Rds Med User Manual V 1 0 1
Rds Med User Manual V 1 0 1
USER MANUAL
RDS-Med™ Survey
RDS-MED™ Meter
SURVEY
METER
NOTE
This document and the information herewith are copyrighted material as mentioned in the
Copyrights Act. No part of this document may be copied without a written authorization from
the manufacturer. The manufacturer withholds the right to make changes in the contents of
this document without prior notice. If you have any comments or suggestions for additions
concerning either the products or this document, please submit your feedback as instructed
at the end of the manual.
ATTENTION
The RDS-Med Radiation Survey Meter does not contain any hazardous or
dangerous substances and can be recycled accordingly. The batteries of the
device must be recycled separately as instructed by the manufacturer of the batteries.
1 INTRODUCTION ............................................................................. 1
2 GETTING STARTED ....................................................................... 2
2.1 OVERVIEW OF THE METER ............................................................................ 2
2.2 INSERTING BATTERIES .................................................................................. 3
2.3 POWER ON....................................................................................................... 3
2.4 POWER OFF ..................................................................................................... 4
2.5 MAIN DISPLAY.................................................................................................. 4
2.6 DISPLAY SYMBOLS ......................................................................................... 5
2.7 KEYPAD USE .................................................................................................... 7
2.8 KEYPAD DIRECT FUNCTIONS ........................................................................ 7
2.9 CONNECTING EXTERNAL PROBES ............................................................... 8
2.10 PARAMETER SETTINGS .................................................................................. 9
3 FUNCTIONALITY .......................................................................... 10
3.1 POWER OFF ................................................................................................... 11
3.2 CUMULATIVE DOSE....................................................................................... 11
3.3 MAXIMUM DOSE RATE RECORD.................................................................. 13
3.4 CHIRP FUNCTION .......................................................................................... 13
3.5 ALARM FUNCTIONS....................................................................................... 14
3.5.1 DOSE RATE ALARM ................................................................................... 14
3.5.2 DOSE ALARM ............................................................................................. 17
3.5.3 EXTERNAL ALARM ..................................................................................... 18
3.6 DIAGNOSTICS ................................................................................................ 19
3.6.1 FIRMWARE VERSION ................................................................................ 19
3.6.2 SELF TEST ROUTINE ................................................................................. 20
3.6.3 CALIBRATION CHECK................................................................................ 20
3.7 DISPLAY ......................................................................................................... 21
3.8 COMMUNICATION .......................................................................................... 21
3.8.1 NET WRM.................................................................................................... 21
3.8.2 BLE .............................................................................................................. 21
3.9 SCALER .......................................................................................................... 22
3.9.1 SCALER FUNCTION INTERNAL DETECTOR/DOSE RATE PROBE .......... 22
3.9.2 SCALER FUNCTION WITH OTHER EXTERNAL PROBES......................... 23
3.10 EXTERNAL PROBES ...................................................................................... 26
3.10.1 CSP (CANBERRA SMART PROBE) SERIES ............................................ 26
3.10.2 OTHER PROBES WITH INTERNAL MEMORY ......................................... 27
3.10.3 PROBES WITHOUT INTERNAL MEMORY ............................................... 28
4 ERROR CODES AND OTHER DISPLAY MESSAGES ................. 29
4.1 LOW BATTERY WARNING ............................................................................. 29
4.2 LOW BATTERY ALARM .................................................................................. 29
4.3 DOSE RATE / DOSE WARNING (BLINKING DISPLAY) ................................. 29
4.4 DOSE RATE / DOSE ALARM (BLINKING DISPLAY) ...................................... 29
4.5 DOSE RATE OVERFLOW ALARM (OFL) ....................................................... 30
4.6 DEFECT ERRORS .......................................................................................... 30
5 MAINTENANCE AND DECONTAMINATION ................................ 31
5.1 FIRMWARE UPDATE ...................................................................................... 31
5.2 BATTERY COVER AND SEAL CHANGE ........................................................ 31
5.3 CALIBRATION ................................................................................................. 32
5.4 MECHANICAL DECONTAMINATION.............................................................. 32
5.5 CHEMICAL DECONTAMINATION................................................................... 32
6 SPECIFICATIONS ......................................................................... 33
7 ACCESSORIES AND PROBES .................................................... 35
7.1 ACCESSORIES ............................................................................................... 35
7.2 COMPATIBLE PROBES GMP/CSP SERIES ................................................... 36
8 FCC COMPLIANCE STATEMENT ................................................ 37
9 ICC COMPLIANCE STATEMENT ................................................. 38
FEEDBACK ......................................................................................... 39
INTRODUCTION
1 INTRODUCTION
The RDS-Med Radiation Survey Meter offers precise monitoring of ionizing radiation,
ensuring the safety of medical professionals and patients. Versatile and user-friendly, the
RDS-Med survey meter is ideal for healthcare applications such as radioisotope production,
radiopharmacies, nuclear medicine, diagnostic imaging, hot labs, and waste disposal.
