LifeStent IFU
LifeStent IFU
2/04-16
5 7 3 6
11
9 10
Figure 1. LIFESTENT® Vascular Stent System
CAUTION: Federal (USA) law restricts this device to sale by or on order of a physician.
This device is supplied in sterile condition. All materials inside the sterile barrier pouch (the delivery system
and stent, as shown in Figure 1, as well as the tray and pouch liner) are sterile. The external surface of the
sterile barrier pouch, as well as the product carton, should not be considered sterile.
A. Device Description
The LIFESTENT® Vascular Stent System is designed to deliver a self-expanding stent to the peripheral vasculature via a sheathed delivery
system. The LIFESTENT® Vascular Stent System is comprised of the following:
An implantable self-expanding nickel-titanium alloy (nitinol) stent (1), as shown in Figure 1 and Figure 2. The stent is a flexible, fine tubular
mesh prosthesis, with a helical design, which achieves its unconstrained diameter upon deployment into the target vessel. Upon deployment,
the stent imparts an outward radial force on the luminal surface of the vessel to establish patency. The stent has a total of 12 tantalum
radiopaque markers (Figure 2, items 1A & 1B) located on the ends of the stent (i.e., 6 at each end).
A delivery system, as shown in Figure 1, is comprised of an inner tubing assembly that contains the guidewire lumen, a stent delivery sheath
(2) and a system stability sheath (3), which are linked together by means of a handle (4). The guidewire lumen terminates distally in an
atraumatic catheter tip (5) and originates proximally in a luer hub (6) designed to accept a compatible guidewire. The self-expanding stent
(1) is constrained in the space between the guidewire lumen and stent delivery sheath. Unintended stent movement during sheath retraction
is restricted by the delivery system. The stent delivery sheath has a radiopaque zone (7) at its distal end. Prior to deployment, the shipping
lock (8) must be removed and discarded.
Refer to “Stent Deployment Procedure, Section 4. Deploy Stent” for directions on deploying the stent with the:
š 7KXPEZKHHO
š )DVW7UDFN'HSOR\PHQW/HYHU
š 5DSLG'HSOR\PHQW5LQJ
1
B. Indication for Use
The LIFESTENT® Vascular Stent System is intended to improve luminal diameter in the treatment of symptomatic de novo or restenotic lesions
up to 240 mm in length in the native superficial femoral artery (SFA) and popliteal artery with reference vessel diameters ranging from
4.0 - 6.5 mm.
C. Contraindications
The LIFESTENT® Vascular Stent System is contraindicated for use in:
š 3DWLHQWVZLWKDNQRZQK\SHUVHQVLWLYLW\WRQLWLQRO QLFNHOWLWDQLXP DQGWDQWDOXP
š 3DWLHQWVZKRFDQQRWUHFHLYHUHFRPPHQGHGDQWLSODWHOHWDQGRUDQWLFRDJXODWLRQWKHUDS\
š 3DWLHQWVZKRDUHMXGJHGWRKDYHDOHVLRQWKDWSUHYHQWVFRPSOHWHLQIJDWLRQRIDQDQJLRSODVW\EDOORRQRUSURSHUSODFHPHQWRIWKHVWHQWRU
stent delivery system.
D. Warnings
š '2127XVHLIWKHWHPSHUDWXUHH[SRVXUHLQGLFDWRU LHVTXDUHODEHOIRXQGRQWKHSRXFK LVEODFNDVWKHXQFRQVWUDLQHGVWHQWGLDPHWHU
may have been compromised. The temperature exposure indicator label should be grey and must be clearly visible on the pouch.
š 7KHLIFESTENTŪ9DVFXODU6WHQW6\VWHPLVVXSSOLHGVWHULOHDQGLVLQWHQGHGIRUVLQJOHXVHRQO\'2127UHVWHULOL]HDQGRUUHXVHWKHGHYLFH
š '2127XVHLISRXFKLVRSHQHGRUGDPDJHG
š '2127XVHWKHGHYLFHDIWHUWKHŗ8VH%\ŘGDWHVSHFLıHGRQWKHODEHO
š 3HUVRQVZLWKDOOHUJLFUHDFWLRQVWRQLFNHOWLWDQLXP QLWLQRO DOOR\PD\VXijHUDQDOOHUJLFUHVSRQVHWRWKLVLPSODQW
š '2127XVHZLWK(THIODOL™ or Lipiodol contrast media.
š '2127H[SRVHWKHGHOLYHU\V\VWHPWRRUJDQLFVROYHQWV HJDOFRKRO
š 7KHVWHQWLVQRWGHVLJQHGIRUUHSRVLWLRQLQJRUUHFDSWXULQJ
š 6WHQWLQJDFURVVDPDMRUEUDQFKFRXOGFDXVHGLĴFXOWLHVGXULQJIXWXUHGLDJQRVWLFRUWKHUDSHXWLFSURFHGXUHV
š If multiple stents are placed in an overlapping fashion, they should be of similar composition (i.e., nitinol).
š 7KHVDIHW\DQGHijHFWLYHQHVVRIVWHQWRYHUODSSLQJLQWKHPLGGOH 3 DQGGLVWDOSRSOLWHDODUWHU\ 3 KDVQRWEHHQHVWDEOLVKHG
š 7KHORQJWHUPRXWFRPHVIROORZLQJUHSHDWGLODWDWLRQRIHQGRWKHOLDOL]HGVWHQWVDUHXQNQRZQ
E. Precautions
š 7KHGHYLFHLVLQWHQGHGIRUXVHE\SK\VLFLDQVZKRKDYHUHFHLYHGDSSURSULDWHWUDLQLQJ
š 7KHGHOLYHU\V\VWHPLVQRWGHVLJQHGIRUXVHZLWKSRZHULQMHFWLRQV\VWHPV
š 5HFURVVLQJDSDUWLDOO\RUIXOO\GHSOR\HGVWHQWZLWKDGMXQFWGHYLFHVPXVWEHSHUIRUPHGZLWKFDXWLRQ
š 3ULRUWRVWHQWGHSOR\PHQWUHPRYHVODFNIURPWKHGHOLYHU\V\VWHPFDWKHWHURXWVLGHWKHSDWLHQW
š ,IH[FHVVLYHIRUFHLVIHOWGXULQJVWHQWGHSOR\PHQWGRQRWIRUFHWKHGHOLYHU\V\VWHP5HPRYHWKHGHOLYHU\V\VWHPDQGUHSODFHZLWKDQHZ
unit.
š 6WRUHLQDFRROGDUNGU\SODFH
š 'RQRWDWWHPSWWREUHDNGDPDJHRUGLVUXSWWKHVWHQWDIWHUSODFHPHQW
š &DVHVRIIUDFWXUHKDYHEHHQUHSRUWHGLQFOLQLFDOXVHRIWKHLIFESTENT® Vascular Stent. Cases of stent fracture occurred in lesions that were
moderate to severely calcified, proximal or distal to an area of stent overlap and in cases where stents experienced >10% elongation at
deployment. Therefore, care should be taken when deploying the stent as manipulation of the delivery system may, in rare instances, lead
to stent elongation and subsequent stent fracture. The long-term clinical implications of these stent fractures have not yet been
established (see section J).
š 7KHVDIHW\DQGHijHFWLYHQHVVRIWKLVGHYLFHIRUXVHLQWUHDWPHQWRILQVWHQWUHVWHQRVLVKDVQRWEHHQHVWDEOLVKHG
2
3.0 Tesla Temperature Rise
Under the scan conditions defined above, the LIFESTENT® Vascular Stent is expected to produce a maximum temperature rise in the patient of
2.7 °C after 15 minutes of continuous scanning.
