the
Veriti® Thermal Cycler User Guide.
Press Print to print results.
Save Test Results 1. Connect your USB memory stick to the USB port on the front of
the thermal cycler.
USB port
Decontaminate 1. Gently wipe the probe tips with a cotton swab that has been
the Multiprobe moistened with bleach solution (10% bleach, 90% water).
Module IMPORTANT! Do not submerge the probe in bleach solution. Do
not disassemble the multiprobe module.
2. Complete a Decontamination Certificate (see page 49) to
prepare for recalibration.
Clean the LCD Use only a commercially available plastic-compatible lens cleaner.
Lens
Chemicals such as alcohol, acetone, and lacquer thinner damage the
lens, the case, and the rubber key pad.
Replace the Battery life is ~20 hours when not connected to an instrument and
Battery ~15 hours when connected to an instrument communication port
(RS-232 interface).
,
To replace the battery:
1. Place the thermometer front side down.
2. Remove the battery door at the bottom of the back side.
Recalibration Services
TVK Support A temperature verification kit is available for use to complete a
customer-performed installation qualification protocol or annual user
maintenance activities. Applied Biosystems recommends that you
have the digital thermometer and multiprobe module calibrated once
a year.
USA
• Fax: (949)-271-2300
2. Ensure that the probe assembly has been decontaminated (see
“Decontaminate the Multiprobe Module” on page 48).
3. Pack the digital thermometer and the probe assembly in the
black case.
4. Create a shipping package with:
Symbols on instruments
Symbol Description
Operating the Ensure that anyone who operates the instrument has:
instrument • Received instructions in both general safety practices for
laboratories and specific safety practices for the instrument.
• Read and understood all applicable Safety Data Sheets (SDSs).
Related documentation
Instruments that use the Multiprobe Temperature Verification Kits
covered in this guide are supported by the following documents:
Document
Instrument
Part Number
Obtaining support
Support information
For the latest services and support information for all locations, go to
www.appliedbiosystems.com, then click the link for Support.
At the Support page, you can:
• Access worldwide telephone and fax numbers to contact
Applied Biosystems Technical Support and Sales facilities.
• Search through frequently asked questions (FAQs)
• Submit a question directly to Technical Support
• Order Applied Biosystems user documents, SDSs (Safety Data
Sheets), certificates of analysis, and other related documents
• Download PDF documents
• Obtain information about customer training
decontaminate 48
description 12 serial number 13
setting up 16, 17, 33
display
test pattern 19
temperature differences 14
TNU/AVG 13
documentation, related 53 O
ON/OFF button 13
E
ending the test 23, 27, 31, 37, 41, 45 P
part numbers, TVK, supported 10
H preparing for tests 16, 32
hazard icons. See safety symbols, on
DRAFT
instruments 51 R
hazard symbols. See safety symbols, on recalibration
instruments frequency 8
T
Temperature Non-Uniformity Test 24, 38
temperature verification kit
instructions for shipping 49
Temperature Verification Kits
components 8
testing frequency 9
Temperature Verification Test 19, 34
test results, saving 22, 26, 30, 36, 40, 44
test wells, tested 19
tests
ending 23, 27, 31, 37, 41, 45
frequency 9
pattern, 96-well plate 19
preparing 16, 32
running 20, 24, 28, 34, 38, 42
types 9
TNU/AVG button 13
training, information on 53
TVK
battery, replacing 48
maintaining 48
part numbers 10
U
upgrade kit, PNs 10
W
WARNING, description 5