Designed to meet stringent health and safety standards, the RDS-Med survey meter is
crucial for medical environments using radiation in diagnosis and treatment.
A clear display and straightforward user interface enable quick assessment of radiation
level. Ergonomic, lightweight design, and an included battery cover with belt clip make the
meter workflow-friendly.
The RDS-Med meter can detect a wide range of ionizing radiation, including gamma and X-
rays, with the ability to connect to external alpha, beta, and neutron probes. Mirion GMP-
12/GMP-25 probes and the full CSP™ probe range can be connected with an adequate
cable.
With both warning and alarm levels, users can know quickly if contamination or dose rates
exceed screening levels for decontamination or patient monitoring.
The RDS-Med is part of the Mirion Technologies RDS-32 survey meter family.
Visit mirion.com to learn more!
1
GETTING STARTED
2 GETTING STARTED
The detector position is indicated with markings on the meter. The triangles on
the meter sides indicate the reference line and the dot on the back of the meter
indicates the GM-tube center point.
2
GETTING STARTED
USB-RDS Cable Link (Part no. 1233-333) can be used to charge NiMH batteries (power
supply max. output power 15 W). Cradle for RDS (Part no. 1241-251) includes power supply
and a wall mounting kit. The cradle can be mounted to wall using the screws provided in the
kit and it is also compatible with most camera mounts.
WARNING! Risk of explosion if the battery is replaced with an incorrect type of battery.
2.3 POWER ON
POWER ON the device by pressing the Select key for two seconds.
The firmware version, time and calibration status are first shown on the display. If the time of
meter’s Real Time Clock is not set, the flashing ‘TIME NOT SET’ message is displayed. If
necessary, the Real Time Clock can be reset with the CSW-Med Configuration Software.
When switched on the meter performs the following self-test functions:
After the initialization phase is complete, the Main display with dose rate will be shown.
3
GETTING STARTED
If the select key is released before the countdown is complete, the device will return to the
menu display.
The two example Main displays above show the measured dose rate, Sv unit with multiplier,
battery capacity, analog dose rate indicator and different states of audio setting: chirp enabled
(left), chirp disabled and key sounds disabled (right).
The basic instrument RDS-Med has two units: Sv-based and rem-based. A suitable sub-
multiplier ‘µ’ or ‘m’ is set to extend the display range. In addition, ‘/h’ is applied to indicate the
dose rate.
With external gamma dose rate probes, the same basic units can be used (Sv/h, Sv, rem/h
and rem). With other external probes, the display can be configured to units cps, cpm, dpm,
Bq or Bq/cm2. Which units are available will vary by probe type. The external probe unit in use
is indicated above the reading.
4
GETTING STARTED
When using external probes, the RDS-Med also displays the dose rate from the internal
detector of the instrument and integrates accumulated dose. The RDS-Med internal detector
dose rate is displayed on the bottom left corner of the Main display. The dose rate and dose
alarms/warnings set for the RDS-Med detector are operational.
Example of the Main display with Example of the Main display with
external dose rate probe connected. alpha/beta probe connected.
NOTE: If speaker and chirp symbols are not visible, the device is muted. Some of
the above settings are only available in the CSW-Med Configuration Software.
5
GETTING STARTED
6
GETTING STARTED
7
GETTING STARTED
Most probes are detected automatically. The display will show CSP.INIT or the symbol
while the connection is being established. For probes with an internal memory (CSP probes,
GMP-25, and GMP-12 series except for GMP-12SD) the radiation detection starts without any
need for additional input. For a probe without internal memory (GMP-12SD) the meter will
prompt to select a detector type before measurement begins. For detailed probe settings by
probe type, see Chapter 3.10 External Probes.