Image Artifact
MR image quality may be compromised if the area of interest is in the exact same area or relatively close to the position of the stent. Artifact
tests were performed according to ASTM F2182-11a. Maximum artifact extended 3 mm beyond the stent for the spin echo sequence and
10 mm for the gradient echo sequence. The lumen was obscured.
Additional Information
The LIFESTENTŪ9DVFXODU6WHQWKDVQRWEHHQHYDOXDWHGLQ05,V\VWHPVRWKHUWKDQRU7HVOD7KHKHDWLQJHijHFWLQWKH05,HQYLURQPHQW
for fractured stents is not known. The presence of other implants or the health state of the patient may require reduction of the MRI limits
listed above.
7KH 5(6,/,(17 SLYRWDO WULDO ZDV D SURVSHFWLYH UDQGRPL]HG PXOWLFHQWHU VWXG\ GHVLJQHG WR FRPSDUH WKH VDIHW\ DQG HijHFWLYHQHVV RI WKH
LIFESTENT® Vascular Stent System to percutaneous transluminal angioplasty (PTA) in the treatment of symptomatic vascular disease of the
VXSHUıFLDOIHPRUDODUWHU\ 6)$ DQGSUR[LPDOSRSOLWHDODUWHU\VXEMHFWVZHUHUDQGRPL]HGLQDIDVKLRQEHWZHHQWKHWHVWDQGFRQWURODUP
DW86DQG(XURSHDQFHQWHUV,QWRWDOVXEMHFWVZHUHUDQGRPL]HGWRWKHWHVWDUP WUHDWPHQWZLWKWKHLIFESTENT® Vascular Stent System)
DQGVXEMHFWVZHUHUDQGRPL]HGWRWKHFRQWURODUP WUHDWPHQWZLWKVWDQGDORQHEDOORRQDQJLRSODVW\ 7KHSULPDU\VDIHW\HQGSRLQWZDVGD\
PRUWDOLW\DQGWKHSULPDU\HijHFWLYHQHVVHQGSRLQWZDVWKHPRQWKUHLQWHUYHQWLRQUDWHGD\GDWDLVDYDLODEOHIRU RIWKH
UDQGRPL]HGVXEMHFWVDQGPRQWKHijHFWLYHQHVVGDWDLVDYDLODEOHIRU RIWKHUDQGRPL]HGVXEMHFWV$OOVXEMHFWVZHUHIROORZHG
for a total of three years following the index procedure.
7KH(7$*,866VXSSRUWLQJWULDOZDVDSURVSHFWLYHQRQUDQGRPL]HGPXOWLFHQWHUVWXG\GHVLJQHGWRDVVHVVWKHDFXWHGHOLYHUDELOLW\RIWKH
LIFESTENT® and LIFESTENTŪ ;/ 9DVFXODU 6WHQW 6\VWHPV VXEMHFWV ZHUH WUHDWHG LQ (XURSHDQ FHQWHUV 7KH SULPDU\ VDIHW\ HQGSRLQW ZDV
GD\PRUWDOLW\DQGWKHSULPDU\HijHFWLYHQHVVHQGSRLQWZDVWKHDVVHVVPHQWRIVWHQWOHQJWKIROORZLQJGHSOR\PHQWGD\PRUWDOLW\GDWD
LVDYDLODEOHIRU RIWKHWUHDWHGVXEMHFWVDQGGHSOR\HGVWHQWOHQJWKGDWDLVDYDLODEOHIRUGHSOR\HGVWHQWV$OOVXEMHFWVZHUH
followed for 30 days following the index procedure.
A retrospective analysis of the performance of the LIFESTENT® Vascular Stent Systems for long-segment lesions was also undertaken.
VXEMHFWV ZHUH LQFOXGHG LQ WKH DQDO\VLV LQ ZKLFK OHVLRQV KDG OHQJWKV Ɗ PP 7KH SULPDU\ HQGSRLQWV RI WKLV DQDO\VLV ZHUH
acute safety (freedom from death, amputation or TVR) at 30-days, long-term safety (freedom from death or amputation) at 12 months
LQSDWLHQWVZLWKWRWDOOHVLRQOHQJWKVƊPPDQGHijHFWLYHQHVV IUHHGRPIURP795 DWPRQWKVLQOHVLRQVRIOHQJWKPPPPPP
200 mm and 240 mm.
7KH 5($/,7< VWXG\ D VLQJOHDUP QRQUDQGRPL]HG SURVSHFWLYH VLQJOHFHQWHU VWXG\ ZDV FRQGXFWHG WR GHPRQVWUDWH WKH VDIHW\ DQG
HijHFWLYHQHVVRIWKHBARD® LIFESTENTŪ9DVFXODU6WHQW6\VWHPVZLWKDPPGLDPHWHUVL]HRijHULQJ7KHSULPDU\REMHFWLYHRIWKLVVWXG\ZDVWR
DVVHVVWKHDFXWHHijHFWLYHQHVVRIWKHBARD® LIFESTENTŪ9DVFXODU6WHQW6\VWHPV3ULPDU\HijHFWLYHQHVVZDVGHıQHGDVVXFFHVVIXOGHSOR\PHQW
and placement accuracy based upon a rating scale completed by the investigators at time of index procedure. Primary safety was defined as
freedom from occurrence of death, amputation and TVR/TLR at 30 days post-index procedure.
$SK\VLFLDQVSRQVRUHGVWXG\WKH(7$3WULDOZDVDSURVSHFWLYHUDQGRPL]HGPXOWLFHQWHUVWXG\GHVLJQHGWRFRPSDUHWKHLIFESTENT® Vascular
Stent Systems to percutaneous transluminal angioplasty (PTA) in the treatment of patients with stenosis and occlusion of the popliteal artery.