With CSP probes we recommend to always turn the meter off when connecting the
probe to ensure correct identification and parameter reading. With the wide
selection of CSP probes you must observe that correct probe type is recognized
when you switch between different types; re-starting the meter is the safest
practice.
8
GETTING STARTED
9
FUNCTIONALITY
3 FUNCTIONALITY
A quick guide to the menu items is provided below. For detailed operation, see the individual
chapters for each item.
The lightbulb symbol will appear from time to time in this user manual to indicate
a helpful tip for the user.
Check and reset the current maximum recorded dose rate since last
MaxDR
reset.
CHIRP Switch ON/OFF the visual and audible pulse indication.
Set dose and dose rate alarm levels and choose the alarm mode:
ALARM
single, dual or off
10
FUNCTIONALITY
When powered off, the instrument does not completely shut down all the internal circuitries.
The RTC-circuit remains active maintaining the time signal. The energy is supplied with the
batteries. The time setting of the instrument RTC is lost in case the batteries are removed for
more than 3 hours. The time can be reset using the CSW-Med Configuration Software.
11
FUNCTIONALITY
To reset the accumulated dose, navigate to DOSE in the menu, press the Select key once so
that the RESET function starts to blink. Press the Select key a second time to see the
confirmation phase SURE? Then press the Select key a second time to perform the reset.
Cancel the dose reset process by pressing the left arrow key.
The RDS-Med collects dose continuously while switched on and saves that value
in the meter memory even if the meter is switched off/on. To follow accumulated
dose by task or mission, reset the dose as indicated above.
Use the DOSE key to quickly check accumulated dose from the Main display.
12
FUNCTIONALITY
The RDS-Med records the measured maximum dose rate while switched on and
saves that measured value even when the meter is switched off/on. To follow the
maximum dose rate by task or mission, reset the record as indicated above.
Use the right arrow key in the Main display to mute or unmute the audible chirp
and key sounds.
13
FUNCTIONALITY
The alarm mode setting allows the user to select the desired alarm functionality. The user can
select to apply a single alarm level only, a dual alarm with an additional warning level or switch
the alarm completely off. The default mode is single alarm where only one alarm level is in
use. The factory set alarm level is 50 µSv/h for dose rate and 500 µSv for accumulated dose.
If the dual mode is taken into use, the factory preset warning levels are 10 µSv/h for dose rate
and 100 µSv for accumulated dose. The alarm levels can be freely set by navigating further
into the menu of each setting as described in the coming chapters.
By default, the activation of an alarm state will activate the audible alarm, the visual alarm
(LED) and the vibration alarm. An active warning can be silenced in the Main menu by
pressing the right arrow key. This will stop the audible alarm, the vibration and the visual (LED)
alarm. If an active warning has been silenced, the display will still indicate the warning
condition. The alarm methods which are in use can be changed from the CSW-Med
Configuration Software.
• Single
• Dual (different level is set for WARNING and ALARM)
• OFF
To switch between the different alarm modes, navigate to DR Mode and set the desired mode
Dual, Single or OFF by pressing the Select key.
To set a new dose rate alarm level, navigate into Dose Rate and set a value by pressing the
up/down arrow keys and then moving right (or left) between the numbers to complete your
desired setting. When the alarm level has been entered, save by pressing the Select key.
14
FUNCTIONALITY
To set the desired alarm level it may be necessary to switch between unit multipliers. To switch
between µSv/h, mSv/h or Sv/h, navigate to the unit (furthest right on the display) and change
the selected unit multiplier with the up/down arrow keys. When the desired alarm level is
selected, save by pressing the Select key.
To ignore the changes, press the left arrow key until the display returns to the previous menu
level. Also waiting 12 seconds time out will switch the display back to Main display without
changing the alarm level.
In dual mode the user can set two levels which will trigger the device alarming functions: a
warning level and an alarm level.
To set the device in dual mode navigate into DR Mode and select Dual from the menu. To set
a dose rate warning and alarm navigate into Dose Rate. The first available setting will be Dose
Rate Warning. Set a suitable level, lower than the intended alarm level, according to the
process described above. Save the desired level by pressing the Select key. The setting will
move automatically to Dose Rate Alarm. Set a suitable dose rate alarm according to the
process described above.