VXEMHFWVZHUHUDQGRPL]HGEHWZHHQWKHWZRVWXG\DUPVDW(XURSHDQFHQWHUV,QWRWDOVXEMHFWVZHUHWUHDWHGZLWKWKHLIFESTENT®
9DVFXODU6WHQWDQGZLWK37$7KHSULPDU\HQGSRLQWZDVWKHUHVWHQRVLVUDWHDWPRQWKV6XEMHFWVZHUHIROORZHGIRUPRQWKV
H. Adverse Events
a. OBSERVED ADVERSE EVENTS
7KHIROORZLQJDGYHUVHHYHQWVZHUHGRFXPHQWHGGXULQJWKHFRXUVHRIWKH5(6,/,(17WULDO 1
3
RESILIENT Trial Adverse Event Summary
RESILIENT Randomized RESILIENT Feasibility
LIFESTENT® (N=134) PTA (N=72) LIFESTENT® (N=20)
Event % (N pts) [N events] % (N pts) [N events] % (N pts) [N events]
In-Hospital Events
0DMRU$GYHUVH(YHQWV 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Death 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Myocardial Infarction 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Target Limb Loss / Amputation 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
TVR 0 (0/134) [0] 41.7 (30/72) [31] 5.0 (1/20) [1]
TLR 0 (0/134) [0] 41.7 (30/72) [30] 0 (0/20) [0]
1RQ7/5 0 (0/134) [0] 1.4 (1/72) [1] 5.0 (1/20) [1]
Stroke/CVA 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
'LVWDO(PEROL]DWLRQ 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Access Site Bleeding / Hematoma 0.7 (1/134) [1] 0 (0/72) [0] 5.0 (1/20) [1]
Blood Loss requiring Transfusion 1.5 (2/134) [2] 1.4 (1/72) [1] 0 (0/20) [0]
Vessel Perforation 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Vessel Aneurysm 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Vessel Pseudo-Aneurysm 0 (0/134) [0] 1.4 (1/72) [1] 5.0 (1/20) [1]
Vessel Dissection 4.5 (6/134) [6] 20.8 (15/72) [16] 5.0 (1/20) [1]
Thrombosis 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Events at 30-Days
0DMRU$GYHUVH(YHQWV 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Death 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Myocardial Infarction 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Target Limb Loss / Amputation 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
TVR 0.7 (1/134) [2] 41.7 (30/72) [31] 5.0 (1/20) [1]
TLR 0.7 (1/134) [1] 41.7 (30/72) [30] 0 (0/20) [0]
1RQ7/5 0.7 (1/134) [1] 1.4 (1/72) [1] 5.0 (1/20) [1]
Stroke/CVA 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
'LVWDO(PEROL]DWLRQ 0 (0/134) [0] 1.4 (1/72) [1] 0 (0/20) [0]
Access Site Bleeding / Hematoma 0.7 (1/134) [1] 1.4 (1/72) [1] 5.0 (1/20) [1]
Blood Loss requiring Transfusion 1.5 (2/134) [2] 2.8 (2/72) [2] 0 (0/20) [0]
Vessel Perforation 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Vessel Aneurysm 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Vessel Pseudo-Aneurysm 0 (0/134) [0] 1.4 (1/72) [1] 5.0 (1/20) [1]
Vessel Dissection 4.5 (6/134) [6] 20.8 (15/72) [16] 5.0 (1/20) [1]
7KURPERVLV +UV'D\V2QO\ 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Events at 12-Months
0DMRU$GYHUVH(YHQWV 8.2 (11/134) [13] 6.9 (5/72) [6] 5.0 (1/20) [1]
Death 3.7 (5/134) [5] 2.8 (2/72) [2] 0 (0/20) [0]
Myocardial Infarction 4.5 (6/134) [8] 1.4 (1/72) [1] 5.0 (1/20) [1]
Target Limb Loss / Amputation 0 (0/134) [0] 4.2 (3/72) [3] 0 (0/20) [0]
TVR 16.4 (22/134) [28] 54.2 (39/72) [54] 15.0 (3/20) [3]
TLR 11.9 (16/134) [16] 54.2 (39/72) [46] 10.0 (2/20) [2]
1RQ7/5 8.2 (11/134) [12] 8.3 (6/72) [8] 5.0 (1/20) [1]
Stroke/CVA 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Vessel Aneurysm 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Vessel Pseudo-Aneurysm 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
/DWH7KURPERVLV !'D\V2QO\ 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
4
RESILIENT Trial Adverse Event Summary
RESILIENT Randomized RESILIENT Feasibility
LIFESTENT® (N=134) PTA (N=72) LIFESTENT® (N=20)
% (N pts) % (N pts) % (N pts)
Event [N events] [N events] [N events]
Events at 24-Months
0DMRU$GYHUVH(YHQWV 13.4 (18/134) [23] 11.1 (8/72) [11] 5.0 (1/20) [1]
Death 7.5 (10/134) [10] 5.6 (4/72) [4] 0 (0/20) [0]
Myocardial Infarction 6.0 (8/134) [11] 5.6 (4/72) [4] 5.0 (1/20) [1]
Target Limb Loss / Amputation 1.5 (2/134) [2] 4.2 (3/72) [3] 0 (0/20) [0]
TVR 25.4 (34/134) [48] 58.3 (42/72) [69] 15.0 (3/20) [4]
TLR 20.1 (27/134) [30] 56.9 (41/72) [53] 10.0 (2/20) [3]
1RQ7/5 12.7 (17/134) [18] 15.3 (11/72) [16] 5.0 (1/20) [1]
Stroke/CVA 0.7 (1/134) [1] 0 (0/72) [0] 0 (0/20) [0]
Vessel Aneurysm 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Vessel Pseudo-Aneurysm 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
/DWH7KURPERVLV !'D\V2QO\ 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Latest Data Available 36-Months 36-Months 36-Months
0DMRU$GYHUVH(YHQWV 15.7 (21/134) [27] 11.1 (8/72) [12] 10.0 (2/20) [2]
Death 9.0 (12/134) [12] 6.9 (5/72) [5] 0 (0/20) [0]
Myocardial Infarction 7.5 (10/134) [13] 5.6 (4/72) [4] 10.0 (2/20) [2]
Target Limb Loss / Amputation 1.5 (2/134) [2] 4.2 (3/72) [3] 0 (0/20) [0]
TVR 28.4 (38/134) [57] 58.3 (42/72) [71] 15.0 (3/20) [4]
TLR 21.6 (29/134) [35] 56.9 (41/72) [54] 10.0 (2/20) [3]
1RQ7/5 15.7 (21/134) [22] 16.7 (12/72) [17] 5.0 (1/20) [1]
Stroke/CVA 1.5 (2/134) [2] 0 (0/72) [0] 0 (0/20) [0]
Vessel Aneurysm 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
Vessel Pseudo-Aneurysm 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
/DWH7KURPERVLV !'D\V2QO\ 0 (0/134) [0] 0 (0/72) [0] 0 (0/20) [0]
7KHIROORZLQJDGYHUVHHYHQWVZHUHGRFXPHQWHGGXULQJWKHFRXUVHRIWKH(7$*,866WULDO 1
5
ETAP Trial Safety Events
P1 P2/P3
Number (%) pts Number (%) pts
p-value p-value
PTA/Stent (N=9) Stent (N=36) PTA/Stent (N=22) Stent (N=85)
Severe
Cardiovascular Events*
12 month 3 (37.5%) 8 (22.9%) 0.4010 6 (31.6%) 19 (25.7%) 0.5770
(YDOXDEOH6XEMHFWVA 1 1 1 1
24 month 4 (50.0%) 9 (31.0%) 0.4132 6 (40.0%) 22 (34.9%) 0.7689
(YDOXDEOH6XEMHFWVA 1 1 1 1
Adverse Events**
12 month 7 (77.8%) 18 (51.4%) 0.2600 13 (61.9%) 43 (56.6%) 0.8510
(YDOXDEOH6XEMHFWVA 1 1 1 1
24 month 7 (77.8%) 23 (76.7%) 1.0000 16 (80.0%) 45 (65.2%) 0.3270
(YDOXDEOH6XEMHFWVA 1 1 1 1
Death***
12 month 0 (0.0%) 1 (3.0%) 1.0000 1 (5.3%) 2 (2.7%) 0.4962
(YDOXDEOHVXEMHFWVA 8 33 19 75
24 month 0 (0.0%) 3 (11.5%) 1.0000 2 (12.5%) 4 (7.0%) 0.6065
(YDOXDEOH6XEMHFWVA 8 26 16 57
GHıQHGZLWKLQWKHUHSRUWDVLQMXU\SRLVRQLQJSURFHGXUDOFRPSOLFDWLRQV
$WOHDVWRQHRFFXUUHQFHRIDQ$(
*** P1 and P2/P3 subset compliance was not stratified at the 24 month interval due to the fact that the deaths verified from Protocol Version
1.0 could not be confirmed to specific patient ID numbers.
A (YDOXDEOHDFFRXQWVIRUPLVVLQJGDWD
6
š ,VFKHPLDLQIDUFWLRQRIWLVVXHRUJDQ
š /LYHUIDLOXUH
š /RFDOLQIHFWLRQ
š 0DOSRVLWLRQ IDLOXUHWRGHOLYHUWKHVWHQWWRWKHLQWHQGHGVLWH
š 2SHQVXUJLFDOUHSDLU
š 3DLQ
š 3DQFUHDWLWLV
š 3XOPRQDU\HPEROLVPHGHPD
š 3QHXPRWKRUD[
š 3VHXGRDQHXU\VP
š 5HQDOIDLOXUH
š 5HVSLUDWRU\DUUHVW
š 5HVWHQRVLV
š 6HSWLFHPLDEDFWHUHPLD
š 6WHQWIUDFWXUH
š 6WHQWPLJUDWLRQ
š 6WURNH
š 9DVRVSDVP
š 9HQRXVRFFOXVLRQWKURPERVLVUHPRWHIURPSXQFWXUHVLWH
š 9HQRXVRFFOXVLRQWKURPERVLVQHDUWKHSXQFWXUHVLWH
I. Clinical Studies
a. RESILIENT FEASIBILITY STUDY
7KH5(6,/,(17VWXG\LQFOXGHGDIHDVLELOLW\VWXG\WRDVVHVVWKHVDIHW\RIWKHLIFESTENT® Vascular Stent System. This feasibility study enrolled
VXEMHFWVDWVL[86LQYHVWLJDWLYHVLWHV5HVXOWVIURPWKLVVWXG\SURYLGHGMXVWLıFDWLRQIRULQLWLDWLRQRIDSLYRWDOVWXG\WRDVVHVVWKHVDIHW\DQG
HijHFWLYHQHVVRIWKHLIFESTENT® Vascular Stent System.
Continuous variables were compared using an independent samples t-test. Dichotomous variables were compared using Fisher‘s exact test.
2UGLQDOYDULDEOHVZHUHFRPSDUHGXVLQJD&KLVTXDUHWHVW7LPHWRHYHQWZDVFRPSDUHGXVLQJDORJUDQNWHVW,QWHUYDOFHQVRUHGGDWDZHUH
analyzed using the Kaplan-Meier method as the primary analysis. A sensitivity analysis for interval censored data was performed using the
:HLEXOOGLVWULEXWLRQ(ijHFWLYHQHVVHQGSRLQWVZHUHDQDO\]HGDVRQHVLGHGWHVWV6DIHW\HQGSRLQWVZHUHDQDO\]HGDVWZRVLGHGWHVWV
7
Demographics
&KDUDFWHULVWLFVRIWKHVXEMHFWVHQUROOHGLQWKHVWXG\LQFOXGLQJDJHJHQGHUPHGLFDOKLVWRU\DVZHOODVOHVLRQFKDUDFWHULVWLFVDUHSURYLGHGLQ
the tables below.
8
Methods
6XEMHFWVXQGHUZHQWHLWKHU37$RU37$SOXVLIFESTENT® Vascular Stent System placement in the target lesion(s). In cases where the PTA only
UHVXOWZDVVXERSWLPDOVWHQWSODFHPHQWZDVSHUIRUPHG7KLVRFFXUUHGLQ RIWKHVXEMHFWVWKDWZHUHUDQGRPL]HGWRWKH37$RQO\
treatment arm. Post procedure medication was suggested as aspirin for 6 months and clopidogrel for 12 weeks.
$OOGDWDZHUHFROOHFWHGRQFDVHUHSRUWIRUPVDWLQYHVWLJDWLYHVLWHV$GYHUVHHYHQWVZHUHDGMXGLFDWHGE\WKHFOLQLFDOHYHQWVFRPPLWWHHDQGWKH
data safety monitoring board routinely reviewed the study outcomes to ensure that the benefits of continuing the study outweighed any
potential risks. Independent core laboratories were utilized to analyze angiographic, x-ray and duplex imaging.
Results
$VVKRZQLQWKHSULQFLSDO6DIHW\DQG(ijHFWLYHQHVVWDEOH 6HFWLRQ- WKHLIFESTENT® Vascular Stent System demonstrated a significantly higher
freedom from intervention rate (freedom from TVR/TLR) at 6 months (LIFESTENTŪFRQWURO PRQWKV LIFESTENTŪFRQWURO
45.2% ), 24 months (LIFESTENTŪFRQWURO DQGPRQWKV LIFESTENTŪFRQWURO WKDQWKH37$FRQWUROJURXS S
$GGLWLRQDOO\DVH[SHFWHGWKHUHZDVQRGLijHUHQFHLQWKHGD\PRUWDOLW\UDWHEHWZHHQWKHWZRVWXG\DUPV
Design
7KH(7$*,866WULDOZDVDSURVSHFWLYHPXOWLFHQWHUFRQıUPDWRU\FOLQLFDOLQYHVWLJDWLRQWRHYDOXDWHWKHLIFESTENT® and LIFESTENT® XL Vascular
6WHQW6\VWHPVLQWKHWUHDWPHQWRIV\PSWRPDWLFYDVFXODUGLVHDVHRIWKH6)$DQGSUR[LPDOSRSOLWHDODUWHU\$WRWDORIVXEMHFWVZHUHWUHDWHG
DW(XURSHDQLQYHVWLJDWLYHVLWHV
Demographics
&KDUDFWHULVWLFVRIWKHVXEMHFWVHQUROOHGLQWKHVWXG\LQFOXGLQJDJHJHQGHUPHGLFDOKLVWRU\DVZHOODVOHVLRQFKDUDFWHULVWLFVDUHSURYLGHGLQ
the tables below.
9
E-TAGIUSS Trial Lesion Characteristics
1 86.5 (32/37)
1XPEHURI/HVLRQV Q1
2 13.5 (5/37)
Methods
6XEMHFWVXQGHUZHQW37$SOXVLIFESTENT® and/or LIFESTENT® XL Vascular Stent placement in the target lesion(s). Post procedure medication was
suggested as aspirin and clopidogrel for a minimum of 30 days.
$OOGDWDZHUHFROOHFWHGRQFDVHUHSRUWIRUPVDWLQYHVWLJDWLYHVLWHV$GYHUVHHYHQWVZHUHDGMXGLFDWHGE\WKHFOLQLFDOHYHQWVFRPPLWWHHDQGWKH
data safety monitoring board reviewed the study outcomes. Independent core laboratories were utilized to analyze angiographic data.
Results
$VVKRZQLQWKHSULQFLSDO6DIHW\DQG(ijHFWLYHQHVVWDEOH 6HFWLRQ- WKHLIFESTENT® and LIFESTENT® XL Vascular Stent Systems were able to
accurately deploy the stent and demonstrated minimal length change (deployment success 100.0%). Additionally, the acute safety and
HijHFWLYHQHVVPHDVXUHVGHPRQVWUDWHGSRVLWLYHUHVXOWV
d. RETROSPECTIVE ANALYSIS OF LIFESTENT® VASCULAR STENT SYSTEMS IN THE TREATMENT OF LONG-SEGMENT LESIONS
Design
7KLVVWXG\FRQVLVWHGRIDSRVWKRFDQDO\VLVRIIRXUVRXUFHVRIGDWD DSLYRWDO,'(FOLQLFDOWULDO 5(6,/,(17,'(*ŗ5(6,/,(17Ř
DPXOWLFHQWHUQRQUDQGRPL]HGREVHUYDWLRQDOVWXG\FRQGXFWHGLQ(XURSH ŗ(/2',(,Ř WKHURXWLQHFOLQLFDOSUDFWLFHRID8QLWHG6WDWHV
86 SK\VLFLDQ ŗ866HULHVŘ DQG WKHURXWLQHFOLQLFDOSUDFWLFHRID(XURSHDQ8QLRQ (8 SK\VLFLDQ ŗ(86HULHVŘ ,QWRWDOWZRKXQGUHGHLJKW\
five (285) patients with one or more implanted LIFESTENT® devices were identified and included in the analysis. There were a total of 46 lesion
segments in this analysis with lesion lengths beyond 160 mm.