15
FUNCTIONALITY
If the dose rate warning is muted from the Main display, the warning audible, vibration and
LED indication will switch off. The display will still continuously show the warning symbol in
the left upper corner and the display will blink.
During a dose rate warning the During a dose rate alarm the
Main display will blink, showing Main display will blink, showing
the measured dose rate. The the measured dose rate. The AL
W symbol remains constantly symbol remains constantly on
on screen. screen.
The configuration software offers additional control over the alarm methods. In the
software the alarm method can be set freely to any single mode or a combination
of audio/visual/vibration.
Use the dual alarm mode to set a dose rate warning level to alert the user to a rising
dose rate field which requires attention. The user is then alerted to the unusual
condition and the meter will next alert the user if the alarm level is exceeded.
16
FUNCTIONALITY
• Single
• Dual (different level is set for WARNING and ALARM)
• OFF
To switch between the different alarm modes, navigate to Dose Mode and set the desired
mode Dual, Single or OFF by pressing the Select key.
To set a new dose alarm level, navigate into Dose and set a value by pressing the up/down
arrow keys and then moving right (or left) between the numbers to complete your desired
setting. When the alarm level has been entered, save by pressing the Select key.
To set the desired alarm level, it may be necessary to switch between unit multipliers. To
switch between µSv, mSv or Sv, navigate to the unit (furthest right on the display) and change
the selected unit multiplier with the up/down arrow keys. When the desired alarm level is
selected, save by pressing the Select key.
To ignore the changes, press the left arrow key until the display returns to the previous menu
level. Also waiting 12 seconds time out will switch the display back to Main display without
changing the alarm level.
In Dual mode the user can set two levels which will trigger the device alarming functions: a
warning level and an alarm level.
To set the device in dual mode navigate into Dose Mode and select Dual from the menu. To
set a dose warning and alarm navigate into Dose. The first available setting will be Dose
Warning. Set a suitable level, lower than the intended alarm level, according to the process
described above. Save the desired level by pressing the Select key. The setting will move
automatically to Dose Alarm. Set a suitable dose alarm according to the process described
above.
If the dose warning is muted from the Main display, the warning audible, vibration and LED
indication will switch off. The display will still continuously show the warning symbol in the left
upper corner and the display will periodically indicate the dose warning level and the text
‘warning’.
17
FUNCTIONALITY
During a dose warning the dose During a dose alarm the dose
warning level will cycle on the alarm level will cycle on the Main
Main display with the measured display with the measured dose
dose rate. The W symbol rate. The AL symbol remains
remains constantly on screen. constantly on screen.
The configuration software offers additional control over the alarm methods. In
the software the alarm method can be set freely to any single mode or a
combination of audio/visual/vibration.
Use the Dual alarm mode to set an accumulated dose warning level to alert the user
to an increasing dose. The meter will next alert the user if the accumulated dose
alarm level is exceeded.
When an external probe is connected to the RDS-Med, the probe alarm setting can be
checked and changed in the EXT.AL meter menu. The external alarms available in the menu
are dependent on the connected probe. The external alarm is enabled by default.
For CSP series probes (product codes NOM*) the available alarm types and levels come from
the probe memory. These vary with probe type, check the probe manual for details. The preset
alarm levels for the CSP probes can be changed using the CSPS software.
For the GMP-12 dose rate probe series, the user can choose any of the preset 8 dose rate
alarm levels listed below.
Preset levels for GMP12 series dose rate alarm:
• OFF, 10, 50, 100, 500, 1 000, 5 000, 10 000 or 50 000 µSv/h or mrem/h.
18
FUNCTIONALITY
For contamination probes such as GMP-25 the preset units of the external alarm depend on
the unit selected in the Probe menu: cps, cpm, dpm, Bq or Bq/cm2.
• CPS: OFF, 10, 15, 20, 60, 100, 60, 1000 or 6000 cps.
• CPM: OFF, 600, 900, 1200, 3600, 6000 cpm, 36 kcpm, 60 kcpm or 360 kcpm
• DPM: OFF, 1200, 1800, 2400, 7200 dpm, 12 kdpm, 72 kpdm, 120 kdpm or 720
kdpm
• Bq: 20, 30, 40, 120, 200, 1200, 2000 Bq or 12kbq
• Bq/cm2: list will be Bq values divided by detector area
Set the probe alarm level by navigating into the EXT.AL menu and pressing the up/down
arrow keys until the desired alarm level is the active selection. Press the Select key to save
the desired alarm level.