Demographics
&KDUDFWHULVWLFVRIWKHVXEMHFWVDQGOHVLRQVDQDO\]HGDUHSURYLGHGLQWKHWDEOHVEHORZ
10
Demographics: Retrospective Analysis of LIFESTENT® Vascular Stent Systems in the Treatment of Long-Segment Lesions
Characteristic RESILIENT ELODIE I US Series EU Series TOTAL
Age at Procedure (years)
1UHSRUWHG 198 11 66 10 285
Mean 68.4 71.8 72.6 73.9 69.7
St Dev 10.2 8.63 10.9 5.53 10.3
Range 20.7 – 88.2 53.9 - 85.6 36.3 - 96.8 63.9 - 83.1 20.7 – 96.8
Gender (% male) 69.2 45.5 60.6 44.4 65.5
1UHSRUWHG 198 11 66 9 284
Race (% Caucasian) 88.9 100 77.3 100 86.6
1UHSRUWHG 198 3 66 10 277
Hypertension (%) 85.4 72.7 84.9 100 85.3
1UHSRUWHG 198 11 66 10 285
Hypercholesterolemia (%) 80.3 54.6 75.8 80.0 78.3
1UHSRUWHG 198 11 66 10 285
Smoking (%) 25.8 36.4 60.6 0.0 33.3
1UHSRUWHG 198 11 66 10 285
CAD (%) 56.6 27.3 57.6 30.0 54.7
1UHSRUWHG 198 11 66 10 285
DM (%) 38.9 0.00 50.0 30.0 39.7
1UHSRUWHG 198 11 66 10 285
Rutherford Category of Target Limb
1UHSRUWHG 198 11 15 10 219
Class 1 (%) 3.5 0 0 3.2
Class 2 (%) 40.4 45.5 10.0 39.3
Class 3 (%) 56.1 36.4 60.0 55.3
Class 4 (%) 0.0 0 0 0
Class 5 (%) 0.0 18.2 30.0 2.3
Indication of Target Limb
1UHSRUWHG 198 11 71 10 290
Claudication (%) 100 90.9 49.3 70.0 86.6
Critical Limb Ischemia (%) 0 9.1 50.7 30.0 13.4
ABI of Target Limb
1UHSRUWHG 183 15 51 10 244
Mean 0.72 0.61 0.41 0.69
St Dev 0.20 0.22 0.18 0.22
Range 0.24 - 1.45 0 - 1.34 0.1 - 0.67 0 – 1.45
2QHSDWLHQWGLGQRWUHSRUWJHQGHU
151RW5HSRUWHG
11
Lesion and Stent Characteristics*
Characteristic RESILIENT ELODIE I US Series EU Series TOTAL
13DWLHQWV 198 11 66 10 285
17UHDWHG/LPEV 198 11 72 10 291
17UHDWHG/HVLRQV 212 16 72 10 310
Individual Lesion Length
1UHSRUWHG 212 16 72 10 310
Mean (mm) 66.0 108.8 152.6 214.0 93.1
St Dev Length 35.7 44.7 104.5 109.6 75.1
0HDQ1SHU/LPE 1.1 1.5 1.1 1.0 1.1
Percent Stenosis (max per limb):
1UHSRUWHG 198 11 0 10 219
Mean 87.8 92.7 96.0 88.5
St Dev 11.3 9.05 6.99 11.2
Range 50 - 100 80 - 100 80 - 100 50 - 100
17RWDO/HVLRQ/HQJWKV
PP 62 1 9 0 72
şPP 93 0 19 0 112
şPP 37 6 15 3 61
şPP 5 1 3 4 13
200 – 240 mm 1 2 8 0 11
ƊPP 0 1 18 3 22
Total Lesion Lengths:
1 198 11 72 10 291
Mean 70.6 158.2 152.6 214 99.15
St Dev 37.7 57.8 104.5 109.6 77.3
Range 10 - 202 30 - 240 16 - 360 140 - 500 10 - 500
17RWDO6WHQWHG/HQJWKV
PP 40 0 15 0 40
şPP 71 0 15 0 71
şPP 73 1 15 1 75
şPP 7 7 15 5 19
şPP 5 0 15 1 6
ƊPP 2 3 15 3 8
Total Stent Lengths:
1 198 11 15 10 219
Mean 104.5 204.5 244.4 115.9
St Dev 55.4 53.2 125.1 69.4
Range 30 - 340 160 - 290 160 - 574 30- 574
TASC Classification
1*UDGH$ 1 (9.1%) 23 (39.0%) 24 (34.3%)
1*UDGH% 15 3 (27.3%) 11(18.6 %) 15 14 (20.0%)
1*UDGH& 7 (63.6%) 6 (10.2%) 13 (18.6%)
1*UDGH' 0 (0%) 19 (32.2%) 19 (27.1%)
Total 11 59 70
)RUOHVLRQFKDUDFWHULVWLFVFRUHODEGDWDZHUHXVHGZKHQDYDLODEOHWKHVLWHUHSRUWHGGDWDZHUHXVHGRWKHUZLVH)LYH SDWLHQWVGLGQRWKDYH
lesion characteristics reported by the core lab
151RW5HSRUWHG
12
Methods
6XEMHFWV UHFHLYHG DW OHDVW RQH FRPPHUFLDOO\ DYDLODEOH LIFESTENTŪ VWHQW LQ WKH FDVH RI WKRVH VXEMHFWV HQUROOHG LQ WKH 5(6,/,(17 VWXG\
,'(* WKH\UHFHLYHGWKHGHYLFHDVGHVFULEHGLQ*ZKLFKZHUHLGHQWLFDOWRWKHFXUUHQWFRPPHUFLDOO\DYDLODEOHLIFESTENT®
device. Specifically, the following analyses were undertaken:
Data for this retrospective analysis were compiled ‚as received‘ from their respective sources.
Results
The rate of freedom from death, amputation, and TVR, at 30 days post-procedure was 99.6% for the combined performance of the LIFESTENT®
and LIFESTENT® ;/ 9DVFXODU 6WHQW 6\VWHPV DQG IRU WKH 9,9$ 23& )XUWKHUPRUH ORQJWHUP VDIHW\ ZDV VKRZQ WR KDYH D FOLQLFDOO\
DFFHSWDEOHIUHHGRPIURPGHDWKDQGDPSXWDWLRQUDWHWKURXJKPRQWKV 0RUHRYHUHijHFWLYHQHVVZDVHYDOXDWHGWKURXJKHVWLPDWLRQ
of patency at 12 months post-procedure for lesion lengths of 50 mm, 100 mm, 160 mm, 200 mm and 240 mm via the lesion-length model.
The patency at 12 months for lesions greater than 160 mm in length is 67%.
e. REALITY STUDY
Design
7KH 5($/,7< VWXG\ ZDV D VLQJOHDUP QRQUDQGRPL]HG SURVSHFWLYH VLQJOH FHQWHU VWXG\ WR DVVHVV WKH GHOLYHUDELOLW\ FOLQLFDO XWLOLW\ DQG
HijHFWLYHQHVV RI WKH b PP GLDPHWHU VL]H RijHULQJ RI WKH LIFESTENT® 9DVFXODU 6WHQW 6\VWHP LQ VXEMHFWV ZLWK OLIHVW\OHOLPLWLQJ FODXGLFDWLRQ
or minor tissue loss (Rutherford Category 2 - 5) who were candidates for PTA and stenting with lesion(s) in the infra-inguinal segment
6)$DQGRUSRSOLWHDODUWHU\ $WRWDORIVXEMHFWVZHUHWUHDWHGDW(XURSHDQLQYHVWLJDWLYHVLWH
6XEMHFWVHOLJLEOHWREHHQUROOHGLQWKLVVWXG\KDGWREH5XWKHUIRUG&DWHJRU\7KHWDUJHWYHVVHOUHIHUHQFHGLDPHWHUZDV E\YLVXDOHVWLPDWH
DSSURSULDWH IRU WUHDWPHQW ZLWK DYDLODEOH VWHQW GLDPHWHU RI PP 7KH UHIHUHQFH YHVVHO GLDPHWHU 59' RI WKH WUHDWHG VXEMHFWV ZDV WR
PPLQGLDPHWHU6XEMHFWVZHUHIROORZHGDWGD\V
Demographics
&KDUDFWHULVWLFVRIWKHVXEMHFWVHQUROOHGLQWKHVWXG\LQFOXGLQJDJHJHQGHUPHGLFDOKLVWRU\DVZHOODVOHVLRQFKDUDFWHULVWLFVDUHSURYLGHGLQ
the tables below.
Subject Demographics
Variable Category Total
Mean 69
Age at Procedure (yrs)
Standard Deviation 10.5
Hyperlipidemia ~57%
Hypercholesteremia ~87%
Diabetes ~37%
13
Lesion Characteristics
Variable Category Total
1XPEHURI/HVLRQV 1 30/32
2 2/32
Left 62.5%
Target Side
Right 37.5%
Lesion Length (mm) Mean, Standard deviation 64.8, 50.0
Target Vessel (RVD) Mean, Standard Deviation 4.5, 0.2
Stenosed 68.8%
Lesion Classification 2FFOXGHG 28.1%
5H2FFOXGHG 3.1%
TASC A 43.8%
TASC B 34.4%
Lesion Severity/TASC Grade
TASC C 15.6%
TASC D 6.3%
1R&DOFLıFDWLRQ 34.4%
Mild Calcification 21.9%
Lesion Calcification
Moderate Calcification 18.8%
Severe Calcification 25.0%
Methods
6XEMHFWVXQGHUZHQW37$SOXVLIFESTENT® Vascular Stent placement in the target lesion(s). All data were collected on case report forms at the
investigative site.
Results
The LIFESTENTŪ9DVFXODU6WHQW6\VWHPLVHijHFWLYHDVWHFKQLFDOVXFFHVVZDVVKRZQLHGHSOR\PHQWDFFXUDF\ZDVJRRGRUH[FHOOHQWDQG
placement accuracy was successful at target site. Additionally, freedom from TLR and/or TVR was achieved. The LIFESTENT® Vascular Stent is
safe in the acute period (index procedure through the 30-day follow-up period) as demonstrated through freedom from occurrence of death,
DPSXWDWLRQDQG7/5DQGRU7951R$'(VZHUHUHSRUWHGGXULQJWKLVSHULRG
Design
7KH(7$3*SK\VLFLDQVSRQVRUHGVWXG\ZDVFRQGXFWHGDWQLQH(XURSHDQFHQWHUVDVDSURVSHFWLYHUDQGRPL]HGFRQWUROOHGVWXG\WRLQYHVWLJDWH
the use of LIFESTENT® Vascular Stent Systems in patients with stenosis and occlusion of the popliteal artery in comparison to percutaneous
transluminal angioplasty (PTA) alone.
A total of 246 patients were recruited and randomized into the two treatment groups, PTA or stent. 119 patients were randomized to the stent
group and 127 patients were randomized to the PTA group. For patients randomized to the PTA group, a balloon angioplasty was performed,
representing standard clinical care of these lesions. If a lesion had a residual stenosis of > 50% after repeated and persistent (5 minutes)
LQIJDWLRQVRUDIJRZOLPLWLQJGLVVHFWLRQDSURYLVLRQDOVWHQWZDVXVHGWRWUHDWWKHOHVLRQ2IWKHSDWLHQWVUHFUXLWHGLQWKHVWXG\SDWLHQWV
received a LIFESTENT® device while 93 patients received PTA alone. Results are provided to individually show the results for the P1 segment
and P2/P3 segments in order to compare the outcomes.
*Rastan A, Krankenberg H, Baumgartner I, et al. Stent placement vs. balloon angioplasty for popliteal artery treatment: Two-year results of a
SURVSHFWLYHPXOWLFHQWHUUDQGRPL]HGWULDO-(QGRYDVF7KHU
14
ETAP Trial Demographics
Characteristic (ITT population) 37$ 1 6WHQW 1 7RWDO 1
Age (years) Median 73 72 72
Range 41 - 89 42 - 89 41 - 89
*HQGHU1
Female 45 (35.4) 43 (36.1) 88 (35.8)
Male 82 (64.6) 76 (63.9) 158 (64.2)
5XWKHUIRUG&DWHJRU\1
Category 1 3 (2.4) 4 (3.4) 7 (2.8)
Category 2 12 (9.4) 24 (20.2) 36 (14.8)
Category 3 76 (59.8) 68 (57.1) 144 (58.5)
Category 4 8 (6.3) 4 (3.4) 12 (4.9)
Category 5 22 (17.3) 16 (13.4) 38 (15.4)
Category 6 - 1 (0.8) 1 (0.4)
Missing 6 (4.7) 2 (1.7) 8 (3.3)
Hypertension (%) 112 (88.2) 98 (82.4) 210 (85.4)
Hypercholesterolemia (%) 104 (81.9) 90 (75.6) 194 (78.9)
Smoking (%) 29 (23) 26 (21.8) 55 (22.4)
Methods
3DWLHQWVXQGHUZHQW37$RUVWHQWLQJDQGUHFHLYHGDFHW\OVDOLF\OLFDFLG $66LIQRWDOUHDG\RQORQJWHUPWUHDWPHQW DQGDGGLWLRQDOO\UHFHLYHG
clopidogrel before the intervention and for a minimum of 4 weeks after the intervention as long-term medication. Patients were followed for
24 months with scheduled visits after 6, 12, and 24 months.
15
Results
Patients in the stent group had a lower restenosis rate than patients in the PTA group, when the crossover procedure was considered to be
D7/5DQGE\GHıQLWLRQDUHVWHQRVLV$OVRDQDO\VLVRIVHFRQGDU\HQGSRLQWVVXJJHVWHGDEHQHıFLDOFOLQLFDOWUHQGLQIDYRURIVWHQWSODFHPHQW
however, conclusions regarding significance of individual endpoints may not be made. Provisional stent placement with a LIFESTENT® Stent
ZDVREVHUYHGGXULQJWKLVVWXG\LQRIWKHUDQGRPL]HG37$SRSXODWLRQ1RFRQFHUQLQJWUHQGVZHUHQRWHGUHJDUGLQJRYHUDOOVDIHW\ZKHQ
the LIFESTENT® Stent was compared to PTA for multiple safety endpoints.
0DMRUDGYHUVHFOLQLFDOHYHQWV 0$&( Any event of death (through 30-days), stroke, myocardial infarction, significant distal embolization,
emergent surgical revascularization of target limb, thrombosis, and/or worsening Rutherford category post procedure at the indicated time
point.
Lesion Success*: Attainment of ≤ 30% residual stenosis of the target lesion using any percutaneous method and/or non-investigational
device.
Hemodynamic Success*: Angiographic evidence of improved flow across the treated area immediately post-procedure. ABI improved from
EDVHOLQHE\ƊDQGQRWGHWHULRUDWHGE\!
16
Procedure Success*: Attainment of ≤ 30% residual stenosis of the target lesion and no in-hospital serious adverse events defined as: death,
stroke, myocardial infarction, emergent surgical revascularization, significant distal embolization in the target limb, and thrombosis of the
target vessel.
Clinical Success*: Relief or improvement of baseline symptoms by Rutherford categories/grades for acute or chronic limb ischemia and the
“definition of improvement”. Improvement must be sustained by one clinical category above the pre-treatment clinical value.
Primary Patency*: The continued flow through the target lesion as evidenced by DUS or angiogram without further/repeat intervention over
time.
Secondary Patency*: The patency history for the target lesion that is sustained or restored (with repeated intervention) over time.