Navigate into the Probe menu to change the probe measuring unit, if necessary. See
Chapter 3.10 External Probes for more details.
3.6 DIAGNOSTICS
The Diagnostics menu has the following features (described in following chapters):
The device firmware can be updated using the CSW-Med Configuration Software. See the
software manual for more details.
19
FUNCTIONALITY
To activate the meter selftest routine, navigate into the DIAG menu, the selftest is the active
selection and press the Select key to initiate.
The selftest routine includes checking the high voltage, the operating voltage, and the battery
capacity. During this process the user may also check the operation of all the display pixels,
the visual alarm LED, the audible alarm, and the vibration alarm.
When using NiMH batteries, set the correct battery type in the CSW-Med
configuration software. Note that with NiMH batteries the battery capacity indication
is less accurate than for alkaline batteries.
Calibration check is available under Diagnostics, the default procedure listed below.
• Calibration check expects to have 1 mSv/h (default) @ CS-137 field and checks
that the response is within ±10%.
• After starting the calibration check, the device shall be placed to 1mSv/h field in 30
seconds.
• Calibration check expects to detect 10000 pulses within 5 minutes. Calibration
check ends after the needed pulses are detected.
• Calibration error is displayed if detected dose rate is not within ±10%.
If the Calibration check is selected by mistake, wait approximately 5 minutes for the routine to
finish (CALIB.CHK, Calibration in progress). The meter will display Calibration error. Press
any button to exit the mode. The meter will restart, no changes or checks made.
The RDS-Med has a fixed calibration validity of 365 days. Performing a successful
calibration check from the DIAG menu will prolong the validity by another 365
days. The prolonging is possible for up to 3 years, after which a full calibration is
needed.
20
FUNCTIONALITY
3.7 DISPLAY
You can turn the source finder mode ON and OFF from the Display menu. The source finder
turns the Main display into a visual histogram to guide the user to find contamination and
radiation sources. The source finder is deactivated in the event of an alarm.
3.8 COMMUNICATION
The WRM functionality is used to send data packets via external data connection, for example
serial line or Local Area Network. The WRM can be set to OFF or to send messages in
intervals between two seconds to one hour.
To activate the Net WRM communication in RDS-Med (or set data send interval in RS
communication or over Local Area Network):
Navigate into the WRM menu and select a suitable send interval from the available options: 2
s, 4 s, 10 s, 20 s, 30 s, 60 s, 5 min, 10 min, 15 min, 30 min and 60 min.
Change the state to OFF to deactivate the WRM communication.
If the WRM is activated the Main display will show the symbol .
3.8.2 BLE
The device has readiness for Bluetooth communication. Contact Mirion for additional
information.
BLE communication can be configured in the BLE menu. The user can manually activate
BLE enumeration for 60 s. The user can also disable to BLE communication in this menu.
21
FUNCTIONALITY
If the BLE is activated the Main display will show the symbol .
3.9 SCALER
The Scaler functionality allows the user to set a measurement integration time and to measure
the current background. This enables net measurements. The scaler function can be used to
improve the measurement MDA (Minimum Detectable Activity).
When an external probe is connected, the scaler functionality is reserved for the external
probe.
The following chapters describe the scaler functionality for the RDS-Med internal detector or
an external dose rate probe and other external probes.
The Scaler menu of the RDS-Med internal detector or an external dose rate probe has the
following features:
• Settings
• Gross Meas.
When the desired integration time has been set, the user can then select to perform a Gross
measurement.
Initiate a scaler measurement by navigating into Gross Meas. and selecting start. The
measurement result remains on the meter display until the Select key is pressed for a new
acquisition or the left arrow key is pressed to exit the mode.
Once a scaler measurement has been initiated it can be stopped at any time by pressing the
Select key or the left arrow key. In that menu the measurement can be restarted, stopped or
aborted. If the scaler measurement is stopped, the result of the measurement will be
displayed. If the scaler measurement is aborted, the display will move to the Scaler menu
without showing a result.
22
FUNCTIONALITY
With an external probe connected, the Scaler menu has the following features:
• Settings
• Gross Meas.
• Background
• Net Meas.