Target Vessel Revascularization (TVR) / Target Lesion Revascularization (TLR): Any “clinically-driven” repeat percutaneous intervention of the
WDUJHWOHVLRQRUE\SDVVVXUJHU\RIWKHWDUJHWYHVVHO,IDFRQWUROVXEMHFWUHTXLUHVDVWHQWSHULSURFHGXUDOO\GXHWRDEDLORXWSURFHGXUHLWZLOO
be considered a TLR/TVR for the control group.
17
Survival Analysis – Freedom from Loss of Primary Patency (at 12 months)
Classification Type
1 Single-strut fracture only
2 0XOWLSOHVLQJOHVWHQWIUDFWXUHVRFFXULQJDWGLijHUHQWVLWHV
3 Multiple stent fractures resulting in complete transverse linear fracture but without stent displacement
4 Complete transverse linear fracture with stent displacement
%DVHGRQ$OOLHHWDOHQGRYDVFXODUWRGD\-XO\$XJXVW
18
* Please note that the fracture analysis in the RESILIENT Study was conducted by an independent core laboratory using the classification
system described by Allie et al., 2004 in accordance with the protocol approved in the IDE prior to study initiation (G040023, 3/19/2004).
This system classifies fractures into four distinct types. Since study initiation, other stent classification systems have been proposed
(Scheinert et al, 2005; Roca-Singh et al., 2007; Popma et al., 2009). The classification system published by Rocha-Singh et al., is currently
used by many core labs in the US, and splits the Type 4 fractures as defined by Allie et al. into “stent fracture(s) with mal-alignment of
components”(Type 4) and “stent fracture(s) in a trans-axial spiral configuration” (Type 5). The Type 4 fractures in the RESILIENT Study were
not sub-categorized according to the system proposed by Rocha-Singh and colleagues.
Conclusion
6WHQWIUDFWXUHVZHUHQRWHGWREHDQXQFRPPRQHYHQWLQWKH5(6,/,(17WULDODQGDSSHDUHGWRQRWLPSDFWWKHVDIHW\DQGSHUIRUPDQFHRIWKH
LIFESTENTŪLPSODQW6WHQWIUDFWXUHVPD\RFFXUZLWKWKHXVHRIRYHUODSSLQJVWHQWVKRZHYHUWKHUHZDVQRFRUUHODWLRQEHWZHHQVWHQWIUDFWXUHV
DQGWKHQXPEHURIVWHQWVLPSODQWHGLQWKH5(6,/,(17WULDO)UDFWXUHVPD\RFFXULQ6)$RUSRSOLWHDOVHJPHQWVWKDWXQGHUJRVLJQLıFDQWPRWLRQ
SDUWLFXODUO\LQDUHDVZLWKVHYHUHDQJXODWLRQDQGWRUWXRVLW\7KH5(6,/,(17WULDOZDVQRWGHVLJQHGWRVKRZDFRUUHODWLRQEHWZHHQVWHQWIUDFWXUHV
and the location, although six (6) fractured stents were observed in areas with severe calcification, and one (1) stent placed across the point
of flexion in the mid-popliteal region resulted in a fracture.
19
b. E-TAGIUSS CONFIRMATORY STUDY
0DMRU DGYHUVH FOLQLFDO HYHQWV 0$&( Any event of death, stroke, myocardial infarction, emergent surgical revascularization, significant
distal embolization in the target limb, amputation of the target limb and thrombosis of the target vessel at the indicated time point.
Deployment Success: Ability to deliver the stent to the intended site with the post deployment stent length being within 10% of the pre-
deployment length.
Lesion Success*: Attainment of ≤ 30% residual stenosis of the target lesion using any percutaneous method and/or non-investigational device.
Procedure Success*: Attainment of ≤ 30% residual stenosis of the target lesion and no in-hospital serious adverse events defined as: death,
stroke, myocardial infarction, emergent surgical revascularization, significant distal embolization in the target limb, and thrombosis of the
target vessel.
c. RETROSPECTIVE ANALYSIS OF LIFESTENT® VASCULAR STENT SYSTEMS IN THE TREATMENT OF LONG-SEGMENT LESIONS
7KHUHVXOWVIRUWKHSULPDU\HijHFWLYHQHVVHQGSRLQWDVGHıQHGE\IUHHGRPIURP7957/5DUHVKRZQLQWDEOHEHORZ
)URPWKH:HLEXOOFRYDULDWHDGMXVWHGDQDO\VLV
1XPEHUVWDUWLQJWKH\HDU
The primary acute safety endpoint of the LIFESTENT® and LIFESTENT® XL Vascular Stent Systems at 30 days post-procedure showed the freedom
IURPUDWHVZHUHKLJKHUWKDQWKH9L9D23& 7KHGD\IUHHGRPIURPGHDWKDPSXWDWLRQDQG795UDWHZDVZLWKDVWDQGDUGHUURU
of 0.34% (95% CI: 97.59% - 99.95%).
The primary long-term safety endpoint was freedom from death/amputation. The Kaplan-Meier analysis showed that the freedom-from-
GHDWKDPSXWDWLRQUDWHDWPRQWKVZDV OHVLRQVPP OHVLRQVPP OHVLRQVPP
(lesions 160 - 200 mm), 90.9% (lesions 200 - 240 mm) and 94.1% (lesions >240 mm).
20
Freedom from Death/Amputation*
12 months (n/N**)
All Lesions 93.8 (17/291)
/HVLRQVbPP 100% (0/72)
/HVLRQVbPP 94.5% (6/112)
/HVLRQVbPP 91.4% (5/61)
/HVLRQVbPP 63.6% (4/13)
/HVLRQVbPP 90.9% (1/11)
/HVLRQV!bPP 94.1% (1/22)
d. REALITY STUDY
7KHVDIHW\DQGHijHFWLYHQHVVUHVXOWVDUHVKRZQEHORZ
Technical Success: All 36 stents deployed in the study were successfully deployed.
Placement Accuracy: The deployment accuracy was evaluated and found to be acceptable in all cases.
3ULPDU\(ijHFWLYHQHVV(QGSRLQW Technical success, defined as successful deployment and placement accuracy based upon a rating scale
completed by the Investigators at the time of the index procedure. Bookend sizes were evaluated for clinical utility of size range.
All stents had good or excellent deployment accuracy with successful placement at the target site. In none of the patients, TLR and/or TVR
was conducted until day 30 from the index procedure.
3ULPDU\6DIHW\(QGSRLQW)UHHGRPIURPRFFXUUHQFHRIGHDWKDPSXWDWLRQDQG795DQGRU7/5DWbGD\VSRVWLQGH[SURFHGXUH
$OOVXEMHFWVVKRZHGIUHHGRPIURPRFFXUUHQFHRIGHDWKDPSXWDWLRQDQG7/5DQGRU795DWGD\SRVWLQGH[SURFHGXUH2YHUDOOWKLV
study demonstrated the safety of the LIFESTENT® Vascular Stent Systems deploying stents of 5 mm in diameter.
21
Freedom from Restenosis for Popliteal Segment 1
Survival Distribution Function
LifeStent
PTA
LifeStent
Survival Distribution Function
PTA
22
Control PTA Test Stent
Survival Subjects Censored Subjects Survival Subjects Censored Subjects
Time
%[95% CI] with Event Subjects at Risk %[95% CI] with Event Subjects at Risk
70.8% 94.7%
180 days 26 6 58 4 9 71
[66.0, 75.6] [92.1, 97.3]
46.5% 75.8%
365 days 45 17 28 18 19 47
[40.9, 52.0] [70.8, 80.7]
39.8% 67.0%
730 days 48 33 9 23 39 22
[33.8, 45.8] [61.3, 72.8]
Safety
7KLUWHHQ SDWLHQWVKDGGLHGE\0RQWKSDWLHQWVZKRZHUHWUHDWHGZLWK37$DQGSDWLHQWVZKRUHFHLYHGDVWHQW1RQHRIWKHDGYHUVH
events causing death were related to LIFESTENT® Stent or procedure.