The Settings menu allows the user to change the probe unit and to set a measurement
integration time or number of counts. The preset time/count levels being:
10 s, 1 min, 3 min, 6 min, 10 min, 30 min and 60 min or 100, 1000, 10000, 20000 counts.
The options available in the scaler menu will depend on probe type. For example, for the
GMP-25 probe there is a time setting available instead of time/count and the scaler menu will
show the current isotope in use.
23
FUNCTIONALITY
When the desired settings have been entered, the user can then select to perform a Gross
measurement by navigating into Gross Meas. and selecting start.
In a Net measurement the background is reduced from the measured counts. Before initiating
a Net measurement, acquire the Background by navigating into the Background menu and
selecting start.
When the background measurement is finished, pressing the Select key will take to the Net
measurement menu. This menu view will also show the current background measurement
results. Initiate a Net measurement by selecting start.
A scaler measurement result remains on the meter display until the select key is pressed for
a new acquisition or the left arrow key is pressed to exit the mode. If the Net Meas. result is
less than background it will be empty.
24
FUNCTIONALITY
Once a scaler measurement has been initiated it can be stopped at any time by pressing the
Select key or the left arrow key. In that menu the measurement can be restarted, stopped or
aborted. If the scaler measurement is stopped, the result of the measurement will be
displayed. If the scaler measurement is aborted, the display will move to the Scaler menu
without showing a result.
25
FUNCTIONALITY
Most probes are detected automatically. The display will show CSP.INIT or the symbol
while the connection is being established. For probes with an internal memory the radiation
detection starts without any need for additional input. For the few probes without internal
memory the meter will prompt to select a detector type before measurement begins. See the
below chapters for detailed instruction by probe series: CSP series, GMP-25/GMP-12 and
probes without internal memory.
The Probe menu is visible only when an external probe is connected. The probe menu
features items depending on the connected probe type.
The CSP series extends the RDS-Med capabilities with a large array of versatile probes. See
Chapter 7.2 for a probe listing. With CSP probes use of lithium batteries is recommended.
With CSP probes we recommend to always turn the meter off when connecting the probe to
ensure correct identification and parameter reading. With the wide selection of CSP probes
you must observe that correct probe type is recognized when you switch between different
types; re-starting the meter is the safest practice.
The CSP probes (part nos beginning with NOM*) have an internal memory and are
automatically recognized by the RDS-Med. The probe menu with CSP probe connected
features:
• PROBE TYPE
• PROBE UNIT
The units which are available will depend on the probe type. See the probe manual for more
details. If the probe unit is changed in the RDS-Med menu, it will be saved into the probe
memory.
To see the isotope coefficients, use the CSPS software or check the probe documentation, if
the factory calibration and isotope settings have not been changed. The isotope coefficients
can be changed in the CSPS software.
26
FUNCTIONALITY
GMP-25 is a contamination probe with internal memory. When these probes are connected
to the meter, they are automatically recognized. The probe menu features will depend on the
probe type. For example, when a GMP-25 is connected to the meter, the Probe menu features:
• PROBE TYPE
• PROBE UNIT
• PROBE Isotope
The GMP-12 probes are a series of dose rate probes with internal memory (except GMP-
12SD). These probes have an internal memory and will be automatically recognized.
However, they don’t have any settings which could be in the Probe menu, so the menu will
not be visible.
The following menu examples are with the GMP-25 pancake probe connected to the RDS-
Med.
The Probe Unit menu shows the probe measurement unit which is currently in use and
allows the user to change the unit. For the GMP-25 probe the available units are cps, cpm,
dpm, Bq and Bq/cm2. To change the measurement unit, navigate into the Probe Unit menu
and select from the list. Note that when selecting Bq or Bq/cm2, the conversion from counts
to Becquerels will be made using the isotope coefficient introduced in the next paragraph.
27
FUNCTIONALITY
The Probe Isotope menu shows the name of the current isotope coefficient in use. To view
the actual numerical coefficient or to change the isotope settings, use the CSW-Med software.
The RDS-Med reads data from the probe memory such as the unit of measurement and the
saved isotopes. For example, the default unit in the GMP-25 is cpm. Even if the unit is changed
from the meter menu, at the next power on the cpm unit will be read again from the probe
memory.
The RDS-Med is compatible with some probes which don’t have internal memory, the most
common being the GMP-12SD dose rate probe with high dose endurance.
When connecting a probe with no memory, the user will be prompted first to select the probe
type. There are three types:
• Gen.CHK.Def is for a probe with no internal memory that is tested for the pulse
generation of the external probe. If no pulses are detected for a given time, the meter
will show an error message. Select this type only when using legacy probes such as
GMP-11-3 or GMP-15-3.
• Gen.No.Def is for a probe with no internal memory that is not tested for pulse
generation. For example, probes which are alpha sensitive only, and there are no
pulses at all in the determined time window.
• GMP-12SD is dedicated to the GMP-12SD dose rate probe. When using this probe
type, select first the probe type and then the probe serial number to use the correct
probe calibration coefficients. If the serial number is not selected, probe type default
coefficients will be used. If the serial number is not available, the probe calibration
coefficients have not been saved into the RDS-Med memory. The CSW-Med
Configuration software is required to do that. See the software manual for instructions
on how to save the probe coefficients.
The probe type can be changed without switching the meter off, by navigating into the PROBE
menu and selecting another type. Selecting the wrong probe type may lead to an error
condition Det.DFE accompanied by an audible alarm, as the pulses from the probe don’t
match to the expectation and the meter detects a detector fault. In this case, switch the meter
off and select another probe type at start up.
The RDS-Med meter will remember the last selected probe type without internal
memory. When connecting a probe without memory, if the user doesn’t make a
selection in the menu when prompted, the RDS-Med will use the type selected last
which will be blinking in the display.
28
ERROR CODES AND OTHER DISPLAY MESSAGES
When necessary, the meter will give display messages regarding battery condition, current
alarms, warnings, and error situations.
29
ERROR CODES AND OTHER DISPLAY MESSAGES
30
MAINTENANCE AND DECONTAMINATION
No specific maintenance is required except for a periodic check of the calibration. Calibration
check is described in Chapter 3.6 Diagnostics.
A simple functional check of the meter can include switching it on and letting it
run though the self-check routine at start up. Leave the meter on additionally for
several minutes to check that the detector receives pulses as expected even at
background level or use a small radiation source to check that the meter
responds.
Note! If the RDS-Med becomes unresponsive during a firmware update, don’t remove
the batteries, simply try to run the firmware update again.
Instructions:
31
MAINTENANCE AND DECONTAMINATION
5.3 CALIBRATION
The CSW-Med Configuration Software is needed for the calibration of RDS-Med. Follow the
calibration procedure outlined in the software manual. A calibration check can be activated
directly from the meter menu under DIAG.
32
SPECIFICATIONS
6 SPECIFICATIONS
Complies with IEC 60846-1 standard, designed to meet ANSI 42.17A, 42.17C standards.
The device is CE and UKCA compliant.
RADIOLOGICAL CHARACTERISTICS
• Radiation detected gamma and X-rays. Alpha, beta, and neutron radiation with
external probes
• Operational quantity: ambient dose equivalent H(10)
• Geiger-Mueller (GM) radiation detector with the possibility of measuring natural
background levels
• IEC Energy Range: 48 keV to 1.8 MeV
• Dose Rate Measurement Range: 0.05 μSv/h to 100 mSv/h (5 µrem/h to 10 rem/h)
• IEC Dose Rate Measurement Range: 0.3 µSv/h to 100 mSv/h (0.03 mrem/h to 10
rem/h)
• Dose Measurement Range:1 µSv to 10 Sv (0.01 mrem to 1000 rem)
• Dose Rate Linearity: -15% to +22% 0.3 μSv/h to 0.1 Sv/h (0.03 mrem/h to 10 rem/h)
FUNCTIONAL CHARACTERISTICS
• Four navigation keys and a select key to operate the instrument
• Three keypad direct functions: Backlight, Mute and Dose and one
• user-defined shortcut
• Direct access to dose screen from keypad: level of dose in percentage of alarm level