The reported fracture rate was 5.4% at 12-months and 11.1% at 24-months for P2/P3 segment treatment. The number of available x-rays
was 37 and 45 x-rays at the 12-month and 24-month time-point respectively (see Table "X-ray Reported Stent Fractures"). Fractures are
counted once, at the first time the fracture was reported.
'XULQJWKH(7$3VWXG\SDWLHQWVLQWKH33JURXSH[SHULHQFHGWKUHH7\SH,RQH7\SH,,RQH7\SH,,,DQGWZR7\SH,9IUDFWXUHVZKLOH
WKH3JURXSKDGRQH7\SH,,,DQGRQH7\SH,,IUDFWXUH1RFRUUHODWLRQFRXOGEHIRXQGEHWZHHQWKHLQFLGHQFHRIVWHQWIUDFWXUHVDQGHLWKHU
restenosis or TLR.
The following information is provided in the packaging for the physician to provide their patients:
š $3DWLHQW*XLGHZKLFKLQFOXGHVLQIRUPDWLRQRQWKHLIFESTENT® Vascular Stent System, peripheral artery occlusive disease, the implantation
procedure and patient care following the implant.
š $3DWLHQW,PSODQW&DUGWKDWLVXVHGWRUHFRUGDQGGLVVHPLQDWHLQIRUPDWLRQDERXWWKHSDWLHQWDQGWKHVWHQW
23
M. How Supplied
STERILE:)256,1*/(86(21/<7KHLIFESTENT® Vascular Stent System is supplied sterile (by ethylene oxide gas) and is nonpyrogenic. Do
not resterilize and/or reuse the device. Do not use if the temperature exposure indicator (i.e., square label found on the pouch) is black as
the unconstrained stent diameter may have been compromised. The temperature exposure indicator label should be grey and must be clearly
visible on the pouch. Do not use if pouch is opened or damaged. Do not use the device after the “Use By” date specified on the label. For
returned product or product issues, please contact Bard Peripheral Vascular at the address below:
Bard Peripheral Vascular, Inc.
1625 West 3rd Street
Tempe, AZ 85281 USA
CONTENTS for one (1) LIFESTENT® Vascular Stent System:
š 2QH LIFESTENT® Vascular Stent System
š 2QH 3DWLHQW,PSODQW&DUG
š 2QH ,QVWUXFWLRQVIRU8VH
š 2QH 3DWLHQW*XLGH
STORAGE: Store in a cool, dark, dry place. Storage temperature should not exceed 60 °C. Use by the “Use By” date specified on the label.
DISPOSAL INSTRUCTIONS: After use, dispose of product and packaging in accordance with hospital, administrative and/or local government
policy.
24
d) Visually inspect the distal end of the stent system to ensure that the stent is contained within the sheath. Do not use if the stent is partially
deployed.
e) Visually inspect the distal end of the delivery system catheter to ensure there is no gap between the delivery system catheter tip (grey
colored) and the primary sheath (braided catheter with light blue colored end) such that the guidewire lumen (orange colored) is visible.
Do not use the device if the orange colored guidewire lumen is visible.
f) Flush the inner lumen of the stent system with heparinized normal saline prior to use.
g) Wipe the usable length portion of the stent system with gauze soaked with heparinized normal saline.
It is recommended to use the 80 cm working length device for ipsilateral procedures. The longer working length of the 130 cm device
may potentially be challenging for the user to keep straight for ipsilateral procedures. Failure to keep the device straight may impede the
optimal deployment of the implant.
2. Dilate Lesion
Predilation of the lesion should be performed using standard techniques. While maintaining site access with a guidewire, remove the
balloon catheter from the patient.
Caution: During dilation, do not expand the balloon such that dissection complication or perforation could occur.
Caution: Any slack in the stent system (outside the patient) could result in deploying the stent beyond the target site.
4. Deploy stent
a) Verify that the distal and proximal stent radiopaque markers are distal and proximal to the target lesion.
b) Confirm that the introducer sheath is secure and will not move during deployment.
c) Remove the shipping lock.
25
d) To ensure the most accurate placement, firmly hold the black system stability sheath throughout deployment.
Note:'R127KROGWKHVLOYHUVWHQWGHOLYHU\VKHDWKDWDQ\WLPHGXULQJGHSOR\PHQW'2127FRQVWULFWWKHVWHQWGHOLYHU\VKHDWKGXULQJVWHQW
deployment.
e) Initiate stent deployment by rotating the thumbwheel in the direction of the arrows while holding the handle in a fixed position.
Note: If excessive force is felt during stent deployment, do not force the stent system. Remove the stent system as possible, and replace
with a new unit.
f) While using fluoroscopy, maintain position of the distal and proximal stent radiopaque markers relative to the targeted site. Watch for the
GLVWDOVWHQWUDGLRSDTXHPDUNHUVWREHJLQVHSDUDWLQJVHSDUDWLRQRIWKHGLVWDOVWHQWUDGLRSDTXHPDUNHUVVLJQDOVWKDWWKHVWHQWLVGHSOR\LQJ
Continue turning the thumbwheel until the distal end of the stent obtains complete wall apposition.
g) With distal end of the stent apposing the vessel wall, final deployment can be continued with the following methods (Fig. 3, 4, 5).
Figure 3. Thumbwheel
h) Deployment of the stent is complete when the proximal stent radiopaque markers appose the vessel wall and the sheath radiopaque zone
is proximal to the proximal stent radiopaque markers.
i) DO NOT attempt to re-sheath stent system prior to removal.
26
Symbols used on labeling DISCLAIMER OF WARRANTY
BARD PERIPHERAL VASCULAR, INC. WARRANTS
TO THE FIRST PURCHASER OF THIS PRODUCT,
Keep away from sunlight
THAT THIS PRODUCT WILL BE FREE FROM
DEFECTS IN MATERIALS AND WORKMANSHIP
FOR A PERIOD OF ONE YEAR FROM THE DATE
OF FIRST PURCHASE AND LIABILITY UNDER THIS
Keep dry
LIMITED PRODUCT WARRANTY WILL BE LIMITED,
TO REPAIR OR REPLACEMENT OF THE DEFECTIVE
PRODUCT, IN BARD‘S SOLE DISCRETION,
The Green Dot OR REFUNDING YOUR NET PRICE PAID. WEAR
AND TEAR FROM NORMAL USE OR DEFECTS
RESULTING FROM MISUSE OF THIS PRODUCT IS
NOT COVERED BY THIS LIMITED WARRANTY.
Recyclable TO THE EXTENT ALLOWABLE BY APPLICABLE LAW,
THIS LIMITED PRODUCT WARRANTY IS IN LIEU OF
ALL OTHER WARRANTIES, WHETHER EXPRESS
OR IMPLIED, INCLUDING, BUT NOT LIMITED TO,
MR Conditional ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE. IN
NO EVENT WILL BARD PERIPHERAL VASCULAR,
INC. BE LIABLE TO YOU FOR ANY INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES
RESULTING FROM YOUR HANDLING OR USE OF
THIS PRODUCT.
Some countries do not allow an exclusion of implied
warranties, incidental or consequential damages.
<RXPD\EHHQWLWOHGWRDGGLWLRQDOUHPHGLHVXQGHUWKH
laws of your country.
27
LIFESTENT® Vascular Stent Systems
Bard and LifeStent are trademarks and/or registered trademarks of C. R. Bard, Inc.
Caution: Federal (USA) law restricts this device to sale by or on order of a physician.
B05678 Rev.2/04-16
28