• Configurable units: Sv(/h), rem(/h), with external detectors cps, cpm, dpm, Bq and
Bq/cm2
• Audible, visual and vibration alarm with configurable levels
• Real-Time Clock (RTC) function with 3 hrs battery back up
• Graphical LCD display; special symbols for alarm, external probe, battery,
communication, vibration alarm, chirp and mute
• Dual display in probe mode; measurements from internal and external detector
simultaneously
• Scaler/time with gross or net measurement (background deduction) for improved
statistics
ELECTRICAL CHARACTERISTICS
• Power supply: Batteries 2 x AA/LR6, alkaline, NiMH or lithium
• Operation time with fresh Alkaline batteries more than 2 months 8h use/24 h (600 h
in background radiation, Bluetooth® Low Energy disabled, 1display backlight off,
LED off)
• Operation time with fully charged NiMH batteries more than 1.5 months 8 h use/24 h
with 2900 mAh capacity (in background radiation, Bluetooth Low Energy disabled,
display backlight off, LED off)
• Contacts for external power and charging of NiMH battery
• Alarm audio level 86 dBA at 30 cm
33
SPECIFICATIONS
ENVIRONMENTAL CHARACTERISTICS
• Operating temperature: -25 °C to +60 °C (-13 °F to 140 °F)
• Storage temperature -40 °C to +70 °C (-40 °F to 158 °F)
• Relative humidity 10% to 95% at +35 °C (95 °F)
• RF-immunity: Fulfills following standards: IEC 61000-4-2 (2008), IEC 61000-4-3
(2006 +A1:2007 + A2:2010), IEC 61000-4-6 (2013), IEC 61000-4-8 (2009)
• RF Emissions: Fulfills following standards: EN 55032B
• FCC approval 2AHI8-RDS-32
• IC Approval 26167-RDS32
• IEC 60846-1
MECHANICAL CHARACTERISTICS
• Case: high impact durable glass fiber reinforced polymer;
drop tested from 1 m height on concrete floor on each side
• Ergonomic design, rubber grip around the case
• Binder-702 series connector
• Enclosure class IP67 (IEC 60529), including battery compartment
• Dimensions: 116 x 72 x 32 mm (4.57 x 2.83 x 1.26 in)
• Weight without batteries/with batteries: 160 g / 210 g (0.35 lb/0.46 lb)
• Wrist strap, battery covers with and without a belt clip
COMMUNICATION PROTOCOLS
• USB communication with suitable adapter
• Bluetooth Low Energy 4.2 protocol, Class 2 communication
The version with Sievert-units is marked with ‘S’ and the version with rem-units with ‘R’.
34
ACCESSORIES AND PROBES
7.1 ACCESSORIES
PROBE CABLES:
Part no. 1233-293 GMP-12SD/GSD/UW/GMP-25i coiled connection cable, 0,7 m - 1,6 m
Part no. 1233-295 GMP-12SD/GSD/UW/GMP-25i straight connection cable, length 20 m
Part no. 1233-318 RDS-CSP probe adapter cable, length 0,4 m
Part no. 1233-319 RDS-CSP probe straight adapter cable, length 1,5 m
Part no. 1233-320 RDS-CSP probe coiled adapter cable, length 0,7 m - 1,6 m
SOFTWARE:
Part no. 1233-345 CSW-Med Configuration Software incl. Cable Link
Part no. 1233-333 USB-RDS-Med Cable Link
OTHER ACCESSORIES:
Part no. 1233-300 Alarm Box without RDS-Med and accessories
Part no. 1233-301 Alarm Box alert kit: signaling set A (fixed on top)
Part no. 1233-302 Alarm Box alert kit: signaling set B (includes wall mounting)
Part no. 1233-303 LAN Adapter to Alarm Box
Part no. 1233-213 Telescopic Pole for RDS meters (incl. carrying strap)
Part no. 1233-311 Carrying bag for telescopic pole
Part no. NOM006819 RDS/CSP Bracket (to fix RDS meters to CSP probe body, excl. cable)
Part no. 1241-251 Cradle for RDS meters including power supply and wall mounting kit
35
ACCESSORIES AND PROBES
36
FCC COMPLIANCE STATEMENT
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference that may cause
undesired operation.
Modifications: Any modifications made to this device that are not approved by Mirion
Technologies Oy may void the authority granted to the user by the FCC to operate this
equipment.
Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
• Consult the dealer or an experienced radio/television technician for help.
37
ICC COMPLIANCE STATEMENT
38
FEEDBACK
FEEDBACK
Email: [email protected]
Tel.: +358 2 4684 600
For additional information on Mirion products and solutions please visit our
webpage www.mirion.com
39
USER MANUAL
RDS-MED™ SURVEY
METER